Gendex Bucky: troubleshooting and error codes
The ServiceForge knowledge base holds 2 documented symptoms for the Gendex Bucky, with 6 ranked causes, 5 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No exposure when bucky selected radiation
X-ray system does not produce exposure when particular bucky interface selected; grid may or may not oscillate
Ranked causes
- 1. Bucky grid not oscillating; interface not functional (common)
- 2. Bucky exposure enable output circuit malfunction (common)
- 3. Cable failure or loose connections (occasional)
- 4. Generator bucky select output circuit failure (occasional)
- 5. Generator exposure enable input circuit failure (occasional)
Observe grid behavior on expose command
- Test point
- Bucky grid
- Expected
- Grid oscillates if interface half functional
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exposure occurs but grid does not move radiation
X-ray exposure fires but bucky grid remains stationary when bucky is selected
Ranked causes
- 1. Incorrect bucky interface wiring (common)
Compare actual interface wiring to connection diagram
- Test point
- Bucky interface connections
- Expected
- All wires connected per diagram; Midwest 14x36 bucky configuration matched