GE Healthcare Vscan Extend: troubleshooting and error codes

The ServiceForge knowledge base holds 37 documented symptoms for the GE Healthcare Vscan Extend, with 71 ranked causes, 45 diagnostic checks with test points and expected values, and 39 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

No connection between Vscan Extend and PC

USB connection fails to establish; device not recognized by computer

Ranked causes

  1. 1. USB door not fully opened or cable not properly inserted (common)
  2. 2. Vscan Extend powered off or not responding (common)
  3. 3. USB drivers not properly installed on PC (occasional)
  4. 4. USB cable defective or incompatible (occasional)
  5. 5. PC USB port defective (occasional)
  6. 6. Vscan Extend device defective (rare)
First check

Inspect and open USB door; reseat USB cable

Test point
Vscan Extend USB port and PC USB connector
Expected
USB door fully open; cable seated firmly on both ends

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit is not charging

Battery does not charge when connected to power source

Ranked causes

  1. 1. Battery not inserted (common)
  2. 2. Battery defect or end of life (occasional)
  3. 3. Broken battery connection (rare)
  4. 4. Defective AC adapter or charger (occasional)
  5. 5. Mains power is down (occasional)
  6. 6. Temperature outside specified limits (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Image noise present

Visible noise artifacts in ultrasound image; may be coherent, intermittent, or continuous depending on source

Ranked causes

  1. 1. EMI from external radio frequency or magnetic field sources (common)
  2. 2. Heat-related noise after prolonged operation (occasional)
  3. 3. Hardware failure in probe or main unit (occasional)
  4. 4. Software issue or settings misalignment (occasional)
First check

Determine if noise is present continuously or appears after equipment use

Test point
Monitor image during normal operation and after extended use
Expected
Noise absent during normal conditions; if appears after use, indicates heat or software issue

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low frame rate caution

Ultrasound system displays reduced frame rate; imaging performance degraded

Ranked causes

  1. 1. Battery temperature out of soft limit (common)
  2. 2. DP Probe temperature out of soft limit (common)
  3. 3. G3S Probe temperature out of soft limit (common)
  4. 4. Power board temperature out of soft limit (occasional)
First check

Check battery temperature status

Test point
Device log (logfile.txt)
Expected
Battery temperature within normal range

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System shutdown - too hot caution

Device powers down due to excessive temperature; battery, probe or power board hard thermal limit exceeded

Ranked causes

  1. 1. Battery temperature out of hard limit (occasional)
  2. 2. DP Probe temperature out of hard limit (occasional)
  3. 3. G3S Probe temperature out of hard limit (occasional)
  4. 4. Power board temperature out of hard limit (rare)
First check

Access device logs via SD Card

Test point
logfile.txt on SD Card
Expected
Log file readable and contains error entry
Tools
Computer with card reader

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit has no power

Unit does not power on, no response to power button

Ranked causes

  1. 1. Battery not inserted (common)
  2. 2. Battery not charged (common)
  3. 3. Battery defect or end of life (occasional)
  4. 4. Broken battery connection (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Color noise in near field

Self-generated noise appearing as color artifacts in near field region of image; does not change when touching scanner or probe

Ranked causes

  1. 1. Heat problems (common)
  2. 2. Hardware problems (occasional)
  3. 3. Software problems (occasional)
First check

Monitor color noise behavior over extended operation period

Test point
Image display during continuous scanning
Expected
Color noise absent or minimal; does not increase with time

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Image does not display

No image shown on display

Ranked causes

  1. 1. Battery insufficient charge (common)
  2. 2. Broken flex cable connection between phone and power board (occasional)
  3. 3. CPU board failure (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No connection to Vscan Extend portal

Unable to connect to Vscan Extend web portal; Internet Explorer or network connectivity issue

Ranked causes

  1. 1. Internet Explorer version too old or not installed (common)
  2. 2. No Internet access or network connection lost (common)
  3. 3. Firewall blocking portal access (occasional)
First check

Verify Internet Explorer version

Test point
IE Help menu or About Internet Explorer
Expected
Version 7 or newer installed

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Software loading or reloading fails

Device does not load application software from microSDHC card or MST firmware upgrade fails to complete

Ranked causes

  1. 1. microSDHC card does not contain application software file (common)
  2. 2. AC/DC adapter disconnected during MST firmware installation (occasional)
  3. 3. Power loss or disconnection during installation (rare)
First check

Verify microSDHC card contents match system software version

Test point
microSDHC card part number
Expected
FRU P/N 5757806-1 (for v1.1.0 and below), 5757806-2 (for v1.1.1 and below), 5793186-1 (for v1.2.1 and v1.2.0), or 5793186-2 (for v1.2.0 and above)

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Device does not power on

Device does not turn on when powered

Ranked causes

  1. 1. No power from mains (common)
  2. 2. Problem with phone or battery (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Device powers off unexpectedly during software upgrade caution

Vscan Extend shuts down or loses power while software or firmware installation is in progress

Ranked causes

  1. 1. Battery not fully charged before upgrade (common)
  2. 2. AC/DC adapter disconnected during application software installation (common)
First check

Verify battery charge level before starting upgrade

Test point
Battery indicator on device
Expected
Fully charged

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Software version not updated after reload

Device still displays old software version number after reload procedure appears to complete

Ranked causes

  1. 1. Installation did not complete successfully (occasional)
  2. 2. Original SD card not reinserted after upgrade (occasional)
First check

Verify software version in Settings menu

Test point
Settings > About screen
Expected
New version number (e.g., 1.2.3 for application software or 1.6 for MST firmware)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

System shutdown after 50 seconds on boot

Battery critically low message appears during boot and device shuts down

Ranked causes

  1. 1. Battery depleted or faulty (common)
  2. 2. CPU/Power board failure (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery critically low

System displays warning that battery is at critical level below 5%

Ranked causes

  1. 1. Battery depleted (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery pack requires replacement

Battery pack (P/N: 5693456) needs to be removed and inserted

Ranked causes

  1. 1. Battery pack end-of-life or malfunction (unknown)
First check

Power off device and position correctly

Test point
Device power state and placement
Expected
Device OFF, placed face down on clean flat soft surface

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Color noise displayed

Vertical color noise stripes appear on display after approximately five minutes of scanning

Ranked causes

  1. 1. CPLD firmware is old (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Connectivity error

System unable to connect, Wi-Fi connection lost

Ranked causes

  1. 1. Wi-Fi network unavailable or weak signal (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Date and time reset

Device date and time settings lost

Ranked causes

  1. 1. Phone settings lost (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Device loses network connectivity during software update

Update process gets cancelled when device loses connection to non-clinical network during update detection or download

Ranked causes

  1. 1. Network connectivity interruption (occasional)
First check

Verify network connection status

Test point
Device Wi-Fi indicator
Expected
Connected to non-clinical network

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

DICOM image server verification failure

DICOM server connection fails during image storage or query

Ranked causes

  1. 1. DICOM endpoint does not support initiated service (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display flashing while scanning

Screen flashes or refresh rate appears reduced during ultrasound scanning

Ranked causes

  1. 1. Automatic frame rate reduction due to temperature rise (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display screen blank when powered on

Screen shows no image or content after power-on

Ranked causes

  1. 1. Connection broken during software loading (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failure to store image

System cannot save or store ultrasound image

Ranked causes

  1. 1. Storage error (cause varies) (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

No image displayed when scanning

Probe connected but no ultrasound image appears on screen

Ranked causes

  1. 1. Defective probe (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Noise when moving probe cable

Audible noise or crackling occurs when probe cable is moved

Ranked causes

  1. 1. Defective probe cable (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Parts of image missing during scan

Ultrasound image displays incomplete or with missing sections

Ranked causes

  1. 1. Channels are missing (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Probe overheats caution

Probe displays warning message indicating overheating

Ranked causes

  1. 1. Prolonged probe use or contact with high-temperature tissue (unknown)

Scanning not allowed with AC connected

System prevents scanning when AC power is connected

Ranked causes

  1. 1. AC adapter connected during scan attempt (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

SD card requires removal or replacement

SD card needs to be removed from or inserted into SD card slot

Ranked causes

  1. 1. SD card replacement or maintenance required (unknown)
First check

Remove battery pack first

Test point
Battery pack status
Expected
Battery pack removed per section 8.2.2

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Software update procedure over internet

Step-by-step process to download and install GE Marketplace, GE Kiosk, and MST Firmware updates via network connection

Ranked causes

  1. 1. Software update installation process (common)
First check

GE Marketplace download and installation

Test point
Update screen
Expected
GE Marketplace updates and device connects to GE server

8 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System memory full

System internal memory usage at 50-60%, backup required

Ranked causes

  1. 1. Insufficient free memory (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

System overheats caution

System displays warning message and will shut down due to overheating

Ranked causes

  1. 1. Prolonged operation or ambient temperature too high (unknown)

System voltage critical caution

System displays warning of critical voltage level

Ranked causes

  1. 1. Power supply malfunction or battery failing (unknown)

Unit not entering scan screen

Unit stays on GE startup screen and does not transition to scanning interface

Ranked causes

  1. 1. Insufficient battery charge (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Voltage out of limit high voltage

Power supply voltage exceeds acceptable operating range; system shuts down

Ranked causes

  1. 1. Input voltage out of hard limits (rare)
First check

Read voltage error from log

Test point
Device logfile.txt
Expected
Voltage hard limit error recorded

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

White screen on startup

Screen remains white and unit does not start after power-on

Ranked causes

  1. 1. Software loading error (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
Abstract syntax not supportedDICOM endpoint does not support requested service typeAttempt to use unsupported DICOM service (e.g., MWL query to non-MWL server)error
Association request failed (or never connected)Initial connection to DICOM server could not be establishedDevice attempts to connect to DICOM endpointerror
Battery criticalBattery is critically low or below 5%Battery charge drops below 5%warning
Connectivity errorUnable to connect, Wi-Fi connection lostLoss of Wi-Fi network connectionwarning
Device not connected to power sourcePower source not detected during update processDevice powered by battery only or no power connection during update attempterror
DUL Association Rejected: App Context Name Not SupportedDICOM endpoint does not recognize application contextDICOM association negotiation fails at context validationerror
DUL Association Rejected: Called AE Title Not RecognizedCalled AE Title mismatchDICOM association request targets unrecognized Called AE Titlewarning
DUL Association Rejected: Calling AE Title Not RecognizedCalling AE Title mismatchDICOM association request includes unrecognized Calling AE Titlewarning
DUL Association Rejected: Local Limit ExceededToo many clients connected to DICOM endpointDevice attempts connection when endpoint client limit reachedwarning
DUL Association Rejected: No Reason (Service Provider)DICOM endpoint rejects association without stated reasonDICOM connection fails during endpoint processingerror
DUL Association Rejected: No Reason (Service User)DICOM endpoint rejects association without reasonDICOM connection attempt failserror
DUL Association Rejected: Protocol Version Not SupportedDICOM protocol version mismatchDevice and endpoint use incompatible DICOM protocol versionserror
DUL Association Rejected: Temporary CongestionNetwork or endpoint resource congestion prevents connectionConnection attempt during high load on DICOM endpointwarning
Failure - Invalid object instanceDICOM file instance is invalid or malformedExport of corrupted or malformed DICOM objecterror
Failure - Processing failureDICOM endpoint cannot process sent fileDevice exports DICOM file type not supported by endpointerror
Identifier does not match SOP classDICOM file dataset does not match its SOP class definitionExport of DICOM file with mismatched dataset and SOP classerror
Internal application errorUnexpected software error during DICOM operationUnhandled exception in DICOM service codecritical
Low frame rateImaging frame rate reduced due to thermal throttlingBattery, DP Probe, G3S Probe or Power board temperature exceeds soft limit (warning threshold)warning
Memory fullSystem internal memory usage at 50-60%Internal memory reaches 50-60% capacitywarning
Network fails after downloading all applicationsNetwork connection lost after all software downloaded but before installation completeNetwork connectivity interruption during software installation phasewarning
Network timeout errorDICOM server does not respond within timeout periodNetwork communication with DICOM endpoint stalledwarning
No network connectionDevice cannot reach networkNetwork interface unavailable or disconnectederror
No ReasonDICOM transaction rejected without stated reasonDICOM endpoint rejects associationerror
No type negotiatedNo DICOM presentation context agreedDICOM association fails to establish any valid contexterror
Refused: Duplicate InvocationMessage ID already in use for another operationRepeated DICOM transaction with same Message IDwarning
Refused: Mistyped argumentInvalid parameter supplied for DICOM serviceDICOM operation uses parameter not negotiated in associationerror
Refused - Server out of resourcesDICOM endpoint cannot accept file due to resource shortageDevice attempts to export DICOM file to full or resource-constrained endpointwarning
Refused: SOP Class Not SupportedSOP Class mismatch with initiated DICOM serviceDevice uses wrong SOP Class for requested DICOM serviceerror
Refused: Unrecognized OperationDICOM endpoint does not recognize requested operationDevice sends unrecognized DICOM operation codeerror
Scan blockedScanning not allowed when AC is connectedUser attempts to scan while AC adapter is plugged ininfo
Scanner app not availableApplication software file missing or not recognized on microSDHC cardPressing Upgrade Software when SD card lacks application software fileerror
SD card with new software not insertedRequired SD card for software installation not present in deviceUpdate process initiated without SD card insertederror
System errorSystem is overheating or other critical conditionSystem overheating or probe overheating conditionerror
Too hot! shut down systemDevice automatically powers down due to excessive temperatureBattery, DP Probe, G3S Probe or Power board temperature exceeds hard limit (shutdown threshold)critical
Transfer syntax not supportedDICOM endpoint rejects proposed transfer syntaxDICOM association negotiation fails on transfer syntaxcritical
User RejectionDICOM presentation context rejected by service providerDICOM association request fails at context negotiationwarning
Voltage criticalSystem voltage at critical levelPower supply voltage drops below safe operating levelerror
Voltage out of limit! shut down systemSupply voltage out of acceptable operating range; system protection activatedInput voltage exceeds hard limits (AC supply or battery voltage)critical
Wi-Fi enabled but no internet connectionDevice cannot connect to internet despite Wi-Fi being enabledWi-Fi enabled, network selected, but no internet access availableerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI