GE Healthcare Voluson E6: troubleshooting and error codes
The ServiceForge knowledge base holds 50 documented symptoms for the GE Healthcare Voluson E6, with 99 ranked causes, 106 diagnostic checks with test points and expected values, and 9 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Excessive leakage current detected high voltage
AC/DC adapter, chassis, probe, or peripheral equipment measures above acceptable leakage current limits during electrical safety testing
Ranked causes
- 1. AC/DC adapter defective (common)
- 2. Power cord or ground connection defective (common)
- 3. Wiring damage or poor connections (common)
- 4. Wall outlet not grounded or faulty (common)
- 5. Neutral and ground wires reversed at outlet (occasional)
- 6. Probe defective or probe connector faulty (occasional)
- 7. System connector pins bent or ground continuity lost (occasional)
- 8. Peripheral equipment ground defective (occasional)
- 9. UPS (Uninterruptible Power Supply) causing leakage (occasional)
- 10. Internal or external peripheral defective (rare)
- 11. ECG cable damaged or poorly connected (rare)
- 12. System hardware defect (new system) (rare)
Visual inspection of AC/DC adapter and cable
- Test point
- AC/DC adapter and cable
- Expected
- No defects, cracks, or damage visible
11 more checks with pass and fail actions, 13 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive leakage current high voltage
Leakage current test fails; potential electrical safety hazard during patient contact or system operation
Ranked causes
- 1. AC/DC adapter defective (occasional)
- 2. Ground connection broken or intermittent on power cord (common)
- 3. Loose or missing ground studs with star washers (common)
- 4. Bad crimps, poor connections, or damaged wiring (occasional)
- 5. Wall outlet not grounded or improperly wired (occasional)
- 6. Probe design or excessive probe leakage (occasional)
- 7. System connector bent pins, poor connections, or ground continuity issue (occasional)
- 8. Peripheral equipment ground connections defective (occasional)
- 9. UPS unit causing leakage current fault (rare)
- 10. Internal or attached peripheral component defective (rare)
Inspect AC/DC adapter and cable for defects
- Test point
- AC/DC adapter connector and cable insulation
- Expected
- No damage, defects, or deterioration visible
9 more checks with pass and fail actions, 11 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
5V Standby voltage not available
LED 1 (5V Standby) on boot diagnostic panel is OFF
Ranked causes
- 1. No mains voltage or power cable not connected (common)
- 2. Circuit breaker OFF (common)
- 3. Power supply (RSP) defective (occasional)
- 4. Standby line interrupted (RSP to RTB to RTT to RTU) (occasional)
Check mains power connection
- Test point
- Power inlet and wall outlet
- Expected
- Power cable firmly connected
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor: no image caution
Display shows no image or blank screen
Ranked causes
- 1. Power cord not connected or loose (common)
- 2. Video cable disconnected or loose (common)
- 3. Video cable pins bent or damaged (occasional)
- 4. No video signal from system backplane (rare)
Verify monitor power cord connection
- Test point
- Power cord
- Expected
- Firmly connected to monitor and wall outlet
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mains power cable damaged high voltage
Power cable shows visible damage, loose connections, or corroded terminals
Ranked causes
- 1. Physical damage to power cable or connectors (occasional)
- 2. Loose or corroded wire terminals (occasional)
- 3. Non-functional inlet connector retainer (rare)
Disconnect and visually inspect mains power cable and connectors
- Test point
- Mains power cable and connectors
- Expected
- No visible damage, corrosion, or bent pins on cable or connectors
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Noise disturbs the image caution
Electromagnetic interference visible in ultrasound image, typically when probe is contacted
Ranked causes
- 1. EMC noise from external power or signal cables (common)
- 2. EMC noise from wall outlet power line (occasional)
- 3. Magnetic field in room or building (occasional)
Verify ultrasound image shows EMC noise with probe contacted
- Test point
- Ultrasound image display
- Expected
- EMC noise visible
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PC Power OK LED OFF
LED 3 (PC Power OK) on boot diagnostic panel is OFF, indicating PC was not started correctly
Ranked causes
- 1. RSP could not send Start PC signal or PC did not send PS_ON signal (common)
- 2. BackEnd (PC board) defective (occasional)
- 3. Power supply (RSP) defective (occasional)
Verify ATX power cables between RSP and PC
- Test point
- ATX connectors and cables
- Expected
- All cables firmly connected
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe connection fails or probe not detected patient safety
Probe does not connect properly, locking handle cannot turn, or system does not recognize probe
Ranked causes
- 1. Debris on probe or socket connectors (common)
- 2. Probe locking handle not at horizontal position before insertion (common)
- 3. Probe connector not fully inserted (common)
Inspect probe and socket for debris
- Test point
- Probe connector tip and socket receptacle opening
- Expected
- Clean, debris-free surfaces on both connector and socket
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Secondary Patient Monitor not displaying image patient safety
Monitor connected but no picture appears on secondary display
Ranked causes
- 1. Incorrect VGA cable connection (common)
- 2. Secondary Monitor power not enabled or timeout not observed (occasional)
- 3. Monitor connected directly to mains instead of isolation transformer (rare)
Verify VGA cable connection
- Test point
- VGA Out connector on external I/O connector panel (GES)
- Expected
- VGA cable from Secondary Monitor securely plugged in
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not complete boot sequence
System powered on but fails to load operating system, boot screen not displayed, or startup stalls
Ranked causes
- 1. Insufficient wait time after power recycling (occasional)
- 2. Encrypted disk locked (BT17 and higher models) (occasional)
- 3. User login required but credentials not entered (occasional)
Verify wait time after last shutdown
- Test point
- Time between power off and current power-on attempt
- Expected
- Minimum 10 seconds elapsed
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not power on high voltage
Standby/Power button does not illuminate or system fails to start after power button press
Ranked causes
- 1. Mains power cable not connected or improperly connected (common)
- 2. Circuit breaker at rear not switched ON (common)
- 3. Fuses blown (2x T10A H/250V) (occasional)
Verify mains power cable connected to system and wall outlet
- Test point
- Rear of system connector and AC wall outlet
- Expected
- Cable fully inserted and secured
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: LCD Monitor Adjustment
Monitor arm positioning and alignment for Voluson E6/E8/E10 systems
Ranked causes
- 1. Monitor arm requires adjustment per model specifications (unknown)
- 2. Monitor arm (Ergotron) requires adjustment per specifications (unknown)
Verify monitor tilt range (MTD20)
- Test point
- Monitor Arm MTD20 tilt axis
- Expected
- + 40° / - 90°
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Isolation Transformer configuration incorrect high voltage
Wrong voltage or fuse settings on isolation transformer may cause damage or non-operation of connected peripherals
Ranked causes
- 1. Input/output voltage selectors not set to match local line power (common)
- 2. Incorrect or missing fuses installed (common)
Verify voltage selector positions match local power supply
- Test point
- Input and output voltage selector switches on isolation transformer
- Expected
- Both selectors set to 115V or 230V matching local line power
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Isolation Transformer physical placement unsafe patient safety
Transformer positioned in water-prone area or too far from monitor/power source, or blocking patient access to safety devices
Ranked causes
- 1. Transformer placed within patient reach or in wet environment (occasional)
- 2. Wall mounting installation performed by unqualified personnel (occasional)
Verify transformer location is dry and out of patient reach
- Test point
- Transformer physical location in operating room or patient area
- Expected
- Transformer on floor or wall, away from patient access, in dry location
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor: picture is fuzzy caution
Monitor display appears blurry or out of focus
Ranked causes
- 1. Contrast or brightness incorrectly adjusted (common)
- 2. SVGA card producing excessive output level (occasional)
Adjust monitor contrast control
- Test point
- Monitor OSD menu
- Expected
- Contrast at optimal level for clear image
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Network: no connection to network
System unable to establish network connectivity
Ranked causes
- 1. Network cable disconnected or unseated (common)
- 2. Network hardware failure in system (rare)
Visually inspect network cable at system connection
- Test point
- Network port
- Expected
- Cable firmly seated in port; indicator shows active (green) or disconnected (gray)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe cleaning procedure not followed patient safety
Probe was cleaned with improper agents or techniques causing damage
Ranked causes
- 1. Probe cleaned with non-approved disinfectants or high-proof alcohol (occasional)
- 2. Probe lens soaked or wiped with incompatible products (rare)
Verify probe was cleaned with approved agents per Care Card
- Test point
- Probe condition after cleaning
- Expected
- Probe is clean and shows no damage from cleaning process
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Standby/Power button LED blinking orange and green
LED alternates between orange and green, system not fully operational
Ranked causes
- 1. FPGA not initialized (FrontEnd or RFM board issue) (common)
- 2. Power Supply (RSP) defect preventing FPGA initialization (occasional)
Verify LED 4 on Power Supply RSP3 (FPGA_CONF_DONE)
- Test point
- LED 4 on ATX board inside RSP3
- Expected
- LED ON (FPGA_CONF_DONE = high)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Standby/Power button LED not illuminated
No light from the LED in the start key on the User Interface console
Ranked causes
- 1. Circuit breaker switched OFF (common)
- 2. Power Supply (RSP) defect (occasional)
Verify circuit breaker position
- Test point
- Main circuit breaker
- Expected
- Breaker in ON position
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Trackball malfunction
Trackball not responding or operating correctly
Ranked causes
- 1. Trackball contamination or debris (common)
- 2. Trackball hardware failure (unknown)
Inspect trackball for visible contamination or debris
- Test point
- Trackball surface
- Expected
- Clean surface
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Activate Windows message displayed
Monitor screen shows 'Activate Windows' message with prompt to go to Settings to activate Windows
Ranked causes
- 1. Activation servers of Microsoft are not reachable (common)
Check license status in System Info
- Test point
- Utilities - Setup - Administration - System Info tab
- Expected
- Licensed status indicates Windows is properly activated
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Configuration done LED OFF
LED 2 (Configuration done) on boot diagnostic panel is OFF, indicating FrontEnd FPGA not initialized correctly
Ranked causes
- 1. RFM board initialization failure (common)
Verify RFM board power supply
- Test point
- RFM board power connectors
- Expected
- All power connectors firmly seated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control console movement without power boot
Procedure to lower or lift UI when system cannot boot
Ranked causes
- 1. System unable to boot (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Data loss when disconnecting USB device caution
Valuable data may be lost if USB device is unplugged or ejected without proper system notification
Ranked causes
- 1. USB device stopped improperly without system notification (common)
Press Eject key on keyboard
- Test point
- Keyboard Eject key
- Expected
- Connect USB and Network Drives dialog displays all connected USB and Network drives
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Date and time settings incorrect after upgrade
System date and time are not correct or not matching the facility time after software upgrade.
Ranked causes
- 1. System clock reset during upgrade or incorrect initial setting (occasional)
Verify system date and time
- Test point
- System Setup - General Settings - General page
- Expected
- Date and time match facility current time
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disk Encryption feature not activated after upgrade when Advanced Security Features enabled
The Disk Encryption feature is not active or not properly configured after software upgrade when Advanced Security Features option is enabled.
Ranked causes
- 1. Disk Encryption not reactivated after software upgrade (common)
Verify Disk Encryption is activated
- Test point
- System Setup - Administration - Disk Encryption
- Expected
- Disk Encryption feature shows active/enabled status
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Distraction during patient scan due to USB device dialogs patient safety
Hardware detection dialogs appear and distract operator during active patient scanning
Ranked causes
- 1. USB device connected or disconnected during active patient scan (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG cable leakage current issue high voltage
Excessive leakage current detected in ECG subsystem
Ranked causes
- 1. ECG cable damaged or poorly connected (common)
Inspect ECG cables for damage and connections
- Test point
- All ECG cable connectors and insulation
- Expected
- No visible damage, all connectors firmly seated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
HDD/SSD Status LED not active
LED 4 (HDD/SSD status) on boot diagnostic panel is OFF, indicating no access to hard disk or solid state drive
Ranked causes
- 1. Software hang-up or HDD/SSD inaccessible (unknown)
Verify HDD/SSD power and data connections
- Test point
- HDD/SSD power and SATA connectors
- Expected
- Cables firmly connected, drive powered
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard layout not correct after upgrade
System keyboard input language or layout does not match the expected or previously configured setting after software upgrade.
Ranked causes
- 1. Keyboard layout reset during upgrade or initial incorrect setting (occasional)
Verify keyboard layout setting
- Test point
- System keyboard and input language configuration
- Expected
- Keyboard layout matches facility standard or previously configured setting
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leakage currents not measured after installation patient safety
Safety-critical electrical safety check not performed after mounting Secondary Monitor
Ranked causes
- 1. Installation procedure incomplete without leakage current measurement (common)
Perform leakage current measurement per IEC standards
- Test point
- Secondary Monitor and Isolation Transformer electrical connections
- Expected
- Leakage currents within limits specified by IEC 60601-1 or IEC 62353
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Memory test does not complete or shows errors
PC Memory Test runs for ~2.5 hours; errors displayed if memory is faulty. Test cycles endlessly unless interrupted.
Ranked causes
- 1. BackEnd Processor memory failure or power supply malfunction (unknown)
Access Boot Screen
- Test point
- System startup
- Expected
- Boot Screen appears with Voluson, GE-Service, Rollback, Memtest options
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor: video test patterns unclear or misaligned caution
Test patterns appear distorted, dim, non-parallel or non-square
Ranked causes
- 1. Monitor hardware failure or defect (common)
Display system video test pattern on monitor
- Test point
- Monitor display
- Expected
- Test pattern is clear, bright, parallel and square
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
New Software Verification report missing after installation
The 'New Software Verification' field does not appear on the left side of the 2D screen after software upgrade completes.
Ranked causes
- 1. Display rendering issue or incomplete boot sequence (unknown)
Verify New Software Verification field is present on screen
- Test point
- 2D screen left side display
- Expected
- New Software Verification field visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Permanent key required for software version upgrade
New software extension version requires new permanent key; system may not upgrade without valid key
Ranked causes
- 1. Software extension version upgrade exceeds current permanent key validity (common)
Verify if upgrade to higher software version is needed
- Test point
- System Setup > Administration > System Info tab
- Expected
- Compare current version to target version; if target version higher, permanent key needed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer not functioning after software upgrade
Network printer is not available or not responding after system software installation or upgrade.
Ranked causes
- 1. Printer driver incompatibility or not reinstalled after upgrade (occasional)
Reconnect external devices and install printers
- Test point
- External device connections and printer installation
- Expected
- All printers installed and functioning
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer Remote Control selection not matched after upgrade
Printer Remote Control settings in System Setup do not match the installed printer configuration after software upgrade.
Ranked causes
- 1. System Setup not reconfigured after software upgrade (common)
Verify Printer Remote Control setting matches installed printer
- Test point
- System Setup - Connectivity - Button Configuration
- Expected
- Printer Remote Control selection matches installed printer configuration
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe disconnected during live scan patient safety
Probe removed while system is in write/live scan mode
Ranked causes
- 1. Probe removed during active scanning (occasional)
Verify system is in freeze or read mode
- Test point
- Touch panel display and probe status
- Expected
- System not in Live Scan Write mode
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe wear or damage evident patient safety
Probe shows visible signs of wear, damage or degradation
Ranked causes
- 1. Probe wear or damage from improper handling or use (occasional)
Visually inspect probe for signs of wear or damage
- Test point
- Probe exterior and connector
- Expected
- No visible signs of wear, cracks, surface damage or corrosion on probe
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Regulatory compliance verification
Ensure regulatory information and tests required by national law are present, documented and performed
Ranked causes
- 1. Regulatory compliance documentation missing or incomplete (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software verification shows failed feature
One or more features listed in the New Software Verification report show 'Failed' status after upgrade.
Ranked causes
- 1. Incomplete or corrupted software installation (unknown)
Verify all features in New Software Verification report
- Test point
- New Software Verification screen items
- Expected
- All features show OK status
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Storing SonoView images from legacy Voluson systems
Need to store SonoView images from Voluson 730/Expert/Pro/ProV to Voluson E-Series systems
Ranked causes
- 1. Archive system is different; SonoView not available on Voluson E-Series (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System boot selection timeout
Boot screen disappears after 3 seconds without user input; system boots to Voluson (normal operation).
Ranked causes
- 1. No menu selection within 3 second timeout window (common)
Monitor for Boot Screen appearance
- Test point
- System startup display
- Expected
- Boot Screen with menu options appears
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System covers or frame panels missing caution
Board covers or frame panels removed during operation
Ranked causes
- 1. Covers removed for service procedures (occasional)
Verify all board covers and frame panels in place
- Test point
- Physical inspection of system enclosure
- Expected
- All covers securely fastened
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System covers or panels not secure caution
Board covers or frame panels are removed, loose, or missing during operation
Ranked causes
- 1. Covers or panels removed for service or inspection (common)
Visually inspect all board covers and frame panels
- Test point
- All accessible enclosure areas
- Expected
- All covers in place, no gaps, all fasteners or latches secure
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not boot after power cycle
System fails to start when power is recycled quickly after shutdown
Ranked causes
- 1. Power recycled too quickly (common)
Wait minimum time after shutdown
- Test point
- System power state
- Expected
- System fully powered down for at least 10 seconds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System shuts down unexpectedly or fails to shut down cleanly caution
System does not respond to shutdown command or power-off procedure hangs
Ranked causes
- 1. Probe disconnected during live scan (Write mode) (occasional)
Verify image is frozen before power-off
- Test point
- Monitor display and control console status
- Expected
- System in freeze or read mode, not in Live Scan Write mode
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System software upgrade required
Need to update system software from one extension version to another (e.g., Ext.2 to Ext.3) on Voluson E-Series BT15-BT21
Ranked causes
- 1. Software extension outdated (common)
Backup all presets and database
- Test point
- System Setup > Backup function
- Expected
- Backup completed successfully
10 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
USB device not recognized after connection
External USB storage device (memory stick or external hard disk) not automatically detected by system after connection to Voluson E-Series
Ranked causes
- 1. Device driver not automatically installed or device not detected by Windows (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Whitelisting feature not activated after upgrade when Advanced Security Features enabled
The Whitelisting feature is not active or not properly configured after software upgrade when Advanced Security Features option is enabled.
Ranked causes
- 1. Whitelisting not reactivated after software upgrade (common)
Verify Whitelisting is activated
- Test point
- System Setup - Administration - Whitelisting
- Expected
- Whitelisting feature shows active/enabled status
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Gray envelope icon | New software update package available for download and installation | System detects available firmware or software update from GE | info |
Gray GE remote icon | Remote access idle; no active connection | Normal state when remote service access is not in use | info |
Gray indicator light | Network cable disconnected or no network signal on connected cable | Network cable unplugged or broken connection | error |
Gray modem icon | Email/MMS function is not configured | Cellular modem not set up for email or MMS use | warning |
Green indicator light | Network cable connected to active network | Network cable properly connected to powered network port | info |
Green modem icon | Email/MMS successfully sent | Cellular modem connection active and message sent successfully | info |
Red GE remote icon | Remote access active in Disrupted State; all functions enabled | GE technician actively connected with full Virtual Console and Service Desktop access | critical |
Red modem icon | Email/MMS not sent (failed) | Cellular modem connection failed or message delivery unsuccessful | error |
Yellow GE remote icon | Remote access pending; connection being established | Remote access initiated but Virtual Console and service functions not yet active | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI