GE Healthcare Vivid iq: troubleshooting and error codes
The ServiceForge knowledge base holds 48 documented symptoms for the GE Healthcare Vivid iq, with 68 ranked causes, 78 diagnostic checks with test points and expected values, and 2 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Excessive leakage current detected patient safety
Leakage current measurement exceeds acceptable limits during safety testing
Ranked causes
- 1. AC/DC adapter or cable defective (common)
- 2. Ground connection broken, frayed, or intermittent on power cord (common)
- 3. Ground studs loose or star washers missing (occasional)
- 4. Wiring damaged, poorly crimped, or poorly connected (occasional)
- 5. Wall outlet not grounded or has wiring abnormalities (occasional)
- 6. Neutral and ground wires reversed at outlet (rare)
- 7. Probe defective or has excessive leakage (common)
- 8. System connector has bent pins or poor connections (occasional)
- 9. Peripheral device defective (occasional)
- 10. Multiple internal or on-board components defective (rare)
Visually inspect AC/DC adapter and cable
- Test point
- AC/DC adapter and cable assembly
- Expected
- No defects, damage, or corrosion visible
9 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe Check fails with element integrity warning patient safety
System displays 'Probe test did not pass' dialog indicating element integrity check failed; image quality may be suboptimal; scanning can continue but requires evaluation
Ranked causes
- 1. Dirty probe lens or surface contamination (common)
- 2. Probe element degradation or weak elements (occasional)
- 3. Probe port malfunction or MST board failure (occasional)
Visual inspection of probe lens and surface
- Test point
- Probe lens
- Expected
- Clean, no gel residue or contamination
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
B-Mode (2D) ultrasound operation not functioning
B-Mode imaging does not display, shows poor image quality, or basic controls do not affect 2D mode operation.
Ranked causes
- 1. Probe not properly recognized or connected (common)
- 2. B-Mode channel or transmit/receive element failure (occasional)
Verify probe is properly connected and recognized
- Test point
- System startup and probe display on screen
- Expected
- Probe type displayed and available in B-Mode selection
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrical shock hazard during setup high voltage
Risk of electrocution to technician during system setup and service
Ranked causes
- 1. AC ground line opened by test instrument (occasional)
- 2. Unit not connected to properly grounded outlet or adapter used (occasional)
Verify power outlet has proper ground
- Test point
- Power outlet
- Expected
- Three-prong grounded outlet
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard or operator panel keys not responding
Physical keys or control panel buttons do not respond to input or function improperly during operation.
Ranked causes
- 1. Keys contaminated with gel, debris, or dried disinfectant (common)
- 2. Keyboard or control panel hardware failure (occasional)
Inspect physical keys for contamination
- Test point
- All physical keys and control panel buttons
- Expected
- Keys clean and free of gel, debris, or disinfectant residue
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD display not functioning properly
LCD screen shows no image, poor image quality, scratches, bad pixels, brightness or contrast issues.
Ranked causes
- 1. LCD physical damage or bad pixels (occasional)
- 2. Contrast or Brightness control malfunction (occasional)
Test Contrast and Brightness control operation
- Test point
- LCD display response to control adjustments
- Expected
- Display brightness and contrast change smoothly with control inputs
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Peripheral or optional device not functioning
Media drive, printer (B/W or Color), DICOM, ECG, Footswitch, or DVD peripheral does not work or produces poor output.
Ranked causes
- 1. Peripheral dirty, contaminated with gel, or needs cleaning (common)
- 2. Peripheral hardware failure (occasional)
Inspect peripheral for visible dirt or gel contamination
- Test point
- Media drive, printer, DVD, or other peripheral
- Expected
- Peripheral surfaces clean, no visible gel or debris
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not power on or fans/peripherals do not start
Ultrasound system fails to power up; fans and peripherals do not activate; warning or error messages appear during startup.
Ranked causes
- 1. Mains cable disconnected or damaged (common)
- 2. Internal battery not charged (occasional)
Inspect mains cable connection at wall and system
- Test point
- Mains cable connectors
- Expected
- Cable securely connected with no visible damage
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not recognize probes
Ultrasound system fails to detect or acknowledge connected probes during startup or operation.
Ranked causes
- 1. Probe connectors dirty or damaged (common)
- 2. Probe hardware failure (occasional)
Visually inspect probe connectors and sockets
- Test point
- Probe connection sockets on system; probe connector pins
- Expected
- No dust, bent pins, or visible damage on connectors or pins
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System overheating or temperature warning caution
System displays overheating warning, thermal shutdown occurs, or temperature log shows concerning trends.
Ranked causes
- 1. Cooling fans blocked or internal airflow obstructed (common)
- 2. Internal thermal sensor failure or temperature calibration error (rare)
Visually verify cooling fan operation
- Test point
- System cooling fan during operation
- Expected
- Cooling fan spinning at normal speed, no grinding or unusual noise
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System safely isolated for high-voltage service high voltage
Procedure to control AC power and discharge stored energy before servicing high-voltage components
Ranked causes
- 1. Need to safely de-energize system before accessing high-voltage circuits (common)
- 2. Capacitor discharge time required before accessing circuits (common)
Turn off the breaker
- Test point
- Equipment breaker
- Expected
- OFF position
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Backup Patient Archive and Presets before software update caution
Procedure to backup system settings and patient data prior to loading software from USB Flash Drive
Ranked causes
- 1. Error or power loss during software installation (occasional)
Document current Vivid iq settings and parameters on paper
- Test point
- System configuration
- Expected
- Complete record of all settings documented
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery absent or not installed
Battery icon indicates battery is absent; system running on AC adapter only
Ranked causes
- 1. Battery not installed or removed (common)
Verify battery is present and properly seated
- Test point
- Battery bay in scanner or cart
- Expected
- Battery fully inserted without gaps; AC adapter icon absent from title bar
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery has reached end of life patient safety
Red sad icon on battery; system displays end-of-life warning; battery disabled for scanning
Ranked causes
- 1. Battery degradation complete; end of service life reached (common)
Check battery information window for end-of-life status
- Test point
- Battery information dialog and warning messages
- Expected
- Red sad icon and warning message requesting battery replacement
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery health below 60%
Yellow exclamation mark appears on battery icon; battery approaching end of life
Ranked causes
- 1. Battery has degraded and requires replacement (common)
View battery health status in battery information window
- Test point
- System title bar battery icon and battery information dialog
- Expected
- Yellow or red health icon displayed; capacity shown in percentage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cable harnesses loose or strain relief damage caution
Internal cable harnesses are loose, improperly seated in connectors, or strain relief boots on cables are damaged or missing.
Ranked causes
- 1. Cable strain relief damage or loose connector seating (occasional)
Visually inspect internal cable harnesses and connectors
- Test point
- All internal cable connections, particularly footswitch and probe assemblies
- Expected
- Cables properly seated, strain reliefs intact, no loose connections
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Height adjustment of mounting platform
Adjust the height of the cart mounting-platform to desired position using the release lever
Ranked causes
- 1. Height adjustment required (unknown)
Raise the Release lever to unlock platform
- Test point
- Release lever
- Expected
- Platform unlocked and movable
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Color Flow (CF) Mode not functioning
CF-Mode imaging does not display, shows poor image quality, or controls do not affect Color Flow mode operation.
Ranked causes
- 1. CF-Mode option not available or hardware failure (occasional)
Verify CF-Mode option is enabled in system
- Test point
- System configuration and mode selection menu
- Expected
- CF-Mode option is present and selectable
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display shows warning or error messages at startup
Ultrasound system displays warning or error messages on screen during power-up sequence.
Ranked causes
- 1. Error log entries indicating system faults (unknown)
Access and review system error log
- Test point
- Diagnostic software error log
- Expected
- Error log shows no active faults or provides information on source of warning message
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Distance measurement inaccuracy
Measurement mode produces incorrect distance results compared to physical measurements using a graduated rule.
Ranked causes
- 1. System calibration drift or error (occasional)
Perform distance measurement accuracy test
- Test point
- Known test object or phantom
- Expected
- Measured distance within ±5% of physical measurement percent, ±5%
- Tools
- Graduated rule, test phantom
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Doppler Mode (PW/CW) not functioning
Doppler imaging including Pulsed Wave (PW) or Continuous Wave (CW) modes do not display, show poor quality, or controls do not affect operation.
Ranked causes
- 1. Doppler mode not available or hardware failure (occasional)
Verify Doppler mode is available in system
- Test point
- System configuration and mode selection menu
- Expected
- Doppler mode (PW and CW if applicable) is present and selectable
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG peripheral high leakage current patient safety
Leakage current from ECG peripheral cable or connector exceeds acceptable limits
Ranked causes
- 1. ECG cable damaged or has poor connections (common)
Visually inspect ECG cable assembly
- Test point
- ECG cable from connector to probe
- Expected
- No visible damage, cuts, fraying, or poor solder connections
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
EMI/RFI interference affecting system or signal quality caution
System exhibits erratic behavior, signal noise, image artifacts, or loss of signal; scanning produces poor quality images or intermittent operation.
Ranked causes
- 1. Missing internal EMI shielding, covers, or hardware (occasional)
Visually inspect internal EMI shielding and covers
- Test point
- All internal EMI shields, covers, air flow panels, and fastening screws
- Expected
- All shielding and covers in place and properly secured
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ESD damage during circuit work caution
Electrostatic discharge may damage circuits if wrist strap worn during live circuit work
Ranked causes
- 1. ESD wrist strap worn during work on live circuits with >30 V (occasional)
Verify circuit voltage before work
- Test point
- Live circuit
- Expected
- Measure voltage to determine if >30 V peak is present
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
External connectors showing damage or corrosion
External input/output connectors show physical damage, corrosion, bent pins, or loose connections affecting peripheral operation.
Ranked causes
- 1. Connector corrosion, oxidation, or physical damage (occasional)
Visually inspect all external I/O connectors
- Test point
- All external connectors (USB, ECG, probe ports, Ethernet, HDMI, etc.)
- Expected
- Connectors clean, pins straight, no visible corrosion or damage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fan failure warning caution
System displays fan failure warning and shuts down automatically when fan speed is extremely low or stopped while temperatures are elevated
Ranked causes
- 1. CPU blower failure or obstruction with elevated temperatures (common)
Verify CPU blower speed
- Test point
- System diagnostics
- Expected
- Fan operating at normal speed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image degradation with dark areas at shallow depths patient safety
Visual inspection shows localized dark area at shallow depths indicating compromised probe elements; affects image quality
Ranked causes
- 1. One or more probe elements non-functional or weak (common)
Visual inspection of ultrasound image at shallow depths
- Test point
- Image display on monitor
- Expected
- Uniform image intensity across field; no dark areas
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image intensity reduced at shallow depths on phased array probe patient safety
Phased array probe shows reduced image intensity particularly at shallow scanning depths despite probe operation
Ranked causes
- 1. Loss of multiple probe elements reduces shallow-depth sensitivity (common)
Examine image intensity pattern across phased array field at shallow depths
- Test point
- Phased array probe ultrasound image
- Expected
- Uniform intensity across field at shallow scanning depths
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
InSite remote communication not functioning
InSite remote communication system is not operating or two-way communications are not established for warranty and contract customers.
Ranked causes
- 1. InSite communication or network connectivity issue (unknown)
Verify InSite software status and network connectivity
- Test point
- InSite application status and system network interface
- Expected
- InSite software running and enabled; network connection active
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD display arm movement not functioning
LCD screen arm cannot swivel left/right, fold to locked position, adjust up/down, or release for adjustment as specified.
Ranked causes
- 1. LCD arm mechanical joint stiff, contaminated, or damaged (occasional)
Test LCD arm movement through all positions
- Test point
- LCD arm swivel left/right, fold locking, forward/backward adjustment, up/down adjustment
- Expected
- Smooth movement in all directions, arm locks securely in folded position
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
M-Mode imaging not functioning
M-Mode (time-motion mode) does not display, shows poor image quality, or controls do not affect M-Mode operation.
Ranked causes
- 1. M-Mode hardware or channel failure (occasional)
Test M-Mode operation with probe
- Test point
- M-Mode display during operation
- Expected
- M-Mode image displays with responsive controls
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
New ultrasound system fails leakage current test patient safety
On a new system, leakage current exceeds limits and cannot be corrected by standard troubleshooting
Ranked causes
- 1. Manufacturing defect in new unit (rare)
Complete all troubleshooting steps
- Test point
- Leakage current measurement
- Expected
- All standard checks performed without resolution
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Operator control panel has damaged or missing items
Control panel shows physical damage, missing buttons, cracked panels, or non-functional mechanical elements.
Ranked causes
- 1. Control panel physical damage or component failure (occasional)
Inspect control panel for physical damage or missing components
- Test point
- All buttons, knobs, switches on operator control panel
- Expected
- All components present, no visible damage or cracks
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Permanent error detected on battery patient safety
Permanent error icon on battery; battery damaged or reached end of life
Ranked causes
- 1. Battery is damaged or has reached end of life (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power and System Status Indicator lights not functioning
Power and System Status Indicator lights on control panel are not illuminating or not responding as expected.
Ranked causes
- 1. Indicator light bulb or LED burned out or disconnected (occasional)
Observe all indicator lights during system startup and operation
- Test point
- All Power and System Status Indicator lights on control panel
- Expected
- Power light illuminates on; System Status lights respond appropriately to system state
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe holders contaminated with ultrasound gel
Probe holders have excess accumulated ultrasound gel that is dried or sticky, affecting probe storage and cleanliness.
Ranked causes
- 1. Buildup of dried or excess ultrasound gel in probe holders (common)
Inspect probe holders for gel contamination
- Test point
- Probe holder surfaces and interior
- Expected
- Holders clean and free of dried or excess gel
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe leakage current test patient safety
Verification of probe leakage current compliance; variations depend on probe design, line voltage, and test lead placement
Ranked causes
- 1. Normal probe leakage variation due to design or test conditions (common)
Determine probe contact type
- Test point
- Probe specification label
- Expected
- Body surface contact or inter-cavity type identified and correct in test checklist
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe showing signs of wear or damage patient safety
Probe has visible wear, bent connector pins, missing pins, cracks, or other physical damage that may indicate non-functional or unsafe condition.
Ranked causes
- 1. Probe physical damage or connector pin failure (occasional)
Perform visual inspection of probe connector and elements
- Test point
- Probe connector pins, cable, and acoustic lens surface
- Expected
- No visible wear, bent pins, cracks, missing pins, or surface damage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe test fails or produces invalid results
Probe test takes longer than expected or returns errors; lens or cap surface contaminated with dirt, water, or coupling gel
Ranked causes
- 1. Lens or cap surface contaminated with dirt, water, or coupling gel (common)
Inspect lens or cap surface for contamination
- Test point
- Probe lens or cap
- Expected
- Clean surface, free of dirt, water, or coupling gel
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System battery damaged or non-functional patient safety
Internal system battery is leaking, deformed, discolored, emitting odor, or not holding charge; system cannot operate on battery power when AC unavailable.
Ranked causes
- 1. Battery degradation, physical damage, or failure after three years (occasional)
Visually inspect internal system battery for damage
- Test point
- System battery (if applicable and accessible)
- Expected
- Battery intact, not leaking, not deformed, not discolored, no odor
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System damaged after cold or hot transport caution
Unit arrives at site at temperature below 3°C or above 40°C; powering on immediately without acclimation causes equipment damage
Ranked causes
- 1. Unit not acclimated to operating temperature after transport (common)
Measure current ambient temperature at installation site
- Test point
- Room thermometer
- Expected
- Between 3°C and 40°C
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System exterior has cracks, dents, or physical damage caution
Visible damage to system enclosure including dents, scratches, or cracks that may affect function or safety.
Ranked causes
- 1. Physical impact or rough handling (occasional)
Visually inspect system exterior for physical damage
- Test point
- All external surfaces of system cabinet
- Expected
- No visible cracks, dents, or damage affecting appearance or function
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System labeling missing or unreadable patient safety
System identification labels or safety markings are missing, faded, or damaged making them unreadable.
Ranked causes
- 1. Label fading, wear, or physical damage (occasional)
Verify readability of all system labels
- Test point
- System identification and safety labels
- Expected
- All labels present, clearly visible, and text legible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System malfunction due to missing covers or panels caution
System performance and cooling compromised when board covers and frame panels are not in place
Ranked causes
- 1. Board covers or frame panels removed or not securely installed (occasional)
Visually inspect all board covers and frame panels
- Test point
- All exposed boards and frame areas
- Expected
- All covers and panels in place and securely fastened
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System overheating warning caution
System displays overheat warning and may shut down automatically when internal temperatures exceed thresholds
Ranked causes
- 1. CPU or MST board temperature exceeds threshold due to inadequate cooling (common)
Verify CPU temperature reading
- Test point
- System diagnostics
- Expected
- Below 100°C (v203+) or 90°C (v204.96.0+), v203+: <100°C; v204.96.0+: <90°C
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System requires cleaning and disinfection after patient contact patient safety
Routine cleaning and disinfection of system surfaces, keyboard, and peripherals needed after patient scanning to prevent infection transmission.
Ranked causes
- 1. Standard infection control protocol requires regular cleaning (common)
Verify system power is off and peripherals disconnected before cleaning
- Test point
- System power state and peripheral connections
- Expected
- System powered down, mains cable disconnected, peripherals removed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ultrasound probe acoustic output exposure hazard patient safety
Technician exposure to ultrasound energy from probe during service or testing
Ranked causes
- 1. Prolonged or unnecessary exposure to ultrasound energy from probe (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unknown error detected on battery
Unknown error icon appears on battery status; system may be unstable
Ranked causes
- 1. Battery sensor error or temporary fault (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Software installation is not started | Application software installation failed to initiate | Software compatibility check failed; system rejected installation package | error |
The package present in media is not compatible | Software installation media contains incompatible package for this Vivid iq system | Attempted installation of software package not compatible with current software version on Vivid iq | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community