GE Healthcare Vivid E9: troubleshooting and error codes
The ServiceForge knowledge base holds 68 documented symptoms for the GE Healthcare Vivid E9, with 151 ranked causes, 145 diagnostic checks with test points and expected values, and 2 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
System pings and verifies but does not send data
Network connectivity verified but data transmission fails; send operation does not complete
Ranked causes
- 1. Target device offline or in error state (occasional)
- 2. Device configuration incorrect (common)
- 3. Print spooler queue blocked (occasional)
- 4. VIVID E9 connectivity configuration incorrect (occasional)
- 5. Printer does not support selected format or image type (occasional)
- 6. Storage device does not support selected image type (occasional)
- 7. Worklist broker configuration incorrect (occasional)
Verify target device status
- Test point
- Target device
- Expected
- Device online and operational
4 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No video on LCD display caution
Display is blank, too dim, has dimmed areas, or shows no video output
Ranked causes
- 1. HV LCD inverter failure or burnt (common)
- 2. Video cabling fault (occasional)
- 3. Digital Video Card on PC damaged by burnt inverter (occasional)
- 4. Video Adapter inside BEP defective (occasional)
- 5. Back End Processor defect (rare)
- 6. Upper Panel Assembly defect (rare)
Visually inspect video cabling
- Test point
- All video signal cables
- Expected
- Cables properly connected, no visible damage
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Video display too dim, has dimmed areas, or no video on LCD panel caution
LCD display dark, uneven illumination, or completely black; no image visible despite system operation
Ranked causes
- 1. HV LCD inverter failure or burn-out (common)
- 2. Damaged or loose display cabling within system (occasional)
- 3. Digital video card on PC damaged by inverter burn-out (occasional)
- 4. Video adapter card in back end processor failure (occasional)
- 5. Back end processor failure or multiple display interface faults (rare)
- 6. Upper panel assembly defect affecting LCD connector or backlight (rare)
Inspect display cable and connections
- Test point
- LCD display cable connectors at BEP and display
- Expected
- Cable fully seated, no bent pins, no visible damage to insulation or connector
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device verify failed
System cannot verify device via network protocol; device not responding to verification request
Ranked causes
- 1. Device does not support Verify protocol (occasional)
- 2. Incorrect port or AE title configuration (occasional)
- 3. Device not running or in error state (occasional)
- 4. Network communication issue (occasional)
- 5. Software issue (occasional)
Verify device supports Verify protocol
- Test point
- Device documentation
- Expected
- Device supports Verify
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image artifacts in one or more areas of the image
White vertical lines, rain, snow cones appearing in B-mode or B-Flow; may be port-related, probe-specific, or system-wide
Ranked causes
- 1. Front end board failure or power supply issue (common)
- 2. Defective probe connector or port (occasional)
- 3. Defective or failed probe (occasional)
- 4. Software corruption in presets or default settings (occasional)
- 5. Artifact may appear only with specific mode, probe, or preset (occasional)
Swap DRX boards between ports
- Test point
- Image display
- Expected
- If artifact moves to different port with board swap, defect is in the board; if artifact stays in same area, defect is in probe or connector
3 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No audio output
No sound from speakers or audio amplifier; audio signal may or may not reach the operator panel
Ranked causes
- 1. BEP software fault or BEP hardware defect (occasional)
- 2. Upper OP Assembly audio amplifier defect (occasional)
- 3. Internal I/O board defect (occasional)
- 4. Speaker impedance mismatch or speaker failure (occasional)
- 5. Audio amplifier on Upper OP Assembly defective (rare)
Verify volume settings
- Test point
- Application and Windows volume controls
- Expected
- Volume enabled and set to adequate level
3 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No audio output from system or speaker
Operator panel or headphones output silent; no audio signal detected during operation or during audio test
Ranked causes
- 1. Audio volume settings too low or muted in software or Windows (common)
- 2. Audio output failure in back end processor (occasional)
- 3. Audio amplifier failure in operator panel upper assembly (occasional)
- 4. Internal I/O board audio amplifier failure (occasional)
- 5. Speaker impedance mismatch or speaker driver failure (occasional)
Check volume settings
- Test point
- Application audio menu and Windows system settings
- Expected
- Volume level above minimum; output not muted in either location
3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Noisy ultrasound image
Image artifacts or noise visible on display, often from electromagnetic interference (EMI) sources in the facility
Ranked causes
- 1. External EMI source (medical laser, scanner, cautery, computer, fan, gel warmer, microwave, light dimmer, portable phone, broadcast station) (common)
- 2. Poor power supply grounding or inadequate power outlet (common)
- 3. Missing or loose RF shielding, covers, gaskets, or screws (common)
- 4. Improper cable routing or non-GE specified interconnect cables/peripherals (occasional)
- 5. Cellular phone transmission near unit (rare)
Try another VIVID E9 at same location
- Test point
- Ultrasound image display
- Expected
- Noise present on new system indicates external source; noise only on original unit indicates unit defect
4 more checks with pass and fail actions, 12 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CD/DVD drive not reading media
System unable to read or write CDs or DVDs; drive may have failed or media compatibility issue.
Ranked causes
- 1. Unsupported media capacity or speed (common)
- 2. CD/DVD drive hardware failure (occasional)
- 3. Defective or corrupted disk media (occasional)
- 4. BEP (Back End Processor) failure (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image artifacts visible in ultrasound display
White vertical lines, rain, or snow cone artifacts appear in one or more areas of the image; may be isolated to specific mode or probe.
Ranked causes
- 1. Front End board or DRX failure (common)
- 2. Probe defect or port-related failure (occasional)
- 3. Software corruption in presets or system files (occasional)
- 4. Reload software (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Network verify not supported or fails
Cannot verify connectivity with network device; device does not respond to verify request
Ranked causes
- 1. Device does not support DICOM Verify (occasional)
- 2. Port or AE title configuration incorrect (common)
- 3. Device offline or in error state (occasional)
- 4. Network configuration or software issue (occasional)
Confirm device supports Verify
- Test point
- Device specification or manual
- Expected
- Device supports DICOM Verify operation
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System pings and verifies but does not send
Network connectivity verified but data transmission fails; images or jobs stuck in queue
Ranked causes
- 1. Device not running or in error state (occasional)
- 2. Device configuration incorrect (occasional)
- 3. Spooler queue full or stuck (occasional)
- 4. VIVID E9 connectivity configuration issue (occasional)
Verify device is powered and running
- Test point
- Device status
- Expected
- Device online and responding
6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit cannot be safely moved or transported patient safety
Console or monitor in free position; unit positioned on unsuitable slope; loose items or probes present
Ranked causes
- 1. Console or monitor not in locked park position (common)
- 2. Brake not engaged or swivel not locked (common)
- 3. Probes not secured; loose items on unit (occasional)
- 4. Unit positioned on ramp steeper than 10 degrees (occasional)
Verify console in locked park position
- Test point
- Keyboard console
- Expected
- Console locked in park position; does not move freely
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CD or DVD drive does not read or write media caution
Drive fails to recognize inserted disc, cannot write to blank media, media not detected by drive
Ranked causes
- 1. Disc media capacity or speed not supported by drive (common)
- 2. CD/DVD drive hardware failure (occasional)
- 3. Back end processor communication or interface failure (occasional)
Verify media type and capacity
- Test point
- Media packaging and disc surface
- Expected
- CD is 700 MB capacity; disc type (CD-R, CD-RW, DVD-R) matches drive specification
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CD or DVD image does not read on external computer
Disc written on system cannot be opened or read when placed in laptop or external drive; disc corruption or writing error
Ranked causes
- 1. Burned disc is corrupted or unreadable (common)
- 2. CD/DVD write drive error or burn speed incompatibility (occasional)
- 3. CD-RW or DVD-R drive failure (occasional)
Test disc on external computer
- Test point
- Laptop or external DVD/CD drive
- Expected
- Image files readable and data integrity verified without errors
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
EMI affecting system after installation or repairs
Electromagnetic interference observed after service work, installation, or repair completed.
Ranked causes
- 1. RF covers, gaskets, or screws not reinstalled or loose after service (common)
- 2. Cable shielding or ferrite protection removed or damaged during service (occasional)
- 3. GE-specified cables and peripherals not used; non-standard interconnects installed (occasional)
Verify all RF covers are installed and all screws tightened
- Test point
- RF shielding covers and fasteners
- Expected
- All covers in place and screws snug
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch panel not responding
Touch panel does not respond to any operator input or touches
Ranked causes
- 1. Touch panel calibration fault (common)
- 2. Software fault (common)
- 3. Upper Panel Assembly hardware defect (occasional)
Calibrate touch panel
- Test point
- Touch panel calibration utility
- Expected
- Calibration routine accessible and completes
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch panel not responding to operator input
Operator touches panel but no response, keys do not activate, touch input ignored
Ranked causes
- 1. Touch panel requires recalibration (common)
- 2. Software fault or communication error (occasional)
- 3. Upper panel assembly defective or touch sensor failure (occasional)
Calibrate touch panel
- Test point
- Touch panel response during calibration
- Expected
- Panel responds to touches during calibration routine and registers coordinates
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to scan patient safety
Scanning function not operational; system unable to acquire ultrasound images.
Ranked causes
- 1. Probe not connected or incorrectly connected (common)
- 2. Application fault requiring restart (occasional)
- 3. Main power supply failure or fault (occasional)
Verify probe connections are correct and fully seated
- Test point
- Probe connectors
- Expected
- All probes properly connected
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit damaged or slides during vehicle transport caution
Unit moves or is not secured during transit; vibration damage occurs
Ranked causes
- 1. Unit not strapped or cushioned during loading (common)
- 2. Unit not secured inside vehicle with straps (common)
- 3. Vehicle driven at excessive speed or aggressively (occasional)
Verify unit is cushioned on lift
- Test point
- Unit surface and loading platform
- Expected
- Unit surrounded by cushioning material before strapping
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
USB Footswitch not functioning
USB footswitch not responding or recognized by system during operation.
Ranked causes
- 1. USB footswitch not plugged into correct rear USB connector (common)
- 2. Footswitch not enabled in system configuration (occasional)
- 3. Footswitch hardware failure (rare)
Verify USB footswitch is connected to rear connector
- Test point
- Rear USB connectors
- Expected
- Footswitch cable connected to rear USB port
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
XY Park Lock not engaging during operation caution
Lock fails to engage when attempting to park console
Ranked causes
- 1. Brakes not applied (common)
- 2. Console not in normal resting position (common)
- 3. Improper engagement technique (common)
Verify brakes are applied
- Test point
- Brake control
- Expected
- Brakes fully applied; device cannot move
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: LCD monitor brightness and display settings
19" LCD monitor brightness, contrast, color temperature, gamma, and smoothing require adjustment for optimal display performance under varying light conditions or after factory reset.
Ranked causes
- 1. Factory defaults require verification or adjustment for operating conditions (common)
- 2. Advanced color and display parameters out of specification (occasional)
Verify monitor model selection in system config
- Test point
- Config > Service dropdown menu
- Expected
- Correct monitor (17" or 19") is selected
3 more checks with pass and fail actions, 11 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot connect via network to device
System cannot send data to or communicate with network devices (printer, PACS, worklist broker)
Ranked causes
- 1. Network connectivity lost (common)
- 2. Device configuration or protocol mismatch (occasional)
Ping remote device
- Test point
- Laptop network, target device IP address
- Expected
- Ping reply received
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot log on to VIVID E9 as ADM user
ADM login fails; cannot access service adjustments or configuration settings
Ranked causes
- 1. Incorrect ADM password entered (common)
- 2. Custom ADM password set by administrator (occasional)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High system temperature caution
System displays high temperature warning or initiates shutdown sequence when internal sensors exceed thermal limits.
Ranked causes
- 1. Card Rack or Front End sensors exceed 78°C (warning) or 80°C (shutdown) (common)
- 2. DRX_TOP sensors exceed 87°C (warning) or 90°C (shutdown) (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image noise or artifacts present
Noisy ultrasound images; visible artifacts or interference in image quality.
Ranked causes
- 1. Electromagnetic interference from external sources (medical lasers, scanners, cauterizing guns, computers, monitors, fans, gel warmers, microwave ovens, light dimmers, portable phones, broadcast stati (common)
- 2. Poor power grounding or shared power circuits (common)
Try different power outlet to verify noise source
- Test point
- Power outlet and facility ground
- Expected
- Noise disappears or significantly reduces at new outlet
6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD monitor cannot be repositioned or locked caution
Monitor position fixed or cannot be moved; cannot secure in parked position
Ranked causes
- 1. Release knob not turned counterclockwise to unlock (common)
- 2. Release knob not turned clockwise to lock (common)
Verify release knob is in counterclockwise position
- Test point
- LCD monitor release knob
- Expected
- Knob fully counterclockwise; monitor moves freely
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe not recognized
System does not recognize ultrasound probe when connected to system port
Ranked causes
- 1. GRLY board defect affecting all probes (occasional)
- 2. Individual probe defect (common)
Test probes on all ports individually
- Test point
- Each probe inserted in each system port
- Expected
- Probe recognized by system
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe not recognized or probe recognition fails on one or more ports
System does not detect probe when plugged in; probe icon missing or marked as fault; affected port or all ports
Ranked causes
- 1. Probe connector failure or port defect (common)
- 2. Defective or failed probe (occasional)
Test all probes on all ports
- Test point
- System probe detection menu
- Expected
- Each probe recognized and displayed in probe list when inserted into each port
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System temperature too high caution
Internal system temperature exceeds normal operating range; cooling system insufficient to maintain proper temperature.
Ranked causes
- 1. Air filters too dusty; insufficient cooling airflow (common)
- 2. Fan worn out or failed (occasional)
Inspect air filters for dust and debris
- Test point
- System air intake filters
- Expected
- Filters clean with good airflow
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
TEE probe temperature too high patient safety
System displays alert that transesophageal echocardiography probe temperature exceeds safe limits.
Ranked causes
- 1. Defective TEE probe (common)
- 2. GFI board malfunction (occasional)
Test temperature with all available TEE probes
- Test point
- TEE probe connector
- Expected
- Temperature normal with each probe
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Top console cannot be lowered caution
Top console moves upward normally but cannot move downward; may require manual assistance.
Ranked causes
- 1. Control switch stuck in up position (common)
- 2. Gas spring failing inside Z mechanism (occasional)
Test top console upward movement
- Test point
- Top console
- Expected
- Moves upward smoothly
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wrong key activated on touch panel
Touching one key on touch panel activates a different key or function; inconsistent response to operator input
Ranked causes
- 1. Touch panel requires calibration (common)
- 2. Upper panel assembly defective or touch sensor failure (occasional)
Calibrate touch panel
- Test point
- Touch panel response during calibration routine
- Expected
- User can touch reference points and panel responds with correct coordinate mapping
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
XY Lock not functioning caution
Mechanical lock mechanism for XY (probe holder) position not engaging or releasing.
Ranked causes
- 1. Motor controller firmware needs reset via Fix Lock function (common)
- 2. Lock mechanism mechanically stuck or failed (occasional)
Run XYZ Motor Controller Test and press Fix Lock button
- Test point
- Lock Engine audible response
- Expected
- Lock Engine engagement sound heard for few seconds
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Z-axis movement limited or abnormal caution
Top Console cannot lower properly or moves with excessive effort; vertical positioning mechanism malfunction.
Ranked causes
- 1. Control switch stuck in Up position (occasional)
- 2. Z Mechanism gas spring failure (common)
Attempt to lower Top Console
- Test point
- Z-axis motor and control
- Expected
- Console lowers smoothly with motor assistance
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
BEP is hanging or unresponsive
BEP (Back End Processor) does not respond to input or commands
Ranked causes
- 1. BEP software hang or freeze (unknown)
Attempt forced restart
- Test point
- ON/OFF switch on Operator Panel
- Expected
- System performs forced restart
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Front End DC Offset
Front End Alignment calibration required after software reload to ensure proper ultrasound transducer offset and signal integrity
Ranked causes
- 1. Software reload performed on system (common)
Execute DC Offset Calibration procedure
- Test point
- Front End electronics
- Expected
- Calibration completed successfully per Section 6-6
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: System settings and parameters not recorded
Settings configuration not backed up before software update or maintenance
Ranked causes
- 1. No documented record of current settings (common)
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Touch screen position
Touch sensitive area misaligned with graphic artwork on display; touches register at incorrect screen locations
Ranked causes
- 1. Touch screen position drift or initial misalignment (unknown)
Access Operator Panel Test Dialog
- Test point
- Config menu
- Expected
- Operator Panel Test Dialog opens
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: XY Lock adjustment caution
Optimize XY Lock tension for minimal slack in transport mode and proper engagement without excessive force
Ranked causes
- 1. XY Lock requires adjustment for optimal tension (common)
Rotate adjustment screw in 1/4 turn increments clockwise
- Test point
- Hex screw behind XY Assembly
- Expected
- Lock engages with minimal slack and reasonable force required
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot connect to network device
System cannot establish connection to networked device such as printer or PACS server
Ranked causes
- 1. Network connectivity issue between devices (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Configure DHCP on Vivid E9
Enable or disable DHCP for automatic or manual IP address assignment on the ultrasound system network interface.
Ranked causes
- 1. DHCP configuration required for network connectivity (common)
Access Network Settings from TCP/IP screen
- Test point
- Network Connections window
- Expected
- Network Connections dialog displayed
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Configure Remote Archive network connection
Set up network communication parameters to connect Vivid E9 to a remote archive server or computer.
Ranked causes
- 1. Remote Archive network information must be configured (common)
Access TCP/IP screen Remote Archive Setup area
- Test point
- TCP/IP configuration screen
- Expected
- Remote Archive Setup fields visible
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Difficult to lock and release alphanumeric keyboard
Alphanumeric keyboard drawer locking and release mechanism requires excessive force or is difficult to operate.
Ranked causes
- 1. Operator pressing on leftmost section where mechanical advantage is poor (common)
Locate and use rightmost section of keyboard for lock/release
- Test point
- Keyboard drawer release mechanism
- Expected
- Mechanism engages/disengages with minimal force
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Difficult to release or lock alphanumeric keyboard caution
Keyboard release/lock mechanism requires excessive force or does not engage properly.
Ranked causes
- 1. Force applied at wrong location on keyboard drawer (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
High system temperature error
System displays high temperature warning or initiates shutdown sequence
Ranked causes
- 1. System temperature exceeds safe operating range (unknown)
Monitor system temperature at first warning threshold
- Test point
- system temperature sensor
- Expected
- 60 degrees centigrade °C
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Image noise from cellular phones or RF transmitters
Image artifacts or noise appearing when cellular phone or RF device is near system.
Ranked causes
- 1. Cellular phone or RF transmitter in proximity to system (common)
Check for cellular phones or portable RF transmitters near system
- Test point
- Area around VIVID E9 system
- Expected
- No active cellular phones within 5 meters; RF transmitters distant
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image or patient data loss patient safety
System experiences loss of ultrasound images or patient data
Ranked causes
- 1. Unknown software or hardware fault (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Install optional Color Video Printer
Connect and configure an external Color Video Printer (Sony UP-D23MD/UP-D25MD or Mitsubishi CP30DW) to the Vivid E9 USB port.
Ranked causes
- 1. Color Video Printer installation and connection required (occasional)
Verify printer is a compatible medical device model
- Test point
- Printer model identification
- Expected
- SONY UP-D23MD, UP-D25MD, or MITSUBISHI CP30DW
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Motor lock does not engage
Lock engine fails to engage when Fix Lock button is pressed during Motor Controller Test
Ranked causes
- 1. Lock mechanism requires adjustment (unknown)
Press Fix Lock button and listen for lock engine engagement
- Test point
- Lock mechanism
- Expected
- Audible lock engine engagement for a few seconds in quiet environment
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Network connectivity not available
System reports no network connection or cannot connect to network
Ranked causes
- 1. Network cabling fault or disconnection (common)
Check network status indicator
- Test point
- Network icon on bottom of screen
- Expected
- Network status shown (connected or not connected)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No network connectivity reported
System reports network disconnected; network icon shows no connection status
Ranked causes
- 1. Network cable disconnected or loose (common)
Verify network cable connection
- Test point
- Network port connector
- Expected
- Cable fully seated and connected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Operator console drops unexpectedly during manual repositioning patient safety
Console falls when Z mechanism is released during power-off repositioning
Ranked causes
- 1. Z mechanism disengaged or drive gear failed (common)
Verify handle movement before repositioning
- Test point
- Up/Down Release Handle
- Expected
- Handle moves freely to right
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Peripheral cables causing interference or image degradation
Poorly routed or dressed external cables introducing noise or signal degradation.
Ranked causes
- 1. Peripheral cables not properly routed or contained (common)
Inspect cable routing in peripheral bays and across Card Rack
- Test point
- Cable layout and management
- Expected
- All cables routed through bays or along frame; none hanging freely or across Card Rack top
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pinch point hazard during monitor adjustment caution
Risk of hand/arm pinch injury at hinges when adjusting LCD Arm or LCD Monitor position.
Ranked causes
- 1. Mechanical pinch hazard at monitor arm and monitor hinges (common)
Ensure no body parts in pinch zone before monitor adjustment
- Test point
- Monitor arm hinges and monitor hinges
- Expected
- Clear view of all hinge areas; no fingers or hands present
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Save network settings and restart system
Persist network configuration changes and activate them on the Vivid E9 system.
Ranked causes
- 1. Network settings must be saved and system restarted for changes to take effect (common)
Press Save Settings button
- Test point
- Network settings configuration screen
- Expected
- Warning dialog displayed
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System appears locked and unresponsive
System stops responding to commands; locked state persists beyond normal operation.
Ranked causes
- 1. Application watchdog timeout trapped system state (unknown)
Wait for watchdog timeout detection
- Test point
- System monitor
- Expected
- Dialog indicating system not responding appears after 60 seconds
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System hangs or does not respond
BEP (Back End Processor) becomes unresponsive and system stops functioning.
Ranked causes
- 1. BEP software halt or deadlock (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System locks up, halts, or experiences intermittent problems caution
System stop, failure to respond to operator input, random freezing; may be software-related or communication error
Ranked causes
- 1. Software fault or corruption in back end processor (common)
Capture system error logs
- Test point
- Scan screen display
- Expected
- Alt+D command opens log capture interface; logs saved to media or exported
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Top console cannot be released without power
- Trackball is sticky and squeaky
- USB devices stopped working
- USB devices stopped working on front connectors
- XY brake motor not functioning
- XY Brake Motors not responding
- XY brakes cannot be released without power
- XY Park Lock not engaging or not responding to Frogleg Controls
8 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
High System Temperature | System temperature has exceeded safe operating threshold | System temperature rises to 60 degrees centigrade or above | warning |
Image or Patient Data Loss | Patient exam data or ultrasound images lost or inaccessible from storage | Data loss detected after image capture, during patient file storage, or on system restart | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.