GE Healthcare Vivid 3 Pro: troubleshooting and error codes
The ServiceForge knowledge base holds 66 documented symptoms for the GE Healthcare Vivid 3 Pro, with 83 ranked causes, 207 diagnostic checks with test points and expected values, and 4 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Front End Diagnostics tests fail or show errors
During Auto Sequence or manual FE diagnostic testing, LED indicators show red (test failed), test results show errors in Comment area, or specific subsystem tests (RFI, IMP, VME, RFT, FEC, Beamformer, FB, MUX) report problems.
Ranked causes
- 1. RFI board malfunction or marginal performance (common)
- 2. IMP board failure or marginal performance (RFT Configuration) (common)
- 3. VME Bus communication error (RFT Configuration) (common)
- 4. Subsystem board malfunction (RFT, FEC, Beamformer, FB, MUX) (occasional)
- 5. System reboot required during diagnostic sequence (common)
Review RFI Full Test results
- Test point
- RFI Tab - Full Test button LED and Comment area
- Expected
- Green LED and no error messages in Comment area
5 more checks with pass and fail actions, 13 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
BF diagnostic test errors identified
Beamformer test completes with error indications, ADC output values out of range, or digital beamformer test failures.
Ranked causes
- 1. BF board malfunction (unknown)
- 2. Marginal BF board performance (unknown)
- 3. System rebooted after CW or PW scanning mode (unknown)
Perform IMP, VME Bus, RFT and FEC tests before BF test
- Test point
- Diagnostic system
- Expected
- IMP, VME Bus, RFT and FEC tests complete successfully
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Moire pattern present on display
Concentric circles or arcs visible on monitor screen indicating electromagnetic interference or display misalignment
Ranked causes
- 1. Horizontal moire pattern caused by display misalignment (occasional)
- 2. Vertical moire pattern caused by display misalignment (occasional)
- 3. Color impurities from magnetic field interference (occasional)
Observe moire pattern orientation on display
- Test point
- Monitor screen
- Expected
- Horizontal or vertical moire pattern clearly visible
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disk/Bus performance below specification
Disk or BUS data transfer speed measured below expected threshold
Ranked causes
- 1. Hard disk performance degraded (occasional)
- 2. BUS interface performance degraded (occasional)
Run Disk Bench test
- Test point
- Hard disk IDE BUS speed
- Expected
- greater than 23MB/sec
- Tools
- HW Diagnostics Disk/Bus Bench
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
FEC diagnostic test errors identified
FEC board test completes with error indications shown by red LED or error messages in Comment area.
Ranked causes
- 1. FEC board malfunction (unknown)
- 2. Marginal FEC board performance (unknown)
Perform IMP, VME Bus and RFT tests before FEC test
- Test point
- Diagnostic system
- Expected
- IMP, VME Bus and RFT tests complete successfully
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Intermittent Front End problems suspected
System exhibits signs of intermittent faults or problems that appear and disappear. Auto Sequence is recommended to help identify root cause of intermittent issues.
Ranked causes
- 1. Cable connections loose or intermittently poor (common)
- 2. Marginal component performance becoming intermittent (occasional)
Use Auto Sequence with Stop on Failure to capture intermittent failures
- Test point
- Auto Sequence dialog - On Failure section
- Expected
- Stop checkbox selected to pause immediately on first failure
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manual FE Diagnostic test required for Front Board Assembly components
Front Board Assembly subsystem requires manual tests not included in Auto Sequence (FB Transmit Test, TR Switch Test, FLA Jumper Test, Pencil Probe Test).
Ranked causes
- 1. Front Board Assembly diagnostics not completed by Auto Sequence (common)
- 2. MUX Board Advanced Board Tests not included in Auto Sequence (common)
Verify Auto Sequence exclusions
- Test point
- Manual testing checklist against Auto Sequence report
- Expected
- FB Transmit Test, TR Switch Test, FLA Jumper Test, Pencil Probe Test, and MUX Advanced Board Tests marked as not included in Auto Sequence
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor display does not show correct colors or patterns
Monitor fails to display correct colors, alignment, or patterns during monitor diagnostic tests
Ranked causes
- 1. Monitor alignment, color depth or resolution issue (common)
- 2. Defective color pixels on panel display (occasional)
Run Combination Test from Monitor Test menu
- Test point
- Monitor display
- Expected
- Four cross hairs visible in each corner, RGB spectrum arrays and horizontal and vertical alignment bars displayed
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
RFT diagnostic test errors identified
RFT board test completes with error indications shown by red LED or error messages in Comment area.
Ranked causes
- 1. RFT board malfunction (unknown)
- 2. Marginal RFT board performance (unknown)
Perform IMP and VME Bus tests before RFT test
- Test point
- Diagnostic system
- Expected
- IMP and VME Bus tests complete successfully
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System displays error messages after boot
Error messages appear during system initialization despite normal operation; may indicate minor calibration changes during normal use
Ranked causes
- 1. Calibration drift during normal operation (common)
- 2. Front End board or component failure (occasional)
Perform system calibration
- Test point
- Calibration menu
- Expected
- Calibration completes without error messages
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
VCR recording or playback not functioning
VCR does not record, recording indicator absent, tape counter not running, or playback fails
Ranked causes
- 1. VCR connection issue or Sony 9600 VCR malfunction (common)
- 2. Audio channel malfunction during CW recording or playback (occasional)
Connect probe and verify VCR recording activation
- Test point
- VCR Record button and on-screen indicators
- Expected
- Recording indicator (red dot) is displayed on-screen and the tape counter is running on-screen
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wheels not moving freely or brake pedal malfunctioning caution
Wheels locked when they should move freely, or wheel movement present when brakes applied
Ranked causes
- 1. Brake pedal position incorrect or brake mechanism stuck (common)
- 2. Wheel wear or damage (occasional)
Test free wheel movement with brake pedal centered
- Test point
- Brake pedal center position; wheel movement in all directions
- Expected
- Ensure that the wheels move freely in all directions.
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio output from speakers unclear or missing
Sound output from one or both speakers is not clear at all frequencies or missing entirely
Ranked causes
- 1. Audio driver or speaker hardware malfunction (common)
Change frequency and speaker activity from right to left
- Test point
- Left and right speaker outputs (as viewed from behind the unit)
- Expected
- Clear sound from both speakers at all frequencies
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio output not functioning
Doppler sound not emitted from speakers during audio test at various PRF and frequencies
Ranked causes
- 1. Speaker malfunction or connectivity issue (unknown)
Right speaker audio output
- Test point
- Right speaker output
- Expected
- Clear tone emitted
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: 17" Monitor Installation caution
Procedure for reinstalling a 17" MAG-type monitor on Vivid 3 console after removal or replacement
Ranked causes
- 1. Monitor requires reinstallation after removal or replacement (common)
Position monitor upside down on soft surface
- Test point
- Monitor case underside
- Expected
- Monitor accessible for base assembly work
16 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: B/W printer image quality
Printed B/W image quality evaluation and brightness/contrast adjustment
Ranked causes
- 1. B/W printer brightness/contrast settings require adjustment (common)
Scan phantom in 2D-Mode
- Test point
- 2D imaging mode; phantom imaging
- Expected
- Phantom scanned in 2D-Mode
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Beamformer Board offset voltage
BF Board ADC offset voltage requires calibration after board replacement to ensure proper digitization of input channels
Ranked causes
- 1. Beamformer Board replaced or offset voltage drift (common)
Verify BF Board is installed in correct slot
- Test point
- Front End Crate third slot from left
- Expected
- BF Board present and seated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CD-R/RW drive media access and formatting
CD-R/RW drive cannot access, read, format, or manage media; media properties cannot be verified or label set
Ranked causes
- 1. CD-R/RW drive initialization or media driver issue (common)
Insert CD-R media and open diagnostic application
- Test point
- CDR drive media slot; UtilMO_CDR_App display
- Expected
- New recordable CD X16 loaded; UtilMO_CDR_App is displayed
11 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Color printer image quality
Printed color image quality evaluation and brightness/contrast adjustment for CFM imaging
Ranked causes
- 1. Color printer brightness/contrast settings require adjustment (common)
Scan blood vessels or CFM phantom and activate CFM mode
- Test point
- 2D imaging and CFM mode; phantom imaging
- Expected
- Blood vessels or CFM phantom scanned with color displayed
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: iLinq IP Save
Save iLinq IP settings and presets to hard disk for backup or transfer
Ranked causes
- 1. Need to backup iLinq IP configuration settings (common)
Open System Configuration window
- Test point
- Configuration interface
- Expected
- Window is accessible and responsive
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Image quality monitor settings
Monitor image quality requires periodic adjustment due to ambient light changes. Display may appear too dark or have excessive contrast. Brightness and contrast controlled via front/side buttons on Samsung 15" or 17" display monitors.
Ranked causes
- 1. Monitor requires recalibration due to ambient light or age (common)
Allow monitor to warm up before calibration
- Test point
- Display monitor warm-up time
- Expected
- 20 minutes from power-on
12 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Keyboard Installation caution
Procedure for reinstalling keyboard assembly on Vivid 3 console after removal or replacement
Ranked causes
- 1. Keyboard requires reinstallation after removal or replacement (common)
Return keyboard to original position on console
- Test point
- Console keyboard bay
- Expected
- Keyboard positioned in original location
11 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Magneto Optical Drive (MOD) media access and formatting
MOD drive cannot access, read, format, or manage media; media properties cannot be verified or modified
Ranked causes
- 1. MOD drive initialization or media driver issue (common)
Verify MOD initialization boot messages
- Test point
- System boot sequence; boot log display
- Expected
- Messages SCSI Adaptor Installation 2271149-2INS and MO are displayed
10 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Monitor geometry adjustment
Adjust monitor display geometry including position, size, linearity, pincushion, trapezoid, parallelogram and rotation for proper image alignment
Ranked causes
- 1. Monitor geometry requires calibration or recalibration (common)
Push menu button to access on-screen menu system
- Test point
- Menu button [1] on monitor side
- Expected
- Main menu displayed on screen
2 more checks with pass and fail actions, 12 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Monitor image quality adjustment
Adjust monitor brightness, contrast, color balance and color temperature for optimal image quality
Ranked causes
- 1. Monitor image quality requires calibration (common)
Measure or observe screen brightness level
- Test point
- Monitor display
- Expected
- Brightness set to appropriate level for clinical viewing
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Console height adjustment unsmooth or difficult caution
Console movement to maximum or minimum height is rough, stiff, or requires excessive force
Ranked causes
- 1. Console lift mechanism friction or damage (common)
Test console lift to maximum height
- Test point
- Front console handle; console position
- Expected
- Console movement is smooth and does not require excessive force
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPU benchmark results below specification
System CPU performance or memory capacity measured below baseline
Ranked causes
- 1. CPU performance degraded or memory insufficient (rare)
Run Benchmark Testing
- Test point
- CPU Description, Memory and Frame Rate
- Expected
- Current values >= baseline values (upper line >= lower line)
- Tools
- HW Diagnostics Computer Testing Benchmark
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG signal not displayed patient safety
No QRS signal shown in Data area during ECG diagnostic test with electrodes connected
Ranked causes
- 1. ECG input connection or electrode issue (unknown)
Internal ECG signal presence
- Test point
- Data area display
- Expected
- Single, clear QRS signal displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Expanded Low Channels diagnostic test channels missing or reduced output
Expanded Low Channels Diagnostic Test identifies missing input channels connected to output channels. Output signal reduced by -6dB if one channel is missing. Input signals from 3S or 7S probe.
Ranked causes
- 1. Input channel signal loss from probe or connection (common)
Run Expanded Low Channels Diagnostic Test
- Test point
- Input and output channels in CW-Mode
- Expected
- All channels present; output at expected level dB
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
FB diagnostic test errors identified
Front Board Assembly test completes with failures in Transmit, Receive, TR Switch, FLA Jumper, Pencil Probe, Port or RAM tests.
Ranked causes
- 1. FB circuit or board malfunction (unknown)
Perform RFI/IMP/VME Bus/RFT/FEC and BF tests before FB tests
- Test point
- Diagnostic system
- Expected
- All prerequisite tests pass
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front End (FE) diagnostic test reveals bad channels
FE Diagnostics identifies one or more channels as bad during channel testing. Output signal reduced by -6dB if channel is missing.
Ranked causes
- 1. Front End channel failure or malfunction (unknown)
Review FE Diagnostics test results for bad channels
- Test point
- FE Diagnostics result display
- Expected
- No channels identified as bad; output signals normal
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front End (FE) Receive Test fails
Receive Test shows errors in amplifier output levels during phased array probe simulation mode
Ranked causes
- 1. Amplifier output not meeting predefined level (unknown)
Disconnect all probes before Receive Test
- Test point
- Probe connectors
- Expected
- All probes disconnected
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Front End (FE) test shows marginal results
TR4 or FLA jumper test completes but results are marginal, uncertain if hardware is fully functional
Ranked causes
- 1. Front End assembly needs recalibration (common)
Review FE diagnostic test result message
- Test point
- Report dialog box
- Expected
- Test result indicates marginal errors
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front End (FE) test shows no errors but problem persists
All diagnostic tests pass with no channels or elements marked as bad, but imaging issue reported by operator continues
Ranked causes
- 1. Intermittent hardware fault or probe quality issue (occasional)
Confirm all FE diagnostic tests passed
- Test point
- Report dialog test result message
- Expected
- No errors identified in all tests
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front End (FE) test shows non-marginal errors and cables connected
TR4 or FLA jumper test shows clear errors and all cables appear properly connected
Ranked causes
- 1. Image port or PCI bridge cable malfunction (common)
Verify all image port and PCI bridge cables connected
- Test point
- Image port, PCI bridge connectors
- Expected
- All cables firmly seated and undamaged
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front End (FE) TR4 channel test fails
TR4 module diagnostic shows red highlighted buttons indicating errors in one or more channels during test
Ranked causes
- 1. TR4 socket or vector channel malfunction (unknown)
Run TR4 Module diagnostic test
- Test point
- FE Diagnostics dialog, F.B. tab
- Expected
- All vector buttons highlighted green; no red highlights
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front End diagnostic test fails with alarm
Front End diagnostic test generates failure alarms during execution
Ranked causes
- 1. Unnecessary probes connected during test (common)
Verify probe connections before running Front End diagnostic
- Test point
- Probe connections
- Expected
- Only required probes for specific test are connected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front End diagnostic test results marginal but no clear errors
FE Diagnostics test completes but results are marginal; no definitive errors identified but problem persists. Calibration may improve results.
Ranked causes
- 1. Front End component drift or calibration drift (common)
Run FE Diagnostics and review for marginal results
- Test point
- FE Diagnostics result display
- Expected
- No errors identified or results marginal
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
iLinq Configuration Utility setup incomplete
OLC checkout communication not established for remote diagnostics via modem connection
Ranked causes
- 1. InSite Checkout tab settings not configured or CHECKOUT NOW not executed (common)
Verify modem device selected as communication device
- Test point
- Device Connection tab
- Expected
- Modem is listed and selected as active communication device
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
iLinq IP settings not operational after software reinstall
iLinq IP settings require restoration from hard disk backup after software reinstallation or configuration loss
Ranked causes
- 1. iLinq IP settings lost or corrupted during software reinstallation (common)
Access System Configuration window
- Test point
- Technical Support Tab
- Expected
- Tab displays iLinq Settings options
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal ECG signal not displaying or unclear patient safety
ECG input does not produce clear QRS signal on display or signal is absent
Ranked causes
- 1. ECG input connection issue or ECG diagnostic failure (common)
Verify clear QRS signal display from ECG input
- Test point
- ECG input connector; on-screen ECG waveform
- Expected
- A single clear QRS signal is displayed.
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard and footswitch controls unresponsive
Buttons, rockers, optical encoders, TGC sliders, or footswitch not activating or responding on diagnostic panel
Ranked causes
- 1. Keyboard or footswitch hardware malfunction or disconnection (common)
Test alphanumeric keyboard backlights
- Test point
- Keyboard control panel
- Expected
- All alphanumeric backlights are off
8 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard buttons not registering
Control panel buttons, rockers, rotaries, sliders or soft keys not activating on-screen in KB Panel tab during manual external keyboard test
Ranked causes
- 1. Keyboard hardware or connection failure (unknown)
Control panel button activation
- Test point
- KB Panel tab on-screen display
- Expected
- Correct buttons activated on-screen
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard key does not respond during test
Key pressed on keyboard does not disappear from test window display during keyboard diagnostic test
Ranked causes
- 1. Keyboard key damaged or defective (common)
Perform keyboard diagnostic test and press each key once
- Test point
- Keyboard keys
- Expected
- Corresponding key on-screen is removed from view
- Tools
- PCDR Keyboard Test window, keyboard
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Media drive not accessible
MO drive or CD Writer drive not responding to access commands or media cannot be loaded/ejected properly during media driver diagnostic test
Ranked causes
- 1. Media drive hardware malfunction or media not inserted (unknown)
MO drive media accessibility
- Test point
- Dialog message display
- Expected
- Message 'Media in drive G is accessible' displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor 15" replacement required caution
Samsung SyncMaster 591S 15" monitor needs removal and installation on Vivid 3 control console
Ranked causes
- 1. Monitor replacement procedure (unknown)
Verify Teflon washer in place on monitor mounting base
- Test point
- Monitor base top surface
- Expected
- Teflon washer present and intact
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor 17" replacement required - Samsung type caution
Samsung 17" monitor needs removal and installation on Vivid 3 control console, Procedure 1
Ranked causes
- 1. Monitor 17" replacement procedure Samsung type (unknown)
Verify Teflon washer on monitor base
- Test point
- Monitor base top surface
- Expected
- Teflon washer present and in place
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Network adapter not detected
Intel PRO/100VE Network Connection not visible in Windows Device Manager
Ranked causes
- 1. Network adapter driver missing or not installed (occasional)
Verify T-plug dongle inserted
- Test point
- Parallel or USB port on rear panel
- Expected
- T-plug (dongle) physically present in port
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Normal channels diagnostic test detects bad channels
MUX Normal Channels test identifies one or more channels as bad, shown by red highlighting in TR4 Module representation
Ranked causes
- 1. Defective TR4 module or amplifier channel (unknown)
Disconnect all probes
- Test point
- Probe connectors
- Expected
- All probes disconnected
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Normal channels test shows marginal results or operator suspects bad channels
Normal channels test shows marginal pass or operator identifies specific channels as problematic; manual marking required
Ranked causes
- 1. Degraded TR4 module or channel performance (occasional)
Visually identify suspected bad channels from operator report or previous observations
- Test point
- TR4 Modules Display area
- Expected
- Channels match suspected problem area
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pencil probe relay test detects bad relays
Front end diagnostics pencil probe test identifies one or more relays as bad, indicated by red highlighting in report dialog box
Ranked causes
- 1. Defective pencil probe relay (unknown)
Run pencil probe test (open)
- Test point
- Front board pencil probe relays
- Expected
- No relays highlighted in red; test result message shows pass
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pencil probe test detects bad channels, problem persists after test
Technician suspects additional bad channels beyond those identified by automatic test; problem continues after testing
Ranked causes
- 1. Additional defective amplifier channels or relays not detected by automatic test (occasional)
Identify suspected bad channels visually or from operator reports
- Test point
- TR4 socket and vector channels in Display area
- Expected
- Channels match suspected problem location
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pencil probe test shows marginal results
Front end diagnostics pencil probe test completes but results are marginal, suggesting potential degradation
Ranked causes
- 1. Marginal relay or channel performance (unknown)
Run pencil probe test (close)
- Test point
- Front board pencil probe relays
- Expected
- Test result message indicates marginal performance
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pencil probe test shows non-marginal errors with connectivity issue
Front end diagnostics pencil probe test identifies errors that are not marginal; cable connectivity check required
Ranked causes
- 1. Loose or defective image port or PCI bridge cable (common)
Inspect image port and PCI bridge cables
- Test point
- Image port connectors and PCI bridge cable connectors
- Expected
- Cables fully seated and secure with no visible damage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Phono microphone signal not displaying or unresponsive patient safety
Microphone input does not produce waveform on display corresponding to microphone clicks or taps
Ranked causes
- 1. Medical microphone not connected or phono diagnostic failure (common)
Tap microphone and verify waveform display
- Test point
- Microphone input connector; on-screen waveform display
- Expected
- The waveform on the screen corresponds to the microphone clicks.
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Phono signal not responding
Microphone waveform not displayed or does not correspond to microphone clicks during Phono diagnostic test
Ranked causes
- 1. Microphone or Phono input board connection failure (unknown)
Phono signal response to microphone input
- Test point
- Data area waveform display
- Expected
- Waveform corresponds to microphone clicks
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Port test detects errors
Front end Port test identifies logic control or port logic errors in Comments area
Ranked causes
- 1. Port logic control malfunction (unknown)
Disconnect all probes before Port test
- Test point
- Probe connectors
- Expected
- All probes disconnected
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power-on or standby mode malfunction caution
System does not enter or exit standby mode correctly, or does not recover to power-on state after boot
Ranked causes
- 1. Power management or boot sequence issue (common)
Enter standby mode
- Test point
- On/Off button; On/Off button LED status
- Expected
- The system enters Standby mode.
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
RAM test detects errors
Front end RAM test identifies errors in RAM storing delay signals for transmitted ultrasound to focal beam
Ranked causes
- 1. Defective RAM or memory controller (unknown)
Access F.B. tab RAM Test
- Test point
- F.B. tab Comments area
- Expected
- RAM test results displayed, no errors
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Solid black line observed during scanning high voltage
A continuous black vertical line appears on ultrasound image during scanning, indicating a potentially bad transmitting channel associated with the pencil probe.
Ranked causes
- 1. Bad transmitting channel (pencil probe relay) (common)
Connect 2D pencil probe
- Test point
- Pencil probe connector
- Expected
- Probe securely connected
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Solid dark line observed during scanning
- Solid line observed during scanning
- UPS not providing backup power
- VIC Correlation Box installation for Samsung 17" monitor
- Video grabbing produces unstable or discolored image
- Video output ports not displaying correctly
6 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Bad Channel Error | Normal Channels diagnostic test identifies one or more TR4 module channels as bad | Normal Channels MUX test identifies red highlighted channel(s) in TR4 Module representation | error |
Pencil Probe Relay Error | One or more front board pencil probe relays detected as bad during pencil probe test (open or close) | Pencil probe test (open or close) identifies relay malfunction | error |
Port Logic Error | Front end Port test detects general logic control or port logic malfunction | Port test execution identifies logic error | error |
RAM Memory Error | RAM test identifies errors in delay signal storage RAM used for transmitted ultrasound focal beam | RAM test execution detects memory malfunction | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI