GE Healthcare STENOSCOP: troubleshooting and error codes
The ServiceForge knowledge base holds 56 documented symptoms for the GE Healthcare STENOSCOP, with 67 ranked causes, 237 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: External laser aiming radiation
Class IIIA laser aiming system requires preadjustment and alignment with X-ray field for accurate targeting during fluoroscopy procedures.
Ranked causes
- 1. Factory preadjustment or laser replacement required (common)
- 2. Reticule positioning required after preadjustment (common)
- 3. Final adjustment before each use (common)
Verify C-arm stability before adjustment
- Test point
- C-arm assembly
- Expected
- Stable without movement
14 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Integrated laser alignment radiation
Alignment of integrated Class IIIA laser diode assembly with image intensifier field using X-ray fluoroscopy and mechanical adjustment screws.
Ranked causes
- 1. Initial laser positioning on image intensifier (common)
- 2. Lateral and longitudinal laser beam alignment (common)
- 3. Final verification and collimator reassembly (common)
Remove collimator cover
- Test point
- Collimator cover assembly
- Expected
- Cover removed to allow laser access
18 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: kV circuit adjustment (Wisconsin cassette method) radiation
Measurement and calibration of high-voltage kV circuit (3A12-R216) using Wisconsin X-ray test cassette to verify HF unit output within tolerance.
Ranked causes
- 1. HF unit kV output out of specification tolerance (occasional)
- 2. VHV measurement outside tolerance range (occasional)
- 3. Unit fault condition (PWB or circuit failure) (rare)
De-energize mobile system before measurements
- Test point
- System power supply
- Expected
- System powered down; no residual voltage present
14 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: T1 delay adjustment on 3A1 Interface DSM radiation
Adjust timing delay T1 for DSM Interface module to ensure proper signal synchronization in fluoroscopy sequence
Ranked causes
- 1. T1 delay timing requires calibration after board installation or accidental uncalibration (common)
- 2. T2 delay verification for automatic control sequence (common)
- 3. T3 delay check for RX control sequence (common)
Measure T1 delay on rising edge of signal at 3A1.E123
- Test point
- 3A1.E123
- Expected
- 1000 +/- 20 ms, +/-20 ms
- Tools
- Oscilloscope, probe x10
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Dose Area Product meter radiation
DAP meter (Model 841-SO) requires calibration to ensure displayed readings match reference instrument measurements. Calibration is necessary when offset adjustment is incorrect or readings drift significantly over time.
Ranked causes
- 1. Incorrect offset adjustment or drift (common)
- 2. Energy response variation (occasional)
Verify display stability after warm-up and reset
- Test point
- Display
- Expected
- Zero for 30 minutes
7 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Phase adjustment for ABD control signal radiation
Adjust oscilloscope-based phase and gain to obtain stable image without overshoot, oscillation, or excessive glare on monitor during fluoroscopy modes
Ranked causes
- 1. Initial phase adjustment required for automatic brightness control stability (common)
- 2. Hysteresis adjustment needed for stable automatic control (occasional)
Measure ABD control signal overshoot and glare on monitor
- Test point
- 3A5.b9
- Expected
- Maximum 33% overshoot with glare; 50% overshoot max for regulation at 45+/-1 kV, 33% max typical, 50% max acceptable
- Tools
- Oscilloscope
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Interface controls alignment check (Imager) radiation
Verification that interface signals from Stenoscop reach the II assembly, including X-RAY ON, CLTV, and PG signals. Optical centering and longitudinal axis alignment are preset at factory.
Ranked causes
- 1. Signal voltage out of tolerance (occasional)
- 2. CCD Interface Board misalignment (rare)
Remove power and access II assembly
- Test point
- II assembly rear cover, rack cover
- Expected
- Components accessible without damage
10 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Back wheel brake cam disassembly caution
Brake cam assembly requires removal and inspection during service
Ranked causes
- 1. Service requirement (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Board disassembly and reassembly from monitor high voltage
Removal and replacement of internal boards: LV power supply, deflection, AT power supply, video, selection, and light sensor boards from monitor assembly
Ranked causes
- 1. Defective monitor board requiring replacement (unknown)
Disconnect monitor from power source
- Test point
- Power supply
- Expected
- Monitor disconnected and de-energized
17 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Brake shoe replacement caution
Worn brake shoe on wheel assembly needs replacement
Ranked causes
- 1. Defective brake shoe (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
C-arm motor PWB removal caution
Motor control board disassembly from mobile system for service or replacement
Ranked causes
- 1. C-arm motor PWB service required (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
C-arm wheel replacement caution
Defective wheel on C-arm chassis requires replacement to restore mobility
Ranked causes
- 1. Defective wheel (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration after X-ray head replacement radiation
Perform preheating and mA adjustment following X-ray head installation
Ranked causes
- 1. System requires recalibration after component replacement (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ABD signal offset adjustment radiation
ABD signal voltage at TP1/TP7 on CCD Video Board requires zeroing without X-rays
Ranked causes
- 1. ABD signal offset drift (unknown)
Measure ABD signal voltage without X-rays
- Test point
- TP1 to TP7, CCD Video Board
- Expected
- 0 V +20/–20 mV, +20/–20 mV
- Tools
- Digital multimeter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Anode cooling pause safety delay adjustment radiation
Adjust safety delay for anode cooling pause function on Interface DSM (PWB 828 988 G035) to prevent premature X-ray exposure
Ranked causes
- 1. Anode cooling pause safety delay requires calibration on DSM Interface (code 828 988 G035) (occasional)
Measure anode cooling pause delay
- Test point
- 3A1.E608
- Expected
- 2 sec +/- 0.4 sec, +/-0.4 sec
- Tools
- Oscilloscope, 0.5 Sec/div sweep
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Automatic fluoroscopy loop gain preadjustment radiation
Preadjustment of regulation loop potentiometers to establish baseline conditions before phase and gain fine-tuning.
Ranked causes
- 1. Loop gain preadjustment for regulation stability (common)
Set all potentiometers to preadjustment positions
- Test point
- 3A5.R67, 3A5.R60, 3A5.R66 positions
- Expected
- R67=fully CCW, R60=fully CW, R66=mid-range
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Automatic fluoroscopy phase and loop gain adjustment radiation
Fine-tuning of regulation loop phase lead and gain for stable automatic brightness control response using appropriate phantom material.
Ranked causes
- 1. Regulation loop stability optimization with material selection (common)
Select appropriate phantom material for adjustment
- Test point
- Phantom material composition
- Expected
- Water or plastic (polyamide/polyethylene) preferred; copper acceptable but requires caution
- Tools
- Phantom materials (water, plastic, copper options)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Automatic fluoroscopy system dose reference radiation
Calibration of dose reference signal for automatic fluoroscopy using operational amplifier output measurement and potentiometer adjustment.
Ranked causes
- 1. Dose reference adjustment for automatic fluoroscopy feedback (common)
Position copper absorption and connect oscilloscope probe
- Test point
- X-ray head outlet, 3A5.A82 pin 6 operational amplifier output
- Expected
- 2mm copper in place, oscilloscope probe clipped to R108 (easy access point)
- Tools
- Oscilloscope, DVM
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Automatic fluoroscopy system X-ray emission reference radiation
Calibration of X-ray emission reference level for automatic fluoroscopy dose control using calibrated copper absorption and oscilloscope measurement.
Ranked causes
- 1. X-ray emission reference level adjustment for automatic fluoroscopy (common)
Place copper absorption and connect oscilloscope
- Test point
- X-ray head outlet, 3A5.b9 signal output
- Expected
- 2mm copper filter in place, oscilloscope connected to ABD signal
- Tools
- Dual beam oscilloscope with double time base
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Black level threshold adjustment radiation
Automatic mode shutter blanking threshold adjustment using ABD window display
Ranked causes
- 1. Black level threshold requires adjustment for shutter blanking (unknown)
Verify shutter blanking in ABD window
- Test point
- ABD window display on CCD Video Board
- Expected
- Opaque shutters visible at image edges
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Brightness and contrast adjustment
Monitor brightness and contrast controls under back door need adjustment; light sensor normally auto-adjusts per ambient light
Ranked causes
- 1. Brightness and contrast require manual calibration (rare)
Set video signal level
- Test point
- video input
- Expected
- 250 mV
- Tools
- probe
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Dose 1 adjustment (low dose) radiation
Low dose fluoroscopy mode iris aperture and voltage adjustment for specified dose rate at image intensifier entry
Ranked causes
- 1. Dose 1 iris aperture requires calibration (unknown)
Check video voltage after dose rate adjustment
- Test point
- TP1 to TP7, CCD Video Board
- Expected
- 2.1 V ±50 mV, ±50 mV
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Dose 2 adjustment (high dose) radiation
High dose fluoroscopy mode iris aperture and voltage adjustment for specified dose rate at image intensifier entry
Ranked causes
- 1. Dose 2 iris aperture requires calibration (unknown)
Verify automatic fluoroscopy video voltage
- Test point
- TP1 to TP7, CCD Video Board
- Expected
- 2.1 V
- Tools
- Digital multimeter
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Dose Area Product Meter display radiation
LCD display setup and functional verification after replacement or power-up
Ranked causes
- 1. Display assembly requires functional check after installation or power-up (common)
Power up X-ray machine and observe LCD display
- Test point
- LCD display
- Expected
- A number will appear on the LCD display
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Fluoroscopy/radiography switch timing radiation
Verification that the switching between fluoroscopy and radiography modes completes within specified time window and focal spot selection is correct.
Ranked causes
- 1. Mode switch timing and focal spot verification (common)
Connect oscilloscope in differential mode to CR5 terminals
- Test point
- 5A1-CR5 anode and cathode (use add mode)
- Expected
- Differential voltage waveform visible; ground to chassis
- Tools
- Dual channel oscilloscope 50MHz, differential probe capability
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Image intensifier assembly alignment radiation
Optical unit lens support alignment with longitudinal axis and circular blanking window centering
Ranked causes
- 1. II assembly misalignment with optical axis detected (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: mAs integrator adjustment radiation
Calibration of mAs (milliampere-seconds) integrator potentiometers R5 and R6 to ensure accurate exposure dose across full mAs range.
Ranked causes
- 1. mAs integrator adjustment required for dose accuracy (common)
Remove X-ray head cover and divider bridge, install mAs meter
- Test point
- 6A1 divider circuit, W33 jumper location
- Expected
- mAs meter inserted in current measuring circuit, circuit continuity verified
- Tools
- mAs meter, wire grip
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Monitor adjustment at 100 Hz high voltage
Adjustment of vertical frequency, size, linearity, and horizontal parameters at 100 Hz refresh rate for proper image display.
Ranked causes
- 1. Monitor requires calibration after assembly or component replacement (common)
Vertical frequency measurement
- Test point
- TP9
- Expected
- Sawtooth 4V amplitude
- Tools
- Oscilloscope
8 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Monitor adjustment at 120 Hz high voltage
Adjustment of vertical frequency, size, and linearity at 120 Hz refresh rate for proper image display.
Ranked causes
- 1. Monitor requires calibration for 120 Hz operation (common)
Vertical frequency measurement at 120 Hz
- Test point
- TP9
- Expected
- Sawtooth 4V amplitude
- Tools
- Oscilloscope
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Monitor brightness and blanking high voltage
Adjustment of blanking timing and horizontal blank position for proper image boundaries.
Ranked causes
- 1. Monitor requires blanking calibration (common)
Blanking timing measurement
- Test point
- TP109 on video board
- Expected
- Square wave with ON time 30 µs
- Tools
- Oscilloscope
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Monitor brightness setting in darkroom
Adjust monitor brightness to achieve correct luminance in darkroom conditions with maximum external adjustment
Ranked causes
- 1. Monitor brightness requires darkroom calibration (rare)
Display black level image
- Test point
- monitor display
- Expected
- black image only
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Monitor focus adjustment
Adjust focus across monitor display using three focus controls for optimal image sharpness
Ranked causes
- 1. Focus requires multi-point calibration (rare)
Set video signal level
- Test point
- video input
- Expected
- 0.7 V
- Tools
- voltmeter
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: T4 delay adjustment for pulsed fluoroscopy radiation
Adjust timing delay T4 generated after first RX pulse in pulsed fluoroscopy mode for proper image capture synchronization
Ranked causes
- 1. T4 delay requires calibration for pulsed fluoroscopy synchronization (common)
Measure T4 delay after first RX pulse in pulsed mode
- Test point
- 3A1.E126
- Expected
- 500 +/- 10 ms, +/-10 ms
- Tools
- Oscilloscope, probe x10
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: T5 minimum delay check for pulsed fluoroscopy radiation
Verify T5 delay timing in pulsed fluoroscopy mode to ensure proper minimum delay after initialization
Ranked causes
- 1. T5 delay verification required for pulsed mode timing validation (common)
Measure T5 minimum delay in pulsed fluoroscopy mode
- Test point
- 3A1.E127
- Expected
- 880 +/- 100 ms, +/-100 ms
- Tools
- Oscilloscope, probe x10
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: T6 pulsed fluoroscopy rate adjustment radiation
Adjust timing for pulsed fluoroscopy pulse rate to achieve 1 X-ray pulse per 1.5 second with T4=500 ms
Ranked causes
- 1. T6 rate adjustment required for pulsed fluoroscopy pulse frequency (common)
Measure T6 pulse rate delay in pulsed fluoroscopy mode
- Test point
- 3A1.E121
- Expected
- 1000 +/- 50 ms; results in 1 X-ray pulse per 1.5 second with T4=500 ms, +/-50 ms
- Tools
- Oscilloscope, probe x10
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Tube current in radiography mode radiation
Adjustment of X-ray tube current (mA) output during radiography exposures using potentiometer control and oscilloscope measurement.
Ranked causes
- 1. Initial tube current adjustment required for radiography mode (common)
Oscilloscope connection to A26.a32 output
- Test point
- A26.a32
- Expected
- Signal capture for current monitoring
- Tools
- Dual channel oscilloscope 50 MHz
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Tube current verification in radiography mode radiation
Verification measurement of actual X-ray tube current across multiple kV settings using inline milliammeter in RAD mode.
Ranked causes
- 1. Tube current verification measurement across kV range (common)
Remove mA jumper and install DC milliammeter
- Test point
- 6A1 divider circuit current measuring path
- Expected
- Circuit continuity verified, milliammeter in series
- Tools
- Milliammeter, wire grip
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Video signal level check radiation
CCD video output signal amplitude verification at test point with copper attenuation filter
Ranked causes
- 1. Video signal level requires verification (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: X-ray on delay adjustment radiation
Automatic mode fluoroscopy startup time delay adjustment to prevent oscillations
Ranked causes
- 1. X-ray on delay requires adjustment for oscillation prevention (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Capacitor discharge procedure for CRT handling high voltage
Safe discharge of CRT capacitance before disconnecting final anode or handling CRT assembly to prevent electrical shock hazard
Ranked causes
- 1. CRT capacitance charge presents electrocution hazard (common)
Disconnect monitor from power source
- Test point
- Power supply connection
- Expected
- Monitor fully disconnected
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collimator adjustment and centering radiation
Reassemble and center collimator in X-ray beam after installation
Ranked causes
- 1. Collimator misalignment (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CRT physical handling and safety patient safety
Safe physical handling procedures for CRT removal and installation to prevent damage and injury from glass implosion
Ranked causes
- 1. CRT glass bottle fragility and implosion hazard (rare)
Wear safety glasses before starting CRT handling
- Test point
- Personal protective equipment
- Expected
- Safety glasses properly fitted
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disassembly/Reassembly: Imageur Head patient safety
Removal and installation of CCD imaging head unit from C-arm gantry
Ranked causes
- 1. Imageur head requires removal for service or repair (occasional)
Remove rear cover from Imageur Head
- Test point
- Imageur rear cover screws
- Expected
- Cover removed: 3x M5 screws (16-cm) or 2x M5 screws (22-cm)
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disassembly/Reassembly: X-ray Image Intensifier Tube radiation
Removal and installation of II tube assembly with high-voltage connector management
Ranked causes
- 1. Image Intensifier tube requires replacement or inspection (rare)
Prepare disassembly area and positioning
- Test point
- Work surface and Stenoscop position
- Expected
- Flat surface prepared, Stenoscop horizontal, Imageur Head above rest surface
17 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
EHV power supply replacement high voltage
Thomson TH7191 EHV power supply requires disassembly and reassembly on STENOSCOP 2 6000/9000 C.C.D. – M.D.A. imaging system
Ranked causes
- 1. EHV power supply failure or end-of-life replacement (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hard disk replacement
Hard disk assembly requires replacement for storage malfunction or upgrade
Ranked causes
- 1. Defective hard disk (unknown)
Remove monitor cart back cover
- Test point
- Monitor cart rear
- Expected
- Back cover removed
12 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Modulator capacitors replacement high voltage
Disassembly and reassembly of modulator capacitors 4C1 to 4C6
Ranked causes
- 1. Defective modulator capacitors (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor cart cover disassembly and reassembly caution
Removal and reinstallation of monitor cart covers A, B, and C for access to internal components
Ranked causes
- 1. Need access to internal monitor components (unknown)
Switch OFF unit and remove mains plug
- Test point
- Power source
- Expected
- Unit powered down and disconnected from outlet
9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Motor belt tension adjustment caution
Belt between motor and pulley requires tension adjustment to ensure proper drive engagement and prevent slippage
Ranked causes
- 1. Belt tension incorrect during assembly or maintenance (common)
Measure belt play between pulleys
- Test point
- Between motor and driven pulley
- Expected
- 10 +/– 2 mm when 5 N force applied, +/– 2 mm
- Tools
- 5 N force gauge
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Motor fuse 1A1F5 blown caution
Motor circuit protection blown; UP/DOWN column motion affected
Ranked causes
- 1. Excessive motor current from mechanical binding or wear (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Operation check: Automatic control stability with varying absorption radiation
Verify automatic brightness control provides stable images without abnormal glare or oscillation across different absorption conditions and filter configurations
Ranked causes
- 1. Operation verification required after calibration to ensure automatic control stability (common)
Check image appearance with various absorptions and filters
- Test point
- Display monitor
- Expected
- No abnormal glare; no oscillation apparent
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replacement: Ion Chamber assembly radiation
Ion chamber unit removal and installation for the dose measurement system
Ranked causes
- 1. Ion chamber requires replacement or service (occasional)
De-energize the mobile system
- Test point
- Power supply
- Expected
- System powered off
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch screen assembly replacement
Touch screen interface needs replacement due to defect or malfunction
Ranked causes
- 1. Defective touch screen assembly (unknown)
Remove monitor cart front cover
- Test point
- Monitor cart front
- Expected
- Front cover removed
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Verification of field size measurement radiation
Confirmation that reference chamber is positioned within radiation beam and field size is measured correctly for DAP calibration calculation.
Ranked causes
- 1. Reference chamber outside beam or field size incorrect (occasional)
Position reference dosemeter within beam
- Test point
- Reference dosemeter location
- Expected
- Totally within radiation beam
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
X-ray head centering radiation
Align X-ray beam and collimator with image intensifier after installation
Ranked causes
- 1. Head misalignment after replacement (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
X-ray head removal and replacement radiation
X-ray head assembly disassembly and reassembly with calibration checks
Ranked causes
- 1. X-ray head service or replacement (unknown)
Continuity of kV and mA measure connections
- Test point
- 6A1.PL2.9 to 3A12.PL1.7
- Expected
- Continuity verified with ohmmeter
- Tools
- Ohmmeter
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.