GE Healthcare Solar 8000M/i: troubleshooting and error codes
The ServiceForge knowledge base holds 22 documented symptoms for the GE Healthcare Solar 8000M/i, with 65 ranked causes, 50 diagnostic checks with test points and expected values, and 20 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Display is blank patient safety
Video display shows no output; no image on screen
Ranked causes
- 1. Video display has no power (common)
- 2. Patient monitor has no power (common)
- 3. Video display not properly connected (occasional)
- 4. Display turned off via control button (occasional)
- 5. SDRAM test failed (rare)
- 6. Boot Code not loaded or corrupt (rare)
- 7. Power supply failure (rare)
- 8. Loose connections to power supply (occasional)
- 9. Excessive loading from malfunctioning peripheral device (rare)
Verify video display power indicator illuminated
- Test point
- Video display power indicator
- Expected
- Illuminated
4 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tram module patient data does not display patient safety
Data from Tram module in Tram-rac housing not appearing on monitor screen
Ranked causes
- 1. Incorrect Tram module leadwire connections (common)
- 2. Tram module not fully inserted or powered off (common)
- 3. Tram-rac housing powered off (common)
- 4. Missing or incompatible software (occasional)
- 5. Loose or faulty Tram-rac housing cable (occasional)
- 6. Tram-rac housing communication failure (occasional)
- 7. Faulty Tram module (rare)
- 8. Faulty Tram-rac housing (rare)
Verify correct leadwire connections to Tram module
- Test point
- Tram module leadwire connectors
- Expected
- Correct patient/simulator leads connected
4 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Remote control does not activate the display caution
Wireless or wired remote control not responding; display cannot be controlled remotely
Ranked causes
- 1. Remote control communication loss or power interruption (common)
- 2. M-Port communication cable disconnected or loose (common)
- 3. M-Port on monitor defective (occasional)
- 4. Remote control requires reset (occasional)
- 5. Remote control malfunction (occasional)
- 6. Processor PCB failure (rare)
Check M-Port LED for remote communication
- Test point
- M-Port LED indicator
- Expected
- Illuminated (indicates remote communication)
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient Data Module does not display patient data patient safety
Patient Data Module not showing data on monitor display
Ranked causes
- 1. Incorrect Patient Data Module leadwire connections (common)
- 2. Patient Data Module not fully inserted or off (common)
- 3. Loose or faulty monitor to module cable (occasional)
- 4. Patient Data Module error condition (occasional)
- 5. Faulty Patient Data Module (rare)
Verify correct leadwire connections to Patient Data Module
- Test point
- Patient Data Module lead connectors
- Expected
- Correct patient/simulator leads connected
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Boot Code starts but Main Code does not patient safety
System starts boot sequence but fails to load main operating code; boot banner shows test results
Ranked causes
- 1. Hardware test failure in boot banner (common)
- 2. Main Code not loaded (occasional)
- 3. Main Code and Boot Code incompatible (occasional)
- 4. Processor PCB failure (rare)
Check boot banner test results
- Test point
- Boot banner display
- Expected
- SRAM test: passed, SDRAM test: passed, Main flash: passed, Boot flash: passed, Real time clock: running
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient alarms are not sounding at desired central station patient safety
Alarms are not transmitted to or heard at the specified central station location
Ranked causes
- 1. Patient not correctly admitted to system (common)
- 2. Alarm destination reprogramming needed after unit move (common)
- 3. Care unit name mismatch between monitor and central station (common)
- 4. Speaker at central station malfunction (occasional)
Check for DISCHARGED message on patient monitor
- Test point
- Monitor display
- Expected
- No DISCHARGED message
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient alarms are not sounding when activated patient safety
Alarm function is disabled or muted; alarms do not produce audible alert
Ranked causes
- 1. Patient not correctly admitted into system (common)
- 2. Alarms not enabled in settings (common)
- 3. Alarm volume turned off (common)
- 4. Speaker malfunction (occasional)
Check for DISCHARGED message and waveform display
- Test point
- Monitor display
- Expected
- No DISCHARGED message; patient waveform present
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient data from monitor does not display at central station patient safety
Central station shows NO COMM message or no patient data from Solar 8000M/i monitor
Ranked causes
- 1. Monitor communication link down (common)
- 2. Loose Ethernet cable connection (common)
- 3. Incorrect LAN address programmed (occasional)
- 4. Processor PCB malfunction (rare)
Verify monitor communication is up
- Test point
- Monitor power and network interface
- Expected
- Monitor powered on; network communication active
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display quality is not good caution
Video output is degraded, distorted, or poor quality image
Ranked causes
- 1. Loose or faulty video display connection (common)
- 2. Faulty video display unit (occasional)
- 3. Processor PCB failure (rare)
Verify video display to monitor connection
- Test point
- Video cable connector
- Expected
- Secure connection, proper seating
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Graphs and alarms are not printed at desired location caution
Central station graph writer or printer not receiving data for specified care unit
Ranked causes
- 1. Central station writer paper incorrect orientation (common)
- 2. Graph destination reprogramming needed after care unit move (common)
- 3. Care unit name mismatch on monitor and central station (occasional)
Check central station writer paper orientation
- Test point
- Central station writer paper tray
- Expected
- Paper loaded correct side up
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display does not appear after power-up
Unit powered on and power LED illuminates but no display appears after 10 seconds
Ranked causes
- 1. Display not turned on (common)
- 2. Unit malfunction (occasional)
Verify display is turned on
- Test point
- Display power control
- Expected
- Display on
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
M-Port LED not green
LED above M-Port is not solid green, indicating communication status issue
Ranked causes
- 1. Device identified but no communication (occasional)
- 2. Too many identical devices connected or device cannot be identified (occasional)
Observe M-Port LED status
- Test point
- M-Port LED indicator
- Expected
- Solid green
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power LED does not illuminate
Green LED indicator above power switch does not light when unit plugged in and power switch turned on
Ranked causes
- 1. Unit not connected to AC power source (common)
- 2. Power switch not turned on (common)
Verify AC outlet is powered
- Test point
- AC wall outlet
- Expected
- AC power available
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC line voltage out of specification high voltage
Wall outlet supplying power to monitor has incorrect voltage measurements
Ranked causes
- 1. Wall outlet improperly wired or damaged (unknown)
Measure voltage between line contact and neutral (120 VAC outlet)
- Test point
- Line contact and neutral
- Expected
- 120 VAC (± 10 VAC), ± 10 VAC
- Tools
- Digital voltmeter
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error logs not accessible or unclear
Service personnel need to review error and event logs from monitor or TRAM module for troubleshooting; logs may also be transferred to central station or diskette
Ranked causes
- 1. Need to access and interpret error/event logs for field troubleshooting (common)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Event log not accessible or unclear
Service personnel need to review event log showing record of each event with index, date, time, event code, event-specific data and textual description
Ranked causes
- 1. Need to access and interpret event logs for system state determination (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Interpret error code severity and impact
Understanding error severity levels to determine impact on system: Continue, Fatal, or Forced Restart
Ranked causes
- 1. Error severity classification needed to assess system impact (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor appears locked up during network download caution
Monitor display becomes unresponsive during Boot Code download; packet or byte numbers stop advancing for two minutes or more
Ranked causes
- 1. Download interrupted or connection failure (common)
Verify cables are properly connected
- Test point
- All network and power cables
- Expected
- All cables fully seated and secure
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Prepare equipment for AC line voltage test high voltage
Initial setup steps before measuring wall outlet voltage
Ranked causes
- 1. Test procedure requires equipment isolation (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software revision window does not list part numbers
Part numbers missing from software revision window for monitor interfaces after update
Ranked causes
- 1. Software update not activated (common)
Verify part numbers appear in revision window
- Test point
- Software revision window display
- Expected
- Part numbers listed for all monitor interfaces
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temporary loss of monitoring during log file retrieval patient safety
Patient monitor monitoring functions may be temporarily unavailable while log files are being copied to a PC via netUpdate tool
Ranked causes
- 1. Log file transfer in progress (common)
Verify patient monitor is in monitoring mode before initiating log transfer
- Test point
- Patient monitor display
- Expected
- Monitor actively displaying waveforms and parameters
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Waveforms do not appear at central station patient safety
Waveform display missing or not transmitting to central station after software update or during operation
Ranked causes
- 1. Communication corrupted between monitor and central station (common)
Check all network and signal cables
- Test point
- Monitor to central station connection
- Expected
- All cables fully connected and undamaged
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
400-4FF | Network errors detected | Network communication failure | warning |
703 | Diagnostic tests completed | Diagnostic test routine executed | info |
70B | Internet address was changed | Monitor network address modification | info |
70C | Video test completed | Video test routine executed | info |
70D | Rear power switch was turned off | Rear panel power switch deactivated | warning |
70E | Time was changed from this monitor | System time modified | info |
70F | Date was changed from this monitor | System date modified | info |
710 | Incompatible software detected | Main processor detects incompatible communication software version | error |
714 | Incompatible TRAM module software detected | Main processor detects TRAM software revision mismatch | error |
Boot Flash test FAILED / Boot Code invalid | Flash memory test failed or Boot Code in Flash is corrupt | Power-on self-test of Boot Flash fails | error |
EC1 | ECG parameter block error | ECG data processing failure | warning |
ERROR: Internal battery failed | Backup battery maintaining monitor defaults has failed | Battery voltage drops below operating threshold | error |
Main Code SDRAM FAILED | SDRAM memory test for Main Code region failed | Power-on self-test detects memory failure | error |
Main Flash test FAILED / Main Code invalid | Flash memory test failed or Main Code in Flash is corrupt | Power-on self-test of Main Flash fails | error |
Real Time Clock FAILED | Real-time clock chip not running | Clock chip fails to initialize or respond | error |
Service the PDM - Error Code | Patient Data Module software/hardware error | PDM encounters fatal error during operation | critical |
Static RAM test FAILED | SRAM memory test failed during startup | Power-on self-test detects SRAM failure | error |
WARNING: Boot Code version incompatible | Boot Code version does not match Main Code version in Flash | Version mismatch detected during boot sequence | warning |
WARNING: EEPROM data invalid or uninitialized | EEPROM contains corrupted or uninitialized data | Boot sequence detects invalid EEPROM contents | warning |
WARNING: Main Code load failed | SDRAM checksum mismatch during Main Code loading | Boot process detects checksum discrepancy | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI