GE Healthcare Solar 8000i: troubleshooting and error codes
The ServiceForge knowledge base holds 18 documented symptoms for the GE Healthcare Solar 8000i, with 18 ranked causes, 84 diagnostic checks with test points and expected values, and 1 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration info not updated in Service Tool
DEVICE INFO in GE Service Tool CALIBRATION INFO does not show appropriate date and time for performed calibrations
Ranked causes
- 1. Calibration data not saved or synchronized to Service Tool (rare)
Verify CALIBRATION INFO date and time matches performed calibrations
- Test point
- Service Tool DEVICE INFO > CALIBRATION INFO
- Expected
- Current date and time of calibrations performed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NBP gain (calibration offset) patient safety
Calibrate NBP sensor gain by inflating cuff to 250 mmHg and matching PDM reading to manometer
Ranked causes
- 1. NBP gain calibration procedure (common)
Connect NBP circuit to PDM with reservoir in circuit
- Test point
- PDM NBP connector, reservoir valve position
- Expected
- NBP circuit connected and reservoir switched into circuit
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NBP zero offset patient safety
Calibrate NBP circuit zero pressure reading to match external manometer
Ranked causes
- 1. NBP zero calibration procedure (common)
Set manometer to 1000 mmHg range and open to atmosphere
- Test point
- Manometer
- Expected
- Set to 1000 mmHg range, all valves open to atmosphere
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiac Output functional test - MedSim 300 simulator
Verify cardiac output reading from simulated thermodilution injection is between 6.65-7.35 L/min
Ranked causes
- 1. Cardiac Output functional test procedure (MedSim 300) (common)
Connect simulator CO test cable to PDM
- Test point
- PDM CO port, simulator CO test cable output
- Expected
- Cable seated securely
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiac Output functional test - SimStim simulator
Verify cardiac output reading from simulated thermodilution injection is between 6.05-7.15 L/min
Ranked causes
- 1. Cardiac Output functional test procedure (SimStim) (common)
Connect CO cable to PDM and catheter connector to patient simulator
- Test point
- PDM CO port, simulator catheter input
- Expected
- Cable seated securely on both ends
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG functional test
Verify all 6 ECG waveforms display correctly at 60 bpm, pacer detection works, and lead failure detection functions
Ranked causes
- 1. ECG functional test procedure (common)
Set simulator to 60 bpm heart rate output
- Test point
- Simulator controls
- Expected
- Simulator set to output 60 bpm
10 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECM message displayed after 395 days operation
ECM reminder message appears on patient monitor to left of ECG parameter block after 395 days of operation
Ranked causes
- 1. Operation cycle timer reached preventive maintenance interval (common)
Perform cold boot sequence to clear ECM message
- Test point
- Patient monitor display and control buttons
- Expected
- Monitor goes blank then boots normally without ECM message
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IBP functional test
Verify invasive blood pressure waveforms, zero calibration, dynamic waveform accuracy, and static pressure readings
Ranked causes
- 1. IBP functional test procedure (common)
Verify PDM cable attached to e-port and power indicator ON (green)
- Test point
- PDM e-port connector, front panel power indicator
- Expected
- Cable seated, power indicator green
9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP calibration out of tolerance patient safety
Pressure reading on patient monitor (CUFF) differs from manometer reading by more than ±3 mmHg or ±2%
Ranked causes
- 1. NBP sensor gain drift or calibration offset (common)
Verify manometer is set to 1000 mmHg range and reads 0 after zero adjustment
- Test point
- Manometer
- Expected
- 0 mmHg (after adjustment if needed)
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PDM CPU board replacement - activation code configuration required
New IBP activation code must be obtained and configured after CPU board replacement before functional testing
Ranked causes
- 1. New CPU board requires IBP activation code matching MAC address and serial number (common)
Verify MAC address from new CPU board
- Test point
- Label on board or GE Service Tool readout
- Expected
- Valid MAC address documented
6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PDM final shutdown
Turn off PDM by removing battery for 10 seconds, then reinstall battery and close battery door
Ranked causes
- 1. PDM power-off procedure (common)
Disconnect all remaining cables from PDM
- Test point
- PDM connectors
- Expected
- All cables removed
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PDM lithium battery will not charge to target capacity
Battery fails to charge to 600% capacity or fails reconditioning after several attempts in Cadex charger
Ranked causes
- 1. PDM battery aged or degraded beyond recovery (unknown)
Attempt reconditioning in Cadex Small Two+ charger with TARGET CAPACITY switch set to 600%
- Test point
- Cadex charger display
- Expected
- Battery charges to READY status
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
PDM software version below minimum required
PDM software version is below 1.2 or SpO2 software version is below 1.6
Ranked causes
- 1. Outdated PDM or SpO2 software installed (occasional)
Check PDM software version on monitor
- Test point
- MONITOR SETUP > REVISION & ID, line below PDM
- Expected
- Version 1.2 or higher
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Possible leak in NBP circuit or PDM patient safety
Manometer reading drops more than 1 mmHg every few seconds, indicating air leak during static pressure test
Ranked causes
- 1. Leak in NBP circuit connections or PDM (occasional)
Verify reservoir is switched out and monitor manometer pressure for drops
- Test point
- Manometer in test circuit without reservoir
- Expected
- Less than 1 mmHg drop every few seconds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Respiration functional test
Verify respiration waveform and respiratory rate display correctly at 30 breaths per minute
Ranked causes
- 1. Respiration functional test procedure (common)
Set simulator respiration rate to 30 breaths per minute using lead II
- Test point
- Simulator respiration controls
- Expected
- Simulator set to 30 breaths/min
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SERVICE PDM message displayed on monitor
Patient monitor shows SERVICE PDM message after PDM connection
Ranked causes
- 1. PDM error condition detected by monitor (unknown)
Confirm no SERVICE PDM message on monitor after PDM connection
- Test point
- Patient monitor display
- Expected
- No SERVICE PDM message displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 functional test (Masimo)
Verify SpO2 waveform, oxygen saturation reading 78-84%, and pulse rate 60-62 bpm from Masimo sensor
Ranked causes
- 1. SpO2 functional test procedure (Masimo) (common)
Connect Masimo SpO2 tester cable to SpO2 sensor adapter cable
- Test point
- SpO2 tester cable and adapter connector
- Expected
- Cable seated securely
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature functional test
Verify temperature sensor readings from T1 and T2 inputs match simulator output between 36.6-37.4°C
Ranked causes
- 1. Temperature functional test procedure (common)
Set simulator to output 37°C and select YSI probe type (400 or 700)
- Test point
- Simulator temperature controls
- Expected
- Simulator set to 37°C output, YSI 400 or 700 selected
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
9993 | Service PDM error logged when ECG Filter Frequency not toggled after CPU board replacement | ECG Filter Frequency not switched from 60 Hz to 50 Hz and back to 60 Hz after entering new IBP activation code | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI