GE Healthcare Senographe Essential: troubleshooting and error codes
The ServiceForge knowledge base holds 49 documented symptoms for the GE Healthcare Senographe Essential, with 99 ranked causes, 5 diagnostic checks with test points and expected values, and 787 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
E10 Heating failure track 1 high voltage
Filament heater track 1 malfunction in X-ray tube
Ranked causes
- 1. Heater presence signal high in stand-by (common)
- 2. Heater signal stays low when started (common)
- 3. Heater signal fails during pre-heater phase (occasional)
- 4. Heater signal fails during first trigger (occasional)
- 5. Heater signal fails during exposure (occasional)
- 6. mA value not significant (occasional)
- 7. Heater signal fails after exposure (occasional)
- 8. Heater signal stays high when stopped (occasional)
E11 Heating failure track 2 high voltage
Filament heater track 2 malfunction in X-ray tube
Ranked causes
- 1. Heater presence signal high in stand-by (common)
- 2. Heater signal stays low when started (common)
- 3. Heater signal fails during pre-heater phase (occasional)
- 4. Heater signal fails during first trigger (occasional)
- 5. Heater signal fails during exposure (occasional)
- 6. mA value not significant track 2 (occasional)
- 7. Heater signal fails after exposure (occasional)
- 8. Heater signal stays high when stopped (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
E03 Rotor failure radiation
X-ray tube anode rotation system malfunction preventing or losing rotation
Ranked causes
- 1. Anode rotation signal high in stand-by (common)
- 2. Rotation not rising during 9000 rpm acceleration (common)
- 3. Rotation not rising during 6000 rpm acceleration (occasional)
- 4. Rotation lost at 9000 rpm (common)
- 5. Rotation lost at 6000 rpm (occasional)
- 6. Rotation lost during deceleration (occasional)
- 7. Rotation signal stays high during braking (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
E02 Focus bias failure high voltage
Filament bias generation system malfunction in X-ray tube
Ranked causes
- 1. Focus bias software error unknown command (rare)
- 2. Bias presence signal high in stand-by (occasional)
- 3. Bias presence signal stays low when generation starts (occasional)
- 4. Bias presence signal goes low after stabilization (occasional)
- 5. Bias presence signal stays high after stopping (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
IRLOCK motion timeout caution
Locking mechanism does not complete motion within expected time; motor stalls or hangs
Ranked causes
- 1. Faulty motor (occasional)
- 2. Mechanism blocked by foreign object (occasional)
- 3. Faulty motor cable (occasional)
- 4. Faulty Breast Support Locking Mechanism Board (occasional)
- 5. Faulty power supply (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IRLOCK motor malfunction caution
Motor fails to operate or responds abnormally; locking mechanism does not move as expected
Ranked causes
- 1. Faulty motor (occasional)
- 2. Mechanism blocked by foreign object (occasional)
- 3. Faulty motor cable (occasional)
- 4. Faulty Breast Support Locking Mechanism Board (occasional)
- 5. Faulty power supply or Arm Distribution Board (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
E04 Arm fatal failure caution
Gantry arm communication or positioning system fatal malfunction
Ranked causes
- 1. Boot communication error with gantry (occasional)
- 2. Bucky positioning error (occasional)
- 3. Gantry error during exposure (occasional)
- 4. No gantry acknowledgment to filter change (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
IRLOCK release left button timeout caution
Left release button press does not trigger expected response; button input times out
Ranked causes
- 1. Right release button faulty (occasional)
- 2. Left release button short-circuiting (occasional)
- 3. Button connector broken or shorted (occasional)
- 4. Button held longer than 1 second (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IRLOCK release right button timeout caution
Right release button press does not trigger expected response; button input times out
Ranked causes
- 1. Left release button faulty (occasional)
- 2. Right release button short-circuiting (occasional)
- 3. Button connector broken or shorted (occasional)
- 4. Button held longer than 1 second (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IR lock/release motion stopped - IR type lost patient safety
Breast support not registered with system via IR code; locking motion halted
Ranked causes
- 1. Faulty breast support unit (occasional)
- 2. Faulty breast support connection cable W502 (occasional)
- 3. Faulty Breast Support Locking Mechanism Board (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IR release fork inactive while lock command received caution
Release fork does not respond during lock command; mechanical or electrical fault
Ranked causes
- 1. Foreign object blocking optical fork or moving parts (occasional)
- 2. Faulty Breast Support Locking Mechanism Board (occasional)
- 3. Faulty Breast Support Locking Mechanism (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
IR type lost and not released patient safety
Breast support not registered via IR code; cannot release mechanism
Ranked causes
- 1. Faulty breast support unit (occasional)
- 2. Faulty breast support connection cable W502 (occasional)
- 3. Faulty Breast Support Locking Mechanism Board (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IRLOCK invalid PDO length caution
PDO with invalid DLC received by Breast Support Locking Mechanism Board; data format error
Ranked causes
- 1. Faulty Breast Support Locking Mechanism Board (occasional)
- 2. Faulty gantry CPU board (occasional)
Replace Breast Support Locking Mechanism Board
- Test point
- PDO communication
- Expected
- PDO received with valid DLC
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lock optical fork test failed caution
Optical fork on Breast Support Locking Mechanism Board fails lock function test
Ranked causes
- 1. Optical fork dirty or obstructed (common)
- 2. Faulty Breast Support Locking Mechanism Board (occasional)
Inspect optical fork for cleanliness and obstruction
- Test point
- Optical fork
- Expected
- Clean, no foreign objects, moves freely
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Release optical fork test failed caution
Optical fork on Breast Support Locking Mechanism Board fails self-test at startup
Ranked causes
- 1. Optical fork dirty or obstructed (common)
- 2. Faulty Breast Support Locking Mechanism Board (occasional)
Inspect optical fork cleanliness
- Test point
- Optical fork
- Expected
- Clean, no foreign objects present
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compression brake low current patient safety
Halt 122/020: Brake current below acceptable limit; screw or manual compression brake failing.
Ranked causes
- 1. Brake power supply failure, wiring break, or brake solenoid malfunction (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compression button not pressed before motion command
Debug-only error 121/005: CPU sent Start Motion without prior button press command received by node.
Ranked causes
- 1. Compression board firmware logic error or synchronization fault (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compression calibration not achieved patient safety
Warning 121/003: Calibration failed; IPMS ignores all non-calibration commands.
Ranked causes
- 1. Compression force or thickness calibration procedure incomplete or failed (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compression correlation error caution
Halt 122/005: Absolute coder (potentiometer) and motor encoder (coding wheel) signals not correlated.
Ranked causes
- 1. Potentiometer or encoder misalignment, mismatch, or sensor failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compression encoder defect caution
Halt 122/003: Motor encoder signals (A, A bar, B, B bar) show complementarity or sequence error.
Ranked causes
- 1. Motor encoder or coding wheel failure or connection issue (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compression force sensor defect patient safety
Halt 122/021: Force sensor reading outside acceptance range.
Ranked causes
- 1. Force sensor failure, excessive reverse force, or signal conditioning problem (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compression motion incoherent patient safety
N/A severity 122/004: Encoder pulses detected when no motion expected; occurs on Lift, Rotation, Compression, Tilt.
Ranked causes
- 1. Encoder signal spurious or motion control synchronization failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compression motor step lost caution
Halt 122/006: Difference detected between commanded and detected steps by absolute encoder.
Ranked causes
- 1. Mechanical jam, stalled motor, or encoder slippage at hardpoint (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compression node firmware incompatibility caution
Halt 122/025: CPLD hardware config, DSP, or MCU firmware mismatch on compression board.
Ranked causes
- 1. Firmware or hardware component revision mismatch after board replacement or update (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compression overcurrent high voltage
Halt 122/008: Integrated current value exceeds limit; current too high for too long.
Ranked causes
- 1. Motor stall, mechanical jam, or power stage short circuit on compression board (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compression overspeed patient safety
Halt 122/007: Measured compression speed exceeds speed profile limits.
Ranked causes
- 1. Motor control or speed regulator failure allowing uncontrolled acceleration (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compression potentiometer defect caution
Halt 122/002: Potentiometer signal outside acceptable range at all times.
Ranked causes
- 1. Potentiometer sensor failure or electrical connection problem (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compression software error (invalid parameter)
Debug-only error 121/002: IPMS received command with parameter values outside acceptable range.
Ranked causes
- 1. Compression board or firmware parameter validation failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compression software error (simultaneous motion)
Debug-only error 122/024: CPU sent motion command while previous motion not finished.
Ranked causes
- 1. Gantry CPU firmware synchronization or timing logic error (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compression software error (unrecognized command)
Debug-only error 121/001: IPMS received unrecognized command identifier from PTB and ignored it.
Ranked causes
- 1. Compression board firmware or command compatibility issue (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compression software watchdog timeout caution
Halt 122/001: Watchdog reset microcontroller after missing 40 ms pulse; node rebooted.
Ranked causes
- 1. Microcontroller firmware hang or clock/CPLD issue on compression board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
E01 Generator initialization failure high voltage
Fatal security input detected during generator initialization tests
Ranked causes
- 1. Fatal security input triggered during initialization (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
E08 Power supply failure high voltage
High voltage power supply timeout or regulation failure
Ranked causes
- 1. Power supply voltage regulation timeout (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
E12 Checksum error caution
Battery-backed memory checksum validation failure
Ranked causes
- 1. Battery backup memory corruption (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Emergency stop motion activated patient safety
N/A severity 122/022: Emergency stop button pressed; compression node cuts stop motion line.
Ranked causes
- 1. Operator intentionally pressed emergency stop button (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
IR lock/release motion stopped due to button press
Locking or release motion interrupted by user button press; operator error condition
Ranked causes
- 1. Operator pressed button during motion (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IR lock/release movement not started caution
Internal motion state machine error; locking mechanism does not respond to commands
Ranked causes
- 1. Internal state machine error (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IRLOCK CAN command not processed caution
CAN command not processed; Gantry CPU sent command before board acknowledged previous command
Ranked causes
- 1. Command timing conflict between Gantry CPU and board (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IRLOCK invalid message caution
Invalid communication between control tasks; locking mechanism command processing error
Ranked causes
- 1. Task communication error (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IRLOCK_CPLD_OPTICAL_FORK_ERROR caution
Hardware error in CPLD affecting optical fork on Breast Support Locking Mechanism Board
Ranked causes
- 1. CPLD hardware failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Left breast support release button still pressed at boot patient safety
System detects left release button state at startup; button may be physically stuck or depressed
Ranked causes
- 1. Left release button stuck or physically pressed (common)
Inspect left release button for obstruction
- Test point
- Release button
- Expected
- Moves freely without resistance
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lift down button still pressed
Warning 120/003-004 or Halt 122/011-012: compression node detects lift down button(s) continuously pressed.
Ranked causes
- 1. Button contacts stuck or signal timing mismatch (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lift up button still pressed
Warning 120/001-002 or Halt 122/009-010: compression node detects lift button(s) continuously pressed on keypad.
Ranked causes
- 1. Button contacts stuck or debounce failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No valid line for compression motion patient safety
Warning 121/004: Gantry CPU requests motion but compression node detects no valid RT line, or valid line falls before movement ends.
Ranked causes
- 1. Compression control node defect or invalid line signal lost (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Preset -MLO button still pressed
Warning 120/011 or Halt 122/019: compression node detects Preset -MLO button continuously pressed.
Ranked causes
- 1. Button contacts stuck or debounce failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Preset +MLO button still pressed
Warning 120/010 or Halt 122/018: compression node detects Preset +MLO button continuously pressed.
Ranked causes
- 1. Button contacts stuck or debounce failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Preset CC button still pressed
Warning 120/009 or Halt 122/017: compression node detects Preset CC button continuously pressed.
Ranked causes
- 1. Button contacts stuck or signal timing mismatch (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Right breast support release button still pressed at boot patient safety
System detects release button state at startup; button may be physically stuck or depressed
Ranked causes
- 1. Release button stuck or physically pressed (common)
Inspect right release button for obstruction
- Test point
- Release button
- Expected
- Moves freely without resistance
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rotation button still pressed
Warning 120/005-008 or Halt 122/013-016: compression node detects rotation button(s) continuously pressed.
Ranked causes
- 1. Button contacts stuck or debounce failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
002/001 | Generator CPU software error: VRTX error during initialization | Detected during generator initialization | error |
002/002 | Generator CPU software error: reboot by watchdog hard or reset button | Detected after CPU hard reset or watchdog reboot | error |
002/003 | Generator CPU software error: reboot by watchdog soft | Detected after soft watchdog reset | error |
002/004 | Generator CPU: reboot after power micro-failure | Detected after power supply micro-interrupt | error |
003/000 | Generator CPU software error: bus error | Detected during CPU operation | error |
003/001 | Generator CPU software error: address error | Detected during CPU operation | error |
003/002 | Generator CPU software error: format error | Detected during CPU operation | error |
003/003 | Generator CPU software error: illegal instruction | Detected during CPU operation | error |
003/004 | Generator CPU software error: privilege violation | Detected during CPU operation | error |
003/005 | Generator CPU software error: trace error | Detected during CPU trace operation | warning |
003/006 | Generator CPU software error: emulation line 1010 error | Detected during CPU operation | error |
003/007 | Generator CPU software error: emulation line 1111 error | Detected during CPU operation | error |
003/008 | Generator CPU software error: CHK error | Detected during CPU operation | error |
003/009 | Generator CPU software error: TRAPV error | Detected during CPU operation | error |
004/001 | Generator interface: reset error | Detected during interface reset | error |
004/002 | Generator interface: read/write error on timer PTM1 | Detected during PTM1 timer access | error |
004/003 | Generator interface: read/write error on timer PTM1 | Detected during PTM1 timer access | error |
004/004 | Generator interface: read/write error on timer PTM2 | Detected during PTM2 timer access | error |
004/005 | Generator interface: read/write error on timer PTM2 | Detected during PTM2 timer access | error |
004/006 | Generator interface: interrupt error on DUART1 | Detected during DUART1 interrupt | error |
004/007 | Generator interface: PTM1 timer 1 defective | Detected during PTM1 timer 1 test | error |
004/008 | Generator interface: PTM1 timer 2 defective | Detected during PTM1 timer 2 test | error |
004/009 | Generator interface: PTM1 timer 3 defective | Detected during PTM1 timer 3 test | error |
004/010 | Generator interface: PTM2 timer 1 defective | Detected during PTM2 timer 1 test | error |
004/011 | Generator interface: PTM2 timer 2 defective | Detected during PTM2 timer 2 test | error |
004/012 | Generator interface: PTM2 timer 3 defective | Detected during PTM2 timer 3 test | error |
004/013 | Generator interface: interrupt error on PTM1 timer 1 | Detected during PTM1 timer 1 interrupt | error |
004/014 | Generator interface: interrupt error on PTM1 timer 2 | Detected during PTM1 timer 2 interrupt | error |
004/015 | Generator interface: interrupt error on PTM1 timer 3 | Detected during PTM1 timer 3 interrupt | error |
004/016 | Generator interface: interrupt error on PTM2 timer 1 | Detected during PTM2 timer 1 interrupt | error |
004/017 | Generator interface: interrupt error on PTM2 timer 2 | Detected during PTM2 timer 2 interrupt | error |
004/018 | Generator interface: interrupt error on PTM2 timer 3 D208 | Detected during PTM2 timer 3 interrupt | error |
006/000 | Generator: fault at exposure command | Detected during exposure command initiation | critical |
006/001 | Generator: fault at exposure command | Detected during exposure command initiation | critical |
006/002 | Generator: signal kV_Inf_5 (Less_than_5kV) not rising | Detected during exposure ramp-up | critical |
006/003 | Generator: Signal En_Pose (Exposure_On) not rising | Detected at exposure start | critical |
006/004 | Generator: fatal security triggered during exposure | Detected during active exposure | critical |
006/005 | Generator: fatal security triggered at end of exposure or signal stays high | Detected at exposure end or during shutdown | critical |
006/011 | Generator: signal Secu_Rattrap (No_Fatal_Security) stays high after arcing recovery | Detected after arc event recovery | error |
007/xxx | E01 Generator failure - one fatal security input active during init | Detected during initialization tests | critical |
011/001 | Focus bias software error: unknown command | Detected during focus bias operation | error |
011/002 | Focus bias software error: unknown task status | Detected during focus bias task monitoring | error |
012/001 | E02 Focus bias failure - Bias_Presence signal high in stand-by | Detected in stand-by mode | error |
012/002 | E02 Focus bias failure - Bias_Presence signal stays low during generation | Detected when bias generation starts | critical |
012/003 | E02 Focus bias failure - Bias_Presence signal goes low after stabilization | Detected after initial bias stabilization | error |
012/004 | E02 Focus bias failure - Bias_Presence signal stays high when stopped | Detected after bias generation is commanded to stop | error |
016/001 | Anode starter software error: unknown command | Detected during anode rotation control | error |
016/002 | Anode starter software error: unknown task status | Detected during anode rotation task monitoring | error |
017/001 | E03 Rotor failure - Rotation_Presence signal high in stand-by | Detected in stand-by mode | error |
017/002 | E03 Rotor failure - Rotation signal not rising during 9000 rpm acceleration | Detected during rotor spin-up to operating speed | critical |
017/003 | E03 Rotor failure - Rotation signal not rising during 6000 rpm acceleration | Detected during initial rotor acceleration phase | critical |
017/004 | E03 Rotor failure - Rotation signal goes low during 9000 rpm operation | Detected during normal rotor operation at speed | critical |
017/005 | E03 Rotor failure - Rotation signal goes low during 6000 rpm operation | Detected during initial operating phase | critical |
017/006 | E03 Rotor failure - Rotation signal goes low during variable-speed rotation | Detected during rotor operation above 6000 rpm | critical |
017/007 | E03 Rotor failure - Rotation signal goes low during low-speed rotation | Detected during rotor operation below 6000 rpm | error |
017/008 | E03 Rotor failure - Rotation signal stays high during 9000 rpm braking | Detected during rotor deceleration from full speed | error |
017/009 | E03 Rotor failure - Rotation signal stays high during 6000 rpm braking | Detected during rotor deceleration from intermediate speed | error |
017/010 | E03 Rotor failure - Rotation signal stays high during rotor failure | Detected after rotor failure event | critical |
017/011 | E03 Rotor failure - Rotation signal stays low during operation | Detected during active rotor operation | critical |
021/001 | Heater software error: unknown command | Detected during heater control | error |
021/002 | Heater software error: unknown task status | Detected during heater task monitoring | error |
021/003 | Heater software error: Heater scaling parameters or values not valid | Detected during heater initialization or control | error |
022/001 | E10 Heating failure track 1 - Heater_Presence signal high in stand-by | Detected in stand-by mode | error |
022/002 | E10 Heating failure track 1 - Heater signal stays low when started | Detected when heater activation begins | critical |
022/003 | E10 Heating failure track 1 - Heater signal goes low during pre-heater phase | Detected during pre-exposure heater warm-up | error |
022/004 | E10 Heating failure track 1 - Heater signal goes low during first trigger | Detected at start of first exposure | error |
022/005 | E10 Heating failure track 1 - Heater signal goes low during exposure | Detected during active X-ray exposure | error |
022/006 | E10 Heating failure track 1 - mA value not significant | Detected during heater current monitoring | error |
022/007 | E10 Heating failure track 1 - Heater signal goes low after exposure | Detected after exposure cessation | error |
022/008 | E10 Heating failure track 1 - Heater signal stays high when stopped | Detected after heater shutdown command | error |
023/001 | E11 Heating failure track 2 - Heater_Presence signal high in stand-by | Detected in stand-by mode | error |
023/002 | E11 Heating failure track 2 - Heater signal stays low when started | Detected when heater activation begins | critical |
023/003 | E11 Heating failure track 2 - Heater signal goes low during pre-heater phase | Detected during pre-exposure heater warm-up | error |
023/004 | E11 Heating failure track 2 - Heater signal goes low during first trigger | Detected at start of first exposure | error |
023/005 | E11 Heating failure track 2 - Heater signal goes low during exposure | Detected during active X-ray exposure | error |
023/006 | E11 Heating failure track 2 - mA value not significant | Detected during heater current monitoring | critical |
023/007 | E11 Heating failure track 2 - Heater signal low after exposure | Detected after exposure cessation | error |
023/008 | E11 Heating failure track 2 - Heater signal stays high when stopped | Detected after heater shutdown command | error |
027/003 | E04 Arm fatal failure - Gantry/optical fiber communication error during boot | Detected during system initialization | critical |
027/005 | E04 Arm fatal failure - Gantry error during exposure | Detected during active X-ray exposure | critical |
027/006 | E04 Arm fatal failure - Gantry error during exposure | Detected during active X-ray exposure | critical |
031/001 | Generator: delay task software error - unknown command | Detected during delay task operation | error |
031/002 | Generator: delay task software error - unknown task status | Detected during delay task monitoring | error |
041/001 | Gene thermal protection software: VRTX error | Detected during thermal protection operation | warning |
041/002 | Gene thermal protection software: VRTX error | Detected during thermal protection operation | warning |
041/003 | Gene thermal protection software: VRTX error | Detected during thermal protection operation | warning |
041/004 | Gene thermal protection software: VRTX error | Detected during thermal protection operation | warning |
041/005 | Gene thermal protection software: unknown anode starter status | Detected during thermal protection monitoring | error |
041/006 | Gene thermal protection software: VRTX error | Detected during thermal protection operation | warning |
050/043 | Unexpected request received in SysCntrl | Detected during system control operation | error |
052/001 | E04 Arm fatal failure - No Gantry acknowledgment to filter change request | Detected during collimator filter change operation | critical |
056/001 | Focus software error: ACQ_NTR or ACQ_OVR received or unknown reply | Detected during focus control operation | error |
056/002 | Focus software error: unknown command | Detected during focus control | error |
056/003 | Focus software error: unknown task status | Detected during focus task monitoring | error |
057/001 | Focus: focus change request not transmitted or focus fault | Detected during focus adjustment operation | error |
061/000 | Gene to gantry communication driver: no reply from gantry after two attempts | Detected during gantry initialization or communication test | critical |
062/000 | Gene to control panel communication driver: no reply from control panel after two attempts | Detected during control panel initialization | critical |
063/006 | Gene init tests: no gantry status feedback after ten requests | Detected during initialization status polling | critical |
066/000 | Gene date/time: access timeout during date initialization | Detected during system date/time setup | warning |
066/001 | Gene date/time: access timeout during read sequence | Detected during date/time read operation | warning |
066/002 | Gene date/time: access timeout during write sequence | Detected during date/time write operation | warning |
066/003 | Gene date/time: access timeout during time write sequence | Detected during time write operation | warning |
067/000 | Generator: Zero divide | Detected during calculation operation | error |
068/xxx | Generator: floating point calculation error, xxx is VRTX task number | Detected during floating-point calculations | error |
071/081 | E12 Checksum error (gen) - checksum error on battery saved memory | Detected during battery memory verification | critical |
071/082 | E08 Power supply failure - timeout on signal Alim_Puis_OK (Inv_Supply_OK) | Detected during power supply status monitoring | critical |
071/083 | Generator: error reading 'kV law table' in 0 point technique (AOP) | Detected during zero-point technique calibration | error |
072/001 | Exposure software error: ACQ_NTR or ACQ_OVR received at start of first trigger | Detected at start of first exposure pulse | error |
072/003 | Exposure software error: ACQ_NTR or ACQ_OVR received at end of first trigger | Detected at end of first exposure pulse | error |
072/004 | Exposure software error: unknown origin of aborted exposure | Software detected exposure was aborted but origin unknown | error |
072/005 | Exposure software error: aborted by PM in 2 pt technique | PM aborted 2-point exposure technique | error |
072/006 | Exposure software error: unknown exposure technique | Software does not recognize the exposure technique being used | error |
072/007 | E06 Software Error (generator CPU) | System task not stable after 7s delay during Prep; Senographe component or IDC not ready | error |
072/008 | Exposure software error: unknown task status | Software cannot determine status of exposure task | error |
072/009 | Exposure software error: 0 pt or 1 pt exposure and gantry absent | Gantry not present for zero or one-point exposure | error |
073/001 | Gene software error (IST, standby): parameter display problem | Problem displaying parameters during standby initialization | error |
073/002 | Gene software error (IST, standby): unknown tube type | System cannot identify X-ray tube type | error |
073/003 | Gene software error (IST, standby): alpha=0.0 when calculating P_NDac_HTPM | Invalid zero value in calibration calculation | error |
073/004 | Gene software error (IST, standby): incorrect thickness for correlation calculation | Invalid thickness value used in correlation calculation | error |
073/005 | Gene software error (IST, standby): bias calibration task problem | Failure during bias calibration task | error |
073/006 | Gene software error (IST, standby): heater scale calibration task problem | Failure during heater scale calibration | error |
073/007 | Gene software error (IST, standby): HV/DAC calibration task problem | Failure during high voltage or DAC calibration | error |
073/008 | Gene software error (IST, standby): FREQ/HV calibration task problem | Failure during frequency or HV calibration | error |
073/009 | Gene software error (IST, standby): PM calibration task problem | Failure during PM (sensor) calibration | error |
073/010 | Gene software error (IST, standby): PM calibration task problem | Failure during PM calibration | error |
073/011 | Gene software error (IST, standby): PM calibration task problem | Failure during PM calibration | error |
073/012 | Gene software error (IST, standby): PM calibration task problem | Failure during PM calibration | error |
073/013 | Gene software error (IST, standby): gantry connection task problem | Failure in gantry connection initialization task | error |
073/014 | Gene software error (IST, standby): gantry connection fault | Communication fault with gantry | error |
073/015 | Gene software error (IST, standby): gantry connection fault | Communication fault with gantry | error |
073/016 | Gene software error (IST, standby): gantry connection fault | Communication fault with gantry | error |
073/017 | Gene software error (IST, standby): mA measurement calibration problem | Failure during mA measurement calibration | error |
074/001 | Gene software error (IST, exposure): unknown exposure end | Exposure end signal not recognized | error |
074/002 | Gene software error (IST, exposure): unknown gantry reply | Gantry response not recognized during exposure | error |
074/003 | Gene software error (IST, exposure): unknown gantry reply | Gantry response not recognized during exposure | error |
074/004 | Gene software error (IST, exposure): unknown exposure status | Exposure status unrecognized by software | error |
075/010 | Gene software error (MNT): anode starter speed impossible | Anode cannot reach required speed | error |
075/011 | Gene software error (MNT): anode starter abnormal test task status | Abnormal status during anode starter test | error |
075/020 | Gene software error (MNT): heater impossible track | Heater cannot track required parameters | error |
075/021 | Gene software error (MNT): heater abnormal test task status | Abnormal status during heater test | error |
075/022 | Gene software error (MNT): heater abnormal test task status | Abnormal status during heater test | error |
075/030 | Gene software error (MNT): bias abnormal test task status | Abnormal status during bias test | error |
075/040 | Gene software error (MNT): HV test abnormal task status | Abnormal status during HV test | error |
075/041 | Gene software error (MNT): gate test abnormal task status | Abnormal status during gate test | error |
075/042 | Gene software error (MNT): inverter test abnormal task status | Abnormal status during inverter test | error |
075/050 | Gene software error (MNT): mA value abnormal test task status | Abnormal status during mA value test | error |
075/051 | Gene software error (MNT): mA measurement abnormal test task status | Abnormal status during mA measurement test | error |
075/060 | Gene software error (MNT): grid sync abnormal test task status | Abnormal status during grid synchronization test | error |
075/070 | Gene software error (MNT): PM test abnormal task status | Abnormal status during PM test | error |
075/071 | Gene software error (MNT): PM test abnormal task status | Abnormal status during PM test | error |
075/072 | Gene software error (MNT): PM test measurement timeout | PM measurement test did not complete in time | error |
075/090 | Gene software error (MNT): error codes display problem | Cannot display error codes | error |
075/091 | Gene software error (MNT): error codes display problem | Cannot display error codes | error |
075/092 | Gene software error (MNT): error codes display problem | Cannot display error codes | error |
075/093 | Gene software error (MNT): error codes display problem | Cannot display error codes | error |
075/100 | Gene software error (MNT): transfer frame code error | Error in transfer frame code | error |
075/101 | Gene software error (MNT): transfer abnormal save task status | Abnormal status during data save transfer | error |
075/102 | Gene software error (MNT): transfer abnormal save task status | Abnormal status during data save transfer | error |
075/103 | Gene software error (MNT): transfer abnormal load task status | Abnormal status during data load transfer | error |
075/104 | Gene software error (MNT): transfer abnormal load task status | Abnormal status during data load transfer | error |
075/105 | Gene software error (MNT): transfer unknown frame | Frame type not recognized | error |
075/106 | Gene software error (MNT): transfer end of loading error | Error at end of loading transfer | error |
075/107 | Gene software error (MNT): transfer abnormal protocol task status | Abnormal status during protocol transfer | error |
078/002 | Gene init tests software error: timeout on system task | System task did not complete within timeout period | error |
078/003 | Gene init tests software error: defective function will self-identify | Defective function detected during initialization | error |
078/004 | Gene init tests software error: defective function will self-identify | Defective function detected during initialization | error |
078/005 | Gene init tests software error: defective function will self-identify | Defective function detected during initialization | error |
078/128 | Gene init tests software error: error following VRTX request | Error occurred after VRTX request | error |
079/001 | Exposure aborted: timeout | Exposure did not complete within timeout period | error |
079/002 | Exposure aborted: heater fault | X-ray tube heater malfunction detected | error |
079/003 | Exposure aborted: anode starter fault | X-ray tube anode starter malfunction | error |
079/004 | Exposure aborted: power fault | Power supply malfunction detected | error |
079/005 | Exposure aborted: generator fault | Generator system malfunction | error |
079/006 | Exposure: no gantry acknowledgement or gantry fault | Gantry not responding at start of first trigger or gantry malfunction | error |
079/008 | Gantry fault detected before exposure start | Gantry malfunction detected during boot or exposure request | error |
079/009 | Exposure: no sync signal from gantry to start exposure | Gantry synchronization signal not received | error |
079/010 | Exposure aborted by mA meter (0 pt or 1 pt technique) | mA sensor aborted exposure | error |
079/011 | Exposure: no gantry acknowledgement at end of first trigger or gantry fault | Gantry not responding at end of first trigger or malfunction | error |
079/014 | E05 Failure during exposure | XRayAuthorizedRT signal not received from IDC | error |
079/015 | E05 Failure during exposure | Brightness value not received from IDC | error |
079/017 | E16 AWS/SENO Synchro fail | Image not correctly transferred from IDC to ADS | error |
080/001 | Gene (IST, exposure): heater fault | X-ray tube heater malfunction during exposure | error |
080/002 | Gene (IST, exposure): anode starter fault | X-ray tube anode starter malfunction during exposure | error |
080/003 | Gene (IST, exposure): power failure | Power supply malfunction during exposure | error |
080/004 | Gene (IST, exposure): interface fault | Interface malfunction during exposure | error |
080/005 | Gene (IST, exposure): no gantry reply timeout | No response from gantry within 1 second during exposure | error |
080/006 | Gene (IST, exposure): gantry fault | Gantry malfunction during exposure | error |
080/007 | Gene (IST, exposure): no gantry reply to exposure command | Gantry not responding to exposure command | error |
080/008 | Gene (IST, exposure): no grid sync feedback after timeout | Grid synchronization feedback not received within timeout | error |
082/000 | Gene: Endurance error code | System endurance threshold exceeded | critical |
082/001 | Gene: Endurance error code | System endurance threshold exceeded | critical |
082/010 | Gene: Endurance error code | System endurance threshold exceeded | critical |
082/020 | Gene: Endurance error code | System endurance threshold exceeded | critical |
082/030 | Gene: Endurance error code | System endurance threshold exceeded | critical |
082/040 | Gene: Endurance error code | System endurance threshold exceeded | critical |
083/002 | I17 AWS/SENO Comm. lost | Senographe/AWS communication failure for 2 minutes | error |
084/008 | E24 Failure before exposure | Signal not correctly received from IDC or Gantry after radiography key pressed but before X-ray emission | error |
084/009 | E24 Failure before exposure | Signal not correctly received from IDC or Gantry after radiography key pressed but before X-ray emission | error |
084/013 | E24 Failure before exposure | Signal not correctly received from IDC or Gantry after radiography key pressed but before X-ray emission | error |
090/000 | Generator: PROM checksum error | PROM memory checksum validation failed | error |
092/000 (E19) | E19 IDC/Seno down, DPS is off | No answer from IDC for 2 minutes; connection lost; detector power supply shut down | error |
092/000 (I19) | I19 IDC/Seno comm. lost | No answer from IDC for 15 seconds; connection lost | error |
094/001 | E18 IDC/Seno synchro failure | Bad command received by IDC task; error in signal/message | error |
094/002 | E18 IDC/Seno synchro failure | Prep command not received by IDC task | error |
094/003 | E18 IDC/Seno synchro failure | Stop command at end of exposure not received by IDC task | error |
094/004 | E18 IDC/Seno synchro failure | XRayAuthorizedRT status not received by IDC task | error |
094/005 | E18 IDC/Seno synchro failure | Stop command not received by IDC task at end of radiography | error |
094/006 | E18 IDC/Seno synchro failure | Prep command not received by exposure automaton task T_APPLI | error |
094/007 | E18 IDC/Seno synchro failure | XRayAuthorizedRT signal active in idle status | error |
096/019 | I34 Thickness out of AOP range | Measured breast thickness outside AOP range | warning |
096/020 | I36 IFF threshold out of range | IFF (IDC algorithm) threshold outside acceptable range | warning |
096/021 | I37 9 param model deviate | Error on Scf factor for pre-exposure; corrected mechanical thickness zero or breast composition 0.6-1.4 range violation | warning |
096/023 | I05 Exposure aborted by mAs level | mAs value for pre-exposure outside acceptable range | warning |
096/024 | I06 Abort: Detect sig. out range | Dose value returned by IDC outside acceptable range | warning |
096/025 | Track/filter change | Track/filter configuration changed between pre-shot and exposure phases | info |
096/033 | Invalid dose value from IDC | IDC sent an invalid dose value to Senographe system | warning |
096/038 | Image signal too low | Image signal below acceptable range in AOP mode between pre-exposure and exposure; dose out of permitted range | error |
096/039 | Image signal too high | Image signal above acceptable range in AOP mode between pre-exposure and exposure; dose out of permitted range | error |
096/040 | No AOP zone established | Exposure aborted; automatic optimization protocol (AOP) zone not established | error |
096/041 | Thermal limit: mAs not optimal | System thermal condition reached; exposure mAs value reduced to maintain image quality limits | warning |
096/042 | Exposure time limit: mAs not optimal | Exposure time limit reached; mAs reduced within image quality limits | warning |
096/043 | Exposure time limit: mAs too low | Maximum exposure time reached; mAs reduced below regular image quality limits | warning |
097/026 | U-turn before end of pre-shot | Pre-shot exposure time exceeds grid motion duration to first turn-back point | info |
097/027 | U-turn before end of exposure | Exposure time in manual or AOP mode exceeds grid motion duration to first turn-back point | info |
100/001 | Lift hardware error | Relay defect at power-up; relay shorted, preventing H-bridge voltage verification | critical |
100/001 | Lift hardware relay defect detected at power-up | Relay mounted parallel to motor H-bridge fails continuity/short test during power-up initialization | critical |
100/002 | Upper optical fork sensor failure | At power-up, LED shutdown does not cause photoresistor to shut down, indicating optical fork failure | critical |
100/003 | Lower optical fork sensor failure | At power-up, LED shutdown does not cause photoresistor to shut down, indicating optical fork failure | critical |
100/004 | Footswitch is pressed during boot | Lift up, lift down, compression up, or compression down button pressed during node boot-up | warning |
100/005 | Footswitch is pressed during boot | Compression or lift button pressed during node boot-up | warning |
100/006 | Footswitch is pressed during boot | Button pressed during node boot-up | warning |
100/007 | Footswitch is pressed during boot | Button pressed during node boot-up | warning |
100/009 | Lift over upper software limit | Lift position not between software limits at power-up, or lift moved manually to upper position | critical |
100/010 | Lift below lower software limit | Lift position not between software limits at power-up, or lift moved manually to lower position | critical |
100/011 | Gantry display failure | LCD failure detected | error |
101/001 | Lift software error - unrecognized command | PTB sends command not in IPMS recognized command list | info |
101/002 | Lift software error - parameter range violation | PTB sends command with parameter values outside valid range | info |
101/003 | Lift calibration not achieved | Lift calibration incomplete or failed | warning |
101/004 | No valid line for lift | Gantry CPU requests motion but node does not see valid line; valid line falls before motion end | error |
101/005 | Keyboard failure - lift motion | CPU sends Start Motion command without preceding CAN command from node; 100ms timeout expires | info |
102/000 | Lift software error - watchdog reset | Watchdog timer detects microcontroller failure; no pulse received for 40ms | critical |
102/001 | Lift position error - potentiometer defect | Potentiometer signal outside acceptable range at power-up or during operation | critical |
102/002 | Lift speed error - encoder defect | Coding wheel defect detected through complemented hardware lines | critical |
102/005 | Uncontrolled lift motion - correlation error | Absolute coder (potentiometer) and coding wheel (encoder) signals not correlated | critical |
102/006 | Lift footswitch defect | Two signals from button do not arrive at same time on CPLD | critical |
102/007 | Lift footswitch defect | Two signals from button do not arrive at same time | critical |
102/008 | Compression footswitch defect | Two signals from button do not arrive at same time | critical |
102/009 | Compression footswitch defect | Two signals from button do not arrive at same time | critical |
102/010 | Lift over upper limit | Lift motion reaches upper electrical stop limit | critical |
102/012 | Lift over lower limit | Lift motion reaches lower electrical stop limit | critical |
102/013 | Lift overspeed | Lift speed measured by encoder exceeds speed profile limit | critical |
102/014 | Lift lowspeed warning | Lift speed measured by encoder below speed profile limit | warning |
102/015 | Lift over-current | Integrated current value too high; current elevated for too long | critical |
102/016 | Movement already in progress | CPU sends motion command while node performing another movement | info |
102/017 | Lift node firmware incompatibility | Hardware configuration (CPLD), DSP, or MCU firmware does not match | critical |
102/018 | Lift node parity error | Parity bit incorrect on on-board bus between CPLD and MCU | critical |
102/019 | Stop line activated due to uncontrolled motion | Same as error 102/004 - uncontrolled motion detected | critical |
102/020 | Stop line activated by CPU | Main CPU detected error and activated or failed to deactivate stop motion line | error |
102/021 | Low voltage error on lift | Power supply voltage falls below 24V | critical |
102/022 | Lift read time-out error | MCU has not reacted on interrupt request (IRQ) from CPLD before 15ms | critical |
102/023 | Lift CPLD memory error | CPLD test data does not match periodic read; communication failure detected | critical |
102/024 | Uncontrolled lift motion detected; stop line activated | Motion performed when lift node should not drive; MCU_STOP_MOTION activated | critical |
102/025 | Lift relative encoder error | Signal from relative encoder does not evolve during motion; absolute encoder signals differ | critical |
102/027 | Lift MCU calibration data corrupt | Checksum on MCU calibration data does not match | warning |
102/028 | Lift DSP calibration data corrupt | Checksum on DSP calibration data does not match | warning |
102/029 | Lift MCU EEPROM failure | Readback of newly written data to EEPROM fails | critical |
102/030 | Lift position error | Motion did not stop at target/limit position; exceeded allowed margin | unknown |
102/031 | Lift deceleration error | Motion began to slow down but did not stop before allowed time | warning |
102/032 | Lift motion started in reverse direction | Motion about to proceed in reverse of requested direction | critical |
102/033 | Lift DSP timeout | Lift DSP did not answer MCU before timeout | critical |
102/034 | Lift CAN command not processed | Gantry CPU sent command before lift node acknowledged previous command | warning |
102/035 | Lift RX CAN buffer overrun | Input queue of incoming CAN messages of MCU is full; message dropped | critical |
102/036 | Lift TX CAN buffer overrun | Transmit queue full or TP1 connector jumper closed | critical |
102/037 | Lift mailbox overrun | Inter-task communication mailbox queue in MCU is full; messages dropped | critical |
102/038 | CPU heartbeat not received by Lift node | Gantry CPU did not send periodic heartbeat CAN message | critical |
102/039 | Heartbeat disabled on Lift node | Gantry CPU attempted to disable heartbeat check of lift node | unknown |
102/04 | Lift low-speed error - encoder phase sequence | Motor encoder signal phase or complementarity check fails | critical |
102/040 | Lift invalid PDO length | PDO with invalid DLC received by lift node | critical |
102/041 | Valid line active on Lift node at power-up | Valid motion line is active at power-up | warning |
102/042 | Lift stopped without stop command | Lift node stopped motion by itself when button/pedal released before timeout | warning |
102/043 | Lift invalid brake defect error | Brake malfunction detected (should not occur) | unknown |
102/044 | Lift invalid clutch malfunction error | Clutch malfunction detected (should not occur) | unknown |
102/045 | Lift invalid user interface error | User interface function error | critical |
102/046 | Lift invalid collimator button error | Collimator button malfunction | unknown |
102/047 | Lift invalid initialization error | Initialization error (should not occur) | unknown |
102/048 | Lift invalid force sensor defect error | Force sensor defect detected (should not occur) | unknown |
102/049 | Lift invalid force sensor range error | Force sensor range error detected (should not occur) | unknown |
102/050 | Lift H-Bridge voltage error | H-Bridge voltage supply fault | critical |
102/051 | Lift unexpected event error | Software error or unexpected condition detected | critical |
102/052 | Lift invalid go rotation failure error | Rotation command failure | unknown |
102/053 | Lift invalid go angulation failure error | Angulation command failure | unknown |
103/001 | Lift end position reached (upper limit) | Start motion command when upper stop limit already active | info |
103/002 | Lift end position reached (lower limit) | Start motion command when lower stop limit already active | info |
110/001 | Rotation relay defect at power-up | At system power-up, relay check fails (relay does not create short-cut of H-bridge or does not re-open properly) | critical |
110/006 | Arm turned right too far | Arm position exceeds software upper limit at power-up (including tolerance) | critical |
110/007 | Arm turned left too far | Arm position exceeds software lower limit at power-up (including tolerance) | critical |
111/001 | Rotation software error - unrecognized command identifier | Command sent by PTB is not recognized by IPMS | info |
111/002 | Rotation software error - parameter out of range | Command parameter values incompatible with range when PTB sends command to IPMS | info |
111/003 | Rotation calibration not achieved | Calibration not completed; IPMS receives non-calibration CAN command | warning |
111/004 | No valid line for rotation | Valid real-time (RT) line not detected during motion command or falls before motion completes | warning |
111/005 | Rotation button not pressed | CAN signal indicating rotation button press not received by rotation node | info |
112/000 | Rotation software error - watchdog reset | Watchdog does not receive pulse from microcontroller every 40 ms; microcontroller resets | critical |
112/001 | Rotation potentiometer defect | Signal from potentiometer out of acceptable range | critical |
112/002 | Rotation encoder defect | Motor encoder signal phases (A, A complementary, B, B complementary) fail complementarity or sequence check | critical |
112/004 | Rotation motion uncoherent | Pulses received from coder when IPMS expects no motion (also applies to Lift, Compression, Tilt) | warning |
112/005 | Rotation correlation error | Signals from absolute coder (potentiometer) and coding wheel (motor encoder) not correlated | critical |
112/008 | Rotation brake defect | Brake current <1.5 A when released or too high when operating | critical |
112/010 | Rotation hardware limit reached | IPMS detects motion reaching electrical stop limit | critical |
112/011 | Rotation overspeed | Speed measured by encoder exceeds speed profile limit (correlation OK) | critical |
112/012 | Rotation low speed | Speed measured by encoder too low relative to speed profile (correlation OK) | warning |
112/014 | Rotation overcurrent | Integrated current value too high over sustained period | critical |
112/015 | Rotation software error - overlapping movement commands | Gantry CPU sends movement command while another movement still in progress | info |
112/016 | Rotation node firmware incompatibility | Hardware configuration (CPLD), DSP, or MCU firmware do not match | critical |
112/017 | Rotation node parity error | Parity bit error on on-board bus between CPLD and MCU | critical |
112/018 | Stop line activated due to uncontrolled rotation motion | Uncontrolled motion detected by DSP; MCU_STOP_MOTION line activated | critical |
112/019 | Stop line activated by CPU | Stop motion line active; CPU detected error or failed to deactivate line | critical |
112/020 | Low voltage error on rotation node | Power supply voltage to rotation board drops below 24 V | critical |
112/021 | Rotation read timeout error | MCU does not respond within 15 ms of receiving interrupt request (IRQ) from CPLD | critical |
112/022 | Rotation CPLD memory error | CPLD test data read during periodic check does not match expected value | critical |
112/023 | Uncontrolled rotation motion detected; stop line activated | Motion occurs when rotation node not driving it; uncontrolled motion detected | critical |
112/024 | Rotation relative encoder error | Relative encoder signal does not evolve during rotation motion while absolute encoder signal does | critical |
112/026 | Rotation MCU calibration data corrupt | Checksum on calibration data stored in MCU does not match | warning |
112/027 | Rotation DSP calibration data corrupt | DSP checksum on stored calibration data does not match | warning |
112/028 | Rotation MCU EEPROM failure | Readback on newly written data to EEPROM has failed | critical |
112/029 | Rotation position error | Rotation motion did not stop at target/limit position and exceeded allowed margin | warning |
112/030 | Rotation deceleration error | Rotation motion began to slow down but did not stop before allowed time | warning |
112/031 | Rotation motion started in reverse direction | Rotation motion about to go in reverse direction to requested direction | critical |
112/032 | Rotation DSP timeout | DSP did not answer to MCU before timeout | critical |
112/033 | Rotation CAN command not processed | CAN command not processed; main CPU sent command before rotation node acknowledged previous command | warning |
112/034 | Rotation RX CAN buffer overrun | Input queue of incoming CAN messages of MCU is full and message dropped, or TP1 connector on Rotation board is closed | critical |
112/035 | Rotation TX CAN buffer overrun | Transmit queue between MCU and DSP is full and message dropped | critical |
112/036 | Rotation mailbox overrun | Mailbox queue used in inter-task communication in MCU is full and mails dropped | critical |
112/037 | CPU heartbeat not received by rotation node | Gantry CPU did not send periodic heartbeat CAN message | critical |
112/038 | Heartbeat disabled on rotation node | Gantry CPU tried to disable heartbeat check of rotation node | warning |
112/039 | Rotation invalid PDO length | PDO with invalid DLC received by rotation node | critical |
112/040 | Valid line active on rotation node | Valid motion line is active at power-up | warning |
112/041 | Rotation stopped without stop command | Rotation node stopped motion by itself because button/pedal released and no stop order received before timeout | warning |
112/042 | Rotation node switch to angulation mode failed | Rotation node failed to switch to angulation mode | warning |
112/043 | Rotation node switch to rotation mode failed | Rotation node failed to switch to rotation mode | warning |
112/044 | Angulation clutch not ready | At power-up, rotation node checks for no current through clutch. Error occurs if multiple angulations in less than 300msec, no current at power-on, or clutch stays locked | warning |
112/045 | Rotation optical fork defect | At power-up, LED of left optical fork shuts down but photoresistor does not shut down | critical |
112/046 | Rotation invalid user interface error | User interface communication or command error detected | info |
112/047 | Rotation invalid lift up slow button error | Lift up slow button interface error detected | info |
112/048 | Rotation invalid up fast button error | Up fast button interface error detected | info |
112/049 | Rotation invalid lift down slow error | Lift down slow button interface error detected | info |
112/050 | Rotation invalid lift down fast button error | Lift down fast button interface error detected | info |
112/051 | Rotation invalid rot left slow button error | Rotate left slow button interface error detected | info |
112/052 | Rotation invalid rot left fast button error | Rotate left fast button interface error detected | info |
112/053 | Rotation invalid rot right slow button error | Rotate right slow button interface error detected | info |
112/054 | Rotation invalid rot right fast button error | Rotate right fast button interface error detected | info |
112/055 | Rotation invalid MLO left button error | MLO left button interface error detected | info |
112/056 | Rotation invalid MLO right button error | MLO right button interface error detected | info |
112/057 | Rotation invalid CC button error | CC button interface error detected | info |
112/058 | Rotation invalid collimator button error | Collimator button interface error detected | info |
112/059 | Rotation invalid initialization error | Initialization sequence error detected | info |
112/060 | Rotation invalid step lost error | Motor step loss detected during operation | info |
112/061 | Rotation invalid force sensor defect | Force sensor hardware failure detected | info |
112/062 | Rotation invalid command error | Invalid command received or processing error | info |
112/063 | Rotation invalid force sensor range error | Force sensor reading exceeds valid range | info |
112/064 | Rotation invalid SW error | Invalid software error detected | critical |
112/065 | Rotation invalid internal SW error | Internal software error detected | critical |
112/066 | Rotation invalid lift down pedal press error | Lift down pedal is depressed | info |
112/067 | Rotation invalid lift up pedal press error | Lift up pedal is depressed | info |
112/068 | Rotation invalid compression pedal press error | Compression pedal is depressed | info |
112/069 | Rotation invalid decompression pedal press error | Lift Up button on Compression Arm or Tube Head is depressed | info |
112/070 | Rotation invalid lift up slow button press error | Lift Up button on Compression Arm or Tube Head is depressed | info |
112/071 | Rotation invalid lift up fast button press error | Down Lift button on Compression Arm or Tube Head is depressed | info |
112/072 | Rotation invalid lift down slow button error | Down Lift button on Compression Arm or Tube Head is depressed | info |
112/073 | Rotation invalid lift down fast button press error | Left Rotation button on Compression Arm or Tube Head is depressed | info |
112/074 | Rotation invalid rot left slow button press error | Left Rotation button on Compression Arm or Tube Head is depressed | info |
112/075 | Rotation invalid rot left fast button press error | Left Rotation button on Compression Arm or Tube Head is depressed | info |
112/076 | Rotation invalid rot right slow button press error | Right Rotation button on Compression Arm or Tube Head is depressed | info |
112/077 | Right rotation button depressed | Right Rotation button on Compression Arm or Tube Head is pressed | info |
112/078 | MLO Left button depressed | MLO Left button on Compression Arm or Tube Head is pressed | info |
112/079 | MLO Right button depressed | MLO Right button on Compression Arm or Tube Head is pressed | info |
112/080 | CC button depressed | CC button on Compression Arm or Tube Head is pressed | info |
112/081 | Collimator button depressed | Collimator button is pressed | info |
113/001 | Rotation end position reached | StartMotion command when software stop limit already reached for rotation direction | warning |
113/002 | Rotation end position reached | StartMotion command when software stop limit already reached for rotation direction | warning |
113/003 | Angulation limit reached | StartMotion command when software stop limit already reached for angulation direction (+45° or -45°) | warning |
113/004 | Angulation limit reached | StartMotion command when software stop limit already reached for angulation direction | warning |
120/001 | Lift up button still pressed | Compression keypad lift up button (V1 or V2) continuously detected pressed | warning |
120/002 | Lift up button still pressed | Compression keypad lift up button continuously detected pressed | warning |
120/003 | Lift down button still pressed | Compression keypad lift down button continuously detected pressed | warning |
120/004 | Lift down button still pressed | Compression keypad lift down button continuously detected pressed | warning |
120/005 | Rotation button still pressed | Compression keypad rotation button continuously detected pressed | warning |
120/006 | Rotation button still pressed | Compression keypad rotation button continuously detected pressed | warning |
120/007 | Rotation button still pressed | Compression keypad rotation button continuously detected pressed | warning |
120/008 | Rotation button still pressed | Compression keypad rotation button continuously detected pressed | warning |
120/009 | CC preset button still pressed | Compression keypad preset CC button continuously detected pressed | warning |
120/010 | Preset +MLO button still pressed | Compression keypad preset +MLO button continuously detected pressed | warning |
120/011 | Preset -MLO button still pressed | Compression keypad preset -MLO button continuously detected pressed | warning |
121/001 | Compression software error - unrecognized command | PTB sent command identifier not in IPMS recognized command list | error |
121/002 | Compression software error - invalid parameter | Command parameter values outside acceptable range for IPMS processing | error |
121/003 | Compression calibration not achieved | Calibration procedure incomplete or failed; system in non-calibrated state | warning |
121/004 | No valid line for compression motion | Gantry CPU requests compression motion but CPLD valid line signal absent or drops before completion | warning |
121/005 | Compression button not pressed before motion | Gantry CPU sent Start Motion command before node received any prior button press on CAN | error |
122/000 | CPU heartbeat not received by compression node | Main CPU did not send periodic heartbeat CAN message | critical |
122/001 | Compression software watchdog timeout | Watchdog did not receive 40 ms pulse from microcontroller; microcontroller reset and node rebooted | critical |
122/002 | Compression potentiometer defect | Potentiometer signal continuously outside acceptable range during operation | critical |
122/003 | Compression encoder defect | Motor encoder (coding wheel) signals fail complementarity or phase sequence check on CPLD | critical |
122/004 | Compression motion incoherent | Encoder pulses detected during movement when no motion commanded by CPU (Lift/Rotation/Compression/Tilt) | unknown |
122/005 | Compression correlation error | Absolute coder (potentiometer) and motor encoder (coding wheel) signals not synchronized | critical |
122/006 | Compression motor step lost | Difference between commanded steps and absolute encoder detected steps exceeds threshold | critical |
122/007 | Compression overspeed | Measured compression speed exceeds speed profile limits during controlled motion | critical |
122/008 | Compression overcurrent | Integrated motor current value exceeds acceptable limit; current excessive for extended duration | critical |
122/009 | Lift up button switch defect (Halt) | 2 signals from lift up button do not arrive simultaneously on CPLD during button press | critical |
122/010 | Lift up button switch defect (Halt) | 2 signals from lift up button do not arrive simultaneously on CPLD | critical |
122/011 | Lift down button switch defect (Halt) | 2 signals from lift down button do not arrive simultaneously on CPLD during press | critical |
122/012 | Lift down button switch defect (Halt) | 2 signals from lift down button do not arrive simultaneously on CPLD | critical |
122/013 | Rotation button switch defect (Halt) | 2 signals from rotation button do not arrive simultaneously on CPLD during press | critical |
122/014 | Rotation button switch defect (Halt) | 2 signals from rotation button do not arrive simultaneously on CPLD | critical |
122/015 | Rotation button switch defect (Halt) | 2 signals from rotation button do not arrive simultaneously on CPLD during press | critical |
122/016 | Rotation button switch defect (Halt) | 2 signals from rotation button do not arrive simultaneously on CPLD | critical |
122/017 | CC preset button switch defect (Halt) | 2 signals from preset CC button do not arrive simultaneously on CPLD during press | critical |
122/018 | Preset +MLO button switch defect (Halt) | 2 signals from preset +MLO button do not arrive simultaneously on CPLD during press | critical |
122/019 | Preset -MLO button switch defect (Halt) | 2 signals from preset -MLO button do not arrive simultaneously on CPLD during press | critical |
122/020 | Compression brake low current | Current supplied to compression brakes (screw brake and manual brake) falls below acceptable limit | critical |
122/021 | Compression force sensor defect | Force sensor reading outside acceptable operating range during compression operation | critical |
122/022 | Emergency stop motion activated | Operator pressed emergency stop button on system; compression node cuts stop motion line | unknown |
122/024 | Compression software error - simultaneous motion | Gantry CPU sent second motion command before previous motion completed on same node | error |
122/025 | Compression node firmware incompatibility | Hardware configuration (CPLD), DSP firmware, or MCU firmware version mismatch detected on startup | critical |
122/026 | Node parity error | Parity bit does not set correctly on onboard bus between CPLD and MCU | critical |
122/027 | Stop line activated by CPU without deactivation | Main CPU detected error and activated or failed to deactivate stop motion line | info |
122/028 | Low voltage error on compression node | Power supply voltage falls below 24V | critical |
122/029 | Compression read timeout error | MCU does not react on IRQ from CPLD before 15ms | critical |
122/030 | Compression CPLD memory error | Test data from CPLD periodically read does not match expected values | critical |
122/031 | Compression relative encoder error | Relative encoder signal does not evolve during motion while absolute encoder signal does | critical |
122/033 | Compression command sequence incorrect | Signal command sequence wrong; incorrect signal sequence to start motor | critical |
122/034 | Compression MCU calibration data corrupt | Checksum on calibration data stored by MCU does not match | warning |
122/035 | Compression MCU EEPROM failure | Readback of newly written data to EEPROM fails | critical |
122/036 | Compression CAN command not processed | Main CPU sends command before compression node acknowledges previous command | warning |
122/037 | Compression CAN buffer overrun | Input queue of incoming CAN messages full; message dropped | critical |
122/038 | Compression mailbox overrun | Mailbox queue for inter-task communication in MCU full; mails dropped | critical |
122/039 | Heartbeat disabled on compression node (debug only) | Main CPU tried to disable heartbeat check of compression node | info |
122/040 | Compression invalid PDO length | Process Data Object length outside specification | critical |
122/041 | Valid line active on compression node at power-up | Valid motion line is active during system startup | warning |
122/042 | Compression stopped without stop command | Compression node stopped motion after button/pedal released; no stop order received before timeout | warning |
122/043 | Compression force sensor failure | No frequency signal detected from compression force sensor | critical |
122/044 | Compression invalid RX_OPTICAL_FORK1 error (debug only) | Optical fork receiver 1 communication error | info |
122/045 | Compression invalid RX_OPTICAL_FORK2 error (debug only) | Optical fork receiver 2 communication error | info |
122/046 | Compression invalid lift down pedal error (debug only) | Lift down pedal sensor communication error | info |
122/047 | Compression invalid lift up pedal error (debug only) | Lift up pedal sensor communication error | info |
122/048 | Compression invalid compression pedal error (debug only) | Compression pedal sensor communication error | info |
122/049 | Compression invalid decompression pedal error (debug only) | Decompression pedal sensor communication error | info |
122/050 | Compression invalid stop motion DSP active (debug only) | Stop motion digital signal processor active during operation | info |
122/051 | Compression invalid clutch malfunction (debug only) | Clutch control communication error | info |
122/052 | Compression invalid user interface error (debug only) | User interface communication error | info |
122/053 | Compression invalid collimator button error (debug only) | Collimator button communication error | info |
122/054 | Compression invalid relay error | Relay control communication failure | critical |
122/055 | Compression invalid upper optical fork reached (debug only) | Upper optical fork limit sensor triggered | info |
122/056 | Compression invalid lower optical fork reached (debug only) | Lower optical fork limit sensor triggered | info |
122/057 | Compression invalid initialization error (debug only) | Node initialization failure | info |
122/058 | Compression invalid uncontrolled motion (debug only) | Motion control failure detected | info |
122/059 | Compression invalid software error | Software execution or logic failure in compression node | critical |
122/060 | Compression invalid go rotation failure (debug only) | Rotation command execution failure | info |
122/061 | Compression invalid go angulation failure (debug only) | Angulation command execution failure | info |
122/062 | Compression invalid underspeed error (debug only) | Motor speed below expected threshold | info |
122/063 | Compression invalid position error (debug only) | Position feedback mismatch | info |
122/064 | Compression invalid deceleration error (debug only) | Deceleration profile failure | info |
122/065 | Compression invalid movement reverse direction (debug only) | Motion direction opposite to commanded | info |
122/066 | Compression invalid internal software error | Internal software execution failure | critical |
122/067 | Compression invalid DSP timeout elapsed (debug only) | Digital signal processor response timeout | info |
122/068 | Compression invalid transmit queue full (debug only) | CAN transmit message queue exceeded | info |
122/069 | Compression invalid lift down pedal press (debug only) | Occurs when lift down pedal is depressed | info |
122/070 | Compression invalid lift up pedal press (debug only) | Occurs when lift up pedal is depressed | info |
122/071 | Invalid compression pedal press | Compression pedal is depressed | info |
122/072 | Invalid decompression pedal press | Decompression pedal is depressed | info |
122/073 | Invalid collimator button press | Collimator button is depressed | info |
123/001 | Maximum compression force already reached | Operator presses compression footswitch when max force is reached | warning |
123/002 | Compression upper position already reached | Operator presses decompression footswitch when upper limit is reached | warning |
123/003 | Compression lower position already reached | Operator presses compression footswitch when lower limit is reached | warning |
123/004 | Negative force limit reached | Decompression motion ongoing or requested when negative force limit (-3daN) reached | warning |
130/001 | UIF Still pressed | One of the 9 buttons on the Compression Arm or Tube Head is still pressed (Lift up, Lift down, Rotation Right, Rotation Left, Light/Field Of View) | warning |
130/002 | UIF Still pressed | One of the 9 buttons on the Compression Arm or Tube Head is still pressed | warning |
130/003 | UIF Still pressed | One of the 9 buttons on the Compression Arm or Tube Head is still pressed | warning |
130/004 | UIF Still pressed | One of the 9 buttons on the Compression Arm or Tube Head is still pressed | warning |
130/005 | UIF Still pressed | One of the 9 buttons on the Compression Arm or Tube Head is still pressed | warning |
130/006 | UIF Still pressed | One of the 9 buttons on the Compression Arm or Tube Head is still pressed | warning |
130/007 | UIF Still pressed | One of the 9 buttons on the Compression Arm or Tube Head is still pressed | warning |
130/008 | UIF Still pressed | One of the 9 buttons on the Compression Arm or Tube Head is still pressed | warning |
130/010 | Tube Tilt initialization failure | At power-up, optical fork shut state mismatch or Tube Tilt motion does not reach expected transition after correct number of steps | error |
130/011 | UIF Still pressed | One of the 9 buttons on the Compression Arm or Tube Head is still pressed | warning |
130/012 | UIF Still pressed | One of the 9 buttons on the Compression Arm or Tube Head is still pressed | warning |
130/013 | UIF Still pressed | One of the 9 buttons on the Compression Arm or Tube Head is still pressed | warning |
130/014 | UIF Still pressed | One of the 9 buttons on the Compression Arm or Tube Head is still pressed | warning |
131/001 | Tube Tilt software error | Corrupt software/firmware on the Tube Tilt board | error |
131/002 | Tube Tilt software error | Corrupt software/firmware on the Tube Tilt board | error |
131/003 | Tube Tilt calibration not achieved | Tube Tilt board has not been calibrated or has lost calibration settings | error |
132/000 | Tilt Invalid PDO length | Invalid protocol data unit length from Tube Tilt board | error |
132/001 | Tube Tilt software error | Corrupt software/firmware on the Tube Tilt board linked to FW problem | error |
132/002 | Tube Tilt fan error | Fan rotation speed not within acceptable range as monitored by third wire | warning |
132/003 | Tube Tilt switch defect | Two signals from button do not arrive at the same time on the CPLD | critical |
132/004 | Tube Tilt switch defect | Two signals from button do not arrive at the same time on the CPLD | critical |
132/005 | Tube Tilt switch defect | Two signals from button do not arrive at the same time on the CPLD | critical |
132/006 | Tube Tilt switch defect | Two signals from button do not arrive at the same time on the CPLD | critical |
132/007 | Tube Tilt switch defect | Two signals from button do not arrive at the same time on the CPLD | critical |
132/008 | Tube Tilt switch defect | Two signals from button do not arrive at the same time on the CPLD | critical |
132/009 | Tube Tilt switch defect | Two signals from button do not arrive at the same time on the CPLD | critical |
132/010 | Tube Tilt switch defect | Two signals from button do not arrive at the same time on the CPLD | critical |
132/012 | Tube Tilt step lost | Tube Tilt does not reach transition after expected number of steps | error |
132/013 | Tilt Node FW incompatibility | Hardware configuration (CPLD) or DSP/MCU firmware mismatch | error |
132/014 | Tilt Node parity error | Parity bit error on on-board bus between CPLD and MCU | error |
132/015 | Low voltage error on Tilt Node | Voltage to Tube Tilt board falls below 24V | error |
132/016 | Tilt read timeout error | MCU does not react within 15ms of receiving interrupt request from CPLD | error |
132/017 | Tilt CPLD memory error | Periodic test data from CPLD does not match on read | error |
132/018 | Tilt command sequence incorrect | Checksum on calibration data stored by MCU does not match | error |
132/019 | Tilt MCU calibration data corrupt | Checksum on calibration data stored by MCU does not match | error |
132/020 | Tilt MCU EEPROM failure | Readback of newly written EEPROM data failed | error |
132/021 | Tilt CAN command not processed | Gantry CPU sent command before compression node acknowledged previous command | error |
132/022 | Tilt CAN buffer overrun | Input queue of incoming CAN messages full and message dropped | error |
132/023 | Tilt mailbox overrun | Mailbox queue for inter-task communication in MCU is full and mails dropped | error |
132/024 | Heartbeat disabled on Tilt Node | Gantry CPU tried to disable the heartbeat check of the tilt node | info |
132/025 | Tube Tilt switch defect | Two signals from button do not arrive at the same time on the CPLD | critical |
132/026 | Tube Tilt switch defect | Two signals from button do not arrive at the same time on the CPLD | critical |
132/027 | Tube Tilt switch defect | Two signals from button do not arrive at the same time on the CPLD | critical |
132/028 | Tilt overcurrent | Motion over current detected | error |
132/029 | CPU heartbeat not received by Tilt Node | Gantry CPU did not send periodic heartbeat CAN message | error |
132/030 | Tilt invalid RX_OPTICAL_FORK1_ERROR | Invalid optical fork sensor 1 signal received | info |
132/031 | Tilt invalid RX_OPTICAL_FORK2_ERROR | Invalid optical fork sensor 2 signal received | info |
132/032 | Tilt invalid LIFT_DOWN_PEDAL_ERROR | Invalid lift down pedal signal | info |
132/033 | Tilt invalid LIFT_UP_PEDAL_ERROR | Invalid lift up pedal signal | info |
132/034 | Tilt invalid COMPRESSION_PEDAL_ERROR | Invalid compression pedal signal | info |
132/035 | Tilt invalid DECOMPRESSION_PEDAL_ERROR | Invalid decompression pedal signal | info |
132/036 | Tilt invalid STOP_MOTION_MCU_ACTIVE_ERROR | MCU stop motion signal invalid or inactive when expected | info |
132/037 | Tilt invalid STOP_MOTION_DSP_ACTIVE_ERROR | DSP stop motion signal invalid or inactive when expected | info |
132/038 | Tilt invalid STOP_MOTION_LINE_ACTIVE_ERROR | Stop motion line signal invalid | info |
132/039 | Tilt invalid BRAKE_MALFUNCTION_ERROR | Brake system malfunction detected | info |
132/040 | Tilt invalid CLUTCH_MALFUNCTION_ERROR | Clutch system malfunction detected | info |
132/041 | Tilt invalid CODER_SEQUENCE_ERROR | Encoder signal sequence error | info |
132/042 | Tilt invalid CODER_SIGNAL_ERROR | Encoder signal invalid | info |
132/043 | Tilt invalid USERINTERFACE_ERROR | User interface signal problem | info |
132/044 | Tilt invalid COLLIMATOR_BUTTON_ERROR | Collimator button signal invalid | info |
132/045 | Tilt invalid RELAY_ERROR | Relay signal invalid or malfunction | info |
132/046 | Tilt invalid CORRELATION_ERROR | Signal correlation check failed | info |
132/047 | Tilt invalid UPPER_OPTICAL_FORK_REACHED_ERROR | Upper optical fork limit reached unexpectedly | info |
132/048 | Tilt invalid LOWER_OPTICAL_FORK_REACHED_ERROR | Lower optical fork limit reached unexpectedly | info |
132/049 | Tilt invalid FORCE_SENSOR_DEFECT_ERROR | Force sensor malfunction detected | info |
132/050 | Tilt invalid UNCONTROLLED_MOTION_ERROR | Motion occurred without control signal | info |
132/051 | Tilt invalid ABSOLUTE_ENCODER_RANGE_ERROR | Absolute encoder position out of valid range | info |
132/052 | Tilt invalid FORCE_SENSOR_RANGE_ERROR | Force sensor reading out of valid range | info |
132/053 | Tilt invalid SOFTWARE_ERROR | General software/firmware error | info |
132/054 | Tilt invalid GO_ROTATION_FAILURE | Rotation motion command execution failed | info |
132/055 | Tilt invalid GO_ANGULATION_FAILURE | Angulation motion command execution failed | info |
132/056 | Tilt invalid OVERSPEED_ERROR | Motion speed exceeded acceptable limit | info |
132/057 | Tilt invalid UNDERSPEED_ERROR | Motion speed below acceptable minimum | info |
132/058 | Tilt invalid POSITION_ERROR | Motor position does not match expected position | info |
132/059 | Tilt invalid DECELERATION_ERROR | Motion deceleration failed or incorrect | info |
132/060 | Tilt invalid MOVEMENT_REVERSE_DIRECTION_ERROR | Motion reversed direction unexpectedly | info |
132/061 | Tilt invalid INTERNAL_SOFTWARE_ERROR | Internal firmware processing error | info |
132/062 | Tilt invalid DSP_TIMEOUT_ELAPSED_ERROR | DSP processing did not complete within timeout | info |
132/063 | Tilt invalid TRANSMIT_QUEUE_FULL_ERROR | Transmit message queue is full | info |
132/064 | Tilt invalid NO_PREVIOUS_UIF_EVENT_ERROR | User interface event expected but not found | info |
132/065 | Tilt invalid STOP_ON_VALID_MOTION_DOWN_ERROR | Valid downward motion stop sequence failed | info |
132/066 | Tilt invalid VALID_MOTION_ACTIVE_ERROR | Valid motion detected when not expected | info |
132/067 | Tilt invalid STOP_ON_UIF_RELEASE_ERROR | Motion stop on UIF button release failed | info |
132/068 | Tilt invalid LIFT_DOWN_PEDAL_PRESS_ERROR | Occurs when the lift down pedal is depressed | info |
132/069 | Tilt invalid LIFT_UP_PEDAL_PRESS_ERROR | Occurs when the lift up pedal is depressed | info |
132/070 | Tilt invalid COMPRESSION_PEDAL_PRESS_ERROR | Occurs when the compression pedal is depressed | info |
132/071 | Tilt invalid DECOMPRESSION_PEDAL_PRESS_ERROR | Occurs when the decompression pedal is depressed | info |
140/001 | Collimator setup table corrupt | Checksum error after loading setup table at power-up | error |
140/002 | Collimator FPGA configuration failed | FPGA configuration failure at power-up from flash memory | error |
140/003 | Collimator front blade homing failure | Axis homing failure at power-up; optical sensor transition not reached | error |
140/004 | Collimator left blade homing failure | Left blade axis homing failure at power-up | error |
140/005 | Collimator rear blade homing failure | Rear blade axis homing failure at power-up | error |
140/006 | Collimator right blade homing failure | Right blade axis homing failure at power-up | error |
140/007 | Collimator filter homing failure | Filter wheel axis homing failure at power-up | error |
141/001 | Collimator front blade not ready | Gantry CPU sends command while axis not in ready status | error |
141/002 | Collimator left blade not ready | Gantry CPU sends command while left blade axis not in ready status | error |
141/003 | Collimator rear blade not ready | Gantry CPU sends command while rear blade axis not in ready status | error |
141/004 | Collimator right blade not ready | Gantry CPU sends command while right blade axis not in ready status | error |
141/005 | Collimator filter not ready | Gantry CPU sends command while filter axis not in ready status | error |
141/006 | Collimator out of FOV (front) | Invalid parameter for axis; wrong FOV value or filter type received | error |
141/007 | Collimator out of FOV (left) | Invalid parameter for left blade axis; wrong FOV value or filter type | error |
141/008 | Collimator out of FOV (rear) | Invalid parameter for rear blade axis; wrong FOV value or filter type | error |
141/009 | Collimator out of FOV (right) | Invalid parameter for right blade axis; wrong FOV value or filter type | error |
141/010 | Collimator filter type not exist | Invalid filter type parameter received by collimator | error |
141/011 | Collimator beyond physical limit (front) | FOV command exceeds blade capability on front axis | error |
141/012 | Collimator beyond physical limit (left) | FOV command exceeds blade capability on left axis | error |
141/013 | Collimator beyond physical limit (rear) | FOV command exceeds blade capability on rear axis | error |
141/014 | Collimator beyond physical limit (right) | FOV command exceeds blade capability on right axis | error |
141/016 | Collimator front blade timeout | Front blade axis does not reach target position within timeout limit | error |
141/017 | Collimator left blade timeout | Left blade axis does not reach target position within timeout limit | error |
141/018 | Collimator rear blade timeout | Rear blade axis does not reach target position within timeout limit | error |
141/019 | Collimator right blade timeout | Right blade axis does not reach target position within timeout limit | error |
141/020 | Collimator filter timeout | Filter wheel axis does not reach target position within timeout limit | error |
141/021 | Collimator PDO1 length error | PDO message from gantry CPU contains incorrect number of data bytes | error |
141/022 | Collimator PDO2 length error | PDO2 (change filter type) message contains incorrect number of data bytes | error |
141/023 | Collimator PDO3 length error | PDO3 (lamp on/off) message contains incorrect number of data bytes | error |
141/024 | Collimator PDO4 length error | PDO4 (change focal spot) message contains incorrect number of data bytes | error |
142/001 | Collimator 12V lamp power failure | 12 VDC power loss to centering lamp | warning |
142/003 | Collimator front blade overcurrent | Motor current exceeds threshold or abnormal noise detected on control board | error |
142/004 | Collimator left blade overcurrent | Motor current exceeds threshold or abnormal noise detected on control board | error |
142/005 | Collimator rear blade overcurrent | Motor current exceeds threshold or abnormal noise detected on control board | error |
142/006 | Collimator right blade overcurrent | Motor current exceeds threshold or abnormal noise detected on control board | error |
142/007 | Collimator filter overcurrent | Motor current exceeds threshold or abnormal noise detected on control board | error |
142/011 | Cannot reach requested filter position | Optical sensor signal not received at expected position during filter position change | error |
152/000 | PDU 230V AC mains is out of spec limits | AC mains voltage exceeds specified operating range | warning |
152/001 | 12VDC generator voltage off | 12V DC from generator not present | warning |
152/002 | 48V DC channel error | AC/DC module turned on but no 48V DC detected | warning |
152/003 | Power fail in AC/DC module | AC/DC module power failure detected | warning |
152/004 | AC/DC module fan failure | Fan in AC/DC module non-operational | info |
152/005 | AC/DC module over temperature | AC/DC module internal temperature exceeds limits | warning |
152/006 | Internal 12VDC voltage out of spec limits | Regulated 12V DC voltage outside specified range | info |
152/007 | Total current consumption too high | PDU total current consumption exceeds 40A | warning |
152/008 | GND offset level PDU ground too high | PDU ground level exceeds acceptable noise threshold | warning |
152/009 | Internal 13.3V voltage out of spec limits | 13.3V regulated supply outside specified range | info |
152/010 | Internal 5V DC voltage out of spec limits | 5V DC regulated supply outside specified range | info |
152/011 | Negative unregulated voltage out of spec limits | Negative unregulated supply from transformer outside limits | info |
152/012 | Positive unregulated voltage out of spec limits | Positive unregulated supply from transformer outside limits | info |
152/013 | Internal -5V DC voltage out of spec limits | Negative 5V regulated supply outside specified range | info |
152/014 | Internal 5V DC voltage out of spec limits | 5V DC regulated supply outside specified range | info |
152/015 | Internal reference voltage for AD conversions out of spec | A/D conversion reference voltage outside limits | info |
152/019 | Internal 0.6V bad, measurement circuit error | 0.6V reference circuit malfunction detected during self-test | info |
152/020 | Internal EEPROM error | PDU MCU EEPROM read/write failure or empty EEPROM on first power-on | info |
152/021 | PDU MCU reset while system on | PDU microcontroller unexpected reset during operation | warning |
152/022 | A/D converter module error on PDU board | A/D converter chip (U13) malfunction detected | error |
152/023 | Diagnostic channel selection circuit failure | PDU internal diagnostic multiplexer failure | info |
152/024 | Internal software error | PDU firmware detects logic or memory access violation | warning |
152/025 | Temperature too high or too low | MPCB box internal temperature outside +9 to +60°C range | info |
152/026 | Low power 2 fuse failure | Fuse F1 to nodes is open or faulty | critical |
152/027 | High power fuse failure | Fuse F2 to motors is open or faulty | critical |
152/028 | Low power 1 fuse failure | Fuse F3 to nodes is open or faulty | critical |
152/029 | CPU fuse failure | Fuse F4 is open or faulty | error |
152/030 | Generator on/off relay failure | Generator control relay or driver detects over/under current on coil | critical |
152/031 | Spare relay failure | Spare relay or its driver absent or faulty | warning |
152/032 | Internal A/D converter failure | Internal A/D converter error detected | error |
152/033 | Generator cannot be turned off by PDU | Generator relay failed to turn off generator or comms broken | critical |
160/000 | Internal software error - out of memory | Functional device, configuration interface, or HAL device creation fails during startup | critical |
160/001 | CORBA extraction failed | CORBA object extraction failure during initialization | critical |
160/002 | Unknown attribute change notification received | Unrecognized HAL signal, Atlas event, or attribute change notification | critical |
160/003 | Configuration file read failure | Gantry CPU unable to read configuration file at startup | critical |
160/004 | Configuration data retrieval failed | Gantry CPU unable to retrieve configuration data | critical |
160/005 | Subscription failure for Atlas events or HAL signals | Atlas event, attribute change, or HAL signal subscription fails | critical |
160/006 | State machine file access failure | Gantry CPU unable to access state machine file | critical |
160/007 | Attribute get failure | Gantry CPU failed to retrieve an attribute value | critical |
160/008 | Attribute set failure | Gantry CPU unable to set an attribute value | critical |
160/009 | Event publish failure | Gantry CPU unable to publish an event | critical |
160/011 | Unhandled error code from known node | Error code received from network node not handled by current Gantry CPU release | critical |
161/001 | Unknown UIF request received | Gantry CPU receives unrecognized UIF (user interface function) request | critical |
161/008 | Lift not operational | Lift UIF request received while lift node is not operational | info |
161/009 | Rotation not operational | Rotation UIF request while rotation node is not operational | info |
161/010 | Compression not operational | Compression UIF request while compression node is not operational | info |
161/011 | Multiple incompatible motions requested | Compression motion stopped due to inconsistent/simultaneous UIF request | info |
161/012 | Multiple incompatible lift motions requested | Lift motion stopped by inconsistent UIF request | info |
161/013 | Multiple incompatible rotation motions requested | Rotation motion stopped by inconsistent UIF request | info |
161/014 | Rotation rejected due to active compression | Rotation UIF rejected while compression motion is active | info |
161/015 | Rotation rejected due to active lift motion | Rotation UIF rejected while lift motion is active | info |
161/016 | Rotation rejected due to applied compression force | Rotation UIF rejected while compression force is applied | info |
161/017 | Lift rejected due to active compression motion | Lift UIF rejected while compression motion is active | info |
161/018 | Lift rejected due to active lift motion | Lift UIF rejected while lift motion is already active | info |
161/019 | Lift rejected due to applied compression force | Lift UIF rejected while compression force is applied | info |
161/020 | Compression rejected due to active rotation | Compression UIF rejected while rotation motion is active | info |
161/021 | Compression rejected due to active rotation lift | Compression UIF rejected while rotation/lift motion is active | info |
161/022 | Rotation already at end position | Rotation UIF rejected because gantry is already at rotation end position | info |
161/023 | Lift already at end position | Lift UIF rejected because lift is already at end position | info |
161/024 | Compression already at upper position | Compression UIF rejected because compression is already in topmost position | info |
161/025 | Maximum power-driven compression force reached | Power-driven compression has reached maximum permissible force | info |
161/026 | Maximum manual compression force reached | Manual compression has reached maximum permissible force | info |
161/027 | Risk of collision with floor - lift motion stopped | Lift motion stopped or UIF request rejected due to potential floor collision | info |
161/028 | Risk of collision with floor - rotation stopped | Rotation motion stopped or UIF rejected due to potential floor collision | info |
161/030 | Compression control failure - node did not start | Compression node fails to send acknowledgement at power-up to Gantry CPU | critical |
161/031 | Lift control failure - node did not start | Lift node fails to send acknowledgement at power-up to Gantry CPU | critical |
161/032 | Rotation control failure - node did not start | Rotation node fails to send acknowledgement at power-up to Gantry CPU | critical |
161/036 | Lift node not responding | Lift node fails to send heartbeat signal (every second) or actual value messages | critical |
161/037 | Rotation node not responding | Rotation node fails to send heartbeat signal (every second) or actual value messages | critical |
161/038 | Compression node not responding | Compression node fails to send heartbeat or actual value messages during motion | critical |
161/039 | Unknown error code from lift node | Gantry CPU receives unrecognized error code from lift node | critical |
161/040 | Unknown error code from rotation node | Gantry CPU receives unrecognized error code from rotation node | critical |
161/041 | Unknown error code from compression node | Gantry CPU receives unrecognized error code from compression node | critical |
161/043 | Rotation node communication failure | Rotation node failed to communicate with Gantry CPU or TP2 connector is closed | critical |
161/050 | Compression node communication failure | Compression node failed to communicate with Gantry CPU | critical |
161/051 | Lift node communication failure | Lift node failed to communicate with Gantry CPU | critical |
161/052 | Breast support locking mechanism node not responding | IR lock node not responding to Gantry CPU | warning |
162/000 | Lift node unknown error code received | Gantry CPU received unknown error code from lift node | warning |
162/001 | Tube tilt can affect image quality | Gantry CPU received unknown error code from tilt node | warning |
162/003 | Tilt node failed to start | Tilt node did not send acknowledgement at power-up or TP1 connector is closed | critical |
162/005 | Grid node failed to start | Grid node did not send acknowledgement at power-up | critical |
162/006 | Tilt node not responding | Heartbeat signal not sent by tilt node or TP1 connector is closed | warning |
162/007 | Grid node not responding | CAN cables faulty or heartbeat signal not sent by grid node | warning |
162/008 | Risk of collision with floor | Active angulation motion stopped or angulation request rejected due to collision detection | info |
162/016 | Move tube head to center position | Angulation not at 0 degrees | info |
162/019 | Compression status is unknown | Compression status cannot be determined | info |
162/020 | Move up the paddle holder | Paddle holder position issue | info |
162/021 | Angulation mode not allowed | Angulation mode request refused | info |
162/022 | Rotation mode not allowed | Rotation mode request refused | info |
162/023 | Risk of collision with puncture tool | Motion stopped due to collision risk with gun | info |
162/024 | Move tube head to center position | Tube head not at center position | info |
162/032 | X-ray source focus status unknown | Current x-ray source focus status not known by Gantry CPU | warning |
162/033 | X-ray source track status unknown | Current x-ray source track status not known by Gantry CPU | warning |
163/000 | Invalid image receptor type | Invalid image receptor type received by Gantry CPU | error |
163/001 | Invalid paddle type detected | Invalid paddle type received by Gantry CPU | error |
163/005 | Invalid button function identity received | Invalid button function identity received by Gantry CPU | critical |
163/006 | Invalid button state received | Invalid button state received by Gantry CPU | critical |
163/007 | Button state inconsistency | Lift or rotation button counter overflow or underflow | critical |
163/105 | Invalid language configuration | Invalid language currently configured; English used as default | warning |
164/001 | Blade format outside field of view | Blade configuration in collimator outside current FOV setting | warning |
164/002 | Machine event with unexpected state | Gantry CPU received machine event with unexpected state | warning |
164/003 | Collimator communication timeout | Timeout waiting for collimator response on Gantry CPU | error |
164/004 | Unable to load collimator parameters | Gantry CPU unable to load collimator parameters | error |
164/005 | Unknown error code from collimator node | Gantry CPU received unknown error code from collimator | error |
164/006 | Collimator node failed to start | Collimator node did not send acknowledgement at power-up | critical |
164/008 | Collimator node not responding | Heartbeat signal not sent by collimator node | warning |
165/000 | Invalid serial port configuration | Invalid serial port configuration prevents Gantry CPU-Generator communication | critical |
165/001 | COM port hardware error | Hardware error with COM port prevents Gantry CPU-Generator communication | critical |
165/002 | Received message number mismatch | Received message number does not match expected | error |
165/003 | Received message BCC mismatch | Received message block check character does not match | error |
165/004 | RX buffer full | Receive buffer overflow | error |
165/005 | TX Buffer FULL | Internal software error | error |
165/006 | GENERATOR DISCONNECTED | Generator not connected or powered down | error |
165/007 | Events count not found | Internal software error | error |
165/008 | Event number from the list not defined | Internal software error | error |
165/009 | Revision description not found in file | Internal software error | warning |
165/010 | Messages count not found | Internal software error | error |
165/011 | Messages number from the list not defined | Internal software error | error |
165/014 | Message count not defined | Internal software error | error |
165/015 | Message list not defined | Internal software error | error |
165/016 | Simulation Message description error | Internal software error | error |
165/017 | Simulation Message Flag description error | Internal software error | error |
165/018 | Message description error | Internal software error | error |
166/000 | OPEN SOCKET FAILED | Open socket operation failure | error |
166/001 | CLOSE SOCKET FAILED | Close socket operation failure | error |
166/002 | ACCEPT CONNECTION FAILED | Incoming connection acceptance failure | error |
166/003 | TCP/IP PACKET TRANSMISSION FAILED | TCP/IP packet transmission failure from gantry | error |
166/004 | TCP/IP PACKET RECEPTION FAILED | TCP/IP packet reception failure to gantry | error |
166/005 | TCP/IP PACKET TOO LONG | TCP/IP packet received but too long | error |
166/006 | TCP/IP PACKET WITH WRONG SIZE | TCP/IP packet received with wrong size | error |
166/007 | ADS IS NOT CONNECTED - TRANSMIT FAILED | Message cannot be sent to ADS | error |
167/001 | PDU NODE NOT STARTED | PDU node did not send acknowledgement at power-up | critical |
167/002 | PDU NODE NOT RESPONDING | Heartbeat signal was not sent by PDU node | critical |
167/003 | INTERNAL SOFTWARE ERROR | Unknown error code received from PDU node | critical |
167/004 | EMCY LINE FAILURE ON COMPRESSION NODE | Stop motion simulation fault or TP1 connector closed | critical |
167/005 | EMCY STOP LINE ACTIVATED | Emergency stop line is activated | critical |
167/006 | NVRAM CHECKSUM ERROR | NVRAM checksum error on Gantry CPU | critical |
167/007 | EMCY LINE FAILURE ON LIFT NODE | Stop motion simulation fault or TP2 connector closed | critical |
167/008 | EMCY LINE FAILURE ON ROTATION NODE | Stop motion simulation fault or TP2 connector closed | critical |
167/009 | INCORRECT GANTRY CONFIGURATION | Field of view setting does not match paddle position | info |
170/000 | Grid node error - invalid command sent by Poseidon | Exposure time out of range or invalid command from Poseidon system | error |
170/001 | Grid node error - command queue delay | Command received while previous command still processing | info |
170/002 | Grid node error - CAN bus overrun | CAN bus overrun condition with data loss | error |
170/003 | Grid node error - invalid PDO length | CAN PDO (Process Data Object) received with incorrect length | error |
170/004 | Grid node error - Gantry CPU heartbeat lost | No heartbeat signal received from Gantry CPU within timeout period | error |
170/005 | Grid node error - correlation error with Bucky step lost | Data correlation failure linked to Bucky step motor loss | warning |
170/006 | Grid node info - exposure time calculation adjustment | Exposure time between 20 and 50 ms | info |
190/000 | Right breast support release button still pressed at boot | Release button pressed when system boots | warning |
191/000 | Release optical fork test failed | Optical fork self-test fails at startup | critical |
191/001 | Lock optical fork test failed | Lock optical fork self-test fails at startup | critical |
191/001 | Left breast support release button still pressed at boot | Left release button pressed when system boots | warning |
191/002 | IRLOCK_CPLD_OPTICAL_FORK_ERROR | Hardware error in CPLD affecting optical fork | critical |
191/003 | IRLOCK motor malfunction | Motor fails to operate or responds abnormally | critical |
191/004 | IRLOCK motion timeout | Locking mechanism does not complete motion within expected time | critical |
191/005 | IRLOCK release left button timeout | Left release button press does not complete within expected time | warning |
191/006 | IRLOCK release right button timeout | Right release button press does not complete within expected time | warning |
191/007 | IR lock/release motion stopped due to button press | User presses button during lock or release motion | info |
191/008 | IR lock/release motion stopped - IR type lost | Breast support not registered via IR code during operation | critical |
191/009 | IR type lost and not released | Breast support not registered via IR code; cannot release mechanism | critical |
191/010 | IR release fork inactive while lock command received | Release fork does not respond while lock command is active | warning |
191/011 | IR release fork inactive while lock command - debug only | Debug-only error indication | info |
191/013 | IR lock/release movement not started | Internal motion state machine error | error |
191/014 | IRLOCK invalid message | Invalid message during inter-task communication | error |
191/015 | IRLOCK CAN command not processed | Gantry CPU sends new command before previous acknowledged | info |
191/016 | IRLOCK invalid PDO length | PDO with invalid DLC received by board | warning |
192/000 | Press both breast support locking buttons to release | Operator held right release button longer than 1 second | unknown |
192/001 | Press both breast support locking buttons to release | Operator held left release button longer than 1 second | unknown |
192/002 | Press both breast support locking buttons to release | IR interlock release failure occurred | info |
192/003 | IR interlock lock failure occurred | IR interlock lock failure during operation | info |
193/000 | IRLOCK node firmware incompatibility | Hardware configuration (CPLD) of node or DSP/MCU firmware do not match | error |
193/001 | IRLOCK node parity error | Parity bit does not set correctly on on-board bus between CPLD and MCU | error |
193/002 | Stop motion safety active (debug only) | System detected stop motion safety condition active | warning |
193/003 | IRLOCK invalid user interface error | Button timeout error reported by the CPLD | warning |
193/004 | Low voltage error on IRLOCK node | Power supply voltage falls below 24V | error |
193/005 | IRLOCK read timeout error | MCU has not reacted on interrupt request (IRQ) from CPLD before 15ms | error |
193/006 | IRLOCK XIRQ error (debug only) | XIRQ error occurred during operation | warning |
193/007 | IRLOCK CPLD memory error | CPLD test data does not match expected values during periodic verification | error |
193/008 | IRLOCK invalid internal software error | Internal software error detected on Breast Support Locking Mechanism Board | warning |
193/009 | IRLOCK CAN task stack low | CAN task stack low software error on Breast Support Locking Mechanism Board | warning |
193/010 | IRLOCK UIF task stack low | UIF task stack low software error on Breast Support Locking Mechanism Board | warning |
193/011 | IRLOCK ACK task stack low | ACK task stack low software error on Breast Support Locking Mechanism Board | warning |
193/012 | IRLOCK MOT task stack low | MOT task stack low software error on Breast Support Locking Mechanism Board | warning |
193/013 | IRLOCK EEPROM task stack low | EEPROM task stack low software error on Breast Support Locking Mechanism Board | warning |
193/014 | IRLOCK BKG task stack low | BKG task stack low software error on Breast Support Locking Mechanism Board | warning |
193/015 | IRLOCK RX CAN buffer overrun | Input queue of incoming CAN messages of the MCU is full; message dropped | error |
193/016 | IRLOCK TX CAN buffer overrun | Transmit queue between MCU and DSP is full; message dropped | error |
193/017 | IRLOCK mailbox overrun | Mailbox queue used in inter-task communication in MCU is full; mails dropped | error |
193/018 | CPU heartbeat not received by IRLOCK node | Gantry CPU did not send periodic heartbeat CAN message to Breast Support Locking Mechanism Board | error |
193/019 | Heartbeat disabled on IRLOCK node (debug only) | Gantry CPU tried to disable heartbeat check of Breast Support Locking Mechanism Board | warning |
E55 | Tube angulation fault | Abnormal tube angle detected during positioning | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.