GE Healthcare SENOGRAPH: troubleshooting and error codes
The ServiceForge knowledge base holds 11 documented symptoms for the GE Healthcare SENOGRAPH, with 16 ranked causes, 21 diagnostic checks with test points and expected values, and 3 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Connection lost to AW
Error message displayed by the AW CCOW Enabler indicating loss of connection to the AW workstation
Ranked causes
- 1. IP address of AW workstation not correctly configured (common)
- 2. Network setup prevents CCOW Enabler from communicating with AW (common)
- 3. CCOW communication ports not enabled in firewall (common)
- 4. User rejected synchronization acknowledgement (common)
Verify AW IP address configuration
- Test point
- AW workstation network settings
- Expected
- IP address correctly configured per Section 5 steps 3 and 4
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
FCC icon not displayed on system tray
The Fusion Context Channel icon is not visible on the Windows system tray
Ranked causes
- 1. CareFx Context Manager services not running (common)
- 2. Security properties incorrectly configured (common)
Verify CareFx Context Manager services status
- Test point
- Windows Services panel
- Expected
- CareFx Context Manager services running
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Synergy connection test fails between Windows computer and AW
Test button in Synergy configuration returns error, connection cannot be established with AW computer.
Ranked causes
- 1. SoftSwitch feature not enabled on AW computer (common)
- 2. Windows computer and AW cannot communicate on network (occasional)
Verify SoftSwitch is enabled on AW computer
- Test point
- AW Service Tools / Utilities / SoftSwitch configuration
- Expected
- SoftSwitch feature enabled
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AW CCOW Enabler fails to connect to AW computer
Error message appears when starting AW CCOW Enabler, indicating the system cannot connect to the configured AW computer.
Ranked causes
- 1. Incorrect or missing IP address of AW4.4/SA2.1 workstation in configuration (common)
Verify IP address of AW4.4 or SA2.1 workstation is entered in configuration
- Test point
- AW CCOW Proxy Configuration window, New AW machine field
- Expected
- Valid IP address of target AW workstation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
AW CCOW Enabler fails to start
Error message appears during startup of AW CCOW Enabler application
Ranked causes
- 1. Java Runtime Environment not properly installed (common)
Verify Java Runtime Environment installation
- Test point
- System environment
- Expected
- Java Runtime Environment properly installed and configured
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Balloon messages not displayed correctly
Some characters are not displayed in the messages
Ranked causes
- 1. Regional and language settings incorrect (common)
Verify Regional and Language settings
- Test point
- Windows Control Panel Regional and Language Options
- Expected
- English (United States) selected for Standards and formats and non-Unicode programs
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Connection lost to Context Manager
AW CCOW Proxy window displays connection lost message
Ranked causes
- 1. Leave Context button clicked in AW CCOW Proxy window (common)
Restore connection to Context Manager
- Test point
- AW CCOW Proxy window
- Expected
- Connection to Context manager restored, similar window displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Data loss in RIS system during context change patient safety
RIS system may not save data modifications when context change request from AW4.4/Seno Advantage 2.1 causes switch to another patient without warning before report completion
Ranked causes
- 1. Context change from AW4.4 initiated while user creating or dictating report (common)
Verify context change initiation is enabled
- Test point
- Service Tools Utilities/CCOW Configuration menu
- Expected
- Enable context change from AW option is checked and working (selection changes in AW Patient List initiate context change)
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No synchronization with CCOW Context Manager
Cannot get the CareFx Context Manager and AW CCOW Enabler to sync up on the PenRad RIS PC
Ranked causes
- 1. RIS PC not part of domain (occasional)
Verify RIS PC domain registration
- Test point
- Windows System Properties Computer Name tab
- Expected
- Domain name is registered
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
RIS system does not synchronize with AW context change patient safety
RIS system blocks patient change without saving modifications or fails to switch to new patient when context change initiated from AW4.4
Ranked causes
- 1. RIS system not synchronized with AW context change (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Softswitch error message at login when not installed
System reports that softswitch key is not installed or wrong in popup message at login, but it has not been intentionally installed
Ranked causes
- 1. Enable Softswitch checkbox inadvertently selected (common)
Check Enable Softswitch checkbox status
- Test point
- CSD/Utilities/softswitch configuration GUI
- Expected
- Enable Softswitch checkbox not selected
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
AW CCOW Enabler startup error | System unable to connect to configured AW computer at startup | AW CCOW Enabler starts without valid AW workstation IP address in configuration or AW workstation unreachable | warning |
Invalid or no license key for CCOW functionality | SoftSwitch license validation failed | At next logout/login after saving SoftSwitch configuration with invalid license key | error |
Synergy connection test failure | Synergy cannot establish connection with AW computer during test | Test button clicked in Synergy configuration and connection cannot be established | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI