GE Healthcare Seno Advantage 2.1: troubleshooting and error codes

The ServiceForge knowledge base holds 7 documented symptoms for the GE Healthcare Seno Advantage 2.1, with 8 ranked causes, 23 diagnostic checks with test points and expected values, and 1 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Monitor serial number mismatch

Serial number displayed by SMfit ACT differs from label on rear panel of CRT or LCD monitor

Ranked causes

  1. 1. EEPROM S/N register not programmed or corrupted (common)
  2. 2. Monitor connection issue or SMfit ACT detection error (occasional)
First check

Compare SMfit ACT displayed serial number with rear panel label for Monitor 2

Test point
Monitor 2 (portrait monitor closest to LCD Browser monitor)
Expected
Serial number from System > Serial Number menu matches rear panel label

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: LCD portrait monitors brightness and contrast

Two Siemens/Eizo LCD portrait monitors used for filming must be calibrated to match brightness and contrast levels and be evenly matched to each other.

Ranked causes

  1. 1. Initial or periodic calibration required (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Luminance meters validation caution

Verification that Luminance meters (SSM and ASLM) are properly calibrated before use on monitors.

Ranked causes

  1. 1. Luminance meters must be verified in calibration before monitor calibration (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Setup Calibration Target Values

Configuration of monitor gamma correction, ambient light values (ALV), and luminance thresholds (Lmin/Lmax) for Siemens/Eizo LCD portrait monitors in SMfitACT service application

Ranked causes

  1. 1. Calibration procedure for monitor display parameters (common)
First check

Verify monitors are powered on

Test point
Monitor power indicators
Expected
Monitors powered on

15 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Siemens/Eizo LCD portrait monitors caution

Calibration of two LCD portrait monitors for brightness and contrast matching to film standards; monitors must be evenly matched to each other.

Ranked causes

  1. 1. Monitors require periodic calibration for imaging accuracy (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: SMfit ACT Preferences for Siemens/Eizo monitors

Detect Portrait Monitor type and S/N in Preferences menu; create backup folder for calibration parameters after system installation or software reload

Ranked causes

  1. 1. System installation or software reload requires calibration setup (common)
First check

Verify SMfit ACT Calibration application starts successfully

Test point
Applications window
Expected
First screen of SMfit ACT Calibration software displays

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: SMfit ACT Preferences setup for portrait monitors

Detect monitor type and serial number in SMfit ACT Calibration software preferences, create backup folder for calibration parameters after system installation or software reload

Ranked causes

  1. 1. First system installation or software reload requires preferences setup (common)
First check

Select monitor type according to installation (CRT or Siemens/Eizo LCD portrait)

Test point
Prerequisites section
Expected
Either Job Card IST010 (CRT) or Job Card IST011A (Siemens/Eizo LCD)

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
SMfit ACT error messagesError or information messages occurring during monitor calibrationDuring calibration processwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI