GE Healthcare Proteus XR/a: troubleshooting and error codes
The ServiceForge knowledge base holds 227 documented symptoms for the GE Healthcare Proteus XR/a, with 383 ranked causes, 498 diagnostic checks with test points and expected values, and 305 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Generator does not reply to system high voltage
System cannot communicate with generator; kV control LEDs may show specific patterns, RESET, or HALT status, or no LEDs lit
Ranked causes
- 1. No power on Generator (common)
- 2. EMC filter failure (occasional)
- 3. AC/DC diode bridge failure (occasional)
- 4. LVPS board failure (occasional)
- 5. CAN cable problem (occasional)
- 6. kV control board failure (occasional)
- 7. Heater board failure (occasional)
- 8. Rotation board failure (occasional)
- 9. AEC board failure (occasional)
- 10. IF board failure (occasional)
Check if kV control LEDs S0-S7 are lit successively
- Test point
- kV control board LEDs S0-S7
- Expected
- LEDs lit successively
10 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
kV regulation error high voltage
Slow speed safety circuit triggered by smooth kV spits or regulation problems
Ranked causes
- 1. Smooth HV tube spits (common)
- 2. kV control board regulation problem (common)
- 3. Too much line impedance (occasional)
- 4. Half of AC/DC capacitors open (occasional)
- 5. Inverter parallel inductor or filtering capacitors (occasional)
- 6. HV tank (occasional)
Error code 0311 reported during inverter short circuit test high voltage
No primary inductor current (Ilp) detected during short circuit diagnostic
Ranked causes
- 1. Parallel inductor disconnected or faulty (common)
- 2. Parallel inductor impedance incorrect (occasional)
- 3. Inverter capacitors broken (occasional)
- 4. Current transformer to gate_cmd board cable faulty (rare)
- 5. HV Tank to kV control cable faulty (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code 0323 reported high voltage
Both Ilr and Ilp currents not detected during inverter diagnostics
Ranked causes
- 1. -15 V DC supply not present on kV control board (common)
- 2. Current transformer cable disconnected (common)
- 3. Inverter inductors disconnected (common)
- 4. Gate_cmd to HV Tank or HV Tank to kV control cable fault (occasional)
- 5. Inverter fault (rare)
Verify -15 V supply LED on kV control board
- Test point
- DS1 LED on kV control board
- Expected
- LED lit
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code 0323 reported during inverter short circuit test high voltage
Both primary and series inductor currents (Ilp and Ilr) not detected
Ranked causes
- 1. -15V supply not present on kV control board (common)
- 2. Current transformer to gate_cmd board cable faulty (occasional)
- 3. Inverter inductors faulty (occasional)
- 4. HV Tank and kV control cabling faulty (occasional)
- 5. Inverter faulty (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
3-phase CB1 breaker trips or AC fuses blown high voltage
Breaker trips or fuses blow in PDU when system powered up, indicating short circuit in generator power distribution
Ranked causes
- 1. IGBT in short circuit (common)
- 2. AC/DC or bridge rectifier in short circuit (common)
- 3. Bridge rectifier in short circuit (occasional)
- 4. EMC filter in short circuit (occasional)
Disconnect DC bus cables between AC/DC and inverter
- Test point
- DC bus cables on AC/DC side
- Expected
- Cables disconnected without short circuit
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AEC diagnostics failed patient safety
Diagnostic procedure to verify AEC (Automatic Exposure Control) function and ion chamber circuit operation
Ranked causes
- 1. AEC board fault (common)
- 2. IF board communication error (occasional)
- 3. kV control board communication fault (occasional)
- 4. Incorrect AEC board wiring or ion chamber connection (occasional)
Verify application mode operating without AEC
- Test point
- Operator console in non-AEC mode
- Expected
- System operates normally
9 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Cassette size display radiation
Calibrate cassette detection and display sizing for correct field-of-view reporting on collimator
Ranked causes
- 1. Cassette sensing system not calibrated (common)
- 2. Baseline voltage offset TP5/TP6 out of spec (common)
- 3. Longitudinal film size sensor signal out of range (common)
- 4. Lateral film size sensor signal out of range (common)
Verify cassette sensing cable is connected
- Test point
- Cassette sensor connector
- Expected
- Cable fully seated and secured
- Tools
- Visual inspection
10 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SG100 Wall Stand SID (Source-to-Image Distance) radiation
Calibrate SID display for SG100 Wall Stand with Bucky positioning and potentiometer adjustment
Ranked causes
- 1. Microswitch S4/S5 Activator Cam positions require adjustment (occasional)
- 2. System Console parameter 9 does not match actual Microswitch position distance (common)
- 3. Potentiometer P4 out of adjustment (SG100 SID) (common)
- 4. Potentiometer PA2 requires auxiliary adjustment (occasional)
Position Bucky for SG100 calibration
- Test point
- Bucky rotation and angulation
- Expected
- Rotate to 0° and angulate to 90° (horizontal)
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SG120 Wall Stand SID (Source-to-Image Distance) radiation
Calibrate SID display for SG120 Wall Stand with Bucky positioning and potentiometer adjustment
Ranked causes
- 1. Microswitch S4/S5 Activator Cam positions require adjustment for SG120 (occasional)
- 2. System Console parameter 9 does not match actual SG120 Microswitch position distance (common)
- 3. Potentiometer P4 out of adjustment (SG120 SID) (common)
- 4. Potentiometer PA2 requires auxiliary adjustment for SG120 (occasional)
Position Bucky for SG120 calibration
- Test point
- Bucky rotation and angulation
- Expected
- Rotate to 0° and angulate to 90° (horizontal)
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code 0311 reported high voltage
No Ilp (parallel inductor current) detected during inverter diagnostics
Ranked causes
- 1. Parallel inductor disconnected or impedance fault (common)
- 2. Parallel inductor impedance not zero (occasional)
- 3. Inverter capacitor set broken (occasional)
- 4. kV control board fault (rare)
Verify parallel inductor connection
- Test point
- Parallel inductor connectors on inverter
- Expected
- Securely connected
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code 0312 reported high voltage
No Ilr (series inductor current) detected during inverter diagnostics
Ranked causes
- 1. Series inductor disconnected or faulty (common)
- 2. Series current transformer impedance fault (occasional)
- 3. Inverter capacitor set broken (occasional)
- 4. kV control board fault (rare)
Verify series inductor connections
- Test point
- Series inductor connectors on inverter
- Expected
- Securely connected
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code 0312 reported during inverter short circuit test high voltage
No series inductor current (Ilr) detected during short circuit diagnostic
Ranked causes
- 1. Series inductors disconnected or faulty (common)
- 2. Series current transformer faulty (occasional)
- 3. Inverter capacitors broken (occasional)
- 4. HV Tank to kV control cable faulty (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code 0314 reported high voltage
Ilr current resonant frequency is lower than expected
Ranked causes
- 1. Series inductors disconnected or faulty (common)
- 2. Inverter capacitor set broken (occasional)
- 3. Current transformer impedance fault (occasional)
- 4. kV control board fault (rare)
Verify inductor connections
- Test point
- Inductor connectors on inverter
- Expected
- Securely connected
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Filament permanent open circuit high voltage
Open circuit detected in filament and 3 restarts attempted without success within 100 ms
Ranked causes
- 1. X-ray tube filament open (common)
- 2. Heater to HV tank cable disconnected (occasional)
- 3. Cathode HV cable or pin contacts failing (occasional)
- 4. Open circuit in filament transformer inside HV Tank (occasional)
Measure filament winding resistance
- Test point
- Heater transformer primary or secondary
- Expected
- Resistance measured (not open)
Lamp cannot be switched on
Lamp fails to light when lamp button on front panel is pushed
Ranked causes
- 1. Lamp failure (common)
- 2. UIF board failure (occasional)
- 3. System lamp power failure (occasional)
- 4. UIF board or UIF panel failure (rare)
Check cable connections
- Test point
- J4, J7, J8 on user interface board
- Expected
- Stable cable connection
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
MAINS_DROP detected (heater) caution
Firmware detected mains_drop signal activation and transmitted error to kV control
Ranked causes
- 1. Interference spikes (occasional)
- 2. Mains power drop (occasional)
- 3. Cable or connector contact in DC bus (occasional)
- 4. Heater board failure (occasional)
Check DC bus voltage stability
- Test point
- DC bus between AC/DC and auxiliary unit
- Expected
- Stable, no dropout
MAINS_DROP detected (rotation) caution
Firmware detected mains_drop signal activation and transmitted error to kV control
Ranked causes
- 1. Interference spikes (occasional)
- 2. Mains power drop (occasional)
- 3. Cable or connector contact issue in DC bus (occasional)
- 4. Rotation board failure (occasional)
Measure mains supply voltage stability
- Test point
- AC mains input
- Expected
- Stable, no dropout
Rotation current openload caution
Rotation board detected no current flowing to the motor
Ranked causes
- 1. Tube stator winding open circuit (common)
- 2. Incorrect wiring open (occasional)
- 3. No DC bus on Rotation board (occasional)
- 4. Rotation board failure (occasional)
Measure rotation motor winding resistance
- Test point
- Tube stator winding terminals
- Expected
- Resistance measured (not open)
AEC diagnostics result 0001000 patient safety
AEC to IF board communication error detected
Ranked causes
- 1. AEC board fault (common)
- 2. IF board communication fault (occasional)
- 3. kV control board communication fault (occasional)
Replace AEC board
- Test point
- AEC board connector
- Expected
- AEC board properly installed and connected
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bucky angulation motion not smooth or lock does not release (SG120) patient safety
SG120 bucky angulation lacks smooth motion or angulation lock fails to release
Ranked causes
- 1. Bearing wear or misalignment at bucky pivot point (occasional)
- 2. Angulation lock board electrical fault (occasional)
- 3. Lock hardware misalignment or wear (occasional)
Press angulation lock pushbutton and attempt bucky tilt
- Test point
- Bucky angulation motion
- Expected
- Bucky tilts forward 20° and backward 90° freely and smoothly
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bucky rotation does not lock or slips during locked state (SG100) patient safety
Bucky rotation lock mechanism fails to prevent rotation or allows rotation when locked
Ranked causes
- 1. Lock hardware misalignment or wear (occasional)
- 2. Brake pad wear or improper adjustment of threaded rods (common)
- 3. Arm assembly wear requires replacement (rare)
Release rotation lock lever and attempt bucky rotation
- Test point
- Bucky rotation range
- Expected
- Bucky rotates easily from 0° to 180° (upright to upside down)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SG100 wall stand SID display radiation
Source-to-image distance (SID) display on OTS console must match actual physical distance when Bucky moves vertically
Ranked causes
- 1. SID sensor output not calibrated (common)
- 2. Microswitch S4/S5 activator cam positions incorrect (common)
- 3. Potentiometer PA3 scaling incorrect (common)
Verify SID display changes when moving OTS vertically
- Test point
- OTS console SID display
- Expected
- Display increments/decrements as OTS moves up and down
- Tools
- OTS console, measure tape on collimator
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SG120 wall stand SID display radiation
Source-to-image distance (SID) display on OTS console for SG120 wall stand must match actual physical distance when Bucky moves vertically
Ranked causes
- 1. SID sensor output not calibrated for SG120 (common)
- 2. Microswitch S4/S5 activator cam positions incorrect on SG120 (common)
- 3. Potentiometer PA3 scaling incorrect for SG120 (common)
Verify SID display changes when moving OTS vertically on SG120
- Test point
- OTS console SID display
- Expected
- Display increments/decrements as OTS moves
- Tools
- OTS console, measure tape
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SID (Source-to-Image Distance) display radiation
Adjust potentiometers and verify SID display on OTS console matches actual table and OTS positions throughout travel range
Ranked causes
- 1. Potentiometer PA1 out of adjustment (table SID) (common)
- 2. Potentiometer PA3 out of adjustment (OTS vertical SID) (common)
- 3. Potentiometer P5 requires calibration adjustment (common)
Ensure DIP Switch settings in red color
- Test point
- DIP Switch positions 1 and 2
- Expected
- On
7 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Workstation Parameters
Configure workstation selection, tube, exposure chamber, screen-film combination, and function key assignments using JEDI service software
Ranked causes
- 1. Workstation parameters not configured or incorrect (common)
- 2. Renard or Normal kV scale not set correctly (occasional)
- 3. Date and time not set (occasional)
Verify workstation type selected
- Test point
- TECH menu display
- Expected
- Correct workstation type shown in reverse video
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code 0001000 reported during AEC diagnostic
AEC board communication error detected
Ranked causes
- 1. AEC board faulty (common)
- 2. IF board faulty (occasional)
- 3. kV control board faulty (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code 0314 reported during inverter short circuit test high voltage
Series current (Ilr) resonant frequency lower than expected
Ranked causes
- 1. Series inductors disconnected or faulty (common)
- 2. Inverter capacitors broken (occasional)
- 3. Current transformer impedance incorrect (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ILP and ILR current not OK high voltage
No inverter current measured at beginning of exposure
Ranked causes
- 1. Inverter LC resonant circuit failure (common)
- 2. Inverter board failure (occasional)
- 3. kV control board failure (occasional)
Measure both ILP and ILR currents at exposure start
- Test point
- Inverter resonant circuit current outputs
- Expected
- Both currents present within 0.5 ms
ILP current not OK high voltage
Current in parallel resonant circuit of inverter did not rise at beginning of exposure
Ranked causes
- 1. Inverter LC resonant circuit failure (common)
- 2. Inverter board failure (occasional)
- 3. kV control failure (occasional)
Measure LC resonant circuit current at exposure start
- Test point
- Inverter parallel resonant circuit
- Expected
- Current rises within 0.5 ms
ILR current not OK high voltage
Current in serial resonant circuit of inverter did not rise at beginning of exposure
Ranked causes
- 1. Inverter LC resonant circuit failure (common)
- 2. Inverter board failure (occasional)
- 3. kV control failure (occasional)
Measure LC resonant circuit current at exposure start
- Test point
- Inverter serial resonant circuit
- Expected
- Current rises within 0.5 ms
ILR Current time out high voltage
Current resonant frequency is lower than expected
Ranked causes
- 1. kV control failure (common)
- 2. Inverter board failure (occasional)
- 3. Current measurement circuit failure (occasional)
Measure resonant circuit frequency
- Test point
- ILR circuit oscillation frequency
- Expected
- Within specification
mA scale error radiation
mA measured too low or too high relative to mA demand 20 ms after exposure start
Ranked causes
- 1. kV control board failure (common)
- 2. Default filament currents not correctly adjusted (occasional)
- 3. HV Tank failure (rare)
Measure actual mA and compare to demand value
- Test point
- mA feedback and mA command
- Expected
- Actual mA within spec of demanded value
No CAN message received (heater board)
Heater board has not received command from kV control main software for 5 seconds, loss of communication
Ranked causes
- 1. kV Control main software lost (common)
- 2. kV control or Heater board driver failure (common)
- 3. Bad contact on CAN bus line connector (occasional)
Verify kV control main software status
- Test point
- Software status display
- Expected
- Software active
No CAN message received (rotation board)
Rotation board has not received any signal from kV control main software for the last 5 seconds, interpreted as loss of communication
Ranked causes
- 1. kV Control main software lost (common)
- 2. kV control or Rotation board driver failure (common)
- 3. Bad contact on CAN bus line connector (occasional)
Verify kV control main software is running
- Test point
- Software status display
- Expected
- Software active
No kV feedback on anode side radiation
kV measured less than 12 kV at 0.5 ms after start of exposure on anode side only
Ranked causes
- 1. HV cable short circuit (occasional)
- 2. HV tank failure (occasional)
- 3. kV control board failure (rare)
Measure anode-side kV 0.5 ms after exposure start
- Test point
- Anode kV feedback circuit
- Expected
- >12 kV
No kV feedback on cathode side radiation
kV measured less than 12 kV at 0.5 ms after start of exposure on cathode side only
Ranked causes
- 1. HV cable short circuit (occasional)
- 2. HV tank failure (occasional)
- 3. kV control board failure (rare)
Measure cathode-side kV 0.5 ms after exposure start
- Test point
- Cathode kV feedback circuit
- Expected
- >12 kV
No mA feedback radiation
No mA feedback detected 20 ms after beginning of exposure
Ranked causes
- 1. HV tank failure (occasional)
- 2. kV control board failure (occasional)
- 3. X-ray tube filament open (occasional)
Measure mA 20 ms after exposure start
- Test point
- mA feedback from HV tank
- Expected
- mA measured above threshold
Rotation Inverter permanent overcurrent high voltage
Overcurrent detected and 3 restart attempts failed unsuccessfully within a single rotation state
Ranked causes
- 1. Rotation board failure (common)
- 2. Tube stator winding short circuit (occasional)
- 3. Wiring shorted (occasional)
Check stator winding continuity and insulation
- Test point
- Tube stator winding terminals
- Expected
- Resistance within spec; no short to ground
Rotation phases unbalanced caution
Amplitude difference of current between main and auxiliary rotation phases is too large
Ranked causes
- 1. One rotation wire missing (occasional)
- 2. Rotation board failure (occasional)
- 3. Wrong tube configuration database (rare)
Measure current amplitude main vs auxiliary phases
- Test point
- Main and auxiliary rotation windings
- Expected
- Amplitude difference within spec
SG100 vertical locks do not function correctly patient safety
Vertical locks either do not engage or remain engaged when operating lock handle
Ranked causes
- 1. No 24 VDC between TP-9 and TP-10 (common)
- 2. Defective handle microswitch or relay K11 (occasional)
- 3. Relay K14 malfunction (lock release delayed) (rare)
Measure voltage at lock board test points
- Test point
- TP-9 and TP-10
- Expected
- 24 VDC
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Three-phase CB1 breaker trips or AC fuses blown high voltage
Main AC input protection device trips or fuses blow on generator power supply, indicating short circuit in power distribution
Ranked causes
- 1. IGBT in short circuit in inverter (common)
- 2. AC/DC or bridge rectifier in short circuit (common)
- 3. EMC filter in short circuit (occasional)
Disconnect DC bus cables between AC/DC and inverter
- Test point
- DC bus cables on AC/DC board side
- Expected
- Cables safely disconnected with no voltage present
5 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Tube spit detection - kV- dropped high voltage
X-ray tube spit with high voltage cathode side voltage dropped
Ranked causes
- 1. Cathode side tube spit (common)
- 2. Cathode HV cable (occasional)
- 3. HV tank (occasional)
Tube spit detection - kV+ dropped high voltage
X-ray tube spit with high voltage anode side voltage dropped
Ranked causes
- 1. Anode side tube spit (common)
- 2. Anode HV cable (occasional)
- 3. HV tank (occasional)
Vertical locks do not engage or remain locked (SG100) patient safety
SG100 wall stand vertical locks fail to engage when lock handle released or remain engaged when handle operated
Ranked causes
- 1. Loss of 24 VDC to electromechanical lock (common)
- 2. Handle microswitch failure (occasional)
- 3. Relay K11 malfunction (occasional)
Verify 24 VDC supply to lock system
- Test point
- TP-9 and TP-10
- Expected
- 24 VDC present when lock handle released
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
AEC ion chamber HV power supply test failed
AEC HV power supply voltage measurement incorrect for ion chamber cells
Ranked causes
- 1. AEC HV power supply faulty (common)
- 2. Ion chamber cell wiring faulty (occasional)
Measure HV power supply with no cell selected
- Test point
- AEC multiplex output
- Expected
- 0 V
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bucky angulation does not lock or releases inadvertently (SG100) patient safety
Bucky angulation lock fails to hold position or binds during angulation motion
Ranked causes
- 1. Lock hardware misalignment or wear (occasional)
- 2. Bearing wear at bucky pivot point (occasional)
Release angulation lock and tilt bucky forward and backward
- Test point
- Bucky angulation range and motion smoothness
- Expected
- Bucky tilts forward 20° and backward 90° freely and smoothly
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bucky angulation lock does not release caution
Angulation lock push button does not release the lock, or Bucky cannot tilt freely forward 20 degrees and backward 90 degrees.
Ranked causes
- 1. Lock board malfunction (common)
- 2. Bearing wear or damage at Bucky pivot point (occasional)
Release angulation lock and verify Bucky tilts freely
- Test point
- Bucky tilt mechanism
- Expected
- Bucky tilts forward 20 degrees and backward 90 degrees freely and smoothly
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: AEC KG parameter (grid correction) radiation
Ionization chamber 2 grid correction calibration for multi-chamber systems using automatic exposure control. Applies only when multiple ionization chambers present.
Ranked causes
- 1. KG calibration required for second ionization chamber in multi-chamber setup (common)
- 2. KG calibration in application mode when SID differs from KVEF setup (occasional)
Select screen-film combination A (slowest preferred)
- Test point
- Calibration menu
- Expected
- Film type A selected
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: AEC KTIME parameter (exposure time correction) radiation
Calibration of exposure time correction parameters for automatic exposure control across 4 or 6 time intervals (10ms to 3.2s).
Ranked causes
- 1. KTIME calibration needed for 4 or 6 exposure time values (common)
- 2. Film darkness correction needed for exposure times below 10 ms (occasional)
Verify KTIME preliminary T values
- Test point
- Service mode MENU4/AEC/F5/KTIME
- Expected
- T0=10ms, T1=100ms, T2=500ms, T3=1000ms, T4=2000ms, T5=3200ms
- Tools
- Service mode access
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Automated AEC Calibration radiation
Perform AEC (Automatic Exposure Control) system calibration using sensitometer and densitometer to ensure consistent film density
Ranked causes
- 1. AEC calibration not performed or requires update (common)
- 2. Positioner, collimator, or X-ray generator not configured (occasional)
Perform tube warmup
- Test point
- X-ray tube
- Expected
- Tube warmed to operating temperature
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Backup & Restore Database Setup Parameters
Backup CPU memory parameters to floppy disk for rapid restoration or recovery after KV Control Board change
Ranked causes
- 1. Need to preserve CPU NVRAM setup parameters for system continuity (common)
- 2. Need to restore CPU NVRAM after KV Control Board replacement (occasional)
Verify service laptop configuration and JEDI software installation
- Test point
- Service laptop
- Expected
- Laptop configured with JEDI service software installed
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Film size detection
Film size display incorrect after cassette insertion; requires voltage adjustment at test points
Ranked causes
- 1. Incorrect voltage at film size sensor TP9 (common)
- 2. Incorrect voltage at film size sensor TP10 (common)
Measure voltage between TP9 and GND
- Test point
- TP9
- Expected
- –0.84V±0.01V, ±0.01V
- Tools
- DMM
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Film size display incorrect
Film size display does not match inserted cassette size after cassette insertion or repositioning
Ranked causes
- 1. Cassette position sensor voltage out of calibration (common)
- 2. Second cassette sensor voltage out of calibration (common)
Measure sensor voltage at TP9
- Test point
- TP9 to GND
- Expected
- -0.84V±0.01V, ±0.01V
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: JEDI 38-2 accuracy and reproducibility radiation
Verify mAs display accuracy across 60/80/120 kVp and reproducibility of exposures in 2-point mode; measure coefficient of variation (C) and linearity (L)
Ranked causes
- 1. Calibration adjustment required (unknown)
- 2. Reproducibility and linearity out of tolerance (unknown)
Verify mAs display accuracy at 60 kVp
- Test point
- Control Console post-display
- Expected
- At 0.5 mAs input, display 0.74 mAs; at 20 mAs input, display 21.72 mAs; at 80 mAs input, display 86.28 mAs
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Light field – X-ray field alignment
Align collimator light field with X-ray field and calibrate line laser position for Proteus XR/a collimator
Ranked causes
- 1. Light field and X-ray field misalignment due to shipment or operational variation (occasional)
- 2. Line laser misaligned from light field center by more than 3mm (occasional)
Power off system
- Test point
- Main power switch
- Expected
- System powered off
8 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: New tube operation verification radiation
Verify X-ray tube function and clear error logs after tube replacement with series of exposures at varying parameters.
Ranked causes
- 1. Tube replacement requires functional verification and error log clearance before clinical use (common)
- 2. Tube replacement documentation and shipment requirements (unknown)
Execute test exposures at low kV and mA settings
- Test point
- Direct workstation exposure controls
- Expected
- Proper X-ray generation at low parameters, tube responds normally
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Tube configuration database selection
Load appropriate JEDI Service Software database file matching installed X-ray tube type and power rating. Includes verification of console display parameters after load.
Ranked causes
- 1. Different tube types and power ratings require matching database to ensure proper operation (common)
- 2. Database mismatch prevents proper system operation and damages components (common)
Identify installed X-ray tube type and power rating
- Test point
- Equipment nameplate or maintenance records
- Expected
- Tube designation (RAD 12/14/21 or MX 100) and power (32/50/65/80 kW) clearly identified
10 more checks with pass and fail actions, 12 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Workstation parameters configuration
Configure JEDI Service Software workstation programming for tube selection, exposure mode settings, AEC chamber selection, and film-screen combinations via TECH menu.
Ranked causes
- 1. Each installation requires workstation-specific parameter configuration via JEDI Service Software (unknown)
- 2. JEDI software preprogrammed values may not match room actual parameters requiring verification (occasional)
Access TECH setup menu from JEDI Service Software
- Test point
- Setup mode menu 3 display
- Expected
- TECH menu accessible via F1 function key from menu 3
7 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Carriage motion not smooth or exhibits side play (SG100) caution
SG100 carriage does not move freely or smoothly over travel range; exhibits excessive side play
Ranked causes
- 1. Counterweights missing, misaligned, or trim covers improperly installed (common)
- 2. Bearing wear or misalignment at bucky pivot (occasional)
Inspect counterweight installation and positioning
- Test point
- Counterweights and trim covers
- Expected
- All counterweights correctly positioned; trim covers fully installed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Current under estimated fork range (low)
- DS100 and/or DS200 LEDs off
- Error code 0301-0304/0309-0314/0319/0323 reported
- Error code 0505 reported
- Error code 0505 reported during inverter short circuit test
- Error codes 0301/0302/0303/0304/0309/0310/0311/0312/0313/0314/0319/0323 reported
- Error codes 0501/0503/0504 reported
- Filament current too high on inverter 1 for Boost
- Filament current too high on inverter 1 for Heat
- Filament current too high on inverter 1 for Pre-Heat
- FPGA problem; Safety hit with unknown reason
- Inverter max ILR current detected
- No kV Feedback on both anode and cathode
- PRD error (heater)
- Re-starting safety triggered
- Rotation current overload
- Rotation function error codes 0103H–0106H
- SG100 bucky angulation lock does not release
- SG100 bucky rotation lock does not engage or slips
- Software or database corrupt
- Vertical locks do not release smoothly
- AEC diagnostic test failed - error code not 1111111
- AEC diagnostics result other combinations
- AEC gain test (x10) failed
- AEC parameters recording
- AEC return measurement test failed
- APR floppy disk recovery from floppy to floppy console
- APR recovery from floppy disk console to USB disk console
- APR recovery from USB disk console to USB disk console
- Boost too long on inverter 1
- Bucky angulation lock does not hold
- Bucky angulation lock does not hold position or binds
- Bucky angulation locks do not engage when releasing lock handle
- Bucky angulation locks remain engaged when operating lock handle
- Bucky angulation microswitches not actuating
- Bucky grid does not oscillate during exposure
- Bucky grid exposure shows shading or objectionable grid lines
- Bucky grid not oscillating or oscillation not audible
- Bucky locks fail to engage when equipment switched on
- Bucky rotation lock does not engage
- Bucky rotation lock does not engage or binds
- Bucky rotation lock does not release
- Bucky rotation lock not engaging properly
- Bucky rotation microswitches not actuating
- Bucky vertical locks remain engaged when operating lock handle
- Bucky vertical motion not smooth or operating force excessive
- Calibration: AEC KF format parameters
- Calibration: AEC KF parameter (cassette sensitivity)
- Calibration: AEC KF parameter (format correction)
- Calibration: AEC KG grid correction parameters
- Calibration: AEC KTIME parameter (time constant)
- Calibration: AEC KVEF parameter (kV efficiency factor)
- Calibration: AEC KVEF parameter acquisition
- Calibration: Anti-collision buzzer disable
- Calibration: Anti-collision switch adjustment
- Calibration: Anti-collision switches adjustment
- Calibration: Automatic Exposure Control (AEC)
- Calibration: Checksum Validation
- Calibration: Collimator Bulb Intensity
- Calibration: Collimator bulb light intensity equality
- Calibration: CPU memory backup
- Calibration: CPU memory restore
- Calibration: DOSE Calibration
- Calibration: DOSE calibration setup
- Calibration: Invasive mA measurement setup
- Calibration: JEDI 38-2 AEC reproducibility
- Calibration: KF parameter entry
- Calibration: KG parameter entry
- Calibration: KTIME parameter entry
- Calibration: KVEF parameter entry
- Calibration: Lateral cell check
- Calibration: Lateral film size
- Calibration: Line laser adjustment
- Calibration: Longitudinal film size
- Calibration: mA measurement procedure
- Calibration: Manual Collimator Bulb Voltage
- Calibration: Manual collimator lamp voltage
- Calibration: Microswitch S4 and S5 Activator Cam adjustment for Auto Exposure positions
- Calibration: Microswitch S4 and S5 positioning
- Calibration: SG100 Wall Stand counterweight
- Calibration: SG120 Wall Stand counterweight
- Calibration: SG120 Wall Stand Vertical Locks
- Calibration: SID (Source-to-Image Distance)
- Calibration: SID (Source-to-Image Distance) measurement
- Calibration: Software restoration for JEDI 7-2
- Calibration: Table height limitation
- Calibration: Touch screen (PN 2259976 with floppy disk)
- Calibration: Touch screen (PN 5462233 with USB)
- Calibration: Touch screen positioning
- Calibration: Tube Configuration
- Calibration: Tube data tracking reset
- Calibration: Wall bucky lateral film size
- Calibration: Wall bucky longitudinal film size
- Calibration: Wall Stand counterweight
- Calibration: X-ray Tube Warm Up
- Carriage motion not smooth or exhibits side play (SG120)
- Cassette size detection verification
- Cassette size not displayed correctly on OTS console and collimator
- Cassette tray pull-out or pull-in movement is not smooth
- Checksum validation
- Collimator bulb replacement
- Current under estimated fork/range (high)
- Database not correct (rotation board)
- DS1 neon not lit
- DS101/DS102/DS201/DS202 LEDs off during normal operation
- DS300 neon not lit
- Error code 0313 reported
- Error code 0313 reported during inverter short circuit test
- Error code 0320 reported
- Error code 0501/0503/0504 reported
- Error message from rotor firmware
- F0 main frequency problem
- Filament Database not correct (tubes 1-4)
- Filament selection error
- Float operation not functioning
- Heater command error
- Heater inverter permanent overcurrent
- Heater Inverter permanent short circuit
- Inverter short circuit diagnostic procedure
- kV detected during kV diagnostics
- kV did not reach 75% after 20 ms
- kV MAX detected
- kV unbalanced detected
- Laser light alignment verification
- Lateral cell response check
- Measured current while inverter OFF
- Microswitch S1 not actuated at vertical position
- Microswitch S1 not actuated at vertical position (SG100)
- Microswitch S1 not actuated at vertical position (SG120)
- Microswitch S2-1 not actuated at 0 degrees rotation
- Microswitch S2-1 not actuated at 0° rotation (SG100)
- Microswitch S2-2 not actuated at 180 degrees rotation
- Microswitch S2-2 not actuated at 180° rotation (SG100)
- Microswitch S3 not actuated at horizontal position
- Microswitch S3 not actuated at horizontal position (SG100)
- Microswitch S3 not actuated at horizontal position (SG120)
- Microswitch S4 not actuated at near vertical SID point
- Microswitch S4 not actuated at near vertical SID point (SG100)
- Microswitch S5 not actuated at far vertical SID point
- Microswitch S5 not actuated at far vertical SID point (SG100)
- No load HV diagnostic procedure
- One of DS101/DS102/DS201/DS202 LEDs off while no error reported
- PRD error (rotation)
- Record AEC parameters mAs0 and table values
- Replacement note: Rotation Capacitor Set when changing database files
- Rotation command error
- Rotation phase error
- SG100 carriage motion not smooth or has excessive side play
- Software download to generator
- Software version verification
- Spit Max error
- Spit retry failed
- System console replacement
- System console replacement - warning labels verification
- Table does not move with Up/Down pedal
- Table height does not meet specification
- Table height out of specification
- Table up/down floats when pedal not depressed
- Tomo-link system installation and calibration
- Tube replacement software procedure
- Tube spit detection - kV+ and kV- dropped
- Unknown heater error
- Unknown Rotation error
- Vertical lock release delayed (SG100)
- Vertical locks do not engage or remain locked (SG120)
- Vertical SID point microswitches not actuating
- Wall bucky does not move smoothly or requires excessive force
167 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0001000 | AEC board communication error | AEC diagnostic test execution | error |
0101H | No CAN message received within 5 secs | Communication failure on CAN bus for rotation function | critical |
0102H | Database not correct | Rotation database validation failure | critical |
0103H | Rotation function error | Detected during rotation diagnostic in high speed mode only | error |
0104H | Rotation function error | Detected during rotation diagnostic in high speed mode only | error |
0105H | Rotation function error | Detected during rotation diagnostic in high speed mode only | error |
0106H | Rotation function error | Detected during rotation diagnostic in high speed mode only | error |
0107H | Rotation Inverter permanent overcurrent | Rotation inverter detects sustained overcurrent condition | critical |
0108H | Rotation command error | Invalid or malformed rotation command received | critical |
0109H | MAINS_DROP has failed | Main power supply monitoring has failed | critical |
0110H | PRD error | Power-on diagnostics failure in rotation circuit | critical |
0111H | F0 main frequency problem | Main power frequency is incorrect or unstable | critical |
0149H | Unknown Rotation error | Unclassified rotation subsystem error detected | critical |
0151H | CAN Domain command number error | CAN Domain command sequence number is incorrect | info |
0152H | CAN Domain request with no transfer init | CAN Domain request received without transfer initialization | info |
0153H | CAN Domain Toggle bit error | CAN Domain toggle bit is incorrect | info |
0154H | CAN Domain : less than 2 data to download | CAN Domain download has insufficient data | info |
0155H | CAN Domain Abort received & applied | CAN Domain transfer was aborted | info |
0156H | Bad index in config upload | Configuration upload contains invalid index value | info |
0157H | Tube switch while rotor not off | Tube selection changed while rotor is still spinning | info |
0158H | Acceleration cmd while no tube selected | Rotor acceleration command issued without tube selection | info |
0159H | Acceleration cmd while database not OK | Rotor acceleration command issued before database is valid | info |
0160H | Database download while rotor speeding | Configuration database download started while rotor is rotating | info |
0161H | Acceleration command not OK | Rotor acceleration command validation has failed | info |
0162H | Rotor acceleration while in error | Rotor acceleration command attempted while system in error state | info |
0163H | No CAN message received within 4 secs | CAN bus timeout during rotation operation | warning |
0164H | Rotation Inverter overcurrent (< 3 times) | Inverter current exceeds threshold less than 3 times | warning |
0180H | Rotor board communication problem | Communication failure with rotor control board | critical |
0181H | Rotor board has reset | Rotor control board reset unexpectedly | critical |
0199H | Unknown rotation warning | Unclassified rotation system warning detected | warning |
0201H | No CAN message received within 5 secs | Communication failure on CAN bus for heater function | critical |
0203H | Heater inverter permanent overcurrent | Heater inverter detects sustained overcurrent condition | critical |
0204H | Filament permanent open circuit | Filament circuit is open or disconnected | critical |
0205H | Heater Inverter permanent short circuit | Heater inverter detects short circuit condition | critical |
0206H | Filament too high for Pre-heat | Filament current exceeds limit during pre-heat phase | critical |
0207H | Filament too high for Boost | Filament current exceeds limit during boost phase | critical |
0208H | Filament too high for Heat | Filament current exceeds limit during heat phase | critical |
0209H | Heater command error | Invalid or malformed heater command received | critical |
0210H | Current under estimated fork | Measured filament current is below expected range | critical |
0211H | Current over estimated fork | Measured filament current is above expected range | critical |
0212H | MAINS_DROP detected | Main power supply has dropped during heater operation | critical |
0213H | PRD error | Power-on diagnostics failure in heater circuit | critical |
0214H | Stay too long in Boost | Filament remains in boost phase beyond allowed time | critical |
0215H | Filament selection error | Invalid filament selection or database mismatch | critical |
0216H | Measured current while inverter OFF | Current detected when heater inverter is off | critical |
0221H | Filament Database not correct | Filament configuration database validation failure | critical |
0248H | Unknown Heater error | Unclassified heater subsystem error detected | critical |
0251H | Received command is not OK | Invalid heater command received | warning |
0252H | Heater command not OK | Heater command validation failure | warning |
0253H | No CAN message received within 4 secs | CAN bus timeout during heater operation | warning |
0254H | Heater inverter overcurrent (inverter1) (<3 times) | Inverter1 current exceeds threshold less than 3 times | warning |
0255H | Filament open circuit (inverter1) (<3 times) | Filament continuity loss in inverter1 less than 3 times detected | warning |
0256H | Heater Inverter short circuit (inverter1) (<3times) | Inverter1 short circuit detected less than 3 times | warning |
0257H | Tube switch while filaments not OFF | Tube switching commanded while filament heating still active | warning |
0258H | CAN Domain command number error | CAN message domain command counter mismatch or corruption | warning |
0259H | CAN Domain request with no transfer init | CAN data transfer requested without proper initialization | warning |
0260H | CAN Domain Toggle bit error | CAN toggle bit sequence mismatch detected | warning |
0261H | CAN Domain : less than 2 data to download | Download initiated with insufficient data segments | warning |
0262H | CAN Domain Abort received & applied | System received and executed abort command on CAN bus | warning |
0263H | Database download while heater not cut | Attempted database update while heater supply still active | warning |
0280H | Heater board communication problem | Communication failure with heater control board | critical |
0281H | Heater board has reset | Heater control board reset unexpectedly | critical |
0299H | Unknown Heater warning | Unclassified heater system warning detected | warning |
0301 | kV control board error during inverter diagnostics | Detected during inverter short circuit or HV diagnostics | error |
0301/0302/0303/0304/0309/0310/0311/0312/0313/0314/0319/0323 | Multiple kV control or inverter faults during short circuit diagnostic | Short circuit diagnostic test sequence | error |
0301H | Tube spit (kV+ and kV- dropped) | High voltage regulation failure during tube operation | error |
0302 | kV control board error during inverter diagnostics | Detected during inverter short circuit or HV diagnostics | error |
0302H | Tube spit (kV+ has dropped) | Positive high voltage has dropped unexpectedly | error |
0303 | kV control board error during inverter diagnostics | Detected during inverter short circuit or HV diagnostics | error |
0303H | Tube spit (kV- has dropped) | Negative high voltage has dropped unexpectedly | error |
0304 | kV control board error during inverter diagnostics | Detected during inverter short circuit or HV diagnostics | error |
0304H | Tube spit (kV regulation error) | High voltage regulation failure during tube operation | error |
0305H | FPGA problem (restarting safety signal) | Field programmable gate array safety signal restarted | error |
0306H | No kV Feedback on anode | High voltage anode feedback signal is missing or invalid | error |
0307H | No kV Feedback on cathode | High voltage cathode feedback signal is missing or invalid | error |
0308H | No kV Feedback on anode and cathode | Both anode and cathode feedback signals are missing | error |
0309 | kV control board error during inverter diagnostics | Detected during inverter short circuit or HV diagnostics | error |
0309H | kV detected during kV diag | High voltage detected when it should be absent during diagnostics | error |
0310 | kV control board error during inverter diagnostics | Detected during inverter short circuit or HV diagnostics | error |
0310H | kV max detected | High voltage exceeds maximum safe limit | error |
0311 | No Ilp (parallel inductor current) detected | During inverter short circuit diagnostic | error |
0311H | ILP current not OK | ILP (left phase) current is out of specification | error |
0312 | No Ilr (series inductor current) detected | During inverter short circuit diagnostic | error |
0312H | ILR current not OK | ILR (right phase) current is out of specification | error |
0313 | kV control board internal error | During inverter short circuit diagnostic | error |
0313H | ILR max current detected | ILR (right phase) current exceeds maximum limit | error |
0314 | Ilr current resonant frequency lower than expected | During inverter short circuit diagnostic | error |
0314H | ILR current timeout | ILR current measurement took too long to stabilize | error |
0316H | Spit Max error | Tube spit magnitude exceeded maximum threshold | error |
0317H | Spit Ratio error | Ratio between spit events indicates abnormal pattern | error |
0318H | kV did not reach 75% after 20ms | High voltage ramp too slow during startup | error |
0319 | kV control board error during inverter diagnostics | Detected during inverter short circuit or HV diagnostics | error |
0319H | kV unbalanced detected | Positive and negative high voltages are not balanced | error |
0320 | Persisting kV control or inverter fault | Short circuit or no load diagnostic test continuation after initial fault | error |
0320 | kV control board error that persists after diagnostics | During inverter diagnostics when problem remains after cable checks | error |
0320H | FPGA problem (safety signal) | Field programmable gate array safety signal failure | error |
0321H | Spit retry failed | Attempt to recover from tube spit was unsuccessful | error |
0321H | kV conversion error | High voltage analog to digital conversion has failed | critical |
0322H | kV ref ADC / DAC failed | Reference or analog conversion circuit has failed | critical |
0323 | Both Ilr and Ilp currents not detected | During inverter short circuit diagnostic | error |
0323H | ILP and ILR current not OK | Both ILP and ILR currents are out of specification | error |
0324H | SPITS (data=1:Anode,2:cathode,4:Both,8:kV regul) | Tube spit event with specified component affected | error |
0401H | No mA feedback | Tube current (mA) feedback signal is missing or invalid | error |
0402H | mA scale error | mA measurement scale or calibration is incorrect | error |
0403H | mA accuracy exceeded 5% | mA measurement error exceeds 5% tolerance | error |
0501 | kV control or inverter fault | During inverter diagnostics | error |
0501/0503/0504 | kV control or inverter fault during diagnostics | Short circuit or no load diagnostic test execution | error |
0501H | DC bus out of range | DC bus voltage is outside safe operating limits | critical |
0503 | kV control or inverter fault | During inverter diagnostics | error |
0503H | Inverter Gate Power Supply error | Gate drive power supply voltage is incorrect | critical |
0504 | kV control or inverter fault | During inverter diagnostics | error |
0504H | Inverter Gate Power Supply failed | Gate drive power supply for inverter has failed | error |
0505 | Isolation fault between inverter components and ground | During inverter diagnostics | error |
0505H | Mains power supply has dropped during exposure | Main power supply failure occurred during exposure | critical |
0506H | DC bus 1 phase precharge error | DC bus precharge failed for single phase mode | critical |
0507H | DC bus 1 phase discharge error | DC bus discharge failed for single phase mode | critical |
0510H | AEC Cell HV supply error | High voltage supply for AEC cell is faulty | critical |
0549H | Unknown LVPS error | Unclassified low voltage power supply error detected | critical |
0550H | No more warn +160V too low or too high | Recovery from +160V voltage warning condition | info |
0551H | Detected +160V too high | +160V supply exceeded high threshold | warning |
0553H | Detected +160V too high | +160V supply voltage is above maximum limit | critical |
0555H | Detected +160V too low | +160V supply dropped below low threshold | warning |
0557H | Detected +160V too low | +160V supply voltage is below minimum limit | critical |
0560H | No more warn +15V too low or too high | Recovery from +15V voltage warning condition | info |
0561H | Detected +15V too high | +15V supply exceeded high threshold | warning |
0563H | Detected +15V too high | +15V supply voltage is above maximum limit | critical |
0565H | Detected +15V too low | +15V supply dropped below low threshold | warning |
0567H | Detected +15V too low | +15V supply voltage is below minimum limit | critical |
0570H | No more warn -15V too low or too high | Recovery from -15V voltage warning condition | info |
0571H | Detected -15V too strong | -15V supply magnitude exceeded high threshold | warning |
0573H | Detected -15V too strong | -15V supply voltage is too high (more negative) | critical |
0575H | Detected -15V too weak | -15V supply magnitude dropped below low threshold | warning |
0577H | Detected -15V too weak | -15V supply voltage is too low (less negative) | critical |
0599H | Unknown LVPS warning | Unclassified low voltage power supply warning detected | warning |
0601H | RTL error | Real-time logic circuit has failed | critical |
0602H | External CAN bus off | External CAN bus communication is not active | critical |
0603H | Debug screen com error | Communication with debug display has failed | critical |
0604H | Database download error | Configuration database download or transfer has failed | critical |
0605H | TAV communication error | Communication with TAV system has failed | critical |
0606H | MPC/Madrid communication error | Communication with MPC/Madrid system has failed | critical |
0701H | Saved RAM checksum pb | Saved RAM checksum verification has failed | critical |
0702H | Software problem | General software or firmware fault detected | critical |
0703H | Watchdog reset has just occurred | Watchdog timer expired causing system reset | error |
0704H | Rotor/Heater hold too long | Rotor or heater command hold state exceeded timeout | critical |
0801H | Exposure backup mAs exceeded | Exposure terminated due to backup mAs limit reached | error |
0802H | Exposure backup time exceeded | Exposure terminated due to backup time limit reached | error |
0803H | Exp cmd while gene not ready | Exposure command received but generator not in ready state | error |
0804H | Tank Thermal Error | Tank temperature has exceeded safe operating limit | warning |
0902H | tube Fan supply error | Tube cooling fan power supply has failed | critical |
0903H | Tube exceeded 70degC | Tube temperature has exceeded 70 degrees Celsius | warning |
100-0603H | Debug screen com. error | Debug screen communication failure | info |
100-0604H | Database download error | Problem detected during database download | error |
100-0605H | TAV communication error | Communication problem between I/F and service laptop | error |
100-0606H | MPC/Madrid Communication error | No reply from the console | error |
100-1301H | AEC communication error | kV Control detects communication problem with AEC board | error |
110-0804H | Tank Thermal Error | HV tank temperature reached 60°C | warning |
110-0903H | 70 degree C thermal safety error | 70° loop detected open | error |
1111111 | AEC diagnostic test passed | AEC diagnostic test completion | info |
120-1500H | Tomo brightness not good | Brightness signal out of range (+/-20%) due to incorrect technique | warning |
120-1501H | Exposure switch released | Handswitch released before end of exposure | warning |
120-1502H | AEC CUT off error | AEC did not cut exposure as expected | warning |
1301H | AEC communication error | Communication with AEC system has failed | critical |
1401H | Saved RAM power supply limit reached | RAM power budget computation exceeds threshold over ~7 year period | info |
1402H | Internal CAN bus off | Internal CAN bus communication is not active | critical |
1403H | Connectic Fault | Connector or connection fault detected | critical |
1404H | FPGA configuration problem | Field programmable gate array configuration has failed | critical |
1405H | Tank sensor problem | Tank temperature or condition sensor has failed | critical |
1407H | mAs counter error | Exposure mAs counter is not operating correctly | error |
1408H | AEC counter error | AEC counter is not operating correctly | error |
1409H | mAs meter saturated | mA meter has reached maximum measurable value | error |
1410H | FPGA Locked | Field programmable gate array is in locked or hung state | error |
1411H | Time counter error | Exposure time counter is not operating correctly | error |
1420H | Tomo exposure cut by tomograph instead generator | Tomography exposure was terminated by tomograph system instead of generator | error |
1500H | Tomo brightness error | Tomography system brightness control error during manipulation | warning |
1501H | Exposure switch released during exposure | Exposure switch was released before exposure completed | warning |
1502H | AEC cut off error | AEC system terminated exposure prematurely | warning |
30-0301H | Tube spit (kV+ and kV- dropped) | Simultaneous voltage drop detected on both anode and cathode HV lines | critical |
30-0302H | Tube spit (kV+ has dropped) | Voltage drop detected on anode (positive) HV line only | critical |
30-0303H | Tube spit (kV- has dropped) | Voltage drop detected on cathode (negative) HV line only | critical |
30-0304H | kV regulation error | Slow voltage drop detected indicating smooth or gradual tube spit condition | error |
30-0305H | Re-starting safety (unknown reason) | Safety interlock activated with no corresponding detected error on primary error inputs | warning |
30-0324H | Cathode or anode spit detected | Data parameter indicates spit type: 1=cathode, 2=anode, 4=both, 8=kV regulation | error |
40-0101H | No CAN message received within 5 seconds | Rotation board communication loss from kV control main software | error |
40-0102H | Rotation board database incorrect | Firmware detects wrong data in database received from kV control board at power-up | error |
40-0103H | Rotation motor current overload | Main or auxiliary rotation current exceeds maximum tube rating | error |
40-0104H | Rotation current open load | No current flowing to rotation motor | error |
40-0105H | Rotation phases unbalanced | Amplitude difference between main and auxiliary current too large | error |
40-0106H | Rotation phase shift error | Anode stator current lacks correct phase relationship | error |
40-0107H | Rotation inverter permanent overcurrent | Overcurrent detected with 3 unsuccessful restart attempts within single rotation state | error |
40-0108H | Rotation command error (debug) | Erroneous rotation command received by board | warning |
40-0109H | Mains drop detected (rotation) | Firmware detected mains_drop signal activation | error |
40-0110H | PRD error on rotation board | Firmware checksum, RAM test, or EPLD access failure at power-up or reset | error |
40-0110H | Rotation board PRD self-test failure | Firmware checksum, RAM, or EPLD test failed at power-up or reset | error |
40-0111H | F0 main frequency problem | EPLD did not apply inverter start command | error |
40-0149H | Unknown rotation error | Main software received unrecognized error from rotation board | warning |
40-111H | Rotor firmware error | Hardware/software conflict detected in rotor | error |
50-0201H | No CAN message received on heater board within 5 seconds | Heater board communication loss from kV control main software | error |
50-0203H | Heater inverter permanent overcurrent | 3 restarts failed within 100 ms | error |
50-0204H | Filament permanent open circuit | Open circuit detected with 3 unsuccessful restart attempts within 100 ms | error |
50-0205H | Heater inverter permanent short circuit | 3 restarts failed within 100 ms | error |
50-0206H | Filament current too high during pre-heat | RMS current exceeds database maximum for tube | warning |
50-0207H | Filament current too high during boost | RMS current exceeds database maximum for tube | warning |
50-0208H | Filament current too high during heat | RMS current exceeds database maximum for tube | warning |
50-0209H | Heater command error (debug) | Erroneous heater command received | warning |
50-0210H | Filament current over-estimated (too high) | RMS current measurement too high in measurement range | warning |
50-0211H | Filament current under-estimated (too low) | RMS current measurement too low in measurement range | warning |
50-0212H | Mains drop detected on heater board | Firmware detects mains_drop signal activation | error |
50-0213H | PRD error on heater board | Firmware checksum, RAM test, or EPLD access failure at power-up or reset | error |
50-0213H | Heater board PRD self-test failure | Firmware checksum, RAM, or EPLD test failed; also -15V missing | error |
50-0214H | Boost phase timeout | Boost command duration exceeds 400 ms | warning |
50-0214H | Boost too long on inverter 1 | Boost command stayed longer than 400 ms | warning |
50-0215H | Filament selection relay error | Relay position does not match filament selection | error |
50-0216H | Current measured with inverter OFF | Inverter current detected when inverter not commanded | warning |
50-0221H | Filament database not correct for tube 1 | Firmware detects erroneous database from kV control at power-up | error |
50-0221H/0222H/0223H/0224H | Filament database error for tube 1, 2, 3, or 4 | Erroneous database parameters received at power-up | error |
50-0222H | Filament database not correct for tube 2 | Firmware detects erroneous database from kV control at power-up | error |
50-0223H | Filament database not correct for tube 3 | Firmware detects erroneous database from kV control at power-up | error |
50-0224H | Filament database not correct for tube 4 | Firmware detects erroneous database from kV control at power-up | error |
50-0248H | Unknown heater error | Main software receives error from heater board with no associated error code | warning |
50-0248H/0249H | Unknown heater error | Unrecognized error received from heater board | warning |
50-0249H | Unknown heater error | Main software receives error from heater board with no associated error code | warning |
60-0306H | No kV feedback on anode side | kV measured <12 kV 0.5 ms after start of exposure on anode side only | error |
60-0307H | No kV feedback on cathode side | kV measured <12 kV 0.5 ms after start of exposure on cathode side only | error |
60-0308H | No kV feedback on both anode and cathode | kV <12 kV at 0.5 ms after exposure on both sides | error |
60-0309H | kV detected during diagnostics | kV measured during inverter diagnostics when should be off | warning |
60-0310H | kV MAX limit exceeded | kV reached 160 kV during exposure | error |
60-0311H | ILP (parallel resonant) current not OK | Current in parallel LC circuit did not rise at exposure start | error |
60-0312H | ILR (serial resonant) current not OK | Current in serial LC circuit did not rise at exposure start | error |
60-0313H | Inverter maximum ILR current detected | Hardware detects maximum current in serial resonant circuit | error |
60-0314H | ILR current timeout | Resonant frequency lower than expected | error |
60-0316H | Spit max error | Number of tube spits during exposure reaches limit | error |
60-0316H | Tube spit count limit exceeded | Number of tube spits reached maximum during exposure | error |
60-0318H | kV did not reach 75% after 20 ms | No HV ON signal generated for exposure time count-up | error |
60-0318H | kV ramp too slow | kV did not reach 75% after 20 ms from exposure start | error |
60-0319H | kV unbalanced detected | More than 12 kV difference between kV+ and kV- | error |
60-0319H | kV imbalance between positive and negative | Difference >12 kV between kV+ and kV- | error |
60-0320H | FPGA safety interlock triggered (unknown cause) | Safety hit detected with no error signal present | error |
60-0321H | Spit recovery retry failed | Retry mechanism unable to clear spit condition | error |
60-0323H | Both ILP and ILR currents absent | No inverter current measured at exposure start | error |
60-0401H | No mA feedback | No mA detected 20 ms after exposure start | error |
60-0402H | mA scale or calibration error | mA too low or too high vs. demand 20 ms after exposure start | error |
60-0403H | mA accuracy exceeded 5% | Measured mA every 50 msec exceeded 5% of mA demand | warning |
60-0504H | Inverter Gate Power Supply error | Gate supply voltage dropped below level required to drive IGBTs properly | error |
60-0801H | Exposure backup mAs exceeded | Exposure command lasted so long that maximum mAs allowed has been reached | error |
60-0802H | Exposure backup time exceeded | Exposure command lasted longer than duration loaded by system (Backup time + 5%) | error |
60-0803H | Exp cmd while gene not ready | Generator received exposure command while not in ready state | error |
60-1407H | mAs counter error | Error found in verifying counter normal operation | error |
60-1408H | AEC counter error | No AEC feedback in AEC station exposure | error |
60-1409H | mAs meter saturated | mAs counter shows unrealistic value at exposure start check | error |
60-1410H | FPGA locked | FPGA detected error and blocked exposure after command received | error |
60-1411H | Time counter error | Error found in verifying counter normal operation | error |
60-1420H | Tomo exposure cut by Tomography | Exposure cut during tomography operation | warning |
70-0501H | DC bus out of range (<400 or >850) | DC bus voltage outside acceptable limits | critical |
70-0503H | Inverter Gate Power Supply error (checked at prep) | Gate supply voltage check fails during preparation | error |
70-0505H | Mains power supply has dropped During exposure | Mains power drops during exposure | error |
70-0510H | AEC cell HV supply error | Power supply 230Vdc for AEC failed | error |
70-0549H | Unknown LVPS error | Main software received LVPS error with no associated error code | error |
70-0553H | Detected +160 V too high | +160 V voltage higher than 225 V at heater board | error |
70-0557H | Detected +160 V too low | +160 V voltage lower than 110 V at heater board | error |
70-0563H | Detected +15 V too high | +15 V voltage higher than 17.3 V at heater board | error |
70-0567H | Detected +15 V too low | +15 V voltage lower than 13 V at heater board | error |
70-0573H | Detected -15 V too high | -15 V voltage lower than -17.3 V at heater board | error |
70-0577H | Detected -15 V too low | -15 V voltage higher than -13 V at heater board | error |
80-0180H | Rotation board communication problem | kV control board communication watch dog timeout with rotation board | error |
80-0181H | Rotation board has reset | kV control detected rotation board reset; reloading database | warning |
80-0280H | Heater board communication problem | kV control board communication watch dog timeout with heater board | error |
80-0321H | KV conversion error | kV feedback measured while no kV is being generated | error |
80-0322H | kV ref ADC / DAC failed | kV control DAC and ADC coherency test failed | error |
80-0601H | RTL error (Real Time Lines show wrong state) | Real Time Lines test detects wrong state during standby | error |
80-0601H | RTL error | Real Time Lines show wrong state in regular standby testing | error |
80-0602H | External CAN bus off | CAN bus communication failure external to generator | error |
80-0902H | Tube Fan supply error | Wrong voltage applied to tube fan (not 115 V) | error |
80-1402H | Internal CAN bus off | CAN device on kV control board detected abnormal level on 2 lines | error |
80-1403H | Connection Fault | One of the flat cable connectors is not connected in Generator | error |
80-1404H | FPGA configuration problem | Detected during power up; main software cannot load FPGA | critical |
80-1405H | Tank temperature sensor problem | HV tank temperature value less than 10°C | warning |
90-0701H | NV RAM checksum problem | Generator kV control board detects NV RAM corruption after power up | critical |
90-0702H | Software problem | Software or database error detected | error |
90-0703H | Watchdog reset has just occurred | CPU of kV control board reset by watchdog timer | error |
90-0704H | Rotation/Heater hold too long | Preparation command maintained longer than 3 minutes | warning |
90-0705H | Configuration error | Software/hardware conflict detected | error |
Cannot open file | Backup or restore operation cannot access specified file on floppy disk | Attempt to backup CPU memory or restore from floppy disk fails at file access | error |
Checksum error (unspecified code) | Stored and computed parameter checksums do not match | Parameter changed or RAM corruption detected on power-up or after settings changes | critical |
Error communication | Data transmission error occurred during backup or restore transfer | Checksum or parity error detected during file transfer between laptop and generator | error |
Error time out | Communication timeout between service laptop and generator during backup or restore | Transfer does not complete within timeout period during memory backup or restore | error |
File contains data. Delete old (y,n) | Target file already exists on floppy disk during backup operation | Backup attempts to create file that already exists with data | warning |
Z00 | RAM D1 error (fatal) | RAM D1 failure detected during self-test | critical |
Z00 | Collimator UIF board error | User Interface board failure detected or not responding | critical |
Z01 | EEPROM D1 error (fatal) | EEPROM D1 failure detected during self-test | critical |
Z01 | Collimator Control Board error | Main collimator control board failure or not responding | critical |
Z02 | Collimator Control board error | Collimator control board detects communication or logic fault | error |
Z02 | CAN error (bus error, error overrun) | CAN bus error frame detected or error counter overrun | error |
Z03 | Collimator UIF Error | User Interface board detected a communication or function error | error |
Z03 | CAN error (receiver overrun) | CAN receiver buffer overflow; messages arriving faster than processor handles | error |
Z04 | Collimator local communication error | Communication failure between collimator submodules over local bus | error |
Z05 | Calculation error - height | Height motor position calculation failure or out-of-range result | error |
Z06 | Collimator Blade out of limit | Blade position outside mechanical limit during operation | error |
Z06 | Calculation error - width | Width motor position calculation failure or out-of-range result | error |
Z10 | Collimator internal error | Collimator processor detected internal logic or state machine error | error |
Z10 | Stepping error - height leaves | Height stepper motor or blade positioning error | error |
Z11 | Stepping error - width leaves | Width stepper motor or blade positioning error | error |
Z11 | Collimator internal error | Collimator processor detected internal logic or state machine error | error |
Z14 | Collimator Homing Error | Collimator failed to find home position during initialization sequence | error |
Z14 | Limit switch - height leaves | Height motor movement detected when limit switch should prevent it | error |
Z15 | Limit switch - width leaves | Width motor movement detected when limit switch should prevent it | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI