GE Healthcare Proteus XR/a: troubleshooting and error codes

The ServiceForge knowledge base holds 227 documented symptoms for the GE Healthcare Proteus XR/a, with 383 ranked causes, 498 diagnostic checks with test points and expected values, and 305 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Generator does not reply to system high voltage

System cannot communicate with generator; kV control LEDs may show specific patterns, RESET, or HALT status, or no LEDs lit

Ranked causes

  1. 1. No power on Generator (common)
  2. 2. EMC filter failure (occasional)
  3. 3. AC/DC diode bridge failure (occasional)
  4. 4. LVPS board failure (occasional)
  5. 5. CAN cable problem (occasional)
  6. 6. kV control board failure (occasional)
  7. 7. Heater board failure (occasional)
  8. 8. Rotation board failure (occasional)
  9. 9. AEC board failure (occasional)
  10. 10. IF board failure (occasional)
First check

Check if kV control LEDs S0-S7 are lit successively

Test point
kV control board LEDs S0-S7
Expected
LEDs lit successively

10 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

kV regulation error high voltage

Slow speed safety circuit triggered by smooth kV spits or regulation problems

Ranked causes

  1. 1. Smooth HV tube spits (common)
  2. 2. kV control board regulation problem (common)
  3. 3. Too much line impedance (occasional)
  4. 4. Half of AC/DC capacitors open (occasional)
  5. 5. Inverter parallel inductor or filtering capacitors (occasional)
  6. 6. HV tank (occasional)

Error code 0311 reported during inverter short circuit test high voltage

No primary inductor current (Ilp) detected during short circuit diagnostic

Ranked causes

  1. 1. Parallel inductor disconnected or faulty (common)
  2. 2. Parallel inductor impedance incorrect (occasional)
  3. 3. Inverter capacitors broken (occasional)
  4. 4. Current transformer to gate_cmd board cable faulty (rare)
  5. 5. HV Tank to kV control cable faulty (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error code 0323 reported high voltage

Both Ilr and Ilp currents not detected during inverter diagnostics

Ranked causes

  1. 1. -15 V DC supply not present on kV control board (common)
  2. 2. Current transformer cable disconnected (common)
  3. 3. Inverter inductors disconnected (common)
  4. 4. Gate_cmd to HV Tank or HV Tank to kV control cable fault (occasional)
  5. 5. Inverter fault (rare)
First check

Verify -15 V supply LED on kV control board

Test point
DS1 LED on kV control board
Expected
LED lit

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error code 0323 reported during inverter short circuit test high voltage

Both primary and series inductor currents (Ilp and Ilr) not detected

Ranked causes

  1. 1. -15V supply not present on kV control board (common)
  2. 2. Current transformer to gate_cmd board cable faulty (occasional)
  3. 3. Inverter inductors faulty (occasional)
  4. 4. HV Tank and kV control cabling faulty (occasional)
  5. 5. Inverter faulty (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

3-phase CB1 breaker trips or AC fuses blown high voltage

Breaker trips or fuses blow in PDU when system powered up, indicating short circuit in generator power distribution

Ranked causes

  1. 1. IGBT in short circuit (common)
  2. 2. AC/DC or bridge rectifier in short circuit (common)
  3. 3. Bridge rectifier in short circuit (occasional)
  4. 4. EMC filter in short circuit (occasional)
First check

Disconnect DC bus cables between AC/DC and inverter

Test point
DC bus cables on AC/DC side
Expected
Cables disconnected without short circuit

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AEC diagnostics failed patient safety

Diagnostic procedure to verify AEC (Automatic Exposure Control) function and ion chamber circuit operation

Ranked causes

  1. 1. AEC board fault (common)
  2. 2. IF board communication error (occasional)
  3. 3. kV control board communication fault (occasional)
  4. 4. Incorrect AEC board wiring or ion chamber connection (occasional)
First check

Verify application mode operating without AEC

Test point
Operator console in non-AEC mode
Expected
System operates normally

9 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Cassette size display radiation

Calibrate cassette detection and display sizing for correct field-of-view reporting on collimator

Ranked causes

  1. 1. Cassette sensing system not calibrated (common)
  2. 2. Baseline voltage offset TP5/TP6 out of spec (common)
  3. 3. Longitudinal film size sensor signal out of range (common)
  4. 4. Lateral film size sensor signal out of range (common)
First check

Verify cassette sensing cable is connected

Test point
Cassette sensor connector
Expected
Cable fully seated and secured
Tools
Visual inspection

10 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: SG100 Wall Stand SID (Source-to-Image Distance) radiation

Calibrate SID display for SG100 Wall Stand with Bucky positioning and potentiometer adjustment

Ranked causes

  1. 1. Microswitch S4/S5 Activator Cam positions require adjustment (occasional)
  2. 2. System Console parameter 9 does not match actual Microswitch position distance (common)
  3. 3. Potentiometer P4 out of adjustment (SG100 SID) (common)
  4. 4. Potentiometer PA2 requires auxiliary adjustment (occasional)
First check

Position Bucky for SG100 calibration

Test point
Bucky rotation and angulation
Expected
Rotate to 0° and angulate to 90° (horizontal)

5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: SG120 Wall Stand SID (Source-to-Image Distance) radiation

Calibrate SID display for SG120 Wall Stand with Bucky positioning and potentiometer adjustment

Ranked causes

  1. 1. Microswitch S4/S5 Activator Cam positions require adjustment for SG120 (occasional)
  2. 2. System Console parameter 9 does not match actual SG120 Microswitch position distance (common)
  3. 3. Potentiometer P4 out of adjustment (SG120 SID) (common)
  4. 4. Potentiometer PA2 requires auxiliary adjustment for SG120 (occasional)
First check

Position Bucky for SG120 calibration

Test point
Bucky rotation and angulation
Expected
Rotate to 0° and angulate to 90° (horizontal)

5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error code 0311 reported high voltage

No Ilp (parallel inductor current) detected during inverter diagnostics

Ranked causes

  1. 1. Parallel inductor disconnected or impedance fault (common)
  2. 2. Parallel inductor impedance not zero (occasional)
  3. 3. Inverter capacitor set broken (occasional)
  4. 4. kV control board fault (rare)
First check

Verify parallel inductor connection

Test point
Parallel inductor connectors on inverter
Expected
Securely connected

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error code 0312 reported high voltage

No Ilr (series inductor current) detected during inverter diagnostics

Ranked causes

  1. 1. Series inductor disconnected or faulty (common)
  2. 2. Series current transformer impedance fault (occasional)
  3. 3. Inverter capacitor set broken (occasional)
  4. 4. kV control board fault (rare)
First check

Verify series inductor connections

Test point
Series inductor connectors on inverter
Expected
Securely connected

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error code 0312 reported during inverter short circuit test high voltage

No series inductor current (Ilr) detected during short circuit diagnostic

Ranked causes

  1. 1. Series inductors disconnected or faulty (common)
  2. 2. Series current transformer faulty (occasional)
  3. 3. Inverter capacitors broken (occasional)
  4. 4. HV Tank to kV control cable faulty (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error code 0314 reported high voltage

Ilr current resonant frequency is lower than expected

Ranked causes

  1. 1. Series inductors disconnected or faulty (common)
  2. 2. Inverter capacitor set broken (occasional)
  3. 3. Current transformer impedance fault (occasional)
  4. 4. kV control board fault (rare)
First check

Verify inductor connections

Test point
Inductor connectors on inverter
Expected
Securely connected

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Filament permanent open circuit high voltage

Open circuit detected in filament and 3 restarts attempted without success within 100 ms

Ranked causes

  1. 1. X-ray tube filament open (common)
  2. 2. Heater to HV tank cable disconnected (occasional)
  3. 3. Cathode HV cable or pin contacts failing (occasional)
  4. 4. Open circuit in filament transformer inside HV Tank (occasional)
First check

Measure filament winding resistance

Test point
Heater transformer primary or secondary
Expected
Resistance measured (not open)

Lamp cannot be switched on

Lamp fails to light when lamp button on front panel is pushed

Ranked causes

  1. 1. Lamp failure (common)
  2. 2. UIF board failure (occasional)
  3. 3. System lamp power failure (occasional)
  4. 4. UIF board or UIF panel failure (rare)
First check

Check cable connections

Test point
J4, J7, J8 on user interface board
Expected
Stable cable connection

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

MAINS_DROP detected (heater) caution

Firmware detected mains_drop signal activation and transmitted error to kV control

Ranked causes

  1. 1. Interference spikes (occasional)
  2. 2. Mains power drop (occasional)
  3. 3. Cable or connector contact in DC bus (occasional)
  4. 4. Heater board failure (occasional)
First check

Check DC bus voltage stability

Test point
DC bus between AC/DC and auxiliary unit
Expected
Stable, no dropout

MAINS_DROP detected (rotation) caution

Firmware detected mains_drop signal activation and transmitted error to kV control

Ranked causes

  1. 1. Interference spikes (occasional)
  2. 2. Mains power drop (occasional)
  3. 3. Cable or connector contact issue in DC bus (occasional)
  4. 4. Rotation board failure (occasional)
First check

Measure mains supply voltage stability

Test point
AC mains input
Expected
Stable, no dropout

Rotation current openload caution

Rotation board detected no current flowing to the motor

Ranked causes

  1. 1. Tube stator winding open circuit (common)
  2. 2. Incorrect wiring open (occasional)
  3. 3. No DC bus on Rotation board (occasional)
  4. 4. Rotation board failure (occasional)
First check

Measure rotation motor winding resistance

Test point
Tube stator winding terminals
Expected
Resistance measured (not open)

AEC diagnostics result 0001000 patient safety

AEC to IF board communication error detected

Ranked causes

  1. 1. AEC board fault (common)
  2. 2. IF board communication fault (occasional)
  3. 3. kV control board communication fault (occasional)
First check

Replace AEC board

Test point
AEC board connector
Expected
AEC board properly installed and connected

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bucky angulation motion not smooth or lock does not release (SG120) patient safety

SG120 bucky angulation lacks smooth motion or angulation lock fails to release

Ranked causes

  1. 1. Bearing wear or misalignment at bucky pivot point (occasional)
  2. 2. Angulation lock board electrical fault (occasional)
  3. 3. Lock hardware misalignment or wear (occasional)
First check

Press angulation lock pushbutton and attempt bucky tilt

Test point
Bucky angulation motion
Expected
Bucky tilts forward 20° and backward 90° freely and smoothly

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bucky rotation does not lock or slips during locked state (SG100) patient safety

Bucky rotation lock mechanism fails to prevent rotation or allows rotation when locked

Ranked causes

  1. 1. Lock hardware misalignment or wear (occasional)
  2. 2. Brake pad wear or improper adjustment of threaded rods (common)
  3. 3. Arm assembly wear requires replacement (rare)
First check

Release rotation lock lever and attempt bucky rotation

Test point
Bucky rotation range
Expected
Bucky rotates easily from 0° to 180° (upright to upside down)

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: SG100 wall stand SID display radiation

Source-to-image distance (SID) display on OTS console must match actual physical distance when Bucky moves vertically

Ranked causes

  1. 1. SID sensor output not calibrated (common)
  2. 2. Microswitch S4/S5 activator cam positions incorrect (common)
  3. 3. Potentiometer PA3 scaling incorrect (common)
First check

Verify SID display changes when moving OTS vertically

Test point
OTS console SID display
Expected
Display increments/decrements as OTS moves up and down
Tools
OTS console, measure tape on collimator

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: SG120 wall stand SID display radiation

Source-to-image distance (SID) display on OTS console for SG120 wall stand must match actual physical distance when Bucky moves vertically

Ranked causes

  1. 1. SID sensor output not calibrated for SG120 (common)
  2. 2. Microswitch S4/S5 activator cam positions incorrect on SG120 (common)
  3. 3. Potentiometer PA3 scaling incorrect for SG120 (common)
First check

Verify SID display changes when moving OTS vertically on SG120

Test point
OTS console SID display
Expected
Display increments/decrements as OTS moves
Tools
OTS console, measure tape

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: SID (Source-to-Image Distance) display radiation

Adjust potentiometers and verify SID display on OTS console matches actual table and OTS positions throughout travel range

Ranked causes

  1. 1. Potentiometer PA1 out of adjustment (table SID) (common)
  2. 2. Potentiometer PA3 out of adjustment (OTS vertical SID) (common)
  3. 3. Potentiometer P5 requires calibration adjustment (common)
First check

Ensure DIP Switch settings in red color

Test point
DIP Switch positions 1 and 2
Expected
On

7 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Workstation Parameters

Configure workstation selection, tube, exposure chamber, screen-film combination, and function key assignments using JEDI service software

Ranked causes

  1. 1. Workstation parameters not configured or incorrect (common)
  2. 2. Renard or Normal kV scale not set correctly (occasional)
  3. 3. Date and time not set (occasional)
First check

Verify workstation type selected

Test point
TECH menu display
Expected
Correct workstation type shown in reverse video

5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error code 0001000 reported during AEC diagnostic

AEC board communication error detected

Ranked causes

  1. 1. AEC board faulty (common)
  2. 2. IF board faulty (occasional)
  3. 3. kV control board faulty (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error code 0314 reported during inverter short circuit test high voltage

Series current (Ilr) resonant frequency lower than expected

Ranked causes

  1. 1. Series inductors disconnected or faulty (common)
  2. 2. Inverter capacitors broken (occasional)
  3. 3. Current transformer impedance incorrect (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

ILP and ILR current not OK high voltage

No inverter current measured at beginning of exposure

Ranked causes

  1. 1. Inverter LC resonant circuit failure (common)
  2. 2. Inverter board failure (occasional)
  3. 3. kV control board failure (occasional)
First check

Measure both ILP and ILR currents at exposure start

Test point
Inverter resonant circuit current outputs
Expected
Both currents present within 0.5 ms

ILP current not OK high voltage

Current in parallel resonant circuit of inverter did not rise at beginning of exposure

Ranked causes

  1. 1. Inverter LC resonant circuit failure (common)
  2. 2. Inverter board failure (occasional)
  3. 3. kV control failure (occasional)
First check

Measure LC resonant circuit current at exposure start

Test point
Inverter parallel resonant circuit
Expected
Current rises within 0.5 ms

ILR current not OK high voltage

Current in serial resonant circuit of inverter did not rise at beginning of exposure

Ranked causes

  1. 1. Inverter LC resonant circuit failure (common)
  2. 2. Inverter board failure (occasional)
  3. 3. kV control failure (occasional)
First check

Measure LC resonant circuit current at exposure start

Test point
Inverter serial resonant circuit
Expected
Current rises within 0.5 ms

ILR Current time out high voltage

Current resonant frequency is lower than expected

Ranked causes

  1. 1. kV control failure (common)
  2. 2. Inverter board failure (occasional)
  3. 3. Current measurement circuit failure (occasional)
First check

Measure resonant circuit frequency

Test point
ILR circuit oscillation frequency
Expected
Within specification

mA scale error radiation

mA measured too low or too high relative to mA demand 20 ms after exposure start

Ranked causes

  1. 1. kV control board failure (common)
  2. 2. Default filament currents not correctly adjusted (occasional)
  3. 3. HV Tank failure (rare)
First check

Measure actual mA and compare to demand value

Test point
mA feedback and mA command
Expected
Actual mA within spec of demanded value

No CAN message received (heater board)

Heater board has not received command from kV control main software for 5 seconds, loss of communication

Ranked causes

  1. 1. kV Control main software lost (common)
  2. 2. kV control or Heater board driver failure (common)
  3. 3. Bad contact on CAN bus line connector (occasional)
First check

Verify kV control main software status

Test point
Software status display
Expected
Software active

No CAN message received (rotation board)

Rotation board has not received any signal from kV control main software for the last 5 seconds, interpreted as loss of communication

Ranked causes

  1. 1. kV Control main software lost (common)
  2. 2. kV control or Rotation board driver failure (common)
  3. 3. Bad contact on CAN bus line connector (occasional)
First check

Verify kV control main software is running

Test point
Software status display
Expected
Software active

No kV feedback on anode side radiation

kV measured less than 12 kV at 0.5 ms after start of exposure on anode side only

Ranked causes

  1. 1. HV cable short circuit (occasional)
  2. 2. HV tank failure (occasional)
  3. 3. kV control board failure (rare)
First check

Measure anode-side kV 0.5 ms after exposure start

Test point
Anode kV feedback circuit
Expected
>12 kV

No kV feedback on cathode side radiation

kV measured less than 12 kV at 0.5 ms after start of exposure on cathode side only

Ranked causes

  1. 1. HV cable short circuit (occasional)
  2. 2. HV tank failure (occasional)
  3. 3. kV control board failure (rare)
First check

Measure cathode-side kV 0.5 ms after exposure start

Test point
Cathode kV feedback circuit
Expected
>12 kV

No mA feedback radiation

No mA feedback detected 20 ms after beginning of exposure

Ranked causes

  1. 1. HV tank failure (occasional)
  2. 2. kV control board failure (occasional)
  3. 3. X-ray tube filament open (occasional)
First check

Measure mA 20 ms after exposure start

Test point
mA feedback from HV tank
Expected
mA measured above threshold

Rotation Inverter permanent overcurrent high voltage

Overcurrent detected and 3 restart attempts failed unsuccessfully within a single rotation state

Ranked causes

  1. 1. Rotation board failure (common)
  2. 2. Tube stator winding short circuit (occasional)
  3. 3. Wiring shorted (occasional)
First check

Check stator winding continuity and insulation

Test point
Tube stator winding terminals
Expected
Resistance within spec; no short to ground

Rotation phases unbalanced caution

Amplitude difference of current between main and auxiliary rotation phases is too large

Ranked causes

  1. 1. One rotation wire missing (occasional)
  2. 2. Rotation board failure (occasional)
  3. 3. Wrong tube configuration database (rare)
First check

Measure current amplitude main vs auxiliary phases

Test point
Main and auxiliary rotation windings
Expected
Amplitude difference within spec

SG100 vertical locks do not function correctly patient safety

Vertical locks either do not engage or remain engaged when operating lock handle

Ranked causes

  1. 1. No 24 VDC between TP-9 and TP-10 (common)
  2. 2. Defective handle microswitch or relay K11 (occasional)
  3. 3. Relay K14 malfunction (lock release delayed) (rare)
First check

Measure voltage at lock board test points

Test point
TP-9 and TP-10
Expected
24 VDC

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Three-phase CB1 breaker trips or AC fuses blown high voltage

Main AC input protection device trips or fuses blow on generator power supply, indicating short circuit in power distribution

Ranked causes

  1. 1. IGBT in short circuit in inverter (common)
  2. 2. AC/DC or bridge rectifier in short circuit (common)
  3. 3. EMC filter in short circuit (occasional)
First check

Disconnect DC bus cables between AC/DC and inverter

Test point
DC bus cables on AC/DC board side
Expected
Cables safely disconnected with no voltage present

5 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Tube spit detection - kV- dropped high voltage

X-ray tube spit with high voltage cathode side voltage dropped

Ranked causes

  1. 1. Cathode side tube spit (common)
  2. 2. Cathode HV cable (occasional)
  3. 3. HV tank (occasional)

Tube spit detection - kV+ dropped high voltage

X-ray tube spit with high voltage anode side voltage dropped

Ranked causes

  1. 1. Anode side tube spit (common)
  2. 2. Anode HV cable (occasional)
  3. 3. HV tank (occasional)

Vertical locks do not engage or remain locked (SG100) patient safety

SG100 wall stand vertical locks fail to engage when lock handle released or remain engaged when handle operated

Ranked causes

  1. 1. Loss of 24 VDC to electromechanical lock (common)
  2. 2. Handle microswitch failure (occasional)
  3. 3. Relay K11 malfunction (occasional)
First check

Verify 24 VDC supply to lock system

Test point
TP-9 and TP-10
Expected
24 VDC present when lock handle released

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

AEC ion chamber HV power supply test failed

AEC HV power supply voltage measurement incorrect for ion chamber cells

Ranked causes

  1. 1. AEC HV power supply faulty (common)
  2. 2. Ion chamber cell wiring faulty (occasional)
First check

Measure HV power supply with no cell selected

Test point
AEC multiplex output
Expected
0 V

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bucky angulation does not lock or releases inadvertently (SG100) patient safety

Bucky angulation lock fails to hold position or binds during angulation motion

Ranked causes

  1. 1. Lock hardware misalignment or wear (occasional)
  2. 2. Bearing wear at bucky pivot point (occasional)
First check

Release angulation lock and tilt bucky forward and backward

Test point
Bucky angulation range and motion smoothness
Expected
Bucky tilts forward 20° and backward 90° freely and smoothly

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bucky angulation lock does not release caution

Angulation lock push button does not release the lock, or Bucky cannot tilt freely forward 20 degrees and backward 90 degrees.

Ranked causes

  1. 1. Lock board malfunction (common)
  2. 2. Bearing wear or damage at Bucky pivot point (occasional)
First check

Release angulation lock and verify Bucky tilts freely

Test point
Bucky tilt mechanism
Expected
Bucky tilts forward 20 degrees and backward 90 degrees freely and smoothly

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AEC KG parameter (grid correction) radiation

Ionization chamber 2 grid correction calibration for multi-chamber systems using automatic exposure control. Applies only when multiple ionization chambers present.

Ranked causes

  1. 1. KG calibration required for second ionization chamber in multi-chamber setup (common)
  2. 2. KG calibration in application mode when SID differs from KVEF setup (occasional)
First check

Select screen-film combination A (slowest preferred)

Test point
Calibration menu
Expected
Film type A selected

5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AEC KTIME parameter (exposure time correction) radiation

Calibration of exposure time correction parameters for automatic exposure control across 4 or 6 time intervals (10ms to 3.2s).

Ranked causes

  1. 1. KTIME calibration needed for 4 or 6 exposure time values (common)
  2. 2. Film darkness correction needed for exposure times below 10 ms (occasional)
First check

Verify KTIME preliminary T values

Test point
Service mode MENU4/AEC/F5/KTIME
Expected
T0=10ms, T1=100ms, T2=500ms, T3=1000ms, T4=2000ms, T5=3200ms
Tools
Service mode access

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Automated AEC Calibration radiation

Perform AEC (Automatic Exposure Control) system calibration using sensitometer and densitometer to ensure consistent film density

Ranked causes

  1. 1. AEC calibration not performed or requires update (common)
  2. 2. Positioner, collimator, or X-ray generator not configured (occasional)
First check

Perform tube warmup

Test point
X-ray tube
Expected
Tube warmed to operating temperature

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Backup & Restore Database Setup Parameters

Backup CPU memory parameters to floppy disk for rapid restoration or recovery after KV Control Board change

Ranked causes

  1. 1. Need to preserve CPU NVRAM setup parameters for system continuity (common)
  2. 2. Need to restore CPU NVRAM after KV Control Board replacement (occasional)
First check

Verify service laptop configuration and JEDI software installation

Test point
Service laptop
Expected
Laptop configured with JEDI service software installed

5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Film size detection

Film size display incorrect after cassette insertion; requires voltage adjustment at test points

Ranked causes

  1. 1. Incorrect voltage at film size sensor TP9 (common)
  2. 2. Incorrect voltage at film size sensor TP10 (common)
First check

Measure voltage between TP9 and GND

Test point
TP9
Expected
–0.84V±0.01V, ±0.01V
Tools
DMM

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Film size display incorrect

Film size display does not match inserted cassette size after cassette insertion or repositioning

Ranked causes

  1. 1. Cassette position sensor voltage out of calibration (common)
  2. 2. Second cassette sensor voltage out of calibration (common)
First check

Measure sensor voltage at TP9

Test point
TP9 to GND
Expected
-0.84V±0.01V, ±0.01V

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: JEDI 38-2 accuracy and reproducibility radiation

Verify mAs display accuracy across 60/80/120 kVp and reproducibility of exposures in 2-point mode; measure coefficient of variation (C) and linearity (L)

Ranked causes

  1. 1. Calibration adjustment required (unknown)
  2. 2. Reproducibility and linearity out of tolerance (unknown)
First check

Verify mAs display accuracy at 60 kVp

Test point
Control Console post-display
Expected
At 0.5 mAs input, display 0.74 mAs; at 20 mAs input, display 21.72 mAs; at 80 mAs input, display 86.28 mAs

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Light field – X-ray field alignment

Align collimator light field with X-ray field and calibrate line laser position for Proteus XR/a collimator

Ranked causes

  1. 1. Light field and X-ray field misalignment due to shipment or operational variation (occasional)
  2. 2. Line laser misaligned from light field center by more than 3mm (occasional)
First check

Power off system

Test point
Main power switch
Expected
System powered off

8 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: New tube operation verification radiation

Verify X-ray tube function and clear error logs after tube replacement with series of exposures at varying parameters.

Ranked causes

  1. 1. Tube replacement requires functional verification and error log clearance before clinical use (common)
  2. 2. Tube replacement documentation and shipment requirements (unknown)
First check

Execute test exposures at low kV and mA settings

Test point
Direct workstation exposure controls
Expected
Proper X-ray generation at low parameters, tube responds normally

5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Tube configuration database selection

Load appropriate JEDI Service Software database file matching installed X-ray tube type and power rating. Includes verification of console display parameters after load.

Ranked causes

  1. 1. Different tube types and power ratings require matching database to ensure proper operation (common)
  2. 2. Database mismatch prevents proper system operation and damages components (common)
First check

Identify installed X-ray tube type and power rating

Test point
Equipment nameplate or maintenance records
Expected
Tube designation (RAD 12/14/21 or MX 100) and power (32/50/65/80 kW) clearly identified

10 more checks with pass and fail actions, 12 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Workstation parameters configuration

Configure JEDI Service Software workstation programming for tube selection, exposure mode settings, AEC chamber selection, and film-screen combinations via TECH menu.

Ranked causes

  1. 1. Each installation requires workstation-specific parameter configuration via JEDI Service Software (unknown)
  2. 2. JEDI software preprogrammed values may not match room actual parameters requiring verification (occasional)
First check

Access TECH setup menu from JEDI Service Software

Test point
Setup mode menu 3 display
Expected
TECH menu accessible via F1 function key from menu 3

7 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Carriage motion not smooth or exhibits side play (SG100) caution

SG100 carriage does not move freely or smoothly over travel range; exhibits excessive side play

Ranked causes

  1. 1. Counterweights missing, misaligned, or trim covers improperly installed (common)
  2. 2. Bearing wear or misalignment at bucky pivot (occasional)
First check

Inspect counterweight installation and positioning

Test point
Counterweights and trim covers
Expected
All counterweights correctly positioned; trim covers fully installed

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Current under estimated fork range (low)
  • DS100 and/or DS200 LEDs off
  • Error code 0301-0304/0309-0314/0319/0323 reported
  • Error code 0505 reported
  • Error code 0505 reported during inverter short circuit test
  • Error codes 0301/0302/0303/0304/0309/0310/0311/0312/0313/0314/0319/0323 reported
  • Error codes 0501/0503/0504 reported
  • Filament current too high on inverter 1 for Boost
  • Filament current too high on inverter 1 for Heat
  • Filament current too high on inverter 1 for Pre-Heat
  • FPGA problem; Safety hit with unknown reason
  • Inverter max ILR current detected
  • No kV Feedback on both anode and cathode
  • PRD error (heater)
  • Re-starting safety triggered
  • Rotation current overload
  • Rotation function error codes 0103H–0106H
  • SG100 bucky angulation lock does not release
  • SG100 bucky rotation lock does not engage or slips
  • Software or database corrupt
  • Vertical locks do not release smoothly
  • AEC diagnostic test failed - error code not 1111111
  • AEC diagnostics result other combinations
  • AEC gain test (x10) failed
  • AEC parameters recording
  • AEC return measurement test failed
  • APR floppy disk recovery from floppy to floppy console
  • APR recovery from floppy disk console to USB disk console
  • APR recovery from USB disk console to USB disk console
  • Boost too long on inverter 1
  • Bucky angulation lock does not hold
  • Bucky angulation lock does not hold position or binds
  • Bucky angulation locks do not engage when releasing lock handle
  • Bucky angulation locks remain engaged when operating lock handle
  • Bucky angulation microswitches not actuating
  • Bucky grid does not oscillate during exposure
  • Bucky grid exposure shows shading or objectionable grid lines
  • Bucky grid not oscillating or oscillation not audible
  • Bucky locks fail to engage when equipment switched on
  • Bucky rotation lock does not engage
  • Bucky rotation lock does not engage or binds
  • Bucky rotation lock does not release
  • Bucky rotation lock not engaging properly
  • Bucky rotation microswitches not actuating
  • Bucky vertical locks remain engaged when operating lock handle
  • Bucky vertical motion not smooth or operating force excessive
  • Calibration: AEC KF format parameters
  • Calibration: AEC KF parameter (cassette sensitivity)
  • Calibration: AEC KF parameter (format correction)
  • Calibration: AEC KG grid correction parameters
  • Calibration: AEC KTIME parameter (time constant)
  • Calibration: AEC KVEF parameter (kV efficiency factor)
  • Calibration: AEC KVEF parameter acquisition
  • Calibration: Anti-collision buzzer disable
  • Calibration: Anti-collision switch adjustment
  • Calibration: Anti-collision switches adjustment
  • Calibration: Automatic Exposure Control (AEC)
  • Calibration: Checksum Validation
  • Calibration: Collimator Bulb Intensity
  • Calibration: Collimator bulb light intensity equality
  • Calibration: CPU memory backup
  • Calibration: CPU memory restore
  • Calibration: DOSE Calibration
  • Calibration: DOSE calibration setup
  • Calibration: Invasive mA measurement setup
  • Calibration: JEDI 38-2 AEC reproducibility
  • Calibration: KF parameter entry
  • Calibration: KG parameter entry
  • Calibration: KTIME parameter entry
  • Calibration: KVEF parameter entry
  • Calibration: Lateral cell check
  • Calibration: Lateral film size
  • Calibration: Line laser adjustment
  • Calibration: Longitudinal film size
  • Calibration: mA measurement procedure
  • Calibration: Manual Collimator Bulb Voltage
  • Calibration: Manual collimator lamp voltage
  • Calibration: Microswitch S4 and S5 Activator Cam adjustment for Auto Exposure positions
  • Calibration: Microswitch S4 and S5 positioning
  • Calibration: SG100 Wall Stand counterweight
  • Calibration: SG120 Wall Stand counterweight
  • Calibration: SG120 Wall Stand Vertical Locks
  • Calibration: SID (Source-to-Image Distance)
  • Calibration: SID (Source-to-Image Distance) measurement
  • Calibration: Software restoration for JEDI 7-2
  • Calibration: Table height limitation
  • Calibration: Touch screen (PN 2259976 with floppy disk)
  • Calibration: Touch screen (PN 5462233 with USB)
  • Calibration: Touch screen positioning
  • Calibration: Tube Configuration
  • Calibration: Tube data tracking reset
  • Calibration: Wall bucky lateral film size
  • Calibration: Wall bucky longitudinal film size
  • Calibration: Wall Stand counterweight
  • Calibration: X-ray Tube Warm Up
  • Carriage motion not smooth or exhibits side play (SG120)
  • Cassette size detection verification
  • Cassette size not displayed correctly on OTS console and collimator
  • Cassette tray pull-out or pull-in movement is not smooth
  • Checksum validation
  • Collimator bulb replacement
  • Current under estimated fork/range (high)
  • Database not correct (rotation board)
  • DS1 neon not lit
  • DS101/DS102/DS201/DS202 LEDs off during normal operation
  • DS300 neon not lit
  • Error code 0313 reported
  • Error code 0313 reported during inverter short circuit test
  • Error code 0320 reported
  • Error code 0501/0503/0504 reported
  • Error message from rotor firmware
  • F0 main frequency problem
  • Filament Database not correct (tubes 1-4)
  • Filament selection error
  • Float operation not functioning
  • Heater command error
  • Heater inverter permanent overcurrent
  • Heater Inverter permanent short circuit
  • Inverter short circuit diagnostic procedure
  • kV detected during kV diagnostics
  • kV did not reach 75% after 20 ms
  • kV MAX detected
  • kV unbalanced detected
  • Laser light alignment verification
  • Lateral cell response check
  • Measured current while inverter OFF
  • Microswitch S1 not actuated at vertical position
  • Microswitch S1 not actuated at vertical position (SG100)
  • Microswitch S1 not actuated at vertical position (SG120)
  • Microswitch S2-1 not actuated at 0 degrees rotation
  • Microswitch S2-1 not actuated at 0° rotation (SG100)
  • Microswitch S2-2 not actuated at 180 degrees rotation
  • Microswitch S2-2 not actuated at 180° rotation (SG100)
  • Microswitch S3 not actuated at horizontal position
  • Microswitch S3 not actuated at horizontal position (SG100)
  • Microswitch S3 not actuated at horizontal position (SG120)
  • Microswitch S4 not actuated at near vertical SID point
  • Microswitch S4 not actuated at near vertical SID point (SG100)
  • Microswitch S5 not actuated at far vertical SID point
  • Microswitch S5 not actuated at far vertical SID point (SG100)
  • No load HV diagnostic procedure
  • One of DS101/DS102/DS201/DS202 LEDs off while no error reported
  • PRD error (rotation)
  • Record AEC parameters mAs0 and table values
  • Replacement note: Rotation Capacitor Set when changing database files
  • Rotation command error
  • Rotation phase error
  • SG100 carriage motion not smooth or has excessive side play
  • Software download to generator
  • Software version verification
  • Spit Max error
  • Spit retry failed
  • System console replacement
  • System console replacement - warning labels verification
  • Table does not move with Up/Down pedal
  • Table height does not meet specification
  • Table height out of specification
  • Table up/down floats when pedal not depressed
  • Tomo-link system installation and calibration
  • Tube replacement software procedure
  • Tube spit detection - kV+ and kV- dropped
  • Unknown heater error
  • Unknown Rotation error
  • Vertical lock release delayed (SG100)
  • Vertical locks do not engage or remain locked (SG120)
  • Vertical SID point microswitches not actuating
  • Wall bucky does not move smoothly or requires excessive force

167 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
0001000AEC board communication errorAEC diagnostic test executionerror
0101HNo CAN message received within 5 secsCommunication failure on CAN bus for rotation functioncritical
0102HDatabase not correctRotation database validation failurecritical
0103HRotation function errorDetected during rotation diagnostic in high speed mode onlyerror
0104HRotation function errorDetected during rotation diagnostic in high speed mode onlyerror
0105HRotation function errorDetected during rotation diagnostic in high speed mode onlyerror
0106HRotation function errorDetected during rotation diagnostic in high speed mode onlyerror
0107HRotation Inverter permanent overcurrentRotation inverter detects sustained overcurrent conditioncritical
0108HRotation command errorInvalid or malformed rotation command receivedcritical
0109HMAINS_DROP has failedMain power supply monitoring has failedcritical
0110HPRD errorPower-on diagnostics failure in rotation circuitcritical
0111HF0 main frequency problemMain power frequency is incorrect or unstablecritical
0149HUnknown Rotation errorUnclassified rotation subsystem error detectedcritical
0151HCAN Domain command number errorCAN Domain command sequence number is incorrectinfo
0152HCAN Domain request with no transfer initCAN Domain request received without transfer initializationinfo
0153HCAN Domain Toggle bit errorCAN Domain toggle bit is incorrectinfo
0154HCAN Domain : less than 2 data to downloadCAN Domain download has insufficient datainfo
0155HCAN Domain Abort received & appliedCAN Domain transfer was abortedinfo
0156HBad index in config uploadConfiguration upload contains invalid index valueinfo
0157HTube switch while rotor not offTube selection changed while rotor is still spinninginfo
0158HAcceleration cmd while no tube selectedRotor acceleration command issued without tube selectioninfo
0159HAcceleration cmd while database not OKRotor acceleration command issued before database is validinfo
0160HDatabase download while rotor speedingConfiguration database download started while rotor is rotatinginfo
0161HAcceleration command not OKRotor acceleration command validation has failedinfo
0162HRotor acceleration while in errorRotor acceleration command attempted while system in error stateinfo
0163HNo CAN message received within 4 secsCAN bus timeout during rotation operationwarning
0164HRotation Inverter overcurrent (< 3 times)Inverter current exceeds threshold less than 3 timeswarning
0180HRotor board communication problemCommunication failure with rotor control boardcritical
0181HRotor board has resetRotor control board reset unexpectedlycritical
0199HUnknown rotation warningUnclassified rotation system warning detectedwarning
0201HNo CAN message received within 5 secsCommunication failure on CAN bus for heater functioncritical
0203HHeater inverter permanent overcurrentHeater inverter detects sustained overcurrent conditioncritical
0204HFilament permanent open circuitFilament circuit is open or disconnectedcritical
0205HHeater Inverter permanent short circuitHeater inverter detects short circuit conditioncritical
0206HFilament too high for Pre-heatFilament current exceeds limit during pre-heat phasecritical
0207HFilament too high for BoostFilament current exceeds limit during boost phasecritical
0208HFilament too high for HeatFilament current exceeds limit during heat phasecritical
0209HHeater command errorInvalid or malformed heater command receivedcritical
0210HCurrent under estimated forkMeasured filament current is below expected rangecritical
0211HCurrent over estimated forkMeasured filament current is above expected rangecritical
0212HMAINS_DROP detectedMain power supply has dropped during heater operationcritical
0213HPRD errorPower-on diagnostics failure in heater circuitcritical
0214HStay too long in BoostFilament remains in boost phase beyond allowed timecritical
0215HFilament selection errorInvalid filament selection or database mismatchcritical
0216HMeasured current while inverter OFFCurrent detected when heater inverter is offcritical
0221HFilament Database not correctFilament configuration database validation failurecritical
0248HUnknown Heater errorUnclassified heater subsystem error detectedcritical
0251HReceived command is not OKInvalid heater command receivedwarning
0252HHeater command not OKHeater command validation failurewarning
0253HNo CAN message received within 4 secsCAN bus timeout during heater operationwarning
0254HHeater inverter overcurrent (inverter1) (<3 times)Inverter1 current exceeds threshold less than 3 timeswarning
0255HFilament open circuit (inverter1) (<3 times)Filament continuity loss in inverter1 less than 3 times detectedwarning
0256HHeater Inverter short circuit (inverter1) (<3times)Inverter1 short circuit detected less than 3 timeswarning
0257HTube switch while filaments not OFFTube switching commanded while filament heating still activewarning
0258HCAN Domain command number errorCAN message domain command counter mismatch or corruptionwarning
0259HCAN Domain request with no transfer initCAN data transfer requested without proper initializationwarning
0260HCAN Domain Toggle bit errorCAN toggle bit sequence mismatch detectedwarning
0261HCAN Domain : less than 2 data to downloadDownload initiated with insufficient data segmentswarning
0262HCAN Domain Abort received & appliedSystem received and executed abort command on CAN buswarning
0263HDatabase download while heater not cutAttempted database update while heater supply still activewarning
0280HHeater board communication problemCommunication failure with heater control boardcritical
0281HHeater board has resetHeater control board reset unexpectedlycritical
0299HUnknown Heater warningUnclassified heater system warning detectedwarning
0301kV control board error during inverter diagnosticsDetected during inverter short circuit or HV diagnosticserror
0301/0302/0303/0304/0309/0310/0311/0312/0313/0314/0319/0323Multiple kV control or inverter faults during short circuit diagnosticShort circuit diagnostic test sequenceerror
0301HTube spit (kV+ and kV- dropped)High voltage regulation failure during tube operationerror
0302kV control board error during inverter diagnosticsDetected during inverter short circuit or HV diagnosticserror
0302HTube spit (kV+ has dropped)Positive high voltage has dropped unexpectedlyerror
0303kV control board error during inverter diagnosticsDetected during inverter short circuit or HV diagnosticserror
0303HTube spit (kV- has dropped)Negative high voltage has dropped unexpectedlyerror
0304kV control board error during inverter diagnosticsDetected during inverter short circuit or HV diagnosticserror
0304HTube spit (kV regulation error)High voltage regulation failure during tube operationerror
0305HFPGA problem (restarting safety signal)Field programmable gate array safety signal restartederror
0306HNo kV Feedback on anodeHigh voltage anode feedback signal is missing or invaliderror
0307HNo kV Feedback on cathodeHigh voltage cathode feedback signal is missing or invaliderror
0308HNo kV Feedback on anode and cathodeBoth anode and cathode feedback signals are missingerror
0309kV control board error during inverter diagnosticsDetected during inverter short circuit or HV diagnosticserror
0309HkV detected during kV diagHigh voltage detected when it should be absent during diagnosticserror
0310kV control board error during inverter diagnosticsDetected during inverter short circuit or HV diagnosticserror
0310HkV max detectedHigh voltage exceeds maximum safe limiterror
0311No Ilp (parallel inductor current) detectedDuring inverter short circuit diagnosticerror
0311HILP current not OKILP (left phase) current is out of specificationerror
0312No Ilr (series inductor current) detectedDuring inverter short circuit diagnosticerror
0312HILR current not OKILR (right phase) current is out of specificationerror
0313kV control board internal errorDuring inverter short circuit diagnosticerror
0313HILR max current detectedILR (right phase) current exceeds maximum limiterror
0314Ilr current resonant frequency lower than expectedDuring inverter short circuit diagnosticerror
0314HILR current timeoutILR current measurement took too long to stabilizeerror
0316HSpit Max errorTube spit magnitude exceeded maximum thresholderror
0317HSpit Ratio errorRatio between spit events indicates abnormal patternerror
0318HkV did not reach 75% after 20msHigh voltage ramp too slow during startuperror
0319kV control board error during inverter diagnosticsDetected during inverter short circuit or HV diagnosticserror
0319HkV unbalanced detectedPositive and negative high voltages are not balancederror
0320Persisting kV control or inverter faultShort circuit or no load diagnostic test continuation after initial faulterror
0320kV control board error that persists after diagnosticsDuring inverter diagnostics when problem remains after cable checkserror
0320HFPGA problem (safety signal)Field programmable gate array safety signal failureerror
0321HSpit retry failedAttempt to recover from tube spit was unsuccessfulerror
0321HkV conversion errorHigh voltage analog to digital conversion has failedcritical
0322HkV ref ADC / DAC failedReference or analog conversion circuit has failedcritical
0323Both Ilr and Ilp currents not detectedDuring inverter short circuit diagnosticerror
0323HILP and ILR current not OKBoth ILP and ILR currents are out of specificationerror
0324HSPITS (data=1:Anode,2:cathode,4:Both,8:kV regul)Tube spit event with specified component affectederror
0401HNo mA feedbackTube current (mA) feedback signal is missing or invaliderror
0402HmA scale errormA measurement scale or calibration is incorrecterror
0403HmA accuracy exceeded 5%mA measurement error exceeds 5% toleranceerror
0501kV control or inverter faultDuring inverter diagnosticserror
0501/0503/0504kV control or inverter fault during diagnosticsShort circuit or no load diagnostic test executionerror
0501HDC bus out of rangeDC bus voltage is outside safe operating limitscritical
0503kV control or inverter faultDuring inverter diagnosticserror
0503HInverter Gate Power Supply errorGate drive power supply voltage is incorrectcritical
0504kV control or inverter faultDuring inverter diagnosticserror
0504HInverter Gate Power Supply failedGate drive power supply for inverter has failederror
0505Isolation fault between inverter components and groundDuring inverter diagnosticserror
0505HMains power supply has dropped during exposureMain power supply failure occurred during exposurecritical
0506HDC bus 1 phase precharge errorDC bus precharge failed for single phase modecritical
0507HDC bus 1 phase discharge errorDC bus discharge failed for single phase modecritical
0510HAEC Cell HV supply errorHigh voltage supply for AEC cell is faultycritical
0549HUnknown LVPS errorUnclassified low voltage power supply error detectedcritical
0550HNo more warn +160V too low or too highRecovery from +160V voltage warning conditioninfo
0551HDetected +160V too high+160V supply exceeded high thresholdwarning
0553HDetected +160V too high+160V supply voltage is above maximum limitcritical
0555HDetected +160V too low+160V supply dropped below low thresholdwarning
0557HDetected +160V too low+160V supply voltage is below minimum limitcritical
0560HNo more warn +15V too low or too highRecovery from +15V voltage warning conditioninfo
0561HDetected +15V too high+15V supply exceeded high thresholdwarning
0563HDetected +15V too high+15V supply voltage is above maximum limitcritical
0565HDetected +15V too low+15V supply dropped below low thresholdwarning
0567HDetected +15V too low+15V supply voltage is below minimum limitcritical
0570HNo more warn -15V too low or too highRecovery from -15V voltage warning conditioninfo
0571HDetected -15V too strong-15V supply magnitude exceeded high thresholdwarning
0573HDetected -15V too strong-15V supply voltage is too high (more negative)critical
0575HDetected -15V too weak-15V supply magnitude dropped below low thresholdwarning
0577HDetected -15V too weak-15V supply voltage is too low (less negative)critical
0599HUnknown LVPS warningUnclassified low voltage power supply warning detectedwarning
0601HRTL errorReal-time logic circuit has failedcritical
0602HExternal CAN bus offExternal CAN bus communication is not activecritical
0603HDebug screen com errorCommunication with debug display has failedcritical
0604HDatabase download errorConfiguration database download or transfer has failedcritical
0605HTAV communication errorCommunication with TAV system has failedcritical
0606HMPC/Madrid communication errorCommunication with MPC/Madrid system has failedcritical
0701HSaved RAM checksum pbSaved RAM checksum verification has failedcritical
0702HSoftware problemGeneral software or firmware fault detectedcritical
0703HWatchdog reset has just occurredWatchdog timer expired causing system reseterror
0704HRotor/Heater hold too longRotor or heater command hold state exceeded timeoutcritical
0801HExposure backup mAs exceededExposure terminated due to backup mAs limit reachederror
0802HExposure backup time exceededExposure terminated due to backup time limit reachederror
0803HExp cmd while gene not readyExposure command received but generator not in ready stateerror
0804HTank Thermal ErrorTank temperature has exceeded safe operating limitwarning
0902Htube Fan supply errorTube cooling fan power supply has failedcritical
0903HTube exceeded 70degCTube temperature has exceeded 70 degrees Celsiuswarning
100-0603HDebug screen com. errorDebug screen communication failureinfo
100-0604HDatabase download errorProblem detected during database downloaderror
100-0605HTAV communication errorCommunication problem between I/F and service laptoperror
100-0606HMPC/Madrid Communication errorNo reply from the consoleerror
100-1301HAEC communication errorkV Control detects communication problem with AEC boarderror
110-0804HTank Thermal ErrorHV tank temperature reached 60°Cwarning
110-0903H70 degree C thermal safety error70° loop detected openerror
1111111AEC diagnostic test passedAEC diagnostic test completioninfo
120-1500HTomo brightness not goodBrightness signal out of range (+/-20%) due to incorrect techniquewarning
120-1501HExposure switch releasedHandswitch released before end of exposurewarning
120-1502HAEC CUT off errorAEC did not cut exposure as expectedwarning
1301HAEC communication errorCommunication with AEC system has failedcritical
1401HSaved RAM power supply limit reachedRAM power budget computation exceeds threshold over ~7 year periodinfo
1402HInternal CAN bus offInternal CAN bus communication is not activecritical
1403HConnectic FaultConnector or connection fault detectedcritical
1404HFPGA configuration problemField programmable gate array configuration has failedcritical
1405HTank sensor problemTank temperature or condition sensor has failedcritical
1407HmAs counter errorExposure mAs counter is not operating correctlyerror
1408HAEC counter errorAEC counter is not operating correctlyerror
1409HmAs meter saturatedmA meter has reached maximum measurable valueerror
1410HFPGA LockedField programmable gate array is in locked or hung stateerror
1411HTime counter errorExposure time counter is not operating correctlyerror
1420HTomo exposure cut by tomograph instead generatorTomography exposure was terminated by tomograph system instead of generatorerror
1500HTomo brightness errorTomography system brightness control error during manipulationwarning
1501HExposure switch released during exposureExposure switch was released before exposure completedwarning
1502HAEC cut off errorAEC system terminated exposure prematurelywarning
30-0301HTube spit (kV+ and kV- dropped)Simultaneous voltage drop detected on both anode and cathode HV linescritical
30-0302HTube spit (kV+ has dropped)Voltage drop detected on anode (positive) HV line onlycritical
30-0303HTube spit (kV- has dropped)Voltage drop detected on cathode (negative) HV line onlycritical
30-0304HkV regulation errorSlow voltage drop detected indicating smooth or gradual tube spit conditionerror
30-0305HRe-starting safety (unknown reason)Safety interlock activated with no corresponding detected error on primary error inputswarning
30-0324HCathode or anode spit detectedData parameter indicates spit type: 1=cathode, 2=anode, 4=both, 8=kV regulationerror
40-0101HNo CAN message received within 5 secondsRotation board communication loss from kV control main softwareerror
40-0102HRotation board database incorrectFirmware detects wrong data in database received from kV control board at power-uperror
40-0103HRotation motor current overloadMain or auxiliary rotation current exceeds maximum tube ratingerror
40-0104HRotation current open loadNo current flowing to rotation motorerror
40-0105HRotation phases unbalancedAmplitude difference between main and auxiliary current too largeerror
40-0106HRotation phase shift errorAnode stator current lacks correct phase relationshiperror
40-0107HRotation inverter permanent overcurrentOvercurrent detected with 3 unsuccessful restart attempts within single rotation stateerror
40-0108HRotation command error (debug)Erroneous rotation command received by boardwarning
40-0109HMains drop detected (rotation)Firmware detected mains_drop signal activationerror
40-0110HPRD error on rotation boardFirmware checksum, RAM test, or EPLD access failure at power-up or reseterror
40-0110HRotation board PRD self-test failureFirmware checksum, RAM, or EPLD test failed at power-up or reseterror
40-0111HF0 main frequency problemEPLD did not apply inverter start commanderror
40-0149HUnknown rotation errorMain software received unrecognized error from rotation boardwarning
40-111HRotor firmware errorHardware/software conflict detected in rotorerror
50-0201HNo CAN message received on heater board within 5 secondsHeater board communication loss from kV control main softwareerror
50-0203HHeater inverter permanent overcurrent3 restarts failed within 100 mserror
50-0204HFilament permanent open circuitOpen circuit detected with 3 unsuccessful restart attempts within 100 mserror
50-0205HHeater inverter permanent short circuit3 restarts failed within 100 mserror
50-0206HFilament current too high during pre-heatRMS current exceeds database maximum for tubewarning
50-0207HFilament current too high during boostRMS current exceeds database maximum for tubewarning
50-0208HFilament current too high during heatRMS current exceeds database maximum for tubewarning
50-0209HHeater command error (debug)Erroneous heater command receivedwarning
50-0210HFilament current over-estimated (too high)RMS current measurement too high in measurement rangewarning
50-0211HFilament current under-estimated (too low)RMS current measurement too low in measurement rangewarning
50-0212HMains drop detected on heater boardFirmware detects mains_drop signal activationerror
50-0213HPRD error on heater boardFirmware checksum, RAM test, or EPLD access failure at power-up or reseterror
50-0213HHeater board PRD self-test failureFirmware checksum, RAM, or EPLD test failed; also -15V missingerror
50-0214HBoost phase timeoutBoost command duration exceeds 400 mswarning
50-0214HBoost too long on inverter 1Boost command stayed longer than 400 mswarning
50-0215HFilament selection relay errorRelay position does not match filament selectionerror
50-0216HCurrent measured with inverter OFFInverter current detected when inverter not commandedwarning
50-0221HFilament database not correct for tube 1Firmware detects erroneous database from kV control at power-uperror
50-0221H/0222H/0223H/0224HFilament database error for tube 1, 2, 3, or 4Erroneous database parameters received at power-uperror
50-0222HFilament database not correct for tube 2Firmware detects erroneous database from kV control at power-uperror
50-0223HFilament database not correct for tube 3Firmware detects erroneous database from kV control at power-uperror
50-0224HFilament database not correct for tube 4Firmware detects erroneous database from kV control at power-uperror
50-0248HUnknown heater errorMain software receives error from heater board with no associated error codewarning
50-0248H/0249HUnknown heater errorUnrecognized error received from heater boardwarning
50-0249HUnknown heater errorMain software receives error from heater board with no associated error codewarning
60-0306HNo kV feedback on anode sidekV measured <12 kV 0.5 ms after start of exposure on anode side onlyerror
60-0307HNo kV feedback on cathode sidekV measured <12 kV 0.5 ms after start of exposure on cathode side onlyerror
60-0308HNo kV feedback on both anode and cathodekV <12 kV at 0.5 ms after exposure on both sideserror
60-0309HkV detected during diagnosticskV measured during inverter diagnostics when should be offwarning
60-0310HkV MAX limit exceededkV reached 160 kV during exposureerror
60-0311HILP (parallel resonant) current not OKCurrent in parallel LC circuit did not rise at exposure starterror
60-0312HILR (serial resonant) current not OKCurrent in serial LC circuit did not rise at exposure starterror
60-0313HInverter maximum ILR current detectedHardware detects maximum current in serial resonant circuiterror
60-0314HILR current timeoutResonant frequency lower than expectederror
60-0316HSpit max errorNumber of tube spits during exposure reaches limiterror
60-0316HTube spit count limit exceededNumber of tube spits reached maximum during exposureerror
60-0318HkV did not reach 75% after 20 msNo HV ON signal generated for exposure time count-uperror
60-0318HkV ramp too slowkV did not reach 75% after 20 ms from exposure starterror
60-0319HkV unbalanced detectedMore than 12 kV difference between kV+ and kV-error
60-0319HkV imbalance between positive and negativeDifference >12 kV between kV+ and kV-error
60-0320HFPGA safety interlock triggered (unknown cause)Safety hit detected with no error signal presenterror
60-0321HSpit recovery retry failedRetry mechanism unable to clear spit conditionerror
60-0323HBoth ILP and ILR currents absentNo inverter current measured at exposure starterror
60-0401HNo mA feedbackNo mA detected 20 ms after exposure starterror
60-0402HmA scale or calibration errormA too low or too high vs. demand 20 ms after exposure starterror
60-0403HmA accuracy exceeded 5%Measured mA every 50 msec exceeded 5% of mA demandwarning
60-0504HInverter Gate Power Supply errorGate supply voltage dropped below level required to drive IGBTs properlyerror
60-0801HExposure backup mAs exceededExposure command lasted so long that maximum mAs allowed has been reachederror
60-0802HExposure backup time exceededExposure command lasted longer than duration loaded by system (Backup time + 5%)error
60-0803HExp cmd while gene not readyGenerator received exposure command while not in ready stateerror
60-1407HmAs counter errorError found in verifying counter normal operationerror
60-1408HAEC counter errorNo AEC feedback in AEC station exposureerror
60-1409HmAs meter saturatedmAs counter shows unrealistic value at exposure start checkerror
60-1410HFPGA lockedFPGA detected error and blocked exposure after command receivederror
60-1411HTime counter errorError found in verifying counter normal operationerror
60-1420HTomo exposure cut by TomographyExposure cut during tomography operationwarning
70-0501HDC bus out of range (<400 or >850)DC bus voltage outside acceptable limitscritical
70-0503HInverter Gate Power Supply error (checked at prep)Gate supply voltage check fails during preparationerror
70-0505HMains power supply has dropped During exposureMains power drops during exposureerror
70-0510HAEC cell HV supply errorPower supply 230Vdc for AEC failederror
70-0549HUnknown LVPS errorMain software received LVPS error with no associated error codeerror
70-0553HDetected +160 V too high+160 V voltage higher than 225 V at heater boarderror
70-0557HDetected +160 V too low+160 V voltage lower than 110 V at heater boarderror
70-0563HDetected +15 V too high+15 V voltage higher than 17.3 V at heater boarderror
70-0567HDetected +15 V too low+15 V voltage lower than 13 V at heater boarderror
70-0573HDetected -15 V too high-15 V voltage lower than -17.3 V at heater boarderror
70-0577HDetected -15 V too low-15 V voltage higher than -13 V at heater boarderror
80-0180HRotation board communication problemkV control board communication watch dog timeout with rotation boarderror
80-0181HRotation board has resetkV control detected rotation board reset; reloading databasewarning
80-0280HHeater board communication problemkV control board communication watch dog timeout with heater boarderror
80-0321HKV conversion errorkV feedback measured while no kV is being generatederror
80-0322HkV ref ADC / DAC failedkV control DAC and ADC coherency test failederror
80-0601HRTL error (Real Time Lines show wrong state)Real Time Lines test detects wrong state during standbyerror
80-0601HRTL errorReal Time Lines show wrong state in regular standby testingerror
80-0602HExternal CAN bus offCAN bus communication failure external to generatorerror
80-0902HTube Fan supply errorWrong voltage applied to tube fan (not 115 V)error
80-1402HInternal CAN bus offCAN device on kV control board detected abnormal level on 2 lineserror
80-1403HConnection FaultOne of the flat cable connectors is not connected in Generatorerror
80-1404HFPGA configuration problemDetected during power up; main software cannot load FPGAcritical
80-1405HTank temperature sensor problemHV tank temperature value less than 10°Cwarning
90-0701HNV RAM checksum problemGenerator kV control board detects NV RAM corruption after power upcritical
90-0702HSoftware problemSoftware or database error detectederror
90-0703HWatchdog reset has just occurredCPU of kV control board reset by watchdog timererror
90-0704HRotation/Heater hold too longPreparation command maintained longer than 3 minuteswarning
90-0705HConfiguration errorSoftware/hardware conflict detectederror
Cannot open fileBackup or restore operation cannot access specified file on floppy diskAttempt to backup CPU memory or restore from floppy disk fails at file accesserror
Checksum error (unspecified code)Stored and computed parameter checksums do not matchParameter changed or RAM corruption detected on power-up or after settings changescritical
Error communicationData transmission error occurred during backup or restore transferChecksum or parity error detected during file transfer between laptop and generatorerror
Error time outCommunication timeout between service laptop and generator during backup or restoreTransfer does not complete within timeout period during memory backup or restoreerror
File contains data. Delete old (y,n)Target file already exists on floppy disk during backup operationBackup attempts to create file that already exists with datawarning
Z00RAM D1 error (fatal)RAM D1 failure detected during self-testcritical
Z00Collimator UIF board errorUser Interface board failure detected or not respondingcritical
Z01EEPROM D1 error (fatal)EEPROM D1 failure detected during self-testcritical
Z01Collimator Control Board errorMain collimator control board failure or not respondingcritical
Z02Collimator Control board errorCollimator control board detects communication or logic faulterror
Z02CAN error (bus error, error overrun)CAN bus error frame detected or error counter overrunerror
Z03Collimator UIF ErrorUser Interface board detected a communication or function errorerror
Z03CAN error (receiver overrun)CAN receiver buffer overflow; messages arriving faster than processor handleserror
Z04Collimator local communication errorCommunication failure between collimator submodules over local buserror
Z05Calculation error - heightHeight motor position calculation failure or out-of-range resulterror
Z06Collimator Blade out of limitBlade position outside mechanical limit during operationerror
Z06Calculation error - widthWidth motor position calculation failure or out-of-range resulterror
Z10Collimator internal errorCollimator processor detected internal logic or state machine errorerror
Z10Stepping error - height leavesHeight stepper motor or blade positioning errorerror
Z11Stepping error - width leavesWidth stepper motor or blade positioning errorerror
Z11Collimator internal errorCollimator processor detected internal logic or state machine errorerror
Z14Collimator Homing ErrorCollimator failed to find home position during initialization sequenceerror
Z14Limit switch - height leavesHeight motor movement detected when limit switch should prevent iterror
Z15Limit switch - width leavesWidth motor movement detected when limit switch should prevent iterror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.