GE Healthcare Panda iRes: troubleshooting and error codes
The ServiceForge knowledge base holds 142 documented symptoms for the GE Healthcare Panda iRes, with 244 ranked causes, 339 diagnostic checks with test points and expected values, and 21 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Baby does not improve during resuscitation patient safety
Infant fails to respond adequately to T-piece resuscitator ventilation or does not reach desired peak inspiratory pressure
Ranked causes
- 1. Face mask not properly positioned on infant (common)
- 2. Gas supply disconnected or improperly connected (common)
- 3. Inadequate flow delivered to circuit (common)
- 4. PEEP and PIP not correctly set (common)
- 5. Leaks in T-piece circuit (occasional)
Visually inspect face mask placement on infant face
- Test point
- Face mask
- Expected
- Proper seal on face with no gaps
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas supply connection improper seal high voltage
Hissing or leaking sounds detected when opening tank valve, indicating inadequate seal between cylinder and regulator
Ranked causes
- 1. Gasket permanently deformed or damaged (common)
- 2. Gasket contaminated with oil, grease, or dirt (occasional)
- 3. Multiple gaskets installed (occasional)
- 4. Gasket improperly seated or misaligned (occasional)
- 5. Regulator or valve damaged (rare)
Visually inspect sealing gasket before installation
- Test point
- Sealing gasket on valve-to-regulator connection
- Expected
- Gasket shows no permanent deformation, cracks, chips, frays, splits, gouges, or contamination
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate temperature readings patient safety
Skin temperature probe not correctly positioned or attached, affecting Baby Mode control
Ranked causes
- 1. Probe positioned between baby and mattress (common)
- 2. Probe not in contact with baby's skin (common)
- 3. Reflective patch removed or damaged (occasional)
- 4. Non-Ohmeda probe in use (occasional)
- 5. Disposable probe cable or tip damaged (rare)
Visually inspect probe position on baby
- Test point
- Baby's abdomen or back
- Expected
- Probe on side of baby facing heater dish, directly in path of radiant heat
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No or inaccurate primary flow
Primary flow absent or does not match expected value; may indicate toggle valve, PIP control, or supply issue
Ranked causes
- 1. Primary flow toggle valve off (common)
- 2. PIP knob not at correct position (occasional)
- 3. Faulty PIP relief valve diaphragm (occasional)
- 4. Contaminated supply gas filters or relief valve leakage (occasional)
- 5. Faulty regulator, toggle valve, or blender (occasional)
Confirm any flow present
- Test point
- Primary flow output
- Expected
- Flow present or absent
14 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No or low PIP pressure
Positive inspiratory pressure (PIP) does not reach expected value; may indicate gauge error, supply issue, or relief valve problem
Ranked causes
- 1. Gauge in unit is inaccurate (occasional)
- 2. Inadequate supply pressure or contaminated filters (occasional)
- 3. Relief valve leakage or faulty check valve (occasional)
- 4. Leak in gas system (occasional)
- 5. Faulty PIP regulator (occasional)
Confirm PIP pressure with external test instrument
- Test point
- PIP system pressure point
- Expected
- Low PIP confirmed with independent measurement
7 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check valve test fails caution
Leakage detected during check valve test at test ports
Ranked causes
- 1. Faulty check valve at C1 or C2 port (common)
- 2. Both C1 and C2 check valves faulty (occasional)
- 3. Leak in threaded fittings or tube connections (occasional)
- 4. High pressure relief valve leakage at poppet (occasional)
Determine if unit has cylinder yokes on back
- Test point
- Unit back panel
- Expected
- Yes or No
- Tools
- Visual inspection
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas supply connection improper seal high voltage
Hissing or leaking sounds detected when opening tank valve, indicating inadequate seal between cylinder and yoke
Ranked causes
- 1. Sealing gasket permanently deformed or damaged (common)
- 2. Gasket contaminated with oil, grease, or dirt (common)
- 3. Multiple gaskets used in sealing connection (occasional)
- 4. Yoke handle over-tightened or under-tightened (occasional)
Visually inspect sealing gasket for damage
- Test point
- Cylinder valve outlet to yoke connection
- Expected
- Gasket shows no permanent deformation, splits, cracks, chips, frays, or gouges; no dirt, oil, grease, or contaminants
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Maximum suction less than 150 mmHg
Suction relief valve does not maintain adequate vacuum when maximum suction control set
Ranked causes
- 1. Suction relief valve faulty or contaminated (occasional)
- 2. Inadequate suction regulator pressure or supply issue (occasional)
- 3. Leak in collection bottle or tubing (occasional)
- 4. Vacuum generator failure (rare)
Occlude suction relief valve opening and slowly increase suction knob
- Test point
- Suction relief valve outlet
- Expected
- Suction reaches 150 mmHg when opening occluded
8 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No or inaccurate auxiliary flow
Auxiliary flow is absent or does not match expected value; may indicate toggle valve issue or supply problem
Ranked causes
- 1. Auxiliary flow toggle valve off (common)
- 2. Contaminated supply gas filters (occasional)
- 3. Faulty check valve (occasional)
- 4. Leak in gas system (occasional)
Confirm any flow present
- Test point
- Auxiliary flow output
- Expected
- Flow present or absent
8 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen analyzer reading does not match blender setting
Analyzer output concentration differs from the blender control setting during operation.
Ranked causes
- 1. Analyzer out of calibration (common)
- 2. Blender out of calibration (common)
- 3. Bleed orifice in outlet fitting restricted (occasional)
- 4. Air leak downstream of blender diluting concentration (occasional)
Calibrate oxygen analyzer and verify readings match known concentrations
- Test point
- Analyzer display
- Expected
- Readings within calibration range
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen delivery system setup and verification patient safety
Configuration of oxygen delivery circuit including free-flow oxygen, manual resuscitator connection, pressure monitoring, flow rate adjustment, and FiO2 concentration verification for resuscitation support.
Ranked causes
- 1. Initial system configuration required (common)
- 2. Flow rate adjustment (common)
- 3. PEEP and PIP adjustment (common)
- 4. FiO2 concentration adjustment (blender-equipped units) (occasional)
Verify oxygen tubing connection to left outlet
- Test point
- Left gas outlet on front panel
- Expected
- Oxygen tubing properly seated
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Suction flow is low caution
Suction outlet flow below expected level
Ranked causes
- 2. Faulty suction regulator (occasional)
- 3. Faulty suction valve (occasional)
- 4. Faulty vacuum generator (occasional)
- 5. System leak (occasional)
Remove rear panel and replace suction filter
- Test point
- Suction filter assembly
- Expected
- Filter replaced; suction flow improves
- Tools
- Screwdriver or wrench
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Suction system not operating or pressure incorrect patient safety
Suction system fails to generate adequate vacuum or pressure reading is outside target range of 60-120 mm Hg
Ranked causes
- 1. Suction canister not properly installed or missing overflow protection (common)
- 2. Suction tubing not properly connected (common)
- 3. Suction control switch off or pressure not adjusted (common)
- 4. Oxygen supply not available or turned off (occasional)
Inspect suction canister installation and overflow protection
- Test point
- Suction canister dovetail rail attachment
- Expected
- Clean canister properly installed with hydrophobic filter or overflow trap present
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baby temperature elevated compared to remote site measurement patient safety
In Baby Mode, temperature measured at remote sites (axilla, rectum, ear) reads up to 2.4°C higher than abdominal skin temperature; warmer skin probe reading matches Set Temp with no alarm triggered; signs include skin redness, warmth to touch, diaphoresis, increased heart rate
Ranked causes
- 1. Warmer temperature probe poor skin contact or suboptimal placement on abdomen (common)
- 2. Extended operation in manual mode at 100% heater power (occasional)
- 3. Probe placement on abdomen does not reflect actual core temperature (occasional)
Observe patient for signs of heat stress
- Test point
- Patient skin, physical assessment
- Expected
- No skin redness, warmth, diaphoresis or elevated heart rate
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baby temperature probe not responding to warmth patient safety
When skin probe is warmed by hand, baby temperature reading does not increase or shows incorrect temperature
Ranked causes
- 1. Temperature probe sensor failure (occasional)
- 2. Temperature probe connector loose or corroded (occasional)
- 3. Temperature input circuit failure (rare)
Warm temperature probe with fingers
- Test point
- Temperature probe sensor
- Expected
- Probe warmed to hand temperature
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed tilt mechanism inoperative or difficult patient safety
Bed tilt control not responsive, head of bed does not raise, or bed will not hold position at intermediate angles
Ranked causes
- 1. Tilt mechanism hydraulic failure or leak (occasional)
- 2. Tilt control damaged or misadjusted (occasional)
- 3. Bubble level indicating mattress not level (occasional)
Squeeze tilt control and push down on foot of bed
- Test point
- Bed head position
- Expected
- Head of bed raises easily
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blender alarming caution
Blender alarm activates during normal operation or when gas supply pressures are within acceptable range.
Ranked causes
- 1. Air and oxygen source pressure differential exceeds 207 kPa (30 psi) (common)
- 2. Alarm system out of calibration (occasional)
- 3. Dirty gas contaminating alarm system (occasional)
Measure air and oxygen supply pressures
- Test point
- Supply pressure gauges or regulators
- Expected
- Differential within 207 kPa (30 psi), ≤ 30 psi differential
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PIP variable relief valve patient safety
Adjustment of the variable PIP adjustable resistor on the yellow line to establish correct pressure relief at the override button position (31-33 cm H2O) and maximum position (45-48 cm H2O).
Ranked causes
- 1. PIP variable relief valve requires calibration (common)
- 2. Second test setup with flow measurement at RV2 test port (common)
- 3. PIP variable relief valve calibration setup verification (unknown)
Measure pressure at primary outlet with variable resistor adjustment
- Test point
- Primary flow outlet
- Expected
- 45-48 cm H2O, ±1.5 cm H2O
- Tools
- Low pressure measurement device
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Casters not in firm floor contact or unit unstable patient safety
Unit mobility and stability compromised; casters not touching floor or unit wobbles; brakes ineffective
Ranked causes
- 1. Casters worn, misaligned, or not fully extended (occasional)
- 2. Caster brakes worn or misadjusted (occasional)
- 3. Floor surface uneven or contaminated (occasional)
Inspect all four casters for floor contact
- Test point
- Each caster wheel
- Expected
- All casters in firm contact with floor
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Confirm Probe Jack Connection alarm activated patient safety
Alarm indicates temperature probe jack is unplugged or probe is disconnected
Ranked causes
- 1. Temperature probe not fully inserted into jack (common)
- 2. Manual Temperature Alarm is enabled without probe in jack (occasional)
- 3. Temperature probe failure or defective jack connection (rare)
Verify temperature probe is fully inserted in jack
- Test point
- Probe jack connection
- Expected
- Probe fully seated and making electrical contact
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Displays or indicators do not illuminate on power up
After power switch activation, displays or indicator lights fail to illuminate, software revision not displayed, or prompt tone absent
Ranked causes
- 1. Power supply not connected or outlet not live (common)
- 2. Main power switch not activated (common)
- 3. Display or LED circuit failure (occasional)
Verify power cord connection
- Test point
- Power inlet on rear of unit and hospital wall outlet
- Expected
- Power cord fully inserted and outlet is live
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Displays or indicators do not light on power-up
When power is switched on, displays and indicators do not illuminate or software revision does not appear
Ranked causes
- 1. Power supply not connected or faulty (common)
- 2. Main switch or standby switch not activated (common)
- 3. Controller or display board failure (occasional)
Verify power cord connection
- Test point
- Power inlet on unit and wall outlet
- Expected
- Power cord firmly connected to both unit and hospital-grade wall outlet
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manometer relief valve not relieving pressure caution
Relief valve poppet stuck or not opening at correct pressure
Ranked causes
- 1. Obstruction blocking relief valve poppet (common)
- 2. Poppet stuck due to foreign material (common)
- 3. Relief valve poppet damaged (occasional)
Apply 100 cmH2O to gauge fitting; observe if poppet lifts
- Test point
- Relief valve poppet, gauge fitting
- Expected
- Poppet lifts from seat cm H2O
- Tools
- Pressure source, observation
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No alarm when temperature out of range in Manual Mode patient safety
Man.Temp Alarm feature not functioning to alert when baby temperature exceeds set limits
Ranked causes
- 1. Man.Temp Alarm not activated in menu (common)
- 2. Temperature probe not connected (common)
- 3. Set point not entered in Baby Mode (occasional)
Verify probe is plugged in
- Test point
- Temperature probe jack on controller
- Expected
- Probe firmly inserted
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe detached during use patient safety
Temperature probe loses contact with baby's skin or disconnects from jack
Ranked causes
- 1. Probe wire pulled or snagged (occasional)
- 2. Adhesive patch degraded (common)
- 3. Probe jack connection loose (occasional)
Inspect probe adhesion to baby's skin
- Test point
- Adhesive patch on baby's abdomen or back
- Expected
- Patch fully adhered with no lifting at edges
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Skin probe does not respond to temperature change patient safety
Baby temperature probe connected but reading does not increase when warmed by hand or does not respond to temperature changes
Ranked causes
- 1. Probe not properly connected to jack (common)
- 2. Probe sensor failure or damage (occasional)
- 3. Temperature measurement circuit failure (rare)
Connect baby temperature probe
- Test point
- Temperature probe jack on right side of bed
- Expected
- Probe fully inserted and secured in jack
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Skin temperature probe placement incorrect patient safety
Temperature probe positioned between baby and mattress or not in radiant heat path
Ranked causes
- 1. Probe placed between baby and mattress (occasional)
- 2. Probe not in direct path of radiant heat (occasional)
- 3. Probe not in contact with baby's skin (occasional)
Verify probe location relative to mattress
- Test point
- Baby's skin surface
- Expected
- Probe directly on skin, not between baby and mattress
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tank circuit leaks caution
Gas escaping from tank cylinder connections or fittings
Ranked causes
- 1. Defective or improperly installed tank cylinder gaskets (common)
- 2. Loose gauge compression fittings (common)
- 3. Leaks elsewhere in system (occasional)
Inspect cylinder gaskets condition and installation
- Test point
- Tank cylinder gaskets
- Expected
- Gaskets in good condition, properly seated
- Tools
- Visual inspection
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms do not trigger in Baby Mode patient safety
Visual and/or audio alarms fail to sound or display when skin probe is disconnected and warmer is in Baby Mode
Ranked causes
- 1. Alarm circuit or speaker failure (occasional)
- 2. Baby Mode not properly entered (occasional)
Disconnect skin temperature probe
- Test point
- Temperature probe jack on right side of bed
- Expected
- Probe fully disconnected from jack
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ancillary equipment not secure or not functional patient safety
Mounted ancillary equipment loose, unstable, or not operating according to its own manual specifications
Ranked causes
- 1. Mounting bracket or fasteners loose or missing (occasional)
- 2. Ancillary equipment malfunction (occasional)
Inspect mounting of all ancillary equipment
- Test point
- All mounted ancillary devices and brackets
- Expected
- All equipment securely fastened and does not move when pushed
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ancillary equipment not securely mounted or not functioning caution
Auxiliary devices loose, rattling, or not operating per their respective manuals
Ranked causes
- 1. Ancillary equipment mounting brackets loose or damaged (unknown)
- 2. Ancillary equipment malfunction per its own operation manual (unknown)
Visually inspect all ancillary equipment mounting
- Test point
- All mounting brackets and fasteners
- Expected
- All equipment secured firmly; no rattling when unit moved
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baby Mode alarms do not trigger when probe unplugged patient safety
Unplugging skin probe while in Baby Mode does not trigger visual and audio alarms as expected
Ranked causes
- 1. Alarm circuit failure or disabled (occasional)
- 2. Warmer not in Baby Mode (common)
Switch warmer to Baby Mode
- Test point
- Mode selector on control panel
- Expected
- Warmer in Baby Mode as indicated on display
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baby temperature elevated compared to independent measurement patient safety
Temperature at remote sites (axilla, rectum, ear) measured up to 2.4°C higher than abdominal skin temperature; warmer skin probe reading matches Set Temp with no alarm; signs of heat stress including skin redness, warmth to touch, diaphoresis, or increased heart rate
Ranked causes
- 1. Warmer temperature probe positioned at abdominal site while core/remote temperature is higher (common)
- 2. Extended operation in manual mode at 100% power (occasional)
Observe patient for signs of heat stress
- Test point
- Patient skin and vital signs
- Expected
- No skin redness, warmth, diaphoresis, or increased heart rate
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baby temperature reading does not increase when skin probe warmed
Temperature probe connected but does not respond to warming by hand
Ranked causes
- 1. Skin probe not connected or connection loose (common)
- 2. Skin probe defective or damaged (occasional)
Verify skin probe connected to temperature jack
- Test point
- Temperature probe jack on right side of bed or above in-bed scale jack
- Expected
- Probe connector fully seated in jack
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed elevation mechanism does not operate smoothly patient safety
Elevating bed does not raise or lower smoothly through full travel range, or pedals do not control height
Ranked causes
- 1. Elevation mechanism actuator or motor faulty, or linkage obstructed (unknown)
- 2. Height adjustment pedals damaged or disconnected (unknown)
Operate bed elevation mechanism through full travel range
- Test point
- Bed frame and elevation mechanism
- Expected
- Bed raises and lowers smoothly without binding or grinding sounds
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed elevation mechanism does not operate smoothly patient safety
Elevating bed raises or lowers unevenly, sticks, or does not respond to pedal controls across full travel range
Ranked causes
- 1. Hydraulic or electric actuator binding or misaligned (occasional)
- 2. Height adjustment pedals misaligned or broken (occasional)
Raise bed through entire travel range
- Test point
- Bed elevation mechanism
- Expected
- Bed raises smoothly and evenly without sticking
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed elevation mechanism inoperative or rough patient safety
Elevating bed does not raise or lower smoothly, or height adjustment pedals do not function
Ranked causes
- 1. Hydraulic or motor mechanism seized or low fluid (occasional)
- 2. Height adjustment pedals damaged or disconnected (occasional)
Raise and lower bed along entire travel range
- Test point
- Bed elevation throughout full range of motion
- Expected
- Bed moves smoothly up and down, no binding or jerking
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed tilt mechanism not functioning patient safety
Tilt control does not operate smoothly, bed does not raise, or does not hold position at any angle
Ranked causes
- 1. Tilt control actuator or linkage worn or obstructed (unknown)
- 2. Bubble levels indicate mattress not level (unknown)
Squeeze tilt control and push down on bed foot
- Test point
- Tilt control lever and bed foot area
- Expected
- Head of bed raises easily without binding
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bedside panel removed or lowered without proper technique caution
Operator raises, lowers, or removes bedside panels using incorrect procedure
Ranked causes
- 1. Incorrect panel lowering technique applied (occasional)
- 2. Incorrect panel removal technique applied (occasional)
Verify correct panel lowering technique
- Test point
- Bedside panel
- Expected
- Panel pulled upward first, then top edge pulled away from bed
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bedside panel will not lock in position caution
Bedside panel does not engage securely in latched position or fails to stay raised
Ranked causes
- 1. Panel not fully swung to upright (common)
- 2. Latch mechanism obstructed or damaged (occasional)
Attempt to raise bedside panel to upright
- Test point
- Bedside panel pivot and latch
- Expected
- Panel swings freely and engages automatically in upright position
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Suction regulator patient safety
Adjustment of the suction regulator pressure to specified ranges depending on unit type (integrated or stand-alone resuscitation units).
Ranked causes
- 1. Suction regulator requires pressure calibration (common)
- 2. Suction checkout procedures required after adjustment (common)
Measure suction vacuum for integrated unit
- Test point
- Test Port SR-Blue
- Expected
- 124-152 kPa (18-22 psi), ±14 kPa
- Tools
- Vacuum measurement device
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Suction valve zero patient safety
Adjustment of the suction control knob to ensure zero position precisely stops suction flow with the knob positioned CCW against the mechanical stop.
Ranked causes
- 1. Suction valve zero position requires calibration (common)
- 2. High flow measurement device connection and air measurement setup (unknown)
Measure suction flow at CW rotation point
- Test point
- Suction barb outlet
- Expected
- Flow detected
- Tools
- High flow measurement device set to Air
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Confirm Probe Jack Connection alarm patient safety
Alarm activates when temperature probe jack is unplugged or disconnected from warmer in Baby Mode or Manual Mode (when Manual Temperature Alarm is set).
Ranked causes
- 1. Temperature probe jack is unplugged or disconnected (common)
- 2. Defective temperature probe (occasional)
Verify probe jack connection
- Test point
- Temperature probe jack connector on warmer
- Expected
- Probe should be fully inserted into jack; no loose connections visible
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Displays or indicators do not light on startup
After power on, displays are dark, LEDs do not illuminate, or prompt tone does not begin
Ranked causes
- 1. Unit not connected to rated power supply (common)
- 2. Mains switch or standby switch not activated (common)
Verify power cord connection
- Test point
- Power cord inlet on unit and wall outlet
- Expected
- Cord fully inserted into unit outlet and connected to hospital grade wall outlet
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Elevated body temperature in remote measurement sites patient safety
Temperature at axilla, rectum, or ear measured up to 2.4°C higher than abdominal skin temperature; warmer Set Temp matches skin probe reading with no alarm; signs of heat stress including skin redness, warmth, diaphoresis, or increased heart rate
Ranked causes
- 1. Warmer temperature probe reading does not reflect true core temperature; abdominal skin probe in thermal gradient (common)
- 2. Improper admission procedure or extended 100% manual power operation (occasional)
Observe patient for signs of heat stress
- Test point
- Patient observation
- Expected
- No skin redness, warmth, diaphoresis, or increased heart rate
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas supply hissing or leaking from tank connection high voltage
Hissing or leaking sounds heard when opening tank valve, indicating improper seal at yoke
Ranked causes
- 1. Gasket missing, damaged, or improperly placed (common)
- 2. Multiple gaskets used or gasket not seated properly (occasional)
Visually inspect gasket before use
- Test point
- Sealing gasket at valve-to-regulator connection
- Expected
- Gasket free of permanent deformation, splits, cracks, chips, frays, deformations, gouges, dirt, oil, grease, or contaminants
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Optional equipment not securely mounted or not operational caution
Optional equipment is loose, missing, or does not function according to its operation manual
Ranked causes
- 1. Optional equipment mounting loose or missing fasteners (occasional)
- 2. Optional equipment malfunction (occasional)
Visually inspect optional equipment mounting
- Test point
- All optional equipment fasteners and brackets
- Expected
- All fasteners tight; no loose or missing components
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Skin temperature probe reflective foil removed patient safety
Reflective patch missing or removed from disposable or reusable probe
Ranked causes
- 1. Reflective foil removed or missing from probe (rare)
- 2. Reflective patch degraded during probe repositioning (occasional)
Inspect probe for reflective foil or patch
- Test point
- Probe tip and adhesive side
- Expected
- Reflective foil present on disposable probe or reusable probe has heat reflecting patch attached
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Suction or gas supply systems leak patient safety
When setup, suction or gas supply systems attached to warmer have detectable leaks
Ranked causes
- 1. Connection loose or fitting corroded (common)
- 2. Supply line damaged or tubing cracked (occasional)
Perform leak check on suction and gas supply systems per respective manuals
- Test point
- All suction and gas supply connections
- Expected
- No visible leaks or pressure drop when systems are activated
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Suction or gas supply systems leaking or not connected patient safety
Suction or gas supply systems not properly connected to warmer or leaking during functional test
Ranked causes
- 1. Supply system not connected to warmer ports (common)
- 2. Leaking connection or tube damage (occasional)
Connect suction and gas supply systems per manuals
- Test point
- Suction and gas supply connection ports on warmer
- Expected
- All required systems connected and secured
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Suction system setup and pressure adjustment caution
Venturi suction system requires configuration with canister, tubing, and pressure control before use in resuscitation procedures
Ranked causes
- 1. Suction system not properly configured (common)
- 2. Oxygen tank depletion from suction switch left on (occasional)
Verify suction canister installation
- Test point
- Dovetail rail connection
- Expected
- Clean canister properly installed and attached to tubing
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Supply Pressure alarm sounds patient safety
Alarm whistle sounds from air oxygen blender when gas supply pressure is inadequate while medical gas switch is on
Ranked causes
- 1. Air or oxygen tank pressure depleted (common)
- 2. Wall gas supply pressure inadequate (occasional)
Observe and verify Supply Pressure alarm condition
- Test point
- Air oxygen blender
- Expected
- Alarm whistle sounds when at least one gas supply is on, pressure is inadequate, and medical gas switch is on
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Supply pressure alarm sounds from air/oxygen blender patient safety
Alarm whistle sounds from the air oxygen blender when gas supply pressure is inadequate
Ranked causes
- 1. Air or oxygen tank pressure depleted (common)
- 2. Inadequate wall gas supply pressure (occasional)
Check tank pressure gauges
- Test point
- Air and O2 tank pressure gauges on back of warmer
- Expected
- Adequate pressure for both air and oxygen tanks
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Supply pressure alarm sounds from air/oxygen blender patient safety
Alarm whistle sounds from the air oxygen blender when at least one gas supply is connected, medical gas switch is on, and there is inadequate pressure from either air or oxygen source.
Ranked causes
- 1. Air or oxygen tank pressure depleted (common)
- 2. Wall gas supply pressure inadequate (occasional)
Verify at least one gas supply is connected and medical gas switch is on
- Test point
- Medical gas switch
- Expected
- ON position
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tank pressure gauge does not register
Pressure gauge on front panel does not show adequate gas supply after opening tank valve
Ranked causes
- 1. Tank valve not opened or opened incorrectly (common)
- 2. Empty or low-pressure tank installed (occasional)
Open tank valve slowly using wrench
- Test point
- Tank valve
- Expected
- Valve turns counter-clockwise without resistance; opens fully
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wall gas supply connection caution
Connection of wall-supplied medical gas sources to resuscitation system inlet ports
Ranked causes
- 1. Wall gas source requires proper connection technique (common)
- 2. Tank valves must be closed when using wall gas (common)
Verify wall gas hose connections are secure
- Test point
- Indexed inlet port connections on back of unit
- Expected
- All wall source hose connectors firmly seated at indexed inlet ports without gaps
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm does not silence using alarm silence key or Hands Free method patient safety
Triggered alarm cannot be silenced by pressing alarm silence key or using Hands Free Alarm silence
Ranked causes
- 1. Alarm silence key or circuit defective (unknown)
Press alarm silence key
- Test point
- Alarm silence key on control panel
- Expected
- Alarm tone stops; visual alarm may remain
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baby Cold - Check Temp Probe alarm patient safety
Alarm activates when baby's skin temperature drops below set temperature. Low priority alarm (blinking red light, intermittent tone, 12 min silence) at >1.0°C below set temp; high priority alarm (continuous tone, 5 min silence) at >2.0°C below set temp.
Ranked causes
- 1. Temperature probe improperly attached to baby's skin (common)
Verify baby's skin temperature reading
- Test point
- Temperature display on warmer screen
- Expected
- Should not be >1.0°C below set temperature (low priority) or >2.0°C below set temperature (high priority)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baby Hot - Check Temp Probe alarm patient safety
Alarm activates when baby's skin temperature exceeds set temperature. Low priority alarm (blinking red light, intermittent tone, 12 min silence) at >1.0°C above set temp; high priority alarm (continuous tone, 5 min silence) at >2.0°C above set temp.
Ranked causes
- 1. Temperature probe improperly attached to baby's skin (common)
Verify baby's skin temperature reading
- Test point
- Temperature display on warmer screen
- Expected
- Should not be >1.0°C above set temperature (low priority) or >2.0°C above set temperature (high priority)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baby placed in bed during warmup mode patient safety
Operator places baby on warmer bed before warmup is complete
Ranked causes
- 1. Operator does not wait for warmup completion (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Baby temperature too high
- Baby temperature too high
- Baby temperature too low
- Baby temperature too low
- Bed tilt mechanism does not operate smoothly
- Bed tilted from horizontal position
- Bedside panel from other warmer model used
- Bedside panels do not lock securely
- Bedside panels do not lock securely
- Calibration: Airway manometer accuracy
- Calibration: Fixed Pressure Controlled Relief Valve - Red Circuit
- Calibration: Gas System Regulator
- Calibration: PIP Valve Zero
- Calibration: Primary flow valve
- Calibration: Primary flow valve accuracy
- Calibration: Secondary Flow
- Calibration: Suction gauge accuracy
- Calibration: Suction Relief Valve
- Casters not in firm contact or unit unstable
- Check Baby - Heat Off alarm
- Check Baby - Heat Off alarm activated
- Check Baby alarm
- Check Baby alarm activated
- Check Baby alarm activated
- Check Baby-Heat Off alarm activated
- Disposable probe reused or cleaned
- Electrical shock hazard during servicing
- Excessive temperature elevation in Baby Mode during admission
- Gas leaks during check valve test
- Heater power drops after seven-minute warmup
- High pressure relief valve fails test
- High pressure relief valve leaking
- High pressure relief valve malfunction
- High pressure relief valve not opening at correct pressure
- High pressure relief valve opens at incorrect pressure
- Improper mode during patient placement
- Low pressure relief valve does not open at correct pressure
- Low pressure relief valve fails test
- Manual temperature alarm not activated in Manual Mode
- Non-Ohmeda temperature probe used
- Patient temperature not adjusted in warmup mode
- Plug retaining brackets missing or loose
- Power cord damaged
- Power cord damaged
- Power outlet brackets missing or loose
- Primary flow valve accuracy verification required
- Probe not removed from storage bag before use
- Probe placement obscured by phototherapy lamp shield
- Probe wire damaged during removal
- Prompt tone does not begin on power-up
- Prompt tone does not silence in Manual Mode
- Rectal temperature used for heater control
- Resuscitation system does not supply accurate FiO2 or adequate pressure
- Resuscitation system fails preoperational checkout
- Resuscitation system FiO2 or pressure inaccurate during Supply Pressure alarm
- Scale overload alarm activated
- Suction or gas supply systems leaking or not functioning
- Suction system checkout
- Suction system setup
- System Failure alarm
- System failure alarm activated
- System Failure alarm activated
- Tank connection leakage after opening valve
- Tank installation improper gasket placement
- Tank installation preparation
- Tank valve opening procedure
- Temperature probe disconnected
- Temperature Probe Failure alarm
- Temperature Probe Failure alarm activated
- Temperature probe failure detected
- Unapproved equipment connected to accessory outlets
- Unit enters power fail alarm condition
- Unit has damaged or missing parts
- Unit has damaged or missing parts
- Unit not fully powered down with standby switch
- Unit not isolated from mains at standby
- Unit not isolated from mains with standby switch
- Unit shuts down after prolonged power loss
- Unit unstable or casters not functioning
- Visual or audio alarms do not trigger in Baby Mode with probe unplugged
- Weight exceeds scale maximum
- Weight on scale above maximum alarm
82 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Baby Cold - Check Temp Probe | Baby's skin temperature falls below set temperature threshold | Baby Mode: temperature >1.0°C below set point; escalates at >2.0°C. Manual Mode: when Man Temp Alarm configured. | warning |
Baby Cold - Check Temp Probe | Baby's skin temperature drops below set temperature threshold | Baby mode: skin temp >1.0°C (or >0.5°C if configured) below set temperature for low priority; >2.0°C below set temperature for high priority | warning |
Baby Hot - Check Temp Probe | Baby's skin temperature exceeds set temperature threshold | Baby Mode: temperature >1.0°C above set point; escalates at >2.0°C. Manual Mode: when Man Temp Alarm configured. | warning |
Baby Hot - Check Temp Probe | Baby's skin temperature exceeds set temperature threshold | Baby mode: skin temp >1.0°C (or >0.5°C if configured) above set temperature for low priority; >2.0°C above set temperature for high priority | warning |
Check Baby | Radiant heater has been operating at high power for extended period without temperature regulation | Baby mode: heater at 100% power for >12 minutes; Manual mode: heater >25% power (or configured limit) for >12 minutes | warning |
Check Baby | Heater sustained high power operation; baby monitoring required | Baby Mode: heater at 100% power for >12 minutes. Manual Mode: heater above configured threshold for >12 minutes. | warning |
Check Baby - Heat Off | Extended high-power operation unresponded to; heater automatically disabled | 100% power for >15 min with Check Baby alarm unanswered for 3 minutes (Baby Mode) or threshold power >15 min unresponded (Manual Mode) | critical |
Check Baby - Heat Off | Heater automatically shut off as safety measure after prolonged high-power operation with ignored warning | Heater at 100% (Baby mode) or >25% (Manual mode) for >15 minutes with unanswered Check Baby alarm for 3 minutes | error |
Check Baby - Heat Off | Heater shut off due to sustained high-power operation with unresolved Check Baby alarm | Baby Mode: 100% heater power >15 min with unanswered Check Baby alarm >3 min. Manual Mode: configured power threshold >15 min with unanswered Check Baby >3 min. | error |
Confirm Probe Jack Connection | Temperature probe is not properly connected to warmer | Probe jack is unplugged in Baby mode; probe missing in Manual mode when Manual Temperature Alarm is set | warning |
Confirm Probe Jack Connection | Temperature probe jack is disconnected or not fully inserted | Baby Mode: probe jack unplugged. Manual Mode: probe jack unplugged when Man Temp Alarm is set. | warning |
Scale Alarm | Weight on scale exceeds maximum safe limit | Total weight on warming bed scale exceeds 8 kilograms (17.6 pounds). | warning |
Scale overload | Weight on integrated scale exceeds safe operating capacity | Weight on scale above 8 kilograms (17.6 lb) | warning |
Supply Pressure Alarm | Inadequate pressure from air or oxygen gas supply to the system | At least one gas supply is supplying gas, there is inadequate pressure from either air or oxygen source, and medical gas switch is on | warning |
Supply Pressure Alarm | Air or oxygen supply pressure is inadequate | At least one gas supply connected, medical gas switch on, and inadequate pressure from either air or oxygen source | warning |
System Failure | Critical electrical system failure; heater shut down automatically | Electrical fault detected in warmer control system. | critical |
System Failure | Internal electrical failure detected by warmer diagnostics | Warmer detects internal electrical malfunction during operation | critical |
Temp Probe Failure | Temperature probe internal thermistor mismatch detected | Two thermistors in probe differ by 0.5°C or more for over 6 minutes. | error |
Temp Probe Failure | Temperature probe sensors have diverged, indicating sensor failure | Difference between two thermistors in probe exceeds 0.5°C for >6 minutes | error |
Temp Probe Failure | Temperature probe internal thermistor mismatch detected | Difference between two thermistors in probe exceeds 0.5°C for more than 6 minutes | error |
Weight on scale above maximum | Scale is overloaded with weight exceeding maximum capacity | Weight on scale platform exceeds 8 kilograms (17.6 lb) | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI