GE Healthcare OEC C-ARM: troubleshooting and error codes
The ServiceForge knowledge base holds 9 documented symptoms for the GE Healthcare OEC C-ARM, with 26 ranked causes, 0 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Vertical column movement button does not produce movement caution
Button press does not cause vertical column position change or operation is intermittent
Ranked causes
- 1. Faulty button (common)
- 2. Faulty wiring between button and Power/Motor Relay PCB (common)
- 3. Loose or broken connection between relay boards (common)
- 4. Fuse F11 blown (occasional)
- 5. +24 VDC not available to Power/Motor Relay PCB (occasional)
- 6. 115 VAC not available to Power/Motor Relay PCB (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fast stop button does not halt column movement patient safety
Vertical column continues to move when fast stop button is pressed
Ranked causes
- 1. Faulty Fast Stop button (common)
- 2. Faulty wiring or connector between button and P4 on Power/Motor Relay PCB (occasional)
- 3. Faulty Power/Motor Relay PCB (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fast Stop button does not halt Vertical Column movement patient safety
Pressing Fast Stop button during vertical movement does not stop the column or does not lock out opposite direction button
Ranked causes
- 1. Fast Stop button electrical failure or loose connection (common)
- 2. Control logic or solenoid malfunction (occasional)
- 3. Motor brake failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vertical Column does not extend to maximum height patient safety
Vertical Column fails to reach full extension or does not extend at all when Up button is pressed
Ranked causes
- 1. Motor malfunction or inadequate power (occasional)
- 2. Limit switch failure or misalignment (occasional)
- 3. Mechanical obstruction or binding in column (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vertical Column does not retract to minimum height patient safety
Vertical Column fails to reach full retraction or does not lower at all when Down button is pressed
Ranked causes
- 1. Motor malfunction or inadequate power (occasional)
- 2. Limit switch failure or misalignment (occasional)
- 3. Mechanical obstruction or binding in column (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vertical Column movement is not smooth or is noisy caution
Vertical Column exhibits jerky, grinding, or audible noise during extension or retraction
Ranked causes
- 1. Insufficient lubrication (common)
- 2. Column bearings worn or damaged (occasional)
- 3. Mechanical contamination or debris in column (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Column movement stops before reaching maximum or minimum height caution
Vertical column reaches intermediate position and stops instead of traveling full range
Ranked causes
- 1. Obstacle in C-Arm's path (common)
- 2. Faulty Vertical Column (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Opposite direction button remains active after Fast Stop patient safety
After pressing Fast Stop button, the opposite direction control button continues to function instead of being locked out
Ranked causes
- 1. Control logic interlock failure (occasional)
- 2. Button contact malfunction or debounce issue (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Column movement is noisy or rough caution
Vertical column produces noise or operates with jerky/rough motion
Ranked causes
- 1. Faulty Vertical Column (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI