GE Healthcare Mobile DaRt: troubleshooting and error codes

The ServiceForge knowledge base holds 2 documented symptoms for the GE Healthcare Mobile DaRt, with 3 ranked causes, 4 diagnostic checks with test points and expected values, and 1 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Tube arm erratic detection when parked caution

Intermittent high/low speed driving transitions when system encounters bumps; occurs more frequently on systems driven longer distances; microswitch SW4 bracket creeps to edge of detection range

Ranked causes

  1. 1. Microswitch SW4 bracket misalignment or creep (occasional)
  2. 2. Tube arm not fully contracted horizontally when parking attempted (rare)
First check

Verify microswitch displacement before and after adjustment

Test point
SW4 microswitch button
Expected
Displacement stops just before button fully depressed when tube arm fully inserted
Tools
Visual inspection

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Parking solenoid/hook not releasing patient safety

Hook latch fails to release or gets stuck at midpoint when parking/unparking tube arm; may cause D11 error if stuck at high position; tube arm can bounce out of parking position

Ranked causes

  1. 1. Solenoid pin misalignment (Evolution only) (rare)
First check

Observe lever position on back of parking assembly cover

Test point
Parking assembly cover rear
Expected
Lever backs up completely when lock release button pressed; returns fully to original position

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
D11Tube arm parking position errorHook latch stuck at high position while tube arm power on; stuck midpoint prevents full stroke releaseerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.