GE Healthcare MAC VU360: troubleshooting and error codes
The ServiceForge knowledge base holds 46 documented symptoms for the GE Healthcare MAC VU360, with 112 ranked causes, 111 diagnostic checks with test points and expected values, and 29 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Software upgrade errors
Upgrade fails with error messages indicating digital signature validation, invalid image, system errors, or file access issues
Ranked causes
- 1. Software image not digitally signed or corrupted (common)
- 2. Invalid software image in USB or shared path (common)
- 3. System error during upgrade (occasional)
- 4. Host software image not digitally signed (rare)
- 5. Software image not present in USB path (common)
- 6. Insufficient file permissions on USB drive files (occasional)
- 7. USB flash drive not detected during upgrade (common)
- 8. Software image not found in USB root directory (common)
- 9. Access to USB flash drive denied in settings (occasional)
- 10. Shared network folder disabled or not configured (common)
- 11. Shared path not configured in device (common)
- 12. Software image not present in shared network path (common)
- 13. Device not connected to network (common)
- 14. Digital signature file not in network shared path (occasional)
- 15. Software image extraction to inactive partition failed (occasional)
15 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
WLAN wireless communication not established or device not connecting
Device unable to establish wireless network connection; WiFi connectivity failure
Ranked causes
- 1. Wireless not configured on device (common)
- 2. Wireless network configuration incorrect in Settings (common)
- 3. Wireless antenna damaged (non-FIPS) (occasional)
- 4. Wireless antenna loosely connected to expansion board (non-FIPS) (occasional)
- 5. Expansion board malfunction (non-FIPS) (occasional)
- 6. Wireless antenna damaged (FIPS) (occasional)
- 7. Wireless antenna loosely connected to FIPS wireless module (occasional)
- 8. FIPS wireless module malfunction (FIPS) (occasional)
- 9. Expansion board malfunction (FIPS) (rare)
Verify Wireless is configured on device
- Test point
- Wireless configuration
- Expected
- Wireless enabled and configured
2 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
USB port not working or not detecting devices
USB flash drive not detected or USB ports not functional for file transfer or upgrades
Ranked causes
- 1. External USB storage option disabled in settings (common)
- 2. All USB ports disabled in hardware settings (common)
- 3. Physical damage to USB port connectors (occasional)
- 4. USB flash drive inserted incorrectly (common)
- 5. USB flash drive defective (occasional)
- 6. Expansion board malfunction (occasional)
- 7. CPU board malfunction (rare)
Verify Enable External USB Storage option enabled
- Test point
- Settings > System > Storage
- Expected
- External USB Storage option enabled
2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG data noise unacceptable levels
Excess noise visible on ECG recordings from environmental interference, improper patient prep, or equipment failure
Ranked causes
- 1. Environmental electromagnetic interference from radios, cell phones, or medical equipment (common)
- 2. Inadequate patient skin preparation or improper electrode placement (common)
- 3. Patient movement during ECG acquisition (common)
- 4. Acquisition module or board malfunction (occasional)
- 5. Defective, broken, or disconnected leadwires (common)
- 6. Defective or expired electrodes (common)
Use Hookup Advisor to monitor ECG signal quality and identify noise source
- Test point
- ECG signal
- Expected
- Clean signal without artifacts, baseline shift, power line interference, baseline wander, muscle artifacts, or electrode noise
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
LDAP server connection error or authentication failure
Unable to connect to LDAP server or login credentials invalid during LDAP configuration
Ranked causes
- 1. LDAP server does not exist or IP address/server name incorrect (common)
- 2. LDAP server down or connection fails for unknown reason (occasional)
- 3. Network connectivity issue causing connection timeout (occasional)
- 4. Distinguished name does not exist in LDAP server (occasional)
- 5. User login credentials invalid (common)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touchscreen not responding to touch input
Touch gestures not sensed on touchscreen; may be localized to specific area or entire screen
Ranked causes
- 1. Cracked or physically damaged touchscreen (common)
- 2. Faulty touchscreen (no visible damage) (occasional)
- 3. Faulty display to expansion board cable (occasional)
- 4. Faulty expansion board (occasional)
- 5. Faulty CPU board (rare)
Visually inspect touchscreen for physical damage
- Test point
- Touchscreen surface
- Expected
- No cracks or visible damage
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard keys not working
Some or all keyboard keys non-functional or no response to key presses
Ranked causes
- 1. Faulty keyboard (common)
- 2. Keyboard membrane physically damaged or deteriorated (common)
- 3. Keyboard to printer PWA cable defective or disconnected (occasional)
- 4. Printer PWA failure (occasional)
Perform Keyboard Diagnostics Test
- Test point
- Keyboard input
- Expected
- All keys responsive
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
LAN communication not established or device not connecting to network
Device unable to establish wired network connection; Ethernet connectivity failure
Ranked causes
- 1. Wired network configuration incorrect in Settings (common)
- 2. LAN cable improperly inserted or disconnected from RJ45 connector (common)
- 3. Expansion board malfunction or defective (occasional)
- 4. CPU board malfunction or defective (rare)
Verify wired network configuration in Settings
- Test point
- Settings screen network config
- Expected
- Correct network configuration
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
WLAN diagnostics fails
WLAN test fails at one or more stages: hardware, software, connectivity, or IP configuration
Ranked causes
- 1. Hardware failure - WLAN hardware not available (unknown)
- 2. Software failure - wireless drivers or interface issues (unknown)
- 3. Connectivity failure - network or supplicant issues (unknown)
- 4. IP configuration failure (unknown)
Verify wireless option enabled
- Test point
- Option Manager
- Expected
- FIPS or WRLS option enabled
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Acquisition Module not communicating or disconnected
Acquisition Module error message displayed on host; LEDs not glowing; buttons not working
Ranked causes
- 1. Loose or damaged Acquisition Module cable connection (common)
- 2. Faulty LEDs or Acquisition Module (occasional)
- 3. Faulty Acquisition Module buttons (occasional)
Inspect Acquisition Module cable connection to host
- Test point
- Acquisition Module cable connector
- Expected
- Cable firmly connected with no visible damage
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot export diagnostic test results to USB drive
Export of diagnostic test results to USB flash drive fails with error message indicating insufficient space or file format corruption.
Ranked causes
- 1. Insufficient space on USB flash drive (common)
- 2. USB flash drive is corrupted or unsupported format (occasional)
- 3. Diagnostic file is corrupted (rare)
Verify available space on USB flash drive
- Test point
- USB flash drive storage status
- Expected
- Sufficient free space available for diagnostic report file
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device not receiving AC power caution
Device fails to power on or charge battery when connected to AC outlet; AC Power LED not illuminated
Ranked causes
- 1. AC power cord not properly connected to device or outlet (common)
- 2. AC power line voltage or frequency does not match unit requirements (occasional)
- 3. Damaged protective earth conductor integrity (rare)
Verify AC power cord connections
- Test point
- AC power connector on device back panel and AC outlet
- Expected
- Female connector fully seated on device; male plug fully inserted in outlet
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software upgrade or downgrade fails
Error message displayed during software upgrade/downgrade process; device may not complete the update or reboot properly.
Ranked causes
- 1. Incorrect software image files or corrupted upgrade package (common)
- 2. Acquisition Module disconnected during upgrade (occasional)
- 3. Device not properly configured for upgrade source (occasional)
Verify all required software image files are present and correct
- Test point
- USB flash drive or shared network folder root directory
- Expected
- Three files present: <sw-filename>.tar.bz2.sign, <sw-filename>.tar.bz2, u-boot.imx
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Acquisition Module barcode reader not scanning
Barcode reader fails to scan barcodes presented to Acquisition Module
Ranked causes
- 1. Loose or damaged Acquisition Module cable connection (common)
- 2. Faulty Acquisition Module barcode reader (occasional)
Inspect Acquisition Module cable connection to host
- Test point
- Acquisition Module cable connector
- Expected
- Cable firmly connected with no visible damage
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Acquisition module software update fails or does not start
Acquisition module software upgrade does not complete successfully; error message may display or upgrade does not initiate automatically on reboot.
Ranked causes
- 1. Acquisition Module disconnected or communication lost during upgrade (common)
- 2. Persistent acquisition module software upgrade failure (rare)
Verify Acquisition Module is physically connected to device
- Test point
- Device physical connection ports
- Expected
- Acquisition Module securely connected via data/power interface
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode reader not scanning or reading barcode correctly
External barcode reader fails to scan barcode displayed on screen; scanned information does not appear on screen.
Ranked causes
- 1. Barcode reader not compatible or not in correct mode; incompatible symbology (occasional)
- 2. Barcode reader hardware malfunction (rare)
Scan barcode displayed on Barcode Diagnostics screen
- Test point
- Barcode on screen; barcode reader
- Expected
- Scanned information appears on screen as text; test result displays Success
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not charging properly caution
Battery charge does not increase when connected to AC power; discharge rate is very low or nonexistent.
Ranked causes
- 1. Battery or charging circuit fault; batteries not fully discharged before test (common)
- 2. Faulty battery pack (occasional)
Verify battery discharge level before charging test
- Test point
- Battery icon on status bar
- Expected
- Battery charge less than 3%
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
KISS pump not responding during diagnostics
KISS pump does not respond or communicate properly when diagnostics are performed after connecting pump to device.
Ranked causes
- 1. KISS pump not connected or powered off (common)
- 2. KISS pump malfunction (rare)
Verify KISS pump connection and power state
- Test point
- KISS pump connector and diagnostics option switch
- Expected
- KISS pump physically connected and diagnostics option turned on to enable power supply
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
LAN diagnostics fails
Network connectivity test does not complete successfully or reports hardware/software failures
Ranked causes
- 1. LAN adapter not available (occasional)
- 2. LAN interface disabled or not found (occasional)
Verify LAN cable is plugged in properly
- Test point
- Network port connector
- Expected
- Cable firmly connected
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper not advancing in printer
Paper jam, misalignment, or failure to advance first sheet of printed ECG
Ranked causes
- 1. Paper inserted with incorrect orientation (Q holes or marks on wrong side) (common)
- 2. Paper spacers not aligned correctly for paper size (occasional)
Inspect paper orientation in tray
- Test point
- Paper position in paper tray
- Expected
- Paper with Q holes on top left side (or Q marks on bottom left side)
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
USB port does not detect connected USB flash drive
USB port fails to detect USB flash drive inserted into slot; port icon remains grey instead of changing to blue.
Ranked causes
- 1. USB port malfunction or disabled (unknown)
- 2. USB flash drive not inserted properly (occasional)
Insert USB flash drive into USB port and observe port icon
- Test point
- USB port slot
- Expected
- Port icon changes from grey to blue and test result displays Success
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC adapter LED on keyboard not responding to power connection
AC adapter LED does not illuminate when device connected to AC power or does not turn off when device disconnected from AC power.
Ranked causes
- 1. AC adapter LED malfunction (unknown)
Insert battery and connect to AC power
- Test point
- AC adapter LED on keyboard
- Expected
- AC adapter LED is on
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Acquisition Module LEDs not functioning
One or more LEDs in Acquisition Module do not illuminate during diagnostic test
Ranked causes
- 1. LED hardware failure (unknown)
Verify Acquisition Module connected
- Test point
- Acquisition Module connection
- Expected
- Module connected to device
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Acquisition Module pushbutton not responding
One or more buttons on Acquisition Module do not respond when pressed during diagnostic test
Ranked causes
- 1. Pushbutton hardware failure (unknown)
Verify Acquisition Module connected
- Test point
- Acquisition Module connection
- Expected
- Module connected to device
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Audio not working or audio output defective
Audio diagnostics test indicates audio malfunction; device does not produce expected sound output.
Ranked causes
- 1. Audio hardware malfunction or speaker failure; audio driver issue (unknown)
Listen for audio output on Audio Diagnostics screen
- Test point
- Device speakers
- Expected
- Audio output heard from device speakers
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode reader does not scan barcode
Barcode reader fails to scan the barcode displayed on the Acquisition Barcode Diagnostics screen; scanned information not shown on screen.
Ranked causes
- 1. Barcode reader malfunction (unknown)
Scan barcode using Acquisition Module barcode reader
- Test point
- Barcode on Acquisition Barcode Diagnostics screen
- Expected
- Scanned information displayed as text and test result shows Success
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charge percentage not increasing during charge observation caution
Battery percentage of charge remains unchanged for up to 5 minutes of observation when connected to AC power; battery current not in expected range.
Ranked causes
- 1. Battery pack faulty or fully discharged (occasional)
Verify initial battery charge percentage and current
- Test point
- Battery Metrics screen or battery icon
- Expected
- Battery charge between 5% and 80%; Battery current between 100 to 2200 mA
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery discharge current varies excessively caution
Battery current value varies more than ±10% from initial recorded value during 5 minute discharge observation; inconsistent discharge behavior.
Ranked causes
- 1. Battery pack faulty or degraded (occasional)
Record initial battery current value and monitor for 5 minutes
- Test point
- Battery Metrics screen
- Expected
- Battery current reads negative (discharging); variation not exceeding ±10% over 5 minutes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery LED on keyboard not responding to battery state
Battery LED does not illuminate when non-fully-charged battery inserted or does not turn off when fully charged battery inserted or battery removed.
Ranked causes
- 1. Battery LED malfunction (unknown)
Insert non-fully-charged battery with AC power connected
- Test point
- Battery LED on keyboard
- Expected
- Battery LED is on
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not charged before first use
Device unable to operate on battery power or unable to perform ECG acquisition due to insufficient battery charge at setup
Ranked causes
- 1. Battery inserted with minimum charge without AC charging (common)
Verify battery charge status before first battery-powered operation
- Test point
- Battery Charging LED on keyboard
- Expected
- LED indicates full charge (green or equivalent per device manual)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not discharging properly caution
Battery discharge rate is below minimum threshold; device may not operate on battery power for expected duration.
Ranked causes
- 1. Battery or discharge circuit fault; batteries not fully charged before test (common)
Verify battery charge level before discharge test
- Test point
- Battery icon on status bar
- Expected
- Battery charge at or near 100%
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Data loss during software upgrade or downgrade patient safety
Patient data and device settings may be permanently lost after software upgrade or downgrade operation completes.
Ranked causes
- 1. Software upgrade/downgrade process erases device memory (common)
Verify patient data backup is complete before starting upgrade
- Test point
- USB flash drive storage status
- Expected
- All patient data and device settings copied to USB flash drive
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device enabled with wrong wireless option code
Device configured with incorrect FIPS versus WRLS-Global Wireless option; Wireless Country of Operation incorrectly set
Ranked causes
- 1. Incorrect wireless option code or country setting during initial configuration (rare)
Disable wireless option in Network tab of Settings screen
- Test point
- Settings > Network > Wireless option
- Expected
- Wireless option disabled
8 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device unable to boot due to software corruption patient safety
Device will not boot; bad sectors detected; software corrupted
Ranked causes
- 1. Software corruption on device storage (rare)
Power off device
- Test point
- Device power state
- Expected
- Device powered off
11 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display brightness control not working
Display brightness does not change or backlight intensity does not vary when brightness slider moved
Ranked causes
- 1. Display brightness control malfunction (unknown)
Slide brightness control slider
- Test point
- Brightness Diagnostics screen
- Expected
- Backlight intensity changes from minimum to maximum
Display does not show all colors correctly
One or more display colors (red, blue, green, black, white) not visible or do not cover entire screen; pixel display failure suspected.
Ranked causes
- 1. Display hardware malfunction or pixel failure; backlight failure (occasional)
Verify red color display covers entire screen
- Test point
- Display screen
- Expected
- Red color visible across full screen
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display does not turn off when horizontal or does not turn on when vertical
Accelerometer malfunction causes display to remain on in horizontal position or remain off in vertical position.
Ranked causes
- 1. Accelerometer malfunction (unknown)
Position display horizontally
- Test point
- Display screen orientation
- Expected
- Display turns off when horizontal
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Improper ECG acquisition due to incorrect leadwire connection patient safety
ECG trace inaccuracies or missing leads caused by mismatched leadwires to electrodes or incorrect connector placement
Ranked causes
- 1. Leadwire not matched to correct electrode label location (common)
Inspect each leadwire connection at acquisition module
- Test point
- Leadwire cable connector (3) on Acquisition Module
- Expected
- All 14 leadwires inserted securely with audible/tactile click
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard keys not responding
Keys on integrated keyboard do not register when pressed; key outline remains gray instead of turning green with check mark on diagnostics screen.
Ranked causes
- 1. Keyboard hardware malfunction or disconnection (unknown)
Visually inspect keyboard key response during diagnostics
- Test point
- Integrated keyboard
- Expected
- Key turns green with white check mark when pressed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LAN ping test fails
Ping test to remote device fails after IP configuration tests complete
Ranked causes
- 1. IP configuration failure (unknown)
Verify IP address format
- Test point
- Enter IP Address field
- Expected
- valid IPv4 address or DNS FQDN (e.g. mydevice.mydomain.myorg.com)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power LED on keyboard not illuminated
Power LED on keyboard is not glowing when device is powered on.
Ranked causes
- 1. Power LED has failed (rare)
Verify Power LED is illuminated
- Test point
- Power LED on keyboard
- Expected
- Power LED is on by default
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer door sensor not detected
Error message does not appear when printer door is opened during writer diagnostics
Ranked causes
- 1. Printer door sensor malfunction or failure (unknown)
Verify error message appears when printer door is open
- Test point
- Diagnostic screen
- Expected
- Error message 'Printer error. Door is open.' appears
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Speaker audio test fails
Device does not emit sound at selected volume or tone during speaker diagnostics test
Ranked causes
- 1. Speaker malfunction or audio system fault (unknown)
Verify device emits sound at selected volume for selected tone
- Test point
- Speaker output
- Expected
- Sound emitted at selected tone and volume
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touchscreen gestures not working or touch input not responding
Touch input non-responsive or gestures not recognized on screen; touchscreen malfunction
Ranked causes
- 1. Touchscreen physically damaged with cracks or damage (common)
Visual inspection for physical damage
- Test point
- Touchscreen surface
- Expected
- No cracks or visible damage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touchscreen not responding to input
Touch sensor does not detect taps or swipes; square areas remain white instead of turning blue when touched.
Ranked causes
- 1. Touchscreen sensor malfunction or poor contact (unknown)
Tap or swipe touchscreen diagnostic areas
- Test point
- Touchscreen squared areas
- Expected
- Square turns blue when touched
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Writer fails to print test patterns
Printer produces no output or incorrect output during writer diagnostics tests at various speeds
Ranked causes
- 1. Writer mechanism malfunction (unknown)
Verify print output at 50 mm/sec
- Test point
- Printer output
- Expected
- 5 pages at 50 mm/sec with header, grids, line crosses, waveforms
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Connect the USB drive to upgrade software | USB drive not detected during upgrade attempt | Software upgrade via USB selected but drive not detected | error |
Copying software upgrade files unsuccessful | File copy operation failed during upgrade | During software upgrade file copy process | error |
Digital signature validation failed | Software image digitally signed validation unsuccessful | During software upgrade when image not properly signed | error |
Failed to download digital signature | Digital signature file download from network failed | Software upgrade via shared folder but signature file missing | error |
Failed to restore configuration due to incompatible database | The database schema is incompatible between software versions | Attempted to restore configuration data with incompatible software version | error |
Failed to restore configuration from PSOC | PSoC does not have backed-up configuration due to error or hardware failure | Attempted to restore configuration data from PSoC memory | error |
Host software image is not digitally signed | Host software image lacks digital signature | During software upgrade when host image not signed | error |
Image is not found in the USB path | Software image file not present on USB drive | Software upgrade via USB selected but image not on drive | error |
Insert USB drive to export logs | No USB flash drive is connected to the device | Attempt to export logs without USB flash drive inserted | warning |
Installation of software image unsuccessful | Software image extraction or installation to partition failed | Software upgrade via shared network failed to install | error |
Invalid software image | Software image file invalid or corrupted | During software upgrade when image is invalid | error |
LDAP authentication failed | LDAP login authentication credentials invalid or rejected | LDAP authentication attempt with wrong credentials | error |
LDAP Server connection timed out | LDAP server connection request timeout occurred | LDAP connection request exceeds timeout period | error |
LDAP Server Distinguished Name does not exist | Configured LDAP distinguished name not found on server | LDAP authentication with non-existent distinguished name | error |
LDAP Server failure | Connection to LDAP server failed for unknown reason | LDAP connection attempt fails unexpectedly | error |
LDAP Server unavailable | LDAP server cannot be reached or located | LDAP authentication connection attempt fails | error |
No USB device detected | USB flash drive is not inserted properly or USB port is not enabled | Attempt to export audit log or event log to USB flash drive | warning |
Save to USB failed | USB flash drive has an unknown error | Attempt to export audit log, event log, or service logs to USB flash drive | warning |
Shared network folder is disabled in Settings screen | Network shared path disabled in system settings | Software upgrade via shared folder selected but path disabled | error |
Shared network not configured | Shared network path not set up on device | Software upgrade via shared folder but path not configured | error |
Software image is not found in the USB | Software image file missing from USB root directory | Software upgrade via USB selected but image not found | error |
Software image not found in shared folder | Software image file not present in shared network path | Software upgrade via shared folder but image missing | error |
Software upgrade failed. Retry. Contact GE Service if proble | Software upgrade process failed for system error reason | System error during upgrade process | error |
Sufficient space is not available | USB flash drive does not have enough free space for export | Attempt to export logs to a full or nearly full USB flash drive | warning |
The device is not connected to the network | Network connectivity lost or not established | Software upgrade via shared folder but no network connection | error |
USB access denied | Access to USB flash drive blocked or denied | Software upgrade via USB selected but storage access denied | error |
USB access denied | External USB storage or USB port access is not enabled in device settings | Attempt to export logs to USB flash drive | error |
USB unknown error | Unknown or unspecified USB error occurred during export | Attempt to export logs to USB flash drive | error |
USB write-protected | USB flash drive has write protection enabled | Attempt to export logs to a write-protected USB flash drive | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.