GE Healthcare Mac-Lab: troubleshooting and error codes
The ServiceForge knowledge base holds 44 documented symptoms for the GE Healthcare Mac-Lab, with 88 ranked causes, 89 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Domain security policy not confirmed on Acquisition systems
Administrator group members (INW\Domain Admins, INW\[withheld], INW\[withheld]) missing from Administrators group on Acquisition system
Ranked causes
- 1. Acquisition system not in correct domain or domain policy not applied (common)
- 2. Domain group policy not updated on Acquisition system (common)
- 3. TCP/IP properties misconfigured on server or Acquisition system (occasional)
- 4. INW server and Acquisition system cannot communicate (occasional)
- 5. DNS setup incorrect on Acquisition system (occasional)
- 6. Time on server and Acquisition system do not match (occasional)
- 7. Domain membership corrupted on Acquisition system (rare)
Verify Acquisition system is in correct domain
- Test point
- Computer Name tab in System Properties
- Expected
- INW Domain is present
10 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Supplies and list items from CVIS not appearing on Mac-Lab
Items added to CVIS system do not display on Mac-Lab interface
Ranked causes
- 1. INW service not configured or started (common)
- 2. CVIS List Sharing Delta export not configured for location (common)
- 3. Corrupt XML export from CVIS system (occasional)
- 4. CVIS List Sharing services not started (common)
- 5. Replication errors or issues with INW Server and Mac-Labs (occasional)
- 6. Network connection between CVIS and INW server lost (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System not ready for operation high voltage
IEB, computer, monitors, or connected equipment not powered on or properly connected; cables frayed, exposed, bent, or damaged
Ranked causes
- 1. IEB not plugged in or power unavailable at wall outlet (common)
- 2. IEB power switches off (common)
- 3. HP computer power switch off (common)
- 4. Monitors not turned on or improperly adjusted (common)
- 5. Loose or damaged cable connections (occasional)
Verify wall outlet power available
- Test point
- wall outlet
- Expected
- AC power present
- Tools
- multimeter
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cases performed on Mac-Lab not appearing in CVIS system
Completed procedures and case data from Mac-Lab do not sync to CVIS
Ranked causes
- 1. Interfaces on CVIS server not running or configured improperly (common)
- 2. Data Export Utility on Mac-Lab not configured correctly (common)
- 3. Data Export not enabled in System Admin File Path setup (common)
- 4. Network connection between Mac-Lab and CVIS Server down (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient search not working with known admitted patients
Patient search returns no results even though patient is known to be admitted in HIS
Ranked causes
- 1. DICOM settings on Mac-Lab not configured correctly (common)
- 2. Network connection lost between Mac-Lab and CVIS (occasional)
- 3. Patient Interface service not running (common)
- 4. HIS feed to ADT Interface is down on CVIS (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to capture image
Image grabber window is blank or displays error (no video found).
Ranked causes
- 1. Loose or incorrectly connected cables (common)
- 2. Odyssey firmware version incorrect (occasional)
- 3. Odyssey driver not installed or updated (occasional)
- 4. Video source not connected or IntelliCam not configured (occasional)
Verify all cables and connectors are tight
- Test point
- Red cable (input 1), blue cable (input 2), left DVI port (channels 1-2), right DVI port (channels 3-4)
- Expected
- Cables connected to correct inputs and ports
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode scanner does not work or system locks after first scan
Barcode scanner malfunction or system hangs following initial scan operation
Ranked causes
- 1. Barcode scanner connected to mouse input instead of keyboard input (common)
- 2. Barcode scanner cable connected to PS/2 keyboard connector while power on (common)
- 3. Incorrect barcode scanner cable type (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Group Policy not applying correctly to local computer
Custom shell not displayed; domain user lacks admin rights; SQL replication failures; auto OS login fails
Ranked causes
- 1. Computer not in correct Organizational Unit (OU) (common)
- 2. Network settings incorrect (DNS, IP, Subnet, Gateway) (common)
- 3. Local machine time not synchronized with Domain Controller (occasional)
Verify computer OU assignment
- Test point
- Active Directory computer object
- Expected
- Computer in correct organizational unit for policy scope
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image buttons grayed out
Live and Freeze buttons on Image window toolbar are grayed out and unavailable.
Ranked causes
- 1. CardioImage or Image capture option not enabled (common)
- 2. Odyssey board firmware outdated or driver not installed (occasional)
- 3. IntelliCam cannot detect Odyssey 0 device (occasional)
Verify software options and monitor configuration
- Test point
- Software options, Image window location
- Expected
- CardioImage enabled, Image capture enabled, window on third monitor
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient not found in Schedule search after CVIS scheduling
Scheduled patient from CVIS system does not appear when searching Mac-Lab schedule
Ranked causes
- 1. DICOM settings not configured correctly on Mac-Lab (common)
- 2. DICOM services not running on CVIS Server (common)
- 3. Network connection lost between Mac-Lab and CVIS (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Test DICOM connectivity with X-ray system
Diagnostic procedure to verify DICOM configuration and connectivity between Mac-Lab and X-ray system
Ranked causes
- 1. MWL MPPS Service not running on Mac-Lab after configuration changes (common)
- 2. X-ray system not configured to use Mac-Lab as MWL/MPPS destination (common)
- 3. Networking problem between Mac-Lab and X-ray system (occasional)
Test network connectivity via PING
- Test point
- Command prompt
- Expected
- PING successful response from X-ray system hostname or IP address
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot obtain proper policies on acquisition or review system
Policies not applied to acquisition or review workstation systems
Ranked causes
- 1. DNS not pointing to correct INW server (common)
- 2. INW server unreachable (occasional)
Verify DNS configuration
- Test point
- DNS settings on workstation
- Expected
- DNS entry directed to INW server IP
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CryoCath to other ablation device switching procedure
Transition from CryoCath to RF ablation device or vice versa; cable and driver switchover sequence
Ranked causes
- 1. Incorrect switchover sequence may cause communication loss or driver failure (occasional)
- 2. Incorrect switchover sequence when switching to CryoCath may cause communication loss (occasional)
Verify current ablation device and cable configuration
- Test point
- Ablation setup screen and physical cable at catheter connector
- Expected
- Software driver and physical cable match; either CryoCath or RF device installed
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DNS resolution by name fails
System cannot ping remote device by name; resolves by IP address only
Ranked causes
- 1. DNS client-side cache out of date (common)
- 2. DNS server-side configuration error (occasional)
Ping device by IP address
- Test point
- Remote Device IP
- Expected
- replies received
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image distorted or partially cut off
Able to capture images using provided DCF files but image is distorted, cropped, vertically elongated, or horizontally elongated.
Ranked causes
- 1. Incorrect pixel clock setting (common)
- 2. Incorrect video timing settings (horizontal/vertical positioning) (common)
Test all provided DCF files to identify best image
- Test point
- IntelliCam Digitizer menu
- Expected
- Identify DCF file with best image quality
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No login through custom shell
User unable to authenticate through the custom shell interface on Windows XP/2003 system
Ranked causes
- 1. Auto-logon sequence interfering with login (common)
- 2. Incorrect group policies applied to system (occasional)
Verify user can reach Windows logon dialog
- Test point
- Start Menu
- Expected
- Shift+logoff shows standard Windows logon dialog
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No video display on review stations with 18 inch LCD monitor
Monitor connected to review station displays no video output
Ranked causes
- 1. Video cable disconnected, damaged, or incorrect type (common)
- 2. Monitor or video card defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replication errors in Mac-Lab system
Database replication between clab database and clients fails; synchronization incomplete
Ranked causes
- 1. Replication tool cannot connect to clab database on acquisition system (common)
- 2. SQLServerAgent service logon credentials incorrect or missing (common)
Verify replication tool connectivity to clab database
- Test point
- Acquisition system, clab database port 1433
- Expected
- Connection successful; replication tool can query database
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System cannot communicate on network
Equipment unable to reach other devices; ping to gateway or remote hosts fails
Ranked causes
- 1. IP addressing misconfigured (common)
- 2. Default Gateway not routing packets correctly (occasional)
Ping loopback address
- Test point
- 127.0.0.1
- Expected
- replies received
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to communicate with TRAM
System cannot connect to TRAM unit; network communication fails during operation
Ranked causes
- 1. System IP address in TRAM reserved range 192.168.0.xxx (common)
- 2. TRAMnet jumper cable disconnected or improperly seated (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to logon to Custom Shell with Hospital User
User cannot authenticate to Custom Shell using hospital account credentials
Ranked causes
- 1. Default_MLCL Computers_Policy missing proper trusted user group information (common)
- 2. Users group missing required MLCL groups for logon rights (common)
Verify Default_MLCL Computers_Policy is listed
- Test point
- Group Policy Objects Applied list
- Expected
- Default_MLCL Computers_Policy present
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
3rd monitor in pan mode after resolution change
Display for 3rd monitor changed to pan mode showing only portion of desktop after display resolution changed to 1600X1200.
Ranked causes
- 1. Monitor does not natively support 1600X1200 resolution (occasional)
Install ATI Control Panel utility
- Test point
- Setup.exe in ATI FireGL driver folder
- Expected
- ATI Control Panel installed and system restarted
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ablation device not receiving data
Ablation device driver does not receive data from cable or device; data transmission fails
Ranked causes
- 1. COM port communication error between device driver and hardware (common)
Click ablation device driver in ablation setup screen to reset COM port
- Test point
- Ablation setup screen device driver button
- Expected
- COM port resets; device reconnects and begins receiving data
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Amplifier communication choppy signals in CardioLab
Signal acquisition from amplifier is degraded, intermittent, or contains noise artifacts
Ranked causes
- 1. PCI Comms board incorrect version number (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Amplifier signals incomplete, intermittent, or noisy
Received signals from amplifier are incomplete, drop out intermittently, or contain noise
Ranked causes
- 1. Fiber optic cable contaminated with dust or debris (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Archive utility user configuration required
GEMS Task Scheduler service must be configured with [withheld] account credentials and set to automatic startup
Ranked causes
- 1. GEMS Task Scheduler not configured with correct user account (common)
Access GEMS Task Scheduler service properties
- Test point
- Services administrative tool
- Expected
- GEMS Task Scheduler service opens in properties dialog
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode scanner disconnected or not connected caution
Barcode scanner cable not connected or disconnected while system powered
Ranked causes
- 1. Barcode scanner cable disconnected while powered (occasional)
Verify barcode scanner cable connected
- Test point
- barcode scanner connector
- Expected
- cable connected securely with power off
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: RF noise during ablation procedure caution
RF noise present during cardiac ablation procedure requires mitigation
Ranked causes
- 1. RF noise interference during ablation (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Clients not in correct Active Directory Organization Unit or group policies not applied
Client systems missing group policy settings or not in correct organization unit; replication cannot proceed
Ranked causes
- 1. Client not joined to domain or organization unit misconfigured (occasional)
Verify all clients are domain members and in correct Organization Unit before starting replication
- Test point
- Active Directory Users and Computers; each client system
- Expected
- All client systems show as domain-joined members in correct OU; group policies applied
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Correct group policies not applied
System does not have Default_MLCL Computers_Policy listed in applied group policy objects
Ranked causes
- 1. System not properly entered into domain (common)
View applied group policy settings
- Test point
- Help > Pick a Task > Use tools > Tools > Advanced System Information
- Expected
- View Group Policy settings applied displays in right window (up to 2 minutes)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Data import does not work on non-English systems
Patient demographics cannot be imported from CARTO XP systems on non-English CardioLab installations
Ranked causes
- 1. Non-English CardioLab system language incompatibility (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Equipment housing damaged or overheating caution
Visible cracks in housing, signs of overheating, or other physical damage to equipment enclosure
Ranked causes
- 1. Equipment housing cracked or damaged (rare)
Visually inspect housing for cracks and damage
- Test point
- equipment enclosure exterior
- Expected
- no visible cracks, overheating signs, or damage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
InSite 2.0 service agent loses connection to network or enterprise server
Service agent loses connection to network or enterprise server
Ranked causes
- 1. Network or enterprise server down (occasional)
Verify network or enterprise server is available
- Test point
- Network connectivity, enterprise server status
- Expected
- Network and enterprise server operational
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
InSite 2.0 service agents fail to start after system restart
Service agents attempt to start three times then shut down when system is restarted
Ranked causes
- 1. Agents require manual restart after system restart (common)
Verify agents have shut down after three startup attempts following system restart
- Test point
- InSite 2.0 agent status
- Expected
- Agents have stopped after three connection attempts
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Option codes become corrupted
Option codes cannot be read or applied correctly; system may not recognize licensed features
Ranked causes
- 1. Corrupted option registry keys (rare)
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replication setup required for Acquisition systems
Each Acquisition system must be added to replication on INW server via Service Console to enable data consolidation
Ranked causes
- 1. Acquisition systems not yet added to replication (common)
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scrambled video display after 6.0 software upgrade
Video output appears garbled or corrupted following system upgrade to version 6.0
Ranked causes
- 1. Incompatible refresh rate between video card and monitor (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Serial number entered incorrectly
System does not recognize or accept serial number entry; options cannot be applied
Ranked causes
- 1. Incorrect serial number value provided (common)
Open command prompt
- Test point
- Windows command line
- Expected
- Command prompt window opens
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Slow network transfer and ping speeds
Network communication experiencing reduced performance; transfer rates below expected
Ranked causes
- 1. NIC not negotiating 100mb/s Full Duplex with switch (common)
Verify NIC negotiation speed and duplex
- Test point
- Network Interface Card settings
- Expected
- 100mb/s Full Duplex
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SQL database not running on clients or service fails to start
SQL Agent service does not run on client systems; service fails when manually started; databases are not replicated to clients
Ranked causes
- 1. SQL user password changed without replication update on all systems (common)
Verify SQL user password change was followed by replication update on all systems
- Test point
- Acquisition system and all client systems SQL Server configuration
- Expected
- All systems running the same SQL user password; replication completed successfully
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SQL Server registration required for replication
SQL Server must be registered in Enterprise Manager before Acquisition systems can be added to replication
Ranked causes
- 1. SQL Server not yet registered in Enterprise Manager (common)
Access SQL Server Enterprise Manager
- Test point
- Server desktop
- Expected
- Enterprise Manager opens and shows SQL Server Group
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Test DICOM connectivity with CVIS
Diagnostic procedure to verify DICOM configuration and connectivity between Mac-Lab and CVIS
Ranked causes
- 1. DICOM configuration settings incorrect on Mac-Lab or CVIS (common)
Perform patient search for common name
- Test point
- Study > Patient Search at Navigator
- Expected
- Enter Patient Last Name of Smith and press Start Search
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to logon to Custom Shell with hospital user account
User cannot authenticate to the Custom Shell using hospital credentials
Ranked causes
- 1. Default_MLCL Computers_Policy missing proper trusted user group information (common)
Verify Default_MLCL Computers_Policy is listed under Applied Group Policy Objects
- Test point
- Group Policy Management Console, Applied Group Policy Objects
- Expected
- Default_MLCL Computers_Policy listed
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
User does not have appropriate rights error
Error message displayed when logged in domain as [withheld]
Ranked causes
- 1. User not in correct organizational unit (OU) (common)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.