GE Healthcare MAC 5500: troubleshooting and error codes
The ServiceForge knowledge base holds 55 documented symptoms for the GE Healthcare MAC 5500, with 107 ranked causes, 102 diagnostic checks with test points and expected values, and 8 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Visual inspection items found caution
Inspection identifies physical damage, loose connections, or component issues
Ranked causes
- 1. Frayed or damaged I/O connectors and cables (occasional)
- 2. Fuse issues (occasional)
- 3. Interface cable damage or loose connections (common)
- 4. Circuit board contamination or damage (occasional)
- 5. Ground wire or wiring issues (occasional)
- 6. Loose or missing mounting hardware (occasional)
- 7. Power source problems (occasional)
Visual inspection of I/O connectors and cables
- Test point
- Connectors, cable ends, plugs
- Expected
- No fraying, damage, bent prongs, cracked housing, or loose screws
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not start or power-up self-test fails
Equipment not functioning properly on power-up; internal self-test fails to display startup screen
Ranked causes
- 1. Unit not turned on (common)
- 2. No AC power supplied (common)
- 3. Battery not installed (occasional)
- 4. Writer door open or loose connections (occasional)
- 5. Equipment configuration change or operator error (occasional)
- 6. Hardware or software modification (rare)
Verify unit power switch is on
- Test point
- Power switch
- Expected
- Power switch in ON position
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG data displays unacceptable noise levels patient safety
Acquired ECG signal contains excessive noise making diagnosis difficult
Ranked causes
- 1. Improper electrode placement (common)
- 2. Improper electrode application or contamination (common)
- 3. Defective or out of date electrodes (occasional)
- 4. Defective, broken, or disconnected leadwires (occasional)
- 5. Patient movement during acquisition (common)
Verify proper electrode placement
- Test point
- electrode positions
- Expected
- per ECG protocol
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Missing ACI-TIPI Report patient safety
ACI-TIPI interpretation report not displayed or generated
Ranked causes
- 1. ACI-TIPI is disabled (common)
- 2. Report selected without interpretation option (occasional)
- 3. Required patient information not entered (common)
- 4. Patient entered as pediatric with invalid age range (rare)
- 5. Original ECG acquired without ACI-TIPI option (rare)
Verify ACI-TIPI is enabled in system settings
- Test point
- software settings menu
- Expected
- ACI-TIPI enabled
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor quality ECGs patient safety
ECG acquisition produces poor signal quality, unreadable traces, or inadequate waveforms
Ranked causes
- 1. Environmental factors (common)
- 2. Inadequate patient preparation (common)
- 3. Hardware failures in acquisition module (occasional)
- 4. Leadwire or cable problems (occasional)
- 5. Unit malfunction (occasional)
Test with simulator
- Test point
- Simulator input
- Expected
- Good ECG report from simulator signal
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No blood pressure reading from external device
Blood pressure data is not received from connected external device
Ranked causes
- 1. Incorrect setup configuration (unknown)
- 2. Suntech protocol misconfiguration (unknown)
- 3. Defective or disconnected cables (common)
- 4. TTL trigger malfunction (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No BP from External Device patient safety
Blood pressure not received or displayed from external BP monitoring device
Ranked causes
- 1. Device setup not configured correctly (common)
- 2. Protocol mismatch (Suntech device) (occasional)
- 3. Serial or TTL cable disconnected or defective (common)
- 4. TTL trigger signal not active (occasional)
Verify external device setup configuration
- Test point
- system setup menu
- Expected
- external BP device enabled and configured
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Treadmill or Ergometer Does Not Move caution
Treadmill or ergometer equipment will not start or respond to control commands
Ranked causes
- 1. Protocol not configured (common)
- 2. Control cables disconnected or defective (common)
- 3. Input/Output settings not configured (occasional)
- 4. Emergency Stop switch engaged (occasional)
Verify equipment protocol configured
- Test point
- protocol settings menu
- Expected
- treadmill or ergometer protocol selected
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Treadmill/Ergometer does not move patient safety
Exercise equipment fails to start or operate during stress testing
Ranked causes
- 1. Incorrect protocol setting (common)
- 2. Defective or disconnected cables (common)
- 3. Incorrect input/output settings (common)
- 4. Emergency Stop switch activated (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Self-test failure on power-up caution
Internal self-test fails on startup, system does not display normal startup screen
Ranked causes
- 1. Hardware or software modified since last use (occasional)
- 2. Environmental or usage changes (occasional)
- 3. Operator error (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermal writer produces poor print quality
Printed output shows distortion, skewing, crushing, curved lines, misalignment, or uneven darkness
Ranked causes
- 1. Print head speed or paper tracking problem (common)
- 2. Flat spot on print roller (occasional)
- 3. Missing print head dot (occasional)
Verify first character printed is not distorted
- Test point
- Printed output start
- Expected
- First character clear and properly formed without distortion
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on
System does not start, no display or startup screen appears
Ranked causes
- 1. Unit turned off (common)
- 2. Battery not installed (occasional)
- 3. No AC power or faulty AC connection (common)
Verify unit power switch is on
- Test point
- Power switch
- Expected
- Switch in ON position
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Acquisition module not responding
Acquisition module does not communicate with system; power test fails; noise levels abnormal
Ranked causes
- 1. Acquisition module power supply failure (common)
- 2. Acquisition module communication circuit defect (occasional)
Run acquisition module power test
- Test point
- Acquisition module power supply
- Expected
- Module powered and responding to test commands
- Tools
- Shorting bar (required)
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Acquisition module test fails
Acquisition module does not respond or communication test fails during diagnostic
Ranked causes
- 1. Acquisition module not powered or communication disabled (common)
- 2. Acquisition module hardware failure (occasional)
Install shorting bar and run acquisition module test
- Test point
- Acquisition module
- Expected
- Module powered and communicating; button presses detected
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery voltage abnormal
Battery voltage outside acceptable range; does not charge or discharge properly; depletes rapidly
Ranked causes
- 1. Battery requires replacement (common)
- 2. Main CPU board failure (occasional)
Run Battery Status test
- Test point
- System battery
- Expected
- Battery voltage between 15 and 24 volts at 80% or higher charge; current between -0.7 and 0.7 amps V, 15–24 V at ≥80% charge
- Tools
- System Diagnostics menu
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ethernet module test fails
Ethernet module IP address and subnet mask display N/A or test is unsuccessful
Ranked causes
- 1. Ethernet module hardware failure (occasional)
- 2. Network configuration issue (common)
Verify Ethernet cable connection
- Test point
- Ethernet port
- Expected
- Cable firmly connected
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
External modem not responding
External modem ID and firmware revision display N/A during External Modem Test
Ranked causes
- 1. Modem not connected to COM 2 or modem power off (common)
- 2. Serial communication failure between unit and modem (occasional)
Verify modem connection to COM 2 and power
- Test point
- COM 2 port
- Expected
- Modem powered on and connected
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal memory test detects bad blocks
Internal Memory Tests display indicates presence of bad blocks in storage
Ranked causes
- 1. Internal storage memory corruption or hardware defect (occasional)
- 2. Internal storage memory circuit failure (rare)
Run Internal Memory Tests and note bad block count
- Test point
- Internal memory storage
- Expected
- Zero bad blocks; free memory displayed
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal modem not responding
Internal modem ID and firmware revision display N/A during Internal Modem Test
Ranked causes
- 1. Internal modem hardware failure or firmware issue (occasional)
- 2. Main CPU modem control circuit failure (rare)
Run Internal Modem Test and check result
- Test point
- Modem response
- Expected
- ID number and firmware rev display (not N/A)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Modem communication unsuccessful
Modem does not respond to test commands; ID number and firmware revision display as N/A
Ranked causes
- 1. External modem not connected or defective (common)
- 2. Internal modem circuit failure (occasional)
Run External Modem Test
- Test point
- COM 2 port (modem connection)
- Expected
- Modem ID number, firmware revision, and parameter settings display correctly
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SD card read/write test failed
SD card cannot be read or written; test returns failure result; memory card appears defective
Ranked causes
- 1. Defective SD card (common)
- 2. CPU board SD card interface failure (occasional)
Insert non-write-protected SD card
- Test point
- SD card slot
- Expected
- Card inserted and detected by system
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SD card test fails
SD Card Test returns Failed status during read/write test
Ranked causes
- 1. SD card defective or incompatible (common)
- 2. SD card controller on main CPU board failure (occasional)
Insert non-write-protected SD card and run SD Card Test
- Test point
- SD card slot
- Expected
- Test displays 'Passed'
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Serial communications port test fails
COM loopback test returns Failed after installing loopback jumpers
Ranked causes
- 1. Serial port hardware defect or loopback jumper not installed (common)
- 2. Main CPU board serial port circuit failure (occasional)
Install loopback jumpers and run COM loopback test
- Test point
- Selected COM port (COM1, COM2, COM3, or COM4)
- Expected
- Test displays 'Passed'
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ambient temperature inside unit excessive caution
Internal ambient temperature more than 10°C above room temperature
Ranked causes
- 1. Cooling fan malfunction or airflow obstruction (common)
Compare internal and external temperatures
- Test point
- Ambient temperature display
- Expected
- Internal temperature ≤ 10°C above room temperature
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Analog input voltage measurement failed
Measured voltage from external DC power supply does not display correctly; analog input circuit not responding
Ranked causes
- 1. Main CPU board analog input circuit defect (unknown)
Run Analog Input Test with external DC power supply
- Test point
- DC input pins (ANA/TTL connector)
- Expected
- Applied voltage correctly displayed on test screen
- Tools
- DC power supply
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Analog input voltage not reading correctly high voltage
Voltage applied to analog input does not display correctly during Analog Input Test
Ranked causes
- 1. Analog input circuit fault on main CPU board (common)
Connect DC power supply to DC input pins and read displayed value
- Test point
- ANA/TTL connector analog input
- Expected
- Displayed voltage matches applied voltage
- Tools
- DC power supply, oscilloscope
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Analog output voltage incorrect
DC output, ECG output, or trigger output voltage does not match expected value when measured
Ranked causes
- 1. Main CPU board DAC (analog output) circuit defect (unknown)
Measure +12V output
- Test point
- Oscilloscope connected to appropriate pins per Analog Output Test
- Expected
- +12 V DC
- Tools
- Oscilloscope
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery capacity below minimum
Discharge capacity measured less than 2000 mAH during battery status check
Ranked causes
- 1. Battery aging or internal chemical degradation (common)
Check battery discharge capacity
- Test point
- Battery status display
- Expected
- Discharge capacity ≥ 2000 mAH
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery current excessive with no AC power
Battery current less than -0.7 amps (more negative) when AC power not applied
Ranked causes
- 1. Main CPU board drawing excessive standby current (occasional)
Monitor battery current with AC power off
- Test point
- Battery status display
- Expected
- Battery current ≥ -0.7 amps A
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery voltage too low
Battery voltage below 15 volts when percent charge is 80% or more
Ranked causes
- 1. Battery internal resistance increase or cell failure (common)
Monitor battery voltage during charge cycle
- Test point
- Battery status display
- Expected
- 15 to 24 volts (when charge ≥ 80%) V, 15-24 V
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication port test failed
Serial communication port does not transmit or receive data correctly; loopback test fails
Ranked causes
- 1. Serial port hardware defect or CPU board failure (unknown)
Install loopback jumper and run COM port test
- Test point
- COM port (1, 2, 3, or 4)
- Expected
- Test result Passed; all ASCII characters transmitted and received correctly
- Tools
- Loopback jumpers
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
DC output loopback test fails high voltage
Loopback test detects incorrect values on analog inputs when DC outputs are sent
Ranked causes
- 1. Analog output or input circuit fault on main CPU board (common)
Connect DC outputs to analog inputs and run loopback test
- Test point
- ANA/TTL connector DC outputs looped to analog inputs
- Expected
- Test displays 'Passed'
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
DC output to analog input loopback test failed
Test cannot confirm correct values read from analog inputs when DC outputs are driven; loopback path broken
Ranked causes
- 1. Main CPU board analog output or input circuit defect (unknown)
Run DCOut Loopback Test
- Test point
- DC output to analog input path (ANA/TTL connector pins)
- Expected
- Pass result; all values transmitted and received correctly
- Tools
- Internal loopback connection per test instructions
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display brightness or contrast out of range
Screen brightness or contrast appears abnormal; grey scale test pattern shows unusual intensity
Ranked causes
- 1. Display settings out of specification (unknown)
Examine brightness and contrast of test patterns
- Test point
- Display screen
- Expected
- Brightness and contrast within normal range; no unusual intensity variation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display pixels not functioning
Missing or non-working pixels visible on screen during pixel verification test
Ranked causes
- 1. Display hardware defect (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display screen pixels not working
Missing or non-functional pixels visible on display during operation
Ranked causes
- 1. Defective display panel (unknown)
Verify all screen pixels illuminate
- Test point
- Display screen
- Expected
- All pixels visible; no missing or dark spots
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG output loopback test fails high voltage
Loopback test detects incorrect ECG output or QRS trigger output values
Ranked causes
- 1. ECG output circuit or QRS trigger circuit fault on main CPU board (common)
Connect ECG output and QRS trigger to analog inputs and run loopback test
- Test point
- ANA/TTL connector ECG and trigger outputs looped to analog inputs
- Expected
- Test displays 'Passed'
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG output to analog input loopback test failed
Test cannot confirm ECG output and QRS trigger output transmit correctly to analog inputs; loopback path broken
Ranked causes
- 1. Main CPU board ECG output or analog input circuit defect (unknown)
Run ECGOut/QRSTrigger Loopback Test
- Test point
- ECG output and QRS trigger output to analog input path (ANA/TTL connector pins)
- Expected
- Pass result; ECG signal and trigger waveforms transmitted and received correctly
- Tools
- Internal loopback connection per test instructions
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Equipment hardware or environmental issues caution
Visible signs of hardware defects, disconnection, or contamination
Ranked causes
- 1. Disconnected cables, debris, missing hardware, loose components (common)
Visual inspection of I/O connectors and cables
- Test point
- Connectors, cable plugs
- Expected
- No fraying, bent prongs, cracked housing, or loose screws
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Equipment malfunction after configuration change
Unit fails after change in use, location, environment, or modification of hardware/software
Ranked causes
- 1. Use, location, or environment change affecting operation (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ethernet module test failed
Ethernet module does not respond; IP address, subnet mask, and MAC address display as N/A
Ranked causes
- 1. Ethernet module circuit failure or disconnection (unknown)
Run Ethernet Module Test
- Test point
- Ethernet module interface
- Expected
- Device IP address, subnet mask, and MAC address display correctly (not N/A)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal memory read/write test failed
Internal storage memory does not read or write correctly; bad blocks detected; memory test returns error
Ranked causes
- 1. Internal memory hardware failure (common)
Run Internal Memory Tests
- Test point
- System internal storage memory
- Expected
- No bad blocks; free memory available; test passes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard key input not recognized
Pressing keys does not produce expected on-screen response; key presses not highlighted or displayed
Ranked causes
- 1. Non-functional keyboard key or keyboard circuit (unknown)
Verify each key produces response on screen
- Test point
- Keyboard keys and arrow pad
- Expected
- Key highlighted and displayed on screen; scan code shown at top; normal dim background may remain
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard key not responding or not displayed
Key press not highlighted on screen or numeric scan code not displayed at top of screen
Ranked causes
- 1. Keyboard hardware defect or connection issue (unknown)
Press individual keys during keyboard test
- Test point
- Keyboard keys
- Expected
- Key highlighted on screen and scan code displayed at top
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software installation fails on -008 main board
Cannot install software versions 9C or 10A on MAC 5500 with 801212-008 main board; boot code prevents installation
Ranked causes
- 1. Incompatible software version for -008 CPU (common)
Verify main board model number
- Test point
- main board
- Expected
- 801212-008 or other model
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Speaker produces no sound
No audible tone output when speaker test is run; speaker appears non-functional
Ranked causes
- 1. Main CPU board failure (speaker integrated) (common)
Run Speaker Test
- Test point
- System speaker output
- Expected
- Audible tone (loud or soft) produced
- Tools
- Ears to detect sound
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Speaker produces no tone
No audible output when speaker test is run with Loud or Soft selection
Ranked causes
- 1. Main CPU board failure (speaker is part of CPU board) (common)
Run speaker test and select Loud or Soft
- Test point
- Speaker output
- Expected
- Audible tone at selected volume level
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermal writer has flat spot on roller caution
Isolated light spots visible in roller test pattern indicating roller defect
Ranked causes
- 1. Roller mechanical wear or flatness defect (occasional)
Run roller test and inspect output pattern
- Test point
- Roller test output
- Expected
- Diagonal light and dark wavy bands without isolated light spots
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermal writer has missing print head dot caution
White line visible across full length of page in roller test output
Ranked causes
- 1. Print head mechanical damage or nozzle failure (occasional)
Run roller test and inspect for white lines
- Test point
- Roller test pattern
- Expected
- No white lines across page length
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermal writer paper travel is not smooth caution
Paper motion during writer test shows hesitation, resistance, or irregular movement
Ranked causes
- 1. Paper path obstruction or roller friction (common)
Observe paper movement during test
- Test point
- Paper feed path
- Expected
- Paper moves smoothly without hesitation or jerking
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermal writer prints curved or wavy lines caution
Large triangles and diagonal lines are curved or wavering instead of straight and uniform
Ranked causes
- 1. Writer motor speed control or mechanical wear (common)
Measure printed line straightness
- Test point
- Printed triangles and diagonal lines
- Expected
- Lines straight and uniform without curves or wavering
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermal writer prints distorted first character caution
First character printed is distorted or unclear at start of writer test
Ranked causes
- 1. Writer motor start-up speed issue (common)
Examine first character in writer test output
- Test point
- First printed character
- Expected
- First character not distorted
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermal writer prints skewed or crushed paper edges caution
Output shows uneven paper edge tracking or paper compression during writing
Ranked causes
- 1. Paper tracking or tension mechanism misalignment (common)
Inspect paper edges in test output
- Test point
- Paper edge alignment
- Expected
- Both edges straight and even, no crushing or skewing
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermal writer test output misaligned with tear bar caution
Perforations do not align with tear bar on writer door after paper cueing
Ranked causes
- 1. Paper cueing mechanism misalignment (occasional)
Inspect perforation alignment after paper cueing
- Test point
- Tear bar alignment
- Expected
- Perforations aligned with tear bar
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermal writer test pattern length incorrect caution
Printout length does not meet specification of 250 mm ± 5 mm for test patterns
Ranked causes
- 1. Motor speed control error at tested speed (common)
Run 50 mm/s test pattern and measure length
- Test point
- Test pattern output length
- Expected
- 250 mm ± 5 mm (use grids on page), ±5 mm
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Battery icon appears on screen | No battery installed in the system | System powered on without battery present | warning |
Battery icon flashes intermittently | Battery charge is low | System operating on battery power with insufficient remaining charge | warning |
Cannot use the system because Device Password does not work | Device password authentication failure | Incorrect device password entered or system password changed | error |
MODEM ERROR. The remote device is not responding. Would you | Modem connection failure; remote device not responding | System attempts to transmit data to remote device but connection cannot be established or maintained | error |
System does not power up when operating from battery power | System fails to start on battery power | Power button pressed while on battery with insufficient charge | error |
System shuts down when operating from battery power | System unexpectedly powers off during battery operation | Battery becomes depleted during operation or automatic shutdown feature activates | warning |
Writer door icon appears on screen | The writer door is open | Writer door detected as open during operation or startup | warning |
X Lead disconnected message | Electrode connection lost | One or more electrodes become physically disconnected from patient or connector | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI