GE Healthcare Invenia ABUS: troubleshooting and error codes
The ServiceForge knowledge base holds 12 documented symptoms for the GE Healthcare Invenia ABUS, with 24 ranked causes, 12 diagnostic checks with test points and expected values, and 80 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Noise in ultrasound images
Electromagnetic interference appearing in ultrasound images; may be present all the time, intermittent, or after system has been running for some time
Ranked causes
- 1. EMI from external sources (radio frequencies, broadcast signals, nearby equipment) (common)
- 2. Poor grounding or shared power outlet (common)
- 3. EMI via external cables (network, power, probe) (common)
- 4. Damaged or missing RF gaskets or covers (occasional)
- 5. Heat problems (fan filters dirty, worn fans, high room temperature) (occasional)
- 6. Hardware malfunction (faulty card or probe) (occasional)
- 7. Software issues or outdated firmware (occasional)
Verify if noise is present all the time or intermittent
- Test point
- Visual inspection of ultrasound image during operation
- Expected
- Noise pattern noted (continuous or intermittent)
6 more checks with pass and fail actions, 17 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blank screen
Display shows nothing; may indicate power, initialization, or startup issues
Ranked causes
- 1. System not powered on or not plugged in (common)
- 2. Probe does not initialize properly (occasional)
- 3. System startup error requiring reboot (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image does not transfer from scan station to workstation
Scanned images fail to transmit to workstation for review
Ranked causes
- 1. Network connection not established (common)
- 2. Workstation or scan station disk full (occasional)
- 3. Network or communication fault (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Intermittent noise pattern
Noise that appears only at certain times or when specific equipment nearby is operating; not consistently present throughout operation
Ranked causes
- 1. Nearby equipment turned on/off causing intermittent EMI (common)
- 2. Time-dependent EMI (broadcast, facility operations) (occasional)
Document when noise occurs
- Test point
- Time log of noise appearance
- Expected
- Clear pattern or specific times noted
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lines appear in image
Artifact lines visible on ultrasound image display
Ranked causes
- 1. Probe orientation incorrect or probe defective (common)
- 2. Acquisition board or TIP board fault (occasional)
Reverse probe and observe if lines move to opposite side
- Test point
- Image display
- Expected
- If lines move to opposite side of screen after probe reversal, probe is defective; if lines remain on same side, board fault
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Articulating arm creeps or drifts caution
Scanner head rises or lowers when not held, indicating insufficient tension in articulating arm
Ranked causes
- 1. Articulating arm tension requires adjustment (common)
Verify arm and head position stability after movement
- Test point
- Scanner head
- Expected
- Arm and head should stay in location where moved to without drifting
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Intermittent operation, system lockups, slow performance, overheating, or poor image quality caution
System exhibits multiple issues: intermittent response, freezing, slow speed, heat, or degraded image
Ranked causes
- 1. Air filter clogged; system thermal throttling (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mechanical noise or thumping sound caution
Unusual mechanical sounds emanate from system during operation
Ranked causes
- 1. Internal mechanical fault (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Multi-axis lock stops working caution
Multi-axis locking mechanism fails to engage or hold
Ranked causes
- 1. Multi-axis lock mechanism fault (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Single-use stabilization membrane not properly removed caution
Membrane fragments or metal plates remain on scanner assembly after use
Ranked causes
- 1. Incomplete removal of single-use stabilization membrane (occasional)
Visually inspect underside of single-use stabilization membrane
- Test point
- Underside of membrane
- Expected
- Two small metal plates should be present
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System overheating caution
System reaches excessive temperature during operation
Ranked causes
- 1. Air filter clogged or dirty (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch screen calibration drift after cleaning caution
Touch screen calibration may be inaccurate following cleaning and disinfection
Ranked causes
- 1. Touch screen requires recalibration after cleaning (occasional)
Verify touch screen calibration accuracy
- Test point
- Touch screen display
- Expected
- Calibration should be accurate and responsive
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Aborted | Scan has been aborted | User cancels scan in progress | info |
Abort finished. Ready for next scan. | Previous scan abort completed successfully | Scan abort operation completes | info |
Aborting | Current scan is being terminated | User initiates scan abort | info |
An application entity can only be set to Print or Store, not | DICOM entity misconfigured for both print and store | User attempts to enable both print and store modes | error |
Are you sure you want to delete all these series/images? | Confirmation prompt before deleting multiple series | User initiates bulk delete | warning |
Are you sure you want to delete these studies? | Confirmation prompt before deleting multiple studies | User initiates bulk study delete | warning |
Are you sure you want to delete this series/image? | Confirmation prompt before deleting series or image | User initiates series/image delete | warning |
Are you sure you want to delete this study? | Confirmation prompt before deleting study | User initiates study delete | warning |
Are you sure you want to log off? | Confirmation prompt before logout | User initiates logout | info |
Are you sure you want to logout? | User confirmation prompt before logout | User initiates logout | info |
Are you sure you want to remove all calipers? | Confirmation prompt before removing all measurement calipers | User initiates bulk caliper removal | info |
Automatic export items can't be deleted. | System-automated exports cannot be manually deleted | User attempts to delete auto-export items | info |
Can't load report data. | Report data retrieval failed | System attempts to load report | error |
Confirm DICOM Send | Check DICOM transmission status | User needs to verify DICOM file transfer | info |
Dataset is not available | Selected patient dataset no longer accessible | User attempts to resend unavailable patient data | error |
Delete confirmation | User confirmation required for deletion | User initiates delete operation | info |
DICOM connection test failed. | DICOM network connectivity test unsuccessful | System performs DICOM connection test | error |
DICOM connection test passed. | DICOM network connectivity test successful | System performs DICOM connection test | info |
Do you want to resume the last saved session? | Prompt to restore previous session | Workstation starts after shutdown | info |
Export from DICOM server is not supported. | DICOM server export not available | User attempts export from server | info |
Failed to export users. | User export operation failed | System attempts user export | error |
Failed to retrieve images from the DICOM server. | Image retrieval from DICOM server failed | System attempts image retrieval | error |
Imaging Engine Initializing.... | System startup in progress | System powers on | info |
InSite installation is corrupt or missing | InSite software installation has failed or is damaged | InSite icon displays corrupt/missing state | error |
InSite not connected to GE | InSite is installed but not connected to GE service servers | InSite icon displays disconnected state | error |
Invalid port number. Port number can only have four digits. | Port number format invalid; must be 4 digits | User enters invalid port number | error |
Invalid printer, or no printer, is installed. | Printer not detected or not properly configured | System attempts to use printer | error |
No Communication Between Scan Station and Workstation | Network link lost between scan and review stations | Communication failure between systems | error |
No DICOM images found at the folder <folder>. | No DICOM files in specified folder | System searches folder for DICOM images | info |
No images are retrieved. | User cannot change local port or AE title; non-service user | Non-service user attempts port or AE title change | warning |
Only Service users can change the local settings. | User permissions insufficient to modify settings | Non-service user attempts configuration change | warning |
Only three lines of text are allowed for Institute Address. | Institute address exceeds 3 lines; input truncated | User enters institute address longer than 3 lines | warning |
[Patient Name] is already selected for this scan | Patient already selected for current scan session | User attempts to select a patient when one is already active | info |
Please check the scantype | Scan type must be selected before scanning | User attempts scan without selecting type | warning |
Please enter a User name. | Username field is required but empty | User attempts operation without username | warning |
Please input patient information. | Patient demographics required before scanning | User attempts scan without complete patient data | warning |
Please mark all nipple positions | Nipple positions must be marked on scan image | User proceeds without marking nipple positions | warning |
Please restart DICOM receive application to make the changes | DICOM configuration changes require app restart | User modifies DICOM settings | info |
Please retrieve the study before you load it. | Study must be retrieved before loading | User attempts to load un-retrieved study | warning |
Please select a location to export the study(s) to. | Destination for study export not specified | User initiates export without selecting destination | warning |
Please select a new patient | Must select new patient to begin scan | User attempts scan without patient selection | warning |
Please select another study. | Current study selection invalid; select alternative | User operates on invalid study selection | warning |
Please select either Export To Disk or Export to DICOM Serve | Export destination type not selected | User initiates export without specifying type | warning |
Please unlock scanner first | System locked; login required | User attempts operation without login | warning |
Please wait until processing is finished | System busy processing current task | User attempts operation while processing active | info |
Press compression button | Compression must be applied before scan start | User attempts scan without compression | warning |
Press Compression button or Start button | Apply compression and start scan | User ready to begin compression scan | info |
Press Scan Button on Scan Head to Start Scanning | User must press scan start button | System ready for scan initiation | info |
Press Start button to start scan | User must press start button to begin scan | System ready for scan initiation | info |
Printer error. | Printer operation failed | System attempts to use printer | error |
Processing... | Image data being processed for display | Scan acquisition complete; processing active | info |
Report file [file name] already exists. | Export filename conflicts with existing file | User exports with duplicate filename | warning |
Retrieve completed | Study retrieval from server completed successfully | System completes study retrieval | info |
Retrieving study <id>. | Study retrieval in progress | System retrieves study from server | info |
Scan finished | Current scan view type acquisition complete | Scan data acquisition ends | info |
Scan limit reached | System storage capacity at limit | Insufficient disk space remaining | warning |
Scanning | System actively acquiring scan data | Scan acquisition in progress | info |
Scan order for this patient already selected | Scan protocol already specified for patient | User attempts to set scan order twice | info |
Scan started | Scan acquisition in progress | User presses scan start button | info |
Scan Station Initializing... | Scan station startup in progress | System powers on | info |
Select scan order | User must specify scan workflow | New exam setup required | info |
Select scan type | User must choose view types for exam | Exam setup before scanning | warning |
Study doesn't exist. | Selected study not found in system | User attempts to access non-existent study | error |
The Invenia ABUS application has encountered a problem. | Application error; restart required | Application encounters unhandled exception | critical |
The Invenia ABUS disk space is below [configured free space] | Available disk space below configured threshold | System monitors disk space during operation | warning |
There are scans missing for this patient. Are you sure you w | Incomplete scans for patient; confirm finish | User attempts to finish exam with pending scans | warning |
The Study doesn't exist. | Referenced study not found | System attempts to access study | error |
The study <id> of the patient <id> already exists. Do you wa | Study import would overwrite existing study; confirm | User imports study with matching ID | warning |
The user name already exists. Please choose a different user | Username already in use; must be unique | User creates account with duplicate username | error |
The user name or password you entered is incorrect. | Login credentials invalid | User submits incorrect username or password | warning |
The view type of the image stored in the Invenia ABUS Storag | Image view type is immutable after storage | User attempts to modify stored image view type | info |
Transducer is moving. Please wait. | Transducer in motion; wait for completion | Transducer initialization or movement | info |
Unable to restore study. Please load it from the study pane. | Study restoration failed; use study pane to load | System fails to restore study | error |
Unauthorized to access the printer. | User lacks permission to use printer | User attempts print operation | warning |
UsysDcmReceiveApp is missing in the installation directory . | DICOM receive application not found; installation incomplete | System attempts to start DICOM service | critical |
You have images in the data pane. | Images remain in data pane; operation may be affected | User performs operation with active images | info |
You have images in the data pane. Are you sure you want to o | Confirm opening new study with unsaved images | User opens new study with active images | warning |
You have images pending int he DICOM queue. | Images queued for DICOM transmission | User operates while DICOM queue active | info |
You must first select a patient | Patient selection required before scanning | User attempts scan without patient selection | warning |
You must restart Invenia ABUS for the configuration changes | System restart required for config changes | User modifies configuration settings | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI