GE Healthcare Invenia ABUS 2.0: troubleshooting and error codes
The ServiceForge knowledge base holds 15 documented symptoms for the GE Healthcare Invenia ABUS 2.0, with 24 ranked causes, 31 diagnostic checks with test points and expected values, and 121 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Electromagnetic interference causing image noise or artifacts caution
Image contains noise patterns attributed to external electromagnetic sources
Ranked causes
- 1. Nearby EMI source (medical laser, scanner, cauterizer, computer, monitor, fan, microwave, light dimmer, mobile phone, wireless device, AC system, HF surgery equipment, AC/DC adapter) (common)
- 2. Poor grounding of power cord or power outlet (common)
- 3. Loose, missing, or damaged RF gaskets or covers; loose metallic parts (occasional)
- 4. Incorrect cable routing or non-GE specified harnesses and peripherals (occasional)
- 5. Network cable acting as EMI antenna (occasional)
Gather information on noise conditions and timing
- Test point
- Operator report and environmental observation
- Expected
- Document presence of noise continuously, intermittently, after delay, at specific times, or at specific locations
8 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Self-generated noise from system
Noise generated internally within Invenia ABUS 2.0 unit during operation
Ranked causes
- 1. Thermal problem due to prolonged operation (common)
- 2. Software or setup issue (occasional)
- 3. Hardware malfunction in transducer or main unit (occasional)
Restart system
- Test point
- System power and boot sequence
- Expected
- System powers down and back up successfully
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unwanted noise in ultrasound image
Noise appears in acquired ultrasound images during scanning or processing
Ranked causes
- 1. Electromagnetic interference via external cables (common)
- 2. Grounding problems in mains power system (common)
- 3. Equipment switched on/off near Invenia ABUS 2.0 (occasional)
Visually inspect cable connections
- Test point
- All external cables
- Expected
- All cables properly connected and shielded
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lines visible in ultrasound image caution
Visible lines or artifacts appearing on the ultrasound image display
Ranked causes
- 1. Transducer orientation incorrect or transducer defective (common)
- 2. Acquisition board or TIP board defect (occasional)
Reverse transducer and observe image
- Test point
- Ultrasound display
- Expected
- Lines move to opposite side of screen or disappear
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blank screen
Display shows no image or content on startup
Ranked causes
- 1. System not powered or powered button not activated (common)
Verify system is plugged in and powered on
- Test point
- Power button
- Expected
- Button lit green
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Touch screen display offset
Touch screen display is noticeably offset from actual tap location; factory calibration requires adjustment
Ranked causes
- 1. Touch screen calibration drift from factory settings (occasional)
Verify touch screen calibration offset is noticeable
- Test point
- Touch screen display
- Expected
- Touch location matches tap point
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image does not transfer to workstation
Ultrasound images fail to transmit from system to connected workstation
Ranked causes
- 1. Network connection not established or disabled (common)
Verify workstation network indicator status
- Test point
- Workstation network indicator light
- Expected
- Green light
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Intermittent noise appearing periodically
Noise appears intermittently during scanning or at specific times, not continuously present
Ranked causes
- 1. External equipment switched on/off near system (occasional)
Monitor noise occurrence pattern
- Test point
- Ultrasound image during scan
- Expected
- Noise appears only during specific times or equipment operation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Intermittent operation, lock-ups, slow response, overheating, poor image quality caution
System exhibits intermittent function, freezes, sluggish response, elevated temperature, or degraded ultrasound image quality
Ranked causes
- 1. Dirty filter restricting airflow and causing thermal stress (common)
Inspect and clean air filter for blockage
- Test point
- System air intake filter
- Expected
- Filter clean with unrestricted airflow
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mechanical noise or thumping sound caution
Audible mechanical noise or thumping from system during operation
Ranked causes
- 1. Internal mechanical defect (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Multi-axis lock not functioning caution
Multi-axis lock feature stops working during operation
Ranked causes
- 1. Lock mechanism failure (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scan Head Assembly repositioning test
Moving Scan Head Assembly to different location to isolate noise source location
Ranked causes
- 1. Noise correlated with physical location (occasional)
Reposition Scan Head Assembly
- Test point
- Scan Head location
- Expected
- Assembly moved to new location at least 6 feet away from original position
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System overheating caution
System temperature exceeds normal operating range
Ranked causes
- 1. Dirty or blocked air filter restricting airflow (common)
Inspect and clean air filter
- Test point
- System air filter
- Expected
- Filter clean and unrestricted
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System powers on with error displayed
System starts but shows error message on screen
Ranked causes
- 1. Software or initialization error (occasional)
Attempt system reboot
- Test point
- Power button
- Expected
- System restarts without error
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System won't shut down normally
System does not respond to standard power button press for shutdown
Ranked causes
- 1. Software or power control unresponsive (occasional)
Attempt normal power button shutdown
- Test point
- Power button
- Expected
- System gracefully shuts down
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
11001 | Failed to initialize acquisition | Missing frontend JSON file or possible transducer disconnection | error |
12001 | Scanhead id does not match | Scanhead identifier mismatch during initialization | error |
13001 | Image renderer initialization failed | Renderer subsystem fails to initialize | error |
14001 | GPU max temperature exceeded | GPU temperature exceeds maximum threshold | critical |
14002 | Power max exceeded | System power consumption exceeds maximum limit | critical |
21001 | Initialization success | ScannerApp initialization completes successfully | info |
21002 | Weight motor init home failed | Weight motor fails to reach home position during initialization | error |
21003 | Scan motor init home failed | Scan motor fails to reach home position during initialization | error |
21004 | Ball motor init home failed | Ball joint motor fails to reach home position during initialization | error |
21005 | Scanhead controller USB COM port not detected | Scanhead controller USB communication port unavailable at startup | error |
21006 | Weight controller USB COM port not detected | Weight controller USB communication port unavailable at startup | error |
21007 | Scanhead and weight controller USB COM port not detected | Both scanhead and weight controller USB ports unavailable at startup | error |
22001 | Scanhead controller USB COM port not detected | Scanhead controller USB port unavailable after initialization | error |
22002 | Weight controller USB COM port not detected | Weight controller USB port unavailable after initialization | error |
31000 | Unknown GEUS error code received | Unrecognized GEUS error code detected | unknown |
31001 | GEUS fatal error requiring system shutdown | Critical GEUS subsystem failure detected | critical |
31002 | GEUS fatal error requiring system restart | Critical GEUS error that requires system restart to recover | critical |
35001 | Transmit overload | Ultrasound transmitter power exceeds safe limits | critical |
35002 | Setup problem occurred | Configuration issue detected during system setup | error |
35003 | Major setup problem occurred | Significant setup configuration failure detected | error |
35004 | Unrecoverable problem occurred | System detects unrecoverable error condition | critical |
35005 | Unknown setup problem occurred | Unidentified setup configuration error detected | error |
35006 | System is overheating | Temperature sensor detects excessive system heat | critical |
35007 | Fan failure detected | Fan tray assembly malfunction detected | critical |
35008 | System voltage fault detected | System voltage supply falls outside safe operating range | critical |
35009 | Power supply voltage fault | Power supply voltage output exceeds or falls below specifications | critical |
35010 | Fan is fault and system is overheating | Both fan failure and overheating condition detected simultaneously | critical |
35011 | BEP monitor abnormal | Backplane electronics package monitoring circuit detects failure | critical |
35012 | Application error | Critical application software error detected | error |
36001 | Transducer at port removed | Transducer physically disconnected from specified port | info |
36002 | Transducer inserted at port | Transducer physically connected to specified port | info |
36003 | Indefinite transducer error | Transducer exhibits undefined or intermittent connection error | warning |
36004 | Transducer not accepted by system | Connected transducer not recognized or not compatible with system | warning |
36005 | System shutdown confirmation requested | System requests user confirmation to shutdown | info |
36006 | Minor setup problem occurred | Non-critical setup issue detected that may affect image quality | warning |
36007 | BEP voltage out of range | Backplane electronics package voltage falls outside specifications | warning |
36008 | BEP fan is abnormal | Backplane electronics package cooling fan exhibits abnormal operation | warning |
36009 | BEP temperature out of range | Backplane electronics package temperature exceeds safe limits | warning |
36010 | Transducer is disconnected | Transducer becomes disconnected during operation or initialization | info |
36011 | Enter diagnostics mode | System enters diagnostic test mode | info |
36012 | Transducer scan head is disconnected | Scan head assembly physically disconnected from system | warning |
36013 | Transducer is not supported | Connected transducer type is not compatible with system version | warning |
36014 | MPSB is broken | Motor power supply board detected as faulty or non-functional | error |
Aborted | Current scan has been aborted | User or system initiated scan abort | info |
Abort finished. Ready for next scan. | Previous scan abort completed successfully | Scan was aborted and abort process finished | info |
Aborting | Scan abort is in progress | Abort process started | info |
An application entity can only be set to Print or Store, not | Application entity configuration must be either Print OR Store, not both | Attempting to set application entity to both Print and Store simultaneously | warning |
Are you sure you want to delete all these series/images? | Deletion confirmation for multiple series or images | User selected delete for multiple series or images | warning |
Are you sure you want to delete these studies? | Deletion confirmation for multiple studies | User selected delete for multiple studies | warning |
Are you sure you want to delete this series/image? | Deletion confirmation for single series or image | User selected delete for one series or image | warning |
Are you sure you want to delete this study? | Deletion confirmation for single study | User selected delete for one study | warning |
Are you sure you want to log off? | Logout confirmation required | User selected logout option | info |
Are you sure you want to logout? | System requesting confirmation to logout current user | User selected logout option | info |
Are you sure you want to remove all calipers? | Caliper removal confirmation for all calipers | User selected remove all calipers option | info |
Automatic export items can't be deleted. | Items set for automatic export cannot be manually deleted | Attempting to delete item with automatic export enabled | info |
Can't load report data. | Report data failed to load | Attempting to open or view report | error |
Dataset is not available | Selected patient dataset cannot be accessed or retrieved | Attempting to access or re-send patient dataset | warning |
Delete confirmation | Confirmation required for delete operation | User initiated delete action | info |
DICOM connection test failed. | DICOM server connection test unsuccessful | Running DICOM connection diagnostic test | error |
DICOM connection test passed. | DICOM server connection test successful | Running DICOM connection diagnostic test | info |
Do you want to resume the last saved session? | Option to resume previous incomplete session | System startup after incomplete session | info |
Export from DICOM server is not supported. | Cannot export data FROM DICOM server; only TO DICOM server supported | Attempting to retrieve/export from DICOM server | info |
Failed to export users. | User export operation failed | Attempting to export user list or user data | error |
Failed to retrieve images from the DICOM server. | Unable to fetch images from DICOM server | Requesting image retrieval from DICOM server | error |
Imaging Engine Initializing.... | System startup sequence in progress | System power-up or imaging module initialization | info |
Invalid port number. Port number can only have four digits. | Port number must be 4 digits or fewer | Entering port number with more than 4 digits | warning |
Invalid printer, or no printer, is installed. | Printer not found or not properly configured | Attempting to print when no valid printer available | error |
Invenia ABUS 2.0 Initializing... | System startup sequence in progress | System power-up | info |
No DICOM images found at the folder <folder>. | Specified folder contains no DICOM images | Attempting to load images from folder without DICOM files | info |
No images are retrieved. | Image retrieval returned no results and non-service user cannot change settings | Non-service user attempting to change local DICOM settings | warning |
Only Service users can change the local settings. | Permission denied: only service-level users can modify local settings | Non-service user attempting to change local Port or AE Title | warning |
Only three lines of text are allowed for Institute Address. | Institute Address field limited to 3 lines maximum | Entering more than 3 lines of text in Institute Address field | warning |
[Patient Name] is already selected for this scan | Patient has already been selected for current scanning session | Attempting to select same patient twice for same scan | info |
Please check the scantype | Scan type has not been selected or is invalid | User attempted to start scan without selecting scan type | warning |
Please enter a User name. | User name field cannot be empty | Attempting to create user or login without username | warning |
Please finish the current scan! | Current scan processing must complete | Attempting action while scan processing in progress | warning |
Please input patient information. | Patient information fields are incomplete | Attempting to scan without complete patient data | warning |
Please mark all nipple positions | Nipple position markers not set on scan | Attempting to process scan without nipple markers | warning |
Please restart DICOM receive application to make the changes | DICOM receiver must be restarted for configuration changes to take effect | Changing DICOM receive application settings | info |
Please retrieve the study before you load it. | Study must be retrieved from DICOM server before loading | Attempting to load study that has not been retrieved | warning |
Please select a location to export the study(s) to. | Destination location must be selected for study export | Attempting to export without selecting export destination | warning |
Please select a new patient | New patient must be selected to begin scan | Attempting to start scan without patient selection or with invalid patient | warning |
Please select another study. | Different study must be selected | Current study selection is invalid or unavailable | warning |
Please select either Export To Disk or Export to DICOM Serve | Export destination type must be specified | Attempting export without selecting export target type | warning |
Please unlock Scan Head Assembly first | Scan Head Assembly is locked; login required | Attempting to scan with locked Scan Head Assembly | warning |
Please wait until processing is finished | System is processing current task; new action delayed | Requesting action while system processing previous task | info |
Press compression button | Compression must be applied before scan can start | Attempting to start scan without applying compression | warning |
Press Compression button or Start button | Apply compression and press Start to begin scanning | At scan start stage before compression applied | warning |
Press Scan Button on Scan Head to Start Scanning | Start button on Scan Head Assembly must be pressed to begin scan | Scan ready to start but button not pressed | info |
Press Start button to start scan | Start button must be pressed to begin scan acquisition | Scan parameters set but acquisition not initiated | info |
Printer error. | Generic printer error occurred | Attempting print operation when printer error occurs | error |
Printer error. Check the printer and try the operation again | Printer error; check printer status and retry | Printer error during print operation | error |
Processing... | System processing acquired image data for display | Scan acquisition completed; processing phase begun | info |
Report file [file name] already exists. | Report file with same name already exists in export location | Attempting to export report with filename that already exists | warning |
Retrieve completed | Study retrieval from DICOM server completed successfully | Retrieval operation finished | info |
Retrieving study <id>. | Study retrieval from DICOM server in progress | Retrieving study from DICOM server | info |
Scan finished | Scanning for current View Type has completed | Requested scan view type acquisition finished | info |
Scan limit reached | Storage or acquisition limit has been reached | System limit (storage space or scan count) exceeded | warning |
Scanning | System is actively performing scan acquisition | Scan in progress | info |
Scan order for this patient already selected | Scan order specification already made for this patient | Attempting to select scan order for patient with existing order | info |
Scan started | Scan acquisition has begun | Start button pressed and scan initiated | info |
Select scan order | Scan order workflow must be specified | Patient selected; scan order not yet specified | info |
Select scan type | Scan type (View Type) must be selected | Attempting to start scan without scan type selection | warning |
Study doesn't exist. | Requested study not found in system | Attempting to access non-existent study | warning |
The Invenia ABUS 2.0 application has encountered a problem. | Application error occurred; restart required | Unexpected application error | critical |
The Invenia ABUS 2.0 disk space is below [configured free sp | Low disk space; old studies auto-deleted to maintain free space threshold | Free disk space falls below configured threshold | warning |
There are scans missing for this patient. Are you sure you w | Selected scans not completed before attempting to finish session | User attempting to finish exam before all scheduled scans done | warning |
The Study doesn't exist. | Requested study not found | Attempting to access non-existent study | warning |
The study <id> of the patient <id> already exists. Do you wa | Study already exists; overwrite confirmation needed | Attempting to create or import study that already exists | warning |
The user name already exists. Please choose a different user | User name is already in use; must select unique name | Attempting to create user with existing user name | warning |
The user name or password you entered is incorrect. | Login credentials are invalid | Entering invalid username or password during login | warning |
The view type of the image stored in the Invenia ABUS 2.0 St | View type cannot be modified for stored images | Attempting to change view type of archived image | info |
Transducer is moving. Please wait. | Transducer in motion; user action delayed | Transducer auto-scan in progress | info |
Unable to restore study. Please load it from the study pane. | Study cannot be restored from backup; must load from study pane | Attempting to restore study backup that cannot be restored | warning |
Unauthorized to access the printer. | User account lacks permission to access printer | Attempting to print with insufficient permissions | warning |
UsysDcmReceiveApp is missing in the installation directory . | DICOM receive application component missing; reinstallation required | System startup or DICOM operation when component missing | critical |
You have images in the data pane. | Images currently present in data pane | Opening new study or closing application with images loaded | info |
You have images in the data pane. Are you sure you want to o | Confirmation needed to open new study; current images will be cleared | Attempting to open new study with images in data pane | info |
You have images pending int he DICOM queue. | Images waiting in DICOM export queue for transmission | DICOM export operations pending | info |
You must first select a patient | Patient must be selected before scanning | Attempting to scan without patient selection | warning |
You must restart Invenia ABUS 2.0 for the configuration chan | System restart required for configuration changes to apply | Configuration settings changed | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI