GE Healthcare Invenia ABUS 2.0: troubleshooting and error codes

The ServiceForge knowledge base holds 15 documented symptoms for the GE Healthcare Invenia ABUS 2.0, with 24 ranked causes, 31 diagnostic checks with test points and expected values, and 121 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Electromagnetic interference causing image noise or artifacts caution

Image contains noise patterns attributed to external electromagnetic sources

Ranked causes

  1. 1. Nearby EMI source (medical laser, scanner, cauterizer, computer, monitor, fan, microwave, light dimmer, mobile phone, wireless device, AC system, HF surgery equipment, AC/DC adapter) (common)
  2. 2. Poor grounding of power cord or power outlet (common)
  3. 3. Loose, missing, or damaged RF gaskets or covers; loose metallic parts (occasional)
  4. 4. Incorrect cable routing or non-GE specified harnesses and peripherals (occasional)
  5. 5. Network cable acting as EMI antenna (occasional)
First check

Gather information on noise conditions and timing

Test point
Operator report and environmental observation
Expected
Document presence of noise continuously, intermittently, after delay, at specific times, or at specific locations

8 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Self-generated noise from system

Noise generated internally within Invenia ABUS 2.0 unit during operation

Ranked causes

  1. 1. Thermal problem due to prolonged operation (common)
  2. 2. Software or setup issue (occasional)
  3. 3. Hardware malfunction in transducer or main unit (occasional)
First check

Restart system

Test point
System power and boot sequence
Expected
System powers down and back up successfully

2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unwanted noise in ultrasound image

Noise appears in acquired ultrasound images during scanning or processing

Ranked causes

  1. 1. Electromagnetic interference via external cables (common)
  2. 2. Grounding problems in mains power system (common)
  3. 3. Equipment switched on/off near Invenia ABUS 2.0 (occasional)
First check

Visually inspect cable connections

Test point
All external cables
Expected
All cables properly connected and shielded

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Lines visible in ultrasound image caution

Visible lines or artifacts appearing on the ultrasound image display

Ranked causes

  1. 1. Transducer orientation incorrect or transducer defective (common)
  2. 2. Acquisition board or TIP board defect (occasional)
First check

Reverse transducer and observe image

Test point
Ultrasound display
Expected
Lines move to opposite side of screen or disappear

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Blank screen

Display shows no image or content on startup

Ranked causes

  1. 1. System not powered or powered button not activated (common)
First check

Verify system is plugged in and powered on

Test point
Power button
Expected
Button lit green

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Touch screen display offset

Touch screen display is noticeably offset from actual tap location; factory calibration requires adjustment

Ranked causes

  1. 1. Touch screen calibration drift from factory settings (occasional)
First check

Verify touch screen calibration offset is noticeable

Test point
Touch screen display
Expected
Touch location matches tap point

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Image does not transfer to workstation

Ultrasound images fail to transmit from system to connected workstation

Ranked causes

  1. 1. Network connection not established or disabled (common)
First check

Verify workstation network indicator status

Test point
Workstation network indicator light
Expected
Green light

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Intermittent noise appearing periodically

Noise appears intermittently during scanning or at specific times, not continuously present

Ranked causes

  1. 1. External equipment switched on/off near system (occasional)
First check

Monitor noise occurrence pattern

Test point
Ultrasound image during scan
Expected
Noise appears only during specific times or equipment operation

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Intermittent operation, lock-ups, slow response, overheating, poor image quality caution

System exhibits intermittent function, freezes, sluggish response, elevated temperature, or degraded ultrasound image quality

Ranked causes

  1. 1. Dirty filter restricting airflow and causing thermal stress (common)
First check

Inspect and clean air filter for blockage

Test point
System air intake filter
Expected
Filter clean with unrestricted airflow

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Mechanical noise or thumping sound caution

Audible mechanical noise or thumping from system during operation

Ranked causes

  1. 1. Internal mechanical defect (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Multi-axis lock not functioning caution

Multi-axis lock feature stops working during operation

Ranked causes

  1. 1. Lock mechanism failure (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scan Head Assembly repositioning test

Moving Scan Head Assembly to different location to isolate noise source location

Ranked causes

  1. 1. Noise correlated with physical location (occasional)
First check

Reposition Scan Head Assembly

Test point
Scan Head location
Expected
Assembly moved to new location at least 6 feet away from original position

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

System overheating caution

System temperature exceeds normal operating range

Ranked causes

  1. 1. Dirty or blocked air filter restricting airflow (common)
First check

Inspect and clean air filter

Test point
System air filter
Expected
Filter clean and unrestricted

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

System powers on with error displayed

System starts but shows error message on screen

Ranked causes

  1. 1. Software or initialization error (occasional)
First check

Attempt system reboot

Test point
Power button
Expected
System restarts without error

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

System won't shut down normally

System does not respond to standard power button press for shutdown

Ranked causes

  1. 1. Software or power control unresponsive (occasional)
First check

Attempt normal power button shutdown

Test point
Power button
Expected
System gracefully shuts down

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
11001Failed to initialize acquisitionMissing frontend JSON file or possible transducer disconnectionerror
12001Scanhead id does not matchScanhead identifier mismatch during initializationerror
13001Image renderer initialization failedRenderer subsystem fails to initializeerror
14001GPU max temperature exceededGPU temperature exceeds maximum thresholdcritical
14002Power max exceededSystem power consumption exceeds maximum limitcritical
21001Initialization successScannerApp initialization completes successfullyinfo
21002Weight motor init home failedWeight motor fails to reach home position during initializationerror
21003Scan motor init home failedScan motor fails to reach home position during initializationerror
21004Ball motor init home failedBall joint motor fails to reach home position during initializationerror
21005Scanhead controller USB COM port not detectedScanhead controller USB communication port unavailable at startuperror
21006Weight controller USB COM port not detectedWeight controller USB communication port unavailable at startuperror
21007Scanhead and weight controller USB COM port not detectedBoth scanhead and weight controller USB ports unavailable at startuperror
22001Scanhead controller USB COM port not detectedScanhead controller USB port unavailable after initializationerror
22002Weight controller USB COM port not detectedWeight controller USB port unavailable after initializationerror
31000Unknown GEUS error code receivedUnrecognized GEUS error code detectedunknown
31001GEUS fatal error requiring system shutdownCritical GEUS subsystem failure detectedcritical
31002GEUS fatal error requiring system restartCritical GEUS error that requires system restart to recovercritical
35001Transmit overloadUltrasound transmitter power exceeds safe limitscritical
35002Setup problem occurredConfiguration issue detected during system setuperror
35003Major setup problem occurredSignificant setup configuration failure detectederror
35004Unrecoverable problem occurredSystem detects unrecoverable error conditioncritical
35005Unknown setup problem occurredUnidentified setup configuration error detectederror
35006System is overheatingTemperature sensor detects excessive system heatcritical
35007Fan failure detectedFan tray assembly malfunction detectedcritical
35008System voltage fault detectedSystem voltage supply falls outside safe operating rangecritical
35009Power supply voltage faultPower supply voltage output exceeds or falls below specificationscritical
35010Fan is fault and system is overheatingBoth fan failure and overheating condition detected simultaneouslycritical
35011BEP monitor abnormalBackplane electronics package monitoring circuit detects failurecritical
35012Application errorCritical application software error detectederror
36001Transducer at port removedTransducer physically disconnected from specified portinfo
36002Transducer inserted at portTransducer physically connected to specified portinfo
36003Indefinite transducer errorTransducer exhibits undefined or intermittent connection errorwarning
36004Transducer not accepted by systemConnected transducer not recognized or not compatible with systemwarning
36005System shutdown confirmation requestedSystem requests user confirmation to shutdowninfo
36006Minor setup problem occurredNon-critical setup issue detected that may affect image qualitywarning
36007BEP voltage out of rangeBackplane electronics package voltage falls outside specificationswarning
36008BEP fan is abnormalBackplane electronics package cooling fan exhibits abnormal operationwarning
36009BEP temperature out of rangeBackplane electronics package temperature exceeds safe limitswarning
36010Transducer is disconnectedTransducer becomes disconnected during operation or initializationinfo
36011Enter diagnostics modeSystem enters diagnostic test modeinfo
36012Transducer scan head is disconnectedScan head assembly physically disconnected from systemwarning
36013Transducer is not supportedConnected transducer type is not compatible with system versionwarning
36014MPSB is brokenMotor power supply board detected as faulty or non-functionalerror
AbortedCurrent scan has been abortedUser or system initiated scan abortinfo
Abort finished. Ready for next scan.Previous scan abort completed successfullyScan was aborted and abort process finishedinfo
AbortingScan abort is in progressAbort process startedinfo
An application entity can only be set to Print or Store, notApplication entity configuration must be either Print OR Store, not bothAttempting to set application entity to both Print and Store simultaneouslywarning
Are you sure you want to delete all these series/images?Deletion confirmation for multiple series or imagesUser selected delete for multiple series or imageswarning
Are you sure you want to delete these studies?Deletion confirmation for multiple studiesUser selected delete for multiple studieswarning
Are you sure you want to delete this series/image?Deletion confirmation for single series or imageUser selected delete for one series or imagewarning
Are you sure you want to delete this study?Deletion confirmation for single studyUser selected delete for one studywarning
Are you sure you want to log off?Logout confirmation requiredUser selected logout optioninfo
Are you sure you want to logout?System requesting confirmation to logout current userUser selected logout optioninfo
Are you sure you want to remove all calipers?Caliper removal confirmation for all calipersUser selected remove all calipers optioninfo
Automatic export items can't be deleted.Items set for automatic export cannot be manually deletedAttempting to delete item with automatic export enabledinfo
Can't load report data.Report data failed to loadAttempting to open or view reporterror
Dataset is not availableSelected patient dataset cannot be accessed or retrievedAttempting to access or re-send patient datasetwarning
Delete confirmationConfirmation required for delete operationUser initiated delete actioninfo
DICOM connection test failed.DICOM server connection test unsuccessfulRunning DICOM connection diagnostic testerror
DICOM connection test passed.DICOM server connection test successfulRunning DICOM connection diagnostic testinfo
Do you want to resume the last saved session?Option to resume previous incomplete sessionSystem startup after incomplete sessioninfo
Export from DICOM server is not supported.Cannot export data FROM DICOM server; only TO DICOM server supportedAttempting to retrieve/export from DICOM serverinfo
Failed to export users.User export operation failedAttempting to export user list or user dataerror
Failed to retrieve images from the DICOM server.Unable to fetch images from DICOM serverRequesting image retrieval from DICOM servererror
Imaging Engine Initializing....System startup sequence in progressSystem power-up or imaging module initializationinfo
Invalid port number. Port number can only have four digits.Port number must be 4 digits or fewerEntering port number with more than 4 digitswarning
Invalid printer, or no printer, is installed.Printer not found or not properly configuredAttempting to print when no valid printer availableerror
Invenia ABUS 2.0 Initializing...System startup sequence in progressSystem power-upinfo
No DICOM images found at the folder <folder>.Specified folder contains no DICOM imagesAttempting to load images from folder without DICOM filesinfo
No images are retrieved.Image retrieval returned no results and non-service user cannot change settingsNon-service user attempting to change local DICOM settingswarning
Only Service users can change the local settings.Permission denied: only service-level users can modify local settingsNon-service user attempting to change local Port or AE Titlewarning
Only three lines of text are allowed for Institute Address.Institute Address field limited to 3 lines maximumEntering more than 3 lines of text in Institute Address fieldwarning
[Patient Name] is already selected for this scanPatient has already been selected for current scanning sessionAttempting to select same patient twice for same scaninfo
Please check the scantypeScan type has not been selected or is invalidUser attempted to start scan without selecting scan typewarning
Please enter a User name.User name field cannot be emptyAttempting to create user or login without usernamewarning
Please finish the current scan!Current scan processing must completeAttempting action while scan processing in progresswarning
Please input patient information.Patient information fields are incompleteAttempting to scan without complete patient datawarning
Please mark all nipple positionsNipple position markers not set on scanAttempting to process scan without nipple markerswarning
Please restart DICOM receive application to make the changesDICOM receiver must be restarted for configuration changes to take effectChanging DICOM receive application settingsinfo
Please retrieve the study before you load it.Study must be retrieved from DICOM server before loadingAttempting to load study that has not been retrievedwarning
Please select a location to export the study(s) to.Destination location must be selected for study exportAttempting to export without selecting export destinationwarning
Please select a new patientNew patient must be selected to begin scanAttempting to start scan without patient selection or with invalid patientwarning
Please select another study.Different study must be selectedCurrent study selection is invalid or unavailablewarning
Please select either Export To Disk or Export to DICOM ServeExport destination type must be specifiedAttempting export without selecting export target typewarning
Please unlock Scan Head Assembly firstScan Head Assembly is locked; login requiredAttempting to scan with locked Scan Head Assemblywarning
Please wait until processing is finishedSystem is processing current task; new action delayedRequesting action while system processing previous taskinfo
Press compression buttonCompression must be applied before scan can startAttempting to start scan without applying compressionwarning
Press Compression button or Start buttonApply compression and press Start to begin scanningAt scan start stage before compression appliedwarning
Press Scan Button on Scan Head to Start ScanningStart button on Scan Head Assembly must be pressed to begin scanScan ready to start but button not pressedinfo
Press Start button to start scanStart button must be pressed to begin scan acquisitionScan parameters set but acquisition not initiatedinfo
Printer error.Generic printer error occurredAttempting print operation when printer error occurserror
Printer error. Check the printer and try the operation againPrinter error; check printer status and retryPrinter error during print operationerror
Processing...System processing acquired image data for displayScan acquisition completed; processing phase beguninfo
Report file [file name] already exists.Report file with same name already exists in export locationAttempting to export report with filename that already existswarning
Retrieve completedStudy retrieval from DICOM server completed successfullyRetrieval operation finishedinfo
Retrieving study <id>.Study retrieval from DICOM server in progressRetrieving study from DICOM serverinfo
Scan finishedScanning for current View Type has completedRequested scan view type acquisition finishedinfo
Scan limit reachedStorage or acquisition limit has been reachedSystem limit (storage space or scan count) exceededwarning
ScanningSystem is actively performing scan acquisitionScan in progressinfo
Scan order for this patient already selectedScan order specification already made for this patientAttempting to select scan order for patient with existing orderinfo
Scan startedScan acquisition has begunStart button pressed and scan initiatedinfo
Select scan orderScan order workflow must be specifiedPatient selected; scan order not yet specifiedinfo
Select scan typeScan type (View Type) must be selectedAttempting to start scan without scan type selectionwarning
Study doesn't exist.Requested study not found in systemAttempting to access non-existent studywarning
The Invenia ABUS 2.0 application has encountered a problem.Application error occurred; restart requiredUnexpected application errorcritical
The Invenia ABUS 2.0 disk space is below [configured free spLow disk space; old studies auto-deleted to maintain free space thresholdFree disk space falls below configured thresholdwarning
There are scans missing for this patient. Are you sure you wSelected scans not completed before attempting to finish sessionUser attempting to finish exam before all scheduled scans donewarning
The Study doesn't exist.Requested study not foundAttempting to access non-existent studywarning
The study <id> of the patient <id> already exists. Do you waStudy already exists; overwrite confirmation neededAttempting to create or import study that already existswarning
The user name already exists. Please choose a different userUser name is already in use; must select unique nameAttempting to create user with existing user namewarning
The user name or password you entered is incorrect.Login credentials are invalidEntering invalid username or password during loginwarning
The view type of the image stored in the Invenia ABUS 2.0 StView type cannot be modified for stored imagesAttempting to change view type of archived imageinfo
Transducer is moving. Please wait.Transducer in motion; user action delayedTransducer auto-scan in progressinfo
Unable to restore study. Please load it from the study pane.Study cannot be restored from backup; must load from study paneAttempting to restore study backup that cannot be restoredwarning
Unauthorized to access the printer.User account lacks permission to access printerAttempting to print with insufficient permissionswarning
UsysDcmReceiveApp is missing in the installation directory .DICOM receive application component missing; reinstallation requiredSystem startup or DICOM operation when component missingcritical
You have images in the data pane.Images currently present in data paneOpening new study or closing application with images loadedinfo
You have images in the data pane. Are you sure you want to oConfirmation needed to open new study; current images will be clearedAttempting to open new study with images in data paneinfo
You have images pending int he DICOM queue.Images waiting in DICOM export queue for transmissionDICOM export operations pendinginfo
You must first select a patientPatient must be selected before scanningAttempting to scan without patient selectionwarning
You must restart Invenia ABUS 2.0 for the configuration chanSystem restart required for configuration changes to applyConfiguration settings changedinfo

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.