GE Healthcare Giraffe: troubleshooting and error codes

The ServiceForge knowledge base holds 154 documented symptoms for the GE Healthcare Giraffe, with 314 ranked causes, 138 diagnostic checks with test points and expected values, and 93 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Device will not power up with alarm sound

Device fails to power up and alarm sounds

Ranked causes

  1. 1. Defective scale (common)
  2. 2. Defective scale harness (occasional)
  3. 3. Defective servo oxygen board (occasional)
  4. 4. Defective servo oxygen cable (occasional)
  5. 5. Defective PF battery (occasional)
  6. 6. Defective power supply (occasional)
  7. 7. Defective power harness to SBC (occasional)
  8. 8. Defective SBC (occasional)
  9. 9. Defective control board (rare)

9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

System Failure 1 displayed

No Timer2 error; clock battery or scale/harness/Servo Oxygen board/cable/ribbon cable defects indicated

Ranked causes

  1. 1. Missing or bad clock battery (common)
  2. 2. Defective scale (occasional)
  3. 3. Defective scale harness (occasional)
  4. 4. Defective Servo Oxygen board (occasional)
  5. 5. Defective Servo Oxygen cable (occasional)
  6. 6. Defective PF Battery (occasional)
  7. 7. Defective 14 pin ribbon cable between control board and SBC (occasional)
  8. 8. Defective SBC (occasional)
  9. 9. Defective control board (occasional)

9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Humidity Failure Alarm

Motor Drive Failure alarm is on; opens safety relay 2 and disables humidifier

Ranked causes

  1. 1. Elevating base does not function; safety relay 2 not energized (common)
  2. 2. Humidifier safety thermostat open (common)
  3. 3. Defective humidifier heater or humidifier safety thermostat (occasional)
  4. 4. No mains voltage at humidifier isolation transformer input (occasional)
  5. 5. Defective fuse on relay board rev 10 or higher (occasional)
  6. 6. Defective fuse on relay board rev 9 or lower (occasional)
  7. 7. Defective relay board (rare)
  8. 8. Humidity sensor failed (common)
First check

Check safety relay 2 control signal continuity

Test point
50-pin ribbon cable, pin 34
Expected
Continuity present

9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

System Failure 15 displayed high voltage

Safety Relay 1 test failure; unable to turn heater on or off; sensor, cable, transformer, or board defect

Ranked causes

  1. 1. If error occurs during calibration, J1, J2, and J4 not unplugged or JP1 not installed properly (common)
  2. 2. Air temperature sensor is above 40C at power-up (occasional)
  3. 3. Defective sensor in compartment air probe (occasional)
  4. 4. Defective compartment sensor cable (occasional)
  5. 5. No output from heater isolation transformer or defective relay board (common)
  6. 6. Defective heater isolation transformer (occasional)
  7. 7. Defective DAC circuit on control board (occasional)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Elevating base will not go up or down

Bed height adjustment mechanism does not respond to controls

Ranked causes

  1. 1. Pedal switches disabled on operator or service screen (common)
  2. 2. Defective toroidal transformer (occasional)
  3. 3. No mains input to toroidal transformer (occasional)
  4. 4. Defective fuse on relay board (occasional)
  5. 5. Defective 50-pin ribbon cable (occasional)
  6. 6. Defective relay board (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen output below setpoint patient safety

Sensor1 reading more than 3% below oxygen set point 7 minutes after change

Ranked causes

  1. 1. Low O2 supply pressure or flow (common)
  2. 2. Air leaks into patient compartment (common)
  3. 3. Supply valve not opening (occasional)
  4. 4. End cap safety valve not opening (occasional)
  5. 5. Calibration valve not closing (occasional)
  6. 6. Patient compartment vents occluded (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed Heater Failure Alarm caution

Software turned on incubator heater but it is not turning on or current sense circuit not detecting it

Ranked causes

  1. 1. Bed safety thermostat open (device shut off when heater was hot) (common)
  2. 2. Defective incubator heater, thermostat, or wire harness (occasional)
  3. 3. Defective incubator solid state relay (SSR) (occasional)
  4. 4. Defective 50-pin ribbon cable between relay and control board (occasional)
  5. 5. Defective relay board (rare)
First check

Allow device to cool if heater was recently hot

Test point
Incubator
Expected
Temperature below 40ºC, below

2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fan Failure Alarm

Fan rotation is not being detected

Ranked causes

  1. 1. Fan not turning (common)
  2. 2. Fan or fan hub not seated properly (common)
  3. 3. Defective fan motor (occasional)
  4. 4. Defective relay board (occasional)
  5. 5. Optical fan sensor not detecting fan (occasional)
First check

Verify fan hub seating on shaft

Test point
Fan hub and shaft
Expected
Hub fully seated down on shaft

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

ADD WATER message stays on after water added

Add water warning message persists on display even after reservoir is refilled

Ranked causes

  1. 1. Defective reservoir switch (common)
  2. 2. Defective 50-pin ribbon cable (occasional)
  3. 3. Defective relay board (occasional)
  4. 4. Defective control board (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Black or dim display

Display is very dark or completely black

Ranked causes

  1. 1. LCD backlight cable not fully seated (common)
  2. 2. Defective LCD backlight cable (occasional)
  3. 3. Defective LCD (occasional)
  4. 4. Defective SBC (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display appears white or has distorted image caution

Display shows white screen or image is distorted with improper colors or pixelation.

Ranked causes

  1. 1. LCD LVDS cable not fully seated (common)
  2. 2. Defective LCD LVDS cable (occasional)
  3. 3. Defective LCD display (occasional)
  4. 4. Defective single board computer (SBC) (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display is black or dim caution

Display shows completely black or very dim/dim image. Brightness is too low to see content.

Ranked causes

  1. 1. LCD backlight cable not fully seated (common)
  2. 2. Defective LCD backlight cable (occasional)
  3. 3. Defective LCD display (occasional)
  4. 4. Defective single board computer (SBC) (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display stuck on Giraffe logo screen

Display shows Giraffe logo and does not proceed to operating screen; alarm LED indicators working at power on

Ranked causes

  1. 1. 14-pin ribbon cable disconnected (common)
  2. 2. Defective ribbon cable (occasional)
  3. 3. Defective SBC (occasional)
  4. 4. Defective control board (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen sensor readings out of range patient safety

Oxygen probe failure - sensors out of range or difference >3%

Ranked causes

  1. 1. Defective sensor (common)
  2. 2. Leak or occlusion in sensor housing (common)
  3. 3. Defective cable or connection (occasional)
  4. 4. Defective Servo O2 board (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scale icon does not appear on control panel

No scale indicator on display

Ranked causes

  1. 1. Scale not plugged in (common)
  2. 2. Defective cable (occasional)
  3. 3. Defective circuit board (occasional)
  4. 4. Defective scale receptacle (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor temperature out of range

Sensor plug thermistor temperature reading outside 15C-55C

Ranked causes

  1. 1. Device is too cold (common)
  2. 2. Cooling fan not running (occasional)
  3. 3. Defective sensor plug thermistor or cable (occasional)
  4. 4. Defective Servo O2 board (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

System Failure 16 displayed high voltage

Incubator heater could not be turned off during Power-on Self Test; SSR or relay board defect

Ranked causes

  1. 1. Defective solid state relay (SSR) (common)
  2. 2. Defective incubator heater SSR (common)
  3. 3. Defective current sense circuit on relay board (occasional)
  4. 4. Defective 50-pin ribbon cable (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

System Failure 3 displayed

ADC calibration detected a channel out of range; scale, harness, or control board defect

Ranked causes

  1. 1. Defective scale (occasional)
  2. 2. Defective scale harness (occasional)
  3. 3. No line frequency signal or Timer2 of microprocessor defective (occasional)
  4. 4. Defective circuit on control board (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touch screen does not function

User interface touch screen not responding to input

Ranked causes

  1. 1. Foreign substance on touch screen surface (common)
  2. 2. Touch controller flex cable not fully seated (common)
  3. 3. Touch controller USB cable not fully seated (common)
  4. 4. Defective touch controller (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

User interface touch screen does not function caution

Touch screen is unresponsive to user input or touches. Interface controls cannot be accessed via touch.

Ranked causes

  1. 1. Foreign substance on touch screen surface (common)
  2. 2. Touch controller flex cable connection not fully seated (common)
  3. 3. Touch controller USB cable connection not fully seated (common)
  4. 4. Defective touch controller (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

White display or distorted image

Display shows white screen or image is garbled/distorted

Ranked causes

  1. 1. LCD LVDS cable not fully seated (common)
  2. 2. Defective LCD LVDS cable (occasional)
  3. 3. Defective LCD (occasional)
  4. 4. Defective SBC (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Baby Hot alarm activates patient safety

Alarm sounds when patient skin temperature rises above set point; audio suspends for 5-12 minutes when silenced depending on temperature deviation

Ranked causes

  1. 1. Baby mode set temperature is incorrect for patient (common)
  2. 2. External heat such as phototherapy or sunlight (occasional)
  3. 3. Probe disconnected from skin (occasional)
First check

Verify baby mode set temperature setting

Test point
Setup menu
Expected
Temperature appropriate for patient

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check SpO2 patient safety

System detects a problem with the SpO2 probe. Display/Light notification with 2 min silence time, medium priority escalating to high after 10 sec.

Ranked causes

  1. 1. Defective probe (occasional)
  2. 2. Interference (occasional)
  3. 3. Too much ambient light (occasional)
First check

Assess ambient light level around probe

Test point
Probe attachment area
Expected
Reduced ambient light exposure

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Device will not turn on (power fail alarm)

Device does not power up and power fail alarm sounds

Ranked causes

  1. 1. No AC power (common)
  2. 2. No 12V DC power (common)
  3. 3. Defective power supply (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fan motor not operating or running slowly caution

Incubator fan does not rotate or rotates slowly; poor air circulation; high internal temperature.

Ranked causes

  1. 1. Fan motor defective or bearing worn (common)
  2. 2. Fan blade or hub stuck or damaged (occasional)
  3. 3. Optical fan sensor malfunction (occasional)
First check

Rotate fan blade by hand (power off)

Test point
Fan blade on motor shaft
Expected
Fan rotates freely without grinding or resistance

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Hands-free alarm silence does not work

Hands-free alarm silence function not responding

Ranked causes

  1. 1. Too much ambient sunlight (common)
  2. 2. Defective alarm board (occasional)
  3. 3. Defective LINCAN cable (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Heater not heating or heating inconsistently caution

Incubator heater or humidifier heater not producing heat or operating erratically

Ranked causes

  1. 1. Safety relay 1 not switching heater isolation transformer primary (occasional)
  2. 2. Heater current sense circuit failure (rare)
  3. 3. Solid state relay (SSR) failure (occasional)
First check

Verify safety relay 1 energization

Test point
Safety relay 1 coil
Expected
Energized when heater demand is active

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No power fail indicator (red LED)

Power fail red LED is not illuminated

Ranked causes

  1. 1. Display power cable disconnected (common)
  2. 2. Defective display power cable (occasional)
  3. 3. Defective alarm board (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Patient Cold alarm patient safety

Temperature measured by skin probe is 1.0°C below set temperature (or 0.5°C if configured); requires Skin Temperature alarm enabled

Ranked causes

  1. 1. Baby Mode set temperature incorrect for patient (common)
  2. 2. Probe disconnected from skin (common)
  3. 3. Patient is cold (common)
First check

Verify probe attachment to skin

Test point
Patient skin
Expected
Probe firmly attached

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Patient Hot alarm patient safety

Temperature measured by skin probe is 1.0°C above set temperature (or 0.5°C if configured); requires Skin Temperature alarm enabled

Ranked causes

  1. 1. Baby Mode set temperature incorrect for patient (common)
  2. 2. External heat such as phototherapy or sunlight (occasional)
  3. 3. Skin temperature probe disconnected from skin (common)
First check

Verify skin temperature probe attachment

Test point
Patient skin
Expected
Probe firmly attached to skin

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power Failure (LED Indicator)

No AC power into device or 5V power supply not present despite AC power

Ranked causes

  1. 1. On/Off switch not turned on (common)
  2. 2. Defective fuse F1 or F2 on relay board (common)
  3. 3. Defective power supply (occasional)
First check

Verify power cord plugged and switch on

Test point
Power cord, on/off switch, back panel circuit breaker
Expected
Cord in live outlet, switch/breaker on

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

RH Sensor calibration failed

Message displayed when RH sensor signal is outside expected range (65–85% RH), indicating either sensor defect or faulty calibration bottle.

Ranked causes

  1. 1. RH sensor signal out of range 65–85% RH (common)
  2. 2. RH sensor defective (occasional)
  3. 3. Calibration bottle defective (occasional)
First check

Verify calibration setup conditions

Test point
RH sensor environment
Expected
Proper sensor placement, ambient conditions stable

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

RH Sensor Calibration Failed error

Calibration procedure fails with error message indicating RH sensor signal outside expected range

Ranked causes

  1. 1. RH sensor signal outside expected range at 65-85% RH (common)
  2. 2. Calibration bottle defective (occasional)
  3. 3. RH sensor defective (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scale does not register weight or SC value does not respond to applied weight patient safety

When standard weight is placed on scale, displayed SC value does not increase or increases by incorrect amount

Ranked causes

  1. 1. Scale cable routed incorrectly, bent, or pinched (common)
  2. 2. Scale tray obstruction from blankets or cushions (common)
  3. 3. Scale sensor malfunction or connector disconnection (occasional)
First check

Verify all four shipping screws are removed from screw holes

Test point
Scale assembly screw holes
Expected
All four screw holes empty, no shipping screws present

5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scale icon disappears/reappears with clicking

Scale indicator flickers with repeated clicking sound

Ranked causes

  1. 1. Defective or disconnected cable (common)
  2. 2. Defective circuit board (occasional)
  3. 3. Defective scale receptacle (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Screen refreshes every few seconds with scale attached

Display constantly refreshes when scale is connected

Ranked causes

  1. 1. Defective circuit on control board (occasional)
  2. 2. Defective scale (occasional)
  3. 3. Defective scale harness (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Servo oxygen system does not calibrate patient safety

System fails to complete oxygen sensor calibration after user initiates procedure; calibration valve may not open or fan may not activate

Ranked causes

  1. 1. Calibration valve stuck closed (unknown)
  2. 2. Calibration fan not activating (unknown)
  3. 3. Oxygen sensors degraded or absent (unknown)
First check

Verify two oxygen sensors are physically present

Test point
Sensor housing beneath bed
Expected
Two fuel cell oxygen sensors installed

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

System Failure 8 displayed

No Line Frequency error; 120 hertz signal missing from control board or relay board circuit defect

Ranked causes

  1. 1. Defective control board (common)
  2. 2. Defective 50-pin ribbon cable between the relay board and control board (occasional)
  3. 3. Defective relay board (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm Board not illuminating or functioning

Red LED not lighting during power fail; beacon lights not responding to commands; buzzer not activating on CAN communication loss

Ranked causes

  1. 1. +5V DC STBY power not available to Alarm Board (common)
  2. 2. Alarm Board CAN bus communication failure (occasional)
First check

Measure +5V DC STBY voltage at Alarm Board

Test point
Alarm Board power input
Expected
+4.5 to +5.5V DC, +/- 0.5V

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Baby Cold alarm activated patient safety

Audio alarm with display and light notification when patient skin temperature falls below set temperature in Baby or Manual mode

Ranked causes

  1. 1. Baby mode set temperature is incorrect for patient (unknown)
  2. 2. Probe disconnected from skin (unknown)
First check

Verify skin temperature probe attachment to patient skin

Test point
skin temperature probe contact point
Expected
Probe firmly attached to patient skin

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Baby Cold alarm activates patient safety

Alarm sounds on display/light when patient skin temperature falls below set temperature in Baby or Manual mode. Audio suspends for 5–12 minutes when silenced depending on temperature deviation.

Ranked causes

  1. 1. Baby mode set temperature is incorrect for patient (common)
  2. 2. Probe disconnected from skin (common)
First check

Measure patient skin temperature with probe

Test point
Skin temperature probe
Expected
1.0°C below set temperature (medium alarm) or 2.0°C below (high priority)

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed Height Pedals Disabled alarm caution

Alarm activates when bed raise or lower pedals are pressed but have been disabled on Service or Setup screen

Ranked causes

  1. 1. Bed pedal disabled in Service Mode (common)
  2. 2. Bed pedal disabled in Setup screen (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed Raise/Lower Pedal Failure Alarm caution

One of the bed Raise/Lower pedal switches is shorted

Ranked causes

  1. 1. Defective bed Raise/Lower pedal switch or harness (common)
  2. 2. Defective relay board (occasional)
First check

Check switch continuity at connectors

Test point
Bed Raise/Lower pedal switch connectors
Expected
Correct continuity state for switch position

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed will not lock in tilt position patient safety

Tilt ball not engaged in finger pocket latch. Bed platform cannot be secured in tilted position.

Ranked causes

  1. 1. Tilt ball not engaged in finger pocket latch (common)
  2. 2. Tilt brake pad failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Black display

Display shows no image or is completely black with high priority audio alarm active at power on

Ranked causes

  1. 1. Display power cable disconnected (common)
  2. 2. Defective display power cable (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check Baby alarm activates patient safety

Alarm sounds on display/light when radiant heater has been running for 12 minutes at specific power setting in Baby or Manual mode without mode or power adjustment.

Ranked causes

  1. 1. Mode has not been changed (occasional)
  2. 2. Power setting has not been changed (occasional)
First check

Monitor patient carefully

Test point
Patient condition
Expected
Patient stable and appropriately warmed

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Compartment Sensor Failure Alarm

Software detected difference in two air thermistors exceeding 0.5°C threshold

Ranked causes

  1. 1. Defective compartment air probe or harness (common)
  2. 2. Defective control board (occasional)
First check

Measure resistance of two air thermistors

Test point
Control board J1, Pins 1-2 and 3-4
Expected
within about 100 ohms
Tools
Resistance meter

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Confirm Probe Jack Connection alarm patient safety

Skin temperature probe connection alarm activates; heater may shut down if Check Baby alarm unaddressed for 15 minutes

Ranked causes

  1. 1. Skin temperature probe is disconnected from probe panel (common)
  2. 2. Malfunctioning probe connector (occasional)
First check

Verify probe connection status

Test point
Probe Panel jack
Expected
Connector fully inserted

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Device will not switch to baby mode

Baby mode cannot be activated

Ranked causes

  1. 1. Skin temperature probe in wrong connector (common)
  2. 2. Baby mode disabled on service screen (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Device with servo-humidity shows RH as 0

Relative humidity reading shows only zero despite servo-humidity capability

Ranked causes

  1. 1. Defective humidity sensor (common)
  2. 2. Defective control board (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display is black caution

Display shows completely black screen. No display output even though system appears to be powered.

Ranked causes

  1. 1. Display power cable disconnected (common)
  2. 2. Defective display power cable (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display stuck on Giraffe logo with base movement possible

Display frozen on logo screen; E-Base can move down or radiant heater may be on even in closed-bed situation

Ranked causes

  1. 1. DC power harnesses disconnected or damaged (common)
  2. 2. DC power harnesses damaged (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Elevating base fails to raise or lower bed height caution

Bed does not respond to raise/lower pedal commands; stuck at one height or slow movement.

Ranked causes

  1. 1. Elevating base motor or actuator defective (common)
  2. 2. Elevating base connector loose or disconnected (occasional)
First check

Visually inspect J-45 connector on relay board for loose contact or corrosion

Test point
J-45 elevating base connector on relay board
Expected
Connector fully seated with no oxidation or loose pins

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Elevating base motor not operating caution

Elevating base motor does not raise or lower bed; no motor response to control commands

Ranked causes

  1. 1. Safety relay 2 not switching elevating base transformer primary (occasional)
  2. 2. Elevating base transformer secondary voltage absent or low (rare)
First check

Verify safety relay 2 energization

Test point
Safety relay 2 coil
Expected
Energized when motor control command issued

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fan Always in High Speed Alarm

Software cannot read temperature of heat sink sensor in incubator heat sink

Ranked causes

  1. 1. Defective heat sink sensor or harness (common)
  2. 2. Defective control board (occasional)
First check

Measure heat sink sensor resistance in service mode

Test point
J3 harness pins 1-2 at control board
Expected
Valid sensor resistance per specifications

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Humidifier heater circuit resistance out of range

Heater resistance measured at J53.1 to J53.3 does not match expected value for device voltage configuration

Ranked causes

  1. 1. Heater element open or shorted (common)
  2. 2. Safety thermostat stuck open or closed (occasional)
First check

Measure heater resistance at J53.1 to J53.3

Test point
J53, pins 1 and 3
Expected
72 ohms for 115V devices or 288 ohms for 230V devices
Tools
ohmmeter

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incubator heater not operating or insufficient heat caution

Bed fails to reach or maintain set temperature; heater cartridge defective or loose connection.

Ranked causes

  1. 1. Heater cartridge defective or failed (common)
  2. 2. Heater electrical connector loose or disconnected (occasional)
First check

Visually inspect heater cartridge for cracks, discoloration, or burn marks

Test point
Incubator heater element in heat sink
Expected
No visible damage; uniform color

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incubator heater resistance measurement high voltage

Verify incubator heater and bed safety thermostat resistance. Two 115V heaters rated 225W at 104V mounted at opposite ends, connected in parallel. Each heater approximately 48 ohms, parallel combination approximately 24 ohms. Bed safety thermostat in series.

Ranked causes

  1. 1. Open circuit or failed heater element (occasional)
  2. 2. Bed safety thermostat failure (rare)
First check

Measure combined heater resistance

Test point
Between harness connector pin 4 and black wire at SSR
Expected
Approximately 24 ohms (two 48-ohm heaters in parallel), 18-30 ohms
Tools
Ohmmeter, DMM

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Insufficient power supply voltage high voltage

Control board power connector voltages below specified thresholds on pins 2, 3, or 4 when measured against pin 1 (ground)

Ranked causes

  1. 1. Harness 6600-0701-700 or 2075837-001 defective (pins 2 and 3) (common)
  2. 2. Harness 6600-0701-700 defective (pin 4) (common)
First check

Measure voltage between pin 1 and pin 2

Test point
4-wire power connector, pin 1 (ground) to pin 2
Expected
more than 4.75V
Tools
digital multimeter

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No alarm LED indicators at power on

Alarm LED lights are not illuminated during power-up sequence

Ranked causes

  1. 1. Disconnected LINCAN cable between SBC and alarm board (common)
  2. 2. Defective LINCAN cable (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • No audio alarm
  • No SpO2 probe
  • Oximetry system failure
  • Oxygen calibration failed
  • Oxygen level elevated in patient compartment
  • Oxygen output above setpoint
  • Oxygen sensors operating temperature exceeds limit
  • Patient Probe 1 or 2 Failure Alarm
  • Patient Probe 1 or Patient Probe 2 Failure Alarm
  • Scale initial power-up SC value is not 10,000
  • Scale measurement error or drift
  • Scale weight reading incorrect
  • Servo oxygen system calibration prompted prematurely
  • System Failure 12 displayed
  • Temperature Out of Calibration Alarm
  • Temperature probe resistance out of specification
  • Toroidal transformer output voltage low or absent
  • Up Pedal Pressed Alarm
  • 50-pin ribbon cable continuity fault
  • Air temperature alarm exceeds 38C or 40C
  • Baby Cold alarm - high priority activation
  • Baby Cold alarm high priority triggered
  • Baby Hot alarm activation criteria and hysteresis
  • Bed Lower Pedal Pressed alarm
  • Bed Raise Pedal Pressed alarm
  • Bed warms up slowly or does not reach temperature
  • Calibration: Electrical safety and calibration
  • Calibration: Line voltage
  • Calibration: Line Voltage after relay board replacement
  • Calibration: Scale weight accuracy
  • Calibration: Scale weight measurement
  • Calibration: Servo oxygen system 21% reference point
  • Calibration: System and Humidifier after control board replacement
  • Calibration: System and line voltage
  • Callback Time Complete alarm activates
  • Center Weight Check
  • Chassis cover difficult to remove or misaligned
  • Check Baby - Heat Off alarm
  • Clock battery replacement
  • Confirm Probe Jack Connection alarm activates (Baby mode)
  • Confirm Probe Jack Connection alarm activates (Manual mode)
  • Device always powers up in service screen mode
  • Electrostatic discharge damage to connectors
  • Fan motor replacement
  • Fan running at high speed without message
  • Heat sink sensor thermistor resistance abnormal
  • High pulse rate alarm
  • High SpO2 alarm
  • Hood removal for replacement
  • Humidifier reservoir crazed
  • Humidifier reservoir has scaling
  • Load Cell Check
  • Loss of mains power
  • Low pulse rate alarm
  • Low SpO2 alarm
  • Motor Drive Failure Alarm
  • No HFAS indicator (green LED)
  • No red alarm indicators
  • No white alarm indicators
  • No yellow alarm indicators
  • Noise from speaker on power down
  • Off Center Weight Check
  • Oxygen system not calibrated
  • Oxygen system sensor replacement
  • Oxygen system supply valve leak test
  • Oxygen system vent screen replacement
  • Patient cold or patient hot alarm does not trigger
  • Patient hot or patient cold alarm in manual mode
  • Power Failure battery replacement
  • Power loss detected
  • Power supply harness cable troubleshooting
  • Power supply not providing correct voltage
  • Power supply voltage measurement
  • Raising and lowering elevating base in service mode
  • Scale cable routing or pinching issue
  • Scale cannot be calibrated - times out
  • Scale does not return weight
  • Scale exceeds maximum weight in run mode
  • Scale gives consistently inaccurate weights
  • Scale gives erratic weight readings
  • Scale slow to return weight
  • Scale stuck on screen >45 seconds
  • Shipping screws not removed from scale
  • Silenced alarm reactivates after silence time expires
  • Software upgrade fails
  • SpO2 probe off baby
  • System Failure 13 displayed
  • System Failure 14 displayed
  • System Failure 2 displayed
  • System Failure 5 displayed
  • System Failure 7 displayed
  • System Failure alarm activates
  • Temp Probe Failure alarm activates
  • Weight displays as SC: 40000 on service screen

94 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
70System Failure 70Unknownunknown
71System Failure 71Unknownunknown
81System Failure 81Unknownunknown
82System Failure 82Unknownunknown
83Unexpected SBC Software problemSBC software errorerror
85System Failure 85Unknownunknown
98OverflowOverflow conditionunknown
99Logic problem on the control boardControl board logic errorerror
Baby ColdPatient skin temperature has fallen below set temperature threshold in Baby or Manual modeMeasured skin temperature 1.0°C below set temperature (medium priority) or 2.0°C below set temperature (high priority)warning
Baby hotPatient skin temperature has risen above the set point thresholdTemperature measured by skin temperature probe is 1.0°C above set temperature (medium priority) or 2.0°C above set temperature (high priority)warning
Bed Heater FailureIncubator heater not turning on or current sense circuit not detecting itHeater commanded on but not operating or current not detectederror
Bed Height Pedals DisabledBed pedal pressed but pedals are disabled in Service or Setup screenRaise or lower pedal pressed while disabledinfo
Bed Lower Pedal PressedBed lower pedal pressed during device power-upLower pedal pressed while turning on deviceinfo
Bed Raise/Lower Pedal FailureBed Raise/Lower pedal switch is shortedSwitch continuity indicates short conditionerror
Bed Raise Pedal PressedBed raise pedal pressed during device power-upRaise pedal pressed while turning on deviceinfo
Calibration failedSensors not reading between 23-55mV after 5 minutes in calibration modeCalibration procedure unable to completeerror
Callback Time CompleteCallback timer has reached zero and intervention is dueCallback timer reaches zerowarning
Check BabyRadiant heater has been running continuously at high power for extended period without adjustmentHeater running at 100% power (Baby mode) or >25% power (Manual mode) for >12 minuteswarning
Check Baby - Heat OffCheck Baby alarm has remained unaddressed for 15 minutes; heater automatically shuts downRadiant heater running at 100% power for >15 minutes with Check Baby alarm unaddressed for 3 minutes (Baby mode) or heater at >25% power for >12 minutes with Check Baby unaddressed for 3 minutes (Manual mode)error
Check O2 SupplyAfter calibration, sensor1 reading not >23% after 20 seconds with both valves openOxygen supply verification failure after calibrationwarning
Check SpO2System detects a problem with the SpO2 probeDefective probe, interference, or too much ambient light detectedwarning
Compartment Sensor FailureSoftware detected difference in two air thermistors exceeding 0.5°CDifference in air thermistors > 0.5ºCwarning
Confirm Probe Jack ConnectionSkin temperature probe connection to Probe Panel should be checkedActivates in Baby mode when skin temperature probe is unplugged; also in Manual mode when probe is unplugged and Man Temp Alarm enabledwarning
Confirm Probe Jack Connection (Baby)No probe detected in Baby modeNo probe plugged into probe jackwarning
Confirm Probe Jack Connection (Manual)No probe detected with Manual temp alarm enabledNo probe plugged into probe jack and Man temp Alarm setwarning
Fan Always in High SpeedSoftware cannot read temperature of heat sink sensorHeat sink sensor not responding or invalid readingerror
Fan FailureFan rotation not being detectedOptical fan sensor not detecting fan rotationerror
FiO2>26%Elevated oxygen levels in patient compartment not controlled by servo O2 systemOxygen level exceeds 26% during servo controlwarning
High OxygenSensor1 reading more than 3% above set pointSensor detects oxygen above desired levelwarning
High Pulse rateBaby's pulse rate is above set limitPulse rate above the set valuewarning
High SpO2Baby's blood oxygen saturation is above set levelBlood oxygen saturation above the set valuewarning
Humidity FailureMotor drive failure or humidifier issue; humidifier disabledVarious humidity system faultserror
Low OxygenSensor1 reading is more than 3% below oxygen set point7 minutes after a set point changewarning
Low Pulse rateBaby's pulse rate falls below set limitPulse rate below the set valuewarning
Low SpO2Baby's blood oxygen saturation falls below set levelBlood oxygen saturation below the set valuewarning
Mains voltage calibration failedEntered line voltage differs by 20% or more from measured valueUser enters calibration value that deviates >20% from actual measured mains voltageerror
Motor Drive FailureElevating base motor running when not commandedMotor detected running without commanderror
No SpO2 probeNo SpO2 probe is detected at the probe panel connectorSystem does not detect an SpO2 probe at the probe panelwarning
O2 Cal Lost-No O2Servo oxygen never calibrated, system will not operateSystem detects no calibration data on startupcritical
Oximetry system failureOximetry system has failedCommunication between SpO2 and warmer system is losterror
Oxygen Probe FailureSensor reading out of range (23-280mV) or difference between sensors >3%Sensor(s) malfunction or connection failureerror
Oxygen System Failure 1Sensor plug thermistor temperature out of range 15C-55CTemperature sensor reading outside valid rangeerror
Oxygen System Failure 2Humidity reading out of valid range 1-99Humidity sensor malfunctionerror
Oxygen System Failure 3Checksum test failed during power-upMicrocontroller self-test failurecritical
Oxygen System Failure 4Analog to digital converter self test failedADC reading of Vtest outside 1.216V-1.254Vcritical
Oxygen System Failure 5Check O2 Supply alarm silenced twice, sensor1 still not >23%Persistent oxygen supply inadequacyerror
Oxygen System Failure 6Giraffe control board cannot communicate with Servo O2 boardCommunication error on system startup or during operationcritical
Oxygen System Failure 7Watchdog circuit problem detected during power-upPower-up self-test failurecritical
Patient Cold - Check Probe 1Temperature measured by skin probe is 1.0°C below set temperatureSkin temperature < set temperature - 1.0ºC (or 0.5ºC if configured)warning
Patient Hot - Check Probe 1Temperature measured by skin probe is 1.0°C above set temperatureSkin temperature > set temperature + 1.0ºC (or 0.5ºC if configured)warning
Patient Probe 1 Failure or Patient Probe 2 FailureProbe thermistors show difference exceeding 0.5°CThermistor difference > 0.5ºCwarning
Power Failure (LED Indicator)No AC power into device or 5V power supply not presentNo AC power or 5V supply failurecritical
Replace Clock BatteryClock battery voltage too lowBattery voltage falls below threshold to maintain timeinfo
SpO2 probe off babySpO2 probe is detected but not attached to the babySystem does not detect a SpO2 probe on the babywarning
System failureElectrical, software, or system failure detected where system cannot continue operationAn electrical failure is detectedcritical
System Failure 0Communication error between SBC and control boardDevice powered on or during operationerror
System Failure 1No Timer2 error; multiple possible hardware defectsDevice powered on or during operationerror
System Failure 11ADC on control board not operating to specification (too slow)Device powered on or during operationerror
System Failure 12Incorrect control board system software checksumDevice powered on or during operationerror
System Failure 13Defective SRAM on control boardDevice powered on or during operationerror
System Failure 14Defective watchdog circuit on control boardDevice powered on or during operationerror
System Failure 15Safety Relay 1 test failure; unable to turn heater on or offDevice powered on or during operation; note: if System Failure 22 appears during operation, System Failure 15 should appear at startuperror
System Failure 16Incubator heater could not be turned off during Power-on Self TestDevice powered on or during operationerror
System Failure 18Defective VariablesDetected during system operationerror
System Failure 19No Audio FrequencyDetected during system operationerror
System Failure 2Timer2 too fast error; microprocessor Timer2 circuit defectiveDevice powered on or during operationerror
System Failure 20Display WDOG Time OutDetected during system operationcritical
System Failure 22Incubator heater will not turn ON during operationHeater fails to activate during operationerror
System Failure 23System unable to turn OFF incubator heater during operationHeater fails to deactivate during operationerror
System Failure 25Fan rotation is not being detectedOptical fan sensor fails to detect fan rotation at power-up or during operationerror
System Failure 26System unable to turn off safety relay 1 during power up testsSafety relay 1 fails to turn off during power-up self-testcritical
System Failure 27Defective non-volatile memoryDetected during system operationerror
System Failure 28SBC board lost communication with control board during operationCommunication failure between SBC and control boardscritical
System Failure 29Defective signal on control boardDetected during system operationerror
System Failure 3ADC calibration detected a channel out of rangeDuring ADC calibration or startuperror
System Failure 30Mains voltage reading LV1 is outside expected rangeMains voltage outside operating range during power-up or operationerror
System Failure 33Defective RH Solid State RelayDetected during system operationerror
System Failure 34Software revision level does not matchControl board and SBC board software versions incompatiblewarning
System Failure 35Wrong dipswitch setupDetected during power-up self-testwarning
System Failure 36LV1 differs from LV2 by more than 10% of LV2Voltage measurement discrepancy detectederror
System Failure 38Door switches indicate heater doors both open and closed simultaneouslyPhysically impossible door switch state detectederror
System Failure 5Volt reference out of acceptable range (1.171V - 1.259V)During system calibration or startuperror
System Failure 55SBC board On Line Self Test FailureDetected during SBC board online self-testerror
System Failure 6Defective overtemp DAC circuit on control boardDevice powered on or during operationerror
System Failure 60Line Circuit FailureDetected during relay board LV2 calibrationerror
System Failure 61Unable to recover from power failDevice cannot recover after power failurecritical
System Failure 62Defective flat ribbon cable between control board and SBCRibbon cable connection or integrity issueerror
System Failure 63Control board has incompatible software versionSoftware version incompatibility detectedwarning
System Failure 71.0 V Thermistor reference voltage out of range (0.951V - 1.049V)During system calibration or startuperror
System Failure 8No Line Frequency signal detectedDevice powered on or during operationerror
Temperature Out of CalibrationCalibration data lost or control board defectiveCalibration data invalid or control board memory failurewarning
Temp Probe FailureSkin probe thermistor mismatch detectedDifference between probe thermistors >0.5 °Cerror
Up Pedal PressedBed up pedal pressed during device power-upUp pedal pressed and held during device startupinfo

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI