GE Healthcare Giraffe: troubleshooting and error codes
The ServiceForge knowledge base holds 154 documented symptoms for the GE Healthcare Giraffe, with 314 ranked causes, 138 diagnostic checks with test points and expected values, and 93 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Device will not power up with alarm sound
Device fails to power up and alarm sounds
Ranked causes
- 1. Defective scale (common)
- 2. Defective scale harness (occasional)
- 3. Defective servo oxygen board (occasional)
- 4. Defective servo oxygen cable (occasional)
- 5. Defective PF battery (occasional)
- 6. Defective power supply (occasional)
- 7. Defective power harness to SBC (occasional)
- 8. Defective SBC (occasional)
- 9. Defective control board (rare)
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System Failure 1 displayed
No Timer2 error; clock battery or scale/harness/Servo Oxygen board/cable/ribbon cable defects indicated
Ranked causes
- 1. Missing or bad clock battery (common)
- 2. Defective scale (occasional)
- 3. Defective scale harness (occasional)
- 4. Defective Servo Oxygen board (occasional)
- 5. Defective Servo Oxygen cable (occasional)
- 6. Defective PF Battery (occasional)
- 7. Defective 14 pin ribbon cable between control board and SBC (occasional)
- 8. Defective SBC (occasional)
- 9. Defective control board (occasional)
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Humidity Failure Alarm
Motor Drive Failure alarm is on; opens safety relay 2 and disables humidifier
Ranked causes
- 1. Elevating base does not function; safety relay 2 not energized (common)
- 2. Humidifier safety thermostat open (common)
- 3. Defective humidifier heater or humidifier safety thermostat (occasional)
- 4. No mains voltage at humidifier isolation transformer input (occasional)
- 5. Defective fuse on relay board rev 10 or higher (occasional)
- 6. Defective fuse on relay board rev 9 or lower (occasional)
- 7. Defective relay board (rare)
- 8. Humidity sensor failed (common)
Check safety relay 2 control signal continuity
- Test point
- 50-pin ribbon cable, pin 34
- Expected
- Continuity present
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System Failure 15 displayed high voltage
Safety Relay 1 test failure; unable to turn heater on or off; sensor, cable, transformer, or board defect
Ranked causes
- 1. If error occurs during calibration, J1, J2, and J4 not unplugged or JP1 not installed properly (common)
- 2. Air temperature sensor is above 40C at power-up (occasional)
- 3. Defective sensor in compartment air probe (occasional)
- 4. Defective compartment sensor cable (occasional)
- 5. No output from heater isolation transformer or defective relay board (common)
- 6. Defective heater isolation transformer (occasional)
- 7. Defective DAC circuit on control board (occasional)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Elevating base will not go up or down
Bed height adjustment mechanism does not respond to controls
Ranked causes
- 1. Pedal switches disabled on operator or service screen (common)
- 2. Defective toroidal transformer (occasional)
- 3. No mains input to toroidal transformer (occasional)
- 4. Defective fuse on relay board (occasional)
- 5. Defective 50-pin ribbon cable (occasional)
- 6. Defective relay board (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen output below setpoint patient safety
Sensor1 reading more than 3% below oxygen set point 7 minutes after change
Ranked causes
- 1. Low O2 supply pressure or flow (common)
- 2. Air leaks into patient compartment (common)
- 3. Supply valve not opening (occasional)
- 4. End cap safety valve not opening (occasional)
- 5. Calibration valve not closing (occasional)
- 6. Patient compartment vents occluded (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed Heater Failure Alarm caution
Software turned on incubator heater but it is not turning on or current sense circuit not detecting it
Ranked causes
- 1. Bed safety thermostat open (device shut off when heater was hot) (common)
- 2. Defective incubator heater, thermostat, or wire harness (occasional)
- 3. Defective incubator solid state relay (SSR) (occasional)
- 4. Defective 50-pin ribbon cable between relay and control board (occasional)
- 5. Defective relay board (rare)
Allow device to cool if heater was recently hot
- Test point
- Incubator
- Expected
- Temperature below 40ºC, below
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fan Failure Alarm
Fan rotation is not being detected
Ranked causes
- 1. Fan not turning (common)
- 2. Fan or fan hub not seated properly (common)
- 3. Defective fan motor (occasional)
- 4. Defective relay board (occasional)
- 5. Optical fan sensor not detecting fan (occasional)
Verify fan hub seating on shaft
- Test point
- Fan hub and shaft
- Expected
- Hub fully seated down on shaft
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ADD WATER message stays on after water added
Add water warning message persists on display even after reservoir is refilled
Ranked causes
- 1. Defective reservoir switch (common)
- 2. Defective 50-pin ribbon cable (occasional)
- 3. Defective relay board (occasional)
- 4. Defective control board (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Black or dim display
Display is very dark or completely black
Ranked causes
- 1. LCD backlight cable not fully seated (common)
- 2. Defective LCD backlight cable (occasional)
- 3. Defective LCD (occasional)
- 4. Defective SBC (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display appears white or has distorted image caution
Display shows white screen or image is distorted with improper colors or pixelation.
Ranked causes
- 1. LCD LVDS cable not fully seated (common)
- 2. Defective LCD LVDS cable (occasional)
- 3. Defective LCD display (occasional)
- 4. Defective single board computer (SBC) (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display is black or dim caution
Display shows completely black or very dim/dim image. Brightness is too low to see content.
Ranked causes
- 1. LCD backlight cable not fully seated (common)
- 2. Defective LCD backlight cable (occasional)
- 3. Defective LCD display (occasional)
- 4. Defective single board computer (SBC) (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display stuck on Giraffe logo screen
Display shows Giraffe logo and does not proceed to operating screen; alarm LED indicators working at power on
Ranked causes
- 1. 14-pin ribbon cable disconnected (common)
- 2. Defective ribbon cable (occasional)
- 3. Defective SBC (occasional)
- 4. Defective control board (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen sensor readings out of range patient safety
Oxygen probe failure - sensors out of range or difference >3%
Ranked causes
- 1. Defective sensor (common)
- 2. Leak or occlusion in sensor housing (common)
- 3. Defective cable or connection (occasional)
- 4. Defective Servo O2 board (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scale icon does not appear on control panel
No scale indicator on display
Ranked causes
- 1. Scale not plugged in (common)
- 2. Defective cable (occasional)
- 3. Defective circuit board (occasional)
- 4. Defective scale receptacle (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor temperature out of range
Sensor plug thermistor temperature reading outside 15C-55C
Ranked causes
- 1. Device is too cold (common)
- 2. Cooling fan not running (occasional)
- 3. Defective sensor plug thermistor or cable (occasional)
- 4. Defective Servo O2 board (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System Failure 16 displayed high voltage
Incubator heater could not be turned off during Power-on Self Test; SSR or relay board defect
Ranked causes
- 1. Defective solid state relay (SSR) (common)
- 2. Defective incubator heater SSR (common)
- 3. Defective current sense circuit on relay board (occasional)
- 4. Defective 50-pin ribbon cable (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System Failure 3 displayed
ADC calibration detected a channel out of range; scale, harness, or control board defect
Ranked causes
- 1. Defective scale (occasional)
- 2. Defective scale harness (occasional)
- 3. No line frequency signal or Timer2 of microprocessor defective (occasional)
- 4. Defective circuit on control board (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch screen does not function
User interface touch screen not responding to input
Ranked causes
- 1. Foreign substance on touch screen surface (common)
- 2. Touch controller flex cable not fully seated (common)
- 3. Touch controller USB cable not fully seated (common)
- 4. Defective touch controller (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
User interface touch screen does not function caution
Touch screen is unresponsive to user input or touches. Interface controls cannot be accessed via touch.
Ranked causes
- 1. Foreign substance on touch screen surface (common)
- 2. Touch controller flex cable connection not fully seated (common)
- 3. Touch controller USB cable connection not fully seated (common)
- 4. Defective touch controller (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
White display or distorted image
Display shows white screen or image is garbled/distorted
Ranked causes
- 1. LCD LVDS cable not fully seated (common)
- 2. Defective LCD LVDS cable (occasional)
- 3. Defective LCD (occasional)
- 4. Defective SBC (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baby Hot alarm activates patient safety
Alarm sounds when patient skin temperature rises above set point; audio suspends for 5-12 minutes when silenced depending on temperature deviation
Ranked causes
- 1. Baby mode set temperature is incorrect for patient (common)
- 2. External heat such as phototherapy or sunlight (occasional)
- 3. Probe disconnected from skin (occasional)
Verify baby mode set temperature setting
- Test point
- Setup menu
- Expected
- Temperature appropriate for patient
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check SpO2 patient safety
System detects a problem with the SpO2 probe. Display/Light notification with 2 min silence time, medium priority escalating to high after 10 sec.
Ranked causes
- 1. Defective probe (occasional)
- 2. Interference (occasional)
- 3. Too much ambient light (occasional)
Assess ambient light level around probe
- Test point
- Probe attachment area
- Expected
- Reduced ambient light exposure
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device will not turn on (power fail alarm)
Device does not power up and power fail alarm sounds
Ranked causes
- 1. No AC power (common)
- 2. No 12V DC power (common)
- 3. Defective power supply (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fan motor not operating or running slowly caution
Incubator fan does not rotate or rotates slowly; poor air circulation; high internal temperature.
Ranked causes
- 1. Fan motor defective or bearing worn (common)
- 2. Fan blade or hub stuck or damaged (occasional)
- 3. Optical fan sensor malfunction (occasional)
Rotate fan blade by hand (power off)
- Test point
- Fan blade on motor shaft
- Expected
- Fan rotates freely without grinding or resistance
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hands-free alarm silence does not work
Hands-free alarm silence function not responding
Ranked causes
- 1. Too much ambient sunlight (common)
- 2. Defective alarm board (occasional)
- 3. Defective LINCAN cable (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heater not heating or heating inconsistently caution
Incubator heater or humidifier heater not producing heat or operating erratically
Ranked causes
- 1. Safety relay 1 not switching heater isolation transformer primary (occasional)
- 2. Heater current sense circuit failure (rare)
- 3. Solid state relay (SSR) failure (occasional)
Verify safety relay 1 energization
- Test point
- Safety relay 1 coil
- Expected
- Energized when heater demand is active
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No power fail indicator (red LED)
Power fail red LED is not illuminated
Ranked causes
- 1. Display power cable disconnected (common)
- 2. Defective display power cable (occasional)
- 3. Defective alarm board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient Cold alarm patient safety
Temperature measured by skin probe is 1.0°C below set temperature (or 0.5°C if configured); requires Skin Temperature alarm enabled
Ranked causes
- 1. Baby Mode set temperature incorrect for patient (common)
- 2. Probe disconnected from skin (common)
- 3. Patient is cold (common)
Verify probe attachment to skin
- Test point
- Patient skin
- Expected
- Probe firmly attached
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient Hot alarm patient safety
Temperature measured by skin probe is 1.0°C above set temperature (or 0.5°C if configured); requires Skin Temperature alarm enabled
Ranked causes
- 1. Baby Mode set temperature incorrect for patient (common)
- 2. External heat such as phototherapy or sunlight (occasional)
- 3. Skin temperature probe disconnected from skin (common)
Verify skin temperature probe attachment
- Test point
- Patient skin
- Expected
- Probe firmly attached to skin
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power Failure (LED Indicator)
No AC power into device or 5V power supply not present despite AC power
Ranked causes
- 1. On/Off switch not turned on (common)
- 2. Defective fuse F1 or F2 on relay board (common)
- 3. Defective power supply (occasional)
Verify power cord plugged and switch on
- Test point
- Power cord, on/off switch, back panel circuit breaker
- Expected
- Cord in live outlet, switch/breaker on
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
RH Sensor calibration failed
Message displayed when RH sensor signal is outside expected range (65–85% RH), indicating either sensor defect or faulty calibration bottle.
Ranked causes
- 1. RH sensor signal out of range 65–85% RH (common)
- 2. RH sensor defective (occasional)
- 3. Calibration bottle defective (occasional)
Verify calibration setup conditions
- Test point
- RH sensor environment
- Expected
- Proper sensor placement, ambient conditions stable
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
RH Sensor Calibration Failed error
Calibration procedure fails with error message indicating RH sensor signal outside expected range
Ranked causes
- 1. RH sensor signal outside expected range at 65-85% RH (common)
- 2. Calibration bottle defective (occasional)
- 3. RH sensor defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scale does not register weight or SC value does not respond to applied weight patient safety
When standard weight is placed on scale, displayed SC value does not increase or increases by incorrect amount
Ranked causes
- 1. Scale cable routed incorrectly, bent, or pinched (common)
- 2. Scale tray obstruction from blankets or cushions (common)
- 3. Scale sensor malfunction or connector disconnection (occasional)
Verify all four shipping screws are removed from screw holes
- Test point
- Scale assembly screw holes
- Expected
- All four screw holes empty, no shipping screws present
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scale icon disappears/reappears with clicking
Scale indicator flickers with repeated clicking sound
Ranked causes
- 1. Defective or disconnected cable (common)
- 2. Defective circuit board (occasional)
- 3. Defective scale receptacle (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen refreshes every few seconds with scale attached
Display constantly refreshes when scale is connected
Ranked causes
- 1. Defective circuit on control board (occasional)
- 2. Defective scale (occasional)
- 3. Defective scale harness (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Servo oxygen system does not calibrate patient safety
System fails to complete oxygen sensor calibration after user initiates procedure; calibration valve may not open or fan may not activate
Ranked causes
- 1. Calibration valve stuck closed (unknown)
- 2. Calibration fan not activating (unknown)
- 3. Oxygen sensors degraded or absent (unknown)
Verify two oxygen sensors are physically present
- Test point
- Sensor housing beneath bed
- Expected
- Two fuel cell oxygen sensors installed
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
System Failure 8 displayed
No Line Frequency error; 120 hertz signal missing from control board or relay board circuit defect
Ranked causes
- 1. Defective control board (common)
- 2. Defective 50-pin ribbon cable between the relay board and control board (occasional)
- 3. Defective relay board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm Board not illuminating or functioning
Red LED not lighting during power fail; beacon lights not responding to commands; buzzer not activating on CAN communication loss
Ranked causes
- 1. +5V DC STBY power not available to Alarm Board (common)
- 2. Alarm Board CAN bus communication failure (occasional)
Measure +5V DC STBY voltage at Alarm Board
- Test point
- Alarm Board power input
- Expected
- +4.5 to +5.5V DC, +/- 0.5V
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baby Cold alarm activated patient safety
Audio alarm with display and light notification when patient skin temperature falls below set temperature in Baby or Manual mode
Ranked causes
- 1. Baby mode set temperature is incorrect for patient (unknown)
- 2. Probe disconnected from skin (unknown)
Verify skin temperature probe attachment to patient skin
- Test point
- skin temperature probe contact point
- Expected
- Probe firmly attached to patient skin
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baby Cold alarm activates patient safety
Alarm sounds on display/light when patient skin temperature falls below set temperature in Baby or Manual mode. Audio suspends for 5–12 minutes when silenced depending on temperature deviation.
Ranked causes
- 1. Baby mode set temperature is incorrect for patient (common)
- 2. Probe disconnected from skin (common)
Measure patient skin temperature with probe
- Test point
- Skin temperature probe
- Expected
- 1.0°C below set temperature (medium alarm) or 2.0°C below (high priority)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed Height Pedals Disabled alarm caution
Alarm activates when bed raise or lower pedals are pressed but have been disabled on Service or Setup screen
Ranked causes
- 1. Bed pedal disabled in Service Mode (common)
- 2. Bed pedal disabled in Setup screen (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed Raise/Lower Pedal Failure Alarm caution
One of the bed Raise/Lower pedal switches is shorted
Ranked causes
- 1. Defective bed Raise/Lower pedal switch or harness (common)
- 2. Defective relay board (occasional)
Check switch continuity at connectors
- Test point
- Bed Raise/Lower pedal switch connectors
- Expected
- Correct continuity state for switch position
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed will not lock in tilt position patient safety
Tilt ball not engaged in finger pocket latch. Bed platform cannot be secured in tilted position.
Ranked causes
- 1. Tilt ball not engaged in finger pocket latch (common)
- 2. Tilt brake pad failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Black display
Display shows no image or is completely black with high priority audio alarm active at power on
Ranked causes
- 1. Display power cable disconnected (common)
- 2. Defective display power cable (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Baby alarm activates patient safety
Alarm sounds on display/light when radiant heater has been running for 12 minutes at specific power setting in Baby or Manual mode without mode or power adjustment.
Ranked causes
- 1. Mode has not been changed (occasional)
- 2. Power setting has not been changed (occasional)
Monitor patient carefully
- Test point
- Patient condition
- Expected
- Patient stable and appropriately warmed
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compartment Sensor Failure Alarm
Software detected difference in two air thermistors exceeding 0.5°C threshold
Ranked causes
- 1. Defective compartment air probe or harness (common)
- 2. Defective control board (occasional)
Measure resistance of two air thermistors
- Test point
- Control board J1, Pins 1-2 and 3-4
- Expected
- within about 100 ohms
- Tools
- Resistance meter
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Confirm Probe Jack Connection alarm patient safety
Skin temperature probe connection alarm activates; heater may shut down if Check Baby alarm unaddressed for 15 minutes
Ranked causes
- 1. Skin temperature probe is disconnected from probe panel (common)
- 2. Malfunctioning probe connector (occasional)
Verify probe connection status
- Test point
- Probe Panel jack
- Expected
- Connector fully inserted
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device will not switch to baby mode
Baby mode cannot be activated
Ranked causes
- 1. Skin temperature probe in wrong connector (common)
- 2. Baby mode disabled on service screen (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device with servo-humidity shows RH as 0
Relative humidity reading shows only zero despite servo-humidity capability
Ranked causes
- 1. Defective humidity sensor (common)
- 2. Defective control board (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display is black caution
Display shows completely black screen. No display output even though system appears to be powered.
Ranked causes
- 1. Display power cable disconnected (common)
- 2. Defective display power cable (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display stuck on Giraffe logo with base movement possible
Display frozen on logo screen; E-Base can move down or radiant heater may be on even in closed-bed situation
Ranked causes
- 1. DC power harnesses disconnected or damaged (common)
- 2. DC power harnesses damaged (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Elevating base fails to raise or lower bed height caution
Bed does not respond to raise/lower pedal commands; stuck at one height or slow movement.
Ranked causes
- 1. Elevating base motor or actuator defective (common)
- 2. Elevating base connector loose or disconnected (occasional)
Visually inspect J-45 connector on relay board for loose contact or corrosion
- Test point
- J-45 elevating base connector on relay board
- Expected
- Connector fully seated with no oxidation or loose pins
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Elevating base motor not operating caution
Elevating base motor does not raise or lower bed; no motor response to control commands
Ranked causes
- 1. Safety relay 2 not switching elevating base transformer primary (occasional)
- 2. Elevating base transformer secondary voltage absent or low (rare)
Verify safety relay 2 energization
- Test point
- Safety relay 2 coil
- Expected
- Energized when motor control command issued
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fan Always in High Speed Alarm
Software cannot read temperature of heat sink sensor in incubator heat sink
Ranked causes
- 1. Defective heat sink sensor or harness (common)
- 2. Defective control board (occasional)
Measure heat sink sensor resistance in service mode
- Test point
- J3 harness pins 1-2 at control board
- Expected
- Valid sensor resistance per specifications
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Humidifier heater circuit resistance out of range
Heater resistance measured at J53.1 to J53.3 does not match expected value for device voltage configuration
Ranked causes
- 1. Heater element open or shorted (common)
- 2. Safety thermostat stuck open or closed (occasional)
Measure heater resistance at J53.1 to J53.3
- Test point
- J53, pins 1 and 3
- Expected
- 72 ohms for 115V devices or 288 ohms for 230V devices
- Tools
- ohmmeter
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incubator heater not operating or insufficient heat caution
Bed fails to reach or maintain set temperature; heater cartridge defective or loose connection.
Ranked causes
- 1. Heater cartridge defective or failed (common)
- 2. Heater electrical connector loose or disconnected (occasional)
Visually inspect heater cartridge for cracks, discoloration, or burn marks
- Test point
- Incubator heater element in heat sink
- Expected
- No visible damage; uniform color
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incubator heater resistance measurement high voltage
Verify incubator heater and bed safety thermostat resistance. Two 115V heaters rated 225W at 104V mounted at opposite ends, connected in parallel. Each heater approximately 48 ohms, parallel combination approximately 24 ohms. Bed safety thermostat in series.
Ranked causes
- 1. Open circuit or failed heater element (occasional)
- 2. Bed safety thermostat failure (rare)
Measure combined heater resistance
- Test point
- Between harness connector pin 4 and black wire at SSR
- Expected
- Approximately 24 ohms (two 48-ohm heaters in parallel), 18-30 ohms
- Tools
- Ohmmeter, DMM
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Insufficient power supply voltage high voltage
Control board power connector voltages below specified thresholds on pins 2, 3, or 4 when measured against pin 1 (ground)
Ranked causes
- 1. Harness 6600-0701-700 or 2075837-001 defective (pins 2 and 3) (common)
- 2. Harness 6600-0701-700 defective (pin 4) (common)
Measure voltage between pin 1 and pin 2
- Test point
- 4-wire power connector, pin 1 (ground) to pin 2
- Expected
- more than 4.75V
- Tools
- digital multimeter
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No alarm LED indicators at power on
Alarm LED lights are not illuminated during power-up sequence
Ranked causes
- 1. Disconnected LINCAN cable between SBC and alarm board (common)
- 2. Defective LINCAN cable (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- No audio alarm
- No SpO2 probe
- Oximetry system failure
- Oxygen calibration failed
- Oxygen level elevated in patient compartment
- Oxygen output above setpoint
- Oxygen sensors operating temperature exceeds limit
- Patient Probe 1 or 2 Failure Alarm
- Patient Probe 1 or Patient Probe 2 Failure Alarm
- Scale initial power-up SC value is not 10,000
- Scale measurement error or drift
- Scale weight reading incorrect
- Servo oxygen system calibration prompted prematurely
- System Failure 12 displayed
- Temperature Out of Calibration Alarm
- Temperature probe resistance out of specification
- Toroidal transformer output voltage low or absent
- Up Pedal Pressed Alarm
- 50-pin ribbon cable continuity fault
- Air temperature alarm exceeds 38C or 40C
- Baby Cold alarm - high priority activation
- Baby Cold alarm high priority triggered
- Baby Hot alarm activation criteria and hysteresis
- Bed Lower Pedal Pressed alarm
- Bed Raise Pedal Pressed alarm
- Bed warms up slowly or does not reach temperature
- Calibration: Electrical safety and calibration
- Calibration: Line voltage
- Calibration: Line Voltage after relay board replacement
- Calibration: Scale weight accuracy
- Calibration: Scale weight measurement
- Calibration: Servo oxygen system 21% reference point
- Calibration: System and Humidifier after control board replacement
- Calibration: System and line voltage
- Callback Time Complete alarm activates
- Center Weight Check
- Chassis cover difficult to remove or misaligned
- Check Baby - Heat Off alarm
- Clock battery replacement
- Confirm Probe Jack Connection alarm activates (Baby mode)
- Confirm Probe Jack Connection alarm activates (Manual mode)
- Device always powers up in service screen mode
- Electrostatic discharge damage to connectors
- Fan motor replacement
- Fan running at high speed without message
- Heat sink sensor thermistor resistance abnormal
- High pulse rate alarm
- High SpO2 alarm
- Hood removal for replacement
- Humidifier reservoir crazed
- Humidifier reservoir has scaling
- Load Cell Check
- Loss of mains power
- Low pulse rate alarm
- Low SpO2 alarm
- Motor Drive Failure Alarm
- No HFAS indicator (green LED)
- No red alarm indicators
- No white alarm indicators
- No yellow alarm indicators
- Noise from speaker on power down
- Off Center Weight Check
- Oxygen system not calibrated
- Oxygen system sensor replacement
- Oxygen system supply valve leak test
- Oxygen system vent screen replacement
- Patient cold or patient hot alarm does not trigger
- Patient hot or patient cold alarm in manual mode
- Power Failure battery replacement
- Power loss detected
- Power supply harness cable troubleshooting
- Power supply not providing correct voltage
- Power supply voltage measurement
- Raising and lowering elevating base in service mode
- Scale cable routing or pinching issue
- Scale cannot be calibrated - times out
- Scale does not return weight
- Scale exceeds maximum weight in run mode
- Scale gives consistently inaccurate weights
- Scale gives erratic weight readings
- Scale slow to return weight
- Scale stuck on screen >45 seconds
- Shipping screws not removed from scale
- Silenced alarm reactivates after silence time expires
- Software upgrade fails
- SpO2 probe off baby
- System Failure 13 displayed
- System Failure 14 displayed
- System Failure 2 displayed
- System Failure 5 displayed
- System Failure 7 displayed
- System Failure alarm activates
- Temp Probe Failure alarm activates
- Weight displays as SC: 40000 on service screen
94 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
70 | System Failure 70 | Unknown | unknown |
71 | System Failure 71 | Unknown | unknown |
81 | System Failure 81 | Unknown | unknown |
82 | System Failure 82 | Unknown | unknown |
83 | Unexpected SBC Software problem | SBC software error | error |
85 | System Failure 85 | Unknown | unknown |
98 | Overflow | Overflow condition | unknown |
99 | Logic problem on the control board | Control board logic error | error |
Baby Cold | Patient skin temperature has fallen below set temperature threshold in Baby or Manual mode | Measured skin temperature 1.0°C below set temperature (medium priority) or 2.0°C below set temperature (high priority) | warning |
Baby hot | Patient skin temperature has risen above the set point threshold | Temperature measured by skin temperature probe is 1.0°C above set temperature (medium priority) or 2.0°C above set temperature (high priority) | warning |
Bed Heater Failure | Incubator heater not turning on or current sense circuit not detecting it | Heater commanded on but not operating or current not detected | error |
Bed Height Pedals Disabled | Bed pedal pressed but pedals are disabled in Service or Setup screen | Raise or lower pedal pressed while disabled | info |
Bed Lower Pedal Pressed | Bed lower pedal pressed during device power-up | Lower pedal pressed while turning on device | info |
Bed Raise/Lower Pedal Failure | Bed Raise/Lower pedal switch is shorted | Switch continuity indicates short condition | error |
Bed Raise Pedal Pressed | Bed raise pedal pressed during device power-up | Raise pedal pressed while turning on device | info |
Calibration failed | Sensors not reading between 23-55mV after 5 minutes in calibration mode | Calibration procedure unable to complete | error |
Callback Time Complete | Callback timer has reached zero and intervention is due | Callback timer reaches zero | warning |
Check Baby | Radiant heater has been running continuously at high power for extended period without adjustment | Heater running at 100% power (Baby mode) or >25% power (Manual mode) for >12 minutes | warning |
Check Baby - Heat Off | Check Baby alarm has remained unaddressed for 15 minutes; heater automatically shuts down | Radiant heater running at 100% power for >15 minutes with Check Baby alarm unaddressed for 3 minutes (Baby mode) or heater at >25% power for >12 minutes with Check Baby unaddressed for 3 minutes (Manual mode) | error |
Check O2 Supply | After calibration, sensor1 reading not >23% after 20 seconds with both valves open | Oxygen supply verification failure after calibration | warning |
Check SpO2 | System detects a problem with the SpO2 probe | Defective probe, interference, or too much ambient light detected | warning |
Compartment Sensor Failure | Software detected difference in two air thermistors exceeding 0.5°C | Difference in air thermistors > 0.5ºC | warning |
Confirm Probe Jack Connection | Skin temperature probe connection to Probe Panel should be checked | Activates in Baby mode when skin temperature probe is unplugged; also in Manual mode when probe is unplugged and Man Temp Alarm enabled | warning |
Confirm Probe Jack Connection (Baby) | No probe detected in Baby mode | No probe plugged into probe jack | warning |
Confirm Probe Jack Connection (Manual) | No probe detected with Manual temp alarm enabled | No probe plugged into probe jack and Man temp Alarm set | warning |
Fan Always in High Speed | Software cannot read temperature of heat sink sensor | Heat sink sensor not responding or invalid reading | error |
Fan Failure | Fan rotation not being detected | Optical fan sensor not detecting fan rotation | error |
FiO2>26% | Elevated oxygen levels in patient compartment not controlled by servo O2 system | Oxygen level exceeds 26% during servo control | warning |
High Oxygen | Sensor1 reading more than 3% above set point | Sensor detects oxygen above desired level | warning |
High Pulse rate | Baby's pulse rate is above set limit | Pulse rate above the set value | warning |
High SpO2 | Baby's blood oxygen saturation is above set level | Blood oxygen saturation above the set value | warning |
Humidity Failure | Motor drive failure or humidifier issue; humidifier disabled | Various humidity system faults | error |
Low Oxygen | Sensor1 reading is more than 3% below oxygen set point | 7 minutes after a set point change | warning |
Low Pulse rate | Baby's pulse rate falls below set limit | Pulse rate below the set value | warning |
Low SpO2 | Baby's blood oxygen saturation falls below set level | Blood oxygen saturation below the set value | warning |
Mains voltage calibration failed | Entered line voltage differs by 20% or more from measured value | User enters calibration value that deviates >20% from actual measured mains voltage | error |
Motor Drive Failure | Elevating base motor running when not commanded | Motor detected running without command | error |
No SpO2 probe | No SpO2 probe is detected at the probe panel connector | System does not detect an SpO2 probe at the probe panel | warning |
O2 Cal Lost-No O2 | Servo oxygen never calibrated, system will not operate | System detects no calibration data on startup | critical |
Oximetry system failure | Oximetry system has failed | Communication between SpO2 and warmer system is lost | error |
Oxygen Probe Failure | Sensor reading out of range (23-280mV) or difference between sensors >3% | Sensor(s) malfunction or connection failure | error |
Oxygen System Failure 1 | Sensor plug thermistor temperature out of range 15C-55C | Temperature sensor reading outside valid range | error |
Oxygen System Failure 2 | Humidity reading out of valid range 1-99 | Humidity sensor malfunction | error |
Oxygen System Failure 3 | Checksum test failed during power-up | Microcontroller self-test failure | critical |
Oxygen System Failure 4 | Analog to digital converter self test failed | ADC reading of Vtest outside 1.216V-1.254V | critical |
Oxygen System Failure 5 | Check O2 Supply alarm silenced twice, sensor1 still not >23% | Persistent oxygen supply inadequacy | error |
Oxygen System Failure 6 | Giraffe control board cannot communicate with Servo O2 board | Communication error on system startup or during operation | critical |
Oxygen System Failure 7 | Watchdog circuit problem detected during power-up | Power-up self-test failure | critical |
Patient Cold - Check Probe 1 | Temperature measured by skin probe is 1.0°C below set temperature | Skin temperature < set temperature - 1.0ºC (or 0.5ºC if configured) | warning |
Patient Hot - Check Probe 1 | Temperature measured by skin probe is 1.0°C above set temperature | Skin temperature > set temperature + 1.0ºC (or 0.5ºC if configured) | warning |
Patient Probe 1 Failure or Patient Probe 2 Failure | Probe thermistors show difference exceeding 0.5°C | Thermistor difference > 0.5ºC | warning |
Power Failure (LED Indicator) | No AC power into device or 5V power supply not present | No AC power or 5V supply failure | critical |
Replace Clock Battery | Clock battery voltage too low | Battery voltage falls below threshold to maintain time | info |
SpO2 probe off baby | SpO2 probe is detected but not attached to the baby | System does not detect a SpO2 probe on the baby | warning |
System failure | Electrical, software, or system failure detected where system cannot continue operation | An electrical failure is detected | critical |
System Failure 0 | Communication error between SBC and control board | Device powered on or during operation | error |
System Failure 1 | No Timer2 error; multiple possible hardware defects | Device powered on or during operation | error |
System Failure 11 | ADC on control board not operating to specification (too slow) | Device powered on or during operation | error |
System Failure 12 | Incorrect control board system software checksum | Device powered on or during operation | error |
System Failure 13 | Defective SRAM on control board | Device powered on or during operation | error |
System Failure 14 | Defective watchdog circuit on control board | Device powered on or during operation | error |
System Failure 15 | Safety Relay 1 test failure; unable to turn heater on or off | Device powered on or during operation; note: if System Failure 22 appears during operation, System Failure 15 should appear at startup | error |
System Failure 16 | Incubator heater could not be turned off during Power-on Self Test | Device powered on or during operation | error |
System Failure 18 | Defective Variables | Detected during system operation | error |
System Failure 19 | No Audio Frequency | Detected during system operation | error |
System Failure 2 | Timer2 too fast error; microprocessor Timer2 circuit defective | Device powered on or during operation | error |
System Failure 20 | Display WDOG Time Out | Detected during system operation | critical |
System Failure 22 | Incubator heater will not turn ON during operation | Heater fails to activate during operation | error |
System Failure 23 | System unable to turn OFF incubator heater during operation | Heater fails to deactivate during operation | error |
System Failure 25 | Fan rotation is not being detected | Optical fan sensor fails to detect fan rotation at power-up or during operation | error |
System Failure 26 | System unable to turn off safety relay 1 during power up tests | Safety relay 1 fails to turn off during power-up self-test | critical |
System Failure 27 | Defective non-volatile memory | Detected during system operation | error |
System Failure 28 | SBC board lost communication with control board during operation | Communication failure between SBC and control boards | critical |
System Failure 29 | Defective signal on control board | Detected during system operation | error |
System Failure 3 | ADC calibration detected a channel out of range | During ADC calibration or startup | error |
System Failure 30 | Mains voltage reading LV1 is outside expected range | Mains voltage outside operating range during power-up or operation | error |
System Failure 33 | Defective RH Solid State Relay | Detected during system operation | error |
System Failure 34 | Software revision level does not match | Control board and SBC board software versions incompatible | warning |
System Failure 35 | Wrong dipswitch setup | Detected during power-up self-test | warning |
System Failure 36 | LV1 differs from LV2 by more than 10% of LV2 | Voltage measurement discrepancy detected | error |
System Failure 38 | Door switches indicate heater doors both open and closed simultaneously | Physically impossible door switch state detected | error |
System Failure 5 | Volt reference out of acceptable range (1.171V - 1.259V) | During system calibration or startup | error |
System Failure 55 | SBC board On Line Self Test Failure | Detected during SBC board online self-test | error |
System Failure 6 | Defective overtemp DAC circuit on control board | Device powered on or during operation | error |
System Failure 60 | Line Circuit Failure | Detected during relay board LV2 calibration | error |
System Failure 61 | Unable to recover from power fail | Device cannot recover after power failure | critical |
System Failure 62 | Defective flat ribbon cable between control board and SBC | Ribbon cable connection or integrity issue | error |
System Failure 63 | Control board has incompatible software version | Software version incompatibility detected | warning |
System Failure 7 | 1.0 V Thermistor reference voltage out of range (0.951V - 1.049V) | During system calibration or startup | error |
System Failure 8 | No Line Frequency signal detected | Device powered on or during operation | error |
Temperature Out of Calibration | Calibration data lost or control board defective | Calibration data invalid or control board memory failure | warning |
Temp Probe Failure | Skin probe thermistor mismatch detected | Difference between probe thermistors >0.5 °C | error |
Up Pedal Pressed | Bed up pedal pressed during device power-up | Up pedal pressed and held during device startup | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.