GE Healthcare Giraffe OmniBed: troubleshooting and error codes

The ServiceForge knowledge base holds 83 documented symptoms for the GE Healthcare Giraffe OmniBed, with 183 ranked causes, 102 diagnostic checks with test points and expected values, and 89 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Unit will not power up, alarm sounds patient safety

Power-up failure with audio alarm

Ranked causes

  1. 1. Defective scale (occasional)
  2. 2. Defective scale harness (occasional)
  3. 3. Defective Servo Oxygen board (occasional)
  4. 4. Defective Servo Oxygen cable (occasional)
  5. 5. Defective 9V battery (occasional)
  6. 6. Defective power supply (occasional)
  7. 7. Defective power harness to display (occasional)
  8. 8. Defective display driver board (rare)
  9. 9. Defective control board (rare)

9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low oxygen level patient safety

Sensor1 reading drops more than 3% below oxygen set point 7 minutes after set point change or canopy closure. Indicated by Low Oxygen alarm.

Ranked causes

  1. 1. Low O2 supply pressure or flow (common)
  2. 2. Inlet screen occluded (common)
  3. 3. Air leaks into infant compartment (occasional)
  4. 4. Supply valve not opening (occasional)
  5. 5. Endcap safety valve not opening (occasional)
  6. 6. Calibration valve stuck closed (occasional)
  7. 7. Infant compartment vents occluded (occasional)
First check

Verify O2 supply pressure

Test point
System inlet
Expected
Greater than 45 psi (310 kPa)

6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System Failure 25: Fan rotation not detected caution

Fan rotation is not being detected during operation or power-up; optical fan sensor not detecting fan motion

Ranked causes

  1. 1. Fan not connected or jack loose (common)
  2. 2. Fan or fan hub not seated properly (common)
  3. 3. Fan speed too low (occasional)
  4. 4. Old design fan (occasional)
  5. 5. Defective or obstructed optical sensor (occasional)
  6. 6. Defective 50 pin ribbon cable (occasional)
  7. 7. Defective relay board (rare)
First check

Fan connected and jack secure?

6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Elevating base AND canopy will not move caution

Neither base elevation nor canopy movement operates

Ranked causes

  1. 1. Pedal switches disabled on setup screen (common)
  2. 2. Defective toroidal transformer (occasional)
  3. 3. No mains input to transformer (occasional)
  4. 4. Defective fuse on relay board (occasional)
  5. 5. Defective 50-pin ribbon cable (occasional)
  6. 6. Defective relay board (rare)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fan failure

Fan rotation not detected; optical fan sensor not detecting fan rotation

Ranked causes

  1. 1. Fan hub not seated properly on fan shaft (common)
  2. 2. Defective fan motor (occasional)
  3. 3. Old design fan (occasional)
  4. 4. Defective or obstructed optical sensor, sensor holder, or gasket (occasional)
  5. 5. Defective 50 pin ribbon cable between relay and control board (rare)
  6. 6. Defective relay board (rare)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

ADD WATER message stays on after water added

Water level indicator remains active despite refilling

Ranked causes

  1. 1. Reservoir switch button jammed (common)
  2. 2. Defective reservoir switch (occasional)
  3. 3. Defective 50-pin ribbon cable (occasional)
  4. 4. Defective relay board (rare)
  5. 5. Defective control board (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Humidifier failure

Humidifier not operating; can be caused by non-humidifier issues including safety relay 2 not energized

Ranked causes

  1. 1. Motor drive failure alarm active (occasional)
  2. 2. Elevating base does not function (occasional)
  3. 3. Humidifier safety thermostat is open (common)
  4. 4. Defective humidifier heater or safety thermostat (occasional)
  5. 5. No mains voltage at humidifier isolation transformer input (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

System Failure 30: Mains voltage out of range caution

Mains voltage reading LV1 outside expected range; voltage not within specified limits

Ranked causes

  1. 1. Mains Voltage Comp Circuit not calibrated correctly (common)
  2. 2. High or low mains voltage (common)
  3. 3. Defective 50 pin ribbon cable (occasional)
  4. 4. Defective relay board (occasional)
  5. 5. Defective control board (rare)
First check

Mains voltage within range? (100V: 80-115V, 115V: 95-132V, 230V: 195-270V)

Expected
100V: 80-115V, 115V: 95-132V, 230V: 195-270V

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Canopy will not move with pedal or hand switches caution

Canopy remains stationary while base works OK

Ranked causes

  1. 1. Canopy motor turns - defective belt or clutch (occasional)
  2. 2. Canopy motor does not turn - defective motor (occasional)
  3. 3. Defective 50-pin ribbon cable (occasional)
  4. 4. Defective relay board (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Heater doors not closed

Alarm activates as canopy reaches halfway down and heater doors are not both closed

Ranked causes

  1. 1. Doors not closing fully (common)
  2. 2. Defective switch or harness (occasional)
  3. 3. Defective 50 pin ribbon cable between relay and control board (rare)
  4. 4. Defective relay board (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Heater doors not opened

Alarm activates as canopy reaches top of travel and heater doors are not both open

Ranked causes

  1. 1. Doors not opening fully (common)
  2. 2. Defective switch or harness (occasional)
  3. 3. Defective 50 pin ribbon cable between relay and control board (rare)
  4. 4. Defective relay board (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen probe failure patient safety

One or both sensors reading out of valid range (23-280 millivolts) or difference between two O2 sensors exceeds 3%. Indicated by Oxygen Probe Failure alarm.

Ranked causes

  1. 1. Defective sensor(s) (occasional)
  2. 2. Leak or occlusion in sensor housing area (occasional)
  3. 3. Defective cable or connection (occasional)
  4. 4. Defective Servo O2 board (occasional)
First check

Verify sensor voltage readings in service mode

Test point
sensor0 and sensor1 display values
Expected
Both 23-280 millivolts; oxygen0 and oxygen1 within 3% of each other

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen system failure - temperature out of range

Sensor plug thermistor temperature reading is out of valid range 15°C-55°C. Indicated by Oxygen System Failure 1 alarm.

Ranked causes

  1. 1. Unit temperature below 15°C (common)
  2. 2. Cooling fan not running above 50°C (occasional)
  3. 3. Defective sensor plug thermistor or cable (occasional)
  4. 4. Defective Servo O2 board (occasional)
First check

Verify ambient temperature and unit location

Test point
Sensor housing temperature sensor (temp reading in service mode)
Expected
15°C or higher for normal operation

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen system failure 5 - persistent low sensor reading after double alarm silence patient safety

Check O2 supply alarm silenced twice and sensor1 reading still not greater than 23%. Indicated by Oxygen System Failure 5 alarm.

Ranked causes

  1. 1. O2 supply not connected or insufficient flow (common)
  2. 2. Calibration valve stuck in calibration position (occasional)
  3. 3. Supply valves not opening (occasional)
  4. 4. Leak or occlusion in sensor housing area (occasional)
First check

Verify O2 supply pressure and flow

Test point
System inlet
Expected
Minimum 45 psi (310 kPa) and 45 L/min

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Radiant heater failure high voltage

Radiant heater not turning on or current sense circuit not detecting it is on; alarm only when canopy at top position

Ranked causes

  1. 1. Defective radiant heater or wire harness (common)
  2. 2. Defective warmer solid state relay (SSR) (occasional)
  3. 3. Defective 50 pin ribbon cable (rare)
  4. 4. Defective relay board (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

System Failure 36: LV1 differs from LV2 by >10% caution

LV1 voltage differs from LV2 voltage by more than 10% of LV2; voltage compensation readings mismatched

Ranked causes

  1. 1. Mains Voltage Comp Circuit not calibrated (common)
  2. 2. Defective 50 pin ribbon cable (occasional)
  3. 3. Defective relay board (occasional)
  4. 4. Defective control board (rare)
First check

Ribbon cable pins 48 and 49 continuity good?

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Canopy works in one direction only caution

Canopy moves up or down but not both directions

Ranked causes

  1. 1. Defective canopy motor (occasional)
  2. 2. Defective 50-pin ribbon cable (occasional)
  3. 3. Defective relay board (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Elevating base will not move, canopy works OK caution

Base elevation non-functional while canopy operates normally

Ranked causes

  1. 1. Pedal switches disabled on setup screen (common)
  2. 2. Defective elevating base motor (occasional)
  3. 3. Defective relay board (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

In transition - heat off

Canopy in transition between incubator and warmer modes; software not detecting end position switch closure

Ranked causes

  1. 1. Defective rail drive belt (occasional)
  2. 2. Spring tube not adjusted correctly (common)
  3. 3. Defective switch or harness (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power failure

No AC power into unit or 5V power supply not present despite AC power available

Ranked causes

  1. 1. Power switch off or cord unplugged (common)
  2. 2. Defective fuses F1 or F2 (occasional)
  3. 3. Defective power supply (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Screen refreshes every few seconds while scale attached

Display flickers or updates excessively

Ranked causes

  1. 1. Defective control board circuit (occasional)
  2. 2. Defective scale (occasional)
  3. 3. Defective scale harness (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Servo-humidity RH setpoint shows 1-10 only

Humidity control range limited to 1-10 instead of full range

Ranked causes

  1. 1. Compartment probe not plugged in at startup (common)
  2. 2. Defective humidity sensor (occasional)
  3. 3. Defective control board (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed will not lock in tilt position caution

Bed tilts but will not stay locked

Ranked causes

  1. 1. Tilt ball not engaged in finger pocket latch (common)
  2. 2. Tilt brake pad failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration failed - sensors not in range patient safety

In calibration mode, after 5 minutes both sensors not reading between 23-55 millivolts. Unit cannot complete calibration. Indicated by Calibration failed alarm.

Ranked causes

  1. 1. Calibration valve not opening or calibration fan not running (common)
  2. 2. Defective sensors (occasional)
First check

Verify calibration valve opens and fan runs

Test point
Service mode Servo O2 screen, VC control
Expected
Click sound when VC opened; calibration fan audible; cooling fan stops

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Canopy motor output voltage incorrect caution

Motor drive output voltage does not match expected control signal or voltage is not reaching motor

Ranked causes

  1. 1. Relay board control signal failure or motor relay fault (common)
  2. 2. Motor relay contact failure or wiring open (occasional)
First check

Measure canopy motor raise output voltage

Test point
Motor output connector J45 pins 1-2
Expected
+24V on pin 1, Gnd on pin 2, 24-32 V acceptable
Tools
DMM

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Canopy raises slowly and sticks near top patient safety

Canopy rises slowly and becomes stuck 4-6 inches from full height when heater doors begin opening

Ranked causes

  1. 1. Rail lift belt failure with broken rungs (common)
  2. 2. Canopy motor coupler, slip clutch or roll pin defective (occasional)
First check

Visually inspect rail lift belt for broken rungs

Test point
Rail lift belt visible through bottom screw slot
Expected
Belt intact with no broken or stripped rungs

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check O2 supply - low sensor reading after calibration patient safety

After calibration, both supply valves opened and after approximately 20 seconds, sensor1 reading not greater than 23%. Indicated by Check O2 supply alarm.

Ranked causes

  1. 1. O2 supply not connected (common)
  2. 2. Kinked hoses in sensor housing (occasional)
First check

Verify O2 supply connection and pressure

Test point
System inlet
Expected
O2 supply connected; pressure minimum 45 psi (310 kPa), flow minimum 45 L/min

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

E-base motor output voltage incorrect caution

E-base lift motor drive voltage does not match expected values

Ranked causes

  1. 1. Relay board control signal failure or motor relay fault (common)
  2. 2. Motor relay contact failure (occasional)
First check

Measure e-base motor raise output voltage

Test point
Motor output connector J45 pins 1-2
Expected
+24V on pin 1, Gnd on pin 2, 24-32 V acceptable
Tools
DMM

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

FiO2 elevated above 26% patient safety

Elevated oxygen levels in patient compartment not controlled by servo O2 system. Oxygen reading above expected 26% baseline. Indicated by FiO2>26% alarm.

Ranked causes

  1. 1. Auxiliary source of oxygen in patient compartment (common)
  2. 2. Leak in servo O2 system supply valve (occasional)
First check

Inspect patient compartment for secondary oxygen source

Test point
Patient compartment, ventilator connection, resuscitation bag
Expected
No auxiliary oxygen delivery devices connected or in use

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

High oxygen level patient safety

Sensor1 reading more than 3% above oxygen set point. Indicated by High Oxygen alarm.

Ranked causes

  1. 1. Oxygen set point recently lowered (common)
  2. 2. Supply valve(s) not closing (occasional)
First check

Verify system stabilization time

Test point
oxygen0 and oxygen1 sensor readings
Expected
Readings approach set point over time; 3% variance acceptable after stabilization

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

In transition alarm sounds continuously patient safety

Alarm activates when canopy not at top or heater doors not fully open, preventing microswitches from activating

Ranked causes

  1. 1. Canopy not reaching full up position due to belt or motor failure (common)
  2. 2. Heater doors not reaching full open due to belt or linkage failure (common)
First check

Verify canopy up detect switch closes when canopy fully raised

Test point
Relay board J36 pins 1-2
Expected
Closed when canopy is up

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incubator heater resistance out of range caution

Measured resistance of incubator heaters and safety thermostat does not match expected values

Ranked causes

  1. 1. Incubator heater element open circuit or thermostat open (common)
  2. 2. One heater element failed or thermostat stuck open (occasional)
First check

Measure total incubator heater resistance

Test point
Between black wire on incubator SSR and white wire on isolation transformer (rev 9-) or J54 connector pin 4 (rev 10+)
Expected
About 24 ohms for parallel combination of two 48 ohm heaters
Tools
Ohmmeter or DMM

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No audio alarm patient safety

Speaker producing no sound output

Ranked causes

  1. 1. Defective speaker or harness (common)
  2. 2. Audio driver circuitry defective (occasional)
First check

Speaker resistance at J40 pins 5-6

Expected
~8 ohms
Tools
Ohmmeter

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 calibration lost patient safety

Servo oxygen unit has not been calibrated or cell voltage is below 33 mV (earlier software versions). Unit will not operate until calibration performed. Indicated by O2 Cal Lost-No O2 alarm.

Ranked causes

  1. 1. Unit never calibrated or calibration data lost (common)
  2. 2. Cell voltage lower than 33 mV (older software) (occasional)
First check

Initiate oxygen calibration

Test point
Cal O2 button on service screen
Expected
Calibration routine starts; sensors read 23-55 mV within 5 minutes

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen system failure - humidity reading out of range

Relative humidity sensor reading is out of valid range 1-99%. Displays 0 in service mode. Indicated by Oxygen System Failure 2 alarm.

Ranked causes

  1. 1. Defective sensor plug (occasional)
  2. 2. Defective Servo O2 board (occasional)
First check

Measure humidity sensor voltage

Test point
J85.2 to J85.1 (humidity sensor connector)
Expected
0.7 to 4.0 Volts

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen system failure 6 - communication failure

Giraffe control board cannot communicate with Servo O2 board. Indicated by Oxygen System Failure 6 alarm.

Ranked causes

  1. 1. Defective Servo O2 board (occasional)
  2. 2. Defective cable between control boards (occasional)
First check

Verify cable connection between control boards

Test point
Cable from Giraffe control board to Servo O2 board connectors
Expected
Cable fully seated on both ends with no visible damage

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power supply voltage out of range caution

Measured DC voltages on power distribution bus do not fall within tolerance

Ranked causes

  1. 1. Power supply or relay board voltage regulator failure (common)
  2. 2. Control board analog voltage regulator failure (occasional)
First check

Measure +5V supply voltage

Test point
J42 pins 1-2 (spare 4-pin connector on power bus)
Expected
4.75 to 5.25 V, 4.75 to 5.25 V
Tools
DMM

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Replace left rail rollers and tension springs

Left rail rollers can wear and require replacement; tension spring may also need service

Ranked causes

  1. 1. Roller wear or bearing failure (unknown)
  2. 2. Tension spring wear or failure (unknown)
First check

Remove roller mounting screw

Test point
Roller hub
Expected
Screw removed with 4mm hex key

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Switch contact resistance excessive or open caution

Membrane switch resistance exceeds 200 ohms or switch does not close when pressed

Ranked causes

  1. 1. Membrane switch contact contamination or wear (occasional)
  2. 2. Membrane switch failed open circuit (rare)
First check

Measure membrane switch contact resistance

Test point
Display Driver board J21 (specific pins depend on switch)
Expected
Less than 200 ohms when switch is pressed, < 200 ohms
Tools
Ohmmeter or DMM

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature out of calibration patient safety

Calibration data lost or defective control board preventing temperature monitoring

Ranked causes

  1. 1. Calibration data lost (common)
  2. 2. Defective control board (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Thermistor resistance out of range caution

Temperature probe or heat sink thermistor resistance does not match expected value for measured temperature

Ranked causes

  1. 1. Thermistor sensor failure or disconnection (common)
  2. 2. Thermistor connector corrosion or loose contact (occasional)
First check

Verify air probe thermistor resistance matches temperature

Test point
Control board J1 pins 1-2 and 3-4 (air probe thermistors)
Expected
Resistance must match temperature reference table for measured compartment temperature
Tools
Ohmmeter, service screen thermometer display

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Toroidal transformer output voltage absent or low high voltage

Secondary output voltage of toroidal transformer is not present or below expected level

Ranked causes

  1. 1. Toroidal transformer primary power failure or disconnection (common)
  2. 2. Toroidal transformer defective (open primary or secondary) (occasional)
First check

Measure toroidal transformer secondary output voltage (service mode)

Test point
Relay board J44 pins 1-2 and J44 pins 3-4
Expected
About 15 volts AC on both pin pairs V AC, ±2V typical
Tools
AC voltmeter or DMM set to AC volts

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit always powers up in service screen mode

Cannot access normal operating mode from startup

Ranked causes

  1. 1. Defective dipswitch setting (common)
  2. 2. Override switch (>37C) shorted (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit will not switch to Baby Control Mode

Baby Control mode unavailable or disabled

Ranked causes

  1. 1. Patient probe plugged into Jack 2 (common)
  2. 2. Baby Control disabled on service screen (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

50-pin ribbon cable continuity fault caution

Communication between control and relay boards interrupted due to ribbon cable fault

Ranked causes

  1. 1. 50-pin ribbon cable broken conductor or poor contact (occasional)
First check

Test 50-pin ribbon cable continuity

Test point
50-pin ribbon cable connectors between control and relay boards
Expected
Low continuity (< 1 ohm) on each conductor; no open circuits
Tools
Ohmmeter, spare 50-pin cable for comparison

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air Temp >38C or >40C alarm active patient safety

Temperature alarm continues even after temperature returns to normal

Ranked causes

  1. 1. Alarm does not auto-reset (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Baby Cold or Hot Alarm does not trigger at setpoint patient safety

Temperature alarm not activating at expected threshold

Ranked causes

  1. 1. Manual Temp Alarm disabled on service screen (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Baby Hot or Baby Cold alarm in Manual Mode

Temperature alarms active in Manual temperature mode

Ranked causes

  1. 1. MANUAL TEMP alarm selected on setup screen (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery replacement due caution

Battery replacement every two years; battery powers failure alarm and memory circuits during power loss and requires charging.

Ranked causes

  1. 1. Two year battery service life expiration (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed warms up very slowly, may not reach set temperature patient safety

Inadequate heating or prolonged warm-up time

Ranked causes

  1. 1. One heater cartridge defective (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Canopy alignment after seal replacement

Realignment procedure for canopy after installation of new enhanced seal kit to ensure proper positioning and operation.

Ranked causes

  1. 1. Canopy alignment required after seal replacement (common)
First check

Install rail locking screws in oblong slot on each rail

Test point
Oblong slots on east and west guide rails
Expected
Rail locking screws fully inserted to unlock rails

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Humidifier

Humidifier requires periodic calibration per maintenance schedule every three years.

Ranked causes

  1. 1. Scheduled maintenance interval reached (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Scale

Periodic scale re-verification procedure required for EU NAWI directive compliance and legal metrology regulations. Requires certified 5 kg test weight (GE part 6600-0209-800).

Ranked causes

  1. 1. Periodic scale calibration required by regulation (common)
First check

Clear scale bed

Test point
Bed surface
Expected
Only mattress and clear plate remain; no other objects present

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Scale weight verification

Annual scale weight check and calibration if scales fail checkout procedures.

Ranked causes

  1. 1. Scale weight check failed (unknown)
First check

Perform scale weight checkout procedures

Test point
Scale
Expected
Weight reading matches reference standard

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Servo Controlled Oxygen patient safety

Initial oxygen system calibration setup and verification using service screen controls (V1, V2, VC, K1) and Cal O2 function to establish baseline sensor readings.

Ranked causes

  1. 1. Unit has never been calibrated (common)
First check

Verify O2 supply pressure and flow

Test point
System inlet
Expected
Minimum 45 psi (310 kPa) and 45 L/min

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Servo Controlled Oxygen System patient safety

Calibration required at least every 24 hours when servo oxygen is in use; system prompts user after 24 hours

Ranked causes

  1. 1. Routine calibration required (common)
First check

Verify two oxygen sensors are present and functioning

Test point
Sensor pair in sensing circuit beneath bed
Expected
Both sensors present and operational

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: System VREF and Mains voltage high voltage

System calibration procedure to set reference voltage and mains voltage on control board after replacement or maintenance

Ranked causes

  1. 1. Control board replacement or calibration requirement (common)
First check

Turn power off

Test point
Main power switch
Expected
Power off

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Canopy descends part way then stops or slows caution

Canopy movement intermittent or slow during descent

Ranked causes

  1. 1. Object fell into radiant heater floppy door opening (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Canopy is clouded

Cloudy or opaque canopy material

Ranked causes

  1. 1. Use of alcohol or hand-disinfectants (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Canopy misalignment caution

Canopy does not seat properly, gaps at corners, misalignment during raise/lower cycles

Ranked causes

  1. 1. Alignment nuts not properly tensioned (common)
First check

Measure gap at canopy corners

Test point
North and south canopy corners
Expected
Approximately 0.15 inch gap, ±0.05 inch

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Canopy rises slowly and stops before doors open fully
  • Canopy seal difficult to remove or reinstall
  • Canopy seals worn or damaged
  • Fan running high speed without indicator
  • Heater doors do not fully open
  • Heater doors do not open fully
  • Heater element electrical failure
  • Humidifier heater resistance out of range
  • Humidifier reservoir crazed - hairline cracks
  • Humidifier reservoir has scaling
  • Motor drive failure
  • Noise from speaker on power down
  • Oxygen sensor replacement due
  • Oxygen sensors require replacement
  • Oxygen system failure - analog to digital converter test failed
  • Oxygen system failure - checksum test failed
  • Oxygen system failure 7 - watchdog circuit error
  • Radiant heater resistance out of range
  • Replace cable carrier links
  • Replace left rail buoyancy springs
  • Supply valve leak test required
  • Vent screen requires replacement
  • Ventilation screen replacement due

23 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
Air Probe FailureThermistors in air probe assembly differ by more than 0.5°CSoftware detected difference in two air thermistors > 0.5°Cwarning
Air Probe FailureAir temperature probe malfunction detectedAir probe disconnected, damaged, or malfunctioningerror
Air Temp >38CAir temperature exceeds 38 degrees CelsiusIncubator air temperature rises above 38°C thresholdwarning
Air Temp >40CAir temperature exceeds 40 degrees CelsiusIncubator air temperature rises above 40°C thresholderror
Baby ColdBaby skin temperature too lowInfant skin temperature falls below safe thresholdwarning
Baby HotBaby skin temperature too highInfant skin temperature exceeds safe thresholdwarning
Baby Mode DisabledBaby mode has been disabled but mode button is pressedBaby mode disabled in service screen and baby mode button pressedinfo
Baby Mode DisabledBaby temperature control mode is inactiveBaby mode control disabled or deactivatederror
Baby Probe 1 FailureBaby temperature probe 1 malfunctionBaby probe 1 disconnected, damaged, or malfunctioningerror
Baby Probe 1 FailurePatient probe 1 thermistors differ by more than 0.5°CProbe thermistors show difference > 0.5°Cwarning
Baby Probe 2 FailureBaby temperature probe 2 malfunctionBaby probe 2 disconnected, damaged, or malfunctioningerror
Baby Probe 2 FailurePatient probe 2 thermistors differ by more than 0.5°CProbe thermistors show difference > 0.5°Cwarning
Bad Membrane SwitchMembrane switch malfunction detectedMembrane switch not responding or damagederror
Bed Heater FailureIncubator heater not functioning or not detectedSoftware turned on heater but it is not on or current sense not detecting iterror
Bed Heater FailureBed heating element malfunctionBed heater not operating or overheatingerror
Bed Up/Down Pedal FailureBed elevation pedal control malfunctionBed pedal not responding to inputerror
Bed Up/Down Pedal FailureBed pedal switch is shorted or defectiveOne of the bed up/down pedal switches is shortedwarning
Bed Up/Down Pedal Switches DisabledBed pedal control disabledBed pedal switches disabled or deactivatedwarning
Calibration failedIn calibration mode, after 5 minutes both sensors not reading between 23-55 mV; unit cannot complete calibration.Cal O2 routine running >5 minutes; sensor0 or sensor1 outside 23-55 mV rangeerror
Canopy Pedal DisabledCanopy control pedal disabledCanopy pedal disabled or deactivatedwarning
Canopy Pedal FailureCanopy control pedal malfunctionCanopy pedal not responding to inputerror
Canopy Pedal FailureCanopy pedal switch is shorted or defectiveOne of the canopy pedal switches is shorted (appears ~100 seconds after short)warning
Canopy Pedal PressedCanopy pedal held down during power-upCanopy pedal switch pressed and held on during power upinfo
Canopy Pedal PressedCanopy control pedal continuously activatedCanopy pedal stuck or held in pressed positionwarning
Check O2 supplyAfter calibration with both supply valves opened, sensor1 reading not greater than 23% after approximately 20 seconds.Cal O2 routine completed; both V1 and V2 controlled to open; sensor1 <23% after 20 secondswarning
Down Pedal PressedBed down pedal held down during power-upBed down pedal switch pressed and held on during power upinfo
Down Pedal PressedBed down control pedal continuously activatedDown pedal stuck or held in pressed positionwarning
Fan Always in High SpeedCooling fan stuck at high speedFan control unable to reduce fan speederror
Fan FailureCooling fan malfunction detectedFan not operating or not responding to controlerror
FiO2>26%Elevated oxygen levels in patient compartment not controlled by servo O2 system.Measured oxygen concentration exceeds 26% and servo system cannot reduce itwarning
Heater Doors Not ClosedHeater door assembly is not fully closedDoor sensor detects open position during operationwarning
Heater Doors Not OpenedHeater door assembly fails to open as commandedDoor does not reach open position during startup or operationerror
High OxygenSensor1 reading more than 3% above oxygen set point.Sensor reading exceeds (oxygen set point + 3%)warning
Low OxygenSensor1 reading more than 3% below oxygen set point 7 minutes after set point change or canopy closed.Sensor reading drops below (oxygen set point - 3%) and remains there for 7 minuteswarning
Mains voltage calibration failedEntered mains voltage differs significantly from measured valueDialed in line voltage is 20% different than measured value (not nominal value)error
Motor Drive FailureMotor drive circuit or motor has failedMotor does not respond to control commands or stallserror
O2 Cal Lost-No O2Servo oxygen unit has never been calibrated or calibration data lost; unit will not operate.No valid calibration data in memory; cell voltage below 33 mV (earlier software)critical
Oxygen Probe FailureOne sensor reading out of range (23-280 mV) or difference between two oxygen sensors exceeds 3%.sensor0 or sensor1 reading outside 23-280 mV range; difference between oxygen0 and oxygen1 exceeds 3%error
Oxygen System Failure 1Sensor plug thermistor temperature reading out of valid range 15°C-55°C.Temperature reading <15°C or >55°C at power-up or during operationerror
Oxygen System Failure 2Relative humidity sensor reading out of valid range 1-99%; displays 0 in service mode.Humidity reading <1% or >99%error
Oxygen System Failure 3Checksum test performed during power-up testing failed.Checksum test failure at power-upcritical
Oxygen System Failure 4Analog to digital converter circuit self-test failed. Vtest reading out of valid range 1.216-1.254 V.A/D converter self-test failure at power-up; Vtest <1.216 V or >1.254 Vcritical
Oxygen System Failure 5Check O2 supply alarm silenced twice and sensor1 still not greater than 23%.Check O2 supply alarm manually silenced twice; sensor1 remains <23%error
Oxygen System Failure 6Giraffe control board cannot communicate with Servo O2 board.Communication failure detected during normal operation or power-uperror
Oxygen System Failure 7Power-up test detected a problem with the watchdog circuit.Watchdog circuit test failure during power-up self-testcritical
Power FailureLoss of main power supplyPower supply voltage drops below operating thresholdcritical
Radiant Heater FailureRadiant heating element or heater control circuit is not functioningHeater output drops or fails to respond to temperature demanderror
RH Sensor Calibration Failed Try AgainHumidity sensor calibration routine failed to complete successfullyRH sensor signal falls outside the expected 65-85% RH range during calibrationerror
System Failure 0Communication error between display driver board and control boardDisplay and control board communication failurecritical
System Failure 1No Timer2Timer2 circuit not functioningcritical
System Failure 11ADC not operating to spec (too slow)ADC conversion speed below specificationerror
System Failure 12Incorrect control board software checksumSystem software PROM checksum mismatchcritical
System Failure 13Defective SRAMSRAM circuit malfunctioncritical
System Failure 14Defective WDOGWatch dog circuit malfunctioncritical
System Failure 15Safety Relay 1 test failureUnable to control heater via safety relay 1 during self-testcritical
System Failure 16Heater could not be turned off during self-testHeater remained on during power-on self-testcritical
System Failure 17Radiant heater could not be turned off during calibrationRadiant heater remained on during calibration self-testerror
System Failure 18Defective VariablesDetected during system startup or operationerror
System Failure 19No Audio FrequencyAudio circuit malfunction detectederror
System Failure 2Timer2 too fastTimer2 running at incorrect speederror
System Failure 20Display WDOG Time OutDisplay watchdog timer expirederror
System Failure 22Radiant heater and incubator heater will not turn ONNeither heater responds during operationwarning
System Failure 23Unable to turn OFF heatersHeaters remain ON when commanded OFFwarning
System Failure 25Fan rotation not detectedOptical fan sensor does not detect fan motion during power-up or operationerror
System Failure 26Unable to turn off safety relay 1Safety relay 1 does not respond to OFF command during power-up testscritical
System Failure 27Defective non-volatile memoryMemory integrity check failederror
System Failure 28Display lost communication with control boardCommunication link interrupted between display and control boardserror
System Failure 29Defective signal on control boardSignal integrity check failederror
System Failure 3ADC channel out of range during calibrationCalibration detected out-of-range ADC channelerror
System Failure 30Mains voltage LV1 outside expected rangeMeasured mains voltage exceeds specified limitswarning
System Failure 33Defective RH Solid State RelayRadiant heater relay failure detectederror
System Failure 34Software revision level does not matchControl board and display board software versions incompatibleerror
System Failure 35Wrong dipswitch setupDipswitch configuration on control board incorrectwarning
System Failure 36LV1 differs from LV2 by >10%Voltage comparison readings diverge beyond 10% tolerancewarning
System Failure 37Defective switches (software 1.60+)Canopy position switches show contradictory statewarning
System Failure 38Door switch contradictionDoor switches show doors both open and closedwarning
System Failure 5Volt reference out of rangeReference voltage outside 1.171V-1.259V rangeerror
System Failure 50Display driver board software checksum errorSystem Software PROM checksum verification failederror
System Failure 51Display defective SRAMSRAM memory test failed on display boarderror
System Failure 52Display no timer 2Timer 2 not responding on display boarderror
System Failure 53Display timer 2 too fastTimer 2 running at incorrect speederror
System Failure 55Display board on-line self-test failureDisplay board failed self-test during power-up or diagnosticscritical
System Failure 6Defective overtemp DAC circuitOver-temperature DAC circuit malfunctionerror
System Failure 60Line circuit failureRelay board LV2 calibration circuit failure during calibrationerror
System Failure 7Thermistor reference voltage out of range1.0V thermistor reference outside 0.951V-1.049Verror
System Failure 8No line frequency120 Hz line frequency signal missingcritical
System Failure 98OverflowData or resource overflow conditioncritical
System Failure 99Logic problem on control boardControl board software logic errorcritical
Temperature out of CalibrationTemperature sensor reading exceeds calibration limitsTemperature measurement drifts outside acceptable calibration rangewarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI