GE Healthcare Giraffe OmniBed: troubleshooting and error codes
The ServiceForge knowledge base holds 83 documented symptoms for the GE Healthcare Giraffe OmniBed, with 183 ranked causes, 102 diagnostic checks with test points and expected values, and 89 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Unit will not power up, alarm sounds patient safety
Power-up failure with audio alarm
Ranked causes
- 1. Defective scale (occasional)
- 2. Defective scale harness (occasional)
- 3. Defective Servo Oxygen board (occasional)
- 4. Defective Servo Oxygen cable (occasional)
- 5. Defective 9V battery (occasional)
- 6. Defective power supply (occasional)
- 7. Defective power harness to display (occasional)
- 8. Defective display driver board (rare)
- 9. Defective control board (rare)
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low oxygen level patient safety
Sensor1 reading drops more than 3% below oxygen set point 7 minutes after set point change or canopy closure. Indicated by Low Oxygen alarm.
Ranked causes
- 1. Low O2 supply pressure or flow (common)
- 2. Inlet screen occluded (common)
- 3. Air leaks into infant compartment (occasional)
- 4. Supply valve not opening (occasional)
- 5. Endcap safety valve not opening (occasional)
- 6. Calibration valve stuck closed (occasional)
- 7. Infant compartment vents occluded (occasional)
Verify O2 supply pressure
- Test point
- System inlet
- Expected
- Greater than 45 psi (310 kPa)
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System Failure 25: Fan rotation not detected caution
Fan rotation is not being detected during operation or power-up; optical fan sensor not detecting fan motion
Ranked causes
- 1. Fan not connected or jack loose (common)
- 2. Fan or fan hub not seated properly (common)
- 3. Fan speed too low (occasional)
- 4. Old design fan (occasional)
- 5. Defective or obstructed optical sensor (occasional)
- 6. Defective 50 pin ribbon cable (occasional)
- 7. Defective relay board (rare)
Fan connected and jack secure?
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Elevating base AND canopy will not move caution
Neither base elevation nor canopy movement operates
Ranked causes
- 1. Pedal switches disabled on setup screen (common)
- 2. Defective toroidal transformer (occasional)
- 3. No mains input to transformer (occasional)
- 4. Defective fuse on relay board (occasional)
- 5. Defective 50-pin ribbon cable (occasional)
- 6. Defective relay board (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fan failure
Fan rotation not detected; optical fan sensor not detecting fan rotation
Ranked causes
- 1. Fan hub not seated properly on fan shaft (common)
- 2. Defective fan motor (occasional)
- 3. Old design fan (occasional)
- 4. Defective or obstructed optical sensor, sensor holder, or gasket (occasional)
- 5. Defective 50 pin ribbon cable between relay and control board (rare)
- 6. Defective relay board (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ADD WATER message stays on after water added
Water level indicator remains active despite refilling
Ranked causes
- 1. Reservoir switch button jammed (common)
- 2. Defective reservoir switch (occasional)
- 3. Defective 50-pin ribbon cable (occasional)
- 4. Defective relay board (rare)
- 5. Defective control board (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Humidifier failure
Humidifier not operating; can be caused by non-humidifier issues including safety relay 2 not energized
Ranked causes
- 1. Motor drive failure alarm active (occasional)
- 2. Elevating base does not function (occasional)
- 3. Humidifier safety thermostat is open (common)
- 4. Defective humidifier heater or safety thermostat (occasional)
- 5. No mains voltage at humidifier isolation transformer input (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System Failure 30: Mains voltage out of range caution
Mains voltage reading LV1 outside expected range; voltage not within specified limits
Ranked causes
- 1. Mains Voltage Comp Circuit not calibrated correctly (common)
- 2. High or low mains voltage (common)
- 3. Defective 50 pin ribbon cable (occasional)
- 4. Defective relay board (occasional)
- 5. Defective control board (rare)
Mains voltage within range? (100V: 80-115V, 115V: 95-132V, 230V: 195-270V)
- Expected
- 100V: 80-115V, 115V: 95-132V, 230V: 195-270V
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Canopy will not move with pedal or hand switches caution
Canopy remains stationary while base works OK
Ranked causes
- 1. Canopy motor turns - defective belt or clutch (occasional)
- 2. Canopy motor does not turn - defective motor (occasional)
- 3. Defective 50-pin ribbon cable (occasional)
- 4. Defective relay board (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heater doors not closed
Alarm activates as canopy reaches halfway down and heater doors are not both closed
Ranked causes
- 1. Doors not closing fully (common)
- 2. Defective switch or harness (occasional)
- 3. Defective 50 pin ribbon cable between relay and control board (rare)
- 4. Defective relay board (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heater doors not opened
Alarm activates as canopy reaches top of travel and heater doors are not both open
Ranked causes
- 1. Doors not opening fully (common)
- 2. Defective switch or harness (occasional)
- 3. Defective 50 pin ribbon cable between relay and control board (rare)
- 4. Defective relay board (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen probe failure patient safety
One or both sensors reading out of valid range (23-280 millivolts) or difference between two O2 sensors exceeds 3%. Indicated by Oxygen Probe Failure alarm.
Ranked causes
- 1. Defective sensor(s) (occasional)
- 2. Leak or occlusion in sensor housing area (occasional)
- 3. Defective cable or connection (occasional)
- 4. Defective Servo O2 board (occasional)
Verify sensor voltage readings in service mode
- Test point
- sensor0 and sensor1 display values
- Expected
- Both 23-280 millivolts; oxygen0 and oxygen1 within 3% of each other
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen system failure - temperature out of range
Sensor plug thermistor temperature reading is out of valid range 15°C-55°C. Indicated by Oxygen System Failure 1 alarm.
Ranked causes
- 1. Unit temperature below 15°C (common)
- 2. Cooling fan not running above 50°C (occasional)
- 3. Defective sensor plug thermistor or cable (occasional)
- 4. Defective Servo O2 board (occasional)
Verify ambient temperature and unit location
- Test point
- Sensor housing temperature sensor (temp reading in service mode)
- Expected
- 15°C or higher for normal operation
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen system failure 5 - persistent low sensor reading after double alarm silence patient safety
Check O2 supply alarm silenced twice and sensor1 reading still not greater than 23%. Indicated by Oxygen System Failure 5 alarm.
Ranked causes
- 1. O2 supply not connected or insufficient flow (common)
- 2. Calibration valve stuck in calibration position (occasional)
- 3. Supply valves not opening (occasional)
- 4. Leak or occlusion in sensor housing area (occasional)
Verify O2 supply pressure and flow
- Test point
- System inlet
- Expected
- Minimum 45 psi (310 kPa) and 45 L/min
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Radiant heater failure high voltage
Radiant heater not turning on or current sense circuit not detecting it is on; alarm only when canopy at top position
Ranked causes
- 1. Defective radiant heater or wire harness (common)
- 2. Defective warmer solid state relay (SSR) (occasional)
- 3. Defective 50 pin ribbon cable (rare)
- 4. Defective relay board (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System Failure 36: LV1 differs from LV2 by >10% caution
LV1 voltage differs from LV2 voltage by more than 10% of LV2; voltage compensation readings mismatched
Ranked causes
- 1. Mains Voltage Comp Circuit not calibrated (common)
- 2. Defective 50 pin ribbon cable (occasional)
- 3. Defective relay board (occasional)
- 4. Defective control board (rare)
Ribbon cable pins 48 and 49 continuity good?
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Canopy works in one direction only caution
Canopy moves up or down but not both directions
Ranked causes
- 1. Defective canopy motor (occasional)
- 2. Defective 50-pin ribbon cable (occasional)
- 3. Defective relay board (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Elevating base will not move, canopy works OK caution
Base elevation non-functional while canopy operates normally
Ranked causes
- 1. Pedal switches disabled on setup screen (common)
- 2. Defective elevating base motor (occasional)
- 3. Defective relay board (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
In transition - heat off
Canopy in transition between incubator and warmer modes; software not detecting end position switch closure
Ranked causes
- 1. Defective rail drive belt (occasional)
- 2. Spring tube not adjusted correctly (common)
- 3. Defective switch or harness (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power failure
No AC power into unit or 5V power supply not present despite AC power available
Ranked causes
- 1. Power switch off or cord unplugged (common)
- 2. Defective fuses F1 or F2 (occasional)
- 3. Defective power supply (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen refreshes every few seconds while scale attached
Display flickers or updates excessively
Ranked causes
- 1. Defective control board circuit (occasional)
- 2. Defective scale (occasional)
- 3. Defective scale harness (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Servo-humidity RH setpoint shows 1-10 only
Humidity control range limited to 1-10 instead of full range
Ranked causes
- 1. Compartment probe not plugged in at startup (common)
- 2. Defective humidity sensor (occasional)
- 3. Defective control board (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed will not lock in tilt position caution
Bed tilts but will not stay locked
Ranked causes
- 1. Tilt ball not engaged in finger pocket latch (common)
- 2. Tilt brake pad failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration failed - sensors not in range patient safety
In calibration mode, after 5 minutes both sensors not reading between 23-55 millivolts. Unit cannot complete calibration. Indicated by Calibration failed alarm.
Ranked causes
- 1. Calibration valve not opening or calibration fan not running (common)
- 2. Defective sensors (occasional)
Verify calibration valve opens and fan runs
- Test point
- Service mode Servo O2 screen, VC control
- Expected
- Click sound when VC opened; calibration fan audible; cooling fan stops
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Canopy motor output voltage incorrect caution
Motor drive output voltage does not match expected control signal or voltage is not reaching motor
Ranked causes
- 1. Relay board control signal failure or motor relay fault (common)
- 2. Motor relay contact failure or wiring open (occasional)
Measure canopy motor raise output voltage
- Test point
- Motor output connector J45 pins 1-2
- Expected
- +24V on pin 1, Gnd on pin 2, 24-32 V acceptable
- Tools
- DMM
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Canopy raises slowly and sticks near top patient safety
Canopy rises slowly and becomes stuck 4-6 inches from full height when heater doors begin opening
Ranked causes
- 1. Rail lift belt failure with broken rungs (common)
- 2. Canopy motor coupler, slip clutch or roll pin defective (occasional)
Visually inspect rail lift belt for broken rungs
- Test point
- Rail lift belt visible through bottom screw slot
- Expected
- Belt intact with no broken or stripped rungs
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check O2 supply - low sensor reading after calibration patient safety
After calibration, both supply valves opened and after approximately 20 seconds, sensor1 reading not greater than 23%. Indicated by Check O2 supply alarm.
Ranked causes
- 1. O2 supply not connected (common)
- 2. Kinked hoses in sensor housing (occasional)
Verify O2 supply connection and pressure
- Test point
- System inlet
- Expected
- O2 supply connected; pressure minimum 45 psi (310 kPa), flow minimum 45 L/min
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
E-base motor output voltage incorrect caution
E-base lift motor drive voltage does not match expected values
Ranked causes
- 1. Relay board control signal failure or motor relay fault (common)
- 2. Motor relay contact failure (occasional)
Measure e-base motor raise output voltage
- Test point
- Motor output connector J45 pins 1-2
- Expected
- +24V on pin 1, Gnd on pin 2, 24-32 V acceptable
- Tools
- DMM
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
FiO2 elevated above 26% patient safety
Elevated oxygen levels in patient compartment not controlled by servo O2 system. Oxygen reading above expected 26% baseline. Indicated by FiO2>26% alarm.
Ranked causes
- 1. Auxiliary source of oxygen in patient compartment (common)
- 2. Leak in servo O2 system supply valve (occasional)
Inspect patient compartment for secondary oxygen source
- Test point
- Patient compartment, ventilator connection, resuscitation bag
- Expected
- No auxiliary oxygen delivery devices connected or in use
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High oxygen level patient safety
Sensor1 reading more than 3% above oxygen set point. Indicated by High Oxygen alarm.
Ranked causes
- 1. Oxygen set point recently lowered (common)
- 2. Supply valve(s) not closing (occasional)
Verify system stabilization time
- Test point
- oxygen0 and oxygen1 sensor readings
- Expected
- Readings approach set point over time; 3% variance acceptable after stabilization
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
In transition alarm sounds continuously patient safety
Alarm activates when canopy not at top or heater doors not fully open, preventing microswitches from activating
Ranked causes
- 1. Canopy not reaching full up position due to belt or motor failure (common)
- 2. Heater doors not reaching full open due to belt or linkage failure (common)
Verify canopy up detect switch closes when canopy fully raised
- Test point
- Relay board J36 pins 1-2
- Expected
- Closed when canopy is up
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incubator heater resistance out of range caution
Measured resistance of incubator heaters and safety thermostat does not match expected values
Ranked causes
- 1. Incubator heater element open circuit or thermostat open (common)
- 2. One heater element failed or thermostat stuck open (occasional)
Measure total incubator heater resistance
- Test point
- Between black wire on incubator SSR and white wire on isolation transformer (rev 9-) or J54 connector pin 4 (rev 10+)
- Expected
- About 24 ohms for parallel combination of two 48 ohm heaters
- Tools
- Ohmmeter or DMM
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No audio alarm patient safety
Speaker producing no sound output
Ranked causes
- 1. Defective speaker or harness (common)
- 2. Audio driver circuitry defective (occasional)
Speaker resistance at J40 pins 5-6
- Expected
- ~8 ohms
- Tools
- Ohmmeter
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 calibration lost patient safety
Servo oxygen unit has not been calibrated or cell voltage is below 33 mV (earlier software versions). Unit will not operate until calibration performed. Indicated by O2 Cal Lost-No O2 alarm.
Ranked causes
- 1. Unit never calibrated or calibration data lost (common)
- 2. Cell voltage lower than 33 mV (older software) (occasional)
Initiate oxygen calibration
- Test point
- Cal O2 button on service screen
- Expected
- Calibration routine starts; sensors read 23-55 mV within 5 minutes
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen system failure - humidity reading out of range
Relative humidity sensor reading is out of valid range 1-99%. Displays 0 in service mode. Indicated by Oxygen System Failure 2 alarm.
Ranked causes
- 1. Defective sensor plug (occasional)
- 2. Defective Servo O2 board (occasional)
Measure humidity sensor voltage
- Test point
- J85.2 to J85.1 (humidity sensor connector)
- Expected
- 0.7 to 4.0 Volts
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen system failure 6 - communication failure
Giraffe control board cannot communicate with Servo O2 board. Indicated by Oxygen System Failure 6 alarm.
Ranked causes
- 1. Defective Servo O2 board (occasional)
- 2. Defective cable between control boards (occasional)
Verify cable connection between control boards
- Test point
- Cable from Giraffe control board to Servo O2 board connectors
- Expected
- Cable fully seated on both ends with no visible damage
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply voltage out of range caution
Measured DC voltages on power distribution bus do not fall within tolerance
Ranked causes
- 1. Power supply or relay board voltage regulator failure (common)
- 2. Control board analog voltage regulator failure (occasional)
Measure +5V supply voltage
- Test point
- J42 pins 1-2 (spare 4-pin connector on power bus)
- Expected
- 4.75 to 5.25 V, 4.75 to 5.25 V
- Tools
- DMM
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replace left rail rollers and tension springs
Left rail rollers can wear and require replacement; tension spring may also need service
Ranked causes
- 1. Roller wear or bearing failure (unknown)
- 2. Tension spring wear or failure (unknown)
Remove roller mounting screw
- Test point
- Roller hub
- Expected
- Screw removed with 4mm hex key
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Switch contact resistance excessive or open caution
Membrane switch resistance exceeds 200 ohms or switch does not close when pressed
Ranked causes
- 1. Membrane switch contact contamination or wear (occasional)
- 2. Membrane switch failed open circuit (rare)
Measure membrane switch contact resistance
- Test point
- Display Driver board J21 (specific pins depend on switch)
- Expected
- Less than 200 ohms when switch is pressed, < 200 ohms
- Tools
- Ohmmeter or DMM
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature out of calibration patient safety
Calibration data lost or defective control board preventing temperature monitoring
Ranked causes
- 1. Calibration data lost (common)
- 2. Defective control board (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermistor resistance out of range caution
Temperature probe or heat sink thermistor resistance does not match expected value for measured temperature
Ranked causes
- 1. Thermistor sensor failure or disconnection (common)
- 2. Thermistor connector corrosion or loose contact (occasional)
Verify air probe thermistor resistance matches temperature
- Test point
- Control board J1 pins 1-2 and 3-4 (air probe thermistors)
- Expected
- Resistance must match temperature reference table for measured compartment temperature
- Tools
- Ohmmeter, service screen thermometer display
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Toroidal transformer output voltage absent or low high voltage
Secondary output voltage of toroidal transformer is not present or below expected level
Ranked causes
- 1. Toroidal transformer primary power failure or disconnection (common)
- 2. Toroidal transformer defective (open primary or secondary) (occasional)
Measure toroidal transformer secondary output voltage (service mode)
- Test point
- Relay board J44 pins 1-2 and J44 pins 3-4
- Expected
- About 15 volts AC on both pin pairs V AC, ±2V typical
- Tools
- AC voltmeter or DMM set to AC volts
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit always powers up in service screen mode
Cannot access normal operating mode from startup
Ranked causes
- 1. Defective dipswitch setting (common)
- 2. Override switch (>37C) shorted (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not switch to Baby Control Mode
Baby Control mode unavailable or disabled
Ranked causes
- 1. Patient probe plugged into Jack 2 (common)
- 2. Baby Control disabled on service screen (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
50-pin ribbon cable continuity fault caution
Communication between control and relay boards interrupted due to ribbon cable fault
Ranked causes
- 1. 50-pin ribbon cable broken conductor or poor contact (occasional)
Test 50-pin ribbon cable continuity
- Test point
- 50-pin ribbon cable connectors between control and relay boards
- Expected
- Low continuity (< 1 ohm) on each conductor; no open circuits
- Tools
- Ohmmeter, spare 50-pin cable for comparison
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air Temp >38C or >40C alarm active patient safety
Temperature alarm continues even after temperature returns to normal
Ranked causes
- 1. Alarm does not auto-reset (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baby Cold or Hot Alarm does not trigger at setpoint patient safety
Temperature alarm not activating at expected threshold
Ranked causes
- 1. Manual Temp Alarm disabled on service screen (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baby Hot or Baby Cold alarm in Manual Mode
Temperature alarms active in Manual temperature mode
Ranked causes
- 1. MANUAL TEMP alarm selected on setup screen (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery replacement due caution
Battery replacement every two years; battery powers failure alarm and memory circuits during power loss and requires charging.
Ranked causes
- 1. Two year battery service life expiration (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed warms up very slowly, may not reach set temperature patient safety
Inadequate heating or prolonged warm-up time
Ranked causes
- 1. One heater cartridge defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Canopy alignment after seal replacement
Realignment procedure for canopy after installation of new enhanced seal kit to ensure proper positioning and operation.
Ranked causes
- 1. Canopy alignment required after seal replacement (common)
Install rail locking screws in oblong slot on each rail
- Test point
- Oblong slots on east and west guide rails
- Expected
- Rail locking screws fully inserted to unlock rails
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Humidifier
Humidifier requires periodic calibration per maintenance schedule every three years.
Ranked causes
- 1. Scheduled maintenance interval reached (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Scale
Periodic scale re-verification procedure required for EU NAWI directive compliance and legal metrology regulations. Requires certified 5 kg test weight (GE part 6600-0209-800).
Ranked causes
- 1. Periodic scale calibration required by regulation (common)
Clear scale bed
- Test point
- Bed surface
- Expected
- Only mattress and clear plate remain; no other objects present
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Scale weight verification
Annual scale weight check and calibration if scales fail checkout procedures.
Ranked causes
- 1. Scale weight check failed (unknown)
Perform scale weight checkout procedures
- Test point
- Scale
- Expected
- Weight reading matches reference standard
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Servo Controlled Oxygen patient safety
Initial oxygen system calibration setup and verification using service screen controls (V1, V2, VC, K1) and Cal O2 function to establish baseline sensor readings.
Ranked causes
- 1. Unit has never been calibrated (common)
Verify O2 supply pressure and flow
- Test point
- System inlet
- Expected
- Minimum 45 psi (310 kPa) and 45 L/min
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Servo Controlled Oxygen System patient safety
Calibration required at least every 24 hours when servo oxygen is in use; system prompts user after 24 hours
Ranked causes
- 1. Routine calibration required (common)
Verify two oxygen sensors are present and functioning
- Test point
- Sensor pair in sensing circuit beneath bed
- Expected
- Both sensors present and operational
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: System VREF and Mains voltage high voltage
System calibration procedure to set reference voltage and mains voltage on control board after replacement or maintenance
Ranked causes
- 1. Control board replacement or calibration requirement (common)
Turn power off
- Test point
- Main power switch
- Expected
- Power off
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Canopy descends part way then stops or slows caution
Canopy movement intermittent or slow during descent
Ranked causes
- 1. Object fell into radiant heater floppy door opening (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Canopy is clouded
Cloudy or opaque canopy material
Ranked causes
- 1. Use of alcohol or hand-disinfectants (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Canopy misalignment caution
Canopy does not seat properly, gaps at corners, misalignment during raise/lower cycles
Ranked causes
- 1. Alignment nuts not properly tensioned (common)
Measure gap at canopy corners
- Test point
- North and south canopy corners
- Expected
- Approximately 0.15 inch gap, ±0.05 inch
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Canopy rises slowly and stops before doors open fully
- Canopy seal difficult to remove or reinstall
- Canopy seals worn or damaged
- Fan running high speed without indicator
- Heater doors do not fully open
- Heater doors do not open fully
- Heater element electrical failure
- Humidifier heater resistance out of range
- Humidifier reservoir crazed - hairline cracks
- Humidifier reservoir has scaling
- Motor drive failure
- Noise from speaker on power down
- Oxygen sensor replacement due
- Oxygen sensors require replacement
- Oxygen system failure - analog to digital converter test failed
- Oxygen system failure - checksum test failed
- Oxygen system failure 7 - watchdog circuit error
- Radiant heater resistance out of range
- Replace cable carrier links
- Replace left rail buoyancy springs
- Supply valve leak test required
- Vent screen requires replacement
- Ventilation screen replacement due
23 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Air Probe Failure | Thermistors in air probe assembly differ by more than 0.5°C | Software detected difference in two air thermistors > 0.5°C | warning |
Air Probe Failure | Air temperature probe malfunction detected | Air probe disconnected, damaged, or malfunctioning | error |
Air Temp >38C | Air temperature exceeds 38 degrees Celsius | Incubator air temperature rises above 38°C threshold | warning |
Air Temp >40C | Air temperature exceeds 40 degrees Celsius | Incubator air temperature rises above 40°C threshold | error |
Baby Cold | Baby skin temperature too low | Infant skin temperature falls below safe threshold | warning |
Baby Hot | Baby skin temperature too high | Infant skin temperature exceeds safe threshold | warning |
Baby Mode Disabled | Baby mode has been disabled but mode button is pressed | Baby mode disabled in service screen and baby mode button pressed | info |
Baby Mode Disabled | Baby temperature control mode is inactive | Baby mode control disabled or deactivated | error |
Baby Probe 1 Failure | Baby temperature probe 1 malfunction | Baby probe 1 disconnected, damaged, or malfunctioning | error |
Baby Probe 1 Failure | Patient probe 1 thermistors differ by more than 0.5°C | Probe thermistors show difference > 0.5°C | warning |
Baby Probe 2 Failure | Baby temperature probe 2 malfunction | Baby probe 2 disconnected, damaged, or malfunctioning | error |
Baby Probe 2 Failure | Patient probe 2 thermistors differ by more than 0.5°C | Probe thermistors show difference > 0.5°C | warning |
Bad Membrane Switch | Membrane switch malfunction detected | Membrane switch not responding or damaged | error |
Bed Heater Failure | Incubator heater not functioning or not detected | Software turned on heater but it is not on or current sense not detecting it | error |
Bed Heater Failure | Bed heating element malfunction | Bed heater not operating or overheating | error |
Bed Up/Down Pedal Failure | Bed elevation pedal control malfunction | Bed pedal not responding to input | error |
Bed Up/Down Pedal Failure | Bed pedal switch is shorted or defective | One of the bed up/down pedal switches is shorted | warning |
Bed Up/Down Pedal Switches Disabled | Bed pedal control disabled | Bed pedal switches disabled or deactivated | warning |
Calibration failed | In calibration mode, after 5 minutes both sensors not reading between 23-55 mV; unit cannot complete calibration. | Cal O2 routine running >5 minutes; sensor0 or sensor1 outside 23-55 mV range | error |
Canopy Pedal Disabled | Canopy control pedal disabled | Canopy pedal disabled or deactivated | warning |
Canopy Pedal Failure | Canopy control pedal malfunction | Canopy pedal not responding to input | error |
Canopy Pedal Failure | Canopy pedal switch is shorted or defective | One of the canopy pedal switches is shorted (appears ~100 seconds after short) | warning |
Canopy Pedal Pressed | Canopy pedal held down during power-up | Canopy pedal switch pressed and held on during power up | info |
Canopy Pedal Pressed | Canopy control pedal continuously activated | Canopy pedal stuck or held in pressed position | warning |
Check O2 supply | After calibration with both supply valves opened, sensor1 reading not greater than 23% after approximately 20 seconds. | Cal O2 routine completed; both V1 and V2 controlled to open; sensor1 <23% after 20 seconds | warning |
Down Pedal Pressed | Bed down pedal held down during power-up | Bed down pedal switch pressed and held on during power up | info |
Down Pedal Pressed | Bed down control pedal continuously activated | Down pedal stuck or held in pressed position | warning |
Fan Always in High Speed | Cooling fan stuck at high speed | Fan control unable to reduce fan speed | error |
Fan Failure | Cooling fan malfunction detected | Fan not operating or not responding to control | error |
FiO2>26% | Elevated oxygen levels in patient compartment not controlled by servo O2 system. | Measured oxygen concentration exceeds 26% and servo system cannot reduce it | warning |
Heater Doors Not Closed | Heater door assembly is not fully closed | Door sensor detects open position during operation | warning |
Heater Doors Not Opened | Heater door assembly fails to open as commanded | Door does not reach open position during startup or operation | error |
High Oxygen | Sensor1 reading more than 3% above oxygen set point. | Sensor reading exceeds (oxygen set point + 3%) | warning |
Low Oxygen | Sensor1 reading more than 3% below oxygen set point 7 minutes after set point change or canopy closed. | Sensor reading drops below (oxygen set point - 3%) and remains there for 7 minutes | warning |
Mains voltage calibration failed | Entered mains voltage differs significantly from measured value | Dialed in line voltage is 20% different than measured value (not nominal value) | error |
Motor Drive Failure | Motor drive circuit or motor has failed | Motor does not respond to control commands or stalls | error |
O2 Cal Lost-No O2 | Servo oxygen unit has never been calibrated or calibration data lost; unit will not operate. | No valid calibration data in memory; cell voltage below 33 mV (earlier software) | critical |
Oxygen Probe Failure | One sensor reading out of range (23-280 mV) or difference between two oxygen sensors exceeds 3%. | sensor0 or sensor1 reading outside 23-280 mV range; difference between oxygen0 and oxygen1 exceeds 3% | error |
Oxygen System Failure 1 | Sensor plug thermistor temperature reading out of valid range 15°C-55°C. | Temperature reading <15°C or >55°C at power-up or during operation | error |
Oxygen System Failure 2 | Relative humidity sensor reading out of valid range 1-99%; displays 0 in service mode. | Humidity reading <1% or >99% | error |
Oxygen System Failure 3 | Checksum test performed during power-up testing failed. | Checksum test failure at power-up | critical |
Oxygen System Failure 4 | Analog to digital converter circuit self-test failed. Vtest reading out of valid range 1.216-1.254 V. | A/D converter self-test failure at power-up; Vtest <1.216 V or >1.254 V | critical |
Oxygen System Failure 5 | Check O2 supply alarm silenced twice and sensor1 still not greater than 23%. | Check O2 supply alarm manually silenced twice; sensor1 remains <23% | error |
Oxygen System Failure 6 | Giraffe control board cannot communicate with Servo O2 board. | Communication failure detected during normal operation or power-up | error |
Oxygen System Failure 7 | Power-up test detected a problem with the watchdog circuit. | Watchdog circuit test failure during power-up self-test | critical |
Power Failure | Loss of main power supply | Power supply voltage drops below operating threshold | critical |
Radiant Heater Failure | Radiant heating element or heater control circuit is not functioning | Heater output drops or fails to respond to temperature demand | error |
RH Sensor Calibration Failed Try Again | Humidity sensor calibration routine failed to complete successfully | RH sensor signal falls outside the expected 65-85% RH range during calibration | error |
System Failure 0 | Communication error between display driver board and control board | Display and control board communication failure | critical |
System Failure 1 | No Timer2 | Timer2 circuit not functioning | critical |
System Failure 11 | ADC not operating to spec (too slow) | ADC conversion speed below specification | error |
System Failure 12 | Incorrect control board software checksum | System software PROM checksum mismatch | critical |
System Failure 13 | Defective SRAM | SRAM circuit malfunction | critical |
System Failure 14 | Defective WDOG | Watch dog circuit malfunction | critical |
System Failure 15 | Safety Relay 1 test failure | Unable to control heater via safety relay 1 during self-test | critical |
System Failure 16 | Heater could not be turned off during self-test | Heater remained on during power-on self-test | critical |
System Failure 17 | Radiant heater could not be turned off during calibration | Radiant heater remained on during calibration self-test | error |
System Failure 18 | Defective Variables | Detected during system startup or operation | error |
System Failure 19 | No Audio Frequency | Audio circuit malfunction detected | error |
System Failure 2 | Timer2 too fast | Timer2 running at incorrect speed | error |
System Failure 20 | Display WDOG Time Out | Display watchdog timer expired | error |
System Failure 22 | Radiant heater and incubator heater will not turn ON | Neither heater responds during operation | warning |
System Failure 23 | Unable to turn OFF heaters | Heaters remain ON when commanded OFF | warning |
System Failure 25 | Fan rotation not detected | Optical fan sensor does not detect fan motion during power-up or operation | error |
System Failure 26 | Unable to turn off safety relay 1 | Safety relay 1 does not respond to OFF command during power-up tests | critical |
System Failure 27 | Defective non-volatile memory | Memory integrity check failed | error |
System Failure 28 | Display lost communication with control board | Communication link interrupted between display and control boards | error |
System Failure 29 | Defective signal on control board | Signal integrity check failed | error |
System Failure 3 | ADC channel out of range during calibration | Calibration detected out-of-range ADC channel | error |
System Failure 30 | Mains voltage LV1 outside expected range | Measured mains voltage exceeds specified limits | warning |
System Failure 33 | Defective RH Solid State Relay | Radiant heater relay failure detected | error |
System Failure 34 | Software revision level does not match | Control board and display board software versions incompatible | error |
System Failure 35 | Wrong dipswitch setup | Dipswitch configuration on control board incorrect | warning |
System Failure 36 | LV1 differs from LV2 by >10% | Voltage comparison readings diverge beyond 10% tolerance | warning |
System Failure 37 | Defective switches (software 1.60+) | Canopy position switches show contradictory state | warning |
System Failure 38 | Door switch contradiction | Door switches show doors both open and closed | warning |
System Failure 5 | Volt reference out of range | Reference voltage outside 1.171V-1.259V range | error |
System Failure 50 | Display driver board software checksum error | System Software PROM checksum verification failed | error |
System Failure 51 | Display defective SRAM | SRAM memory test failed on display board | error |
System Failure 52 | Display no timer 2 | Timer 2 not responding on display board | error |
System Failure 53 | Display timer 2 too fast | Timer 2 running at incorrect speed | error |
System Failure 55 | Display board on-line self-test failure | Display board failed self-test during power-up or diagnostics | critical |
System Failure 6 | Defective overtemp DAC circuit | Over-temperature DAC circuit malfunction | error |
System Failure 60 | Line circuit failure | Relay board LV2 calibration circuit failure during calibration | error |
System Failure 7 | Thermistor reference voltage out of range | 1.0V thermistor reference outside 0.951V-1.049V | error |
System Failure 8 | No line frequency | 120 Hz line frequency signal missing | critical |
System Failure 98 | Overflow | Data or resource overflow condition | critical |
System Failure 99 | Logic problem on control board | Control board software logic error | critical |
Temperature out of Calibration | Temperature sensor reading exceeds calibration limits | Temperature measurement drifts outside acceptable calibration range | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.