GE Healthcare Giraffe Incubator: troubleshooting and error codes
The ServiceForge knowledge base holds 68 documented symptoms for the GE Healthcare Giraffe Incubator, with 142 ranked causes, 79 diagnostic checks with test points and expected values, and 70 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Unit will not power up, alarm sounds
Unit fails to power on and produces audible alarm sound
Ranked causes
- 1. Defective scale (unknown)
- 2. Defective scale harness (unknown)
- 3. Defective Servo Oxygen board (unknown)
- 4. Defective Servo Oxygen cable (unknown)
- 5. Defective battery (unknown)
- 6. Defective power supply (unknown)
- 7. Defective power harness to display board (unknown)
- 8. Defective display driver board (unknown)
- 9. Defective control board (unknown)
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Elevating base will not go up or down
Bed height adjustment pedal not raising or lowering bed platform
Ranked causes
- 1. Pedal switches disabled in nurse or service setup screen (unknown)
- 2. Defective toroidal transformer (unknown)
- 3. No mains input to toroidal transformer (unknown)
- 4. Defective fuse on relay board (unknown)
- 5. Defective 50 pin ribbon cable between relay and control board (unknown)
- 6. Defective elevating base motor (unknown)
- 7. Defective relay board (unknown)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low oxygen level patient safety
Sensor reading drops more than 3% below oxygen set point 7 minutes after a set point change
Ranked causes
- 1. Low O2 supply pressure or flow (common)
- 2. Inlet screen occluded (common)
- 3. Air leaks into infant compartment (occasional)
- 4. Supply valve not opening (occasional)
- 5. End cap safety valve not opening (rare)
- 6. Calibration valve stuck closed (occasional)
- 7. Infant compartment vents occluded (occasional)
Verify O2 supply pressure and flow rate
- Test point
- O2 inlet connector
- Expected
- Pressure greater than 45 psi (310 kPa), flow rate greater than 45 L/min
3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fan not rotating detection failure
Optical sensor cannot detect fan rotation
Ranked causes
- 1. Fan hub not seated properly on shaft (common)
- 2. Defective fan motor (occasional)
- 3. Old design fan without grooves (occasional)
- 4. Defective or obstructed optical sensor (occasional)
- 5. Defective ribbon cable (occasional)
- 6. Defective relay board (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ADD WATER message stays on even after water added
Water level indicator message does not clear after refilling reservoir
Ranked causes
- 1. Reservoir switch button jammed (old style units with button) (unknown)
- 2. Defective reservoir switch (unknown)
- 3. Defective 50 pin ribbon cable between relay and control board (unknown)
- 4. Defective relay board (unknown)
- 5. Defective control board (unknown)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed heater failure or no current sense high voltage
Heater commanded on but not firing or current not detected
Ranked causes
- 1. Bed safety thermostat open from heat shutdown (common)
- 2. Defective heater, thermostat, or wire harness (occasional)
- 3. Defective solid state relay (SSR) (occasional)
- 4. Defective 50-pin ribbon cable (occasional)
- 5. Defective relay board (rare)
Allow unit to cool if recently powered off hot
- Expected
- Temperature drops below 40C
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compartment probe assembly malfunction
Compartment air probe, humidity sensor, or temperature sensor requires repair or replacement due to failure or damage.
Ranked causes
- 1. Humidity sensor failure or degradation (unknown)
- 2. Temperature sensor failure (unknown)
- 3. Probe housing cover damage or connector issues (unknown)
- 4. Disconnected or damaged electrical connectors (unknown)
Inspect jack cable clip attachment
- Test point
- Hood back wall
- Expected
- Cable clip secured with 2.5 mm hex key screw
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen probe failure patient safety
One or both sensors reading out of valid range or sensors differ by more than 3%
Ranked causes
- 1. Defective sensor (occasional)
- 2. Leak or occlusion in sensor housing area (occasional)
- 3. Defective cable or connection (occasional)
- 4. Defective Servo O2 board (rare)
Verify sensor voltage readings are in valid range
- Test point
- sensor0, sensor1 in service mode
- Expected
- 23-280 millivolts for each sensor
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen system failure 1 - sensor temperature out of range patient safety
Sensor plug thermistor temperature reading outside valid range 15°C to 55°C
Ranked causes
- 1. Unit temperature too cold (common)
- 2. Cooling fan not running when temperature exceeds 50°C (occasional)
- 3. Defective sensor plug thermistor or cable (occasional)
- 4. Defective servo O2 board (rare)
Check unit ambient temperature
- Test point
- Room/incubator temperature
- Expected
- Greater than 15°C
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen system failure 5 - persistent low oxygen after alarm silenced twice patient safety
Check O2 Supply alarm silenced twice and sensor1 reading still not greater than 23%
Ranked causes
- 1. O2 supply not connected (common)
- 2. Calibration valve stuck in calibration position (occasional)
- 3. Supply valves not opening (occasional)
- 4. Leak or occlusion in sensor housing area (occasional)
Verify O2 supply connection, pressure and flow
- Test point
- O2 inlet
- Expected
- Supply connected with minimum 45 psi (310 kPa) and 45 L/min flow, >= 45 psi, >= 45 L/min
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No AC power to unit high voltage
Unit not receiving mains AC power
Ranked causes
- 1. Power cord not plugged in or switch off (common)
- 2. Mains present but 5V supply missing (occasional)
- 3. Defective power supply (occasional)
Verify wall power available
- Expected
- Live outlet with correct voltage
- Tools
- Voltmeter
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No mains voltage at humidifier transformer high voltage
Isolation transformer not receiving AC power input
Ranked causes
- 1. Defective fuse on relay board (common)
- 2. Defective fuse on humidifier interface board (rev 9 or lower) (common)
- 3. Defective relay board (rare)
Verify mains at transformer input (rev 10+)
- Test point
- J55 pins 1-2
- Expected
- Mains voltage present VAC
- Tools
- Voltmeter
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen refreshes every few seconds while scale attached
Display screen refresh rate increases when scale is connected, causing flickering or stuttering
Ranked causes
- 1. Defective circuit on control board (unbiased RS485 during idle state) (unknown)
- 2. Defective scale (unknown)
- 3. Defective scale harness (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit equipped with servo-humidity but RH setpoint shows only 1-10
Humidity control setpoint display limited to values 1-10 instead of full range
Ranked causes
- 1. Unit powered on without compartment probe plugged in (old style units) (unknown)
- 2. Defective humidity sensor (unknown)
- 3. Defective control board (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air probe sensor mismatch
Two air thermistors report temperature difference greater than 0.5C
Ranked causes
- 1. Defective compartment air probe (common)
- 2. Defective control board (rare)
Measure air thermistor resistance at harness
- Test point
- J1 pins 1-2 and 3-4
- Expected
- Within 100 ohms
- Tools
- Ohmmeter
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed down pedal pressed during startup
Bed down pedal held during power-up
Ranked causes
- 1. Pedal held during power-up (common)
- 2. Bed down pedal switch failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed up/down pedal switch failure
One or more bed pedal switches shorted or open
Ranked causes
- 1. Defective pedal switch or harness (common)
- 2. Defective relay board (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed will not lock in tilt position patient safety
Bed platform cannot be secured when tilted, does not remain at angle
Ranked causes
- 1. Tilt ball not engaged in finger pocket latch (unknown)
- 2. Tilt brake pad failure (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration failed - sensors not in valid range patient safety
In calibration mode, after 5 minutes both sensors not reading between 23 and 55 millivolts
Ranked causes
- 1. Calibration valve not opening or calibration fan not running (common)
- 2. Defective sensors (occasional)
Verify calibration fan operation when VC opened
- Test point
- VC in service mode
- Expected
- Calibration fan running and calibration valve open
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Elevating base does not respond to pedal controls
Pedals for raising or lowering the elevating base are non-functional during normal operation or system failure.
Ranked causes
- 1. System failure preventing normal pedal operation (common)
- 2. Pedal function disabled in service mode page 2 (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Elevating base motor not raising or lowering patient safety
Motor fails to respond to footswitch or control commands; base stuck in raised or lowered position.
Ranked causes
- 1. Loss of 24V DC motor supply voltage (common)
- 2. Motor control signal malfunction (incorrect J37 relay board signals) (occasional)
Verify 24V DC supply for raise function
- Test point
- J45 pins 1-2
- Expected
- +24V on pin 1, Ground on pin 2, 24 to 32 V
- Tools
- DMM
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fan running at high speed continuously
Fan cannot read heat sink sensor; stays in high speed mode
Ranked causes
- 1. Defective heat sink sensor or harness (common)
- 2. Defective control board (rare)
Measure heat sink sensor resistance
- Test point
- J3 pins 1-2
- Tools
- Ohmmeter
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
High oxygen level patient safety
Sensor reading exceeds set point by more than 3%
Ranked causes
- 1. Oxygen set point recently lowered (common)
- 2. Supply valve not closing (occasional)
Wait for oxygen stabilization after set point change
- Test point
- oxygen0, oxygen1 readings
- Expected
- Readings stabilize at new set point within several minutes
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incubator heaters not functioning high voltage
Heaters fail to maintain temperature or do not heat at all. Two 115V, 225W heaters connected in parallel powered by isolation transformer.
Ranked causes
- 1. Open circuit in bed safety thermostat or heater element (common)
- 2. Open circuit in heater circuit (revision 10 or higher) (common)
Measure incubator heater and bed safety thermostat resistance
- Test point
- Between black wire on SSR and white wire on transformer output (rev 9 or lower) or between J54 harness pin 4 and black w
- Expected
- 24 ohms (two 48 ohm heaters in parallel)
- Tools
- DMM
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No audio alarm patient safety
Audible alarm not sounding when expected during alarm condition
Ranked causes
- 1. Defective speaker or harness (unknown)
- 2. Audio driver circuitry on control board is defective (unknown)
Measure speaker resistance at relay board connector
- Test point
- J40 pins 5-6
- Expected
- about 8 ohms
- Tools
- Ohmmeter
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 Cal Lost alarm appears patient safety
Alarm displays after unit power-on or power-fail recovery, indicating oxygen sensor calibration lost
Ranked causes
- 1. Oxygen sensors require calibration after power event (common)
- 2. Oxygen sensor fault during normal operation (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 calibration lost - no O2 patient safety
Servo oxygen unit has never been calibrated; unit will not operate until initial calibration performed
Ranked causes
- 1. Initial calibration never performed (common)
- 2. Cell voltage less than 33 mV (rare)
Attempt oxygen calibration
- Test point
- Cal O2 in service mode
- Expected
- Calibration routine completes successfully
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen system failure 2 - humidity sensor out of range
Relative humidity reading out of valid range (1-99%; displays 0 in service mode when out of range)
Ranked causes
- 1. Defective sensor plug (occasional)
- 2. Defective servo O2 board (rare)
Measure humidity sensor voltage
- Test point
- J85.2 to J85.1
- Expected
- 0.7 to 4.0 Volts, 0.7-4.0 V
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen system failure 6 - communication lost with Servo O2 board
Giraffe control board cannot communicate with Servo O2 board
Ranked causes
- 1. Defective Servo O2 board (occasional)
- 2. Defective cable (occasional)
Verify control board to Servo O2 board cable connection
- Test point
- Cable connectors between boards
- Expected
- Cable fully seated and secure
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient probe thermistor mismatch
Baby probe thermistors show difference greater than 0.5C
Ranked causes
- 1. Defective patient skin probe (common)
- 2. Defective patient probe jack wiring (occasional)
Test probe in jack 1
- Expected
- Difference < 0.5C
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply voltages absent or out of range high voltage
Critical supply voltages (+5V, +5VSTBY, +12V) are not present or outside acceptable range at power bus connectors.
Ranked causes
- 1. Power supply malfunction (common)
- 2. Relay board +5VSTBY generation failure (occasional)
Measure +5V supply voltage
- Test point
- J42 pins 1-2 (orange and blue wires)
- Expected
- 4.75 to 5.25 V, 4.75 to 5.25
- Tools
- DMM
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Switch or thermostat malfunction
Membrane or mechanical switches (display panel, reservoir, elevating base, water level) do not register presses or are stuck.
Ranked causes
- 1. Open membrane switch contact or loose connector (common)
- 2. Defective water level or reservoir thermostat (occasional)
Measure alarm silence switch resistance (membrane)
- Test point
- Display Driver bd J21 pins 1-2
- Expected
- Closed (< 200 ohms when pressed), < 200
- Tools
- DMM
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Toroidal transformer no AC output voltage high voltage
Transformer secondary output at J44 absent or below 15 VAC; unit loses power to downstream circuits.
Ranked causes
- 1. Toroidal transformer primary winding or core failure (common)
- 2. Loss of AC mains or relay board failure preventing primary voltage (occasional)
Measure toroidal transformer secondary output voltage (winding 1)
- Test point
- J44 pins 1-2
- Expected
- About 15 V AC, ±1 V
- Tools
- AC voltmeter
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit always powers up in service screen mode
System boots directly into service mode instead of normal operating mode
Ranked causes
- 1. Defective dipswitch setting on control board (unknown)
- 2. Override switch (>37°C) on control panel shorted (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not switch to Baby Control Mode
System unable to activate Baby Control mode for single-patient operation
Ranked causes
- 1. Patient probe plugged into Jack 2 on probe panel (unknown)
- 2. Baby Control disabled in Service screen (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
50 pin ribbon cable intermittent or degraded
Communication errors or intermittent failures between control and relay boards; difficulty isolating cable as cause.
Ranked causes
- 1. Connector pin contamination, poor seating, or internal cable damage (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air temperature alarm activated although compartment below setpoint
Temperature alarm triggered even though measured temperature is below alarm threshold
Ranked causes
- 1. Alarm not designed to automatically reset after condition resolved (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air temperature high alarm no auto-reset
Air Temp >38C or >40C alarm remains active after temperature drops below threshold
Ranked causes
- 1. Alarm latch does not auto-reset (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baby Cold or Hot alarm does not trigger at setpoint offset patient safety
Temperature alarm fails to activate when set to 0.5°C or 1.0°C offset in Air Control mode
Ranked causes
- 1. Manual Temp Alarm disabled in Service Screen (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baby Hot or Baby Cold alarms activated in Manual Mode
Temperature alarm triggering when unit is set to Manual mode
Ranked causes
- 1. MANUAL TEMP alarm selected in Setup screen (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baby mode disabled when button pressed
Baby mode button pressed but mode is disabled in configuration
Ranked causes
- 1. Baby mode disabled in service configuration (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baby temperature alarm in air mode patient safety
Baby Hot or Baby Cold alarm activates while in Air Mode
Ranked causes
- 1. Manual temperature alarm enabled in configuration (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed up/down pedal switches disabled alarm
Pedal pressed when switches disabled in configuration
Ranked causes
- 1. Switches disabled in configuration (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed warms up slowly, may not reach set temperature
Bed temperature rises more slowly than expected or fails to reach programmed setpoint
Ranked causes
- 1. One bed heater cartridge defective (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Line voltage high voltage
Service procedure to calibrate mains voltage measurement after relay board replacement or during maintenance
Ranked causes
- 1. Relay board replaced or calibration required (common)
Access service menu
- Test point
- Override button (>37)
- Expected
- Service screen appears
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Servo Controlled Oxygen patient safety
Initial or routine calibration of the oxygen sensor system using the service mode calibration routine
Ranked causes
- 1. Calibration required during startup (common)
Verify calibration valve operation
- Test point
- VC (calibration valve)
- Expected
- Opens calibration valve and turns on calibration fan
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Servo Controlled Oxygen System patient safety
System must be calibrated at least every 24 hours when servo oxygen is in use. After 24 hours, unit prompts operator to perform calibration.
Ranked causes
- 1. Routine calibration required (common)
Calibration valve opens and calibration fan activates
- Test point
- Calibration valve and calibration fan
- Expected
- Valve open, fan running
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check O2 supply - valves opened but sensor low patient safety
After calibration, both supply valves opened and after ~20 seconds, sensor1 reading not greater than 23%
Ranked causes
- 1. O2 supply not connected (common)
Verify O2 supply connection and pressure
- Test point
- O2 inlet
- Expected
- Supply connected with minimum 45 psi (310 kPa) pressure, >= 45 psi
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Elevated oxygen level - FiO2 > 26% with Servo O2 off patient safety
Oxygen level elevated but Servo O2 system is not turned on
Ranked causes
- 1. Leak in Servo O2 system (common)
Verify Servo O2 is off and check for oxygen source
- Test point
- Servo O2 control and oxygen supply
- Expected
- Servo O2 off and no other oxygen source in use
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fan running in high speed with no ALWAYS HIGH message
Fan operating at maximum speed but system display shows no indication of high speed mode
Ranked causes
- 1. Normal condition during 90 minutes after power up or after bed mode transition (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ground resistance check high voltage
Verification of ground continuity between line cord ground pin and electronic enclosure
Ranked causes
- 1. Ground resistance exceeds safe limit (unknown)
Measure ground resistance between line cord ground pin and enclosure
- Test point
- ground pin on line cord plug to exposed metal enclosure
- Expected
- Less than 0.2 ohms, <0.2 Ω
- Tools
- electrical safety analyzer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Humidifier disabled - elevating base not functioning
Elevating base does not work; safety relay 2 not energized
Ranked causes
- 1. Defective ribbon cable to relay board (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Humidifier disabled by motor drive failure
Humidifier cannot operate due to motor drive alarm
Ranked causes
- 1. Motor drive failure active (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Humidifier heater malfunction high voltage
Humidifier does not heat or overheats. Two 225W heater elements with safety thermostat rated at 104V.
Ranked causes
- 1. Open circuit in heater element or safety thermostat (common)
Measure humidifier heater resistance (115V unit)
- Test point
- J53.1 to J53.3
- Expected
- 72 ohms (two heaters in parallel), ±5 ohms
- Tools
- DMM
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Humidifier heater or thermostat defective
Humidifier heating element or safety thermostat failure
Ranked causes
- 1. Defective heater or thermostat (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Humidifier reservoir has hairline cracks
Humidifier water container shows visible crazing or hairline fractures
Ranked causes
- 1. Sterilization damage to plastic reservoir (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Humidifier reservoir has scaling
Mineral deposits or scaling visible on inner surfaces of humidifier reservoir
Ranked causes
- 1. Use of tap water instead of distilled water (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Humidifier safety thermostat open
Humidifier heater safety thermostat open circuit
Ranked causes
- 1. Thermostat opened from overheating (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incubator heat sink sensor thermistor malfunction
Heat sink sensor temperature reading incorrect or absent; thermistor resistance out of range for operating temperature.
Ranked causes
- 1. Thermistor degraded or open circuit (common)
Measure heat sink sensor thermistor resistance at 25°C and compare to table
- Test point
- Control bd J3 pins 1-2
- Expected
- 20000 ohms at 25°C (or per table for actual temperature)
- Tools
- DMM or service screen
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leakage current measurement high voltage
Verification of leakage current at exposed metal surfaces to ensure safety compliance
Ranked causes
- 1. Leakage current exceeds safe limits (unknown)
Measure leakage current at exposed metal surfaces (120 VAC equipment)
- Test point
- exposed metal surface
- Expected
- Less than 300 microamperes, <300 µA
- Tools
- approved electrical safety analyzer
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Membrane switch malfunction
- Motor drive failure - elevating base malfunction
- Noise from speaker on power down
- Oxygen system failure 3 - checksum test failed
- Oxygen system failure 4 - A/D converter self-test failed
- Oxygen system failure 7 - watchdog circuit error
- Scale calibration required
- Temperature probe reading incorrect or absent
8 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
19 | No Audio Frequency | Audio circuit fault during system test | error |
20 | Display WDOG Time Out | Display driver board not responding during watchdog timer | error |
22 | During operation, the incubator heater will not turn ON | Heater control circuit failed or relay not engaging | critical |
23 | During operation the system was unable to turn the incubator heater OFF | Heater remains on or relay stuck engaged | critical |
25 | Fan rotation is not being detected | Optical fan sensor not detecting fan rotation; occurs at power-up if fan not rotating | warning |
26 | During the power up tests, the system was unable to turn off safety relay 1 | Safety relay 1 not disengaging during power-up test | critical |
27 | Defective non-volatile memory | Non-volatile memory circuit failure detected | error |
28 | During operation, the display driver board lost communication with control board | Serial communication failure between display and control boards | error |
29 | Defective signal on control board | Control board signal processing circuit failure | error |
30 | Mains voltage reading LV1 is outside of the expected range | Line voltage out of safe operating range or calibration circuit fault | warning |
33 | Defective RH Solid State Relay | RH solid state relay not functioning | error |
34 | Software revision level does not match | Control board and display board software versions incompatible | error |
35 | Wrong dipswitch set up | Control board dipswitch configuration incorrect | warning |
50 | Incorrect display driver board System Software checksum | Display driver board PROM checksum error | error |
51 | Display Defective SRAM | SRAM circuit on display driver board failed | error |
52 | Display No Timer2 | Timer2 function on display driver board not available | error |
53 | Display Timer2 Too Fast | Timer2 on display driver board running at incorrect speed | error |
Air Probe Failure | Air thermistor sensor mismatch | Temperature difference > 0.5C between two air probes | warning |
Air Temp >38C | Compartment temperature exceeds 38 degrees C | Ambient temperature rises above 38C | warning |
Air Temp >40C | Compartment temperature exceeds 40 degrees C | Ambient temperature rises above 40C | critical |
Baby Cold | Baby skin temperature too low | Skin probe detects temperature below setpoint | warning |
Baby Hot | Baby skin temperature too high | Skin probe detects temperature above setpoint | warning |
Baby Mode Disabled | Baby mode is disabled in service configuration | Baby mode button pressed when disabled | info |
Baby Probe 1 Failure | Patient probe 1 thermistor mismatch | Temperature difference > 0.5C in probe 1 | warning |
Baby Probe 2 Failure | Patient probe 2 thermistor mismatch | Temperature difference > 0.5C in probe 2 | warning |
Bad Membrane Switch | Front panel membrane switch panel malfunction | Switch test fails in service mode | warning |
Bed Heater Failure | Incubator bed heater not operating or not detected | Heater commanded but current not sensed | error |
Bed Up/Down Pedal Failure | Bed pedal switch shorted or open | Pedal switch stuck or not responding | warning |
Bed Up/Down Pedal Switches Disabled | Pedal switches disabled by configuration | Pedal pressed when disabled in setup | info |
Calibration failed | Calibration cannot complete; sensors not in valid range after 5 minutes | Both sensors not reading 23-55 mV during 5-minute calibration attempt | error |
Check O2 supply | After calibration and supply valve opening, sensor not detecting oxygen | Both supply valves opened and after ~20 seconds, sensor1 <23% | warning |
Down Pedal Pressed | Bed down pedal held during power-up | Pedal switch activated during startup | warning |
Fan Always in High Speed | Cooling fan locked at high speed | Heat sink sensor temperature not readable | warning |
Fan Failure | Fan rotation not detected by optical sensor | Fan not spinning or sensor malfunction | error |
FiO2>26% | Oxygen level elevated with Servo O2 system off, indicating system leak | Sensor reading >26% oxygen when Servo O2 off and no other oxygen source in use | warning |
High Oxygen | Oxygen level exceeds set point by more than 3% | Sensor1 reading more than 3% above set point | warning |
Humidifier Failure | Humidifier system malfunction | Heater on but not heating or sensor fault | warning |
Low O2 alarm | Oxygen level below acceptable threshold | During oxygen rise time test procedure | warning |
Low Oxygen | Oxygen level below set point by more than 3% | Sensor1 reading is >3% below oxygen set point 7 minutes after set point change | warning |
Mains voltage calibration failed. Please enter the mains vol | Dialed line voltage differs from measured value by 20% or more | Entered mains voltage value is 20% different than the measured value (not nominal value) | warning |
Motor Drive Failure | Elevating base motor running unintended | Motor detected running when not commanded | error |
O2 Cal Lost – No O2 | Oxygen sensor calibration data has been lost or reset | Unit power-on or return from power-fail condition | warning |
O2 Cal Lost-No O2 | Servo oxygen unit never calibrated; unit will not operate until calibrated | System powered up without prior calibration | error |
Oxygen Probe Failure | Sensor malfunction or inconsistency | One sensor reading out of range (23-280 mV) or sensors differ by >3% | error |
Oxygen System Failure 1 | Sensor temperature out of valid range 15°C to 55°C | Thermistor temperature reading outside 15-55°C range | error |
Oxygen System Failure 2 | Humidity sensor out of valid range 1-99% RH | Humidity reading out of range (displays 0 in service mode when invalid) | error |
Oxygen System Failure 3 | Microcontroller checksum test failed during power up | Checksum verification fails during system initialization | error |
Oxygen System Failure 4 | A/D converter self-test failed or Vtest out of valid range | Vtest reading outside 1.216V to 1.254V during self-test | error |
Oxygen System Failure 5 | Persistent low oxygen after Check O2 Supply alarm silenced twice | Alarm silenced twice and sensor1 still not >23% | error |
Oxygen System Failure 6 | Loss of communication between Giraffe control board and Servo O2 board | System unable to establish communication during initialization | error |
Oxygen System Failure 7 | Watchdog circuit malfunction detected during power up | Watchdog self-test fails during system initialization | error |
Power Failure | No AC power to unit or 5V supply missing | Loss of mains AC or 5V DC supply | critical |
System Failure 0 | Communication error between display driver board and control board | Self-test during power-up | critical |
System Failure 1 | No Timer2 | Self-test during power-up | critical |
System Failure 11 | ADC on control board not operating to spec (too slow) | Self-test during power-up | error |
System Failure 12 | Incorrect control board system software checksum | Self-test during power-up | critical |
System Failure 13 | Defective SRAM | Self-test during power-up | critical |
System Failure 14 | Defective WDOG | Self-test during power-up | critical |
System Failure 15 | Safety Relay 1 test failure - unable to turn incubator heater on or off | Self-test during power-up or system calibration | critical |
System Failure 16 | Incubator heater could not be turned off during power-on self test | Power-on self test | critical |
System Failure 18 | Defective Variables | Self-test during power-up | error |
System Failure 2 | Timer2 too fast | Self-test during power-up | critical |
System Failure 3 | During ADC calibration, software detected a channel out of range | During system calibration | warning |
System Failure 5 | Volt reference is out of range | Self-test during power-up | error |
System failure 55 | Display board On Line Self Test Failure | Display board self-test fails during power-up sequence | critical |
System Failure 6 | Defective overtemp DAC circuit on control board | Self-test during power-up | error |
System Failure 7 | 1.0V Thermistor reference voltage is out of range | Self-test during power-up | error |
System Failure 8 | No Line Frequency | Self-test during power-up | critical |
System failure 98 | Overflow condition detected | Software overflow or buffer overrun | error |
System failure 99 | Logic problem on control board | Control board software logic error | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.