GE Healthcare E-PP: troubleshooting and error codes
The ServiceForge knowledge base holds 28 documented symptoms for the GE Healthcare E-PP, with 50 ranked causes, 68 diagnostic checks with test points and expected values, and 26 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Abnormally low invasive pressure reading patient safety
Invasive pressure measurement displays lower than expected value
Ranked causes
- 1. Transducer wrongly positioned (common)
- 2. Defective transducer (occasional)
- 3. Module not connected (common)
- 4. Transducer adapter cable not connected (common)
- 5. Invasive pressure channel not configured or inactive (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormally low pressure reading
Invasive pressure measurement shows abnormally low values on display despite proper setup.
Ranked causes
- 1. Transducer wrongly positioned (common)
- 2. Defective transducer (occasional)
- 3. Module not connected (occasional)
- 4. Transducer adapter cable not connected (occasional)
- 5. Channel not configured with adequate priority (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No temperature displayed
Temperature measurement shows no reading or value on display.
Ranked causes
- 1. Temperature channel not configured with adequate priority (occasional)
- 2. Incompatible temperature probe (occasional)
- 3. STP setting incorrect (occasional)
- 4. Faulty temperature probe (occasional)
- 5. Temperature out of measurable range (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invasive pressure measurement over range patient safety
Displayed pressure reading exceeds 320 mmHg or 43 kPa; alarm displayed in parameter window
Ranked causes
- 1. Measurement is over range (occasional)
- 2. Transducer not zeroed correctly (common)
- 3. Sensor or cable fault (occasional)
Verify transducer zeroing procedure
- Test point
- Transducer zeroing status
- Expected
- Transducer zeroed correctly with stopcock open
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invasive pressure measurement under range patient safety
Displayed pressure reading below -40 mmHg or -5 kPa; alarm displayed in parameter window
Ranked causes
- 1. Measurement is under range (occasional)
- 2. Transducer not zeroed correctly (common)
- 3. Sensor or cable fault (occasional)
Verify transducer zeroing procedure
- Test point
- Transducer zeroing status
- Expected
- Transducer zeroed correctly; no pulsating waveform during zeroing
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure calibration error or timeout patient safety
Calibration of invasive pressure channel fails; timeout during calibration process or gain out of limits
Ranked causes
- 1. Timeout during calibration (occasional)
- 2. Pulsating waveform detected during calibration (occasional)
- 3. Gain beyond limits (rare)
Verify static pressure during calibration
- Test point
- Manometer reading
- Expected
- Steady 100-300 mmHg without pulsation
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature measurement error patient safety
T3, T4 or general temperature error; message in alarm area and parameter window
Ranked causes
- 1. Calibration error (common)
- 2. Front panel connector disconnection or poor contact (occasional)
- 3. STP board hardware failure (rare)
Perform temperature calibration
- Test point
- Temperature measurement circuit
- Expected
- Calibration completed without errors
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Functional check of invasive pressure measurement patient safety
Test invasive pressure channels (P5/P6 on E-PP, P3/P7 on E-PT) using patient simulator; verify zeroing, static pressure, and waveform accuracy
Ranked causes
- 1. Module installed or returned from maintenance; functional verification required (common)
- 2. Pressure reading out of specification; recalibration needed (occasional)
Zero the invasive pressure channel
- Test point
- Module Zero key for tested channel
- Expected
- Zeroing message followed by Zeroed message in parameter window mmHg
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature measurement removed patient safety
Temperature measurement channel has become inactive; message in alarm area
Ranked causes
- 1. Acquisition module removed (occasional)
- 2. Temperature cable disconnected (occasional)
Verify acquisition module connection
- Test point
- Module connector
- Expected
- Module fully seated in connector with no gaps
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Arterial pressure line disconnected patient safety
Pressure reading lost; disconnection alarm displayed for arterial pressure channels Art 1 through Art 8 or Fem 1 through Fem 8
Ranked causes
- 1. No arterial invasive pressure detected (common)
Verify physical connection of pressure line
- Test point
- Arterial line connector at module
- Expected
- Line connected and seated properly; no visible disconnection
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: E-PP/E-PT PT/PP setting configuration
Module PT/PP setting must be checked and reconfigured after STP board replacement to ensure correct parameter set capability.
Ranked causes
- 1. STP board replacement requires PT/PP setting verification (common)
Disconnect all other E-PP, E-PT, E-P modules except the one being configured
- Test point
- Physical connections to monitor
- Expected
- Only target module connected
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: E-PP/E-PT PT/PP setting reconfiguration
After STP board replacement, PT/PP setting must be reconfigured if current setting does not match module capability.
Ranked causes
- 1. Incorrect PT/PP setting after STP board replacement (occasional)
Access P/PT/PP Settings configuration field
- Test point
- Configuration menu
- Expected
- Setting editable
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Functional check patient safety
Verify E-PP and E-PT modules operate correctly after maintenance or repair using patient simulator
Ranked causes
- 1. Module requires functional verification after service (common)
Visual inspection of front cover
- Test point
- Module front cover
- Expected
- Intact with no damage
7 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Invasive pressure patient safety
Invasive pressure transducer readings are inaccurate or out of specification; required after transducer replacement or failed calibration check
Ranked causes
- 1. Pressure transducer replaced or calibration drift (common)
Verify module is connected to monitor
- Test point
- Module connection status
- Expected
- Module connected
12 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Patient simulator connection setup
Connect patient simulator with appropriate adapter cables for temperature and invasive blood pressure testing
Ranked causes
- 1. Patient simulator required for functional testing (common)
Prepare patient simulator and cables
- Test point
- Patient simulator equipment
- Expected
- All cables and adapters available and functional
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Temperature patient safety
Temperature probe readings are inaccurate or out of specification; required after failed calibration check or when reading outside specification limits
Ranked causes
- 1. Temperature calibration drift or check failure (common)
Verify module is connected to monitor
- Test point
- Module connection status
- Expected
- Module connected
11 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Channel zeroing in progress
IP channel is currently being zeroed; message in parameter window
Ranked causes
- 1. Normal zeroing operation underway (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Functional check of temperature measurement patient safety
Test temperature channels (T3 and T4 on E-PT) using patient simulator; verify temperature reading accuracy
Ranked causes
- 1. Module installed or returned from maintenance; temperature channel verification required (common)
Verify temperature reading accuracy
- Test point
- T3 and T4 parameter windows
- Expected
- Temperature matches configured simulator value ± 0.1 °C / ± 0.2 °F, ±0.1 °C / ±0.2 °F
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ICP channel must be zeroed separately patient safety
ICP channel requires independent zeroing from other invasive pressures; message in alarm area
Ranked causes
- 1. ICP channel grouped with other pressure channels during zeroing (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invasive pressure channel not zeroed patient safety
Alarm indicates at least one invasive pressure channel has not been zeroed before measurement
Ranked causes
- 1. Required zeroing not performed (common)
Identify channels requiring zeroing
- Test point
- Pressure channel status display
- Expected
- All active channels show zeroing status completed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Module not responding
Module does not open menu or start activity when key is pressed
Ranked causes
- 1. Loose keypad cable or module connection problem (common)
Press module key and observe menu or activity
- Test point
- Module keypad
- Expected
- Correct menu opens or activity starts
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No invasive pressure displayed patient safety
Monitor shows no pressure value or reading; pressure channel inactive or not zeroed
Ranked causes
- 1. Invasive pressure channel not zeroed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No pressure displayed
Invasive pressure measurement shows no reading or zero value on display.
Ranked causes
- 1. Invasive pressure channel not zeroed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature module calibration in progress
Module is performing self-test calibration; message in parameter window
Ranked causes
- 1. Normal calibration procedure underway (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transducer zeroing failed with high offset patient safety
Zeroing operation fails; offset error exceeds 150 mmHg indicating defective transducer
Ranked causes
- 1. Defective transducer with offset exceeding 150 mmHg (common)
Measure transducer offset value
- Test point
- Transducer offset during zeroing
- Expected
- Offset value < 150 mmHg
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Visual inspection of E-PP/E-PT pressure module caution
Check module casing, connectors, front cover, patient cables for damage, cleanliness, and proper attachment
Ranked causes
- 1. Module installed or replaced; inspection required before operation (common)
Inspect front cover for damage
- Test point
- Module front cover
- Expected
- Cover is intact with no cracks or separation
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Zero adjustment exceeds 100 mmHg patient safety
Offset during zeroing has exceeded 100 mmHg; message appears in parameter window
Ranked causes
- 1. Transducer offset error during zeroing (common)
Verify transducer zeroing procedure completed correctly
- Test point
- Parameter window
- Expected
- Zeroing offset within 100 mmHg range
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Zeroing successful
Transducer zeroing completed successfully; message in parameter window
Ranked causes
- 1. Normal completion of zeroing procedure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Art 1 disconnect to Art 8 disconnect | Arterial invasive pressure line is disconnected from module | Loss of signal from arterial pressure transducer; line not properly connected | warning |
Calibrated | Invasive pressure channel calibration completed successfully | Calibration process finished within time limit with valid pressure applied | info |
Calibrating | Calibration of invasive pressure channel is in progress | Calibration process initiated by operator or automatic procedure | info |
Calibration Error | Invasive pressure calibration procedure failed or timed out | If pressure inflation not started within 45 seconds after automatic zeroing completion, or if calibration process fails | error |
Fem 1 disconnect to Fem 8 disconnect | Femoral arterial invasive pressure line is disconnected from module | Loss of signal from femoral pressure transducer; line not properly connected | warning |
Identical PP modules | Two or more E-PP (Dual Pressure) modules detected in same system | Multiple identical pressure modules installed simultaneously | error |
Identical PT modules | Two or more E-PT (Pressure Temperature) modules detected in same system | Multiple identical PT modules installed simultaneously | error |
IP's not zeroed | One or more invasive pressure channels have not been zeroed before measurement | Measurement attempt without prior zeroing of all active channels | warning |
P1 standby to P8 standby | Invasive pressure channel set to standby mode and not active | Channel deactivated or placed in standby by user action | info |
P1 zeroing failed to P8 zeroing failed | Zeroing operation for invasive pressure channel failed due to transducer fault | Zeroing attempted with defective transducer; offset exceeds 150 mmHg threshold | error |
P5 over range to P8 over range | Invasive pressure measurement channels P5 through P8 reading exceeds 320 mmHg or 43 kPa | Pressure transducer output exceeds upper measurement limit | warning |
P5 under range to P8 under range | Invasive pressure measurement channels P5 through P8 reading below -40 mmHg or -5 kPa | Pressure transducer output below lower measurement limit | warning |
Performing temp test | Temperature module self-test calibration in progress | Module calibration sequence started | info |
Pressure measurement removed | Pressure measurement acquisition device has been disconnected or removed from system | Module or transducer physically disconnected from monitor | warning |
Pressure Sensed | Pulsating waveform detected during zeroing operation | Dynamic pressure pulsation present when zeroing procedure initiated | warning |
Sensor disconnected | Invasive pressure sensor is physically disconnected or communication lost | Sensor cable disconnected from module; transducer detection lost | warning |
T3 temperature error | T3 temperature measurement sensor or circuit failure | Temperature measurement error detected on T3 channel | error |
T4 temperature error | T4 temperature measurement sensor or circuit failure | Temperature measurement error detected on T4 channel | error |
Temperature error | General temperature measurement failure | Temperature measurement circuit malfunction detected | error |
Temp measurement removed | Temperature measurement channel has become inactive | Acquisition module or temperature cable disconnected during operation | warning |
UAC 1 disconnect to UAC 4 disconnect | Umbilical arterial catheter invasive pressure line is disconnected | Loss of signal from UAC transducer; line not properly connected | warning |
Zero adj >100 mmHg | Zero adjustment exceeds 100 mmHg limit | Transducer offset during zeroing procedure exceeds 100 mmHg | warning |
Zeroed | Transducer zeroing completed successfully | Zeroing procedure completed without exceeding 100 mmHg offset | info |
Zero Failure | Automatic zeroing of invasive pressure channel failed | During invasive pressure calibration zeroing procedure when transducer preparation is incorrect | error |
Zero ICP separately | ICP channel must be zeroed independently from other invasive pressures | Attempt to zero ICP with other pressure channels | warning |
Zeroing | ICP channel currently being zeroed | Zeroing procedure in progress | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI