GE Healthcare E-PP: troubleshooting and error codes

The ServiceForge knowledge base holds 28 documented symptoms for the GE Healthcare E-PP, with 50 ranked causes, 68 diagnostic checks with test points and expected values, and 26 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Abnormally low invasive pressure reading patient safety

Invasive pressure measurement displays lower than expected value

Ranked causes

  1. 1. Transducer wrongly positioned (common)
  2. 2. Defective transducer (occasional)
  3. 3. Module not connected (common)
  4. 4. Transducer adapter cable not connected (common)
  5. 5. Invasive pressure channel not configured or inactive (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Abnormally low pressure reading

Invasive pressure measurement shows abnormally low values on display despite proper setup.

Ranked causes

  1. 1. Transducer wrongly positioned (common)
  2. 2. Defective transducer (occasional)
  3. 3. Module not connected (occasional)
  4. 4. Transducer adapter cable not connected (occasional)
  5. 5. Channel not configured with adequate priority (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No temperature displayed

Temperature measurement shows no reading or value on display.

Ranked causes

  1. 1. Temperature channel not configured with adequate priority (occasional)
  2. 2. Incompatible temperature probe (occasional)
  3. 3. STP setting incorrect (occasional)
  4. 4. Faulty temperature probe (occasional)
  5. 5. Temperature out of measurable range (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invasive pressure measurement over range patient safety

Displayed pressure reading exceeds 320 mmHg or 43 kPa; alarm displayed in parameter window

Ranked causes

  1. 1. Measurement is over range (occasional)
  2. 2. Transducer not zeroed correctly (common)
  3. 3. Sensor or cable fault (occasional)
First check

Verify transducer zeroing procedure

Test point
Transducer zeroing status
Expected
Transducer zeroed correctly with stopcock open

2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invasive pressure measurement under range patient safety

Displayed pressure reading below -40 mmHg or -5 kPa; alarm displayed in parameter window

Ranked causes

  1. 1. Measurement is under range (occasional)
  2. 2. Transducer not zeroed correctly (common)
  3. 3. Sensor or cable fault (occasional)
First check

Verify transducer zeroing procedure

Test point
Transducer zeroing status
Expected
Transducer zeroed correctly; no pulsating waveform during zeroing

2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pressure calibration error or timeout patient safety

Calibration of invasive pressure channel fails; timeout during calibration process or gain out of limits

Ranked causes

  1. 1. Timeout during calibration (occasional)
  2. 2. Pulsating waveform detected during calibration (occasional)
  3. 3. Gain beyond limits (rare)
First check

Verify static pressure during calibration

Test point
Manometer reading
Expected
Steady 100-300 mmHg without pulsation

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature measurement error patient safety

T3, T4 or general temperature error; message in alarm area and parameter window

Ranked causes

  1. 1. Calibration error (common)
  2. 2. Front panel connector disconnection or poor contact (occasional)
  3. 3. STP board hardware failure (rare)
First check

Perform temperature calibration

Test point
Temperature measurement circuit
Expected
Calibration completed without errors

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Functional check of invasive pressure measurement patient safety

Test invasive pressure channels (P5/P6 on E-PP, P3/P7 on E-PT) using patient simulator; verify zeroing, static pressure, and waveform accuracy

Ranked causes

  1. 1. Module installed or returned from maintenance; functional verification required (common)
  2. 2. Pressure reading out of specification; recalibration needed (occasional)
First check

Zero the invasive pressure channel

Test point
Module Zero key for tested channel
Expected
Zeroing message followed by Zeroed message in parameter window mmHg

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature measurement removed patient safety

Temperature measurement channel has become inactive; message in alarm area

Ranked causes

  1. 1. Acquisition module removed (occasional)
  2. 2. Temperature cable disconnected (occasional)
First check

Verify acquisition module connection

Test point
Module connector
Expected
Module fully seated in connector with no gaps

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Arterial pressure line disconnected patient safety

Pressure reading lost; disconnection alarm displayed for arterial pressure channels Art 1 through Art 8 or Fem 1 through Fem 8

Ranked causes

  1. 1. No arterial invasive pressure detected (common)
First check

Verify physical connection of pressure line

Test point
Arterial line connector at module
Expected
Line connected and seated properly; no visible disconnection

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: E-PP/E-PT PT/PP setting configuration

Module PT/PP setting must be checked and reconfigured after STP board replacement to ensure correct parameter set capability.

Ranked causes

  1. 1. STP board replacement requires PT/PP setting verification (common)
First check

Disconnect all other E-PP, E-PT, E-P modules except the one being configured

Test point
Physical connections to monitor
Expected
Only target module connected

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: E-PP/E-PT PT/PP setting reconfiguration

After STP board replacement, PT/PP setting must be reconfigured if current setting does not match module capability.

Ranked causes

  1. 1. Incorrect PT/PP setting after STP board replacement (occasional)
First check

Access P/PT/PP Settings configuration field

Test point
Configuration menu
Expected
Setting editable

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Functional check patient safety

Verify E-PP and E-PT modules operate correctly after maintenance or repair using patient simulator

Ranked causes

  1. 1. Module requires functional verification after service (common)
First check

Visual inspection of front cover

Test point
Module front cover
Expected
Intact with no damage

7 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Invasive pressure patient safety

Invasive pressure transducer readings are inaccurate or out of specification; required after transducer replacement or failed calibration check

Ranked causes

  1. 1. Pressure transducer replaced or calibration drift (common)
First check

Verify module is connected to monitor

Test point
Module connection status
Expected
Module connected

12 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Patient simulator connection setup

Connect patient simulator with appropriate adapter cables for temperature and invasive blood pressure testing

Ranked causes

  1. 1. Patient simulator required for functional testing (common)
First check

Prepare patient simulator and cables

Test point
Patient simulator equipment
Expected
All cables and adapters available and functional

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Temperature patient safety

Temperature probe readings are inaccurate or out of specification; required after failed calibration check or when reading outside specification limits

Ranked causes

  1. 1. Temperature calibration drift or check failure (common)
First check

Verify module is connected to monitor

Test point
Module connection status
Expected
Module connected

11 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Channel zeroing in progress

IP channel is currently being zeroed; message in parameter window

Ranked causes

  1. 1. Normal zeroing operation underway (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Functional check of temperature measurement patient safety

Test temperature channels (T3 and T4 on E-PT) using patient simulator; verify temperature reading accuracy

Ranked causes

  1. 1. Module installed or returned from maintenance; temperature channel verification required (common)
First check

Verify temperature reading accuracy

Test point
T3 and T4 parameter windows
Expected
Temperature matches configured simulator value ± 0.1 °C / ± 0.2 °F, ±0.1 °C / ±0.2 °F

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ICP channel must be zeroed separately patient safety

ICP channel requires independent zeroing from other invasive pressures; message in alarm area

Ranked causes

  1. 1. ICP channel grouped with other pressure channels during zeroing (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invasive pressure channel not zeroed patient safety

Alarm indicates at least one invasive pressure channel has not been zeroed before measurement

Ranked causes

  1. 1. Required zeroing not performed (common)
First check

Identify channels requiring zeroing

Test point
Pressure channel status display
Expected
All active channels show zeroing status completed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Module not responding

Module does not open menu or start activity when key is pressed

Ranked causes

  1. 1. Loose keypad cable or module connection problem (common)
First check

Press module key and observe menu or activity

Test point
Module keypad
Expected
Correct menu opens or activity starts

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

No invasive pressure displayed patient safety

Monitor shows no pressure value or reading; pressure channel inactive or not zeroed

Ranked causes

  1. 1. Invasive pressure channel not zeroed (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

No pressure displayed

Invasive pressure measurement shows no reading or zero value on display.

Ranked causes

  1. 1. Invasive pressure channel not zeroed (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature module calibration in progress

Module is performing self-test calibration; message in parameter window

Ranked causes

  1. 1. Normal calibration procedure underway (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Transducer zeroing failed with high offset patient safety

Zeroing operation fails; offset error exceeds 150 mmHg indicating defective transducer

Ranked causes

  1. 1. Defective transducer with offset exceeding 150 mmHg (common)
First check

Measure transducer offset value

Test point
Transducer offset during zeroing
Expected
Offset value < 150 mmHg

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Visual inspection of E-PP/E-PT pressure module caution

Check module casing, connectors, front cover, patient cables for damage, cleanliness, and proper attachment

Ranked causes

  1. 1. Module installed or replaced; inspection required before operation (common)
First check

Inspect front cover for damage

Test point
Module front cover
Expected
Cover is intact with no cracks or separation

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Zero adjustment exceeds 100 mmHg patient safety

Offset during zeroing has exceeded 100 mmHg; message appears in parameter window

Ranked causes

  1. 1. Transducer offset error during zeroing (common)
First check

Verify transducer zeroing procedure completed correctly

Test point
Parameter window
Expected
Zeroing offset within 100 mmHg range

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Zeroing successful

Transducer zeroing completed successfully; message in parameter window

Ranked causes

  1. 1. Normal completion of zeroing procedure (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
Art 1 disconnect to Art 8 disconnectArterial invasive pressure line is disconnected from moduleLoss of signal from arterial pressure transducer; line not properly connectedwarning
CalibratedInvasive pressure channel calibration completed successfullyCalibration process finished within time limit with valid pressure appliedinfo
CalibratingCalibration of invasive pressure channel is in progressCalibration process initiated by operator or automatic procedureinfo
Calibration ErrorInvasive pressure calibration procedure failed or timed outIf pressure inflation not started within 45 seconds after automatic zeroing completion, or if calibration process failserror
Fem 1 disconnect to Fem 8 disconnectFemoral arterial invasive pressure line is disconnected from moduleLoss of signal from femoral pressure transducer; line not properly connectedwarning
Identical PP modulesTwo or more E-PP (Dual Pressure) modules detected in same systemMultiple identical pressure modules installed simultaneouslyerror
Identical PT modulesTwo or more E-PT (Pressure Temperature) modules detected in same systemMultiple identical PT modules installed simultaneouslyerror
IP's not zeroedOne or more invasive pressure channels have not been zeroed before measurementMeasurement attempt without prior zeroing of all active channelswarning
P1 standby to P8 standbyInvasive pressure channel set to standby mode and not activeChannel deactivated or placed in standby by user actioninfo
P1 zeroing failed to P8 zeroing failedZeroing operation for invasive pressure channel failed due to transducer faultZeroing attempted with defective transducer; offset exceeds 150 mmHg thresholderror
P5 over range to P8 over rangeInvasive pressure measurement channels P5 through P8 reading exceeds 320 mmHg or 43 kPaPressure transducer output exceeds upper measurement limitwarning
P5 under range to P8 under rangeInvasive pressure measurement channels P5 through P8 reading below -40 mmHg or -5 kPaPressure transducer output below lower measurement limitwarning
Performing temp testTemperature module self-test calibration in progressModule calibration sequence startedinfo
Pressure measurement removedPressure measurement acquisition device has been disconnected or removed from systemModule or transducer physically disconnected from monitorwarning
Pressure SensedPulsating waveform detected during zeroing operationDynamic pressure pulsation present when zeroing procedure initiatedwarning
Sensor disconnectedInvasive pressure sensor is physically disconnected or communication lostSensor cable disconnected from module; transducer detection lostwarning
T3 temperature errorT3 temperature measurement sensor or circuit failureTemperature measurement error detected on T3 channelerror
T4 temperature errorT4 temperature measurement sensor or circuit failureTemperature measurement error detected on T4 channelerror
Temperature errorGeneral temperature measurement failureTemperature measurement circuit malfunction detectederror
Temp measurement removedTemperature measurement channel has become inactiveAcquisition module or temperature cable disconnected during operationwarning
UAC 1 disconnect to UAC 4 disconnectUmbilical arterial catheter invasive pressure line is disconnectedLoss of signal from UAC transducer; line not properly connectedwarning
Zero adj >100 mmHgZero adjustment exceeds 100 mmHg limitTransducer offset during zeroing procedure exceeds 100 mmHgwarning
ZeroedTransducer zeroing completed successfullyZeroing procedure completed without exceeding 100 mmHg offsetinfo
Zero FailureAutomatic zeroing of invasive pressure channel failedDuring invasive pressure calibration zeroing procedure when transducer preparation is incorrecterror
Zero ICP separatelyICP channel must be zeroed independently from other invasive pressuresAttempt to zero ICP with other pressure channelswarning
ZeroingICP channel currently being zeroedZeroing procedure in progressinfo

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.