GE Healthcare E-COP: troubleshooting and error codes
The ServiceForge knowledge base holds 38 documented symptoms for the GE Healthcare E-COP, with 50 ranked causes, 56 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Abnormally low invasive pressure patient safety
Invasive pressure readings lower than expected; may indicate transducer or system error
Ranked causes
- 1. Transducer wrongly positioned (common)
- 2. Defective or disconnected transducer (occasional)
- 3. Module not connected or cable disconnected (common)
- 4. Invasive pressure channel not configured or inactive (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Invasive pressure patient safety
Invasive pressure transducer requires zeroing and pressure calibration to ensure accurate measurement within specification limits
Ranked causes
- 1. New pressure transducer installed (common)
- 2. Calibration check failed (occasional)
- 3. Measured value outside specification limits (occasional)
Module is connected to monitor
- Test point
- Monitor connection status
- Expected
- Module connected and powered
10 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invasive pressure measurement over range patient safety
Alarm message '> 320 mmHg or > 43 kPa' or 'P5 over range to P8 over range' displayed in parameter window
Ranked causes
- 1. Measurement over range or sensor/cable faulty (common)
- 2. Pre-zeroing with stopcock closed creating high fluid bag pressure (occasional)
- 3. Transducer not zeroed correctly (common)
Inspect cable and connections for damage or looseness
- Test point
- Invasive pressure sensor cable and module connections
- Expected
- All connections secure and undamaged
1 more check with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Module functionality not working
Module does not respond to keypad input; menu does not open or activity does not start
Ranked causes
- 1. Loose keypad cable or module connection (common)
- 2. Incompatible or malfunctioning module (occasional)
- 3. Faulty accessory or simulator connection (occasional)
Check for error messages in message field
- Test point
- Monitor display message field
- Expected
- No error messages
6 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invasive pressure measurement under range patient safety
Alarm message '< -40 mmHg or < -5 kPa' or 'P5 under range to P8 under range' displayed
Ranked causes
- 1. Measurement under range or sensor/cable faulty (common)
- 2. Transducer not zeroed correctly (common)
Inspect cable and connections
- Test point
- Invasive pressure sensor cable and module connections
- Expected
- All connections secure and undamaged
1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure calibration error within 45 seconds patient safety
Calibration error occurs when pressure application does not start within 45 seconds after automatic zeroing completes, or calibration fails during pressure application
Ranked causes
- 1. Pressure pump not activated within 45 second timeout (common)
- 2. Pressure application or measurement failure (occasional)
Time from Zero Ok message to pressure pump activation
- Test point
- Monitor display and pump operation
- Expected
- Pressure pump begins within 45 seconds of Zero Ok message
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure transducer zeroing fails patient safety
Automatic zeroing of invasive pressure channel does not complete successfully; Zero Failure message displayed
Ranked causes
- 1. Transducer not properly opened to room air (common)
- 2. Transducer connector or cable issue (occasional)
Dome stopcock fully open to room air
- Test point
- Transducer dome stopcock position
- Expected
- Stopcock lever perpendicular to pressure line, fully open
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Arterial invasive pressure disconnected patient safety
Alarm 'Art 1 disconnect to Art 8 disconnect' displayed in alarm area and parameter window
Ranked causes
- 1. No arterial invasive pressure detected (cable disconnected) (common)
Verify arterial pressure cable is connected to module
- Test point
- Arterial pressure cable connector at invasive pressure module
- Expected
- Cable fully seated and locked
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Electrical safety tests high voltage
Perform patient leakage current tests per monitor service manual Checkout procedures
Ranked causes
- 1. Post-maintenance electrical safety verification (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Functional check patient safety
Verify E-COP and E-COPSv modules operate correctly using multiparameter patient simulator and required accessories
Ranked causes
- 1. Initial setup or post-maintenance verification required (common)
Front cover inspection
- Test point
- Front cover
- Expected
- Intact with no damage
7 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Invasive pressure check failed patient safety
Recalibration required when invasive pressure calibration check does not pass
Ranked causes
- 1. Failed calibration check (occasional)
Perform invasive pressure calibration check
- Test point
- Pressure measurement output
- Expected
- Calibration check passes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiac output calibration failed patient safety
Message 'Calibration fail' displayed in C.O. menu and parameter window; calibration did not complete successfully
Ranked causes
- 1. Calibration procedure unsuccessful (unknown)
Verify calibration procedure steps were followed correctly
- Test point
- C.O. module and monitor
- Expected
- Calibration completed without error
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiac output calibration in progress
Message 'Calibrating' displayed in parameter window; C.O./CCO calibration is in progress
Ranked causes
- 1. Cardiac output calibration procedure active (common)
Allow calibration to complete
- Test point
- C.O. measurement parameter
- Expected
- Calibration completes; cardiac output value displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiac output module disconnected patient safety
Alarm 'CO measurement removed' displayed in alarm area; C.O. module has been disconnected from system
Ranked causes
- 1. C.O. module physically disconnected (common)
Verify C.O. module is physically connected to monitor
- Test point
- C.O. module connector port on monitor
- Expected
- Module firmly seated in connector
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiac output module service required patient safety
Message 'Service CO module Not available with host software version 3.1 or earlier' displayed in alarm area; blood temperature value inaccurate
Ranked causes
- 1. E-COP(Sv) module with inaccurate blood temperature measurement (rare)
Verify blood temperature reading accuracy during C.O. measurement
- Test point
- C.O. module and displayed blood temperature value
- Expected
- Blood temperature reading accurate and within expected range
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Femoral invasive pressure disconnected patient safety
Alarm 'Fem 1 disconnect to Fem 8 disconnect' displayed in alarm area and parameter window
Ranked causes
- 1. No femoral invasive pressure detected (cable disconnected) (common)
Verify femoral pressure cable is connected
- Test point
- Femoral pressure cable connector at module
- Expected
- Cable fully seated and locked
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incompatible cardiac output module patient safety
Message 'Incompatible device: COP(Sv) module Not available with host software version 3.1 or earlier' displayed in alarm area
Ranked causes
- 1. Connected C.O. module incompatible with host software version (occasional)
Verify host software version compatibility with C.O. module
- Test point
- Monitor software version and module model number
- Expected
- Module compatible with installed software version (later than 3.1)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Intracranial pressure must be zeroed separately patient safety
Alarm 'Zero ICP separately' displayed in alarm area; ICP channel must be zeroed separately from other invasive pressures
Ranked causes
- 1. ICP attempted to be zeroed with other pressures (common)
Access ICP channel setup menu
- Test point
- Monitor ICP channel menu
- Expected
- Zero option available in setup menu
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invasive pressure channel successfully zeroed
Message 'Zeroed' displayed in parameter window; zeroing was completed successfully
Ranked causes
- 1. Successful zeroing completion (common)
Verify 'Zeroed' message displayed
- Test point
- Invasive pressure channel parameter window
- Expected
- 'Zeroed' message displayed; offset within limits
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invasive pressure channel zeroing in progress
Message 'Zeroing' displayed in parameter window; channel is currently being zeroed
Ranked causes
- 1. Zeroing procedure in progress (common)
Wait for zeroing procedure to complete
- Test point
- Invasive pressure channel
- Expected
- 'Zeroing' message replaced by 'Zeroed' or error message
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invasive pressure channels not zeroed patient safety
Alarm 'IP's not zeroed' or 'PX not zeroed' (where X=1-8) displayed in alarm area and parameter window
Ranked causes
- 1. Invasive pressure channel not zeroed after setup (common)
Perform zeroing on all invasive pressure channels
- Test point
- Each invasive pressure channel (P1-P8)
- Expected
- All channels zeroed successfully; alarm clears
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invasive pressure zeroing failed patient safety
Alarm 'P1 zeroing failed to P8 zeroing failed' displayed in parameter window
Ranked causes
- 1. Defective transducer (common)
Open transducer venting stopcock to room air
- Test point
- Transducer venting stopcock
- Expected
- Stopcock fully open to atmospheric pressure
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Multiple identical cardiac output modules detected patient safety
Alarm 'Identical C.O. modules' displayed in alarm area; two or more E-COP/E-COPSv/E-PiCCO modules in system or conflicting C.O. sources
Ranked causes
- 1. Duplicate C.O. modules or conflicting sources (common)
Count active C.O. modules physically connected
- Test point
- Rear panel module connectors
- Expected
- Only one C.O. module (E-COP or E-COPSv) connected
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Multiple identical invasive pressure modules detected patient safety
Alarm 'Identical IP8 modules' displayed in alarm area; two or more E-PiCCO or E-COPSv modules mapping to same channel
Ranked causes
- 1. Duplicate IP module mapping to same channel (common)
Count active invasive pressure modules in system
- Test point
- Physical inspection of all IP modules connected to monitor
- Expected
- Only one IP module per channel
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No cardiac output module connected patient safety
Message 'No module' displayed in C.O. menu; no C.O. module detected in system
Ranked causes
- 1. Cardiac output module not connected (common)
Verify C.O. module is installed in monitor connector
- Test point
- C.O. module connector ports
- Expected
- C.O. module (E-COP or E-COPSv) physically connected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No invasive pressure reading patient safety
Invasive pressure measurement zero or absent; no pressure displayed
Ranked causes
- 1. Invasive pressure channel not zeroed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure sensed during zeroing patient safety
Message 'Pressure Sensed' displayed in parameter window; pulsation detected during zeroing procedure
Ranked causes
- 1. Venting stopcock not fully open to air during zeroing (common)
Verify venting stopcock is fully open to air
- Test point
- Transducer venting stopcock
- Expected
- Stopcock fully open; no pressure reading on gauge
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure zeroing offset exceeds limit patient safety
Message 'Zero adj >100 mmHg' displayed in parameter window; offset during zeroing exceeded 100 mmHg
Ranked causes
- 1. Excessive offset during zeroing (>100 mmHg) (common)
Repeat transducer zeroing procedure
- Test point
- Invasive pressure channel offset
- Expected
- Offset <= 100 mmHg
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ScvO2 or SvO2 measurement cable faulty patient safety
Measurement removed from display or faulty cable detected by monitor
Ranked causes
- 1. Cable connection failure or defect (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ScvO2 or SvO2 measurement not calibrated patient safety
Module connected but catheter has not been calibrated; measurement not available
Ranked causes
- 1. Catheter not calibrated after module connection (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ScvO2 or SvO2 measurement removed patient safety
Module disconnected from monitor; measurement not available
Ranked causes
- 1. Module disconnected (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ScvO2 or SvO2 optical module error patient safety
Factory calibration corrupted or red/infrared transmitter error; monitor unable to adjust transmitter currents to factory defaults
Ranked causes
- 1. Optical module calibration corrupted or transmitter failure (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ScvO2 or SvO2 signal poor patient safety
Signal pulsation, catheter touching vessel wall, or signal intensity shift; measurement unreliable
Ranked causes
- 1. Catheter malposition or signal quality degradation (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor disconnected from module patient safety
Message '[Invasive pressure channel label] X sensor disconnected' (Art/CVP/ICP/etc.) displayed; sensor disconnected or cable unplugged from module
Ranked causes
- 1. Sensor cable disconnected from module (common)
Verify sensor cable is fully seated at module connector
- Test point
- Invasive pressure sensor cable connector at module
- Expected
- Cable fully inserted and locked
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SvO2/ScvO2 calibration overdue patient safety
Alarm 'Re-calibrate ScvO2' or 'Re-calibrate SvO2' displayed in alarm area and parameter window; calibration is over 24 hours old
Ranked causes
- 1. Venous oxygenation calibration older than 24 hours (common)
Perform in vivo calibration procedure
- Test point
- ScvO2 or SvO2 measurement
- Expected
- Calibration completed; message clears; accurate readings resume
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Umbilical arterial catheter disconnected patient safety
Alarm 'UAC 1 disconnect to UAC 4 disconnect' displayed in alarm area and parameter window
Ranked causes
- 1. Invasive pressure line disconnected (common)
Verify UAC cable is connected to module
- Test point
- UAC cable connector at invasive pressure module
- Expected
- Cable fully seated and locked
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to calibrate SvO2 measurement patient safety
Calibration procedure fails; measurement cannot be initialized or recalibrated
Ranked causes
- 1. Module and optical unit incompatibility (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Venous oxygenation cable disconnected patient safety
Alarm 'ScvO2 cable off' or 'SvO2 cable off' displayed in alarm area; cable disconnected from module
Ranked causes
- 1. Venous oxygenation cable disconnected from module (common)
Verify venous oxygenation cable is connected to module
- Test point
- ScvO2 or SvO2 cable connector at module
- Expected
- Cable fully seated and locked
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI