GE Healthcare Dinamap Pro 100: troubleshooting and error codes
The ServiceForge knowledge base holds 27 documented symptoms for the GE Healthcare Dinamap Pro 100, with 43 ranked causes, 83 diagnostic checks with test points and expected values, and 17 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Incorrect, illegible, or no LCD output
LCD display shows incorrect information, is difficult to read, or shows no output at all.
Ranked causes
- 1. LCD fault (common)
- 2. Main board fault with audible alarm (occasional)
- 3. Negative supply to LCD (-9.6 Vdc) out of tolerance (occasional)
Swap LCD module with a known good unit
- Test point
- LCD display output
- Expected
- Display shows correct and legible information
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP pump sounds labored; inflation timeout or overpressure at start
At NIBP start, pump operates with labored sound and generates N33 (Inflation Timeout) or N00 (Overpressure) message.
Ranked causes
- 1. Kinked air hose or faulty cuff (common)
- 2. Blocked or kinked hose from pump output to front panel (occasional)
- 3. Pneumatic assembly failure (occasional)
Test NIBP measurement with known good cuff and hose
- Test point
- NIBP connector with test cuff
- Expected
- NIBP measurement completes without timeout or overpressure error; pump sounds normal
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 probe disconnected error
LCD displays "SpO2 probe disconnected" error message.
Ranked causes
- 1. Faulty finger probe or extender cable (common)
- 2. Faulty internal SpO2 cabling (occasional)
- 3. Faulty SpO2 PWA (occasional)
Connect known good SpO2 probe and extender cable
- Test point
- SpO2 connector port on monitor
- Expected
- Error clears; no "SpO2 probe disconnected" message
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System error P105, P110, or P115 after power-up patient safety
Failsafe alarm sounds during monitor power-up sequence and LCD displays System error code P105, P110, or P115.
Ranked causes
- 1. Faulty SpO2 PWA (common)
- 2. Isolated power supply module failure on Main board (occasional)
- 3. Main board processor SpO2 RS-232 communications failure (occasional)
Swap SpO2 PWA with a known good unit
- Test point
- SpO2 PWA installation
- Expected
- Power-up completes without failsafe alarm or system error messages
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect, missing segments, or no LED display output
LED display shows missing segments, is illegible, or shows no output.
Ranked causes
- 1. Main board fault with possible audible alarm (common)
- 2. No or low +4 Vdc LED supply (occasional)
Verify presence of audible alarm
- Test point
- Speaker/buzzer output
- Expected
- Audible alarm indicates Main board processor failure
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Menus skipped when rotor is turned
Rotor control does not properly step through menu items; selections are skipped.
Ranked causes
- 1. Rotor failure (common)
- 2. Main board failure (occasional)
Swap rotor assembly with a known good unit
- Test point
- Rotor assembly connection
- Expected
- All menu items selectable without skipping
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No LCD backlight
LCD display is not illuminated; backlight does not function.
Ranked causes
- 1. LCD faulty (common)
- 2. Main board fault (occasional)
Swap LCD module with a known good unit
- Test point
- LCD backlight illumination
- Expected
- Backlight illuminates and display is visible
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No rotor response
Rotor does not respond to user input or menu navigation commands.
Ranked causes
- 1. Rotor failure (common)
- 2. Main board failure (occasional)
Swap rotor assembly with a known good unit
- Test point
- Rotor assembly connection
- Expected
- Rotor responds to button presses and rotation
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overpressure warnings at NIBP initialization
LCD displays overpressure warning when NIBP is initialized, regardless of whether cuff and hose are connected.
Ranked causes
- 1. Kinked air hose or faulty cuff (common)
- 2. Partial or full blockage of internal pneumatic hosing (occasional)
Visually inspect external cuff and hose for kinks and blockages
- Test point
- Cuff and external hose
- Expected
- No visible kinks; hose passes air when manually inflated
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump over-current condition detected
LCD displays error message "Pump Over Current 0 (code line failure), 0, 0" indicating over-current condition on pump circuit.
Ranked causes
- 1. Pump current sensing circuitry failure on Main board (occasional)
- 2. Pump failure (occasional)
Swap Main board with a known good unit
- Test point
- Main board installation and pump control circuit
- Expected
- Pump operates normally; no over-current error displayed
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on from battery
Monitor fails to power up when powered by internal battery.
Ranked causes
- 1. Battery may be discharged (common)
- 2. PSU board fault (occasional)
Measure battery terminal voltage
- Test point
- Battery terminals
- Expected
- Voltage within normal range
- Tools
- Voltmeter
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on from battery or external AC source
Monitor fails to power up regardless of whether powered by battery or AC mains.
Ranked causes
- 1. On/Off switch/rotor faulty (common)
- 2. +5Vdc supply low (occasional)
Attempt to power on by pressing rotor switch
- Test point
- Rotor On/Off switch
- Expected
- Monitor powers on or shows menu selection capability
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Broken probe sensing failure
Temperature display does not read 106.0°F ±0.2°F when probe simulator is set to B.P., or LED display does not change to four dashes when BROKEN PROBE is pressed on simulator.
Ranked causes
- 1. Main PWA failure (common)
Verify temperature display reading with probe simulator set to B.P.
- Test point
- Temperature display
- Expected
- 106.0°F ±0.2°F, ±0.2°F
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Neonatal overpressure point verification patient safety
Verify pressure at which overpressure detection system prevents cuff inflation and generates alarm for neonatal setting
Ranked causes
- 1. Neonatal overpressure setpoint requires verification or adjustment (occasional)
Verify neonatal overpressure detection activates
- Test point
- Cuff pressure during pump inflation
- Expected
- Monitor prevents further inflation and generates alarm at neonatal overpressure setpoint
- Tools
- Manometer and pump
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP linearity and zero offset caution
Verify that pressure transducers PT1 and PT2 accurately measure across the full measurement range from 0 to 250 mmHg.
Ranked causes
- 1. Pressure transducers out of calibration or linearity error (occasional)
Set up calibration equipment and power on monitor
- Test point
- Main menu
- Expected
- Monitor displays Main menu; calibration equipment at 0 mmHg
8 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP overpressure adult caution
Verify and adjust the pressure at which the dump valve opens for adult measurements. Equipment must open valve and release pressure at specified point.
Ranked causes
- 1. Overpressure potentiometer set incorrectly or adult setting derived from incorrect neonatal setting (common)
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP overpressure neonatal caution
Verify and adjust the pressure at which the dump valve opens for neonatal measurements. Equipment must open valve and release pressure at specified point.
Ranked causes
- 1. Overpressure potentiometer (VR1) set incorrectly (common)
Verify calibration equipment at zero pressure
- Test point
- Calibration pressure measurement system
- Expected
- 0 mmHg (0 mb)
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP system adjustment
Full NIBP calibration procedure using two-point calibration at 0 and 200 mmHg; stores readings to system memory and vents to atmosphere
Ranked causes
- 1. NIBP calibration requires update or failed previous attempt (occasional)
Establish zero pressure baseline
- Test point
- Manometer and system reference
- Expected
- 0 mmHg
- Tools
- Manometer
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP system verification
Verify accuracy of NIBP readings against calibrated manometer; systolic and diastolic display pressures must match reference within tolerance
Ranked causes
- 1. NIBP calibration drift or inaccuracy (common)
Verify calibration equipment baseline
- Test point
- Manometer zero reference
- Expected
- 0 mmHg (0 mb)
- Tools
- NIST calibrated digital manometer (±0.5 mm Hg accuracy)
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Temperature system verification (PRO 200 & 400 only)
Verify that temperature measurement functions are operating correctly using IVAC probe simulator across temperature range.
Ranked causes
- 1. Temperature probe malfunction or simulator not connected (occasional)
Power on monitor and remove temperature probe from holster
- Test point
- Temperature probe holster; LCD temperature display
- Expected
- Temperature reading displayed; LED numbers flashing (real-time mode)
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressing rotor switch fails to power on unit; menu selection is possible
Rotor On/Off switch does not activate unit power despite showing menu selection capability.
Ranked causes
- 1. Rotor On/Off switch assembly faulty (common)
Swap rotor switch assembly with a known good unit
- Test point
- Rotor assembly seating and connection
- Expected
- Monitor powers on when rotor is pressed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature system self-calibration failure
LCD displays error message "TSH_HW: PIC" indicating temperature system has failed self-calibration.
Ranked causes
- 1. Monitor has failed temperature self-calibration; Main board fault (common)
Swap Main board with a known good unit
- Test point
- Main board installation and temperature calibration at startup
- Expected
- Temperature calibration completes without error; TSH_HW: PIC message does not appear
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not respond to host communications
Monitor does not receive or respond to commands from external host or communications interface.
Ranked causes
- 1. PSU board unresponsive (common)
Swap PSU board with a known good unit
- Test point
- PSU board installation and host communication port
- Expected
- Monitor responds to host commands after PSU replacement
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Verification: NIBP neonatal overpressure check caution
Verify that neonatal NIBP overpressure setting is within specified range and pressure release time is acceptable.
Ranked causes
- 1. Overpressure potentiometer set outside specification (occasional)
Set up calibration equipment and power on monitor
- Test point
- Calibration equipment and monitor
- Expected
- Setup as in Figure 4-1; monitor displays Main menu
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Verification: SpO2 probe disconnected detection
Verify that monitor correctly detects removal of SpO2 sensor and displays appropriate messages.
Ranked causes
- 1. SpO2 sensor disconnect detection malfunction (occasional)
Operate with SpO2 sensor connected at 90 BPM
- Test point
- NELLCOR simulator and LCD display
- Expected
- Normal SpO2 and pulse readings displayed
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Verification: SpO2 pulse data and saturation data
Verify SpO2 sensor operation and that monitor readings are within tolerance across full range of saturation and pulse rate settings.
Ranked causes
- 1. SpO2 sensor malfunction or simulator not connected (occasional)
Connect SpO2 simulator and power on monitor
- Test point
- SpO2 sensor input on monitor front panel
- Expected
- Monitor powers up and displays Main Menu
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Verification: SpO2 pulse search function
Verify that monitor correctly detects loss of SpO2 signal and displays appropriate message.
Ranked causes
- 1. SpO2 signal detection or alarm messaging malfunction (occasional)
Set NELLCOR simulator pulse rate to zero
- Test point
- NELLCOR simulator pulse rate setting
- Expected
- Pulse rate = 0 BPM
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
E00 | Temperature: Fail | Predictive temperature determination takes longer than 60 seconds or axillary temperature measurement attempted | warning |
E33 | Temperature: Fail | Temperature probe not connected or inoperable; temperature measurement unavailable | error |
HIGH TEMP | High Temperature Alarm | Predictive temperature reading exceeds upper range limit for patient | warning |
LOW BATTERY | Low Battery | Battery voltage drops below operational threshold | warning |
LOW BATTERY SYSTEM DISABLED | Low Battery - System Disabled | Battery voltage drops to critical level; system cannot continue operation | critical |
N00 | Overpressure | Cuff pressure exceeds safe threshold; overpressure state detected | critical |
N33 | Timeout: Inflation | Inflation time exceeds 40 seconds or air leak detected during cuff inflation | error |
N44 | Timeout: Total | Determination time exceeds 2 minutes; motion artifact detected | warning |
N55 | Timeout: Pressure | Cuff pressure elevation exceeds 1 minute; motion artifact detected | warning |
N99 | NIBP Failed | Unable to register NIBP measurement due to insufficient signal | error |
NIBP RANGE ERROR | NIBP Range Error | NIBP algorithm produces values outside target range | warning |
N, P, E, I, S (other) | Internal System Fault | System detects internal error condition | critical |
P00 | No SpO2 Sensor | SpO2 sensor not connected or not detected by monitor | error |
P55 | SpO2 No Signal | No or very weak SpO2 signal detected; inadequate sensor signal quality | error |
PRINTER NO PAPER | Printer Out of Paper | Printer detects no paper or paper door is open | warning |
SpO2 CABLE? | SpO2 Cable Defect | SpO2 sensor or cable possibly defective; cable not connected properly | warning |
SpO2 PLACEMENT? | SpO2 Sensor Placement Issue | SpO2 signal is weak or noisy; sensor position is questionable | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.