GE Healthcare Dinamap Pro 100: troubleshooting and error codes

The ServiceForge knowledge base holds 27 documented symptoms for the GE Healthcare Dinamap Pro 100, with 43 ranked causes, 83 diagnostic checks with test points and expected values, and 17 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Incorrect, illegible, or no LCD output

LCD display shows incorrect information, is difficult to read, or shows no output at all.

Ranked causes

  1. 1. LCD fault (common)
  2. 2. Main board fault with audible alarm (occasional)
  3. 3. Negative supply to LCD (-9.6 Vdc) out of tolerance (occasional)
First check

Swap LCD module with a known good unit

Test point
LCD display output
Expected
Display shows correct and legible information

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP pump sounds labored; inflation timeout or overpressure at start

At NIBP start, pump operates with labored sound and generates N33 (Inflation Timeout) or N00 (Overpressure) message.

Ranked causes

  1. 1. Kinked air hose or faulty cuff (common)
  2. 2. Blocked or kinked hose from pump output to front panel (occasional)
  3. 3. Pneumatic assembly failure (occasional)
First check

Test NIBP measurement with known good cuff and hose

Test point
NIBP connector with test cuff
Expected
NIBP measurement completes without timeout or overpressure error; pump sounds normal

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 probe disconnected error

LCD displays "SpO2 probe disconnected" error message.

Ranked causes

  1. 1. Faulty finger probe or extender cable (common)
  2. 2. Faulty internal SpO2 cabling (occasional)
  3. 3. Faulty SpO2 PWA (occasional)
First check

Connect known good SpO2 probe and extender cable

Test point
SpO2 connector port on monitor
Expected
Error clears; no "SpO2 probe disconnected" message

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System error P105, P110, or P115 after power-up patient safety

Failsafe alarm sounds during monitor power-up sequence and LCD displays System error code P105, P110, or P115.

Ranked causes

  1. 1. Faulty SpO2 PWA (common)
  2. 2. Isolated power supply module failure on Main board (occasional)
  3. 3. Main board processor SpO2 RS-232 communications failure (occasional)
First check

Swap SpO2 PWA with a known good unit

Test point
SpO2 PWA installation
Expected
Power-up completes without failsafe alarm or system error messages

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incorrect, missing segments, or no LED display output

LED display shows missing segments, is illegible, or shows no output.

Ranked causes

  1. 1. Main board fault with possible audible alarm (common)
  2. 2. No or low +4 Vdc LED supply (occasional)
First check

Verify presence of audible alarm

Test point
Speaker/buzzer output
Expected
Audible alarm indicates Main board processor failure

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Menus skipped when rotor is turned

Rotor control does not properly step through menu items; selections are skipped.

Ranked causes

  1. 1. Rotor failure (common)
  2. 2. Main board failure (occasional)
First check

Swap rotor assembly with a known good unit

Test point
Rotor assembly connection
Expected
All menu items selectable without skipping

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No LCD backlight

LCD display is not illuminated; backlight does not function.

Ranked causes

  1. 1. LCD faulty (common)
  2. 2. Main board fault (occasional)
First check

Swap LCD module with a known good unit

Test point
LCD backlight illumination
Expected
Backlight illuminates and display is visible

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No rotor response

Rotor does not respond to user input or menu navigation commands.

Ranked causes

  1. 1. Rotor failure (common)
  2. 2. Main board failure (occasional)
First check

Swap rotor assembly with a known good unit

Test point
Rotor assembly connection
Expected
Rotor responds to button presses and rotation

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Overpressure warnings at NIBP initialization

LCD displays overpressure warning when NIBP is initialized, regardless of whether cuff and hose are connected.

Ranked causes

  1. 1. Kinked air hose or faulty cuff (common)
  2. 2. Partial or full blockage of internal pneumatic hosing (occasional)
First check

Visually inspect external cuff and hose for kinks and blockages

Test point
Cuff and external hose
Expected
No visible kinks; hose passes air when manually inflated

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump over-current condition detected

LCD displays error message "Pump Over Current 0 (code line failure), 0, 0" indicating over-current condition on pump circuit.

Ranked causes

  1. 1. Pump current sensing circuitry failure on Main board (occasional)
  2. 2. Pump failure (occasional)
First check

Swap Main board with a known good unit

Test point
Main board installation and pump control circuit
Expected
Pump operates normally; no over-current error displayed

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not power on from battery

Monitor fails to power up when powered by internal battery.

Ranked causes

  1. 1. Battery may be discharged (common)
  2. 2. PSU board fault (occasional)
First check

Measure battery terminal voltage

Test point
Battery terminals
Expected
Voltage within normal range
Tools
Voltmeter

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not power on from battery or external AC source

Monitor fails to power up regardless of whether powered by battery or AC mains.

Ranked causes

  1. 1. On/Off switch/rotor faulty (common)
  2. 2. +5Vdc supply low (occasional)
First check

Attempt to power on by pressing rotor switch

Test point
Rotor On/Off switch
Expected
Monitor powers on or shows menu selection capability

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Broken probe sensing failure

Temperature display does not read 106.0°F ±0.2°F when probe simulator is set to B.P., or LED display does not change to four dashes when BROKEN PROBE is pressed on simulator.

Ranked causes

  1. 1. Main PWA failure (common)
First check

Verify temperature display reading with probe simulator set to B.P.

Test point
Temperature display
Expected
106.0°F ±0.2°F, ±0.2°F

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Neonatal overpressure point verification patient safety

Verify pressure at which overpressure detection system prevents cuff inflation and generates alarm for neonatal setting

Ranked causes

  1. 1. Neonatal overpressure setpoint requires verification or adjustment (occasional)
First check

Verify neonatal overpressure detection activates

Test point
Cuff pressure during pump inflation
Expected
Monitor prevents further inflation and generates alarm at neonatal overpressure setpoint
Tools
Manometer and pump

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: NIBP linearity and zero offset caution

Verify that pressure transducers PT1 and PT2 accurately measure across the full measurement range from 0 to 250 mmHg.

Ranked causes

  1. 1. Pressure transducers out of calibration or linearity error (occasional)
First check

Set up calibration equipment and power on monitor

Test point
Main menu
Expected
Monitor displays Main menu; calibration equipment at 0 mmHg

8 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: NIBP overpressure adult caution

Verify and adjust the pressure at which the dump valve opens for adult measurements. Equipment must open valve and release pressure at specified point.

Ranked causes

  1. 1. Overpressure potentiometer set incorrectly or adult setting derived from incorrect neonatal setting (common)

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: NIBP overpressure neonatal caution

Verify and adjust the pressure at which the dump valve opens for neonatal measurements. Equipment must open valve and release pressure at specified point.

Ranked causes

  1. 1. Overpressure potentiometer (VR1) set incorrectly (common)
First check

Verify calibration equipment at zero pressure

Test point
Calibration pressure measurement system
Expected
0 mmHg (0 mb)

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: NIBP system adjustment

Full NIBP calibration procedure using two-point calibration at 0 and 200 mmHg; stores readings to system memory and vents to atmosphere

Ranked causes

  1. 1. NIBP calibration requires update or failed previous attempt (occasional)
First check

Establish zero pressure baseline

Test point
Manometer and system reference
Expected
0 mmHg
Tools
Manometer

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: NIBP system verification

Verify accuracy of NIBP readings against calibrated manometer; systolic and diastolic display pressures must match reference within tolerance

Ranked causes

  1. 1. NIBP calibration drift or inaccuracy (common)
First check

Verify calibration equipment baseline

Test point
Manometer zero reference
Expected
0 mmHg (0 mb)
Tools
NIST calibrated digital manometer (±0.5 mm Hg accuracy)

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Temperature system verification (PRO 200 & 400 only)

Verify that temperature measurement functions are operating correctly using IVAC probe simulator across temperature range.

Ranked causes

  1. 1. Temperature probe malfunction or simulator not connected (occasional)
First check

Power on monitor and remove temperature probe from holster

Test point
Temperature probe holster; LCD temperature display
Expected
Temperature reading displayed; LED numbers flashing (real-time mode)

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pressing rotor switch fails to power on unit; menu selection is possible

Rotor On/Off switch does not activate unit power despite showing menu selection capability.

Ranked causes

  1. 1. Rotor On/Off switch assembly faulty (common)
First check

Swap rotor switch assembly with a known good unit

Test point
Rotor assembly seating and connection
Expected
Monitor powers on when rotor is pressed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature system self-calibration failure

LCD displays error message "TSH_HW: PIC" indicating temperature system has failed self-calibration.

Ranked causes

  1. 1. Monitor has failed temperature self-calibration; Main board fault (common)
First check

Swap Main board with a known good unit

Test point
Main board installation and temperature calibration at startup
Expected
Temperature calibration completes without error; TSH_HW: PIC message does not appear

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not respond to host communications

Monitor does not receive or respond to commands from external host or communications interface.

Ranked causes

  1. 1. PSU board unresponsive (common)
First check

Swap PSU board with a known good unit

Test point
PSU board installation and host communication port
Expected
Monitor responds to host commands after PSU replacement

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Verification: NIBP neonatal overpressure check caution

Verify that neonatal NIBP overpressure setting is within specified range and pressure release time is acceptable.

Ranked causes

  1. 1. Overpressure potentiometer set outside specification (occasional)
First check

Set up calibration equipment and power on monitor

Test point
Calibration equipment and monitor
Expected
Setup as in Figure 4-1; monitor displays Main menu

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Verification: SpO2 probe disconnected detection

Verify that monitor correctly detects removal of SpO2 sensor and displays appropriate messages.

Ranked causes

  1. 1. SpO2 sensor disconnect detection malfunction (occasional)
First check

Operate with SpO2 sensor connected at 90 BPM

Test point
NELLCOR simulator and LCD display
Expected
Normal SpO2 and pulse readings displayed

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Verification: SpO2 pulse data and saturation data

Verify SpO2 sensor operation and that monitor readings are within tolerance across full range of saturation and pulse rate settings.

Ranked causes

  1. 1. SpO2 sensor malfunction or simulator not connected (occasional)
First check

Connect SpO2 simulator and power on monitor

Test point
SpO2 sensor input on monitor front panel
Expected
Monitor powers up and displays Main Menu

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Verification: SpO2 pulse search function

Verify that monitor correctly detects loss of SpO2 signal and displays appropriate message.

Ranked causes

  1. 1. SpO2 signal detection or alarm messaging malfunction (occasional)
First check

Set NELLCOR simulator pulse rate to zero

Test point
NELLCOR simulator pulse rate setting
Expected
Pulse rate = 0 BPM

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
E00Temperature: FailPredictive temperature determination takes longer than 60 seconds or axillary temperature measurement attemptedwarning
E33Temperature: FailTemperature probe not connected or inoperable; temperature measurement unavailableerror
HIGH TEMPHigh Temperature AlarmPredictive temperature reading exceeds upper range limit for patientwarning
LOW BATTERYLow BatteryBattery voltage drops below operational thresholdwarning
LOW BATTERY SYSTEM DISABLEDLow Battery - System DisabledBattery voltage drops to critical level; system cannot continue operationcritical
N00OverpressureCuff pressure exceeds safe threshold; overpressure state detectedcritical
N33Timeout: InflationInflation time exceeds 40 seconds or air leak detected during cuff inflationerror
N44Timeout: TotalDetermination time exceeds 2 minutes; motion artifact detectedwarning
N55Timeout: PressureCuff pressure elevation exceeds 1 minute; motion artifact detectedwarning
N99NIBP FailedUnable to register NIBP measurement due to insufficient signalerror
NIBP RANGE ERRORNIBP Range ErrorNIBP algorithm produces values outside target rangewarning
N, P, E, I, S (other)Internal System FaultSystem detects internal error conditioncritical
P00No SpO2 SensorSpO2 sensor not connected or not detected by monitorerror
P55SpO2 No SignalNo or very weak SpO2 signal detected; inadequate sensor signal qualityerror
PRINTER NO PAPERPrinter Out of PaperPrinter detects no paper or paper door is openwarning
SpO2 CABLE?SpO2 Cable DefectSpO2 sensor or cable possibly defective; cable not connected properlywarning
SpO2 PLACEMENT?SpO2 Sensor Placement IssueSpO2 signal is weak or noisy; sensor position is questionablewarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI