GE Healthcare DINAMAP PRO 100-400V2: troubleshooting and error codes
The ServiceForge knowledge base holds 35 documented symptoms for the GE Healthcare DINAMAP PRO 100-400V2, with 53 ranked causes, 27 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
System error P105, P110, or P115 displayed after power-up patient safety
Fail-safe alarm sounds during monitor power-up sequence; one of system error messages appears (P105, P110, or P115).
Ranked causes
- 1. Failed SpO2 sensor or cable (common)
- 2. SpO2 PWA faulty (occasional)
- 3. Isolated +5 VDC power supply out of specification (occasional)
- 4. Main board fault (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Illegible or no LCD display
LCD screen shows garbled text, blank, or unreadable content.
Ranked causes
- 1. LCD failure (common)
- 2. Main board fault (usually with audible alarm) (occasional)
- 3. Negative supply to LCD (-9.6 VDC) out of tolerance (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overpressure warnings displayed when NIBP initialized
Monitor shows overpressure error message when NIBP measurement starts; occurs with or without cuff and hose.
Ranked causes
- 1. Occluded air hose or cuff (common)
- 2. Partial or full blockage of internal pneumatic tubing (occasional)
- 3. Valve assembly faulty (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump sounds labored at NIBP start with inflation timeout or overpressure messages
NIBP pump produces labored sound during initialization; N33 (Inflation Timeout) or N00 (Overpressure) error messages appear.
Ranked causes
- 1. Reversed connections on pneumatic assembly failure (common)
- 2. Blocked or kinked hose from pump output (common)
- 3. Pneumatic assembly failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External DC power or battery charging failure
Monitor does not charge battery when external power connected; external power indicator flashes instead of glowing continuously
Ranked causes
- 1. Battery not installed (occasional)
- 2. External DC input voltage too low (occasional)
Verify battery is installed
- Test point
- Battery compartment
- Expected
- Battery present and properly seated
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
External power LED flashes instead of steady glow
External power indicator LED flashes every four seconds instead of continuously glowing green
Ranked causes
- 1. Battery not installed (common)
- 2. External DC input voltage too low (common)
Verify battery installation
- Test point
- Battery compartment
- Expected
- Battery present and properly seated
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Incorrect, missing segments, or no LED display output
LED displays show missing segments, garbled characters, or no output at all.
Ranked causes
- 1. Main board fault (usually with audible alarm) (common)
- 2. No or low +4 VDC LED supply (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor will not switch on when powered by battery
Monitor does not power on when battery is installed; no LED or display activity.
Ranked causes
- 1. Battery may be discharged (common)
- 2. PSU board fault (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No LCD backlight
LCD screen is completely dark with no illumination; display content not visible.
Ranked causes
- 1. LCD faulty (common)
- 2. Main board fault (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump Over Current condition detected caution
Error message 'Pump Over Current 0, (code line failure), 0, 0' displayed; pump drawing excessive current.
Ranked causes
- 1. Pump over-current condition signaled to main board (common)
- 2. Pump failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 probe disconnected error displayed patient safety
Monitor displays 'SpO2 probe disconnected' error message; SpO2 measurement unavailable.
Ranked causes
- 1. Faulty probe (common)
- 2. SpO2 PWA failed (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not respond to host communications
Monitor does not communicate with external host system or interface; no data exchange.
Ranked causes
- 1. PSU board unresponsive (common)
- 2. Broken continuity of PSU board to host comm port cable (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not switch on from battery or AC; no AC mains light
Monitor does not power on from either battery or external AC source; AC power indicator LED is dark.
Ranked causes
- 1. Locked processor (common)
- 2. PSU Module fault (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery LED flashes and battery icon changes on LCD
Battery LED flashes every four seconds and battery icon on LCD changes state; indicates low battery power
Ranked causes
- 1. Battery power depleted or battery nearing end of life (common)
Reconnect external power supply
- Test point
- AC/DC power input
- Expected
- Battery LED becomes off; battery charges
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery LED off when battery connected and external power disconnected
Battery LED does not glow yellow or battery icon does not appear on LCD after battery installation and external power removal
Ranked causes
- 1. Battery not properly installed or defective (unknown)
Verify battery installation
- Test point
- Battery compartment
- Expected
- Battery LED glows yellow; battery icon appears and toggles with time indicator on LCD
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not charging or not holding charge
Battery fails to charge or rapidly loses charge after charging; does not maintain power level.
Ranked causes
- 1. PSU PWA battery charging function faulty (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery will not hold charge
Battery repeatedly fails to maintain charge even after full charging cycles
Ranked causes
- 1. Battery end-of-life; depleted storage battery (common)
Verify battery has been fully charged
- Test point
- Battery LED and LCD battery icon
- Expected
- Battery LED off; battery icon stable on LCD (not flashing)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP pressure
Monitor enters service mode on power-up if NIBP system is uncalibrated; requires recalibration procedure
Ranked causes
- 1. NIBP system uncalibrated (common)
Observe power-up self-test sequence
- Test point
- LCD display
- Expected
- Software version displayed in inverted format (white on black)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP system patient safety
Check and calibrate blood pressure transducers (PT1, PT2) and overpressure circuitry via Service Mode menu
Ranked causes
- 1. Routine NIBP calibration maintenance (common)
Connect Digital Pressure Gauge to NIBP system
- Test point
- NIBP connector
- Expected
- Gauge reads 0-375 mmHg with 0.2% accuracy
- Tools
- Digital Pressure Gauge 0-375 mmHg (Digitron P200L or similar)
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SpO2 system check patient safety
Verify SpO2 sensor integrity and algorithm response for Nellcor or Masimo equipped monitors
Ranked causes
- 1. SpO2 simulator settings cause algorithm to miss pulse rate (occasional)
Set up Nellcor SpO2 simulator test
- Test point
- SpO2 connector
- Expected
- Monitor displays detected pulse rate and saturation
- Tools
- NELLCOR SpO2 Simulator Model BIO-TEK INDEX 2 MF or SRC-MAX Portable Tester
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD display or backlight malfunction
LCD pixels do not illuminate or backlight does not power on during initialization
Ranked causes
- 1. LCD or backlight circuit failure (unknown)
Observe LCD and backlight during power-up in dimly lit room
- Test point
- LCD display
- Expected
- All pixels light momentarily; backlight powered; software version displayed inverted (white on black)
LCD display or backlight not functioning at power-up
LCD does not display software version in inverted format or backlight does not power during initialization
Ranked causes
- 1. LCD or backlight module failure (unknown)
Observe LCD display during power-up in dimly lit room
- Test point
- LCD display and backlight
- Expected
- All pixels lit momentarily; software version displayed in inverted format (white on black); backlight powers on
LED display malfunction
LED digits do not illuminate properly during power-up self-test sequence
Ranked causes
- 1. LED display failure or connection issue (unknown)
Verify LED segment illumination during power-up
- Test point
- LED digit display
- Expected
- All seven segments illuminate in sequence beginning with upper left segment
LED display segments not illuminated during power-up
Seven-segment LED digits fail to light or do not illuminate in proper sequence during initialization
Ranked causes
- 1. LED display failure or internal connection issue (unknown)
Observe LED display segment illumination during power-up
- Test point
- LED display digits
- Expected
- All seven segments light in sequence beginning with upper left segment
Mains power switching failure high voltage
Monitor will not switch from battery power to mains power when plugged into AC power source; remains on battery only
Ranked causes
- 1. Open mains fuse on AC mains converter board (common)
Confirm mains fuse is open
- Test point
- AC mains converter board
- Expected
- Fuse element not visible or resistance reads open (infinite ohms)
- Tools
- Ohmmeter or multimeter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not change from battery to mains power
Unit fails to switch to mains power when plugged into AC source; remains on battery power
Ranked causes
- 1. Mains fuse blown (non-serviceable) (common)
Verify unit operation with mains power only
- Test point
- AC power input
- Expected
- Unit operates
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor will not retain date and time
Date and time settings reset after power cycle; monitor does not keep time when powered off.
Ranked causes
- 1. Backup battery low or disconnected (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No rotor selection or menu skips
Menu navigation does not respond to rotor input; menu items skip or do not advance.
Ranked causes
- 1. Main board failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pneumatic system contamination caution
Foreign substances or moisture introduced into pneumatic system causing calibration errors or measurement failures
Ranked causes
- 1. Improper cleaning or immersion of hoses, cuffs, or equipment (occasional)
Inspect pneumatic hoses and cuff connectors for moisture or debris
- Test point
- Hose interior and connection points
- Expected
- Hoses and connectors dry and free of visible contamination
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressing rotor fails to power on unit; menu selection possible
Power button (rotor) does not activate unit; menu navigation works but unit remains in low-power state.
Ranked causes
- 1. Rotor ON/OFF switch malfunction (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Remote alarm switch does not change state at power-up
Remote alarm switching accessory fails to clear alarm state or does not respond to monitor power cycles
Ranked causes
- 1. Remote alarm switch module failure or communication link issue (unknown)
Observe remote alarm switch state during monitor power-up and power-down
- Test point
- Remote alarm switch accessory
- Expected
- Alarm state when monitor is off; clears during monitor power-up
Remote alarm switch malfunction
Remote alarm switch does not change states during power-up initialization
Ranked causes
- 1. Remote alarm switch hardware failure (unknown)
Verify remote alarm state changes during power-up
- Test point
- Remote alarm switch
- Expected
- Alarm state clears when Monitor powers up (transitions from alarm state to normal state)
Speaker audio distortion or no sound
Speaker produces distorted audio or fails to produce sound during power-up self-test
Ranked causes
- 1. Speaker failure or audio circuit malfunction (unknown)
Listen to speaker output during power-up
- Test point
- Speaker output
- Expected
- Three short beeps followed by brief pause then three more beeps
Speaker generates distorted or no sound at power-up
Audio system produces distorted sounds or is silent during three-beep startup sequence
Ranked causes
- 1. Speaker unit failure or audio amplifier malfunction (unknown)
Observe audio output during power-up sequence
- Test point
- Speaker output
- Expected
- Three short beeps followed by brief pause, then three more beeps; clear undistorted sound
TSH_HW:PIC displayed on main LCD
Message 'TSH_HW:PIC' appears on monitor LCD indicating temperature self-calibration failure.
Ranked causes
- 1. Monitor has failed temperature self-calibration (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.