GE Healthcare dicom configuration um lu17719 encore v18 1: troubleshooting and error codes

The ServiceForge knowledge base holds 4 documented symptoms for the GE Healthcare dicom configuration um lu17719 encore v18 1, with 6 ranked causes, 7 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

DICOM Store or Print transmission fails

Reports cannot be sent to PACS or print destinations; jobs remain queued or fail to transmit

Ranked causes

  1. 1. AE Title, IP address or Port number incorrect (common)
  2. 2. Network connection failed (common)
First check

Verify DICOM node configuration parameters

Test point
DICOM Queue window Nodes/Edit Node List
Expected
AE Title, IP address and Port match PACS configuration

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

DICOM Worklist or MPPS query fails

Worklist or MPPS connection cannot establish; DICOM Echo Test fails

Ranked causes

  1. 1. Network connection failed (common)
  2. 2. AE Title, IP address or Port incorrect (common)
First check

Run DICOM Echo Test

Test point
DICOM Worklist Configure Connection window
Expected
Echo Test passes with HIS/RIS server

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: DICOM MPPS Configuration

Setup and configuration of DICOM Modality Performed Procedure Step (MPPS) service for HIS/RIS/PACS system integration

Ranked causes

  1. 1. MPPS service requires configuration for exam start/completion notification (common)
First check

Verify DICOM communications between enCORE and MPPS SCP

Test point
DICOM Echo Test button
Expected
Successful connection test

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: MPPS Attributes Mapping

Map HIS/RIS coding standards to enCORE MPPS attributes for correct information exchange

Ranked causes

  1. 1. HIS/RIS uses custom codes requiring mapping to enCORE attributes (common)
First check

Select attributes to be sent in MPPS N-CREATE message (exam start)

Test point
Configure Attributes dialog
Expected
Required attributes identified and enabled

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI