GE Healthcare Definium AMX 700: troubleshooting and error codes
The ServiceForge knowledge base holds 42 documented symptoms for the GE Healthcare Definium AMX 700, with 63 ranked causes, 85 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No 12V output from voltage converters high voltage
12V tap from main converter shows very low or 0V voltage; may indicate current overload or converter failure
Ranked causes
- 1. Current overload on 12V outputs (common)
- 2. Converter over-temperature (occasional)
- 3. Input over-voltage condition (occasional)
- 4. Input under-voltage condition (occasional)
- 5. Converter failure (occasional)
Measure voltage at debug connector pin 2 (12V tap from converter output)
- Test point
- Debug connector pin 2
- Expected
- 12V nominal
- Tools
- DMM
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No 24V output from voltage converters high voltage
24V tap from main converter shows very low or 0V voltage; may indicate current overload or converter failure
Ranked causes
- 1. Current overload on 24V outputs (common)
- 2. Converter over-temperature (occasional)
- 3. Input over-voltage condition (occasional)
- 4. Input under-voltage condition (occasional)
- 5. Converter failure (occasional)
Measure voltage at debug connector pin 1 (24V tap from converter output)
- Test point
- Debug connector pin 1
- Expected
- 24V nominal
- Tools
- DMM
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Detector vertical bar, bad pixels and gain radiation
Correct vertical bar artifacts, detect defective pixels, and calibrate gain coefficients for uniform detector response. Three-step procedure in sequence: vertical bar, bad pixel, then gain calibration.
Ranked causes
- 1. Vertical bar artifacts from temperature transient variation (common)
- 2. Defective pixels in detector (common)
- 3. Non-uniform pixel gain across detector (common)
Execute vertical bar calibration step
- Test point
- Software calibration tool
- Expected
- On-screen instructions complete
6 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Beetle interface board LED not illuminated caution
Beetle interface board shows no lights on
Ranked causes
- 1. Loss of power to Beetle interface board (common)
- 2. Serial or ribbon cables reversed (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Monitor brightness and contrast (DTX)
Calibrate DTX monitor brightness and contrast using SMPTE test pattern with spot meter measurements. Brightness should display 95% white visible in 100% white, contrast should display 5% black visible in 0% black.
Ranked causes
- 1. Monitor brightness needs adjustment for ambient light conditions (common)
- 2. Monitor contrast needs adjustment (common)
Verify X-ray tube in un-parked position and monitor powered on for 20 minutes minimum
- Test point
- X-ray tube, monitor
- Expected
- Tube un-parked, monitor on and stabilized
6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Monitor brightness and contrast (GVISION)
Calibrate GVISION monitor brightness and contrast using SMPTE test pattern with spot meter measurements. Brightness and contrast ranges same as DTX monitor.
Ranked causes
- 1. Monitor brightness needs adjustment for ambient light conditions (common)
- 2. Monitor contrast needs adjustment (common)
Verify X-ray tube in un-parked position and monitor powered on for 20 minutes minimum
- Test point
- X-ray tube, monitor
- Expected
- Tube un-parked, monitor on and stabilized
6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PACS LUT functionality test
Verify PACS can select and apply different curve modes (Normal/Harder/Softer) and that CR-Fallback mode operates correctly when DX images are rejected
Ranked causes
- 1. PACS cannot select alternative curve modes (unknown)
- 2. CR-Fallback mode configured incorrectly when DX modality rejected (unknown)
Verify Images #1, 4, 5 appear visually identical
- Test point
- PACS review workstation display of Images #1, 4, 5
- Expected
- All three images show identical appearance and band/square counts
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PACS monitor display quality verification
Verify PACS monitor calibration and ability to display correct number of gray-scale bands and intensity squares from test images
Ranked causes
- 1. PACS monitor not calibrated to DICOM standards (unknown)
- 2. SMPTE 5% and 95% intensity squares not equally visible (unknown)
Verify PACS monitor DICOM calibration status
- Test point
- PACS monitor and calibration records
- Expected
- Monitor calibrated per DICOM standards
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mobile PC LED not illuminated caution
Mobile PC has no lights on; indicates power or connection issue
Ranked causes
- 1. Power switch not ON (common)
- 2. Loss of power or connection issue to Mobile PC (common)
Verify Mobile PC power switch is ON
- Test point
- Mobile PC power switch
- Expected
- Switch in ON position
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor and touch screen not functioning caution
Monitor and/or touch screen has no power indicated by lack of LEDs
Ranked causes
- 1. Power switch not ON (common)
- 2. Loss of power to monitor (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No 12V at LCD display (LCD_12V) caution
12V feed to LCD shows no voltage despite green 12V LED illuminated and red LCD 12V LED on; likely circuit breaker trip
Ranked causes
- 1. Circuit breaker tripped due to LCD over-current (common)
- 2. LCD display short circuit or failure (common)
Measure voltage at debug connector pin 5 (LCD_12V)
- Test point
- Debug connector pin 5
- Expected
- 12V
- Tools
- DMM
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No 24V at Beetle Board (IF_24V) high voltage
24V feed to Beetle Board shows no voltage despite green 24V LED illuminated; likely circuit breaker trip
Ranked causes
- 1. Circuit breaker tripped due to over-current (common)
- 2. Beetle Board short circuit or excessive load (common)
Measure voltage at debug connector pin 3 (IF_24V to Beetle Board)
- Test point
- Debug connector pin 3
- Expected
- 24V
- Tools
- DMM
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No 24V at PC power supply (PC_24V) high voltage
24V feed to PC shows no voltage despite green 24V LED illuminated and red PC 24V LED on; likely circuit breaker trip
Ranked causes
- 1. Circuit breaker tripped due to PC over-current (common)
- 2. PC short circuit or power failure (common)
Measure voltage at debug connector pin 4 (PC_24V)
- Test point
- Debug connector pin 4
- Expected
- 24V
- Tools
- DMM
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
USB cables not properly connected caution
USB cables may have improper connections or lack strain relief
Ranked causes
- 1. USB cable lacks strain relief (occasional)
- 2. USB cable connected to incorrect port (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Detector calibration radiation
Periodic calibration of X-ray detector required as part of planned maintenance schedule
Ranked causes
- 1. Scheduled maintenance requirement at 12-month interval (common)
Perform detector calibration
- Test point
- Detector assembly
- Expected
- Detector calibrated and functional
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Detector Vertical Bar radiation
Correct vertical bar artifacts due to temperature transient variation in detector response
Ranked causes
- 1. Temperature transient variation in detector (common)
Launch Service Tools from application desktop
- Test point
- Desktop Service Tools icon
- Expected
- Service Tools window opens
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Digital Detector
Perform detector calibration as part of planned maintenance every 12 months
Ranked causes
- 1. Scheduled preventive maintenance (common)
Inspect detector for damage
- Test point
- Detector surface, grid holder, tether
- Expected
- No sharp edges, no visible damage
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Hand Switch
Clean, disinfect, and perform visual/functional checks of hand switch every 12 months; replace every 3 years
Ranked causes
- 1. Scheduled preventive maintenance (common)
Visual inspection of cord
- Test point
- Hand switch cord and RJ11 connector
- Expected
- No nicks, cuts, severe pinching, or exposed wires at connector
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Monitor
Perform monitor calibration and touch screen calibration as part of planned maintenance every 12 months
Ranked causes
- 1. Scheduled preventive maintenance (common)
Clean monitor screen
- Test point
- Monitor display surface
- Expected
- Clean screen without streaks or residue
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Monitor calibration
Periodic calibration of display monitor required as part of planned maintenance schedule
Ranked causes
- 1. Scheduled maintenance requirement at 12-month interval (common)
Perform monitor calibration
- Test point
- Display monitor
- Expected
- Monitor calibrated and image quality verified
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PACS VOI-LUT support verification
Determine if PACS system correctly supports Value-of-Interest Look-up-Table (VOI-LUT) for proper image display. Send test images with and without Burn-on-Send mode, compare visual appearance and DICOM headers on PACS review workstation.
Ranked causes
- 1. PACS does not support VOI-LUT or applies it incorrectly (unknown)
Load PACS Test Images onto Definium AMX 700 system
- Test point
- Image Tools or DICOM-formatted CDROM
- Expected
- Patient Name: GE_PACS_Test_Image Definium, Patient ID: 2000002
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Printer
Execute printer calibration following on-screen instructions from the calibration tool.
Ranked causes
- 1. Printer output needs color and density calibration (unknown)
Execute printer calibration tool
- Test point
- Printer calibration software
- Expected
- On-screen instructions appear and complete
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Test image all bands visibility verification
Verify ability to see all 5 intensity squares in top band of test images
Ranked causes
- 1. Monitor contrast insufficient to display all top-band squares (unknown)
Verify all 5 squares visible in Image #6 top band
- Test point
- Image #6 left side top dark band squares
- Expected
- All 5 intensity squares distinctly visible in top band
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Touch screen
Execute touch screen calibration following on-screen instructions from the calibration tool.
Ranked causes
- 1. Touch screen coordinates need recalibration (unknown)
Execute touch screen calibration tool
- Test point
- Touch screen calibration software
- Expected
- On-screen instructions appear and complete
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Touchscreen calibration
Periodic calibration of touchscreen interface required as part of planned maintenance schedule
Ranked causes
- 1. Scheduled maintenance requirement at 12-month interval (common)
Perform touchscreen calibration
- Test point
- Touchscreen panel
- Expected
- Touchscreen calibrated and responsive
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Clean Barcode Reader Window caution
Clean barcode reader window and housing as part of planned maintenance every 12 months
Ranked causes
- 1. Scheduled preventive maintenance or degraded reading performance (common)
Inspect barcode reader window and housing
- Test point
- Scanner window and housing surface
- Expected
- No visible dirt, scratches, or damage
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Clean Tech Switch and check cable connections caution
Inspect and clean tech switch as part of planned maintenance every 12 months
Ranked causes
- 1. Scheduled preventive maintenance (common)
Inspect tech switch cable and connector
- Test point
- Cable, RJ11 connector, connection points
- Expected
- No visible wear, cuts, pinching, exposed wires, or damage
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Complete system software load/upgrade patient safety
Re-install or upgrade system's complete software including operating system and application software on workstation PC
Ranked causes
- 1. Need to re-install or upgrade system software (unknown)
Archive all protocols, patient files, and backup calibration and configuration data
- Test point
- System storage
- Expected
- All critical data backed up before installation begins
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detector power supply LED not illuminated caution
Detector power supply has no lights on; indicates power issue
Ranked causes
- 1. Loss of power to detector power supply (common)
Observe detector power supply LEDs
- Test point
- Detector power supply front panel
- Expected
- LEDs illuminated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
DVD drive malfunction caution
DVD drive has reported problems; tray may not function properly
Ranked causes
- 1. Manual closure of DVD drive tray (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hospital network and detector ethernet cables reversed caution
Hospital network and detector ethernet connections may be reversed causing communication issues
Ranked causes
- 1. Hospital network and detector ethernet cables reversed (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspection: Cable ties and cable retention caution
Verify that all cable ties remain tight and secure, and that detector power supply cables and retention clips are properly seated
Ranked causes
- 1. Scheduled mechanical inspection at 12-month interval (common)
Verify all cable ties are tight and secure
- Test point
- Cable ties throughout system
- Expected
- All cable ties tight and secure
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspection: Detector, tether and grid condition caution
Visual inspection to identify damage, wear or deterioration of detector assembly and associated components
Ranked causes
- 1. Scheduled mechanical inspection at 12-month interval (common)
Visual inspection of detector, tether and grid
- Test point
- Detector assembly, tether cable, grid
- Expected
- No visible damage, wear or deterioration
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspection: Monitor condition and mounting caution
Visual inspection of monitor for damage or wear and verification of secure mounting to top cover
Ranked causes
- 1. Scheduled mechanical inspection at 12-month interval (common)
Inspect monitor for damage or wear
- Test point
- Monitor assembly
- Expected
- No visible damage or wear
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspection: PC mounting security caution
Verify that PC is still securely mounted and has not loosened during operation
Ranked causes
- 1. Scheduled mechanical inspection at 12-month interval (common)
Verify PC is mounted securely
- Test point
- PC mounting bolts and fasteners
- Expected
- All mounting fasteners tight and secure
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspection: PC shock mounts condition caution
Visual inspection to identify damage or wear on PC shock mount components
Ranked causes
- 1. Scheduled mechanical inspection at 12-month interval (common)
Inspect PC shock mounts for damage or wear
- Test point
- PC shock mounts
- Expected
- Mounts intact and undamaged
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mobile power supply assembly LED not illuminated caution
Mobile power supply assembly shows no lights; indicates power issue
Ranked causes
- 1. Loss of power or connection to mobile power supply assembly (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No voltage after 2A fuse (12V converter input) high voltage
2A_FUSE_SCALED output shows 0V when proper input voltage is present; fuse is likely blown
Ranked causes
- 1. 2A fuse blown (common)
Measure voltage at debug connector pin 6 (2AMP_FUSE_SCALED)
- Test point
- Debug connector pin 6
- Expected
- 9.8-12.3V, 9.8-12.3V
- Tools
- DMM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No voltage after 3A fuse (detector power supply output) high voltage
3A_FUSE_SCALED output shows 0V when proper input voltage is present; fuse is likely blown
Ranked causes
- 1. 3A fuse blown (common)
Measure voltage at debug connector pin 7 (3AMP_FUSE_SCALED)
- Test point
- Debug connector pin 7
- Expected
- 9.8-12.3V, 9.8-12.3V
- Tools
- DMM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No voltage after 5A fuse (24V converter input) high voltage
5A_FUSE_SCALED output shows 0V when proper input voltage is present; fuse is likely blown
Ranked causes
- 1. 5A fuse blown (common)
Measure voltage at debug connector pin 8 (5AMP_FUSE_SCALED)
- Test point
- Debug connector pin 8
- Expected
- 9.8-12.3V, 9.8-12.3V
- Tools
- DMM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
TRAD detector LED not illuminated caution
TRAD detector has no lights on; indicates power or connection issue
Ranked causes
- 1. Loss of power to TRAD detector (common)
Observe TRAD detector LEDs
- Test point
- TRAD detector front panel
- Expected
- LEDs illuminated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Workstation PC operating system corruption risk caution
Incorrect power cycling during software installation can corrupt Linux operating file system on workstation PC, requiring OS reload
Ranked causes
- 1. Power cycled at wrong time during software installation (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI