GE Healthcare Definium 8000: troubleshooting and error codes
The ServiceForge knowledge base holds 56 documented symptoms for the GE Healthcare Definium 8000, with 75 ranked causes, 150 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Collimator light field intensity low radiation
Light output from collimator is below minimum acceptable level (less than 15 foot candles / 160 lux at 100 cm from focal spot)
Ranked causes
- 1. Lamp, mirror, Mylar window dirty or discolored (common)
- 2. Lamp voltage too low (common)
- 3. Mirror installed backwards (occasional)
- 4. Supply cable too long (occasional)
- 5. Faulty collimator bulb (common)
- 6. Faulty lamp socket (occasional)
Measure lamp voltage
- Test point
- directly across lamp or bulb socket
- Expected
- Ultranet GE lamp: 23.0 VAC min; Ultranet Philips lamp: 21.5 VAC min; AMX: 20.0 VAC min; Siemens: 22.2 VAC min; Eclipse Proteus: 13.4 VDC min, + meter tolerance
- Tools
- Fluke Digital Multimeter on 200V scale or equivalent
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Technique accuracy or indirect linearity test fails radiation
Pass flag is not intact after exposure testing at specified kVp and mA settings; one or more flags indicate measurement deviation
Ranked causes
- 1. Meter calibration out of tolerance (common)
- 2. mA station out of calibration (common)
- 3. kV calibration out of tolerance (common)
- 4. Linearity error in measurement system (occasional)
Take minimum of 6 exposures for technique accuracy
- Test point
- RAD tube and SF tube in R013
- Expected
- One exposure each at 60, 80, and 130 kVp for each filament; use XS=100 mA and XL=500 mA at 20 mAs
- Tools
- Keithley non-invasive kV meter, software version 3382 form
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Meter calibration overdue or non-compliant radiation
Radiation measuring instruments without current calibration or not calibrated per regulatory requirements
Ranked causes
- 1. Calibration interval has expired (common)
- 2. Meter requires recalibration after repair or malfunction (common)
- 3. Suspected meter damage from drop or abuse (rare)
Verify calibration sticker shows current calibration date within interval
- Test point
- Meter calibration sticker
- Expected
- Next calibration date is in the future; current date is before that date
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auto collimator does not enter or remain in AUTO mode at correct SID radiation
Collimator fails to automatically control shutters or enters manual mode outside the required SID range of 90–130 cm
Ranked causes
- 1. SID detection switch or sensor malfunction (unknown)
- 2. Control logic or mode selection circuit failure (unknown)
Verify SID scale and SID indication present and readable
- Test point
- Collimator SID scale
- Expected
- SID scale visible and SID indicator present
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Collimator blades not visible or misaligned with X-ray field radiation
Four collimator blades or rectangular blades not visible on monitor during fluoroscopy, or X-ray field extends beyond display by more than tolerance
Ranked causes
- 1. Collimator gain incorrectly set (common)
- 2. Collimator blade mechanical misalignment or decentering (occasional)
Verify collimator is in automatic mode
- Test point
- Collimator blades joystick button
- Expected
- Lights ON beside joystick
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collimator corner blades misaligned (Discovery IGS systems) radiation
Corner blades of octagonal X-ray field not parallel to monitor edges or decentered during head-up display mode
Ranked causes
- 1. Corner blade decentering or rotation relative to system (occasional)
- 2. Collimator corner blade gain setting incorrect (common)
Verify head-up display mode activation
- Test point
- DL console rotation button
- Expected
- Rotation button highlighted (active)
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
In-table fluoro collimator does not limit beam to cassette or film size radiation
Shutters fail to automatically limit beam size when cassette is inserted or during transfer between fluoroscopy and record modes
Ranked causes
- 1. Cassette detection switch or positioning sensor failure (common)
- 2. Shutter motor or solenoid malfunction (occasional)
Insert cassette and switch SFD selector to AUTO
- Test point
- SFD mode selector switch, cassette position
- Expected
- Shutters limit beam to visual area in fluoroscopy mode
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Inaccurate or unstable radiation meter readings radiation
Meters showing noisy, unstable, or unreliable measurements due to condensation or environmental factors
Ranked causes
- 1. Condensation on meter due to temperature exposure (occasional)
- 2. Meter malfunction or damage (rare)
Inspect meter for visible condensation on display or probe window
- Test point
- Meter exterior and probe entrance
- Expected
- No condensation visible; clear optics
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manual collimator installed in unsupported system configuration patient safety
Manual collimator used in system type not listed in HHS certification requirements or without required documentation
Ranked causes
- 1. Collimator model not certified for installed system type (rare)
- 2. Missing on-site documentation permitting manual collimator use (occasional)
Identify radiographic system type and compare to Table 3-2
- Test point
- System type nameplate, service manual
- Expected
- System type matches one of the approved entries in Table 3-2 (Mobile, Rapid Film Changer, Single Size Cassette & Fixed SID, Room W/O Permanent Bucky, Single purpose systems, or documented other types)
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manual fluoro collimator does not limit beam to maximum receptor size radiation
Shutters exceed maximum beam size when manually controlled or fail to prevent opening beyond format size after cassette transfer
Ranked causes
- 1. Shutter limit stop or mechanical stop worn or misaligned (occasional)
- 2. Format size sensor or automatic size-change logic failure (occasional)
Switch SFD selector to MANUAL
- Test point
- SFD mode selector
- Expected
- Collimator enters manual shutter control mode
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Octagonal X-ray field compliance test failed (Discovery IGS with Head up display) radiation
Test Data Results report shows fail in compliance of octagonal X-ray field measurements, discrepancies, overlap percentage, or maximum extent for Discovery IGS systems with Head up display mode
Ranked causes
- 1. Blade not visible or too visible on monitor (common)
- 2. Alignment of detector or collimator fail (common)
Activate Head up display mode and measure octagonal X-ray field
- Test point
- Monitor display with rotation button highlighted
- Expected
- Measurements c1', c2', d1', d2' recorded in cm on monitor
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Square X-ray field compliance test failed radiation
Test Data Results report shows fail in compliance of square X-ray field measurements or discrepancies between X-field and image reception area
Ranked causes
- 1. Blade not visible or too visible on monitor (common)
- 2. Alignment of detector (common)
Verify all measurement data points recorded
- Test point
- Test Data Results report
- Expected
- All c1, c2, d1, d2 values recorded and formulas correctly applied
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
600 mAs/2000 mAs backup not actuating or not resetting radiation
Exposure does not terminate at 600 mAs (≥50 kVp) or 2000 mAs (<50 kVp); manual reset not required or visual indication absent after backup activation
Ranked causes
- 1. Backup circuit malfunction or not configured per specification (unknown)
Verify 600 mAs/2000 mAs backup terminates exposure and manual reset is required
- Test point
- Generator output; exposure timer
- Expected
- Exposure terminates at 600 mAs when kVp ≥50 kVp or at 2000 mAs when kVp <50 kVp; manual reset required before next automatic exposure
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible tone or buzzer not functioning during exposure termination patient safety
No audible signal when radiographic exposure ends
Ranked causes
- 1. Buzzer or tone circuit failure (unknown)
Verify audible tone or buzzer sounds at radiographic exposure termination
- Test point
- Generator audible output; expose without film
- Expected
- Audible tone or buzzer sounds when exposure ends
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible tones not sounding at exposure termination patient safety
Radiographic or fluoroscopic exposure does not produce audible tone or buzzer at termination
Ranked causes
- 1. Audible tone/buzzer system malfunction (unknown)
Verify audible tone sounds at radiographic exposure termination
- Test point
- Radiographic exposure termination
- Expected
- Audible tone or buzzer sounds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration accuracy accounting for meter tolerance radiation
Setting exposure rate limits to account for radiation meter calibration error when measuring table top output
Ranked causes
- 1. Meter has specified calibration tolerance that affects measurement accuracy (common)
Read calibration accuracy specification from meter or probe
- Test point
- Meter calibration sticker or probe specification label
- Expected
- Accuracy listed as percentage (e.g., ±5%, ±10%)
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: AEC dose table verification radiation
Verify automatic exposure control dose output for table position meets specification
Ranked causes
- 1. AEC dose calibration drift (unknown)
Measure AEC dose output at table position
- Test point
- AEC detector output
- Expected
- Pass
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: AEC dose wall stand verification radiation
Verify automatic exposure control dose output for wall stand position meets specification
Ranked causes
- 1. AEC dose calibration drift (unknown)
Measure AEC dose output at wall stand position
- Test point
- AEC detector output
- Expected
- Pass
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Beam quality HVL measurement radiation
Verify half-value layer (HVL) meets specification for beam quality
Ranked causes
- 1. Beam quality out of specification (unknown)
Measure beam quality HVL
- Test point
- X-ray beam
- Expected
- Pass
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Collimator alignment at 100 cm large focal spot radiation
Verify collimator alignment meets specification at 100 cm SID with large focal spot
Ranked causes
- 1. Collimator misalignment (unknown)
Collimator alignment at 100 cm SID large focal spot
- Test point
- Collimator output
- Expected
- Pass
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Collimator alignment at 100 cm small focal spot radiation
Verify collimator alignment meets specification at 100 cm SID with small focal spot
Ranked causes
- 1. Collimator misalignment (unknown)
Collimator alignment at 100 cm SID small focal spot
- Test point
- Collimator output
- Expected
- Pass
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Fluoro Collimator to Spot-film Device centering radiation
Center-to-center displacement between fluoro collimator and spot-film device (SFD) image centers exceeds acceptable limits across different format and angulation combinations
Ranked causes
- 1. Fluoro collimator and SFD misalignment (unknown)
Retrieve factory data for Fluoro Collimator to SFD centering
- Test point
- Factory data form
- Expected
- Two worst-case format and angulation combinations identified
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Fluoro Collimator to Spot-film Device field size radiation
Field size (width and length) of spot-film device image does not match factory specification across different format and angulation combinations
Ranked causes
- 1. Fluoro collimator and SFD field size mismatch (unknown)
Verify factory material thickness specification
- Test point
- Factory form for SFD test
- Expected
- Total thickness of material between table hard surface and test tool bottom specified by Factory
6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Image overlap test for fixed cones or cylinders radiation
Verify X-ray field alignment and overlap with image receptor for fixed cones or cylinders used at fixed SID with single image receptor size
Ranked causes
- 1. Image overlap or misalignment with receptor (unknown)
Align X-ray source to receptor
- Test point
- Source to receptor alignment
- Expected
- Normal operator alignment procedures completed
8 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: kVp/mA & Generator mAs specification accuracy radiation
Verify that kVp and mAs (or mA and time) match selected values within specified accuracy at installation, preventative maintenance, and repair.
Ranked causes
- 1. Calibration verification required (common)
Verify half value layer before testing
- Test point
- Equipment setup
- Expected
- Half value layer verified
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Light field size
Light field size adjustment to 25 cm square (12.5 cm per direction)
Ranked causes
- 1. Default light field requires adjustment (unknown)
Verify current light field size setting
- Test point
- Collimator display or control panel
- Expected
- Current field size displayed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: mR/mAs value verification radiation
Verify exposure rate per milliampere-second meets specification
Ranked causes
- 1. Tube output drift (unknown)
Measure exposure rate mR/mAs
- Test point
- X-ray tube output
- Expected
- Pass
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Power and leakage test USA high voltage
Measure leakage current and voltage at ground for USA critical care and wet location installations
Ranked causes
- 1. Excessive leakage current or improper grounding (unknown)
Measure leakage voltage and current per USA regulatory requirements
- Test point
- System grounds and chassis
- Expected
- Within regulatory limits
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Quality assurance phantom test radiation
Verify image quality using QA phantom meets specification
Ranked causes
- 1. Image quality degradation (unknown)
Evaluate QA phantom image quality
- Test point
- Phantom image
- Expected
- Pass
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Center to Center misalignment exceeds tolerance radiation
X-ray field center does not align with image receptor center; detected during collimator alignment test procedures
Ranked causes
- 1. Tube or receptor mechanical misalignment (unknown)
Insert cassette and establish SID
- Test point
- Bucky or cassette holder
- Expected
- 12 x 14 inch cassette centered in Bucky
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collimator blades do not move to new position within required time radiation
Shutters or blades respond slowly to mode changes or SID adjustments, taking longer than expected to reach final position
Ranked causes
- 1. Collimator mechanical or drive system malfunction (unknown)
Measure time for collimator blades to move to new position
- Test point
- Collimator shutters
- Expected
- less than 5 seconds, <5 seconds
Collimator blades not visible in magnification mode radiation
One or more collimator blades are not visible in the intensifier image during fluoro magnification modes
Ranked causes
- 1. Collimator to intensifier sizing misalignment (unknown)
Verify all four collimator blades visible in each magnification mode
- Test point
- Intensifier image display
- Expected
- All four collimator blades visible
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collimator does not inhibit exposures when exposure inhibit condition is detected radiation
System allows X-ray firing when an inhibit condition (such as missing cassette or incorrect mode) is present
Ranked causes
- 1. Exposure inhibit safety circuit or interlock relay failure (unknown)
Create exposure inhibit condition (e.g. remove cassette, change SID incorrectly)
- Test point
- Cassette position sensor, mode selector, SID interlock
- Expected
- Collimator indicates inhibit condition
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Collimator gain value out of limits after alignment
After performing detector alignment, test still fails and Maximum Collimator Gain Allowed needs to be set to enable test passage
Ranked causes
- 1. Actual collimator gain exceeds maximum allowed value for primary barrier coverage (occasional)
Retrieve Max Collimator Gain Allowed from Test Data Results Report
- Test point
- Test Data Results Report
- Expected
- Maximum gain value for current detector and FOV configuration
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collimator has minimum 30 cm or 38 cm focal spot distance not maintained patient safety
Focal spot to patient distance falls below HHS minimum of 30 cm for radiography or 38 cm for fluoroscopy due to missing or detached spacer
Ranked causes
- 1. Skin spacer or handle extension missing or detached from collimator (occasional)
Inspect collimator for attached skin spacer or handle extension
- Test point
- Collimator body, spacer attachment points
- Expected
- Spacer or extension is visibly attached and secure (if designed into collimator)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Coolant hoses require labeling and disconnection from conditioner caution
Hoses must be identified before disconnection to ensure correct reconnection during reassembly
Ranked causes
- 1. Need to identify outlet and inlet hoses for proper reassembly (common)
Coolant hose outlet connection identified and labeled
- Test point
- Conditioner outlet hose connection
- Expected
- Hose labeled '1' with provided marker
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Coolant replacement required caution
Table detector coolant conditioner requires removal and replacement
Ranked causes
- 1. Scheduled coolant maintenance or coolant degradation (unknown)
Power off the system
- Test point
- Main power switch
- Expected
- System powered down
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Coolant replacement required on table detector caution
Table conditioner coolant system requires maintenance or replacement per service schedule or kit availability
Ranked causes
- 1. Scheduled coolant maintenance or replacement (unknown)
System is powered off before beginning procedure
- Test point
- Main power switch
- Expected
- Off position confirmed
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Coolant reservoir refill after purging caution
System requires coolant refilling into the conditioner reservoir after table and detector purge procedures have been completed.
Ranked causes
- 1. Routine maintenance: coolant level depletion during purge procedure (common)
Verify distilled water fully removed from system hoses and detector
- Test point
- System hoses and detector
- Expected
- No distilled water present
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Coolant spill during refilling caution
Coolant spilled onto electronics in table during filling procedure
Ranked causes
- 1. Improper funnel placement or overfilling (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Coolant spill on electronics caution
Risk of coolant contacting electronic components during refill
Ranked causes
- 1. Improper funnel placement or overfilling during coolant addition (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Coolant system leak or contamination caution
Visible coolant dripping from conditioner or hoses during removal procedure; fluid appears discolored or cloudy
Ranked causes
- 1. Old or degraded coolant requires replacement (common)
Inspect coolant hoses and connections for leaks
- Test point
- Hose connections and conditioner outlet
- Expected
- No active leaks or drips visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Definium 8000 kVp output above calibration range radiation
K-edge measurement indicates kVp reading near or exceeding K-edge pack limit, rejected from calibration
Ranked causes
- 1. K-edge pack reading near upper limit of measurement range (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluoro collimator to image intensifier misalignment radiation
Visible area edges on monitor do not correspond to X-ray field edges; collimator opening exceeds minimum specification or alignment percentages out of tolerance.
Ranked causes
- 1. Collimator blades misaligned relative to image intensifier (unknown)
Collimator minimum opening size
- Test point
- Collimator blades at minimum
- Expected
- less than 5 cm square
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image overlap test: Field size verification radiation
Verify calculated X-ray field size does not exceed receptor nominal dimensions when centers are misaligned
Ranked causes
- 1. Field size exceeds receptor dimensions or misalignment exceeds 2% of SID (rare)
Compare measured center misalignment to 2% SID threshold
- Test point
- Form entry for center misalignment
- Expected
- Misalignment within 2% of SID
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Light to X-Ray field misalignment radiation
X-ray field does not align with light field; misalignment in lateral or longitudinal directions
Ranked causes
- 1. Collimator misalignment requiring adjustment (unknown)
Set beam angulation and select manual collimator mode
- Test point
- Collimator control panel
- Expected
- Manual mode selected; if remote table, may require AUTO then format selection then override/OBL mode
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement data verification radiation
Tests for Stenoscop performed simultaneously; results recorded on Form F3382SS data record chart.
Ranked causes
- 1. Simultaneous test execution for Stenoscop (common)
Confirm all measurements recorded on Form F3382SS
- Test point
- Data record chart
- Expected
- All kVp and mAs values documented
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Missing X-ray warning label patient safety
X-ray warning label not present on equipment
Ranked causes
- 1. Warning label missing or damaged (unknown)
Verify X-ray warning label presence
- Test point
- Below the lock for rotational motion of the C arm
- Expected
- Warning label present and readable
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Preventive maintenance: Chiller air filter cleaning
Clean chiller air filter to maintain cooling efficiency
Ranked causes
- 1. Air filter clogging reduces cooling efficiency (unknown)
Inspect and clean air filter
- Test point
- Chiller air filter
- Expected
- Complete
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Preventive maintenance: Chiller refill caution
Refill chiller fluid to specified level
Ranked causes
- 1. Low chiller fluid level (unknown)
Verify chiller fluid level
- Test point
- Chiller reservoir
- Expected
- Complete
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Preventive maintenance: Visual inspection
Complete visual inspection of system components per maintenance checklist
Ranked causes
- 1. Component wear or damage (unknown)
Visual inspection of all listed components
- Test point
- Cabling, grease cup, screw, table, bearings, tracks, bumpers, hand grips
- Expected
- Complete
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe saturation during beam quality measurements radiation
Incorrect low readings when probe ion collection efficiency drops below 95% during high-intensity short pulse exposures, especially during beam quality testing
Ranked causes
- 1. Peak field intensity exceeds probe intensity limit during short radiation pulses (common)
Determine if exposure rate will exceed probe saturation limit
- Test point
- Probe type and exposure parameters
- Expected
- Peak exposure rate below probe maximum per Table 5-1 or 5-2
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SID indicator accuracy out of tolerance radiation
Difference between indicated SID on operator scale and actual measured SID exceeds 2% (rejection limit 1.8%)
Ranked causes
- 1. Tube position indicator miscalibrated (unknown)
Set tube to common SID and measure actual distance
- Test point
- focal spot locator decal to film plane
- Expected
- Actual distance matches operator indicator scale
- Tools
- pocket tape measure, Sears tape measure
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vascular collimator permits X-ray exposure at incorrect SID radiation
System allows exposure when SID is outside the correct or specified range, creating patient and operator safety hazard
Ranked causes
- 1. SID interlock switch or detector malfunction (common)
Position collimator at incorrect SID and attempt to fire X-ray
- Test point
- SID position indicator, exposure enable circuit
- Expected
- Exposure is prevented; no X-ray output
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
X-ray field exceeds image receptor on lateral or inner edges radiation
X-ray field boundary extends beyond image receptor on left, right, or inner wall edge
Ranked causes
- 1. X-ray collimator misalignment (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
X-ray warning label missing or incorrect patient safety
Warning label not present on master control or does not bear required statement
Ranked causes
- 1. Warning label not installed or damaged (unknown)
Verify X-ray warning label present on master control with correct statement
- Test point
- Master control panel
- Expected
- Label present bearing: 'WARNING: This x-ray unit may be dangerous to patient and operator unless safe exposure factors and operating instructions are observed'
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.