GE Healthcare Dash 3000: troubleshooting and error codes
The ServiceForge knowledge base holds 129 documented symptoms for the GE Healthcare Dash 3000, with 214 ranked causes, 178 diagnostic checks with test points and expected values, and 35 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No waveform at central station
Waveform display missing at central station despite patient monitor active
Ranked causes
- 1. Central station software incompatible (occasional)
- 2. Cable connection loose or faulty (common)
- 3. Incorrect alarm or print location configuration (occasional)
- 4. Care unit name mismatch between devices (occasional)
- 5. IP addresses configured incorrectly (common)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure waveforms do not appear on display
BP parameter labels, readings, and waveforms do not display; AR1 or PA2 not showing
Ranked causes
- 1. Faulty patient simulator or test cables (common)
- 2. Bent or broken pins on BP side panel connectors (occasional)
- 3. Broken continuity in BP signal path (occasional)
- 4. DAS assembly defect (occasional)
Verify patient simulator and test cables on working monitor
- Test point
- Working monitor BP display
- Expected
- BP waveforms and parameters appear correctly
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG waveforms do not display at all
No ECG waveforms appear on monitor display despite simulator connected
Ranked causes
- 1. Faulty ECG patient cables (common)
- 2. Faulty patient simulator (common)
- 3. Acquisition PCB defect (occasional)
- 4. Processor PCB or power supply PCB defect (rare)
Test ECG patient cables
- Test point
- Working monitor ECG connector
- Expected
- ECG signal appears on working monitor
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invasive blood pressure waveforms do not appear on display patient safety
AR1 or PA2 parameter labels, readings, and waveforms not displayed on monitor; BP side panel shows no signal
Ranked causes
- 1. Patient simulator or cables faulty (common)
- 2. BP side panel connector pins bent or broken (occasional)
- 3. Continuity fault in BP signal path (occasional)
- 4. DAS assembly faulty (occasional)
Verify patient simulator and BP cables on working monitor
- Test point
- Working monitor with suspect simulator and cables
- Expected
- AR1 and PA2 waveforms and parameters display normally
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Review runtime error logs
Access and review software error logs stored in monitor for troubleshooting. View input/output errors with timestamps and parameters.
Ranked causes
- 1. Advanced troubleshooting required (occasional)
- 2. View output software errors (occasional)
- 3. View input software errors (occasional)
- 4. Clear error logs (occasional)
Navigate to REVIEW ERRORS in Service Mode
- Test point
- Service Mode Menu display
- Expected
- REVIEW ERRORS option visible and selectable
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
802.11b association status incorrect (not 01)
Wireless LAN ID field shows association status other than 01, indicating monitor not connected to network or incorrect credentials
Ranked causes
- 1. SSID/PASS KEY incorrect (common)
- 2. Wireless card firmware requires reset (occasional)
- 3. Monitor not connected to network (occasional)
Display WIRELESS LAN ID field
- Test point
- Monitor Setup Menu, REVISION AND ID, HARDWARE ID DISPLAY
- Expected
- WIRELESS LAN ID field displays status code (e.g., 01 G3.93-73e 55 04)
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
802.11b signal quality below threshold
WIRELESS LAN ID field shows signal quality status less than 20, indicating weak wireless signal or network coverage issue
Ranked causes
- 1. Monitor out of range from access point (common)
- 2. RF interference from nearby equipment (occasional)
- 3. Access point radio issue (occasional)
Verify signal quality status value
- Test point
- WIRELESS LAN ID field third field value
- Expected
- greater than 20
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm audio does not sound caution
No audio output when alarm is triggered or during audio test
Ranked causes
- 1. Speaker cable loose or disconnected (common)
- 2. Speaker failed (occasional)
- 3. Audio circuit on processor PCB failed (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Buzz signal during ECG or BP patient safety
Unwanted buzz signal output during ECG or blood pressure measurement
Ranked causes
- 1. BP not zeroed (common)
- 2. BP not available (common)
- 3. ECG leads failed (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Buzz signal output instead of tone for ECG or BP alarms patient safety
Defib sync monitor outputs a buzz or buzzer tone instead of normal alarm tone for ECG or blood pressure measurements
Ranked causes
- 1. Blood pressure not zeroed (common)
- 2. Blood pressure signal not available or disconnected (occasional)
- 3. ECG leads disconnected from patient or monitor (occasional)
Check BP zeroing status
- Test point
- Blood pressure module and monitor display
- Expected
- BP reference zeroed and displayed as baseline value
- Tools
- Monitor interface and BP module controls
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Defib Sync: Buzz signal for ECG or BP patient safety
Incorrect buzz signal output during defibrillator sync when ECG or BP is active
Ranked causes
- 1. BP not zeroed (common)
- 2. BP not available (occasional)
- 3. ECG leads fail (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG waveforms display incorrectly with noise
Displayed ECG waveforms contain significant amount of noise instead of clean signal
Ranked causes
- 1. Faulty ECG patient cables (common)
- 2. Faulty patient simulator (occasional)
- 3. Acquisition PCB defect (rare)
Verify calibration pulses are correct
- Test point
- Monitor display
- Expected
- Cal pulse amplitude: Lead I 0.5 mV, Lead II 1 mV, Lead III 0.5 mV, Lead V -0.5 mV, +/-20%
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal lithium battery voltage low patient safety
Processor PCB internal lithium battery depleted; monitor powers up but one or more power-up tests fail; may be unable to maintain date/time
Ranked causes
- 1. Lithium battery depleted from age or excessive drain (common)
- 2. Battery switch S1 on processor PCB in OFF position (occasional)
- 3. Processor PCB defective or battery circuit failure (rare)
Allow boot sequence to complete power-up tests
- Test point
- Boot diagnostics screen
- Expected
- One or more power-up tests fail; message 'DO YOU WISH TO CONTINUE MONITORING ANYWAY?' appears
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor cannot be viewed at central station or view other monitors
Monitor is not communicating with other monitors or central station over Wireless LAN; no network connectivity visible.
Ranked causes
- 1. Monitor not communicating with internal RF card (common)
- 2. Wireless LAN RF card disabled or inoperative (occasional)
- 3. Incorrect country configuration code mismatch (occasional)
Check Wireless LAN communication status
- Test point
- Monitor Setup Menu REVISION AND ID
- Expected
- WIRELESS LAN ID displays with country, status, revision and date code
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Network communication present but no data transfer
802.11 wireless communication status is 02 (connected to access point) but network data transfer not occurring or intermittent
Ranked causes
- 1. Network infrastructure or wired network unavailable (common)
- 2. Monitor out of range or RF interference (occasional)
- 3. Wireless card failure (rare)
Verify association status in WIRELESS LAN STATUS screen
- Test point
- Service Mode, MONITOR SETTINGS, WIRELESS LAN STATUS, ASSOCIATION field
- Expected
- CONNECTED
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Network Connector: no network connection with cable connected
Network cable connected to monitor's network connector but no network communication established
Ranked causes
- 1. Dash connected to docking station; network connector inactive (common)
- 2. Network disabled in boot code (occasional)
- 3. Optional Unity Network option not enabled (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No display output
Monitor powers on but display screen remains completely dark with no image; backlight may or may not be active
Ranked causes
- 1. Display in standby mode (common)
- 2. Backlight inverter defective (occasional)
- 3. Display assembly failure (occasional)
Press Power button and wait for display activation
- Test point
- Display screen and backlight
- Expected
- Display image appears within 10 seconds of pressing Power button; backlight illuminates
- Tools
- Monitor power button and visual inspection
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Noninvasive blood pressure (NBP) functions inadequate after calibration caution
NBP calibration unsuccessful or cannot be properly performed; potential pneumatic circuit leaks suspected
Ranked causes
- 1. Leaks in NBP cuff or tubing (common)
- 2. Internal pneumatic circuit plumbing leaks (occasional)
- 3. NBP pump assembly faulty (occasional)
Inspect NBP cuff and tubing for visible leaks or cracks
- Test point
- NBP cuff and flexible tubing
- Expected
- No visible cracks, leaks, or deformations
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Noninvasive blood pressure calibration or checkout unsuccessful
NBP checkout procedure shows calibration failure or NBP functions not working as designed
Ranked causes
- 1. Leaks in pneumatic circuit plumbing or NBP cuff/tubing (common)
- 2. Internal pneumatic circuit leaks or connection problems (occasional)
- 3. NBP pump assembly defect (occasional)
Test NBP cuff and tubing on working monitor
- Test point
- Working monitor NBP connector
- Expected
- NBP functions and calibration complete successfully on working monitor
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer does not print
No output from external writer or printer device
Ranked causes
- 1. Writer or printer is turned OFF (common)
- 2. Cables not properly connected (common)
- 3. Incorrect alarm or print configuration (occasional)
Verify writer/printer power state
- Test point
- Power switch on writer/printer
- Expected
- ON
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Respiration waveform or rate does not appear on display
No respiration waveform or respiration rate displayed despite simulator connected
Ranked causes
- 1. Incorrect respiration simulator settings (common)
- 2. Faulty patient simulator or cables (common)
- 3. Acquisition PCB defect (occasional)
Verify respiration rate displayed and accurate
- Test point
- Monitor display RR parameter
- Expected
- Respiration rate displayed and matches simulator setting of 30 BPM
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Saving message displayed and missing graph segments caution
Message displays during graph saving; printed output shows gaps or missing segments in waveform data
Ranked causes
- 1. Writer is busy processing previous graph (common)
- 2. Print head dirty or defective (occasional)
- 3. Writer is no longer available or graph location incorrectly set (rare)
Verify writer not processing previous output
- Test point
- Writer mechanism and display
- Expected
- Writer quiet and idle between graph outputs; no mechanical noise or pending messages
- Tools
- Visual and audio observation
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery error or failure patient safety
CHECK BATT STATUS or BATTERY ERROR message displayed in ECG waveform area indicating minor or serious battery/charger failure
Ranked causes
- 1. Battery hardware failure (occasional)
- 2. Charger communications failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery error serious patient safety
BATTERY ERROR message in ECG waveform area, system WARNING alarm, serious battery failure or incompatible battery installed
Ranked causes
- 1. Serious battery failure (occasional)
- 2. Incompatible battery installed (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery failure message displayed
Display shows 'CHECK BATTERY STATUS' or 'BATTERY ERROR' message indicating battery has experienced failure during use or charging
Ranked causes
- 1. Minor battery failure during charge/use cycle (occasional)
- 2. Serious battery failure (occasional)
Observe battery status window for error indication
- Test point
- Battery Status Information Window
- Expected
- 'CHECK BATTERY STATUS' or 'BATTERY ERROR' message displayed
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery fuel gauge shows ERROR patient safety
ERROR message displayed in Battery Fuel Gauge icon, battery is asleep, faulty, or unrecognized
Ranked causes
- 1. Battery in sleep mode (occasional)
- 2. Battery faulty or unrecognized (occasional)
Check battery status screen in Service Menu
- Test point
- Service Menu battery status
- Expected
- Battery health status displayed
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery stuck in sleep mode or faulty battery
Battery fuel gauge icon displays 'ERROR'; battery does not respond to normal charge operations; battery may not have been used for extended period
Ranked causes
- 1. Battery in sleep mode from disuse (common)
- 2. Battery faulty and unrepairable (occasional)
Observe battery fuel gauge icon
- Test point
- Battery fuel gauge icon display
- Expected
- Battery icon shows normal charge level or 'ERROR'
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery will not wake up
Battery fuel gauge displays ERROR; battery is asleep or faulty after prolonged non-use
Ranked causes
- 1. Battery in sleep mode (common)
- 2. Battery faulty (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure waveforms display with noise or distortion
Invasive BP waveforms appear noisy, distorted, or fail to match expected clean display
Ranked causes
- 1. Faulty patient simulator or test cables (common)
- 2. Acquisition PCB defect (occasional)
Verify BP waveforms are noise-free
- Test point
- Monitor display for AR1 and PA2
- Expected
- Both AR1 and PA2 BP waveforms noise-free as shown in reference figure
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG waveforms display incorrectly patient safety
Displayed ECG waveforms contain significant noise; calibration pulses not correct
Ranked causes
- 1. ECG patient cables faulty (common)
- 2. Acquisition PCB faulty (occasional)
Measure calibration pulse amplitude
- Test point
- ECG display with 1.0 mV input
- Expected
- Lead I: 0.5 mV, Lead II: 1 mV, Lead III: 0.5 mV, Lead V: –0.5 mV, +/–20%
- Tools
- Patient simulator (MARQII), monitor display
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
EEPROM data invalid or uninitialized patient safety
Monitor detects EEPROM data corruption on startup; GE factory defaults will be loaded; network and site configuration lost
Ranked causes
- 1. EEPROM chip data corruption or initialization failure (rare)
- 2. Processor PCB defective or EEPROM chip failure (occasional)
Restore Ethernet and IP address configuration from Boot Code prompts
- Test point
- Boot Code configuration interface
- Expected
- Network address successfully entered and confirmed
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Intermittent wireless communication
Monitor exhibits intermittent communication over Wireless LAN
Ranked causes
- 1. Monitor out of range from Access Point (occasional)
- 2. RF interference from external sources (occasional)
Assess monitor location relative to Access Point range
- Test point
- Monitor physical location and Access Point coverage area
- Expected
- Monitor location within Access Point coverage; communication stable
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Intermittent Wireless LAN communication
Monitor loses network connectivity intermittently; communication drops in and out.
Ranked causes
- 1. Monitor out of range of Access Point (common)
- 2. RF interference from other equipment or sources (occasional)
Verify monitor is within Access Point range
- Test point
- Monitor location and Access Point proximity
- Expected
- Monitor within documented range of at least one Access Point
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal writer does not print
No output from internal writer mechanism
Ranked causes
- 1. No paper in internal writer (common)
- 2. Paper not loaded properly (common)
Verify paper is present in internal writer
- Test point
- Internal writer paper slot
- Expected
- Paper present
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invasive blood pressure waveforms display incorrectly patient safety
AR1 and/or PA2 BP waveforms appear noisy or distorted; BP parameter readings inaccurate
Ranked causes
- 1. Patient simulator or test cables faulty (common)
- 2. Acquisition PCB faulty (occasional)
Generate dynamic BP waveforms at WAVE setting
- Test point
- AR1 and PA2 waveform displays on monitor
- Expected
- Both waveforms noise-free as per reference figure; AR1: Systolic 116–124 mmHg, Diastolic 78–82 mmHg; PA2: Systolic 28–32 mmHg, Diastolic 9–11 mmHg
- Tools
- Patient simulator (MARQII), BP cuff, monitor display
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No network connection when cable connected
Network connectivity fails despite network cable being properly connected
Ranked causes
- 1. Monitor connected to docking station (common)
- 2. Unity Network option not enabled (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Optional Alarm Light Indicator: red or yellow lights not lit on boot caution
Red or yellow alarm indicator lights do not illuminate when monitor boots up
Ranked causes
- 1. Cable loose or disconnected (common)
- 2. LEDs burned out (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Optional Wireless Connection: intermittent or no waveform patient safety
Waveforms intermittently appear or do not display over wireless connection
Ranked causes
- 1. Network connection bad, wireless not enabled, or card not communicating (common)
- 2. Outside coverage area (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Optional Writer: paper output but no graph data caution
Paper comes out from writer but graph data is not printed on it
Ranked causes
- 1. Paper loaded incorrectly (common)
- 2. Print head dirty or defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Optional Writer: Saving Message displayed
Saving message appears, indicating writer is busy or no longer available
Ranked causes
- 1. Writer is busy (common)
- 2. Writer no longer available (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacemaker detection does not work properly patient safety
Pacemaker spikes not detected at correct amplitude or heart rate display inaccurate after VP1, VP2, or AVS input changes
Ranked causes
- 1. Patient simulator faulty (common)
- 2. Acquisition PCB faulty (occasional)
Set patient simulator to VP1 output and enable pace detect PACE 2
- Test point
- ECG display with pacemaker detection enabled
- Expected
- Heart rate approximately 80 bpm, pacemaker spikes detected
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacemaker detection not functioning patient safety
Pacemaker spikes not detected or displayed at correct amplitude; heart rate detection unreliable
Ranked causes
- 1. Faulty patient simulator (common)
- 2. Acquisition PCB defect (occasional)
Verify VP1 waveform at approximately 80 bpm
- Test point
- Monitor heart rate display
- Expected
- Heart rate approximately 80 bpm
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper output but no graph data shown
Paper comes out of writer but no graph data is visible on it
Ranked causes
- 1. Paper loaded incorrectly (common)
- 2. Printhead dirty or defective (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper output without graph data display caution
Paper advances and prints but no graph trace or waveform data appears on printed page
Ranked causes
- 1. Paper loaded incorrectly (common)
- 2. Print head dirty or defective (occasional)
Verify paper installation and feed
- Test point
- Paper path through writer assembly
- Expected
- Paper correctly threaded through roller and printhead area; paper advances smoothly
- Tools
- Visual inspection
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power cord continuity failure caution
Power cord does not provide continuous electrical path through line, neutral, or ground conductors, or conductors are short-circuited
Ranked causes
- 1. Power cord strain relief damaged or broken conductor (common)
- 2. Line, neutral, or ground conductor not properly connected to plug (occasional)
Test continuity through power cord conductors
- Test point
- Power cord plug terminals and connector
- Expected
- Continuous (no breaks)
- Tools
- DMM continuity function
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power cord damaged or intermittent high voltage
Physical damage to power cord, strain relief failure, or broken internal conductors causing intermittent power or no power
Ranked causes
- 1. Power cord strain relief failure (common)
- 2. Broken or loose conductor connections at plug (occasional)
Test continuity through each power cord conductor
- Test point
- Power cord plug pins to connector at monitor power input
- Expected
- Continuity present on all three conductors (line, neutral, ground)
- Tools
- DMM continuity mode or ohmmeter
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Real time clock failure at startup caution
Real time clock chip not running on boot; system cannot load Main Code until clock is started; may be unable to start clock
Ranked causes
- 1. Real time clock chip powered off or not responding to start command (occasional)
- 2. Real time clock chip defective; unable to start (rare)
Monitor boot sequence for real time clock start message
- Test point
- Boot diagnostics display
- Expected
- One of two outcomes: clock started successfully OR unable to start clock error
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red or yellow alarm light does not illuminate on boot up
Monitor boots normally but indicator LEDs (red or yellow) do not light during power-up sequence
Ranked causes
- 1. Alarm light cable loose or disconnected (common)
- 2. LEDs burned out (occasional)
Verify alarm light cable connections
- Test point
- Alarm light connector and cable
- Expected
- Cable fully seated and locked in connector; no loose, corroded, or missing pins
- Tools
- Visual inspection
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red or yellow alarm lights do not light on boot
Alarm light indicator LEDs do not illuminate during monitor startup
Ranked causes
- 1. Cable loose or disconnected (common)
- 2. LEDs burned out (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Respiration functions work on patient simulator but not on actual patient patient safety
Respiration waveform noisy or rate inaccurate when monitoring actual patient; functions work correctly with simulator
Ranked causes
- 1. Inadequate patient preparation for ECG electrode placement (common)
- 2. Patient has excessively high baseline chest impedance (rare)
Verify ECG electrode placement and patient skin preparation
- Test point
- Patient chest ECG electrode sites
- Expected
- Clean, dry skin; electrodes properly adhered; good electrical contact
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Respiration waveform functions correctly on simulator but noisy on actual patient patient safety
Clean respiration waveform on patient simulator but noisy or unreliable on patient
Ranked causes
- 1. Improper patient preparation or ECG electrode placement (common)
- 2. Excessively high baseline chest impedance in patient (occasional)
Verify clean respiration waveform versus noisy patient waveform
- Test point
- Comparison of simulator waveform to patient waveform on display
- Expected
- Patient waveform should be similarly clean to simulator waveform if proper preparation completed
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Saving message displayed during graph output
Saving message appears, preventing graph output
Ranked causes
- 1. Writer is busy (common)
- 2. Writer is no longer available (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not switch from battery to AC power
Unit plugged in but fails to transition from battery to AC power operation
Ranked causes
- 1. Power supply output voltage inadequate (common)
- 2. Processor/power management assembly failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not switch to AC from battery power
Unit is plugged in but does not switch to AC power from battery operation
Ranked causes
- 1. Power supply output voltage inadequate (common)
- 2. Processor/power management assembly failed (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Video Display: no display output caution
Display screen remains blank or does not turn on
Ranked causes
- 1. Display in standby mode (common)
- 2. Backlight inverter defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Waveform display not showing patient safety
Patient waveforms not displaying on monitor screen
Ranked causes
- 1. Patient cables are bad (common)
- 2. Acquisition PCB, DAS Assembly or power supply PCB failed (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wireless LAN communication loss or delayed response
Dash 3000/4000/5000 monitor exhibits loss of waveforms, delayed communication response, or no wireless connectivity
Ranked causes
- 1. Wireless feature not enabled (common)
- 2. Wireless card not present or driver not installed (occasional)
Identify wireless technology type
- Test point
- Monitor Setup Menu, REVISION AND ID, HARDWARE ID DISPLAY
- Expected
- WIRELESS LAN ID field visible with 802.11b or 802.11 technology
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
802.11 country code mismatch
802.11 wireless monitor configured for wrong country abbreviation code (international roaming), preventing communication with access point
Ranked causes
- 1. Country code incorrect for deployed location (rare)
Display WIRELESS LAN ID field and verify country code
- Test point
- Monitor Setup Menu REVISION AND ID, HARDWARE ID DISPLAY
- Expected
- Country code matches current location (e.g., US 02 for United States)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
802.11 SSID incorrect or not matching access point
802.11 wireless monitor communication status is 01 but not communicating with access point due to SSID mismatch
Ranked causes
- 1. SSID does not match access point NET ID (common)
Verify SSID configuration matches access point
- Test point
- Service Mode, MONITOR SETTINGS, CONFIGURE WIRELESS LAN, SSID field
- Expected
- GEMS or access point NET ID as configured
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Access Boot Loader service menu patient safety
Technician needs to enter Boot Loader menu for software downloads or serious equipment failures. Activated via front panel button sequence.
Ranked causes
- 1. Need to download Boot Code or Main Code software (common)
Hold NBP GO/STOP and ZERO ALL buttons
- Test point
- Front panel controls
- Expected
- Buttons held down without release
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Access Service Mode menu from Main Menu
- Alarm broadcast text in English when Japanese language selected
- Alarm pause duration three minutes with French language setting
- Bars or strips of pixels missing on display
- Battery condition – battery requesting conditioning cycle
- Battery condition – incompatible battery installed
- Battery failure minor
- Battery low or empty
- Battery low warning
- Battery low warning message displayed
- Boot code and main code version mismatch
- Boot Flash test failure
- Calibration: Analog output BP
- Calibration: Analog output ECG
- Calibration: Blood pressure analog output
- Calibration: CPU assembly
- Calibration: Display assembly
- Calibration: ECG analog output
- Calibration: NBP assembly
- Calibration: NBP software
- Calibration: Noninvasive blood pressure (NBP)
- Centralscope central station not visible to monitor on network
- Change Internet (IP) address via Boot Loader
- Charger failure message displayed
- Checkout: DAS assembly
- Checkout: Keypad and Trim Knob assembly
- Checkout: NBP assembly
- Checkout: Power supply assembly
- Checkout: Speaker assembly
- Checkout: Wireless LAN module
- Checkout: Writer and writer flex
- Continuous NBP alarms occur
- Ground-neutral loop resistance excessive
- Ground-to-neutral loop resistance out of tolerance
- Handle assembly removal and replacement
- Internal charger failed
- Lead fail detection does not function properly
- Lead fail detection not functioning
- Main Flash test failure
- Markers missing on respiration waveform; respiration rate inaccurate
- Masimo SET or Nellcor SPO2: probe or module malfunction
- Masimo SET SpO2 probe or module malfunction
- Missing pixels or bars on display in rows or columns
- Missing segments in graph data
- Monitor appears locked up during network download
- Monitor appears locked up during PC download
- Monitor powering down due to empty battery
- NBP alarms occurring continuously
- No paper output from optional writer
- No paper output from optional writer despite graphing display
- Optional Writer: missing segments in graph data
- Optional Writer: no paper output
- Remote control does not activate display
- Respiration rate inaccurate or markers missing from waveform
- RF LAN card communicating with Access Point but network unavailable
- RF LAN card not communicating with Access Point
- Run graph test pattern
- Set monitor bed number for network
- Set monitor Internet (IP) address for network
- Set monitor patient type configuration
- Set monitor unit name for network
- Software revision window does not list part numbers
- Special characters not displayed in network messages
- Static RAM test failure
- Text input restricted to English with Japanese language setting
- Unit powering down due to empty battery
- Video Display: missing rows or columns of pixels
- Waveforms do not appear at central station
- WIRELESS LAN STATUS screen shows abnormal values
69 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
400-4FF | Network errors detected | Network communication failure or configuration issue | error |
703 | Diagnostic tests were completed | Successful completion of system diagnostic test procedure | info |
70B | Internet address was changed from this patient monitor | Network address modified on this monitor | info |
70C | Video test was completed | Video display diagnostic test performed | info |
70E | Time was changed from this patient monitor | System time modified on this monitor | info |
70F | Date was changed from this patient monitor | System date modified on this monitor | info |
710 | Incompatible software was detected | Main processor software detected communication software incompatibility | error |
BATTERY ERROR | Serious battery failure | Serious battery failure during use or charging | error |
BATTERY LOW | Low battery warning | 10 minutes (single battery) or 20 minutes (dual battery) of run time remaining | warning |
Boot Code version incompatible with Main Code | VERSION OF BOOT CODE IS NOT COMPATIBLE WITH VERSION OF MAIN CODE CURRENTLY STORED IN FLASH | Boot Code version does not match Main Code version during system startup | error |
Boot Flash test FAILED | Boot Code stored in Flash memory is not valid; Main Code cannot be loaded | Flash memory self-test detects invalid Boot Code during startup | critical |
Boot Flash test FAILED / Boot Code invalid | Boot Code stored in Flash memory is invalid or corrupted; cannot load Main Code | Boot Flash memory test fails during system startup | error |
CHECK BATT STATUS | Battery or charger failure detected | Minor battery failure during use or charging, or charger communications failure | warning |
CONDITION | Battery conditioning cycle requested | Battery requires conditioning cycle | info |
EEPROM data INVALID or uninitialized | EEPROM memory contains invalid data or has not been initialized; GE factory defaults will be stored | System detects EEPROM data corruption or uninitialized state during boot | warning |
ERROR COPYING UNIT DEFAULTS - COUNTRY CODE MISMATCH | The server and client patient monitors are set to different country codes (i.e. DEFAULT or FRANCE). | Server and client monitors configured with different country codes | error |
ERROR COPYING UNIT DEFAULTS - DATA CORRUPTION ERROR | The patient monitor defaults transferred have become corrupted, possibly due to a memory error. | Data corruption during defaults transfer | error |
ERROR COPYING UNIT DEFAULTS - NETWORK ERROR | Either an error has occurred while the defaults were being transferred, or defaults have not been saved on the server patient monitor. | Error during defaults transfer or defaults not saved on server | error |
ERROR COPYING UNIT DEFAULTS - SOFTWARE FEATURE LEVEL MISMATC | The server and client patient monitors are set to different software levels (i.e. Basic, Cardiac, Cardiopulmonary). | Server and client monitors configured with different software feature levels | error |
ERROR COPYING UNIT DEFAULTS - SOFTWARE VERSION MISMATCH | The server and client patient monitors have different software versions (i.e. V2C, V3A). | Server and client monitors running different software versions | error |
ERROR COPYING UNIT DEFAULTS - UNIT TYPE MISMATCH | The server and client patient monitors are set to different unit types (i.e. ADULT, NEO, or OR) | Server and client monitors configured with different unit types | error |
ERROR: THE INTERNAL BATTERY THAT MAINTAINS THE MONITOR'S DEF | Internal backup battery for default settings has failed; monitor configuration at risk | Battery voltage drops below operating threshold during power-off retention test | critical |
ERROR - UNIT DEFAULTS HAVE NOT BEEN SAVED | Patient monitor defaults were not saved on the server patient monitor due to a memory error in the patient monitor. | Memory error in patient monitor during defaults save | error |
INTERNAL CHARGER FAILED, CALL SERVICE | Charger communications failure | Charger unable to communicate with monitor | error |
Internal lithium battery is LOW | Processor PCB backup battery voltage is low; may impact RTC and configuration retention | Battery voltage drops below operating threshold during power-up test | warning |
Main Flash test FAILED | Main Code stored in Flash is not valid; system cannot boot properly | Flash memory self-test detects invalid Main Code during startup | critical |
Main Flash test FAILED / Main Code invalid | Main Code stored in Flash memory is invalid or corrupted; cannot load | Main Flash memory test fails during system startup | error |
POWERING DOWN | Empty battery warning | No battery run time remaining | critical |
Real Time Clock FAILED | Real time clock chip is not running; Main Code cannot load until chip is started | Real time clock chip non-functional during boot sequence | error |
Real Time Clock FAILED - will not start | Real time clock chip not running; cannot load Main Code until started or confirmed defective | Real time clock not responding during boot initialization | warning |
Static RAM test FAILED | SRAM memory test failed; cannot load Main Code until test passes | Static RAM memory test fails during boot diagnostics | error |
WARNING: Boot Code incompatible with Main Code version | Version mismatch between Boot Code and Main Code in Flash memory | Boot Code version does not match Main Code version during boot sequence | warning |
WARNING: EEPROM data INVALID or uninitialized | EEPROM memory corrupted or not initialized; GE factory defaults loaded | Monitor detects invalid EEPROM data during boot initialization | warning |
WARNING: The EEPROM data was found to be either INVALID or u | EEPROM memory corrupted or not initialized; factory defaults will be loaded | EEPROM checksum or format error detected at startup | warning |
WARNING: THIS VERSION OF BOOT CODE IS NOT COMPATIBLE WITH TH | Boot Code and Main Code versions are incompatible; system cannot execute | Version mismatch detected during Boot Code initialization | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.