Patient monitors · Dash family

GE Healthcare Dash 3000: troubleshooting and error codes

The ServiceForge knowledge base holds 129 documented symptoms for the GE Healthcare Dash 3000, with 214 ranked causes, 178 diagnostic checks with test points and expected values, and 35 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

No waveform at central station

Waveform display missing at central station despite patient monitor active

Ranked causes

  1. 1. Central station software incompatible (occasional)
  2. 2. Cable connection loose or faulty (common)
  3. 3. Incorrect alarm or print location configuration (occasional)
  4. 4. Care unit name mismatch between devices (occasional)
  5. 5. IP addresses configured incorrectly (common)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Blood pressure waveforms do not appear on display

BP parameter labels, readings, and waveforms do not display; AR1 or PA2 not showing

Ranked causes

  1. 1. Faulty patient simulator or test cables (common)
  2. 2. Bent or broken pins on BP side panel connectors (occasional)
  3. 3. Broken continuity in BP signal path (occasional)
  4. 4. DAS assembly defect (occasional)
First check

Verify patient simulator and test cables on working monitor

Test point
Working monitor BP display
Expected
BP waveforms and parameters appear correctly

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG waveforms do not display at all

No ECG waveforms appear on monitor display despite simulator connected

Ranked causes

  1. 1. Faulty ECG patient cables (common)
  2. 2. Faulty patient simulator (common)
  3. 3. Acquisition PCB defect (occasional)
  4. 4. Processor PCB or power supply PCB defect (rare)
First check

Test ECG patient cables

Test point
Working monitor ECG connector
Expected
ECG signal appears on working monitor

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invasive blood pressure waveforms do not appear on display patient safety

AR1 or PA2 parameter labels, readings, and waveforms not displayed on monitor; BP side panel shows no signal

Ranked causes

  1. 1. Patient simulator or cables faulty (common)
  2. 2. BP side panel connector pins bent or broken (occasional)
  3. 3. Continuity fault in BP signal path (occasional)
  4. 4. DAS assembly faulty (occasional)
First check

Verify patient simulator and BP cables on working monitor

Test point
Working monitor with suspect simulator and cables
Expected
AR1 and PA2 waveforms and parameters display normally

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Review runtime error logs

Access and review software error logs stored in monitor for troubleshooting. View input/output errors with timestamps and parameters.

Ranked causes

  1. 1. Advanced troubleshooting required (occasional)
  2. 2. View output software errors (occasional)
  3. 3. View input software errors (occasional)
  4. 4. Clear error logs (occasional)
First check

Navigate to REVIEW ERRORS in Service Mode

Test point
Service Mode Menu display
Expected
REVIEW ERRORS option visible and selectable

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

802.11b association status incorrect (not 01)

Wireless LAN ID field shows association status other than 01, indicating monitor not connected to network or incorrect credentials

Ranked causes

  1. 1. SSID/PASS KEY incorrect (common)
  2. 2. Wireless card firmware requires reset (occasional)
  3. 3. Monitor not connected to network (occasional)
First check

Display WIRELESS LAN ID field

Test point
Monitor Setup Menu, REVISION AND ID, HARDWARE ID DISPLAY
Expected
WIRELESS LAN ID field displays status code (e.g., 01 G3.93-73e 55 04)

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

802.11b signal quality below threshold

WIRELESS LAN ID field shows signal quality status less than 20, indicating weak wireless signal or network coverage issue

Ranked causes

  1. 1. Monitor out of range from access point (common)
  2. 2. RF interference from nearby equipment (occasional)
  3. 3. Access point radio issue (occasional)
First check

Verify signal quality status value

Test point
WIRELESS LAN ID field third field value
Expected
greater than 20

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm audio does not sound caution

No audio output when alarm is triggered or during audio test

Ranked causes

  1. 1. Speaker cable loose or disconnected (common)
  2. 2. Speaker failed (occasional)
  3. 3. Audio circuit on processor PCB failed (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Buzz signal during ECG or BP patient safety

Unwanted buzz signal output during ECG or blood pressure measurement

Ranked causes

  1. 1. BP not zeroed (common)
  2. 2. BP not available (common)
  3. 3. ECG leads failed (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Buzz signal output instead of tone for ECG or BP alarms patient safety

Defib sync monitor outputs a buzz or buzzer tone instead of normal alarm tone for ECG or blood pressure measurements

Ranked causes

  1. 1. Blood pressure not zeroed (common)
  2. 2. Blood pressure signal not available or disconnected (occasional)
  3. 3. ECG leads disconnected from patient or monitor (occasional)
First check

Check BP zeroing status

Test point
Blood pressure module and monitor display
Expected
BP reference zeroed and displayed as baseline value
Tools
Monitor interface and BP module controls

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Defib Sync: Buzz signal for ECG or BP patient safety

Incorrect buzz signal output during defibrillator sync when ECG or BP is active

Ranked causes

  1. 1. BP not zeroed (common)
  2. 2. BP not available (occasional)
  3. 3. ECG leads fail (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG waveforms display incorrectly with noise

Displayed ECG waveforms contain significant amount of noise instead of clean signal

Ranked causes

  1. 1. Faulty ECG patient cables (common)
  2. 2. Faulty patient simulator (occasional)
  3. 3. Acquisition PCB defect (rare)
First check

Verify calibration pulses are correct

Test point
Monitor display
Expected
Cal pulse amplitude: Lead I 0.5 mV, Lead II 1 mV, Lead III 0.5 mV, Lead V -0.5 mV, +/-20%

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Internal lithium battery voltage low patient safety

Processor PCB internal lithium battery depleted; monitor powers up but one or more power-up tests fail; may be unable to maintain date/time

Ranked causes

  1. 1. Lithium battery depleted from age or excessive drain (common)
  2. 2. Battery switch S1 on processor PCB in OFF position (occasional)
  3. 3. Processor PCB defective or battery circuit failure (rare)
First check

Allow boot sequence to complete power-up tests

Test point
Boot diagnostics screen
Expected
One or more power-up tests fail; message 'DO YOU WISH TO CONTINUE MONITORING ANYWAY?' appears

1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor cannot be viewed at central station or view other monitors

Monitor is not communicating with other monitors or central station over Wireless LAN; no network connectivity visible.

Ranked causes

  1. 1. Monitor not communicating with internal RF card (common)
  2. 2. Wireless LAN RF card disabled or inoperative (occasional)
  3. 3. Incorrect country configuration code mismatch (occasional)
First check

Check Wireless LAN communication status

Test point
Monitor Setup Menu REVISION AND ID
Expected
WIRELESS LAN ID displays with country, status, revision and date code

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Network communication present but no data transfer

802.11 wireless communication status is 02 (connected to access point) but network data transfer not occurring or intermittent

Ranked causes

  1. 1. Network infrastructure or wired network unavailable (common)
  2. 2. Monitor out of range or RF interference (occasional)
  3. 3. Wireless card failure (rare)
First check

Verify association status in WIRELESS LAN STATUS screen

Test point
Service Mode, MONITOR SETTINGS, WIRELESS LAN STATUS, ASSOCIATION field
Expected
CONNECTED

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Network Connector: no network connection with cable connected

Network cable connected to monitor's network connector but no network communication established

Ranked causes

  1. 1. Dash connected to docking station; network connector inactive (common)
  2. 2. Network disabled in boot code (occasional)
  3. 3. Optional Unity Network option not enabled (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No display output

Monitor powers on but display screen remains completely dark with no image; backlight may or may not be active

Ranked causes

  1. 1. Display in standby mode (common)
  2. 2. Backlight inverter defective (occasional)
  3. 3. Display assembly failure (occasional)
First check

Press Power button and wait for display activation

Test point
Display screen and backlight
Expected
Display image appears within 10 seconds of pressing Power button; backlight illuminates
Tools
Monitor power button and visual inspection

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Noninvasive blood pressure (NBP) functions inadequate after calibration caution

NBP calibration unsuccessful or cannot be properly performed; potential pneumatic circuit leaks suspected

Ranked causes

  1. 1. Leaks in NBP cuff or tubing (common)
  2. 2. Internal pneumatic circuit plumbing leaks (occasional)
  3. 3. NBP pump assembly faulty (occasional)
First check

Inspect NBP cuff and tubing for visible leaks or cracks

Test point
NBP cuff and flexible tubing
Expected
No visible cracks, leaks, or deformations

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Noninvasive blood pressure calibration or checkout unsuccessful

NBP checkout procedure shows calibration failure or NBP functions not working as designed

Ranked causes

  1. 1. Leaks in pneumatic circuit plumbing or NBP cuff/tubing (common)
  2. 2. Internal pneumatic circuit leaks or connection problems (occasional)
  3. 3. NBP pump assembly defect (occasional)
First check

Test NBP cuff and tubing on working monitor

Test point
Working monitor NBP connector
Expected
NBP functions and calibration complete successfully on working monitor

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer does not print

No output from external writer or printer device

Ranked causes

  1. 1. Writer or printer is turned OFF (common)
  2. 2. Cables not properly connected (common)
  3. 3. Incorrect alarm or print configuration (occasional)
First check

Verify writer/printer power state

Test point
Power switch on writer/printer
Expected
ON

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Respiration waveform or rate does not appear on display

No respiration waveform or respiration rate displayed despite simulator connected

Ranked causes

  1. 1. Incorrect respiration simulator settings (common)
  2. 2. Faulty patient simulator or cables (common)
  3. 3. Acquisition PCB defect (occasional)
First check

Verify respiration rate displayed and accurate

Test point
Monitor display RR parameter
Expected
Respiration rate displayed and matches simulator setting of 30 BPM

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Saving message displayed and missing graph segments caution

Message displays during graph saving; printed output shows gaps or missing segments in waveform data

Ranked causes

  1. 1. Writer is busy processing previous graph (common)
  2. 2. Print head dirty or defective (occasional)
  3. 3. Writer is no longer available or graph location incorrectly set (rare)
First check

Verify writer not processing previous output

Test point
Writer mechanism and display
Expected
Writer quiet and idle between graph outputs; no mechanical noise or pending messages
Tools
Visual and audio observation

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery error or failure patient safety

CHECK BATT STATUS or BATTERY ERROR message displayed in ECG waveform area indicating minor or serious battery/charger failure

Ranked causes

  1. 1. Battery hardware failure (occasional)
  2. 2. Charger communications failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery error serious patient safety

BATTERY ERROR message in ECG waveform area, system WARNING alarm, serious battery failure or incompatible battery installed

Ranked causes

  1. 1. Serious battery failure (occasional)
  2. 2. Incompatible battery installed (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery failure message displayed

Display shows 'CHECK BATTERY STATUS' or 'BATTERY ERROR' message indicating battery has experienced failure during use or charging

Ranked causes

  1. 1. Minor battery failure during charge/use cycle (occasional)
  2. 2. Serious battery failure (occasional)
First check

Observe battery status window for error indication

Test point
Battery Status Information Window
Expected
'CHECK BATTERY STATUS' or 'BATTERY ERROR' message displayed

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery fuel gauge shows ERROR patient safety

ERROR message displayed in Battery Fuel Gauge icon, battery is asleep, faulty, or unrecognized

Ranked causes

  1. 1. Battery in sleep mode (occasional)
  2. 2. Battery faulty or unrecognized (occasional)
First check

Check battery status screen in Service Menu

Test point
Service Menu battery status
Expected
Battery health status displayed

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery stuck in sleep mode or faulty battery

Battery fuel gauge icon displays 'ERROR'; battery does not respond to normal charge operations; battery may not have been used for extended period

Ranked causes

  1. 1. Battery in sleep mode from disuse (common)
  2. 2. Battery faulty and unrepairable (occasional)
First check

Observe battery fuel gauge icon

Test point
Battery fuel gauge icon display
Expected
Battery icon shows normal charge level or 'ERROR'

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery will not wake up

Battery fuel gauge displays ERROR; battery is asleep or faulty after prolonged non-use

Ranked causes

  1. 1. Battery in sleep mode (common)
  2. 2. Battery faulty (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Blood pressure waveforms display with noise or distortion

Invasive BP waveforms appear noisy, distorted, or fail to match expected clean display

Ranked causes

  1. 1. Faulty patient simulator or test cables (common)
  2. 2. Acquisition PCB defect (occasional)
First check

Verify BP waveforms are noise-free

Test point
Monitor display for AR1 and PA2
Expected
Both AR1 and PA2 BP waveforms noise-free as shown in reference figure

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG waveforms display incorrectly patient safety

Displayed ECG waveforms contain significant noise; calibration pulses not correct

Ranked causes

  1. 1. ECG patient cables faulty (common)
  2. 2. Acquisition PCB faulty (occasional)
First check

Measure calibration pulse amplitude

Test point
ECG display with 1.0 mV input
Expected
Lead I: 0.5 mV, Lead II: 1 mV, Lead III: 0.5 mV, Lead V: –0.5 mV, +/–20%
Tools
Patient simulator (MARQII), monitor display

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

EEPROM data invalid or uninitialized patient safety

Monitor detects EEPROM data corruption on startup; GE factory defaults will be loaded; network and site configuration lost

Ranked causes

  1. 1. EEPROM chip data corruption or initialization failure (rare)
  2. 2. Processor PCB defective or EEPROM chip failure (occasional)
First check

Restore Ethernet and IP address configuration from Boot Code prompts

Test point
Boot Code configuration interface
Expected
Network address successfully entered and confirmed

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Intermittent wireless communication

Monitor exhibits intermittent communication over Wireless LAN

Ranked causes

  1. 1. Monitor out of range from Access Point (occasional)
  2. 2. RF interference from external sources (occasional)
First check

Assess monitor location relative to Access Point range

Test point
Monitor physical location and Access Point coverage area
Expected
Monitor location within Access Point coverage; communication stable

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Intermittent Wireless LAN communication

Monitor loses network connectivity intermittently; communication drops in and out.

Ranked causes

  1. 1. Monitor out of range of Access Point (common)
  2. 2. RF interference from other equipment or sources (occasional)
First check

Verify monitor is within Access Point range

Test point
Monitor location and Access Point proximity
Expected
Monitor within documented range of at least one Access Point

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Internal writer does not print

No output from internal writer mechanism

Ranked causes

  1. 1. No paper in internal writer (common)
  2. 2. Paper not loaded properly (common)
First check

Verify paper is present in internal writer

Test point
Internal writer paper slot
Expected
Paper present

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invasive blood pressure waveforms display incorrectly patient safety

AR1 and/or PA2 BP waveforms appear noisy or distorted; BP parameter readings inaccurate

Ranked causes

  1. 1. Patient simulator or test cables faulty (common)
  2. 2. Acquisition PCB faulty (occasional)
First check

Generate dynamic BP waveforms at WAVE setting

Test point
AR1 and PA2 waveform displays on monitor
Expected
Both waveforms noise-free as per reference figure; AR1: Systolic 116–124 mmHg, Diastolic 78–82 mmHg; PA2: Systolic 28–32 mmHg, Diastolic 9–11 mmHg
Tools
Patient simulator (MARQII), BP cuff, monitor display

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No network connection when cable connected

Network connectivity fails despite network cable being properly connected

Ranked causes

  1. 1. Monitor connected to docking station (common)
  2. 2. Unity Network option not enabled (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Optional Alarm Light Indicator: red or yellow lights not lit on boot caution

Red or yellow alarm indicator lights do not illuminate when monitor boots up

Ranked causes

  1. 1. Cable loose or disconnected (common)
  2. 2. LEDs burned out (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Optional Wireless Connection: intermittent or no waveform patient safety

Waveforms intermittently appear or do not display over wireless connection

Ranked causes

  1. 1. Network connection bad, wireless not enabled, or card not communicating (common)
  2. 2. Outside coverage area (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Optional Writer: paper output but no graph data caution

Paper comes out from writer but graph data is not printed on it

Ranked causes

  1. 1. Paper loaded incorrectly (common)
  2. 2. Print head dirty or defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Optional Writer: Saving Message displayed

Saving message appears, indicating writer is busy or no longer available

Ranked causes

  1. 1. Writer is busy (common)
  2. 2. Writer no longer available (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pacemaker detection does not work properly patient safety

Pacemaker spikes not detected at correct amplitude or heart rate display inaccurate after VP1, VP2, or AVS input changes

Ranked causes

  1. 1. Patient simulator faulty (common)
  2. 2. Acquisition PCB faulty (occasional)
First check

Set patient simulator to VP1 output and enable pace detect PACE 2

Test point
ECG display with pacemaker detection enabled
Expected
Heart rate approximately 80 bpm, pacemaker spikes detected

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pacemaker detection not functioning patient safety

Pacemaker spikes not detected or displayed at correct amplitude; heart rate detection unreliable

Ranked causes

  1. 1. Faulty patient simulator (common)
  2. 2. Acquisition PCB defect (occasional)
First check

Verify VP1 waveform at approximately 80 bpm

Test point
Monitor heart rate display
Expected
Heart rate approximately 80 bpm

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Paper output but no graph data shown

Paper comes out of writer but no graph data is visible on it

Ranked causes

  1. 1. Paper loaded incorrectly (common)
  2. 2. Printhead dirty or defective (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Paper output without graph data display caution

Paper advances and prints but no graph trace or waveform data appears on printed page

Ranked causes

  1. 1. Paper loaded incorrectly (common)
  2. 2. Print head dirty or defective (occasional)
First check

Verify paper installation and feed

Test point
Paper path through writer assembly
Expected
Paper correctly threaded through roller and printhead area; paper advances smoothly
Tools
Visual inspection

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power cord continuity failure caution

Power cord does not provide continuous electrical path through line, neutral, or ground conductors, or conductors are short-circuited

Ranked causes

  1. 1. Power cord strain relief damaged or broken conductor (common)
  2. 2. Line, neutral, or ground conductor not properly connected to plug (occasional)
First check

Test continuity through power cord conductors

Test point
Power cord plug terminals and connector
Expected
Continuous (no breaks)
Tools
DMM continuity function

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power cord damaged or intermittent high voltage

Physical damage to power cord, strain relief failure, or broken internal conductors causing intermittent power or no power

Ranked causes

  1. 1. Power cord strain relief failure (common)
  2. 2. Broken or loose conductor connections at plug (occasional)
First check

Test continuity through each power cord conductor

Test point
Power cord plug pins to connector at monitor power input
Expected
Continuity present on all three conductors (line, neutral, ground)
Tools
DMM continuity mode or ohmmeter

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Real time clock failure at startup caution

Real time clock chip not running on boot; system cannot load Main Code until clock is started; may be unable to start clock

Ranked causes

  1. 1. Real time clock chip powered off or not responding to start command (occasional)
  2. 2. Real time clock chip defective; unable to start (rare)
First check

Monitor boot sequence for real time clock start message

Test point
Boot diagnostics display
Expected
One of two outcomes: clock started successfully OR unable to start clock error

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Red or yellow alarm light does not illuminate on boot up

Monitor boots normally but indicator LEDs (red or yellow) do not light during power-up sequence

Ranked causes

  1. 1. Alarm light cable loose or disconnected (common)
  2. 2. LEDs burned out (occasional)
First check

Verify alarm light cable connections

Test point
Alarm light connector and cable
Expected
Cable fully seated and locked in connector; no loose, corroded, or missing pins
Tools
Visual inspection

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Red or yellow alarm lights do not light on boot

Alarm light indicator LEDs do not illuminate during monitor startup

Ranked causes

  1. 1. Cable loose or disconnected (common)
  2. 2. LEDs burned out (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Respiration functions work on patient simulator but not on actual patient patient safety

Respiration waveform noisy or rate inaccurate when monitoring actual patient; functions work correctly with simulator

Ranked causes

  1. 1. Inadequate patient preparation for ECG electrode placement (common)
  2. 2. Patient has excessively high baseline chest impedance (rare)
First check

Verify ECG electrode placement and patient skin preparation

Test point
Patient chest ECG electrode sites
Expected
Clean, dry skin; electrodes properly adhered; good electrical contact

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Respiration waveform functions correctly on simulator but noisy on actual patient patient safety

Clean respiration waveform on patient simulator but noisy or unreliable on patient

Ranked causes

  1. 1. Improper patient preparation or ECG electrode placement (common)
  2. 2. Excessively high baseline chest impedance in patient (occasional)
First check

Verify clean respiration waveform versus noisy patient waveform

Test point
Comparison of simulator waveform to patient waveform on display
Expected
Patient waveform should be similarly clean to simulator waveform if proper preparation completed

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Saving message displayed during graph output

Saving message appears, preventing graph output

Ranked causes

  1. 1. Writer is busy (common)
  2. 2. Writer is no longer available (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not switch from battery to AC power

Unit plugged in but fails to transition from battery to AC power operation

Ranked causes

  1. 1. Power supply output voltage inadequate (common)
  2. 2. Processor/power management assembly failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not switch to AC from battery power

Unit is plugged in but does not switch to AC power from battery operation

Ranked causes

  1. 1. Power supply output voltage inadequate (common)
  2. 2. Processor/power management assembly failed (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Video Display: no display output caution

Display screen remains blank or does not turn on

Ranked causes

  1. 1. Display in standby mode (common)
  2. 2. Backlight inverter defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Waveform display not showing patient safety

Patient waveforms not displaying on monitor screen

Ranked causes

  1. 1. Patient cables are bad (common)
  2. 2. Acquisition PCB, DAS Assembly or power supply PCB failed (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Wireless LAN communication loss or delayed response

Dash 3000/4000/5000 monitor exhibits loss of waveforms, delayed communication response, or no wireless connectivity

Ranked causes

  1. 1. Wireless feature not enabled (common)
  2. 2. Wireless card not present or driver not installed (occasional)
First check

Identify wireless technology type

Test point
Monitor Setup Menu, REVISION AND ID, HARDWARE ID DISPLAY
Expected
WIRELESS LAN ID field visible with 802.11b or 802.11 technology

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

802.11 country code mismatch

802.11 wireless monitor configured for wrong country abbreviation code (international roaming), preventing communication with access point

Ranked causes

  1. 1. Country code incorrect for deployed location (rare)
First check

Display WIRELESS LAN ID field and verify country code

Test point
Monitor Setup Menu REVISION AND ID, HARDWARE ID DISPLAY
Expected
Country code matches current location (e.g., US 02 for United States)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

802.11 SSID incorrect or not matching access point

802.11 wireless monitor communication status is 01 but not communicating with access point due to SSID mismatch

Ranked causes

  1. 1. SSID does not match access point NET ID (common)
First check

Verify SSID configuration matches access point

Test point
Service Mode, MONITOR SETTINGS, CONFIGURE WIRELESS LAN, SSID field
Expected
GEMS or access point NET ID as configured

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Access Boot Loader service menu patient safety

Technician needs to enter Boot Loader menu for software downloads or serious equipment failures. Activated via front panel button sequence.

Ranked causes

  1. 1. Need to download Boot Code or Main Code software (common)
First check

Hold NBP GO/STOP and ZERO ALL buttons

Test point
Front panel controls
Expected
Buttons held down without release

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Access Service Mode menu from Main Menu
  • Alarm broadcast text in English when Japanese language selected
  • Alarm pause duration three minutes with French language setting
  • Bars or strips of pixels missing on display
  • Battery condition – battery requesting conditioning cycle
  • Battery condition – incompatible battery installed
  • Battery failure minor
  • Battery low or empty
  • Battery low warning
  • Battery low warning message displayed
  • Boot code and main code version mismatch
  • Boot Flash test failure
  • Calibration: Analog output BP
  • Calibration: Analog output ECG
  • Calibration: Blood pressure analog output
  • Calibration: CPU assembly
  • Calibration: Display assembly
  • Calibration: ECG analog output
  • Calibration: NBP assembly
  • Calibration: NBP software
  • Calibration: Noninvasive blood pressure (NBP)
  • Centralscope central station not visible to monitor on network
  • Change Internet (IP) address via Boot Loader
  • Charger failure message displayed
  • Checkout: DAS assembly
  • Checkout: Keypad and Trim Knob assembly
  • Checkout: NBP assembly
  • Checkout: Power supply assembly
  • Checkout: Speaker assembly
  • Checkout: Wireless LAN module
  • Checkout: Writer and writer flex
  • Continuous NBP alarms occur
  • Ground-neutral loop resistance excessive
  • Ground-to-neutral loop resistance out of tolerance
  • Handle assembly removal and replacement
  • Internal charger failed
  • Lead fail detection does not function properly
  • Lead fail detection not functioning
  • Main Flash test failure
  • Markers missing on respiration waveform; respiration rate inaccurate
  • Masimo SET or Nellcor SPO2: probe or module malfunction
  • Masimo SET SpO2 probe or module malfunction
  • Missing pixels or bars on display in rows or columns
  • Missing segments in graph data
  • Monitor appears locked up during network download
  • Monitor appears locked up during PC download
  • Monitor powering down due to empty battery
  • NBP alarms occurring continuously
  • No paper output from optional writer
  • No paper output from optional writer despite graphing display
  • Optional Writer: missing segments in graph data
  • Optional Writer: no paper output
  • Remote control does not activate display
  • Respiration rate inaccurate or markers missing from waveform
  • RF LAN card communicating with Access Point but network unavailable
  • RF LAN card not communicating with Access Point
  • Run graph test pattern
  • Set monitor bed number for network
  • Set monitor Internet (IP) address for network
  • Set monitor patient type configuration
  • Set monitor unit name for network
  • Software revision window does not list part numbers
  • Special characters not displayed in network messages
  • Static RAM test failure
  • Text input restricted to English with Japanese language setting
  • Unit powering down due to empty battery
  • Video Display: missing rows or columns of pixels
  • Waveforms do not appear at central station
  • WIRELESS LAN STATUS screen shows abnormal values

69 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
400-4FFNetwork errors detectedNetwork communication failure or configuration issueerror
703Diagnostic tests were completedSuccessful completion of system diagnostic test procedureinfo
70BInternet address was changed from this patient monitorNetwork address modified on this monitorinfo
70CVideo test was completedVideo display diagnostic test performedinfo
70ETime was changed from this patient monitorSystem time modified on this monitorinfo
70FDate was changed from this patient monitorSystem date modified on this monitorinfo
710Incompatible software was detectedMain processor software detected communication software incompatibilityerror
BATTERY ERRORSerious battery failureSerious battery failure during use or chargingerror
BATTERY LOWLow battery warning10 minutes (single battery) or 20 minutes (dual battery) of run time remainingwarning
Boot Code version incompatible with Main CodeVERSION OF BOOT CODE IS NOT COMPATIBLE WITH VERSION OF MAIN CODE CURRENTLY STORED IN FLASHBoot Code version does not match Main Code version during system startuperror
Boot Flash test FAILEDBoot Code stored in Flash memory is not valid; Main Code cannot be loadedFlash memory self-test detects invalid Boot Code during startupcritical
Boot Flash test FAILED / Boot Code invalidBoot Code stored in Flash memory is invalid or corrupted; cannot load Main CodeBoot Flash memory test fails during system startuperror
CHECK BATT STATUSBattery or charger failure detectedMinor battery failure during use or charging, or charger communications failurewarning
CONDITIONBattery conditioning cycle requestedBattery requires conditioning cycleinfo
EEPROM data INVALID or uninitializedEEPROM memory contains invalid data or has not been initialized; GE factory defaults will be storedSystem detects EEPROM data corruption or uninitialized state during bootwarning
ERROR COPYING UNIT DEFAULTS - COUNTRY CODE MISMATCHThe server and client patient monitors are set to different country codes (i.e. DEFAULT or FRANCE).Server and client monitors configured with different country codeserror
ERROR COPYING UNIT DEFAULTS - DATA CORRUPTION ERRORThe patient monitor defaults transferred have become corrupted, possibly due to a memory error.Data corruption during defaults transfererror
ERROR COPYING UNIT DEFAULTS - NETWORK ERROREither an error has occurred while the defaults were being transferred, or defaults have not been saved on the server patient monitor.Error during defaults transfer or defaults not saved on servererror
ERROR COPYING UNIT DEFAULTS - SOFTWARE FEATURE LEVEL MISMATCThe server and client patient monitors are set to different software levels (i.e. Basic, Cardiac, Cardiopulmonary).Server and client monitors configured with different software feature levelserror
ERROR COPYING UNIT DEFAULTS - SOFTWARE VERSION MISMATCHThe server and client patient monitors have different software versions (i.e. V2C, V3A).Server and client monitors running different software versionserror
ERROR COPYING UNIT DEFAULTS - UNIT TYPE MISMATCHThe server and client patient monitors are set to different unit types (i.e. ADULT, NEO, or OR)Server and client monitors configured with different unit typeserror
ERROR: THE INTERNAL BATTERY THAT MAINTAINS THE MONITOR'S DEFInternal backup battery for default settings has failed; monitor configuration at riskBattery voltage drops below operating threshold during power-off retention testcritical
ERROR - UNIT DEFAULTS HAVE NOT BEEN SAVEDPatient monitor defaults were not saved on the server patient monitor due to a memory error in the patient monitor.Memory error in patient monitor during defaults saveerror
INTERNAL CHARGER FAILED, CALL SERVICECharger communications failureCharger unable to communicate with monitorerror
Internal lithium battery is LOWProcessor PCB backup battery voltage is low; may impact RTC and configuration retentionBattery voltage drops below operating threshold during power-up testwarning
Main Flash test FAILEDMain Code stored in Flash is not valid; system cannot boot properlyFlash memory self-test detects invalid Main Code during startupcritical
Main Flash test FAILED / Main Code invalidMain Code stored in Flash memory is invalid or corrupted; cannot loadMain Flash memory test fails during system startuperror
POWERING DOWNEmpty battery warningNo battery run time remainingcritical
Real Time Clock FAILEDReal time clock chip is not running; Main Code cannot load until chip is startedReal time clock chip non-functional during boot sequenceerror
Real Time Clock FAILED - will not startReal time clock chip not running; cannot load Main Code until started or confirmed defectiveReal time clock not responding during boot initializationwarning
Static RAM test FAILEDSRAM memory test failed; cannot load Main Code until test passesStatic RAM memory test fails during boot diagnosticserror
WARNING: Boot Code incompatible with Main Code versionVersion mismatch between Boot Code and Main Code in Flash memoryBoot Code version does not match Main Code version during boot sequencewarning
WARNING: EEPROM data INVALID or uninitializedEEPROM memory corrupted or not initialized; GE factory defaults loadedMonitor detects invalid EEPROM data during boot initializationwarning
WARNING: The EEPROM data was found to be either INVALID or uEEPROM memory corrupted or not initialized; factory defaults will be loadedEEPROM checksum or format error detected at startupwarning
WARNING: THIS VERSION OF BOOT CODE IS NOT COMPATIBLE WITH THBoot Code and Main Code versions are incompatible; system cannot executeVersion mismatch detected during Boot Code initializationwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI