GE Healthcare CT ProSpeed CT5: troubleshooting and error codes
The ServiceForge knowledge base holds 158 documented symptoms for the GE Healthcare CT ProSpeed CT5, with 215 ranked causes, 548 diagnostic checks with test points and expected values, and 21 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: Software loading and system recovery
Post-ATINST procedure to restore communication data, patient image data, raw data files, and user settings for CBF and Functional Image options, followed by software version updates (Qj, Qj-Base, Qj-Hlt, Zj-Adv)
Ranked causes
- 1. System software upgrade required (common)
- 2. Communication data restoration required for ELINK option (occasional)
- 3. Patient image data restoration required (occasional)
- 4. Raw data files restoration required (occasional)
- 5. CBF user setting parameters restoration required (occasional)
- 6. Functional Image user setting parameters restoration required (occasional)
- 7. Quick Filming Delay Time parameter restoration required (occasional)
Verify system starts up after software load
- Test point
- CRT monitor display
- Expected
- VER. 06.02 and logo appear on screen (for Qj v6.00 load)
2 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Custom parameters configuration caution
Configure custom parameters including Recon Kernel Type, Collimator Type, Network Option, and Tube selection
Ranked causes
- 1. Custom parameters require manual restoration due to software bug (common)
- 2. Recon Kernel Type must match detector type (common)
- 3. Collimator Type parameter must be set correctly (common)
- 4. Network Option parameter must match installed hardware (common)
- 5. Tube parameter selection must match installed tube (common)
Verify Recon Kernel Type selection
- Test point
- CHANGE PARAMETERS menu
- Expected
- Normal for Xe Detector; Normal or Enhanced for Hilight Detector; parameter absent for Version 6.xx/4.xx with Xe
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gantry panel switch verification - table and movement functions
Verify that table up/down, height display, and in/out movement functions operate correctly via gantry panel switches and foot switches.
Ranked causes
- 1. Table and movement controls require functional verification (common)
- 2. Table in/out and fast movement require verification (common)
- 3. Table scan position and zero function require verification (common)
- 4. Gantry tilt range function requires verification (common)
Tilt gantry to 0 degrees
- Test point
- Gantry tilt control
- Expected
- Gantry at 0 deg position
7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image performance out of specification
Measured image performance values exceed acceptable limits during periodic maintenance check
Ranked causes
- 1. KV output incorrect (unknown)
- 2. mA output incorrect (unknown)
- 3. Detector or system calibration drift (unknown)
- 4. CT number accuracy incorrect (unknown)
Perform Image Performance Check
- Test point
- SYSTEM tab
- Expected
- Within specification
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Image performance radiation
Image performance check shows measured values out of specification
Ranked causes
- 1. KV output incorrect (unknown)
- 2. mA output incorrect (unknown)
- 3. CT number calibration drift (unknown)
Perform Image Performance Check
- Test point
- Per SYSTEM tab FUNCTIONAL CHECK / ADJUSTMENT Manual
- Expected
- Within specification
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Image performance verification radiation
Verify CT image quality and noise levels across various scan parameters (kV, mA, FOV, slice thickness) using water and QA phantoms
Ranked causes
- 1. Image calibration and quality assurance required (common)
- 2. Image performance check at 120 kV, 2 sec, 350 mA, large focus, 5mm slice (common)
- 3. Image performance check at 120 kV, 2 sec, 350 mA, large focus, 3mm slice (common)
Scan 25cm water phantom and measure mean CT number (MC)
- Test point
- 25cm water phantom
- Expected
- MC 0 ± 2 HU, ±2
- Tools
- CT scanner, water phantom
10 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Software Loading (ATINST)
Load system software and restore parameters to CT ProSpeed system. Procedure involves boot switch settings, ATINST program execution, and post-installation steps.
Ranked causes
- 1. Software installation and parameter restoration required (common)
- 2. ATINST program error during execution (occasional)
- 3. Retry ATINST after failure (occasional)
Verify SW2 and SW4 BOOT switches on FMP1 Board set to ON
- Test point
- FMP1 Board Slot #30
- Expected
- SW2 ON, SW4 ON
7 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gantry rotation control unavailable or abnormal patient safety
Gantry does not respond to rotation commands or behaves abnormally during offline control via service switches
Ranked causes
- 1. SYS/OFF/MNL switch set to OFF during gantry rotation (occasional)
- 2. Gantry rotated by hand causing abnormal behavior (occasional)
- 3. Repeated rapid acceleration/deceleration causing servo overload (occasional)
Verify SYS/OFF/MNL switch position
- Test point
- TGP Board SYS/OFF/MNL switch
- Expected
- Switch in ON or MNL position (OFF only in emergency)
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Load-From-Cold procedure flowchart execution
Step-by-step execution of Load-From-Cold flowchart including DIP switch configuration, boot floppy insertion, reset cycles, ATINST execution, data and file operations, and parameter changes.
Ranked causes
- 1. Initial setup phase (unknown)
- 2. ATINST program execution and data management (unknown)
- 3. Finalization and parameter configuration (unknown)
Execute load-from-cold standard procedure steps in sequence
- Test point
- Flowchart execution
- Expected
- Each step completes without errors; ATINST program runs successfully
12 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tiewrap Base adhesion to Gantry caution
Epoxy adhesive application procedure for bonding Tiewrap Base to CT gantry surface. Critical timing and placement requirements to ensure proper adhesion without damage to center hole.
Ranked causes
- 1. Improper epoxy mixing ratio or insufficient mixing (common)
- 2. Epoxy poured into center hole causing damage (occasional)
- 3. Inadequate surface preparation (common)
Verify surface preparation
- Test point
- Tiewrap Base and Gantry bonding surfaces
- Expected
- Abraded and cleaned with alcohol
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: -5 VDC power supply output (Zj) high voltage
Output voltage out of range or excessive ripple on -5 VDC rail measured at terminal strip
Ranked causes
- 1. Potentiometer adjustment needed (common)
- 2. Power supply faulty (excessive ripple) (occasional)
Measure -5 VDC output voltage
- Test point
- Terminal strip on front of power supply
- Expected
- –4.9 VDC ∼ –5.1 VDC, +/-0.1 VDC
- Tools
- Digital multimeter (voltage measure mode)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: +12 VDC power supply output (Qj, Qj-Base, Qj-Hlt) high voltage
Output voltage out of range or excessive ripple on +12 VDC rail measured at terminal strip on front of power supply
Ranked causes
- 1. Potentiometer adjustment needed (common)
- 2. Power supply faulty (excessive ripple) (occasional)
Measure +12 VDC output voltage
- Test point
- Terminal strip on front of power supply
- Expected
- +11.8 VDC ∼ +12.4 VDC, +/-0.4 VDC
- Tools
- Digital multimeter (voltage measure mode)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: +12 VDC power supply output (Zj-Adv) high voltage
Output voltage out of range or excessive ripple on +12 VDC rail
Ranked causes
- 1. Potentiometer adjustment needed (common)
- 2. Power supply faulty (excessive ripple) (occasional)
Measure +12 VDC output voltage
- Test point
- Terminal strip on front of power supply
- Expected
- +11.8 VDC ∼ +12.4 VDC, +/-0.4 VDC
- Tools
- Digital multimeter (voltage measure mode)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: +12 VDC power supply output (Zj) high voltage
Output voltage out of range or excessive ripple on +12 VDC rail
Ranked causes
- 1. Potentiometer adjustment needed (common)
- 2. Power supply faulty (excessive ripple) (occasional)
Measure +12 VDC output voltage
- Test point
- Terminal strip on front of power supply
- Expected
- +11.8 VDC ∼ +12.4 VDC, +/-0.4 VDC
- Tools
- Digital multimeter (voltage measure mode)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: +24 VDC power supply output (Zj-Adv) high voltage
Output voltage out of range or excessive ripple on +24 VDC rail
Ranked causes
- 1. Potentiometer adjustment needed (common)
- 2. Power supply faulty (excessive ripple) (occasional)
Measure +24 VDC output voltage
- Test point
- Terminal strip on front of power supply
- Expected
- +23.7 VDC ∼ +24.7 VDC, +/-0.7 VDC
- Tools
- Digital multimeter (voltage measure mode)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: +24 VDC power supply output (Zj) high voltage
Output voltage out of range or excessive ripple on +24 VDC rail
Ranked causes
- 1. Potentiometer adjustment needed (common)
- 2. Power supply faulty (excessive ripple) (occasional)
Measure +24 VDC output voltage
- Test point
- Terminal strip on front of power supply
- Expected
- +23.7 VDC ∼ +24.7 VDC, +/-0.7 VDC
- Tools
- Digital multimeter (voltage measure mode)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: +5 VDC power supply output (Qj, Qj-Base, Qj-Hlt) high voltage
Output voltage out of range or excessive ripple on +5 VDC rail measured at IPU2 Board slot #2 capacitor C1
Ranked causes
- 1. Potentiometer adjustment needed (common)
- 2. Power supply faulty (excessive ripple) (occasional)
Measure +5 VDC output voltage
- Test point
- Capacitor C1 at front edge of IPU2 Board Slot #2
- Expected
- +5.0 VDC ∼ +5.1 VDC, +0.1 VDC
- Tools
- Digital multimeter (voltage measure mode)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: +5 VDC power supply output (Zj-Adv) high voltage
Output voltage out of range or excessive ripple on +5 VDC rail measured at IPU2 Board slot #32 capacitor C1
Ranked causes
- 1. Potentiometer adjustment needed (common)
- 2. Power supply faulty (excessive ripple) (occasional)
Measure +5 VDC output voltage
- Test point
- Capacitor C1 at front edge of IPU2 Board Slot #32
- Expected
- +4.9 VDC ∼ +5.1 VDC, +/-0.1 VDC
- Tools
- Digital multimeter (voltage measure mode)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: +5 VDC power supply output (Zj) high voltage
Output voltage out of range or excessive ripple on +5 VDC rail measured at FCP2A Board slot #36 capacitor
Ranked causes
- 1. Potentiometer adjustment needed (common)
- 2. Power supply faulty (excessive ripple) (occasional)
Measure +5 VDC output voltage
- Test point
- Capacitor at front edge of FCP2A Board Slot #36
- Expected
- +4.9 VDC ∼ +5.1 VDC, +/-0.1 VDC
- Tools
- Digital multimeter (voltage measure mode)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Air and Phantom Calibrations radiation
Obtain baseline calibration files using air and phantom scans after CAM linearity check and Hilight calibrations (Q-CAL, XT-CAL) are completed. These calibrations adjust CT number accuracy.
Ranked causes
- 1. System requires baseline calibration after maintenance (common)
- 2. CT number scaling accuracy required (common)
Confirm Q-CAL and XT-CAL procedures completed before air/phantom calibrations
- Test point
- System calibration status
- Expected
- Q-CAL and XT-CAL calibration files present
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Azimuth belt tension caution
Adjust rotational mechanism belt tension to prevent over-tension damage and ensure proper gantry rotation
Ranked causes
- 1. Belt tension requires adjustment per maintenance procedure (unknown)
- 2. Over tension condition exists or must be prevented (unknown)
Visual inspection: gantry switches and covers
- Test point
- Gantry rear base
- Expected
- Rotate and XG switches in OFF position, rear and side covers removable
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Gravity SAG measurement and storage radiation
Procedure to measure and store detector gravity sag deflection data under varying x-ray parameters
Ranked causes
- 1. Baseline gravity SAG data acquisition and recording (common)
- 2. Gravity SAG data processing and mean file storage (common)
Verify x-ray tube warm-up before air scans
- Test point
- Touch screen case heat indicator
- Expected
- 20% or higher heat level
- Tools
- CT system display
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Rotation deceleration
Adjustment of gantry stop deceleration at specified speeds using VR5 potentiometer on Servo Amp to achieve shortest stop time without overshoot at home position.
Ranked causes
- 1. Gantry overshoots home position during deceleration stop (occasional)
- 2. Gantry does not stop quickly or overshoots after adjustment (occasional)
Remove gantry left side cover
- Test point
- Left side cover
- Expected
- Cover removed
9 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Rotation velocity offset
Adjustment of gantry rotation to maintain correct azimuth position with no drift during continuous rotation using VZERO potentiometer.
Ranked causes
- 1. Gantry drifts CCW during continuous rotation (occasional)
- 2. Gantry drifts CW during continuous rotation (occasional)
Switch OFF Rotate switch
- Test point
- Rotate switch at gantry rear base
- Expected
- Rotate switch OFF
9 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Software loading from MOD and floppy disks
Load-From-Cold (LFC) procedure to rebuild operating software on hard disk after corruption or to apply software revisions. Requires setting CPW DIP switches, inserting boot floppy, pulling reset switch, running ATINST program, initializing disk, and loading software/data files.
Ranked causes
- 1. Hard disk corrupted by failure or crash (occasional)
- 2. Software requires update for improvements or bug fixes (occasional)
Save existing patient data before LFC procedure
- Test point
- Pre-ATINST Procedure menu
- Expected
- All patient data saved to external medium
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: System software loading
Load-From-Cold (LFC) procedure to rebuild operating software on hard disk from MOD and floppy disks after corruption or software updates
Ranked causes
- 1. Hard disk corruption due to failure or crash (occasional)
- 2. Software requires update to incorporate improvements or fixes (occasional)
Save patient image data before proceeding
- Test point
- Image File Management function
- Expected
- All patient data archived to external media
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: X-ray beam detector alignment radiation
Procedure to verify and adjust x-ray beam centering to detector window for two detector configurations (Qj/Qj-Base/Zj and Qj-Hlt/Zj-Adv)
Ranked causes
- 1. Detector Z-axis misalignment from specification (common)
- 2. Detector bowl position misalignment (advanced configuration) (common)
Warm up x-ray tube to operating temperature
- Test point
- Case heat indicator on touch screen
- Expected
- 20% or higher heat level
- Tools
- CT system touch screen display
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cradle fails to move or moves with resistance patient safety
Cradle does not respond to motion commands or exhibits restricted movement during travel
Ranked causes
- 1. Cradle bearing misalignment or inadequate gap (occasional)
- 2. Cradle-to-support bearing contact (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 537 or 538 occurs during scan high voltage
Error 537 or 538 displayed during CT scan operation
Ranked causes
- 1. HV Tank or kV board issue (common)
- 2. Tube or HV Tank failure (occasional)
Run kV control board service tests S33, S34
- Test point
- kV control board
- Expected
- Tests complete without errors
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error X41 with spit noise high voltage
Error X41 occurs during scan with audible spit-like noise; indicates anode-cathode arcing or HV tank issue
Ranked causes
- 1. Spit (arcing) between anode and cathode in X-ray tube (common)
- 2. HV tank oil leakage or air bubble (occasional)
Detect spit issue using oscilloscope
- Test point
- TP9 on kV board (Anode-kV) and TP32 on kV board (Cathode-kV)
- Expected
- No spit waveform observed during high kV scans
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error X41 without spit noise - intermittent high voltage
Error X41 occurs during scan without audible spit noise; indicates HV tank or kV board issue
Ranked causes
- 1. HV tank intermittent fault (occasional)
- 2. kV board or inverter malfunction (occasional)
Run kV board service test
- Test point
- kV board test point diagnostic
- Expected
- No error during kV board test
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gantry cover safety bracket switches disengaged or non-functional patient safety
Safety switches on gantry right side cover are disengaged or not functioning, affecting safety loop closure, gantry rotation control, or alarm sounds
Ranked causes
- 1. Gantry right side cover removed or misaligned (occasional)
- 2. Safety switch mechanical failure or misalignment (rare)
Verify gantry right side cover is properly installed
- Test point
- Gantry right side cover seating and attachment
- Expected
- Cover fully seated and secured with all fasteners tight
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High voltage discharge to ground high voltage
System console displays XG Error codes 533, 537, 538 or 541 indicating possible HV tank or X-ray tube failure with high-voltage to ground discharge
Ranked causes
- 1. HV Tank failure (unknown)
- 2. X-ray tube failure (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
HV Tank or X-ray tube failure indicated high voltage
System console displays XG Error codes 533, 537, 538, or 541; high voltage to ground discharge detected but source unclear between tank and tube
Ranked causes
- 1. X-ray tube high voltage failure (unknown)
- 2. HV Tank high voltage failure (unknown)
Follow troubleshooting tree HV smart diag procedure; may require oscilloscope for differentiation
- Test point
- HV Tank and X-ray tube discharge points
- Expected
- Determine source of high voltage to ground discharge
HV tank or X-ray tube failure with error codes high voltage
System console displays XG error codes 533, 537, 538, or 541 indicating high voltage to ground discharge
Ranked causes
- 1. HV tank failure (unknown)
- 2. X-ray tube failure (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
HV Tank or X-Ray Tube failure with XG errors high voltage
System console displays error codes 533, 537, 538, or 541 indicating high voltage to ground discharge; need to differentiate between HV Tank failure and X-Ray Tube failure to avoid removing good tube while retaining faulty tank
Ranked causes
- 1. HV Tank failure (unknown)
- 2. X-Ray Tube failure (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software Version Upgrade Loading (V411)
Load revision 4.01 system software and perform revision update after ATINST completion.
Ranked causes
- 1. Post-ATINST software version upgrade required (common)
- 2. Custom parameters and data files restoration (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Spits (Error 533 / Synergy 28021) high voltage
Voltage instability or arcing events detected; indicates potential HV tank or tube failure
Ranked causes
- 1. HV tank high voltage to ground discharge (unknown)
- 2. X-ray tube high voltage discharge (unknown)
Follow HV smart diag troubleshooting tree
- Expected
- Procedure outcome per HV smart diag
XG Error displayed on system console high voltage
System console displays XG Error, indicating possible high voltage tank or X-ray tube failure with high probability of HV Tank or X Ray Tube issue
Ranked causes
- 1. HV Tank failure (unknown)
- 2. X Ray Tube failure (unknown)
Follow HV smart diag troubleshooting tree to differentiate Tank versus Tube failures
- Test point
- System console error display
- Expected
- XG Error code (533, 537, 538, or 541)
ATINST Error Menu Response
If ATINST program terminates unexpectedly, error menu appears allowing exit, retry, or diagnostic file inspection.
Ranked causes
- 1. Program error or hardware failure during ATINST execution (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: 25cm water phantom protocol ProSpeed series radiation
Verify image quality at multiple kV and mA settings on 25cm water phantom for ProSpeed models
Ranked causes
- 1. Multi-parameter image quality baseline required (common)
Scan at 80 kV, 3 sec, 350 mA, large focus
- Test point
- 25cm water phantom
- Expected
- MC 0 ± 2
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: After Glow (AV) compensation radiation
Lumex Crystal light emission persists after x-ray stops. Creates After Glow Vector file for compensation.
Ranked causes
- 1. After Glow artifact from Lumex Crystal (common)
Rotate switch OFF
- Test point
- Rotate switch
- Expected
- OFF
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Air and phantom (Qj, Qj-Base, Qj-Hlt, Zj-Adv) radiation
Automatic calibration using air and phantom scans to establish baseline CT reference standards.
Ranked causes
- 1. System requires air and phantom baseline calibration (common)
Verify bow-tie filter installation
- Test point
- Collimator
- Expected
- Bow-tie filter installed
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Air and phantom (Zj) with phantoms radiation
Air calibration initialization followed by phantom calibrations using 35 cm and 25 cm phantoms at specified FOVs.
Ranked causes
- 1. System requires air and phantom baseline calibration with heating precondition (common)
Verify anode heat level before procedure
- Test point
- Touch screen 'Anode' indicator upper right
- Expected
- 10% or below
11 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Aperture ratio verification
Verify that collimator aperture ratios for five focal spot configurations fall within specification 0.10–0.16 (1mm) to 0.10–0.16 (10mm) by acquiring raw data, calculating mean files, and graphical profile analysis via SUPPORT DISPLAY.
Ranked causes
- 1. Aperture ratio out of specification (unknown)
Insert BOOT floppy and access Maintenance menu
- Test point
- Gantry control panel
- Expected
- MAINTENANCE MENU displays; CALIBRATION option visible
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Bow-tie filter curve verification caution
Verify bow-tie filter curve is smooth and properly centered by acquiring two air stationary scans (filter installed and removed), computing filter transmission curve, and inspecting graphical profile for channels 1–500 and 400–end.
Ranked causes
- 1. Filter curve irregular or out of center (unknown)
Power down and prepare gantry for filter access
- Test point
- Gantry rear base rotate switch and covers
- Expected
- Rotate switch OFF; front and both side covers removed
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: BPU switch configuration
Configure BPU board switch settings (SWI5A1) for correct operation
Ranked causes
- 1. Factory default configuration required (unknown)
Verify BPU switch SWI5A1 position matches normal setting
- Test point
- BPU SWI5A1
- Expected
- As illustrated in Illustration 2–30
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: BPU2 switch configuration
Configure BPU2 board switch settings (SW15A1) for correct operation on ProSpeed Plus and ProSpeed EF models
Ranked causes
- 1. Factory default configuration required (unknown)
Verify BPU2 switch SW15A1 position matches normal setting
- Test point
- BPU2 SW15A1
- Expected
- As illustrated in Illustration 2–24
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CAM Amplifier Linearity Check radiation
Verify detector amplifier response linearity after X-ray tube, detector, collimator, filter, or alignment changes. Two offline air scans at 100 mA and 200 mA are performed and mean files generated to check linearity ratio.
Ranked causes
- 1. Amplifier non-linearity detected (unknown)
Verify CAM amplifier ratio between 100 mA and 200 mA scans
- Test point
- Calculated mean file (M file 3)
- Expected
- Linearity ratio 0.99 to 1.01 across channels 1–500 and 300–end, ±0.01
- Tools
- SUPPORT DISPLAY software
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Collimator thickness configuration for GEMS-Europe caution
Configure collimator thickness setting for ProSpeed systems in GEMS-Europe
Ranked causes
- 1. 2 mm thickness incompatible with certain ProSpeed models in GEMS-Europe (common)
Verify collimator thickness setting for GEMS-Europe systems
- Test point
- Collimator thickness configuration menu
- Expected
- 3 mm selected for ProSpeed S Fast, SX Power, SX Advantage, or Advantage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Compact Siting Mode Dip Switch Configuration
Set TGP dip switches for Compact Siting (Short Foot Print) Mode 1 or 2 when ROM part number 2106128–2 or later is installed
Ranked causes
- 1. Space-constrained installation requires Compact Siting mode (rare)
Verify TP ROM part number is 2106128–2 or later
- Test point
- TP ROM location U3P
- Expected
- Part number 2106128–2 or later
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CPW Board DIP switches caution
Configure DIP switches on CPW board for correct test mode, RS232C baud rate, boot source, and HDD configuration
Ranked causes
- 1. DIP switch configuration required during installation or service (common)
Verify DIP switch 5A1 position
- Test point
- CPW board DIP switch 5A1
- Expected
- All positions set to 0 (not used)
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CPW Board jumpers caution
Configure jumpers on CPW board for SRAM battery backup, Power ON Reset, Bus Timeout, and INTVL settings
Ranked causes
- 1. Jumper configuration required for proper operation (common)
Verify jumper 30D1 SRAM battery backup position
- Test point
- CPW board jumper 30D1
- Expected
- STBY for battery backup (normal setting)
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Cradle motor maximum torque caution
Adjustment of motor driver RUN potentiometer to set maximum torque via CP+/CP– voltage
Ranked causes
- 1. Motor torque requires calibration adjustment (unknown)
Verify cradle latch switches position
- Test point
- Cradle latch switches
- Expected
- Both set to 'Latch'
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CT number adjustment (Qj, Qj-Base) radiation
Modifies CAL files so water CT=0 and air CT=-1000. Measures CT values from air and phantom scans at all FOVs (25-50 cm) and kV (80-140). Takes ~3 hours. Temperature affects water CT by -0.4/°C.
Ranked causes
- 1. Air and phantom calibration scans show CT deviations from standard (common)
Preset scan parameters for all FOVs and thicknesses
- Test point
- [Preset Prmtr] key while [New Patient] flashes
- Expected
- SCAN PARAMETER menu; select I-EXTR and MULTI/DYNAMIC; create params for each FOV/TH combo
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: DBM and M32W/M64W jumper configuration
Configure DBM board jumpers for correct operation with M32W or M64W memory modules
Ranked causes
- 1. Factory default jumper configuration required (unknown)
Verify DBM board jumpers match normal setting
- Test point
- DBM jumpers
- Expected
- As illustrated in Illustration 2–29
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Detector beam-on-window alignment radiation
Verify X-ray beam exposure pattern on detector using Polaroid film imaging to ensure proper beam window positioning
Ranked causes
- 1. Detector window beam alignment requires verification (common)
Verify X-ray tube anode heat level before detector beam check
- Test point
- Touch screen 'A' or 'Anode' indicator upper right
- Expected
- Anode heat at 40% or higher
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: DTRE Board Jumper Configuration
Set jumper settings on DTRE Board for Qj–Base, Qj–Hlt, or Zj–Adv variants
Ranked causes
- 1. DTRE Board jumper configuration required for system variant (common)
Verify DTRE Board jumper configuration for system variant
- Test point
- DTRE Board jumpers (location shown in Illustration 3–2)
- Expected
- Jumpers set correctly for Qj–Base, Qj–Hlt, or Zj–Adv variant
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Dynamic gain (DG) compensation radiation
Compensates for CT number variations caused by CAM Board gain changes during different scan times.
Ranked causes
- 1. CAM Board gain changes from different scan times (common)
Rotate switch OFF
- Test point
- Rotate switch
- Expected
- OFF
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: FCP3B switch configuration
Configure FCP3B board switch settings (SW15A1) for correct operation on ProSpeed VX, ProSpeed, Accell, and ProSpeed models
Ranked causes
- 1. Factory default configuration required (unknown)
Verify FCP3B switch SW15A1 position matches normal setting
- Test point
- FCP3B SW15A1
- Expected
- As illustrated in Illustration 2–25
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Calibration: FCP4 jumper configuration
- Calibration: FCP4 switch configuration
- Calibration: Floppy disk initialization for data save
- Calibration: Floppy disk initialization for system data backup
- Calibration: FMD switch configuration
- Calibration: FPA + FPAM configuration
- Calibration: FPA2 switch configuration
- Calibration: Gantry tilt range display
- Calibration: Gantry tilt rotation
- Calibration: Gravity SAG data review via display utility
- Calibration: Gravity SAG measurement
- Calibration: Hilight Calibration
- Calibration: IPU2 Board memory configuration
- Calibration: IPU2 Board recon processor selection
- Calibration: ISC board switch configuration
- Calibration: Iso Center Alignment
- Calibration: ISO Center Alignment (Qj-Hlt, Zj-Adv)
- Calibration: ISO Center Alignment (Qj, Qj-Base, Zj)
- Calibration: KV and MA output verification
- Calibration: KV check with HV divider
- Calibration: M64W memory assembly switch configuration
- Calibration: MA check
- Calibration: MISCE board jumper and switch configuration (option)
- Calibration: Mixed-material screw tightening
- Calibration: MTC board jumper and switch configuration (option)
- Calibration: ODC2 board jumper and switch configuration (option)
- Calibration: OGP Board Dip Switch Configuration
- Calibration: Operator Console switch and jumper settings
- Calibration: PDC board jumper and switch configuration
- Calibration: PDC jumper JP1B2 configuration
- Calibration: PDC switch SWG6A1 configuration
- Calibration: Phantom Calibration
- Calibration: Positioning lights alignment
- Calibration: Q Calibration (Hilight)
- Calibration: Rotation speed
- Calibration: Screw tightening torque
- Calibration: SCSI board jumper and switch configuration
- Calibration: Servo Amplifier Dip Switch Configuration
- Calibration: Set cursor for image measurement
- Calibration: Set cursor on reconstruction display
- Calibration: Software loading post-ATINST procedure
- Calibration: Software loading V/R 5.01 and 5.02
- Calibration: Special startup procedure (FMP1 board E-Link configuration)
- Calibration: STL board jumper and switch configuration (option)
- Calibration: Table perpendicularity verification
- Calibration: Table right-left positioning
- Calibration: Table safety rotation control
- Calibration: TGP Board Biopsy Mode Dip Switch Setup
- Calibration: TGP Board Dip Switch Configuration
- Calibration: Tilt angle
- Calibration: Tilt guide roller gap
- Calibration: Tilt roller operation
- Calibration: Tilt speed
- Calibration: Video signal for multiformat camera
- Calibration: Water phantom surface alignment to scan plane
- Calibration: X-ray alignment after detector and gravity procedures
- Calibration: X-ray alignment plane of rotation
- Calibration: XT Calibration (Crosstalk)
- Communication lost between OC and DC III
- Cradle motor RUN potentiometer adjustment
- Cradle motor torque not in range 17-18 kg
- CRT monitor display controls out of adjustment
- CT-MAX Tube to CT-Plus Tube replacement requires X-Ray Generator modification
- Detector bias voltage failure
- Error 537 or 538 during scan
- Error beep volume incorrect
- Fan defective or inoperative
- Gantry does not rotate
- Gantry Error 4 or Error 6 occurs during rotation
- Gantry Home button operation verification
- Gantry home position malfunction
- Gantry positioning lights control verification
- Gantry rotation hazard during position adjustment
- Hoist hook cannot catch tube hanging bar
- Hoist hook cannot catch tube hanging bar during replacement
- Hoist hook cannot engage tube hanging bar during replacement
- HV system intermittent noise or arcing
- Oil leak from cylinder, pump or hoses
- Oil leaks from cylinder, pump or hoses
- Plasma display calibration required
- Power supply ground connection poor
- Power supply ground connection verification required
- Preparation: Remove gantry covers and collimator filter
- Scan box LED indicator lights not illuminated
- SCAN BOX LED indicators not illuminated
- Scan box LED switches not illuminated
- SCAN BOX LEDs broken or not illuminated
- STOP potentiometer factory setting
- System hang-up prevention with ELINK option
- Table and Gantry interlock verification
- Table lowers on its own when unoperated
- TGP Board relay replacement required
- Tiewrap base detaches from gantry cover
- Tiewrap base falls from gantry cover
- Tiewrap base falls off gantry cover
- Tiewrap base loose on gantry cover
- Tiewrap base on gantry cover comes loose or falls off
- Touch panel tilt angle misaligned
98 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
28021 | Spits (Synergy series equivalent of code 533) | Voltage instability or arcing in HV circuit | error |
28025 | Anode Over KV (Synergy series equivalent of code 537) | Anode voltage exceeds normal level | error |
28026 | Cathode Over KV (Synergy series equivalent of code 538) | Cathode voltage exceeds normal level | error |
28029 | Anode Overcurrent (Synergy series equivalent of code 541) | Anode current exceeds normal limit | error |
533 | Spits (Error Code 28021 for Synergy) | High voltage to ground discharge | error |
533 | Spits | Voltage instability or arcing events in HV tank or X-ray tube | error |
537 | HV system error related to error condition x41 patterns | Intermittent or persistent high-voltage fault detection | warning |
537 | High voltage or kV board error | Occurs during CT scan operation | error |
537 | Anode Over KV (Synergy: 28025) | Anode high voltage threshold exceeded | error |
537 | Anode Over KV (Error Code 28025 for Synergy) | High voltage to ground discharge in anode circuit | error |
538 | Cathode Over KV (Error Code 28026 for Synergy) | High voltage to ground discharge in cathode circuit | error |
538 | HV system error related to error condition x41 patterns | Intermittent or persistent high-voltage fault detection | warning |
538 | High voltage or kV board error | Occurs during CT scan operation | error |
541 | Anode Overcurrent (Error Code 28029 for Synergy) | Excessive anode current detected | error |
C2 | FHS error | Error condition during ATINST operations | error |
CALIBRATION FILE INCOMPATIBILITY | Calibration files of old version and new version have no compatibility | Approximately 30 minutes after ATINST start when incompatible calibration file versions detected | warning |
Current Task Aborted Unexpectedly | ATINST program encountered error during execution | Program abort due to program bug, hardware failure, or deviation between program execution and expected execution (Text File vs Log File mismatch) | error |
Error 4 | Gantry rotation malfunction | Gantry rotation function commanded; AZ CONN board relay failure detected | error |
Error 6 | Gantry rotation failure | Gantry rotation function commanded; AZ CONN board relay failure detected | error |
X41 | High voltage tank or X-ray tube fault; arcing or critical hardware failure | Error occurs during X-ray scan operations | critical |
x41 | Generic high-voltage system error condition | Various HV circuit faults or anomalies | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.