GE Healthcare CT HiSpeed: troubleshooting and error codes
The ServiceForge knowledge base holds 301 documented symptoms for the GE Healthcare CT HiSpeed, with 363 ranked causes, 1051 diagnostic checks with test points and expected values, and 19 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
HV chain fault detection high voltage
Fast and slow safeties monitor inverter performance; fast safeties cut HV power on error; slow safeties measure signals every 1 ms for fault detection
Ranked causes
- 1. Over kV condition (common)
- 2. No kV condition (common)
- 3. kV regulation error at 7th error in exposure (occasional)
- 4. Anode spit exceeds max allowed count (occasional)
- 5. Cathode spit exceeds max allowed count (occasional)
- 6. Max resonant current exceeded (occasional)
- 7. Gate supply out of range (occasional)
Monitor kV measurement
- Test point
- kV measure signal
- Expected
- Within safe operating range
3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
TGP board does not initialize on power-on
After powering on or resetting TGP board, processor fails to complete ROM and RAM tests; ERRG, LPG, RQG, ERRT, LPT, RQT, MNT, ERRM, LPM, RQM, or MNM LEDs light and remain on
Ranked causes
- 1. Gantry processor RAM check failure (U87, U92, U77) (rare)
- 2. Gantry processor ROM check failure (U112) (rare)
- 3. Table processor RAM check failure (U42, U64, U48) (rare)
- 4. Table processor ROM check failure (U68) (rare)
- 5. Management processor RAM check failure (U88, U62, U48, U77) (rare)
- 6. Management processor ROM check failure (U111) (rare)
Observe gantry processor LED status on TGP board after power-on
- Test point
- ERRG, LPG, RQG LEDs on TGP board
- Expected
- LPG LED should blink after successful RAM and ROM checks; ERRG LED indicates hang up state
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Miscellaneous Hardware Tests patient safety
Operator Console miscellaneous tests including Hardware Inventory, SCSI Test, System Log viewing, NPR LED Test, Safety Loop test, and Shutdown test to verify system configuration and relay operation.
Ranked causes
- 1. Hardware configuration change or SCSI device malfunction (occasional)
- 2. NPR LED display malfunction (rare)
- 3. Safety loop relay malfunction (occasional)
- 4. System shutdown failure (rare)
Access Misc Test from diagnostics menu
- Test point
- H/W Diagnostics menu
- Expected
- Misc test selection window appears after selecting [Service] -> [Diagnostics & analysis] -> [H/W Diagnostics] -> [Interactive Test] -> [Misc Test]
6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electromagnetic interference with other equipment caution
CT system may cause EMC issues affecting connected or nearby electronic devices through air or cables
Ranked causes
- 1. Improper installation or cable routing (common)
- 2. System covers left open during operation (occasional)
- 3. Loose fasteners or damaged gaskets (occasional)
- 4. Unauthorized modifications to system (rare)
Verify installation location distance from other equipment
- Test point
- General equipment area
- Expected
- Equipment positioned as far as possible from other electronic devices
5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Positioning lights halogen horizontal alignment (right-left) caution
Inner and outer right-left lights horizontally misaligned; top light horizontal misalignment; vertical light offset from phantom surface
Ranked causes
- 1. Inner right-left light horizontal adjustment screw (B) out of tolerance (common)
- 2. Outer right-left light horizontal adjustment screw (B) out of tolerance (common)
- 3. Top light horizontal adjustment screw (B) out of tolerance (common)
Verify inner right-left light horizontal alignment
- Test point
- Vertical positioning light relative to phantom surface
- Expected
- Vertical light and phantom surface aligned within ±0.5 mm, ±0.5 mm
- Tools
- Positioning lights, adjustment screw B, lever D, water phantom
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CAM amplifier linearity out of specification
Detector channel response linearity fails verification during no-X-ray test scans
Ranked causes
- 1. Amplifier gain drift or non-linearity (occasional)
- 2. Detector channel response mismatch (occasional)
- 3. Insufficient warm-up before test (common)
Verify vector file 7 linearity ratio within tolerance
- Test point
- View Vector plot for Vector File No. 7
- Expected
- 0.9995 < ratio < 1.0005
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Component ejection hazard during gantry rotation with covers removed patient safety
Risk of components or objects being ejected at high velocity when gantry rotates with side covers removed
Ranked causes
- 1. Side covers removed during gantry operation or rotation (common)
- 2. Board chassis left open during scanning (occasional)
- 3. Components not properly secured with correct torque (occasional)
Inspect gantry side covers for removal or damage
- Test point
- Right and left side covers of gantry
- Expected
- Covers installed and secured in place
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DAS offset data out of specification
Offset data averages outside acceptable range without offset file correction
Ranked causes
- 1. Detector or DAS channel drift or malfunction (unknown)
- 2. Offset file correction needed (common)
- 3. Detector response drift after warm-up (occasional)
Verify Avg of Mean of Active Views without offset correction
- Test point
- Fan Data Averages display
- Expected
- 3000 ≤ Avg of Mean ≤ 6000
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrocution hazard from 400VAC slip ring supply high voltage
Risk of electrocution when servicing X-ray generator or slip ring components due to active 400VAC supply
Ranked causes
- 1. XG Power Switch (400VAC) not turned OFF before service (common)
- 2. CB1 and CB2 circuit breakers not turned OFF (common)
- 3. Capacitors in XG units not discharged (occasional)
Verify XG Power Switch is OFF
- Test point
- XG Power Switch at gantry rear base
- Expected
- Switch in OFF position; System Safety Loop open
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gantry rotates unintentionally during service patient safety
Gantry continues rotating or can rotate when it should be disabled, posing impact hazard to personnel
Ranked causes
- 1. Axial Drive Switch not turned OFF (common)
- 2. Right maintenance cover removed without reinstalling before other covers installed (occasional)
- 3. Service brackets not engaged when covers removed (occasional)
Verify Axial Drive Enable Buzzer status
- Test point
- Gantry rear base
- Expected
- Buzzer is silent
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gantry rotation fails during power-on test patient safety
Gantry does not rotate or rotation check terminates prematurely with ERRG LED lighting during power-on initialization
Ranked causes
- 1. Servo amplifier not powered (common)
- 2. Service switch (SW1) set to OFF (occasional)
- 3. Gantry azimuth counter initialization timeout (> 25 sec) (occasional)
Verify service switch SW1 position on TGP board
- Test point
- Service switch SW1 on TGP board
- Expected
- Set to SYS (normal) or MNL (manual)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Memory module installation and configuration incomplete
System not properly configured after installing IPU2 board and memory modules; requires MD EXT switch settings and P-ON test verification before operation
Ranked causes
- 1. MD EXT switch (S7A1) not set to correct configuration for kit type (common)
- 2. IPU2 board not installed in OC nest slot 2 (common)
- 3. P-ON test level 2 not run to verify system startup (common)
Run P-ON test level 2 to verify system startup
- Test point
- CPW Board slot 1, Dip SW 8A1
- Expected
- Startup screen appears after approximately 14 minutes with no errors
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power line voltage disturbances not monitored high voltage
Input power line voltage requires monitoring for mean voltage, surges, sags, impulses, and frequency stability during installation to detect correlation with system malfunctions
Ranked causes
- 1. Line voltage out of specification or unstable (occasional)
- 2. Mean line voltage outside acceptable range (rare)
- 3. Line frequency drifting outside tolerance (rare)
Measure current line voltage at PDU input terminals L1, L2, L3
- Test point
- Power Distribution Box terminals L1, L2, L3
- Expected
- 200/208/380/400/415/440/460/480 VAC at 50/60 Hz
- Tools
- Line analyzer (Dranetz 606-3 or equivalent)
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software loading: Load-From-Cold (LFC) procedure
Full system software loading (OS + NP application) to system hard disk from CD-ROMs and MODs. Reinitializes all system data disks, erasing all images and scan data.
Ranked causes
- 1. System software update or downgrade required (common)
- 2. Patch software available for NP application (occasional)
- 3. System downgrade to older version (rare)
All patient images and scan data saved to MOD
- Test point
- MOD optical disk
- Expected
- Images and raw data archived successfully
10 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature environment out of specification caution
Scan room and equipment room temperature and humidity do not meet installation specifications; excessive temperature may overstress components
Ranked causes
- 1. Scan room temperature outside range 20–28 °C (common)
- 2. Equipment room temperature outside range 15–30 °C (occasional)
- 3. Relative humidity outside safe range (occasional)
Record scan room temperature
- Test point
- Scan room with thermometer/recorder
- Expected
- 20 °C to 28 °C with at least 5 °C/hr gradient, to 28 °C
- Tools
- Thermometer, temperature/humidity recorder
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
X-ray exposure risk during improper equipment operation radiation
Potential radiation injury to operators and bystanders from inadequate shielding or lack of protective precautions during CT scanner use and maintenance
Ranked causes
- 1. Operator or technician not trained in radiation safety protocols (common)
- 2. Inadequate use of available protective materials and devices (common)
- 3. Unqualified personnel performing electrical or maintenance work (occasional)
Verify radiation safety training credentials for all personnel
- Test point
- Training documentation
- Expected
- Current NCRP and ICRP training certification on file
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormal gantry rotation noise or movement caution
Gantry movement is not smooth or abnormal noise detected during manual rotation
Ranked causes
- 1. Brush Block Assy not properly restored to original configuration (common)
- 2. Brush assembly installation issue (occasional)
Manual gantry rotation test
- Test point
- Gantry
- Expected
- Smooth movement with no abnormal noise through at least one complete rotation
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Audio Function Test
Operator Console audio functional check including OC and table speaker tests for Auto Voice, CD player, X-ray ON buzzer, and Alert. Tests verify volume control and sound output.
Ranked causes
- 1. Audio output or volume control malfunction (occasional)
- 2. Auto Voice Playback still active (occasional)
Access Audio Function Test from diagnostics menu
- Test point
- H/W Diagnostics menu
- Expected
- Audio test screen appears after selecting [Service] -> [Diagnostics & analysis] -> [H/W Diagnostics] -> [Interactive Test] -> [Audio Test]
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Azimuth belt tension caution
Adjust axial belt tension to specified gap between tensioner components to maintain drive without over-tensioning
Ranked causes
- 1. Belt tension incorrect or over-tensioned (common)
- 2. Belt over-tensioned due to incorrect adjustment (occasional)
Loosen mounting and adjustment hardware
- Test point
- Mounting plate bolts and set nut
- Expected
- All bolts and nut loose enough to allow adjustment
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CAM Amplifier Linearity radiation
Check for non-linearity in detector amplification response across different tube currents. Performed after X-ray tube, detector, DAS, or CAM board replacement.
Ranked causes
- 1. Detector or amplifier non-linearity after component replacement (common)
- 2. Detector crystal response drift (occasional)
Perform 100 mA air stationary scan
- Test point
- Offline Scan service menu
- Expected
- Scan 1: Azimuth 0°, Thickness 10 mm, Focus Large, KV 120, MA 100 mA
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Halogen positioning lights alignment patient safety
Five halogen positioning lights (inner right/left, outer right/left, outer top) must be aligned with scan center and phantom surface for accurate patient positioning
Ranked causes
- 1. Positioning lights misaligned vertically or horizontally (common)
- 2. Water phantom surface not parallel to scan plane (occasional)
Place fine pin at phantom holder center and position perpendicular to scan plane
- Test point
- M3 hexagonal wrench at phantom holder center
- Expected
- Pin aligned with vertical line in magnified (x4) image
7 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Positioning lights halogen vertical alignment (right-left) caution
Inner and outer right-left positioning lights vertically misaligned with pin; horizontal light offset from pin image
Ranked causes
- 1. Inner right-left light vertical adjustment screw (A) out of tolerance (common)
- 2. Outer right-left light vertical adjustment screw (A) out of tolerance (common)
Position pin perpendicular to scan plane
- Test point
- Pin image versus vertical reference line
- Expected
- Pin and vertical line aligned within ±0.5 mm, ±0.5 mm
- Tools
- M3 hexagonal wrench or fine pin, phantom holder, cursor/magnification tools
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Remove gantry front cover caution
Access to internal gantry components by removing front cover with service angles and connector management
Ranked causes
- 1. Front cover removal procedure for standard installation (common)
- 2. Front cover removal for small room (short-footprint mode) (occasional)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Remove gantry side covers caution
Access to side gantry components by removing left and right side covers with connector disconnection as needed
Ranked causes
- 1. Left side cover removal (common)
- 2. Right side cover removal (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: RMT CNT jumper configuration
Setting JP1 jumper plug on RMT CNT (Remote Control) module according to TGP board part number and revision
Ranked causes
- 1. RMT CNT part 2175741 with TGP 2156510-3 or earlier (common)
- 2. RMT CNT part 2175741-2 or later with TGP 2156510-4 or later (common)
Identify RMT CNT part number
- Test point
- RMT CNT module label
- Expected
- Part No. 2175741 or 2175741-2 or later
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Rotation balance caution
Achieve balance of rotating gantry assembly through rough and fine weight adjustment to minimize resistance and ensure smooth rotation
Ranked causes
- 1. Gantry unbalanced in CW or CCW direction (common)
- 2. Main weight unbalanced (common)
Rough balance check with main weight at 6 o'clock
- Test point
- Main weight position
- Expected
- Gantry does not rotate on its own
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Saving system state data
Save system configuration, preferences, calibration files, and operational data to MOD before performing software installation to enable restoration after LFC/LFW procedure.
Ranked causes
- 1. System state data must be preserved before software re-initialization (common)
- 2. Option key information needs backup (common)
Verify system powered ON
- Test point
- Main switch of OC
- Expected
- System on and main screen displayed
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Emergency stop switch not accessible or not activated during emergency patient safety
Inability to stop gantry rotation quickly during emergency situation due to inaccessible or misused stop switches
Ranked causes
- 1. Using Stop Scan switch instead of Emergency Stop switch (common)
- 2. Emergency stop switches on gantry frame not connected when covers removed (occasional)
Identify correct Emergency Stop switch location
- Test point
- Operator console or gantry
- Expected
- Emergency Stop Switch clearly visible and accessible (not Stop Scan switch on keyboard)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gantry rotation check not performed patient safety
Gantry mechanical rotation, absence of interference, and proper home position return must be verified before operation
Ranked causes
- 1. Gantry mechanical obstruction or malfunction (occasional)
- 2. Gantry fails to stop at home position (rare)
Switch OFF Rotate switch
- Test point
- Gantry rear base Rotate switch
- Expected
- Switch in OFF position
10 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gantry tilt or table up/down relay malfunction patient safety
Relay circuit designed with dual series relays fails to properly disable pump/valve mechanism when relay should be open; relay remains closed or fails to activate on command
Ranked causes
- 1. Direct relay stuck closed (enable relay fails to isolate when open) (occasional)
- 2. Enable relay stuck closed (processor cannot command relay to open) (occasional)
Activate enable relay and verify pump/valve mechanism does not operate
- Test point
- Gantry tilt and table up/down pump/valve mechanism
- Expected
- No operation; no fluid flow; no movement
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gantry tilting hazard during service with covers loose patient safety
Risk of gantry falling or unexpected movement during tilting if covers and support mechanisms are not properly secured
Ranked causes
- 1. Covers not secured before tilting gantry (common)
- 2. Azimuth lock pin not used to hold gantry during component removal (common)
Verify all gantry covers are secured
- Test point
- Side covers, front cover, rear cover, tube hoist rail
- Expected
- All covers installed, latched, and secure
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Management processor communications failure
Management processor unable to communicate with operator console or OGP board on gantry rotative frame; IMS (Intermediate Support) not responding
Ranked causes
- 1. Communication interface fault between management processor and console/OGP board (unknown)
- 2. IMS (Intermediate Support) interface fault (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Memory capacity verification required after upgrade
System memory must be validated through scanning, paging, and post-reconstruction tests to ensure proper function after memory module installation
Ranked causes
- 1. Memory not verified after module installation (common)
- 2. Post reconstruction test not completed (common)
Run POST RECON test with image subtraction
- Test point
- Image processing, Subtraction function
- Expected
- Subtracted image (A.B.C;1 minus A.B.C;2) shows nil image at Window Width 2, Window Level 0
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
OGP ROM not verified or updated
SP ROM on OGP board requires verification of part number and potential replacement with newer version to ensure gantry operation compatibility
Ranked causes
- 1. OGP ROM part number is 2121676-2 or lower (outdated) (occasional)
- 2. OGP ROM part number is 2121676-3 or later (current) (occasional)
Verify SP ROM part number on OGP board at location U15E
- Test point
- OGP board, location U15E
- Expected
- Part number 2121676-3 or later (current version) part number
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Signal brush contact failure or intermittency patient safety
Signal brushes or slip rings show poor electrical contact causing signal loss or intermittent data transmission during gantry rotation
Ranked causes
- 1. Brush wear below check line, reduced contact pressure (common)
- 2. Slip ring contamination, dust, or grease coating reducing contact (common)
Perform signal brush contact voltage check on problematic ring pair
- Test point
- Oscilloscope measurement across 120 Ω resistor while rotating
- Expected
- No voltage drops greater than 0.5 V, smooth continuous signal without fluctuations, +/- 0.5 V
- Tools
- Digital storage oscilloscope in envelope mode, Ring Service Tool (2206131)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software loading: Load-From-Warm (LFW) procedure
NP application loading only from CD-ROMs and MODs to system hard disk. System remains powered on. Does NOT reinitialize system data disks.
Ranked causes
- 1. NP application software update required (no OS change) (common)
- 2. NP patch software available (occasional)
NP application software installation in LFW mode completed
- Test point
- Display after automatic reboot
- Expected
- IRIS main menu appears without OS reinstall
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
X-ray generator KV stabilization and mA linearity check radiation
Verify KV and mA output remain within specification across multiple stations during functional check
Ranked causes
- 1. KV deviation exceeds ±5 kV specification (occasional)
- 2. mA deviation exceeds ±20 mA specification (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC/DC 3 Phase Mid Power Assembly failure high voltage
DC bus voltage loss; rectifier block malfunction; AC/DC conversion failure; board overheating or component failure
Ranked causes
- 1. AC/DC board electronic failure or rectifier block failure (unknown)
DC bus voltage
- Test point
- +DC and −DC terminals on board
- Expected
- Correct DC voltage per system specification
- Tools
- DMM
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Access XG CPU board for firmware check caution
Procedural steps to safely access CPU board in electronic rack for inspection
Ranked causes
- 1. Routine maintenance procedure for firmware version verification (unknown)
Turn Rotate, XG, Slip Ring switches OFF
- Test point
- Rear base of Gantry
- Expected
- All three switches in OFF position
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Aperture assembly requires replacement caution
Aperture assembly needs removal and installation with ratio check verification
Ranked causes
- 1. Aperture assembly wear or malfunction (unknown)
Set gantry to 0° tilt and HOME position
- Test point
- Gantry tilt angle and position
- Expected
- 0° tilt, HOME position
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Aperture motor driver failure
Aperture motor driver not responding or not controlling aperture position
Ranked causes
- 1. Defective aperture motor driver assembly (common)
Verify aperture motor driver response to control signals
- Test point
- Aperture Motor Driver Assy connectors
- Expected
- Aperture opens and closes smoothly in response to commands
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio function test - Gantry speaker Auto Voice and CD player
Verify Gantry speaker outputs Auto Voice and CD player audio with volume control
Ranked causes
- 1. Gantry speaker or audio output circuit failure (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio function test - OC speaker alert
Verify alert sound outputs once and alert width and frequency controls function
Ranked causes
- 1. Alert sound circuit or audio output circuit failure (unknown)
Alert sound output
- Test point
- OC speaker
- Expected
- Alert sounds once when [ON] button clicked once
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio function test - OC speaker Auto Voice
Verify OC speaker outputs Auto Voice message and volume controls function
Ranked causes
- 1. OC speaker or audio output circuit failure (unknown)
Auto Voice message output and volume control
- Test point
- OC speaker
- Expected
- Auto Voice message audible; volume adjustable via L and R sliders
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio function test - OC speaker CD player
Verify OC CD player outputs audio and volume controls function
Ranked causes
- 1. OC CD player or audio output circuit failure (unknown)
CD player audio output and volume control
- Test point
- OC speaker
- Expected
- CD audio audible when music CD played; volume adjustable via L and R sliders
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio function test - OC speaker X-ray ON buzzer patient safety
Verify X-ray ON buzzer sounds and buzzer modulation and frequency controls function
Ranked causes
- 1. X-ray ON buzzer or audio output circuit failure (unknown)
X-ray ON buzzer sound continuity
- Test point
- OC speaker
- Expected
- Buzzer sounds continuously while [ON/OFF] button is ON; stops when OFF
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auto voice not heard or unintelligible patient safety
Autovoice announcements for scan start or end not heard or not understandable
Ranked causes
- 1. Audio output circuit malfunction or speaker failure (unknown)
Listen for autovoice scan start announcement
- Test point
- Audio speaker output
- Expected
- Clear, understandable audio announcement for scan start
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Axial drive belt requires replacement caution
Axial drive belt needs removal and installation after removing covers and setting azimuth lock pin
Ranked causes
- 1. Axial drive belt wear or damage (unknown)
Switch OFF Rotate, XG Power, Slip Ring115V switches
- Test point
- Gantry rear base switches
- Expected
- All three switches OFF
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Axial scan does not complete or errors occur
Axial scan with specified parameters fails to complete without error
Ranked causes
- 1. DAS/Detector malfunction or X-ray generator instability (unknown)
Perform axial scan with test parameters
- Test point
- Detector and X-ray generator during scan
- Expected
- Scan completes without error kV, as set
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Balun assembly requires replacement caution
Balun assembly with coil needs removal and installation with data transfer check and gantry restoration
Ranked causes
- 1. Balun assembly failure or coil damage (unknown)
Switch OFF Rotate, XG Power, Slip Ring115V switches
- Test point
- Gantry rear base switches
- Expected
- All three switches OFF
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Brush block assembly misalignment caution
Brushes not centered between slip ring track barriers after installation; potential contact issues
Ranked causes
- 1. Brush Block Assy pin not aligned to corresponding hole (common)
Visual inspection of brush position
- Test point
- Brush Block Assy
- Expected
- Brushes centered between slip ring track barriers
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Air and Phantom Calibration Sequence radiation
Automated sequential execution of Air Calibration followed by Phantom Calibration using calibration station phantoms. Establishes baseline CT number reference values. Precondition: bow-tie filter must be installed. Performed via Auto Sequence 1.
Ranked causes
- 1. Air and phantom calibration required after Hilight calibration or major component replacement (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Air and Phantom Calibrations radiation
Establish baseline detector response for water and air reference standards. Bow-tie filter must be installed. Performed after Hilight calibrations (Q, XT, AV, DG as required). Includes Auto Sequence 1 (Air Cal → Phantom Cal).
Ranked causes
- 1. Detector sensitivity baseline drift after calibration-triggering component replacement or alignment (common)
Verify bow-tie filter is installed on collimator
- Test point
- Gantry collimator assembly
- Expected
- Bow-tie filter present and properly seated
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: AV Calibration (AV-CAL) radiation
Compensate for Lumex crystal afterglow persistence following X-ray pulse cessation. Performed when detector or alignment is changed. Takes ~2 minutes. Gantry does not rotate.
Ranked causes
- 1. Detector afterglow (phosphorescence) from Lumex crystals after detector replacement or alignment (common)
Verify X-ray tube is sufficiently cool before starting
- Test point
- Tube case temperature indicator
- Expected
- Below 10% heat level
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: AV-CAL (Afterglow Compensation) radiation
Create After Glow Vector file to compensate for persistent light emission from Lumex crystal after x-ray radiation stops. Gantry does not rotate during this procedure. Approximately 2 minutes duration.
Ranked causes
- 1. Detector afterglow artifact after detector replacement or beam-on-window alignment (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Axial scan data collection parameters radiation
System collects 972 views of actual data during axial scan at view period synchronized with gantry azimuth; scan speeds vary 0.8 to 3.0 sec
Ranked causes
- 1. Axial scan data collection with speed-dependent view period (common)
Verify correct view period for selected scan speed
- Test point
- Scan speed setting
- Expected
- 0.8 sec: 823 µsec/view; 1.0 sec: 1029 µsec/view; 1.5 sec: 1543 µsec/view; 2.0 sec: 2058 µsec/view; 3.0 sec: 3086 µsec/view
- Tools
- System configuration
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Axial scan geometry and offset correction radiation
Axial scans perform 360-degree rotation collecting 972 views of data at variable scan speeds. Offset data (64 views) collected prior to x-ray exposure for correction of actual scan data.
Ranked causes
- 1. Offset data required for x-ray data correction (common)
Verify gantry acceleration to constant scan speed
- Test point
- Gantry status
- Expected
- Gantry accelerated and at constant scan speed
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Axial scan offset data collection radiation
System collects 64 views of offset data from DAS prior to actual x-ray exposure during axial scans to correct x-ray data
Ranked causes
- 1. Offset data collection required for axial scan calibration (common)
Verify 64 offset data views collected
- Test point
- DAS offset data collection
- Expected
- 64 views
- Tools
- System monitoring
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Brush block assembly cleaning caution
Remove and clean brush block assembly and surrounding components using dry Kim wipes and HEPA vacuum cleaner to remove brush powder and metal filings. Required during periodic maintenance.
Ranked causes
- 1. Accumulation of brush powder and metal filings contaminates brushes and slip rings (common)
Power down gantry
- Test point
- Gantry rear base switches
- Expected
- 'Rotate', 'XG Power', 'Slip Ring 115V' switches OFF
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Brush debris cleaning caution
Remove brush powder and debris from gantry covers, rear base, slip ring cover, tracks, detector surface, and mylar ring using vacuum and dry wipes
Ranked causes
- 1. Brush debris and powder accumulation in gantry housing and components (common)
Switch OFF safety switches
- Test point
- Gantry rear base
- Expected
- Rotate, XG Power, Slip Ring 115V switches OFF
9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Calibration: Brush debris cleaning from gantry interior
- Calibration: Brush wear inspection
- Calibration: CD-ROM Drive SCSI Configuration
- Calibration: Cleaning of Brush Block assy
- Calibration: Cradle absolute position (in limit)
- Calibration: Cradle absolute position (out limit)
- Calibration: Cradle compensation (position stability vs table height)
- Calibration: Cradle IN limit
- Calibration: Cradle repeatability (In-to-Out-to-In)
- Calibration: Cradle repeatability (Out-to-In-to-Out)
- Calibration: CRT Monitor Factory Preset Reset
- Calibration: CRT Monitor Horizontal Moire Reduction
- Calibration: CRT monitor input port selection
- Calibration: CRT Monitor Input Select
- Calibration: CRT Monitor Language Selection
- Calibration: CRT Monitor Manual Degaussing
- Calibration: CRT Monitor User Color
- Calibration: CRT monitor user color adjustment
- Calibration: CRT Monitor Vertical Moire Reduction
- Calibration: CRT Monitor Video Input Level
- Calibration: DAS Count Check
- Calibration: DASIFN Board Switch Configuration
- Calibration: DC Power Supplies
- Calibration: DC power supplies (+5V, +12V, -12V)
- Calibration: DC power supplies on rotative frame
- Calibration: DC power supplies on stationary frame
- Calibration: Detector
- Calibration: Detector Beam-on-Window
- Calibration: DG Calibration (DG-CAL)
- Calibration: DG-CAL (Scan Time Gain Correction)
- Calibration: Display Japanese messages on scan screen
- Calibration: Display SMPTE image
- Calibration: DTRF jumper configuration
- Calibration: Fan operation
- Calibration: Filter center alignment and CT number
- Calibration: Foot switch position adjustment
- Calibration: FRONT PNL Power OFF Delay Time
- Calibration: G. Pulse position
- Calibration: Gantry initialize movement
- Calibration: Gantry movement check
- Calibration: Gantry rear base switches
- Calibration: Gap between cradle and cradle support
- Calibration: Gravity SAG
- Calibration: Gravity SAG measurement
- Calibration: Hard Disk Drive 2205368 SCSI Configuration
- Calibration: Hard Disk Drive SCSI Configuration (2205368)
- Calibration: Hard disk drive system software load
- Calibration: Heat Soak
- Calibration: Heat soak procedure
- Calibration: HV Tank flat cable transfer
- Calibration: IMS (Intermediate Support) position (electrical limits)
- Calibration: IMS position (mechanical limits verification)
- Calibration: Install gantry covers
- Calibration: Install option software
- Calibration: Installing NP/NP Plus application software
- Calibration: Installing system software core OS
- Calibration: Keyboard Function Test
- Calibration: KV and MA output verification
- Calibration: Laser positioning lights alignment
- Calibration: Laser positioning lights alignment (NP+)
- Calibration: Longitudinal accuracy (In-to-Out travel)
- Calibration: Longitudinal accuracy (Out-to-In travel)
- Calibration: MA check with load resistor
- Calibration: Memory module (DIMM) installation verification
- Calibration: MOD Drive 2194318 SCSI Configuration
- Calibration: MOD Drive 2198923 SCSI Configuration
- Calibration: MOD Drive SCSI Configuration (2194318)
- Calibration: MOD Drive SCSI Configuration (2198923)
- Calibration: Monitor Function Test
- Calibration: NAA1 assembly audio function
- Calibration: NAA1 Audio Board Input Impedance
- Calibration: NAA1 Audio Input and Speaker Volume
- Calibration: NAA1 Ground Connection
- Calibration: NAA1 Speaker Volume
- Calibration: NAA2 ground connection
- Calibration: NAA2 input impedance
- Calibration: NPR LED test
- Calibration: NPRIF Board Jumper Configuration
- Calibration: NPRM Board Memory and NPRS Configuration
- Calibration: NPRS Board Jumper Settings
- Calibration: OC power supplies (+12VDC, -12VDC, +5VDC)
- Calibration: OC Power-up Sequence
- Calibration: OGP DIP switches
- Calibration: Operator Console jumper and switch settings
- Calibration: Operator Console Power Supply Ground Connection
- Calibration: Operator Console power-up sequence verification
- Calibration: Parallel setting of cradle
- Calibration: Patient intercom volume adjustment
- Calibration: PCI Host Card Connector Configuration
- Calibration: PCI Host Card Jumper Configuration
- Calibration: PCI tray installation verification
- Calibration: Photo sensor
- Calibration: Photo sensor and plate alignment
- Calibration: Plane of Rotation
- Calibration: Positioning lights halogen top light right-left alignment
- Calibration: Power Off Delay Time
- Calibration: Powering off optional Pioneer MOD drive (LFC only)
- Calibration: Q Calibration (Q-CAL)
- Calibration: Q-CAL (Focus Position)
- Calibration: Radial Alignment
- Calibration: Rating plate attachment
- Calibration: REAR CN1 Fan Alarm Enable/Disable
- Calibration: REAR CN1 Intercom Minimum Volume
- Calibration: REAR CN1 Safety Loop Control
- Calibration: REAR CN1 Talk On Switch
- Calibration: REAR CN1 TGP Communication
- Calibration: Recon cycle verification
- Calibration: Remove gantry rear cover
- Calibration: Remove OC covers
- Calibration: Remove PDU covers
- Calibration: RF Shoe position and alignment
- Calibration: Rotation speed
- Calibration: Rotation velocity offset
- Calibration: Rotational mechanism clearance
- Calibration: Saving patient images to MOD
- Calibration: Scout scan cradle positioning and data collection
- Calibration: Scout scan data collection
- Calibration: Scout scan offset data collection
- Calibration: Screw tightening torque
- Calibration: Serial Port Expander SCSI Configuration (ST-1800)
- Calibration: Serial Port Expander ST-1800 Configuration
- Calibration: Set gantry safety switches
- Calibration: Signal and power brush wear inspection
- Calibration: Signal brush contact verification
- Calibration: Slip ring cleaning
- Calibration: Slip ring cleaning for noise reduction
- Calibration: Slip ring radial position
- Calibration: Software installation and system configuration
- Calibration: Software loading preparation
- Calibration: Speaker assembly audio function
- Calibration: SUB BD service switches
- Calibration: System configuration
- Calibration: System configuration and desktop menu
- Calibration: System module LED verification
- Calibration: System revision upgrade procedure
- Calibration: System shutdown before OS installation
- Calibration: Table board IMS jumper configuration
- Calibration: Table DOWN limit
- Calibration: Table positioning in right-left direction
- Calibration: Table positioning right-left alignment
- Calibration: TGP board DIP switches S1
- Calibration: TGP board DIP switches S2
- Calibration: TGP board DIP switches S3
- Calibration: TGP board jumpers
- Calibration: Tube Calibration
- Calibration: Unlatch switches
- Calibration: Workstation Default Test
- Calibration: X-ray tube automatic learning
- Calibration: X-ray tube DAS count verification
- Calibration: X-ray tube warm-up
- Calibration: XT Calibration (XT-CAL)
- Calibration: XT-CAL (Crosstalk Compensation)
- Capacitor assembly screws loose
- Capacitor explosion hazard in gantry
- Capacitor Inverter Assembly failure
- Collimator filter requires replacement
- Component assembly procedure: Primary coil flat cable connection
- Cooling check failure – COOLING OVER LIMIT message
- Cradle stops before reaching scan position
- CRT monitor manual degaussing for color correction
- CRT monitor reset to factory preset
- DAS data transfer errors occur
- Detector and collimator/filter maintenance note
- DTRF assembly requires replacement
- Dust filter condition not checked
- Electrical hazard during gantry service
- Electrocution hazard from gantry slip rings
- Electromagnetic interference from other equipment affecting CT system
- Electrostatic discharge damage to circuit boards
- Electrostatic discharge precaution during board service
- EMC Board 3 Phases failure
- Emergency stop function not tested
- Emergency stop switch does not halt X-ray
- External landmark function malfunction
- Eye damage from laser positioning light
- Fan operation check
- Foot pedal contact with front cover
- Functional check: Table and gantry interlock
- Functional check: Table and gantry panel switches
- Gantry and rotor do not stop after End Exam
- Gantry cannot rotate with covers removed
- Gantry rotation hazard with wrong installation sequence
- Hardware inventory check
- HV cable (anode/cathode) failure
- HV Tank replacement required
- HV Tank temperature overheat condition
- Initialization and safety checks before adjustment
- Inverter Coils Mid Power Assembly failure
- Inverter Mid Power Assembly replacement required
- Keyboard function test - 103 keyboard input
- Keyboard function test - input response
- Keyboard function test - LED response
- KV Control Board failure
- Laser positioning light malfunction
- Load distribution and DC supply voltage not verified
- Monitor function test - color output
- Monitor function test - display pattern
- Move to Scan button does not blink after confirm
- Nonvolatile Parameters RAM requires replacement
- OC filter clogged or dirty
- OGP assembly requires replacement
- Operator Console power-up fails or stops during self-test
- Optical fiber cable degradation
- Optical fiber cable disconnection
- Pause function does not stop scan at next ISD
- Perform functional scan test after memory installation
- Personal injury from loose clothing or jewelry during gantry servicing
- Power distribution unit input voltage not verified
- Power supply PS L module malfunction
- Power system lockout procedure
- Pre-adjustment setup: Access power supplies and speaker panel
- Pre-adjustment setup: Remove operator console covers
- QA phantom quality assessment check
- Replace detector assembly
- RF transmitter assembly requires replacement
- Rotation Board High Speed requires replacement
- Rotation Capacitors Set 23/23 requires replacement
- Safety loop relay test and reset
- Scout scan phantom diameter measurement incorrect
- Scout scan tilted gantry not inhibited
- SCSI test - storage device access verification
- Slip ring damage from contamination
- Software upgrade not installed
- Stop Scan does not immediately halt
- Stud bolt installation with locktite
- System installation checks not completed
- System shutdown test
- Thermal burn hazard from X-ray tube
- Top positioning light replacement
- Verify main memory size after installation
- Warm-up error detected
- Warm-up scan does not complete
- Workstation default test - Audio
- Workstation default test - Monitor
- Workstation default test - Mouse
- X-ray generator power and gantry servo voltage not verified
- X-ray ON indicator or alarm not functioning
- X-ray radiation exposure during service
- X-ray tube overheat protection not functioning
- XG Error 371 during continuous scanning
- XG ERROR 371 occurs during helical scanning
241 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
100xx | External communication errors | Communication failure with external system or network | unknown |
110xx | Thermal errors | Generator or tube temperature limit reached | warning |
30xx | Tube spits errors | Errors generated by X-ray tube during operation | unknown |
40xx | Rotation errors | Rotation function malfunction or failure | unknown |
50xx | Heater errors | Heater function malfunction or failure | unknown |
60xx | Exposure errors | HV inverter, mA measurement, or exposure control malfunction | unknown |
70xx | Power supply errors | Low voltage supply or DC bus voltage failure | unknown |
80xx | Hardware errors | Internal communication failure or cable fault | unknown |
90xx | Application errors | Saved RAM corruption or software failure | unknown |
ALERT: SCSI hard error on (1,3) | SCSI hard drive error reported on controller 1, drive 3 | Occurs during OS installation on systems with DASM (Direct Attached Storage Module) until configuration setting is completed | warning |
CHANGE | Adjustment data not stored during gantry or table adjustment | Calibration or adjustment operation fails to save data | warning |
CONERR | FPGA Configuration error occurred | FPGA configuration fails or becomes corrupted | error |
ERRG | Gantry rotation operation error detected | Gantry rotation fails to respond or completes incorrectly | error |
ERRM | Processor communication or scan operation error detected | Internal processor communication failure or scan operation error occurs | error |
ERRT | Table or tilt operation error detected | Table or gantry tilt fails to respond or completes incorrectly | error |
FECERR | FEC encode error detected | Forward Error Correction encoding fails during data transmission | warning |
SVALM | Servo amplifier in alarm state or AC 200V power loss | Servo amplifier detects fault or loses mains AC power supply | error |
WARM-UP ERROR | Warm-up scan data out of acceptable range | Data collected during warm-up scan sequence falls outside specified range limits | warning |
XG ERROR 371 | Scan cancellation error in XG CPU systems | Helical scanning with 140kV/160mA/10mm or larger thickness/90 sec or longer scan time on 90 slice option systems with XG CPU firmware V3.10 or lower | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.