GE Healthcare CT HiSpeed: troubleshooting and error codes

The ServiceForge knowledge base holds 301 documented symptoms for the GE Healthcare CT HiSpeed, with 363 ranked causes, 1051 diagnostic checks with test points and expected values, and 19 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

HV chain fault detection high voltage

Fast and slow safeties monitor inverter performance; fast safeties cut HV power on error; slow safeties measure signals every 1 ms for fault detection

Ranked causes

  1. 1. Over kV condition (common)
  2. 2. No kV condition (common)
  3. 3. kV regulation error at 7th error in exposure (occasional)
  4. 4. Anode spit exceeds max allowed count (occasional)
  5. 5. Cathode spit exceeds max allowed count (occasional)
  6. 6. Max resonant current exceeded (occasional)
  7. 7. Gate supply out of range (occasional)
First check

Monitor kV measurement

Test point
kV measure signal
Expected
Within safe operating range

3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

TGP board does not initialize on power-on

After powering on or resetting TGP board, processor fails to complete ROM and RAM tests; ERRG, LPG, RQG, ERRT, LPT, RQT, MNT, ERRM, LPM, RQM, or MNM LEDs light and remain on

Ranked causes

  1. 1. Gantry processor RAM check failure (U87, U92, U77) (rare)
  2. 2. Gantry processor ROM check failure (U112) (rare)
  3. 3. Table processor RAM check failure (U42, U64, U48) (rare)
  4. 4. Table processor ROM check failure (U68) (rare)
  5. 5. Management processor RAM check failure (U88, U62, U48, U77) (rare)
  6. 6. Management processor ROM check failure (U111) (rare)
First check

Observe gantry processor LED status on TGP board after power-on

Test point
ERRG, LPG, RQG LEDs on TGP board
Expected
LPG LED should blink after successful RAM and ROM checks; ERRG LED indicates hang up state

2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Miscellaneous Hardware Tests patient safety

Operator Console miscellaneous tests including Hardware Inventory, SCSI Test, System Log viewing, NPR LED Test, Safety Loop test, and Shutdown test to verify system configuration and relay operation.

Ranked causes

  1. 1. Hardware configuration change or SCSI device malfunction (occasional)
  2. 2. NPR LED display malfunction (rare)
  3. 3. Safety loop relay malfunction (occasional)
  4. 4. System shutdown failure (rare)
First check

Access Misc Test from diagnostics menu

Test point
H/W Diagnostics menu
Expected
Misc test selection window appears after selecting [Service] -> [Diagnostics & analysis] -> [H/W Diagnostics] -> [Interactive Test] -> [Misc Test]

6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Electromagnetic interference with other equipment caution

CT system may cause EMC issues affecting connected or nearby electronic devices through air or cables

Ranked causes

  1. 1. Improper installation or cable routing (common)
  2. 2. System covers left open during operation (occasional)
  3. 3. Loose fasteners or damaged gaskets (occasional)
  4. 4. Unauthorized modifications to system (rare)
First check

Verify installation location distance from other equipment

Test point
General equipment area
Expected
Equipment positioned as far as possible from other electronic devices

5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Positioning lights halogen horizontal alignment (right-left) caution

Inner and outer right-left lights horizontally misaligned; top light horizontal misalignment; vertical light offset from phantom surface

Ranked causes

  1. 1. Inner right-left light horizontal adjustment screw (B) out of tolerance (common)
  2. 2. Outer right-left light horizontal adjustment screw (B) out of tolerance (common)
  3. 3. Top light horizontal adjustment screw (B) out of tolerance (common)
First check

Verify inner right-left light horizontal alignment

Test point
Vertical positioning light relative to phantom surface
Expected
Vertical light and phantom surface aligned within ±0.5 mm, ±0.5 mm
Tools
Positioning lights, adjustment screw B, lever D, water phantom

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CAM amplifier linearity out of specification

Detector channel response linearity fails verification during no-X-ray test scans

Ranked causes

  1. 1. Amplifier gain drift or non-linearity (occasional)
  2. 2. Detector channel response mismatch (occasional)
  3. 3. Insufficient warm-up before test (common)
First check

Verify vector file 7 linearity ratio within tolerance

Test point
View Vector plot for Vector File No. 7
Expected
0.9995 < ratio < 1.0005

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Component ejection hazard during gantry rotation with covers removed patient safety

Risk of components or objects being ejected at high velocity when gantry rotates with side covers removed

Ranked causes

  1. 1. Side covers removed during gantry operation or rotation (common)
  2. 2. Board chassis left open during scanning (occasional)
  3. 3. Components not properly secured with correct torque (occasional)
First check

Inspect gantry side covers for removal or damage

Test point
Right and left side covers of gantry
Expected
Covers installed and secured in place

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

DAS offset data out of specification

Offset data averages outside acceptable range without offset file correction

Ranked causes

  1. 1. Detector or DAS channel drift or malfunction (unknown)
  2. 2. Offset file correction needed (common)
  3. 3. Detector response drift after warm-up (occasional)
First check

Verify Avg of Mean of Active Views without offset correction

Test point
Fan Data Averages display
Expected
3000 ≤ Avg of Mean ≤ 6000

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Electrocution hazard from 400VAC slip ring supply high voltage

Risk of electrocution when servicing X-ray generator or slip ring components due to active 400VAC supply

Ranked causes

  1. 1. XG Power Switch (400VAC) not turned OFF before service (common)
  2. 2. CB1 and CB2 circuit breakers not turned OFF (common)
  3. 3. Capacitors in XG units not discharged (occasional)
First check

Verify XG Power Switch is OFF

Test point
XG Power Switch at gantry rear base
Expected
Switch in OFF position; System Safety Loop open

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gantry rotates unintentionally during service patient safety

Gantry continues rotating or can rotate when it should be disabled, posing impact hazard to personnel

Ranked causes

  1. 1. Axial Drive Switch not turned OFF (common)
  2. 2. Right maintenance cover removed without reinstalling before other covers installed (occasional)
  3. 3. Service brackets not engaged when covers removed (occasional)
First check

Verify Axial Drive Enable Buzzer status

Test point
Gantry rear base
Expected
Buzzer is silent

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gantry rotation fails during power-on test patient safety

Gantry does not rotate or rotation check terminates prematurely with ERRG LED lighting during power-on initialization

Ranked causes

  1. 1. Servo amplifier not powered (common)
  2. 2. Service switch (SW1) set to OFF (occasional)
  3. 3. Gantry azimuth counter initialization timeout (> 25 sec) (occasional)
First check

Verify service switch SW1 position on TGP board

Test point
Service switch SW1 on TGP board
Expected
Set to SYS (normal) or MNL (manual)

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Memory module installation and configuration incomplete

System not properly configured after installing IPU2 board and memory modules; requires MD EXT switch settings and P-ON test verification before operation

Ranked causes

  1. 1. MD EXT switch (S7A1) not set to correct configuration for kit type (common)
  2. 2. IPU2 board not installed in OC nest slot 2 (common)
  3. 3. P-ON test level 2 not run to verify system startup (common)
First check

Run P-ON test level 2 to verify system startup

Test point
CPW Board slot 1, Dip SW 8A1
Expected
Startup screen appears after approximately 14 minutes with no errors

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power line voltage disturbances not monitored high voltage

Input power line voltage requires monitoring for mean voltage, surges, sags, impulses, and frequency stability during installation to detect correlation with system malfunctions

Ranked causes

  1. 1. Line voltage out of specification or unstable (occasional)
  2. 2. Mean line voltage outside acceptable range (rare)
  3. 3. Line frequency drifting outside tolerance (rare)
First check

Measure current line voltage at PDU input terminals L1, L2, L3

Test point
Power Distribution Box terminals L1, L2, L3
Expected
200/208/380/400/415/440/460/480 VAC at 50/60 Hz
Tools
Line analyzer (Dranetz 606-3 or equivalent)

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Software loading: Load-From-Cold (LFC) procedure

Full system software loading (OS + NP application) to system hard disk from CD-ROMs and MODs. Reinitializes all system data disks, erasing all images and scan data.

Ranked causes

  1. 1. System software update or downgrade required (common)
  2. 2. Patch software available for NP application (occasional)
  3. 3. System downgrade to older version (rare)
First check

All patient images and scan data saved to MOD

Test point
MOD optical disk
Expected
Images and raw data archived successfully

10 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature environment out of specification caution

Scan room and equipment room temperature and humidity do not meet installation specifications; excessive temperature may overstress components

Ranked causes

  1. 1. Scan room temperature outside range 20–28 °C (common)
  2. 2. Equipment room temperature outside range 15–30 °C (occasional)
  3. 3. Relative humidity outside safe range (occasional)
First check

Record scan room temperature

Test point
Scan room with thermometer/recorder
Expected
20 °C to 28 °C with at least 5 °C/hr gradient, to 28 °C
Tools
Thermometer, temperature/humidity recorder

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

X-ray exposure risk during improper equipment operation radiation

Potential radiation injury to operators and bystanders from inadequate shielding or lack of protective precautions during CT scanner use and maintenance

Ranked causes

  1. 1. Operator or technician not trained in radiation safety protocols (common)
  2. 2. Inadequate use of available protective materials and devices (common)
  3. 3. Unqualified personnel performing electrical or maintenance work (occasional)
First check

Verify radiation safety training credentials for all personnel

Test point
Training documentation
Expected
Current NCRP and ICRP training certification on file

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Abnormal gantry rotation noise or movement caution

Gantry movement is not smooth or abnormal noise detected during manual rotation

Ranked causes

  1. 1. Brush Block Assy not properly restored to original configuration (common)
  2. 2. Brush assembly installation issue (occasional)
First check

Manual gantry rotation test

Test point
Gantry
Expected
Smooth movement with no abnormal noise through at least one complete rotation

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Audio Function Test

Operator Console audio functional check including OC and table speaker tests for Auto Voice, CD player, X-ray ON buzzer, and Alert. Tests verify volume control and sound output.

Ranked causes

  1. 1. Audio output or volume control malfunction (occasional)
  2. 2. Auto Voice Playback still active (occasional)
First check

Access Audio Function Test from diagnostics menu

Test point
H/W Diagnostics menu
Expected
Audio test screen appears after selecting [Service] -> [Diagnostics & analysis] -> [H/W Diagnostics] -> [Interactive Test] -> [Audio Test]

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Azimuth belt tension caution

Adjust axial belt tension to specified gap between tensioner components to maintain drive without over-tensioning

Ranked causes

  1. 1. Belt tension incorrect or over-tensioned (common)
  2. 2. Belt over-tensioned due to incorrect adjustment (occasional)
First check

Loosen mounting and adjustment hardware

Test point
Mounting plate bolts and set nut
Expected
All bolts and nut loose enough to allow adjustment

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: CAM Amplifier Linearity radiation

Check for non-linearity in detector amplification response across different tube currents. Performed after X-ray tube, detector, DAS, or CAM board replacement.

Ranked causes

  1. 1. Detector or amplifier non-linearity after component replacement (common)
  2. 2. Detector crystal response drift (occasional)
First check

Perform 100 mA air stationary scan

Test point
Offline Scan service menu
Expected
Scan 1: Azimuth 0°, Thickness 10 mm, Focus Large, KV 120, MA 100 mA

5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Halogen positioning lights alignment patient safety

Five halogen positioning lights (inner right/left, outer right/left, outer top) must be aligned with scan center and phantom surface for accurate patient positioning

Ranked causes

  1. 1. Positioning lights misaligned vertically or horizontally (common)
  2. 2. Water phantom surface not parallel to scan plane (occasional)
First check

Place fine pin at phantom holder center and position perpendicular to scan plane

Test point
M3 hexagonal wrench at phantom holder center
Expected
Pin aligned with vertical line in magnified (x4) image

7 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Positioning lights halogen vertical alignment (right-left) caution

Inner and outer right-left positioning lights vertically misaligned with pin; horizontal light offset from pin image

Ranked causes

  1. 1. Inner right-left light vertical adjustment screw (A) out of tolerance (common)
  2. 2. Outer right-left light vertical adjustment screw (A) out of tolerance (common)
First check

Position pin perpendicular to scan plane

Test point
Pin image versus vertical reference line
Expected
Pin and vertical line aligned within ±0.5 mm, ±0.5 mm
Tools
M3 hexagonal wrench or fine pin, phantom holder, cursor/magnification tools

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Remove gantry front cover caution

Access to internal gantry components by removing front cover with service angles and connector management

Ranked causes

  1. 1. Front cover removal procedure for standard installation (common)
  2. 2. Front cover removal for small room (short-footprint mode) (occasional)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Remove gantry side covers caution

Access to side gantry components by removing left and right side covers with connector disconnection as needed

Ranked causes

  1. 1. Left side cover removal (common)
  2. 2. Right side cover removal (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: RMT CNT jumper configuration

Setting JP1 jumper plug on RMT CNT (Remote Control) module according to TGP board part number and revision

Ranked causes

  1. 1. RMT CNT part 2175741 with TGP 2156510-3 or earlier (common)
  2. 2. RMT CNT part 2175741-2 or later with TGP 2156510-4 or later (common)
First check

Identify RMT CNT part number

Test point
RMT CNT module label
Expected
Part No. 2175741 or 2175741-2 or later

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Rotation balance caution

Achieve balance of rotating gantry assembly through rough and fine weight adjustment to minimize resistance and ensure smooth rotation

Ranked causes

  1. 1. Gantry unbalanced in CW or CCW direction (common)
  2. 2. Main weight unbalanced (common)
First check

Rough balance check with main weight at 6 o'clock

Test point
Main weight position
Expected
Gantry does not rotate on its own

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Saving system state data

Save system configuration, preferences, calibration files, and operational data to MOD before performing software installation to enable restoration after LFC/LFW procedure.

Ranked causes

  1. 1. System state data must be preserved before software re-initialization (common)
  2. 2. Option key information needs backup (common)
First check

Verify system powered ON

Test point
Main switch of OC
Expected
System on and main screen displayed

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Emergency stop switch not accessible or not activated during emergency patient safety

Inability to stop gantry rotation quickly during emergency situation due to inaccessible or misused stop switches

Ranked causes

  1. 1. Using Stop Scan switch instead of Emergency Stop switch (common)
  2. 2. Emergency stop switches on gantry frame not connected when covers removed (occasional)
First check

Identify correct Emergency Stop switch location

Test point
Operator console or gantry
Expected
Emergency Stop Switch clearly visible and accessible (not Stop Scan switch on keyboard)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gantry rotation check not performed patient safety

Gantry mechanical rotation, absence of interference, and proper home position return must be verified before operation

Ranked causes

  1. 1. Gantry mechanical obstruction or malfunction (occasional)
  2. 2. Gantry fails to stop at home position (rare)
First check

Switch OFF Rotate switch

Test point
Gantry rear base Rotate switch
Expected
Switch in OFF position

10 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gantry tilt or table up/down relay malfunction patient safety

Relay circuit designed with dual series relays fails to properly disable pump/valve mechanism when relay should be open; relay remains closed or fails to activate on command

Ranked causes

  1. 1. Direct relay stuck closed (enable relay fails to isolate when open) (occasional)
  2. 2. Enable relay stuck closed (processor cannot command relay to open) (occasional)
First check

Activate enable relay and verify pump/valve mechanism does not operate

Test point
Gantry tilt and table up/down pump/valve mechanism
Expected
No operation; no fluid flow; no movement

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gantry tilting hazard during service with covers loose patient safety

Risk of gantry falling or unexpected movement during tilting if covers and support mechanisms are not properly secured

Ranked causes

  1. 1. Covers not secured before tilting gantry (common)
  2. 2. Azimuth lock pin not used to hold gantry during component removal (common)
First check

Verify all gantry covers are secured

Test point
Side covers, front cover, rear cover, tube hoist rail
Expected
All covers installed, latched, and secure

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Management processor communications failure

Management processor unable to communicate with operator console or OGP board on gantry rotative frame; IMS (Intermediate Support) not responding

Ranked causes

  1. 1. Communication interface fault between management processor and console/OGP board (unknown)
  2. 2. IMS (Intermediate Support) interface fault (unknown)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Memory capacity verification required after upgrade

System memory must be validated through scanning, paging, and post-reconstruction tests to ensure proper function after memory module installation

Ranked causes

  1. 1. Memory not verified after module installation (common)
  2. 2. Post reconstruction test not completed (common)
First check

Run POST RECON test with image subtraction

Test point
Image processing, Subtraction function
Expected
Subtracted image (A.B.C;1 minus A.B.C;2) shows nil image at Window Width 2, Window Level 0

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

OGP ROM not verified or updated

SP ROM on OGP board requires verification of part number and potential replacement with newer version to ensure gantry operation compatibility

Ranked causes

  1. 1. OGP ROM part number is 2121676-2 or lower (outdated) (occasional)
  2. 2. OGP ROM part number is 2121676-3 or later (current) (occasional)
First check

Verify SP ROM part number on OGP board at location U15E

Test point
OGP board, location U15E
Expected
Part number 2121676-3 or later (current version) part number

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Signal brush contact failure or intermittency patient safety

Signal brushes or slip rings show poor electrical contact causing signal loss or intermittent data transmission during gantry rotation

Ranked causes

  1. 1. Brush wear below check line, reduced contact pressure (common)
  2. 2. Slip ring contamination, dust, or grease coating reducing contact (common)
First check

Perform signal brush contact voltage check on problematic ring pair

Test point
Oscilloscope measurement across 120 Ω resistor while rotating
Expected
No voltage drops greater than 0.5 V, smooth continuous signal without fluctuations, +/- 0.5 V
Tools
Digital storage oscilloscope in envelope mode, Ring Service Tool (2206131)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Software loading: Load-From-Warm (LFW) procedure

NP application loading only from CD-ROMs and MODs to system hard disk. System remains powered on. Does NOT reinitialize system data disks.

Ranked causes

  1. 1. NP application software update required (no OS change) (common)
  2. 2. NP patch software available (occasional)
First check

NP application software installation in LFW mode completed

Test point
Display after automatic reboot
Expected
IRIS main menu appears without OS reinstall

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

X-ray generator KV stabilization and mA linearity check radiation

Verify KV and mA output remain within specification across multiple stations during functional check

Ranked causes

  1. 1. KV deviation exceeds ±5 kV specification (occasional)
  2. 2. mA deviation exceeds ±20 mA specification (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

AC/DC 3 Phase Mid Power Assembly failure high voltage

DC bus voltage loss; rectifier block malfunction; AC/DC conversion failure; board overheating or component failure

Ranked causes

  1. 1. AC/DC board electronic failure or rectifier block failure (unknown)
First check

DC bus voltage

Test point
+DC and −DC terminals on board
Expected
Correct DC voltage per system specification
Tools
DMM

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Access XG CPU board for firmware check caution

Procedural steps to safely access CPU board in electronic rack for inspection

Ranked causes

  1. 1. Routine maintenance procedure for firmware version verification (unknown)
First check

Turn Rotate, XG, Slip Ring switches OFF

Test point
Rear base of Gantry
Expected
All three switches in OFF position

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Aperture assembly requires replacement caution

Aperture assembly needs removal and installation with ratio check verification

Ranked causes

  1. 1. Aperture assembly wear or malfunction (unknown)
First check

Set gantry to 0° tilt and HOME position

Test point
Gantry tilt angle and position
Expected
0° tilt, HOME position

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Aperture motor driver failure

Aperture motor driver not responding or not controlling aperture position

Ranked causes

  1. 1. Defective aperture motor driver assembly (common)
First check

Verify aperture motor driver response to control signals

Test point
Aperture Motor Driver Assy connectors
Expected
Aperture opens and closes smoothly in response to commands

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audio function test - Gantry speaker Auto Voice and CD player

Verify Gantry speaker outputs Auto Voice and CD player audio with volume control

Ranked causes

  1. 1. Gantry speaker or audio output circuit failure (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audio function test - OC speaker alert

Verify alert sound outputs once and alert width and frequency controls function

Ranked causes

  1. 1. Alert sound circuit or audio output circuit failure (unknown)
First check

Alert sound output

Test point
OC speaker
Expected
Alert sounds once when [ON] button clicked once

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audio function test - OC speaker Auto Voice

Verify OC speaker outputs Auto Voice message and volume controls function

Ranked causes

  1. 1. OC speaker or audio output circuit failure (unknown)
First check

Auto Voice message output and volume control

Test point
OC speaker
Expected
Auto Voice message audible; volume adjustable via L and R sliders

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audio function test - OC speaker CD player

Verify OC CD player outputs audio and volume controls function

Ranked causes

  1. 1. OC CD player or audio output circuit failure (unknown)
First check

CD player audio output and volume control

Test point
OC speaker
Expected
CD audio audible when music CD played; volume adjustable via L and R sliders

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audio function test - OC speaker X-ray ON buzzer patient safety

Verify X-ray ON buzzer sounds and buzzer modulation and frequency controls function

Ranked causes

  1. 1. X-ray ON buzzer or audio output circuit failure (unknown)
First check

X-ray ON buzzer sound continuity

Test point
OC speaker
Expected
Buzzer sounds continuously while [ON/OFF] button is ON; stops when OFF

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Auto voice not heard or unintelligible patient safety

Autovoice announcements for scan start or end not heard or not understandable

Ranked causes

  1. 1. Audio output circuit malfunction or speaker failure (unknown)
First check

Listen for autovoice scan start announcement

Test point
Audio speaker output
Expected
Clear, understandable audio announcement for scan start

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Axial drive belt requires replacement caution

Axial drive belt needs removal and installation after removing covers and setting azimuth lock pin

Ranked causes

  1. 1. Axial drive belt wear or damage (unknown)
First check

Switch OFF Rotate, XG Power, Slip Ring115V switches

Test point
Gantry rear base switches
Expected
All three switches OFF

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Axial scan does not complete or errors occur

Axial scan with specified parameters fails to complete without error

Ranked causes

  1. 1. DAS/Detector malfunction or X-ray generator instability (unknown)
First check

Perform axial scan with test parameters

Test point
Detector and X-ray generator during scan
Expected
Scan completes without error kV, as set

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Balun assembly requires replacement caution

Balun assembly with coil needs removal and installation with data transfer check and gantry restoration

Ranked causes

  1. 1. Balun assembly failure or coil damage (unknown)
First check

Switch OFF Rotate, XG Power, Slip Ring115V switches

Test point
Gantry rear base switches
Expected
All three switches OFF

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Brush block assembly misalignment caution

Brushes not centered between slip ring track barriers after installation; potential contact issues

Ranked causes

  1. 1. Brush Block Assy pin not aligned to corresponding hole (common)
First check

Visual inspection of brush position

Test point
Brush Block Assy
Expected
Brushes centered between slip ring track barriers

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Air and Phantom Calibration Sequence radiation

Automated sequential execution of Air Calibration followed by Phantom Calibration using calibration station phantoms. Establishes baseline CT number reference values. Precondition: bow-tie filter must be installed. Performed via Auto Sequence 1.

Ranked causes

  1. 1. Air and phantom calibration required after Hilight calibration or major component replacement (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Air and Phantom Calibrations radiation

Establish baseline detector response for water and air reference standards. Bow-tie filter must be installed. Performed after Hilight calibrations (Q, XT, AV, DG as required). Includes Auto Sequence 1 (Air Cal → Phantom Cal).

Ranked causes

  1. 1. Detector sensitivity baseline drift after calibration-triggering component replacement or alignment (common)
First check

Verify bow-tie filter is installed on collimator

Test point
Gantry collimator assembly
Expected
Bow-tie filter present and properly seated

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AV Calibration (AV-CAL) radiation

Compensate for Lumex crystal afterglow persistence following X-ray pulse cessation. Performed when detector or alignment is changed. Takes ~2 minutes. Gantry does not rotate.

Ranked causes

  1. 1. Detector afterglow (phosphorescence) from Lumex crystals after detector replacement or alignment (common)
First check

Verify X-ray tube is sufficiently cool before starting

Test point
Tube case temperature indicator
Expected
Below 10% heat level

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AV-CAL (Afterglow Compensation) radiation

Create After Glow Vector file to compensate for persistent light emission from Lumex crystal after x-ray radiation stops. Gantry does not rotate during this procedure. Approximately 2 minutes duration.

Ranked causes

  1. 1. Detector afterglow artifact after detector replacement or beam-on-window alignment (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Axial scan data collection parameters radiation

System collects 972 views of actual data during axial scan at view period synchronized with gantry azimuth; scan speeds vary 0.8 to 3.0 sec

Ranked causes

  1. 1. Axial scan data collection with speed-dependent view period (common)
First check

Verify correct view period for selected scan speed

Test point
Scan speed setting
Expected
0.8 sec: 823 µsec/view; 1.0 sec: 1029 µsec/view; 1.5 sec: 1543 µsec/view; 2.0 sec: 2058 µsec/view; 3.0 sec: 3086 µsec/view
Tools
System configuration

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Axial scan geometry and offset correction radiation

Axial scans perform 360-degree rotation collecting 972 views of data at variable scan speeds. Offset data (64 views) collected prior to x-ray exposure for correction of actual scan data.

Ranked causes

  1. 1. Offset data required for x-ray data correction (common)
First check

Verify gantry acceleration to constant scan speed

Test point
Gantry status
Expected
Gantry accelerated and at constant scan speed

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Axial scan offset data collection radiation

System collects 64 views of offset data from DAS prior to actual x-ray exposure during axial scans to correct x-ray data

Ranked causes

  1. 1. Offset data collection required for axial scan calibration (common)
First check

Verify 64 offset data views collected

Test point
DAS offset data collection
Expected
64 views
Tools
System monitoring

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Brush block assembly cleaning caution

Remove and clean brush block assembly and surrounding components using dry Kim wipes and HEPA vacuum cleaner to remove brush powder and metal filings. Required during periodic maintenance.

Ranked causes

  1. 1. Accumulation of brush powder and metal filings contaminates brushes and slip rings (common)
First check

Power down gantry

Test point
Gantry rear base switches
Expected
'Rotate', 'XG Power', 'Slip Ring 115V' switches OFF

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Brush debris cleaning caution

Remove brush powder and debris from gantry covers, rear base, slip ring cover, tracks, detector surface, and mylar ring using vacuum and dry wipes

Ranked causes

  1. 1. Brush debris and powder accumulation in gantry housing and components (common)
First check

Switch OFF safety switches

Test point
Gantry rear base
Expected
Rotate, XG Power, Slip Ring 115V switches OFF

9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Calibration: Brush debris cleaning from gantry interior
  • Calibration: Brush wear inspection
  • Calibration: CD-ROM Drive SCSI Configuration
  • Calibration: Cleaning of Brush Block assy
  • Calibration: Cradle absolute position (in limit)
  • Calibration: Cradle absolute position (out limit)
  • Calibration: Cradle compensation (position stability vs table height)
  • Calibration: Cradle IN limit
  • Calibration: Cradle repeatability (In-to-Out-to-In)
  • Calibration: Cradle repeatability (Out-to-In-to-Out)
  • Calibration: CRT Monitor Factory Preset Reset
  • Calibration: CRT Monitor Horizontal Moire Reduction
  • Calibration: CRT monitor input port selection
  • Calibration: CRT Monitor Input Select
  • Calibration: CRT Monitor Language Selection
  • Calibration: CRT Monitor Manual Degaussing
  • Calibration: CRT Monitor User Color
  • Calibration: CRT monitor user color adjustment
  • Calibration: CRT Monitor Vertical Moire Reduction
  • Calibration: CRT Monitor Video Input Level
  • Calibration: DAS Count Check
  • Calibration: DASIFN Board Switch Configuration
  • Calibration: DC Power Supplies
  • Calibration: DC power supplies (+5V, +12V, -12V)
  • Calibration: DC power supplies on rotative frame
  • Calibration: DC power supplies on stationary frame
  • Calibration: Detector
  • Calibration: Detector Beam-on-Window
  • Calibration: DG Calibration (DG-CAL)
  • Calibration: DG-CAL (Scan Time Gain Correction)
  • Calibration: Display Japanese messages on scan screen
  • Calibration: Display SMPTE image
  • Calibration: DTRF jumper configuration
  • Calibration: Fan operation
  • Calibration: Filter center alignment and CT number
  • Calibration: Foot switch position adjustment
  • Calibration: FRONT PNL Power OFF Delay Time
  • Calibration: G. Pulse position
  • Calibration: Gantry initialize movement
  • Calibration: Gantry movement check
  • Calibration: Gantry rear base switches
  • Calibration: Gap between cradle and cradle support
  • Calibration: Gravity SAG
  • Calibration: Gravity SAG measurement
  • Calibration: Hard Disk Drive 2205368 SCSI Configuration
  • Calibration: Hard Disk Drive SCSI Configuration (2205368)
  • Calibration: Hard disk drive system software load
  • Calibration: Heat Soak
  • Calibration: Heat soak procedure
  • Calibration: HV Tank flat cable transfer
  • Calibration: IMS (Intermediate Support) position (electrical limits)
  • Calibration: IMS position (mechanical limits verification)
  • Calibration: Install gantry covers
  • Calibration: Install option software
  • Calibration: Installing NP/NP Plus application software
  • Calibration: Installing system software core OS
  • Calibration: Keyboard Function Test
  • Calibration: KV and MA output verification
  • Calibration: Laser positioning lights alignment
  • Calibration: Laser positioning lights alignment (NP+)
  • Calibration: Longitudinal accuracy (In-to-Out travel)
  • Calibration: Longitudinal accuracy (Out-to-In travel)
  • Calibration: MA check with load resistor
  • Calibration: Memory module (DIMM) installation verification
  • Calibration: MOD Drive 2194318 SCSI Configuration
  • Calibration: MOD Drive 2198923 SCSI Configuration
  • Calibration: MOD Drive SCSI Configuration (2194318)
  • Calibration: MOD Drive SCSI Configuration (2198923)
  • Calibration: Monitor Function Test
  • Calibration: NAA1 assembly audio function
  • Calibration: NAA1 Audio Board Input Impedance
  • Calibration: NAA1 Audio Input and Speaker Volume
  • Calibration: NAA1 Ground Connection
  • Calibration: NAA1 Speaker Volume
  • Calibration: NAA2 ground connection
  • Calibration: NAA2 input impedance
  • Calibration: NPR LED test
  • Calibration: NPRIF Board Jumper Configuration
  • Calibration: NPRM Board Memory and NPRS Configuration
  • Calibration: NPRS Board Jumper Settings
  • Calibration: OC power supplies (+12VDC, -12VDC, +5VDC)
  • Calibration: OC Power-up Sequence
  • Calibration: OGP DIP switches
  • Calibration: Operator Console jumper and switch settings
  • Calibration: Operator Console Power Supply Ground Connection
  • Calibration: Operator Console power-up sequence verification
  • Calibration: Parallel setting of cradle
  • Calibration: Patient intercom volume adjustment
  • Calibration: PCI Host Card Connector Configuration
  • Calibration: PCI Host Card Jumper Configuration
  • Calibration: PCI tray installation verification
  • Calibration: Photo sensor
  • Calibration: Photo sensor and plate alignment
  • Calibration: Plane of Rotation
  • Calibration: Positioning lights halogen top light right-left alignment
  • Calibration: Power Off Delay Time
  • Calibration: Powering off optional Pioneer MOD drive (LFC only)
  • Calibration: Q Calibration (Q-CAL)
  • Calibration: Q-CAL (Focus Position)
  • Calibration: Radial Alignment
  • Calibration: Rating plate attachment
  • Calibration: REAR CN1 Fan Alarm Enable/Disable
  • Calibration: REAR CN1 Intercom Minimum Volume
  • Calibration: REAR CN1 Safety Loop Control
  • Calibration: REAR CN1 Talk On Switch
  • Calibration: REAR CN1 TGP Communication
  • Calibration: Recon cycle verification
  • Calibration: Remove gantry rear cover
  • Calibration: Remove OC covers
  • Calibration: Remove PDU covers
  • Calibration: RF Shoe position and alignment
  • Calibration: Rotation speed
  • Calibration: Rotation velocity offset
  • Calibration: Rotational mechanism clearance
  • Calibration: Saving patient images to MOD
  • Calibration: Scout scan cradle positioning and data collection
  • Calibration: Scout scan data collection
  • Calibration: Scout scan offset data collection
  • Calibration: Screw tightening torque
  • Calibration: Serial Port Expander SCSI Configuration (ST-1800)
  • Calibration: Serial Port Expander ST-1800 Configuration
  • Calibration: Set gantry safety switches
  • Calibration: Signal and power brush wear inspection
  • Calibration: Signal brush contact verification
  • Calibration: Slip ring cleaning
  • Calibration: Slip ring cleaning for noise reduction
  • Calibration: Slip ring radial position
  • Calibration: Software installation and system configuration
  • Calibration: Software loading preparation
  • Calibration: Speaker assembly audio function
  • Calibration: SUB BD service switches
  • Calibration: System configuration
  • Calibration: System configuration and desktop menu
  • Calibration: System module LED verification
  • Calibration: System revision upgrade procedure
  • Calibration: System shutdown before OS installation
  • Calibration: Table board IMS jumper configuration
  • Calibration: Table DOWN limit
  • Calibration: Table positioning in right-left direction
  • Calibration: Table positioning right-left alignment
  • Calibration: TGP board DIP switches S1
  • Calibration: TGP board DIP switches S2
  • Calibration: TGP board DIP switches S3
  • Calibration: TGP board jumpers
  • Calibration: Tube Calibration
  • Calibration: Unlatch switches
  • Calibration: Workstation Default Test
  • Calibration: X-ray tube automatic learning
  • Calibration: X-ray tube DAS count verification
  • Calibration: X-ray tube warm-up
  • Calibration: XT Calibration (XT-CAL)
  • Calibration: XT-CAL (Crosstalk Compensation)
  • Capacitor assembly screws loose
  • Capacitor explosion hazard in gantry
  • Capacitor Inverter Assembly failure
  • Collimator filter requires replacement
  • Component assembly procedure: Primary coil flat cable connection
  • Cooling check failure – COOLING OVER LIMIT message
  • Cradle stops before reaching scan position
  • CRT monitor manual degaussing for color correction
  • CRT monitor reset to factory preset
  • DAS data transfer errors occur
  • Detector and collimator/filter maintenance note
  • DTRF assembly requires replacement
  • Dust filter condition not checked
  • Electrical hazard during gantry service
  • Electrocution hazard from gantry slip rings
  • Electromagnetic interference from other equipment affecting CT system
  • Electrostatic discharge damage to circuit boards
  • Electrostatic discharge precaution during board service
  • EMC Board 3 Phases failure
  • Emergency stop function not tested
  • Emergency stop switch does not halt X-ray
  • External landmark function malfunction
  • Eye damage from laser positioning light
  • Fan operation check
  • Foot pedal contact with front cover
  • Functional check: Table and gantry interlock
  • Functional check: Table and gantry panel switches
  • Gantry and rotor do not stop after End Exam
  • Gantry cannot rotate with covers removed
  • Gantry rotation hazard with wrong installation sequence
  • Hardware inventory check
  • HV cable (anode/cathode) failure
  • HV Tank replacement required
  • HV Tank temperature overheat condition
  • Initialization and safety checks before adjustment
  • Inverter Coils Mid Power Assembly failure
  • Inverter Mid Power Assembly replacement required
  • Keyboard function test - 103 keyboard input
  • Keyboard function test - input response
  • Keyboard function test - LED response
  • KV Control Board failure
  • Laser positioning light malfunction
  • Load distribution and DC supply voltage not verified
  • Monitor function test - color output
  • Monitor function test - display pattern
  • Move to Scan button does not blink after confirm
  • Nonvolatile Parameters RAM requires replacement
  • OC filter clogged or dirty
  • OGP assembly requires replacement
  • Operator Console power-up fails or stops during self-test
  • Optical fiber cable degradation
  • Optical fiber cable disconnection
  • Pause function does not stop scan at next ISD
  • Perform functional scan test after memory installation
  • Personal injury from loose clothing or jewelry during gantry servicing
  • Power distribution unit input voltage not verified
  • Power supply PS L module malfunction
  • Power system lockout procedure
  • Pre-adjustment setup: Access power supplies and speaker panel
  • Pre-adjustment setup: Remove operator console covers
  • QA phantom quality assessment check
  • Replace detector assembly
  • RF transmitter assembly requires replacement
  • Rotation Board High Speed requires replacement
  • Rotation Capacitors Set 23/23 requires replacement
  • Safety loop relay test and reset
  • Scout scan phantom diameter measurement incorrect
  • Scout scan tilted gantry not inhibited
  • SCSI test - storage device access verification
  • Slip ring damage from contamination
  • Software upgrade not installed
  • Stop Scan does not immediately halt
  • Stud bolt installation with locktite
  • System installation checks not completed
  • System shutdown test
  • Thermal burn hazard from X-ray tube
  • Top positioning light replacement
  • Verify main memory size after installation
  • Warm-up error detected
  • Warm-up scan does not complete
  • Workstation default test - Audio
  • Workstation default test - Monitor
  • Workstation default test - Mouse
  • X-ray generator power and gantry servo voltage not verified
  • X-ray ON indicator or alarm not functioning
  • X-ray radiation exposure during service
  • X-ray tube overheat protection not functioning
  • XG Error 371 during continuous scanning
  • XG ERROR 371 occurs during helical scanning

241 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
100xxExternal communication errorsCommunication failure with external system or networkunknown
110xxThermal errorsGenerator or tube temperature limit reachedwarning
30xxTube spits errorsErrors generated by X-ray tube during operationunknown
40xxRotation errorsRotation function malfunction or failureunknown
50xxHeater errorsHeater function malfunction or failureunknown
60xxExposure errorsHV inverter, mA measurement, or exposure control malfunctionunknown
70xxPower supply errorsLow voltage supply or DC bus voltage failureunknown
80xxHardware errorsInternal communication failure or cable faultunknown
90xxApplication errorsSaved RAM corruption or software failureunknown
ALERT: SCSI hard error on (1,3)SCSI hard drive error reported on controller 1, drive 3Occurs during OS installation on systems with DASM (Direct Attached Storage Module) until configuration setting is completedwarning
CHANGEAdjustment data not stored during gantry or table adjustmentCalibration or adjustment operation fails to save datawarning
CONERRFPGA Configuration error occurredFPGA configuration fails or becomes corruptederror
ERRGGantry rotation operation error detectedGantry rotation fails to respond or completes incorrectlyerror
ERRMProcessor communication or scan operation error detectedInternal processor communication failure or scan operation error occurserror
ERRTTable or tilt operation error detectedTable or gantry tilt fails to respond or completes incorrectlyerror
FECERRFEC encode error detectedForward Error Correction encoding fails during data transmissionwarning
SVALMServo amplifier in alarm state or AC 200V power lossServo amplifier detects fault or loses mains AC power supplyerror
WARM-UP ERRORWarm-up scan data out of acceptable rangeData collected during warm-up scan sequence falls outside specified range limitswarning
XG ERROR 371Scan cancellation error in XG CPU systemsHelical scanning with 140kV/160mA/10mm or larger thickness/90 sec or longer scan time on 90 slice option systems with XG CPU firmware V3.10 or lowererror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI