GE Healthcare Corometrics 250cx: troubleshooting and error codes
The ServiceForge knowledge base holds 97 documented symptoms for the GE Healthcare Corometrics 250cx, with 215 ranked causes, 149 diagnostic checks with test points and expected values, and 19 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Recorder light printing caution
Recorder prints too lightly (intermittently), fetal chart prints barely legible
Ranked causes
- 1. Inappropriate or wrong chart paper used (common)
- 2. Paper residue, dust, or material buildup on thermal print head (common)
- 3. Thermal print head pressure uneven; springs not seated properly (occasional)
- 4. Recorder door assembly broken or loose causing roller misalignment (occasional)
- 5. Thermal print head not aligned properly with roller (occasional)
- 6. Print head voltage incorrect (occasional)
- 7. Thermal print head defective (rare)
- 8. Recorder board defective (rare)
Load GE-approved chart paper
- Test point
- Recorder paper slot
- Expected
- GE-approved paper (2009828-CAO, DAO, FAO) installed
2 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder prints too lightly
Intermittent or consistent light printing on fetal chart paper making prints barely legible.
Ranked causes
- 1. Wrong chart paper type used (common)
- 2. Paper residue or dust on thermal print head (common)
- 3. Thermal print head springs not seated properly (occasional)
- 4. Recorder door assembly loose or broken (occasional)
- 5. Thermal print head misaligned with roller (occasional)
- 6. Print head voltage incorrect (occasional)
- 7. Thermal print head defective (rare)
- 8. Recorder board defective (rare)
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dashes displayed in maternal SpO2 area
Dashes (---) shown in MSpO2 area of display
Ranked causes
- 1. Sensor improperly applied on patient finger (common)
- 2. Maternal movement (common)
- 3. Excessive ambient light (occasional)
- 4. Insufficient signal strength (occasional)
- 5. Wrong SpO2 sensor type connected (occasional)
- 6. Sensor defective (occasional)
- 7. SpO2 module or carrier board defective (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dashes displayed in MSpO2 area patient safety
Dashes (---) shown in MSpO2 display area instead of SpO2 value, indicating signal loss or measurement failure
Ranked causes
- 1. Sensor improperly applied on patient's finger (common)
- 2. Maternal movement artifact (common)
- 3. Excessive ambient light (occasional)
- 4. Intermediate cable not firmly attached (occasional)
- 5. Wrong SpO2 sensor type connected (occasional)
- 6. Sensor defective (occasional)
- 7. SpO2 module or carrier board defective (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal pressure parameter measured incorrectly patient safety
Internal uterine pressure (IUP) signal not accurate or not being measured correctly during monitoring
Ranked causes
- 1. Transducer not properly connected to monitor unit (common)
- 2. Air bubble in dome or catheter blocked (common)
- 3. Dome cracked (occasional)
- 4. Catheter fallen out of place (occasional)
- 5. Catheter or strain gauge not zeroed (occasional)
- 6. UA board defective (rare)
Verify transducer connection to monitor
- Test point
- IUP transducer connector port
- Expected
- Transducer securely seated
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal pressure parameter not measured correctly patient safety
Internal uterine activity (IUA) pressure reading is inaccurate or not functioning; displayed as ID S5.1
Ranked causes
- 1. Transducer not properly connected to monitor (common)
- 2. Air bubble in dome or blocked catheter (common)
- 3. Dome cracked or damaged (occasional)
- 4. Catheter out of place (occasional)
- 5. Catheter or strain gauge not zeroed (occasional)
- 6. UA board defective (rare)
Inspect transducer connection
- Test point
- Transducer input connector on monitor
- Expected
- Connector fully seated and secure
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK CUFF message displayed patient safety
CHECK CUFF message shown in NIBP area of display
Ranked causes
- 1. Improper cuff position or loose cuff (common)
- 2. Maternal movement (common)
- 3. Hose not properly connected to monitor (air pressure error) (occasional)
- 4. Neonatal cuff used instead of adult cuff (occasional)
- 5. NIBP board defective (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK CUFF message displayed in NIBP area patient safety
Error message 'CHECK CUFF' appears in non-invasive blood pressure display area on screen
Ranked causes
- 1. Improper cuff position or loose cuff (common)
- 2. Maternal movement (common)
- 3. Hose not properly connected to monitor (occasional)
- 4. Neonatal cuff used instead of adult cuff (occasional)
- 5. NIBP board defective (rare)
Verify cuff position and tension on maternal arm
- Test point
- Blood pressure cuff
- Expected
- Cuff snug on upper arm at heart level
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fetal heart rate not displayed properly patient safety
Monitor does not show FHR on display or display is incorrect when ultrasound transducer connected
Ranked causes
- 1. Ultrasound transducer not properly connected (common)
- 2. Transducer not properly positioned on patient (common)
- 3. Insufficient ultrasound gel applied (common)
- 4. Ultrasound transducer defective (occasional)
- 5. Signal cannot be captured due to patient/fetus conditions (occasional)
Verify transducer physical connection to monitor
- Test point
- Transducer connector port
- Expected
- Transducer securely seated in port
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fetal heart rate not displayed properly with ultrasound transducer connected patient safety
Monitor does not show FHR on display when ultrasound transducer is connected
Ranked causes
- 1. Ultrasound transducer not properly connected (common)
- 2. Ultrasound transducer not properly positioned on patient (common)
- 3. Too little ultrasound gel applied (common)
- 4. Ultrasound transducer defective (occasional)
- 5. Ultrasound signal not captured due to patient or fetal conditions (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal FECG parameter erratic or not recorded patient safety
FECG parameter shown on display erratically or not being recorded properly on strip chart.
Ranked causes
- 1. FECG cable not properly connected to monitor unit (common)
- 2. Attachment pad or leg plate not securely attached to patient (common)
- 3. Electrode wire not secure in leg plate post (common)
- 4. Electrode defective or not properly attached (occasional)
- 5. Attachment pad defective (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal FECG parameter erratic or not recording patient safety
FECG signal shown erratically on display or not being recorded to strip chart properly
Ranked causes
- 1. FECG cable not properly connected to monitor unit (common)
- 2. Attachment pad or leg plate not securely attached to patient (common)
- 3. Electrode wire not secure in leg plate post (occasional)
- 4. Electrode defective or improperly attached (occasional)
- 5. Attachment pad defective (occasional)
Verify FECG cable connection to monitor unit
- Test point
- FECG connector port
- Expected
- Cable securely seated
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Static noise exists on ultrasound signal
Ultrasound signal shows static or noise interference during operation
Ranked causes
- 1. Inappropriate or refurbished ultrasound transducer used (common)
- 2. Environmental noise (common)
- 3. Active fetus (occasional)
- 4. Ultrasound transducer defective (occasional)
- 5. Maternal movement (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
TOCO transducer not recording contractions patient safety
Uterine activity (TOCO) signal not appearing on display or strip chart during maternal contractions
Ranked causes
- 1. TOCO transducer not properly connected to monitor unit (common)
- 2. TOCO transducer not properly positioned on patient (common)
- 3. UA reference range exceeded (occasional)
- 4. No maternal contractions occurring (rare)
- 5. TOCO transducer defective (occasional)
Verify TOCO transducer connection to monitor
- Test point
- TOCO connector port
- Expected
- Transducer securely connected
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
TOCO transducer not recording uterine contractions patient safety
Fetal monitor external uterine activity (UA) display shows no contraction signal; transducer inactive
Ranked causes
- 1. TOCO transducer not properly connected to monitor unit (common)
- 2. TOCO transducer not properly positioned on patient (common)
- 3. Uterine activity reference range exceeded (occasional)
- 4. No maternal contractions occurring (rare)
- 5. TOCO transducer is defective (occasional)
Verify TOCO transducer connector seated at monitor
- Test point
- Transducer connector at monitor unit
- Expected
- Connector fully inserted and securely latched
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK TOCO message displayed in UA area patient safety
Error message 'CHECK TOCO' appears in uterine activity display area when UA Reference button is pressed
Ranked causes
- 1. UA Reference button pressed before circuitry stabilized (common)
- 2. UA reference range exceeded due to over-tightened belt (common)
- 3. TOCO transducer defective (occasional)
- 4. UA board defective (rare)
Verify UA circuitry stabilization delay observed
- Test point
- Monitor startup or transducer connection timing
- Expected
- 10 seconds elapsed before pressing UA Reference button
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK TOCO message displayed when UA Reference button pressed patient safety
Monitor displays 'CHECK TOCO' message in UA area when technician attempts to reset UA reference
Ranked causes
- 1. UA Reference button pressed before UA circuitry stabilized after power-on (common)
- 2. Uterine activity reference range exceeded due to belt over-tightening (common)
- 3. TOCO transducer is defective (occasional)
- 4. UA board is defective (occasional)
Wait 10 seconds after power-on or transducer connection before pressing UA Reference
- Test point
- Monitor power state and transducer connection timing
- Expected
- UA circuitry stabilized; no CHECK TOCO message when reference button pressed after 10-second wait
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No monitoring functions and power indicator off patient safety
Monitor does not power on; no display or LED illumination when power switch is set to On
Ranked causes
- 1. Power cord not properly connected (common)
- 2. No power at outlet (common)
- 3. Power cord defective (occasional)
- 4. Fuses open or missing (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK IUP message displayed patient safety
Monitor displays CHECK IUP alert in uterine activity area; indicates pressure measurement problem
Ranked causes
- 1. Blockage in fluid-filled catheter (common)
- 2. Fetus pressing on catheter (common)
- 3. Defective catheter (occasional)
Flush catheter system
- Test point
- Catheter fluid pathway
- Expected
- Clear fluid flow without resistance
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK IUP message displayed in UA area patient safety
Error message 'CHECK IUP' shown in uterine activity display area on screen
Ranked causes
- 1. Blockage in fluid-filled catheter (common)
- 2. Fetus pressing directly on catheter (occasional)
- 3. Catheter defective (occasional)
Flush and zero the catheter
- Test point
- Catheter fluid path and calibration
- Expected
- Blockage cleared, zero reading verified
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect date and time displayed caution
Monitor displays wrong date and/or time in system clock
Ranked causes
- 1. Date and time set incorrectly (common)
- 2. Timekeeping RAM chip defective (rare)
- 3. Main board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
MECG parameter erratic or not properly recorded patient safety
MECG signal displayed inconsistently or not recording properly on monitor; erratic or missing data
Ranked causes
- 2. Electrode gel is dried (common)
- 6. MECG cable is defective (rare)
- 7. MECG board is defective (rare)
Verify MECG cable is securely connected to monitor
- Test point
- MECG cable connector to monitor unit
- Expected
- Cable securely connected
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not generate heartbeats or pulse sounds patient safety
No audio output from monitor during operation despite having transducer signal
Ranked causes
- 1. Audio volume set too low (common)
- 2. Transducer not connected or loosely connected (common)
- 3. Inadequate ultrasound gel (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor shows incorrect date and time
System displays wrong date or time after power-on or during operation
Ranked causes
- 1. Date and time set incorrectly (common)
- 2. Timekeeping RAM chip defective (occasional)
- 3. Main board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No heartbeat or pulse sounds generated patient safety
Monitor does not emit audio tones for detected heartbeats or pulses
Ranked causes
- 1. Audio volume set too low (common)
- 2. Transducer not connected or loose (common)
- 3. Inadequate ultrasound gel (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
OVERPRESSURE message displayed patient safety
OVERPRESSURE message shown in NIBP area indicating cuff pressure exceeded over-pressure limit
Ranked causes
- 1. Cuff pressure exceeded 315 mmHg limit (common)
- 2. Hose kinked (occasional)
- 3. Hose blocked (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
OVERPRESSURE message displayed in NIBP area patient safety
Error message 'OVERPRESSURE' shown in non-invasive blood pressure display area, indicating pressure limit exceeded
Ranked causes
- 1. Cuff pressure exceeded over-pressure limit of 315 mmHg ± 15 mmHg (common)
- 2. Hose kinked (occasional)
- 3. Hose blocked (occasional)
Restrict maternal limb movement
- Test point
- Patient arm position
- Expected
- Limb immobilized and supported
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder does not function and Record Indicator is off
Recorder inactive with Record Indicator not lit when monitor is powered on
Ranked causes
- 1. Recorder is off or out of paper (common)
- 2. Paper-out photosensor out of calibration (occasional)
- 3. Paper-out photosensor disconnected or defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder does not function; indicator off caution
Recorder does not operate and Record Indicator LED is not illuminated
Ranked causes
- 1. Recorder off or out of paper (common)
- 2. Paper-out photosensor out of calibration (occasional)
- 3. Paper-out photosensor disconnected or defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder functions but Record Indicator flashes caution
Recorder operates but Record Indicator LED flashes on and off every second
Ranked causes
- 1. Paper supply low or incorrectly loaded (common)
- 2. Paper-low photosensor out of calibration (occasional)
- 3. Paper-low photosensor disconnected or defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder Record Indicator flashes on and off every second
Recorder functions but Record Indicator flashes intermittently
Ranked causes
- 1. Paper supply low or incorrectly loaded (common)
- 2. Paper-low photosensor out of calibration (occasional)
- 3. Paper-low photosensor disconnected or defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Recorder horizontal and vertical offset
Verify recorder trace position and quality on strip chart paper
Ranked causes
- 1. Recorder horizontal offset out of specification (occasional)
- 2. Recorder vertical line quality or paper curl defect (occasional)
Run recorder calibration test on strip chart paper
- Test point
- Calibration traces on paper
- Expected
- First horizontal trace and vertical lines printed within specification
- Tools
- GE-approved strip chart paper, Trim Knob Control
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dashes shown in MHR/P display area patient safety
MECG reading displays as dashes (– – –) indicating inability to determine maternal heart rate due to insufficient signal or signal loss
Ranked causes
- 1. Monitor unable to determine MECG due to insufficient signal (common)
- 2. MECG board is defective (occasional)
Confirm patient is not asystolic and electrodes are firmly secured
- Test point
- Patient condition and electrode attachment
- Expected
- Patient has cardiac activity; electrodes firmly adhered to skin
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flashing '+' sign in UA field patient safety
Plus sign flashing in uterine activity display area on screen
Ranked causes
- 1. Relative pressure exceeds 100 (common)
- 2. UA board defective (rare)
Reset UA reference pressure
- Test point
- UA Reference button
- Expected
- Flashing '+' clears after button press between contractions
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flashing plus sign in UA field on display patient safety
Fetal monitor shows flashing '+' character in uterine activity display field
Ranked causes
- 1. Relative pressure reading exceeds 100 units (common)
- 2. UA board is defective (occasional)
Press UA Reference button between contractions
- Test point
- UA Reference button and display response
- Expected
- Flashing + disappears; UA display returns to normal reading
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder does not function and Record Indicator is on
Recorder inactive with Record Indicator illuminated when monitor is powered on
Ranked causes
- 1. Recorder stepper motor defective (occasional)
- 2. Recorder roller defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder not functioning; Record Indicator on caution
Recorder does not operate but Record Indicator LED remains illuminated
Ranked causes
- 1. Recorder stepper motor defective (occasional)
- 2. Recorder roller defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
REPAIR message displayed patient safety
REPAIR message shown in NIBP area
Ranked causes
- 1. NIBP pump defective (occasional)
- 2. System error or self-test failure in NIBP or main board (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
REPAIR message displayed in NIBP area patient safety
Repair request message 'REPAIR' shown in non-invasive blood pressure display area indicating service needed
Ranked causes
- 1. NIBP pump defective (occasional)
- 2. System error or self-test failure (occasional)
Verify NIBP pump functionality
- Test point
- NIBP pneumatic system
- Expected
- Pump cycles during BP measurement attempt
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SYSTEM FAULT:SOFTWARE error message displayed patient safety
Monitor displays error message 'SYSTEM FAULT:SOFTWARE error' on startup
Ranked causes
- 1. SW1 switch #7 incorrectly set (common)
- 2. Software error (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SYSTEM FAULT:UI KEYPAD error message displayed patient safety
Monitor displays error message 'SYSTEM FAULT:UI KEYPAD error' on startup
Ranked causes
- 1. Main-keypad communication problem (occasional)
- 2. UI keypad board defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Volume of ultrasound tone is low
Ultrasound audio output volume is insufficient during operation
Ranked causes
- 1. Tone volume set low (common)
- 2. Dual ultrasound board out of calibration (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure high reading patient safety
Blood pressure measurement displays higher than expected; measurement artifact or measurement taken during uterine contraction
Ranked causes
- 1. Measurement taken during uterine contraction (common)
Verify measurement was taken between uterine contractions
- Test point
- Contraction monitor display and BP reading correlation
- Expected
- BP reading taken during low-contraction period (baseline)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure reading high patient safety
NIBP displays elevated pressure value
Ranked causes
- 1. Measurement taken during uterine contraction (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: MHR/P alarm limits and response
Verify maternal heart rate/pulsatile alarm thresholds (60–120 bpm) and alarm tone behavior (silence, re-alarm timing) using fetal simulator
Ranked causes
- 1. Alarm limit calibration and functional verification required (common)
Test high alarm threshold activation
- Test point
- Monitor display; rear panel speaker
- Expected
- No alarm tone at 119 bpm, no alarm at 120 bpm, alarm tone at 121 bpm
- Tools
- CorometricsTM 325 simulator with Manual Adjustment knob
6 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP overpressure detection caution
Verify NIBP overpressure detection triggers at correct pressure range (300–330 mmHg)
Ranked causes
- 1. Overpressure detection out of specification (occasional)
Set NIBP to overpressure detection test mode
- Test point
- Mode selector in NIBP Calibration screen
- Expected
- OVP Test mode selected; monitor closes valves
- Tools
- Trim Knob Control
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP system patient safety
Non-invasive blood pressure subsystem requires periodic verification and checkout after service work
Ranked causes
- 1. Routine maintenance interval or post-service verification required (common)
Perform NIBP calibration check
- Test point
- NIBP subsystem
- Expected
- Calibration within specification
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP system leakage
Verify NIBP pneumatic circuit does not leak more than specification
Ranked causes
- 1. System leakage exceeds 6 mmHg/min specification (occasional)
Set NIBP to leakage test mode
- Test point
- Mode selector in NIBP Calibration screen
- Expected
- Leak Test mode selected
- Tools
- Trim Knob Control
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP transducers
NIBP pressure readings inaccurate; verification shows values outside ±2 mmHg pass range compared to external manometer
Ranked causes
- 1. NIBP transducers out of calibration (common)
Verify NIBP transducer accuracy
- Test point
- PT1 and PT2 on monitor display vs external manometer
- Expected
- Within ± 2 mmHg (0.3 kPa) of external manometer reading at ~200 mmHg cuff inflation, ±2 mmHg
- Tools
- External manometer capable of reading to 350 mmHg, adult cuff, 3 inch rigid cylinder, 12-foot NIBP hose, hand pump bulb
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Recorder offsets caution
Verification of recorder paper feed calibration and signal offset values for accurate waveform display
Ranked causes
- 1. Annual maintenance or post-service verification of recorder assembly (occasional)
Verify recorder calibration offsets
- Test point
- Recorder offset adjustment
- Expected
- Offsets within calibration specification
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Recorder paper-load photosensor caution
Verify and adjust paper-load photosensor voltage output to specification
Ranked causes
- 1. Paper-load photosensor voltage out of specification (occasional)
Load GE-approved paper without black squares, minimum 9 sheets
- Test point
- Recorder paper loaded in recorder
- Expected
- No black squares visible; at least 9 sheets of strip chart paper loaded
- Tools
- GE-approved strip chart paper
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Recorder paper-low photosensor caution
Verify and adjust paper-low photosensor voltage output to specification
Ranked causes
- 1. Paper-low photosensor voltage out of specification (occasional)
Load GE-approved paper without black squares
- Test point
- Recorder paper surface
- Expected
- No black squares visible; paper advanced for tension
- Tools
- GE-approved strip chart paper
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Recorder paper-out photosensor caution
Verify and adjust paper-out photosensor voltage output to specification
Ranked causes
- 1. Paper-out photosensor voltage out of specification (occasional)
Load GE-approved paper without black squares
- Test point
- Recorder paper surface
- Expected
- No black squares visible; paper advanced for tension
- Tools
- GE-approved strip chart paper
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Recorder photosensors caution
Optical sensor verification for recorder assembly to ensure accurate paper feed and signal detection
Ranked causes
- 1. Scheduled maintenance or post-service checkout of recorder assembly (occasional)
Perform recorder photosensors check
- Test point
- Recorder photosensor assembly
- Expected
- Sensors operating and properly aligned
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Recorder vertical offset adjustment caution
Recorder prints horizontal traces with offset outside acceptable range; adjust using trim knob and set screws
Ranked causes
- 1. Recorder assembly misaligned (common)
Remove top cover
- Test point
- Top cover fasteners
- Expected
- Cover removed
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Ultrasound input signal verification
Verify ultrasound signal input to monitor via simulator cable, FHR1 display, heartbeat indicator, audio output, and recorder strip chart output at baseline and variable heart rates.
Ranked causes
- 1. Ultrasound input signal not properly configured or verified (unknown)
Verify FHR1 value display
- Test point
- FHR1 display on monitor
- Expected
- 120 bpm
7 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
COMM message displayed patient safety
COMM message shown in NIBP area indicating communication error
Ranked causes
- 1. Communication error between NIBP module and monitor circuitry (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
COMM message displayed in MSpO2 area patient safety
Communication error message shown in MSpO2 display area, preventing pulse oximetry reading
Ranked causes
- 1. SpO2 module or main board defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
COMM message displayed in NIBP area patient safety
Communication error message 'COMM' shown in non-invasive blood pressure display area
Ranked causes
- 1. Communication error between NIBP module and monitor circuitry (rare)
Verify NIBP module and main board communication
- Test point
- NIBP board-to-main board interface
- Expected
- Communication link established
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication error in maternal SpO2 module
COMM message shown in MSpO2 area of display
Ranked causes
- 1. SpO2 module or main board defective (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Dashes displayed in NIBP area
- Dashes shown in NIBP display area
- ESD precautions during FECG/UA and recorder procedures
- ESD precautions for front bezel replacement
- Fan replacement
- FECG/UA board voltage adjustment
- FHR rate and strip chart trend do not correlate
- FHR Rate displayed and FHR trend on strip chart paper do not correlate
- Front bezel replacement
- High blood pressure reading
- Main board power supply voltage out of specification
- Main power supply replacement
- MHR/P pulse source blank when MSpO2 selected
- Monitor does not show MSpO2 value - SENSOR message displayed
- MOTION message displayed
- MOTION message displayed in NIBP area
- No monitoring functions and power indicator does not illuminate
- No MSpO2 value with SENSOR message
- Paper-low photosensor replacement
- Paper-out photosensor replacement
- Recorder horizontal offset adjustment
- Recorder message: PAPER INCORRECTLY LOADED, RELOAD WITH BLACK SQUARES DOWN
- Recorder not functioning; incorrect paper loaded message
- Recorder paper curls or printing unequal
- Recorder paper-loading photosensor replacement required
- Recorder prints with vertical offset
- Recorder vertical offset adjustment
- REPAIR message in maternal SpO2 area
- REPAIR message shown in SpO2 area (Nellcor only)
- Simulator setup verification
- SYSTEM FAULT:ALERT error message displayed
- SYSTEM FAULT:DEFAULTS error message displayed
- SYSTEM FAULT:DSP error message displayed
- SYSTEM FAULT:RAM error message displayed
- SYSTEM FAULT:ROM error message displayed
- Thermal print head replacement and voltage adjustment
- Weak signal on blood pressure measurement
37 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
CHECK CUFF | Cuff position, connection, or pressure measurement issue preventing accurate reading | System detects improper cuff application, disconnection, or other setup error during measurement | warning |
COMM | Communication error between built-in NIBP module and main monitor circuitry | NIBP module unable to communicate with monitor during measurement cycle | error |
MOTION | Excessive patient movement detected during blood pressure measurement | System detects limb movement exceeding acceptable threshold during measurement | warning |
OVERPRESSURE | Cuff pressure exceeded safe operating limit of 315 mmHg ± 15 mmHg | System detects cuff pressure above over-pressure threshold during measurement attempt | warning |
REPAIR | NIBP pump failure or system self-test error detected | Pump malfunction or system self-test failure during initialization or measurement | error |
S11.1 | Recorder prints too lightly | Fetal chart prints are barely readable or intermittently light | error |
S7.8 | WEAK SIGNAL message on blood pressure display | Monitor unable to determine blood pressure due to insufficient signal strength | warning |
S8.1 | COMM message in MSpO2 area | Communication error between built-in MSpO2 module and monitor circuitry | error |
S8.2 | Dashes (---) in MSpO2 area | Monitor unable to determine pulse oximetry due to insufficient or erratic signal | warning |
S8.3 | MHR/P pulse source blank when MSpO2 selected | MSpO2 pulse source mode set to Normal instead of Fast | info |
S8.4 | REPAIR message in MSpO2 area (Nellcor only) | System error or self-test failure detected | error |
S8.5 | No MSpO2 value displayed with SENSOR message | Monitor detects wrong sensor type or sensor not recognized | error |
SYSTEM FAULT:ALERT | Software error detected during initialization | Software exception or fault during startup | error |
SYSTEM FAULT:DEFAULTS | Default configuration initialization failure | Failure to initialize default parameters on startup | critical |
SYSTEM FAULT:DSP | DSP board communication or hardware failure | DSP board failure detected during initialization | critical |
SYSTEM FAULT:RAM | Main board SRAM memory defective | SRAM memory test failure on startup | critical |
SYSTEM FAULT:ROM | Main board flash memory defective | Flash memory read failure on startup | critical |
SYSTEM FAULT:SOFTWARE | Software configuration error or incorrect switch setting | Invalid software configuration detected on startup | error |
SYSTEM FAULT:UI KEYPAD | Communication failure between main board and keypad | Keypad communication bus failure detected | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI