GE Healthcare CIC: troubleshooting and error codes
The ServiceForge knowledge base holds 72 documented symptoms for the GE Healthcare CIC, with 153 ranked causes, 169 diagnostic checks with test points and expected values, and 1 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Erroneous noninvasive blood pressure measurement patient safety
Monitor unable to detect adequate pulsations or detecting excessive movement, producing inaccurate NBP values
Ranked causes
- 1. Incorrect cuff size (common)
- 2. Residual air in cuff from previous measurement (occasional)
- 3. Cuff fit too tight or too loose (common)
- 4. Cuff and heart at different levels (occasional)
- 5. Artery marker misaligned with artery (occasional)
- 6. Excessive patient movement during measurement (common)
- 7. Cuff or tubing leak (occasional)
- 8. Patient has weak pulse (rare)
- 9. Calibration drift (occasional)
Confirm NBP reading using auscultatory method
- Test point
- Patient arm
- Expected
- Auscultatory reading matches or closely approximates monitor reading
4 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate cardiac output values patient safety
Cardiac output readings are not accurate or reliable. Multiple factors can contribute including injection technique, patient respiration, catheter placement, baseline temperature stability, and IV solution infusion.
Ranked causes
- 1. Improper solution temperature management (common)
- 2. Incorrect injection technique (common)
- 3. Inadequate interval between injections (common)
- 4. Inadequate temperature differential (occasional)
- 5. Respiratory variation interference (common)
- 6. Defective Swan-Ganz catheter (occasional)
- 7. Patient arrhythmias (occasional)
- 8. Rapid IV solution infusion interference (occasional)
- 9. Injectate temperature cable malfunction (occasional)
Verify injectate solution storage conditions
- Test point
- Solution bag
- Expected
- Not exposed to heat sources or other equipment
7 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate or unreliable noninvasive blood pressure measurements patient safety
Erroneous NBP readings obtained during patient monitoring due to improper cuff selection or application
Ranked causes
- 1. Single-hose cuff used with dual-hose tubing (occasional)
- 2. Cuff placed on limb with compromised circulation (occasional)
- 3. Incorrect cuff size selected for patient limb circumference (common)
- 4. Cuff not positioned snugly or artery marking misaligned (common)
- 5. Cuff placed too far from elbow or tubing kinked (occasional)
- 6. External pressure applied against cuff during measurement (occasional)
- 7. Arm placement incorrect relative to heart level (occasional)
- 8. Measurement performed on dominant arm or with clothing obstruction (occasional)
Verify cuff size matches patient limb circumference
- Test point
- Cuff label
- Expected
- Limb circumference marked on cuff
10 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate SpO2 reading patient safety
Pulse oximeter displays SpO2 value that does not match blood gas analyzer or clinical expectation; reading may be artificially high or low depending on patient condition or interference
Ranked causes
- 1. Dysfunctional hemoglobins (carboxyhemoglobin, methemoglobin) (common)
- 2. Nail polish or artificial nails (common)
- 3. Dyes or pigmentation-altering substances (occasional)
- 4. Ambient light interference (common)
- 5. Electrosurgical unit interference (occasional)
- 6. Motion artifact (common)
- 7. Sensor placement on extremity with blood pressure cuff or catheter (occasional)
- 8. Patient conditions: anemia, sickle cell crisis, smoke inhalation, jaundice (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Displayed invasive pressure values differ from expected patient safety
Arterial or pressure readings shown on monitor do not match clinical expectation or are inconsistent with patient condition
Ranked causes
- 1. Patient lying on tubing or tubing kinked (common)
- 2. Air bubbles in tubing (common)
- 3. Excess tubing length (occasional)
- 4. Transducer not at correct anatomical level (common)
- 5. Pressure reference not zeroed (common)
- 6. IABP mode not enabled on monitor when patient on IABP (occasional)
Inspect tubing for kinks or compression
- Test point
- Patient tubing path
- Expected
- No kinks; tubing free of patient contact
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low light or no light message displayed patient safety
Monitor shows low light or no light indicator during SvO2 monitoring
Ranked causes
- 1. Disconnected or loose optical module connections (common)
- 2. Disconnected or loose catheter connections (common)
- 3. Catheter tip not in optical reference block (packaged) (common)
- 4. Optical module malfunction (occasional)
- 5. Catheter position in patient improper (occasional)
Verify optical module to SvO2 module connection
- Test point
- Connection between optical module and SvO2 module
- Expected
- Secure connection with no gaps
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Preinsertion calibration failure message displayed patient safety
Monitor displays calibration failure message during SvO2 module setup
Ranked causes
- 1. Disconnected or loose optical module connections (common)
- 2. Disconnected or loose catheter connections (common)
- 3. Catheter tip not in optical reference block (common)
- 4. Optical module malfunction (occasional)
- 5. Damaged catheter (occasional)
Verify optical module to SvO2 module connection
- Test point
- Connection between optical module and SvO2 module
- Expected
- Secure connection with no gaps
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect end-tidal CO2 values displayed patient safety
ETCO2 readings are considerably lower than expected or do not match blood-gas analysis
Ranked causes
- 1. Leak in respiratory circuit (common)
- 2. Hypothermia without blood-gas correction (occasional)
- 3. Anesthesia gases affecting readings (occasional)
- 4. Clinical factors: dead-space ventilation, V/Q mismatch, low cardiac output, alveolar shunts, incomplete alveolar emptying (occasional)
Inspect respiratory circuit for leaks
- Test point
- entire circuit
- Expected
- No leaks, circuit intact
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor cannot detect stable baseline temperature patient safety
Monitor displays message indicating inability to find stable baseline temperature. Message clears when stable baseline is detected.
Ranked causes
- 1. Respiratory variation interference (common)
- 2. Inadequate injectate temperature differential (common)
- 3. IT cable malfunction (occasional)
- 4. Defective Swan-Ganz catheter (occasional)
Verify patient and CO setup
- Test point
- Monitor settings, cables, catheter placement
- Expected
- All connections secure, settings correct, catheter properly positioned
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacemaker activity not detected; double-counting heart rate or false bradycardia alarm patient safety
Monitor fails to detect pacemaker spikes; displays double heart rate count, alarming for low rate, or shows no pace markers on ECG
Ranked causes
- 1. Pacemaker detection program is turned off (common)
- 2. Pacemaker signal too weak for detection (occasional)
- 3. ECG signal too weak for monitor detection (occasional)
- 4. Monitor detecting atrial pacemaker artifact or non-QRS features as beats (rare)
Verify pacemaker detection is enabled
- Test point
- Monitor settings menu
- Expected
- Pacemaker detection mode ON
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Anesthetic agent sample line connection error or leak patient safety
Patient sample line (PE/PVC type) is incorrect size, loose, kinked, or leaking, causing inaccurate anesthetic agent readings or gas escape
Ranked causes
- 1. Sample line has incorrect inner diameter (not 1.19 mm) (common)
- 2. Luer-Lok fitting connection is loose or not seated (common)
- 3. Sample line is kinked, crimped, or obstructed (occasional)
Verify sample line inner diameter
- Test point
- PE/PVC patient sample line
- Expected
- Inner diameter 1.19 mm (0.047 inches)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
BIS sensor placement and connection patient safety
Proper installation of three-electrode or four-electrode BIS forehead sensors for continuous monitoring
Ranked causes
- 1. Three-electrode sensor improper placement (common)
- 2. Four-electrode sensor improper placement (common)
- 3. Sensor cable disconnection (occasional)
Verify three-electrode sensor placement
- Test point
- Circle 1, circle 3 position
- Expected
- Circle 1 at center of forehead 1.5 inches (4 cm) above nose; circle 3 on temple
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor alarming for arterial disconnect patient safety
Disconnect alarm activates when mean arterial pressure drops or waveform is lost during zeroing
Ranked causes
- 1. Catheter dislodged from vessel (rare)
- 2. Mean pressure below threshold during normal zeroing procedure (common)
- 3. Alarm threshold set too high or waveform loss during measurement (occasional)
Assess patient condition and catheter site
- Test point
- Catheter insertion site
- Expected
- Catheter in place; no visible bleeding or dislodgement
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor displaying mean arterial pressure (MAP) only measurement patient safety
Systolic and diastolic values not displayed; MAP displayed as safety mechanism when algorithm cannot reliably determine other values
Ranked causes
- 1. Small measured oscillations in cuff with excessive noise (occasional)
- 2. Calculated systolic/diastolic values too close to MAP value (occasional)
- 3. Low blood flow state (rare)
Verify cuff size and placement
- Test point
- Cuff dimensions and position
- Expected
- Proper size per diagram on cuff; positioned with artery marker aligned to brachial artery
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 waveform shows excessive artifact or noise patient safety
SpO2 waveform display contains noise spikes and artifact; signal quality degraded by motion or poor placement
Ranked causes
- 1. Photodetector not flush with tissue due to poor sensor placement (common)
- 2. Motion artifact from patient or sensor movement (common)
- 3. Inadequate or insecure sensor attachment (common)
Examine SpO2 waveform for artifact and motion spikes
- Test point
- SpO2 waveform display on monitor
- Expected
- Smooth waveform corresponding to arterial pulse; minimal noise spikes
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unclear or degraded ECG signal quality patient safety
ECG waveform appears noisy, artifact-prone, or signal strength is insufficient for reliable beat detection across leads I, II, III, or V
Ranked causes
- 1. Electrode gel dried out or adhesive aged after 48+ hours of wear (common)
- 2. Leadwire tugging or rotation about electrode snap causing ECG artifact (occasional)
- 3. Electrosurgical unit (ESU) interference present in operating environment (occasional)
Visually inspect electrodes for gel dryness and adhesive condition
- Test point
- All electrode contact surfaces
- Expected
- Gel moist and adhesive firmly adhered
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Airway adapter connection error or incomplete assembly patient safety
Patient sample line Luer-Lok fitting is not connected to airway adapter in upward position, or adapter is loose, causing gas leakage or sampling errors
Ranked causes
- 1. Sample line Luer-Lok fitting not in upward position at airway adapter (common)
- 2. Airway adapter connection is loose or disconnected (occasional)
Verify sample line Luer-Lok fitting orientation at airway adapter
- Test point
- Airway adapter Luer-Lok connector and sample line fitting position
- Expected
- Sample line Luer-Lok fitting positioned upward on airway adapter; connection is secure
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Room air
SAM module periodically calibrates to room air; process takes about 8 seconds. During calibration, host monitor displays progress message, last measured values remain displayed, CO2 waveform drops to baseline.
Ranked causes
- 1. Calibration procedure - standard operation (common)
- 2. Warm-up period after insertion into Rac (common)
Observe calibration procedure
- Test point
- Host monitor display
- Expected
- Air calibration in progress message displayed; last measured parameter values visible; CO2 waveform at baseline
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiac output value higher than expected patient safety
Calculated cardiac output is higher than clinically expected for the patient condition.
Ranked causes
- 1. Excessive injectate volume (common)
- 2. Injectate temperature too low (common)
Verify injectate volume used
- Test point
- Syringe volume
- Expected
- Reduced from previous measurement
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiac output value lower than expected patient safety
Calculated cardiac output is lower than clinically expected for the patient condition.
Ranked causes
- 1. Insufficient injectate volume (common)
- 2. Injectate temperature too high (common)
Verify injectate volume used
- Test point
- Syringe volume
- Expected
- Standard volume for patient size and measurement protocol
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Displayed PA wedge measurement differs from expected patient safety
PA wedge pressure value shown on monitor does not match clinical expectation
Ranked causes
- 1. Measurement obtained without minimum three respiratory cycles or at wrong respiratory phase (common)
- 2. PA wedge cursor not positioned at end-expiratory value (common)
Obtain minimum three respiratory cycles of wedge data
- Test point
- Respiratory waveform and PA pressure waveform
- Expected
- At least 3 complete respiratory cycles visible; stable values
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhaust line back pressure causes inaccurate readings or module damage patient safety
Exhaust line is kinked, blocked, or connected to unregulated high vacuum source, causing back pressure that damages internal module components or corrupts measurements
Ranked causes
- 1. Exhaust line is kinked or blocked (common)
- 2. Exhaust line connected to unregulated high vacuum source (occasional)
Inspect exhaust line for kinks or blockages
- Test point
- Entire length of exhaust line from EXHAUST outlet to scavenging system
- Expected
- Exhaust line is straight and patent with no visible kinks, crimps, or blockages
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow sensor performance degradation patient safety
Sensor performance degrades or contamination occurs when flow sensor is reused or module is powered off for prolonged periods
Ranked causes
- 1. Reused disposable flow sensor (common)
- 2. Module powered off with sensor in circuit (occasional)
Verify flow sensor packaging or history
- Test point
- Flow sensor label or medical record
- Expected
- Single-use sensor, not previously used
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate 12-lead analysis with 10-leadwire cable patient safety
12-lead ECG waveforms or measurements do not match expected values or previous patient studies
Ranked causes
- 1. Incorrect or misplaced V2–V6 leadwire block in 10-leadwire cable (common)
- 2. Electrode placement does not match prior configuration for serial comparison (occasional)
Visually inspect V2–V6 leadwire block color code and seating in cable
- Test point
- 10-leadwire cable block connector and color marking
- Expected
- Brown-coded V2–V6 block firmly seated in correct side of cable
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lead analysis fails or displays limited options with 3-leadwire cable
Monitor software forces single-lead analysis, restricts lead selection to I, II, III only, or disables multi-lead analysis and optional ECG features
Ranked causes
- 1. 3-leadwire electrode configuration connected instead of 5-leadwire or 10-leadwire (common)
- 2. 3-leadwire cable with fixed reference connected to incompatible Tram module (occasional)
Identify leadwire cable type connected to monitor
- Test point
- Patient cable connector and visual wire count/labeling
- Expected
- 5-leadwire (RA, LA, RL, LL, V) or 10-leadwire if multi-lead analysis required
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Module warm-up delay after storage
When SAM module is stored outside Rac and then reinserted, a warm-up period is required before operation, duration depends on module temperature.
Ranked causes
- 1. Module temperature above 60°F at reinsertion (common)
- 2. Module temperature below 60°F at reinsertion (occasional)
Check module temperature
- Test point
- SAM module
- Expected
- >60 degrees F (16 degrees C)
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor alarming for asystole, bradycardia, pause, or inaccurate heart rate with visible QRS patient safety
Monitor triggers false alarms for asystole, bradycardia, or pause despite visible QRS complexes on waveform; or displays inaccurate heart rate
Ranked causes
- 1. Insufficient QRS amplitude in analyzed leads (common)
- 2. Analyzed leads not optimal for detection (occasional)
Measure QRS amplitude in all analyzed leads
- Test point
- All ECG leads (I, II, III, V)
- Expected
- Minimum 0.5 mV amplitude in one direction; 1.0 mV recommended, minimum
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor calculates cardiac output without injection patient safety
Monitor displays cardiac output calculation even though solution was not injected. Monitor senses temperature change in patient blood consistent with an injection.
Ranked causes
- 1. Respiratory variation mimicking injection (common)
- 2. Rapid IV solution infusion (common)
Verify patient and CO setup
- Test point
- Monitor settings and cable connections
- Expected
- All settings correct, cables secure
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor incorrectly calling VTach (ventricular tachycardia) patient safety
Monitor displays VTach alarm when patient is not in VTach; false positive arrhythmia detection
Ranked causes
- 1. Wider QRS complex or artifact in analyzed leads (common)
- 2. V lead polarity changes causing misinterpretation (occasional)
Assess QRS width across all analyzed leads
- Test point
- ECG leads I, II, III, V
- Expected
- Normal QRS width; no abnormally wide complexes
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 reading unstable or varying rapidly patient safety
Displayed SpO2 values fluctuate significantly; stability indicator suggests questionable signal validity
Ranked causes
- 1. Signal artifact mimicking valid reading; adverse conditions unnoticed (occasional)
- 2. Hemodynamic disturbances affecting pulse waveform (occasional)
Assess stability of displayed SpO2 values over time
- Test point
- SpO2 parameter display
- Expected
- Reading stable within physiological variation; tracks expected arterial saturation
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 signal strength weak or unstable patient safety
SpO2 parameter window shows fewer than 3 asterisks indicating weak signal; waveform amplitude low
Ranked causes
- 1. Poor sensor attachment or positioning (common)
- 2. Non-ideal environmental conditions (occasional)
Observe signal strength indicator asterisks in SpO2 window
- Test point
- SpO2 parameter window on monitor display
- Expected
- Display shows 3 asterisks (strongest signal)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unstable or slow TC sensor reading response patient safety
Sensor takes longer than expected to stabilize or reading continues to drift after expected stabilization time
Ranked causes
- 1. Sensor attachment not correct (occasional)
- 2. Measurement site not appropriate (occasional)
Verify sensor locking and fixation ring seal
- Test point
- sensor position, fixation ring rim contact
- Expected
- Sensor locked with 1/4 turn; rim fully sealed
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
X displayed in SvO2 parameter window patient safety
SvO2 parameter window displays X symbol instead of numerical value
Ranked causes
- 1. Optical module just connected with catheter in package (common)
- 2. Optical module just connected with catheter in patient (common)
Determine if catheter is in package or in patient
- Test point
- Visual inspection of catheter location
- Expected
- Catheter visibly in package or in patient
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Aqua-Knot water trap directional flow error patient safety
Water trap is inserted backward or at incorrect orientation so arrow does not point toward module, causing gas flow obstruction or condensation trapping
Ranked causes
- 1. Aqua-Knot water trap inserted with arrow pointing away from module (common)
Verify Aqua-Knot water trap arrow orientation
- Test point
- Arrow label on Aqua-Knot water trap
- Expected
- Arrow points toward SAM module (direction of gas flow into module)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Arrhy Suspend message displayed patient safety
Continuous alarm displays 'Arrhy Suspend' message; arrhythmia analysis suspended and alarms do not trigger
Ranked causes
- 1. ECG signal quality poor; 20 of last 30 seconds below threshold (common)
Verify ECG lead placement is correct
- Test point
- All ECG electrode positions
- Expected
- Leads placed per protocol; no loose or displaced electrodes
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Arterial, noninvasive, and auscultated blood pressure readings differ patient safety
Monitor displays different systolic/diastolic values across three measurement methods (direct arterial, oscillometric, auscultation)
Ranked causes
- 1. Different measurement methods inherently use different technologies and measurement sites (common)
Verify measurement locations
- Test point
- Arterial line site vs NBP cuff site
- Expected
- Direct arterial from radial/femoral; NBP from brachial artery
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SvO2 signal strength indicator patient safety
Monitor displays signal strength indicator (asterisks or dashes) in SvO2 parameter window and calibration history to assess reliability of saturation readings from Abbott catheter
Ranked causes
- 1. Inadequate light intensity from catheter (unknown)
Observe signal strength indicator display
- Test point
- SvO2 parameter window or calibration history
- Expected
- Three asterisks (***) = optimum light intensity
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Capnostat sensor reading out of range patient safety
Monitor displays CO2 value outside expected range or message indicating sensor not calibrated
Ranked causes
- 1. Sensor not calibrated to zero cell (common)
Place sensor on REF cell and verify reading
- Test point
- REF cell
- Expected
- 38 mmHg ± 2 mmHg, ± 2 mmHg
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiac artifact detected as breaths patient safety
Monitor identifies cardiac artifact as respiratory breaths; incorrect inspiration and expiration markers; waveform shows artifact marked as breath
Ranked causes
- 1. Breath detection threshold set too low (common)
Verify breath and artifact markers are correct
- Test point
- Display waveform and markers
- Expected
- Cardiac artifact and breaths correctly distinguished with separate markers
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Change primary ECG lead for improved detection
Need to optimize which ECG lead is displayed and analyzed for arrhythmia detection
Ranked causes
- 1. Current lead not optimal for pacemaker spike visualization (common)
Display alternative ECG lead
- Test point
- Monitor ECG waveform display
- Expected
- Uniform, upright pacemaker enhancement spikes visible in ECG data
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disconnect alarm triggers patient safety
Warning alarm sounds and 'DISCONNECTED' message displayed in values window with parameter name when mean pressure falls below threshold
Ranked causes
- 1. Catheter dislodged or disconnected (common)
Verify mean pressure reading
- Test point
- ART or FEM pressure display
- Expected
- 25 mmHg or higher
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG monitoring difficult with transcutaneous pacemaker patient safety
Large electromagnetic energy from external transcutaneous pacemaker interferes with ECG monitoring on the CIC monitor.
Ranked causes
- 1. Transcutaneous pacemaker energy exceeds ECG signal filtering capability (occasional)
Assess ECG waveform quality with monitor
- Test point
- Monitor ECG display
- Expected
- Clear, interpretable waveform
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
False low heart rate indicator or false asystole call patient safety
Monitor displays incorrect low heart rate or asystole alarm with pacemaker patient due to electrical overshoots in certain pacemakers.
Ranked causes
- 1. Certain pacemaker types produce electrical overshoots (occasional)
Verify pacemaker detection is activated on monitor
- Test point
- Pace program setting
- Expected
- Pacemaker detection enabled
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heart rate double counting patient safety
Monitor counts both pacemaker spikes and QRS complexes, displaying artificially elevated heart rate.
Ranked causes
- 1. Suboptimal electrode placement or pacemaker detection mode (common)
Assess heart rate display and trend
- Test point
- Monitor ECG rate meter and waveform
- Expected
- Heart rate matches actual paced rate without doubling
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ICG cable improperly connected to patient
ICG cable connectors not properly attached to sensors; left/right bundles reversed; incorrect connector sequence
Ranked causes
- 1. Connectors attached in wrong order or reversed left/right (common)
Verify left/right cable bundle orientation
- Test point
- ICG cable left and right bundles
- Expected
- Right bundle on patient's right side, left bundle on patient's left side
- Tools
- None
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ICG cable not connected to module
ICG cable not plugged into module connector; incomplete setup
Ranked causes
- 1. Cable not inserted into module connector (common)
Verify cable insertion into module connector
- Test point
- Module front connector and ICG cable
- Expected
- Cable fully inserted and latched into module connector
- Tools
- None
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ICG sensor connector snap attachment incorrect
Neck or thorax connectors snapped in wrong order or not fully secured
Ranked causes
- 1. Connector snap sequence not followed (common)
Verify neck connector sequence and engagement
- Test point
- Blue and purple neck connectors
- Expected
- Blue connector snapped above purple; both fully engaged
- Tools
- None
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate flow sensor readings or anesthetic agent leakage patient safety
Faulty connections in RM-M patient circuit cause inaccurate readings and/or leakage of anesthetic gases that can contaminate room air
Ranked causes
- 1. Improper flow sensor connection setup (common)
Inspect thin line marking on flow sensor tube
- Test point
- Patient circuit tube
- Expected
- Line positioned proximal to patient
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect ICG sensor placement
ICG sensors not positioned properly; not opposite each other or misaligned relative to anatomical landmarks
Ranked causes
- 1. Misplaced neck or thoracic sensors (common)
Verify sensor placement geometry
- Test point
- Neck and thoracic sensor positions relative to patient anatomy
- Expected
- Neck sensors below earlobe, thoracic sensors at xiphoid line, all 180° opposite, heart icons oriented toward heart
- Tools
- None
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor displays inflate balloon message after balloon was inflated patient safety
Monitor continues to prompt for balloon inflation even though balloon has already been inflated by clinician
Ranked causes
- 1. Waveform amplitude has not decreased by 30% after balloon inflation (occasional)
Measure waveform amplitude change with balloon inflated
- Test point
- PA waveform on monitor display
- Expected
- Amplitude decrease of 30% or greater after inflation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor displays processing wedge message without balloon inflation patient safety
Monitor indicates it is detecting and processing a wedge condition when clinician has not inflated the balloon
Ranked causes
- 1. Respiratory artifact on PA waveform mimics wedge condition (occasional)
Verify balloon is not inflated
- Test point
- Balloon volume indicator
- Expected
- Balloon at zero volume
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor pacemaker patient without pace mode activated patient safety
Attempted ECG monitoring of pacemaker patient without enabling pacemaker detection software.
Ranked causes
- 1. Pace program not activated at monitor (common)
Verify pace program activation status
- Test point
- Monitor pace detection setting display
- Expected
- Pacemaker detection enabled and indicated on display
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No SpO2 data displayed after connecting interface cable and sensor patient safety
SpO2 window does not appear on monitor after connecting SpO2 interface cable and sensor; no data shown
Ranked causes
- 1. Hardware failure or unrecognized/defective sensor (unknown)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No SpO2 display on monitor after Masimo sensor site change patient safety
SpO2 window does not appear on monitor after repositioning Masimo sensor to new site
Ranked causes
- 1. Masimo sensor not recognized after repositioning (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacemaker detection mode selection for optimal QRS detection patient safety
Choosing between Pace 1 and Pace 2 algorithms to minimize artifact counting while maintaining paced beat detection.
Ranked causes
- 1. Patient pacemaker characteristics vary; no single mode optimal for all (common)
Assess paced QRS detection adequacy in Pace 2 mode
- Test point
- Monitor ECG display and rate meter
- Expected
- Paced QRS complexes recognized; rate meter shows patient pacing rate
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacemaker spikes not recognized by software patient safety
Software fails to detect or recognize pacemaker spikes on the ECG display.
Ranked causes
- 1. Pace mode setting not optimal for patient pacemaker type (common)
Confirm pacemaker detection mode in use
- Test point
- Monitor pace mode setting
- Expected
- Pace 1 or Pace 2 mode selected
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
pCO2 readings lower than expected at altitude patient safety
pCO2 sensor readings appear low when device is used at high altitude locations
Ranked causes
- 1. Normal response to altitude-dependent barometric pressure (common)
Determine local barometric pressure
- Test point
- Local altitude and weather station data
- Expected
- Barometric pressure in mmHg or kPa
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor ICG signal quality at sensor
Low signal quality due to inadequate skin preparation; oily residue, dead skin cells, or abrasive particles present on sensor site
Ranked causes
- 1. Improper skin preparation (common)
Inspect sensor site for cleanliness
- Test point
- Skin surface at ICG sensor placement sites
- Expected
- No hair, dead skin cells, oily residue, or abrasive particles visible
- Tools
- None
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reduction in pCO2 sensitivity patient safety
Transcutaneous carbon dioxide (pCO2) sensor shows reduced sensitivity or inaccurate readings
Ranked causes
- 1. Residual electrolyte solution after membraning (common)
Clean sensor surface
- Test point
- Sensor surface
- Expected
- No residual electrolyte or contaminants
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reduction in pO2 electrode sensitivity patient safety
Transcutaneous oxygen (pO2) sensor shows reduced sensitivity or inaccurate readings
Ranked causes
- 1. Electrode deposits or membrane contamination (common)
Prepare fresh zero solution
- Test point
- Zero solution preparation
- Expected
- Fresh zero solution in clean beaker
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Respiration lead fail message without ECG lead fail
- Respiration monitoring on pacemaker patients interaction risk
- Respiration monitoring unreliable with limb electrodes
- Respiration monitoring with fixed-lead cables
- Service module message displayed
- Shallow breaths not detected
- Signal quality concern - monitor may not alarm on invalid reading
- Skin blister or burn from transcutaneous sensor
- Smart BP senses artifact without flush, draw, or zero
- TC sensor readings affected by anesthetic gases
- Transcutaneous pO2/pCO2 sensor application procedure
- Unable to detect PA wedge
12 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
DISCONNECTED | Catheter disconnected or dislodged from patient | Mean pressure falls below 25 mmHg when disconnect alarm feature is enabled | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI