GE Healthcare CIC Pro: troubleshooting and error codes
The ServiceForge knowledge base holds 60 documented symptoms for the GE Healthcare CIC Pro, with 74 ranked causes, 51 diagnostic checks with test points and expected values, and 27 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Time does not advance or alternates
System time jumps between multiple values or does not progress forward on CARESCAPE Network.
Ranked causes
- 1. Abrupt power cycle or UPS not connected (common)
- 2. Time zone mismatch between devices (common)
- 3. New CIC Pro center added without time sync (occasional)
- 4. Non-GE equipment on MC/IX network (rare)
- 5. DST checkbox enabled incorrectly (occasional)
Verify IP addresses and time zone settings
- Test point
- Check Centrals utility, time zone, DST, IP address
- Expected
- No errors reported; IP addresses valid; time zone consistent; DST status appropriate
10 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blank display screen caution
Monitor displays completely blank or black with no content visible.
Ranked causes
- 1. Abrupt power failure (occasional)
- 2. No UPS connected or functional (occasional)
- 3. Loose cables or connectors (common)
- 4. Internal component is loose or failed (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to print to Digital Writer
System cannot send output to digital display writer device.
Ranked causes
- 1. Digital writer not powered up (common)
- 2. Printer door is open (common)
- 3. Printer is out of paper (common)
- 4. Printer not configured correctly (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of browser functionality
Web browser no longer operates properly; applications fail to launch.
Ranked causes
- 1. Microsoft Internet Explorer parsing defect (occasional)
- 2. Security restrictions prevent application execution (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer button dimmed when viewing stored patient data
Printer control becomes unavailable in patient data viewing mode.
Ranked causes
- 1. Laser printer not configured correctly (common)
- 2. Restricted support from Bedside monitor (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not boot from USB stick
Computer fails to boot from USB-HDD device during reload process.
Ranked causes
- 1. Incorrect boot order (common)
- 2. Software image on USB stick does not match hardware (occasional)
Verify boot order sequence
- Test point
- BIOS Setup Utility boot order setting
- Expected
- USB-HDD, HDD-2, IBA GE Slot 010
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to use MultiKM feature
MultiKM (Multimouse) functionality is not available or not working.
Ranked causes
- 1. MultiKM license is missing (common)
- 2. MultiKM feature is not configured (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms silenced for out-of-unit beds at central station patient safety
Alarms for patients outside monitored unit do not sound at CIC Pro center despite enabled setflags.
Ranked causes
- 1. Alarm silencing capability exercised unintentionally despite setflags all_mvaudio enabled (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio alarm test fails or no beeping heard
CIC Alarm Sounds do not play or beeping tone not heard from speaker during audio test
Ranked causes
- 1. Speaker failure or audio signal path malfunction (unknown)
Execute Audio Test for CIC Alarm Sounds
- Test point
- Diagnostics > Preventative Maintenance > Audio Test dropdown selection
- Expected
- Beeping tone heard from speaker
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
BIOS version incorrect or unexpected
BIOS Version field displays unexpected or incorrect version string
Ranked causes
- 1. BIOS update or corruption (unknown)
Check BIOS version in Webmin
- Test point
- Diagnostics > Preventative Maintenance > Bios Information screen
- Expected
- Phoenix - AwardBIOS v6.00PG
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blue display screen
Monitor displays blue screen indicating Windows operating system error (BSOD).
Ranked causes
- 1. Windows operating system functional error (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Power leakage test high voltage
Verify device under test power leakage does not exceed specification limits
Ranked causes
- 1. Leakage current exceeds specification (unknown)
Measure power leakage current on device under test
- Test point
- Device power outlet
- Expected
- Less than or equal to specification
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot locate control/focus screen when using Multimouse
User is unable to find the control/focus screen in a Multimouse environment.
Ranked causes
- 1. Physical display layout makes screen difficult to locate visually (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Chassis Fan 1 speed out of range caution
MP100 platform fan speed outside acceptable range limits
Ranked causes
- 1. Fan obstruction, bearing wear, or speed controller malfunction (unknown)
Measure Chassis Fan 1 speed
- Test point
- Diagnostics > Logfiles
- Expected
- Within specified range (X - X) rpm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Chassis Fan 2 speed out of range caution
MP100 platform second fan speed outside acceptable limits
Ranked causes
- 1. Fan obstruction, bearing wear, or speed control failure (unknown)
Measure Chassis Fan 2 speed
- Test point
- Diagnostics > Logfiles
- Expected
- Within specified range (X - X) rpm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CIC Pro continuously reboots during USB reload
System enters reboot loop when USB reload procedure is initiated.
Ranked causes
- 1. Network cables not disconnected before reload start (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
COMM ports not properly configured
Communication ports fail to configure correctly when tested via COMM Port Test utility
Ranked causes
- 1. COMM port driver or configuration error (unknown)
Run COMM Port Test from Webmin
- Test point
- Diagnostics > Preventative Maintenance > COMM Port Test
- Expected
- All ports configured properly
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPU or enclosure temperature out of range caution
Internal CPU or enclosure temperature measured outside acceptable limits
Ranked causes
- 1. Cooling system failure or thermal sensor malfunction (unknown)
Measure CPU temperature
- Test point
- Webmin Diagnostics > Runtime Diagnostics > Temp/Voltage Info
- Expected
- 5 — 90 C, 5–90 C
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPU temperature out of range caution
CPU temperature exceeds specified operating limits; shown in system log files
Ranked causes
- 1. CPU thermal issues or sensor error (unknown)
Monitor CPU temperature in diagnostics log
- Test point
- Diagnostics > Logfiles
- Expected
- Within specified range (X - X)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display does not meet operational requirements
Display contrast, color, focus, linearity, or position does not meet proper operation specifications for the environment.
Ranked causes
- 1. Display settings misalignment or misconfiguration (common)
Verify display contrast and color
- Test point
- Display
- Expected
- Proper contrast and color for best viewing in the environment where used
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drive operation or status information abnormal
Flash drive or hard drive operation test shows abnormal status or invalid information
Ranked causes
- 1. Drive failure or firmware corruption (unknown)
Run Drive Test from Webmin
- Test point
- Diagnostics > Preventative Maintenance > Drive Test
- Expected
- Flash drive and hard drive show normal operation status
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error response from picklist server
Server returns invalid information that cannot be processed by the system.
Ranked causes
- 1. Server communication error or invalid data format (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
External speaker is unplugged
External speaker not connected or connection lost
Ranked causes
- 1. External speaker connector disconnected (common)
Verify external speaker connection
- Test point
- External speaker connector
- Expected
- Firmly seated and connected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fan A speed out of range caution
Legacy platform fan speed falls outside acceptable limits
Ranked causes
- 1. Fan obstruction or speed controller failure (unknown)
Measure Fan A speed
- Test point
- Diagnostics > Logfiles
- Expected
- Within specified range (X - X) rpm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fan B speed out of range caution
Legacy platform second fan speed falls outside acceptable limits
Ranked causes
- 1. Fan obstruction or speed controller failure (unknown)
Measure Fan B speed
- Test point
- Diagnostics > Logfiles
- Expected
- Within specified range (X - X) rpm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flash Disk Drive failure caution
MP100 platform flash disk drive has failed
Ranked causes
- 1. Flash disk electronic failure or corruption (rare)
Verify flash disk status
- Test point
- Diagnostics > Logfiles
- Expected
- No flash disk failure indication
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hard Disk Drive failure caution
Hard disk drive has failed and is not operational
Ranked causes
- 1. Hard disk mechanical or electronic failure (rare)
Monitor system operation after drive failure notice
- Test point
- System resource indicator
- Expected
- System operating in limited data store mode
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hard disk temperature out of range caution
Monitored temperature exceeds specified limits; displayed in log files during diagnostics
Ranked causes
- 1. Hard disk overheating or thermal sensor malfunction (unknown)
Verify hard disk temperature reading in log files
- Test point
- Diagnostics > Logfiles
- Expected
- Within specified range (X - X)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hard drive read/write errors detected
Disk errors found when running chkdsk on C, D, F, or G drive partitions during checkout procedure
Ranked causes
- 1. Disk corruption or bad sectors on drive (unknown)
Run chkdsk utility on all drives
- Test point
- C, D, F, G drive partitions
- Expected
- No errors found
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect display colors
Monitor displays wrong or distorted colors; color fidelity is degraded.
Ranked causes
- 1. Video card or driver issue (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect printer IP address
Printer configuration fails due to invalid or missing IP address.
Ranked causes
- 1. Printer IP address entered incorrectly or field left blank (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal speaker is unplugged
Internal speaker not connected or disconnected from system
Ranked causes
- 1. Speaker connector loose or disconnected (common)
Verify internal speaker connection
- Test point
- Internal speaker connector
- Expected
- Firmly seated and connected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid picklist query
Search request format is not valid and server cannot process it.
Ranked causes
- 1. Improperly formatted search criteria (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard keys stick or are non-functional
One or more keys on the keyboard are stuck or do not respond to key presses.
Ranked causes
- 1. Keyboard mechanical failure or contamination (common)
Test each key on the keyboard for proper function
- Test point
- Keyboard
- Expected
- All keys respond to presses and do not stick
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard keys stick or malfunction
Keys stick or do not respond to presses during normal operation
Ranked causes
- 1. Defective keyboard (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Liquid spilled into device high voltage
Device may malfunction after liquid exposure; risk of electric shock.
Ranked causes
- 1. Accidental spill of liquid into device (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mouse cursor does not follow movements or buttons malfunction
Mouse cursor movement is not smooth relative to mouse movements, or mouse buttons do not function properly.
Ranked causes
- 1. Mouse mechanical or optical failure (common)
Verify mouse cursor follows movements smoothly
- Test point
- Mouse
- Expected
- Cursor movement is smooth and follows mouse movements
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mouse cursor does not track smoothly
Cursor does not follow mouse movements smoothly or mouse buttons do not function
Ranked causes
- 1. Defective mouse (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
MultiKM license activation fails
License activation fails on one or more CIC Pro centers; red slash icon appears; application works for max one hour before timeout
Ranked causes
- 1. Failed MultiKM license on any center in group (common)
Look for failed MultiKM license icon with red slash in bottom right of display screen
- Test point
- Display screen bottom right corner
- Expected
- Failed MultiKM license icon visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
MultiKM worked for only one hour
MultiKM feature operates briefly then fails after approximately one hour of use.
Ranked causes
- 1. MultiKM enable has failed (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No matches found for picklist request
Search returns no valid results matching the entered criteria.
Ranked causes
- 1. Search criteria does not match any records (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Operating system does not boot from USB
System fails to boot from USB memory stick during reload procedure.
Ranked causes
- 1. Incorrect boot order in system BIOS (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Operating system missing after USB insert
System displays error when a non-reload USB stick is connected during boot.
Ranked causes
- 1. Wrong USB memory stick connected to CIC Pro center (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Picklist server not available
Server is not present on network or network connectivity is lost.
Ranked causes
- 1. Network connectivity failure or server offline (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer name not specified
System rejects printer configuration due to missing printer name entry.
Ranked causes
- 1. Printer name field left blank during configuration (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Processor fan speed out of acceptable range caution
Chassis Fan 1 or Fan 2 operating speed outside acceptable RPM limits
Ranked causes
- 1. Fan motor failure or speed controller malfunction (unknown)
Measure Chassis Fan 1 speed
- Test point
- Diagnostics > Preventative Maintenance > Fan Test
- Expected
- 2700 — 4000 RPM, 2700–4000 RPM
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red display screen
Monitor displays red screen during system startup or initialization.
Ranked causes
- 1. CIC Pro Application is starting up (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Request time out, cancelling request
Communication to picklist server fails and search request is cancelled.
Ranked causes
- 1. Server communication timeout or network latency (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Server off network
Hospital Information System (HIS) is not available or not connected.
Ranked causes
- 1. HIS offline or network unreachable (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Service Monitor Battery low
CIC Pro center CPU battery charge is low and requires replacement soon.
Ranked causes
- 1. CPU battery aging and low charge (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Speaker not properly connected
External or internal speaker detected as unplugged or not connected to correct socket
Ranked causes
- 1. Speaker cable unplugged or loose connection (common)
Verify external speaker connection
- Test point
- Webmin Diagnostics > Runtime Diagnostics > Closed-loop Audio Info
- Expected
- External Speaker Status = Plugged (1)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System file integrity check detects errors
Store Integrity Test finds extra, invalid, or missing system files
Ranked causes
- 1. File system corruption or software update failure (unknown)
Execute Store Integrity Test
- Test point
- Diagnostics > Preventative Maintenance > Store Integrity Test
- Expected
- No extra, invalid, or missing files detected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System resources critically low; restart required patient safety
Available system resources critically low; system restart required to restore function; patients not monitored during restart
Ranked causes
- 1. Severe resource depletion from excessive logs or memory leak (occasional)
Verify system resource indicator status
- Test point
- System resource indicator
- Expected
- Red state indicating critical condition
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System resources running low caution
Available system resources approaching limit; preventive maintenance required
Ranked causes
- 1. Accumulated log data, temporary files, or memory constraints (occasional)
Monitor system resource indicator color
- Test point
- System resource indicator
- Expected
- Green or normal state
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System voltage out of acceptable range caution
5V or 12V internal power subsystem voltage measured outside acceptable limits
Ranked causes
- 1. Power supply or voltage regulation failure (unknown)
Measure 5V system voltage
- Test point
- Webmin Diagnostics > Runtime Diagnostics > Temp/Voltage Info
- Expected
- 4850 — 5250 mV, 4850–5250 mV
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to access CIC Pro center within Multimouse group
Cannot communicate with or access a CIC Pro center in a Multimouse configuration.
Ranked causes
- 1. Incompatible software versions within Multimouse group (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to communicate with a device
CIC Pro center cannot establish communication with a connected network device.
Ranked causes
- 1. Network connectivity failure or device offline (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
USB devices not detected in loopback test
Connected USB devices (keyboard, mouse, printer) do not appear in USB Loopback Test results
Ranked causes
- 1. USB device connection or port malfunction (unknown)
Log into Webmin and run USB Loopback Test
- Test point
- Diagnostics > Preventative Maintenance > USB Loopback Test
- Expected
- All connected USB devices appear in test results
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Video test color verification fails
Video function test detects color display error or video card/driver malfunction
Ranked causes
- 1. Video card failure or display driver corruption (unknown)
Run Video Test color verification
- Test point
- Diagnostics > Preventative Maintenance > Video Test
- Expected
- All colors pass the test
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Watchdog timer test fails or system does not reboot caution
Watchdog countdown does not complete or system fails to reboot when timer reaches zero
Ranked causes
- 1. Watchdog circuit or reboot mechanism failure (unknown)
Run Watchdog Test countdown
- Test point
- Diagnostics > Preventative Maintenance > Watchdog Test screen
- Expected
- Timer counts to zero and system reboots to normal state
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
A Printer Name must be specified. | Printer configuration missing required name field | Attempt to configure printer without entering name | error |
Chassis Fan 1 speed out of range | MP100 platform Chassis Fan 1 speed outside acceptable range | Fan speed measurement falls outside specified limits during system operation | warning |
Chassis Fan 2 speed out of range | MP100 platform Chassis Fan 2 speed outside acceptable range | Fan speed measurement falls outside specified limits during system operation | warning |
CPU temperature out of range | CPU temperature has exceeded acceptable operating limits | CPU temperature rises above or falls below specified range during operation | warning |
Error response returned from picklist server | Server returned invalid or unusable data | Search query submitted to picklist server | warning |
External speaker is unplugged | External speaker is disconnected or not detected by system | System detects missing external speaker connection during startup or operation | warning |
Fan A speed out of range | Legacy platform Fan A speed outside acceptable range | Fan A speed measurement falls outside specified limits during operation | warning |
Fan B speed out of range | Legacy platform Fan B speed outside acceptable range | Fan B speed measurement falls outside specified limits during operation | warning |
Flash Disk Drive failure | MP100 platform flash disk drive has failed | Flash disk failure detected during system operation; drive not accessible | error |
Hard Disk Drive failure | Hard disk drive has failed and is not accessible | Hard disk failure detected during system boot or operation; drive no longer responds | error |
Hard Disk temperature out of range | Hard disk temperature has exceeded acceptable operating limits | Hard disk temperature rises above or falls below specified range during normal operation | warning |
image cannot be restored | System cannot restore from USB image due to mismatch or corruption | Boot from USB stick fails with restoration error | error |
Internal speaker is unplugged | Internal speaker is disconnected or not detected by system | System detects missing internal speaker connection during startup or operation | warning |
Invalid picklist query | Search request format is invalid | User submits improperly formatted search criteria | error |
No matches found for the picklist request | Search produced no results | Search query executed but found no matching records | info |
Operating system does not boot from the USB memory stick | System fails to load OS from USB device | System attempts to boot from USB | critical |
Operating system missing | System cannot find or boot operating system | Non-reload USB stick connected during boot | critical |
Operating system not found | Boot process cannot locate operating system on target device | System attempts boot and finds no valid OS | error |
Picklist server is not available | Server cannot be reached on network | Search query attempts to reach picklist server | error |
Platform type test failed | Software image on USB stick does not match CIC Pro hardware | System attempts to boot from USB stick with incompatible image | error |
Request time out, cancelling request | Communication to server exceeded time limit | Search request sent to picklist server | warning |
Server off network | Hospital Information System not available | System attempts to communicate with HIS | error |
Service Monitor Battery | CPU battery low charge requiring replacement | Periodic battery check by system | warning |
System resources critically low; restart required | Available system resources critically depleted; immediate system restart required | System resource usage reaches critical level; patients are not monitored during restart period | critical |
System resources running low | Available system resources approaching critical limit; preventive maintenance needed | System resource usage exceeds warning threshold during normal operation | warning |
There is a problem with your IP address. | Printer IP address invalid or missing | Attempt to configure printer network | error |
unknown platform image | System detects image with unrecognized platform type | Image validation check fails during boot | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI