GE Healthcare CASE: troubleshooting and error codes
The ServiceForge knowledge base holds 47 documented symptoms for the GE Healthcare CASE, with 73 ranked causes, 72 diagnostic checks with test points and expected values, and 1 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
ESD damage to internal components during service caution
Component failure or malfunction caused by electrostatic discharge exposure when servicing the device. External connectors have ESD protection, but internal components are susceptible.
Ranked causes
- 1. Improper handling of semiconductors or assemblies without ESD protection (common)
- 2. Use of ungrounded soldering or test equipment (occasional)
- 3. Removing semiconductors from antistatic containers prematurely (occasional)
- 4. Power applied to assembly during semiconductor removal or insertion (occasional)
Verify antistatic wristband is worn
- Test point
- Technician wrist
- Expected
- Grounded antistatic wristband (3M 2046 or equivalent) in contact with skin
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image is unstable, unfocused, or swimming
Display appears blurry, unstable, or with visible distortion and swimming motion
Ranked causes
- 1. Signal cable not fully attached (common)
- 2. OSM Image Adjust settings need adjustment (common)
- 3. Monitor not set to recommended signal timings (occasional)
- 4. Interlaced video mode or low refresh rate (occasional)
Verify signal cable fully attached
- Test point
- Monitor and computer signal cable connectors
- Expected
- Cable completely and firmly seated
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No picture on monitor
Monitor display is blank or dark with no image visible
Ranked causes
- 1. Signal cable not fully connected to monitor (common)
- 2. Vacation Switch in OFF position (common)
- 3. Power switches in OFF position (common)
- 4. Monitor not set to recommended video settings (occasional)
Verify signal cable is fully connected
- Test point
- Monitor and computer signal cable connectors
- Expected
- Cable fully seated with no bent or pushed-in pins
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error when reading examinations or decompression error
Error messages appear when opening examinations, examinations get lost, or decompression errors occur when retrieving tests from archive
Ranked causes
- 1. Read error on SSD (occasional)
- 2. Virus corruption on SSD (occasional)
- 3. Defective SSD (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
First Setup of Windows 8.1 Pro for CASE after repair or new installation
Initial configuration steps required after Windows 8.1 Pro installation or repair on CASE system
Ranked causes
- 1. New or repaired Windows installation requires configuration (common)
- 2. Admin login required (common)
- 3. System post-configuration verification and application installation (common)
Wait for initial system startup and setup screen
- Test point
- System startup
- Expected
- Country/region, App language, and Keyboard layout selection screen displays
7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Laser printer prints numbers instead of grid
Laser printer output shows numbers instead of expected grid patterns.
Ranked causes
- 1. Printer scheduling not set to direct printing (common)
- 2. Grid removal setting not configured in LOC_WIN.INI (common)
- 3. Incompatible character sets enabled (occasional)
Verify printer scheduling mode
- Test point
- Printer properties menu
- Expected
- Printer scheduling set to Print direct
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Windows 8.1 Pro running with major functionality issues
Windows operating system is running but exhibiting major performance or functionality problems
Ranked causes
- 1. System restore point available (common)
- 2. Windows installation corrupted and system bootable (occasional)
- 3. SSD defective (rare)
Attempt System Restore to earlier point in time
- Test point
- Windows system restore functionality
- Expected
- System restores to stable earlier configuration
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Windows 8.1 Pro system requires repair and software reinstallation
System needs CASE 6.73 application software extracted to SSD and Windows 8.1 Pro reinstalled with proper configuration
Ranked causes
- 1. System requires Windows 8.1 Pro repair procedure (unknown)
- 2. CASE V6.73 application installation required (unknown)
- 3. InSite software installation needed after successful repair (unknown)
Verify DVD or USB stick contents match original source
- Test point
- CRCUTIL.EXE checksum comparison
- Expected
- Checksums match between source and copied media
- Tools
- CRCUTIL.EXE
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Btrieve database error on patient test storage
System unable to store patient test data; may follow improper shutdown or power interruption
Ranked causes
- 1. Improper system shutdown (common)
- 2. Facility power interruption (common)
Log on as Administrator
- Test point
- System login
- Expected
- Desktop with icons (My Computer, Network Neighborhood, etc.) appears
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display image not sized properly
Image appears too large, too small, or not scaled to fill the screen properly
Ranked causes
- 1. OSM coarse adjustment needs adjustment (common)
- 2. Unsupported graphics mode selected (occasional)
Adjust OSM coarse sizing control
- Test point
- OSM Image Adjust menu on monitor
- Expected
- Image resized to fill screen properly
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG acquisition module malfunction
ECG signal not being acquired or displayed correctly in CASE system
Ranked causes
- 1. Disconnected or faulty ECG lead (common)
- 2. CAM-USB acquisition driver failure (occasional)
Check lead impedance measurement
- Test point
- Lead Status window in Acquisition Test
- Expected
- Impedance range 0-999
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No video output to monitor
No image appears on screen; monitor shows no video signal despite being powered on
Ranked causes
- 1. Monitor power needs reset (common)
- 2. Computer in power-saving mode (common)
Power cycle the monitor
- Test point
- Monitor Power button
- Expected
- Monitor turns off then back on and video appears
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power button does not respond
Monitor power button is unresponsive to presses
Ranked causes
- 1. Monitor needs power reset (common)
- 2. Vacation Switch preventing power-on (occasional)
Power cycle the monitor
- Test point
- Power cord and outlet
- Expected
- Monitor resets and power button becomes responsive
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thermal writer not operating
Thermal writer fails to print test graphs or does not respond during Print or Roller test
Ranked causes
- 1. Registry error when running writer test as Admin or GEService user (common)
- 2. Print head resistance out of adjustment (occasional)
Run Speed Test to verify print operation
- Test point
- Writer Speed setting in WRTR TEST
- Expected
- Test graph prints at selected speed with consistent density
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit displays BTRV errors
System shows BTRV error messages, typically after data corruption or power failure
Ranked causes
- 1. Corrupted BTRIEVE database files (occasional)
- 2. Irreparable database corruption (rare)
Verify sufficient free SSD space for reorganization
- Test point
- SSD storage
- Expected
- Free space ≥ 2 × (sum of all BTR file sizes)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Administrator password entry blocked
Password dialog does not accept administrator password due to non-Latin keyboard language settings
Ranked causes
- 1. Non-Latin keyboard language active during CASE startup (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Application appears to lock up at Test End
Program becomes unresponsive during test end in Exercise Test; error message hidden behind acquisition window
Ranked causes
- 1. Damaged database with hidden error message box (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Application appears to lock up while printing
System becomes unresponsive during printing to thermal printer or taskbar shows 'document failed to print'
Ranked causes
- 1. Print job stuck in queue on Marquette thermal printer (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CASE V6.73 application installation
Installation of CASE ADVANTECH V6.73 application software on the system after Windows setup
Ranked causes
- 1. CASE V6.73 application installation required (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Catastrophic system failure requiring SSD replacement
System has failed catastrophically and requires replacement of the Solid State Drive (SSD) with the SSD FRU containing CASE Windows 8.1 custom image and CASE ADVANTECH v6.73 application
Ranked causes
- 1. SSD failure (unknown)
Verify SSD FRU kit contents before beginning repair
- Test point
- FRU kit PN 2102160-001
- Expected
- Kit contains 2104417-001 (Pre-programmed SSD with Windows 8.1) and 2022587-025 (DVD CASE ADVANTECH V6.73)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Characters overlap in printout
Some characters overlap the characters that follow them in printed output.
Ranked causes
- 1. Print Optimizations enabled in printer settings (common)
Verify Print Optimizations setting
- Test point
- Printer Properties > General > Printing Preferences > Advanced > Document Options
- Expected
- Print Optimizations disabled
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Desk light malfunction
Desk light under monitor does not turn on or off when desk light button is pressed
Ranked causes
- 1. Faulty desk light button on keypad (unknown)
Verify desk light toggles on and off
- Test point
- Desk light button and monitor light
- Expected
- Light under monitor turns on when button pressed, turns off when pressed again
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display Arm replacement needed caution
Procedure to replace the Display Arm assembly on the CASE V6.7 Exercise Testing System
Ranked causes
- 1. Display Arm component failure or damage (unknown)
Verify wheels are locked before beginning work
- Test point
- Cart wheels
- Expected
- Locked position
8 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrical safety testing caution
Electrical safety test verification performed after system setup to ensure compliance and safety
Ranked causes
- 1. Safety compliance verification required (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error [11007] starting print job
Error message [11007] appears when attempting to print from CASE
Ranked causes
- 1. No default printer defined (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error message when CASE patient list invoked with MUSE Web Server
Error message displays when attempting to invoke CASE patient list even though MUSE Web Server is running
Ranked causes
- 1. MUSE Web user account not created or not assigned to MUSE group (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ESD precautions during Display Arm replacement caution
Electrostatic discharge prevention measures during semiconductor and electronic component handling
Ranked causes
- 1. Electrostatic discharge (ESD) damage to semiconductors and electronic assemblies (occasional)
Verify antistatic protection during storage and handling
- Test point
- Parts and assemblies
- Expected
- All semiconductors in antistatic bags or boxes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Functional checkout and verification
Complete functional testing to verify system operation after SSD replacement and configuration
Ranked causes
- 1. System functionality verification required (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image persistence visible on screen
Residual or ghost image of previous display remains visible; faint remnant of old image lingers on LCD monitor
Ranked causes
- 1. Static image displayed for extended time on LCD (common)
Determine how long previous image was displayed
- Test point
- User observation of display history
- Expected
- Duration of previous static image identified
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
InSite installation and configuration
Installation of InSite configuration management software after CASE application installation
Ranked causes
- 1. InSite software installation required (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad button malfunction
One or more keypad buttons not responding when pressed during operation
Ranked causes
- 1. Physical keypad contact failure (unknown)
Test each keypad button response
- Test point
- Keypad and interactive screen display
- Expected
- Each keypad button produces corresponding response on interactive screen
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Logon to repaired CASE system and verify installation
After system reboot, user must logon and verify CASE 6.73 application launch successfully
Ranked causes
- 1. System reboot requires logon with correct credentials (unknown)
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of configurations after SSD replacement caution
Custom configurations, user names, network settings, and option codes will be lost during SSD replacement procedure
Ranked causes
- 1. SSD replacement procedure erases custom settings (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Metabolic cart displays higher heart rate than CASE
Heart rate value shown on metabolic cart equipment is higher than the corresponding value in CASE during exercise tests
Ranked causes
- 1. Different heart rate calculation and transmission intervals (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor LED indicator not lit
LED on monitor displays no green or amber light indicator; power status light is dark
Ranked causes
- 1. Power switch in OFF position or power cord not connected (common)
Verify power cord is connected
- Test point
- Monitor power cord at monitor and AC outlet
- Expected
- Power cord fully seated at both ends
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor picture too dark in dark environment
22 inch monitor display becomes excessively dark when used in dark ambient lighting conditions
Ranked causes
- 1. Auto EcoView mode enabled (common)
Disable Auto EcoView mode
- Test point
- Leftmost EcoView button on front of 22 inch monitor
- Expected
- Auto EcoView mode set to Off
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor replacement needed caution
Procedure to remove and prepare flat panel monitor for replacement on CASE V6.7 system
Ranked causes
- 1. Monitor failure or damage requiring replacement (unknown)
Disconnect all cables from monitor back
- Test point
- Back of Monitor
- Expected
- All cables disconnected
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper out message displayed from laser printer
Paper out message appears on display from laser printer, causing delay or timeout during printing.
Ranked causes
- 1. Transmission Retry Timeout set too high (common)
Adjust Transmission Retry Timeout
- Test point
- Windows Printer Settings > Port tab > Configure Port
- Expected
- Transmission Retry Timeout reduced to minimize paper out delay
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printout incomplete
Printer output is incomplete or truncated.
Ranked causes
- 1. Laser printer insufficient memory (common)
Check printer memory capacity
- Test point
- Printer specifications or settings
- Expected
- Minimum 8 MB memory available
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Remote View tool unavailable for simultaneous use
More than one user attempts to view the same stress test from separate client PCs using Remote View tool
Ranked causes
- 1. Remote View single-access limitation by design (unknown)
RL-electrode always indicated as OK
RL-electrode status display shows green/OK even when checking electrode connections
Ranked causes
- 1. RL-electrode is output electrode, not input electrode (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Small differences in measurement values after data transfer from MAC 1200
Measurement values from MAC 1200 transferred to CASE show slight differences from original values
Ranked causes
- 1. Data conversion uses different amplitude units (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SSD replacement and Windows 8.1 setup
System requires SSD replacement with pre-programmed image and Windows 8.1 setup followed by CASE application installation and configuration
Ranked causes
- 1. SSD FRU installation required (unknown)
Prepare tools and materials
- Test point
- Tools required
- Expected
- External DVD drive and Phillips-head screwdriver available
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Target load not calculated
Program fails to calculate target load for stress test
Ranked causes
- 1. Patient demographics outside acceptable ranges (occasional)
Verify patient age is within range
- Test point
- Patient date of birth
- Expected
- age 15–84 years
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Test appears open at another system
System displays message indicating test is open on another CASE or client PC when users confirm it is not
Ranked causes
- 1. Stale network lock entry (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Windows 8.1 Pro fails to start correctly
System does not boot into Windows 8.1 Pro properly
Ranked causes
- 1. Startup corruption (unknown)
Attempt Windows 8.1 Pro startup
- Test point
- System boot process
- Expected
- Windows 8.1 Pro starts and reaches login screen
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Windows 8.1 Pro initial setup for CASE
First-time setup of Windows 8.1 Pro operating system on the CASE system after SSD replacement
Ranked causes
- 1. Windows 8.1 Pro initial configuration required (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
11007 | Error starting print job on CASE | Attempt to print from CASE application | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community