GE Healthcare Carestation 650: troubleshooting and error codes
The ServiceForge knowledge base holds 187 documented symptoms for the GE Healthcare Carestation 650, with 291 ranked causes, 434 diagnostic checks with test points and expected values, and 145 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Vent engine component seal degradation caution
Loss of gas tightness or component leakage due to worn seals or o-rings
Ranked causes
- 1. Gas inlet valve solenoid seal worn or damaged (occasional)
- 2. Gas inlet valve o-ring worn or damaged (occasional)
- 3. Interface manifold o-rings degraded (occasional)
- 4. Drive gas regulator o-rings worn (occasional)
- 5. Inspiratory flow control valve o-rings worn (occasional)
- 6. Flow sensor o-rings degraded (occasional)
- 7. Free breathing check valve o-ring worn (occasional)
- 8. Reservoir o-rings degraded (occasional)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Breathing system leak from canister or valves patient safety
Gas leaking from absorber canister, check valves, APL valve, breathing tubes, or negative pressure relief valve
Ranked causes
- 1. Absorber canister open or missing (common)
- 2. Damaged or missing canister o-ring or seals (occasional)
- 3. APL valve not closed (common)
- 4. Breathing system tubes damaged or connectors loose (occasional)
- 5. Check valve of inspiratory or expiratory ports not correctly installed (occasional)
- 6. Negative pressure relief valve leak (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check flow sensor failure patient safety
During mechanical breaths, measured flow to and from patient does not meet criteria; no or negative flow on inspiration or expiration sensor.
Ranked causes
- 1. Flow sensor connection loose or disconnected (common)
- 2. Breathing circuit issue (common)
- 3. SIB sensor tubing leak (occasional)
- 4. Flow sensor calibration out of range (occasional)
- 5. Insp/Exp check valve failure (occasional)
- 6. Flow sensor defective (occasional)
Verify flow sensor connections are seated
- Test point
- Flow sensor connectors
- Expected
- Fully plugged in
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High pressure leak detected high voltage
Gas leaking from high-pressure system components (pipeline, O2 flush valve, cylinders, transducers, gaskets, relief valves)
Ranked causes
- 1. Pipeline leak (common)
- 2. O2 flush valve malfunction (common)
- 3. Cylinder not installed properly (occasional)
- 4. Cylinder transducer leak (occasional)
- 5. Cylinder gasket failure (occasional)
- 6. Relief valve malfunction (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator leak test – cannot empty bellows patient safety
System cannot empty bellows during ventilator leak test (manifold pressure must be ≥60 cmH2O).
Ranked causes
- 1. Patient Y not open (common)
- 2. Drive gas leak from bellows housing to atmosphere (occasional)
- 3. Bellows installation error or bellows leak (occasional)
- 4. Pop-off valve leak (occasional)
- 5. Ventilator exhalation valve leak (occasional)
- 6. Drive Gas Check Valve stuck open (rare)
Set Bag/Vent switch to Vent, open patient Y, select Confirm
- Test point
- Bellows emptying
- Expected
- Bellows empty completely
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Circuit leak test – cannot pressurize patient safety
Breathing circuit leak test cannot pressurize circuit during testing.
Ranked causes
- 1. Test bag not properly occluded or test bag leaks (common)
- 2. Leak in canister or CO2 bypass (occasional)
- 3. Leak in O2 cell (occasional)
- 4. Leak in patient port or Y piece (occasional)
- 5. Leak in APL assembly (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
DCB procedure com fail patient safety
Anesthesia Computer loses communication with Display Computer; procedure messages missing or mismatched for >15 seconds.
Ranked causes
- 1. ACB to FIB or SIB communication cable disconnected (common)
- 2. Display Unit cable connection loose (occasional)
- 3. Display Controller board failure (occasional)
- 4. ACB failure or disconnection (occasional)
Reboot system
- Test point
- System startup
- Expected
- ACB and DCB re-establish communication
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas inlet valve bootup fail patient safety
Gas inlet valve solenoid self-test fails at boot-up; ventilator cannot control gas delivery.
Ranked causes
- 1. Gas inlet valve solenoid connector disconnected or loose (common)
- 2. Gas inlet valve solenoid defective (occasional)
- 3. Gas inlet valve mechanical failure (occasional)
- 4. ACB solenoid drive circuit failure (rare)
Verify gas inlet valve solenoid connector is seated
- Test point
- Gas inlet valve solenoid connector on ACB
- Expected
- Fully plugged in, no loose pins
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leak in both bag and vent modes patient safety
Gas escapes from breathing circuit in both bag and vent operational modes
Ranked causes
- 1. Canister seal, body, or cover defective (common)
- 2. Flow sensor or Paw transducer port leakage on SIB board (common)
- 3. Patient ports assembly defective (common)
- 4. Canister or circuit board connectors, filters, or tubes defective (occasional)
Perform Test 7 for canister leakage
- Test point
- Canister handle port
- Expected
- Airway Pressure rises to >=30 cmH2O at 0.1 L/min O2 flow
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of ventilator parameter settings patient safety
Ventilator parameter settings from Display Computer not received for >10 seconds; ventilator operates in monitoring-only mode.
Ranked causes
- 1. ACB to FIB or SIB communication cable disconnected (common)
- 2. Display Unit cable connection loose (occasional)
- 3. Display Controller board failure (occasional)
- 4. ACB connection or failure (occasional)
Reboot system
- Test point
- System startup and communication re-establishment
- Expected
- ACB receives vent parameters within 10 seconds
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low pressure leak with vaporizer mounted caution
Gas leaking from low-pressure system when vaporizer is installed; includes vaporizer installation, o-ring, fill port issues
Ranked causes
- 1. Vaporizer not installed properly (common)
- 2. Vaporizer leaks (occasional)
- 3. Missing or damaged o-ring on vaporizer manifold (occasional)
- 4. Loose fill port (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator circuit leak exceeds 750 ml/min patient safety
During ventilator leak test, detected leak rate exceeds 750 ml/min threshold.
Ranked causes
- 1. Leak in canister or CO2 bypass (common)
- 2. Leak in O2 cell (common)
- 3. Leak in patient port or Y piece (common)
- 4. Leak in sample gas return line or gas return port inside bellows (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator leak test – cannot pressurize circuit patient safety
During ventilator leak test, circuit cannot be pressurized to required level.
Ranked causes
- 1. Leak in canister or CO2 bypass (common)
- 2. Leak in O2 cell (common)
- 3. Leak in patient port or Y piece (common)
- 4. Leak in sample gas return or gas return port inside bellows (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bellows leak or dysfunction patient safety
Bellows not sealing properly, mounting issues, or physical damage affecting ventilation
Ranked causes
- 1. Pop-off valve diaphragm not sealing properly (occasional)
- 2. Bellows mounting rim loose (occasional)
- 3. Bellows improperly mounted or has hole or tear (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bellows over inflate patient safety
Bellows bulging excessively during operation due to exhaust blockage or O2 flush malfunction
Ranked causes
- 1. Exhaust outlet is blocked (occasional)
- 2. O2 flush button is malfunctioning (occasional)
- 3. Flow mixer module is malfunctioning (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Circuit leak test – timed out patient safety
Breathing circuit leak test times out without establishing required pressure.
Ranked causes
- 1. Unsuitable test bag (too large or too soft) (common)
- 2. Leak inside O2 flush valve manifold (occasional)
- 3. Leak from O2 bypass, latch, N2O selector, or Air selector valve (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Com error ventilator to ACB patient safety
No communication between Ventilator boundary object and SIB CPU; loss of ventilator function.
Ranked causes
- 1. Communication cable disconnected or damaged (common)
- 2. SIB module failure (occasional)
- 3. ACB failure (occasional)
Reboot system
- Test point
- System startup
- Expected
- System boots and establishes communication within 35ms
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPU fan speed fail caution
CPU cooling fan speed drops below 2500 RPM (50% nominal), indicating fan failure or obstruction.
Ranked causes
- 1. CPU heatsink fan not operating (common)
- 2. Fan connector disconnected (common)
- 3. Fan defective (common)
Verify CPU fan is spinning
- Test point
- CPU heatsink fan visual inspection
- Expected
- Fan rotating
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPU overheat caution
CPU thermal diode temperature exceeds 80°C, indicating insufficient cooling.
Ranked causes
- 1. CPU fan not operating (common)
- 2. Fan connector disconnected (common)
- 3. Fan defective (common)
Check CPU fan airflow
- Test point
- Back of Display Unit fan outlet
- Expected
- Warm air flowing from fan filter opening
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display arm positioning incorrect caution
Display arm not properly positioned side-to-side or at wrong tilt angle; may move with finger pressure or drop by own weight
Ranked causes
- 1. Arm mounting nut loose (common)
- 2. Display mount attachment nut loose (common)
- 3. Axis bar clamping screws insufficiently tight (common)
Display moves when finger pressure applied to left side
- Test point
- Left side of display
- Expected
- Display does not move
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fan fail caution
Cooling fan power status bit is low; fan needs service but system operation OK.
Ranked causes
- 1. Fan connector loose or disconnected (common)
- 2. Cooling fan defective (occasional)
- 3. Power Management Board (PMB) failure (rare)
Check cooling fan connector
- Test point
- Fan power connector on motherboard or PMB
- Expected
- Fully seated, no loose pins
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leak in bag mode only patient safety
Gas escapes from breathing circuit during bag mode operation
Ranked causes
- 1. APL valve or diaphragm assembly defective (common)
- 2. BTV assembly, bag port, or negative valve defective (common)
- 3. Breathing circuit components improperly installed (common)
Verify breathing circuit assembly installation
- Test point
- APL valve, Bag/Vent Switch, Cross Rubber
- Expected
- All components properly seated and connected
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leak in vent mode only patient safety
Gas escapes from breathing circuit during vent mode operation
Ranked causes
- 1. U-Cup seal defective in bellows housing (common)
- 2. Pop-off diaphragm defective or bellows base port damaged (common)
- 3. BTV rubber, manifold, or O-ring seals defective (occasional)
Perform Test 6 for bellows housing leakage
- Test point
- BTV port on upper BC assembly base
- Expected
- Airway Pressure rises to >=30 cmH2O at 0.1 L/min O2 flow
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low pressure leak with or without vaporizer caution
Gas leaking from low-pressure system; includes port valve, gas control module, flush valve, bellows issues
Ranked causes
- 1. Leaking port valve on vaporizer manifold (common)
- 2. Leak in gas control module (occasional)
- 3. Leaking flush valve (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Right top task light not functioning
Right task light does not illuminate, dimmer switch inoperative, or light is defective
Ranked causes
- 1. Defective right top task light board (common)
- 2. Defective task lights dimmer switch (occasional)
- 3. Defective right top task light harness (occasional)
Verify right top task light illuminates
- Test point
- Right top task light
- Expected
- Light is lit
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to begin mechanical ventilation patient safety
System cannot initiate ventilation mode; breathing system not engaged or drive gas unavailable
Ranked causes
- 1. Breathing system not fully engaged (common)
- 2. No drive gas supply (occasional)
- 3. Defective Bag/Vent switch (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vaporizer leak test failure patient safety
Vaporizer leak test detects leak rate exceeding 750 ml/min.
Ranked causes
- 1. Vaporizer port O-ring damaged or missing (common)
- 2. Nipple or vaporizer manifold port leak (occasional)
- 3. Vaporizer internal leak (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator leak test – circuit pressure too high patient safety
During ventilator leak test, circuit pressure is higher than expected when patient Y is occluded.
Ranked causes
- 1. Drive gas leak from bellows into bellows due to pop-off or bellows failure (occasional)
- 2. Leak inside O2 flush valve manifold (occasional)
- 3. Leak from O2 bypass, O2 latch, N2O selector, or Air selector valve (occasional)
Occlude patient Y and select Confirm during ventilator leak test
- Test point
- Circuit pressure
- Expected
- Within normal range
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bag/Vent switch or CO2 bypass defective patient safety
Bag/vent mode not switching, ventilation mode malfunction, or CO2 bypass stuck
Ranked causes
- 1. Bag/Vent switch assembly defective (occasional)
- 2. CO2 bypass stuck or defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bellows fills slowly or rapidly
O2 Flush button depressed continuously; bellows fill time outside expected range (1.3–3 seconds)
Ranked causes
- 1. Large leak in breathing system (common)
- 2. ACGO selector valve not correctly connected (occasional)
Measure bellows fill time with O2 Flush button fully depressed
- Test point
- patient port
- Expected
- 1.3 to 3 seconds
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Canister holder assembly damaged
Canister release lever broken, canister support base cracked, or holder not securing canister
Ranked causes
- 1. Canister release lever broken or bent (occasional)
- 2. Canister support base cracked or deformed (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drive gas flow sensor calibration zero fail
Drive gas flow sensor power-on self-test failed; sensor unable to calibrate zero offset.
Ranked causes
- 1. Flow sensor connection loose or disconnected (common)
- 2. Flow sensor defective (occasional)
Verify drive gas flow sensor connector is seated
- Test point
- Drive gas flow sensor connector
- Expected
- Fully plugged in, no loose pins
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drive gas flow sensor fail
Anesthesia Computer detects drive gas flow sensor failure; ventilator operates in monitoring-only mode.
Ranked causes
- 1. Flow sensor connection loose or disconnected (common)
- 2. Flow sensor defective (common)
Verify flow sensor connector is seated
- Test point
- Drive gas flow sensor connector
- Expected
- Fully plugged in
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drive gas lost patient safety
Ventilator has no drive gas; O2 or Air supply pressure is low during mechanical ventilation.
Ranked causes
- 1. O2 supply disconnected or depleted (common)
- 2. Air supply disconnected or depleted (common)
Verify O2 or Air supply is connected and pressurized
- Test point
- O2 or Air supply inlet
- Expected
- Pressure gauge above minimum (see O2 PRESS LOW or AIR PRES LOW troubleshooting)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
DU-ACB communication failure patient safety
System Malfunction screen displayed; ACB COM FAIL or ACB DCB COM FAIL error; loss of communication between Display Unit and Anesthesia Control Board
Ranked causes
- 1. Cable connection between ACB and DU loose or faulty (common)
- 2. ACB TXD/RXD LED failure or ACB internal communication error (occasional)
Visually inspect cable connection between ACB and DU
- Test point
- RXD (DS5), TXD (DS4) connectors
- Expected
- Connectors seated firmly and undamaged
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow control system out of specification patient safety
Flow control test fails; gas proportioning or individual flow outputs do not match expected values
Ranked causes
- 1. Flow control needle valve calibration drift (occasional)
- 2. Link proportioning system calibration drift (occasional)
Increase N2O flow from minimum and verify O2 concentration or flow tracking
- Test point
- O2 concentration display or Mixer Output subsystem
- Expected
- O2 concentration 21% minimum at 0.2 l/min N2O; 21-34% at 1.6 l/min; 21-34% at 5.0 l/min; 21-34% at 9.0 l/min
8 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow sensor reading out of specification caution
O2, Air, or N2O flow sensor readings outside expected limits during service test
Ranked causes
- 1. Flowmeter test device not backpressure compensated (occasional)
- 2. Flow sensor calibration drift (occasional)
Set O2 flow to 1.5 l/min and verify sensor reading
- Test point
- O2 flow sensor display
- Expected
- 1.3 to 1.7 l/min, 1.3 to 1.7 l/min
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas flow > 300 ml/min during checkout caution
Gas flow exceeds 300 ml/min while system is in Checkout: General sub-state; operator must turn off gas flow.
Ranked causes
- 1. Mixer flow sensor zero offset out of calibration (common)
- 2. Needle valve adjustment out of spec (occasional)
Turn off gas flow
- Test point
- Gas inlet to mixer
- Expected
- Gas flow stops
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas inlet valve shuttle assembly degradation caution
Gas inlet valve shuttle not functioning smoothly due to missing or degraded seals
Ranked causes
- 1. Upper or lower u-cup seal missing or degraded on shuttle (occasional)
- 2. Shuttle o-ring seal worn or missing (occasional)
Inspect shuttle o-rings and u-cup seals
- Test point
- Gas inlet valve shuttle assembly
- Expected
- Seals intact and not degraded
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas module sensor zeroing in progress or not responding caution
Gas Monitor service page displays warming message, no message, or no-gas-module-detected message indicating sensor status during initialization
Ranked causes
- 1. Gas module connected but sensors still warming or zeroing (occasional)
- 2. Gas module not detected or disconnected (occasional)
Access Gas Monitor Service and observe sensor warming status message
- Test point
- Gas Monitor Service page, top left corner
- Expected
- Either warming message or no message (connected), not no-detected message
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ground resistance exceeds limit high voltage
Resistance between exposed metal surface and ground pin on power cord is ≥ 0.20 Ω
Ranked causes
- 1. Corroded or loose ground connection on power cord (common)
- 2. Defective power cord (occasional)
Set safety analyzer to measure line voltage
- Test point
- safety analyzer
- Expected
- normal polarity indicated
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspiratory/expiratory check valve stuck or defective patient safety
One-way valve not sealing or stuck closed, causing improper gas flow direction
Ranked causes
- 1. Inspiratory check valve defective (rare)
- 2. Expiratory check valve defective (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leak test device integrity verification failure
Leak test tool kit is leaking or cannot hold pressure, invalidating test results
Ranked causes
- 1. Low-pressure leak test device bulb does not inflate or leaks (common)
- 2. Tool kit (2091916-001) has leaking seals (common)
Verify low-pressure leak test device integrity
- Test point
- Leak test device inlet and bulb
- Expected
- Bulb takes more than 60 seconds to fully inflate when hand seals inlet and bulb is squeezed empty, >60 sec
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Left top task light not functioning
Left task light does not illuminate or is defective
Ranked causes
- 1. Defective left top task light board (common)
- 2. Defective left top task light harness (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manifold PAW sensor fail
Calibration failure at boot-up; manifold pressure sensor unable to zero offset properly.
Ranked causes
- 1. Manifold pressure sensor tubing occluded (common)
- 2. ACB manifold sensor circuit failure (rare)
Verify manifold flow sensor counts in Service Software
- Test point
- Manifold Flow counts in Vent Flow and Pressure section 12.9.2
- Expected
- Between 2575 and 4447 counts
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
N2O mixer hardware failure
Gas control test detects N2O mixer hardware failure.
Ranked causes
- 1. O2 bypass or latch valve failure (common)
- 2. O2 latch or N2O selector valve failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 flow decreases excessively when vaporizer adjusted caution
O2 flow drops more than 1 l/min when vaporizer concentration is adjusted from 0 to 1%
Ranked causes
- 1. Malfunction in vaporizer back pressure mechanism (common)
- 2. Malfunction in anesthesia system vaporizer manifold port valve (occasional)
Set O2 flow to 6 l/min and adjust vaporizer from 0 to 1%
- Test point
- O2 flowmeter
- Expected
- O2 flow must not decrease more than 1 l/min, ≤1 l/min decrease
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SIB cable connection improper caution
SIB cable loose, disconnected, or interfering with breathing system assembly
Ranked causes
- 1. SIB cable not fully seated or connected (common)
- 2. SIB cable routing interferes with breathing system assembly (occasional)
Verify SIB-to-ACB cable is fully seated and secure
- Test point
- SIB cable connector at ACB
- Expected
- Cable fully connected with no play or movement
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SIB filter installation incorrect caution
Filters not installed in proper sequence or orientation
Ranked causes
- 1. Filters installed out of sequence (common)
- 2. Filter installed with incorrect orientation (occasional)
Verify filter installation sequence
- Test point
- SIB filter assembly
- Expected
- Filters a and c installed first, then b, d, e in sequence
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vaporizer back pressure exceeds limit
O2 flow decreases more than 1 l/min when vaporizer concentration adjusted from 0 to 1%
Ranked causes
- 1. Defective vaporizer installed (common)
- 2. Malfunction in anesthesia system vaporizer manifold (occasional)
Verify O2 flow at 6 l/min baseline
- Test point
- O2 outlet
- Expected
- 6 l/min
- Tools
- flow meter
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ACB +5VA power supply out of range patient safety
No data; Ventilator failure alarm; system unable to operate normally due to power supply voltage excursion
Ranked causes
- 1. ACB +5VA power supply voltage out of specification (< 4.85 V or > 5.15 V) (occasional)
Measure ACB +5VA power supply voltage
- Test point
- ACB +5VA supply rail
- Expected
- Between 4.85 VDC and 5.15 VDC, +/-0.15V
- Tools
- Digital multimeter
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Access rear electronics enclosure for repair high voltage
Need to access rear panels to repair or replace machine components including electronics
Ranked causes
- 1. Component repair or replacement required (unknown)
Bleed all gas pressure from the machine
- Test point
- System pressurization state
- Expected
- All pressure bled
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ACGO pressure relief valve not functioning patient safety
ACGO pressure relief weight does not relieve or pop off from seat; no purring sound heard during test
Ranked causes
- 1. ACGO pressure relief valve stuck or blocked (occasional)
Set ACGO selector switch position
- Test point
- ACGO selector switch
- Expected
- ACGO position
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Anesthesia Control Board power supply voltage out of range high voltage
5 volt digital or analog supply measured outside 4.85–5.15 Vdc range, displayed in red text on diagnostics screen
Ranked causes
- 1. +5 VD or +5 VA supply voltage outside 4.85–5.15 Vdc (unknown)
Measure +5 VD Supply voltage on Anesthesia Control Board
- Test point
- Anesthesia Control Board
- Expected
- 4.85 to 5.15 Vdc, 4.85–5.15
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
APL valve assembly defective patient safety
APL valve not opening/closing properly or pressure not relieving
Ranked causes
- 1. APL valve assembly defective (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auxiliary O2 flow rate outside specified limits
Flow test device reading does not match expected values for selected flowmeter settings during accuracy test
Ranked causes
- 1. Flowmeter calibration drift or internal wear (occasional)
Verify flow test device capability
- Test point
- Flow test device specifications
- Expected
- Device measures 0-15 l/min with ±2% accuracy
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auxiliary O2 flowmeter ball does not respond correctly
Flowmeter ball does not rise smoothly when flow control is rotated counterclockwise, does not stay at bottom when closed, or does not maintain steady position at set flow
Ranked causes
- 1. Flowmeter internal mechanism stiction or contamination (occasional)
Rotate flow control clockwise to minimum
- Test point
- Flowmeter ball and control
- Expected
- Ball rests at bottom of flow tube and does not move
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auxiliary O2 flowmeter not operating smoothly or ball not moving
Ball in flowmeter does not rise smoothly when control rotated counterclockwise or does not maintain steady position at set flow
Ranked causes
- 1. Flowmeter obstruction, sticky ball, or control valve blockage (common)
Verify O2 supply available
- Test point
- O2 cylinder valve or pipeline connection
- Expected
- O2 cylinder valve ON or pipeline connected
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery does not maintain charge or runs down too quickly patient safety
System shuts down in less than 50 minutes when running on battery power under standard load
Ranked causes
- 1. Battery capacity degradation or failed battery cell (common)
Configure system for battery test
- Test point
- System settings
- Expected
- Anesthetic Agent: None, Task Light: max, LCD Brightness: 5, Keypad brightness: 5
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bellows assembly damaged or stuck patient safety
Bellows housing not rotating freely or bellows cracked/leaking
Ranked causes
- 1. Bellows housing stuck or bellows damaged (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Bellows fills slowly during flush test
- Bellows fills too quickly during flush test
- Bottom task light not functioning
- Calibration: Adjust drive gas regulator
- Calibration: Airway gas module
- Calibration: Airway Pressure and ACGO Span
- Calibration: Airway Pressure Span (Paw Span)
- Calibration: Bellows drop test
- Calibration: Bleed Resistor
- Calibration: Circuit leak test
- Calibration: Flow control test
- Calibration: Flow sensing test
- Calibration: Inspiratory Flow Valve
- Calibration: Low pressure leak test with ACGO
- Calibration: Low pressure leak test without ACGO
- Calibration: Manifold P Span
- Calibration: Mixer Flow Sensor Zero
- Calibration: O2 needle valve minimum flow
- Calibration: O2 supply alarm test
- Calibration: Paw and ACGO Span
- Calibration: Paw Span
- Calibration: Pipeline and cylinder high-pressure leak test
- Calibration: Primary regulator output pressure
- Calibration: Secondary regulator output pressure
- Calibration: Touch screen misaligned
- Calibration: Zero ACGO/AGSS
- Calibration: Zero Gas Transducers
- Carrier and Com Express board assembly replacement
- Carrier board overheat
- Caster replacement procedure
- Caster wheel damaged or non-functional (without central brake)
- Central brake system does not operate smoothly
- Central vaporizer task light not functioning
- Chassis leakage current exceeds limit
- CPLD test fail
- CPU coin battery replacement on display unit
- Customize checklist items
- Cylinder check valve defective
- Cylinder inlet filter blocked or contaminated
- Display arm replacement
- Display unit disassembly and component access
- Display unit encoder replacement
- Display unit front assembly and LCD/Touch bezel replacement
- Display unit power and data cable replacement
- Display Unit power supply voltage out of range
- Display unit removal from machine
- Display unit speaker replacement
- Exhalation valve diaphragm orientation incorrect
- Expiratory flow sensor EEPROM failure
- FIB +5VA fail
- FIB +5VD fail
- Final machine checks incomplete
- FiO2 high alarm does not activate or deactivate
- FiO2 high alarm does not trigger
- FiO2 low alarm does not activate or deactivate
- FiO2 low alarm does not trigger
- Flow sensor calibration error
- Flow sensors defective or misaligned
- Flow valve leak test fail
- Frame Interface Board power supply voltage out of range
- Front caster damaged or non-functional (with new central brake linkage)
- Front caster damaged or non-functional (with old central brake linkage)
- Front panel com fail
- Gas mixer output flow out of range
- Gas supply pressure out of range
- High pressure leak in cylinder supply
- Inspiration stopped during ventilation
- Inspiratory flow control valve orientation incorrect
- Inspiratory flow sensor EEPROM failure
- Inspiratory/expiratory ports assembly damaged
- Leakage current exceeds safety limits
- Machine serial number or options mismatch after board replacement
- MGAS inlet filter residue > 40 sec
- MGAS line blocked > 20 sec
- MGAS sample flow deviation
- MGAS sample gas out blocked > 20 sec
- MGAS sample line not connected > 40 sec
- MGAS sensor inop
- Minute volume low alarm does not trigger
- MOPV pressure relief valve not functioning
- Multiple vaporizers can be enabled simultaneously
- Multiple vaporizers can be turned on simultaneously
- MVexp low alarm does not activate
- N2O and O2 linkage failure
- N2O flow does not decrease with O2 flow
- O2 bypass mixer hardware failure
- O2 flush valve malfunction
- O2 sensor reading inaccurate
- Peak airway pressure high alarm does not trigger
- Peak airway pressure low leak alarm does not trigger
- PEEP blockage alarm does not trigger
- PEEP high blockage alarm does not activate
- Pipeline inlet check valve stuck or leaking
- Pipeline inlet filter contamination or blockage
- Pipeline inlet manifold pressure transducer failure
- Pneumatic fittings cannot be safely disconnected
- Power failure message not displayed correctly
- Power Management Board voltage out of range
- Ppeak high alarm does not activate
- Ppeak low leak alarm does not activate
- Pressure limit circuit not functioning
- Primary regulator module defective
- Scavenger positive relief not functioning correctly
- Sensor Interface Board power supply voltage out of range
- Set all alarms to clinical limits
- SIB pneumatic sense tubing leak
- SIB pneumatic sense tubing leak test
- Suction flow rate less than minimum
- Suction flow rate low
- Suction gauge does not bleed to zero when turned off
- Suction gauge does not read accurately
- Suction gauge needle drifts when turned off with knob not at Off
- Suction gauge reading inaccuracy
- Suction regulator does not hold set pressure
- Suction regulator does not maintain set vacuum
- Suction system leaks with valve Off
- Suction vacuum does not bleed off when mode selector turned Off
- Sustained high airway pressure
- System does not power on or power alarm does not activate
- Touchscreen unresponsive or fails
- Transformer requires replacement
- Vent engine drive gas check valve adhesion
- Vent flow or pressure measurement out of range
- Vent subsystem status parameter abnormal
- Ventilator failure with low airway pressure
- Ventilator flow valve current feedback failure
- Ventilator flow valve DAC feedback failure
127 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
12 HR TEST | System has been operating more than 12 hours without power cycle self-test | System uptime exceeds 12 hours in Checkout state without a power on/off cycle | info |
ACB +5VD FAIL | Ventilator failure due to voltage out of range | VACB +5VD < 4.85 VDC or > 5.15 VDC | error |
ACB CLOCK SPEED | Anesthesia Control Board clock speed out of specification | ACB oscillator or timing circuit failure detected during self-test | critical |
ACB CLOCK SPEED | System malfunction due to abnormal clock frequency | Clock frequency > 1.1× or < 0.9× expected | error |
ACB COM FAIL | Communication failure between Display Unit and Anesthesia Control Board | DU unable to receive data from ACB or handshake protocol failure | critical |
ACB COM FAIL | System malfunction due to communication loss between computers | Display Computer receives no messages from Anesthesia Computer for 10 seconds | error |
ACB CPU TEST FAIL | Anesthesia Control Board CPU self-test failed | CPU executes invalid instruction or core logic failure detected | critical |
ACB DCB COM FAIL | System malfunction due to Display Computer communication loss | Anesthesia Computer receives no state messages from Display Computer for 10 seconds | error |
ACB DCB COM FAIL | Communication failure between ACB and Display Control Board (DCB) | ACB unable to communicate with DCB subsystem or protocol error | critical |
ACB EEPROM FAIL | Memory failure on ACB | Read/write or CRC failure of EEPROM on ACB | warning |
ACB FIB COMMUNICATION FAIL | Gas controls failure, O2 only | Interval between valid messages between ACB and FIB exceeds 100 milliseconds | error |
ACB FLASH FAIL | ACB flash memory read/write failure | Flash memory integrity check failed or data verify error during read/write operation | critical |
ACB FLASH FAIL | System malfunction due to code memory CRC failure | CRC failure occurs in code space | error |
ACB HW WATCHDOG | ACB hardware watchdog timer expired | ACB main processor did not service watchdog within timeout period | critical |
ACB MICROPROC ERROR | ACB microprocessor detected an internal error condition | Microprocessor exception, parity error, or undefined instruction trap | critical |
ACB MICROPROC ERROR | System malfunction due to microcontroller exception | Unexpected microcontroller exception (bus error, address error, etc.) | error |
ACB PROCEDURE COM FAIL | System malfunction due to procedure message communication failure | No procedure messages for 15±1 sec or message mismatch for 15±1 sec | error |
ACB PROCEDURE COM FAIL | Communication failure during ACB procedure execution or command handshake | ACB failed to acknowledge procedure command or lost connection mid-procedure | critical |
ACB RAM ERROR | ACB RAM memory test or parity error detected | RAM self-test failed or uncorrectable memory parity error during operation | critical |
ACB RAM ERROR | System malfunction due to memory test failure | Multiple bit errors detected during memory test | error |
ACB REDUNDANT MEMORY FAIL | System malfunction due to redundant parameter storage failure | Redundantly stored parameter could not be stored or was corrupted | error |
ACB REDUNDANT MEMORY FAIL | ACB redundant or backup memory verification failed | Redundant memory mirror check or backup memory integrity check failure | critical |
ACB SW ERROR | ACB software detected an internal error or exception condition | Firmware exception handler invoked due to invalid state or logic error | critical |
ACB SW ERROR | System malfunction due to unexpected software error | Unexpected software error occurs | error |
ACB SW WATCHDOG | System malfunction due to software watchdog failure | Software watchdog failed power-up test, timed out, or function delinquent | error |
ACB SW WATCHDOG | ACB software watchdog timer expired during firmware execution | Firmware task did not complete or report status within watchdog timeout period | critical |
ACB UNEXPECTED RESET | ACB experienced an unexpected reset or reboot without operator action | ACB core or watchdog forced a system reset during operation | critical |
ACB UNEXPECTED RESET | System malfunction due to unexpected ACB reset | Unexpected reset of anesthesia computer occurs | error |
ACGO | Vol and Apnea monitoring off | Non Circle (ACGO) selected | info |
ACMains POWER FAIL | AC mains power loss, system on battery | ACMains_GOOD goes low and stays low for ≥3 seconds | warning |
AGSS HIGH | Scavenging flow is too high | AGSS flow exceeds threshold based on configuration (>36, >50, or >80 l/min) | info |
AGSS LOW | Scavenging flow is too low | AGSS flow below threshold based on configuration (<21 or <42 l/min) | info |
AIR FRESH GAS FLOW SENSOR FAIL | Fresh gas flow sensor for air inlet failed or out of range | Air flow sensor reports zero or invalid value during gas delivery check | error |
AIR FRESH GAS FLOW SENSOR FAIL | Gas controls failure, O2 only | FIB reports air fresh gas flow sensor fail to ACB | warning |
AIR MIXER VALVE FAIL | Gas controls failure, O2 only | Air mixer valve feedback current check failed | warning |
AIR MIXER VALVE FAIL | Air mixer valve failure or stuck condition detected | Air mixer valve does not respond to command or position feedback invalid | error |
AIR PIPE INVALID | Cannot monitor air pipeline due to invalid pressure | Air pipeline pressure reading is invalid | warning |
AIR PRESS LOW | Air supply pressure low | Air pipeline <252 kPa and air cylinder <2633 kPa for 1 second | warning |
AUX | Vol and Apnea monitoring off | Auxiliary gas turned on | info |
BATTERY CHARGE FAIL | No battery backup due to charging failure | Standby: charge current >2.0A; powered on: battery current >1.3A | warning |
BATTERY EMPTY | System shutdown in less than 5 minutes due to battery depletion | Available battery power 1-5 minutes | critical |
BATTERY FAIL | No battery backup, battery voltage critical | Battery <10.5V, or charging anomalies (duration >12-16h, voltage >15.5V, current <0.15A) | warning |
BATTERY LOW | Plug in power cable, system on battery | Available battery power 40-60 minutes | warning |
BATTERY MISSING | No battery backup, battery not connected | Any battery voltage <+2.0 VDC | warning |
BATTERY REVERSED CONNECTIONS | No battery backup due to reversed battery polarity | Any battery voltage <–10.0 VDC | warning |
BATTERY V LOW | Plug in power cable, battery power 20-40 minutes remaining | Available battery power 20-40 minutes | warning |
BATT V VERY LOW | Critical battery depletion, plug in power cable | Available battery power decreases to 5-20 minutes | error |
BELLOWS COLLAPSED | Unable to drive bellows, manifold pressure excessive | Manifold pressure > Paw + 10 + [0.25×(Insp valve flow)] | info |
BREATHING SYSTEM NOT LATCHED | Breathing system loose, detection switch indicates not latched | Breathing system detection switch indicates not latched | info |
CAL DATA FAILURE IN EEPROM | Service calibration advised, default cal data in use | Corrupt data in calibration region of EEPROM | warning |
CARRIER BOARD OVERHEAT | Cooling fans failed, system may overheat | Both Display Unit thermistors > 65°C, or neither thermistor reading in valid range 0-85°C | warning |
CHECK FLOW SENSOR | Flow sensor measurements do not meet criteria during mechanical breaths | No or negative flow on inspiratory flow sensor during inspiration, or negative flow on expiratory flow sensor for 6+ consecutive breaths | warning |
COM ERROR VENT TO ACB | Loss of communication between Ventilator boundary and SIB CPU | No messages received from SIB CPU for 35 milliseconds after initial communication established | critical |
CPLD TEST FAIL | Complex Programmable Logic Device self-test failure | CPLD internal logic or integrity check failed during startup or periodic test | critical |
CPU FAN SPEED FAIL | CPU cooling fan speed < 50% nominal (2500 RPM) | CPU fan speed drops below 2500 RPM during operation | warning |
CPU OVERHEAT | CPU thermal diode temperature exceeds 80°C | CPU thermal diode temperature > 80°C sustained during operation | warning |
DCB PROCEDURE COM FAIL | Anesthesia Computer loses procedure message communication with Display Computer | No procedure messages for 15s +/- 0.5s, or messages do not match for 15s +/- 0.5s, during Therapy state | critical |
DRIVE GAS LOST | Ventilator has no drive gas; pressure too low | O2 PRESS LOW alarm active when O2 is drive gas, or Air PRESS LOW active when Air is drive gas, during mechanical ventilation | critical |
DRV GAS FLOW SENSOR CAL ZERO FAIL | Drive gas flow sensor self-test calibration failed | Drive gas flow sensor zero calibration fails at power-on self-test | warning |
DRV GAS FLOW SENSOR FAIL | Drive gas flow sensor hardware failure detected | Anesthesia Computer detects drive gas flow sensor failure during operation | warning |
***ERROR: Cannot load DOS! Any Key to retry | DOS environment failed to load; boot sequence interrupted | System startup attempting to initialize DOS environment | error |
ERROR!! INCOMPATIBLE SOFTWARE | Software version on USB Media is not compatible with installed subsystem | Software Download process detects incompatible subsystem during compatibility check | error |
***ERROR: No bootable device available. | System cannot locate or access bootable storage device | System startup attempting to locate bootable device | critical |
***ERROR: Program load failed - CRC. | Cyclic redundancy check error during software load; data corruption detected | System startup attempting to load software from CF card | critical |
***ERROR: RAM memory error. | Random access memory hardware failure detected | System startup performing memory diagnostics | critical |
***ERROR: SMM init failure. Press any key to continue. | System management mode initialization failed; firmware initialization error | System startup attempting to initialize SMM | error |
***ERROR: System reset: ECxx xx xx . . . | Unexpected system reset with error code; general software failure | System startup detecting previous unexpected reset event | error |
***ERROR: System reset: FFFF FF FF | CMOS battery has lost contact or voltage below minimum threshold | System startup detecting CMOS battery contact or voltage failure | critical |
***ERROR: The normal Windows CE application is not available | Windows CE application not found or failed to load | System startup attempting to launch Windows CE application | error |
***ERROR: Watchdog circuit failed. | Hardware watchdog timer circuit malfunction; system safety feature failed | System startup detecting watchdog circuit failure | critical |
EXP FLOW SENSOR EEPROM FAILURE | Expiratory flow sensor calibration data cannot be read | Expiratory flow sensor EEPROM read failure during initialization | error |
FAN FAIL | Cooling fan power status bit indicates fan failure | Fan power status bit is low (FAN1_GOOD bit = 0) | warning |
FIB +5VA FAIL | FIB +5VA power supply voltage out of tolerance | VFIB +5VA < 4.85 VDC or > 5.15 VDC and FIB +12.5V within range | warning |
FIB +5VD FAIL | Frame Interface Board +5VD digital power supply out of specification | FIB +5VD supply voltage out of acceptable range detected | error |
FLOW SENSOR CAL ERROR | Flow sensor calibration out of range | Inspiratory or expiratory flow sensor or airway/manifold pressure sensor zero offset calibration fails | error |
FLOW VALVE LEAK TEST FAIL | Flow control valve leakage test failed during POST state | Flow control valve fails leakage test during power-on self-test | warning |
FRONT PANEL COM FAIL | Display panel keypad controller communication failure | Keypad controller fails to send life tick signal for >10 seconds | warning |
GAS FLOW > 300ML/MIN | Gas flow exceeds 300 ml/min during Checkout: General sub-state | System detects gas flow > 300 ml/min during Checkout: General mode | warning |
GAS INLET VALVE BOOTUP | Gas inlet valve solenoid self-test failed at boot-up | Gas inlet valve solenoid fails test during system initialization | critical |
INSP FLOW SENSOR EEPROM FAILURE | Inspiratory flow sensor calibration data cannot be read | Inspiratory flow sensor EEPROM read failure during initialization | error |
LOSS OF VENT PARAMETER SETTINGS | Ventilator parameter settings communication lost from Display Computer | Ventilator parameters not received from Display Computer for >10 seconds during ventilation mode | warning |
MANIFOLD PAW SENSOR FAIL | Manifold pressure sensor calibration failed at boot-up | Manifold pressure sensor zero calibration fails during system initialization | warning |
MGAS CHECK SAMPLE GAS OUT >20 SEC | Airway Module sample line blocked or occluded >20 seconds | Airway Module reports continuous occlusion alarm lasting >20 seconds | warning |
MGAS INLET FILTER RESIDUE >40 SEC | D-Fend water trap residue buildup reducing airflow >40 seconds | Airway Module reports water trap residue alarm lasting >40 seconds | warning |
MGAS LINE BLOCKED >20 SEC | Airway Module sample tubing blocked or water trap occluded >20 seconds | Airway Module reports sample tubing occlusion alarm lasting >20 seconds | warning |
MGAS SAMPLE FLOW DEVIATION | Sample flow < 80% of Airway Module nominal flow (120 ml/min for PROLITE) | Sample flow stays below 80% of module nominal value for 20+ consecutive seconds | info |
MGAS SAMPLE LINE NOT CONNECTED >40 S | Airway Module sample line or D-Fend module not installed >40 seconds | Airway Module reports open circuit alarm lasting >40 seconds | warning |
MGAS SENSOR INOP > XX | Airway Module hardware failure or initialization fault | Airway Module reports hardware error (RAM/ROM/EEPROM failure, voltage error, lamp failure) or UPI initialization failure | error |
MODULE NOT COMPATIBLE | Monitoring Module detected is not compatible with system software | Incompatible airway module connected | warning |
N2O FLOW > 100 ML/MIN | N2O flow exceeds 100 ml/min at auxiliary outlet | N2O flow > 100 ml/min in Auxiliary Gas Delivery mode | warning |
N2O FRESH GAS FLOW SENSOR FAIL | Nitrous oxide fresh gas flow sensor failure or out of range | N2O flow sensor reports invalid or zero reading during checkout or delivery | error |
N2O FRESH GAS FLOW SENSOR FAIL | Gas controls failure, N2O only | FIB reports N2O Fresh Gas Flow Sensor Fail alarm to ACB | warning |
N2O MIXER VALVE FAIL | Gas controls failure, N2O only | N2O mixer valve feedback current check failed | warning |
N2O MIXER VALVE FAIL | Nitrous oxide mixer valve failed or does not respond to commands | N2O mixer valve stuck or position feedback invalid during gas mixing operation | error |
N2O PIPE INVALID | Cannot monitor N2O pipeline | N2O Pipeline pressure is invalid | warning |
N2O PRESS LOW | N2O supply pressure low | N2O pipeline < 252 kPa and cylinder < 2633 kPa with N2O selected | info |
NO EXPIRATORY FLOW SENSOR | No expiratory sensor connected and not calibrating | Expiratory flow sensor disconnected or missing during calibration | warning |
NO INSPIRATORY FLOW SENSOR | No inspiratory sensor connected and not calibrating | Inspiratory flow sensor disconnected or missing during calibration | warning |
***NOTE: Alarm speaker not detected. Check connection. | Speaker connection failure or speaker not responding | System startup detecting no audio output device | warning |
***NOTE: CMOS battery is weak. Please replace. | CMOS battery voltage insufficient for proper real-time clock and settings retention | System startup detecting low CMOS battery voltage | warning |
***NOTE: CPU Board Supply Voltage Out of Range. | One or more processor board supply voltages (1.5V_CB, 5V_CB, or 3.3V_CB) outside acceptable range | System startup performing voltage range validation on Carrier board | critical |
***NOTE: RTC date/time error. Battery may be weak. | Real-time clock error detected; system time and date not maintained | System startup validating RTC function and battery status | warning |
O2 BYPASS VALVE FAIL | Oxygen bypass valve failure or stuck condition | O2 bypass valve does not open/close as commanded or position feedback invalid | error |
O2 BYPASS VALVE FAIL | Gas controls failure, O2 only | O2 bypass valve feedback current check failed | warning |
O2 CAL ERROR | Calibrate O2 sensor | O2 sensor offset, slope, cell voltage out of range or O2 > 110% | info |
O2 FLOW < 150ML/MIN | O2 flow low | O2 flow < 150 ml/min in Therapy state | info |
O2 FLUSH FAILURE | O2 flush stuck on? | O2 Flush Switch detected on continuously > 30 sec | info |
O2 FRESH GAS FLOW SENSOR FAIL | Gas controls failure, O2 only | FIB reports O2 Fresh Gas Flow Sensor Fail alarm to ACB | warning |
O2 FRESH GAS FLOW SENSOR FAIL | Oxygen fresh gas flow sensor failure or out of range reading | O2 flow sensor reports zero or invalid value during system checkout | error |
O2 LATCHING VALVE FAIL | System Malfunction | O2 latching valve feedback current check failed | error |
O2 LATCHING VALVE FAIL | Oxygen latching valve failure or unable to maintain state | O2 latching valve does not latch or unlatch as commanded during operation | error |
O2 PIPE INVALID | Cannot monitor O2 pipeline | O2 Pipeline pressure is invalid | warning |
O2 PRESS LOW | O2 supply pressure low | O2 pipeline < 252 kPa and O2 cylinder < 2633 kPa for one second | error |
O2 SENSOR FAILURE | Replace O2 sensor | O2 < 5% | info |
ON/STANDBY SWITCH TO STANDBY | Confirm to shutdown | On/Standby switch activated in Therapy state | error |
PATIENT VOLUME MISMATCH OCCURRED | Calibrate, dry, or replace flow sensors | PATIENT VOLUME MISMATCH alarm occurred during Checkout | info |
PAW OVERPRESSURE SIGNAL DISCONNECTED | Vent fail. Monitoring only | SIB reports PAW Overpressure Signal Disconnected alarm to ACB | warning |
PEEP PCV NOT AVAILABLE | Vol vent only. No PEEP or PSV | Paw data in range but Pmanifold <= -15 cmH2O | warning |
PORT PLUG POST FAIL | Vent fail. Monitoring only | Paw > 100 cmH2O for 1 second in POST state | warning |
POWER CONTROLLER COM FAIL | Internal failure. System may shut down | Communications with PC and DC cannot be established for 10 seconds | warning |
POWER CONTROLLER SELF TEST | Internal failure. System may shut down | Power Controller fails self tests | error |
POWER SUPPLY BOOST > 75C | Circuitry >75C shutdown possible | Power supply temperature near boost converter exceeds 75C | info |
POWER SUPPLY BUCK > 75C | Circuitry >75C shutdown possible | Power supply temperature exceeds 75C | warning |
PWR CNTRL DC-DC BOOST FAIL | Internal failure. System may shut down | Boost converter failure | warning |
PWR CNTRL DC-DC BUCK FAIL | No battery backup | Battery not connected or buck converter failure | warning |
PWR CNTRL OUTPUT ERROR SWITCHES AC MAIN TO BACKUP POWER | Power switched to Battery | Output power over/under voltage, switches from AC Mains to backup power | warning |
PWR CNTRL SPL1 CHANNEL ERROR | System Malfunction | SPL channel 1 current > 0.5A or voltage out of range (>13.2V or <12.1V) 3+ times | error |
PWR CNTRL SPL2 CHANNEL ERROR | Vent fail. Monitoring only | SPL channel 2 current > 0.85A or voltage out of range 3+ times | error |
PWR CNTRL SPL3 CHANNEL ERROR | No data. Ventilator failure | SPL channel 3 current > 0.62A or voltage out of range 3+ times | error |
PWR CNTRL SPL4 CHANNEL ERROR | Power Controller supply channel 4 (SPL4) error or out of specification | Power Controller SPL4 output voltage or current out of range | error |
PWR CNTRL SPL4 CHANNEL ERROR | Gas controls failure, O2 only | SPL channel 4 current > 1.3A or voltage out of range 3+ times | warning |
PWR CNTRL SPL5 CHANNEL ERROR | Power Controller supply channel 5 (SPL5) error or out of specification | Power Controller SPL5 output voltage or current exceeds limits | error |
PWR CNTRL SPL5 CHANNEL ERROR | Gas controls failure, O2 only | SPL channel 5 current > 0.7A or voltage out of range 3+ times | warning |
PWR CNTRL SPL6 CHANNEL ERROR | Gas monitoring not available | SPL channel 6 current > 0.7A or voltage out of range 3+ times | warning |
REVERSE EXPIRATORY FLOW | Reverse exp flow. Check valves OK? | Volume >= 20 ml toward patient on expiratory sensor and >= 5 ml on inspiratory during inspiration for 6 consecutive breaths | warning |
SIB +5VA FAIL | No data. Ventilator failure | VSIB +5VA < 4.85 VDC or > 5.15 VDC | error |
SIB +5VD FAIL | No data. Ventilator failure | VSIB +5VD < 4.85 VDC or > 5.15 VDC | error |
SIB +5VE2 FAIL | No data. Ventilator failure | SIB +5VE2 voltage < 4.85 VDC or > 5.15 VDC | error |
STANDBY PATIENT DETECTION | No fresh gas flow! | 3 volume breaths or 3 CO2 breaths detected within 30 seconds during Checkout | error |
TOUCHSCREEN FAILURE | Touchscreen panel failure | Touch panel controller fails to send a life tick for more than 10 seconds | error |
VENT AIRWAY OVERPRESS SIGNAL | High airway overpressure signal during ventilation | High airway overpressure signal is set during mechanical ventilation | warning |
VENT AIRWAY OVERPRESS SIGNAL FAIL | Ventilator failure with airway overpressure signal and low pressure | High Airway overpressure signal set with Paw < 90 cmH2O and Pmanifold < 80 cmH2O | warning |
VENT FLOW VALVE FAIL CURRENT | Ventilator flow valve current feedback error | Current feedback check failed for flow control valve | error |
VENT FLOW VALVE FAIL DAC | Ventilator flow valve DAC feedback error | Incorrect DAC feedback check failed for flow control valve | error |
VENT SUSTAINED PAW SDOWN | Sustained high airway pressure alert | Airway pressure exceeds 100 cmH2O for 10 seconds | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.