GE Healthcare Carestation 650: troubleshooting and error codes

The ServiceForge knowledge base holds 187 documented symptoms for the GE Healthcare Carestation 650, with 291 ranked causes, 434 diagnostic checks with test points and expected values, and 145 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Vent engine component seal degradation caution

Loss of gas tightness or component leakage due to worn seals or o-rings

Ranked causes

  1. 1. Gas inlet valve solenoid seal worn or damaged (occasional)
  2. 2. Gas inlet valve o-ring worn or damaged (occasional)
  3. 3. Interface manifold o-rings degraded (occasional)
  4. 4. Drive gas regulator o-rings worn (occasional)
  5. 5. Inspiratory flow control valve o-rings worn (occasional)
  6. 6. Flow sensor o-rings degraded (occasional)
  7. 7. Free breathing check valve o-ring worn (occasional)
  8. 8. Reservoir o-rings degraded (occasional)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Breathing system leak from canister or valves patient safety

Gas leaking from absorber canister, check valves, APL valve, breathing tubes, or negative pressure relief valve

Ranked causes

  1. 1. Absorber canister open or missing (common)
  2. 2. Damaged or missing canister o-ring or seals (occasional)
  3. 3. APL valve not closed (common)
  4. 4. Breathing system tubes damaged or connectors loose (occasional)
  5. 5. Check valve of inspiratory or expiratory ports not correctly installed (occasional)
  6. 6. Negative pressure relief valve leak (rare)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check flow sensor failure patient safety

During mechanical breaths, measured flow to and from patient does not meet criteria; no or negative flow on inspiration or expiration sensor.

Ranked causes

  1. 1. Flow sensor connection loose or disconnected (common)
  2. 2. Breathing circuit issue (common)
  3. 3. SIB sensor tubing leak (occasional)
  4. 4. Flow sensor calibration out of range (occasional)
  5. 5. Insp/Exp check valve failure (occasional)
  6. 6. Flow sensor defective (occasional)
First check

Verify flow sensor connections are seated

Test point
Flow sensor connectors
Expected
Fully plugged in

3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

High pressure leak detected high voltage

Gas leaking from high-pressure system components (pipeline, O2 flush valve, cylinders, transducers, gaskets, relief valves)

Ranked causes

  1. 1. Pipeline leak (common)
  2. 2. O2 flush valve malfunction (common)
  3. 3. Cylinder not installed properly (occasional)
  4. 4. Cylinder transducer leak (occasional)
  5. 5. Cylinder gasket failure (occasional)
  6. 6. Relief valve malfunction (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ventilator leak test – cannot empty bellows patient safety

System cannot empty bellows during ventilator leak test (manifold pressure must be ≥60 cmH2O).

Ranked causes

  1. 1. Patient Y not open (common)
  2. 2. Drive gas leak from bellows housing to atmosphere (occasional)
  3. 3. Bellows installation error or bellows leak (occasional)
  4. 4. Pop-off valve leak (occasional)
  5. 5. Ventilator exhalation valve leak (occasional)
  6. 6. Drive Gas Check Valve stuck open (rare)
First check

Set Bag/Vent switch to Vent, open patient Y, select Confirm

Test point
Bellows emptying
Expected
Bellows empty completely

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Circuit leak test – cannot pressurize patient safety

Breathing circuit leak test cannot pressurize circuit during testing.

Ranked causes

  1. 1. Test bag not properly occluded or test bag leaks (common)
  2. 2. Leak in canister or CO2 bypass (occasional)
  3. 3. Leak in O2 cell (occasional)
  4. 4. Leak in patient port or Y piece (occasional)
  5. 5. Leak in APL assembly (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

DCB procedure com fail patient safety

Anesthesia Computer loses communication with Display Computer; procedure messages missing or mismatched for >15 seconds.

Ranked causes

  1. 1. ACB to FIB or SIB communication cable disconnected (common)
  2. 2. Display Unit cable connection loose (occasional)
  3. 3. Display Controller board failure (occasional)
  4. 4. ACB failure or disconnection (occasional)
First check

Reboot system

Test point
System startup
Expected
ACB and DCB re-establish communication

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gas inlet valve bootup fail patient safety

Gas inlet valve solenoid self-test fails at boot-up; ventilator cannot control gas delivery.

Ranked causes

  1. 1. Gas inlet valve solenoid connector disconnected or loose (common)
  2. 2. Gas inlet valve solenoid defective (occasional)
  3. 3. Gas inlet valve mechanical failure (occasional)
  4. 4. ACB solenoid drive circuit failure (rare)
First check

Verify gas inlet valve solenoid connector is seated

Test point
Gas inlet valve solenoid connector on ACB
Expected
Fully plugged in, no loose pins

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Leak in both bag and vent modes patient safety

Gas escapes from breathing circuit in both bag and vent operational modes

Ranked causes

  1. 1. Canister seal, body, or cover defective (common)
  2. 2. Flow sensor or Paw transducer port leakage on SIB board (common)
  3. 3. Patient ports assembly defective (common)
  4. 4. Canister or circuit board connectors, filters, or tubes defective (occasional)
First check

Perform Test 7 for canister leakage

Test point
Canister handle port
Expected
Airway Pressure rises to >=30 cmH2O at 0.1 L/min O2 flow

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of ventilator parameter settings patient safety

Ventilator parameter settings from Display Computer not received for >10 seconds; ventilator operates in monitoring-only mode.

Ranked causes

  1. 1. ACB to FIB or SIB communication cable disconnected (common)
  2. 2. Display Unit cable connection loose (occasional)
  3. 3. Display Controller board failure (occasional)
  4. 4. ACB connection or failure (occasional)
First check

Reboot system

Test point
System startup and communication re-establishment
Expected
ACB receives vent parameters within 10 seconds

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low pressure leak with vaporizer mounted caution

Gas leaking from low-pressure system when vaporizer is installed; includes vaporizer installation, o-ring, fill port issues

Ranked causes

  1. 1. Vaporizer not installed properly (common)
  2. 2. Vaporizer leaks (occasional)
  3. 3. Missing or damaged o-ring on vaporizer manifold (occasional)
  4. 4. Loose fill port (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ventilator circuit leak exceeds 750 ml/min patient safety

During ventilator leak test, detected leak rate exceeds 750 ml/min threshold.

Ranked causes

  1. 1. Leak in canister or CO2 bypass (common)
  2. 2. Leak in O2 cell (common)
  3. 3. Leak in patient port or Y piece (common)
  4. 4. Leak in sample gas return line or gas return port inside bellows (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ventilator leak test – cannot pressurize circuit patient safety

During ventilator leak test, circuit cannot be pressurized to required level.

Ranked causes

  1. 1. Leak in canister or CO2 bypass (common)
  2. 2. Leak in O2 cell (common)
  3. 3. Leak in patient port or Y piece (common)
  4. 4. Leak in sample gas return or gas return port inside bellows (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bellows leak or dysfunction patient safety

Bellows not sealing properly, mounting issues, or physical damage affecting ventilation

Ranked causes

  1. 1. Pop-off valve diaphragm not sealing properly (occasional)
  2. 2. Bellows mounting rim loose (occasional)
  3. 3. Bellows improperly mounted or has hole or tear (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bellows over inflate patient safety

Bellows bulging excessively during operation due to exhaust blockage or O2 flush malfunction

Ranked causes

  1. 1. Exhaust outlet is blocked (occasional)
  2. 2. O2 flush button is malfunctioning (occasional)
  3. 3. Flow mixer module is malfunctioning (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Circuit leak test – timed out patient safety

Breathing circuit leak test times out without establishing required pressure.

Ranked causes

  1. 1. Unsuitable test bag (too large or too soft) (common)
  2. 2. Leak inside O2 flush valve manifold (occasional)
  3. 3. Leak from O2 bypass, latch, N2O selector, or Air selector valve (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Com error ventilator to ACB patient safety

No communication between Ventilator boundary object and SIB CPU; loss of ventilator function.

Ranked causes

  1. 1. Communication cable disconnected or damaged (common)
  2. 2. SIB module failure (occasional)
  3. 3. ACB failure (occasional)
First check

Reboot system

Test point
System startup
Expected
System boots and establishes communication within 35ms

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CPU fan speed fail caution

CPU cooling fan speed drops below 2500 RPM (50% nominal), indicating fan failure or obstruction.

Ranked causes

  1. 1. CPU heatsink fan not operating (common)
  2. 2. Fan connector disconnected (common)
  3. 3. Fan defective (common)
First check

Verify CPU fan is spinning

Test point
CPU heatsink fan visual inspection
Expected
Fan rotating

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CPU overheat caution

CPU thermal diode temperature exceeds 80°C, indicating insufficient cooling.

Ranked causes

  1. 1. CPU fan not operating (common)
  2. 2. Fan connector disconnected (common)
  3. 3. Fan defective (common)
First check

Check CPU fan airflow

Test point
Back of Display Unit fan outlet
Expected
Warm air flowing from fan filter opening

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display arm positioning incorrect caution

Display arm not properly positioned side-to-side or at wrong tilt angle; may move with finger pressure or drop by own weight

Ranked causes

  1. 1. Arm mounting nut loose (common)
  2. 2. Display mount attachment nut loose (common)
  3. 3. Axis bar clamping screws insufficiently tight (common)
First check

Display moves when finger pressure applied to left side

Test point
Left side of display
Expected
Display does not move

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fan fail caution

Cooling fan power status bit is low; fan needs service but system operation OK.

Ranked causes

  1. 1. Fan connector loose or disconnected (common)
  2. 2. Cooling fan defective (occasional)
  3. 3. Power Management Board (PMB) failure (rare)
First check

Check cooling fan connector

Test point
Fan power connector on motherboard or PMB
Expected
Fully seated, no loose pins

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Leak in bag mode only patient safety

Gas escapes from breathing circuit during bag mode operation

Ranked causes

  1. 1. APL valve or diaphragm assembly defective (common)
  2. 2. BTV assembly, bag port, or negative valve defective (common)
  3. 3. Breathing circuit components improperly installed (common)
First check

Verify breathing circuit assembly installation

Test point
APL valve, Bag/Vent Switch, Cross Rubber
Expected
All components properly seated and connected

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Leak in vent mode only patient safety

Gas escapes from breathing circuit during vent mode operation

Ranked causes

  1. 1. U-Cup seal defective in bellows housing (common)
  2. 2. Pop-off diaphragm defective or bellows base port damaged (common)
  3. 3. BTV rubber, manifold, or O-ring seals defective (occasional)
First check

Perform Test 6 for bellows housing leakage

Test point
BTV port on upper BC assembly base
Expected
Airway Pressure rises to >=30 cmH2O at 0.1 L/min O2 flow

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low pressure leak with or without vaporizer caution

Gas leaking from low-pressure system; includes port valve, gas control module, flush valve, bellows issues

Ranked causes

  1. 1. Leaking port valve on vaporizer manifold (common)
  2. 2. Leak in gas control module (occasional)
  3. 3. Leaking flush valve (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Right top task light not functioning

Right task light does not illuminate, dimmer switch inoperative, or light is defective

Ranked causes

  1. 1. Defective right top task light board (common)
  2. 2. Defective task lights dimmer switch (occasional)
  3. 3. Defective right top task light harness (occasional)
First check

Verify right top task light illuminates

Test point
Right top task light
Expected
Light is lit

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to begin mechanical ventilation patient safety

System cannot initiate ventilation mode; breathing system not engaged or drive gas unavailable

Ranked causes

  1. 1. Breathing system not fully engaged (common)
  2. 2. No drive gas supply (occasional)
  3. 3. Defective Bag/Vent switch (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Vaporizer leak test failure patient safety

Vaporizer leak test detects leak rate exceeding 750 ml/min.

Ranked causes

  1. 1. Vaporizer port O-ring damaged or missing (common)
  2. 2. Nipple or vaporizer manifold port leak (occasional)
  3. 3. Vaporizer internal leak (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ventilator leak test – circuit pressure too high patient safety

During ventilator leak test, circuit pressure is higher than expected when patient Y is occluded.

Ranked causes

  1. 1. Drive gas leak from bellows into bellows due to pop-off or bellows failure (occasional)
  2. 2. Leak inside O2 flush valve manifold (occasional)
  3. 3. Leak from O2 bypass, O2 latch, N2O selector, or Air selector valve (occasional)
First check

Occlude patient Y and select Confirm during ventilator leak test

Test point
Circuit pressure
Expected
Within normal range

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bag/Vent switch or CO2 bypass defective patient safety

Bag/vent mode not switching, ventilation mode malfunction, or CO2 bypass stuck

Ranked causes

  1. 1. Bag/Vent switch assembly defective (occasional)
  2. 2. CO2 bypass stuck or defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bellows fills slowly or rapidly

O2 Flush button depressed continuously; bellows fill time outside expected range (1.3–3 seconds)

Ranked causes

  1. 1. Large leak in breathing system (common)
  2. 2. ACGO selector valve not correctly connected (occasional)
First check

Measure bellows fill time with O2 Flush button fully depressed

Test point
patient port
Expected
1.3 to 3 seconds

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Canister holder assembly damaged

Canister release lever broken, canister support base cracked, or holder not securing canister

Ranked causes

  1. 1. Canister release lever broken or bent (occasional)
  2. 2. Canister support base cracked or deformed (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Drive gas flow sensor calibration zero fail

Drive gas flow sensor power-on self-test failed; sensor unable to calibrate zero offset.

Ranked causes

  1. 1. Flow sensor connection loose or disconnected (common)
  2. 2. Flow sensor defective (occasional)
First check

Verify drive gas flow sensor connector is seated

Test point
Drive gas flow sensor connector
Expected
Fully plugged in, no loose pins

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Drive gas flow sensor fail

Anesthesia Computer detects drive gas flow sensor failure; ventilator operates in monitoring-only mode.

Ranked causes

  1. 1. Flow sensor connection loose or disconnected (common)
  2. 2. Flow sensor defective (common)
First check

Verify flow sensor connector is seated

Test point
Drive gas flow sensor connector
Expected
Fully plugged in

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Drive gas lost patient safety

Ventilator has no drive gas; O2 or Air supply pressure is low during mechanical ventilation.

Ranked causes

  1. 1. O2 supply disconnected or depleted (common)
  2. 2. Air supply disconnected or depleted (common)
First check

Verify O2 or Air supply is connected and pressurized

Test point
O2 or Air supply inlet
Expected
Pressure gauge above minimum (see O2 PRESS LOW or AIR PRES LOW troubleshooting)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

DU-ACB communication failure patient safety

System Malfunction screen displayed; ACB COM FAIL or ACB DCB COM FAIL error; loss of communication between Display Unit and Anesthesia Control Board

Ranked causes

  1. 1. Cable connection between ACB and DU loose or faulty (common)
  2. 2. ACB TXD/RXD LED failure or ACB internal communication error (occasional)
First check

Visually inspect cable connection between ACB and DU

Test point
RXD (DS5), TXD (DS4) connectors
Expected
Connectors seated firmly and undamaged

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flow control system out of specification patient safety

Flow control test fails; gas proportioning or individual flow outputs do not match expected values

Ranked causes

  1. 1. Flow control needle valve calibration drift (occasional)
  2. 2. Link proportioning system calibration drift (occasional)
First check

Increase N2O flow from minimum and verify O2 concentration or flow tracking

Test point
O2 concentration display or Mixer Output subsystem
Expected
O2 concentration 21% minimum at 0.2 l/min N2O; 21-34% at 1.6 l/min; 21-34% at 5.0 l/min; 21-34% at 9.0 l/min

8 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flow sensor reading out of specification caution

O2, Air, or N2O flow sensor readings outside expected limits during service test

Ranked causes

  1. 1. Flowmeter test device not backpressure compensated (occasional)
  2. 2. Flow sensor calibration drift (occasional)
First check

Set O2 flow to 1.5 l/min and verify sensor reading

Test point
O2 flow sensor display
Expected
1.3 to 1.7 l/min, 1.3 to 1.7 l/min

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gas flow > 300 ml/min during checkout caution

Gas flow exceeds 300 ml/min while system is in Checkout: General sub-state; operator must turn off gas flow.

Ranked causes

  1. 1. Mixer flow sensor zero offset out of calibration (common)
  2. 2. Needle valve adjustment out of spec (occasional)
First check

Turn off gas flow

Test point
Gas inlet to mixer
Expected
Gas flow stops

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gas inlet valve shuttle assembly degradation caution

Gas inlet valve shuttle not functioning smoothly due to missing or degraded seals

Ranked causes

  1. 1. Upper or lower u-cup seal missing or degraded on shuttle (occasional)
  2. 2. Shuttle o-ring seal worn or missing (occasional)
First check

Inspect shuttle o-rings and u-cup seals

Test point
Gas inlet valve shuttle assembly
Expected
Seals intact and not degraded

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gas module sensor zeroing in progress or not responding caution

Gas Monitor service page displays warming message, no message, or no-gas-module-detected message indicating sensor status during initialization

Ranked causes

  1. 1. Gas module connected but sensors still warming or zeroing (occasional)
  2. 2. Gas module not detected or disconnected (occasional)
First check

Access Gas Monitor Service and observe sensor warming status message

Test point
Gas Monitor Service page, top left corner
Expected
Either warming message or no message (connected), not no-detected message

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ground resistance exceeds limit high voltage

Resistance between exposed metal surface and ground pin on power cord is ≥ 0.20 Ω

Ranked causes

  1. 1. Corroded or loose ground connection on power cord (common)
  2. 2. Defective power cord (occasional)
First check

Set safety analyzer to measure line voltage

Test point
safety analyzer
Expected
normal polarity indicated

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inspiratory/expiratory check valve stuck or defective patient safety

One-way valve not sealing or stuck closed, causing improper gas flow direction

Ranked causes

  1. 1. Inspiratory check valve defective (rare)
  2. 2. Expiratory check valve defective (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Leak test device integrity verification failure

Leak test tool kit is leaking or cannot hold pressure, invalidating test results

Ranked causes

  1. 1. Low-pressure leak test device bulb does not inflate or leaks (common)
  2. 2. Tool kit (2091916-001) has leaking seals (common)
First check

Verify low-pressure leak test device integrity

Test point
Leak test device inlet and bulb
Expected
Bulb takes more than 60 seconds to fully inflate when hand seals inlet and bulb is squeezed empty, >60 sec

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Left top task light not functioning

Left task light does not illuminate or is defective

Ranked causes

  1. 1. Defective left top task light board (common)
  2. 2. Defective left top task light harness (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Manifold PAW sensor fail

Calibration failure at boot-up; manifold pressure sensor unable to zero offset properly.

Ranked causes

  1. 1. Manifold pressure sensor tubing occluded (common)
  2. 2. ACB manifold sensor circuit failure (rare)
First check

Verify manifold flow sensor counts in Service Software

Test point
Manifold Flow counts in Vent Flow and Pressure section 12.9.2
Expected
Between 2575 and 4447 counts

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

N2O mixer hardware failure

Gas control test detects N2O mixer hardware failure.

Ranked causes

  1. 1. O2 bypass or latch valve failure (common)
  2. 2. O2 latch or N2O selector valve failure (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 flow decreases excessively when vaporizer adjusted caution

O2 flow drops more than 1 l/min when vaporizer concentration is adjusted from 0 to 1%

Ranked causes

  1. 1. Malfunction in vaporizer back pressure mechanism (common)
  2. 2. Malfunction in anesthesia system vaporizer manifold port valve (occasional)
First check

Set O2 flow to 6 l/min and adjust vaporizer from 0 to 1%

Test point
O2 flowmeter
Expected
O2 flow must not decrease more than 1 l/min, ≤1 l/min decrease

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SIB cable connection improper caution

SIB cable loose, disconnected, or interfering with breathing system assembly

Ranked causes

  1. 1. SIB cable not fully seated or connected (common)
  2. 2. SIB cable routing interferes with breathing system assembly (occasional)
First check

Verify SIB-to-ACB cable is fully seated and secure

Test point
SIB cable connector at ACB
Expected
Cable fully connected with no play or movement

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SIB filter installation incorrect caution

Filters not installed in proper sequence or orientation

Ranked causes

  1. 1. Filters installed out of sequence (common)
  2. 2. Filter installed with incorrect orientation (occasional)
First check

Verify filter installation sequence

Test point
SIB filter assembly
Expected
Filters a and c installed first, then b, d, e in sequence

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Vaporizer back pressure exceeds limit

O2 flow decreases more than 1 l/min when vaporizer concentration adjusted from 0 to 1%

Ranked causes

  1. 1. Defective vaporizer installed (common)
  2. 2. Malfunction in anesthesia system vaporizer manifold (occasional)
First check

Verify O2 flow at 6 l/min baseline

Test point
O2 outlet
Expected
6 l/min
Tools
flow meter

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ACB +5VA power supply out of range patient safety

No data; Ventilator failure alarm; system unable to operate normally due to power supply voltage excursion

Ranked causes

  1. 1. ACB +5VA power supply voltage out of specification (< 4.85 V or > 5.15 V) (occasional)
First check

Measure ACB +5VA power supply voltage

Test point
ACB +5VA supply rail
Expected
Between 4.85 VDC and 5.15 VDC, +/-0.15V
Tools
Digital multimeter

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Access rear electronics enclosure for repair high voltage

Need to access rear panels to repair or replace machine components including electronics

Ranked causes

  1. 1. Component repair or replacement required (unknown)
First check

Bleed all gas pressure from the machine

Test point
System pressurization state
Expected
All pressure bled

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ACGO pressure relief valve not functioning patient safety

ACGO pressure relief weight does not relieve or pop off from seat; no purring sound heard during test

Ranked causes

  1. 1. ACGO pressure relief valve stuck or blocked (occasional)
First check

Set ACGO selector switch position

Test point
ACGO selector switch
Expected
ACGO position

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Anesthesia Control Board power supply voltage out of range high voltage

5 volt digital or analog supply measured outside 4.85–5.15 Vdc range, displayed in red text on diagnostics screen

Ranked causes

  1. 1. +5 VD or +5 VA supply voltage outside 4.85–5.15 Vdc (unknown)
First check

Measure +5 VD Supply voltage on Anesthesia Control Board

Test point
Anesthesia Control Board
Expected
4.85 to 5.15 Vdc, 4.85–5.15

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

APL valve assembly defective patient safety

APL valve not opening/closing properly or pressure not relieving

Ranked causes

  1. 1. APL valve assembly defective (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Auxiliary O2 flow rate outside specified limits

Flow test device reading does not match expected values for selected flowmeter settings during accuracy test

Ranked causes

  1. 1. Flowmeter calibration drift or internal wear (occasional)
First check

Verify flow test device capability

Test point
Flow test device specifications
Expected
Device measures 0-15 l/min with ±2% accuracy

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Auxiliary O2 flowmeter ball does not respond correctly

Flowmeter ball does not rise smoothly when flow control is rotated counterclockwise, does not stay at bottom when closed, or does not maintain steady position at set flow

Ranked causes

  1. 1. Flowmeter internal mechanism stiction or contamination (occasional)
First check

Rotate flow control clockwise to minimum

Test point
Flowmeter ball and control
Expected
Ball rests at bottom of flow tube and does not move

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Auxiliary O2 flowmeter not operating smoothly or ball not moving

Ball in flowmeter does not rise smoothly when control rotated counterclockwise or does not maintain steady position at set flow

Ranked causes

  1. 1. Flowmeter obstruction, sticky ball, or control valve blockage (common)
First check

Verify O2 supply available

Test point
O2 cylinder valve or pipeline connection
Expected
O2 cylinder valve ON or pipeline connected

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery does not maintain charge or runs down too quickly patient safety

System shuts down in less than 50 minutes when running on battery power under standard load

Ranked causes

  1. 1. Battery capacity degradation or failed battery cell (common)
First check

Configure system for battery test

Test point
System settings
Expected
Anesthetic Agent: None, Task Light: max, LCD Brightness: 5, Keypad brightness: 5

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bellows assembly damaged or stuck patient safety

Bellows housing not rotating freely or bellows cracked/leaking

Ranked causes

  1. 1. Bellows housing stuck or bellows damaged (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Bellows fills slowly during flush test
  • Bellows fills too quickly during flush test
  • Bottom task light not functioning
  • Calibration: Adjust drive gas regulator
  • Calibration: Airway gas module
  • Calibration: Airway Pressure and ACGO Span
  • Calibration: Airway Pressure Span (Paw Span)
  • Calibration: Bellows drop test
  • Calibration: Bleed Resistor
  • Calibration: Circuit leak test
  • Calibration: Flow control test
  • Calibration: Flow sensing test
  • Calibration: Inspiratory Flow Valve
  • Calibration: Low pressure leak test with ACGO
  • Calibration: Low pressure leak test without ACGO
  • Calibration: Manifold P Span
  • Calibration: Mixer Flow Sensor Zero
  • Calibration: O2 needle valve minimum flow
  • Calibration: O2 supply alarm test
  • Calibration: Paw and ACGO Span
  • Calibration: Paw Span
  • Calibration: Pipeline and cylinder high-pressure leak test
  • Calibration: Primary regulator output pressure
  • Calibration: Secondary regulator output pressure
  • Calibration: Touch screen misaligned
  • Calibration: Zero ACGO/AGSS
  • Calibration: Zero Gas Transducers
  • Carrier and Com Express board assembly replacement
  • Carrier board overheat
  • Caster replacement procedure
  • Caster wheel damaged or non-functional (without central brake)
  • Central brake system does not operate smoothly
  • Central vaporizer task light not functioning
  • Chassis leakage current exceeds limit
  • CPLD test fail
  • CPU coin battery replacement on display unit
  • Customize checklist items
  • Cylinder check valve defective
  • Cylinder inlet filter blocked or contaminated
  • Display arm replacement
  • Display unit disassembly and component access
  • Display unit encoder replacement
  • Display unit front assembly and LCD/Touch bezel replacement
  • Display unit power and data cable replacement
  • Display Unit power supply voltage out of range
  • Display unit removal from machine
  • Display unit speaker replacement
  • Exhalation valve diaphragm orientation incorrect
  • Expiratory flow sensor EEPROM failure
  • FIB +5VA fail
  • FIB +5VD fail
  • Final machine checks incomplete
  • FiO2 high alarm does not activate or deactivate
  • FiO2 high alarm does not trigger
  • FiO2 low alarm does not activate or deactivate
  • FiO2 low alarm does not trigger
  • Flow sensor calibration error
  • Flow sensors defective or misaligned
  • Flow valve leak test fail
  • Frame Interface Board power supply voltage out of range
  • Front caster damaged or non-functional (with new central brake linkage)
  • Front caster damaged or non-functional (with old central brake linkage)
  • Front panel com fail
  • Gas mixer output flow out of range
  • Gas supply pressure out of range
  • High pressure leak in cylinder supply
  • Inspiration stopped during ventilation
  • Inspiratory flow control valve orientation incorrect
  • Inspiratory flow sensor EEPROM failure
  • Inspiratory/expiratory ports assembly damaged
  • Leakage current exceeds safety limits
  • Machine serial number or options mismatch after board replacement
  • MGAS inlet filter residue > 40 sec
  • MGAS line blocked > 20 sec
  • MGAS sample flow deviation
  • MGAS sample gas out blocked > 20 sec
  • MGAS sample line not connected > 40 sec
  • MGAS sensor inop
  • Minute volume low alarm does not trigger
  • MOPV pressure relief valve not functioning
  • Multiple vaporizers can be enabled simultaneously
  • Multiple vaporizers can be turned on simultaneously
  • MVexp low alarm does not activate
  • N2O and O2 linkage failure
  • N2O flow does not decrease with O2 flow
  • O2 bypass mixer hardware failure
  • O2 flush valve malfunction
  • O2 sensor reading inaccurate
  • Peak airway pressure high alarm does not trigger
  • Peak airway pressure low leak alarm does not trigger
  • PEEP blockage alarm does not trigger
  • PEEP high blockage alarm does not activate
  • Pipeline inlet check valve stuck or leaking
  • Pipeline inlet filter contamination or blockage
  • Pipeline inlet manifold pressure transducer failure
  • Pneumatic fittings cannot be safely disconnected
  • Power failure message not displayed correctly
  • Power Management Board voltage out of range
  • Ppeak high alarm does not activate
  • Ppeak low leak alarm does not activate
  • Pressure limit circuit not functioning
  • Primary regulator module defective
  • Scavenger positive relief not functioning correctly
  • Sensor Interface Board power supply voltage out of range
  • Set all alarms to clinical limits
  • SIB pneumatic sense tubing leak
  • SIB pneumatic sense tubing leak test
  • Suction flow rate less than minimum
  • Suction flow rate low
  • Suction gauge does not bleed to zero when turned off
  • Suction gauge does not read accurately
  • Suction gauge needle drifts when turned off with knob not at Off
  • Suction gauge reading inaccuracy
  • Suction regulator does not hold set pressure
  • Suction regulator does not maintain set vacuum
  • Suction system leaks with valve Off
  • Suction vacuum does not bleed off when mode selector turned Off
  • Sustained high airway pressure
  • System does not power on or power alarm does not activate
  • Touchscreen unresponsive or fails
  • Transformer requires replacement
  • Vent engine drive gas check valve adhesion
  • Vent flow or pressure measurement out of range
  • Vent subsystem status parameter abnormal
  • Ventilator failure with low airway pressure
  • Ventilator flow valve current feedback failure
  • Ventilator flow valve DAC feedback failure

127 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
12 HR TESTSystem has been operating more than 12 hours without power cycle self-testSystem uptime exceeds 12 hours in Checkout state without a power on/off cycleinfo
ACB +5VD FAILVentilator failure due to voltage out of rangeVACB +5VD < 4.85 VDC or > 5.15 VDCerror
ACB CLOCK SPEEDAnesthesia Control Board clock speed out of specificationACB oscillator or timing circuit failure detected during self-testcritical
ACB CLOCK SPEEDSystem malfunction due to abnormal clock frequencyClock frequency > 1.1× or < 0.9× expectederror
ACB COM FAILCommunication failure between Display Unit and Anesthesia Control BoardDU unable to receive data from ACB or handshake protocol failurecritical
ACB COM FAILSystem malfunction due to communication loss between computersDisplay Computer receives no messages from Anesthesia Computer for 10 secondserror
ACB CPU TEST FAILAnesthesia Control Board CPU self-test failedCPU executes invalid instruction or core logic failure detectedcritical
ACB DCB COM FAILSystem malfunction due to Display Computer communication lossAnesthesia Computer receives no state messages from Display Computer for 10 secondserror
ACB DCB COM FAILCommunication failure between ACB and Display Control Board (DCB)ACB unable to communicate with DCB subsystem or protocol errorcritical
ACB EEPROM FAILMemory failure on ACBRead/write or CRC failure of EEPROM on ACBwarning
ACB FIB COMMUNICATION FAILGas controls failure, O2 onlyInterval between valid messages between ACB and FIB exceeds 100 millisecondserror
ACB FLASH FAILACB flash memory read/write failureFlash memory integrity check failed or data verify error during read/write operationcritical
ACB FLASH FAILSystem malfunction due to code memory CRC failureCRC failure occurs in code spaceerror
ACB HW WATCHDOGACB hardware watchdog timer expiredACB main processor did not service watchdog within timeout periodcritical
ACB MICROPROC ERRORACB microprocessor detected an internal error conditionMicroprocessor exception, parity error, or undefined instruction trapcritical
ACB MICROPROC ERRORSystem malfunction due to microcontroller exceptionUnexpected microcontroller exception (bus error, address error, etc.)error
ACB PROCEDURE COM FAILSystem malfunction due to procedure message communication failureNo procedure messages for 15±1 sec or message mismatch for 15±1 secerror
ACB PROCEDURE COM FAILCommunication failure during ACB procedure execution or command handshakeACB failed to acknowledge procedure command or lost connection mid-procedurecritical
ACB RAM ERRORACB RAM memory test or parity error detectedRAM self-test failed or uncorrectable memory parity error during operationcritical
ACB RAM ERRORSystem malfunction due to memory test failureMultiple bit errors detected during memory testerror
ACB REDUNDANT MEMORY FAILSystem malfunction due to redundant parameter storage failureRedundantly stored parameter could not be stored or was corruptederror
ACB REDUNDANT MEMORY FAILACB redundant or backup memory verification failedRedundant memory mirror check or backup memory integrity check failurecritical
ACB SW ERRORACB software detected an internal error or exception conditionFirmware exception handler invoked due to invalid state or logic errorcritical
ACB SW ERRORSystem malfunction due to unexpected software errorUnexpected software error occurserror
ACB SW WATCHDOGSystem malfunction due to software watchdog failureSoftware watchdog failed power-up test, timed out, or function delinquenterror
ACB SW WATCHDOGACB software watchdog timer expired during firmware executionFirmware task did not complete or report status within watchdog timeout periodcritical
ACB UNEXPECTED RESETACB experienced an unexpected reset or reboot without operator actionACB core or watchdog forced a system reset during operationcritical
ACB UNEXPECTED RESETSystem malfunction due to unexpected ACB resetUnexpected reset of anesthesia computer occurserror
ACGOVol and Apnea monitoring offNon Circle (ACGO) selectedinfo
ACMains POWER FAILAC mains power loss, system on batteryACMains_GOOD goes low and stays low for ≥3 secondswarning
AGSS HIGHScavenging flow is too highAGSS flow exceeds threshold based on configuration (>36, >50, or >80 l/min)info
AGSS LOWScavenging flow is too lowAGSS flow below threshold based on configuration (<21 or <42 l/min)info
AIR FRESH GAS FLOW SENSOR FAILFresh gas flow sensor for air inlet failed or out of rangeAir flow sensor reports zero or invalid value during gas delivery checkerror
AIR FRESH GAS FLOW SENSOR FAILGas controls failure, O2 onlyFIB reports air fresh gas flow sensor fail to ACBwarning
AIR MIXER VALVE FAILGas controls failure, O2 onlyAir mixer valve feedback current check failedwarning
AIR MIXER VALVE FAILAir mixer valve failure or stuck condition detectedAir mixer valve does not respond to command or position feedback invaliderror
AIR PIPE INVALIDCannot monitor air pipeline due to invalid pressureAir pipeline pressure reading is invalidwarning
AIR PRESS LOWAir supply pressure lowAir pipeline <252 kPa and air cylinder <2633 kPa for 1 secondwarning
AUXVol and Apnea monitoring offAuxiliary gas turned oninfo
BATTERY CHARGE FAILNo battery backup due to charging failureStandby: charge current >2.0A; powered on: battery current >1.3Awarning
BATTERY EMPTYSystem shutdown in less than 5 minutes due to battery depletionAvailable battery power 1-5 minutescritical
BATTERY FAILNo battery backup, battery voltage criticalBattery <10.5V, or charging anomalies (duration >12-16h, voltage >15.5V, current <0.15A)warning
BATTERY LOWPlug in power cable, system on batteryAvailable battery power 40-60 minuteswarning
BATTERY MISSINGNo battery backup, battery not connectedAny battery voltage <+2.0 VDCwarning
BATTERY REVERSED CONNECTIONSNo battery backup due to reversed battery polarityAny battery voltage <–10.0 VDCwarning
BATTERY V LOWPlug in power cable, battery power 20-40 minutes remainingAvailable battery power 20-40 minuteswarning
BATT V VERY LOWCritical battery depletion, plug in power cableAvailable battery power decreases to 5-20 minuteserror
BELLOWS COLLAPSEDUnable to drive bellows, manifold pressure excessiveManifold pressure > Paw + 10 + [0.25×(Insp valve flow)]info
BREATHING SYSTEM NOT LATCHEDBreathing system loose, detection switch indicates not latchedBreathing system detection switch indicates not latchedinfo
CAL DATA FAILURE IN EEPROMService calibration advised, default cal data in useCorrupt data in calibration region of EEPROMwarning
CARRIER BOARD OVERHEATCooling fans failed, system may overheatBoth Display Unit thermistors > 65°C, or neither thermistor reading in valid range 0-85°Cwarning
CHECK FLOW SENSORFlow sensor measurements do not meet criteria during mechanical breathsNo or negative flow on inspiratory flow sensor during inspiration, or negative flow on expiratory flow sensor for 6+ consecutive breathswarning
COM ERROR VENT TO ACBLoss of communication between Ventilator boundary and SIB CPUNo messages received from SIB CPU for 35 milliseconds after initial communication establishedcritical
CPLD TEST FAILComplex Programmable Logic Device self-test failureCPLD internal logic or integrity check failed during startup or periodic testcritical
CPU FAN SPEED FAILCPU cooling fan speed < 50% nominal (2500 RPM)CPU fan speed drops below 2500 RPM during operationwarning
CPU OVERHEATCPU thermal diode temperature exceeds 80°CCPU thermal diode temperature > 80°C sustained during operationwarning
DCB PROCEDURE COM FAILAnesthesia Computer loses procedure message communication with Display ComputerNo procedure messages for 15s +/- 0.5s, or messages do not match for 15s +/- 0.5s, during Therapy statecritical
DRIVE GAS LOSTVentilator has no drive gas; pressure too lowO2 PRESS LOW alarm active when O2 is drive gas, or Air PRESS LOW active when Air is drive gas, during mechanical ventilationcritical
DRV GAS FLOW SENSOR CAL ZERO FAILDrive gas flow sensor self-test calibration failedDrive gas flow sensor zero calibration fails at power-on self-testwarning
DRV GAS FLOW SENSOR FAILDrive gas flow sensor hardware failure detectedAnesthesia Computer detects drive gas flow sensor failure during operationwarning
***ERROR: Cannot load DOS! Any Key to retryDOS environment failed to load; boot sequence interruptedSystem startup attempting to initialize DOS environmenterror
ERROR!! INCOMPATIBLE SOFTWARESoftware version on USB Media is not compatible with installed subsystemSoftware Download process detects incompatible subsystem during compatibility checkerror
***ERROR: No bootable device available.System cannot locate or access bootable storage deviceSystem startup attempting to locate bootable devicecritical
***ERROR: Program load failed - CRC.Cyclic redundancy check error during software load; data corruption detectedSystem startup attempting to load software from CF cardcritical
***ERROR: RAM memory error.Random access memory hardware failure detectedSystem startup performing memory diagnosticscritical
***ERROR: SMM init failure. Press any key to continue.System management mode initialization failed; firmware initialization errorSystem startup attempting to initialize SMMerror
***ERROR: System reset: ECxx xx xx . . .Unexpected system reset with error code; general software failureSystem startup detecting previous unexpected reset eventerror
***ERROR: System reset: FFFF FF FFCMOS battery has lost contact or voltage below minimum thresholdSystem startup detecting CMOS battery contact or voltage failurecritical
***ERROR: The normal Windows CE application is not availableWindows CE application not found or failed to loadSystem startup attempting to launch Windows CE applicationerror
***ERROR: Watchdog circuit failed.Hardware watchdog timer circuit malfunction; system safety feature failedSystem startup detecting watchdog circuit failurecritical
EXP FLOW SENSOR EEPROM FAILUREExpiratory flow sensor calibration data cannot be readExpiratory flow sensor EEPROM read failure during initializationerror
FAN FAILCooling fan power status bit indicates fan failureFan power status bit is low (FAN1_GOOD bit = 0)warning
FIB +5VA FAILFIB +5VA power supply voltage out of toleranceVFIB +5VA < 4.85 VDC or > 5.15 VDC and FIB +12.5V within rangewarning
FIB +5VD FAILFrame Interface Board +5VD digital power supply out of specificationFIB +5VD supply voltage out of acceptable range detectederror
FLOW SENSOR CAL ERRORFlow sensor calibration out of rangeInspiratory or expiratory flow sensor or airway/manifold pressure sensor zero offset calibration failserror
FLOW VALVE LEAK TEST FAILFlow control valve leakage test failed during POST stateFlow control valve fails leakage test during power-on self-testwarning
FRONT PANEL COM FAILDisplay panel keypad controller communication failureKeypad controller fails to send life tick signal for >10 secondswarning
GAS FLOW > 300ML/MINGas flow exceeds 300 ml/min during Checkout: General sub-stateSystem detects gas flow > 300 ml/min during Checkout: General modewarning
GAS INLET VALVE BOOTUPGas inlet valve solenoid self-test failed at boot-upGas inlet valve solenoid fails test during system initializationcritical
INSP FLOW SENSOR EEPROM FAILUREInspiratory flow sensor calibration data cannot be readInspiratory flow sensor EEPROM read failure during initializationerror
LOSS OF VENT PARAMETER SETTINGSVentilator parameter settings communication lost from Display ComputerVentilator parameters not received from Display Computer for >10 seconds during ventilation modewarning
MANIFOLD PAW SENSOR FAILManifold pressure sensor calibration failed at boot-upManifold pressure sensor zero calibration fails during system initializationwarning
MGAS CHECK SAMPLE GAS OUT >20 SECAirway Module sample line blocked or occluded >20 secondsAirway Module reports continuous occlusion alarm lasting >20 secondswarning
MGAS INLET FILTER RESIDUE >40 SECD-Fend water trap residue buildup reducing airflow >40 secondsAirway Module reports water trap residue alarm lasting >40 secondswarning
MGAS LINE BLOCKED >20 SECAirway Module sample tubing blocked or water trap occluded >20 secondsAirway Module reports sample tubing occlusion alarm lasting >20 secondswarning
MGAS SAMPLE FLOW DEVIATIONSample flow < 80% of Airway Module nominal flow (120 ml/min for PROLITE)Sample flow stays below 80% of module nominal value for 20+ consecutive secondsinfo
MGAS SAMPLE LINE NOT CONNECTED >40 SAirway Module sample line or D-Fend module not installed >40 secondsAirway Module reports open circuit alarm lasting >40 secondswarning
MGAS SENSOR INOP > XXAirway Module hardware failure or initialization faultAirway Module reports hardware error (RAM/ROM/EEPROM failure, voltage error, lamp failure) or UPI initialization failureerror
MODULE NOT COMPATIBLEMonitoring Module detected is not compatible with system softwareIncompatible airway module connectedwarning
N2O FLOW > 100 ML/MINN2O flow exceeds 100 ml/min at auxiliary outletN2O flow > 100 ml/min in Auxiliary Gas Delivery modewarning
N2O FRESH GAS FLOW SENSOR FAILNitrous oxide fresh gas flow sensor failure or out of rangeN2O flow sensor reports invalid or zero reading during checkout or deliveryerror
N2O FRESH GAS FLOW SENSOR FAILGas controls failure, N2O onlyFIB reports N2O Fresh Gas Flow Sensor Fail alarm to ACBwarning
N2O MIXER VALVE FAILGas controls failure, N2O onlyN2O mixer valve feedback current check failedwarning
N2O MIXER VALVE FAILNitrous oxide mixer valve failed or does not respond to commandsN2O mixer valve stuck or position feedback invalid during gas mixing operationerror
N2O PIPE INVALIDCannot monitor N2O pipelineN2O Pipeline pressure is invalidwarning
N2O PRESS LOWN2O supply pressure lowN2O pipeline < 252 kPa and cylinder < 2633 kPa with N2O selectedinfo
NO EXPIRATORY FLOW SENSORNo expiratory sensor connected and not calibratingExpiratory flow sensor disconnected or missing during calibrationwarning
NO INSPIRATORY FLOW SENSORNo inspiratory sensor connected and not calibratingInspiratory flow sensor disconnected or missing during calibrationwarning
***NOTE: Alarm speaker not detected. Check connection.Speaker connection failure or speaker not respondingSystem startup detecting no audio output devicewarning
***NOTE: CMOS battery is weak. Please replace.CMOS battery voltage insufficient for proper real-time clock and settings retentionSystem startup detecting low CMOS battery voltagewarning
***NOTE: CPU Board Supply Voltage Out of Range.One or more processor board supply voltages (1.5V_CB, 5V_CB, or 3.3V_CB) outside acceptable rangeSystem startup performing voltage range validation on Carrier boardcritical
***NOTE: RTC date/time error. Battery may be weak.Real-time clock error detected; system time and date not maintainedSystem startup validating RTC function and battery statuswarning
O2 BYPASS VALVE FAILOxygen bypass valve failure or stuck conditionO2 bypass valve does not open/close as commanded or position feedback invaliderror
O2 BYPASS VALVE FAILGas controls failure, O2 onlyO2 bypass valve feedback current check failedwarning
O2 CAL ERRORCalibrate O2 sensorO2 sensor offset, slope, cell voltage out of range or O2 > 110%info
O2 FLOW < 150ML/MINO2 flow lowO2 flow < 150 ml/min in Therapy stateinfo
O2 FLUSH FAILUREO2 flush stuck on?O2 Flush Switch detected on continuously > 30 secinfo
O2 FRESH GAS FLOW SENSOR FAILGas controls failure, O2 onlyFIB reports O2 Fresh Gas Flow Sensor Fail alarm to ACBwarning
O2 FRESH GAS FLOW SENSOR FAILOxygen fresh gas flow sensor failure or out of range readingO2 flow sensor reports zero or invalid value during system checkouterror
O2 LATCHING VALVE FAILSystem MalfunctionO2 latching valve feedback current check failederror
O2 LATCHING VALVE FAILOxygen latching valve failure or unable to maintain stateO2 latching valve does not latch or unlatch as commanded during operationerror
O2 PIPE INVALIDCannot monitor O2 pipelineO2 Pipeline pressure is invalidwarning
O2 PRESS LOWO2 supply pressure lowO2 pipeline < 252 kPa and O2 cylinder < 2633 kPa for one seconderror
O2 SENSOR FAILUREReplace O2 sensorO2 < 5%info
ON/STANDBY SWITCH TO STANDBYConfirm to shutdownOn/Standby switch activated in Therapy stateerror
PATIENT VOLUME MISMATCH OCCURREDCalibrate, dry, or replace flow sensorsPATIENT VOLUME MISMATCH alarm occurred during Checkoutinfo
PAW OVERPRESSURE SIGNAL DISCONNECTEDVent fail. Monitoring onlySIB reports PAW Overpressure Signal Disconnected alarm to ACBwarning
PEEP PCV NOT AVAILABLEVol vent only. No PEEP or PSVPaw data in range but Pmanifold <= -15 cmH2Owarning
PORT PLUG POST FAILVent fail. Monitoring onlyPaw > 100 cmH2O for 1 second in POST statewarning
POWER CONTROLLER COM FAILInternal failure. System may shut downCommunications with PC and DC cannot be established for 10 secondswarning
POWER CONTROLLER SELF TESTInternal failure. System may shut downPower Controller fails self testserror
POWER SUPPLY BOOST > 75CCircuitry >75C shutdown possiblePower supply temperature near boost converter exceeds 75Cinfo
POWER SUPPLY BUCK > 75CCircuitry >75C shutdown possiblePower supply temperature exceeds 75Cwarning
PWR CNTRL DC-DC BOOST FAILInternal failure. System may shut downBoost converter failurewarning
PWR CNTRL DC-DC BUCK FAILNo battery backupBattery not connected or buck converter failurewarning
PWR CNTRL OUTPUT ERROR SWITCHES AC MAIN TO BACKUP POWERPower switched to BatteryOutput power over/under voltage, switches from AC Mains to backup powerwarning
PWR CNTRL SPL1 CHANNEL ERRORSystem MalfunctionSPL channel 1 current > 0.5A or voltage out of range (>13.2V or <12.1V) 3+ timeserror
PWR CNTRL SPL2 CHANNEL ERRORVent fail. Monitoring onlySPL channel 2 current > 0.85A or voltage out of range 3+ timeserror
PWR CNTRL SPL3 CHANNEL ERRORNo data. Ventilator failureSPL channel 3 current > 0.62A or voltage out of range 3+ timeserror
PWR CNTRL SPL4 CHANNEL ERRORPower Controller supply channel 4 (SPL4) error or out of specificationPower Controller SPL4 output voltage or current out of rangeerror
PWR CNTRL SPL4 CHANNEL ERRORGas controls failure, O2 onlySPL channel 4 current > 1.3A or voltage out of range 3+ timeswarning
PWR CNTRL SPL5 CHANNEL ERRORPower Controller supply channel 5 (SPL5) error or out of specificationPower Controller SPL5 output voltage or current exceeds limitserror
PWR CNTRL SPL5 CHANNEL ERRORGas controls failure, O2 onlySPL channel 5 current > 0.7A or voltage out of range 3+ timeswarning
PWR CNTRL SPL6 CHANNEL ERRORGas monitoring not availableSPL channel 6 current > 0.7A or voltage out of range 3+ timeswarning
REVERSE EXPIRATORY FLOWReverse exp flow. Check valves OK?Volume >= 20 ml toward patient on expiratory sensor and >= 5 ml on inspiratory during inspiration for 6 consecutive breathswarning
SIB +5VA FAILNo data. Ventilator failureVSIB +5VA < 4.85 VDC or > 5.15 VDCerror
SIB +5VD FAILNo data. Ventilator failureVSIB +5VD < 4.85 VDC or > 5.15 VDCerror
SIB +5VE2 FAILNo data. Ventilator failureSIB +5VE2 voltage < 4.85 VDC or > 5.15 VDCerror
STANDBY PATIENT DETECTIONNo fresh gas flow!3 volume breaths or 3 CO2 breaths detected within 30 seconds during Checkouterror
TOUCHSCREEN FAILURETouchscreen panel failureTouch panel controller fails to send a life tick for more than 10 secondserror
VENT AIRWAY OVERPRESS SIGNALHigh airway overpressure signal during ventilationHigh airway overpressure signal is set during mechanical ventilationwarning
VENT AIRWAY OVERPRESS SIGNAL FAILVentilator failure with airway overpressure signal and low pressureHigh Airway overpressure signal set with Paw < 90 cmH2O and Pmanifold < 80 cmH2Owarning
VENT FLOW VALVE FAIL CURRENTVentilator flow valve current feedback errorCurrent feedback check failed for flow control valveerror
VENT FLOW VALVE FAIL DACVentilator flow valve DAC feedback errorIncorrect DAC feedback check failed for flow control valveerror
VENT SUSTAINED PAW SDOWNSustained high airway pressure alertAirway pressure exceeds 100 cmH2O for 10 secondscritical

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI