GE Healthcare CARESCAPE R860: troubleshooting and error codes

The ServiceForge knowledge base holds 206 documented symptoms for the GE Healthcare CARESCAPE R860, with 426 ranked causes, 300 diagnostic checks with test points and expected values, and 326 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

System Check fails

One or more automated checks display red X instead of green check mark during System Check procedure on power up or after use

Ranked causes

  1. 1. Paw transducer check failure (unknown)
  2. 2. Barometric pressure check failure (unknown)
  3. 3. Relief valve check failure (unknown)
  4. 4. Exhalation valve check failure (unknown)
  5. 5. Expiratory flow sensor check failure (unknown)
  6. 6. Air flow sensor check failure (unknown)
  7. 7. Oxygen sensor check failure (unknown)
  8. 8. O2 flow sensor check failure (unknown)
  9. 9. Neonatal Flow Sensor check failure (unknown)
  10. 10. Resistance check failure (unknown)
  11. 11. Circuit measurements check failure (unknown)
First check

Select New Patient from Standby menu

Test point
Standby menu
Expected
Confirm or select Previous Patient

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System Check with O2 Therapy (Dual limb) fails

One or more checks display red X instead of green check mark during dual limb System Check procedure with O2 therapy option enabled

Ranked causes

  1. 1. Paw transducer check failure (unknown)
  2. 2. Barometric pressure check failure (unknown)
  3. 3. Relief valve check failure (unknown)
  4. 4. Exhalation valve check failure with calibration (unknown)
  5. 5. Expiratory flow sensor check failure (unknown)
  6. 6. Air flow sensor check failure (unknown)
  7. 7. Oxygen sensor check failure (unknown)
  8. 8. O2 flow sensor check failure (unknown)
  9. 9. Neonatal Flow Sensor check failure (unknown)
  10. 10. Resistance check failure (unknown)
  11. 11. Circuit measurements check failure (unknown)
First check

Select New Patient from Standby menu

Test point
Standby menu
Expected
Confirm or select Previous Patient

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exhalation valve calibration cannot achieve 80 cmH2O pressure

Exhalation valve calibration fails to generate 80 cmH2O pressure in the breathing circuit

Ranked causes

  1. 1. Insufficient flow from flow control valve (common)
  2. 2. Breathing circuit disconnected or leaking (common)
  3. 3. Low pressure vent engine system leaking (common)
  4. 4. Exhalation housing seat/seal contaminated or damaged (common)
  5. 5. No power to exhalation valve actuator (occasional)
  6. 6. Harness failure or poor connection (occasional)
  7. 7. VCB (valve control board) failure (rare)
First check

Measure flow from flow control valve

Test point
FCV outlet
Expected
At least 1 l/min

3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Vent engine leak test fails patient safety

Unable to build 70 cmH2O pressure or leak rate exceeds 10 ml/min during low pressure vent engine leak test

Ranked causes

  1. 1. Air and O2 flow path leaks or blockages (common)
  2. 2. Effort Valve leak (common)
  3. 3. Free Breathing Valve diaphragm or o-ring seal failure (occasional)
  4. 4. Safety valve seal failure (occasional)
  5. 5. Individual Total Flow path component leaks (occasional)
  6. 6. O2 Flow path component leaks (occasional)
  7. 7. Air Flow path component leaks (occasional)
First check

Close Air and O2 flow valves and remove DAC count checkmarks

Test point
Air and O2 flow valve controls
Expected
Valves closed, DAC count boxes unchecked

7 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot achieve 40 cmH2O pressure during exhalation valve calibration

Calibration running field displays 'Can't Achieve 40 cmH2O' message; exhalation valve check cannot generate required pressure

Ranked causes

  1. 1. Flow control valve (FCV) flow insufficient (occasional)
  2. 2. Breathing circuit disconnected or leaking (common)
  3. 3. Exhalation housing seat or seal dirty or damaged (occasional)
  4. 4. No power to exhalation valve actuator (rare)
  5. 5. Harness defective or disconnected (occasional)
  6. 6. Ventilation Control Board (VCB) failure (rare)
First check

Verify FCV flow during calibration

Test point
FCV output
Expected
Flow greater than 1 l/min

4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 and air flow disagree

O2 sensor check failed; difference between O2 sensor and calculated O2% from flow measurement exceeds 5% O2

Ranked causes

  1. 1. Air pressure regulator leak (occasional)
  2. 2. O2 pressure regulator leak (occasional)
  3. 3. Air flow sensor malfunction (occasional)
  4. 4. O2 flow sensor malfunction (occasional)
  5. 5. Total flow sensor malfunction (occasional)
  6. 6. O2 sensor malfunction (occasional)

9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit fails to boot with green LED on

Display unit will not boot; green power LED is illuminated

Ranked causes

  1. 1. Lithium coin cell at J17 not properly installed (common)
  2. 2. Defective lithium coin cell at J17 (common)
  3. 3. Internal Flash card at J10 not properly seated (occasional)
  4. 4. Defective Compact Flash card (occasional)
  5. 5. External Compact Flash card issue (occasional)
  6. 6. Failed display unit main PCB (rare)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exhalation valve calibration check fails

Checkout test fails during exhalation valve calibration; system sets flow at 4 l/min and pressure at 34 cmH2O, then opens valve and verifies expiratory pressure drops below 2 cmH2O within 250 ms. All ventilation modes remain available.

Ranked causes

  1. 1. Total flow too low or too high (common)
  2. 2. Significant leak in system (occasional)
  3. 3. Pexp sensor not returning to zero (occasional)
  4. 4. Exhalation valve out of calibration (occasional)
  5. 5. Exhalation flow sensor not detected (rare)
First check

Verify supply gas connection and flow capability

Test point
System output at 10 l/min calibration point
Expected
At least 4 l/min

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Safety relief valve cannot relieve pressure (single limb system) patient safety

Single limb system Safety Relief Valve test failure. Cannot achieve 2 cmH2O inspiratory pressure with 80 l/min flow, or valve cannot drop pressure >0.75 cmH2O within 250 msec, or inspiratory back pressure exceeds 1 cmH2O minimum.

Ranked causes

  1. 1. Inspiratory Safety Guard (ISG) not installed (common)
  2. 2. Insufficient supply gas or regulator/valve malfunction (common)
  3. 3. System leak (occasional)
  4. 4. Safety relief valve defective or stuck closed (occasional)
  5. 5. Manifold or inspiratory valve assembly malfunction (rare)
First check

Verify Inspiratory Safety Guard (ISG) is installed

Test point
Single limb system inlet
Expected
ISG present and secure

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to build expiratory or auxiliary pressure patient safety

Able to build 70 cmH2O inspiratory pressure but unable to build 70 cmH2O of either expiratory pressure or auxiliary pressure despite marginal leak rate

Ranked causes

  1. 1. Exhalation valve or expiratory pressure circuit issue (common)
  2. 2. Exhalation Valve Assembly pressure port seal failure (occasional)
  3. 3. Restrictor blockage or excessive restriction (occasional)
  4. 4. Restrictor gasket, zero valve gasket, or sensor o-ring failure (occasional)
  5. 5. Expiratory pressure sensor assembly or manifold failure (rare)
First check

Verify air shut-off valve is open and supply available

Test point
Air shut-off valve
Expected
Valve open and supply available

5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot achieve 2 cmH2O inspiratory pressure (single limb system)

Single limb system Paw transducer check failure. Air/O2 flow set to approximately 4 l/min, increased until 2 cmH2O inspiratory pressure achieved. If 1 cmH2O cannot be achieved with up to 60 l/min, test fails.

Ranked causes

  1. 1. Inspiratory Safety Guard (ISG) not installed (common)
  2. 2. Supply gas not connected or insufficient pressure (common)
  3. 3. Internal leak in system (occasional)
  4. 4. Regulator or Flow Control Valve malfunction (occasional)
First check

Verify Inspiratory Safety Guard (ISG) is installed

Test point
Single limb system inlet
Expected
ISG present and secure

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot achieve 34 cmH2O pressure

Pressure too low or total flow too low during Paw transducer check. Air/O2 flow valve set to approximately 8 l/min. If 33 cmH2O pressure cannot be achieved, test fails without alarm.

Ranked causes

  1. 1. Supply gas not connected or low pressure (common)
  2. 2. Internal leak in system (occasional)
  3. 3. Flow valve calibration drift (occasional)
  4. 4. Regulator or Flow Control Valve malfunction (occasional)
First check

Verify supply gas is connected

Test point
Gas inlet
Expected
Supply gas present

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot establish pressure (insufficient flow)

During checkout step, unable to achieve 8 Lpm flow required to attempt reaching 34 cmH2O pressure

Ranked causes

  1. 1. Breathing circuit disconnected or leaking (common)
  2. 2. Low pressure vent engine system leaking (common)
  3. 3. Exhalation valve improperly connected or seals missing (common)
  4. 4. Inadequate gas flow from flow control valves (occasional)
First check

Measure flow from system during checkout

Test point
System flow output
Expected
Achieve 8 Lpm

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot pressurize breathing circuit to 33 cmH2O (Paw transducer check)

During Paw transducer check, breathing circuit cannot reach 33 cmH2O pressure

Ranked causes

  1. 1. Breathing circuit disconnected or leaking (common)
  2. 2. Low pressure vent engine system leaking (common)
  3. 3. Exhalation valve improperly connected or seals missing (common)
  4. 4. Inadequate gas flow from flow control valves (occasional)
First check

Verify breathing circuit connection and integrity

Test point
All circuit connections and tubing
Expected
Connected, sealed, and leak-free

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display unit will not power on high voltage

Display unit fails to power up; no LED indication or system response

Ranked causes

  1. 1. Mains power not connected (common)
  2. 2. Power inlet problem (occasional)
  3. 3. On/Standby switch signal failure (occasional)
  4. 4. Power Controller Board failure (rare)
First check

Verify +12V Green LED illuminated at back of display behind cable cover

Test point
Back of display unit
Expected
Green LED illuminated

7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Excessive fan noise caution

Display unit fan is making excessive noise

Ranked causes

  1. 1. Display case fan inlet filter clogged or dirty (common)
  2. 2. Obstruction in case fan (common)
  3. 3. Com Express fan noise (occasional)
  4. 4. Failed fan assembly (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Excessive leak in breathing circuit

Leak value measured by checkout exceeds 0.5 Lpm, causing expiratory flow sensor normalization to be disabled

Ranked causes

  1. 1. Breathing circuit disconnected or leaking (common)
  2. 2. Low pressure vent engine leaking (common)
  3. 3. Exhalation valve improperly connected or seals missing (common)
  4. 4. Inadequate gas flow from flow control valves (occasional)
First check

Measure leak rate during checkout

Test point
Breathing circuit and vent engine
Expected
Leak less than 0.5 Lpm

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exhalation flow sensor pressure stabilization timeout

During exhalation valve calibration check, expiratory pressure does not stabilize within 0.1 cmH2O within 12 seconds

Ranked causes

  1. 1. Breathing circuit disconnected or leaking (common)
  2. 2. Low pressure vent engine system leaking (common)
  3. 3. Exhalation valve improperly connected or seals missing (common)
  4. 4. Inadequate gas flow from flow control valves (occasional)
First check

Monitor pressure stabilization time

Test point
Breathing circuit pressure
Expected
Stabilize within 0.1 cmH2O within 12 seconds

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Guaranteed flow not achieved

Cannot bring flow down to target flow with minimum flow command during checkout

Ranked causes

  1. 1. Breathing circuit disconnected or leaking (common)
  2. 2. Low pressure vent engine leaking (occasional)
  3. 3. Exhalation valve not properly connected or seals missing (common)
  4. 4. Flow control valve malfunction (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Insufficient flow during checkout step patient safety

Could not achieve minimum flow required for a checkout step

Ranked causes

  1. 1. Breathing circuit disconnected or leaking (common)
  2. 2. Low pressure vent engine leaking (common)
  3. 3. Exhalation valve improperly connected or unsealed (common)
  4. 4. Flow control valve malfunction or incorrect calibration (occasional)
First check

Verify breathing circuit connected and not leaking

Test point
Breathing circuit connections
Expected
All connections tight, no leaks

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low pressure leak check pressure inadequate (first half)

During first half of low pressure leak check, breathing circuit could not reach 34 cmH2O

Ranked causes

  1. 1. Breathing circuit disconnected or leaking (common)
  2. 2. Low pressure vent engine system leaking (common)
  3. 3. Exhalation valve improperly connected or seals missing (common)
  4. 4. Inadequate gas flow from flow control valves (occasional)
First check

Monitor breathing circuit pressure during first half of leak check

Test point
Breathing circuit pressure
Expected
Reach 34 cmH2O

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low pressure leak check pressure inadequate (second half)

During second half of low pressure leak check, breathing circuit could not reach 25 cmH2O

Ranked causes

  1. 1. Breathing circuit disconnected or leaking (common)
  2. 2. Low pressure vent engine system leaking (common)
  3. 3. Exhalation valve improperly connected or seals missing (common)
  4. 4. Inadequate gas flow from flow control valves (occasional)
First check

Monitor breathing circuit pressure during second half of leak check

Test point
Breathing circuit pressure
Expected
Reach 25 cmH2O

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Minimum flow not achieved

Could not achieve minimum flow required for checkout step

Ranked causes

  1. 1. Breathing circuit disconnected or leaking (common)
  2. 2. Low pressure vent engine leaking (occasional)
  3. 3. Exhalation valve not properly connected or seals missing (common)
  4. 4. Flow control valve malfunction (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Neonatal checkout failed at 20 Lpm

System was unable to achieve 20 Lpm of flow during neonatal checkout

Ranked causes

  1. 1. Breathing circuit disconnected or leaking (common)
  2. 2. Low pressure vent engine leaking (occasional)
  3. 3. Exhalation valve not properly connected or seals missing (common)
  4. 4. Flow control valve malfunction (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Neonatal checkout failed at 30 Lpm

System was unable to achieve 30 Lpm of flow during neonatal checkout

Ranked causes

  1. 1. Breathing circuit disconnected or leaking (common)
  2. 2. Low pressure vent engine leaking (occasional)
  3. 3. Exhalation valve not properly connected or seals missing (common)
  4. 4. Flow control valve malfunction (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Neonatal zero check failed at 4 Lpm

Neonatal flow sensor zero check failed because system unable to achieve 4 Lpm

Ranked causes

  1. 1. Breathing circuit disconnected or leaking (common)
  2. 2. Low pressure vent engine leaking (occasional)
  3. 3. Exhalation valve not properly connected or seals missing (common)
  4. 4. Flow control valve malfunction (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Normalized flow comparison failed

Normalized expiratory flow and total flow are more than 10% (or 200 ml) apart

Ranked causes

  1. 1. Breathing circuit disconnected or leaking (common)
  2. 2. Low pressure vent engine leaking (occasional)
  3. 3. Exhalation valve not properly connected or seals missing (common)
  4. 4. Flow control valve malfunction (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen flow sensor check failed - insufficient flow patient safety

System unable to achieve 30 Lpm during oxygen flow sensor check

Ranked causes

  1. 1. Breathing circuit disconnected or leaking (common)
  2. 2. Low pressure vent engine leaking (common)
  3. 3. Exhalation valve improperly connected or unsealed (common)
  4. 4. Flow control valve malfunction or incorrect calibration (occasional)
First check

Verify breathing circuit connected and not leaking

Test point
Breathing circuit connections
Expected
All connections tight, no leaks

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Paracube check failed - insufficient airflow for 50/50 air mix patient safety

Failed to achieve 15 Lpm airflow during 50/50 calibration

Ranked causes

  1. 1. Breathing circuit disconnected or leaking (common)
  2. 2. Low pressure vent engine leaking (common)
  3. 3. Exhalation valve improperly connected or unsealed (common)
  4. 4. Flow control valve malfunction or incorrect calibration (occasional)
First check

Verify breathing circuit connected and not leaking

Test point
Breathing circuit connections
Expected
All connections tight, no leaks

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Paracube check failed - insufficient airflow for 50/50 O2 mix patient safety

Failed to achieve 15 Lpm oxygen flow during 50/50 calibration

Ranked causes

  1. 1. Breathing circuit disconnected or leaking (common)
  2. 2. Low pressure vent engine leaking (common)
  3. 3. Exhalation valve improperly connected or unsealed (common)
  4. 4. Flow control valve malfunction or incorrect calibration (occasional)
First check

Verify breathing circuit connected and not leaking

Test point
Breathing circuit connections
Expected
All connections tight, no leaks

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Paracube check failed - insufficient oxygen flow patient safety

Failed to achieve 30 Lpm oxygen flow during O2 calibration

Ranked causes

  1. 1. Breathing circuit disconnected or leaking (common)
  2. 2. Low pressure vent engine leaking (common)
  3. 3. Exhalation valve improperly connected or unsealed (common)
  4. 4. Flow control valve malfunction or incorrect calibration (occasional)
First check

Verify breathing circuit connected and not leaking

Test point
Breathing circuit connections
Expected
All connections tight, no leaks

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Safety relief valve cannot relieve pressure (dual limb system) patient safety

Safety Relief Valve test failure during dual limb system check. Cannot relieve 30 cmH2O within 250 msec or back pressure with 75 l/min flow exceeds 15 cmH2O. Ventilation not allowed.

Ranked causes

  1. 1. Insufficient supply gas pressure or flow (common)
  2. 2. Relief valve opened during compliance check; exhalation diaphragm fault (occasional)
  3. 3. Regulator or Flow Control Valve malfunction (occasional)
  4. 4. Internal leak in system (occasional)
First check

Ensure supply gas is connected and set Air/O2 flow to 8 l/min

Test point
Gas inlet and flow setting
Expected
Supply gas present; 8 l/min flow established

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System unable to maintain minimum pressure during relief valve check patient safety

System pressure dropped below 25 cmH2O during relief valve check

Ranked causes

  1. 1. Breathing circuit disconnected or leaking (common)
  2. 2. Low pressure vent engine leaking (common)
  3. 3. Exhalation valve improperly connected or unsealed (common)
  4. 4. Flow control valve malfunction or incorrect calibration (occasional)
First check

Verify breathing circuit connected and not leaking

Test point
Breathing circuit connections
Expected
All connections tight, no leaks

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System unable to pressurize circuit to 10 cmH2O patient safety

Circuit could not be pressurized to minimum 10 cmH2O pressure

Ranked causes

  1. 1. Breathing circuit disconnected or leaking (common)
  2. 2. Low pressure vent engine leaking (common)
  3. 3. Exhalation valve improperly connected or unsealed (common)
  4. 4. Flow control valve malfunction or incorrect calibration (occasional)
First check

Verify breathing circuit connected and not leaking

Test point
Breathing circuit connections
Expected
All connections tight, no leaks

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Target flow not reached at minimum flow command patient safety

Flow remains greater than target flow with minimum command during checkout

Ranked causes

  1. 1. Breathing circuit disconnected or leaking (common)
  2. 2. Low pressure vent engine leaking (common)
  3. 3. Exhalation valve improperly connected or unsealed (common)
  4. 4. Flow control valve malfunction or incorrect calibration (occasional)
First check

Verify breathing circuit connected and not leaking

Test point
Breathing circuit connections
Expected
All connections tight, no leaks

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touchscreen fails to work patient safety

Touchscreen is not responding to touch input or non-functional

Ranked causes

  1. 1. Screen lock is enabled (common)
  2. 2. Internal cable or connection failure (occasional)
  3. 3. Failed Touch Controller Board (occasional)
  4. 4. Failed Touch Bezel Assembly (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Auxiliary pressure sensor zero out of range

Auxiliary pressure sensor zero DAC is outside normal range (631 to 968)

Ranked causes

  1. 1. Paux port blocked or not open to atmosphere (common)
  2. 2. Zero valve failure (occasional)
  3. 3. Pressure transducer failure (occasional)
First check

Verify Paux port is open to atmosphere

Test point
Paux port
Expected
Open to atmosphere directly or via open tube

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Barometric pressure calculation and measurement disagree

Measured barometric pressure differs from calculated pressure based on user-set altitude by more than 20%. If check fails, all ventilation modes remain available using altitude-based calculation.

Ranked causes

  1. 1. Incorrect altitude setting (common)
  2. 2. Barometric sensor calibration drift (occasional)
  3. 3. Vent Monitor board (VMB) failure (rare)
First check

Verify altitude setting matches actual location

Test point
System configuration
Expected
Altitude setting correct for installation location

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery Status abnormal

Battery Status shows Fail, Over Chg, or other non-float condition unexpectedly

Ranked causes

  1. 1. Battery charging circuit failure or degradation (occasional)
  2. 2. Battery disconnected and in Trickle mode >12 hours (occasional)
  3. 3. Severely degraded battery (occasional)
First check

Check Battery Status field in Power Diagnostics

Test point
Power Controller Readings Page 1, Battery Status
Expected
Float Chg (normal operation) or Discharge (on battery power)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Board Temperature out of range caution

Power Controller board temperature excessive on Power Diagnostics Page 2

Ranked causes

  1. 1. Cooling fan failure or malfunction (common)
  2. 2. Blocked or restricted airflow (occasional)
  3. 3. Excessive system load or ambient temperature (occasional)
First check

Monitor Board Temperature in Power Diagnostics

Test point
Power Controller Readings Page 2, Board Temperature field
Expected
<65C (green), <65

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exhalation flow sensor calibration test fails

Exhalation Flow Sensor output differs from Total Flow Sensor output by more than 20% at 10 l/min. All ventilator functions remain available; Expiratory Flow Sensor Error message may display.

Ranked causes

  1. 1. Exhalation flow sensor not installed or faulty (common)
  2. 2. Inspiratory and expiratory flows do not match (common)
  3. 3. Leak too large for normalization (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exhalation valve calibration pressure too low at high point

100 cmH2O calibration point reads greater than 65000 counts; test fails with no alarm displayed.

Ranked causes

  1. 1. FCV not generating sufficient flow (common)
  2. 2. Exhalation valve shaft binding or dirty (occasional)
  3. 3. Faulty exhalation valve or voice coil (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exhalation valve calibration slope too steep

Difference between 30 and 40 cmH2O calibration points greater than 6500 counts; test fails with no alarm displayed.

Ranked causes

  1. 1. FCV not generating sufficient flow (common)
  2. 2. Exhalation valve shaft binding or dirty (occasional)
  3. 3. Faulty exhalation valve or voice coil (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Expiratory pressure sensor zero out of range

Expiratory Pressure Sensor Zero Failure during Paw transducer check. Raw counts from Expiratory Pressure Sensor must be between 631 and 968 counts when Zero valve is actuated; otherwise test fails.

Ranked causes

  1. 1. Expiratory Pressure Zero valve defective (common)
  2. 2. Expiratory Pressure Sensor Board malfunction (occasional)
  3. 3. Vent Monitor board (VMB) failure (rare)
First check

Verify Expiratory Pressure Sensor raw counts with Zero valve actuated

Test point
Expiratory Pressure Sensor
Expected
631 to 968 counts

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flow sensors disagree during FCV calibration patient safety

Air or O2 flow sensor differs from total flow sensor by >20% or >1500 ml/min

Ranked causes

  1. 1. Regulator oscillation or instability (common)
  2. 2. Individual flow sensor defective (common)
  3. 3. Total flow sensor defective (occasional)
First check

Check gas regulator for oscillation

Test point
Regulator pressure output
Expected
Stable pressure without oscillation

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gas sampling system leak detected

Pressure drops excessively during leak test, indicating possible disconnection or damage to gas sampling tubing or connectors.

Ranked causes

  1. 1. Loose or disconnected tubing connections (common)
  2. 2. Damaged or worn tubing (occasional)
  3. 3. Water trap seal failure (occasional)
First check

Perform gas sampling leak test on E-series module

Test point
Gas sampling system
Expected
Pressure drop not more than 4 mmHg during one minute, ≤4 mmHg per minute
Tools
Pressure manometer (±1% accuracy at 25 PSI), pressure pump

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

High pressure air leak test failed due to barometric pressure

Vent engine high air pressure leak test failed because of invalid (low) barometric pressure reading

Ranked causes

  1. 1. Incorrect altitude setting (common)
  2. 2. Barometric sensor malfunction (occasional)
  3. 3. Vent Monitor Board malfunction (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

High pressure O2 leak test failed due to barometric pressure

Vent engine high O2 pressure leak test failed because of invalid (low) barometric pressure reading

Ranked causes

  1. 1. Incorrect altitude setting (common)
  2. 2. Barometric sensor malfunction (occasional)
  3. 3. Vent Monitor Board malfunction (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inspiratory Paw transducer check failed - low pressure

Single Limb Paw Transducer check failed; inspiratory pressure could not reach 1 cmH2O

Ranked causes

  1. 1. Inspiratory Safety Guard (ISG) not connected or leaking (common)
  2. 2. Low pressure vent engine system leaking (occasional)
  3. 3. Insufficient gas flow from flow control valves (occasional)
First check

Verify Inspiratory Safety Guard connection and integrity

Test point
ISG connector and seals
Expected
ISG connected and no visible leaks

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inspiratory Paw zero valve fails to generate minimum pressure (single limb system)

Inspiratory pressure transducer zero valve closed while constant flow provided. Check for minimum 1 cmH2O pressure. If pressure does not achieve minimum, test fails.

Ranked causes

  1. 1. Inspiratory Pressure Zero valve defective (common)
  2. 2. Inspiratory Pressure Sensor Board malfunction (occasional)
  3. 3. Vent Control board (VCB) failure (rare)
First check

Close inspiratory zero valve and provide constant flow

Test point
Pinsp sensor
Expected
Minimum 1 cmH2O, Minimum

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inspiratory Paw zero valve fails to open (single limb system)

After establishing pressure, inspiratory pressure transducer zero valve is opened to atmosphere and checked for zero cmH2O ±0.5 cmH2O. If pressure is out of range, test fails.

Ranked causes

  1. 1. Inspiratory Pressure Zero valve defective or stuck (common)
  2. 2. Inspiratory Pressure Sensor Board malfunction (occasional)
  3. 3. Vent Control board (VCB) failure (rare)
First check

Establish pressure then open inspiratory zero valve to atmosphere

Test point
Pinsp sensor
Expected
Zero cmH2O ±0.5 cmH2O, ±0.5

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inspiratory pressure sensor zero out of range

Inspiratory Pressure Sensor Zero Failure during Paw transducer check. Raw counts from Inspiratory Pressure Sensor must be between 631 and 968 counts when Zero valve is actuated; otherwise test fails.

Ranked causes

  1. 1. Inspiratory Pressure Zero valve defective (common)
  2. 2. Inspiratory Pressure Sensor Board malfunction (occasional)
  3. 3. Vent Control board (VCB) failure (rare)
First check

Verify Inspiratory Pressure Sensor raw counts with Zero valve actuated

Test point
Inspiratory Pressure Sensor
Expected
631 to 968 counts

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low pressure leak test failed due to barometric pressure

Vent engine low pressure leak test failed because of invalid (low) barometric pressure reading

Ranked causes

  1. 1. Incorrect altitude setting (common)
  2. 2. Barometric sensor malfunction (occasional)
  3. 3. Vent Monitor Board malfunction (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 flow sensor differs from total flow sensor patient safety

O2 flow sensor differs >15% from total flow sensor at 30 L/min nominal flow

Ranked causes

  1. 1. Gas regulator oscillating or unstable (common)
  2. 2. O2 flow sensor defective (common)
  3. 3. Total flow sensor defective (occasional)
First check

Check oxygen regulator for oscillation

Test point
O2 regulator pressure output
Expected
Stable pressure, no oscillation

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 measurement too low for 100% calibration patient safety

Measured O2 concentration below 41% during 100% span calibration

Ranked causes

  1. 1. Gas supply not properly connected (common)
  2. 2. Paracube sensor out of calibration range (occasional)
  3. 3. Paracube sensor defective (occasional)
First check

Verify gas supply connections

Test point
O2 inlet connection
Expected
Firmly connected

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 measurement too low for 21% calibration patient safety

Measured O2 concentration below 8% during 21% calibration

Ranked causes

  1. 1. Gas supply not properly connected (common)
  2. 2. Paracube sensor out of calibration range (occasional)
  3. 3. Paracube sensor defective (occasional)
First check

Verify gas supply connections

Test point
Air/O2 inlet connections
Expected
Firmly connected, no leaks

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pressure not maintained at calibration points

During calibration at 30, 20, 10, 7.5, 5, 2.5, and 1.25 cmH2O, pressure drops more than 1.25 cmH2O during a calibration point; test fails with no alarm displayed.

Ranked causes

  1. 1. FCV not generating sufficient flow (common)
  2. 2. Exhalation valve shaft binding or dirty (occasional)
  3. 3. Faulty exhalation valve or voice coil (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pressure too low in exhalation valve calibration

Exhalation valve cannot close to achieve 40 cmH2O when 4 l/min flow is present; test fails with no alarm displayed.

Ranked causes

  1. 1. FCV not generating sufficient flow (common)
  2. 2. Exhalation valve shaft binding or dirty (occasional)
  3. 3. Faulty exhalation valve or voice coil assembly (occasional)
First check

Verify FCV generates required flow

Test point
FCV output via PC Service Application
Expected
At least 4 l/min

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Rotary encoder fails to work

Rotary encoder is not responding or operational

Ranked causes

  1. 1. Rotary encoder cable not properly seated (common)
  2. 2. Failed rotary encoder (occasional)
  3. 3. Failed Carrier Board or Com Express Board Assembly (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

System will not boot from USB media during software installation

Ventilator display unit fails to boot from USB media during software installation process

Ranked causes

  1. 1. Incorrect or improperly inserted USB media (common)
  2. 2. Defective backup USB media (occasional)
  3. 3. Failed Carrier Board or Com Express Board Assembly (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • 'App Not Loaded' message is displayed
  • Air flow sensor temperature disagreement
  • Air inspiratory flow sensor calibration fails
  • Airway Module Gas Data unavailable
  • Battery 1 voltage out of range
  • Battery 2 voltage out of range
  • Battery not connected
  • Breathing circuit resistance measurement fails
  • D-fend water trap disconnection or failure
  • Exhalation flow sensor fails checkout
  • Exhalation valve cannot open sufficiently
  • Expiratory flow sensor calibration failed
  • Fall time measurements out of specification
  • Flow valve calibration point extrapolation failed (190 ml/min)
  • Flow valve DAC too low
  • Flow Valve excessive flow or stuck open
  • High pressure checkout aborted
  • Inspiratory flow below target
  • Low pressure excessive drop (exhalation valve calibration)
  • Low pressure leak and compliance check fails
  • Neonatal flow sensor calibration failed
  • Neonatal zero check failed with non-zero flow
  • No expiratory flow sensor detected
  • O2 inspiratory flow sensor calibration fails
  • O2 sensor test and calibration fails
  • Paw inspiratory and expiratory pressures disagree
  • Pressure too high in exhalation valve calibration
  • Sample flow rate out of specification
  • Sample line occluded alarm
  • Spirometry system leak detected
  • Spirometry volume reading error
  • System power does not turn on
  • Total flow sensor reading disagrees with neonatal flow
  • Total resistance invalid during inspiratory limb resistance check
  • Zero valve not responding to command
  • 1.5 Vdc supply voltage out of range
  • 12 Vdc supply voltage out of range
  • 3.3 Vdc supply voltage out of range
  • AC Status failure
  • Air flow sensor disagrees with total flow sensor reading
  • Airway module CO2 and O2 measurements inaccurate
  • Airway module communication failure
  • Airway pressure transducer check failed - sensors not zeroed
  • Airway pressure transducer inspiratory/expiratory difference exceeds tolerance
  • Alarm speaker not detected
  • Ambient pressure sensor reading incorrect
  • Backup buzzer does not sound
  • Bad air flow during exhalation valve calibration
  • Bad oxygen flow during exhalation valve calibration
  • Barometric pressure sensor check fails
  • Battery in use alarm does not sound on power loss
  • Battery minimum capacity verification
  • Bios upgrade error during software download
  • BIOS Upgrade Error during software download of 10.SP04
  • Calibration: Air Flow Control Valve
  • Calibration: Air flow valve
  • Calibration: Airway module
  • Calibration: Airway Module (Touch-based)
  • Calibration: airway pressure transducers
  • Calibration: barometric pressure transducer
  • Calibration: Battery performance test
  • Calibration: Exhalation valve
  • Calibration: FAS flow control valve
  • Calibration: flow control valves
  • Calibration: O2 Flow Control Valve
  • Calibration: O2 flow valve
  • Calibration: Paw span pressure adjustment
  • Calibration: Touchscreen
  • Cannot establish pressure during checkout - flow insufficient
  • Cannot establish pressure during checkout - pressure insufficient
  • Cannot pressurize breathing circuit during Paw transducer check
  • Checkout calculation error during resistance check
  • CMOS battery is weak
  • Configuration menu access denied
  • CPU Board Supply Voltage Out of Range
  • Degraded touchscreen performance
  • Dual Limb checkout skipped - not in checkout state
  • ECC failed
  • Exhalation flow sensor calibration difference excessive
  • Exhalation flow sensor removed during checkout
  • Exhalation valve calibration offset too high
  • Exhalation valve calibration point out of range (1.25 cmH2O)
  • Exhalation valve calibration point too high
  • Exhalation valve relief pressure relief time excessive
  • External device connection failure
  • Fall Time Measurement execution
  • Fan 1 status fail
  • Fan 2 status fail
  • Fan voltage out of specification
  • Flow difference between total and neonatal flow sensors
  • Flow too high during FCV calibration
  • Flow valve calibration cannot achieve 120 l/min
  • Flow valve calibration cannot hold 120 l/min flow
  • Flow valve calibration extrapolation failure
  • Free-breathing valve operation test
  • Inspiratory effort valve operation test
  • Inspiratory Paw transducer check failed - no gas supply
  • Inspiratory Paw transducer pressure zero calibration failed
  • Inspiratory Paw transducer zero error
  • Low air pressure
  • Low exhalation pressure during valve calibration
  • Low maximum flow
  • Low O2 pressure
  • Manage Service Password
  • Mechanical overpressure valve test
  • Multiple airway modules installed
  • Multiple failures in calibrations or system checks
  • Neonatal fail no gas supply
  • Neonatal flow sensor calibration fails
  • Neonatal reversed flow at 20 Lpm
  • Neonatal reversed flow at 30 Lpm
  • Neonatal zero check failed no gas supply
  • No air supply
  • No bootable device available
  • No gas supply
  • No oxygen supply
  • O2 Sensor calibration failed at 100% oxygen
  • O2 Sensor calibration failed at 21% oxygen
  • O2 sensor communication failed
  • Occlusion Valve Control adjustment
  • Oxygen flow sensor check failed - no O2 supply detected
  • Paracube O2 sensor check failed - 21% calibration incomplete
  • Pressure transducer zero ADC counts out of range
  • Pressure unstable during exhalation valve checkout
  • Program load failed - CRC
  • Pump Control adjustment
  • RAM memory error
  • Reference gas flow rate out of range
  • Regulator pressure unstable or exceeds limit
  • RTC date/time error
  • Safety valve relief check failed
  • Sample Gain adjustment
  • Single Limb checkout skipped - not in checkout state
  • Single Limb Relief Valve check failed
  • Single Limb Relief Valve check time measurement
  • Software download fails or shows 'Fail' status
  • System reset: ECxx xx xx
  • System reset: FFFF FF FF
  • Total flow too low during circuit occlusion test
  • Touch points are misaligned (out of calibration)
  • Touch points are not accessible (out of calibration)
  • Upstream inspiratory flow differs from total flow sensor
  • USB service stick license consumption on major upgrade
  • VCB buzzer operation test
  • Watchdog circuit failed
  • Zero Valve Control adjustment

146 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
120 Lpm flow testPre-calibration system flow verification message during exhalation valve calibrationDisplayed while system attempts to flow 120 Lpm prior to flow valve calibrationinfo
125 Cal Point Too High: ValueExhalation valve calibration point at 1.25 cmH2O exceeds maximum allowable value of 1.30 cmH2ODuring exhalation valve calibration when measurement at 1.25 cmH2O cal point is above 1.30 cmH2Oerror
190 point too highFlow valve calibration point at 190 ml/min cannot be properly extrapolatedDuring flow valve calibration when low flow point cannot be extrapolated accuratelyerror
AbortedCalibration or test in progress has been abortedUser or system aborts ongoing calibration or test procedurewarning
Air Flow Sensor ChkAir flow sensor check is in progressSystem initiates air flow sensor verification during checkoutinfo
Air Flow Sensor Chk CompleteAir flow sensor check passed successfullyAir flow sensor check completes without errorsinfo
Air Flow Sensor Chk FailAir flow sensor check failed during checkoutAir flow sensor check encounters error or fails verificationerror
Air Flow Sensor Communications FailureLoss of communication with Air Flow SensorSerial/communication link lost between vent engine and air flow sensorerror
Air Flow Sensor Fail: No AirAir flow sensor step of checkout failed due to lack of air supply pressureDuring air flow sensor check when low or no air supply pressure detectederror
Air Flush Time: Value secTime required to flush breathing system with air until oxygen concentration stabilizes to 0.1%Displayed during air flush procedure before oxygen-related checksinfo
Air Supply Pres HighAir supply pressure exceeds maximum thresholdAir supply pressure > 95 psi for > 0.5 secondswarning
Air Supply Pres LowAir supply pressure below minimum thresholdAir supply pressure < 24.3 psi for > 0.5 secondswarning
Air Supply Pressure Sensor Out of RangeAir supply pressure sensor data outside acceptable rangeSensor data outside 195-4760 mV rangewarning
Air Temperature Sensor FailureAir temperature sensor data outside acceptable rangeSensor data outside 0-60°C rangeinfo
Air temperature Sensor FailureVentilator unable to measure air flow temperature; delivered FiO2 may not match setAir flow temperature sensor data unavailable or out of rangeinfo
Air Valve CalAir valve calibration is in progressSystem begins air valve calibration during checkoutinfo
Air Valve Cal CompleteAir valve calibration passed successfullyAir valve calibration completes without errorsinfo
Air Valve Cal FailAir valve calibration failedDuring air valve calibration when verification failserror
Airway Module Gas Sampling ErrorMGAS module has detected a gas sampling fault conditionGas sampling malfunction detected by MGAS modulewarning
Alarm Confirm ErrorDU detects speaker current sense bit false when attempting to sound speakerSpeaker fails to produce audible output when alarm activatederror
Alarm Light FailureAlarm light not functioningAlarm light not workingwarning
Alarm speaker not detectedSpeaker connection is faultyStartup (POST) when speaker not detectederror
Already reset: file D:\ELOCAL.DAT not foundTouchscreen has been reset previouslyPerforming Reset Touchscreen Cal a second timeinfo
Aux Span Counts: ValueAuxiliary pressure sensor DAC counts at 100 cmH2O reference point (normal range: 3026 to 3589)Displayed during auxiliary pressure sensor calibrationinfo
Aux ZeroAuxiliary pressure sensor zero calibration is in progressSystem initiates zero-point calibration for auxiliary pressure sensorinfo
Aux Zero CompleteAuxiliary pressure sensor zero calibration passedAux zero calibration completes successfullyinfo
Aux Zero Fail: ValueAuxiliary pressure sensor zero DAC reading is outside normal range (631 to 968)During auxiliary zero calibration when DAC reading exceeds specified rangeerror
Backup Buzzer POSTCurrent to buzzer indicates audio is not soundingBackup buzzer POST test detects no current to buzzer or audio not soundingerror
Backup Mode Active - ApneaBackup ventilation activated due to insufficient spontaneous breathingSpontaneous breathing insufficientwarning
Backup Mode Active - MVexp LowBackup ventilation activated due to low expired minute volumeMeasured expired minute volume < 50% of MVexp low alarm limitwarning
Bad Air FlowCannot achieve proper 4 l/min air flow during exhalation valve calibrationDuring exhalation valve calibration when air flow control valve cannot deliver sufficient flowerror
Bad Low Pressure: ValueLowest pressure achieved during exhalation valve calibration exceeds 2.5 cmH2O maximum allowableDuring exhalation valve calibration when minimum pressure cannot be reached low enougherror
Bad O2 FlowCannot achieve proper 4 l/min oxygen flow during exhalation valve calibrationDuring exhalation valve calibration when oxygen flow control valve cannot deliver sufficient flowerror
Barometric ChkBarometric pressure sensor check is in progressSystem initiates barometric pressure verification during checkoutinfo
Barometric Chk CompleteBarometric pressure check passed successfullyBarometric check completes without errorsinfo
Barometric Chk FailBarometric pressure sensor reading and calculated pressure differ by more than 20%During barometric check when sensor reading deviates excessively from altitude-based calculationerror
Barometric Pressure Sensor Out of RangeBarometric pressure sensor data outside acceptable rangeSensor data outside 2163-4340 mV or 732-1196 cmH2O rangeinfo
Battery Charge Fail or Standby Current HighBattery charging failure or excessive standby power drawSystem powered on with battery current > 1.2A, or standby charge current > 1.7Awarning
Battery Fail or Bulk Charge > 24 HrsBattery short circuit or excessive charging timeAny battery < 10 VDC while charging, or charging > 24 hourswarning
Battery in useSystem is operating on internal battery power due to AC power lossAC power cord unplugged or power supply failure with system switch Onwarning
Battery in use – Low Internal Battery – 30 MinMains power unavailable; internal battery remaining time less than 30 minutesMains power lost and battery discharge algorithm indicates <30 min remainingwarning
Battery MissingBattery not detected or disconnectedBattery voltage between -1.0 and +1.0 VDCwarning
Battery Not ConnectedMains power lost and system operating on batteryMains power unavailable; system powered by internal battery > 300 mswarning
Battery Reversed ConnectionBattery polarity reversed or severely dischargedAny battery voltage < -1.0 VDCwarning
Calibrating Flow ValveFlow valve calibration procedure is currently runningSystem initiates flow valve calibration during checkoutinfo
Cal Offset Too HighExhalation valve calibration offset exceeds acceptable rangeCalibration point at 1.25 cmH2O is greater than 10000 counts during exhalation valve checkerror
Cal Point Too HighExhalation valve calibration points exceed acceptable rangeCalibration points are greater than 65000 counts during exhalation valve checkerror
Cal Slope Too SteepExhalation valve calibration slope exceeds acceptable rangeCalibration points between 30 cmH2O and 40 cmH2O are greater than 6500 countserror
Can't Achieve 40 cmH2OExhalation valve calibration cannot reach required pressure thresholdExhalation valve check calibration cannot generate 40 cmH2O pressure in breathing circuiterror
Can't Achieve 80 cmH2OExhalation valve calibration cannot generate 80 cmH2O pressure in breathing circuitDuring exhalation valve calibration when 80 cmH2O pressure target cannot be reachederror
Can't Establish Pressure: FlowUnable to achieve 8 Lpm flow required to attempt reaching 34 cmH2O during checkoutDuring checkout pressure establishment step when flow target cannot be meterror
Can't Establish Pressure: FlowCannot achieve required flow during Dual Limb or Single Limb checkout stepDuring Dual Limb checkout, unable to achieve 8 Lpm; during Single Limb checkout, unable to achieve required pressureerror
Can't Establish Pressure: PressureCannot achieve required pressure during Paw Transducer or Relief Valve checkSingle Limb Paw check: inspiratory pressure cannot reach 1 cmH2O with 60 l/min; Single Limb Relief check: cannot reach 2 cmH2O with 80 l/minerror
Can't Extrapolate TableFlow valve calibration extrapolation results in DAC counts exceeding 65,000 maximumDuring flow valve calibration when weak supply pressure forces extrapolation beyond acceptable limitserror
Can't Pressurize1: ValueDuring first half of low pressure leak check, breathing circuit cannot reach 34 cmH2ODuring low pressure leak check part 1 when target pressure not achievederror
Can't Pressurize2: ValueDuring second half of low pressure leak check, breathing circuit cannot reach 25 cmH2ODuring low pressure leak check part 2 when target pressure not achievederror
Can't Pressurize: ValueBreathing circuit cannot reach required pressure during various checkout steps (33 cmH2O for Paw, 30 cmH2O for relief valve, stable exp pressure for exhalation valve)During Paw check, relief valve check, or exhalation valve check when target pressure not achievableerror
Carrier Board OverheatAn internal DU board has overheatedDU board temperature excessiveerror
Case Fan Speed FailCase fan speed less than 800 RPMCase fan speed drops below 800 RPM thresholderror
Checking Supply Press1System is testing gas supply source and integrity during FCV calibration step 1During FCV calibration when supply pressure verification beginsinfo
Checking Supply Press2System is testing gas supply source and integrity during FCV calibration step 2During FCV calibration when second supply pressure verification occursinfo
Checking Supply Press3: ValueSystem is testing gas supply source and displaying measured supply pressure in psi during FCV calibrationDuring FCV calibration when supply pressure is measured and displayedinfo
Checkout AbortedCheckout procedure has been abortedUser initiates abort or system detects error requiring procedure restartwarning
Checkout DoneCheckout procedure completed successfullyCheckout procedure finishes all tests without critical failuresinfo
Ci*1000=ValueIntermediate compliance calculation value during low pressure leak checkDisplayed during low pressure leak check compliance computationinfo
Circ Unparked Patient DetectPatient detected connected during standby statePatient detected connected during standby stateerror
CMOS battery is weakCarrier board battery voltage is below acceptable levelStartup (POST) when battery voltage detected as weakerror
CO2 inaccurateCO2 values from module may be inaccurateCO2 sensor reading outside acceptable accuracy rangeinfo
CPU Board Supply Voltage Out of RangeOne or more CPU board supply voltages are outside specificationStartup (POST) when 1.5V_CB, 5V_CB, or 3.3V_CB is out of rangecritical
CPU Fan Speed FailCPU fan speed less than 800 RPMCPU fan speed drops below 800 RPM thresholderror
CPU OverheatDU thermistor temperature reading exceeds 80°CTemperature sensor reading >80°C on DU boarderror
Degraded BatteryBattery voltage or charge state abnormalBattery voltage > 16.5 VDC and current >= 0.25A, or not in float charge, or trickle charging > 12 hourswarning
Disabling normalization - invalid leakExpiratory flow sensor normalization is disabled because leak exceeds 0.5 LpmDuring checkout when leak measurement exceeds 0.5 Lpm thresholdwarning
Disagree Temp: ValueAir (or oxygen) flow sensor temperature differs from total flow sensor temperature by more than 9°CDuring air or oxygen flow sensor check when temperature discrepancy exceeds thresholderror
Disagree Tot: Value1 Air: Value2Air flow sensor reading differs from total flow sensor by more than 15% at nominal 30 l/min flowDuring air flow sensor check when measured values diverge excessivelyerror
Downstream Flow Diff: ValueFlow difference between total flow sensor and neonatal expiratory flow sensor wire measured in ml/minMeasured during checkout as informational parameterinfo
DU RAM ERRORSelf test failure or multi bit error detectedSelf test failure or multi bit error detectederror
DU to VCB Com Error FailedCommunication from Display Unit to Ventilation Control Board failedCommunications not completed within 90 seconds after startup or 10 second window after establishederror
DU to VMB Com Error FailedCommunication from Display Unit to Ventilation Management Board failedCommunications not completed within 90 seconds after startup or 10 second window after establishederror
ECC failedRAM memory error correction code circuitry is faultyStartup (POST) when ECC check failscritical
End Pressure: ValueTime required to relieve breathing circuit pressure from 34 cmH2O to below 2 cmH2O during exhalation valve checkDuring exhalation valve check when pressure relief time is measurederror
ERROR!! INCOMPATIBLE SOFTWAREThe software version on the USB Media is not compatible with the installed subsystem.Software Download detects an incompatible subsystem during download process.error
Exhalation Flow Sensor Comm FailureFailed communications with exhalation flow sensorLoss of serial/communication link between VMB and exhalation flow sensor PCBcritical
Exhalation Flow Sensor ComparisonExpelled tidal volume differs excessively from inspired tidal volumeAdult/Ped: (TVexp–TVinsp) > (0.3*TVinsp) or 100ml for 6 breaths; Neo: >0.3*TVinsp or 18mlerror
Exh Flow Sensor ChkExhalation flow sensor check is in progressSystem initiates exhalation flow sensor verification during checkoutinfo
Exh Flow Sensor Chk CompleteExhalation flow sensor check passed successfullyExhalation flow sensor check completes without errorsinfo
Exh Flow Sensor Chk FailExhalation flow sensor check failed; may also indicate total flow sensor failureDuring exhalation flow sensor check when verification failserror
Exh Valve CalExhalation valve calibration is in progressSystem begins exhalation valve calibration during checkoutinfo
Exh Valve Cal ChkExhalation valve check (post-calibration verification) is in progressSystem performs exhalation valve verification after calibrationinfo
Exh Valve Cal Chk CompleteExhalation valve calibration check passed successfullyExhalation valve check completes without errorsinfo
Exh Valve Cal Chk FailExhalation valve calibration check failedDuring exhalation valve check when verification failserror
Exh Valve Cal CompleteExhalation valve calibration completed successfullyExhalation valve calibration completes without errorsinfo
Exh Valve Cal FailExhalation valve calibration failedDuring exhalation valve calibration when target parameters cannot be achievederror
Exp Flow CalExpiratory flow sensor calibration is in progressSystem initiates flow sensor calibration during checkoutinfo
Exp Valve Step Sweep DoneExhalation valve step sweep calibration has completedAfter exhalation valve sweep procedure finishesinfo
Exp Valve SweepExhalation valve step sweep calibration is in progressSystem begins exhalation valve sweep calibration procedureinfo
Failed 120l/min flow: ValueDuring flow control valve calibration, unable to maintain 120 l/min flow for 1 second before supply pressure drops below 30 psiDuring FCV calibration when supply pressure drops below minimum during high flow testerror
Failed Supply Press2: ValueDuring FCV calibration step 2, gas supply pressure is inadequateDuring second FCV calibration supply pressure check when reading is insufficienterror
Failed Supply Press3: ValueDuring FCV calibration step 3, gas supply pressure is inadequateDuring third FCV calibration supply pressure check when reading is insufficienterror
Failed Supply Press4: ValueDuring FCV calibration step 4, gas supply pressure is inadequateDuring fourth FCV calibration supply pressure check when reading is insufficienterror
Failed Supply Press5: ValueDuring FCV calibration step 5, gas supply pressure is inadequateDuring fifth FCV calibration supply pressure check when reading is insufficienterror
Failed Supply Press6: ValueDuring FCV calibration step 6, gas supply pressure is inadequateDuring sixth FCV calibration supply pressure check when reading is insufficienterror
Failed Supply Press7: ValueDuring FCV calibration step 7, gas supply pressure is inadequateDuring seventh FCV calibration supply pressure check when reading is insufficienterror
Failed Supply Press8: ValueDuring FCV calibration step 8, gas supply pressure is inadequateDuring eighth FCV calibration supply pressure check when reading is insufficienterror
Failed Supply Press9: ValueDuring FCV calibration step 9, gas supply pressure is inadequateDuring ninth FCV calibration supply pressure check when reading is insufficienterror
Failed Supply Press: ValueDuring FCV calibration, gas supply pressure is inadequateDuring flow control valve calibration when initial supply pressure check failserror
FAIL No 120 Min: ValueFlow valve calibration times out without reaching 120 l/min target flowDuring flow valve calibration when 120 l/min target cannot be achieved within timeout perioderror
Fan FailFan Power Status Bits low; main fan or DC-DC converter fan is lowMain fan or DC-DC converter fan power status indicates low/faulterror
fCompliance*1000= ValueIntermediate flow compliance calculation value during low pressure leak checkDisplayed during low pressure leak check compliance computationinfo
fCorrectedLeak*1000 = ValueIntermediate corrected leak calculation value during low pressure leak checkDisplayed during low pressure leak check leak calculationinfo
Final Flow Achieved: ValueFlow value in ml/min achieved during current checkout stepDisplayed at end of flow-related checkout stepinfo
Final Rising Flow Achieved: ValueFlow rise rate value in ml/min achieved during current checkout stepDisplayed during rising flow checkout proceduresinfo
fLeak*1000 = ValueIntermediate leak calculation value during low pressure leak check in ml/minDisplayed during low pressure leak check leak computationinfo
Flow Cal - Waiting for SensironTotal flow sensor is being configured to read correct gas type for current valve calibrationDuring flow calibration when total flow sensor setup is in progressinfo
Flow Diff: ValueFlow difference between inspiratory (total) and expiratory flow sensors during exhalation flow sensor check (must be less than 20%)During exhalation flow sensor check when difference is calculatederror
Flow/O2 Disagree: Value1/Value2O2 sensor check failed; O2 and air flow set to 15 l/min; difference exceeds 5% O2During O2 sensor checkout procedureerror
Flow Too High: ValueFCV leaks more than 1150 ml/min during calibrationDuring FCV calibration routineerror
Flow Valve Cal CompleteFlow valve calibration passedAfter flow valve calibration routineinfo
Flow Valve Cal FailFlow valve calibration failedDuring flow valve calibration routineerror
Flow Valve DAC Too LowFCV-DAC counts went too low without corresponding decrease in flowDuring flow valve operationerror
fPressureDifference1 10 00 = ValueIntermediate value in low pressure leak check compliance calculationDuring leak check calibrationinfo
fPressureDifference2 10 00 = ValueIntermediate value in low pressure leak check compliance calculationDuring leak check calibrationinfo
FRC InvalidFRC measurement unable to be completed within last 5 secondsFRC measurement started but not finished within 5-second windowinfo
FRC missedFRC series measurement unable to start within last 5 secondsAnother procedure or condition prevented FRC measurement from startinginfo
Front Panel Com FailDisplay CPU board fails to send life tick for >10 secondsLoss of communication heartbeat from DU CPU board to main controller for >10 seccritical
Guaranteed Flow Not Achieved: ValueCould not bring flow down to target flow with minimum flow commandDuring checkout flow control testerror
High Pres Air Leak Test Fail (Baro)Vent engine high air pressure leak test failed; invalid (low) barometric pressure readingDuring high pressure air leak testerror
High Pres O2 Leak Test Fail (Baro)Vent engine high O2 pressure leak test failed; invalid (low) barometric pressure readingDuring high pressure O2 leak testerror
High Pressure: AbortedCheckout step in progress was aborted due to high pressureWhen system pressure exceeds safe limits during checkoutwarning
High Pressure Air Leak TestVent engine high air pressure leak test in progressDuring checkout procedureinfo
High Pressure Air Leak Test CompleteVent engine high air pressure leak test passedAfter high pressure air leak testinfo
High Pressure Air Leak Test FailVent engine high air pressure leak test failed; measured leak was too highDuring high pressure air leak testerror
High Pressure O2 Leak TestVent engine high O2 pressure leak test in progressDuring checkout procedureinfo
High Pressure O2 Leak Test CompleteVent engine high O2 pressure leak test passedAfter high pressure O2 leak testinfo
High Pressure O2 Leak Test FailVent engine high O2 pressure leak test failed; measured leak was too highDuring high pressure O2 leak testerror
InspFlowAccumulator = Value1, _jj=Value2Intermediate values in low pressure leak check compliance calculationDuring leak check calibrationinfo
Insp flow below target: >Target = Value1 >Insp = Value2Flow cannot be controlled to target; measured flow already lower than targetDuring inspiratory flow control testerror
Insp Paw Xducer CheckSingle Limb Paw Transducer check in progressPaw transducer calibration check initiated and runninginfo
Insp Paw Xducer Check CompleteSingle Limb Paw Transducer check has finished executionPaw transducer check completed all stepsinfo
Insp Paw Xducer Check Fail (Aborted)Single Limb Paw Transducer check was abortedPaw transducer check was interrupted or aborted during executionwarning
Insp Paw Xducer Check Fail (Low Pressure)Single Limb Paw Transducer check failed; inspiratory pressure too lowInspiratory pressure could not reach 1 cmH2O during checkerror
Insp Paw Xducer Check Fail (No Gas Supply)Single Limb Paw Transducer check failed due to absent gas supplyNo gas supply detected during Paw transducer checkerror
Insp Paw Xducer Check Fail (PressureZeroCal)Inspiratory pressure sensor zero calibration failedInspiratory pressure sensor was not successfully zeroed during calibrationerror
Insp Paw Xducer OKSingle Limb Paw Transducer check passed successfullyPaw transducer check completed all tests within acceptable parametersinfo
Insp Paw Xducer Zero Error: ValueInspiratory pressure sensor zero valve operation failedInspiratory pressure sensor zero valve was not successfully opened and closederror
Insp Span Counts: ValueInspiratory pressure sensor DAC counts for 100 cmH2O; range 3026-3589During inspiratory pressure calibrationinfo
Low Air Pressure: ValueAir supply pressure below 30 psiDuring system operation or calibrationwarning
Low Exp Pressure: Value1, Value2During exh valve cal, pressure dropped more than 1.25 cmH2O in 4 msecDuring exhalation valve calibrationerror
Low Internal Battery – 10 MinMains power not available; internal battery supply less than 10 minutesBattery discharge algorithm calculates less than 10 minutes remaining and mains power not availableerror
Low Internal Battery – 1 MinMains power not available; internal battery supply less than 1 minuteBattery discharge algorithm calculates less than 1 minute remaining and mains power not availableerror
Low Internal Battery – 20 MinMains power not available; internal battery supply less than 20 minutesBattery discharge algorithm calculates less than 20 minutes remaining and mains power not availableerror
Low Internal Battery – 5 MinMains power not available; internal battery supply less than 5 minutesBattery discharge algorithm calculates less than 5 minutes remaining and mains power not availableerror
Low Max Flow: ValueFCV indicator that system will use degraded wall supply routine for FCV calibrationWhen gas supply pressure is marginalwarning
Low O2 Pressure: ValueO2 supply pressure below 30 psiDuring system operation or calibrationwarning
Low Pres Leak Test Fail (Baro)Vent engine low pressure leak test failed; invalid (low) barometric pressure readingDuring low pressure leak testerror
Low Pressure Leak ChkBreathing circuit leak check in progressDuring checkout procedureinfo
Low Pressure Leak Chk CompleteBreathing circuit leak check passedAfter low pressure leak checkinfo
Low Pressure Leak Chk FailBreathing circuit leak check failedDuring breathing circuit leak checkerror
Low Pressure Leak TestVent engine low pressure leak test in progressDuring checkout procedureinfo
Low Pressure Leak Test CompleteVent engine low pressure leak test passedAfter low pressure leak testinfo
Low Pressure Leak Test FailVent engine low pressure leak test failed; leak was too highDuring low pressure leak testerror
Measured Resistance: ValueResistance measured during CheckoutDuring checkout procedureinfo
Measure Insp Resistance CheckInspiratory resistance check in progressDuring checkout procedureinfo
Measure Insp Resistance Check CompleteInspiratory resistance check passesAfter inspiratory resistance checkinfo
Measure Insp Resistance Check FailInspiratory resistance check failedDuring inspiratory resistance checkerror
Measure Insp Resistance Check ValueMeasured inspiratory resistance valueDuring inspiratory resistance measurementinfo
Measure Resis ChkBreathing circuit resistance measurement in progressDuring checkout procedureinfo
Measure Resis Chk CompleteBreathing circuit resistance measurement completed successfullyAfter resistance measurementinfo
Measure Resis Chk FailBreathing circuit resistance measurement failedDuring resistance measurementerror
Measuring P 0.1P0.1 procedure in progressDuring checkout procedureinfo
Minimum Flow not achieved: ValueCould not achieve minimum flow required for checkout stepDuring flow control test in checkouterror
Module Check Sample Gas Out > 20 SecOcclusion in airway module sample gas outletSample gas outlet blocked for > 20 secondswarning
Module D-fend Not Connected > 40 SecD-fend module not detected or disconnectedD-fend module not detected for > 40 secondswarning
Module Inlet Filter Residue > 40 SecResidue buildup on water trap membrane reducing airflowResidue detected on water trap membrane for > 40 secondsinfo
Module Line Blocked > 20 SecGas sampling line or water trap blockedOcclusion detected in sampling line or water trap for > 20 secondswarning
Module Not CompatibleMonitoring module software incompatible with systemModule detected is not compatible with system softwarewarning
Module Sample Flow DeviationSample gas flow rate too slow for extended periodSample flow too slow for 20 secondsinfo
Module sensor Inop > XXAirway module hardware failure detectedAirway module reports RAM failure, checksum error, EEPROM error, voltage error, or lamp control failurewarning
Module Source Dlite offAirway module peak pressure significantly below ventilator readingPeak pressure measured by module < 3 cmH2O and >= 5 cmH2O below ventilator peakcritical
Nebulizer not ConnectedNebulizer disconnected while nebulization procedure runningNebulizer connection lost during active nebulizationinfo
Negative PressureAirway pressure during inspiration <–10 cmH₂O for >50 continuous msInspiratory airway pressure drops below –10 cmH₂O for >50 mscritical
Neo – Exh Flow Sensor ComparisonDifference between exhalation and neonatal expired tidal volume exceeds thresholdNeo: (Exh TVexp–Neo TVexp) > (0.3*Neo TVexp) or 30 ml for 6 breathscritical
Neo Flow Sensor CalNeonatal flow sensor calibration in progressDuring neonatal flow sensor calibrationinfo
Neo Flow Sensor Cal FailNeonatal flow sensor calibration failedDuring neonatal flow sensor calibrationerror
Neo Flow Sensor Cal PassNeonatal flow sensor calibration passedAfter neonatal flow sensor calibrationinfo
Neo Flow Sensor ContaminatedNeonatal flow sensor reporting flow >32 l/min for 4 continuous secondsNeo flow sensor sustained reading exceeds 32 l/min for ≥4 secondserror
Neo Flow Sensor FailureOne hot wire reporting less than 20 ADC, other reporting normalNeo flow sensor with one hot wire reporting less than 20 ADC and other hot wire reporting normal as measured by VCB or VMBerror
Neo module not compatibleIncompatible module installed in neo patient modeAny airway module installed in neo patient type modewarning
Neonatal Check Failed (20 Lpm)System unable to achieve 20 Lpm of flow during neonatal checkoutDuring neonatal flow test at 20 Lpmerror
Neonatal Check Failed (Downstream)Expiratory flow from neonatal sensor differs from total flow sensor by more than 25%During neonatal flow sensor comparisonerror
Neonatal Check Failed (Reaching 30)System unable to achieve 30 Lpm of flow during neonatal checkoutDuring neonatal flow test at 30 Lpmerror
Neonatal Check Failed (Reversed @20)Neonatal flow sensor indicated reversed flow during 20 Lpm checkDuring neonatal flow test at 20 Lpmerror
Neonatal Check Failed (Reversed @ 30)Neonatal flow sensor indicated reversed flow during 30 Lpm checkDuring neonatal flow test at 30 Lpmerror
Neonatal Fail FailedNeonatal flow sensor check failed because no gas supply detectedDuring neonatal flow sensor testerror
Neonatal Flow CheckNeonatal flow sensor flow check is in progressDuring neonatal flow sensor testinfo
Neonatal Flow Comparison - SystemNeo flow sensor comparison alarm active or neo-total flow sensor comparison alarm active or neoexh flow sensor comparison alarm active.Flow sensor comparison fault detectederror
Neonatal Flow Sensor ComparisonNeonatal expired tidal volume differs excessively from inspired tidal volumeNeo: (TVexp–TVinsp) > (0.3*TVinsp) or 5 ml for 6 consecutive breathscritical
Neonatal Flow Sensor OffPatient Type is Neonatal but Data Source not set to Neonatal Flow SensorPatient Type is Neonatal, but Data Source is not set to Neonatal Flow Sensorerror
Neonatal Flow Sensor ReversedNeonatal flow sensor not installed correctlyThe neonatal flow sensor is not installed correctlyerror
Neonatal ZeroNeonatal flow sensor zero check is in progressDuring neonatal sensor zero calibrationinfo
Neonatal Zero FailedNeonatal flow sensor zero check failed; no gas supply detectedDuring neonatal sensor zero calibrationerror
Neonatal Zero Failed (4 Lpm)Neonatal flow sensor zero check failed; unable to achieve 4 LpmDuring neonatal sensor zero calibrationerror
Neonatal Zero Failed (Expected 0 Flow)Neonatal flow sensor zero check failed; measured non-zero flow while occludedDuring neonatal sensor zero calibrationerror
Neo – Total Flow Sensor ComparisonNeonatal expired minus inspired tidal volume exceeds threshold vs total flow sensorNeo: (TVexp–TVinsp) > (0.3*TVinsp) or 30 ml for 6 consecutive breathscritical
NFS Checkout CompleteNeonatal flow sensor flow check completedAfter neonatal flow sensor testinfo
NFS reports < 0 flow during > 0 flow (20 l/min)Neonatal flow sensor reports negative flow during positive flow testOpen circuit flow test at 20 l/min targeterror
NFS reports < 0 flow during > 0 flow (30 l/min)Neonatal flow sensor reports negative flow during positive flow testOpen circuit flow test at 30 l/min targeterror
NFS reports nonzero flow during 0 flowNeonatal flow sensor reports flow when circuit is occluded and should read zeroCircuit occlusion test at 4 l/min targeterror
No AirSupply pressure for Air is below 30 psiDuring system operationwarning
No bootable device availableSystem cannot find a bootable deviceStartup (POST) when internal CF card not accessible or corruptedcritical
No Exhal Flow SensorNo Exhalation Flow Sensor connectedExhalation flow sensor not detected at startup or during operationerror
No Exp Flow SensorExpiratory flow sensor is not detected by systemDuring system initialization or testerror
No Gas SupplySupply pressure for both Air and O2 is below 30 psiDuring system operationcritical
NoneNo calibrations, zero measurements, or checkout tests are runningDuring normal operationinfo
No Neonatal Flow Sensor - StandbyBoth neonatal flow sensor hot wires reporting less than 20 ADCBoth neonatal flow sensor hot wires reporting less than 20 ADC as measured by VCB or VMBwarning
No Neonatal Flow Sensor - TherapyBoth neonatal flow sensor hot wires reporting less than 20 ADCBoth neonatal flow sensor hot wires reporting less than 20 ADC as measured by VCB or VMB during therapyerror
No OxygenSupply pressure for O2 is below 30 psiDuring system operationwarning
Normalized Flow Comparison Failed: >Target = Value1 >I = ValNormalized expiratory flow and total flow differ by >10% (or >200 ml)During flow sensor normalization checkerror
No Supply PressureSupply pressure for both Air and O2 is below 30 psiDuring system operationcritical
No Supply PressureO2 and air supply pressure below minimum thresholdO2 and air supply pressure < 24.3 psi for > 0.5 secondscritical
O2 Flow Sensor Communications FailureLoss of communication with O2 Flow SensorSerial/communication link lost between vent engine and O2 flow sensorerror
O2 Flush Time: Value secTime needed to flush system with O2 until stability of 0.1% is reachedDuring O2 sensor calibrationinfo
O2 inaccurateO2 values from the module may be inaccurateO2 values from the module detected as inaccuratewarning
O2 Sensor FailOut of range O2 sensor data (>103%) or communication errorO2 sensor reading exceeds 103% or communication loss with sensorerror
O2 Sensor FailFailed Paracube communications during O2 sensor checkDuring O2 sensor checkouterror
O2 Sensor: Failed 100% CalParacube O2 sensor calibration timed out at 100% oxygenParacube calibration did not complete within timeout period during 100% O2 span calibrationerror
O2 Sensor: Failed 21% CalParacube O2 sensor calibration timed out at 21% oxygenParacube calibration did not complete within timeout period during 21% O2 span calibrationerror
O2 Supply Pres HighO2 supply pressure greater than 95 psi for more than 0.5 secondsO2 supply pressure > 95 psi for more than 0.5 secondswarning
O2 Supply Pres LowO2 supply pressure less than 24.3 psi for more than 0.5 secondsO2 supply pressure < 24.3 psi for more than 0.5 secondserror
O2 Supply Pressure Sensor Out of RangeOut of range O2 supply pressure sensor dataO2 supply pressure sensor data outside range of 195 to 4760 mVwarning
O2 Temperature Sensor FailureOut of range O2 temperature sensor data (0–60°C range)O2 flow temperature reading outside 0–60°C or communication lossinfo
O2 Too Low for 100% CalMeasured O2 concentration below 41% during 100% calibrationO2 measurement <41% during 100% span calibration attemptwarning
O2 Too Low for 21% CalMeasured O2 concentration below 8% during 21% calibrationO2 measurement <8% during 21% span calibration attemptwarning
O2 Valve Cal CompleteO2 flow control valve calibration passed successfullyO2 Valve Cal completed without errorsinfo
O2 Valve Cal FailO2 flow control valve calibration failedO2 Valve Cal did not complete successfullyerror
Occluded Patient Interface DetectedPressure detected at the patient inlet portPressure detected at the patient inlet porterror
On Battery – Low or MediumMains power not available; unit running on internal batteryAC mains power disconnected or failedwarning
Oxygen Flow Sensor Chk CompleteOxygen flow sensor check passed successfullyOxygen flow sensor check completed without errorsinfo
Oxygen Flow Sensor Chk FailOxygen flow sensor check failedOxygen flow sensor check did not complete successfullyerror
Oxygen Flow Sensor Chk Fail: No O2Oxygen flow sensor check failed due to no oxygen supply detectedSystem unable to detect oxygen supply during sensor checkerror
Oxygen Flow Sensor FailUnable to achieve 30 Lpm during oxygen flow sensor checkMaximum oxygen flow during check remains below 30 Lpmerror
Paracube Cal CompleteO2 sensor calibration passed successfullyParacube calibration completed without errorsinfo
Paracube Cal FailO2 sensor calibration failedParacube calibration did not complete successfullyerror
Paracube Chk CompleteParacube O2 sensor check passed successfullyParacube sensor check completed without errorsinfo
Paracube Chk FailParacube O2 sensor check failedParacube check did not complete successfullyerror
Paracube Chk Fail (50/50 Air)Failed to achieve 15 Lpm airflow for 50/50 calibrationAirflow remained below 15 Lpm during 50/50 air calibration steperror
Paracube Chk Fail (Air)Paracube 21% air calibration did not completeParacube failed to complete 21% air calibrationerror
Paracube Chk Fail (O2)Failed to achieve 30 Lpm O2 flow for O2 calibrationO2 flow remained below 30 Lpm during oxygen calibration steperror
Paracube FailFailed to achieve 30 Lpm air flow during Paracube checkAirflow remained below 30 Lpm during checkerror
Paracube Sensor Fail (50/50 O2)Failed to achieve 15 Lpm O2 flow for 50/50 calibrationO2 flow remained below 15 Lpm during 50/50 oxygen calibration steperror
Paux HighPaux pressure greater than 100 cmH2OPaux > 100 cmH2Oerror
Paux Sensor Out Of RangePaux sensor data out of rangePaux sensor data outside range of 10-4095 mVwarning
Paux Zeroing ErrorPaux zeroing failed; zero pressure range 631-968 mVPaux zeroing failed; measurement outside range 631-968 mVwarning
Paw Cross CheckDifference between expiratory and inspiratory airway pressure excessiveDifference > 10 cmH2O for > 350 mscritical
Paw Exp Sensor Out of RangeExpiratory airway pressure sensor data outside acceptable rangeSensor data outside 10-4095 mV rangecritical
Paw Exp Zeroing ErrorExpiratory pressure sensor zero offset out of range during calibrationZero pressure reading outside 631-968 mV rangeinfo
Paw Insp Sensor Out of RangeInspiratory airway pressure sensor data outside acceptable rangeSensor data >= 50 ms or inspiratory zero valve energized >= 1 second; range 10-4095 mVcritical
Paw Insp Zeroing ErrorInspiratory pressure sensor zero offset out of range during calibrationZero pressure reading outside 631-968 mV rangeinfo
Paw Xducer Check CompleteAirway pressure transducer check passed successfullyPressure sensor check completed without errorsinfo
Paw Xducer Check FailAirway pressure transducer check failedPressure sensor check did not complete successfullyerror
Paw Xducer Check Fail (PressureZeroCal)Pressure sensors failed to zero during calibrationZero calibration of pressure sensors did not complete successfullyerror
Paw Xducer OKAirway pressure transducer check passed successfullyPressure sensor check completed successfullyinfo
PMB DC-DC FailInternal power failure; AC supply OK but batteries being dischargedDC-DC converter on Power Management Board failureerror
PMB Self Test FailPower controller failed self tests (memory, voltages, or CPU)Power Management Board internal test failureerror
PMB to DU Com ErrorCommunication failure between power controller and display unitCommunications cannot be established for 10 secondswarning
Power Supply FailMain power supply is not available; ventilator powered by batteryMain power supply not availableerror
Power Supply Temp HighVicor power supply temperature exceeds 75°CPower supply internal temperature rises above 75°Cerror
Pressure not maintained: ValueSystem unable to maintain minimum 25 cmH2O pressure during relief valve checkSystem pressure fell below 25 cmH2O during pressure maintenance testerror
Pressure too lowSystem pressure insufficient during resistance measurementInspiratory limb pressure drops below required threshold during resistance checkerror
Processor Reset FailedSequential watchdogs caused processor reset on VCB or VMBWatchdog resets indicate loss of processor controlerror
Program load failed - CRCSoftware file CRC check failed during loadStartup (POST) when software CRC verification failserror
PSM Power FailureExhalation valve heater plugged into incorrect portExhalation valve heater connected to port 2error
RAM memory errorRAM memory hardware failure detectedStartup (POST) when RAM test failscritical
Relief Valve Chk CompleteRelief valve check passed successfullyRelief valve check completed without errorsinfo
Relief Valve Chk FailRelief valve check failedRelief valve check did not complete successfully or relief valve out of specificationerror
Rising Flow Not Achieved: ValueCould not achieve minimum required flow for checkout stepFlow remained below minimum during checkout step, Value = maximum flow achieved in ml/minerror
RTC date/time errorReal-time clock error detected; battery may be weakStartup (POST) when RTC error detectederror
Safety Valve Checkout FailureFailed Safety Relief Valve System CheckPaw reaches 30 cmH2O within 4 seconds of check startcritical
SBT CompletedSpontaneous Breathing Trial time remaining expired and completed successfullySBT countdown time expiresinfo
SBT ended - ApneaSpontaneous Breathing Trial ended due to apneaSBT countdown time reaches 2 minuteswarning
SBT ended - MVexp HighSpontaneous Breathing Trial ended due to high minute volume expirationMVexp exceeds threshold during SBTerror
SBT ended - MVexp LowSpontaneous Breathing Trial ended due to low minute volume expirationMVexp below threshold during SBTwarning
SBT ended - RR HighSpontaneous Breathing Trial ended due to high respiratory rateRespiratory rate exceeds threshold during SBTwarning
SBT ended - RR LowSpontaneous Breathing Trial ended due to low respiratory rateRespiratory rate below threshold during SBTwarning
Sensor Disagree: ValueAir or O2 flow sensor differs from total flow sensor by >20% or >1500 ml/minFlow sensor difference exceeds 20% of total flow or 1500 ml/min, whichever is greatererror
Sensor RemovedExhalation flow sensor was removed or disconnected during checkoutExhalation flow sensor not detected during checkout procedureerror
Single Limb Relief Valve check CompleteSingle Limb Relief Valve check has finished executionRelief valve check completed all stepsinfo
Single Limb Relief Valve ChkSingle Limb Relief Valve check in progressRelief valve calibration check initiated and runninginfo
Single Limb Relief Valve Chk FailSingle Limb Relief Valve check did not passRelief valve check failed to meet acceptance criteriaerror
Single Limb Relief Valve Time: ValueRelief valve response time measurement during checkTime measured to relieve 0.75 cmH2O pressure within 250 mswarning
Skipping CHECKOUT - not in checkout stateCheckout command received while not in Checkout stateCheckout start command sent to Vent Control Board not in Checkout statewarning
Skipping Dual Limb CHECKOUT - not in checkout stateDual Limb Checkout command rejected; VCB not in checkout stateDual Limb Checkout command received while VCB not in checkout statewarning
Skipping Single Limb CHECKOUT - not in checkout stateSingle Limb Checkout command rejected; VCB not in checkout stateSingle Limb Checkout command received while VCB not in checkout statewarning
Standby Bulk Charge > 12HrsSystem in Standby with bulk charging duration exceeding 12 hoursBulk charge mode active for more than 12 hours while in Standbywarning
System reset: ECxx xx xxUnexpected system reset; usually indicates software failureStartup (POST) after system reset eventerror
System reset: FFFF FF FFSystem reset due to Carrier board contact loss or low battery voltageStartup (POST) after system reset with FFFF FF FF signatureerror
System shutdown in less than X minutesSystem will shut down if batteries not fully charged when AC power lostAC power loss with batteries below full chargecritical
Target flow not reached @ 300Flow exceeds target at minimum flow command during checkoutFlow greater than target despite minimum flow commanderror
Temp HighTotal flow sensor temperature exceeds safe limitTotal flow sensor temperature > 51°Celsiuscritical
Therapy Power OffPower switch is set to OffPower switch in Off positioncritical
Total Flow Sensor Communications FailureOut of range Total Flow Sensor data or loss of communicationTotal flow sensor reading out of range or communication link lostcritical
Total Flow Temperature Sensor FailureOut of range total temperature sensor data or loss of communicationTemperature reading outside 0–60°C or communication loss from total flow sensorinfo
Total flow too high (1)Flow measurement during circuit occlusion exceeds acceptable rangeCircuit occlusion test reads flow higher than expected at 4 l/minerror
Total flow too high (2)Flow measurement during open circuit test exceeds acceptable rangeOpen circuit flow measurement at 20 l/min or 30 l/min reads higher than expectederror
Total flow too lowSystem cannot establish required flow for resistance measurementInspiratory limb resistance check cannot achieve 60 l/min flowerror
Total flow too low (1)Flow measurement during circuit occlusion is insufficientCircuit occlusion test at 4 l/min target fails to achieve zero flow readingerror
Total flow too low (2)Flow measurement during open circuit test is insufficientOpen circuit flow measurement at 20 l/min or 30 l/min fails to reach targeterror
Total/Neo Exp flows disagreeTotal and neonatal expiratory flow sensor readings do not correlateFlow comparison during circuit open state shows >25% deviationerror
Total/Neo Insp flows disagreeTotal and neonatal inspiratory flow sensor readings do not correlateFlow comparison during circuit open state shows >15% deviationerror
Total resistance invalidResistance check failed and invalid result returnedCalculation errors during resistance measurement (negative values or divide by zero)error
Total: Value1 O2: Value2O2 flow sensor differs >15% from total flow sensor at 30 L/min nominalO2 flow measurement differs more than 15% from total flow sensor at 30 L/minerror
Touchscreen FailureDisplay unit touchscreen malfunctionTouchscreen input failure detectederror
Unable to pressurizeCircuit could not be pressurized to minimum 10 cmH2OSystem unable to generate 10 cmH2O pressure in breathing circuiterror
Upstream Flow Failed (Upstream)Neonatal flow sensor inspiratory flow differs >15% from total flow sensorNeonatal flow sensor inspiratory measurement differs >15% from total flow sensor readingerror
VCB: Expansion Power FailExhalation valve heater or cable failureExhalation valve heater or cable failure, or external device connected to rear portserror
VCB to DU Com Error FailedCommunication from Ventilation Control Board to Display Unit failedCommunications not completed within 10 second window (timing starts after 90 sec power-up)error
VMB Exp Flow Sensor Cal FailExpiratory flow sensor calibration failedExpiratory flow sensor calibration did not complete successfullyerror
VMB Exp Flow Sensor Cal PassExpiratory flow sensor calibration passed successfullyExpiratory flow sensor calibration completed without errorsinfo
VMB to DU Com Error FailedCommunication from Ventilation Management Board to Display Unit failedCommunications not completed within 10 second window (timing starts after 90 sec power-up)error
VMB to VCB Com Error FailedHigh-speed sensor data communication from VMB to VCB failedCommunication not completed within 50 millisecond windowerror
Volume Delivery ErrorVentilator is not delivering flowVCB detects flow delivery failure or blockagecritical
VT not achievedTidal volume not delivered to patientTVinsp < 80% of set TV, or set TV - 5 ml (whichever greater) for six consecutive breathswarning
VT not achieved - NeoTidal volume not delivered to neonatal patientTVinsp < 80% of set TV, or set TV - 1 ml (whichever greater) for six consecutive breathswarning
watchdog circuit failedWatchdog circuit hardware failureStartup (POST) when watchdog circuit test failscritical
Xducer Error: ValueInspiratory/expiratory pressure difference exceeds 4 cmH2O during Paw Xducer checkPressure difference between inspiratory and expiratory sensors greater than 4 cmH2Oerror
Zero out of rangeADC counts out of acceptable zero pressure range for pressure sensorsInspiratory and/or expiratory sensor ADC counts outside range 631-968error

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.