GE Healthcare CARESCAPE PDM: troubleshooting and error codes

The ServiceForge knowledge base holds 100 documented symptoms for the GE Healthcare CARESCAPE PDM, with 136 ranked causes, 168 diagnostic checks with test points and expected values, and 42 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Abnormally low invasive pressure reading patient safety

Invasive pressure measurement displays lower values than expected, below physiological range

Ranked causes

  1. 1. Transducer wrongly positioned (common)
  2. 2. Defective transducer (occasional)
  3. 3. Module not connected (occasional)
  4. 4. Transducer adapter cable not connected (occasional)
  5. 5. Invasive pressure channel not configured or low priority (occasional)
First check

Verify transducer position

Test point
Patient transducer placement
Expected
Transducer at mid-heart level

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Communication loss with NIBP module

ePort interface not communicating with monitor; module not powering or responding

Ranked causes

  1. 1. ePort cable damaged (occasional)
  2. 2. ePort connector on monitor damaged (occasional)
  3. 3. ePort connector on module damaged (occasional)
  4. 4. Module main acquisition board defective (rare)
First check

Inspect ePort to host interface cable

Test point
ePort cable connectors and insulation
Expected
Cable intact, no visible damage, connectors secure

5 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

No power at the module

Module does not receive power when connected to monitor or docking station

Ranked causes

  1. 1. ePort host interface cable is damaged (common)
  2. 2. ePort connector on monitor is damaged (common)
  3. 3. ePort connector on module is damaged (occasional)
  4. 4. Module battery is not charging (occasional)
First check

Inspect ePort host interface cable for damage

Test point
ePort to host interface cable
Expected
No visible damage or breaks

3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No temperature displayed patient safety

Temperature measurement channel shows no value or missing from display completely

Ranked causes

  1. 1. Temperature channel not configured or low priority (common)
  2. 2. Incompatible temperature probe (occasional)
  3. 3. Faulty temperature probe (occasional)
  4. 4. Temperature out of measurable range (rare)
First check

Verify temperature channel configuration

Test point
PDM channel configuration screen
Expected
Temperature channel configured with adequate priority and active

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery not charging patient safety

Module battery does not charge or charge level remains low

Ranked causes

  1. 1. Battery tray damaged (rare)
  2. 2. Battery connector damaged (occasional)
  3. 3. Battery defective (common)
First check

Verify module is powered by host monitor

Test point
Module power connection
Expected
Module powered by host monitor ePort connection

1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No parameter data from module patient safety

Parameter data is not being received or displayed from the connected parameter module

Ranked causes

  1. 1. Parameter cable disconnected or loose (common)
  2. 2. Parameter interface damaged (occasional)
  3. 3. Related parameter license not activated (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Noisy waveforms or missing markers in parameter data patient safety

Parameter waveforms contain excessive noise, markers are missing, or patient data shows inaccuracy on display

Ranked causes

  1. 1. ECG filter disabled (common)
  2. 2. Parameter cable or interface damaged (occasional)
  3. 3. PDM and host monitor power frequency mismatch (occasional)
First check

Verify ECG filter enabled setting

Test point
PDM settings menu
Expected
ECG Filter setting = Enabled

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

PDM battery cannot be charged caution

Battery charging is disabled due to temperature or power condition

Ranked causes

  1. 1. Internal monitor temperature too high (occasional)
  2. 2. Power fault preventing charge (occasional)
  3. 3. Battery temperature exceeds limit (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Artifact detected

Artifact signal detected at measurement sensor site during SpO2 measurement.

Ranked causes

  1. 1. Poor sensor contact or placement (common)
  2. 2. Patient movement or shivering (common)
First check

Inspect SpO2 sensor contact with patient skin

Test point
Sensor site
Expected
Firm contact, clean, properly positioned

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Artifact on ECG waveform

Muscle artifact, high or low frequency noise, or pacemaker artifact visible on waveform display.

Ranked causes

  1. 1. Poor electrode contact or placement (common)
  2. 2. Electrical interference from nearby devices (occasional)
First check

Inspect electrode contact and lead placement

Test point
All ECG electrodes and leadwires
Expected
Secure connections, proper lead routing

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration fail patient safety

Calibration of a temperature or C.O. measurement channel failed. Message appears in parameter window or alarm area.

Ranked causes

  1. 1. Poor or disconnected connections (common)
  2. 2. Unsuccessful calibration procedure (occasional)
First check

Verify all connections to acquisition module and temperature/C.O. measurement cables

Test point
Module connectors and cable connections
Expected
All connections fully seated and secure

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check SpO2 probe

No detectable SpO2 signal. Sensor may be faulty, detached from patient, or have connectivity issues.

Ranked causes

  1. 1. Sensor not connected or detached from patient (common)
  2. 2. Faulty sensor (occasional)
First check

Check sensor physical connection

Test point
Sensor and acquisition module interface
Expected
Sensor firmly connected to module and patient

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Defib/Sync analog output test failure patient safety

Defibrillator synchronization or analog output signal not detected

Ranked causes

  1. 1. ECG or invasive pressure cable not properly connected for Defib/Sync output (common)
  2. 2. Defib/Sync tester or analog output cable not configured (occasional)
First check

Verify ECG cable connection for Defib/Sync

Test point
Module ECG connector
Expected
5-lead ECG cable securely connected

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG/Impedance respiration test failure patient safety

ECG waveform not displayed or respiration measurement not detected

Ranked causes

  1. 1. ECG cable or leadwire not properly connected (common)
  2. 2. Simulator not configured for ECG/respiration (occasional)
First check

Verify ECG trunk cable connection

Test point
Module ECG connector
Expected
10-lead ECG trunk cable securely seated

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Functional check configuration failure patient safety

Measurement parameters not displaying or simulator not properly configured for testing

Ranked causes

  1. 1. Module or monitor not properly connected (common)
  2. 2. Measurement parameter not configured with adequate priority (occasional)
First check

Verify module is connected to monitor

Test point
Module connection status
Expected
Module connected and powered on

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invasive pressure channel not reading correctly patient safety

Pressure waveform not displayed or reading does not match configured value within tolerance

Ranked causes

  1. 1. Channel not zeroed or simulator not configured to correct pressure (common)
  2. 2. Pressure waveform generation disabled or sensor malfunction (occasional)
First check

Configure simulator to zero static pressure

Test point
Simulator output
Expected
0 mmHg static pressure

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invasive pressure measurement failure patient safety

Invasive pressure waveform not displayed or pressure values not accurate

Ranked causes

  1. 1. Invasive pressure cables not properly connected (common)
  2. 2. Simulator invasive pressure not configured (occasional)
First check

Verify invasive pressure cable connection

Test point
Module invasive pressure connector
Expected
Dual invasive pressure adapter cable securely connected

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Module will not dock to docking station caution

Module cannot be physically docked to the docking station

Ranked causes

  1. 1. Mounting rails on module are damaged (occasional)
  2. 2. Docking station is damaged (occasional)
First check

Inspect module mounting rails for damage

Test point
Module mounting rails
Expected
Rails undamaged and aligned

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP measurement failure patient safety

Non-invasive blood pressure reading not obtained or pressure values incorrect

Ranked causes

  1. 1. NIBP hose or cuff not properly connected or assembled (common)
  2. 2. Manometer not properly configured (occasional)
First check

Verify NIBP hose connection

Test point
Module NIBP connector
Expected
NIBP hose securely connected

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP module not functioning patient safety

Non-invasive blood pressure measurement system not operational

Ranked causes

  1. 1. NIBP system not calibrated (common)
  2. 2. NIBP cuff pressure not maintained; cuff leaking (occasional)
First check

Verify calibration required

Expected
Calibration status indicator shows need for recalibration

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

NIBP tubing system leaking caution

Pressure does not stabilize or drops more than 1 mmHg per 5 seconds during NIBP calibration check

Ranked causes

  1. 1. External NIBP test setup leaking (common)
  2. 2. Internal NIBP tubing leaking (occasional)
First check

Select NIBP calibration from service menu

Test point
Monitor Setup > Defaults & Service > Service Calibrations menu
Expected
Calibrations menu displays

5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 measurement failure patient safety

Oxygen saturation or pulse rate not displayed or showing incorrect values

Ranked causes

  1. 1. SpO2 simulator/cable not properly connected (common)
  2. 2. Simulator SpO2 not configured (occasional)
First check

Verify SpO2 simulator connection

Test point
Module SpO2 connector
Expected
SpO2 simulator and cable securely connected

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature measurement failure patient safety

Temperature readings not displayed or showing incorrect values

Ranked causes

  1. 1. Temperature cables not properly connected (common)
  2. 2. Simulator not configured for 400 series temperature probes (occasional)
First check

Verify temperature adapter cable connection

Test point
Module Temp/CO connector
Expected
Dual temperature adapter cable securely connected

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Analog output and defib sync marker not functioning high voltage

Analog output or defibrillator synchronization marker signals not present or incorrect

Ranked causes

  1. 1. Analog output/defib sync cable disconnected or improperly connected (common)
First check

Connect analog output/defib sync cable to module

Test point
Defib/Sync connector
Expected
Cable seated and connected

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Apnea alarm not activated patient safety

Monitor does not trigger apnea alarm when respiration rate is zero

Ranked causes

  1. 1. Simulator not configured for apnea simulation or alarm threshold issue (common)
First check

Verify apnea alarm activation

Test point
Monitor screen and alarm
Expected
Apnea alarm activated when respiration stops

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Arrhythmia paused patient safety

ECG channels unavailable for analysis due to excessive artifact (20 seconds), internal HR calculation not updated (30 seconds), or single lead analysis ECG 1 inactive (5 seconds). Appears in alarm area or waveform area.

Ranked causes

  1. 1. Excessive artifact on ECG signal (common)
First check

Inspect electrode contact quality

Test point
All ECG electrode sites
Expected
Firm contact, no visible gaps or debris

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Asystole alarm not triggered patient safety

Monitor does not display asystole alarm when asystole is configured on simulator

Ranked causes

  1. 1. Simulator not configured to asystole output (common)
First check

Verify asystole alarm appears

Test point
Monitor display
Expected
Asystole alarm shown on screen

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bottom housing is damaged caution

Cracks or other damage visible on bottom housing

Ranked causes

  1. 1. Bottom housing cracked or damaged (common)
First check

Inspect bottom housing for cracks or damage

Test point
Bottom housing surface
Expected
Housing intact without cracks

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

C.O. measurement removed patient safety

Cardiac output measurement module has been disconnected. Message appears in alarm area.

Ranked causes

  1. 1. C.O. measurement module disconnected (common)
First check

Verify C.O. measurement module is physically connected

Test point
C.O. module connectors
Expected
Module fully seated and connected

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration error

Calibration error message appears in parameter window when calibration procedure fails

Ranked causes

  1. 1. Calibration procedure failed (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Call service message displayed

Call service message appears in parameter window indicating a technical fault

Ranked causes

  1. 1. Technical fault detected (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Channel calibration in progress

Calibrating message appears in parameter window during active calibration of a channel

Ranked causes

  1. 1. Calibration procedure active (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check device malfunction

SpO2 acquisition module has failed and is unable to process signal.

Ranked causes

  1. 1. Acquisition module failure (occasional)
First check

Perform planned maintenance on acquisition module

Test point
Acquisition module
Expected
Module functions normally after maintenance

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check NIBP message displayed

Check NIBP message appears in alarm area when systolic, diastolic, and/or mean results are missing

Ranked causes

  1. 1. Systolic, diastolic, and/or mean results missing (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Control measurement pressure alarm limit exceeded

Control measurement message appears in parameter window when pressure alarm limit is exceeded during measurement

Ranked causes

  1. 1. Pressure alarm limit exceeded during measurement (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cuff loose

Cuff loose message appears in parameter window when cuff connection is loose or cuff hose is disconnected

Ranked causes

  1. 1. Loose cuff or cuff hose (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cuff occlusion detected patient safety

Cuff occlusion message appears in parameter window when occlusion occurs during measurement or cuff is overpressured

Ranked causes

  1. 1. Occlusion during measurement or overpressured cuff (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cuff overpressure detected patient safety

Cuff overpressure message appears in parameter window when NIBP cuff is squeezed during measurement and exceeds pressure safety limits

Ranked causes

  1. 1. NIBP cuff squeezed during measurement exceeding pressure safety limits (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cuff size not selected

Cuff size has not been selected in NIBP setup

Ranked causes

  1. 1. Cuff size not selected from setup menu (common)
First check

Verify cuff size selection

Test point
NIBP setup menu
Expected
Cuff size selected (e.g., adult, child, infant)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Device malfunction or performance issue patient safety

General troubleshooting guidance for technical problems; patient disconnected from monitor before service

Ranked causes

  1. 1. Visual inspection incomplete or findings missed (common)
First check

Visual inspection of module exterior

Test point
Front cover, connectors, casing, latch, patient cables
Expected
Front cover intact, connectors intact and properly attached, casing and latch clean and intact, cables clean and intact

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG measurement not displaying correctly patient safety

Monitor does not display ECG leads II, VI & aVL, or waveforms contain noise; heart rate not 80±5 bpm or QRS tone absent

Ranked causes

  1. 1. Simulator not configured for normal sinus rhythm (common)
First check

Verify ECG leads II, VI & aVL displayed and noise-free

Test point
Monitor display
Expected
Leads II, VI & aVL visible, no noise

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Electrical safety test failure patient safety

Patient leakage current exceeds acceptable limits during source or sink testing

Ranked causes

  1. 1. Electrical isolation or patient circuit defect (unknown)
First check

Patient (source) leakage current test

Test point
Patient circuit source
Expected
Within acceptable limits per service manual

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Faulty SpO2 probe

The SpO2 sensor has failed and is no longer functional.

Ranked causes

  1. 1. Sensor failure (occasional)
First check

Attempt sensor reactivation or test with another sensor

Test point
SpO2 sensor
Expected
SpO2 signal returns or test sensor functions

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incompatible SpO2 probe

The sensor connected is not compatible with the PDM system.

Ranked causes

  1. 1. Incompatible sensor type installed (occasional)
First check

Verify sensor compatibility with PDM system

Test point
Sensor model and type
Expected
Sensor is listed as compatible with CARESCAPE PDM

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Lead off - single or multiple leads disconnected patient safety

One or more leadwires or electrodes disconnected. Message appears in alarm area and waveform area with specific lead designation (e.g., LA/L, LL/F, RA/R, RL/N, V2/C2-V6/C6, Va/Ca, Vb/Cb). Other ECG leads available for arrhythmia detection.

Ranked causes

  1. 1. Leadwire or electrode connection loose or disconnected (common)
First check

Verify connection of identified lead

Test point
Specific lead shown in alarm message (LA, LL, RA, RL, V2-V6, Va, Vb, etc.)
Expected
Lead connected and signal present

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Lead off detection not working patient safety

Detaching leadwire does not trigger lead off message or automatic lead switching

Ranked causes

  1. 1. Leadwire connection issue or monitor lead detection malfunction (common)
First check

Verify Lead II disappears when RA/R detached

Test point
ECG1 waveform field
Expected
Lead II disappears, RA/R lead off message shown

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Leads off - multiple leads disconnected, arrhythmia detection impossible patient safety

One or more connected electrodes disconnected and arrhythmia detection is not possible. Appears in alarm area and waveform area.

Ranked causes

  1. 1. Multiple electrode/leadwire disconnections preventing arrhythmia analysis (common)
First check

Inspect all leadwire and electrode connections

Test point
All ECG lead connection points
Expected
All leads connected with secure seating and good skin contact

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Long measurement time

NIBP measurement takes longer than expected: >2 min for adult/child, 85 s for infant

Ranked causes

  1. 1. Patient status issue or equipment malfunction (unknown)
First check

Verify measurement time within specification

Test point
NIBP measurement display
Expected
>2 min for adult/child, 85 s for infant

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low perfusion at measurement point patient safety

Blood perfusion at the sensor measurement site is insufficient for reliable signal detection.

Ranked causes

  1. 1. Poor perfusion at sensor site (common)
First check

Assess blood perfusion at current sensor site

Test point
Measurement site
Expected
Palpable pulse, warm skin, good color

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low signal quality

Signal quality is questionable or poor. Patient may be shivering or sensor placement is suboptimal.

Ranked causes

  1. 1. Poor sensor placement or patient shivering (common)
First check

Evaluate sensor placement on patient

Test point
Sensor site
Expected
Proper contact, minimal motion artifact, adequate perfusion

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Module power or LED not illuminated

Communication LED on module not lighting when connected to monitor

Ranked causes

  1. 1. Module defective (common)
First check

Connect known good module to monitor

Expected
Known good module LED lights

NIBP alarm limits changed patient safety

Alarm limits have changed due to cuff size or profile change

Ranked causes

  1. 1. Cuff size or profile changed (common)
First check

Confirm cuff size is correct

Test point
NIBP setup menu
Expected
Cuff size matches patient and clinical order

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP Auto mode stopped patient safety

NIBP Auto mode has stopped due to loose cuff, loose hose, or cuff size changed to not selected

Ranked causes

  1. 1. Loose cuff or cuff hose (common)
First check

Inspect cuff and hose connections

Test point
Cuff connector and hose attachment points
Expected
Firm connection with no visible damage or kinks

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP calibration failed or gain correction exceeded

Unable to complete NIBP calibration; Confirm button inactive or gain correction exceeds +/-10%

Ranked causes

  1. 1. NIBP main board components not functioning properly (occasional)
First check

Note gain correction value required

Expected
Gain correction within +/-10% of factory calibration

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

NIBP cuff loose patient safety

Loose cuff or cuff hose detected

Ranked causes

  1. 1. Loose cuff or cuff hose connection (common)
First check

Verify cuff and hose are secure

Test point
Cuff connector and hose attachment
Expected
Firm, tight connection with no movement

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP cuff occlusion during measurement patient safety

Occlusion detected during measurement or cuff overpressure condition

Ranked causes

  1. 1. Cuff occlusion or overpressure (common)
First check

Verify cuff condition and placement

Test point
Cuff on patient limb
Expected
Cuff properly placed, not kinked, pressure appropriate

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP measurement functional check failed caution

NIBP tubing system has leaks or module is out of calibration; pressure readings inaccurate or unstable

Ranked causes

  1. 1. Combined NIBP tubing leak and/or calibration issue (common)
First check

Check NIBP tubing system and calibration status

Test point
Full NIBP calibration check procedure per steps 1-7
Expected
Pass status: pressure stable ±1 mmHg, no >1 mmHg drop per 5 sec

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP out of calibration caution

NIBP readings differ by more than ±1 mmHg between manometer and calibration menu

Ranked causes

  1. 1. NIBP module requires recalibration (common)
First check

Compare manometer and calibration menu readings

Test point
Manometer display vs. Calibration Check menu
Expected
Readings match ±1 mmHg, ±1 mmHg

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP pressure value outside alarm limits patient safety

NIBP systolic, diastolic, or mean pressure equal to or outside set alarm limits

Ranked causes

  1. 1. Patient hemodynamic change (common)
First check

Review measured pressure values

Test point
NIBP display (systolic, diastolic, mean)
Expected
Values within set alarm limits

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP STAT mode stopped patient safety

STAT mode has stopped due to loose cuff, loose hose, or cuff size changed to not selected

Ranked causes

  1. 1. Loose cuff or cuff hose (common)
First check

Inspect cuff and hose connections

Test point
Cuff connector and hose attachment points
Expected
Firm connection with no visible damage

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • NIBP zeroing in progress
  • NIBP zeroing successful
  • No pressure reading displayed
  • No SpO2 probe connected
  • Pacemaker spikes not detected
  • Parameter interface module damaged
  • Parameter not displayed on host monitor
  • Planned maintenance procedure
  • PR(SpO2) high or low
  • Probe off
  • Pulse search in progress
  • Respiration rate incorrect
  • Service the PDM - Voltage rail errors
  • Service the PDM Calibrate ECG Analog Out
  • Service the PDM Calibrate IP Analog Out
  • Service the PDM Calibrate NIBP
  • Service the PDM Configure Power Frequency
  • Service the PDM Error Code 0xPDM1000
  • Service the PDM Error Code 0xPDM1001
  • Service the PDM Error Code 0xPDM1002
  • Service the PDM Error Code 0xPDM1003
  • Service the PDM Error Code 0xPDM1004
  • Service the PDM Error Code 0xPDM1005
  • Service the PDM Error Code 0xPDM1006
  • Service the PDM Error Code 0xPDM100C
  • Service the PDM Error Code 0xPDM1100
  • Service the PDM Error Code 0xPDM1101
  • Service the PDM Error Code 0xPDM1102
  • Service the PDM Error Code 0xPDM1103
  • Service the PDM Error Code 0xPDM1104
  • Service the PDM Error Code 0xPDM1105
  • Service the PDM Install License
  • SpO2 high or low
  • SpO2 measurement not displaying
  • SpO2 measurement removed
  • Temperature measurement removed
  • Temperature reading incorrect
  • Top housing is damaged
  • Visual inspection failure
  • Wrong cable connected

40 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
0xPDM1000Secure storage data is corruptSystem detects corrupted data in secure storage on startup or runtimecritical
0xPDM1001Flash file system partition is fullSystem detects one or more partitions at capacity during operationcritical
0xPDM1002Error writing data to secure storageWrite operation to secure storage fails during normal operationcritical
0xPDM1003Error writing to the SPI busSPI bus write operation fails during normal operationcritical
0xPDM1004Error reading the network interface informationNetwork interface data read fails during startup or operationerror
0xPDM1005Error reading flash file system partition informationFlash partition read operation fails during startup or operationcritical
0xPDM1006Error using the TWI busTWI (Two-Wire Interface) bus operation fails during startup or runtimecritical
0xPDM100CMAC address not readable from hardwareSystem cannot read MAC address from hardware on startuperror
0xPDM1100Main CPU board 1.8V rail voltage out of specificationPower monitoring detects 1.8V rail outside acceptable rangecritical
0xPDM1101Main CPU board 3.3V rail voltage out of specificationPower monitoring detects 3.3V rail outside acceptable rangecritical
0xPDM1102Main CPU board 5.4V rail voltage out of specificationPower monitoring detects 5.4V rail outside acceptable rangecritical
0xPDM1103Main CPU board 12V rail voltage out of specificationPower monitoring detects 12V rail outside acceptable rangecritical
0xPDM1104Main CPU board –12V rail voltage out of specificationPower monitoring detects –12V rail outside acceptable rangecritical
0xPDM1105Main CPU board power bus voltage out of specificationPower monitoring detects main power bus outside acceptable rangecritical
CalibratedChannel calibrated successfullySuccessful completion of channel calibrationinfo
CalibratingCalibration of a channel is in progressCalibration procedure initiatedinfo
Calibration errorCalibration has failedCalibration procedure failed to completeerror
Calibration failTemperature or C.O. measurement channel calibration failedCalibration procedure initiated but failed to complete successfullyerror
Call serviceTechnical faultSystem detected an unspecified technical faultcritical
Check NIBPSystolic, diastolic, and/or mean results missingMeasurement completed without all required valueswarning
CO measurement removedCardiac output measurement module disconnectedC.O. module detected as missing during operationwarning
Control measurementPressure alarm limit exceededPressure during measurement exceeded alarm limitwarning
Cuff looseLoose cuff or cuff hoseLoss of pressure signal or cuff connection failurewarning
Cuff occlusionOcclusion during measurement or overpressured cuffOcclusion detected or cuff pressure exceeds safe limits during measurementwarning
Cuff overpressureNIBP cuff is squeezed during measurement and exceeded the pressure safety limitsCuff pressure exceeded safety limits during measurementwarning
PDM battery lowBattery charge is insufficient, charging failure detectedBattery cannot be charged due to power faultwarning
PDM battery temp highBattery temperature exceeds safe operating limitBattery temperature sensor detects excessive temperaturewarning
PDM charging is deniedBattery charging is disabled due to excessive internal temperatureMonitor internal temperature exceeds safe charging thresholdwarning
Service the PDM Error Code 0xPDM1106ECG DAS board 1.26V rail voltage out of specificationPower supply rail voltage deviation detected on ECG data acquisition boarderror
Service the PDM Error Code 0xPDM1107ECG DAS board 3.3V rail voltage out of specificationPower supply rail voltage deviation detected on ECG data acquisition boarderror
Service the PDM Error Code 0xPDM1108ECG DAS board 5V rail voltage out of specificationPower supply rail voltage deviation detected on ECG data acquisition boarderror
Service the PDM Error Code 0xPDM1109ECG DAS board 7.5V rail voltage out of specificationPower supply rail voltage deviation detected on ECG data acquisition boarderror
Service the PDM Error Code 0xPDM110AECG DAS board negative 7.5V rail voltage out of specificationPower supply rail voltage deviation detected on ECG data acquisition boarderror
Service the PDM Error Code 0xPDM110BMain DAS board 1.26V rail voltage out of specificationPower supply rail voltage deviation detected on main data acquisition boarderror
Service the PDM Error Code 0xPDM110CMain DAS board 3.3V rail voltage out of specificationPower supply rail voltage deviation detected on main data acquisition boarderror
Service the PDM Error Code 0xPDM110DMain DAS board 5V rail voltage out of specificationPower supply rail voltage deviation detected on main data acquisition boarderror
Service the PDM Error Code 0xPDM110EMain DAS board 2.5V rail voltage out of specificationPower supply rail voltage deviation detected on main data acquisition boarderror
Service the PDM Error Code 0xPDM110FMain DAS board Nellcor 5V rail voltage out of specificationPower supply rail voltage deviation detected on main data acquisition board Nellcor channelerror
Service the PDM Error Code 0xPDM1110Main DAS board Nellcor negative 3V rail voltage out of specificationPower supply rail voltage deviation detected on main data acquisition board Nellcor channelerror
Service the PDM Error Code 0xPDM1120Temperature sensor reading out of specificationInternal temperature measurement exceeds acceptable rangeerror
T1 calibration fail / T2 Calibration fail / Tblood calibratiTemperature channel (T1, T2, or Tblood) calibration failedCalibration procedure attempted for specified temperature channelerror
Temp measurement removedTemperature measurement module or cable disconnectedAcquisition module or temperature cable detected as missing during operationwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI