GE Healthcare CARESCAPE PDM: troubleshooting and error codes
The ServiceForge knowledge base holds 100 documented symptoms for the GE Healthcare CARESCAPE PDM, with 136 ranked causes, 168 diagnostic checks with test points and expected values, and 42 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Abnormally low invasive pressure reading patient safety
Invasive pressure measurement displays lower values than expected, below physiological range
Ranked causes
- 1. Transducer wrongly positioned (common)
- 2. Defective transducer (occasional)
- 3. Module not connected (occasional)
- 4. Transducer adapter cable not connected (occasional)
- 5. Invasive pressure channel not configured or low priority (occasional)
Verify transducer position
- Test point
- Patient transducer placement
- Expected
- Transducer at mid-heart level
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication loss with NIBP module
ePort interface not communicating with monitor; module not powering or responding
Ranked causes
- 1. ePort cable damaged (occasional)
- 2. ePort connector on monitor damaged (occasional)
- 3. ePort connector on module damaged (occasional)
- 4. Module main acquisition board defective (rare)
Inspect ePort to host interface cable
- Test point
- ePort cable connectors and insulation
- Expected
- Cable intact, no visible damage, connectors secure
5 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
No power at the module
Module does not receive power when connected to monitor or docking station
Ranked causes
- 1. ePort host interface cable is damaged (common)
- 2. ePort connector on monitor is damaged (common)
- 3. ePort connector on module is damaged (occasional)
- 4. Module battery is not charging (occasional)
Inspect ePort host interface cable for damage
- Test point
- ePort to host interface cable
- Expected
- No visible damage or breaks
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No temperature displayed patient safety
Temperature measurement channel shows no value or missing from display completely
Ranked causes
- 1. Temperature channel not configured or low priority (common)
- 2. Incompatible temperature probe (occasional)
- 3. Faulty temperature probe (occasional)
- 4. Temperature out of measurable range (rare)
Verify temperature channel configuration
- Test point
- PDM channel configuration screen
- Expected
- Temperature channel configured with adequate priority and active
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not charging patient safety
Module battery does not charge or charge level remains low
Ranked causes
- 1. Battery tray damaged (rare)
- 2. Battery connector damaged (occasional)
- 3. Battery defective (common)
Verify module is powered by host monitor
- Test point
- Module power connection
- Expected
- Module powered by host monitor ePort connection
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No parameter data from module patient safety
Parameter data is not being received or displayed from the connected parameter module
Ranked causes
- 1. Parameter cable disconnected or loose (common)
- 2. Parameter interface damaged (occasional)
- 3. Related parameter license not activated (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Noisy waveforms or missing markers in parameter data patient safety
Parameter waveforms contain excessive noise, markers are missing, or patient data shows inaccuracy on display
Ranked causes
- 1. ECG filter disabled (common)
- 2. Parameter cable or interface damaged (occasional)
- 3. PDM and host monitor power frequency mismatch (occasional)
Verify ECG filter enabled setting
- Test point
- PDM settings menu
- Expected
- ECG Filter setting = Enabled
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PDM battery cannot be charged caution
Battery charging is disabled due to temperature or power condition
Ranked causes
- 1. Internal monitor temperature too high (occasional)
- 2. Power fault preventing charge (occasional)
- 3. Battery temperature exceeds limit (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Artifact detected
Artifact signal detected at measurement sensor site during SpO2 measurement.
Ranked causes
- 1. Poor sensor contact or placement (common)
- 2. Patient movement or shivering (common)
Inspect SpO2 sensor contact with patient skin
- Test point
- Sensor site
- Expected
- Firm contact, clean, properly positioned
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Artifact on ECG waveform
Muscle artifact, high or low frequency noise, or pacemaker artifact visible on waveform display.
Ranked causes
- 1. Poor electrode contact or placement (common)
- 2. Electrical interference from nearby devices (occasional)
Inspect electrode contact and lead placement
- Test point
- All ECG electrodes and leadwires
- Expected
- Secure connections, proper lead routing
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration fail patient safety
Calibration of a temperature or C.O. measurement channel failed. Message appears in parameter window or alarm area.
Ranked causes
- 1. Poor or disconnected connections (common)
- 2. Unsuccessful calibration procedure (occasional)
Verify all connections to acquisition module and temperature/C.O. measurement cables
- Test point
- Module connectors and cable connections
- Expected
- All connections fully seated and secure
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check SpO2 probe
No detectable SpO2 signal. Sensor may be faulty, detached from patient, or have connectivity issues.
Ranked causes
- 1. Sensor not connected or detached from patient (common)
- 2. Faulty sensor (occasional)
Check sensor physical connection
- Test point
- Sensor and acquisition module interface
- Expected
- Sensor firmly connected to module and patient
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Defib/Sync analog output test failure patient safety
Defibrillator synchronization or analog output signal not detected
Ranked causes
- 1. ECG or invasive pressure cable not properly connected for Defib/Sync output (common)
- 2. Defib/Sync tester or analog output cable not configured (occasional)
Verify ECG cable connection for Defib/Sync
- Test point
- Module ECG connector
- Expected
- 5-lead ECG cable securely connected
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG/Impedance respiration test failure patient safety
ECG waveform not displayed or respiration measurement not detected
Ranked causes
- 1. ECG cable or leadwire not properly connected (common)
- 2. Simulator not configured for ECG/respiration (occasional)
Verify ECG trunk cable connection
- Test point
- Module ECG connector
- Expected
- 10-lead ECG trunk cable securely seated
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Functional check configuration failure patient safety
Measurement parameters not displaying or simulator not properly configured for testing
Ranked causes
- 1. Module or monitor not properly connected (common)
- 2. Measurement parameter not configured with adequate priority (occasional)
Verify module is connected to monitor
- Test point
- Module connection status
- Expected
- Module connected and powered on
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invasive pressure channel not reading correctly patient safety
Pressure waveform not displayed or reading does not match configured value within tolerance
Ranked causes
- 1. Channel not zeroed or simulator not configured to correct pressure (common)
- 2. Pressure waveform generation disabled or sensor malfunction (occasional)
Configure simulator to zero static pressure
- Test point
- Simulator output
- Expected
- 0 mmHg static pressure
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invasive pressure measurement failure patient safety
Invasive pressure waveform not displayed or pressure values not accurate
Ranked causes
- 1. Invasive pressure cables not properly connected (common)
- 2. Simulator invasive pressure not configured (occasional)
Verify invasive pressure cable connection
- Test point
- Module invasive pressure connector
- Expected
- Dual invasive pressure adapter cable securely connected
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Module will not dock to docking station caution
Module cannot be physically docked to the docking station
Ranked causes
- 1. Mounting rails on module are damaged (occasional)
- 2. Docking station is damaged (occasional)
Inspect module mounting rails for damage
- Test point
- Module mounting rails
- Expected
- Rails undamaged and aligned
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP measurement failure patient safety
Non-invasive blood pressure reading not obtained or pressure values incorrect
Ranked causes
- 1. NIBP hose or cuff not properly connected or assembled (common)
- 2. Manometer not properly configured (occasional)
Verify NIBP hose connection
- Test point
- Module NIBP connector
- Expected
- NIBP hose securely connected
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP module not functioning patient safety
Non-invasive blood pressure measurement system not operational
Ranked causes
- 1. NIBP system not calibrated (common)
- 2. NIBP cuff pressure not maintained; cuff leaking (occasional)
Verify calibration required
- Expected
- Calibration status indicator shows need for recalibration
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
NIBP tubing system leaking caution
Pressure does not stabilize or drops more than 1 mmHg per 5 seconds during NIBP calibration check
Ranked causes
- 1. External NIBP test setup leaking (common)
- 2. Internal NIBP tubing leaking (occasional)
Select NIBP calibration from service menu
- Test point
- Monitor Setup > Defaults & Service > Service Calibrations menu
- Expected
- Calibrations menu displays
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 measurement failure patient safety
Oxygen saturation or pulse rate not displayed or showing incorrect values
Ranked causes
- 1. SpO2 simulator/cable not properly connected (common)
- 2. Simulator SpO2 not configured (occasional)
Verify SpO2 simulator connection
- Test point
- Module SpO2 connector
- Expected
- SpO2 simulator and cable securely connected
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature measurement failure patient safety
Temperature readings not displayed or showing incorrect values
Ranked causes
- 1. Temperature cables not properly connected (common)
- 2. Simulator not configured for 400 series temperature probes (occasional)
Verify temperature adapter cable connection
- Test point
- Module Temp/CO connector
- Expected
- Dual temperature adapter cable securely connected
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Analog output and defib sync marker not functioning high voltage
Analog output or defibrillator synchronization marker signals not present or incorrect
Ranked causes
- 1. Analog output/defib sync cable disconnected or improperly connected (common)
Connect analog output/defib sync cable to module
- Test point
- Defib/Sync connector
- Expected
- Cable seated and connected
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Apnea alarm not activated patient safety
Monitor does not trigger apnea alarm when respiration rate is zero
Ranked causes
- 1. Simulator not configured for apnea simulation or alarm threshold issue (common)
Verify apnea alarm activation
- Test point
- Monitor screen and alarm
- Expected
- Apnea alarm activated when respiration stops
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Arrhythmia paused patient safety
ECG channels unavailable for analysis due to excessive artifact (20 seconds), internal HR calculation not updated (30 seconds), or single lead analysis ECG 1 inactive (5 seconds). Appears in alarm area or waveform area.
Ranked causes
- 1. Excessive artifact on ECG signal (common)
Inspect electrode contact quality
- Test point
- All ECG electrode sites
- Expected
- Firm contact, no visible gaps or debris
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Asystole alarm not triggered patient safety
Monitor does not display asystole alarm when asystole is configured on simulator
Ranked causes
- 1. Simulator not configured to asystole output (common)
Verify asystole alarm appears
- Test point
- Monitor display
- Expected
- Asystole alarm shown on screen
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bottom housing is damaged caution
Cracks or other damage visible on bottom housing
Ranked causes
- 1. Bottom housing cracked or damaged (common)
Inspect bottom housing for cracks or damage
- Test point
- Bottom housing surface
- Expected
- Housing intact without cracks
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
C.O. measurement removed patient safety
Cardiac output measurement module has been disconnected. Message appears in alarm area.
Ranked causes
- 1. C.O. measurement module disconnected (common)
Verify C.O. measurement module is physically connected
- Test point
- C.O. module connectors
- Expected
- Module fully seated and connected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration error
Calibration error message appears in parameter window when calibration procedure fails
Ranked causes
- 1. Calibration procedure failed (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Call service message displayed
Call service message appears in parameter window indicating a technical fault
Ranked causes
- 1. Technical fault detected (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Channel calibration in progress
Calibrating message appears in parameter window during active calibration of a channel
Ranked causes
- 1. Calibration procedure active (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check device malfunction
SpO2 acquisition module has failed and is unable to process signal.
Ranked causes
- 1. Acquisition module failure (occasional)
Perform planned maintenance on acquisition module
- Test point
- Acquisition module
- Expected
- Module functions normally after maintenance
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check NIBP message displayed
Check NIBP message appears in alarm area when systolic, diastolic, and/or mean results are missing
Ranked causes
- 1. Systolic, diastolic, and/or mean results missing (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control measurement pressure alarm limit exceeded
Control measurement message appears in parameter window when pressure alarm limit is exceeded during measurement
Ranked causes
- 1. Pressure alarm limit exceeded during measurement (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cuff loose
Cuff loose message appears in parameter window when cuff connection is loose or cuff hose is disconnected
Ranked causes
- 1. Loose cuff or cuff hose (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cuff occlusion detected patient safety
Cuff occlusion message appears in parameter window when occlusion occurs during measurement or cuff is overpressured
Ranked causes
- 1. Occlusion during measurement or overpressured cuff (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cuff overpressure detected patient safety
Cuff overpressure message appears in parameter window when NIBP cuff is squeezed during measurement and exceeds pressure safety limits
Ranked causes
- 1. NIBP cuff squeezed during measurement exceeding pressure safety limits (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cuff size not selected
Cuff size has not been selected in NIBP setup
Ranked causes
- 1. Cuff size not selected from setup menu (common)
Verify cuff size selection
- Test point
- NIBP setup menu
- Expected
- Cuff size selected (e.g., adult, child, infant)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device malfunction or performance issue patient safety
General troubleshooting guidance for technical problems; patient disconnected from monitor before service
Ranked causes
- 1. Visual inspection incomplete or findings missed (common)
Visual inspection of module exterior
- Test point
- Front cover, connectors, casing, latch, patient cables
- Expected
- Front cover intact, connectors intact and properly attached, casing and latch clean and intact, cables clean and intact
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG measurement not displaying correctly patient safety
Monitor does not display ECG leads II, VI & aVL, or waveforms contain noise; heart rate not 80±5 bpm or QRS tone absent
Ranked causes
- 1. Simulator not configured for normal sinus rhythm (common)
Verify ECG leads II, VI & aVL displayed and noise-free
- Test point
- Monitor display
- Expected
- Leads II, VI & aVL visible, no noise
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrical safety test failure patient safety
Patient leakage current exceeds acceptable limits during source or sink testing
Ranked causes
- 1. Electrical isolation or patient circuit defect (unknown)
Patient (source) leakage current test
- Test point
- Patient circuit source
- Expected
- Within acceptable limits per service manual
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Faulty SpO2 probe
The SpO2 sensor has failed and is no longer functional.
Ranked causes
- 1. Sensor failure (occasional)
Attempt sensor reactivation or test with another sensor
- Test point
- SpO2 sensor
- Expected
- SpO2 signal returns or test sensor functions
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incompatible SpO2 probe
The sensor connected is not compatible with the PDM system.
Ranked causes
- 1. Incompatible sensor type installed (occasional)
Verify sensor compatibility with PDM system
- Test point
- Sensor model and type
- Expected
- Sensor is listed as compatible with CARESCAPE PDM
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lead off - single or multiple leads disconnected patient safety
One or more leadwires or electrodes disconnected. Message appears in alarm area and waveform area with specific lead designation (e.g., LA/L, LL/F, RA/R, RL/N, V2/C2-V6/C6, Va/Ca, Vb/Cb). Other ECG leads available for arrhythmia detection.
Ranked causes
- 1. Leadwire or electrode connection loose or disconnected (common)
Verify connection of identified lead
- Test point
- Specific lead shown in alarm message (LA, LL, RA, RL, V2-V6, Va, Vb, etc.)
- Expected
- Lead connected and signal present
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lead off detection not working patient safety
Detaching leadwire does not trigger lead off message or automatic lead switching
Ranked causes
- 1. Leadwire connection issue or monitor lead detection malfunction (common)
Verify Lead II disappears when RA/R detached
- Test point
- ECG1 waveform field
- Expected
- Lead II disappears, RA/R lead off message shown
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leads off - multiple leads disconnected, arrhythmia detection impossible patient safety
One or more connected electrodes disconnected and arrhythmia detection is not possible. Appears in alarm area and waveform area.
Ranked causes
- 1. Multiple electrode/leadwire disconnections preventing arrhythmia analysis (common)
Inspect all leadwire and electrode connections
- Test point
- All ECG lead connection points
- Expected
- All leads connected with secure seating and good skin contact
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Long measurement time
NIBP measurement takes longer than expected: >2 min for adult/child, 85 s for infant
Ranked causes
- 1. Patient status issue or equipment malfunction (unknown)
Verify measurement time within specification
- Test point
- NIBP measurement display
- Expected
- >2 min for adult/child, 85 s for infant
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low perfusion at measurement point patient safety
Blood perfusion at the sensor measurement site is insufficient for reliable signal detection.
Ranked causes
- 1. Poor perfusion at sensor site (common)
Assess blood perfusion at current sensor site
- Test point
- Measurement site
- Expected
- Palpable pulse, warm skin, good color
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low signal quality
Signal quality is questionable or poor. Patient may be shivering or sensor placement is suboptimal.
Ranked causes
- 1. Poor sensor placement or patient shivering (common)
Evaluate sensor placement on patient
- Test point
- Sensor site
- Expected
- Proper contact, minimal motion artifact, adequate perfusion
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Module power or LED not illuminated
Communication LED on module not lighting when connected to monitor
Ranked causes
- 1. Module defective (common)
Connect known good module to monitor
- Expected
- Known good module LED lights
NIBP alarm limits changed patient safety
Alarm limits have changed due to cuff size or profile change
Ranked causes
- 1. Cuff size or profile changed (common)
Confirm cuff size is correct
- Test point
- NIBP setup menu
- Expected
- Cuff size matches patient and clinical order
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP Auto mode stopped patient safety
NIBP Auto mode has stopped due to loose cuff, loose hose, or cuff size changed to not selected
Ranked causes
- 1. Loose cuff or cuff hose (common)
Inspect cuff and hose connections
- Test point
- Cuff connector and hose attachment points
- Expected
- Firm connection with no visible damage or kinks
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP calibration failed or gain correction exceeded
Unable to complete NIBP calibration; Confirm button inactive or gain correction exceeds +/-10%
Ranked causes
- 1. NIBP main board components not functioning properly (occasional)
Note gain correction value required
- Expected
- Gain correction within +/-10% of factory calibration
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
NIBP cuff loose patient safety
Loose cuff or cuff hose detected
Ranked causes
- 1. Loose cuff or cuff hose connection (common)
Verify cuff and hose are secure
- Test point
- Cuff connector and hose attachment
- Expected
- Firm, tight connection with no movement
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP cuff occlusion during measurement patient safety
Occlusion detected during measurement or cuff overpressure condition
Ranked causes
- 1. Cuff occlusion or overpressure (common)
Verify cuff condition and placement
- Test point
- Cuff on patient limb
- Expected
- Cuff properly placed, not kinked, pressure appropriate
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP measurement functional check failed caution
NIBP tubing system has leaks or module is out of calibration; pressure readings inaccurate or unstable
Ranked causes
- 1. Combined NIBP tubing leak and/or calibration issue (common)
Check NIBP tubing system and calibration status
- Test point
- Full NIBP calibration check procedure per steps 1-7
- Expected
- Pass status: pressure stable ±1 mmHg, no >1 mmHg drop per 5 sec
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP out of calibration caution
NIBP readings differ by more than ±1 mmHg between manometer and calibration menu
Ranked causes
- 1. NIBP module requires recalibration (common)
Compare manometer and calibration menu readings
- Test point
- Manometer display vs. Calibration Check menu
- Expected
- Readings match ±1 mmHg, ±1 mmHg
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP pressure value outside alarm limits patient safety
NIBP systolic, diastolic, or mean pressure equal to or outside set alarm limits
Ranked causes
- 1. Patient hemodynamic change (common)
Review measured pressure values
- Test point
- NIBP display (systolic, diastolic, mean)
- Expected
- Values within set alarm limits
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP STAT mode stopped patient safety
STAT mode has stopped due to loose cuff, loose hose, or cuff size changed to not selected
Ranked causes
- 1. Loose cuff or cuff hose (common)
Inspect cuff and hose connections
- Test point
- Cuff connector and hose attachment points
- Expected
- Firm connection with no visible damage
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- NIBP zeroing in progress
- NIBP zeroing successful
- No pressure reading displayed
- No SpO2 probe connected
- Pacemaker spikes not detected
- Parameter interface module damaged
- Parameter not displayed on host monitor
- Planned maintenance procedure
- PR(SpO2) high or low
- Probe off
- Pulse search in progress
- Respiration rate incorrect
- Service the PDM - Voltage rail errors
- Service the PDM Calibrate ECG Analog Out
- Service the PDM Calibrate IP Analog Out
- Service the PDM Calibrate NIBP
- Service the PDM Configure Power Frequency
- Service the PDM Error Code 0xPDM1000
- Service the PDM Error Code 0xPDM1001
- Service the PDM Error Code 0xPDM1002
- Service the PDM Error Code 0xPDM1003
- Service the PDM Error Code 0xPDM1004
- Service the PDM Error Code 0xPDM1005
- Service the PDM Error Code 0xPDM1006
- Service the PDM Error Code 0xPDM100C
- Service the PDM Error Code 0xPDM1100
- Service the PDM Error Code 0xPDM1101
- Service the PDM Error Code 0xPDM1102
- Service the PDM Error Code 0xPDM1103
- Service the PDM Error Code 0xPDM1104
- Service the PDM Error Code 0xPDM1105
- Service the PDM Install License
- SpO2 high or low
- SpO2 measurement not displaying
- SpO2 measurement removed
- Temperature measurement removed
- Temperature reading incorrect
- Top housing is damaged
- Visual inspection failure
- Wrong cable connected
40 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0xPDM1000 | Secure storage data is corrupt | System detects corrupted data in secure storage on startup or runtime | critical |
0xPDM1001 | Flash file system partition is full | System detects one or more partitions at capacity during operation | critical |
0xPDM1002 | Error writing data to secure storage | Write operation to secure storage fails during normal operation | critical |
0xPDM1003 | Error writing to the SPI bus | SPI bus write operation fails during normal operation | critical |
0xPDM1004 | Error reading the network interface information | Network interface data read fails during startup or operation | error |
0xPDM1005 | Error reading flash file system partition information | Flash partition read operation fails during startup or operation | critical |
0xPDM1006 | Error using the TWI bus | TWI (Two-Wire Interface) bus operation fails during startup or runtime | critical |
0xPDM100C | MAC address not readable from hardware | System cannot read MAC address from hardware on startup | error |
0xPDM1100 | Main CPU board 1.8V rail voltage out of specification | Power monitoring detects 1.8V rail outside acceptable range | critical |
0xPDM1101 | Main CPU board 3.3V rail voltage out of specification | Power monitoring detects 3.3V rail outside acceptable range | critical |
0xPDM1102 | Main CPU board 5.4V rail voltage out of specification | Power monitoring detects 5.4V rail outside acceptable range | critical |
0xPDM1103 | Main CPU board 12V rail voltage out of specification | Power monitoring detects 12V rail outside acceptable range | critical |
0xPDM1104 | Main CPU board –12V rail voltage out of specification | Power monitoring detects –12V rail outside acceptable range | critical |
0xPDM1105 | Main CPU board power bus voltage out of specification | Power monitoring detects main power bus outside acceptable range | critical |
Calibrated | Channel calibrated successfully | Successful completion of channel calibration | info |
Calibrating | Calibration of a channel is in progress | Calibration procedure initiated | info |
Calibration error | Calibration has failed | Calibration procedure failed to complete | error |
Calibration fail | Temperature or C.O. measurement channel calibration failed | Calibration procedure initiated but failed to complete successfully | error |
Call service | Technical fault | System detected an unspecified technical fault | critical |
Check NIBP | Systolic, diastolic, and/or mean results missing | Measurement completed without all required values | warning |
CO measurement removed | Cardiac output measurement module disconnected | C.O. module detected as missing during operation | warning |
Control measurement | Pressure alarm limit exceeded | Pressure during measurement exceeded alarm limit | warning |
Cuff loose | Loose cuff or cuff hose | Loss of pressure signal or cuff connection failure | warning |
Cuff occlusion | Occlusion during measurement or overpressured cuff | Occlusion detected or cuff pressure exceeds safe limits during measurement | warning |
Cuff overpressure | NIBP cuff is squeezed during measurement and exceeded the pressure safety limits | Cuff pressure exceeded safety limits during measurement | warning |
PDM battery low | Battery charge is insufficient, charging failure detected | Battery cannot be charged due to power fault | warning |
PDM battery temp high | Battery temperature exceeds safe operating limit | Battery temperature sensor detects excessive temperature | warning |
PDM charging is denied | Battery charging is disabled due to excessive internal temperature | Monitor internal temperature exceeds safe charging threshold | warning |
Service the PDM Error Code 0xPDM1106 | ECG DAS board 1.26V rail voltage out of specification | Power supply rail voltage deviation detected on ECG data acquisition board | error |
Service the PDM Error Code 0xPDM1107 | ECG DAS board 3.3V rail voltage out of specification | Power supply rail voltage deviation detected on ECG data acquisition board | error |
Service the PDM Error Code 0xPDM1108 | ECG DAS board 5V rail voltage out of specification | Power supply rail voltage deviation detected on ECG data acquisition board | error |
Service the PDM Error Code 0xPDM1109 | ECG DAS board 7.5V rail voltage out of specification | Power supply rail voltage deviation detected on ECG data acquisition board | error |
Service the PDM Error Code 0xPDM110A | ECG DAS board negative 7.5V rail voltage out of specification | Power supply rail voltage deviation detected on ECG data acquisition board | error |
Service the PDM Error Code 0xPDM110B | Main DAS board 1.26V rail voltage out of specification | Power supply rail voltage deviation detected on main data acquisition board | error |
Service the PDM Error Code 0xPDM110C | Main DAS board 3.3V rail voltage out of specification | Power supply rail voltage deviation detected on main data acquisition board | error |
Service the PDM Error Code 0xPDM110D | Main DAS board 5V rail voltage out of specification | Power supply rail voltage deviation detected on main data acquisition board | error |
Service the PDM Error Code 0xPDM110E | Main DAS board 2.5V rail voltage out of specification | Power supply rail voltage deviation detected on main data acquisition board | error |
Service the PDM Error Code 0xPDM110F | Main DAS board Nellcor 5V rail voltage out of specification | Power supply rail voltage deviation detected on main data acquisition board Nellcor channel | error |
Service the PDM Error Code 0xPDM1110 | Main DAS board Nellcor negative 3V rail voltage out of specification | Power supply rail voltage deviation detected on main data acquisition board Nellcor channel | error |
Service the PDM Error Code 0xPDM1120 | Temperature sensor reading out of specification | Internal temperature measurement exceeds acceptable range | error |
T1 calibration fail / T2 Calibration fail / Tblood calibrati | Temperature channel (T1, T2, or Tblood) calibration failed | Calibration procedure attempted for specified temperature channel | error |
Temp measurement removed | Temperature measurement module or cable disconnected | Acquisition module or temperature cable detected as missing during operation | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI