GE Healthcare CARESCAPE ONE: troubleshooting and error codes

The ServiceForge knowledge base holds 35 documented symptoms for the GE Healthcare CARESCAPE ONE, with 73 ranked causes, 60 diagnostic checks with test points and expected values, and 151 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

No parameter data displayed

CARESCAPE parameters not shown on monitor screen

Ranked causes

  1. 1. Parameter not configured for display (common)
  2. 2. Parameter cable disconnected or loose (common)
  3. 3. Parameter device faulty (occasional)
  4. 4. Parameter cable damaged (occasional)
  5. 5. Parameter device physically damaged (occasional)
  6. 6. Patient block connector damaged (rare)

12 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No temperature displayed

Temperature measurement not showing on monitor screen

Ranked causes

  1. 1. Temperature channel not configured for display (common)
  2. 2. Incompatible temperature probe (occasional)
  3. 3. Faulty temperature probe (occasional)
  4. 4. Faulty CARESCAPE Parameter cable (occasional)
  5. 5. Faulty CARESCAPE Temperature module (occasional)
  6. 6. Temperature out of measurement range (rare)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No invasive pressure displayed patient safety

Invasive pressure measurement not showing on screen

Ranked causes

  1. 1. Defective transducer (occasional)
  2. 2. Module not connected (common)
  3. 3. Transducer adapter cable not connected (common)
  4. 4. Invasive pressure channel not configured (common)
  5. 5. Invasive pressure channel not zeroed (common)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Troubleshooting CO2 calibration issues

Diagnostics for specific error messages and conditions encountered during Microstream CO2 calibration check or calibration procedure.

Ranked causes

  1. 1. Wrong gas message displayed (occasional)
  2. 2. Unstable gas flow message displayed (occasional)
  3. 3. Device is not ready message displayed (occasional)
  4. 4. Zero error message displayed during calibration (rare)
  5. 5. Calibration error message displayed (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ground leakage current exceeds limit high voltage

Ground leakage current measured greater than 1000 µA with NEUTRAL switch CLOSED or OPEN

Ranked causes

  1. 1. Defective power cord (common)
  2. 2. Incorrectly wired power outlet (occasional)
  3. 3. Incorrect test setup configuration (occasional)
  4. 4. Internal device leakage fault (rare)
First check

Verify power cord integrity

Test point
Power cord surface and connectors
Expected
No wear or damage

5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to turn on monitor when connected to CARESCAPE Dock F0

Monitor does not power on when connected to AC-powered CARESCAPE Dock F0

Ranked causes

  1. 1. Power cord is loose (common)
  2. 2. Power cord is faulty (occasional)
  3. 3. Power outlet does not meet specified requirements (occasional)
  4. 4. AC/DC power supply is faulty (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarms do not work patient safety

Monitor does not produce audible alarm sounds

Ranked causes

  1. 1. Audible alarms turned off (common)
  2. 2. Alarm volume too low (common)
  3. 3. Defective speaker (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Chassis leakage current to ePort exceeds limit high voltage

Chassis leakage current measured at ePort connection greater than 100 µA (NEUTRAL CLOSED/GND CLOSED or OPEN) or 500 µA (NEUTRAL OPEN)

Ranked causes

  1. 1. Defective power cord (common)
  2. 2. Incorrectly wired power outlet (occasional)
  3. 3. Internal device leakage fault (rare)
First check

Perform chassis leakage test to ePort with NEUTRAL CLOSED and GND CLOSED

Test point
Chassis to ePort connection via safety test cable
Expected
Less than 100 µA, < 100 µA
Tools
Fluke 612/615/620 or equivalent; ePort safety test cable P/N per manual

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touchscreen inoperative

Touchscreen does not respond to touch input

Ranked causes

  1. 1. Touchscreen cable loose (common)
  2. 2. Faulty touchscreen sensor (occasional)
  3. 3. Touchscreen dirty (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery not charging caution

Battery does not charge when inserted in monitor

Ranked causes

  1. 1. Battery connector damaged (occasional)
  2. 2. Battery defective (common)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Analog output caution

Analog output calibration adjusts ECG and invasive pressure analog outputs using a calibrated digital voltmeter after main board replacement or out-of-spec measurements.

Ranked causes

  1. 1. Main Board Assembly replaced (common)
  2. 2. Measured value not within specification limits (occasional)
First check

Turn on CARESCAPE ONE and wait for normal screen

Expected
Normal screen appears

8 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Microstream CO2 gas caution

CO2 calibration adjusts measured CO2 value to show exactly 5% using reference calibration gas. Required after 1200 hours or 1 year post-installation, then every 4000 hours or annually.

Ranked causes

  1. 1. Calibration required per operating hour schedule (common)
  2. 2. Calibration required message displayed on screen (common)
First check

Verify correct date and time on monitor

Test point
Monitor system clock
Expected
Current date and time displayed

11 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: NIBP patient safety

NIBP calibration must be performed when calibration check fails or measured values are outside specification limits.

Ranked causes

  1. 1. NIBP calibration check failed (common)
  2. 2. Measured value not within specification limits (occasional)
First check

Turn on CARESCAPE ONE and wait for normal screen

Expected
Normal screen appears

5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Dock power supply or cord damaged high voltage

Power cord shows wear, damage, or continuity failure; external power supply not properly mounted or secured

Ranked causes

  1. 1. Power cord wear or physical damage (occasional)
  2. 2. Power supply not properly mounted or secured (occasional)
First check

Inspect power cord for wear or damage

Test point
Power cord surface and connectors
Expected
No visible wear or damage

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor starts but display remains white

Monitor powers on but LCD display backlight illuminates without showing any content, no error messages

Ranked causes

  1. 1. Display cable is damaged or loose (common)
  2. 2. LCD display is faulty (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP not functioning patient safety

Non-invasive blood pressure measurement not working

Ranked causes

  1. 1. NIBP not calibrated (common)
  2. 2. Cuff pressure not maintained (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 reading not appearing on screen

SpO2 parameter value and plethysmographic waveform do not display after connecting sensor

Ranked causes

  1. 1. SpO2 sensor not detecting pulse (common)
  2. 2. SpO2 interconnect cable not connected (occasional)
First check

SpO2 reading appears in parameter window

Test point
CARESCAPE ONE display
Expected
SpO2 value displayed

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Abnormally low invasive pressure reading patient safety

Pressure measurement lower than expected in invasive pressure monitoring

Ranked causes

  1. 1. Transducer wrongly positioned (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot view feature despite enabled license

Feature not accessible on monitor even though license is enabled

Ranked causes

  1. 1. Software package does not include feature (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

CARESCAPE SpO2-Masimo measurement values inaccurate patient safety

SpO2 measurement readings do not match expected values

Ranked causes

  1. 1. CARESCAPE SpO2-Masimo device faulty (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Dock fasteners loose or insecure caution

Mounting solution fasteners not properly tightened or correctly secured on CARESCAPE Dock F0

Ranked causes

  1. 1. Fasteners not properly tightened (occasional)
First check

Inspect mounting fasteners

Test point
All dock mounting fasteners
Expected
Intact, properly tightened, and correctly secured

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

LoFlo CO2 accuracy test failed

CO2 value displayed does not match expected calibration gas concentration during accuracy check

Ranked causes

  1. 1. LoFlo Sidestream Module calibration drift (occasional)
First check

Wait for LoFlo Sidestream Module warm-up

Test point
LoFlo Sidestream Module
Expected
Warm up time exceeded minutes

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

LoFlo CO2 sample flow rate out of specification

Measured sample flow rate does not meet 50 ml/min ± 10 ml specification during functional check

Ranked causes

  1. 1. LoFlo Sidestream Module malfunction (common)
First check

Wait for LoFlo Sidestream Module warm-up

Test point
LoFlo Sidestream Module
Expected
Warm up messages cleared minutes

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low audio volume or no audio

Speaker output is quiet or silent

Ranked causes

  1. 1. Speaker covers blocked by dust or debris (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Masimo rainbow measurements not displayed on CARESCAPE

Masimo rainbow measurement data not visible on CARESCAPE ONE screen

Ranked causes

  1. 1. Measurement only visible on host monitor when used as acquisition platform (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Masimo rainbow measurements not displayed on host monitor

Masimo rainbow measurement data not visible on host monitor

Ranked causes

  1. 1. Masimo rainbow measurement licenses not activated (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor starts with FRU software display

Monitor displays Field Replacement Unit, SW version, and IP address at startup instead of CARESCAPE software

Ranked causes

  1. 1. CARESCAPE software not installed after main board FRU replacement (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Parameter cables or strain reliefs damaged caution

Abrasion or damage observed on CARESCAPE Parameter cables or cable strain reliefs

Ranked causes

  1. 1. Cable wear or environmental abrasion (occasional)
First check

Inspect all Parameter cables and strain reliefs

Test point
All cable lengths and connector areas
Expected
No abrasion or damage visible

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Parameter mounting bracket or latch damaged caution

CARESCAPE rSO2—INVOS or CO2—LoFlo mounting bracket broken or CARESCAPE SpO2—Nellcor latch damaged or missing

Ranked causes

  1. 1. Bracket or latch damaged or broken (occasional)
First check

Inspect Parameter mounting brackets and latches

Test point
rSO2/CO2 mounting brackets and SpO2 Nellcor latch
Expected
Intact and undamaged

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Parameter not properly connected or mounted caution

CARESCAPE Parameter disconnected from CARESCAPE ONE or not properly mounted to Parameter Docks or mounting solutions

Ranked causes

  1. 1. Parameter disconnected or loose (occasional)
First check

Verify Parameter connection to CARESCAPE ONE

Test point
Parameter connector and receptacle
Expected
Properly connected and latched

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to activate new monitor software

Cannot activate or upgrade monitor software; invalid activation code error

Ranked causes

  1. 1. Missing or invalid activation code (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to complete NIBP calibration

Confirm button inactive during NIBP calibration procedure

Ranked causes

  1. 1. Gain correction exceeds ±10% limit (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not power on

Start-up sequence does not advance beyond GE HealthCare logo screen; error messages may appear; unable to read software from memory

Ranked causes

  1. 1. Software not properly installed after main board replacement (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Visual damage or contamination detected caution

Cracks, breaks, dirt, loose connectors, corroded pins, bent pins, or fluid contamination on enclosure, display misalignment, or battery not fully inserted

Ranked causes

  1. 1. Physical damage or environmental exposure (unknown)
First check

Inspect product labeling, markings and symbols

Test point
All visible labels
Expected
Intact and readable

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Wrong software package activated

Incorrect software package enabled on monitor

Ranked causes

  1. 1. Wrong software package activated for device (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
0x11CFCARESCAPE CO2 device surface temperature too high (LoFlo devices)CARESCAPE CO2 device surface temperature exceeds safe operating limitwarning
0x11D5CARESCAPE CO2 device surface temperature too high (Microstream)CARESCAPE CO2 Microstream device surface temperature exceeds safe operating limitwarning
Acquiring signal / rSO2 Acquiring signalrSO2 signal acquisition in progress from patientMeasurement startup sequenceinfo
Amber LED flashing on, green LED off on CARESCAPE SpO2-MasimSoftware failure or startup failureDevice powered but not starting or running properlyerror
Amber LED off, Green LED on on CARESCAPE SpO2-MasimoDevice running and operationalDevice successfully started and communicatinginfo
Amber LED on, Green LED off on CARESCAPE SpO2-MasimoDevice starting upPower applied during initial startup sequenceinfo
Analog output malfunctionCARESCAPE ONE analog output voltage failureHardware or software fault in analog output circuitryerror
Apnea deactivatedApnea alarm start-up conditions not reachedApnea alarm conditions not metinfo
Auto detecting sourceAutomatic gas measurement source detection is ongoingCO2 source detection in progressinfo
Battery empty!Battery charge critically low, less than 5 minutes remainingBattery state of charge drops below critical thresholdcritical
Battery failureBattery is faulty and cannot hold chargeBattery self-test or charging circuit detects battery faulterror
Battery lowBattery charge low, less than 20 minutes remainingBattery state of charge drops below warning thresholdwarning
Battery temp highBattery temperature errorBattery internal temperature exceeds safe operating limiterror
Both amber and green LEDs off on CARESCAPE SpO2-MasimoNo power or software failurePower not applied or critical software failurecritical
CalibratedChannel calibrated successfullyCompletion of channel calibrationinfo
CalibratingCalibration of a channel is in progressUser initiates channel calibrationinfo
Calibration errorUnsuccessful CO2 calibrationFailed calibration attemptwarning
Calibration errorCalibration has failedUnsuccessful calibration attemptwarning
Calibration failCalibration of temperature channel has failedFailed temperature calibrationerror
Calibration requiredCARESCAPE CO2 — Microstream needs calibrationCalibration not performed or expiredwarning
Call Service: Text(s) missingSoftware text file missing for current languageText resource file not found or corrupted during language initializationerror
Check CO2 adapterObstruction in LoFlo receptacle sample cell or cell not in optimal operating rangeSample cell obstruction or misalignmentwarning
Check deviceCARESCAPE Pressure has malfunctionedHardware or communication fault detectederror
Check deviceHardware or software fault in CARESCAPE TemperatureCable fault, no updated data, data sample drift, or data out of rangeerror
Check device / ECG malfunctionHardware or software fault in CARESCAPE ECGCable fault, no updated data, data sample drift, or configuration failureerror
Check device / Imped. malfunctionHardware or software fault in CARESCAPE ECG impedance measurementCable fault, configuration failure, no updated data, or data sample drifterror
Check device / rSO2 malfunctionHardware or software fault in CARESCAPE rSO2 — INVOS; no rSO2 detectedDevice fault; signal losserror
Check device / SpO2 malfunctionCARESCAPE SpO2 device malfunction; no SpO2 data detectedDevice hardware or software fault; no signal detectederror
Check gas flowSample flow rate is lower or higher than nominal due to obstruction or leakAbnormal flow rate detectedwarning
Check NIBPSystolic and/or diastolic results missingIncomplete NIBP measurementwarning
Check probe / Check SpO2 probeNo detectable SpO2 signal; sensor faulty or detached from patientSignal loss; sensor disconnection; sensor malfunction; pulse rate lost (Nellcor); saturation update exceeds 30 secondserror
Check rSO2 labelsrSO2 measurement labels need verification after patient transportPatient transport completed; label verification requiredinfo
Check sample lineSample flow abnormal due to obstruction, cut, split, or saturated filterAbnormal flow rate detectedwarning
Check SpHbNo adequate signal for SpHb measurement with CARESCAPE SpO2 — MasimoPoor sensor connection or defective sensorwarning
CO2 device failure: Call serviceFront end processor startup failure, incorrect initialization, or no keep alive responseDevice startup, initialization, or communication failureerror
CO2 device malfunctionCARESCAPE CO2 device startup, runtime failure, or in software download modeDevice startup or runtime failureerror
CO2 device overheated. Shutting downCARESCAPE CO2 device surface temperature too high, measurement turned off. Error codes: 0x120F, 0x1215Temperature threshold exceedederror
CO2 inaccurateMeasured values outside range or inaccurate due to high respiration rateHigh respiration rate or out-of-range measurementswarning
CO2 measurement purgingCARESCAPE CO2 — Microstream self-cleaning blockage in sample lineBlockage detected in sample lineinfo
CO2 measurement removedCARESCAPE CO2 device has been disconnected from CARESCAPE ONEDevice disconnectionwarning
CO2 sensor overheated. Shutting downCARESCAPE CO2 — Microstream sensor temperature too high, measurement offSensor temperature threshold exceedederror
CO2 sensor removedLoFlo Sidestream Module disconnected from CARESCAPE CO2 — LoFloModule disconnection detectedwarning
CO2 zero requiredCARESCAPE CO2 — LoFlo needs zeroingZeroing not performed or expiredwarning
Condition batteryBattery needs conditioning before useBattery detected that requires initial conditioning cyclewarning
Configuration error(s)One or more configuration errors detectedConfiguration validation failure at startup or during operationerror
Continuous blockage. Check sample line.CARESCAPE CO2 — Microstream sampling line is blockedSustained blockage detectedwarning
CS ONE Faulty Device ConnectedConnected CARESCAPE Parameter has communication failureCommunication error detected between monitor and parametererror
CS ONE Faulty Device in Port #xCARESCAPE Parameter in port x has detected failureOver current condition, repeated over temperature faults, or repeated communication failureserror
CS ONE Speaker FailureAlarm speaker is inoperativeSpeaker hardware failure or audio output circuit faulterror
CS ONE Unknown Device in Port #xUnauthorized device connected to CARESCAPE ONE USB portUnrecognized USB device detected on one of eight portswarning
Cuff looseLoose cuff or cuff hoseLoose connection detectedwarning
Cuff occlusionOcclusion during measurement or overpressured cuffObstruction detected during NIBP measurementwarning
Cuff overpressureNIBP cuff exceeded pressure safety limitsCuff squeezed during measurement exceeding limitswarning
Demo mode / SpO2 Demo modeCARESCAPE SpO2 — Masimo is operating in demonstration modeDemo mode enabledinfo
Device not readyDevice is still warming up or there is a technical conditionDevice startup or initialization in progressinfo
Device overheatingCARESCAPE ECG surface temperature is too highInadequate ventilation or obstruction of ECG devicewarning
ECG analog output malfunctionHardware or software fault in CARESCAPE ECG analog outputECG analog output circuit failure or data communication errorerror
ECG device failure: Call serviceCARESCAPE ECG startup or initialization failureDevice unable to initialize or start upcritical
ECG device overheated. Shutting downCARESCAPE ECG surface temperature exceeded safe operating limits; device is being disabledSurface temperature too high; monitor automatically shuts down ECG to prevent damageerror
ECG measurements removedCARESCAPE ECG has been disconnected or removed from systemECG device physically removed or connection lostwarning
Expiring cableCARESCAPE SpO2 — Masimo interconnect cable approaching expiration dateCable near end of service lifewarning
Expiring probeCARESCAPE SpO2 — Masimo sensor or adhesive sensor approaching expiration dateSensor near end of service lifewarning
Faulty cable / SpO2 faulty cableCARESCAPE SpO2 — Masimo interconnect cable is defectiveCable malfunction detected by deviceerror
Faulty probe / SpO2 faulty probe / Faulty adhesive SpO2 probSpO2 sensor or adhesive sensor has failedSensor malfunction detectederror
Identical CO2 deviceMore than one CARESCAPE CO2 device in the system. Error codes: 0x110F, 0x1115Multiple CO2 devices detectedwarning
Identical ECG deviceTwo CARESCAPE ECG devices are connected to the systemMultiple ECG modules detectedwarning
Identical pressure devicesThree or more CARESCAPE Pressures connectedMultiple pressure devices detectedwarning
Identical rSO2 devicesTwo or more CARESCAPE rSO2 — INVOS devices are connectedMultiple rSO2 devices detectedwarning
Identical SpO2 deviceMultiple CARESCAPE SpO2 devices are connectedMore than one SpO2 device detectedwarning
Identical Temp deviceTwo CARESCAPE Temperatures connected. Error code: 0x110DMultiple temperature devices detectedwarning
Incompatible cable / Incompatible SpO2 cableIncompatible or unrecognized interconnect cable connected to CARESCAPE SpO2 — MasimoUnrecognized cable detectederror
Incompatible probe / Incompatible SpO2 probe / IncompatibleSensor is not compatible or not recognized by CARESCAPE SpO2 — MasimoIncompatible or unrecognized sensor detectederror
Incompatible rSO2 deviceCARESCAPE rSO2 — INVOS is not compatible with systemDevice compatibility check failederror
Incompatible SpO2 deviceCARESCAPE SpO2 device startup, initialization failure, or hardware faultDevice incompatibility or initialization failure detectederror
Insufficient power for ECG deviceNot enough power available to operate CARESCAPE ECGPower supply insufficient for ECG operationerror
Insufficient power for rSO2 deviceNot enough power available for CARESCAPE rSO2 — INVOS operationInsufficient power supplyerror
Insufficient power for SpO2 deviceNot enough power available for CARESCAPE SpO2 device operationInsufficient power supplyerror
IP analog output malfunctionHardware or software fault in CARESCAPE Pressure analog outputPressure analog output circuit failure or data communication errorerror
Long meas. timeThe measurement time is longMeasurement duration exceeds thresholdwarning
Measurement errorTechnical fault in monitor measurement circuitryHardware fault detected in measurement subsystemerror
Measurement off / OFFImpedance respiration measurement is disabled or not activeECG leads not connected to patient or measurement turned off from setup menuinfo
Move SpO2 device to a cooler locationCARESCAPE SpO2 device surface temperature is too highExcessive device temperature detectedwarning
NIBP auto stoppedNIBP Auto mode stopped due to loose cuff, hose, communication failure, or cuff size changeDetection of loose cuff, hose, module communication failure, or cuff size changewarning
NIBP cuff looseLoose cuff or cuff hoseLoose connection detectedwarning
NIBP cuff occlusionOcclusion during measurement or overpressured cuffObstruction detected during NIBP measurementwarning
NIBP hardware fault in CARESCAPE ONENIBP hardware fault detectedHardware malfunction in NIBP moduleerror
NIBP malfunctionNIBP hardware fault in CARESCAPE ONEHardware malfunction detectederror
NIBP STAT stoppedSTAT mode stopped due to loose cuff, hose, communication failure, or cuff size changeDetection of loose cuff, hose, module communication failure, or cuff size changewarning
No battery backupNo battery installed in CARESCAPE ONEBattery compartment empty or battery not detected on startupwarning
No cable / No SpO2 cableInterconnect cable not connected to CARESCAPE SpO2 — MasimoMissing or disconnected cableerror
No probe / No SpO2 probe / No adhesive SpO2 probeSensor not connected to CARESCAPE SpO2 deviceSensor disconnectionerror
Over rangeCO2 value outside measurable rangeCO2 concentration exceeds measurement capabilityerror
Over scaleGas signal exceeds maximum waveform areaGas concentration signal exceeds upper limit of current waveform scalewarning
P1 standby to P4 standbyInvasive pressure channel set to standbyIP channel standby mode activatedinfo
P1 zeroing failed to P4 zeroing failedChannel zeroing not successfulFailed zeroing attemptwarning
Powering down!Battery charge critically low, less than 1 minute remainingBattery state of charge at critically low levelcritical
Pressure device failure: Call serviceFaulty CARESCAPE Pressure detectedHardware failure detectederror
Pressure device initializingNew software being downloaded to CARESCAPE PressureSoftware download in progressinfo
Pressure device overheated. Shutting it downCARESCAPE Pressure turned off due to excessive temperatureSurface temperature threshold exceedederror
Pressure measurement removedCARESCAPE Pressure removed or active pressure channel became inactivePressure device disconnected or channel deactivatedwarning
Pressure SensedPressure has been sensed during zeroingPressure detected during zeroing procedurewarning
Probe initializingCARESCAPE SpO2 — Masimo sensor is initializingSensor initialization in progressinfo
PX not zeroedInvasive pressure channel not zeroedInvasive pressure channel not zeroed before measurementwarning
Replace batteryBattery capacity degraded below acceptable thresholdBattery full capacity less than 60% of new battery capacitywarning
Replace cable / Replace SpO2 cableCARESCAPE SpO2 — Masimo interconnect cable has expiredCable expiration date reachederror
Replace probe / Replace SpO2 probe / Replace adhesive SpO2 pCARESCAPE SpO2 — Masimo sensor or adhesive sensor has expiredSensor expiration date reachederror
Replace X / rSO2 X replace sensorrSO2 sensor is defective or not compatibleSensor malfunction or incompatibility detectederror
rSO2 device failure: Call serviceCARESCAPE rSO2 — INVOS startup or initialization failureDevice initialization failure; startup errorcritical
rSO2 measurement disconnectedCARESCAPE ONE disconnected from host monitor while rSO2 — INVOS connected to acquisition platformHost monitor connection lostwarning
rSO2 measurement removedCARESCAPE rSO2 — INVOS has been disconnected from acquisition platformrSO2 device disconnectionwarning
Sample line blockedCARESCAPE CO2 Microstream sampling line is blockedObstruction detected in sampling line preventing gas flowerror
Sample line disconnectedSample line is disconnected from deviceDisconnection of sampling line from sensor or monitor porterror
Select agePatient age must be selected before CO2 measurementAttempt to start CARESCAPE CO2 Microstream measurement without age selectioninfo
Service CO2 sensorProblem detected with CARESCAPE CO2 deviceInternal fault detected in CO2 measurement deviceerror
Service CS ONE 0xHOST 3001Internal temperature of CARESCAPE ONE out of specificationOne of internal temperature sensors indicates out-of-range temperatureerror
Service CS ONE 0xHOST 3002Internal supply voltage out of specificationOne or more internal power supply voltages exceed specification limitserror
Service CS ONE 0xHOST 3004Disk utilization exceeded 90%Storage partition usage exceeds 90% capacitywarning
Service CS ONE 0xHOST 3100Main board battery voltage out of specificationBattery backup voltage on main board below or above acceptable rangeerror
Service CS ONE 0xPDM1000Data in secure storage corruptChecksum or validation failure of stored configuration dataerror
Service CS ONE 0xPDM1002Error occurred writing data to secured storageWrite operation to secure storage failederror
Service CS ONE 0xPDM1200Initialization of platform security library returned invalid directory errorSecurity library initialization failure during startuperror
Service CS ONE activation failedSettings activation failedConfiguration or settings activation process encountered errorerror
Service CS ONE Calibrate NIBPNIBP calibration values could not be read from secure storageNIBP calibration data missing or unreadable from storageerror
Service CS ONE ECG Analog OutECG calibration values could not be read from secure storageECG analog output calibration data missing or unreadableerror
Service CS ONE Faulty Port #xFault detected on CARESCAPE Parameter portHardware failure detected on monitor porterror
Service CS ONE IP Analog OutInvasive pressure calibration values could not be read from storageIP calibration data missing or unreadable from storageerror
Service CS ONE Power FailureLost communication with power management controllerCommunication failure between main CPU and power management moduleerror
SpO2 device failure: Call serviceCARESCAPE SpO2 device startup or initialization failureDevice unable to start or initialize; possible hardware failurecritical
SpO2 device overheated Shutting downCARESCAPE SpO2 surface temperature too high; CARESCAPE ONE is disabling the deviceExcessive surface temperature detected; monitor auto-shutoff activatederror
SpO2 measurement removedCARESCAPE SpO2 device has been disconnected from systemSpO2 device disconnectionwarning
SW download failed on CS ONE port xSoftware update to CARESCAPE Parameter failedSoftware transfer to connected parameter failed or was interruptederror
SW update in progress on CS ONE port xSoftware update in progress to connected CARESCAPE ParameterFirmware update transfer from CARESCAPE ONE to parameter activeinfo
Temp device failure: Call serviceCARESCAPE Temperature startup or initialization failure. Error codes: 0x108D, 0x124DDevice startup or initialization failureerror
Temp device overheated. Shutting downCARESCAPE Temperature turned off due to excessive temperature. Error code: 0x120DSurface temperature threshold exceedederror
Temp measurement removedCARESCAPE Temperature removed or active temperature channel became inactiveTemperature device disconnected or channel deactivatedwarning
Unable to read licensesLicense file could not be readLicense file not found, corrupted, or unreadableerror
Unstable gas flowGas flow is unstable during calibrationVariation or instability detected in calibration gas flowwarning
Unsupported X / rSO2 X unsupported sensorrSO2 sensor type is not supported by systemUnsupported sensor detectederror
Warming upCARESCAPE CO2 device is in warm-up phaseDevice startup or power-on initialization sequenceinfo
Wrong gasIncorrect calibration gas in useDetected gas composition does not match expected calibration gaserror
X excess light / rSO2 excess lightExcessive light detected on rSO2 sensorHigh ambient light conditions detected at sensorwarning
X interference / rSO2 interferencerSO2 measurement is disturbed by electromagnetic interferenceElectromagnetic noise or interference detectedwarning
X no sensor / rSO2 X sensor disconnectedrSO2 sensor has been disconnected from systemSensor disconnection detectederror
X poor signal / rSO2 poor signalrSO2 measurement cannot be performed due to poor signal qualityInsufficient signal quality detectedwarning
X sensor off / rSO2 X sensor offrSO2 sensor is not connected to patientSensor not in contact with patientwarning
ZeroedZeroing was successfulCompletion of zeroing procedureinfo
Zero ICP separatelyICP channel must be zeroed separately from other invasive pressuresAttempt to zero ICP with other invasive pressureswarning
ZeroingInvasive pressure channel currently being zeroedZeroing procedure in progressinfo
ZeroingZeroing procedure in progressZeroing calibration sequence activeinfo
Zeroing errorZeroing failed during calibration check processZeroing procedure unable to complete successfullyerror
Zero OKZeroing was successfulCompletion of zeroing procedureinfo

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.