GE Healthcare CARESCAPE ONE: troubleshooting and error codes
The ServiceForge knowledge base holds 35 documented symptoms for the GE Healthcare CARESCAPE ONE, with 73 ranked causes, 60 diagnostic checks with test points and expected values, and 151 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No parameter data displayed
CARESCAPE parameters not shown on monitor screen
Ranked causes
- 1. Parameter not configured for display (common)
- 2. Parameter cable disconnected or loose (common)
- 3. Parameter device faulty (occasional)
- 4. Parameter cable damaged (occasional)
- 5. Parameter device physically damaged (occasional)
- 6. Patient block connector damaged (rare)
12 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No temperature displayed
Temperature measurement not showing on monitor screen
Ranked causes
- 1. Temperature channel not configured for display (common)
- 2. Incompatible temperature probe (occasional)
- 3. Faulty temperature probe (occasional)
- 4. Faulty CARESCAPE Parameter cable (occasional)
- 5. Faulty CARESCAPE Temperature module (occasional)
- 6. Temperature out of measurement range (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No invasive pressure displayed patient safety
Invasive pressure measurement not showing on screen
Ranked causes
- 1. Defective transducer (occasional)
- 2. Module not connected (common)
- 3. Transducer adapter cable not connected (common)
- 4. Invasive pressure channel not configured (common)
- 5. Invasive pressure channel not zeroed (common)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Troubleshooting CO2 calibration issues
Diagnostics for specific error messages and conditions encountered during Microstream CO2 calibration check or calibration procedure.
Ranked causes
- 1. Wrong gas message displayed (occasional)
- 2. Unstable gas flow message displayed (occasional)
- 3. Device is not ready message displayed (occasional)
- 4. Zero error message displayed during calibration (rare)
- 5. Calibration error message displayed (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ground leakage current exceeds limit high voltage
Ground leakage current measured greater than 1000 µA with NEUTRAL switch CLOSED or OPEN
Ranked causes
- 1. Defective power cord (common)
- 2. Incorrectly wired power outlet (occasional)
- 3. Incorrect test setup configuration (occasional)
- 4. Internal device leakage fault (rare)
Verify power cord integrity
- Test point
- Power cord surface and connectors
- Expected
- No wear or damage
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to turn on monitor when connected to CARESCAPE Dock F0
Monitor does not power on when connected to AC-powered CARESCAPE Dock F0
Ranked causes
- 1. Power cord is loose (common)
- 2. Power cord is faulty (occasional)
- 3. Power outlet does not meet specified requirements (occasional)
- 4. AC/DC power supply is faulty (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarms do not work patient safety
Monitor does not produce audible alarm sounds
Ranked causes
- 1. Audible alarms turned off (common)
- 2. Alarm volume too low (common)
- 3. Defective speaker (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Chassis leakage current to ePort exceeds limit high voltage
Chassis leakage current measured at ePort connection greater than 100 µA (NEUTRAL CLOSED/GND CLOSED or OPEN) or 500 µA (NEUTRAL OPEN)
Ranked causes
- 1. Defective power cord (common)
- 2. Incorrectly wired power outlet (occasional)
- 3. Internal device leakage fault (rare)
Perform chassis leakage test to ePort with NEUTRAL CLOSED and GND CLOSED
- Test point
- Chassis to ePort connection via safety test cable
- Expected
- Less than 100 µA, < 100 µA
- Tools
- Fluke 612/615/620 or equivalent; ePort safety test cable P/N per manual
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touchscreen inoperative
Touchscreen does not respond to touch input
Ranked causes
- 1. Touchscreen cable loose (common)
- 2. Faulty touchscreen sensor (occasional)
- 3. Touchscreen dirty (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not charging caution
Battery does not charge when inserted in monitor
Ranked causes
- 1. Battery connector damaged (occasional)
- 2. Battery defective (common)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Analog output caution
Analog output calibration adjusts ECG and invasive pressure analog outputs using a calibrated digital voltmeter after main board replacement or out-of-spec measurements.
Ranked causes
- 1. Main Board Assembly replaced (common)
- 2. Measured value not within specification limits (occasional)
Turn on CARESCAPE ONE and wait for normal screen
- Expected
- Normal screen appears
8 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Microstream CO2 gas caution
CO2 calibration adjusts measured CO2 value to show exactly 5% using reference calibration gas. Required after 1200 hours or 1 year post-installation, then every 4000 hours or annually.
Ranked causes
- 1. Calibration required per operating hour schedule (common)
- 2. Calibration required message displayed on screen (common)
Verify correct date and time on monitor
- Test point
- Monitor system clock
- Expected
- Current date and time displayed
11 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP patient safety
NIBP calibration must be performed when calibration check fails or measured values are outside specification limits.
Ranked causes
- 1. NIBP calibration check failed (common)
- 2. Measured value not within specification limits (occasional)
Turn on CARESCAPE ONE and wait for normal screen
- Expected
- Normal screen appears
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dock power supply or cord damaged high voltage
Power cord shows wear, damage, or continuity failure; external power supply not properly mounted or secured
Ranked causes
- 1. Power cord wear or physical damage (occasional)
- 2. Power supply not properly mounted or secured (occasional)
Inspect power cord for wear or damage
- Test point
- Power cord surface and connectors
- Expected
- No visible wear or damage
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor starts but display remains white
Monitor powers on but LCD display backlight illuminates without showing any content, no error messages
Ranked causes
- 1. Display cable is damaged or loose (common)
- 2. LCD display is faulty (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP not functioning patient safety
Non-invasive blood pressure measurement not working
Ranked causes
- 1. NIBP not calibrated (common)
- 2. Cuff pressure not maintained (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 reading not appearing on screen
SpO2 parameter value and plethysmographic waveform do not display after connecting sensor
Ranked causes
- 1. SpO2 sensor not detecting pulse (common)
- 2. SpO2 interconnect cable not connected (occasional)
SpO2 reading appears in parameter window
- Test point
- CARESCAPE ONE display
- Expected
- SpO2 value displayed
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormally low invasive pressure reading patient safety
Pressure measurement lower than expected in invasive pressure monitoring
Ranked causes
- 1. Transducer wrongly positioned (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot view feature despite enabled license
Feature not accessible on monitor even though license is enabled
Ranked causes
- 1. Software package does not include feature (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CARESCAPE SpO2-Masimo measurement values inaccurate patient safety
SpO2 measurement readings do not match expected values
Ranked causes
- 1. CARESCAPE SpO2-Masimo device faulty (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dock fasteners loose or insecure caution
Mounting solution fasteners not properly tightened or correctly secured on CARESCAPE Dock F0
Ranked causes
- 1. Fasteners not properly tightened (occasional)
Inspect mounting fasteners
- Test point
- All dock mounting fasteners
- Expected
- Intact, properly tightened, and correctly secured
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LoFlo CO2 accuracy test failed
CO2 value displayed does not match expected calibration gas concentration during accuracy check
Ranked causes
- 1. LoFlo Sidestream Module calibration drift (occasional)
Wait for LoFlo Sidestream Module warm-up
- Test point
- LoFlo Sidestream Module
- Expected
- Warm up time exceeded minutes
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LoFlo CO2 sample flow rate out of specification
Measured sample flow rate does not meet 50 ml/min ± 10 ml specification during functional check
Ranked causes
- 1. LoFlo Sidestream Module malfunction (common)
Wait for LoFlo Sidestream Module warm-up
- Test point
- LoFlo Sidestream Module
- Expected
- Warm up messages cleared minutes
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low audio volume or no audio
Speaker output is quiet or silent
Ranked causes
- 1. Speaker covers blocked by dust or debris (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Masimo rainbow measurements not displayed on CARESCAPE
Masimo rainbow measurement data not visible on CARESCAPE ONE screen
Ranked causes
- 1. Measurement only visible on host monitor when used as acquisition platform (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Masimo rainbow measurements not displayed on host monitor
Masimo rainbow measurement data not visible on host monitor
Ranked causes
- 1. Masimo rainbow measurement licenses not activated (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor starts with FRU software display
Monitor displays Field Replacement Unit, SW version, and IP address at startup instead of CARESCAPE software
Ranked causes
- 1. CARESCAPE software not installed after main board FRU replacement (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Parameter cables or strain reliefs damaged caution
Abrasion or damage observed on CARESCAPE Parameter cables or cable strain reliefs
Ranked causes
- 1. Cable wear or environmental abrasion (occasional)
Inspect all Parameter cables and strain reliefs
- Test point
- All cable lengths and connector areas
- Expected
- No abrasion or damage visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Parameter mounting bracket or latch damaged caution
CARESCAPE rSO2—INVOS or CO2—LoFlo mounting bracket broken or CARESCAPE SpO2—Nellcor latch damaged or missing
Ranked causes
- 1. Bracket or latch damaged or broken (occasional)
Inspect Parameter mounting brackets and latches
- Test point
- rSO2/CO2 mounting brackets and SpO2 Nellcor latch
- Expected
- Intact and undamaged
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Parameter not properly connected or mounted caution
CARESCAPE Parameter disconnected from CARESCAPE ONE or not properly mounted to Parameter Docks or mounting solutions
Ranked causes
- 1. Parameter disconnected or loose (occasional)
Verify Parameter connection to CARESCAPE ONE
- Test point
- Parameter connector and receptacle
- Expected
- Properly connected and latched
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to activate new monitor software
Cannot activate or upgrade monitor software; invalid activation code error
Ranked causes
- 1. Missing or invalid activation code (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to complete NIBP calibration
Confirm button inactive during NIBP calibration procedure
Ranked causes
- 1. Gain correction exceeds ±10% limit (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on
Start-up sequence does not advance beyond GE HealthCare logo screen; error messages may appear; unable to read software from memory
Ranked causes
- 1. Software not properly installed after main board replacement (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Visual damage or contamination detected caution
Cracks, breaks, dirt, loose connectors, corroded pins, bent pins, or fluid contamination on enclosure, display misalignment, or battery not fully inserted
Ranked causes
- 1. Physical damage or environmental exposure (unknown)
Inspect product labeling, markings and symbols
- Test point
- All visible labels
- Expected
- Intact and readable
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wrong software package activated
Incorrect software package enabled on monitor
Ranked causes
- 1. Wrong software package activated for device (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0x11CF | CARESCAPE CO2 device surface temperature too high (LoFlo devices) | CARESCAPE CO2 device surface temperature exceeds safe operating limit | warning |
0x11D5 | CARESCAPE CO2 device surface temperature too high (Microstream) | CARESCAPE CO2 Microstream device surface temperature exceeds safe operating limit | warning |
Acquiring signal / rSO2 Acquiring signal | rSO2 signal acquisition in progress from patient | Measurement startup sequence | info |
Amber LED flashing on, green LED off on CARESCAPE SpO2-Masim | Software failure or startup failure | Device powered but not starting or running properly | error |
Amber LED off, Green LED on on CARESCAPE SpO2-Masimo | Device running and operational | Device successfully started and communicating | info |
Amber LED on, Green LED off on CARESCAPE SpO2-Masimo | Device starting up | Power applied during initial startup sequence | info |
Analog output malfunction | CARESCAPE ONE analog output voltage failure | Hardware or software fault in analog output circuitry | error |
Apnea deactivated | Apnea alarm start-up conditions not reached | Apnea alarm conditions not met | info |
Auto detecting source | Automatic gas measurement source detection is ongoing | CO2 source detection in progress | info |
Battery empty! | Battery charge critically low, less than 5 minutes remaining | Battery state of charge drops below critical threshold | critical |
Battery failure | Battery is faulty and cannot hold charge | Battery self-test or charging circuit detects battery fault | error |
Battery low | Battery charge low, less than 20 minutes remaining | Battery state of charge drops below warning threshold | warning |
Battery temp high | Battery temperature error | Battery internal temperature exceeds safe operating limit | error |
Both amber and green LEDs off on CARESCAPE SpO2-Masimo | No power or software failure | Power not applied or critical software failure | critical |
Calibrated | Channel calibrated successfully | Completion of channel calibration | info |
Calibrating | Calibration of a channel is in progress | User initiates channel calibration | info |
Calibration error | Unsuccessful CO2 calibration | Failed calibration attempt | warning |
Calibration error | Calibration has failed | Unsuccessful calibration attempt | warning |
Calibration fail | Calibration of temperature channel has failed | Failed temperature calibration | error |
Calibration required | CARESCAPE CO2 — Microstream needs calibration | Calibration not performed or expired | warning |
Call Service: Text(s) missing | Software text file missing for current language | Text resource file not found or corrupted during language initialization | error |
Check CO2 adapter | Obstruction in LoFlo receptacle sample cell or cell not in optimal operating range | Sample cell obstruction or misalignment | warning |
Check device | CARESCAPE Pressure has malfunctioned | Hardware or communication fault detected | error |
Check device | Hardware or software fault in CARESCAPE Temperature | Cable fault, no updated data, data sample drift, or data out of range | error |
Check device / ECG malfunction | Hardware or software fault in CARESCAPE ECG | Cable fault, no updated data, data sample drift, or configuration failure | error |
Check device / Imped. malfunction | Hardware or software fault in CARESCAPE ECG impedance measurement | Cable fault, configuration failure, no updated data, or data sample drift | error |
Check device / rSO2 malfunction | Hardware or software fault in CARESCAPE rSO2 — INVOS; no rSO2 detected | Device fault; signal loss | error |
Check device / SpO2 malfunction | CARESCAPE SpO2 device malfunction; no SpO2 data detected | Device hardware or software fault; no signal detected | error |
Check gas flow | Sample flow rate is lower or higher than nominal due to obstruction or leak | Abnormal flow rate detected | warning |
Check NIBP | Systolic and/or diastolic results missing | Incomplete NIBP measurement | warning |
Check probe / Check SpO2 probe | No detectable SpO2 signal; sensor faulty or detached from patient | Signal loss; sensor disconnection; sensor malfunction; pulse rate lost (Nellcor); saturation update exceeds 30 seconds | error |
Check rSO2 labels | rSO2 measurement labels need verification after patient transport | Patient transport completed; label verification required | info |
Check sample line | Sample flow abnormal due to obstruction, cut, split, or saturated filter | Abnormal flow rate detected | warning |
Check SpHb | No adequate signal for SpHb measurement with CARESCAPE SpO2 — Masimo | Poor sensor connection or defective sensor | warning |
CO2 device failure: Call service | Front end processor startup failure, incorrect initialization, or no keep alive response | Device startup, initialization, or communication failure | error |
CO2 device malfunction | CARESCAPE CO2 device startup, runtime failure, or in software download mode | Device startup or runtime failure | error |
CO2 device overheated. Shutting down | CARESCAPE CO2 device surface temperature too high, measurement turned off. Error codes: 0x120F, 0x1215 | Temperature threshold exceeded | error |
CO2 inaccurate | Measured values outside range or inaccurate due to high respiration rate | High respiration rate or out-of-range measurements | warning |
CO2 measurement purging | CARESCAPE CO2 — Microstream self-cleaning blockage in sample line | Blockage detected in sample line | info |
CO2 measurement removed | CARESCAPE CO2 device has been disconnected from CARESCAPE ONE | Device disconnection | warning |
CO2 sensor overheated. Shutting down | CARESCAPE CO2 — Microstream sensor temperature too high, measurement off | Sensor temperature threshold exceeded | error |
CO2 sensor removed | LoFlo Sidestream Module disconnected from CARESCAPE CO2 — LoFlo | Module disconnection detected | warning |
CO2 zero required | CARESCAPE CO2 — LoFlo needs zeroing | Zeroing not performed or expired | warning |
Condition battery | Battery needs conditioning before use | Battery detected that requires initial conditioning cycle | warning |
Configuration error(s) | One or more configuration errors detected | Configuration validation failure at startup or during operation | error |
Continuous blockage. Check sample line. | CARESCAPE CO2 — Microstream sampling line is blocked | Sustained blockage detected | warning |
CS ONE Faulty Device Connected | Connected CARESCAPE Parameter has communication failure | Communication error detected between monitor and parameter | error |
CS ONE Faulty Device in Port #x | CARESCAPE Parameter in port x has detected failure | Over current condition, repeated over temperature faults, or repeated communication failures | error |
CS ONE Speaker Failure | Alarm speaker is inoperative | Speaker hardware failure or audio output circuit fault | error |
CS ONE Unknown Device in Port #x | Unauthorized device connected to CARESCAPE ONE USB port | Unrecognized USB device detected on one of eight ports | warning |
Cuff loose | Loose cuff or cuff hose | Loose connection detected | warning |
Cuff occlusion | Occlusion during measurement or overpressured cuff | Obstruction detected during NIBP measurement | warning |
Cuff overpressure | NIBP cuff exceeded pressure safety limits | Cuff squeezed during measurement exceeding limits | warning |
Demo mode / SpO2 Demo mode | CARESCAPE SpO2 — Masimo is operating in demonstration mode | Demo mode enabled | info |
Device not ready | Device is still warming up or there is a technical condition | Device startup or initialization in progress | info |
Device overheating | CARESCAPE ECG surface temperature is too high | Inadequate ventilation or obstruction of ECG device | warning |
ECG analog output malfunction | Hardware or software fault in CARESCAPE ECG analog output | ECG analog output circuit failure or data communication error | error |
ECG device failure: Call service | CARESCAPE ECG startup or initialization failure | Device unable to initialize or start up | critical |
ECG device overheated. Shutting down | CARESCAPE ECG surface temperature exceeded safe operating limits; device is being disabled | Surface temperature too high; monitor automatically shuts down ECG to prevent damage | error |
ECG measurements removed | CARESCAPE ECG has been disconnected or removed from system | ECG device physically removed or connection lost | warning |
Expiring cable | CARESCAPE SpO2 — Masimo interconnect cable approaching expiration date | Cable near end of service life | warning |
Expiring probe | CARESCAPE SpO2 — Masimo sensor or adhesive sensor approaching expiration date | Sensor near end of service life | warning |
Faulty cable / SpO2 faulty cable | CARESCAPE SpO2 — Masimo interconnect cable is defective | Cable malfunction detected by device | error |
Faulty probe / SpO2 faulty probe / Faulty adhesive SpO2 prob | SpO2 sensor or adhesive sensor has failed | Sensor malfunction detected | error |
Identical CO2 device | More than one CARESCAPE CO2 device in the system. Error codes: 0x110F, 0x1115 | Multiple CO2 devices detected | warning |
Identical ECG device | Two CARESCAPE ECG devices are connected to the system | Multiple ECG modules detected | warning |
Identical pressure devices | Three or more CARESCAPE Pressures connected | Multiple pressure devices detected | warning |
Identical rSO2 devices | Two or more CARESCAPE rSO2 — INVOS devices are connected | Multiple rSO2 devices detected | warning |
Identical SpO2 device | Multiple CARESCAPE SpO2 devices are connected | More than one SpO2 device detected | warning |
Identical Temp device | Two CARESCAPE Temperatures connected. Error code: 0x110D | Multiple temperature devices detected | warning |
Incompatible cable / Incompatible SpO2 cable | Incompatible or unrecognized interconnect cable connected to CARESCAPE SpO2 — Masimo | Unrecognized cable detected | error |
Incompatible probe / Incompatible SpO2 probe / Incompatible | Sensor is not compatible or not recognized by CARESCAPE SpO2 — Masimo | Incompatible or unrecognized sensor detected | error |
Incompatible rSO2 device | CARESCAPE rSO2 — INVOS is not compatible with system | Device compatibility check failed | error |
Incompatible SpO2 device | CARESCAPE SpO2 device startup, initialization failure, or hardware fault | Device incompatibility or initialization failure detected | error |
Insufficient power for ECG device | Not enough power available to operate CARESCAPE ECG | Power supply insufficient for ECG operation | error |
Insufficient power for rSO2 device | Not enough power available for CARESCAPE rSO2 — INVOS operation | Insufficient power supply | error |
Insufficient power for SpO2 device | Not enough power available for CARESCAPE SpO2 device operation | Insufficient power supply | error |
IP analog output malfunction | Hardware or software fault in CARESCAPE Pressure analog output | Pressure analog output circuit failure or data communication error | error |
Long meas. time | The measurement time is long | Measurement duration exceeds threshold | warning |
Measurement error | Technical fault in monitor measurement circuitry | Hardware fault detected in measurement subsystem | error |
Measurement off / OFF | Impedance respiration measurement is disabled or not active | ECG leads not connected to patient or measurement turned off from setup menu | info |
Move SpO2 device to a cooler location | CARESCAPE SpO2 device surface temperature is too high | Excessive device temperature detected | warning |
NIBP auto stopped | NIBP Auto mode stopped due to loose cuff, hose, communication failure, or cuff size change | Detection of loose cuff, hose, module communication failure, or cuff size change | warning |
NIBP cuff loose | Loose cuff or cuff hose | Loose connection detected | warning |
NIBP cuff occlusion | Occlusion during measurement or overpressured cuff | Obstruction detected during NIBP measurement | warning |
NIBP hardware fault in CARESCAPE ONE | NIBP hardware fault detected | Hardware malfunction in NIBP module | error |
NIBP malfunction | NIBP hardware fault in CARESCAPE ONE | Hardware malfunction detected | error |
NIBP STAT stopped | STAT mode stopped due to loose cuff, hose, communication failure, or cuff size change | Detection of loose cuff, hose, module communication failure, or cuff size change | warning |
No battery backup | No battery installed in CARESCAPE ONE | Battery compartment empty or battery not detected on startup | warning |
No cable / No SpO2 cable | Interconnect cable not connected to CARESCAPE SpO2 — Masimo | Missing or disconnected cable | error |
No probe / No SpO2 probe / No adhesive SpO2 probe | Sensor not connected to CARESCAPE SpO2 device | Sensor disconnection | error |
Over range | CO2 value outside measurable range | CO2 concentration exceeds measurement capability | error |
Over scale | Gas signal exceeds maximum waveform area | Gas concentration signal exceeds upper limit of current waveform scale | warning |
P1 standby to P4 standby | Invasive pressure channel set to standby | IP channel standby mode activated | info |
P1 zeroing failed to P4 zeroing failed | Channel zeroing not successful | Failed zeroing attempt | warning |
Powering down! | Battery charge critically low, less than 1 minute remaining | Battery state of charge at critically low level | critical |
Pressure device failure: Call service | Faulty CARESCAPE Pressure detected | Hardware failure detected | error |
Pressure device initializing | New software being downloaded to CARESCAPE Pressure | Software download in progress | info |
Pressure device overheated. Shutting it down | CARESCAPE Pressure turned off due to excessive temperature | Surface temperature threshold exceeded | error |
Pressure measurement removed | CARESCAPE Pressure removed or active pressure channel became inactive | Pressure device disconnected or channel deactivated | warning |
Pressure Sensed | Pressure has been sensed during zeroing | Pressure detected during zeroing procedure | warning |
Probe initializing | CARESCAPE SpO2 — Masimo sensor is initializing | Sensor initialization in progress | info |
PX not zeroed | Invasive pressure channel not zeroed | Invasive pressure channel not zeroed before measurement | warning |
Replace battery | Battery capacity degraded below acceptable threshold | Battery full capacity less than 60% of new battery capacity | warning |
Replace cable / Replace SpO2 cable | CARESCAPE SpO2 — Masimo interconnect cable has expired | Cable expiration date reached | error |
Replace probe / Replace SpO2 probe / Replace adhesive SpO2 p | CARESCAPE SpO2 — Masimo sensor or adhesive sensor has expired | Sensor expiration date reached | error |
Replace X / rSO2 X replace sensor | rSO2 sensor is defective or not compatible | Sensor malfunction or incompatibility detected | error |
rSO2 device failure: Call service | CARESCAPE rSO2 — INVOS startup or initialization failure | Device initialization failure; startup error | critical |
rSO2 measurement disconnected | CARESCAPE ONE disconnected from host monitor while rSO2 — INVOS connected to acquisition platform | Host monitor connection lost | warning |
rSO2 measurement removed | CARESCAPE rSO2 — INVOS has been disconnected from acquisition platform | rSO2 device disconnection | warning |
Sample line blocked | CARESCAPE CO2 Microstream sampling line is blocked | Obstruction detected in sampling line preventing gas flow | error |
Sample line disconnected | Sample line is disconnected from device | Disconnection of sampling line from sensor or monitor port | error |
Select age | Patient age must be selected before CO2 measurement | Attempt to start CARESCAPE CO2 Microstream measurement without age selection | info |
Service CO2 sensor | Problem detected with CARESCAPE CO2 device | Internal fault detected in CO2 measurement device | error |
Service CS ONE 0xHOST 3001 | Internal temperature of CARESCAPE ONE out of specification | One of internal temperature sensors indicates out-of-range temperature | error |
Service CS ONE 0xHOST 3002 | Internal supply voltage out of specification | One or more internal power supply voltages exceed specification limits | error |
Service CS ONE 0xHOST 3004 | Disk utilization exceeded 90% | Storage partition usage exceeds 90% capacity | warning |
Service CS ONE 0xHOST 3100 | Main board battery voltage out of specification | Battery backup voltage on main board below or above acceptable range | error |
Service CS ONE 0xPDM1000 | Data in secure storage corrupt | Checksum or validation failure of stored configuration data | error |
Service CS ONE 0xPDM1002 | Error occurred writing data to secured storage | Write operation to secure storage failed | error |
Service CS ONE 0xPDM1200 | Initialization of platform security library returned invalid directory error | Security library initialization failure during startup | error |
Service CS ONE activation failed | Settings activation failed | Configuration or settings activation process encountered error | error |
Service CS ONE Calibrate NIBP | NIBP calibration values could not be read from secure storage | NIBP calibration data missing or unreadable from storage | error |
Service CS ONE ECG Analog Out | ECG calibration values could not be read from secure storage | ECG analog output calibration data missing or unreadable | error |
Service CS ONE Faulty Port #x | Fault detected on CARESCAPE Parameter port | Hardware failure detected on monitor port | error |
Service CS ONE IP Analog Out | Invasive pressure calibration values could not be read from storage | IP calibration data missing or unreadable from storage | error |
Service CS ONE Power Failure | Lost communication with power management controller | Communication failure between main CPU and power management module | error |
SpO2 device failure: Call service | CARESCAPE SpO2 device startup or initialization failure | Device unable to start or initialize; possible hardware failure | critical |
SpO2 device overheated Shutting down | CARESCAPE SpO2 surface temperature too high; CARESCAPE ONE is disabling the device | Excessive surface temperature detected; monitor auto-shutoff activated | error |
SpO2 measurement removed | CARESCAPE SpO2 device has been disconnected from system | SpO2 device disconnection | warning |
SW download failed on CS ONE port x | Software update to CARESCAPE Parameter failed | Software transfer to connected parameter failed or was interrupted | error |
SW update in progress on CS ONE port x | Software update in progress to connected CARESCAPE Parameter | Firmware update transfer from CARESCAPE ONE to parameter active | info |
Temp device failure: Call service | CARESCAPE Temperature startup or initialization failure. Error codes: 0x108D, 0x124D | Device startup or initialization failure | error |
Temp device overheated. Shutting down | CARESCAPE Temperature turned off due to excessive temperature. Error code: 0x120D | Surface temperature threshold exceeded | error |
Temp measurement removed | CARESCAPE Temperature removed or active temperature channel became inactive | Temperature device disconnected or channel deactivated | warning |
Unable to read licenses | License file could not be read | License file not found, corrupted, or unreadable | error |
Unstable gas flow | Gas flow is unstable during calibration | Variation or instability detected in calibration gas flow | warning |
Unsupported X / rSO2 X unsupported sensor | rSO2 sensor type is not supported by system | Unsupported sensor detected | error |
Warming up | CARESCAPE CO2 device is in warm-up phase | Device startup or power-on initialization sequence | info |
Wrong gas | Incorrect calibration gas in use | Detected gas composition does not match expected calibration gas | error |
X excess light / rSO2 excess light | Excessive light detected on rSO2 sensor | High ambient light conditions detected at sensor | warning |
X interference / rSO2 interference | rSO2 measurement is disturbed by electromagnetic interference | Electromagnetic noise or interference detected | warning |
X no sensor / rSO2 X sensor disconnected | rSO2 sensor has been disconnected from system | Sensor disconnection detected | error |
X poor signal / rSO2 poor signal | rSO2 measurement cannot be performed due to poor signal quality | Insufficient signal quality detected | warning |
X sensor off / rSO2 X sensor off | rSO2 sensor is not connected to patient | Sensor not in contact with patient | warning |
Zeroed | Zeroing was successful | Completion of zeroing procedure | info |
Zero ICP separately | ICP channel must be zeroed separately from other invasive pressures | Attempt to zero ICP with other invasive pressures | warning |
Zeroing | Invasive pressure channel currently being zeroed | Zeroing procedure in progress | info |
Zeroing | Zeroing procedure in progress | Zeroing calibration sequence active | info |
Zeroing error | Zeroing failed during calibration check process | Zeroing procedure unable to complete successfully | error |
Zero OK | Zeroing was successful | Completion of zeroing procedure | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.