GE Healthcare Carescape Modules: troubleshooting and error codes

The ServiceForge knowledge base holds 343 documented symptoms for the GE Healthcare Carescape Modules, with 553 ranked causes, 489 diagnostic checks with test points and expected values, and 117 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

NIBP hardware error

Parameter window message 'Call Service: Error x' indicating NIBP module hardware fault; message stays on screen until problem is fixed

Ranked causes

  1. 1. RAM failure (error x=0) (rare)
  2. 2. ROM checksum failure (error x=1) (rare)
  3. 3. +15V supply failure (error x=2) (rare)
  4. 4. -15V supply failure (error x=3) (rare)
  5. 5. ADC error (error x=6) (rare)
  6. 6. Watchdog timer too short (error x=7) (rare)
  7. 7. Watchdog timer too long (error x=8) (rare)
  8. 8. Watchdog activated (error x=9) (rare)
  9. 9. EEPROM checksum error (error x=10) (rare)
  10. 10. Autozero range exceeded (error x=11) (rare)
  11. 11. Communication watchdog activated (error x=12) (rare)
  12. 12. Too early autocycle start (error x=14) (rare)

12 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP air leakage patient safety

Alarm or parameter window message indicating air is leaking from cuff, hose, internal tubes, or module components during pressure measurement

Ranked causes

  1. 1. Cuff physically damaged (occasional)
  2. 2. Cuff or hose connector damaged or O-ring missing (occasional)
  3. 3. NIBP cuff hose double connector damaged (rare)
  4. 4. Leakage in patient connector panel tubing (occasional)
  5. 5. Internal module tubing leakage (rare)
  6. 6. Tube disconnected or damaged (occasional)
  7. 7. Manifold leakage (rare)
  8. 8. Valves or connecting tubes damaged (rare)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Module not appearing on monitor display

Parameter module connected to frame but measurements not shown on display; either specific module data missing or all non-PSM/PDM modules not visible

Ranked causes

  1. 1. Module not connected properly (common)
  2. 2. Module connector defective (occasional)
  3. 3. EMBC internal connectors defective or loose (occasional)
  4. 4. EMBC board defective (occasional)
  5. 5. Cable between frame and host not connected or defective (occasional)
  6. 6. DC/DC board defective (occasional)
  7. 7. Monitor display not configured for parameter (common)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unstable O2 delivery affecting VO2 calculation patient safety

VO2 readings invalid or unreliable due to unstable oxygen delivery or gas mixer issues

Ranked causes

  1. 1. Unstable O2 delivery or gas mixer malfunction (occasional)
  2. 2. Respiratory rate over 35/min (occasional)
  3. 3. Reference gas inlet port blocked (occasional)
  4. 4. Gas sampling line longer than 2 m (occasional)
  5. 5. Dead space of Y-piece greater than 8 ml (occasional)
  6. 6. Gas sampling line connected to HME instead of D-lite (occasional)
  7. 7. D-lite incorrectly placed in circuit (occasional)
First check

Display oxygram and assess curve stability

Test point
Monitor VO2 waveform display
Expected
Stable, consistent oxygram curve

4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Measurement(s) removed from display

Alarm area shows 'XXXX measurement(s) removed'; module data no longer visible on monitor

Ranked causes

  1. 1. Module not connected properly (common)
  2. 2. Module connector defective (occasional)
  3. 3. Frame not properly connected to host (occasional)
  4. 4. EMBC board connectors loose inside frame (occasional)
  5. 5. EMBC board defective (occasional)
  6. 6. DC/DC board defective (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP cuff loose or disconnected patient safety

Alarm or parameter window message indicating blood pressure cuff is not properly connected, loosely wrapped, or has internal air leakage

Ranked causes

  1. 1. Cuff or hose not connected to module (common)
  2. 2. Cuff loosely wrapped on patient arm (common)
  3. 3. Air leakage in shielding or connectors (occasional)
  4. 4. Cuff or hose air leakage (occasional)
  5. 5. Internal module tubing leak (rare)
  6. 6. NIBP pump failure (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP weak pulsation signal patient safety

Parameter window message indicating weak oscillation or poor signal quality due to weak blood circulation, improper cuff position, or small air leakage

Ranked causes

  1. 1. Weak or abnormal blood circulation (common)
  2. 2. Improper cuff position or attachment (common)
  3. 3. Slow heart rate with artifacts (occasional)
  4. 4. Diastolic pressure measurement problems (occasional)
  5. 5. Small air leakage (occasional)
  6. 6. Insufficient measurement data due to artifacts (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gas sensor indicates INOP patient safety

Waveform field or parameter window displays INOP; gas measurement unavailable from respiratory module

Ranked causes

  1. 1. MiniOM unit internal failure (temperature, supply voltage, other) (common)
  2. 2. MiniTPX unit failure (temperature, supply voltage, IR lamp) (common)
  3. 3. MiniPVX unit failure (pressure sensor, supply voltage) (occasional)
  4. 4. CPU failure (supply voltage, pressure sensor, A/D converter) (occasional)
  5. 5. Other failures (fan, pump, valve, zeroing, CO2 reference) (occasional)
First check

Check flex cable connection

Test point
Module flex cable
Expected
Secure connection

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low gas sample flow rate patient safety

Sample flow below 80% of module nominal value, affecting measurement accuracy

Ranked causes

  1. 1. Gas sampling line blocked (common)
  2. 2. Water trap blocked or full (common)
  3. 3. Sample gas output blocked (common)
  4. 4. Internal tubing blocked (occasional)
  5. 5. Pump unit failure (occasional)
First check

Measure sample flow rate in calibration menu

Test point
Flow rate display
Expected
At least 80% of module nominal flow value

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No temperature displayed on monitor patient safety

Temperature parameter window not showing any temperature value or waveform; measurement completely absent from display

Ranked causes

  1. 1. Temperature channel not configured to screen display (common)
  2. 2. Incompatible temperature probe installed (occasional)
  3. 3. Temperature probe type setting (STP/TP/ST) incorrect (occasional)
  4. 4. Faulty temperature probe (occasional)
  5. 5. Temperature measurement out of range (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Too high Ppeak patient safety

Peak inspiratory pressure measurement reads abnormally elevated

Ranked causes

  1. 1. Bronchospasm (common)
  2. 2. Patient coughing (common)
  3. 3. Patient breathing against ventilator (common)
  4. 4. Obstruction in airways (occasional)
  5. 5. HME obstructed (occasional)
First check

Assess for acute bronchospasm

Test point
Patient clinical status and breath sounds
Expected
Normal breath sounds, no wheezing or stridor

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check sample line message

Message displayed in parameter window or alarm area indicating sampling line issue.

Ranked causes

  1. 1. Water, mucus, or contamination in sampling line (common)
  2. 2. Sampling line crimped, pinched, or damaged (common)
  3. 3. Gas outlet port obstructed (occasional)
  4. 4. Pressure changes or gas leak (occasional)
First check

Visual inspection for contamination in sampling line

Test point
Sampling line interior
Expected
Clear, no water, mucus, or debris visible

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Communication LED not lit

Communication LED indicator on module does not illuminate

Ranked causes

  1. 1. ePort to host interface cable damaged (common)
  2. 2. ePort connector on monitor damaged (occasional)
  3. 3. ePort interface on module damaged (occasional)
  4. 4. Module defective (rare)
First check

Inspect ePort cable

Test point
ePort to host interface cable
Expected
No visible damage, cuts, or bent pins

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gas sampling line blocked patient safety

Blockage in sampling system preventing gas flow from patient to sensor

Ranked causes

  1. 1. External gas sampling line blocked (common)
  2. 2. Water trap occluded (common)
  3. 3. Internal tubing blocked (common)
  4. 4. Sampling pump failure (occasional)
First check

Inspect external gas sampling line

Test point
Patient sampling line tubing
Expected
No visible blockages, kinks or condensation

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inspired tidal volume exceeds expired tidal volume (insp TV > exp TV) patient safety

Spirometry indicates inspired volume greater than expired volume; leak suspected in lungs or breathing circuit

Ranked causes

  1. 1. Leak in patient lungs, endotracheal tube cuff, or spirometry system (common)
  2. 2. Water inside D-lite or tubing (occasional)
  3. 3. Another side-stream gas sampling line between D-lite and patient causing bypass (occasional)
  4. 4. D-fend Pro filter leaking (occasional)
First check

Perform spirometry leak test

Test point
Spirometry circuit with patient connected
Expected
No leak detected; TVexp ≥ TVinsp

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP cuff occlusion patient safety

Parameter window message indicating air path from cuff or tubing is blocked or kinked, preventing proper inflation and measurement

Ranked causes

  1. 1. Cuff or hose tubing kinked or occluded (common)
  2. 2. Pressure transducer failure (occasional)
  3. 3. A/D converter fault (occasional)
  4. 4. NIBP calibration fault (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No invasive pressure signal patient safety

No invasive pressure waveform or values displayed on monitor; measurement completely absent

Ranked causes

  1. 1. Defective invasive pressure transducer (occasional)
  2. 2. Invasive pressure module not connected to monitor (common)
  3. 3. Transducer adapter cable not connected to module (common)
  4. 4. Invasive pressure channel not configured or displayed (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No parameter data from module

Parameter window shows no data output from module due to cable or interface issue

Ranked causes

  1. 1. Parameter cable disconnected (common)
  2. 2. Parameter interface damaged (occasional)
  3. 3. Related parameter license not activated (occasional)
  4. 4. Monitor not configured to display parameter (occasional)
First check

Verify cable connection

Test point
Parameter cable connector
Expected
Cable fully seated and tight

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No power at module

Module does not power on when connected to monitor

Ranked causes

  1. 1. ePort cable damaged (common)
  2. 2. Monitor ePort connector damaged (occasional)
  3. 3. Module ePort interface damaged (occasional)
  4. 4. Module battery not charging (occasional)
First check

Inspect ePort cable integrity

Test point
ePort to host interface cable
Expected
No damage visible

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No pressure measurement available patient safety

No invasive pressure waveform or value displayed. No signal detected.

Ranked causes

  1. 1. Defective transducer (occasional)
  2. 2. Module not connected (common)
  3. 3. Transducer adapter cable not connected to module (common)
  4. 4. Invasive pressure channel not configured on display (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pulse search NIBP measurement patient safety

Parameter window displays pulse search message during NIBP initialization or due to sensor/cable fault

Ranked causes

  1. 1. Normal NIBP measurement startup (common)
  2. 2. Defective or damaged sensor or cable (occasional)
  3. 3. Sensor off patient (common)
  4. 4. Repeatable pulse detection failed (occasional)
First check

Wait for pulse detection

Test point
NIBP measurement display
Expected
Waveform and pressure values displayed

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sample gas outflow blocked patient safety

Exhaust line from module is occluded, preventing return of sampled gas to patient or scavenging system

Ranked causes

  1. 1. Sample gas out connector or exhaust line blocked (common)
  2. 2. Return tubing to patient circuit occluded (occasional)
  3. 3. Scavenging system not open to room pressure (occasional)
  4. 4. Internal tubing blocked or pump failure (occasional)
First check

Inspect sample gas out connector and exhaust line

Test point
Front panel outlet and exhaust tubing
Expected
No visible blockages or obstructions

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sample line blocked patient safety

Continuous blockage message in parameter window, waveform field, or alarm area

Ranked causes

  1. 1. Gas sampling line blocked (common)
  2. 2. Water trap container full (common)
  3. 3. Water trap occluded (occasional)
  4. 4. Internal tubing blocked (occasional)
First check

Inspect external sampling line

Test point
External gas sampling line
Expected
Clear, no blockages

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Strongly vibrating loop patient safety

Loop display shows excessive vibration or oscillation, volumes appear too large or too small

Ranked causes

  1. 1. Mucus in endotracheal tube (common)
  2. 2. Water or secretions in hoses or D-lite (common)
  3. 3. Wrong mode or sensor selection (occasional)
  4. 4. Incorrect sensor type selected (occasional)
First check

Assess patient airway status and secretion level

Test point
Patient ET tube and airway
Expected
Patent airway, minimal secretions

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Too high Raw patient safety

Airway resistance measurement reads abnormally high, indicating increased airway obstruction

Ranked causes

  1. 1. Kink in tubing (common)
  2. 2. Mucus in airway (common)
  3. 3. Asthmatic patient or bronchospasm (occasional)
  4. 4. Spontaneous or patient-triggered breaths with breathing effort against ventilator (occasional)
First check

Visually inspect all tubing for kinks or compression

Test point
Entire spirometry and patient circuit tubing
Expected
Straight, unobstructed tubing; free flow

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Water trap blockage or leakage patient safety

Messages appear for water trap issues, sampling system not functioning properly

Ranked causes

  1. 1. Water trap not connected (common)
  2. 2. Air leak inside internal tubing (common)
  3. 3. Water trap container full (occasional)
  4. 4. Defective or contaminated Mini D-fend water trap (occasional)
First check

Inspect water trap connection and condition

Test point
Water trap connector on front panel
Expected
Water trap properly connected, no visible damage

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery not charging

Module battery does not charge when connected to charger or docking station

Ranked causes

  1. 1. Battery tray damaged (occasional)
  2. 2. Battery connector damaged (occasional)
  3. 3. Battery defective (occasional)
First check

Inspect battery tray

Test point
Battery compartment tray
Expected
Tray undamaged and properly seated

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration error

Calibration error message in parameter window or calibration menu

Ranked causes

  1. 1. Calibration gas not fed within 1 minute after zeroing (common)
  2. 2. Calibration failed due to large gain adjustment (occasional)
  3. 3. Wrong calibration gas used (common)
First check

Verify calibration gas supply started within 1 minute

Test point
Gas supply timing
Expected
Gas fed within 60 seconds of zeroing completion

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration error - gas not supplied or gain adjustment issue

Calibration failed due to calibration gas not supplied within 1 minute or excessive gain adjustment

Ranked causes

  1. 1. Calibration gas not supplied within 1 minute of zeroing (common)
  2. 2. Calibration failed due to excessive gain adjustment (occasional)
  3. 3. Wrong calibration gas used (occasional)
First check

Verify correct calibration gas bottle connected and pressurized

Test point
Calibration gas supply at module inlet
Expected
Specified calibration gas connected and pressure available

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration failed patient safety

Message appears in parameter window. Pressure calibration did not complete successfully.

Ranked causes

  1. 1. Pulsating waveform detected during calibration (occasional)
  2. 2. Gain beyond limits (±20% of default) (occasional)
  3. 3. Defective transducer (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration not available during alarm condition

Gas calibration unavailable when alarm conditions present (sample line blocked, D-fend check, gas out check)

Ranked causes

  1. 1. Sample line blocked alarm active (common)
  2. 2. D-fend check alarm active (common)
  3. 3. Sample gas out check alarm active (common)
First check

Check for active alarms in alarm area

Test point
Alarm area display
Expected
No active alarm messages; clear alarms before calibration

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Invasive pressure patient safety

Module invasive pressure measurement requires calibration when transducer is replaced, calibration check fails, or reading exceeds specification limits.

Ranked causes

  1. 1. New pressure transducer installed (common)
  2. 2. Previous calibration check failed (occasional)
  3. 3. Measured value outside specification limits (occasional)
First check

Verify module is connected to monitor

Test point
Module connection port
Expected
Module actively connected

10 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG waveforms not displayed or noisy patient safety

ECG leads II, V1, aVL and all 12-lead waveforms not appearing or contain excessive noise

Ranked causes

  1. 1. Leadwire not properly connected or disconnected (common)
  2. 2. ECG signal path noise or interference (occasional)
  3. 3. Simulator not configured correctly or not generating ECG signal (occasional)
First check

Verify ECG leads II, V1, aVL display with waveforms

Test point
ECG waveform fields on monitor screen
Expected
All three leads displayed simultaneously with noise-free waveforms and 80±5 bpm heart rate, ±5 bpm
Tools
Multiparameter patient simulator, 12-lead ECG trunk cable, 5-leadwire sets

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Expired volume much smaller than inspired volume (MVexp<<MVinsp) patient safety

Spirometry indicates expired volume ≤70% of inspired volume; likely leak in breathing circuit

Ranked causes

  1. 1. Leaking or occluded spirometry tube (common)
  2. 2. Water in spirometry tubing (occasional)
  3. 3. MiniPVX sensor failure (occasional)
First check

Perform spirometry leak test

Test point
Spirometry circuit
Expected
No leak detected; TVexp ≥70% TVinsp

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fluctuating Raw patient safety

Airway resistance (Raw) value fluctuates or varies abnormally during breath cycle

Ranked causes

  1. 1. Mucus in airways or tubings (common)
  2. 2. Breathing effort against ventilator; patient-triggered or spontaneous breaths (common)
  3. 3. Ventilator expiratory valve causing fluctuations during expiratory flow (occasional)
First check

Inspect spirometry circuit for secretions

Test point
Tubing and D-fend Pro sensor
Expected
Clear of mucus and secretions

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invasive pressure calibration failure patient safety

Automatic calibration did not complete; timeout or error condition detected during pressure calibration

Ranked causes

  1. 1. Pulsating waveform during calibration (common)
  2. 2. Gain beyond limits (outside ±20% of default) (occasional)
  3. 3. Defective transducer (occasional)
First check

Verify static pressure applied during calibration

Test point
Manometer reading during calibration
Expected
100-300 mmHg static pressure (no pulsation)

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loop overscale patient safety

Monitored volumes displayed less than set volumes, loop scaling appears incorrect

Ranked causes

  1. 1. Wrong scale selected (common)
  2. 2. Wrong TV base selected (common)
  3. 3. Leak between ventilator and D-lite (occasional)
First check

Verify volume scale setting on display

Test point
Monitor display scaling menu
Expected
Scale matches current tidal volume range

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Measurement off (impedance respiration)

Parameter window or waveform displays 'Measurement off'; ECG leads not connected or measurement disabled in setup menu, or ECG source is telemetry

Ranked causes

  1. 1. ECG leads disconnected from patient (common)
  2. 2. Impedance respiration measurement disabled in setup menu (occasional)
  3. 3. ECG source is telemetry (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Module not functioning

Module fails to respond to key press, parameters not displayed, or communication error indicated

Ranked causes

  1. 1. Loose keypad cable or internal module connection (common)
  2. 2. Module incompatibility or configuration error (occasional)
  3. 3. Defective accessory or sensor (occasional)
First check

Visual inspection of module and connectors

Test point
Front cover, front panel, connectors, module box and latch
Expected
Intact, clean, properly attached

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP not functioning patient safety

NIBP measurement does not operate or calibration cannot be completed

Ranked causes

  1. 1. NIBP not calibrated (common)
  2. 2. Cuff pressure not maintained (leak) (common)
  3. 3. NIBP assembly malfunction (occasional)
First check

Check calibration status

Test point
NIBP module status
Expected
Calibration completed within acceptable timeframe

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No EntEEG waveform or Entropy display on screen patient safety

Entropy module not connected or parameter not configured/visible on display.

Ranked causes

  1. 1. Entropy module not connected (common)
  2. 2. Parameter not configured or priority too low (occasional)
  3. 3. Entropy not supported by active software (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No pressure reading displayed patient safety

Invasive pressure parameter window shows no value; pressure waveform absent from display

Ranked causes

  1. 1. Defective pressure transducer (occasional)
  2. 2. Module not connected (common)
  3. 3. Transducer adapter cable not connected (common)
First check

Verify module is connected to monitor

Test point
module connectors
Expected
Module fully seated on monitor

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Out of range VO2 or VCO2 in gas exchange patient safety

Parameter window displays Out of range; VO2 or VCO2 measurement is <0 or >999 ml/min and exceeds measurement range

Ranked causes

  1. 1. Gas sampling or spirometry lines incorrectly connected to airway or module (common)
  2. 2. Incorrect sensor type (D-lite vs Pedi Lite) selected in monitor menu (occasional)
  3. 3. FiO2 recently changed; VO2 out of range temporarily (occasional)
First check

Verify gas sampling and spirometry line connections

Test point
Airway connectors and module ports
Expected
All lines securely connected to airway and gas module

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Response signal too weak

Response too weak message in parameter window; maximum gain insufficient to amplify response signal to measurable level

Ranked causes

  1. 1. Stimulus electrodes loose or incorrect placement (common)
  2. 2. Response electrodes disconnected (common)
  3. 3. Dry or degraded electrodes (occasional)
First check

Verify stimulus electrode placement and tightness

Test point
Stimulus electrode connections
Expected
Electrodes firmly seated on skin at correct anatomical location

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor INOP patient safety

Sensor INOP message in waveform field

Ranked causes

  1. 1. IR lamp failure (occasional)
  2. 2. Ambient pressure too high or low (occasional)
  3. 3. CPU failure or module communication issue (rare)
First check

Check miniTPX flex cable connection

Test point
miniTPX flex cable connector
Expected
Securely connected

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor INOP (inoperative)

Gas sensor inoperative; sensor not responding to gas measurement

Ranked causes

  1. 1. Ambient pressure too high or low (occasional)
  2. 2. No response from gas module (occasional)
  3. 3. High temperature inside module or EEPROM checksum failure (rare)
First check

Check ambient pressure reading in gas calibration menu

Test point
Calibration menu ambient pressure display
Expected
Pressure within normal operating range (typically 600-800 mmHg)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor INOP in spirometry (MiniPVX) patient safety

Spirometry waveform field displays INOP; no response from MiniPVX measurement board or EEPROM checksum failure

Ranked causes

  1. 1. MiniPVX flex cable loose or disconnected (occasional)
  2. 2. Zero valve broken or internal failure (occasional)
  3. 3. EEPROM checksum failure or measurement board failure (rare)
First check

Verify MiniPVX flex cable connection

Test point
MiniPVX flex cable connector
Expected
Fully seated, secure

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Strange or nonphysiological VO2 readings patient safety

Gas exchange (VO2) values appear abnormal or inconsistent with patient physiology

Ranked causes

  1. 1. Ventilation mode BiPAP/CPAP with high continuous bypass flow (occasional)
  2. 2. Presence of N2O or anesthetic agents in ICU applications (occasional)
  3. 3. Gas exchange not measurable (occasional)
First check

Select oxygram and verify curve stability

Test point
Monitor VO2 waveform display
Expected
Stable oxygram curve consistent with patient condition

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Too low ETCO2 reading patient safety

Measured end-tidal CO2 value abnormally low compared to expected clinical range

Ranked causes

  1. 1. Leak in sampling system (common)
  2. 2. Calibration error (common)
  3. 3. High by-pass flow from ventilator (occasional)
First check

Check all gas connections for leaks

Test point
Water trap, sampling line, outlet connections
Expected
No hissing or leakage detected; tight connections

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Transducer zeroing failed patient safety

Message 'Zeroing failed' displays in alarm area. Zeroing procedure could not complete. Possible causes: pulsating waveform detected, defective transducer, or offset exceeds 150 mmHg.

Ranked causes

  1. 1. Pulsating waveform detected during zeroing (occasional)
  2. 2. Defective transducer (occasional)
  3. 3. Offset exceeds 150 mmHg (rare)
First check

Verify transducer venting

Test point
Transducer venting stopcock or port
Expected
Transducer open to room air (0 mmHg reference)

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Abnormally high or low CO2 response or sudden occlusion

Abnormal CO2 reading or sudden occlusion message. Response to CO2 is abnormally high or low.

Ranked causes

  1. 1. Optical blockage or dirt in adapter (common)
  2. 2. LoFlo Sidestream Module malfunction (occasional)
First check

Inspect and clean CO2 adapter

Test point
Adapter optical window
Expected
Adapter clean, no optical blockage

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Apnea alarm not detected by monitor patient safety

Monitor does not activate apnea alarm when respiration stops or falls below alarm threshold for configured duration

Ranked causes

  1. 1. Simulator apnea simulation not activated or duration not set correctly (common)
  2. 2. Monitor apnea alarm disabled or threshold set too high (occasional)
First check

Configure simulator apnea simulation to 32 seconds

Test point
Simulator Apnea Simulation setting
Expected
Apnea Simulation set to '32 sec' and activated

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Asystole alarm not appearing on monitor patient safety

Asystole (cardiac standstill) signal from simulator does not trigger alarm display

Ranked causes

  1. 1. Simulator not configured to output asystole signal (common)
  2. 2. Asystole alarm disabled or alarm volume muted on monitor (occasional)
First check

Configure simulator to output asystole

Test point
Simulator ECG waveform settings
Expected
Simulator configured to output 'Asystole' signal

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

BP (Arterial) analog output signal incorrect

Arterial BP waveform on oscilloscope does not match P1 waveform on monitor

Ranked causes

  1. 1. BP analog output cable disconnected or faulty (common)
  2. 2. Main board analog output circuit failure (occasional)
First check

Arterial BP analog output signal waveform comparison

Test point
Red wire (BP_ANALOG_OUT), Green wire (ANALOG_RETURN)
Expected
Oscilloscope waveform matches P1 waveform on monitor screen
Tools
Oscilloscope, unterminated Defib/Sync cable, simulator

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

C.O./CCO measurement calibration failed

Message 'Calibration fail' or 'CCO meas. fail' displays in C.O. menu or parameter window. Calibration procedure was unsuccessful.

Ranked causes

  1. 1. Calibration procedure unsuccessful (occasional)
  2. 2. C.O./CCO measurement failure (occasional)
First check

Verify module connections and status

Test point
C.O./CCO module and cable connections
Expected
Secure connections; module powered and responsive

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration failed: Invasive pressure

Pressure calibration failure detected; system message 'Calibration failed' in parameter window

Ranked causes

  1. 1. Pulsating waveform detected during calibration (occasional)
  2. 2. Gain beyond ±20% of default (occasional)
First check

Check manometer reading during calibration

Test point
Pressure calibration reference manometer
Expected
Static 100–300 mmHg pressure maintained throughout calibration

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Analog outputs caution

Analog outputs calibration performed after main board replacement or when measured values fall outside specification. Involves measuring DC voltages across ECG and invasive pressure (BP) analog output pins using digital voltmeter connected to unterminated defib sync cable.

Ranked causes

  1. 1. Main board replaced (unknown)
  2. 2. Measured analog output value outside specification limits (unknown)
First check

Turn monitor on and wait for normal screen

Test point
Monitor display
Expected
Normal operating screen displayed

14 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Invasive pressure (Calibration error) patient safety

Monitor displays Calibration Error message when pressure is not applied within 45 seconds after zeroing completion or if calibration process fails for other reasons.

Ranked causes

  1. 1. Pressure application delayed beyond 45 seconds after zeroing (common)
  2. 2. Calibration process failure for other reasons (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: NIBP caution

NIBP calibration must be performed after failed calibration check or when measured values fall outside specification limits. Involves connecting adult cuff and manometer to rigid cylinder, pressurizing to ~250 mmHg, then adjusting gain against manometer reading.

Ranked causes

  1. 1. NIBP calibration check failed (unknown)
  2. 2. Measured NIBP value outside specification limits (unknown)
First check

Turn monitor on and wait for normal screen

Test point
Monitor display
Expected
Normal operating screen displayed

15 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CCO measurement stopped or not calibrated patient safety

Cardiac Output parameter window shows 'Not calibrated' or 'No CCO catheter'; CO waveform absent

Ranked causes

  1. 1. CCO not calibrated after module connection (common)
  2. 2. CCO catheter disconnected (occasional)
First check

Verify catheter and cable connections

Test point
CCO catheter connector, module CCO port
Expected
Catheter and cable securely connected

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • CO2 accuracy out of calibration
  • Defibrillator sync marker signal incorrect
  • ECG analog output signal incorrect
  • ECG artifact or noise on waveform
  • ECG lead off detection not functioning
  • ECG sync signal not available
  • EEG electrode impedance too high
  • EEG signal corrupted by muscle activity or external noise
  • Entropy sensor expired
  • ETCO2 over scale above 20%
  • ETCO2 over scale greater than 20%
  • EtCO2 overscale (>15% or >20%)
  • EtCO2 value too high
  • EtCO2 value too low
  • Expiratory tidal volume greater than inspiratory tidal volume
  • Incompatible or duplicate gas modules detected
  • Invasive pressure zeroing failure
  • Low gas sample flow
  • Marker in synchronization not working
  • Module cannot be placed properly in frame
  • Module will not dock
  • NIBP tubing is leaking
  • No P8 transducer detected
  • No response to patient breathing
  • No temperature displayed
  • Noisy waveforms and missing markers
  • P8 over range
  • P8 under range
  • Pacemaker spikes not detected on ECG
  • Pressure measurement artifact detected
  • Pressure measurement over range
  • Pressure over range condition
  • Replace D-fend Pro water trap
  • Replace D-fend water trap message
  • Respiration rate not displayed or incorrect
  • Sample flow rate out of specification
  • Sampling line disconnected
  • Sensor electrode poor contact
  • SpO2 reading not displayed or plethysmographic waveform missing
  • Temperature measurement error
  • Temperature module error or calibration failure
  • Too high ETCO2 reading
  • VO2 value invalid or no VO2 reading
  • Water trap not connected or air leak
  • Wrong sensor type connected
  • Zeroing error in spirometry
  • Abnormal CO2 response or sudden occlusion
  • Abnormally low invasive pressure
  • Abnormally low pressure invasive
  • Abnormally low pressure reading
  • Ambient pressure sensor reading incorrect
  • Apnea alarm deactivated
  • Apnea deactivated (impedance respiration)
  • Apnea detected (impedance respiration)
  • Artifact (muscle or noise)
  • Artifact detected (cardiac artifact on respiration)
  • Artifact detected in SpO2
  • Artifact in gas exchange measurement
  • Artifact in signal
  • Automatic sensor check disabled
  • BIS measurement removed
  • Bottom housing damaged
  • Bypass flow too high in gas exchange
  • C.O. measurement cable disconnected
  • C.O./CCO measurement calibration in progress
  • C.O./CCO module disconnected
  • Calibrating gas sensor - warm-up in progress
  • Calibration in progress
  • Calibration not available during warm-up period
  • Calibration verification
  • Calibration: BIS Module electrical safety
  • Calibration: Cardiac Output Module maintenance check
  • Calibration: Gas Calibration
  • Calibration: Invasive pressure (P1/P2)
  • Calibration: Invasive pressure (Zeroing failure)
  • Calibration: Masimo Module E-MASIMO
  • Calibration: Module Frame and Modules F5, F7
  • Calibration: NIBP (non-invasive blood pressure)
  • Calibration: Pressure Modules E-PT, E-PP
  • Calibration: Sample Flow Rate
  • Calibration: Sample flow rate adjustment
  • Calibration: Single-width Airway Module E-miniC
  • Calibration: Spirometry Flow Calibration
  • Calibration: Temperature
  • Calibration: Temperature (T1/T2)
  • Check SpO2 probe
  • Checking sensor message stays more than 2 minutes
  • CO cable off
  • CO measurement removed
  • CO2 adapter disconnected or faulty
  • CO2 pump goes off after prolonged occlusion
  • CO2 pump stopped
  • CO2 sensor overheating
  • CO2 sensor removed or disconnected
  • CO2 waveform clipped
  • Compliance value invalid
  • Confirm calibration message displayed
  • Corrective maintenance: Pressure Modules E-PT, E-PP
  • Demo data displayed
  • DSC error message displayed
  • Duplicate invasive pressure modules detected
  • E-LoFlo module and LoFlo Sidestream module not recognized
  • ECG measurement removed
  • ECG measurements removed
  • ECG module error
  • EEG headbox cable disconnected
  • EEG leads disconnected
  • EEG module disconnected from monitor
  • Electrical safety verification - Masimo Module
  • Electromagnetic interference detected
  • EMG recording electrodes disconnected or broken
  • Entropy cable disconnected
  • Entropy measurement starting up
  • Entropy module disconnected
  • Entropy sensor not attached to patient
  • Entropy simulator connected
  • EtCO2 > PaCO2 (arterial discrepancy)
  • ETCO2 higher than PaCO2
  • EtCO2 waveform clipped
  • Failure in Agent ID
  • Faulty ECG cable
  • Faulty SpO2 probe
  • Gas measurements removed
  • Gas measurements removed from display
  • Gas sensor warm-up delay
  • High BIS electrode impedance
  • High electrode impedance variation between leads
  • High-frequency noise during measurement
  • Humidity setting affects EtCO2 measurement accuracy
  • ICP channel requires separate zeroing
  • Identical C.O. modules message
  • Identical CO modules connected
  • Identical EEG modules connected
  • Identical gas modules detected
  • Identical IP8 modules connected
  • Identical SpO2 modules
  • Identical SpO2(2) sources detected
  • Identical temperature modules detected
  • Incompatible DSC detected
  • Incompatible gas module detected
  • Incompatible sensor detected
  • Incompatible SpO2 probe
  • Incompatible SpO2 probe installed
  • Invasive pressure channel over range
  • Invasive pressure channel under range
  • Invasive pressure channels not zeroed
  • Invasive pressure line disconnected
  • Invasive pressure not zeroed
  • Invasive pressure zero offset too high
  • Invasive pressure zeroing fails
  • Isoelectric (flatline) EEG detected
  • Lead I failed (impedance respiration)
  • Lead II failed (impedance respiration)
  • Lead off (single or multiple leads)
  • Lead RL-LL failed (impedance respiration)
  • Leads off (arrhythmia detection lost)
  • Low EEG signal amplitude
  • Low perfusion at measurement site
  • Low perfusion SpO2
  • Low signal quality SpO2
  • Low signal SpO2
  • Low SpO2 signal quality
  • Low volumes detected in spirometry
  • Measurement artifact detected
  • Measurement removed from display
  • Module error message displayed
  • Module malfunction detected
  • Module not responding to key press
  • Module unable to find pulse
  • Multiple ECG modules detected
  • Multiple Entropy modules connected
  • Multiple identical NMT modules connected
  • Multiple identical SpO2 modules connected
  • NIBP artifacts detected during measurement
  • NIBP Auto stopped
  • NIBP calibration failed
  • NIBP calibration protection mode left active
  • NIBP control measurement triggered
  • NIBP cuff ID not automatically detected
  • NIBP cuff overpressure
  • NIBP diastolic pressure measurement failed
  • NIBP incorrect inflation limits
  • NIBP is out of calibration
  • NIBP measurement incomplete
  • NIBP measurement time exceeds limit
  • NIBP overpressure relief not working
  • NIBP unable to measure systolic pressure
  • NIBP unstable zero pressure
  • NIBP zeroing failed
  • NMT measurement stopped
  • NMT module disconnected from monitor
  • NMT sensor cable disconnected
  • No BIS waveforms on screen
  • No C.O. module message
  • No CCO catheter message
  • No Entropy sensor detected
  • No pressure transducer connected
  • No response to breathing in CO2 measurement
  • No sensor detected
  • No SpO2 probe connected
  • No SpO2 probe message
  • No VO2 (FiN2O too high) in gas exchange
  • No VO2 (FiO2 >85%) in gas exchange
  • Noisy sampling pump operation
  • Not calibrated message
  • Over range measured FiO2
  • Over scale in spirometry (flow/pressure waveform)
  • Poor signal quality
  • Pressure channel calibration failed
  • Pressure channel calibration in progress
  • Pressure channel not zeroed
  • Pressure channel standby mode active
  • Pressure measurement out of range
  • Pressure measurement under range
  • Pressure under range condition
  • Printing spirometry loop
  • Pulse not detected
  • Pulse rate alarm SpO2 high or low
  • Re-calibrate CCO message
  • Recalibrate SvO2
  • Reference level setting fails
  • Reference level setting in progress
  • Regional block stimulation in progress
  • Relearning respiration
  • RF interference detected
  • Sample line disconnected message displayed
  • Saving spirometry loop
  • Scale changed notification in spirometry
  • Sensor check failed
  • Sensor expired or validity time unknown
  • Sensor impedance check fails
  • Sensor impedance check not available in menus
  • Sensor is faulty - Service CO2
  • Sensor warming up
  • Service CO2 module - multiple failures
  • Service CO2 sensor - sensor faulty
  • Service Monitor Error Code 0xHOST1002
  • Signal artifact detected
  • Signal contains excess noise or artifact
  • Spirometry measurements removed
  • SpO2 high or low NIBP
  • SpO2 malfunction check device
  • SpO2 measurement removed
  • SpO2 module disconnected from bus
  • SpO2 module disconnected from slot
  • SpO2 probe LEDs not functioning
  • SpO2 probe LEDs not operating
  • SpO2 probe not connected to module
  • SpO2 probe not connected to patient
  • SpO2 probe not detected
  • SpO2 probe off
  • SpO2 probe off patient
  • SpO2 probe signal imbalance
  • Stimulating electrodes disconnected or cable broken
  • Supramaximal stimulus current not found
  • Supramaximal stimulus search in progress
  • SvO2 cable off
  • SvO2 faulty cable
  • SvO2 measurement removed
  • SvO2 not calibrated
  • SvO2 signal poor
  • Temperature measurement out of range
  • Temperature measurement removed
  • Temperature module calibration in progress
  • Temperature out of measurable range
  • Testing DSC in progress
  • Tetanic stimulation in progress
  • Top housing damaged
  • Tubing contamination inside module
  • Unable to complete NIBP calibration
  • Visual inspection of EEG module and headbox
  • Visual inspections
  • Waveform clipped
  • Waveform over scale
  • Wrong cable Nellcor PDM Masimo
  • Wrong cable NIBP Masimo
  • Wrong cable use Masimo Set
  • Wrong cable use Nellcor OxiMax
  • Zero error - autozeroing failed
  • Zero error during calibration
  • Zero recommended
  • Zero required - measurement needs zeroing
  • Zeroing in progress

283 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
0xGAS0001Service gas module: CO2, N2O and anesthetic agent sensor failedMiniTPX measuring unit failure detected during sensor initialization or operationerror
0xGAS0010Service gas module: O2 sensor failedMiniOM oxygen sensor failure detected during sensor initialization or operationerror
0xGAS0011Service gas module: CO2, N2O, AA and O2 sensors failed simultaneouslySimultaneous MiniTPX and MiniOM failure detected during initializationcritical
0xGAS0100Service gas module: spirometry sensor failedSpirometry measurement unit (MiniPVX) failure detectederror
0xGAS0101Service gas module: CO2, N2O, AA and spirometry sensors failed simultaneouslySimultaneous MiniTPX and MiniPVX failure detectedcritical
0xGAS0110Service gas module: O2 and spirometry sensors failed simultaneouslySimultaneous MiniOM and MiniPVX failure detectedcritical
0xGAS0111Service gas module: CO2, N2O, AA, O2 and spirometry sensors failed simultaneouslyAll three sensor units (MiniTPX, MiniOM, MiniPVX) failed concurrentlycritical
0xGAS1000Service gas module: general gas measurement failedGas measurement system failure not attributable to individual sensorserror
0xGAS1001Service gas module: CO2, N2O, AA sensor and general gas measurement failedSimultaneous MiniTPX and general measurement system failurecritical
0xGAS1010Service gas module: O2 sensor and general gas measurement failedSimultaneous MiniOM and general measurement system failurecritical
0xGAS1011Service gas module: CO2, N2O, AA, O2 and general gas measurement failedSimultaneous MiniTPX, MiniOM and general measurement system failurecritical
0xGAS1100Service gas module: spirometry sensor and general gas measurement failedSimultaneous MiniPVX and general measurement system failurecritical
0xGAS1101Service gas module: CO2, N2O, AA, spirometry and general gas measurement failedSimultaneous MiniTPX, MiniPVX and general measurement system failurecritical
0xGAS1110Service gas module: O2, spirometry and general gas measurement failedSimultaneous MiniOM, MiniPVX and general measurement system failurecritical
0xGAS1111Service gas module: all sensors and general gas measurement failedCatastrophic module failure affecting all sensors and measurement systemscritical
Apnea alarmPatient respiratory apnea detectedAbsence of impedance respiration signal for duration exceeding configured apnea limit (default 3 seconds)critical
Apnea deactivatedApnea alarm detection not currently active; startup conditions not metPatient breathing not yet established at module power-up or reconnectioninfo
ArtifactNoise or movement artifact detected in SpO2 signalSignal quality degradation due to motion or sensor issueswarning
ArtifactsSignal contaminated with noise or artifactEEG waveform contains excessive noise during measurementwarning
Asystole alarmCardiac asystole detected on ECGMonitor detects absence of organized cardiac electrical activity (flat-line ECG)critical
Automatic check offAutomatic sensor impedance check is disabledUser turned off automatic check in menuinfo
Cable offEntropy sensor cable disconnectedCable unplugged from module or poor connectionwarning
CalibratedInvasive pressure calibration completed successfullyCalibration routine completed without errorsinfo
CalibratingCO2 sensor calibration procedure in progressUser initiated gas calibration sequenceinfo
Calibrating errorGas calibration procedure failed to complete successfullyCalibration gas not supplied, incorrect gas used, or excessive gain adjustment detectederror
Calibrating gas sensorGas sensor warm-up procedure in progress; measurements not availableModule initialization after connection or power-up; sensor requires stabilizationinfo
CalibrationCalibration in progressInvasive pressure channel calibration initiated and runninginfo
Calibration ErrorGas calibration failed during feed-gas phase or gain adjustmentNo calibration gas supplied within 1 minute after zeroing, or gain adjustment too largeerror
Calibration ErrorInvasive pressure calibration process failed during pressure inflation phaseNo pressure inflation within 45 seconds after automatic zeroing completion or calibration fails during inflationerror
Calibration ErrorCalibration procedure failed to completeInflation of pressure not started within 45 seconds after automatic zeroing, or calibration fails during the procedureerror
Calibration failedInvasive pressure channel calibration did not complete successfullyTime-out during calibration; pulsating waveform or incorrect gain detectederror
Check device (SpO2 malfunction)SpO2 acquisition module hardware failureAcquisition module internal failure detectederror
Checking sensorSensor impedance check in progressSensor check initiated (initial, manual, or periodic)info
Check probe (SpO2)No detectable SpO2 signal or sensor disconnectionSpO2 signal loss or sensor detachmenterror
Check sampling gas outSample gas outflow from module is blocked or pump failure detectedUnable to return sampled gas to patient circuit or scavenger systemwarning
Check water trapWater trap not connected or internal tubing has air leakCondition detected during sampling system operationwarning
CO cable offCO measurement cable disconnected from moduleCO measurement cable unplugged from module connectorerror
CO measurement removedCardiac output measurement module disconnectedCO module unplugged or disconnected from monitorerror
Confirm calibrationCCO measurement data requires confirmation before exiting calibration menuUser attempts to exit Cardiac Output/CCO Calibrate menu without confirming measurementwarning
Confirm sensor electrodesMultiple electrode contact loss detectedAutomatic impedance check detects high impedance on multiple electrodeswarning
Confirm sensor electrode XSingle electrode contact loss (electrode 1, 2, or 3)Automatic impedance check detects high impedance on one electrodewarning
Demo dataEntropy simulator detected; demo/test data displayedSimulator device connected instead of patient sensorinfo
Entropy measurement removedEntropy module disconnected from monitorModule unplugged or lost connection to module slotwarning
Failure in Agent IDAgent ID measurement has failedAgent ID sensor failure detectederror
Faulty probe (SpO2)SpO2 sensor hardware failureSensor self-test or communication failureerror
Gas calibration is not available during first 5 minutesCalibration cannot start immediately after module connectionAttempt to start gas calibration within 5 minutes of module connectionwarning
Gas calibration is not available during gas sampling warningCalibration unavailable due to active alarm conditionSample line blocked, Check Dfend, or Check sample gas out alarm condition activewarning
Gas measurements removedGas parameter display removed from monitor screen; monitoring no longer activeModule disconnected or measurement channel deactivated by systemwarning
Identical CO modulesMultiple identical CO modules connected simultaneouslyTwo or more of E-COP, E-COPSv, E-PiCCO modules connected at same timeerror
Identical Entropy modulesMultiple Entropy modules connected simultaneouslyMore than one Entropy module detected on module buserror
Identical gas modulesMonitor detects CO2 measurements from multiple identical modules; ambiguity in measurement sourceTwo or more identical CO2 modules connected to monitor providing duplicate measurementswarning
Identical IP5 modulesTwo or more modules providing the same IP5 (invasive pressure channel 5) detectedMonitor detects identical IP5 channel from multiple modules during initializationwarning
Identical IP6 modulesTwo or more modules providing the same IP6 (invasive pressure channel 6) detectedMonitor detects identical IP6 channel from multiple modules during initializationwarning
Identical IP7 modulesTwo or more modules providing the same IP7 (invasive pressure channel 7) detectedMonitor detects identical IP7 channel from multiple modules during initializationwarning
Identical IP8 modulesMultiple identical invasive pressure modules detectedTwo or more E-PiCCO or E-COPSv modules providing same IP channel connected simultaneouslyerror
Identical SpO2(2) modulesMultiple identical SpO2 modules in systemTwo or more identical SpO2 modules connectederror
Identical SpO2(2) sourcesMultiple SpO2 measurement sources active simultaneouslyE-module and UNID SpO2 sources detected at same timeerror
Incompatible gas moduleGas module detected as incompatible with current monitor systemMonitor firmware recognizes module type as incompatible with installed software versionerror
Incompatible probe (SpO2)SpO2 sensor not compatible with moduleWrong sensor type installed for module varianterror
Isoelectric EEGFlatline EEG detected during Entropy measurementNo EEG signal detected (isoelectric condition)critical
Low gas sample flowSample flow rate below 80% of module nominal specified valueSampling system flow measurement indicates reduced flow during operationwarning
Low perfusionInsufficient blood flow at SpO2 measurement siteWeak or absent pulsatile perfusion detectedwarning
Low signalEEG amplitude too low for reliable measurementEEG signal amplitude below minimum thresholdwarning
Low signalSpO2 signal quality questionableSignal strength below quality thresholdwarning
Low signal qualitySpO2 signal quality degradedSignal quality metrics below acceptable levelwarning
NIBP Auto stoppedNIBP automatic measurement stopped unexpectedlyLoose cuff or hose connection detected during measurementwarning
No CCO catheterCCO catheter or cable disconnected from systemCatheter disconnects from cable or cable disconnects from moduleerror
NoiseUnreliable Entropy calculation; distorted waveform from HF interferenceElectrosurgery or high-frequency noise during measurementwarning
No module (C.O.)No cardiac output module connectedCCO module not connected when C.O. measurement attemptederror
No P5 transducerPressure transducer P5 not detected or not connectedPressure transducer and/or transducer adapter cable disconnected from modulewarning
No P6 transducerPressure transducer P6 not detected or not connectedPressure transducer and/or transducer adapter cable disconnected from modulewarning
No P7 transducerPressure transducer P7 not detected or not connectedPressure transducer and/or transducer adapter cable disconnected from modulewarning
No P8 transducerPressure transducer disconnectedTransducer or adapter cable disconnected from moduleerror
No probe (SpO2)SpO2 sensor not connected to moduleSensor connector not detected by acquisition moduleerror
No sensorNo Entropy sensor detectedCable connected but sensor not attached to cablewarning
Not calibratedCCO measurement not calibratedModule and cable connected but calibration not performedwarning
Not Entropy sensorIncompatible or simulator device connectedNon-Entropy device detected at sensor porterror
Not zeroedInvasive pressure channel not zeroedAt least one invasive pressure channel requires zeroing before measurementswarning
Over rangeMeasured FiO2 exceeds valid rangeFiO2 measurement exceeds 103%warning
Over ScaleDisplayed CO2 waveform clipped; ETCO2 value exceeds upper limit of current display scaleMeasured ETCO2 exceeds maximum value displayable on selected scalewarning
P8 over rangeMeasured pressure exceeds upper limit (>320 mmHg / >43 kPa)Measured pressure value exceeds maximum rangeerror
P8 under rangeMeasured pressure below lower limit (<-40 mmHg / <-5 kPa)Measured pressure value below minimum rangeerror
Pressure measurement removedInvasive pressure module has been disconnected from monitorModule physically disconnected or communication lost with pressure moduleinfo
Probe off (SpO2)SpO2 sensor not functioning or not on patientSensor inactive or disconnected from patienterror
PR(SpO2) high / PR(SpO2) lowPulse rate measurement outside alarm limitsMeasured pulse rate equals or exceeds configured alarm thresholdswarning
Pulse search (NIBP)NIBP measurement initializing or pulse detection in progressNIBP measurement cycle starting or prolonged pulse detectioninfo
RA/R Lead offRight arm or right leg ECG lead disconnected or failedLoss of signal from RA/R lead wire contactwarning
Re-calibrate CCOCCO calibration expired or requires refreshMore than 8 hours elapsed since last calibrationwarning
Recalibrate SvO2SvO2 calibration older than 24 hoursElapsed time since last calibration exceeds 24 hourswarning
Replace D-FendWater trap filter exhausted, saturated with condensate, or contaminatedWater trap monitoring indicates saturation or contamination detectedwarning
Replace water trapD-fend Pro water trap requires replacement or internal tubing occludedDefective water trap detected or internal occlusionwarning
Sample line blockedContinuous blockage detected in gas sampling line, water trap, or internal tubingObstruction detected in sampling gas delivery pathwarning
Sensor check failedSensor impedance exceeds acceptable thresholdElectrode impedance measurement exceeds limit during checkerror
Sensor expiredEntropy sensor exceeded usable lifespanSensor used > 24 hours or past expiration datewarning
Sensor INOPGas sensor inoperative or not responding to gas measurementsModule unable to establish communication with sensor or sensor failure detectederror
Sensor offSensor cable connected but not attached to patientSensor not placed on patient skinwarning
SpO2 high / SpO2 low (NIBP)SpO2 measurement outside alarm limits during NIBP measurementSpO2 reading equals or exceeds alarm thresholdswarning
SpO2 measurement removedSpO2 acquisition module disconnected from systemModule connection lost or removederror
Starting upEntropy measurement initialization in progressSensor check passed; measurement startinginfo
SvO2 cable offSvO2 measurement cable disconnectedSvO2 cable unplugged from module connectorerror
SvO2 faulty cableOptical module factory calibration corrupted or transmitter errorRed/infrared transmitter malfunction during operationerror
SvO2 measurement removedSvO2 module disconnectedSvO2 module unplugged or disconnected from monitorerror
SvO2 not calibratedSvO2 catheter not calibrated after module connectionOptical module connected without catheter calibrationwarning
SvO2 signal poorSvO2 signal quality degraded or pulsation detectedSignal pulsation or intensity shift in signal qualitywarning
Temperature measurement removedTemperature module has been disconnected from monitorModule physically disconnected or communication lost with temperature moduleinfo
Tx temperature errorTemperature channel calibration or hardware failure detectedAutomatic self-check detects calibration failure or hardware fault on temperature channelerror
Wrong cable (NIBP/Masimo)Incorrect cable type connected to moduleNellcor cable connected to Masimo PDM or vice versaerror
Wrong cable. Use Masimo SetNellcor cable installed on Masimo PDMNellcor OxiMax cable detected connected to Masimo moduleerror
Wrong cable. Use Nellcor OxiMaxMasimo cable installed on Nellcor PDMMasimo Set cable detected connected to Nellcor moduleerror
Zero adj > 100 mmHgPressure zeroing offset exceeds 100 mmHg but less than 150 mmHgZeroing procedure detects offset > 100 mmHg from absolute zero of module with default gainwarning
ZeroedInvasive pressure zeroing completed successfullyZeroing procedure completed without errorsinfo
Zero ErrorAutomatic zeroing of gas sensor failedDuring automatic zeroing phase of gas calibrationerror
Zero errorAutozeroing failed during measurement start or gas calibration initiationZero valve malfunction or blockage preventing baseline zeroingerror
Zero FailureAutomatic zeroing of invasive pressure channel failedDuring automatic zeroing step of invasive pressure calibrationwarning
Zero ICP separatelyICP channel requires separate zeroingZero All command executed while ICP is active IP channelwarning
ZeroingCO2 sensor autozeroing procedure in progressSystem performing baseline zeroing before measurement or calibrationinfo
Zeroing failedInvasive pressure channel zeroing procedure did not complete successfullyPulsating waveform detected; offset exceeds 150 mmHg during zeroingerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI