GE Healthcare Carescape Modules: troubleshooting and error codes
The ServiceForge knowledge base holds 343 documented symptoms for the GE Healthcare Carescape Modules, with 553 ranked causes, 489 diagnostic checks with test points and expected values, and 117 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
NIBP hardware error
Parameter window message 'Call Service: Error x' indicating NIBP module hardware fault; message stays on screen until problem is fixed
Ranked causes
- 1. RAM failure (error x=0) (rare)
- 2. ROM checksum failure (error x=1) (rare)
- 3. +15V supply failure (error x=2) (rare)
- 4. -15V supply failure (error x=3) (rare)
- 5. ADC error (error x=6) (rare)
- 6. Watchdog timer too short (error x=7) (rare)
- 7. Watchdog timer too long (error x=8) (rare)
- 8. Watchdog activated (error x=9) (rare)
- 9. EEPROM checksum error (error x=10) (rare)
- 10. Autozero range exceeded (error x=11) (rare)
- 11. Communication watchdog activated (error x=12) (rare)
- 12. Too early autocycle start (error x=14) (rare)
12 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP air leakage patient safety
Alarm or parameter window message indicating air is leaking from cuff, hose, internal tubes, or module components during pressure measurement
Ranked causes
- 1. Cuff physically damaged (occasional)
- 2. Cuff or hose connector damaged or O-ring missing (occasional)
- 3. NIBP cuff hose double connector damaged (rare)
- 4. Leakage in patient connector panel tubing (occasional)
- 5. Internal module tubing leakage (rare)
- 6. Tube disconnected or damaged (occasional)
- 7. Manifold leakage (rare)
- 8. Valves or connecting tubes damaged (rare)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Module not appearing on monitor display
Parameter module connected to frame but measurements not shown on display; either specific module data missing or all non-PSM/PDM modules not visible
Ranked causes
- 1. Module not connected properly (common)
- 2. Module connector defective (occasional)
- 3. EMBC internal connectors defective or loose (occasional)
- 4. EMBC board defective (occasional)
- 5. Cable between frame and host not connected or defective (occasional)
- 6. DC/DC board defective (occasional)
- 7. Monitor display not configured for parameter (common)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unstable O2 delivery affecting VO2 calculation patient safety
VO2 readings invalid or unreliable due to unstable oxygen delivery or gas mixer issues
Ranked causes
- 1. Unstable O2 delivery or gas mixer malfunction (occasional)
- 2. Respiratory rate over 35/min (occasional)
- 3. Reference gas inlet port blocked (occasional)
- 4. Gas sampling line longer than 2 m (occasional)
- 5. Dead space of Y-piece greater than 8 ml (occasional)
- 6. Gas sampling line connected to HME instead of D-lite (occasional)
- 7. D-lite incorrectly placed in circuit (occasional)
Display oxygram and assess curve stability
- Test point
- Monitor VO2 waveform display
- Expected
- Stable, consistent oxygram curve
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement(s) removed from display
Alarm area shows 'XXXX measurement(s) removed'; module data no longer visible on monitor
Ranked causes
- 1. Module not connected properly (common)
- 2. Module connector defective (occasional)
- 3. Frame not properly connected to host (occasional)
- 4. EMBC board connectors loose inside frame (occasional)
- 5. EMBC board defective (occasional)
- 6. DC/DC board defective (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP cuff loose or disconnected patient safety
Alarm or parameter window message indicating blood pressure cuff is not properly connected, loosely wrapped, or has internal air leakage
Ranked causes
- 1. Cuff or hose not connected to module (common)
- 2. Cuff loosely wrapped on patient arm (common)
- 3. Air leakage in shielding or connectors (occasional)
- 4. Cuff or hose air leakage (occasional)
- 5. Internal module tubing leak (rare)
- 6. NIBP pump failure (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP weak pulsation signal patient safety
Parameter window message indicating weak oscillation or poor signal quality due to weak blood circulation, improper cuff position, or small air leakage
Ranked causes
- 1. Weak or abnormal blood circulation (common)
- 2. Improper cuff position or attachment (common)
- 3. Slow heart rate with artifacts (occasional)
- 4. Diastolic pressure measurement problems (occasional)
- 5. Small air leakage (occasional)
- 6. Insufficient measurement data due to artifacts (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas sensor indicates INOP patient safety
Waveform field or parameter window displays INOP; gas measurement unavailable from respiratory module
Ranked causes
- 1. MiniOM unit internal failure (temperature, supply voltage, other) (common)
- 2. MiniTPX unit failure (temperature, supply voltage, IR lamp) (common)
- 3. MiniPVX unit failure (pressure sensor, supply voltage) (occasional)
- 4. CPU failure (supply voltage, pressure sensor, A/D converter) (occasional)
- 5. Other failures (fan, pump, valve, zeroing, CO2 reference) (occasional)
Check flex cable connection
- Test point
- Module flex cable
- Expected
- Secure connection
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low gas sample flow rate patient safety
Sample flow below 80% of module nominal value, affecting measurement accuracy
Ranked causes
- 1. Gas sampling line blocked (common)
- 2. Water trap blocked or full (common)
- 3. Sample gas output blocked (common)
- 4. Internal tubing blocked (occasional)
- 5. Pump unit failure (occasional)
Measure sample flow rate in calibration menu
- Test point
- Flow rate display
- Expected
- At least 80% of module nominal flow value
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No temperature displayed on monitor patient safety
Temperature parameter window not showing any temperature value or waveform; measurement completely absent from display
Ranked causes
- 1. Temperature channel not configured to screen display (common)
- 2. Incompatible temperature probe installed (occasional)
- 3. Temperature probe type setting (STP/TP/ST) incorrect (occasional)
- 4. Faulty temperature probe (occasional)
- 5. Temperature measurement out of range (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Too high Ppeak patient safety
Peak inspiratory pressure measurement reads abnormally elevated
Ranked causes
- 1. Bronchospasm (common)
- 2. Patient coughing (common)
- 3. Patient breathing against ventilator (common)
- 4. Obstruction in airways (occasional)
- 5. HME obstructed (occasional)
Assess for acute bronchospasm
- Test point
- Patient clinical status and breath sounds
- Expected
- Normal breath sounds, no wheezing or stridor
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check sample line message
Message displayed in parameter window or alarm area indicating sampling line issue.
Ranked causes
- 1. Water, mucus, or contamination in sampling line (common)
- 2. Sampling line crimped, pinched, or damaged (common)
- 3. Gas outlet port obstructed (occasional)
- 4. Pressure changes or gas leak (occasional)
Visual inspection for contamination in sampling line
- Test point
- Sampling line interior
- Expected
- Clear, no water, mucus, or debris visible
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication LED not lit
Communication LED indicator on module does not illuminate
Ranked causes
- 1. ePort to host interface cable damaged (common)
- 2. ePort connector on monitor damaged (occasional)
- 3. ePort interface on module damaged (occasional)
- 4. Module defective (rare)
Inspect ePort cable
- Test point
- ePort to host interface cable
- Expected
- No visible damage, cuts, or bent pins
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas sampling line blocked patient safety
Blockage in sampling system preventing gas flow from patient to sensor
Ranked causes
- 1. External gas sampling line blocked (common)
- 2. Water trap occluded (common)
- 3. Internal tubing blocked (common)
- 4. Sampling pump failure (occasional)
Inspect external gas sampling line
- Test point
- Patient sampling line tubing
- Expected
- No visible blockages, kinks or condensation
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspired tidal volume exceeds expired tidal volume (insp TV > exp TV) patient safety
Spirometry indicates inspired volume greater than expired volume; leak suspected in lungs or breathing circuit
Ranked causes
- 1. Leak in patient lungs, endotracheal tube cuff, or spirometry system (common)
- 2. Water inside D-lite or tubing (occasional)
- 3. Another side-stream gas sampling line between D-lite and patient causing bypass (occasional)
- 4. D-fend Pro filter leaking (occasional)
Perform spirometry leak test
- Test point
- Spirometry circuit with patient connected
- Expected
- No leak detected; TVexp ≥ TVinsp
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP cuff occlusion patient safety
Parameter window message indicating air path from cuff or tubing is blocked or kinked, preventing proper inflation and measurement
Ranked causes
- 1. Cuff or hose tubing kinked or occluded (common)
- 2. Pressure transducer failure (occasional)
- 3. A/D converter fault (occasional)
- 4. NIBP calibration fault (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No invasive pressure signal patient safety
No invasive pressure waveform or values displayed on monitor; measurement completely absent
Ranked causes
- 1. Defective invasive pressure transducer (occasional)
- 2. Invasive pressure module not connected to monitor (common)
- 3. Transducer adapter cable not connected to module (common)
- 4. Invasive pressure channel not configured or displayed (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No parameter data from module
Parameter window shows no data output from module due to cable or interface issue
Ranked causes
- 1. Parameter cable disconnected (common)
- 2. Parameter interface damaged (occasional)
- 3. Related parameter license not activated (occasional)
- 4. Monitor not configured to display parameter (occasional)
Verify cable connection
- Test point
- Parameter cable connector
- Expected
- Cable fully seated and tight
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No power at module
Module does not power on when connected to monitor
Ranked causes
- 1. ePort cable damaged (common)
- 2. Monitor ePort connector damaged (occasional)
- 3. Module ePort interface damaged (occasional)
- 4. Module battery not charging (occasional)
Inspect ePort cable integrity
- Test point
- ePort to host interface cable
- Expected
- No damage visible
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No pressure measurement available patient safety
No invasive pressure waveform or value displayed. No signal detected.
Ranked causes
- 1. Defective transducer (occasional)
- 2. Module not connected (common)
- 3. Transducer adapter cable not connected to module (common)
- 4. Invasive pressure channel not configured on display (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pulse search NIBP measurement patient safety
Parameter window displays pulse search message during NIBP initialization or due to sensor/cable fault
Ranked causes
- 1. Normal NIBP measurement startup (common)
- 2. Defective or damaged sensor or cable (occasional)
- 3. Sensor off patient (common)
- 4. Repeatable pulse detection failed (occasional)
Wait for pulse detection
- Test point
- NIBP measurement display
- Expected
- Waveform and pressure values displayed
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sample gas outflow blocked patient safety
Exhaust line from module is occluded, preventing return of sampled gas to patient or scavenging system
Ranked causes
- 1. Sample gas out connector or exhaust line blocked (common)
- 2. Return tubing to patient circuit occluded (occasional)
- 3. Scavenging system not open to room pressure (occasional)
- 4. Internal tubing blocked or pump failure (occasional)
Inspect sample gas out connector and exhaust line
- Test point
- Front panel outlet and exhaust tubing
- Expected
- No visible blockages or obstructions
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sample line blocked patient safety
Continuous blockage message in parameter window, waveform field, or alarm area
Ranked causes
- 1. Gas sampling line blocked (common)
- 2. Water trap container full (common)
- 3. Water trap occluded (occasional)
- 4. Internal tubing blocked (occasional)
Inspect external sampling line
- Test point
- External gas sampling line
- Expected
- Clear, no blockages
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Strongly vibrating loop patient safety
Loop display shows excessive vibration or oscillation, volumes appear too large or too small
Ranked causes
- 1. Mucus in endotracheal tube (common)
- 2. Water or secretions in hoses or D-lite (common)
- 3. Wrong mode or sensor selection (occasional)
- 4. Incorrect sensor type selected (occasional)
Assess patient airway status and secretion level
- Test point
- Patient ET tube and airway
- Expected
- Patent airway, minimal secretions
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Too high Raw patient safety
Airway resistance measurement reads abnormally high, indicating increased airway obstruction
Ranked causes
- 1. Kink in tubing (common)
- 2. Mucus in airway (common)
- 3. Asthmatic patient or bronchospasm (occasional)
- 4. Spontaneous or patient-triggered breaths with breathing effort against ventilator (occasional)
Visually inspect all tubing for kinks or compression
- Test point
- Entire spirometry and patient circuit tubing
- Expected
- Straight, unobstructed tubing; free flow
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Water trap blockage or leakage patient safety
Messages appear for water trap issues, sampling system not functioning properly
Ranked causes
- 1. Water trap not connected (common)
- 2. Air leak inside internal tubing (common)
- 3. Water trap container full (occasional)
- 4. Defective or contaminated Mini D-fend water trap (occasional)
Inspect water trap connection and condition
- Test point
- Water trap connector on front panel
- Expected
- Water trap properly connected, no visible damage
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not charging
Module battery does not charge when connected to charger or docking station
Ranked causes
- 1. Battery tray damaged (occasional)
- 2. Battery connector damaged (occasional)
- 3. Battery defective (occasional)
Inspect battery tray
- Test point
- Battery compartment tray
- Expected
- Tray undamaged and properly seated
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration error
Calibration error message in parameter window or calibration menu
Ranked causes
- 1. Calibration gas not fed within 1 minute after zeroing (common)
- 2. Calibration failed due to large gain adjustment (occasional)
- 3. Wrong calibration gas used (common)
Verify calibration gas supply started within 1 minute
- Test point
- Gas supply timing
- Expected
- Gas fed within 60 seconds of zeroing completion
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration error - gas not supplied or gain adjustment issue
Calibration failed due to calibration gas not supplied within 1 minute or excessive gain adjustment
Ranked causes
- 1. Calibration gas not supplied within 1 minute of zeroing (common)
- 2. Calibration failed due to excessive gain adjustment (occasional)
- 3. Wrong calibration gas used (occasional)
Verify correct calibration gas bottle connected and pressurized
- Test point
- Calibration gas supply at module inlet
- Expected
- Specified calibration gas connected and pressure available
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration failed patient safety
Message appears in parameter window. Pressure calibration did not complete successfully.
Ranked causes
- 1. Pulsating waveform detected during calibration (occasional)
- 2. Gain beyond limits (±20% of default) (occasional)
- 3. Defective transducer (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration not available during alarm condition
Gas calibration unavailable when alarm conditions present (sample line blocked, D-fend check, gas out check)
Ranked causes
- 1. Sample line blocked alarm active (common)
- 2. D-fend check alarm active (common)
- 3. Sample gas out check alarm active (common)
Check for active alarms in alarm area
- Test point
- Alarm area display
- Expected
- No active alarm messages; clear alarms before calibration
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Invasive pressure patient safety
Module invasive pressure measurement requires calibration when transducer is replaced, calibration check fails, or reading exceeds specification limits.
Ranked causes
- 1. New pressure transducer installed (common)
- 2. Previous calibration check failed (occasional)
- 3. Measured value outside specification limits (occasional)
Verify module is connected to monitor
- Test point
- Module connection port
- Expected
- Module actively connected
10 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG waveforms not displayed or noisy patient safety
ECG leads II, V1, aVL and all 12-lead waveforms not appearing or contain excessive noise
Ranked causes
- 1. Leadwire not properly connected or disconnected (common)
- 2. ECG signal path noise or interference (occasional)
- 3. Simulator not configured correctly or not generating ECG signal (occasional)
Verify ECG leads II, V1, aVL display with waveforms
- Test point
- ECG waveform fields on monitor screen
- Expected
- All three leads displayed simultaneously with noise-free waveforms and 80±5 bpm heart rate, ±5 bpm
- Tools
- Multiparameter patient simulator, 12-lead ECG trunk cable, 5-leadwire sets
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Expired volume much smaller than inspired volume (MVexp<<MVinsp) patient safety
Spirometry indicates expired volume ≤70% of inspired volume; likely leak in breathing circuit
Ranked causes
- 1. Leaking or occluded spirometry tube (common)
- 2. Water in spirometry tubing (occasional)
- 3. MiniPVX sensor failure (occasional)
Perform spirometry leak test
- Test point
- Spirometry circuit
- Expected
- No leak detected; TVexp ≥70% TVinsp
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluctuating Raw patient safety
Airway resistance (Raw) value fluctuates or varies abnormally during breath cycle
Ranked causes
- 1. Mucus in airways or tubings (common)
- 2. Breathing effort against ventilator; patient-triggered or spontaneous breaths (common)
- 3. Ventilator expiratory valve causing fluctuations during expiratory flow (occasional)
Inspect spirometry circuit for secretions
- Test point
- Tubing and D-fend Pro sensor
- Expected
- Clear of mucus and secretions
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invasive pressure calibration failure patient safety
Automatic calibration did not complete; timeout or error condition detected during pressure calibration
Ranked causes
- 1. Pulsating waveform during calibration (common)
- 2. Gain beyond limits (outside ±20% of default) (occasional)
- 3. Defective transducer (occasional)
Verify static pressure applied during calibration
- Test point
- Manometer reading during calibration
- Expected
- 100-300 mmHg static pressure (no pulsation)
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loop overscale patient safety
Monitored volumes displayed less than set volumes, loop scaling appears incorrect
Ranked causes
- 1. Wrong scale selected (common)
- 2. Wrong TV base selected (common)
- 3. Leak between ventilator and D-lite (occasional)
Verify volume scale setting on display
- Test point
- Monitor display scaling menu
- Expected
- Scale matches current tidal volume range
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement off (impedance respiration)
Parameter window or waveform displays 'Measurement off'; ECG leads not connected or measurement disabled in setup menu, or ECG source is telemetry
Ranked causes
- 1. ECG leads disconnected from patient (common)
- 2. Impedance respiration measurement disabled in setup menu (occasional)
- 3. ECG source is telemetry (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Module not functioning
Module fails to respond to key press, parameters not displayed, or communication error indicated
Ranked causes
- 1. Loose keypad cable or internal module connection (common)
- 2. Module incompatibility or configuration error (occasional)
- 3. Defective accessory or sensor (occasional)
Visual inspection of module and connectors
- Test point
- Front cover, front panel, connectors, module box and latch
- Expected
- Intact, clean, properly attached
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP not functioning patient safety
NIBP measurement does not operate or calibration cannot be completed
Ranked causes
- 1. NIBP not calibrated (common)
- 2. Cuff pressure not maintained (leak) (common)
- 3. NIBP assembly malfunction (occasional)
Check calibration status
- Test point
- NIBP module status
- Expected
- Calibration completed within acceptable timeframe
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No EntEEG waveform or Entropy display on screen patient safety
Entropy module not connected or parameter not configured/visible on display.
Ranked causes
- 1. Entropy module not connected (common)
- 2. Parameter not configured or priority too low (occasional)
- 3. Entropy not supported by active software (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No pressure reading displayed patient safety
Invasive pressure parameter window shows no value; pressure waveform absent from display
Ranked causes
- 1. Defective pressure transducer (occasional)
- 2. Module not connected (common)
- 3. Transducer adapter cable not connected (common)
Verify module is connected to monitor
- Test point
- module connectors
- Expected
- Module fully seated on monitor
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Out of range VO2 or VCO2 in gas exchange patient safety
Parameter window displays Out of range; VO2 or VCO2 measurement is <0 or >999 ml/min and exceeds measurement range
Ranked causes
- 1. Gas sampling or spirometry lines incorrectly connected to airway or module (common)
- 2. Incorrect sensor type (D-lite vs Pedi Lite) selected in monitor menu (occasional)
- 3. FiO2 recently changed; VO2 out of range temporarily (occasional)
Verify gas sampling and spirometry line connections
- Test point
- Airway connectors and module ports
- Expected
- All lines securely connected to airway and gas module
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Response signal too weak
Response too weak message in parameter window; maximum gain insufficient to amplify response signal to measurable level
Ranked causes
- 1. Stimulus electrodes loose or incorrect placement (common)
- 2. Response electrodes disconnected (common)
- 3. Dry or degraded electrodes (occasional)
Verify stimulus electrode placement and tightness
- Test point
- Stimulus electrode connections
- Expected
- Electrodes firmly seated on skin at correct anatomical location
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor INOP patient safety
Sensor INOP message in waveform field
Ranked causes
- 1. IR lamp failure (occasional)
- 2. Ambient pressure too high or low (occasional)
- 3. CPU failure or module communication issue (rare)
Check miniTPX flex cable connection
- Test point
- miniTPX flex cable connector
- Expected
- Securely connected
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor INOP (inoperative)
Gas sensor inoperative; sensor not responding to gas measurement
Ranked causes
- 1. Ambient pressure too high or low (occasional)
- 2. No response from gas module (occasional)
- 3. High temperature inside module or EEPROM checksum failure (rare)
Check ambient pressure reading in gas calibration menu
- Test point
- Calibration menu ambient pressure display
- Expected
- Pressure within normal operating range (typically 600-800 mmHg)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor INOP in spirometry (MiniPVX) patient safety
Spirometry waveform field displays INOP; no response from MiniPVX measurement board or EEPROM checksum failure
Ranked causes
- 1. MiniPVX flex cable loose or disconnected (occasional)
- 2. Zero valve broken or internal failure (occasional)
- 3. EEPROM checksum failure or measurement board failure (rare)
Verify MiniPVX flex cable connection
- Test point
- MiniPVX flex cable connector
- Expected
- Fully seated, secure
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Strange or nonphysiological VO2 readings patient safety
Gas exchange (VO2) values appear abnormal or inconsistent with patient physiology
Ranked causes
- 1. Ventilation mode BiPAP/CPAP with high continuous bypass flow (occasional)
- 2. Presence of N2O or anesthetic agents in ICU applications (occasional)
- 3. Gas exchange not measurable (occasional)
Select oxygram and verify curve stability
- Test point
- Monitor VO2 waveform display
- Expected
- Stable oxygram curve consistent with patient condition
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Too low ETCO2 reading patient safety
Measured end-tidal CO2 value abnormally low compared to expected clinical range
Ranked causes
- 1. Leak in sampling system (common)
- 2. Calibration error (common)
- 3. High by-pass flow from ventilator (occasional)
Check all gas connections for leaks
- Test point
- Water trap, sampling line, outlet connections
- Expected
- No hissing or leakage detected; tight connections
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transducer zeroing failed patient safety
Message 'Zeroing failed' displays in alarm area. Zeroing procedure could not complete. Possible causes: pulsating waveform detected, defective transducer, or offset exceeds 150 mmHg.
Ranked causes
- 1. Pulsating waveform detected during zeroing (occasional)
- 2. Defective transducer (occasional)
- 3. Offset exceeds 150 mmHg (rare)
Verify transducer venting
- Test point
- Transducer venting stopcock or port
- Expected
- Transducer open to room air (0 mmHg reference)
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormally high or low CO2 response or sudden occlusion
Abnormal CO2 reading or sudden occlusion message. Response to CO2 is abnormally high or low.
Ranked causes
- 1. Optical blockage or dirt in adapter (common)
- 2. LoFlo Sidestream Module malfunction (occasional)
Inspect and clean CO2 adapter
- Test point
- Adapter optical window
- Expected
- Adapter clean, no optical blockage
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Apnea alarm not detected by monitor patient safety
Monitor does not activate apnea alarm when respiration stops or falls below alarm threshold for configured duration
Ranked causes
- 1. Simulator apnea simulation not activated or duration not set correctly (common)
- 2. Monitor apnea alarm disabled or threshold set too high (occasional)
Configure simulator apnea simulation to 32 seconds
- Test point
- Simulator Apnea Simulation setting
- Expected
- Apnea Simulation set to '32 sec' and activated
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Asystole alarm not appearing on monitor patient safety
Asystole (cardiac standstill) signal from simulator does not trigger alarm display
Ranked causes
- 1. Simulator not configured to output asystole signal (common)
- 2. Asystole alarm disabled or alarm volume muted on monitor (occasional)
Configure simulator to output asystole
- Test point
- Simulator ECG waveform settings
- Expected
- Simulator configured to output 'Asystole' signal
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
BP (Arterial) analog output signal incorrect
Arterial BP waveform on oscilloscope does not match P1 waveform on monitor
Ranked causes
- 1. BP analog output cable disconnected or faulty (common)
- 2. Main board analog output circuit failure (occasional)
Arterial BP analog output signal waveform comparison
- Test point
- Red wire (BP_ANALOG_OUT), Green wire (ANALOG_RETURN)
- Expected
- Oscilloscope waveform matches P1 waveform on monitor screen
- Tools
- Oscilloscope, unterminated Defib/Sync cable, simulator
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
C.O./CCO measurement calibration failed
Message 'Calibration fail' or 'CCO meas. fail' displays in C.O. menu or parameter window. Calibration procedure was unsuccessful.
Ranked causes
- 1. Calibration procedure unsuccessful (occasional)
- 2. C.O./CCO measurement failure (occasional)
Verify module connections and status
- Test point
- C.O./CCO module and cable connections
- Expected
- Secure connections; module powered and responsive
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration failed: Invasive pressure
Pressure calibration failure detected; system message 'Calibration failed' in parameter window
Ranked causes
- 1. Pulsating waveform detected during calibration (occasional)
- 2. Gain beyond ±20% of default (occasional)
Check manometer reading during calibration
- Test point
- Pressure calibration reference manometer
- Expected
- Static 100–300 mmHg pressure maintained throughout calibration
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Analog outputs caution
Analog outputs calibration performed after main board replacement or when measured values fall outside specification. Involves measuring DC voltages across ECG and invasive pressure (BP) analog output pins using digital voltmeter connected to unterminated defib sync cable.
Ranked causes
- 1. Main board replaced (unknown)
- 2. Measured analog output value outside specification limits (unknown)
Turn monitor on and wait for normal screen
- Test point
- Monitor display
- Expected
- Normal operating screen displayed
14 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Invasive pressure (Calibration error) patient safety
Monitor displays Calibration Error message when pressure is not applied within 45 seconds after zeroing completion or if calibration process fails for other reasons.
Ranked causes
- 1. Pressure application delayed beyond 45 seconds after zeroing (common)
- 2. Calibration process failure for other reasons (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP caution
NIBP calibration must be performed after failed calibration check or when measured values fall outside specification limits. Involves connecting adult cuff and manometer to rigid cylinder, pressurizing to ~250 mmHg, then adjusting gain against manometer reading.
Ranked causes
- 1. NIBP calibration check failed (unknown)
- 2. Measured NIBP value outside specification limits (unknown)
Turn monitor on and wait for normal screen
- Test point
- Monitor display
- Expected
- Normal operating screen displayed
15 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CCO measurement stopped or not calibrated patient safety
Cardiac Output parameter window shows 'Not calibrated' or 'No CCO catheter'; CO waveform absent
Ranked causes
- 1. CCO not calibrated after module connection (common)
- 2. CCO catheter disconnected (occasional)
Verify catheter and cable connections
- Test point
- CCO catheter connector, module CCO port
- Expected
- Catheter and cable securely connected
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- CO2 accuracy out of calibration
- Defibrillator sync marker signal incorrect
- ECG analog output signal incorrect
- ECG artifact or noise on waveform
- ECG lead off detection not functioning
- ECG sync signal not available
- EEG electrode impedance too high
- EEG signal corrupted by muscle activity or external noise
- Entropy sensor expired
- ETCO2 over scale above 20%
- ETCO2 over scale greater than 20%
- EtCO2 overscale (>15% or >20%)
- EtCO2 value too high
- EtCO2 value too low
- Expiratory tidal volume greater than inspiratory tidal volume
- Incompatible or duplicate gas modules detected
- Invasive pressure zeroing failure
- Low gas sample flow
- Marker in synchronization not working
- Module cannot be placed properly in frame
- Module will not dock
- NIBP tubing is leaking
- No P8 transducer detected
- No response to patient breathing
- No temperature displayed
- Noisy waveforms and missing markers
- P8 over range
- P8 under range
- Pacemaker spikes not detected on ECG
- Pressure measurement artifact detected
- Pressure measurement over range
- Pressure over range condition
- Replace D-fend Pro water trap
- Replace D-fend water trap message
- Respiration rate not displayed or incorrect
- Sample flow rate out of specification
- Sampling line disconnected
- Sensor electrode poor contact
- SpO2 reading not displayed or plethysmographic waveform missing
- Temperature measurement error
- Temperature module error or calibration failure
- Too high ETCO2 reading
- VO2 value invalid or no VO2 reading
- Water trap not connected or air leak
- Wrong sensor type connected
- Zeroing error in spirometry
- Abnormal CO2 response or sudden occlusion
- Abnormally low invasive pressure
- Abnormally low pressure invasive
- Abnormally low pressure reading
- Ambient pressure sensor reading incorrect
- Apnea alarm deactivated
- Apnea deactivated (impedance respiration)
- Apnea detected (impedance respiration)
- Artifact (muscle or noise)
- Artifact detected (cardiac artifact on respiration)
- Artifact detected in SpO2
- Artifact in gas exchange measurement
- Artifact in signal
- Automatic sensor check disabled
- BIS measurement removed
- Bottom housing damaged
- Bypass flow too high in gas exchange
- C.O. measurement cable disconnected
- C.O./CCO measurement calibration in progress
- C.O./CCO module disconnected
- Calibrating gas sensor - warm-up in progress
- Calibration in progress
- Calibration not available during warm-up period
- Calibration verification
- Calibration: BIS Module electrical safety
- Calibration: Cardiac Output Module maintenance check
- Calibration: Gas Calibration
- Calibration: Invasive pressure (P1/P2)
- Calibration: Invasive pressure (Zeroing failure)
- Calibration: Masimo Module E-MASIMO
- Calibration: Module Frame and Modules F5, F7
- Calibration: NIBP (non-invasive blood pressure)
- Calibration: Pressure Modules E-PT, E-PP
- Calibration: Sample Flow Rate
- Calibration: Sample flow rate adjustment
- Calibration: Single-width Airway Module E-miniC
- Calibration: Spirometry Flow Calibration
- Calibration: Temperature
- Calibration: Temperature (T1/T2)
- Check SpO2 probe
- Checking sensor message stays more than 2 minutes
- CO cable off
- CO measurement removed
- CO2 adapter disconnected or faulty
- CO2 pump goes off after prolonged occlusion
- CO2 pump stopped
- CO2 sensor overheating
- CO2 sensor removed or disconnected
- CO2 waveform clipped
- Compliance value invalid
- Confirm calibration message displayed
- Corrective maintenance: Pressure Modules E-PT, E-PP
- Demo data displayed
- DSC error message displayed
- Duplicate invasive pressure modules detected
- E-LoFlo module and LoFlo Sidestream module not recognized
- ECG measurement removed
- ECG measurements removed
- ECG module error
- EEG headbox cable disconnected
- EEG leads disconnected
- EEG module disconnected from monitor
- Electrical safety verification - Masimo Module
- Electromagnetic interference detected
- EMG recording electrodes disconnected or broken
- Entropy cable disconnected
- Entropy measurement starting up
- Entropy module disconnected
- Entropy sensor not attached to patient
- Entropy simulator connected
- EtCO2 > PaCO2 (arterial discrepancy)
- ETCO2 higher than PaCO2
- EtCO2 waveform clipped
- Failure in Agent ID
- Faulty ECG cable
- Faulty SpO2 probe
- Gas measurements removed
- Gas measurements removed from display
- Gas sensor warm-up delay
- High BIS electrode impedance
- High electrode impedance variation between leads
- High-frequency noise during measurement
- Humidity setting affects EtCO2 measurement accuracy
- ICP channel requires separate zeroing
- Identical C.O. modules message
- Identical CO modules connected
- Identical EEG modules connected
- Identical gas modules detected
- Identical IP8 modules connected
- Identical SpO2 modules
- Identical SpO2(2) sources detected
- Identical temperature modules detected
- Incompatible DSC detected
- Incompatible gas module detected
- Incompatible sensor detected
- Incompatible SpO2 probe
- Incompatible SpO2 probe installed
- Invasive pressure channel over range
- Invasive pressure channel under range
- Invasive pressure channels not zeroed
- Invasive pressure line disconnected
- Invasive pressure not zeroed
- Invasive pressure zero offset too high
- Invasive pressure zeroing fails
- Isoelectric (flatline) EEG detected
- Lead I failed (impedance respiration)
- Lead II failed (impedance respiration)
- Lead off (single or multiple leads)
- Lead RL-LL failed (impedance respiration)
- Leads off (arrhythmia detection lost)
- Low EEG signal amplitude
- Low perfusion at measurement site
- Low perfusion SpO2
- Low signal quality SpO2
- Low signal SpO2
- Low SpO2 signal quality
- Low volumes detected in spirometry
- Measurement artifact detected
- Measurement removed from display
- Module error message displayed
- Module malfunction detected
- Module not responding to key press
- Module unable to find pulse
- Multiple ECG modules detected
- Multiple Entropy modules connected
- Multiple identical NMT modules connected
- Multiple identical SpO2 modules connected
- NIBP artifacts detected during measurement
- NIBP Auto stopped
- NIBP calibration failed
- NIBP calibration protection mode left active
- NIBP control measurement triggered
- NIBP cuff ID not automatically detected
- NIBP cuff overpressure
- NIBP diastolic pressure measurement failed
- NIBP incorrect inflation limits
- NIBP is out of calibration
- NIBP measurement incomplete
- NIBP measurement time exceeds limit
- NIBP overpressure relief not working
- NIBP unable to measure systolic pressure
- NIBP unstable zero pressure
- NIBP zeroing failed
- NMT measurement stopped
- NMT module disconnected from monitor
- NMT sensor cable disconnected
- No BIS waveforms on screen
- No C.O. module message
- No CCO catheter message
- No Entropy sensor detected
- No pressure transducer connected
- No response to breathing in CO2 measurement
- No sensor detected
- No SpO2 probe connected
- No SpO2 probe message
- No VO2 (FiN2O too high) in gas exchange
- No VO2 (FiO2 >85%) in gas exchange
- Noisy sampling pump operation
- Not calibrated message
- Over range measured FiO2
- Over scale in spirometry (flow/pressure waveform)
- Poor signal quality
- Pressure channel calibration failed
- Pressure channel calibration in progress
- Pressure channel not zeroed
- Pressure channel standby mode active
- Pressure measurement out of range
- Pressure measurement under range
- Pressure under range condition
- Printing spirometry loop
- Pulse not detected
- Pulse rate alarm SpO2 high or low
- Re-calibrate CCO message
- Recalibrate SvO2
- Reference level setting fails
- Reference level setting in progress
- Regional block stimulation in progress
- Relearning respiration
- RF interference detected
- Sample line disconnected message displayed
- Saving spirometry loop
- Scale changed notification in spirometry
- Sensor check failed
- Sensor expired or validity time unknown
- Sensor impedance check fails
- Sensor impedance check not available in menus
- Sensor is faulty - Service CO2
- Sensor warming up
- Service CO2 module - multiple failures
- Service CO2 sensor - sensor faulty
- Service Monitor Error Code 0xHOST1002
- Signal artifact detected
- Signal contains excess noise or artifact
- Spirometry measurements removed
- SpO2 high or low NIBP
- SpO2 malfunction check device
- SpO2 measurement removed
- SpO2 module disconnected from bus
- SpO2 module disconnected from slot
- SpO2 probe LEDs not functioning
- SpO2 probe LEDs not operating
- SpO2 probe not connected to module
- SpO2 probe not connected to patient
- SpO2 probe not detected
- SpO2 probe off
- SpO2 probe off patient
- SpO2 probe signal imbalance
- Stimulating electrodes disconnected or cable broken
- Supramaximal stimulus current not found
- Supramaximal stimulus search in progress
- SvO2 cable off
- SvO2 faulty cable
- SvO2 measurement removed
- SvO2 not calibrated
- SvO2 signal poor
- Temperature measurement out of range
- Temperature measurement removed
- Temperature module calibration in progress
- Temperature out of measurable range
- Testing DSC in progress
- Tetanic stimulation in progress
- Top housing damaged
- Tubing contamination inside module
- Unable to complete NIBP calibration
- Visual inspection of EEG module and headbox
- Visual inspections
- Waveform clipped
- Waveform over scale
- Wrong cable Nellcor PDM Masimo
- Wrong cable NIBP Masimo
- Wrong cable use Masimo Set
- Wrong cable use Nellcor OxiMax
- Zero error - autozeroing failed
- Zero error during calibration
- Zero recommended
- Zero required - measurement needs zeroing
- Zeroing in progress
283 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0xGAS0001 | Service gas module: CO2, N2O and anesthetic agent sensor failed | MiniTPX measuring unit failure detected during sensor initialization or operation | error |
0xGAS0010 | Service gas module: O2 sensor failed | MiniOM oxygen sensor failure detected during sensor initialization or operation | error |
0xGAS0011 | Service gas module: CO2, N2O, AA and O2 sensors failed simultaneously | Simultaneous MiniTPX and MiniOM failure detected during initialization | critical |
0xGAS0100 | Service gas module: spirometry sensor failed | Spirometry measurement unit (MiniPVX) failure detected | error |
0xGAS0101 | Service gas module: CO2, N2O, AA and spirometry sensors failed simultaneously | Simultaneous MiniTPX and MiniPVX failure detected | critical |
0xGAS0110 | Service gas module: O2 and spirometry sensors failed simultaneously | Simultaneous MiniOM and MiniPVX failure detected | critical |
0xGAS0111 | Service gas module: CO2, N2O, AA, O2 and spirometry sensors failed simultaneously | All three sensor units (MiniTPX, MiniOM, MiniPVX) failed concurrently | critical |
0xGAS1000 | Service gas module: general gas measurement failed | Gas measurement system failure not attributable to individual sensors | error |
0xGAS1001 | Service gas module: CO2, N2O, AA sensor and general gas measurement failed | Simultaneous MiniTPX and general measurement system failure | critical |
0xGAS1010 | Service gas module: O2 sensor and general gas measurement failed | Simultaneous MiniOM and general measurement system failure | critical |
0xGAS1011 | Service gas module: CO2, N2O, AA, O2 and general gas measurement failed | Simultaneous MiniTPX, MiniOM and general measurement system failure | critical |
0xGAS1100 | Service gas module: spirometry sensor and general gas measurement failed | Simultaneous MiniPVX and general measurement system failure | critical |
0xGAS1101 | Service gas module: CO2, N2O, AA, spirometry and general gas measurement failed | Simultaneous MiniTPX, MiniPVX and general measurement system failure | critical |
0xGAS1110 | Service gas module: O2, spirometry and general gas measurement failed | Simultaneous MiniOM, MiniPVX and general measurement system failure | critical |
0xGAS1111 | Service gas module: all sensors and general gas measurement failed | Catastrophic module failure affecting all sensors and measurement systems | critical |
Apnea alarm | Patient respiratory apnea detected | Absence of impedance respiration signal for duration exceeding configured apnea limit (default 3 seconds) | critical |
Apnea deactivated | Apnea alarm detection not currently active; startup conditions not met | Patient breathing not yet established at module power-up or reconnection | info |
Artifact | Noise or movement artifact detected in SpO2 signal | Signal quality degradation due to motion or sensor issues | warning |
Artifacts | Signal contaminated with noise or artifact | EEG waveform contains excessive noise during measurement | warning |
Asystole alarm | Cardiac asystole detected on ECG | Monitor detects absence of organized cardiac electrical activity (flat-line ECG) | critical |
Automatic check off | Automatic sensor impedance check is disabled | User turned off automatic check in menu | info |
Cable off | Entropy sensor cable disconnected | Cable unplugged from module or poor connection | warning |
Calibrated | Invasive pressure calibration completed successfully | Calibration routine completed without errors | info |
Calibrating | CO2 sensor calibration procedure in progress | User initiated gas calibration sequence | info |
Calibrating error | Gas calibration procedure failed to complete successfully | Calibration gas not supplied, incorrect gas used, or excessive gain adjustment detected | error |
Calibrating gas sensor | Gas sensor warm-up procedure in progress; measurements not available | Module initialization after connection or power-up; sensor requires stabilization | info |
Calibration | Calibration in progress | Invasive pressure channel calibration initiated and running | info |
Calibration Error | Gas calibration failed during feed-gas phase or gain adjustment | No calibration gas supplied within 1 minute after zeroing, or gain adjustment too large | error |
Calibration Error | Invasive pressure calibration process failed during pressure inflation phase | No pressure inflation within 45 seconds after automatic zeroing completion or calibration fails during inflation | error |
Calibration Error | Calibration procedure failed to complete | Inflation of pressure not started within 45 seconds after automatic zeroing, or calibration fails during the procedure | error |
Calibration failed | Invasive pressure channel calibration did not complete successfully | Time-out during calibration; pulsating waveform or incorrect gain detected | error |
Check device (SpO2 malfunction) | SpO2 acquisition module hardware failure | Acquisition module internal failure detected | error |
Checking sensor | Sensor impedance check in progress | Sensor check initiated (initial, manual, or periodic) | info |
Check probe (SpO2) | No detectable SpO2 signal or sensor disconnection | SpO2 signal loss or sensor detachment | error |
Check sampling gas out | Sample gas outflow from module is blocked or pump failure detected | Unable to return sampled gas to patient circuit or scavenger system | warning |
Check water trap | Water trap not connected or internal tubing has air leak | Condition detected during sampling system operation | warning |
CO cable off | CO measurement cable disconnected from module | CO measurement cable unplugged from module connector | error |
CO measurement removed | Cardiac output measurement module disconnected | CO module unplugged or disconnected from monitor | error |
Confirm calibration | CCO measurement data requires confirmation before exiting calibration menu | User attempts to exit Cardiac Output/CCO Calibrate menu without confirming measurement | warning |
Confirm sensor electrodes | Multiple electrode contact loss detected | Automatic impedance check detects high impedance on multiple electrodes | warning |
Confirm sensor electrode X | Single electrode contact loss (electrode 1, 2, or 3) | Automatic impedance check detects high impedance on one electrode | warning |
Demo data | Entropy simulator detected; demo/test data displayed | Simulator device connected instead of patient sensor | info |
Entropy measurement removed | Entropy module disconnected from monitor | Module unplugged or lost connection to module slot | warning |
Failure in Agent ID | Agent ID measurement has failed | Agent ID sensor failure detected | error |
Faulty probe (SpO2) | SpO2 sensor hardware failure | Sensor self-test or communication failure | error |
Gas calibration is not available during first 5 minutes | Calibration cannot start immediately after module connection | Attempt to start gas calibration within 5 minutes of module connection | warning |
Gas calibration is not available during gas sampling warning | Calibration unavailable due to active alarm condition | Sample line blocked, Check Dfend, or Check sample gas out alarm condition active | warning |
Gas measurements removed | Gas parameter display removed from monitor screen; monitoring no longer active | Module disconnected or measurement channel deactivated by system | warning |
Identical CO modules | Multiple identical CO modules connected simultaneously | Two or more of E-COP, E-COPSv, E-PiCCO modules connected at same time | error |
Identical Entropy modules | Multiple Entropy modules connected simultaneously | More than one Entropy module detected on module bus | error |
Identical gas modules | Monitor detects CO2 measurements from multiple identical modules; ambiguity in measurement source | Two or more identical CO2 modules connected to monitor providing duplicate measurements | warning |
Identical IP5 modules | Two or more modules providing the same IP5 (invasive pressure channel 5) detected | Monitor detects identical IP5 channel from multiple modules during initialization | warning |
Identical IP6 modules | Two or more modules providing the same IP6 (invasive pressure channel 6) detected | Monitor detects identical IP6 channel from multiple modules during initialization | warning |
Identical IP7 modules | Two or more modules providing the same IP7 (invasive pressure channel 7) detected | Monitor detects identical IP7 channel from multiple modules during initialization | warning |
Identical IP8 modules | Multiple identical invasive pressure modules detected | Two or more E-PiCCO or E-COPSv modules providing same IP channel connected simultaneously | error |
Identical SpO2(2) modules | Multiple identical SpO2 modules in system | Two or more identical SpO2 modules connected | error |
Identical SpO2(2) sources | Multiple SpO2 measurement sources active simultaneously | E-module and UNID SpO2 sources detected at same time | error |
Incompatible gas module | Gas module detected as incompatible with current monitor system | Monitor firmware recognizes module type as incompatible with installed software version | error |
Incompatible probe (SpO2) | SpO2 sensor not compatible with module | Wrong sensor type installed for module variant | error |
Isoelectric EEG | Flatline EEG detected during Entropy measurement | No EEG signal detected (isoelectric condition) | critical |
Low gas sample flow | Sample flow rate below 80% of module nominal specified value | Sampling system flow measurement indicates reduced flow during operation | warning |
Low perfusion | Insufficient blood flow at SpO2 measurement site | Weak or absent pulsatile perfusion detected | warning |
Low signal | EEG amplitude too low for reliable measurement | EEG signal amplitude below minimum threshold | warning |
Low signal | SpO2 signal quality questionable | Signal strength below quality threshold | warning |
Low signal quality | SpO2 signal quality degraded | Signal quality metrics below acceptable level | warning |
NIBP Auto stopped | NIBP automatic measurement stopped unexpectedly | Loose cuff or hose connection detected during measurement | warning |
No CCO catheter | CCO catheter or cable disconnected from system | Catheter disconnects from cable or cable disconnects from module | error |
Noise | Unreliable Entropy calculation; distorted waveform from HF interference | Electrosurgery or high-frequency noise during measurement | warning |
No module (C.O.) | No cardiac output module connected | CCO module not connected when C.O. measurement attempted | error |
No P5 transducer | Pressure transducer P5 not detected or not connected | Pressure transducer and/or transducer adapter cable disconnected from module | warning |
No P6 transducer | Pressure transducer P6 not detected or not connected | Pressure transducer and/or transducer adapter cable disconnected from module | warning |
No P7 transducer | Pressure transducer P7 not detected or not connected | Pressure transducer and/or transducer adapter cable disconnected from module | warning |
No P8 transducer | Pressure transducer disconnected | Transducer or adapter cable disconnected from module | error |
No probe (SpO2) | SpO2 sensor not connected to module | Sensor connector not detected by acquisition module | error |
No sensor | No Entropy sensor detected | Cable connected but sensor not attached to cable | warning |
Not calibrated | CCO measurement not calibrated | Module and cable connected but calibration not performed | warning |
Not Entropy sensor | Incompatible or simulator device connected | Non-Entropy device detected at sensor port | error |
Not zeroed | Invasive pressure channel not zeroed | At least one invasive pressure channel requires zeroing before measurements | warning |
Over range | Measured FiO2 exceeds valid range | FiO2 measurement exceeds 103% | warning |
Over Scale | Displayed CO2 waveform clipped; ETCO2 value exceeds upper limit of current display scale | Measured ETCO2 exceeds maximum value displayable on selected scale | warning |
P8 over range | Measured pressure exceeds upper limit (>320 mmHg / >43 kPa) | Measured pressure value exceeds maximum range | error |
P8 under range | Measured pressure below lower limit (<-40 mmHg / <-5 kPa) | Measured pressure value below minimum range | error |
Pressure measurement removed | Invasive pressure module has been disconnected from monitor | Module physically disconnected or communication lost with pressure module | info |
Probe off (SpO2) | SpO2 sensor not functioning or not on patient | Sensor inactive or disconnected from patient | error |
PR(SpO2) high / PR(SpO2) low | Pulse rate measurement outside alarm limits | Measured pulse rate equals or exceeds configured alarm thresholds | warning |
Pulse search (NIBP) | NIBP measurement initializing or pulse detection in progress | NIBP measurement cycle starting or prolonged pulse detection | info |
RA/R Lead off | Right arm or right leg ECG lead disconnected or failed | Loss of signal from RA/R lead wire contact | warning |
Re-calibrate CCO | CCO calibration expired or requires refresh | More than 8 hours elapsed since last calibration | warning |
Recalibrate SvO2 | SvO2 calibration older than 24 hours | Elapsed time since last calibration exceeds 24 hours | warning |
Replace D-Fend | Water trap filter exhausted, saturated with condensate, or contaminated | Water trap monitoring indicates saturation or contamination detected | warning |
Replace water trap | D-fend Pro water trap requires replacement or internal tubing occluded | Defective water trap detected or internal occlusion | warning |
Sample line blocked | Continuous blockage detected in gas sampling line, water trap, or internal tubing | Obstruction detected in sampling gas delivery path | warning |
Sensor check failed | Sensor impedance exceeds acceptable threshold | Electrode impedance measurement exceeds limit during check | error |
Sensor expired | Entropy sensor exceeded usable lifespan | Sensor used > 24 hours or past expiration date | warning |
Sensor INOP | Gas sensor inoperative or not responding to gas measurements | Module unable to establish communication with sensor or sensor failure detected | error |
Sensor off | Sensor cable connected but not attached to patient | Sensor not placed on patient skin | warning |
SpO2 high / SpO2 low (NIBP) | SpO2 measurement outside alarm limits during NIBP measurement | SpO2 reading equals or exceeds alarm thresholds | warning |
SpO2 measurement removed | SpO2 acquisition module disconnected from system | Module connection lost or removed | error |
Starting up | Entropy measurement initialization in progress | Sensor check passed; measurement starting | info |
SvO2 cable off | SvO2 measurement cable disconnected | SvO2 cable unplugged from module connector | error |
SvO2 faulty cable | Optical module factory calibration corrupted or transmitter error | Red/infrared transmitter malfunction during operation | error |
SvO2 measurement removed | SvO2 module disconnected | SvO2 module unplugged or disconnected from monitor | error |
SvO2 not calibrated | SvO2 catheter not calibrated after module connection | Optical module connected without catheter calibration | warning |
SvO2 signal poor | SvO2 signal quality degraded or pulsation detected | Signal pulsation or intensity shift in signal quality | warning |
Temperature measurement removed | Temperature module has been disconnected from monitor | Module physically disconnected or communication lost with temperature module | info |
Tx temperature error | Temperature channel calibration or hardware failure detected | Automatic self-check detects calibration failure or hardware fault on temperature channel | error |
Wrong cable (NIBP/Masimo) | Incorrect cable type connected to module | Nellcor cable connected to Masimo PDM or vice versa | error |
Wrong cable. Use Masimo Set | Nellcor cable installed on Masimo PDM | Nellcor OxiMax cable detected connected to Masimo module | error |
Wrong cable. Use Nellcor OxiMax | Masimo cable installed on Nellcor PDM | Masimo Set cable detected connected to Nellcor module | error |
Zero adj > 100 mmHg | Pressure zeroing offset exceeds 100 mmHg but less than 150 mmHg | Zeroing procedure detects offset > 100 mmHg from absolute zero of module with default gain | warning |
Zeroed | Invasive pressure zeroing completed successfully | Zeroing procedure completed without errors | info |
Zero Error | Automatic zeroing of gas sensor failed | During automatic zeroing phase of gas calibration | error |
Zero error | Autozeroing failed during measurement start or gas calibration initiation | Zero valve malfunction or blockage preventing baseline zeroing | error |
Zero Failure | Automatic zeroing of invasive pressure channel failed | During automatic zeroing step of invasive pressure calibration | warning |
Zero ICP separately | ICP channel requires separate zeroing | Zero All command executed while ICP is active IP channel | warning |
Zeroing | CO2 sensor autozeroing procedure in progress | System performing baseline zeroing before measurement or calibration | info |
Zeroing failed | Invasive pressure channel zeroing procedure did not complete successfully | Pulsating waveform detected; offset exceeds 150 mmHg during zeroing | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI