GE Healthcare Carescape Gateway: troubleshooting and error codes
The ServiceForge knowledge base holds 97 documented symptoms for the GE Healthcare Carescape Gateway, with 174 ranked causes, 104 diagnostic checks with test points and expected values, and 42 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
ADT query fails - service unavailable patient safety
ADT query fails some or all of the time due to service or licensing issues
Ranked causes
- 1. Multiple ADT service providers (occasional)
- 2. No ADT licenses available or license count is zero (common)
- 3. ADT service is disabled (common)
- 4. License file may have been deleted (occasional)
- 5. License key may be invalid (occasional)
Verify only one ADT service provider
- Test point
- CIC Pro center query or Webmin ADT service configuration
- Expected
- Only one ADT service provider offering service
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot connect to Webmin
Remote computer cannot establish connection to CARESCAPE Gateway Webmin web interface.
Ranked causes
- 1. Internet Explorer proxy server settings incorrect (common)
- 2. Incorrect network settings on remote computer (common)
- 3. Remote computer on wrong network (MC or S/5) (occasional)
- 4. Network problems on Gateway (occasional)
- 5. Problem with CARESCAPE Gateway (occasional)
Verify Internet Explorer proxy server settings
- Test point
- Internet Explorer proxy settings dialog
- Expected
- Proxy settings configured per network administrator requirements
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
HSDI outbound waveform and numeric data not sent to all beds
Client not receiving real-time numeric and waveform data for one or more beds on the system
Ranked causes
- 1. CARESCAPE Gateway to client network is down (common)
- 2. Client is down or not operational (common)
- 3. No High Speed Data Interface Client Connection Count licenses available (occasional)
- 4. Too many clients are connected (occasional)
- 5. Reserved client connections have wrong IP address (rare)
Verify network connectivity
- Test point
- CARESCAPE Gateway to client network
- Expected
- Network functioning and connected
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect patient data at monitor patient safety
Patient data ADT workflow mapping error or missing ADT events
Ranked causes
- 1. Incorrect ADT workflow setting (common)
- 2. Missing ADT^A02 transfer event (occasional)
- 3. ADT^A08 update re-admits patient when should just update (occasional)
- 4. Missing ADT^A02 transfer, new transfer to non-vacated bed (rare)
- 5. Missing ADT^A03 discharge, new transfer to non-vacated bed (rare)
Verify ADT workflow mapping
- Test point
- Webmin Configuration > ADT workflow configuration
- Expected
- ADT workflow set to current customer customization
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No numeric and waveform data collected from admitted bed
Bed is admitted but not sending numeric or waveform data; bed offline or not on network
Ranked causes
- 1. Bed was never admitted from the monitor (common)
- 2. Bed does not exist on the network (common)
- 3. Network problems or monitor communication issues (common)
- 4. Bed was never added to filter list via Webmin (occasional)
- 5. Bed is not supported by CARESCAPE Gateway (rare)
Check device compatibility
- Test point
- Appendix C device list
- Expected
- Bed model listed as supported
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
VOD client fails to connect to CARESCAPE Gateway
VOD client application unable to establish connection to CARESCAPE Gateway system
Ranked causes
- 1. Ping from VOD client to CARESCAPE Gateway fails (common)
- 2. Incorrect network setup with invalid IP configurations (common)
- 3. Ping succeeds but Cloverleaf/VOD not configured for correct ports (occasional)
- 4. Firewall on VOD client host blocking traffic on VOD port (occasional)
- 5. CARESCAPE Gateway firewall incorrectly configured blocking incoming VOD connections (occasional)
Test network connectivity with ping
- Test point
- VOD client host to CARESCAPE Gateway
- Expected
- Ping response successful
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CARESCAPE Gateway does not send VOD response message
VOD client connects and sends requests but receives no response messages from system
Ranked causes
- 1. VOD request and response queues backed up with messages (occasional)
- 2. Cloverleaf not configured properly or has become corrupt (occasional)
- 3. VOD request message not well-formed HL7 or improper query type (occasional)
- 4. Cloverleaf recovery database purge delayed request (rare)
Monitor VOD queue status
- Test point
- Cloverleaf queue monitoring
- Expected
- Request and response queues not backed up
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CARESCAPE Gateway responds with VOD error message
VOD client receives error response instead of valid data; various error messages indicate specific issues
Ranked causes
- 1. VOD request message not formatted properly (occasional)
- 2. Requested device name or session not found (common)
- 3. Device found but data not available from monitor (occasional)
- 4. VOD feature not licensed (rare)
Verify device in bed list and admitted
- Test point
- CARESCAPE Gateway bedlist configuration
- Expected
- Device online and in Admitted state
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No data collected from admitted bed patient safety
CARESCAPE Gateway is not collecting data from a bed that has been admitted
Ranked causes
- 1. Bed was never admitted from the monitor (common)
- 2. Bed does not exist on network or network problems (occasional)
- 3. Bed not supported by CARESCAPE Gateway or does not offer trends (occasional)
- 4. Bed not added to filter list via Webmin (occasional)
Verify device compatibility
- Test point
- Device specifications against Appendix C
- Expected
- Bed model supported by CARESCAPE Gateway
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Total network connection failure
CARESCAPE Gateway cannot communicate on any network interface; no connectivity to remote computers or services.
Ranked causes
- 1. Swapped or loose network cables (common)
- 2. Network LEDs indicating no connection (common)
- 3. Internet Explorer cookies disabled (occasional)
- 4. Failed Network Interface Card (NIC) (occasional)
Verify network cable connections are secure and not swapped
- Test point
- RJ-45 Ethernet connectors for MC Network, IX Network, S/5 Network, iLO
- Expected
- All cables firmly seated in correct ports
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ADT request times out with no error message
Query times out and returns no matching patient records, no error displayed
Ranked causes
- 1. CCG Patient Data Cache and/or PostgreSQL Database services not running (common)
- 2. CCG_PROD not running in cloverleaf (occasional)
- 3. Permission errors in CCG_prod site log (occasional)
Verify service status
- Test point
- System Manager Webmin page
- Expected
- CCG Patient Data Cache and PostgreSQL Database services running
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot add beds to bed list
Unable to add beds to the configured bed list
Ranked causes
- 1. No licenses available or available license count is zero (common)
- 2. License file may have been deleted (occasional)
- 3. License key may be invalid (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
HL7 outbound not being sent to HIS
HL7 messages are not reaching the HIS system
Ranked causes
- 1. HIS software upgrade not compatible with Cloverleaf setup (occasional)
- 2. HIS is down or not operational (common)
- 3. Cloverleaf response times greater than 40 seconds (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
License key cannot be added via Webmin
Operation fails with message stating key was not able to be added, it is invalid in the given context
Ranked causes
- 1. License key may be invalid (common)
- 2. Access denied or CARESCAPE Gateway lacks license file access (occasional)
- 3. License bed count lower than current bed list (occasional)
Verify license key was entered correctly
- Test point
- Webmin license entry field
- Expected
- License key matches shipped documentation
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of HL7 inbound/outbound data
System may be in continuous reboot cycle; no HL7 messages flowing in or out
Ranked causes
- 1. License keys became invalid (occasional)
- 2. Disconnected CARESCAPE Network IX or MC interface cable (common)
- 3. Bad IP address configuration (occasional)
Inspect network cables
- Test point
- CARESCAPE Network IX or MC interface
- Expected
- Cables connected and undamaged
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Network connectivity lost or degraded
System cannot communicate on network; connectivity LEDs off or not showing activity; network applications unavailable
Ranked causes
- 1. Network cable disconnected or improperly connected (common)
- 2. Network interface card failure or malfunction (occasional)
- 3. Network switch or infrastructure issue (occasional)
Inspect network cable connections at back panel
- Test point
- All RJ-45 ports on CARESCAPE Gateway rear panel
- Expected
- All cables fully inserted with click engagement; no loose connectors
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient record cannot be found patient safety
ADT HL7 inbound data not being stored in Patient data cache
Ranked causes
- 1. Required services not running (Cloverleaf, Cache, PostgreSQL) (common)
- 2. Cloverleaf processes not running (common)
- 3. ADT message translation failures (occasional)
Verify Cloverleaf Engine is running
- Test point
- System Manager Webmin module
- Expected
- Cloverleaf Engine service status: running
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power is not supplied to system high voltage
System does not power on; no lights visible on front panel; system is unresponsive
Ranked causes
- 1. Power cord not plugged in or loose connection (common)
- 2. Wall outlet is not functional (occasional)
- 3. UPS device malfunction or power failure (occasional)
Verify power cord is fully connected to wall outlet and CARESCAPE Gateway
- Test point
- Power cord inlet and wall outlet
- Expected
- Power cord firmly seated at both ends; no visible gaps or loose connections
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Suspected CARESCAPE Network IX problem
CARESCAPE Network IX port is unresponsive to ping or network communication from remote computers.
Ranked causes
- 1. Network configuration incorrect on Gateway or hospital switch (common)
- 2. Failed Network Interface Card (NIC) (occasional)
- 3. Network IX port or cable defective (occasional)
Ping CARESCAPE Network IX port from remote computer
- Test point
- CARESCAPE Gateway CARESCAPE Network IX IP address
- Expected
- Successful ping response from Gateway
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Suspected CARESCAPE Network MC problem
CARESCAPE Network MC port is unresponsive to ping or network communication from remote computers.
Ranked causes
- 1. Network configuration incorrect on Gateway or hospital switch (common)
- 2. Failed Network Interface Card (NIC) (occasional)
- 3. Network MC port or cable defective (occasional)
Ping CARESCAPE Network MC port from remote computer
- Test point
- CARESCAPE Gateway CARESCAPE Network MC IP address
- Expected
- Successful ping response from Gateway
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Suspected S/5 Network problem
S/5 Network port is unresponsive to ping or network communication from remote computers.
Ranked causes
- 1. Network configuration incorrect on Gateway or hospital switch (common)
- 2. Failed Network Interface Card (NIC) (occasional)
- 3. S/5 Network port or cable defective (occasional)
Ping S/5 Network port from remote computer
- Test point
- S/5 Network IP address
- Expected
- Successful ping response from Gateway
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not boot
CARESCAPE Gateway fails to start up or enter operational mode after power application.
Ranked causes
- 1. Loose or disconnected cables and connectors (common)
- 2. Incorrect BIOS boot order configuration (occasional)
- 3. Hardware failure detected by diagnostics (occasional)
Verify power connection and cable seating
- Test point
- Power connector and all component connections
- Expected
- All cables properly connected and seated
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Account number not correct in HL7 outbound message
Account number reconciliation not populating account number in HL7 outbound message
Ranked causes
- 1. Domain field does not match ADT inbound feed from HIS (common)
- 2. Integrated ADT data feature is disabled (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Account number not in HL7 outbound message
HL7 outbound messages missing correct account number; field not populated properly
Ranked causes
- 1. Account number reconciliation not populating field (occasional)
- 2. Domain field does not match ADT inbound feed (common)
Compare CARESCAPE Gateway domain with HIS ADT feed
- Test point
- CARESCAPE Gateway configuration vs HIS system
- Expected
- Domain values match
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ADT information has incorrect PID patient safety
ADT data received with wrong patient ID from domain or typecode mismatch
Ranked causes
- 1. Incorrect Cloverleaf configuration for domain and patient ID typecode (common)
- 2. Incorrect CARESCAPE Gateway domain or patient ID typecode (common)
Compare CARESCAPE Gateway domain setting
- Test point
- Properties configuration page 7-20
- Expected
- Domain matches Cloverleaf translation
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed not appearing in collection filter list via Webmin
Bed does not appear in collection/filter list when viewed via Webmin
Ranked causes
- 1. Device not compatible or does not offer trends (occasional)
- 2. Bed disconnected from network (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed not appearing in collection/filter list
Bed does not appear in Webmin bed list or filter list even though bed is on network
Ranked causes
- 1. Bed may be disconnected from network (common)
- 2. Bed may not be supported by CARESCAPE Gateway (occasional)
Verify bed is supported model
- Test point
- Appendix C compatibility list
- Expected
- Bed model listed as compatible
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed offline in CARESCAPE Gateway
Bed appears in Webmin bed list but shows as Offline; not communicating with system
Ranked causes
- 1. Bed may be disconnected from network (common)
- 2. Bed may not be supported by CARESCAPE Gateway (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CARESCAPE Gateway does not adjust from NTP server
No time adjustment being made from NTP server
Ranked causes
- 1. NTP Server not properly configured (common)
- 2. Time Master properties not configured correctly (occasional)
Enter NTP server network address in Webmin
- Test point
- Configuration > System Time > NTPD Settings
- Expected
- Network address of valid NTP server
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CARESCAPE Gateway Service communication failure with bed/monitor
Multiple communication errors logged including socket timeouts, failed responses, device offline messages when trying to poll beds/monitors for trend data
Ranked causes
- 1. Monitor removed from network or patient discharged (common)
- 2. Network communication problems or connectivity issue (occasional)
Review log messages for communication pattern
- Test point
- CSG.log file
- Expected
- Brief errors on monitor disconnect; errors stop when monitor reconnects
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
HIS does not acknowledge HL7 message
HIS system not acknowledging HL7 outbound messages
Ranked causes
- 1. HIS software upgrade incompatible with Cloverleaf setup (occasional)
- 2. HIS is down or not operational (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No matching patient found when querying by location patient safety
Query by Unit or Unit and Bed from central station or monitor returns no results
Ranked causes
- 1. Multiple ADT service providers active (occasional)
- 2. ADT HL7 inbound data not in Patient data cache (occasional)
Determine if multiple ADT providers offering service
- Test point
- CIC Pro Clinical Information Center
- Expected
- Only one ADT service provider (CARESCAPE Gateway, MUSE Gateway, or Aware Gateway)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Not sending HL7 from some beds
Selected beds are not transmitting HL7 data
Ranked causes
- 1. Beds no longer on bed list (common)
- 2. HL7 license key file is missing or corrupt (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Not sending numeric and waveform data from some beds
Numeric and waveform data collection failing for beds that were previously on bed list; offline status
Ranked causes
- 1. Beds are no longer on the bed list (occasional)
- 2. HL7 license key file is missing or corrupt (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Performance degrades over time
Cloverleaf system performance degradation observed over extended operation
Ranked causes
- 1. HL7 outbound messages backing up at inter-process communication ports (common)
- 2. Excessive translations, data manipulations, or procedural calls (occasional)
Monitor CARESCAPE Gateway service queue
- Test point
- Webmin > Configuration > CARESCAPE Gateway Configuration > Dashboard HL7/Cloverleaf section
- Expected
- Message queue not building up
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System health indicator shows critical or degraded status
Red or amber Health LED illuminated on front panel; system status page shows fault conditions
Ranked causes
- 1. Hardware fault detected by monitoring system (unknown)
- 2. iLO management system detecting component failure (unknown)
Observe front panel Health LED color
- Test point
- Health LED on front panel
- Expected
- Green indicates system health is normal
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Too many records returned from ADT query
Querying by last name or unit name only returns more records than desired
Ranked causes
- 1. Many patient records in database (common)
- 2. Old patient records in database not cleaned (occasional)
Evaluate query specificity
- Test point
- Query input fields
- Expected
- Last name AND first name or unit AND bed specified
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to request ADT data - no data server present patient safety
Message displayed at monitoring device indicates no data server
Ranked causes
- 1. Invalid or corrupt ADT license (common)
- 2. CARESCAPE Network MC misconfigured or network connectivity lost (occasional)
Verify ADT license is valid and current
- Test point
- Webmin License screen, ADT license configuration page 7-28
- Expected
- Valid ADT license for current MAC address present and active
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Webmin properties page displays Serious Error
Properties page malfunctions showing Serious Error and referring to log files
Ranked causes
- 1. ugsUser.properties file size is zero (common)
- 2. Restore configuration defaults (occasional)
Check ugsUser.properties file size
- Test point
- /usr/gehc/csg/resources/ugsUser.properties
- Expected
- File size > 0 bytes
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Administrator login to iLO password not accepted
iLO authentication fails for administrator account despite correct password entry.
Ranked causes
- 1. iLO password corrupted or needs reset (occasional)
Reset iLO password through Webmin
- Test point
- Webmin Lights-Out Management interface
- Expected
- Password reset successfully; administrator login accepted
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ADT functionality disabled due to missing or invalid license
ADT service turned off when no valid license key available
Ranked causes
- 1. No valid ADT license key configured (common)
Verify ADT license in Webmin
- Test point
- Webmin Configuration > License section
- Expected
- Valid ADT license key present
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ADT information has default PID patient safety
No matching domain or patient ID typecode found in ADT Query Response
Ranked causes
- 1. No matching domain or patient ID typecode in response (occasional)
ADT query returns data with question mark characters
Data returned contains question mark characters in place of expected values
Ranked causes
- 1. Non-ASCII character in patient data in database cannot map (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ADT query returns intermittent unit/bed not found error
Some messages returned indicate unit or bed not found intermittently
Ranked causes
- 1. More than one ADT service provider (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Advanced iLO does not work
Advanced iLO functionality unavailable after factory default reset or battery replacement with server unplugged.
Ranked causes
- 1. iLO reset to factory defaults or lost power during battery replacement (occasional)
Restart or reset iLO functionality
- Test point
- Webmin Lights-Out Management interface
- Expected
- iLO restarted or reset; advanced features available
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
All Cloverleaf threads stopped
All threads are in stopped state, system requires manual intervention to restart processes
Ranked causes
- 1. Threads stopped for unknown reason (unknown)
Verify thread status in Interface Monitor
- Test point
- Interface Monitor
- Expected
- All threads running
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Applications or subsystems not running patient safety
CARESCAPE Gateway applications show status Off; services not responding; system appears to be running but no patient monitoring or data available
Ranked causes
- 1. Application services stopped or disabled (common)
Verify all CARESCAPE applications show status On in Webmin
- Test point
- Webmin Start/Stop CARESCAPE Gateway page
- Expected
- All required applications and subsystems showing status On
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Applications will not start
Attempts to start Cloverleaf Engine fail
Ranked causes
- 1. /var or /tmp are full (common)
Check disk space on /var and /tmp
- Test point
- /var, /tmp filesystems
- Expected
- Less than 100% full
- Tools
- df command
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed offline; CARESCAPE Gateway not detecting patient safety
Bed appears in collection list via Webmin but is noted as Offline
Ranked causes
- 1. Device not compatible or does not offer trends (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot connect to iLO, Webmin or VNC
Remote management interfaces (iLO, Webmin, VNC) are not accessible from remote computer or KVM.
Ranked causes
- 1. iLO network configuration not set up (common)
Local iLO configuration via keyboard and monitor
- Test point
- iLO setup during system boot
- Expected
- iLO network settings configured and saved
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot resend failed message via Configuration Network Monitor
Error displayed when attempting to resend failed message; cannot retrieve process name for CCG_Inbound
Ranked causes
- 1. Communication problem between service interface and Cloverleaf (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CARESCAPE Gateway application will not start
System Manager page shows CARESCAPE Gateway application as stopped and multiple restart attempts have failed
Ranked causes
- 1. Corrupt configuration file (common)
Verify application configuration integrity
- Test point
- CARESCAPE Gateway configuration files
- Expected
- Valid configuration present and readable
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CARESCAPE Gateway not sending network data
System configuration reviewed and correct, services restarted but data not transmitted
Ranked causes
- 1. Firewall or router blocking network connections (common)
Restart CARESCAPE Gateway firewall
- Test point
- Configuration > System Manager in Webmin
- Expected
- Firewall service restarted
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CARESCAPE Gateway OS time wrong
Service disabled or time outside 8 hour maximum adjustment window
Ranked causes
- 1. Service disabled or time outside 8 hour window (occasional)
Verify time difference from NTP is within 8 hours
- Test point
- System clock vs NTP server time
- Expected
- Time difference less than 8 hours
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CARESCAPE Gateway Service exit with code 20
Service shuts down unable to establish TCP connections to Enterprise Gateway
Ranked causes
- 1. Enterprise Gateway system down or unreachable (common)
Verify Enterprise Gateway TCP connectivity
- Test point
- Network connectivity to EG IP address and port
- Expected
- TCP connection successful on expected port
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CARESCAPE Gateway Service exit with code 30
Service shuts down unable to establish TCP connections to Cloverleaf
Ranked causes
- 1. Cloverleaf system down or unreachable (common)
Verify Cloverleaf TCP connectivity
- Test point
- Network connectivity to Cloverleaf IP address and port
- Expected
- TCP connections established; default is 20 connections
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CARESCAPE Gateway Service exit with code 50
Service shuts down due to Cloverleaf being in indeterminate state and unable to communicate
Ranked causes
- 1. Cloverleaf internal error or crash (occasional)
Monitor Cloverleaf connection status
- Test point
- CSG.log Cloverleaf Connection messages
- Expected
- Cloverleaf Connection xx created messages appear for all expected connections
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CARESCAPE Gateway service not running
Configuration link in Webmin module returns 'Failed to connect to 127.0.0.1:9093: Connection refused'
Ranked causes
- 1. CARESCAPE Gateway service stopped (common)
Verify CARESCAPE Gateway service status
- Test point
- Webmin service management
- Expected
- Service running and responding on 127.0.0.1:9093
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CARESCAPE Gateway Service shutdown due to command queue overflow
Service exits with code 60 when internal queue depth exceeds maximum allowed commands
Ranked causes
- 1. Cloverleaf slow or unable to process HL7 messages (common)
Monitor command queue depth in logs
- Test point
- CSG.log Total queue depth message
- Expected
- Queue depth below maximum allowed commands threshold
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CARESCAPE Gateway time set incorrectly
Monitor time wrong, time service not running
Ranked causes
- 1. Time Master properties configured incorrectly (common)
Verify Time Master Properties in Webmin
- Test point
- Configuration > CARESCAPE Gateway Configuration > Setup > Properties
- Expected
- All Time Master settings correct per configuration document
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- CARESCAPE Network device clocks fluctuate between times
- Cloverleaf IDE becomes unresponsive
- Cloverleaf IDE unable to launch separate process
- Cloverleaf interface threads not restarting
- Cloverleaf Monitor daemon restarts automatically
- Cloverleaf response times exceed 40 seconds
- Desktop GUI cannot perform all functions
- Duplicate UNIT|BED name detected on network
- HL7 Outbound data not being sent to HIS
- HL7 outbound HIS connection failure after HIS restart
- HSDI outbound numeric data not sent to all beds
- HSDI outbound waveform data not sent to all beds
- Incorrect time propagated to network
- Interface Monitor/Management unavailable
- Internal database vista error recovery
- Internal thread errors in down state
- Invalid license key in license file
- Left mouse click does not work in VNC
- Logging has stopped
- Monitor time wrong after daylight savings
- No time update from CARESCAPE Network master
- No time update from NTP server
- One or few Cloverleaf threads stopped
- Patient data with HIS mapping causing wrong location query
- Patient record lookup in ADT database fails
- PCD-01 HL7 trend messages missing EEG site information
- Response is not in requested HL7 version
- Some beds not being collected in real time
- Some beds not collected in real time
- Statistics subsystem unable to record performance data
- Threads and processes in down state refuse to start
- Time at CARESCAPE Gateway does not match network time
- Translation or database error on ADT admit event
- Trended data not showing after HIS mapping update
- Trended PA values not being delivered
- Unit name lookup in ADT database fails
- Unsupported combination of patient query fields
37 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
ADT-Communication-Error-CSG-PDC | The CSG application is not communicating with Patient Data Cache (PDC), resulting in an ADT failure | CSG application loses communication with PDC | error |
ADT-Communication-Recovered-CSG-PDC | The CSG application starts communicating with the Patient Data Cache (PDC), having recovered from the previous failed state | CSG application resumes communication with PDC | info |
Cannot process ADT request | ADT service disabled due to missing or invalid license key | ADT data request received when ADT license not available or invalid | warning |
CIS-Connection-Error | The system shall notify when the Clinical Information System has been disconnected for 20 minutes unexpectedly | CIS disconnects for 20 minutes | warning |
Cloverleaf License Issue | Cloverleaf license not valid or not installed | System startup or runtime when Cloverleaf module is accessed | warning |
cpqDa6LogDrvStatusChange | Logical drive status changes | Logical drive status changes | error |
cpqDa7PhyDrvStatusChange | Physical drive status changes | Physical drive status changes | error |
cpqHe3FltTolFanDegraded | Fan degraded | Fan performance degrades | warning |
cpqHe3FltTolFanFailed | Fan failed | Fan fails | warning |
cpqHe3FltTolPowerRedundancyLost | Power redundancy lost | Power redundancy fails | warning |
cpqHe3FltTolPowerRedundancyRestored | Power supply redundancy is restored | Power redundancy restored after loss | info |
cpqHe3FltTolPowerSupplyInserted | Power supply inserted | Power supply inserted into system | info |
cpqHe3FltTolPowerSupplyRemoved | Power supply removed | Power supply removed from system | warning |
cpqHe3ThermalCpuFanFailed | CPU fan failed | CPU fan fails | error |
cpqHe3ThermalCpuFanOk | CPU fan transitions from failed to OK | CPU fan recovers from failed state | info |
cpqHe3ThermalSystemFanDegraded | System fan degraded | System fan performance degrades | warning |
cpqHe3ThermalSystemFanFailed | System fan failed | System fan fails | warning |
cpqHe3ThermalSystemFanOk | System fan transitions from failed to OK | System fan recovers from failed state | info |
cpqHe4FltTolPowerSupplyACpowerloss | Power supply AC power loss | AC power input is lost | warning |
cpqHe4FltTolPowerSupplyDegraded | Power supply degraded | Power supply performance degrades | warning |
cpqHe4FltTolPowerSupplyFailed | Power supply failed | Power supply fails | warning |
cpqHe4FltTolPowerSupplyOk | Power supply transitions from failed to OK | Power supply recovers from failed state | info |
cpqHo2GenericTrap | Generic trap | Hardware fault condition detected | info |
cpqHo2NicStatusFailed | NIC has failed | Network interface card fails | error |
cpqHo2NicStatusOk | NIC status transitions from failed to OK | NIC recovers from failed state | info |
cpqNic3ConnectivityLost | Network interface has lost connectivity | Network connectivity is lost | error |
cpqNic3ConnectivityRestored | Network connectivity restored | Network connectivity restored after loss | info |
cpqNic3RedundancyIncreased | Ethernet interface team's redundancy restored | Ethernet interface redundancy restored | info |
cpqNic3RedundancyReduced | Ethernet interface team loses redundancy | Ethernet interface redundancy fails | warning |
Critical-Process-Restart-Threshold | Watchdog fails to restart a process | Watchdog unable to restart critical process | error |
Duplicate UNIT|BED detected | Another device with same unit/bed name found on network | CARESCAPE Gateway discovers duplicate unit|bed identifier during device scanning | warning |
Exit Code 20 | Unable to make TCP socket connections to Enterprise Gateway (EG down) | CARESCAPE Gateway cannot establish TCP connection to Enterprise Gateway system at startup or runtime | critical |
Exit Code 30 | Unable to make TCP socket connections to Cloverleaf (Cloverleaf down) | CARESCAPE Gateway cannot establish TCP connections to Cloverleaf system at startup or runtime | critical |
Exit Code 50 | Cloverleaf in indeterminate state unable to communicate with CARESCAPE Gateway Service (Cloverleaf down) | CARESCAPE Gateway detects Cloverleaf communication failure indicating system crash or internal error | critical |
Exit Code 60 | CARESCAPE Gateway collecting trend data faster than can be sent to Cloverleaf (Cloverleaf down or slow) | Internal command queue depth exceeds predefined maximum threshold | critical |
HL7/ADT License Issue | HL7 or ADT license not valid or not installed | System startup or runtime when HL7/ADT module is accessed | warning |
HL7-Communication-Error-CSG-Cloverleaf | The CSG application is not communicating with Cloverleaf, resulting in an HL7 failure | CSG application loses communication with Cloverleaf | error |
HL7-Communication-Recovered-CSG-Cloverleaf | The CSG application starts communicating with Cloverleaf, having recovered from the previous failed state | CSG application resumes communication with Cloverleaf | info |
HSDI-Client-Connection-Error | A client unexpectedly disconnected from the High Speed Data Interface (HDSI) | Client disconnects from HDSI | warning |
INVALID license key | License keys in licensefile.txt do not match MAC address in macaddress.txt | CARESCAPE Gateway starts and validates license keys against stored MAC address | critical |
Time-Out-of-Sync-Recovered | The time update to the time master has recovered after a previous error | Time synchronization restored after failure | info |
Time-Out-of-Sync-Threshold | The system is unable to update the time master's time for an hour | Time synchronization fails for one hour | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.