GE Healthcare CARESCAPE Canvas: troubleshooting and error codes
The ServiceForge knowledge base holds 20 documented symptoms for the GE Healthcare CARESCAPE Canvas, with 40 ranked causes, 24 diagnostic checks with test points and expected values, and 1 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Display does not power on
Display unit will not turn on when connected to power or remains unresponsive to power input despite connection
Ranked causes
- 1. Power cord loose at wall outlet or display connection (common)
- 2. Power cord damaged or faulty (occasional)
- 3. Power outlet does not meet specification requirements (occasional)
- 4. Button board defective or cable disconnected (rare)
- 5. Power supply cable loose or faulty between power supply and display (rare)
- 6. Power supply unit is faulty (rare)
- 7. Display board is faulty (rare)
Verify power cord connection to wall outlet and display unit
- Expected
- Power cord fully seated and snug at both ends
- Tools
- Visual inspection
6 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Unable to turn on the display caution
Monitor display does not power on or illuminate
Ranked causes
- 1. Power cord loose (common)
- 2. Power cord faulty (occasional)
- 3. Power outlet does not meet specified requirements (occasional)
- 4. Button board defective or cable loose (occasional)
- 5. Power supply cable loose or faulty (occasional)
- 6. Power supply faulty (rare)
- 7. Display board faulty (rare)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Module functionality problem
Monitor displays error messages or module malfunction; system modules not identified correctly; parameters not configured to display.
Ranked causes
- 1. Module not connected or not identified (common)
- 2. Identical modules connected (occasional)
- 3. Accessory incompatibility or damage (occasional)
- 4. Screen configuration incorrect (occasional)
Verify monitor power and module connections
- Test point
- All modules
- Expected
- Monitor on; all modules connected
7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarms do not work patient safety
Monitor produces no audible alert or alarm sounds
Ranked causes
- 1. Defective speaker (common)
- 2. Speaker cable loose or faulty (common)
- 3. Audio amplifier failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to configure D19 display independently
Cannot configure D19 as standalone display without parent monitor
Ranked causes
- 1. Dual Video license not purchased (common)
- 2. Dual Video trial license expired (occasional)
- 3. Dual Video license not installed properly (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarms do not function patient safety
Speaker does not produce audible tones for alarm notifications or system alerts
Ranked causes
- 1. Speaker unit is defective (common)
- 2. Speaker cable loose or faulty connection (occasional)
Inspect speaker cable for damage and verify connection
- Expected
- Speaker cable intact, properly seated, no visible damage or corrosion
- Tools
- Visual inspection
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Check device or SpO2 malfunction patient safety
Acquisition module has failed; no SpO2 signal or device error displayed in parameter window or alarm area
Ranked causes
- 1. SpO2 acquisition module failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check probe or Check SpO2 probe patient safety
No detectable SpO2 signal; sensor faulty, loose, or detached from patient; messages in parameter window or alarm area
Ranked causes
- 1. Sensor disconnected or faulty (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Faulty probe or SpO2 faulty probe patient safety
Sensor has failed; message displayed in parameter window or alarm area
Ranked causes
- 1. SpO2 sensor failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Identical SpO2(2) modules patient safety
Two or more identical SpO2 modules installed in system; message in alarm area
Ranked causes
- 1. Multiple identical SpO2 modules installed (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Identical SpO2(2) sources patient safety
E-module and UNID SpO2 sources detected simultaneously; message in alarm area
Ranked causes
- 1. Duplicate SpO2 signal sources configured (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incompatible probe or Incompatible SpO2 probe patient safety
Sensor is not compatible with acquisition module; message in parameter window or alarm area
Ranked causes
- 1. Incompatible sensor installed (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No probe or No SpO2(2) probe patient safety
Sensor not connected to acquisition module; message in parameter window or alarm area
Ranked causes
- 1. SpO2 sensor not connected (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Planned maintenance procedure patient safety
Routine maintenance every 2 years to ensure operational and functional order, essential performance and basic safety
Ranked causes
- 1. Maintenance not performed at recommended 2-year interval (common)
Visual inspection
- Test point
- Module exterior and components
- Expected
- No visible damage or wear
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power cord and display cable strain relief not secured caution
Strain relief for power cord, display cable, and USB cables not properly secured using cable holder.
Ranked causes
- 1. Strain relief not properly fastened to cable holder (common)
Strain relief fastened at power cord entry
- Test point
- Power cord strain relief and cable holder
- Expected
- Strain relief properly secured in cable holder
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PR(SpO2(2)) high or PR(SpO2(2)) low patient safety
Pulse rate measurement values equal to or outside alarm limits; message in alarm area
Ranked causes
- 1. Pulse rate measurement outside normal range (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe off or SpO2(2) probe off patient safety
Sensor may be defective or not connected to patient; message in parameter window or alarm area
Ranked causes
- 1. Sensor disconnected from patient or defective (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pulse search patient safety
Measurement starting normally; defective sensor or cable, or sensor off patient; pulse detection has stopped; message in parameter window
Ranked causes
- 1. Sensor or cable defective; sensor off patient; pulse lost (common)
Wait for waveforms and parameter values to display
- Test point
- Monitor display
- Expected
- SpO2 waveforms and values appear on screen
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2(2) high or SpO2(2) low patient safety
SpO2 measurement values equal to or outside alarm limits; message in alarm area
Ranked causes
- 1. SpO2 measurement outside normal range (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2(2) measurement removed patient safety
Acquisition module is disconnected; message in alarm area
Ranked causes
- 1. SpO2 acquisition module disconnected (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Error messages in message field | System detected error or alarm condition | Module malfunction, incompatibility, or configuration issue | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI