GE Healthcare Carescape Canvas 1000: troubleshooting and error codes
The ServiceForge knowledge base holds 319 documented symptoms for the GE Healthcare Carescape Canvas 1000, with 527 ranked causes, 414 diagnostic checks with test points and expected values, and 63 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Acquisition module data not displayed on monitor patient safety
Module data not shown on monitor display; parameter or waveform missing from screen
Ranked causes
- 1. Acquisition module incompatible with system (occasional)
- 2. Module not inserted properly into rails (common)
- 3. Monitor not configured to display parameter (common)
- 4. Faulty or improperly connected frame-to-host cable (occasional)
- 5. Internal frame or EMBC connector defective or loose (occasional)
- 6. Module connector or EMBC board defective (occasional)
- 7. DC/DC board defective (rare)
- 8. Module defective (see module troubleshooting) (occasional)
Verify module compatibility
- Test point
- Module and system documentation
- Expected
- Module listed as compatible with Carescape Canvas 1000
4 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Acquisition module data not shown on monitor display
Parameter data from acquisition module (CARESCAPE ONE or E-module) does not appear on host monitor screen
Ranked causes
- 1. Acquisition module incompatible (occasional)
- 2. Module not connected properly (common)
- 3. Monitor not configured to display parameter (occasional)
- 4. Cable connecting F2 Frame to host defective or loose (occasional)
- 5. F2 Frame internal connectors defective or loose (rare)
- 6. F2 Frame main board defective (rare)
- 7. AC/DC module defective (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on high voltage
Monitor fails to turn on; no display or indicators active
Ranked causes
- 1. Power cord is loose (common)
- 2. Power cord is faulty (occasional)
- 3. Power outlet does not meet specifications (occasional)
- 4. Button board is defective or cable is loose (rare)
- 5. Power supply cable loose or faulty (rare)
- 6. Power supply is faulty (rare)
- 7. CPU assembly is faulty (rare)
Verify power cord connection
- Test point
- Wall outlet and monitor connector
- Expected
- Cord fully inserted and seated
1 more check with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
E-module or PDM does not work correctly when connected
Module fails to function or communicate properly after connection to monitor frame
Ranked causes
- 1. Identical modules connected simultaneously (common)
- 2. PDM and CS ONE connected simultaneously (common)
- 3. Configuration issue (occasional)
- 4. Compatibility issue with module or accessories (occasional)
- 5. Frame-CPU cable loose or defective (occasional)
- 6. Defective acquisition module (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement removed from display (XXXX measurement(s) removed) patient safety
One or more measurements removed from display; module or measurement parameters no longer displayed
Ranked causes
- 1. Module not connected properly (common)
- 2. Module connector defective (occasional)
- 3. Frame not properly connected to host (occasional)
- 4. EMBC board connectors loose or defective (occasional)
- 5. EMBC board defective (occasional)
- 6. DC/DC board defective (occasional)
Verify module seating in frame rails
- Test point
- Module and frame rails
- Expected
- Module fully inserted; connectors aligned; latches engaged
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurements removed from display
System message shows XXXX measurement(s) removed, where XXXX is module name. Module data not displaying on monitor screen.
Ranked causes
- 1. Acquisition module not connected properly (common)
- 2. Module connector defective (occasional)
- 3. Frame to host cable defective or loose (occasional)
- 4. EMBC board connectors loose or defective (occasional)
- 5. EMBC board defective (rare)
- 6. DC/DC board defective (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nonphysiological VO2 readings patient safety
Oxygen consumption (VO2) measurements display clinically implausible values
Ranked causes
- 1. Unstable O2 delivery or gas mixer issue (common)
- 2. Respiratory rate over 35/min (occasional)
- 3. Reference gas inlet port blocked (occasional)
- 4. Gas sampling line longer than 2 m (occasional)
- 5. Dead space of Y-piece exceeds 8 ml (occasional)
- 6. Gas sampling line connected to HME or D-lite incorrectly placed (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormally low invasive pressure reading patient safety
Pressure measurement is lower than expected or abnormally low
Ranked causes
- 1. Transducer wrongly positioned (common)
- 2. Defective transducer (occasional)
- 3. Module not connected (occasional)
- 4. Transducer adapter cable not connected to module (occasional)
- 5. Invasive pressure channel not configured to screen (occasional)
Verify transducer position
- Test point
- Transducer location
- Expected
- Transducer at mid-heart level
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Acquisition platform not recognized by host monitor
CARESCAPE ONE, CARESCAPE Dock F0, or parameters not displayed on host monitor; platform not communicating
Ranked causes
- 1. PDM and CARESCAPE ONE connected simultaneously (common)
- 2. CARESCAPE parameters not shown due to configuration issue (occasional)
- 3. Cable between dock/frame and host monitor is loose or defective (occasional)
- 4. CARESCAPE Dock F0 or F2 Frame is defective (rare)
- 5. CPU assembly in host monitor is defective (rare)
Verify CARESCAPE ONE works standalone
- Test point
- CARESCAPE ONE disconnected from host or F2 Frame
- Expected
- CARESCAPE ONE operates normally as standalone monitor
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarms do not work patient safety
Alarm sounds not activating or not heard during alarm conditions
Ranked causes
- 1. Audible alarms are turned off (common)
- 2. Alarm volume is too low (common)
- 3. Defective speaker (occasional)
- 4. Speaker cable is loose or faulty (occasional)
- 5. Tone generator or audio amplifier failure (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loose parts or cable connections inside module caution
Module may have internal loose parts, improperly tightened screws, disconnected cables, pinched or bent tubes, or loose objects rattling inside.
Ranked causes
- 1. Screws not tightened properly (common)
- 2. Cables disconnected or improperly seated (common)
- 3. Tubes pinched or sharply bent (common)
- 4. Tubes disconnected (common)
- 5. Loose objects inside module (occasional)
Verify all screws are tightened
- Test point
- All screws in module
- Expected
- Tight, no rotation
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Module functionality test fails patient safety
BIS module does not function correctly; error messages appear or test DSC fails.
Ranked causes
- 1. BISx unit defective (common)
- 2. Incompatible system component (occasional)
- 3. Identical modules connected (occasional)
- 4. Defective or incompatible accessories (occasional)
- 5. Loose keypad cable or module connector problem (rare)
Verify system component compatibility
- Test point
- Monitor and all connected modules
- Expected
- All components listed in supplemental information
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No IX network traffic
Monitor has no network communication on IX port; unable to transmit or receive data
Ranked causes
- 1. Incorrect configuration (common)
- 2. Cable connections do not match network type (common)
- 3. IX network traffic overload detected (rare)
- 4. IX port authentication enabled but monitor misconfigured (rare)
- 5. CA certificate has expired (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No MC network traffic
Monitor has no network communication on MC port; unable to transmit or receive data
Ranked causes
- 1. Incorrect configuration (common)
- 2. Cable connections do not match network type (common)
- 3. MC network traffic overload detected (rare)
- 4. MC port authentication enabled but monitor misconfigured (rare)
- 5. CA certificate has expired (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sample gas outflow blocked caution
Gas exhaust line or sample output path is obstructed; measurements may be affected.
Ranked causes
- 1. Sample gas out connector or exhaust line blocked (common)
- 2. Patient circuit tubing occluded (occasional)
- 3. Scavenging system improperly configured (occasional)
- 4. Internal tubing blocked (occasional)
- 5. Pump failure (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration error
Gas sensor calibration failed; measurements are unavailable until recalibration succeeds.
Ranked causes
- 1. Calibration gas not fed within one minute (common)
- 2. CO2 trapped in sensor during transport (occasional)
- 3. Calibration gain adjustment too large (rare)
- 4. Wrong calibration gas used (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Continuous gas sampling line blockage caution
Gas sampling line or water trap is persistently blocked; continuous message shown.
Ranked causes
- 1. Gas sampling line blocked (common)
- 2. Water trap container full (common)
- 3. Water trap occluded (occasional)
- 4. Internal tubing blocked (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrical safety verification high voltage
Comprehensive electrical safety testing including power outlet, power cord, ground integrity, earth leakage, and touch current measurements under normal and single fault conditions.
Ranked causes
- 1. Power outlet improperly wired or damaged power cord (common)
- 2. Inadequate ground continuity or impedance (occasional)
- 3. Excessive earth leakage current (occasional)
- 4. Excessive touch current (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas sampling system leaks or blockages patient safety
Faults in gas sampling system including leaks, blockages, pump failure, valve failure, reduced flow rates, or sensor issues
Ranked causes
- 1. Leaks in tubing or blockages (occasional)
- 2. Sampling pump or magnetic valve failure (occasional)
- 3. Dirt or matter accumulation in tubing reducing flow (occasional)
- 4. Pressure sensor failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspiratory TV greater than expiratory TV patient safety
Spirometry measurement shows inspiratory tidal volume larger than expiratory, indicating system leak
Ranked causes
- 1. Leak in lungs or ET tube cuff (common)
- 2. Spirometry tube leak (common)
- 3. Water inside D-lite or tubings (occasional)
- 4. D-fend Pro leaks (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
MiniOM oxygen measuring unit malfunction patient safety
No response to O2 signal or strong signal drift; incorrect oxygen readings or unstable measurements
Ranked causes
- 1. Loose tubing connections (common)
- 2. Tubing blockage or occlusion (common)
- 3. Reference gas filter assembly contaminated or clogged (occasional)
- 4. Internal measurement unit fault (rare)
Inspect all tubing connections on oxygen measuring unit
- Test point
- Tubing connectors
- Expected
- All connections seated firmly with no visible gaps
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Module physical damage or contamination
Front cover cracked, connectors dirty or loose, casing damaged, latch broken, patient cables worn or dirty
Ranked causes
- 1. Front cover not intact (unknown)
- 2. Connectors dirty, damaged or loose (unknown)
- 3. Module casing or latch damaged (unknown)
- 4. Patient cables compromised (unknown)
Inspect front cover for cracks, breaks or damage
- Test point
- Front cover exterior
- Expected
- Intact, no cracks or breaks
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Module physical inspection before troubleshooting
Visual inspection procedure to verify module structural integrity, connectors, latch mechanism, and D-fend connector condition before detailed diagnostics.
Ranked causes
- 1. Damaged front cover or housing (unknown)
- 2. Connector damage, contamination, or improper seating (unknown)
- 3. Latch mechanism damage or contamination (unknown)
- 4. Mini D-fend latch or connector problems (unknown)
Remove module from system
- Test point
- Module mounting
- Expected
- Module removed without damage
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Response too weak message
Parameter window shows 'Response too weak'; maximum gain insufficient to measure response signal amplitude
Ranked causes
- 1. Stimulus electrodes are loose (common)
- 2. Response electrodes attached to wrong place (common)
- 3. Electrodes are dry (occasional)
- 4. Skin at electrode site not properly prepared (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sample line blocked message displayed
Alarm area, parameter window, and waveform area display 'Sample line blocked' message indicating continuous blockage in the CO2 sampling pathway.
Ranked causes
- 1. Gas sampling line is blocked (common)
- 2. Water trap container is full (common)
- 3. Water trap is occluded (occasional)
- 4. Internal tubing is blocked (occasional)
Inspect external gas sampling line for blockages
- Test point
- External gas sampling line
- Expected
- Clear, unobstructed pathway
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sampling system leaks or blockages
Gas sampling system may have leaks in tubing, blockages, or flow rate diminishing due to dirt or contamination.
Ranked causes
- 1. Leaks in sampling system tubing (common)
- 2. Blockages in tubing or internal passages (common)
- 3. Sampling pump failure or wear (occasional)
- 4. Magnetic valve failure (occasional)
Listen for sampling pump noise
- Test point
- Sampling pump
- Expected
- Quiet operation, no grinding or rattling
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to send or retrieve MUSE 12SL reports
Cannot send or retrieve ECG reports to/from MUSE Web Server
Ranked causes
- 1. Incorrect MUSE Web Server username or password (common)
- 2. Incorrect MUSE Web Server address (common)
- 3. AMSE certificate not installed (occasional)
- 4. MUSE certificate not installed or invalid (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Visual inspection for frame and module condition
Routine inspection of physical frame integrity, connectors, and attached modules before troubleshooting
Ranked causes
- 1. Damaged cover or module rails (occasional)
- 2. Loose or dirty connectors (occasional)
- 3. Loose or damaged cables (occasional)
- 4. Modules not properly locked to frame (occasional)
Inspect frame cover and module rails
- Test point
- Frame exterior
- Expected
- Cover and rails intact, no visible damage
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
BIS sensor module degradation or malfunction
Cover stickers peeling, panel damaged, cables loose, internal components dirty or corroded
Ranked causes
- 1. Cover or panel stickers not intact (unknown)
- 2. Cables or connections not intact on BISx unit (unknown)
- 3. Module or applied parts not clean (unknown)
Inspect BISx unit cover and panel stickers for damage or peeling
- Test point
- BISx unit cover and sticker areas
- Expected
- Intact, not peeling or damaged
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
BIS sensor not responding or failing electrode check patient safety
Sensor checking message appears but green checkmarks not shown for electrodes; sensor test fails
Ranked causes
- 1. BIS sensor simulator not connected or faulty (common)
- 2. PIC Plus cable not intact or poorly seated (occasional)
- 3. BISx unit internal test failing (occasional)
Connect BIS sensor simulator to PIC Plus cable and wait for electrode checking to start
- Test point
- BIS sensor simulator and PIC Plus cable connection
- Expected
- Checking sensor message appears in BIS parameter window
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration error message displayed
Parameter window and calibration menu display 'Calibration error' message indicating calibration procedure failed due to timing, gain, or gas issues.
Ranked causes
- 1. Calibration gas not started within 1 minute of autozero completion (common)
- 2. Calibration failed due to too large gain adjustment (occasional)
- 3. Wrong calibration gas used (occasional)
Verify calibration gas was started within 1 minute of autozero completion
- Test point
- Calibration gas inlet
- Expected
- Gas flow initiated within 1 minute of zero completion
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration error on invasive pressure channel patient safety
Message displayed as Calibration error in parameter window during channel calibration
Ranked causes
- 1. Pressure calibration failure due to time-out (common)
- 2. Pulsating waveform detected during calibration (occasional)
- 3. Gain beyond limits (±20% of default gain) (occasional)
Start calibration within time window
- Test point
- pressure calibration timer
- Expected
- Inflation starts within 45 seconds of auto-zeroing completion
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Invasive pressure patient safety
Invasive pressure measurement requires calibration when transducer is replaced, calibration check fails, or measured values exceed specification limits
Ranked causes
- 1. Pressure transducer replaced (common)
- 2. Invasive pressure calibration check failed (occasional)
- 3. Measured value outside specification limits (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check sample gas out message displayed
Alarm area and waveform area display 'Check sample gas out' message indicating blockage in the sample gas outflow path.
Ranked causes
- 1. The sample gas outflow is blocked (common)
- 2. Internal tubing is blocked (occasional)
- 3. Pump failure (occasional)
Inspect sample gas out connector and exhaust line
- Test point
- Front panel sample gas out connector, exhaust/scavenging line
- Expected
- No blockages, clear pathway for gas return or scavenging
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 waveform and values not updating patient safety
CO2 waveform does not appear or EtCO2/FiCO2 values do not update when breathing into sampling line
Ranked causes
- 1. Gas sensor warm-up not complete (common)
- 2. Gas calibration required (common)
- 3. CO2 measurement source not configured (occasional)
Gas sensor warm-up status
- Test point
- Parameter window message display
- Expected
- No 'Calibrating Gas Sensor' message; waveform field shows CO2 waveform
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ePort pod to host cable connection issue
PDM module connected via ePort connector not communicating or recognized
Ranked causes
- 1. ePort cable loose or defective (common)
- 2. Defective PDM or accessories (occasional)
- 3. Defective CPU assembly (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Expired volume much smaller than inspired volume
MVexp << MVinsp; Ppeak - PEEP < 2 cmH2O; TVinsp and TVexp are DATA_INVALID. Leak in spirometry system causing expired volume 70% or less of inspired volume.
Ranked causes
- 1. Leaking spirometry tube (common)
- 2. Water in tubing (common)
- 3. MiniPVX sensor failure (occasional)
Perform spirometry leak test
- Test point
- Spirometry system
- Expected
- No leaks detected
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid or unreachable monitor address
Monitor cannot be contacted due to incorrect IP address, hostname, or network connectivity issues
Ranked causes
- 1. Incorrect IP address or hostname (common)
- 2. Monitor turned off or not connected to IX network (common)
- 3. Network configuration error or network failure (occasional)
Verify IP address or hostname
- Test point
- Monitor Addresses field
- Expected
- Valid IP address or hostname matching target monitor
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invasive pressure measurement over range patient safety
Pressure reading displayed as >320 mmHg or >43 kPa in parameter window
Ranked causes
- 1. Measurement is over range or sensor/cable faulty (common)
- 2. Transducer not zeroed correctly (common)
- 3. Pre-zero with stopcock closed (occasional)
Verify measurement value
- Test point
- parameter window
- Expected
- >320 mmHg or >43 kPa
2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loop overscale in spirometry patient safety
Spirometry waveform display exceeds the selected vertical scale limits
Ranked causes
- 1. Wrong scale selected (common)
- 2. Wrong TV base selected (occasional)
- 3. Leak between ventilator and D-lite (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low gas sample flow detected patient safety
Sample flow drops below 80% of module nominal specification
Ranked causes
- 1. Gas sampling line or tubing blocked (common)
- 2. Water trap blocked or full (common)
- 3. Pump failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement unit failure or malfunction caution
CPU board, MiniOM measuring unit, or MiniTPX measuring unit malfunction; these units are not field serviceable.
Ranked causes
- 1. CPU board failure (rare)
- 2. MiniOM measuring unit failure (rare)
- 3. MiniTPX measuring unit failure (rare)
Identify failed component
- Test point
- Affected measurement unit or circuit board
- Expected
- Normal operation or no error codes
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NMT measurement output incorrect or waveforms not displaying
TOF% value not within 95-105, Count not equal to 4, T1% not within 95-105, or stimulus pulses/responses not visible on waveform
Ranked causes
- 1. Incorrect monitor configuration settings (common)
- 2. Simulator not transmitting proper response signal (occasional)
- 3. Module hardware malfunction (rare)
Verify monitor software version and apply correct configuration
- Test point
- Monitor software version display
- Expected
- Settings match software version (3.1 or earlier: Start with New patient, Current Supra, TOF, Volume 6, Cycle 20s; 3.2+: AUTO, Supramax Selected, TOF, Volume 6, Cycle 20s)
6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No response to breathing patient safety
Module does not detect or display respiratory activity when patient breathes
Ranked causes
- 1. Gas sample line disconnected (common)
- 2. Sampling line or water trap loose or blocked (common)
- 3. Sample gas out blocked (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
PDM data not displayed on monitor
Cardiac output, invasive pressure, or other PDM measurements missing from monitor screen
Ranked causes
- 1. Missing cardiac output license (common)
- 2. Missing PDM license (common)
- 3. Configuration issue with parameter priority (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder does not work
B1X5-REC recorder unit non-functional
Ranked causes
- 1. Recorder connect cable loose or faulty (common)
- 2. Recorder unit failure (occasional)
- 3. Host or frame issue (unknown)
Verify recorder connect cable integrity and connection
- Test point
- Recorder connector
- Expected
- Cable intact and properly seated
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor check failed patient safety
BIS sensor impedance check fails; one or more electrode impedances exceed acceptable threshold.
Ranked causes
- 1. Sensor improperly attached to patient (common)
- 2. Faulty sensor (occasional)
- 3. PIC+ cable or BISx unit faulty (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Supramaximal current search fails or value exceeds limit
Detected supramaximal current is greater than 70 mA, or search does not complete
Ranked causes
- 1. Incorrect simulator response knob setting (common)
- 2. Simulator Fade setting not correct (common)
- 3. Sensor cable or ElectroSensor defective (occasional)
Verify simulator response knob position
- Test point
- NMT Simulator response control
- Expected
- Response knob set to maximum
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Too high EtCO2 value patient safety
End-tidal carbon dioxide concentration above expected for patient
Ranked causes
- 1. Water trap contamination or malfunction (common)
- 2. Calibration error (common)
- 3. Incorrect humidity compensation setting (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to perform function or feature not available
Feature or function is unavailable or grayed out in menu
Ranked causes
- 1. License not purchased for feature (common)
- 2. Trial license expired (occasional)
- 3. License not installed properly (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Zero error message displayed
Message appears in parameter window indicating autozero failure during measurement or gas calibration startup
Ranked causes
- 1. Zero valve operation failure (common)
- 2. Defective zero absorber or dehumidification tube (common)
- 3. Module internal failure (rare)
Verify zero valve operation
- Test point
- Zero valve
- Expected
- Valve opens and closes properly
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Artifacts message on entropy measurement patient safety
Entropy parameter display shows Artifacts; signals contain noise or artifact
Ranked causes
- 1. Patient movement, eye blinks, shivering (common)
- 2. Deep breathing or arrhythmia (common)
Wait for signal to stabilize
- Test point
- Entropy parameter display
- Expected
- Artifacts message clears and measurement proceeds
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode reader does not read multi-field barcode correctly
Multi-field barcode information not populated correctly to fields in Admit menu
Ranked causes
- 1. Parser configuration is incorrect (common)
- 2. Parser configuration is incompatible with barcode format (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
BIS waveform not appearing on screen
BIS parameter and waveform missing from monitor display; No sensor message shown instead
Ranked causes
- 1. BISx unit not connected to BIS module (common)
- 2. BIS EEG waveform not configured on monitor screen (occasional)
Verify BISx unit is connected to BIS module with patient interface cable
- Test point
- BIS module connection port and BISx cable
- Expected
- Cable firmly connected, connection secure
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Gas sampling system and CO2 functional check
Functional verification of gas sampling system and CO2 measurement after visual inspection, O-ring replacement, or module case opening
Ranked causes
- 1. Routine calibration and functional verification required (common)
- 2. Gas sampling system leakage or flow anomaly (occasional)
Visually inspect front cover
- Test point
- Module front cover
- Expected
- Front cover is intact
6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CARESCAPE ONE works but E-modules data not displayed
CARESCAPE ONE acquisition module operates normally but connected E-modules do not send data to monitor
Ranked causes
- 1. E-module defective (occasional)
- 2. Module connector defective (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check electrodes message displayed
Parameter window shows 'Check electrodes' message; measuring electrodes are off or stimulus electrode connection is broken
Ranked causes
- 1. Measuring electrodes are off (common)
- 2. Broken connection of stimulating electrode or cable (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Water Trap message displayed
Alarm area and waveform area display 'Check Water Trap' message. Water trap connection or internal tubing issue.
Ranked causes
- 1. Water trap is not connected (common)
- 2. Air leak inside the internal tubing (occasional)
Verify water trap is physically connected to the module
- Test point
- Water trap connector
- Expected
- Water trap firmly connected
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication LED off (Ethernet Activity)
Communication (Ethernet Activity) LED is off, indicating loss of communication between frame and host or module malfunction.
Ranked causes
- 1. EMBC (Frame) faulty and not communicating (common)
- 2. PDM module faulty and not communicating (common)
Verify EMBC and host communication from service interface
- Test point
- Service interface communication status
- Expected
- EMBC communicating with host
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication LED off or not blinking
Ethernet Activity LED is off, indicating no communication between frame (EMBC) and host, or PDM module and host
Ranked causes
- 1. EMBC (Frame) faulty or not communicating (occasional)
- 2. PDM module faulty or not communicating (occasional)
Verify Communication LED blinking when EMBC and host are connected
- Test point
- Communication LED (Ethernet Activity LED)
- Expected
- Blinking
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Connected device software activation fails or shows update failed message
- Contact Monitors button is inactive
- CPU board malfunction
- Erroneous gas readings after maintenance
- EtCO2 over scale >15%
- EtCO2 over scale above 20 percent
- Expiratory TV greater than inspiratory TV
- F2 Frame is not recognized by monitor
- Fluctuating airway resistance (Raw)
- Gas sampling system leakage suspected
- Incompatible or duplicate modules connected
- Invasive pressure channel zeroing failed
- Invasive pressure measurement under range
- Loose parts or internal cable connections
- Low gas sample flow
- Low volumes message displayed
- MiniOM oxygen signal noisy or unstable
- Modules cannot be placed properly in F2 Frame
- NMT module not recognized or not communicating
- No Entropy waveform or parameter on screen
- No picture on screen despite monitor powered on
- On/Standby LED off when monitor powered on
- On/Standby LED off while monitor is on
- PDM module works but E-module data not visible
- Recorder works but nothing appears on paper
- Reference not stable message
- Replace Water Trap message displayed
- Sensor impedance check not available in menus
- Sensor INOP (inoperative)
- Settings file password incorrect or file corrupted
- Startup procedure does not advance beyond welcome screen
- Strange or nonphysiological gas exchange values
- Strongly vibrating spirometry loop
- System time is incorrect without network connection
- Testing DSC message displayed
- Too high airway resistance (Raw)
- UAC disconnect message appears
- Unable to activate new monitor software
- Unable to perform gas calibration
- Unable to upload license file
- Water trap connection issue
- Water trap filter failure
- Wrong character displayed when barcode is read
- Abnormally high or low CO2 response
- Absolute pressure sensor error
- Accessory failure or degradation
- Apnea alarm not activated
- Apnea detection not working
- Apply sensor
- Arterial invasive pressure disconnected
- Artifact
- Artifact message
- Automatic check off
- BIS cable off
- BIS DSC error
- BIS measurement removed
- BIS module error
- BIS sensor check failed
- BIS sensor expired
- Bypass flow high
- Cable off
- Cable off message on entropy measurement
- Calibrating gas sensor (warm-up in progress)
- Calibrating message appears
- Calibrating message displayed
- Calibration: Agent identification
- Calibration: Air leak detection
- Calibration: Airway gases measurement
- Calibration: Ambient pressure verification
- Calibration: Apnea detection
- Calibration: BIS module functional check
- Calibration: CO2 waveform and FiO2 level setup
- Calibration: D19 display maintenance check
- Calibration: E-NSATX module maintenance check
- Calibration: E-PiCCO module functional check
- Calibration: Entropy module functional check
- Calibration: Fan operation
- Calibration: Flow waveform measurement
- Calibration: Frame F5-01 and F7-01 maintenance check
- Calibration: Gas calibration
- Calibration: Gas exhaust blockage detection
- Calibration: Gas sampling system leak test
- Calibration: Module keys operation
- Calibration: NMT module functional check
- Calibration: Occlusion detection
- Calibration: Pulse contour cardiac output
- Calibration: Reference gas flow rate check
- Calibration: Respiratory module functional check
- Calibration: Sample flow rate check
- Calibration: Spirometry system leak test
- Calibration: Visual inspection before functional check
- Calibration: Zero valve operation
- CCO Calibration fail message
- CCO meas. fail message
- Checking sensor
- Checking sensor message on entropy measurement
- Checking sensor message stays longer than 2 minutes
- CO measurement removed message
- Complete functional check and reset monitor
- Compliance value invalid
- Confirm calibration message
- Confirm electrode contact issue on entropy measurement
- Demo data
- Demo data message on entropy measurement
- Device authentication failed
- DSC error
- Duplicate monitor addresses detected
- E-module cannot be placed properly in frame
- E-module showing erroneous readings
- E-musb and CARESCAPE ONE simultaneously connected
- Electrical safety test
- Electromagnetic interference detected
- EMG electrodes off message
- Entropy cable off alarm on entropy measurement
- Entropy measurement removed message
- Entropy module not connected
- Entropy sensor check failed
- Entropy sensor off
- Erroneous readings after maintenance
- Erroneous readings after module cover reassembly
- Error messages displayed in message field
- EtCO2 exceeds PaCO2 arterial value
- EtCO2 greater than PaCO2
- Ethernet Link LED blinking
- Ethernet Link LED on (solid green)
- F2 Frame communication failure with host monitor
- F5/F7 Frame is not recognized by monitor
- F7 Frame used standalone causing ventilation blockage
- Failure in Agent ID
- Femoral invasive pressure disconnected
- Flow or Paw waveform exceeds scale
- Gas exhaust blockage not detected
- Gas flow out of measurement range
- Gas measurement failed
- Gas measurement source detection ongoing
- Gas measurements removed
- Gas module disconnected
- Gas module in standby mode
- Gas sensor calibration in progress
- High BIS impedance
- Host software activation completes but monitor does not restart
- Host software activation takes longer than expected
- Host software upgrade activation code rejected or not accepted
- Identical BIS modules
- Identical C.O. modules message
- Identical Entropy modules
- Identical frame detected
- Identical frame detected in monitoring system
- Identical gas modules installed
- Identical IP8 modules detected
- Identical modules causing patient safety risk
- Identical NMT modules message
- Incompatible device: BIS module
- Incompatible DSC
- Incompatible gas module detected
- Incompatible sensor
- Invalid software image file
- Invasive pressure calibration required
- Invasive pressure channel in standby
- Invasive pressure channels not zeroed
- Invasive pressure sensor disconnected
- Isoelectric EEG detected
- Loop is being saved
- Low EEG signal
- Low perfusion at measurement site
- Measurement off message
- Measurement parameters not displaying correctly
- MiniPVX measuring unit malfunction
- MiniTPX measuring unit malfunction
- Module box loose parts or cable connections
- Module error
- Module frame in wrong position causing malfunction
- Module frame not properly grounded
- Module not identified correctly
- Module not responding
- Monitor authentication failed
- Monitor displays Field Replacement Unit startup message
- Monitor not connected to mains power during activation
- Multiple electrodes poor contact on entropy measurement
- Multiple identical gas modules detected
- Multiple identical SpO2 modules in system
- Multiple identical SpO2 sources detected
- N2O detected
- NMT cable removed message
- NMT measurement removed message
- NMT sensor removed or Sensor off message
- No BIS sensor
- No BIS waveform displayed on screen
- No CCO catheter message
- No Entropy sensor detected
- No invasive pressure displayed
- No license for F5/F7 frame
- No module message
- No SpO2 signal or sensor problem
- No uploaded software version available to activate
- Noise interference detected
- Not calibrated message
- Occlusion detection not working
- Over range FiO2 measurement
- Over scale waveform
- Over Scale waveform clipping
- Parameter not displayed or not active
- PDM not working when connected to ePort
- Perform electrical safety tests
- Perform SpO2 functional check
- Perform visual inspection after service
- Performing visual inspection
- Poor signal
- Pressure pulsation sensed during zeroing
- Pulse rate high or low alarm
- Re-calibrate CCO required
- Recorder is printing but nothing appears on paper
- Reference set message
- Regional Block message
- Replace sensor message displayed
- Sample flow rate out of specification
- Scale changed
- Sensor impedance check fails during operation
- Service CO module required
- Setting reference message
- Settings cannot be activated on older software versions
- Software activation does not start after upload
- Software activation requires base license
- Software upload cannot be initiated
- Software upload fails or does not complete
- SpO2 acquisition module malfunction
- SpO2 measurement module disconnected
- SpO2 measurement searching for pulse
- SpO2 sensor has failed
- SpO2 sensor not compatible
- SpO2 sensor not connected
- SpO2 sensor off patient
- SpO2 value high or low alarm
- Start measurement message
- Supramax not found message
- Supramax search message
- System time is incorrect with network connection
- TETANIC message
- Too high peak inspiratory pressure (Ppeak)
- Too low EtCO2 value
- Tubing contamination detected
- Unable to view feature despite enabled license
- Verify measuring equipment is calibrated
- Verify network device connectivity
- VO2 not measurable, high FiO2
- VO2 value invalid or no VO2 with high FiO2
- Wait to continue message
- Water trap not detected
- Wrong character displayed when keyboard key is pressed
- Wrong software package is activated
- Zero adj >100 mmHg message
- Zero ICP separately message
- Zero valve not functioning
- Zeroed message appears
- Zeroing error
- Zeroing in progress
- Zeroing in progress message displayed
- Zeroing message appears
- Zeroing message displayed
259 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0xGAS0001 | Service gas module. CO2, N2O, AA sensor failed | MiniTPX unit failure | error |
0xGAS0010 | Service gas module. O2 sensor failed | MiniOM oxygen sensor failure | error |
0xGAS0011 | Simultaneous MiniTPX and MiniOM oxygen sensor failure | Both CO2/N2O/AA and O2 sensor failures | error |
0xGAS0100 | Service gas module. Spirometry sensor failed | Spirometry sensor failure | error |
0xGAS0101 | Simultaneous MiniTPX and spirometry sensor failure | Both CO2/N2O/AA sensor and spirometry sensor failures | error |
0xGAS0110 | Simultaneous O2 sensor and spirometry sensor failure | Both MiniOM and spirometry sensor failures | error |
0xGAS0111 | Simultaneous MiniTPX, O2, and spirometry sensor failure | All three sensor types failed | error |
0xGAS1000 | Service gas module. Gas measurement failed | General gas measurement failure | error |
0xGAS1001 | Simultaneous MiniTPX unit and general gas measurement failure | MiniTPX and general measurement failure | error |
0xGAS1010 | Simultaneous O2 sensor and general gas measurement failure | MiniOM and general measurement failure | error |
0xGAS1011 | Simultaneous MiniTPX, O2, and general measurement failure | All three component failures with general measurement failure | error |
0xGAS1100 | Simultaneous spirometry sensor and general gas measurement failure | Spirometry sensor and general measurement failure | error |
0xGAS1101 | Simultaneous MiniTPX, spirometry, and general gas measurement failure | All measurement and sensor failures combined | error |
0xGAS1110 | Service gas module gas measurement failed | Simultaneous MiniOM oxygen sensor, spirometry sensor, and general gas measurement failure | error |
0xGAS1111 | Service gas module gas measurement failed | Simultaneous MiniTPX measuring unit, MiniOM oxygen sensor, spirometry sensor, and general gas measurement failure | error |
>320 mmHg or >43 kPa | Invasive pressure measurement over range | Measurement exceeds maximum range or sensor/cable fault | warning |
<-40 mmHg or <-5 kPa | Invasive pressure measurement under range | Measurement below minimum range or sensor/cable fault | warning |
Apnea deactivated | Apnea alarm is not active or has not met activation criteria. | Apnea alarm start-up conditions not reached. | warning |
Art 1-8 disconnect | Arterial invasive pressure disconnected | No arterial invasive pressure detected | warning |
Auto detecting source | System is automatically detecting gas measurement source. | Module initialization or gas source change. | info |
Calibrated | Channel calibrated successfully | Calibration procedure completed without errors | info |
Calibrating | Gas sensor calibration is in progress. | Manual calibration started or scheduled warm-up calibration. | info |
Calibrating gas sensor | Module is warming up and gas measurements are not yet available. | Gas module connected or powered on. | info |
Calibration error | Gas sensor calibration failed and must be repeated. | Calibration gas not fed within one minute; wrong gas used; gain adjustment error; trapped CO2. | error |
CCO Calibration fail | Unsuccessful calibration | Calibration procedure fails to complete or validate | error |
CCO meas. fail | Calibration has failed | Previous calibration validation failed | error |
Check sample gas out | Sample gas output path is blocked or water trap has an issue. | Blockage in sample line or water trap connection failure. | warning |
Check Water Trap | Water trap is not connected or has internal leakage. | Water trap disconnection or internal tubing leak detected. | warning |
CO measurement removed | The module is disconnected | C.O./CCO module becomes disconnected from monitor | warning |
Confirm calibration | Measurement data has not been confirmed before trying to exit Cardiac Output/CCO Calibrate menu | User attempts to exit Calibrate menu without confirming measurement | warning |
Continuous blockage. Check sample line and water trap. | Persistent blockage in gas sampling path despite previous checks. | Ongoing obstruction in sample line, water trap or internal tubing. | error |
F2 frame not authenticated: Call service | F2 Frame authentication required | F2 Frame fails authentication check during initialization | error |
Failure in Agent ID | Anesthetic agent identification system malfunction. | Agent ID sensor or circuit failure. | error |
Fem 1-8 disconnect | Femoral invasive pressure disconnected | No femoral arterial invasive pressure detected | warning |
Gas measurements removed | Gas acquisition module has been disconnected. | Module unplugged or removed from system. | info |
Gas module standby | Gas module pump is shut down and module is in standby mode. | Module pump manually stopped or auto-shutdown triggered. | info |
Identical C.O. modules | Two or more identical C.O./CCO modules detected in system | Multiple E-COP, E-COPSv, or E-PiCCO modules connected simultaneously | error |
Identical frame detected | Two or more identical module frames detected in system | Multiple F2, F5, or F7 frames connected simultaneously to monitoring system | error |
Identical gas modules | System detected two or more identical gas modules installed. | Multiple identical modules connected to system simultaneously. | warning |
Identical IP8 modules | Duplicate IP modules on same channel | Two or more E-PiCCO or E-COPSv modules map to same channel | error |
IP's not zeroed | One or more invasive pressure channels not zeroed | System detects unzeroed invasive pressure channels | warning |
Low gas sample flow | Sample flow rate is below minimum specification (less than 80% of nominal). | Flow rate measurement below 80% of module nominal flow. | warning |
Low volumes | Tidal volume detection failure or circuit leak | I:E detection failure for 20 seconds without apnea or low measured volumes | warning |
No CCO catheter | The catheter is disconnected from the cable, or the cable is disconnected from the module | Catheter or cable disconnection detected | warning |
No license for F2 frame | F2 Frame support license not enabled in configuration | F2 Frame license not purchased or activated in system configuration | warning |
No module | No C.O. module connected | User accesses C.O. menu without module installed | warning |
Not calibrated | The module and cable are connected, but the CCO measurement has not been calibrated | User attempts to measure without prior calibration | warning |
Over range | Measured FiO2 exceeds maximum valid range (>103%). | FiO2 measurement reading above 103%. | error |
Over scale | Waveform display clipped due to gas concentration exceeding current scale upper limit. | Gas concentration exceeds scale maximum for selected waveform scale. | warning |
P1-8 standby | Invasive pressure channel in standby mode | User or system placed IP channel in standby | info |
P1-8 zeroing failed | Invasive pressure channel zeroing failed | Zeroing procedure could not complete successfully | error |
Pressure measurement removed | Invasive pressure acquisition device removed | IP measurement device physically removed from system | warning |
Pressure Sensed | Pressure pulsation detected during zeroing | Dynamic pressure detected when static zeroing should occur | warning |
Replace Water Trap | Water trap is defective, contaminated, or occluded | Water trap assembly failure or contamination detected | warning |
Sample line blocked | Gas sampling line or water trap system is blocked or full | Blockage detected in sampling line or water trap container full condition | warning |
Sensor disconnected | Invasive pressure sensor cable disconnected | Transducer detects disconnection or cable removed from module | warning |
UAC 1 disconnect to UAC 4 disconnect | Invasive pressure line is disconnected | Invasive pressure line becomes disconnected | warning |
Zero adj >100 mmHg | Offset during zeroing exceeded 100 mmHg | Zeroing process shows offset greater than 100 mmHg | error |
Zeroed | Zeroing was successful | Zeroing process completed successfully | info |
Zero error | Autozero calibration has failed | Autozero failure during measurement or at start of gas calibration | error |
Zero ICP separately | ICP channel must be zeroed separately from other invasive pressures | User attempts to zero all channels together including ICP | warning |
Zeroing | IP channel is currently being zeroed | Zeroing process in progress | info |
Zeroing | Zeroing procedure is in progress | System is performing automatic zero calibration | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI