FR FieldReady Academy
Patient monitors · Carescape Canvas family

GE Healthcare Carescape Canvas 1000: troubleshooting and error codes

The ServiceForge knowledge base holds 319 documented symptoms for the GE Healthcare Carescape Canvas 1000, with 527 ranked causes, 414 diagnostic checks with test points and expected values, and 63 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Acquisition module data not displayed on monitor patient safety

Module data not shown on monitor display; parameter or waveform missing from screen

Ranked causes

  1. 1. Acquisition module incompatible with system (occasional)
  2. 2. Module not inserted properly into rails (common)
  3. 3. Monitor not configured to display parameter (common)
  4. 4. Faulty or improperly connected frame-to-host cable (occasional)
  5. 5. Internal frame or EMBC connector defective or loose (occasional)
  6. 6. Module connector or EMBC board defective (occasional)
  7. 7. DC/DC board defective (rare)
  8. 8. Module defective (see module troubleshooting) (occasional)
First check

Verify module compatibility

Test point
Module and system documentation
Expected
Module listed as compatible with Carescape Canvas 1000

4 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Acquisition module data not shown on monitor display

Parameter data from acquisition module (CARESCAPE ONE or E-module) does not appear on host monitor screen

Ranked causes

  1. 1. Acquisition module incompatible (occasional)
  2. 2. Module not connected properly (common)
  3. 3. Monitor not configured to display parameter (occasional)
  4. 4. Cable connecting F2 Frame to host defective or loose (occasional)
  5. 5. F2 Frame internal connectors defective or loose (rare)
  6. 6. F2 Frame main board defective (rare)
  7. 7. AC/DC module defective (rare)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not power on high voltage

Monitor fails to turn on; no display or indicators active

Ranked causes

  1. 1. Power cord is loose (common)
  2. 2. Power cord is faulty (occasional)
  3. 3. Power outlet does not meet specifications (occasional)
  4. 4. Button board is defective or cable is loose (rare)
  5. 5. Power supply cable loose or faulty (rare)
  6. 6. Power supply is faulty (rare)
  7. 7. CPU assembly is faulty (rare)
First check

Verify power cord connection

Test point
Wall outlet and monitor connector
Expected
Cord fully inserted and seated

1 more check with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

E-module or PDM does not work correctly when connected

Module fails to function or communicate properly after connection to monitor frame

Ranked causes

  1. 1. Identical modules connected simultaneously (common)
  2. 2. PDM and CS ONE connected simultaneously (common)
  3. 3. Configuration issue (occasional)
  4. 4. Compatibility issue with module or accessories (occasional)
  5. 5. Frame-CPU cable loose or defective (occasional)
  6. 6. Defective acquisition module (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Measurement removed from display (XXXX measurement(s) removed) patient safety

One or more measurements removed from display; module or measurement parameters no longer displayed

Ranked causes

  1. 1. Module not connected properly (common)
  2. 2. Module connector defective (occasional)
  3. 3. Frame not properly connected to host (occasional)
  4. 4. EMBC board connectors loose or defective (occasional)
  5. 5. EMBC board defective (occasional)
  6. 6. DC/DC board defective (occasional)
First check

Verify module seating in frame rails

Test point
Module and frame rails
Expected
Module fully inserted; connectors aligned; latches engaged

2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Measurements removed from display

System message shows XXXX measurement(s) removed, where XXXX is module name. Module data not displaying on monitor screen.

Ranked causes

  1. 1. Acquisition module not connected properly (common)
  2. 2. Module connector defective (occasional)
  3. 3. Frame to host cable defective or loose (occasional)
  4. 4. EMBC board connectors loose or defective (occasional)
  5. 5. EMBC board defective (rare)
  6. 6. DC/DC board defective (rare)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Nonphysiological VO2 readings patient safety

Oxygen consumption (VO2) measurements display clinically implausible values

Ranked causes

  1. 1. Unstable O2 delivery or gas mixer issue (common)
  2. 2. Respiratory rate over 35/min (occasional)
  3. 3. Reference gas inlet port blocked (occasional)
  4. 4. Gas sampling line longer than 2 m (occasional)
  5. 5. Dead space of Y-piece exceeds 8 ml (occasional)
  6. 6. Gas sampling line connected to HME or D-lite incorrectly placed (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Abnormally low invasive pressure reading patient safety

Pressure measurement is lower than expected or abnormally low

Ranked causes

  1. 1. Transducer wrongly positioned (common)
  2. 2. Defective transducer (occasional)
  3. 3. Module not connected (occasional)
  4. 4. Transducer adapter cable not connected to module (occasional)
  5. 5. Invasive pressure channel not configured to screen (occasional)
First check

Verify transducer position

Test point
Transducer location
Expected
Transducer at mid-heart level

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Acquisition platform not recognized by host monitor

CARESCAPE ONE, CARESCAPE Dock F0, or parameters not displayed on host monitor; platform not communicating

Ranked causes

  1. 1. PDM and CARESCAPE ONE connected simultaneously (common)
  2. 2. CARESCAPE parameters not shown due to configuration issue (occasional)
  3. 3. Cable between dock/frame and host monitor is loose or defective (occasional)
  4. 4. CARESCAPE Dock F0 or F2 Frame is defective (rare)
  5. 5. CPU assembly in host monitor is defective (rare)
First check

Verify CARESCAPE ONE works standalone

Test point
CARESCAPE ONE disconnected from host or F2 Frame
Expected
CARESCAPE ONE operates normally as standalone monitor

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarms do not work patient safety

Alarm sounds not activating or not heard during alarm conditions

Ranked causes

  1. 1. Audible alarms are turned off (common)
  2. 2. Alarm volume is too low (common)
  3. 3. Defective speaker (occasional)
  4. 4. Speaker cable is loose or faulty (occasional)
  5. 5. Tone generator or audio amplifier failure (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loose parts or cable connections inside module caution

Module may have internal loose parts, improperly tightened screws, disconnected cables, pinched or bent tubes, or loose objects rattling inside.

Ranked causes

  1. 1. Screws not tightened properly (common)
  2. 2. Cables disconnected or improperly seated (common)
  3. 3. Tubes pinched or sharply bent (common)
  4. 4. Tubes disconnected (common)
  5. 5. Loose objects inside module (occasional)
First check

Verify all screws are tightened

Test point
All screws in module
Expected
Tight, no rotation

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Module functionality test fails patient safety

BIS module does not function correctly; error messages appear or test DSC fails.

Ranked causes

  1. 1. BISx unit defective (common)
  2. 2. Incompatible system component (occasional)
  3. 3. Identical modules connected (occasional)
  4. 4. Defective or incompatible accessories (occasional)
  5. 5. Loose keypad cable or module connector problem (rare)
First check

Verify system component compatibility

Test point
Monitor and all connected modules
Expected
All components listed in supplemental information

5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No IX network traffic

Monitor has no network communication on IX port; unable to transmit or receive data

Ranked causes

  1. 1. Incorrect configuration (common)
  2. 2. Cable connections do not match network type (common)
  3. 3. IX network traffic overload detected (rare)
  4. 4. IX port authentication enabled but monitor misconfigured (rare)
  5. 5. CA certificate has expired (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No MC network traffic

Monitor has no network communication on MC port; unable to transmit or receive data

Ranked causes

  1. 1. Incorrect configuration (common)
  2. 2. Cable connections do not match network type (common)
  3. 3. MC network traffic overload detected (rare)
  4. 4. MC port authentication enabled but monitor misconfigured (rare)
  5. 5. CA certificate has expired (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sample gas outflow blocked caution

Gas exhaust line or sample output path is obstructed; measurements may be affected.

Ranked causes

  1. 1. Sample gas out connector or exhaust line blocked (common)
  2. 2. Patient circuit tubing occluded (occasional)
  3. 3. Scavenging system improperly configured (occasional)
  4. 4. Internal tubing blocked (occasional)
  5. 5. Pump failure (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration error

Gas sensor calibration failed; measurements are unavailable until recalibration succeeds.

Ranked causes

  1. 1. Calibration gas not fed within one minute (common)
  2. 2. CO2 trapped in sensor during transport (occasional)
  3. 3. Calibration gain adjustment too large (rare)
  4. 4. Wrong calibration gas used (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Continuous gas sampling line blockage caution

Gas sampling line or water trap is persistently blocked; continuous message shown.

Ranked causes

  1. 1. Gas sampling line blocked (common)
  2. 2. Water trap container full (common)
  3. 3. Water trap occluded (occasional)
  4. 4. Internal tubing blocked (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Electrical safety verification high voltage

Comprehensive electrical safety testing including power outlet, power cord, ground integrity, earth leakage, and touch current measurements under normal and single fault conditions.

Ranked causes

  1. 1. Power outlet improperly wired or damaged power cord (common)
  2. 2. Inadequate ground continuity or impedance (occasional)
  3. 3. Excessive earth leakage current (occasional)
  4. 4. Excessive touch current (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gas sampling system leaks or blockages patient safety

Faults in gas sampling system including leaks, blockages, pump failure, valve failure, reduced flow rates, or sensor issues

Ranked causes

  1. 1. Leaks in tubing or blockages (occasional)
  2. 2. Sampling pump or magnetic valve failure (occasional)
  3. 3. Dirt or matter accumulation in tubing reducing flow (occasional)
  4. 4. Pressure sensor failure (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inspiratory TV greater than expiratory TV patient safety

Spirometry measurement shows inspiratory tidal volume larger than expiratory, indicating system leak

Ranked causes

  1. 1. Leak in lungs or ET tube cuff (common)
  2. 2. Spirometry tube leak (common)
  3. 3. Water inside D-lite or tubings (occasional)
  4. 4. D-fend Pro leaks (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

MiniOM oxygen measuring unit malfunction patient safety

No response to O2 signal or strong signal drift; incorrect oxygen readings or unstable measurements

Ranked causes

  1. 1. Loose tubing connections (common)
  2. 2. Tubing blockage or occlusion (common)
  3. 3. Reference gas filter assembly contaminated or clogged (occasional)
  4. 4. Internal measurement unit fault (rare)
First check

Inspect all tubing connections on oxygen measuring unit

Test point
Tubing connectors
Expected
All connections seated firmly with no visible gaps

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Module physical damage or contamination

Front cover cracked, connectors dirty or loose, casing damaged, latch broken, patient cables worn or dirty

Ranked causes

  1. 1. Front cover not intact (unknown)
  2. 2. Connectors dirty, damaged or loose (unknown)
  3. 3. Module casing or latch damaged (unknown)
  4. 4. Patient cables compromised (unknown)
First check

Inspect front cover for cracks, breaks or damage

Test point
Front cover exterior
Expected
Intact, no cracks or breaks

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Module physical inspection before troubleshooting

Visual inspection procedure to verify module structural integrity, connectors, latch mechanism, and D-fend connector condition before detailed diagnostics.

Ranked causes

  1. 1. Damaged front cover or housing (unknown)
  2. 2. Connector damage, contamination, or improper seating (unknown)
  3. 3. Latch mechanism damage or contamination (unknown)
  4. 4. Mini D-fend latch or connector problems (unknown)
First check

Remove module from system

Test point
Module mounting
Expected
Module removed without damage

5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Response too weak message

Parameter window shows 'Response too weak'; maximum gain insufficient to measure response signal amplitude

Ranked causes

  1. 1. Stimulus electrodes are loose (common)
  2. 2. Response electrodes attached to wrong place (common)
  3. 3. Electrodes are dry (occasional)
  4. 4. Skin at electrode site not properly prepared (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sample line blocked message displayed

Alarm area, parameter window, and waveform area display 'Sample line blocked' message indicating continuous blockage in the CO2 sampling pathway.

Ranked causes

  1. 1. Gas sampling line is blocked (common)
  2. 2. Water trap container is full (common)
  3. 3. Water trap is occluded (occasional)
  4. 4. Internal tubing is blocked (occasional)
First check

Inspect external gas sampling line for blockages

Test point
External gas sampling line
Expected
Clear, unobstructed pathway

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sampling system leaks or blockages

Gas sampling system may have leaks in tubing, blockages, or flow rate diminishing due to dirt or contamination.

Ranked causes

  1. 1. Leaks in sampling system tubing (common)
  2. 2. Blockages in tubing or internal passages (common)
  3. 3. Sampling pump failure or wear (occasional)
  4. 4. Magnetic valve failure (occasional)
First check

Listen for sampling pump noise

Test point
Sampling pump
Expected
Quiet operation, no grinding or rattling

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to send or retrieve MUSE 12SL reports

Cannot send or retrieve ECG reports to/from MUSE Web Server

Ranked causes

  1. 1. Incorrect MUSE Web Server username or password (common)
  2. 2. Incorrect MUSE Web Server address (common)
  3. 3. AMSE certificate not installed (occasional)
  4. 4. MUSE certificate not installed or invalid (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Visual inspection for frame and module condition

Routine inspection of physical frame integrity, connectors, and attached modules before troubleshooting

Ranked causes

  1. 1. Damaged cover or module rails (occasional)
  2. 2. Loose or dirty connectors (occasional)
  3. 3. Loose or damaged cables (occasional)
  4. 4. Modules not properly locked to frame (occasional)
First check

Inspect frame cover and module rails

Test point
Frame exterior
Expected
Cover and rails intact, no visible damage

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

BIS sensor module degradation or malfunction

Cover stickers peeling, panel damaged, cables loose, internal components dirty or corroded

Ranked causes

  1. 1. Cover or panel stickers not intact (unknown)
  2. 2. Cables or connections not intact on BISx unit (unknown)
  3. 3. Module or applied parts not clean (unknown)
First check

Inspect BISx unit cover and panel stickers for damage or peeling

Test point
BISx unit cover and sticker areas
Expected
Intact, not peeling or damaged

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

BIS sensor not responding or failing electrode check patient safety

Sensor checking message appears but green checkmarks not shown for electrodes; sensor test fails

Ranked causes

  1. 1. BIS sensor simulator not connected or faulty (common)
  2. 2. PIC Plus cable not intact or poorly seated (occasional)
  3. 3. BISx unit internal test failing (occasional)
First check

Connect BIS sensor simulator to PIC Plus cable and wait for electrode checking to start

Test point
BIS sensor simulator and PIC Plus cable connection
Expected
Checking sensor message appears in BIS parameter window

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration error message displayed

Parameter window and calibration menu display 'Calibration error' message indicating calibration procedure failed due to timing, gain, or gas issues.

Ranked causes

  1. 1. Calibration gas not started within 1 minute of autozero completion (common)
  2. 2. Calibration failed due to too large gain adjustment (occasional)
  3. 3. Wrong calibration gas used (occasional)
First check

Verify calibration gas was started within 1 minute of autozero completion

Test point
Calibration gas inlet
Expected
Gas flow initiated within 1 minute of zero completion

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration error on invasive pressure channel patient safety

Message displayed as Calibration error in parameter window during channel calibration

Ranked causes

  1. 1. Pressure calibration failure due to time-out (common)
  2. 2. Pulsating waveform detected during calibration (occasional)
  3. 3. Gain beyond limits (±20% of default gain) (occasional)
First check

Start calibration within time window

Test point
pressure calibration timer
Expected
Inflation starts within 45 seconds of auto-zeroing completion

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Invasive pressure patient safety

Invasive pressure measurement requires calibration when transducer is replaced, calibration check fails, or measured values exceed specification limits

Ranked causes

  1. 1. Pressure transducer replaced (common)
  2. 2. Invasive pressure calibration check failed (occasional)
  3. 3. Measured value outside specification limits (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check sample gas out message displayed

Alarm area and waveform area display 'Check sample gas out' message indicating blockage in the sample gas outflow path.

Ranked causes

  1. 1. The sample gas outflow is blocked (common)
  2. 2. Internal tubing is blocked (occasional)
  3. 3. Pump failure (occasional)
First check

Inspect sample gas out connector and exhaust line

Test point
Front panel sample gas out connector, exhaust/scavenging line
Expected
No blockages, clear pathway for gas return or scavenging

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 waveform and values not updating patient safety

CO2 waveform does not appear or EtCO2/FiCO2 values do not update when breathing into sampling line

Ranked causes

  1. 1. Gas sensor warm-up not complete (common)
  2. 2. Gas calibration required (common)
  3. 3. CO2 measurement source not configured (occasional)
First check

Gas sensor warm-up status

Test point
Parameter window message display
Expected
No 'Calibrating Gas Sensor' message; waveform field shows CO2 waveform

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ePort pod to host cable connection issue

PDM module connected via ePort connector not communicating or recognized

Ranked causes

  1. 1. ePort cable loose or defective (common)
  2. 2. Defective PDM or accessories (occasional)
  3. 3. Defective CPU assembly (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Expired volume much smaller than inspired volume

MVexp << MVinsp; Ppeak - PEEP < 2 cmH2O; TVinsp and TVexp are DATA_INVALID. Leak in spirometry system causing expired volume 70% or less of inspired volume.

Ranked causes

  1. 1. Leaking spirometry tube (common)
  2. 2. Water in tubing (common)
  3. 3. MiniPVX sensor failure (occasional)
First check

Perform spirometry leak test

Test point
Spirometry system
Expected
No leaks detected

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invalid or unreachable monitor address

Monitor cannot be contacted due to incorrect IP address, hostname, or network connectivity issues

Ranked causes

  1. 1. Incorrect IP address or hostname (common)
  2. 2. Monitor turned off or not connected to IX network (common)
  3. 3. Network configuration error or network failure (occasional)
First check

Verify IP address or hostname

Test point
Monitor Addresses field
Expected
Valid IP address or hostname matching target monitor

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invasive pressure measurement over range patient safety

Pressure reading displayed as >320 mmHg or >43 kPa in parameter window

Ranked causes

  1. 1. Measurement is over range or sensor/cable faulty (common)
  2. 2. Transducer not zeroed correctly (common)
  3. 3. Pre-zero with stopcock closed (occasional)
First check

Verify measurement value

Test point
parameter window
Expected
>320 mmHg or >43 kPa

2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loop overscale in spirometry patient safety

Spirometry waveform display exceeds the selected vertical scale limits

Ranked causes

  1. 1. Wrong scale selected (common)
  2. 2. Wrong TV base selected (occasional)
  3. 3. Leak between ventilator and D-lite (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low gas sample flow detected patient safety

Sample flow drops below 80% of module nominal specification

Ranked causes

  1. 1. Gas sampling line or tubing blocked (common)
  2. 2. Water trap blocked or full (common)
  3. 3. Pump failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Measurement unit failure or malfunction caution

CPU board, MiniOM measuring unit, or MiniTPX measuring unit malfunction; these units are not field serviceable.

Ranked causes

  1. 1. CPU board failure (rare)
  2. 2. MiniOM measuring unit failure (rare)
  3. 3. MiniTPX measuring unit failure (rare)
First check

Identify failed component

Test point
Affected measurement unit or circuit board
Expected
Normal operation or no error codes

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NMT measurement output incorrect or waveforms not displaying

TOF% value not within 95-105, Count not equal to 4, T1% not within 95-105, or stimulus pulses/responses not visible on waveform

Ranked causes

  1. 1. Incorrect monitor configuration settings (common)
  2. 2. Simulator not transmitting proper response signal (occasional)
  3. 3. Module hardware malfunction (rare)
First check

Verify monitor software version and apply correct configuration

Test point
Monitor software version display
Expected
Settings match software version (3.1 or earlier: Start with New patient, Current Supra, TOF, Volume 6, Cycle 20s; 3.2+: AUTO, Supramax Selected, TOF, Volume 6, Cycle 20s)

6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No response to breathing patient safety

Module does not detect or display respiratory activity when patient breathes

Ranked causes

  1. 1. Gas sample line disconnected (common)
  2. 2. Sampling line or water trap loose or blocked (common)
  3. 3. Sample gas out blocked (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

PDM data not displayed on monitor

Cardiac output, invasive pressure, or other PDM measurements missing from monitor screen

Ranked causes

  1. 1. Missing cardiac output license (common)
  2. 2. Missing PDM license (common)
  3. 3. Configuration issue with parameter priority (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Recorder does not work

B1X5-REC recorder unit non-functional

Ranked causes

  1. 1. Recorder connect cable loose or faulty (common)
  2. 2. Recorder unit failure (occasional)
  3. 3. Host or frame issue (unknown)
First check

Verify recorder connect cable integrity and connection

Test point
Recorder connector
Expected
Cable intact and properly seated

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor check failed patient safety

BIS sensor impedance check fails; one or more electrode impedances exceed acceptable threshold.

Ranked causes

  1. 1. Sensor improperly attached to patient (common)
  2. 2. Faulty sensor (occasional)
  3. 3. PIC+ cable or BISx unit faulty (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Supramaximal current search fails or value exceeds limit

Detected supramaximal current is greater than 70 mA, or search does not complete

Ranked causes

  1. 1. Incorrect simulator response knob setting (common)
  2. 2. Simulator Fade setting not correct (common)
  3. 3. Sensor cable or ElectroSensor defective (occasional)
First check

Verify simulator response knob position

Test point
NMT Simulator response control
Expected
Response knob set to maximum

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Too high EtCO2 value patient safety

End-tidal carbon dioxide concentration above expected for patient

Ranked causes

  1. 1. Water trap contamination or malfunction (common)
  2. 2. Calibration error (common)
  3. 3. Incorrect humidity compensation setting (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to perform function or feature not available

Feature or function is unavailable or grayed out in menu

Ranked causes

  1. 1. License not purchased for feature (common)
  2. 2. Trial license expired (occasional)
  3. 3. License not installed properly (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Zero error message displayed

Message appears in parameter window indicating autozero failure during measurement or gas calibration startup

Ranked causes

  1. 1. Zero valve operation failure (common)
  2. 2. Defective zero absorber or dehumidification tube (common)
  3. 3. Module internal failure (rare)
First check

Verify zero valve operation

Test point
Zero valve
Expected
Valve opens and closes properly

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Artifacts message on entropy measurement patient safety

Entropy parameter display shows Artifacts; signals contain noise or artifact

Ranked causes

  1. 1. Patient movement, eye blinks, shivering (common)
  2. 2. Deep breathing or arrhythmia (common)
First check

Wait for signal to stabilize

Test point
Entropy parameter display
Expected
Artifacts message clears and measurement proceeds

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Barcode reader does not read multi-field barcode correctly

Multi-field barcode information not populated correctly to fields in Admit menu

Ranked causes

  1. 1. Parser configuration is incorrect (common)
  2. 2. Parser configuration is incompatible with barcode format (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

BIS waveform not appearing on screen

BIS parameter and waveform missing from monitor display; No sensor message shown instead

Ranked causes

  1. 1. BISx unit not connected to BIS module (common)
  2. 2. BIS EEG waveform not configured on monitor screen (occasional)
First check

Verify BISx unit is connected to BIS module with patient interface cable

Test point
BIS module connection port and BISx cable
Expected
Cable firmly connected, connection secure

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Gas sampling system and CO2 functional check

Functional verification of gas sampling system and CO2 measurement after visual inspection, O-ring replacement, or module case opening

Ranked causes

  1. 1. Routine calibration and functional verification required (common)
  2. 2. Gas sampling system leakage or flow anomaly (occasional)
First check

Visually inspect front cover

Test point
Module front cover
Expected
Front cover is intact

6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CARESCAPE ONE works but E-modules data not displayed

CARESCAPE ONE acquisition module operates normally but connected E-modules do not send data to monitor

Ranked causes

  1. 1. E-module defective (occasional)
  2. 2. Module connector defective (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check electrodes message displayed

Parameter window shows 'Check electrodes' message; measuring electrodes are off or stimulus electrode connection is broken

Ranked causes

  1. 1. Measuring electrodes are off (common)
  2. 2. Broken connection of stimulating electrode or cable (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check Water Trap message displayed

Alarm area and waveform area display 'Check Water Trap' message. Water trap connection or internal tubing issue.

Ranked causes

  1. 1. Water trap is not connected (common)
  2. 2. Air leak inside the internal tubing (occasional)
First check

Verify water trap is physically connected to the module

Test point
Water trap connector
Expected
Water trap firmly connected

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Communication LED off (Ethernet Activity)

Communication (Ethernet Activity) LED is off, indicating loss of communication between frame and host or module malfunction.

Ranked causes

  1. 1. EMBC (Frame) faulty and not communicating (common)
  2. 2. PDM module faulty and not communicating (common)
First check

Verify EMBC and host communication from service interface

Test point
Service interface communication status
Expected
EMBC communicating with host

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Communication LED off or not blinking

Ethernet Activity LED is off, indicating no communication between frame (EMBC) and host, or PDM module and host

Ranked causes

  1. 1. EMBC (Frame) faulty or not communicating (occasional)
  2. 2. PDM module faulty or not communicating (occasional)
First check

Verify Communication LED blinking when EMBC and host are connected

Test point
Communication LED (Ethernet Activity LED)
Expected
Blinking

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Connected device software activation fails or shows update failed message
  • Contact Monitors button is inactive
  • CPU board malfunction
  • Erroneous gas readings after maintenance
  • EtCO2 over scale >15%
  • EtCO2 over scale above 20 percent
  • Expiratory TV greater than inspiratory TV
  • F2 Frame is not recognized by monitor
  • Fluctuating airway resistance (Raw)
  • Gas sampling system leakage suspected
  • Incompatible or duplicate modules connected
  • Invasive pressure channel zeroing failed
  • Invasive pressure measurement under range
  • Loose parts or internal cable connections
  • Low gas sample flow
  • Low volumes message displayed
  • MiniOM oxygen signal noisy or unstable
  • Modules cannot be placed properly in F2 Frame
  • NMT module not recognized or not communicating
  • No Entropy waveform or parameter on screen
  • No picture on screen despite monitor powered on
  • On/Standby LED off when monitor powered on
  • On/Standby LED off while monitor is on
  • PDM module works but E-module data not visible
  • Recorder works but nothing appears on paper
  • Reference not stable message
  • Replace Water Trap message displayed
  • Sensor impedance check not available in menus
  • Sensor INOP (inoperative)
  • Settings file password incorrect or file corrupted
  • Startup procedure does not advance beyond welcome screen
  • Strange or nonphysiological gas exchange values
  • Strongly vibrating spirometry loop
  • System time is incorrect without network connection
  • Testing DSC message displayed
  • Too high airway resistance (Raw)
  • UAC disconnect message appears
  • Unable to activate new monitor software
  • Unable to perform gas calibration
  • Unable to upload license file
  • Water trap connection issue
  • Water trap filter failure
  • Wrong character displayed when barcode is read
  • Abnormally high or low CO2 response
  • Absolute pressure sensor error
  • Accessory failure or degradation
  • Apnea alarm not activated
  • Apnea detection not working
  • Apply sensor
  • Arterial invasive pressure disconnected
  • Artifact
  • Artifact message
  • Automatic check off
  • BIS cable off
  • BIS DSC error
  • BIS measurement removed
  • BIS module error
  • BIS sensor check failed
  • BIS sensor expired
  • Bypass flow high
  • Cable off
  • Cable off message on entropy measurement
  • Calibrating gas sensor (warm-up in progress)
  • Calibrating message appears
  • Calibrating message displayed
  • Calibration: Agent identification
  • Calibration: Air leak detection
  • Calibration: Airway gases measurement
  • Calibration: Ambient pressure verification
  • Calibration: Apnea detection
  • Calibration: BIS module functional check
  • Calibration: CO2 waveform and FiO2 level setup
  • Calibration: D19 display maintenance check
  • Calibration: E-NSATX module maintenance check
  • Calibration: E-PiCCO module functional check
  • Calibration: Entropy module functional check
  • Calibration: Fan operation
  • Calibration: Flow waveform measurement
  • Calibration: Frame F5-01 and F7-01 maintenance check
  • Calibration: Gas calibration
  • Calibration: Gas exhaust blockage detection
  • Calibration: Gas sampling system leak test
  • Calibration: Module keys operation
  • Calibration: NMT module functional check
  • Calibration: Occlusion detection
  • Calibration: Pulse contour cardiac output
  • Calibration: Reference gas flow rate check
  • Calibration: Respiratory module functional check
  • Calibration: Sample flow rate check
  • Calibration: Spirometry system leak test
  • Calibration: Visual inspection before functional check
  • Calibration: Zero valve operation
  • CCO Calibration fail message
  • CCO meas. fail message
  • Checking sensor
  • Checking sensor message on entropy measurement
  • Checking sensor message stays longer than 2 minutes
  • CO measurement removed message
  • Complete functional check and reset monitor
  • Compliance value invalid
  • Confirm calibration message
  • Confirm electrode contact issue on entropy measurement
  • Demo data
  • Demo data message on entropy measurement
  • Device authentication failed
  • DSC error
  • Duplicate monitor addresses detected
  • E-module cannot be placed properly in frame
  • E-module showing erroneous readings
  • E-musb and CARESCAPE ONE simultaneously connected
  • Electrical safety test
  • Electromagnetic interference detected
  • EMG electrodes off message
  • Entropy cable off alarm on entropy measurement
  • Entropy measurement removed message
  • Entropy module not connected
  • Entropy sensor check failed
  • Entropy sensor off
  • Erroneous readings after maintenance
  • Erroneous readings after module cover reassembly
  • Error messages displayed in message field
  • EtCO2 exceeds PaCO2 arterial value
  • EtCO2 greater than PaCO2
  • Ethernet Link LED blinking
  • Ethernet Link LED on (solid green)
  • F2 Frame communication failure with host monitor
  • F5/F7 Frame is not recognized by monitor
  • F7 Frame used standalone causing ventilation blockage
  • Failure in Agent ID
  • Femoral invasive pressure disconnected
  • Flow or Paw waveform exceeds scale
  • Gas exhaust blockage not detected
  • Gas flow out of measurement range
  • Gas measurement failed
  • Gas measurement source detection ongoing
  • Gas measurements removed
  • Gas module disconnected
  • Gas module in standby mode
  • Gas sensor calibration in progress
  • High BIS impedance
  • Host software activation completes but monitor does not restart
  • Host software activation takes longer than expected
  • Host software upgrade activation code rejected or not accepted
  • Identical BIS modules
  • Identical C.O. modules message
  • Identical Entropy modules
  • Identical frame detected
  • Identical frame detected in monitoring system
  • Identical gas modules installed
  • Identical IP8 modules detected
  • Identical modules causing patient safety risk
  • Identical NMT modules message
  • Incompatible device: BIS module
  • Incompatible DSC
  • Incompatible gas module detected
  • Incompatible sensor
  • Invalid software image file
  • Invasive pressure calibration required
  • Invasive pressure channel in standby
  • Invasive pressure channels not zeroed
  • Invasive pressure sensor disconnected
  • Isoelectric EEG detected
  • Loop is being saved
  • Low EEG signal
  • Low perfusion at measurement site
  • Measurement off message
  • Measurement parameters not displaying correctly
  • MiniPVX measuring unit malfunction
  • MiniTPX measuring unit malfunction
  • Module box loose parts or cable connections
  • Module error
  • Module frame in wrong position causing malfunction
  • Module frame not properly grounded
  • Module not identified correctly
  • Module not responding
  • Monitor authentication failed
  • Monitor displays Field Replacement Unit startup message
  • Monitor not connected to mains power during activation
  • Multiple electrodes poor contact on entropy measurement
  • Multiple identical gas modules detected
  • Multiple identical SpO2 modules in system
  • Multiple identical SpO2 sources detected
  • N2O detected
  • NMT cable removed message
  • NMT measurement removed message
  • NMT sensor removed or Sensor off message
  • No BIS sensor
  • No BIS waveform displayed on screen
  • No CCO catheter message
  • No Entropy sensor detected
  • No invasive pressure displayed
  • No license for F5/F7 frame
  • No module message
  • No SpO2 signal or sensor problem
  • No uploaded software version available to activate
  • Noise interference detected
  • Not calibrated message
  • Occlusion detection not working
  • Over range FiO2 measurement
  • Over scale waveform
  • Over Scale waveform clipping
  • Parameter not displayed or not active
  • PDM not working when connected to ePort
  • Perform electrical safety tests
  • Perform SpO2 functional check
  • Perform visual inspection after service
  • Performing visual inspection
  • Poor signal
  • Pressure pulsation sensed during zeroing
  • Pulse rate high or low alarm
  • Re-calibrate CCO required
  • Recorder is printing but nothing appears on paper
  • Reference set message
  • Regional Block message
  • Replace sensor message displayed
  • Sample flow rate out of specification
  • Scale changed
  • Sensor impedance check fails during operation
  • Service CO module required
  • Setting reference message
  • Settings cannot be activated on older software versions
  • Software activation does not start after upload
  • Software activation requires base license
  • Software upload cannot be initiated
  • Software upload fails or does not complete
  • SpO2 acquisition module malfunction
  • SpO2 measurement module disconnected
  • SpO2 measurement searching for pulse
  • SpO2 sensor has failed
  • SpO2 sensor not compatible
  • SpO2 sensor not connected
  • SpO2 sensor off patient
  • SpO2 value high or low alarm
  • Start measurement message
  • Supramax not found message
  • Supramax search message
  • System time is incorrect with network connection
  • TETANIC message
  • Too high peak inspiratory pressure (Ppeak)
  • Too low EtCO2 value
  • Tubing contamination detected
  • Unable to view feature despite enabled license
  • Verify measuring equipment is calibrated
  • Verify network device connectivity
  • VO2 not measurable, high FiO2
  • VO2 value invalid or no VO2 with high FiO2
  • Wait to continue message
  • Water trap not detected
  • Wrong character displayed when keyboard key is pressed
  • Wrong software package is activated
  • Zero adj >100 mmHg message
  • Zero ICP separately message
  • Zero valve not functioning
  • Zeroed message appears
  • Zeroing error
  • Zeroing in progress
  • Zeroing in progress message displayed
  • Zeroing message appears
  • Zeroing message displayed

259 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
0xGAS0001Service gas module. CO2, N2O, AA sensor failedMiniTPX unit failureerror
0xGAS0010Service gas module. O2 sensor failedMiniOM oxygen sensor failureerror
0xGAS0011Simultaneous MiniTPX and MiniOM oxygen sensor failureBoth CO2/N2O/AA and O2 sensor failureserror
0xGAS0100Service gas module. Spirometry sensor failedSpirometry sensor failureerror
0xGAS0101Simultaneous MiniTPX and spirometry sensor failureBoth CO2/N2O/AA sensor and spirometry sensor failureserror
0xGAS0110Simultaneous O2 sensor and spirometry sensor failureBoth MiniOM and spirometry sensor failureserror
0xGAS0111Simultaneous MiniTPX, O2, and spirometry sensor failureAll three sensor types failederror
0xGAS1000Service gas module. Gas measurement failedGeneral gas measurement failureerror
0xGAS1001Simultaneous MiniTPX unit and general gas measurement failureMiniTPX and general measurement failureerror
0xGAS1010Simultaneous O2 sensor and general gas measurement failureMiniOM and general measurement failureerror
0xGAS1011Simultaneous MiniTPX, O2, and general measurement failureAll three component failures with general measurement failureerror
0xGAS1100Simultaneous spirometry sensor and general gas measurement failureSpirometry sensor and general measurement failureerror
0xGAS1101Simultaneous MiniTPX, spirometry, and general gas measurement failureAll measurement and sensor failures combinederror
0xGAS1110Service gas module gas measurement failedSimultaneous MiniOM oxygen sensor, spirometry sensor, and general gas measurement failureerror
0xGAS1111Service gas module gas measurement failedSimultaneous MiniTPX measuring unit, MiniOM oxygen sensor, spirometry sensor, and general gas measurement failureerror
>320 mmHg or >43 kPaInvasive pressure measurement over rangeMeasurement exceeds maximum range or sensor/cable faultwarning
<-40 mmHg or <-5 kPaInvasive pressure measurement under rangeMeasurement below minimum range or sensor/cable faultwarning
Apnea deactivatedApnea alarm is not active or has not met activation criteria.Apnea alarm start-up conditions not reached.warning
Art 1-8 disconnectArterial invasive pressure disconnectedNo arterial invasive pressure detectedwarning
Auto detecting sourceSystem is automatically detecting gas measurement source.Module initialization or gas source change.info
CalibratedChannel calibrated successfullyCalibration procedure completed without errorsinfo
CalibratingGas sensor calibration is in progress.Manual calibration started or scheduled warm-up calibration.info
Calibrating gas sensorModule is warming up and gas measurements are not yet available.Gas module connected or powered on.info
Calibration errorGas sensor calibration failed and must be repeated.Calibration gas not fed within one minute; wrong gas used; gain adjustment error; trapped CO2.error
CCO Calibration failUnsuccessful calibrationCalibration procedure fails to complete or validateerror
CCO meas. failCalibration has failedPrevious calibration validation failederror
Check sample gas outSample gas output path is blocked or water trap has an issue.Blockage in sample line or water trap connection failure.warning
Check Water TrapWater trap is not connected or has internal leakage.Water trap disconnection or internal tubing leak detected.warning
CO measurement removedThe module is disconnectedC.O./CCO module becomes disconnected from monitorwarning
Confirm calibrationMeasurement data has not been confirmed before trying to exit Cardiac Output/CCO Calibrate menuUser attempts to exit Calibrate menu without confirming measurementwarning
Continuous blockage. Check sample line and water trap.Persistent blockage in gas sampling path despite previous checks.Ongoing obstruction in sample line, water trap or internal tubing.error
F2 frame not authenticated: Call serviceF2 Frame authentication requiredF2 Frame fails authentication check during initializationerror
Failure in Agent IDAnesthetic agent identification system malfunction.Agent ID sensor or circuit failure.error
Fem 1-8 disconnectFemoral invasive pressure disconnectedNo femoral arterial invasive pressure detectedwarning
Gas measurements removedGas acquisition module has been disconnected.Module unplugged or removed from system.info
Gas module standbyGas module pump is shut down and module is in standby mode.Module pump manually stopped or auto-shutdown triggered.info
Identical C.O. modulesTwo or more identical C.O./CCO modules detected in systemMultiple E-COP, E-COPSv, or E-PiCCO modules connected simultaneouslyerror
Identical frame detectedTwo or more identical module frames detected in systemMultiple F2, F5, or F7 frames connected simultaneously to monitoring systemerror
Identical gas modulesSystem detected two or more identical gas modules installed.Multiple identical modules connected to system simultaneously.warning
Identical IP8 modulesDuplicate IP modules on same channelTwo or more E-PiCCO or E-COPSv modules map to same channelerror
IP's not zeroedOne or more invasive pressure channels not zeroedSystem detects unzeroed invasive pressure channelswarning
Low gas sample flowSample flow rate is below minimum specification (less than 80% of nominal).Flow rate measurement below 80% of module nominal flow.warning
Low volumesTidal volume detection failure or circuit leakI:E detection failure for 20 seconds without apnea or low measured volumeswarning
No CCO catheterThe catheter is disconnected from the cable, or the cable is disconnected from the moduleCatheter or cable disconnection detectedwarning
No license for F2 frameF2 Frame support license not enabled in configurationF2 Frame license not purchased or activated in system configurationwarning
No moduleNo C.O. module connectedUser accesses C.O. menu without module installedwarning
Not calibratedThe module and cable are connected, but the CCO measurement has not been calibratedUser attempts to measure without prior calibrationwarning
Over rangeMeasured FiO2 exceeds maximum valid range (>103%).FiO2 measurement reading above 103%.error
Over scaleWaveform display clipped due to gas concentration exceeding current scale upper limit.Gas concentration exceeds scale maximum for selected waveform scale.warning
P1-8 standbyInvasive pressure channel in standby modeUser or system placed IP channel in standbyinfo
P1-8 zeroing failedInvasive pressure channel zeroing failedZeroing procedure could not complete successfullyerror
Pressure measurement removedInvasive pressure acquisition device removedIP measurement device physically removed from systemwarning
Pressure SensedPressure pulsation detected during zeroingDynamic pressure detected when static zeroing should occurwarning
Replace Water TrapWater trap is defective, contaminated, or occludedWater trap assembly failure or contamination detectedwarning
Sample line blockedGas sampling line or water trap system is blocked or fullBlockage detected in sampling line or water trap container full conditionwarning
Sensor disconnectedInvasive pressure sensor cable disconnectedTransducer detects disconnection or cable removed from modulewarning
UAC 1 disconnect to UAC 4 disconnectInvasive pressure line is disconnectedInvasive pressure line becomes disconnectedwarning
Zero adj >100 mmHgOffset during zeroing exceeded 100 mmHgZeroing process shows offset greater than 100 mmHgerror
ZeroedZeroing was successfulZeroing process completed successfullyinfo
Zero errorAutozero calibration has failedAutozero failure during measurement or at start of gas calibrationerror
Zero ICP separatelyICP channel must be zeroed separately from other invasive pressuresUser attempts to zero all channels together including ICPwarning
ZeroingIP channel is currently being zeroedZeroing process in progressinfo
ZeroingZeroing procedure is in progressSystem is performing automatic zero calibrationinfo

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.