GE Healthcare CardioLab: troubleshooting and error codes
The ServiceForge knowledge base holds 102 documented symptoms for the GE Healthcare CardioLab, with 221 ranked causes, 163 diagnostic checks with test points and expected values, and 26 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Excessive noise on intracardiac signal traces patient safety
Traces appear fuzzy, thick, or dull; noise appears on one or all intracardiac channels during or after cardiac procedures
Ranked causes
- 1. Power line frequency noise (60Hz or 50Hz) (common)
- 2. Bad electrical grounds (power outlets, cords, equipment) (common)
- 3. Bad ablation device cabling or personality module (occasional)
- 4. Poor routing of wires around other equipment (occasional)
- 5. Poor electrode connection or Right Leg lead issue (common)
- 6. High gain amplification (occasional)
- 7. Bad or worn patient cables or lead wires (common)
- 8. Bad grounding pad (occasional)
Measure period of sine wave on trace to confirm 60Hz or 50Hz power line frequency
- Test point
- Intracardiac trace
- Expected
- 16.7ms for 60Hz or 20ms for 50Hz, ±0.5ms
- Tools
- Oscilloscope or display speed measurement
12 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ablation noise on ECG
Large spikes and noise on waveform occur while RF generator actively ablates heart; disappears when generator turned off
Ranked causes
- 1. Bad lead wires (common)
- 2. Bad patient cables (common)
- 3. Poor electrode connections with patient (common)
- 4. Bad ablation grounding pad (occasional)
- 5. Proximity of electrodes/lead wires to RF source (occasional)
- 6. Improper grounding of equipment (occasional)
Inspect lead wires for damage or wear
- Test point
- All ECG lead wires
- Expected
- No visible damage, corrosion, or wear
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ablation noise on intracardiac signal high voltage
Noise occurs while RF generator actively ablates; disappears when generator is turned off. Higher power settings produce higher amplitude noise. Most difficult to eliminate on Ablation Distal signal; noise expected above 35W.
Ranked causes
- 1. High power setting on RF generator (above 35W) (common)
- 2. Poor electrode connections with patient (common)
- 3. Bad ablation grounding pad (occasional)
- 4. Proximity of electrodes or lead wires to RF source (occasional)
- 5. Equipment grounding issues (occasional)
- 6. Old EPT personality module without Ablate and Record capability (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Supplies added to CVIS do not appear on Mac-Lab
List items added in CVIS system are not visible in Mac-Lab/CardioLab
Ranked causes
- 1. INW service that processes lists not configured or started (common)
- 2. CVIS List Sharing Delta export not configured for the location (common)
- 3. Corrupt XML export from CVIS system (occasional)
- 4. CVIS List Sharing services not started on CVIS server (common)
- 5. Replication errors or INW server issues (occasional)
- 6. Network connection between CVIS and INW server lost (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot connect to remote DICOM host
Mac-Lab/CardioLab system fails to establish DICOM connection to CVIS, EMR, or scheduling system; no patient records returned or error messages displayed
Ranked causes
- 1. Incorrect hostname or IP address configured (common)
- 2. Incorrect port number configured (common)
- 3. Incorrect AE Title configured (common)
- 4. Transport Layer Security (TLS) mode mismatch (occasional)
- 5. Remote DICOM host not actively listening (occasional)
Resolve hostname to IP address
- Test point
- Command Prompt on Mac-Lab/CardioLab
- Expected
- Valid IP address resolved or configured IP address is reachable
- Tools
- DICOMDiag tool
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot establish connection to patient monitor patient safety
Mac-Lab/CardioLab application reports connection lost or cannot be established to patient monitor at pre/post-workstation
Ranked causes
- 1. Physical connection issue between workstation and patient monitor (common)
- 2. Patient monitor configuration mismatch (occasional)
- 3. Patient monitor not powered on or devices not connected (common)
- 4. CARESCAPE B450 software version outdated (occasional)
- 5. Review workstation needs reboot (occasional)
Verify physical cable connections
- Test point
- Connection between workstation and patient monitor
- Expected
- All cables properly connected and tight
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Intracardiac signal noise or degraded waveform display patient safety
Lost or degraded display and accuracy of intracardiac waveforms due to electromagnetic disturbances; intracardiac signal noise may occur
Ranked causes
- 1. Use with high-frequency surgical equipment (common)
- 2. Portable RF communications equipment used too close (common)
- 3. Incorrect HDMI video cables installed (occasional)
- 4. Non-approved accessories, transducers or cables used (occasional)
- 5. Strong intentional radio transmitters in vicinity (occasional)
Verify distance of portable RF communications equipment from system
- Test point
- All parts of CardioLab system and its cables
- Expected
- All RF equipment and antennas at least 30 cm (12 in) away
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacing artifact noise on ECG patient safety
Noise with spikes occurs while pace generator actively paces heart; disappears when pacing pulses turned off
Ranked causes
- 1. Excessively high pacing current settings (common)
- 2. Bad stimulator cables (common)
- 3. Loose fitting or bad stimulator connections (common)
- 4. Stimulator Isolation/Interface box low batteries (occasional)
- 5. Charge build-up on pacing poles from extended pacing (occasional)
Verify pacing current settings
- Test point
- Stimulator milliamp output and duration controls
- Expected
- Current set to 1-5mA @ 2.0ms or clinically appropriate levels
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Send Movies option disabled in EP recording preferences
Send Movie Option grayed out when configuring EP recording preferences for CARTO 3 System movies
Ranked causes
- 1. MapImageImport directory not shared on CardioLab (common)
- 2. MapImageImport directory permissions not set for CARTO 3 interface user (common)
- 3. CARTO 3 interface user credentials mismatch (common)
- 4. CARTO 3 interface user account password changed (occasional)
- 5. CARTO 3 System moved to unconfigured CardioLab (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to capture X-ray image in image capture window
Image grabber window is blank or displays 'Failed to capture an image' error
Ranked causes
- 1. DVI cable not connected to correct OrionHD board connector (common)
- 2. Incorrect .dcf configuration file selected (common)
- 3. OrionHD board driver not installed or not detected (occasional)
- 4. MLCL or ImgTest application still running (occasional)
- 5. Non-standard video source format not supported (rare)
Verify DVI cable connection
- Test point
- DVI connector #3 on OrionHD board
- Expected
- DVI-to-5BNC cable securely connected; green and blue BNC inputs available
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baseline drift noise on ECG
ECG trace moves vertically up and down; noise frequency varies and correlates with physical movement near patient
Ranked causes
- 1. Bad or worn lead wires (common)
- 2. Bad or worn patient cables (common)
- 3. Poor electrode connection with patient (common)
- 4. Patient table not grounded properly (occasional)
Inspect lead wires for damage or wear
- Test point
- All ECG lead wires
- Expected
- No visible damage, corrosion, or wear
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baseline drift noise on intracardiac signal
Entire trace moves vertically; low frequency noise similar to ECG baseline drift noise
Ranked causes
- 1. Bad lead wires or patient cables (common)
- 2. Poor electrode connection or location (common)
- 3. Patient movement or drape movement (occasional)
- 4. Patient table not grounded properly (occasional)
Adjust high pass filter to higher setting
- Test point
- High pass filter setting
- Expected
- Higher than normal (normal: 30 Hz for intracardiac signal)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CardioLab amplifier does not send signals to system
Amplifier is not responding when a new study is started; system displays 'Prucka 3 Amplifier is not responding' or 'Clab II Amplifier is not responding' message
Ranked causes
- 1. Amplifier not plugged in or turned off (common)
- 2. External cables not connected or not secure (common)
- 3. Fiber optic cable (FOC) defective or dirty (occasional)
- 4. FOC connector not properly mated or no light output (occasional)
Verify amplifier power status
- Test point
- Amplifier power switch
- Expected
- Amplifier plugged in and powered on
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cases performed on Mac-Lab do not appear in CVIS
Study data acquired on Mac-Lab/CardioLab system does not transfer to CVIS archive
Ranked causes
- 1. Interfaces on CVIS server not running or configured incorrectly (common)
- 2. Data Export Utility on Mac-Lab not configured correctly (common)
- 3. Enable file export on study close not checked (common)
- 4. Network connection between Mac-Lab and CVIS server down (common)
Verify file export enabled on study close
- Test point
- Administration > System Settings > File Paths
- Expected
- Enable file export on study close option is checked
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
DICOM interface connectivity not established
Mac-Lab/CardioLab system cannot communicate with remote DICOM hosts (X-ray, ultrasound, CVIS/EMR)
Ranked causes
- 1. Incorrect DICOM AE Title configuration (common)
- 2. Incorrect Listen Port or remote host port configuration (common)
- 3. Remote host DICOM configuration incomplete or incorrect (common)
- 4. Security mode mismatch between systems (occasional)
Verify local system AE Title is configured
- Test point
- Administration > System Settings > Connectivity > Local System > DICOM Settings
- Expected
- AE Title filled with alphanumeric/underscore, ≤16 chars, case-sensitive
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mapping system noise on intracardiac signal
Noise typically associated with power line frequency noise or magnet-generated noise. Can be measured by display speed to determine if 60Hz or 50Hz.
Ranked causes
- 1. Power line frequency noise or magnet interference (common)
- 2. Poor electrode connections with patient (common)
- 3. Equipment grounding issues (occasional)
- 4. 5.6 KHz End Guide Signal interference (rare)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mapping system noise on intracardiac signals patient safety
Noise visible on intracardiac mapping signals; typically power line frequency or magnet-generated noise
Ranked causes
- 1. Power line frequency interference (60 Hz or 50 Hz) (common)
- 2. Poor electrode connections or improper electrode placement (common)
- 3. Magnet interference or 5.6 KHz End Guide Signal interference (occasional)
- 4. Inadequate grounding of equipment and patient (occasional)
Measure noise frequency with CardioLab calipers
- Test point
- Intracardiac ECG display
- Expected
- 16 ms time base peak-to-peak for 60 Hz power line; magnet noise differs
1 more check with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Muscle artifact noise on ECG
Variable noise riding on ECG trace(s) caused by patient movement such as shivering or snoring
Ranked causes
- 1. Bad or worn lead wires (common)
- 2. Bad or worn patient cables (common)
- 3. Poor electrode connection with patient (common)
- 4. Patient movement caused by cold, medications, or snoring (common)
Inspect lead wires for damage or wear
- Test point
- All ECG lead wires
- Expected
- No visible damage, corrosion, or wear
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacing artifact noise on intracardiac signal
Noise occurs while pace generator actively paces heart; disappears after pacing pulses are turned off. Amplitude increases with higher pacing current.
Ranked causes
- 1. Excessively high pacing current settings (common)
- 2. Bad stimulator cables or loose connections (common)
- 3. Stimulator isolation box with low batteries (occasional)
- 4. Charge build-up on pacing poles from prolonged pacing (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient search returns no results
CVIS patient search fails to return scheduled or admitted patients from Mac-Lab/CardioLab system
Ranked causes
- 1. AE Title mismatch between Mac-Lab/CardioLab and CVIS (common)
- 2. DICOM settings not configured correctly on Mac-Lab (common)
- 3. DICOM services not running on CVIS server (common)
- 4. Network connection lost between Mac-Lab and CVIS (common)
Verify AE Title configuration matches between systems
- Test point
- Administrative > System Management > DICOM Maintenance Location drop-down
- Expected
- AE Title in Mac-Lab matches configured value in CVIS (e.g., MLCL_Lab1)
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power line frequency noise on ECG traces
ECG traces appear fuzzy, thick, or dull due to 60 Hz or 50 Hz power line noise
Ranked causes
- 1. Power line filter turned off (common)
- 2. Bad or worn lead wires (common)
- 3. Bad or worn patient cables (common)
- 4. Poor electrode connection with patient (common)
Verify power line filter setting
- Test point
- ECG filter menu option
- Expected
- Power line filter set to Fixed or Adaptive
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CARTO 3 System button inactive
CARTO 3 System interface button is not clickable or grayed out in CardioLab application
Ranked causes
- 1. CARTO 3 System interface not enabled on CardioLab (common)
- 2. Interface not accessed from Real-Time window (common)
- 3. Software option disabled (occasional)
Verify CARTO 3 interface is enabled
- Test point
- CardioLab Administration settings
- Expected
- CARTO 3 System interface enabled and checkbox is not grayed out
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CARTO 3 System movies not transferring to CardioLab
Movie files from CARTO 3 System fail to transfer to CardioLab system despite proper configuration attempts
Ranked causes
- 1. CARTO 3 System configuration incorrect (common)
- 2. CARTO System interface user account not created (occasional)
- 3. File sharing disabled on CARTO adapter (common)
Verify file sharing setting for CARTO adapter
- Test point
- CARTO Connection Properties
- Expected
- File and Printer Sharing for Microsoft Networks enabled
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mapping points unsynchronized or missing patient safety
CARTO 3 System mapping points not synchronized with CardioLab or missing from display
Ranked causes
- 1. CARTO 3 System point filter filtering out points (occasional)
- 2. Points manually deselected in report (occasional)
- 3. Waveform synchronization off by more than 1 second (occasional)
Verify waveform synchronization timing
- Test point
- CARTO 3 System waveform and CardioLab recording timestamps
- Expected
- Synchronization within 1 second or less
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mapping system noise on surface ECG traces patient safety
ECG traces appear fuzzy, thick, or dull; power line frequency or magnet noise visible on one or all traces
Ranked causes
- 1. Power line frequency interference (60 Hz or 50 Hz) (common)
- 2. Improper grounding of equipment and patient table (common)
- 3. Inadequate right leg lead connection (common)
Measure power line frequency period with time calipers
- Test point
- ECG display at high speed
- Expected
- 16.7ms peak-to-peak for 60Hz or 20ms for 50Hz
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Over-filtered blocky ECG signals
ECG signal appears over-filtered with squared-off waveform appearance
Ranked causes
- 1. ECG filter setting too high (e.g., 5 Hz high-pass) (common)
- 2. Low sampling rate (occasional)
- 3. Non-optimal screen resolution or contrast/brightness (occasional)
Verify ECG high-pass filter setting
- Test point
- ECG filter menu option
- Expected
- Filter set to 0.05Hz or 0.5Hz
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient search returns no results or error message
When attempting Patient Search in Navigator with Centricity Cardiology DMS or Workflow, no patient records are returned or error message displayed indicating configuration or connectivity issue
Ranked causes
- 1. DICOM configuration settings incorrect on Mac-Lab/CardioLab (common)
- 2. CVIS system configuration settings incorrect (common)
- 3. CVIS system not actively listening for connections (occasional)
Run DICOMDiag tool to validate client connectivity
- Test point
- All configured remote DICOM hosts
- Expected
- No failures reported; all connection steps successful
- Tools
- DICOMDiag command-line tool
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Random spikes or square waveforms in signals when not stimulating
Unwanted spikes or square wave patterns appear in signal traces during non-stimulation periods
Ranked causes
- 1. Dirty or defective fiber optic cable (common)
- 2. Defective catheter, catheter extender, or catheter junction box (occasional)
- 3. Defective catheter input module (occasional)
Clean fiber optic cable contacts
- Test point
- FOC connectors at amplifier and computer
- Expected
- Contacts clean and free of debris
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to connect to remote server via Web service
System cannot establish connection to remote server through Mac-Lab/CardioLab Web service
Ranked causes
- 1. World Wide Web Publishing Service is not running (common)
- 2. Internet Information Services (IIS) is not started (common)
- 3. Default Web service is not running (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unipolar high frequency noise on intracardiac signal
Difference in noise level between catheter pole and channel reference (surface WCT or Auxiliary Reference). Unipolar signals more susceptible due to common mode rejection issues.
Ranked causes
- 1. Bad lead wires or patient cables (common)
- 2. Poor electrode connection with patient (common)
- 3. Intracardiac cable proximity to other devices (WCT as reference) (occasional)
Adjust low pass filter to lower setting
- Test point
- Low pass filter frequency
- Expected
- 150 Hz (changed from 1000 Hz)
- Tools
- System software filter control
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
X-ray image Live and Freeze buttons are grayed out
Live and Freeze buttons on Image 1 or Image 2 toolbar are unavailable and cannot be activated
Ranked causes
- 1. Cardio Image option not enabled; only Image Capture option installed (common)
- 2. OrionHD image capture board not present or not detected (occasional)
- 3. OrionHD board driver not properly installed or functioning (occasional)
Verify Cardio Image software option is installed and enabled
- Test point
- System software options configuration
- Expected
- Cardio Image option enabled (not Image Capture option only)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Amplifier communication or register readback failure
CardioLab II Plus Amplifier shows failures during Register Readback or Walking Ones test, indicating data communication errors between amplifier boards and computer.
Ranked causes
- 1. Defective amplifier board or PCIe connection (occasional)
- 2. Firmware version mismatch or corruption (rare)
Read amplifier registers and verify no communication errors
- Test point
- Amplifier Test > Register Readback tab
- Expected
- No register read errors; all values stable
- Tools
- Amplifier Test software
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: MLLPService debug logging patient safety
Enable and disable detailed debug logging for troubleshooting MLLPService communication failures
Ranked causes
- 1. Need to diagnose MLLPService transfer failures in detail (occasional)
- 2. Patient health information exposure risk from logs (common)
Enable TRACE logging in MLLPService.config
- Test point
- [Drive]\GEData\config\MLLPService.config first line
- Expected
- log4cplus.logger.MLCL.Server.MLLPService=TRACE, STDOUT, MLLP
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CARTO 3 System connection failure
No communication between CardioLab and CARTO 3 System
Ranked causes
- 1. Defective or disconnected crossover cable (common)
- 2. Ethernet lights not active (common)
Inspect crossover cable connection
- Test point
- Ethernet ports on acquisition system and CARTO 3 System
- Expected
- Crossover cable firmly connected with no visible damage
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Computer name resolution fails on GE HealthCare system
Computer name resolution fails even when DNS is properly configured and flushed
Ranked causes
- 1. NetBIOS disabled by default on GE HealthCare acquisition or review system (common)
- 2. LLMNR disabled on GE HealthCare system (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
DNS name resolution fails
System cannot resolve computer names via DNS
Ranked causes
- 1. DNS cache is obsolete or corrupted (common)
- 2. DNS problem on server side (occasional)
Flush local DNS cache
- Test point
- Local DNS resolver
- Expected
- DNS cache cleared successfully
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
EtCO2 check airway adapter patient safety
Airway adapter removed, optically blocked, or adapter type changed without zeroing
Ranked causes
- 1. Airway adapter removed or contaminated with secretions (common)
- 2. Adapter type changed without re-zeroing sensor (occasional)
Inspect airway adapter for blockage or contamination
- Test point
- Adapter optical windows
- Expected
- Clean windows, no mucus or moisture
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
EtCO2 sensor faulty patient safety
Sensor reports capnostat source current failure, EEPROM checksum error, or hardware malfunction detected
Ranked causes
- 1. Sensor improperly seated or disconnected (common)
- 2. Sensor hardware failure (rare)
Verify sensor is properly connected
- Test point
- Sensor connector
- Expected
- Sensor seated firmly in connector
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
EtCO2 sensor over temperature patient safety
End Tidal CO2 sensor temperature exceeds operating limit, sensor may not function correctly
Ranked causes
- 1. Sensor exposed to extreme heat source (common)
- 2. Sensor hardware malfunction (rare)
Check for nearby heat sources
- Test point
- Sensor environment
- Expected
- No heat lamps or extreme heat sources within sensor area
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
EtCO2 zero required patient safety
Sensor requires zeroing due to required zero condition or zero error detected during operation
Ranked causes
- 1. Airway adapter contaminated or requires zeroing after adapter change (common)
- 2. Hardware error developing (rare)
Inspect airway adapter for contamination
- Test point
- Adapter windows
- Expected
- Clean, unobstructed optical windows
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
INNOVA CENTRAL/Touch screen connection status shows Listening
IGS system shows Listening on port message instead of Connected status
Ranked causes
- 1. INNOVA CENTRAL/Touch screen physically disconnected (common)
- 2. Port setting is not 889 (common)
Verify port number in INNOVA CENTRAL configuration
- Test point
- Administration > System Settings > Connectivity > INNOVA CENTRAL/Touch Screen section
- Expected
- Port = 889
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid number of groups message when creating or continuing study
Message 'Invalid number of groups' appears when creating or continuing a CardioLab study in the application
Ranked causes
- 1. CardioLab II Plus Amplifier PCI board not properly installed in slot 6 (common)
- 2. Study configuration has no amplifier groups configured (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lost connection to Study Host during active study patient safety
Error message 'Lost connection to Study Host. No further data can be saved in this Study' displays during active study on acquisition systems and review workstations
Ranked causes
- 1. Network connection to study host broken (common)
- 2. NetMsgSender service failed on study host (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
MLLPService not functioning properly
MLLP export to external CVIS/EMR systems is not working or data is not being sent
Ranked causes
- 1. EMR or receiving system is down (common)
- 2. Network outage preventing data export (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Network Messaging not active when not in a study
Error message 'Network Messaging is not active. Study catalog changes may not be visible' displays at Navigator on acquisition systems and review workstations when not in a study
Ranked causes
- 1. NetMsgSender background service not running or improperly configured (common)
- 2. Network connectivity loss to INW server (occasional)
Verify systems can ping each other across network
- Test point
- All systems on network
- Expected
- Ping responses successful
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient Search or Schedule Search operation fails
Unable to retrieve patient information or scheduled procedures from CVIS/EMR system
Ranked causes
- 1. Modality Worklist Host not configured (common)
- 2. Incorrect patient demographics field locking configuration (occasional)
Verify Modality Worklist Host configuration is complete
- Test point
- System Settings > Connectivity > Modality Worklist Host section
- Expected
- AE Title, Hostname/IP, Port number all filled; Cath Modality=HD by default
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paucal communication packet loss on amplifier board
Amplifier Test Status section shows COMM_QI values below 1.0, indicating packet loss between amplifier boards, affecting signal integrity.
Ranked causes
- 1. Loose or defective PCIe connection between amplifier boards (occasional)
- 2. Defective amplifier board or trace damage (rare)
Verify COMM_QI values for first ten communication cycles
- Test point
- Amplifier Test > Registers > Status section
- Expected
- All COMM_QI values equal 1
- Tools
- Amplifier Test software
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System cannot establish SSL/TLS secure channel trust relationship
Web service fails to establish secure channel, typically during client connection or data transfer
Ranked causes
- 1. Mac-Lab/CardioLab certificate is expired (common)
- 2. Mac-Lab/CardioLab certificate permission issue (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to communicate with cath amplifier
System cannot establish or maintain communication with PDM (Portable Device Module)
Ranked causes
- 1. System IP address in 192.168.0.xxx range (common)
- 2. PDM not registered on system (occasional)
Verify system IP address
- Test point
- Network Properties IP address
- Expected
- Not in 192.168.0.xxx range
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vital signs not displayed in Patient Status Area patient safety
Expected vital parameters do not appear in the Patient Status Area during study
Ranked causes
- 1. Patient monitor connection issue (common)
- 2. Respiration rate or temperature display disabled (common)
Verify patient monitor connection cabling
- Test point
- Connection between patient monitor and workstation
- Expected
- Proper cabling for monitor family
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Windows scaling level warning message displayed
System displays a warning message indicating the need to change Windows scaling level
Ranked causes
- 1. Windows scaling level not set to 125% (common)
- 2. User session needs refresh after scaling change (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Acquisition system not powered on
White LED not illuminated on acquisition system front panel
Ranked causes
- 1. Acquisition system power switch off (common)
Verify acquisition system power status
- Test point
- Front panel white LED
- Expected
- LED illuminated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode scanner triggers NIBP/NBP measurement
Scanning barcodes during a study causes unwanted NIBP or NBP measurements
Ranked causes
- 1. Barcode scanner not configured properly (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charge level indicator shows critical low patient safety
PDM battery LED test button shows one LED flashing, indicating less than 10% capacity remaining.
Ranked causes
- 1. Battery charge depleted below 10% (common)
Test battery charge status
- Test point
- TEST button on PDM battery
- Expected
- 1 LED flashing indicates less than 10% capacity
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CARTO 3 System IP address conflict
IP addressing issue between CardioLab and CARTO 3 System prevents communication
Ranked causes
- 1. IP address mismatch or conflict on network (common)
Use ping to test communication between systems
- Test point
- Network connectivity from CardioLab to CARTO 3 System
- Expected
- Ping successful from 192.168.5.2 to 192.168.5.1
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cath amplifier battery error reported caution
PDM displays battery error message indicating battery problem
Ranked causes
- 1. Battery overheating or not inserted properly (occasional)
Verify PDM battery insertion and condition
- Test point
- PDM battery compartment
- Expected
- Battery properly installed and not overheated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cath amplifier battery is low caution
PDM battery charge level is low and requires charging or replacement
Ranked causes
- 1. PDM battery insufficient charge (common)
Connect PDM to charging station
- Test point
- PDM Base Station Plus or PDM Slim Connect
- Expected
- PDM properly seated in charger
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cath amplifier hardware error reported
PDM displays hardware error message and system cannot operate
Ranked causes
- 1. PDM hardware failure or malfunction (rare)
Check Windows event log for PDM error code
- Test point
- Windows event log
- Expected
- PDM error code visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Default Web Site binding configuration issue
Web site binding not properly configured for SSL/TLS communication
Ranked causes
- 1. SSL certificate not properly selected in binding (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
DirectSight website cannot be reached
DirectSight website is inaccessible to clients
Ranked causes
- 1. DirectSight website is not running (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Enclosure leakage current exceeds limit
- EP recording disabled on CARTO 3 System
- Equipment failure or improper operation from adjacent equipment
- EtCO2 device connection attempting
- EtCO2 no breaths detected
- EtCO2 out of range high
- EtCO2 pump life exceeded
- EtCO2 pump off
- EtCO2 sample line disconnected
- EtCO2 sampling line check required
- EtCO2 sensor warm-up in progress
- Getting Windows ready message appears during startup
- HL7 export to CVIS fails
- INW server does not pull studies
- Isolation transformer not powered on
- Matrox Orion HD card firmware update required
- MLLPService not running or disabled
- Monitor not powered on
- Network failures on standalone acquisition system
- No data displayed on DirectSight or DirectSnap
- Patient lead leakage current exceeds limit - CardioLab amplifier
- Patient lead leakage current exceeds limit - PDM
- Patient leakage current exceeds limit
- Patient leakage current exceeds limit on CardioLab amplifier
- Patient monitor connection lost intermittently
- Patient monitor not powered on
- Patient sink leakage current exceeds limit - CardioLab amplifier
- PDM battery does not hold charge
- PDM battery requires replacement
- PDM does not support invasive pressures
- Peripheral devices not powered on
- Ping test for network connectivity
- Poor color output on acquisition system
- Pressure calibration offset detected
- Repeated and sporadic network disconnections
- Review workstation not powered on
- Slow network transfer and ping speed
- Studies not transferring to customer system
- System does not receive ablation device data
- System locks up after first barcode scan
- UPS not powered on
- Wireless scanner battery unable to hold adequate charge
42 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Attempting to connect to CO2 Device | System trying to establish communication link with CO2 device | Power-up, device reconnection, or communication loss recovery | info |
Capnostat in Sleep Mode | Sensor has been placed into sleep/standby mode unintentionally | Accidental sensor mode change or software command | warning |
Cath Amplifier battery error | PDM battery problem detected | PDM reports battery error condition | warning |
Cath Amplifier battery is low | PDM battery charge level insufficient | PDM detects low battery charge during operation or at startup | warning |
Cath Amplifier hardware error | PDM hardware failure detected | PDM reports hardware error condition | error |
Check Airway Adapter | Airway adapter removed, optically blocked, or adapter type changed without re-zeroing | Adapter disconnection, occlusion with secretions, or adapter type change without calibration | warning |
Check Sampling Line | Sampling line pneumatic pressure outside expected operating range | Kinked, occluded, or disconnected sampling line | warning |
CO2 No Breaths Detected | No respiratory cycles detected for extended timeout period | Timeout condition when breathing ceases or sampling line disconnected | warning |
CO2 Out of Range | Measured CO2 value exceeds upper limit of 150 mmHg (20.0 kPa, 19.7%) | Patient CO2 level exceeds maximum measurement range | warning |
COM+ Network Access (DCOM-In) | COM+ component object model network access error preventing remote connection | Review system attempts to connect to remote event logs during diagnostic process | warning |
EventDicomPatientSearchFailedFault | DICOM patient database search operation failed | DICOM query timeout or database unavailable | warning |
EventDicomScheduleSearchFailedFault | DICOM schedule database search operation failed | DICOM schedule query timeout or database unavailable | warning |
EventNetMsgSenderNetworkFault | Failure to connect NetMsgSender | NetMsgSender connection failure | warning |
EventNetMsgSenderNetworkFault | NetMsgSender communication failure detected | Network connectivity loss or NetMsgSender service crash | error |
EventPdmNotRegisteredOrConnectedFault | PDM (Patient Data Module) not registered with system or communication disconnected | PDM power-off, network disconnection, or registration loss | error |
EventStudyPullerError64Fault | StudyPuller service encountered error 64 during operation | Study pull/synchronization operation failure | error |
Initializing CO2 Device | CO2 device initialization in progress | Device startup or reconnection after power cycle | info |
Pump life has been exceeded | Pump has reached manufacturer end-of-life after extended operation | Cumulative pump operation hours exceeded rated lifespan | critical |
Pump Off | CO2 sampling pump is not operating | Pump button toggled off or pump failure | warning |
Sample Line Disconnected | No sidestream sampling set connected to LoFlo Capnostat | Sampling adapter not attached or became disconnected during operation | error |
Sensor Faulty | Sensor hardware failure detected: source current failure, EEPROM checksum error, or internal hardware error | Capnostat source current fails, EE-PROM checksum mismatch, or hardware error condition | error |
Sensor Over Temp | Sensor temperature exceeds 40°C operating limit | Sensor exposed to extreme heat or internal temperature sensor malfunction | warning |
Sensor Warm Up | Sensor temperature or source current stabilizing during startup | Power-up of sensor or device after power-down | info |
The Cath Amplifier on this system does not support Invasive | PDM lacks invasive blood pressure measurement capability | System opens Mac-Lab study and PDM does not have IBP option installed | error |
Zero In Progress | Capnostat zero calibration procedure currently running | User initiated zero command or automatic zero sequence started | info |
Zero Required | Sensor requires zeroing to recalibrate measurement range | Required zero condition or zero error detected during operation | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI