GE Healthcare CapnoFlex LF: troubleshooting and error codes
The ServiceForge knowledge base holds 9 documented symptoms for the GE Healthcare CapnoFlex LF, with 19 ranked causes, 20 diagnostic checks with test points and expected values, and 7 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Display shows CHECK ADAPTER or ADAPTER CAL message patient safety
Module detected failure in CO2 circuit or internal board and displays adapter error
Ranked causes
- 1. Cannula or sample line not inserted or not properly installed (common)
- 2. Sample line fouled or needs zero calibration (common)
- 3. Defective or contaminated cannula/sample line (occasional)
- 4. CO2 module board failure (occasional)
Verify cannula or sample line is inserted into module inlet
- Expected
- Line fully inserted
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Display shows CAL SENSOR TO ZERO CELL or CALIBRATE SAMPLE LINE message patient safety
Module requires zero point calibration for accurate CO2 measurement
Ranked causes
- 1. Zero point calibration required for normal operation (common)
- 2. Sample line not properly installed (common)
- 3. Defective sample line (occasional)
Perform zero cell calibration per module procedure
- Expected
- Message clears after calibration
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Display shows WARNING UP message patient safety
Module is in warm-up phase or warning state for warm-up condition
Ranked causes
- 1. Module warm-up cycle in progress (common)
- 2. Cannula or sample line not properly installed (common)
- 3. Defective cannula or sample line (occasional)
Wait 1 minute for module warm-up to complete
- Expected
- Message clears after warm-up
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Display shows CAL ERROR SENSOR message patient safety
Sensor calibration failed or sensor has an electrical or functional error
Ranked causes
- 1. Defective or contaminated cannula/sample line (common)
- 2. CO2 module board or sensor failure (occasional)
Replace cannula/sample line with new one
- Expected
- Error clears with new line
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Display shows CANNOT CALIBRATE message patient safety
Module calibration cannot proceed because a breath is detected or timing is wrong
Ranked causes
- 1. Module detected breath signal during calibration attempt (common)
- 2. Calibration timing issue or sensor failure (occasional)
Wait 2 minutes for any breath signal to clear before retrying
- Expected
- Breath signal cleared
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Module test setup configuration
Proper connection of CapnoFlex LF module to patient monitor before calibration verification
Ranked causes
- 1. Module must be correctly connected to monitor type (common)
- 2. Dual CO2 module pump must be disabled for sidestream mode (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CO2 module verification
Yearly calibration verification of CapnoFlex LF CO2 module to ensure accurate CO2 measurement
Ranked causes
- 1. Calibration verification required annually (common)
Verify equipment availability
- Test point
- Equipment kit
- Expected
- Gas scavenging adapter pn 9504-016, CO2 module calibration kit pn 405910-001, water beaker, flow meter 0-100 ml/min
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display shows SERVICE CO2 SENSOR message patient safety
CO2 sensor requires service or has failed and needs replacement
Ranked causes
- 1. CO2 sensor has reached end of service life or failed (common)
Return module to service for sensor replacement
- Expected
- Module returned with new sensor installed
Display shows SERVICE CO2 TEMP message patient safety
Temperature sensor for CO2 measurement has failed
Ranked causes
- 1. Temperature sensor failure requiring replacement (common)
Return module to service for temperature sensor replacement
- Expected
- Module serviced with new sensor
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
CAL ERROR SENSOR | Sensor calibration failed or sensor has error | Detected during calibration or after line replacement | error |
CAL SENSOR TO ZERO CELL or CALIBRATE SAMPLE LINE | Module requires zero point calibration to establish baseline | Appears when sensor zero point needs to be set or reset | warning |
CANNOT CALIBRATE | Calibration cannot proceed due to detected breath or timing issue | Appears when breath is detected during calibration attempt | warning |
CHECK ADAPTER ADAPTER CAL | CO2 measurement failure in module board or flow circuit | Detected during inlet occlusion test or normal operation | error |
SERVICE CO2 SENSOR | CO2 sensor has reached end of life or requires replacement | Appears when sensor failure is detected | error |
SERVICE CO2 TEMP | Temperature sensor failure | Detected when temperature measurement circuit fails | error |
WARNING UP | Module is in warm-up phase or warm-up condition detected | Appears during initial power-up or after service | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI