GE Healthcare C ARM: troubleshooting and error codes

The ServiceForge knowledge base holds 118 documented symptoms for the GE Healthcare C ARM, with 175 ranked causes, 118 diagnostic checks with test points and expected values, and 84 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Boot sequence stalls or fails during initialization

C-arm control panel 20-character display shows squares or arrows pattern that stops, indicating failure during boot process before reaching completion

Ranked causes

  1. 1. Hardware problem preventing IDE hard drive boot (occasional)
  2. 2. Operating system or software load failure (common)
  3. 3. ARCNet connection failure (common)
  4. 4. CMOS data corruption or battery failure (occasional)
  5. 5. Flash memory or file corruption (occasional)
  6. 6. Stator circuit failure (rare)
  7. 7. Precharge circuit failure (rare)
  8. 8. Filament drive circuit failure (rare)
  9. 9. Battery voltage low or battery failure (occasional)
  10. 10. Workstation boot failure or ARCNet communication loss (occasional)
First check

Observe boot pattern on control panel display

Test point
C-Arm control panel 20-character display
Expected
Five, ten, 15, 20 squares followed by arrow progression

20 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Film Mode Performance Check failed radiation

C-Arm system fails Film Mode diagnostic test; specific failure mode must be identified from fault chart

Ranked causes

  1. 1. No anode whirring sound (unknown)
  2. 2. No X-RAY ON indicator light (unknown)
  3. 3. No beeps when footswitch pressed (unknown)
  4. 4. Exposure unsuitable (unknown)
  5. 5. Mechanical movement fault (unknown)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

ARCNet communication failure between C-Arm and Workstation

Data pathway disruption or signal degradation on ARCNet bus affecting system control

Ranked causes

  1. 1. Fault on Fluoro Functions PCB ARCNet node (unknown)
  2. 2. Fault on X-Ray Controller PCB ARCNet node (unknown)
  3. 3. Fault on System Interface PCB Workstation node (unknown)
  4. 4. Interconnect cable defect or poor connection (unknown)
First check

Observe differential ARCNet signal using digital oscilloscope

Test point
ARCNet bus lines at PCB node
Expected
Clean differential signal per ARCNet specification
Tools
Digital oscilloscope

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Hexadecimal display boot code indicates fatal error or memory fault

2-digit hexadecimal display on Fluoro Functions PCB shows FE (fatal error), 0D/FF (out of memory), E0/E4 (memory/stack warning), or E0,XX,FF (ARCNET transmit pipe full)

Ranked causes

  1. 1. Fatal error: stack overflow, no memory, or unhandled interrupt (rare)
  2. 2. System halted: out of memory condition (rare)
  3. 3. Stack memory dangerously low (occasional)
  4. 4. ARCNET transmit pipe temporarily full (occasional)
First check

Observe hex display codes during bootup sequence

Test point
Fluoro Functions PCB 2-digit hexadecimal display
Expected
Sequential codes 01-0A during initialization; display shows 01=Flash check, 02=Processor init, 03-05=Memory clear, 06=App init start, 07=Arch util init, 08=Network init, 09=Timers init, 0A=Tasks init

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

High Voltage Supply Regulator fault or shutdown high voltage

LEDs on HV Supply Regulator PCB show fault condition (FAULT LED lit, SHUTDOWN active, LOOP A failure, MA HIGH condition, or KV ramp abnormality)

Ranked causes

  1. 1. FAULT LATCH activated (rare)
  2. 2. Ground integrity failure (rare)
  3. 3. KV regulation loop failure (occasional)
  4. 4. KV ramp timing abnormality (occasional)
First check

Verify filament drive status on HV Supply Regulator PCB

Test point
HV Supply Regulator PCB: FILAMENT DRIVE OK LED
Expected
FILAMENT DRIVE OK LED ON under normal operation

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AD CHANNEL # FAIL message displayed

System displays 'AD CHANNEL # FAIL' on C-Arm Control Panel indicating failure in specific analog-to-digital converter channel on X-Ray Controller PCB.

Ranked causes

  1. 1. A/D converter channel component failure (unknown)
  2. 2. A/D reference voltage U36 failure (unknown)
  3. 3. A/D data buffer U27 failure (unknown)
First check

Probe A/D channels on X-Ray Controller PCB

Test point
A/D converter input channels
Expected
Within valid signal range

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Boot sequence failure

Workstation fails to boot or produces beeping sounds, interconnect cable not recognized, or green Battery light does not illuminate.

Ranked causes

  1. 1. Power supply or AC wall voltage issue (common)
  2. 2. Interconnect cable not properly connected (common)
  3. 3. C-Arm battery or internal electronics fault (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Data error

System displays DATA ERROR, indicating corruption in program memory, data storage, or system configuration

Ranked causes

  1. 1. Program area memory corruption (rare)
  2. 2. Data area memory corruption (rare)
  3. 3. Flash memory corrupted (rare)
First check

Identify which memory area is corrupted

Test point
Error data field 1 and 2
Expected
Specific corruption type indicated (0-13)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator Driver PCB fault during X-ray exposure high voltage

LEDs on Generator Driver PCB not lighting as expected during X-ray exposure or showing fault conditions (overload, saturation, timing faults)

Ranked causes

  1. 1. HV generator drive signal timing fault (rare)
  2. 2. Filament drive signal failure (occasional)
  3. 3. X-ray enable or shutdown circuit failure (occasional)
First check

Check Generator Driver PCB filament drive status before X-ray

Test point
Filament Driver PCB LED DS1
Expected
ON during 16-18 arrow bootup stage, OFF after bootup, ON at power-off then decay

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Image out of focus

Image displayed on monitor is not in focus or appears blurry

Ranked causes

  1. 1. Image intensifier power supply problem (unknown)
  2. 2. Image intensifier tube problems (unknown)
  3. 3. CCD camera problem (unknown)
First check

Verify II and camera can be focused

Test point
II and camera focus control
Expected
Image comes into focus with adjustment

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Interlock failure patient safety

System displays INTERLOCK FAILURE error, indicating the safety interlock system is not functioning properly

Ranked causes

  1. 1. Interlock solenoid failed to close (occasional)
  2. 2. Interlock stuck in closed position (occasional)
  3. 3. Interlock IO task missing from software (rare)
First check

Identify interlock failure type

Test point
Error data field 1
Expected
0=failed to close; 1=already closed; 3=task missing

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Lift switch stuck caution

C-Arm lift switch held ON for 30+ seconds; all generator functions inhibited until key pressed

Ranked causes

  1. 1. C-Arm lift switch held in ON position (occasional)
  2. 2. Object pressing on lift switch (occasional)
  3. 3. Faulty lift switch (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No click heard when Interconnect cable unplugged from C-Arm

Audible relay click not heard when disconnecting cable from C-arm unit

Ranked causes

  1. 1. Workstation not plugged into wall receptacle (common)
  2. 2. Receptacle not powered - fuse blown or circuit breaker tripped (common)
  3. 3. Incorrect voltage Workstation connected to wrong outlet (occasional)
First check

Verify wall outlet is powered

Test point
Wall receptacle where Workstation plugged in
Expected
AC voltage present (100-125V or 200-250V as rated)
Tools
Multimeter or AC outlet tester

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No click heard when Workstation plugged into AC power high voltage

Audible relay click not heard when connecting Workstation to AC power

Ranked causes

  1. 1. Voltage mismatch - 100-125V unit plugged into 200-250V outlet (common)
  2. 2. Circuit breaker CB1 or CB2 tripped (common)
  3. 3. Outlet power off or facility breaker tripped (common)
First check

Verify outlet has AC power

Test point
Wall receptacle
Expected
AC voltage present at outlet
Tools
Multimeter or outlet tester

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Workstation totally inoperative high voltage

No power, lights, or response; unit does not function at all

Ranked causes

  1. 1. Power and interconnect cables not connected or faulty (common)
  2. 2. Voltage mismatch - wrong outlet voltage (common)
  3. 3. Isolation Transformer strapped incorrectly (occasional)
First check

Verify power cord connection and outlet power

Test point
Power cord connector and wall outlet
Expected
Cord fully seated; outlet has AC voltage
Tools
Multimeter, visual inspection

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

X-ray image not appearing on monitor radiation

After pressing X-ray on switch, no digitized fluoro image appears on left monitor or image disappears immediately after exposure.

Ranked causes

  1. 1. X-ray generation not initiating (common)
  2. 2. Image intensifier or detector fault (occasional)
  3. 3. Image acquisition or display system malfunction (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

X-RAY SWITCH SECURITY ERROR displayed

System displays error when mismatch occurs between Security line on footswitch or handswitch and Security line on matching connector.

Ranked causes

  1. 1. Footswitch incorrect wiring (unknown)
  2. 2. Handswitch incorrect wiring (unknown)
  3. 3. X-Ray Controller PCB fault (unknown)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

X-RAY SWITCH STUCK message displayed

System displays error and shuts down when X-ray control (Left or Right) is detected in ON state during startup, indicating switch being held down or internal fault.

Ranked causes

  1. 1. Footswitch or handswitch held down during startup (occasional)
  2. 2. Footswitch or handswitch requires service (occasional)
  3. 3. Internal fault in X-ray control circuit (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

All indicators did not illuminate; error message appears

System front panel indicator squares and arrows fail to light; error message displayed on screen

Ranked causes

  1. 1. Interlock or control logic failure (unknown)
  2. 2. Boot or startup sequence error (unknown)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery Charger light did not turn on

Battery charger indicator light remains off; no charging indicator displayed

Ranked causes

  1. 1. Interconnect cable problem (common)
  2. 2. Connector loose or corroded (common)
First check

Visually inspect Interconnect cable for damage

Test point
Cable entire length and connectors
Expected
No visible damage, kinks, cuts, or corrosion
Tools
Visual inspection

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Filament driver overcurrent or saturation fault high voltage

Overload Fault or Saturation Fault LED on Generator Driver PCB lit, indicating filament drive circuit malfunction or excessive current draw

Ranked causes

  1. 1. Filament drive circuit overload (occasional)
  2. 2. Filament drive transistor saturation or control circuit fault (occasional)
First check

Observe Overload Fault and Saturation Fault LEDs during normal operation

Test point
Generator Driver PCB: Overload Fault LED, Saturation Fault LED
Expected
Both LEDs OFF during normal X-ray exposure and idle

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Housing overheating caution

System displays HOUSING IS HOT, HOUSING IS WARM, or HOUSING OVERHEATED error, indicating internal temperature approaching or exceeding safe limits

Ranked causes

  1. 1. Internal heat accumulation due to sustained X-ray exposure (common)
  2. 2. Cooling fan malfunction (occasional)
First check

Monitor housing temperature percentage

Test point
Control panel display
Expected
Percentage of capacity shown

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Irregularities when field size changes

Image control anomalies occur during field size adjustment in fluoro mode

Ranked causes

  1. 1. Control Panel Processor PCB failure (common)
  2. 2. Fluoro Functions PCB failure (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Key switch in standby position caution

X-ray disable keyswitch turned on (vertical); prevents vertical column movement and X-ray generation; X-rays and lift disabled

Ranked causes

  1. 1. X-ray disable keyswitch in ON position (common)
  2. 2. Keyswitch or associated hardware fault (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No image on monitor

Monitor displays no image output from the C-arm system

Ranked causes

  1. 2. Bad CCD camera (common)
  2. 3. Bad monitor (occasional)
First check

Verify kV tracking to nominal level

Test point
kV output
Expected
120

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Video abnormalities, artifacts, or drop-outs

Video signal exhibits artifacts, abnormalities, or intermittent drop-outs in image

Ranked causes

  1. 1. CCD Camera failure (common)
  2. 2. Fluoro Functions PCB failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

WARNING LOW MA displayed patient safety

System displays 'WARNING LOW MA - PRESS ANY KEY' message on C-Arm Control Panel when actual mA output is lower than display shows and not within tolerance. Results in poor image quality.

Ranked causes

  1. 1. Filament drive circuitry fault (unknown)
  2. 2. mA Sense circuitry fault (unknown)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

24 hour recharge required caution

System displays 24 HOUR RECHARGE REQUIRED – X-RAYS DISABLED, indicating battery charge is critically low and system requires extended charging

Ranked causes

  1. 1. Battery discharged below safe operating level (common)
First check

Monitor battery charge status during extended charging

Test point
Battery voltage and charge indicator
Expected
Battery voltage gradually rising during 24-hour period

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

4 hour warm up required caution

System displays 4 HOUR WARM UP REQUIRED on control panel, indicating stator/rotor requires warm-up period before high-power operation

Ranked causes

  1. 1. System powered on after extended shutdown requiring thermal stabilization (common)
First check

Monitor warm-up timer

Test point
Control panel display
Expected
Timer counts down from 4 hours to zero

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

AD channel measurement failure caution

System reports AD CHANNEL FAIL error on control panel display, indicating analog-to-digital conversion failure on specified channel

Ranked causes

  1. 1. Analog-to-digital converter malfunction on specified channel (common)
First check

Verify channel number from error message

Test point
Control panel display
Expected
Channel number displayed in error message

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Anode overheating radiation

System displays ANODE IS HOT or ANODE IS WARM error on control panel, indicating X-ray tube anode temperature approaching or exceeding safe limits

Ranked causes

  1. 1. Anode thermal capacity exceeded or approaching limit (common)
First check

Monitor anode heat capacity percentage

Test point
Control panel display
Expected
Percentage of capacity below safe threshold

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ARCNET communications failure

Control panel button pushes do not cause intended changes on workstation; ARCNET network communications failure

Ranked causes

  1. 1. Fluoro Functions PCB failure (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery charge below minimum caution

System displays BATTERY CHARGE - ##% error on control panel, indicating battery state of charge is insufficient for operation

Ranked causes

  1. 1. Battery discharged below minimum operating level (common)
First check

Monitor battery charge percentage

Test point
Control panel display or error log
Expected
Percentage of capacity shown

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Boot sequence takes longer than normal

System takes considerably longer than 80-120 seconds to complete bootup sequence

Ranked causes

  1. 1. Systems with WJ98 or later software require 128 Meg RAM instead of 64 Meg (common)
First check

Verify system software version and required RAM configuration

Test point
SBC memory configuration
Expected
128 Meg RAM for WJ98+ software, 64 Meg for earlier versions

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bright pixels present in image

Camera cooler failure causes bright pixels in image due to offset correction failure

Ranked causes

  1. 1. Camera Cooler failure (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Film mode exposure radiation

Verify film mode X-ray exposure at 60 kVp @ 2.5 mAs produces acceptable image on developed film cassette

Ranked causes

  1. 1. System calibration verification required (unknown)
First check

Listen for anode acceleration whirring sound before X-ray indicator

Test point
X-ray switch indicator and audio output
Expected
Whirring sound approximately 2 seconds before indicator illuminates
Tools
Footswitch

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Mechanical brake operation

Verify Wig-Wag and Flip Flop brake engagement prevents unwanted movement

Ranked causes

  1. 1. Brake system calibration verification (unknown)
First check

Test Wig-Wag brake prevents movement

Test point
Wig-Wag arm with brake engaged
Expected
No Wig-Wag movement when brake is engaged

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Camera iris mechanism failure caution

System displays CAMERA IRIS FAILURE error, indicating iris control mechanism is malfunctioning

Ranked causes

  1. 1. Iris control mechanism stuck or damaged (occasional)
First check

Determine iris position state from error data

Test point
Error message data field 1
Expected
0 = open state; 1 = closed state

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

CCD camera rotation binds or exhibits jerky motion

Camera rotation shows binding or jerky motion during operation

Ranked causes

  1. 1. Fluoro Functions PCB failure (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Charger failure caution

System displays CHARGER FAILED or CHARGER FAILED with shutdown, indicating battery charger circuit malfunction

Ranked causes

  1. 1. Battery charging circuit failure (common)
First check

Measure battery voltage via inverter

Test point
Inverter B A/D
Expected
Within charging specification

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Collimation failure radiation

Collimator jaws fail to open/close or position incorrectly; X-ray field size uncontrollable

Ranked causes

  1. 1. Collimator motor, solenoid, or control circuit failure (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Collimator calibration required caution

System displays COLLIMATOR CAL REQUIRED error on control panel, indicating collimator leaves or iris needs recalibration

Ranked causes

  1. 1. Collimator leaf or iris position drift out of calibration (occasional)
First check

Identify which component requires calibration

Test point
Error data field 1
Expected
0 = Leaves; 1 = Iris; 2 = Leaves and Iris

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Collimator iris potentiometer error caution

System displays COLL IRIS POTENTIOMETER ERROR, indicating the iris position feedback sensor is malfunctioning

Ranked causes

  1. 1. Iris potentiometer sensor failure or connection issue (common)
First check

Check potentiometer output reading

Test point
Pot reading from error message
Expected
Reading outside normal range or erratic

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Collimator iris too large caution

System displays COLL IRIS TOO LARGE error, indicating iris opening exceeds maximum safe aperture

Ranked causes

  1. 1. Iris potentiometer reading indicates opening beyond specification (occasional)
First check

Check iris potentiometer reading

Test point
Pot reading from error
Expected
Reading above maximum aperture threshold

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Collimator iris unstable caution

System displays COLL IRIS UNSTABLE error, indicating collimator iris position is drifting or fluctuating

Ranked causes

  1. 1. Collimator iris motor or potentiometer malfunction (occasional)
First check

Check iris potentiometer reading

Test point
Pot reading from error data
Expected
Stable reading within specification

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Collimator movements are jerky or sticky

Collimator exhibits sticky or jerky motion during operation

Ranked causes

  1. 1. Collimator lubrication insufficient (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Collimator or camera motor not responding

Camera iris, optics, collimator leaves or iris functions fail to operate

Ranked causes

  1. 1. Fluoro Functions PCB fault affecting motor control (common)
First check

Measure +24VDC supply to II power supply at J7

Test point
J7 pins 15-8
Expected
+24VDC

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Collimator stuck caution

System displays COLLIMATOR STUCK error, indicating collimator leaves cannot move to commanded position

Ranked causes

  1. 1. Collimator leaf mechanical jam or motor failure (common)
First check

Monitor potentiometer reading

Test point
Pot reading during leaf movement attempt
Expected
Reading changes when motor is commanded

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Communication failure

System displays COMMUNICATION FAILED error, indicating loss of communications between system components

Ranked causes

  1. 1. Communication bus error or component disconnection (occasional)
First check

Verify system startup completes

Test point
Control panel displays
Expected
System initializes without error

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Control panel error

System displays CONTROL PANEL ERROR on display, indicating control panel microprocessor or interface failure

Ranked causes

  1. 1. Control panel microprocessor or communications interface failure (rare)
First check

Verify control panel responsiveness

Test point
Control panel buttons and display
Expected
Display shows message and accepts input

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Control panel key stuck caution

System senses key closure during boot-up; possible stuck key detected

Ranked causes

  1. 1. Control panel key stuck or object pressing key (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Dark images with low pulse rate

Pulse rates less than 4 pps produce dark images due to failure of integrating data over multiple frame times; auto brightness stabilization incorrect

Ranked causes

  1. 1. Camera control failure (Manchester Encoder/Decoder) (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error message displayed during bootup

Error message appears on C-Arm Control Panel after boot sequence completion

Ranked causes

  1. 1. Hardware problem preventing IDE hard drive software load (occasional)
First check

Observe boot display pattern on 20-character control panel

Test point
C-Arm Control Panel display
Expected
Five squares, ten squares, 15 squares, 20 squares pattern, then arrows

Fast stop activated caution

System displays FAST STOP ACTIVATED and powers down, indicating emergency shutdown was triggered

Ranked causes

  1. 1. Emergency stop button pressed or safety condition detected (unknown)
First check

Check high voltage register

Test point
High Voltage Input Register
Expected
HV present at time of fast stop

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Field size selection failure

Image intensifier field size control unresponsive or produces incorrect zoom levels

Ranked causes

  1. 1. Image intensifier tube or power supply control malfunction (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Filament calibration required caution

System displays FILAMENT CAL REQUIRED error on control panel, indicating filament characteristics need recalibration

Ranked causes

  1. 1. Filament parameters drift out of calibration (occasional)
First check

Run filament calibration procedure

Test point
Generator control system
Expected
Calibration completes without error

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Filament regulator failure high voltage

System displays FILAMENT REGULATOR FAILED error, indicating filament voltage regulation circuit is malfunctioning

Ranked causes

  1. 1. Filament voltage regulation circuit failure (common)
First check

Measure filament voltage

Test point
Filament supply output
Expected
Measured filament voltage from error data

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Filament select error caution

System displays FILAMENT SELECT ERROR, indicating incorrect filament is selected or filament selection circuit failure

Ranked causes

  1. 1. Filament selector switch or control circuit failure (occasional)
First check

Check which filament is selected

Test point
Error data field 1
Expected
0 = Small filament; 1 = Large filament

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fluoro Functions PCB microcontroller not running

LED segment 6 on DS1/DS2 does not blink at 1 Hz after bootup, indicating microcontroller startup failure or crash

Ranked causes

  1. 1. Microcontroller initialization failure (unknown)
First check

Observe LED segment 6 on DS1 blinking after power-up

Test point
DS1 LED segment 6
Expected
Blink at 1 Hz during bootup and after

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fluoro mode performance check failed patient safety

System fluoro mode performance test did not pass; image quality, exposure, or mode parameters out of spec

Ranked causes

  1. 1. One or more fluoro subsystem parameter out of specification (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Footswitch security error
  • Generator filament calibration running
  • High kV warning
  • High kVp warning
  • High mA warning
  • High voltage generator error
  • High voltage generator unexpected on
  • High voltage register failure at startup
  • High-level fluoroscopy overtime
  • HV register failure
  • Image brightness and contrast not compensating with added filtration
  • Image quality unacceptable in fluoroscopy
  • Image rotation or reversal in fluoroscopy
  • Key stuck
  • Key switch in standby
  • KV turned on in error
  • Low dose ABS table loading
  • Low kV warning
  • Low kVp warning
  • Low mA warning
  • mA sense on without X-ray switch
  • mA sensor failure
  • mA turned on in error
  • No click heard when Interconnect cable plugged into C-Arm
  • No fluoroscopy image output
  • Overload fault
  • Overvoltage fault
  • Performance check: Image quality unacceptable
  • Performance check: No Fluoro image
  • Power Supply PS2 output voltage incorrect
  • Pre-charge voltage error
  • Precharge circuit timeout
  • Precharge voltage error
  • Room door open
  • Saturation fault
  • Standard ABS table loading
  • Stator not on
  • Stator standby wait
  • System processing in progress
  • System shuts down unexpectedly without error message
  • Temperature sensor failure
  • Vascular mode control failure
  • Verify ARCNet signal integrity at nodes
  • Workstation beeps and does not boot when switched on
  • X-ray control released early in film mode
  • X-ray Controller PCB initialization failure
  • X-ray overtime
  • X-RAY OVERTIME message displayed
  • X-ray room door open
  • X-ray switch security failure
  • X-ray switch stuck
  • X-ray tube anode not powered
  • X-ray tube current on without control
  • X-ray tube housing at elevated temperature
  • X-ray tube housing at warm temperature
  • X-ray tube housing overheated
  • X-RAY TUBE TEMPERATURE SENSOR FAILURE displayed
  • X-rays disabled

58 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
24 HOUR RECHARGE REQUIREDBattery charge critically low; system requires 24-hour recharge cycleBattery voltage below minimum safe operating levelerror
24-HOUR RECHARGE REQUIRED - X RAYS DISABLEDBattery charge depleted after long-term storageC-Arm batteries depleted; typically after six months or more storagewarning
4 HOUR WARM UP REQUIREDStator and tube require 4-hour warm-up before full-power operationSystem powered on after extended shutdownwarning
4-HOUR WARM-UP REQUIRED - PRESS ANY KEYSystem inactive for more than 60 days; ion removal from Image Intensifier requiredSystem has been inactive for more than 60 dayswarning
AD CHANNEL # FAILAnalog-to-digital converter channel failureA/D channel functional test failure at startup or during operationerror
AD CHANNEL %s FAILAnalog-to-digital conversion failure on specified channelA/D converter unable to process signal from specified channelerror
AD CHANNEL %s FAIL – POWER OFF – WAIT 5 SECONDSAnalog to digital converter channel failureAD converter channel malfunction detectederror
AD CHANNEL %s FAIL – POWER OFF. WAIT 5 SECONDSAnalog-to-digital channel failure at startupA/D channel fails during system start-upcritical
ANODE IS HOTX-ray tube anode is at high temperature near safe limitAnode heat accumulation from sustained X-ray exposurewarning
ANODE IS HOTAnode heat unit capacity criticalAnode heat units reach critical level approaching maximum capacitycritical
ANODE IS HOT - XX%X-ray tube anode at 80% or greater of rated capacityX-ray tube reaches 80% (+2%) of rated thermal capacityerror
ANODE IS WARMX-ray tube anode is at elevated temperature but safeAnode heat accumulation approaching high thresholdwarning
ANODE IS WARMAnode heat unit capacity elevatedAnode heat units elevated but not at critical levelwarning
ANODE IS WARM - XX%X-ray tube anode at 70% of rated capacity; warning onlyX-ray tube reaches 70% (+2%) of rated thermal capacitywarning
BATTERY CHARGE - ##%Battery state of charge below minimum safe levelBattery voltage drops below minimum operating thresholderror
BATTERY CHARGE - XX%Battery charge between 70% and 40% of required levelC-Arm battery charge drops to 70%-40% rangewarning
BATTERY CHARGE - XX% /WAITBattery charge below 40% of required level; limited functionalityC-Arm battery charge drops below 40% of required chargewarning
CAMERA IRIS FAILURECamera iris mechanism failureCamera iris motor or sensor malfunctionwarning
CHARGER FAILEDBattery charging circuit is malfunctioningCharger unable to deliver proper charge voltage or currenterror
CHARGER FAILEDBattery charger failure detected during bootupBattery charger malfunction at system startuperror
COLLIMATOR CAL REQUIREDCollimator leaves or iris require position recalibrationCollimator position feedback drift out of toleranceerror
COLLIMATOR CAL REQUIRED - PRESS ANY KEYCollimator iris and leaf position values do not match stored reference valuesCollimator position values drift from stored reference values in memorycritical
COLLIMATOR STUCKCollimator iris motor cannot open or close irisCollimator iris motor mechanical failure or jamerror
COLL IRIS POTENTIOMETER ERRORIris position feedback potentiometer is malfunctioningPotentiometer output outside expected range or erraticerror
COLL IRIS POTENTIOMETER ERROR - PRESS ANY KEYCollimator iris position potentiometer failure or disconnectionPosition potentiometer on collimator not working or disconnectedcritical
COLL IRIS TOO LARGECollimator iris aperture larger than displayed value on control panelCollimator iris aperture size exceeds indicated valuewarning
COLL IRIS UNSTABLECollimator iris motor cannot maintain position toleranceCollimator iris motor position instability, typically jittery operationwarning
COMMUNICATION FAILEDWorkstation and C-Arm communication loss for more than one secondNo communication between Workstation and C-Arm exceeds one-second timeout; Watchdog circuit triggerscritical
COMMUNICATION FAILEDCommunication between system components lostData transmission error on system communication buserror
CONTROL PANEL ERRORSystem cannot communicate with C-Arm Control PanelsLoss of communication between generator and control panelscritical
CONTROL PANEL ERRORControl panel microprocessor or interface failurePanel processor malfunction or bus errorerror
CONTROL PANEL KEY STUCK - RELEASE, THEN PRESS ANY KEY TO CONC-Arm Control Panel key stuck in pressed position at bootupControl panel key mechanically stuck or electrically held at bootupcritical
DATA ERRORSystem detects invalid data in memory or processingInvalid ABS table write/read, corrupted non-volatile memory, or software malfunctioncritical
DATA ERRORSystem memory or flash storage contains corrupted dataMemory CRC check failure or data range errorerror
FAST STOP ACTIVATEDEmergency shutdown was triggeredEmergency stop button pressed or safety condition detectedwarning
FAST STOP ACTIVATED – POWER OFF. WAIT 5 SECONDS.Interlocks opened; fast stop switch activatedFAST STOP switch pressed on front of C-Arm Control Panel or interlock circuit openederror
FILAMENT CAL REQUIREDFilament characteristics require recalibrationFilament parameters drift beyond acceptable toleranceerror
FILAMENT CAL REQUIRED - PRESS ANY KEYFilament calibration not performed since system startup or software updateSoftware detects missing or expired filament calibration at startuperror
FILAMENT REGULATOR FAILEDFilament voltage regulation circuit has failedFilament regulator unable to maintain set voltageerror
FILAMENT REGULATOR FAILURE - PRESS ANY KEYX-ray tube filament current out of toleranceFilament current measurement exceeds tolerance limitserror
FILAMENT SELECT ERRORIncorrect filament selected or filament selector failureMismatch between commanded and actual filamenterror
FILAMENT SELECT ERROR - PRESS ANY KEYSelected filament size in software does not match filament in PIO hardwareFilament selection mismatch between software and hardware configurationerror
FILAMENT STANDBYFilament in standby modeFilament heater energized but X-ray production not activeinfo
FLUORO ALARMFluoroscopy alarm condition triggeredFluoroscopy mode parameter or time limit exceededwarning
HLF OVERTIMEHigh-level fluoroscopy overtime condition detectedProlonged high-level fluoroscopy operationwarning
HLF OVERTIMEGenerator operating in overtime modeGenerator heat limit exceeded during operationwarning
HLF OVERTIME (digital cine)High-level fluoroscopy overtime during cine acquisitionProlonged high-speed recording mode exceeds overtime limitwarning
HOUSING IS HOTGenerator internal housing at elevated temperatureInternal component temperature high due to X-ray exposurewarning
HOUSING OVERHEATEDGenerator internal housing at or above maximum safe temperatureInternal temperature exceeds maximum safe thresholderror
HOUSING WARMGenerator internal housing at elevated but safe temperatureInternal temperature high but below maximum limitwarning
HV GENERATOR ONHigh voltage power supply is active when it should be offUnintended HV output or control failureerror
HV REGISTER FAILHigh voltage monitoring register has failedHV register cannot be read or writtenerror
INTERLOCK FAILURESafety interlock system is not functioning properlyInterlock solenoid, latch, or sensor malfunctioncritical
KEY STUCKA physical control button or key is mechanically stuckButton does not return to normal position after depressionerror
KEY SWITCH IN STANDBYKey switch is in standby position disabling system operationKey switch manually set to standby by operatorinfo
KV ON IN ERRORKilovoltage became active unexpectedlyUnintended kV supply activationcritical
LIFT SWITCH STUCKVertical positioning switch is mechanically stuckLift control switch cannot change stateerror
MA ON IN ERRORTube current became active when it should be offUnintended tube current activationcritical
MA SENSOR FAILTube current measurement sensor is malfunctioningmA sensor output out of range or erraticerror
OVERLOAD FAULTElectrical load or power consumption exceeds safe limitPower draw exceeds system capacityerror
OVERVOLTAGE FAULTHigh voltage supply exceeds maximum safe levelHV regulator failure allowing excessive outputcritical
PLEASE WAITSystem is performing initialization or processing operationSystem startup or internal operation in progressinfo
PRECHARGE CIRCUIT TIMEOUTHV power supply cannot charge to operating voltage within timeout periodPrecharge circuit failure or insufficient battery voltageerror
PRE-CHARGE VOLTAGE ERRORBattery inverter voltage insufficient for HV power supply prechargeInverter B battery voltage below precharge thresholderror
PRE-CHARGE VOLTAGE ERROR – POWER OFFPre-charge inverter voltage errorInverter pre-charge voltage out of specificationerror
RELEASED EARLYX-ray exposure terminated before expected completionOperator released exposure button earlyinfo
ROOM DOOR OPENProcedure room safety door is openRoom door sensor indicates open statewarning
SATURATION FAULTPower supply or transformer has reached magnetic saturationExcessive current through transformer coreerror
STATOR NOT ONX-ray tube rotor/stator motor is not runningStator motor failed to start or power lost to motorerror
STATOR STANDBYStator motor is in standby state and not rotatingSystem standby mode or motor disabled during initializationinfo
TEMP. SENSOR FAILTemperature measurement sensor is malfunctioningTemperature sensor output out of rangeerror
WARNING – HIGH kVMeasured kilovoltage exceeds desired setpointkV regulation failure or setpoint errorwarning
WARNING, HIGH MAMeasured tube current exceeds desired setpointmA regulation failure or setpoint errorwarning
WARNING – LOW kVMeasured kilovoltage is below desired setpointkV supply unable to achieve setpointwarning
WARNING, LOW MAMeasured tube current is below desired setpointmA supply unable to reach setpointwarning
WARNING LOW MAActual mA output lower than displayed value and not within tolerancemA output falls below expected range during operationwarning
X-RAY OVERTIMEFilm shot exposure time exceeded required durationExposure continues beyond required time limitwarning
XRAY OVERTIMECumulative daily X-ray exposure has exceeded safe limitTotal exposure time exceeds daily maximumerror
XRAYS DISABLEDX-ray output has been disabled by safety systemSafety interlock engaged or control system failurewarning
X-RAY SWITCH SECURITY ERRORMismatch between Security line on footswitch/handswitch and connectorSecurity line signal mismatch detectederror
X-RAY SWITCH SECURITY FAILUREX-ray enable switch security check has failedX-ray switch state does not match expected control signalerror
X-RAY SWITCH STUCKX-ray enable switch cannot change from current stateSwitch solenoid or mechanical latch unable to moveerror
X-RAY SWITCH STUCKX-ray control switch detected in ON state during system startupLeft or Right X-ray control switch held ON at startuperror
X-RAY TUBE TEMPERATURE SENSOR FAILUREX-ray tube temperature thermistor fails functional testing at startupThermistor functional test failure during system initializationcritical

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.