GE Healthcare C ARM: troubleshooting and error codes
The ServiceForge knowledge base holds 118 documented symptoms for the GE Healthcare C ARM, with 175 ranked causes, 118 diagnostic checks with test points and expected values, and 84 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Boot sequence stalls or fails during initialization
C-arm control panel 20-character display shows squares or arrows pattern that stops, indicating failure during boot process before reaching completion
Ranked causes
- 1. Hardware problem preventing IDE hard drive boot (occasional)
- 2. Operating system or software load failure (common)
- 3. ARCNet connection failure (common)
- 4. CMOS data corruption or battery failure (occasional)
- 5. Flash memory or file corruption (occasional)
- 6. Stator circuit failure (rare)
- 7. Precharge circuit failure (rare)
- 8. Filament drive circuit failure (rare)
- 9. Battery voltage low or battery failure (occasional)
- 10. Workstation boot failure or ARCNet communication loss (occasional)
Observe boot pattern on control panel display
- Test point
- C-Arm control panel 20-character display
- Expected
- Five, ten, 15, 20 squares followed by arrow progression
20 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Film Mode Performance Check failed radiation
C-Arm system fails Film Mode diagnostic test; specific failure mode must be identified from fault chart
Ranked causes
- 1. No anode whirring sound (unknown)
- 2. No X-RAY ON indicator light (unknown)
- 3. No beeps when footswitch pressed (unknown)
- 4. Exposure unsuitable (unknown)
- 5. Mechanical movement fault (unknown)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ARCNet communication failure between C-Arm and Workstation
Data pathway disruption or signal degradation on ARCNet bus affecting system control
Ranked causes
- 1. Fault on Fluoro Functions PCB ARCNet node (unknown)
- 2. Fault on X-Ray Controller PCB ARCNet node (unknown)
- 3. Fault on System Interface PCB Workstation node (unknown)
- 4. Interconnect cable defect or poor connection (unknown)
Observe differential ARCNet signal using digital oscilloscope
- Test point
- ARCNet bus lines at PCB node
- Expected
- Clean differential signal per ARCNet specification
- Tools
- Digital oscilloscope
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hexadecimal display boot code indicates fatal error or memory fault
2-digit hexadecimal display on Fluoro Functions PCB shows FE (fatal error), 0D/FF (out of memory), E0/E4 (memory/stack warning), or E0,XX,FF (ARCNET transmit pipe full)
Ranked causes
- 1. Fatal error: stack overflow, no memory, or unhandled interrupt (rare)
- 2. System halted: out of memory condition (rare)
- 3. Stack memory dangerously low (occasional)
- 4. ARCNET transmit pipe temporarily full (occasional)
Observe hex display codes during bootup sequence
- Test point
- Fluoro Functions PCB 2-digit hexadecimal display
- Expected
- Sequential codes 01-0A during initialization; display shows 01=Flash check, 02=Processor init, 03-05=Memory clear, 06=App init start, 07=Arch util init, 08=Network init, 09=Timers init, 0A=Tasks init
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
High Voltage Supply Regulator fault or shutdown high voltage
LEDs on HV Supply Regulator PCB show fault condition (FAULT LED lit, SHUTDOWN active, LOOP A failure, MA HIGH condition, or KV ramp abnormality)
Ranked causes
- 1. FAULT LATCH activated (rare)
- 2. Ground integrity failure (rare)
- 3. KV regulation loop failure (occasional)
- 4. KV ramp timing abnormality (occasional)
Verify filament drive status on HV Supply Regulator PCB
- Test point
- HV Supply Regulator PCB: FILAMENT DRIVE OK LED
- Expected
- FILAMENT DRIVE OK LED ON under normal operation
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AD CHANNEL # FAIL message displayed
System displays 'AD CHANNEL # FAIL' on C-Arm Control Panel indicating failure in specific analog-to-digital converter channel on X-Ray Controller PCB.
Ranked causes
- 1. A/D converter channel component failure (unknown)
- 2. A/D reference voltage U36 failure (unknown)
- 3. A/D data buffer U27 failure (unknown)
Probe A/D channels on X-Ray Controller PCB
- Test point
- A/D converter input channels
- Expected
- Within valid signal range
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Boot sequence failure
Workstation fails to boot or produces beeping sounds, interconnect cable not recognized, or green Battery light does not illuminate.
Ranked causes
- 1. Power supply or AC wall voltage issue (common)
- 2. Interconnect cable not properly connected (common)
- 3. C-Arm battery or internal electronics fault (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Data error
System displays DATA ERROR, indicating corruption in program memory, data storage, or system configuration
Ranked causes
- 1. Program area memory corruption (rare)
- 2. Data area memory corruption (rare)
- 3. Flash memory corrupted (rare)
Identify which memory area is corrupted
- Test point
- Error data field 1 and 2
- Expected
- Specific corruption type indicated (0-13)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator Driver PCB fault during X-ray exposure high voltage
LEDs on Generator Driver PCB not lighting as expected during X-ray exposure or showing fault conditions (overload, saturation, timing faults)
Ranked causes
- 1. HV generator drive signal timing fault (rare)
- 2. Filament drive signal failure (occasional)
- 3. X-ray enable or shutdown circuit failure (occasional)
Check Generator Driver PCB filament drive status before X-ray
- Test point
- Filament Driver PCB LED DS1
- Expected
- ON during 16-18 arrow bootup stage, OFF after bootup, ON at power-off then decay
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image out of focus
Image displayed on monitor is not in focus or appears blurry
Ranked causes
- 1. Image intensifier power supply problem (unknown)
- 2. Image intensifier tube problems (unknown)
- 3. CCD camera problem (unknown)
Verify II and camera can be focused
- Test point
- II and camera focus control
- Expected
- Image comes into focus with adjustment
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Interlock failure patient safety
System displays INTERLOCK FAILURE error, indicating the safety interlock system is not functioning properly
Ranked causes
- 1. Interlock solenoid failed to close (occasional)
- 2. Interlock stuck in closed position (occasional)
- 3. Interlock IO task missing from software (rare)
Identify interlock failure type
- Test point
- Error data field 1
- Expected
- 0=failed to close; 1=already closed; 3=task missing
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lift switch stuck caution
C-Arm lift switch held ON for 30+ seconds; all generator functions inhibited until key pressed
Ranked causes
- 1. C-Arm lift switch held in ON position (occasional)
- 2. Object pressing on lift switch (occasional)
- 3. Faulty lift switch (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No click heard when Interconnect cable unplugged from C-Arm
Audible relay click not heard when disconnecting cable from C-arm unit
Ranked causes
- 1. Workstation not plugged into wall receptacle (common)
- 2. Receptacle not powered - fuse blown or circuit breaker tripped (common)
- 3. Incorrect voltage Workstation connected to wrong outlet (occasional)
Verify wall outlet is powered
- Test point
- Wall receptacle where Workstation plugged in
- Expected
- AC voltage present (100-125V or 200-250V as rated)
- Tools
- Multimeter or AC outlet tester
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No click heard when Workstation plugged into AC power high voltage
Audible relay click not heard when connecting Workstation to AC power
Ranked causes
- 1. Voltage mismatch - 100-125V unit plugged into 200-250V outlet (common)
- 2. Circuit breaker CB1 or CB2 tripped (common)
- 3. Outlet power off or facility breaker tripped (common)
Verify outlet has AC power
- Test point
- Wall receptacle
- Expected
- AC voltage present at outlet
- Tools
- Multimeter or outlet tester
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Workstation totally inoperative high voltage
No power, lights, or response; unit does not function at all
Ranked causes
- 1. Power and interconnect cables not connected or faulty (common)
- 2. Voltage mismatch - wrong outlet voltage (common)
- 3. Isolation Transformer strapped incorrectly (occasional)
Verify power cord connection and outlet power
- Test point
- Power cord connector and wall outlet
- Expected
- Cord fully seated; outlet has AC voltage
- Tools
- Multimeter, visual inspection
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
X-ray image not appearing on monitor radiation
After pressing X-ray on switch, no digitized fluoro image appears on left monitor or image disappears immediately after exposure.
Ranked causes
- 1. X-ray generation not initiating (common)
- 2. Image intensifier or detector fault (occasional)
- 3. Image acquisition or display system malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
X-RAY SWITCH SECURITY ERROR displayed
System displays error when mismatch occurs between Security line on footswitch or handswitch and Security line on matching connector.
Ranked causes
- 1. Footswitch incorrect wiring (unknown)
- 2. Handswitch incorrect wiring (unknown)
- 3. X-Ray Controller PCB fault (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
X-RAY SWITCH STUCK message displayed
System displays error and shuts down when X-ray control (Left or Right) is detected in ON state during startup, indicating switch being held down or internal fault.
Ranked causes
- 1. Footswitch or handswitch held down during startup (occasional)
- 2. Footswitch or handswitch requires service (occasional)
- 3. Internal fault in X-ray control circuit (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
All indicators did not illuminate; error message appears
System front panel indicator squares and arrows fail to light; error message displayed on screen
Ranked causes
- 1. Interlock or control logic failure (unknown)
- 2. Boot or startup sequence error (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery Charger light did not turn on
Battery charger indicator light remains off; no charging indicator displayed
Ranked causes
- 1. Interconnect cable problem (common)
- 2. Connector loose or corroded (common)
Visually inspect Interconnect cable for damage
- Test point
- Cable entire length and connectors
- Expected
- No visible damage, kinks, cuts, or corrosion
- Tools
- Visual inspection
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Filament driver overcurrent or saturation fault high voltage
Overload Fault or Saturation Fault LED on Generator Driver PCB lit, indicating filament drive circuit malfunction or excessive current draw
Ranked causes
- 1. Filament drive circuit overload (occasional)
- 2. Filament drive transistor saturation or control circuit fault (occasional)
Observe Overload Fault and Saturation Fault LEDs during normal operation
- Test point
- Generator Driver PCB: Overload Fault LED, Saturation Fault LED
- Expected
- Both LEDs OFF during normal X-ray exposure and idle
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Housing overheating caution
System displays HOUSING IS HOT, HOUSING IS WARM, or HOUSING OVERHEATED error, indicating internal temperature approaching or exceeding safe limits
Ranked causes
- 1. Internal heat accumulation due to sustained X-ray exposure (common)
- 2. Cooling fan malfunction (occasional)
Monitor housing temperature percentage
- Test point
- Control panel display
- Expected
- Percentage of capacity shown
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Irregularities when field size changes
Image control anomalies occur during field size adjustment in fluoro mode
Ranked causes
- 1. Control Panel Processor PCB failure (common)
- 2. Fluoro Functions PCB failure (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Key switch in standby position caution
X-ray disable keyswitch turned on (vertical); prevents vertical column movement and X-ray generation; X-rays and lift disabled
Ranked causes
- 1. X-ray disable keyswitch in ON position (common)
- 2. Keyswitch or associated hardware fault (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No image on monitor
Monitor displays no image output from the C-arm system
Ranked causes
- 2. Bad CCD camera (common)
- 3. Bad monitor (occasional)
Verify kV tracking to nominal level
- Test point
- kV output
- Expected
- 120
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Video abnormalities, artifacts, or drop-outs
Video signal exhibits artifacts, abnormalities, or intermittent drop-outs in image
Ranked causes
- 1. CCD Camera failure (common)
- 2. Fluoro Functions PCB failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
WARNING LOW MA displayed patient safety
System displays 'WARNING LOW MA - PRESS ANY KEY' message on C-Arm Control Panel when actual mA output is lower than display shows and not within tolerance. Results in poor image quality.
Ranked causes
- 1. Filament drive circuitry fault (unknown)
- 2. mA Sense circuitry fault (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
24 hour recharge required caution
System displays 24 HOUR RECHARGE REQUIRED – X-RAYS DISABLED, indicating battery charge is critically low and system requires extended charging
Ranked causes
- 1. Battery discharged below safe operating level (common)
Monitor battery charge status during extended charging
- Test point
- Battery voltage and charge indicator
- Expected
- Battery voltage gradually rising during 24-hour period
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
4 hour warm up required caution
System displays 4 HOUR WARM UP REQUIRED on control panel, indicating stator/rotor requires warm-up period before high-power operation
Ranked causes
- 1. System powered on after extended shutdown requiring thermal stabilization (common)
Monitor warm-up timer
- Test point
- Control panel display
- Expected
- Timer counts down from 4 hours to zero
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
AD channel measurement failure caution
System reports AD CHANNEL FAIL error on control panel display, indicating analog-to-digital conversion failure on specified channel
Ranked causes
- 1. Analog-to-digital converter malfunction on specified channel (common)
Verify channel number from error message
- Test point
- Control panel display
- Expected
- Channel number displayed in error message
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Anode overheating radiation
System displays ANODE IS HOT or ANODE IS WARM error on control panel, indicating X-ray tube anode temperature approaching or exceeding safe limits
Ranked causes
- 1. Anode thermal capacity exceeded or approaching limit (common)
Monitor anode heat capacity percentage
- Test point
- Control panel display
- Expected
- Percentage of capacity below safe threshold
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ARCNET communications failure
Control panel button pushes do not cause intended changes on workstation; ARCNET network communications failure
Ranked causes
- 1. Fluoro Functions PCB failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charge below minimum caution
System displays BATTERY CHARGE - ##% error on control panel, indicating battery state of charge is insufficient for operation
Ranked causes
- 1. Battery discharged below minimum operating level (common)
Monitor battery charge percentage
- Test point
- Control panel display or error log
- Expected
- Percentage of capacity shown
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Boot sequence takes longer than normal
System takes considerably longer than 80-120 seconds to complete bootup sequence
Ranked causes
- 1. Systems with WJ98 or later software require 128 Meg RAM instead of 64 Meg (common)
Verify system software version and required RAM configuration
- Test point
- SBC memory configuration
- Expected
- 128 Meg RAM for WJ98+ software, 64 Meg for earlier versions
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bright pixels present in image
Camera cooler failure causes bright pixels in image due to offset correction failure
Ranked causes
- 1. Camera Cooler failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Film mode exposure radiation
Verify film mode X-ray exposure at 60 kVp @ 2.5 mAs produces acceptable image on developed film cassette
Ranked causes
- 1. System calibration verification required (unknown)
Listen for anode acceleration whirring sound before X-ray indicator
- Test point
- X-ray switch indicator and audio output
- Expected
- Whirring sound approximately 2 seconds before indicator illuminates
- Tools
- Footswitch
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Mechanical brake operation
Verify Wig-Wag and Flip Flop brake engagement prevents unwanted movement
Ranked causes
- 1. Brake system calibration verification (unknown)
Test Wig-Wag brake prevents movement
- Test point
- Wig-Wag arm with brake engaged
- Expected
- No Wig-Wag movement when brake is engaged
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Camera iris mechanism failure caution
System displays CAMERA IRIS FAILURE error, indicating iris control mechanism is malfunctioning
Ranked causes
- 1. Iris control mechanism stuck or damaged (occasional)
Determine iris position state from error data
- Test point
- Error message data field 1
- Expected
- 0 = open state; 1 = closed state
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CCD camera rotation binds or exhibits jerky motion
Camera rotation shows binding or jerky motion during operation
Ranked causes
- 1. Fluoro Functions PCB failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charger failure caution
System displays CHARGER FAILED or CHARGER FAILED with shutdown, indicating battery charger circuit malfunction
Ranked causes
- 1. Battery charging circuit failure (common)
Measure battery voltage via inverter
- Test point
- Inverter B A/D
- Expected
- Within charging specification
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collimation failure radiation
Collimator jaws fail to open/close or position incorrectly; X-ray field size uncontrollable
Ranked causes
- 1. Collimator motor, solenoid, or control circuit failure (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collimator calibration required caution
System displays COLLIMATOR CAL REQUIRED error on control panel, indicating collimator leaves or iris needs recalibration
Ranked causes
- 1. Collimator leaf or iris position drift out of calibration (occasional)
Identify which component requires calibration
- Test point
- Error data field 1
- Expected
- 0 = Leaves; 1 = Iris; 2 = Leaves and Iris
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collimator iris potentiometer error caution
System displays COLL IRIS POTENTIOMETER ERROR, indicating the iris position feedback sensor is malfunctioning
Ranked causes
- 1. Iris potentiometer sensor failure or connection issue (common)
Check potentiometer output reading
- Test point
- Pot reading from error message
- Expected
- Reading outside normal range or erratic
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collimator iris too large caution
System displays COLL IRIS TOO LARGE error, indicating iris opening exceeds maximum safe aperture
Ranked causes
- 1. Iris potentiometer reading indicates opening beyond specification (occasional)
Check iris potentiometer reading
- Test point
- Pot reading from error
- Expected
- Reading above maximum aperture threshold
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collimator iris unstable caution
System displays COLL IRIS UNSTABLE error, indicating collimator iris position is drifting or fluctuating
Ranked causes
- 1. Collimator iris motor or potentiometer malfunction (occasional)
Check iris potentiometer reading
- Test point
- Pot reading from error data
- Expected
- Stable reading within specification
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collimator movements are jerky or sticky
Collimator exhibits sticky or jerky motion during operation
Ranked causes
- 1. Collimator lubrication insufficient (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collimator or camera motor not responding
Camera iris, optics, collimator leaves or iris functions fail to operate
Ranked causes
- 1. Fluoro Functions PCB fault affecting motor control (common)
Measure +24VDC supply to II power supply at J7
- Test point
- J7 pins 15-8
- Expected
- +24VDC
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collimator stuck caution
System displays COLLIMATOR STUCK error, indicating collimator leaves cannot move to commanded position
Ranked causes
- 1. Collimator leaf mechanical jam or motor failure (common)
Monitor potentiometer reading
- Test point
- Pot reading during leaf movement attempt
- Expected
- Reading changes when motor is commanded
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication failure
System displays COMMUNICATION FAILED error, indicating loss of communications between system components
Ranked causes
- 1. Communication bus error or component disconnection (occasional)
Verify system startup completes
- Test point
- Control panel displays
- Expected
- System initializes without error
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control panel error
System displays CONTROL PANEL ERROR on display, indicating control panel microprocessor or interface failure
Ranked causes
- 1. Control panel microprocessor or communications interface failure (rare)
Verify control panel responsiveness
- Test point
- Control panel buttons and display
- Expected
- Display shows message and accepts input
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control panel key stuck caution
System senses key closure during boot-up; possible stuck key detected
Ranked causes
- 1. Control panel key stuck or object pressing key (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dark images with low pulse rate
Pulse rates less than 4 pps produce dark images due to failure of integrating data over multiple frame times; auto brightness stabilization incorrect
Ranked causes
- 1. Camera control failure (Manchester Encoder/Decoder) (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error message displayed during bootup
Error message appears on C-Arm Control Panel after boot sequence completion
Ranked causes
- 1. Hardware problem preventing IDE hard drive software load (occasional)
Observe boot display pattern on 20-character control panel
- Test point
- C-Arm Control Panel display
- Expected
- Five squares, ten squares, 15 squares, 20 squares pattern, then arrows
Fast stop activated caution
System displays FAST STOP ACTIVATED and powers down, indicating emergency shutdown was triggered
Ranked causes
- 1. Emergency stop button pressed or safety condition detected (unknown)
Check high voltage register
- Test point
- High Voltage Input Register
- Expected
- HV present at time of fast stop
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Field size selection failure
Image intensifier field size control unresponsive or produces incorrect zoom levels
Ranked causes
- 1. Image intensifier tube or power supply control malfunction (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Filament calibration required caution
System displays FILAMENT CAL REQUIRED error on control panel, indicating filament characteristics need recalibration
Ranked causes
- 1. Filament parameters drift out of calibration (occasional)
Run filament calibration procedure
- Test point
- Generator control system
- Expected
- Calibration completes without error
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Filament regulator failure high voltage
System displays FILAMENT REGULATOR FAILED error, indicating filament voltage regulation circuit is malfunctioning
Ranked causes
- 1. Filament voltage regulation circuit failure (common)
Measure filament voltage
- Test point
- Filament supply output
- Expected
- Measured filament voltage from error data
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Filament select error caution
System displays FILAMENT SELECT ERROR, indicating incorrect filament is selected or filament selection circuit failure
Ranked causes
- 1. Filament selector switch or control circuit failure (occasional)
Check which filament is selected
- Test point
- Error data field 1
- Expected
- 0 = Small filament; 1 = Large filament
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluoro Functions PCB microcontroller not running
LED segment 6 on DS1/DS2 does not blink at 1 Hz after bootup, indicating microcontroller startup failure or crash
Ranked causes
- 1. Microcontroller initialization failure (unknown)
Observe LED segment 6 on DS1 blinking after power-up
- Test point
- DS1 LED segment 6
- Expected
- Blink at 1 Hz during bootup and after
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluoro mode performance check failed patient safety
System fluoro mode performance test did not pass; image quality, exposure, or mode parameters out of spec
Ranked causes
- 1. One or more fluoro subsystem parameter out of specification (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Footswitch security error
- Generator filament calibration running
- High kV warning
- High kVp warning
- High mA warning
- High voltage generator error
- High voltage generator unexpected on
- High voltage register failure at startup
- High-level fluoroscopy overtime
- HV register failure
- Image brightness and contrast not compensating with added filtration
- Image quality unacceptable in fluoroscopy
- Image rotation or reversal in fluoroscopy
- Key stuck
- Key switch in standby
- KV turned on in error
- Low dose ABS table loading
- Low kV warning
- Low kVp warning
- Low mA warning
- mA sense on without X-ray switch
- mA sensor failure
- mA turned on in error
- No click heard when Interconnect cable plugged into C-Arm
- No fluoroscopy image output
- Overload fault
- Overvoltage fault
- Performance check: Image quality unacceptable
- Performance check: No Fluoro image
- Power Supply PS2 output voltage incorrect
- Pre-charge voltage error
- Precharge circuit timeout
- Precharge voltage error
- Room door open
- Saturation fault
- Standard ABS table loading
- Stator not on
- Stator standby wait
- System processing in progress
- System shuts down unexpectedly without error message
- Temperature sensor failure
- Vascular mode control failure
- Verify ARCNet signal integrity at nodes
- Workstation beeps and does not boot when switched on
- X-ray control released early in film mode
- X-ray Controller PCB initialization failure
- X-ray overtime
- X-RAY OVERTIME message displayed
- X-ray room door open
- X-ray switch security failure
- X-ray switch stuck
- X-ray tube anode not powered
- X-ray tube current on without control
- X-ray tube housing at elevated temperature
- X-ray tube housing at warm temperature
- X-ray tube housing overheated
- X-RAY TUBE TEMPERATURE SENSOR FAILURE displayed
- X-rays disabled
58 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
24 HOUR RECHARGE REQUIRED | Battery charge critically low; system requires 24-hour recharge cycle | Battery voltage below minimum safe operating level | error |
24-HOUR RECHARGE REQUIRED - X RAYS DISABLED | Battery charge depleted after long-term storage | C-Arm batteries depleted; typically after six months or more storage | warning |
4 HOUR WARM UP REQUIRED | Stator and tube require 4-hour warm-up before full-power operation | System powered on after extended shutdown | warning |
4-HOUR WARM-UP REQUIRED - PRESS ANY KEY | System inactive for more than 60 days; ion removal from Image Intensifier required | System has been inactive for more than 60 days | warning |
AD CHANNEL # FAIL | Analog-to-digital converter channel failure | A/D channel functional test failure at startup or during operation | error |
AD CHANNEL %s FAIL | Analog-to-digital conversion failure on specified channel | A/D converter unable to process signal from specified channel | error |
AD CHANNEL %s FAIL – POWER OFF – WAIT 5 SECONDS | Analog to digital converter channel failure | AD converter channel malfunction detected | error |
AD CHANNEL %s FAIL – POWER OFF. WAIT 5 SECONDS | Analog-to-digital channel failure at startup | A/D channel fails during system start-up | critical |
ANODE IS HOT | X-ray tube anode is at high temperature near safe limit | Anode heat accumulation from sustained X-ray exposure | warning |
ANODE IS HOT | Anode heat unit capacity critical | Anode heat units reach critical level approaching maximum capacity | critical |
ANODE IS HOT - XX% | X-ray tube anode at 80% or greater of rated capacity | X-ray tube reaches 80% (+2%) of rated thermal capacity | error |
ANODE IS WARM | X-ray tube anode is at elevated temperature but safe | Anode heat accumulation approaching high threshold | warning |
ANODE IS WARM | Anode heat unit capacity elevated | Anode heat units elevated but not at critical level | warning |
ANODE IS WARM - XX% | X-ray tube anode at 70% of rated capacity; warning only | X-ray tube reaches 70% (+2%) of rated thermal capacity | warning |
BATTERY CHARGE - ##% | Battery state of charge below minimum safe level | Battery voltage drops below minimum operating threshold | error |
BATTERY CHARGE - XX% | Battery charge between 70% and 40% of required level | C-Arm battery charge drops to 70%-40% range | warning |
BATTERY CHARGE - XX% /WAIT | Battery charge below 40% of required level; limited functionality | C-Arm battery charge drops below 40% of required charge | warning |
CAMERA IRIS FAILURE | Camera iris mechanism failure | Camera iris motor or sensor malfunction | warning |
CHARGER FAILED | Battery charging circuit is malfunctioning | Charger unable to deliver proper charge voltage or current | error |
CHARGER FAILED | Battery charger failure detected during bootup | Battery charger malfunction at system startup | error |
COLLIMATOR CAL REQUIRED | Collimator leaves or iris require position recalibration | Collimator position feedback drift out of tolerance | error |
COLLIMATOR CAL REQUIRED - PRESS ANY KEY | Collimator iris and leaf position values do not match stored reference values | Collimator position values drift from stored reference values in memory | critical |
COLLIMATOR STUCK | Collimator iris motor cannot open or close iris | Collimator iris motor mechanical failure or jam | error |
COLL IRIS POTENTIOMETER ERROR | Iris position feedback potentiometer is malfunctioning | Potentiometer output outside expected range or erratic | error |
COLL IRIS POTENTIOMETER ERROR - PRESS ANY KEY | Collimator iris position potentiometer failure or disconnection | Position potentiometer on collimator not working or disconnected | critical |
COLL IRIS TOO LARGE | Collimator iris aperture larger than displayed value on control panel | Collimator iris aperture size exceeds indicated value | warning |
COLL IRIS UNSTABLE | Collimator iris motor cannot maintain position tolerance | Collimator iris motor position instability, typically jittery operation | warning |
COMMUNICATION FAILED | Workstation and C-Arm communication loss for more than one second | No communication between Workstation and C-Arm exceeds one-second timeout; Watchdog circuit triggers | critical |
COMMUNICATION FAILED | Communication between system components lost | Data transmission error on system communication bus | error |
CONTROL PANEL ERROR | System cannot communicate with C-Arm Control Panels | Loss of communication between generator and control panels | critical |
CONTROL PANEL ERROR | Control panel microprocessor or interface failure | Panel processor malfunction or bus error | error |
CONTROL PANEL KEY STUCK - RELEASE, THEN PRESS ANY KEY TO CON | C-Arm Control Panel key stuck in pressed position at bootup | Control panel key mechanically stuck or electrically held at bootup | critical |
DATA ERROR | System detects invalid data in memory or processing | Invalid ABS table write/read, corrupted non-volatile memory, or software malfunction | critical |
DATA ERROR | System memory or flash storage contains corrupted data | Memory CRC check failure or data range error | error |
FAST STOP ACTIVATED | Emergency shutdown was triggered | Emergency stop button pressed or safety condition detected | warning |
FAST STOP ACTIVATED – POWER OFF. WAIT 5 SECONDS. | Interlocks opened; fast stop switch activated | FAST STOP switch pressed on front of C-Arm Control Panel or interlock circuit opened | error |
FILAMENT CAL REQUIRED | Filament characteristics require recalibration | Filament parameters drift beyond acceptable tolerance | error |
FILAMENT CAL REQUIRED - PRESS ANY KEY | Filament calibration not performed since system startup or software update | Software detects missing or expired filament calibration at startup | error |
FILAMENT REGULATOR FAILED | Filament voltage regulation circuit has failed | Filament regulator unable to maintain set voltage | error |
FILAMENT REGULATOR FAILURE - PRESS ANY KEY | X-ray tube filament current out of tolerance | Filament current measurement exceeds tolerance limits | error |
FILAMENT SELECT ERROR | Incorrect filament selected or filament selector failure | Mismatch between commanded and actual filament | error |
FILAMENT SELECT ERROR - PRESS ANY KEY | Selected filament size in software does not match filament in PIO hardware | Filament selection mismatch between software and hardware configuration | error |
FILAMENT STANDBY | Filament in standby mode | Filament heater energized but X-ray production not active | info |
FLUORO ALARM | Fluoroscopy alarm condition triggered | Fluoroscopy mode parameter or time limit exceeded | warning |
HLF OVERTIME | High-level fluoroscopy overtime condition detected | Prolonged high-level fluoroscopy operation | warning |
HLF OVERTIME | Generator operating in overtime mode | Generator heat limit exceeded during operation | warning |
HLF OVERTIME (digital cine) | High-level fluoroscopy overtime during cine acquisition | Prolonged high-speed recording mode exceeds overtime limit | warning |
HOUSING IS HOT | Generator internal housing at elevated temperature | Internal component temperature high due to X-ray exposure | warning |
HOUSING OVERHEATED | Generator internal housing at or above maximum safe temperature | Internal temperature exceeds maximum safe threshold | error |
HOUSING WARM | Generator internal housing at elevated but safe temperature | Internal temperature high but below maximum limit | warning |
HV GENERATOR ON | High voltage power supply is active when it should be off | Unintended HV output or control failure | error |
HV REGISTER FAIL | High voltage monitoring register has failed | HV register cannot be read or written | error |
INTERLOCK FAILURE | Safety interlock system is not functioning properly | Interlock solenoid, latch, or sensor malfunction | critical |
KEY STUCK | A physical control button or key is mechanically stuck | Button does not return to normal position after depression | error |
KEY SWITCH IN STANDBY | Key switch is in standby position disabling system operation | Key switch manually set to standby by operator | info |
KV ON IN ERROR | Kilovoltage became active unexpectedly | Unintended kV supply activation | critical |
LIFT SWITCH STUCK | Vertical positioning switch is mechanically stuck | Lift control switch cannot change state | error |
MA ON IN ERROR | Tube current became active when it should be off | Unintended tube current activation | critical |
MA SENSOR FAIL | Tube current measurement sensor is malfunctioning | mA sensor output out of range or erratic | error |
OVERLOAD FAULT | Electrical load or power consumption exceeds safe limit | Power draw exceeds system capacity | error |
OVERVOLTAGE FAULT | High voltage supply exceeds maximum safe level | HV regulator failure allowing excessive output | critical |
PLEASE WAIT | System is performing initialization or processing operation | System startup or internal operation in progress | info |
PRECHARGE CIRCUIT TIMEOUT | HV power supply cannot charge to operating voltage within timeout period | Precharge circuit failure or insufficient battery voltage | error |
PRE-CHARGE VOLTAGE ERROR | Battery inverter voltage insufficient for HV power supply precharge | Inverter B battery voltage below precharge threshold | error |
PRE-CHARGE VOLTAGE ERROR – POWER OFF | Pre-charge inverter voltage error | Inverter pre-charge voltage out of specification | error |
RELEASED EARLY | X-ray exposure terminated before expected completion | Operator released exposure button early | info |
ROOM DOOR OPEN | Procedure room safety door is open | Room door sensor indicates open state | warning |
SATURATION FAULT | Power supply or transformer has reached magnetic saturation | Excessive current through transformer core | error |
STATOR NOT ON | X-ray tube rotor/stator motor is not running | Stator motor failed to start or power lost to motor | error |
STATOR STANDBY | Stator motor is in standby state and not rotating | System standby mode or motor disabled during initialization | info |
TEMP. SENSOR FAIL | Temperature measurement sensor is malfunctioning | Temperature sensor output out of range | error |
WARNING – HIGH kV | Measured kilovoltage exceeds desired setpoint | kV regulation failure or setpoint error | warning |
WARNING, HIGH MA | Measured tube current exceeds desired setpoint | mA regulation failure or setpoint error | warning |
WARNING – LOW kV | Measured kilovoltage is below desired setpoint | kV supply unable to achieve setpoint | warning |
WARNING, LOW MA | Measured tube current is below desired setpoint | mA supply unable to reach setpoint | warning |
WARNING LOW MA | Actual mA output lower than displayed value and not within tolerance | mA output falls below expected range during operation | warning |
X-RAY OVERTIME | Film shot exposure time exceeded required duration | Exposure continues beyond required time limit | warning |
XRAY OVERTIME | Cumulative daily X-ray exposure has exceeded safe limit | Total exposure time exceeds daily maximum | error |
XRAYS DISABLED | X-ray output has been disabled by safety system | Safety interlock engaged or control system failure | warning |
X-RAY SWITCH SECURITY ERROR | Mismatch between Security line on footswitch/handswitch and connector | Security line signal mismatch detected | error |
X-RAY SWITCH SECURITY FAILURE | X-ray enable switch security check has failed | X-ray switch state does not match expected control signal | error |
X-RAY SWITCH STUCK | X-ray enable switch cannot change from current state | Switch solenoid or mechanical latch unable to move | error |
X-RAY SWITCH STUCK | X-ray control switch detected in ON state during system startup | Left or Right X-ray control switch held ON at startup | error |
X-RAY TUBE TEMPERATURE SENSOR FAILURE | X-ray tube temperature thermistor fails functional testing at startup | Thermistor functional test failure during system initialization | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.