GE Healthcare B125/B105: troubleshooting and error codes
The ServiceForge knowledge base holds 133 documented symptoms for the GE Healthcare B125/B105, with 189 ranked causes, 128 diagnostic checks with test points and expected values, and 63 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Unit does not power on from AC mains high voltage
Mains voltage indicator not lit, monitor connected to AC power, no battery installed
Ranked causes
- 1. Power cord loose (common)
- 2. Blown fuses (common)
- 3. Power cord faulty (occasional)
- 4. Power outlet does not meet specifications (occasional)
- 5. Cable loose or faulty between AC/DC module and AC inlet (occasional)
- 6. Cable loose or faulty between AC/DC module and carrier board (occasional)
- 7. AC/DC module defective (rare)
- 8. Carrier board defective (rare)
Verify power cord connection
- Test point
- Wall outlet and monitor power inlet
- Expected
- Power cord fully seated
4 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Parameter module data not displayed
Parameter values from acquisition module not shown on monitor display
Ranked causes
- 1. Parameter module defective (occasional)
- 2. Module not connected properly (common)
- 3. Cable between frame and host not connected or defective (occasional)
- 4. Monitor display not configured for parameter (common)
- 5. Frame defective (rare)
Verify module connection
- Test point
- Module connector at frame
- Expected
- Module fully seated and connected
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on from battery
Monitor not connected to AC mains, powered from battery, monitor will not turn on
Ranked causes
- 1. Battery empty (common)
- 2. Battery failure (occasional)
- 3. Battery not installed (common)
- 4. Battery interface board loose or faulty (occasional)
- 5. Carrier board defective (rare)
Check battery charge status
- Test point
- Battery and monitor
- Expected
- Battery fully charged
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarms do not work patient safety
No sound from speaker during alarm conditions
Ranked causes
- 1. Audible alarms disabled (common)
- 2. Speaker failure (occasional)
- 3. Speaker cable loose or faulty (occasional)
- 4. Tone generator or audio amplifier failure (rare)
Verify audible alarm setting
- Test point
- Alarms Setup menu
- Expected
- Audible alarms enabled
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery LED indicator flashing orange caution
Orange flashing LED on monitor front panel battery indicator
Ranked causes
- 1. AC power input overvoltage (occasional)
- 2. Battery overvoltage (occasional)
- 3. Battery precharge timeout fault (occasional)
- 4. Battery fast charge timeout fault (occasional)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display remains black or white after startup
Monitor starts with beep sound but display shows no image (black or white)
Ranked causes
- 1. Display cables loose (common)
- 2. Display panel defective (occasional)
- 3. LCD interface board defective (occasional)
- 4. Battery issue (rare)
Verify display cable connection
- Test point
- Display cable connectors
- Expected
- Display cable fully seated and intact
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
IBP no pressure reading patient safety
Invasive blood pressure shows no reading or zero pressure
Ranked causes
- 1. Defective transducer (common)
- 2. Hemo module not connected (common)
- 3. Transducer adapter cable not connected (common)
- 4. Invasive pressure channel not configured to screen (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Acquisition module incompatible
Module does not work properly with patient monitor
Ranked causes
- 1. Incompatible module (occasional)
- 2. Connection issue (occasional)
- 3. Carrier board faulty (rare)
Verify module compatibility
- Test point
- Module type and model
- Expected
- Module listed in compatible modules
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Invasive pressure transducer patient safety
Invasive pressure calibration procedure for pressure transducer replacement, functional check failure, or values outside specification limits
Ranked causes
- 1. Pressure transducer replaced with new type (common)
- 2. Invasive pressure functional check failed (occasional)
- 3. Measured value not within specification limits (occasional)
Calibration: Pressure channel patient safety
Channel calibration procedure with success, failure, and specific error conditions for invasive blood pressure measurement
Ranked causes
- 1. Pulsating waveform detected during calibration (common)
- 2. Gain beyond limits (±20% of default gain) (occasional)
- 3. Calibration reference value less than 100 mmHg (common)
Monitor for pulsating waveform during calibration
- Test point
- pressure channel
- Expected
- Static pressure reading without pulsation
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Module visual inspection for E-COP caution
Physical integrity check of module housing, connectors, and cleanliness prior to functional testing
Ranked causes
- 1. Module damaged during service or shipping (rare)
- 2. Connector corrosion or poor seating (occasional)
- 3. Contamination affecting electrical performance (occasional)
Inspect module front cover for cracks or damage
- Test point
- Front cover surface
- Expected
- Intact, no visible damage
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder does not work
Recorder unit fails to operate or produce output
Ranked causes
- 1. Recorder connect cable loose or faulty (common)
- 2. Recorder unit failure (occasional)
- 3. Carrier board issue (rare)
Check recorder cable connection
- Test point
- Recorder connect cable
- Expected
- Cable fully seated and intact
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature not displayed
No temperature reading or value shown on monitor display
Ranked causes
- 1. Temperature channel not configured to screen (common)
- 2. Temperature out of measurable range (occasional)
- 3. Incompatible temperature probe (occasional)
Verify temperature is within measurable range
- Test point
- Temperature reading
- Expected
- 10-45 °C, 10 to 45 °C
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on with AC mains and charged battery installed high voltage
Monitor connected to AC mains with fully charged battery installed but will not turn on
Ranked causes
- 1. On/Off key faulty (occasional)
- 2. LCD interface board defective (occasional)
- 3. Carrier board defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Zeroing: Invasive blood pressure channel patient safety
Zeroing procedure for IBP channels with status messages and failure conditions
Ranked causes
- 1. Pulsating waveform detected during zeroing (common)
- 2. Defective transducer (occasional)
- 3. Offset greater than 150 mmHg (occasional)
Monitor for pulsating waveform during zeroing
- Test point
- pressure channel
- Expected
- No pulsation; static baseline
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm light does not illuminate patient safety
Alarm light not lit on monitor front panel; audible alarms work and alarm message visible
Ranked causes
- 1. Alarm interface cable loose or faulty (common)
- 2. Alarm LED board faulty (occasional)
Check alarm interface cable
- Test point
- Alarm cable connection to LCD interface board
- Expected
- Cable intact and properly connected
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas exhaust blockage detected caution
Monitor displays 'Sample gas out' message in digit field when gas exhaust connector is blocked
Ranked causes
- 1. Gas exhaust connector blocked or restricted (common)
- 2. Backpressure detection sensor malfunction (occasional)
Block gas exhaust connector with finger
- Test point
- Gas exhaust connector outlet
- Expected
- Exhaust path blocked
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas sampling line occlusion detected patient safety
Sampling line blocked; monitor displays 'Sample line blocked' message in digit field within 30 seconds of occlusion
Ranked causes
- 1. Sampling line tip blocked or kinked (common)
- 2. Occlusion detection sensor malfunction (occasional)
Block sampling line tip with finger
- Test point
- Gas sampling line inlet
- Expected
- Line physically blocked
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect system time when not network connected
System clock shows wrong date and time when monitor is not connected to network
Ranked causes
- 1. CPU battery empty (common)
- 2. Time not configured properly (common)
Check system date and time
- Test point
- Monitor clock display
- Expected
- Correct date and time displayed
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard not working or partially working
Keyboard keys unresponsive or only some keys work
Ranked causes
- 1. Keyboard cable loose or faulty (common)
- 2. Membrane keypad faulty (occasional)
Check keyboard cable connection
- Test point
- Keyboard cable to LCD interface board
- Expected
- Cable fully seated and intact
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
MC network port authentication issue
Monitor cannot authenticate on IEEE 802.1X port-based authentication enabled port
Ranked causes
- 1. 802.1X authentication not enabled on monitor (common)
- 2. Incorrect authentication credentials (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP air leakage
NIBP module exhibits air leakage or loss of pressure integrity
Ranked causes
- 1. Hose or cuff leaking (common)
- 2. Leakage inside the module (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No MC network traffic
Monitor not communicating on MC network; no traffic detected
Ranked causes
- 1. Incorrect network configuration (common)
- 2. MC network traffic overload (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No transducer connected or sensor faulty patient safety
Missing or non-functional pressure transducer detected on indicated channel
Ranked causes
- 1. Transducer not connected (common)
- 2. Sensor faulty (occasional)
Verify transducer connection
- Test point
- transducer connector on indicated channel
- Expected
- Transducer physically connected
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sample flow rate out of specification
Measured gas sample flow rate does not meet 150 ± 25 ml/min specification; indicates pump or flow restriction issue
Ranked causes
- 1. Sample gas flow rate too low or high (common)
- 2. Sampling pump motor bearing wear or malfunction (occasional)
Connect module and allow 5 minute warm-up
- Test point
- Module operation
- Expected
- Module powered and stabilized
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touchscreen inoperative
Touchscreen does not respond to touches or clicks
Ranked causes
- 1. Touchscreen cable loose (common)
- 2. Touchscreen sensor faulty (occasional)
Check touchscreen cable connection
- Test point
- Touchscreen cable to LCD interface board
- Expected
- Cable fully seated and intact
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Trim knob not working
Trim knob control is unresponsive
Ranked causes
- 1. Trim knob cable loose or faulty (common)
- 2. Trim knob defective (occasional)
Check trim knob cable connection
- Test point
- Trim knob cable to LCD interface board
- Expected
- Cable fully seated and intact
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Water trap disconnected or missing
Monitor displays 'Check Water Trap' message in digit field; water trap detached from sampling system
Ranked causes
- 1. Mini D-fend water trap disconnected or unseated (common)
- 2. Water trap detection sensor malfunction (occasional)
Detach Mini D-fend water trap from module
- Test point
- Water trap gas connector
- Expected
- Water trap disconnected from module
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ambient pressure sensor reading error
Absolute pressure sensor reading differs from barometer by more than ± 10 mmHg; indicates sensor malfunction
Ranked causes
- 1. Pressure sensor calibration drift or failure (occasional)
Access Gas service menu and note ambient pressure reading
- Test point
- Ambient pressure value in Gas service menu
- Expected
- Numeric pressure value displayed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Arrhythmia paused patient safety
ECG channels not available for analysis for 20+ seconds or internal HR not updated for 30+ seconds due to excessive artifact.
Ranked causes
- 1. Excessive artifact or poor electrode contact (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Artifact detected in ECG patient safety
Muscle artifact or high/low frequency noise detected in waveform.
Ranked causes
- 1. Poor electrode contact or lead placement, patient movement (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery empty
Monitor is battery powered with less than 5 minutes of monitoring time remaining.
Ranked causes
- 1. Battery charge depleted (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery low
Monitor is battery powered with less than 20 minutes of monitoring time remaining.
Ranked causes
- 1. Battery charge low (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery temperature high caution
Battery temperature is too high.
Ranked causes
- 1. Battery overheated (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibrate NIBP message displayed
NIBP calibration required. Message in digit field.
Ranked causes
- 1. NIBP calibration required (unknown)
Calibration: CPU battery replacement
Replacement of FRU B40B20 CPU battery at the recommended 5-year interval
Ranked causes
- 1. CPU battery maintenance interval reached (common)
CPU battery age verification
- Test point
- Installation records
- Expected
- 5 years or more since last replacement
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Display brightness adjustment
Set display brightness level according to operational needs
Ranked causes
- 1. Display brightness requires adjustment (common)
Verify display brightness adjustment is accessible
- Test point
- Menu > Brightness control
- Expected
- Brightness control present and functional
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: E-COP module maintenance check high voltage
Preventive and corrective maintenance checklist for E-COP module including visual inspection and electrical safety validation
Ranked causes
- 1. Routine E-COP module maintenance procedure (common)
Module front cover integrity
- Test point
- E-COP module front cover
- Expected
- Front cover is intact
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: E-Entropy module maintenance check patient safety
Verify E-Entropy module functionality, sensor recognition, and electrical safety prior to clinical use
Ranked causes
- 1. Module or sensor malfunction requiring verification (unknown)
Visual inspection of module front cover
- Test point
- Module front cover
- Expected
- Cover is intact
8 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Invasive blood pressure channel P4
Calibrate pressure transducer on P4 invasive blood pressure measurement channel
Ranked causes
- 1. P4 pressure channel calibration drift (unknown)
Verify P4 calibration menu available
- Test point
- Service interface COP calibrations
- Expected
- Calibrate P4 option accessible
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Invasive blood pressure channels P1 and P2
Calibrate pressure transducers on invasive blood pressure measurement channels
Ranked causes
- 1. Pressure channel calibration drift (unknown)
Verify P1/P2 calibration menu available
- Test point
- Service interface STP calibration
- Expected
- Calibrate P1/P2 option accessible
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP pressure measurement patient safety
NIBP calibration procedure to verify or adjust non-invasive blood pressure measurement accuracy using a digital manometer and reference pressure source.
Ranked causes
- 1. NIBP measurement out of specification or calibration check failure (unknown)
Connect adult NIBP hose to NIBP connector
- Test point
- NIBP connector
- Expected
- Secure connection established
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Post-FRU functional verification patient safety
Verification procedure after detaching, replacing or upgrading field replaceable units (FRU) in patient monitor to ensure proper installation and system functionality.
Ranked causes
- 1. FRU installation requires functional verification (common)
Check all product labelling, marking and symbols
- Test point
- Device exterior
- Expected
- Intact and readable
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Post-maintenance verification patient safety
Visual inspection and functional checks required after detaching, replacing or upgrading FRU components or maintenance parts in the B125/B105 patient monitor
Ranked causes
- 1. FRU replacement or upgrade completed (common)
Product labelling marking and symbols intact and readable
- Test point
- Monitor exterior
- Expected
- All labels present and legible
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Sample flow gain
Adjust sampling pump gain value for gas measurement accuracy
Ranked causes
- 1. Sample flow measurement drift (unknown)
Verify sample flow gain adjustment available
- Test point
- Service interface Gas Control menu
- Expected
- Sample Flow Gain parameter accessible
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Sample flow rate adjustment caution
Sample flow rate must be adjusted if the sample flow rate check failed during checkout procedures.
Ranked causes
- 1. Sample flow rate check failed (unknown)
Stop breathing to gas sampling line
- Test point
- Gas sampling line inlet
- Expected
- Apnea alarm appears in message field within 30 seconds
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Set national requirements
Activate France specific defaults for ECG heart rate adjustment range and reminder beep behavior
Ranked causes
- 1. National requirements configuration needed (occasional)
Access Service menu with authentication
- Test point
- Service menu
- Expected
- Service menu accessible after entering credentials
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Set time zone
Configure time zone, daylight savings, time sync source and NTP settings on patient monitor when not connected to network and patient discharged
Ranked causes
- 1. Time zone configuration required (common)
Verify monitor is not connected to network
- Test point
- Network connection status
- Expected
- Monitor not connected to network
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Slave display picture adjustment
Adjust picture quality on slave display using OSD menu
Ranked causes
- 1. Slave display picture requires adjustment (occasional)
Slave display picture quality acceptable
- Test point
- Slave display screen
- Expected
- Picture quality clear and properly adjusted per display user manual
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Temperature channels T1 and T2
Calibrate invasive temperature sensor channels T1 and T2 for accurate measurements
Ranked causes
- 1. Temperature channel calibration drift (unknown)
Verify T1/T2 calibration menu available
- Test point
- Service interface STP calibration
- Expected
- Calibrate T1/T2 option accessible
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Temperature measurement
Temperature channel self-calibration procedure
Ranked causes
- 1. Temperature calibration in progress (common)
Monitor temperature self-check status
- Test point
- temperature channel (DF)
- Expected
- Self-check completes with no error messages
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Wired CARESCAPE Network configuration
Configure monitor network settings including unit name, bed name, and MCS IP address
Ranked causes
- 1. Monitor requires network configuration for first-time use or maintenance (common)
Enter Unit Name in Network Config
- Test point
- Basic Service > Network menu
- Expected
- Unit Name field populated and saved
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiac output measurement not responding or reading incorrectly patient safety
C.O. manual measurement fails to start, displays error, or returns unexpected values; temperature readings (T injectate, T blood) not close to simulator set values
Ranked causes
- 1. Simulator temperature settings not matching configured values or measurement initialization issue (common)
Verify monitor C.O. configuration
- Test point
- Cardiac output menu
- Expected
- Catheter Type: User defined, Computation Constant: 0.542, Measurement Type: Manual, Injectate Volume: 10 ml, Ref Measurement: Deselect
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Device message displayed
Hemo module malfunction, Masimo type only. Message appears in digit field.
Ranked causes
- 1. Hemo module malfunction (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check NIBP message displayed
Systolic and/or diastolic results missing during NIBP measurement. Message in message field.
Ranked causes
- 1. Missing NIBP measurement results (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Probe message displayed
No detectable SpO2 signal. Sensor is faulty or detached from patient. Message in digit or message field.
Ranked causes
- 1. Faulty or detached SpO2 sensor (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check zero drift compensation NIBP
Monitor and assess zero drift compensation in NIBP pneumatics system
Ranked causes
- 1. NIBP zero drift exceeds ±10 mmHg limit (occasional)
Display and review zero drift value
- Test point
- Service interface NIBP module Zero parameter
- Expected
- Zero drift value between +20 to -20 mmHg, ideally within ±10 mmHg
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Condition battery
Battery is not working properly and requires conditioning.
Ranked causes
- 1. Battery requires conditioning or replacement (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
COP module gain adjustment
Adjust gain coefficient to compensate for gain error in cardiac output measurement
Ranked causes
- 1. Gain value outside normal range 17000-25000 (occasional)
Monitor gain coefficient value
- Test point
- Service interface COP module Gain parameter
- Expected
- Gain value between 17000 and 25000
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
COP module zero offset adjustment
Adjust zero offset compensation value for A/D converter in cardiac output measurement
Ranked causes
- 1. Zero offset value outside ±1000 range (occasional)
Monitor zero offset compensation value
- Test point
- Service interface COP module Zero parameter
- Expected
- Zero offset within ±1000
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Cuff loose message displayed
- Cuff occlusion message displayed
- Cuff overpressure message displayed
- E-COP module test result documentation
- ECG measurements removed
- ECG module error
- Faulty Probe message displayed
- Frame has power issue
- Frame temperature high
- Gas sampling system leakage detected
- Hemo parameters removed
- IBP abnormally low pressure
- IBP channel zeroed to over 100 mmHg
- ICP channel must be zeroed separately
- Identical IP address noticed
- Identical SpO2 modules message displayed
- Identical unit and bed name noticed
- Incompatible NIBP message displayed
- Incompatible Probe message displayed
- Incorrect system time when network connected
- Interference message displayed
- Invasive pressure channel not zeroed
- Invasive pressure measurement not zeroing or reading incorrectly
- Lead off - LA/L, LL/F, RA/R, RL/N, or V/C
- Leads off
- License invalid
- Loading failed
- Long measurement time message displayed
- Low Perfusion message displayed
- Low signal quality message displayed
- Maintenance: CPU battery replacement
- Measurement off - impedance respiration
- Memory ERROR
- Mode data reset
- Network down
- Network down: HL7
- Network storm causes temporary data loss
- NIBP cuff loose message displayed
- NIBP cuff occlusion message displayed
- NIBP manual message displayed
- NIBP measurement removed message displayed
- No battery backup
- No SpO2 probe message displayed
- No SpO2 pulse message displayed
- P1/P2/P4 over range message displayed
- Parameter module cannot be placed in frame properly
- Poor Signal message displayed
- Probe Off message displayed
- Pulse Search message displayed
- Px/P2/P4 under range message displayed
- Recorder cover open
- Recorder input voltage high or low
- Recorder out of paper
- Recorder system error
- Recorder thermal array overheat
- Recorder works but no output on paper
- Replace battery
- Replacement of planned maintenance parts
- Restart needed
- Select cuff size message displayed
- Small respiration curve
- Software activation incomplete or unsuccessful
- SpO2 faulty probe message displayed
- SpO2 measurement removed message displayed
- SpO2 module error message displayed
- SpO2 probe off message displayed
- Temperature error / T1 temperature error / T2 temperature error
- Tubing contamination check
- UMBC communication error
- Unit does not display Condition battery or Battery failure message after startup
- Unstable zero pressure message displayed
- Weak pulsation message displayed
- Zero valve not functioning
73 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Alarm setup changed from Central | Alarm setup retrieved from central station or ECG alarms modified from Central | Connection to central station established or ECG alarm configuration changed remotely | info |
Alarms reset from Central | Alarms were remotely paused from central station | Central station sends pause command | info |
Apnea deactivated | Monitor recently admitted patient or measurement just started | Patient admission or measurement initiation | info |
Arrhythmia Paused | ECG channels unavailable for analysis or HR not updated due to excessive artifact | Artifact present for 20+ seconds or HR not updated for 30+ seconds | warning |
Artifact | Muscle artifact or high/low frequency noise in ECG signal | Artifact detected in waveform | warning |
Calibrated | Channel calibrated successfully | Calibration procedure completed without errors | info |
Calibrating | Calibration of a channel is in progress | User initiates calibration procedure | info |
Call service: Error 0 | RAM test failure | Detected during startup self-test | critical |
Call service: Error 1 | ROM checksum failure | Detected during startup self-test | critical |
Call service: Error 10 | RTK 400Hz timer re-entry | Detected in real-time kernel | critical |
Call service: Error 11 | RTK 50Hz timer re-entry | Detected in real-time kernel | critical |
Call service: Error 13 | RTK overrun | Detected in real-time kernel | critical |
Call service: Error 14 | Too early AUTO START according to module check | AUTO START command issued prematurely | error |
Call service: Error 15 | Calibration data invalid on initialization or unit never calibrated | Device startup when calibration data is invalid | error |
Call service: Error 16 | Communication timeout between main and safety CPU | Detected during operation when CPUs fail to communicate | critical |
Call service: Error 17 | Safety CPU report communication timeout | Detected when safety CPU detects communication loss | critical |
Call service: Error 18 | Wrong message rate in communication between main and safety CPU | Message rate deviates >±2% from 480 msg/s | error |
Call service: Error 2 | Pump on during idle or over current detected | Detected during normal operation | error |
Call service: Error 3 | Startup communication failure with safety CPU | Detected at device startup | critical |
Call service: Error 5 | Calibration Not Protected | Detected during calibration procedure | error |
Call service: Error 6 | Valve stuck closed during cuff typing | During NIBP cuff inflation | error |
Call service: Error 7 | Could not save calibration data | After calibration procedure | error |
Call service: Error 8 | Pressure exceeds limits while idle | PT2 >150 mmHg for >15 s while idle, or pressure >15 mmHg for >180 s (Adult) or >5 mmHg for >90 s (Neo) | error |
Call service: Error 9 | Determination time too long | NIBP measurement takes excessive time | error |
Call service UMBC error | UMBC communication error and UMBC communication disabled | UMBC communication failure | error |
Default settings returned | Monitor settings have been reset to default | Factory reset or settings recovery activated | info |
DEMO MODE | DEMO mode has been enabled | DEMO mode activated | warning |
Entering standby | Standby mode activation selected | Standby activation command issued | info |
Failed | Pressure calibration failure due to time-out or waveform/gain issues | Calibration procedure does not complete within time limit or detects pulsation or gain out of range | error |
Failed: P<100 | Calibration reference value entered is less than 100 mmHg | User enters pressure reference below 100 mmHg | error |
InvBP's not zeroed | At least one invasive pressure channel has not been zeroed | Attempt to measure with unzeroed IBP channel | warning |
Lead changed | Monitor auto-switched ECG1 waveform to measurable ECG lead | Current ECG1 waveform not measurable | info |
Learning | ST algorithm in learning phase | ST algorithm startup or ECG measurement start | info |
License invalid | License is invalid during startup | Monitor startup with invalid license | error |
Loading failed | Loading mode from network interrupted | Network data transfer interrupted | error |
Loading from network | Patient data being loaded from the network | Data transfer from network initiated | info |
Memory ERROR | Memory error after software download | Software download completion | error |
Mode data reset | Error loading settings from flash file | Cold start, warm start, upgrade, or revert mode | error |
Network down | CARESCAPE Network connection failed | Network connectivity loss | error |
Network down: HL7 | No HL7 TCP Client configured to connect to monitor | HL7 connection attempt with no client configured | error |
Network: HL7 | HL7 TCP Client application network connection established | HL7 connection established | info |
Network made | CARESCAPE network connection established | Network connection successful | info |
No breath deactivated | One of the electrodes is off | Electrode disconnection detected | warning |
No x transducer | No transducer connected to indicated channel or sensor is faulty | Channel detects missing or non-responsive transducer | error |
Patient admitted | Current patient has been admitted | Patient admission recorded | info |
Patient discharged | Patient has been discharged | Patient discharge recorded | info |
Performing temp test | Temperature is calibrating (self-check in progress) | Temperature channel self-calibration initiated | info |
Printing... | Printing is occurring | Print operation initiated | info |
Printing Alarm | Alarm has triggered printing | Alarm condition activates print | warning |
Recorder: cover open | Recorder cover is open | Recorder cover opened | warning |
Recorder: input voltage high / Recorder: input voltage low | Problems with recorder input voltage | Recorder voltage out of specification | error |
Recorder module removed | Recorder module has been removed | Recorder module disconnected | info |
Recorder: out of paper | Recorder out of paper or cover open | No paper detected or cover open | warning |
Recorder system error {0d} | Local recorder not working | Recorder communication failure | error |
Recorder thermal array overheat | Problems with recorder temperature | Recorder temperature exceeds limit during operation | warning |
Saving | Printing device not available; records saved for later printing | Print request with no printer available | warning |
Saving to network | Saving trend and information to Network | Network save operation initiated | info |
STP measurements removed | Hemodynamic (Hemo) parameters have been removed | Hemo module disconnected or communication lost | warning |
Zero adj > 100 mmHg | IBP channel zeroed to over 100 mmHg pressure | Zero adjustment value exceeds 100 mmHg limit | error |
Zeroed | Zeroing was successful | Zeroing procedure completed without errors | info |
Zero ICP separately | ICP channel must be zeroed separately from other invasive pressures | Attempt to zero ICP with other pressure channels | warning |
Zeroing | IBP channel is currently being zeroed | User initiates zeroing procedure | info |
Zeroing failed | Zeroing procedure failed due to pulsation, defective transducer, or excessive offset | Zeroing encounters pulsating waveform, detects defective transducer, or offset exceeds 150 mmHg | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.