GE Healthcare B125/B105: troubleshooting and error codes

The ServiceForge knowledge base holds 133 documented symptoms for the GE Healthcare B125/B105, with 189 ranked causes, 128 diagnostic checks with test points and expected values, and 63 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Unit does not power on from AC mains high voltage

Mains voltage indicator not lit, monitor connected to AC power, no battery installed

Ranked causes

  1. 1. Power cord loose (common)
  2. 2. Blown fuses (common)
  3. 3. Power cord faulty (occasional)
  4. 4. Power outlet does not meet specifications (occasional)
  5. 5. Cable loose or faulty between AC/DC module and AC inlet (occasional)
  6. 6. Cable loose or faulty between AC/DC module and carrier board (occasional)
  7. 7. AC/DC module defective (rare)
  8. 8. Carrier board defective (rare)
First check

Verify power cord connection

Test point
Wall outlet and monitor power inlet
Expected
Power cord fully seated

4 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Parameter module data not displayed

Parameter values from acquisition module not shown on monitor display

Ranked causes

  1. 1. Parameter module defective (occasional)
  2. 2. Module not connected properly (common)
  3. 3. Cable between frame and host not connected or defective (occasional)
  4. 4. Monitor display not configured for parameter (common)
  5. 5. Frame defective (rare)
First check

Verify module connection

Test point
Module connector at frame
Expected
Module fully seated and connected

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not power on from battery

Monitor not connected to AC mains, powered from battery, monitor will not turn on

Ranked causes

  1. 1. Battery empty (common)
  2. 2. Battery failure (occasional)
  3. 3. Battery not installed (common)
  4. 4. Battery interface board loose or faulty (occasional)
  5. 5. Carrier board defective (rare)
First check

Check battery charge status

Test point
Battery and monitor
Expected
Battery fully charged

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarms do not work patient safety

No sound from speaker during alarm conditions

Ranked causes

  1. 1. Audible alarms disabled (common)
  2. 2. Speaker failure (occasional)
  3. 3. Speaker cable loose or faulty (occasional)
  4. 4. Tone generator or audio amplifier failure (rare)
First check

Verify audible alarm setting

Test point
Alarms Setup menu
Expected
Audible alarms enabled

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery LED indicator flashing orange caution

Orange flashing LED on monitor front panel battery indicator

Ranked causes

  1. 1. AC power input overvoltage (occasional)
  2. 2. Battery overvoltage (occasional)
  3. 3. Battery precharge timeout fault (occasional)
  4. 4. Battery fast charge timeout fault (occasional)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display remains black or white after startup

Monitor starts with beep sound but display shows no image (black or white)

Ranked causes

  1. 1. Display cables loose (common)
  2. 2. Display panel defective (occasional)
  3. 3. LCD interface board defective (occasional)
  4. 4. Battery issue (rare)
First check

Verify display cable connection

Test point
Display cable connectors
Expected
Display cable fully seated and intact

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

IBP no pressure reading patient safety

Invasive blood pressure shows no reading or zero pressure

Ranked causes

  1. 1. Defective transducer (common)
  2. 2. Hemo module not connected (common)
  3. 3. Transducer adapter cable not connected (common)
  4. 4. Invasive pressure channel not configured to screen (common)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Acquisition module incompatible

Module does not work properly with patient monitor

Ranked causes

  1. 1. Incompatible module (occasional)
  2. 2. Connection issue (occasional)
  3. 3. Carrier board faulty (rare)
First check

Verify module compatibility

Test point
Module type and model
Expected
Module listed in compatible modules

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Invasive pressure transducer patient safety

Invasive pressure calibration procedure for pressure transducer replacement, functional check failure, or values outside specification limits

Ranked causes

  1. 1. Pressure transducer replaced with new type (common)
  2. 2. Invasive pressure functional check failed (occasional)
  3. 3. Measured value not within specification limits (occasional)

Calibration: Pressure channel patient safety

Channel calibration procedure with success, failure, and specific error conditions for invasive blood pressure measurement

Ranked causes

  1. 1. Pulsating waveform detected during calibration (common)
  2. 2. Gain beyond limits (±20% of default gain) (occasional)
  3. 3. Calibration reference value less than 100 mmHg (common)
First check

Monitor for pulsating waveform during calibration

Test point
pressure channel
Expected
Static pressure reading without pulsation

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Module visual inspection for E-COP caution

Physical integrity check of module housing, connectors, and cleanliness prior to functional testing

Ranked causes

  1. 1. Module damaged during service or shipping (rare)
  2. 2. Connector corrosion or poor seating (occasional)
  3. 3. Contamination affecting electrical performance (occasional)
First check

Inspect module front cover for cracks or damage

Test point
Front cover surface
Expected
Intact, no visible damage

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Recorder does not work

Recorder unit fails to operate or produce output

Ranked causes

  1. 1. Recorder connect cable loose or faulty (common)
  2. 2. Recorder unit failure (occasional)
  3. 3. Carrier board issue (rare)
First check

Check recorder cable connection

Test point
Recorder connect cable
Expected
Cable fully seated and intact

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature not displayed

No temperature reading or value shown on monitor display

Ranked causes

  1. 1. Temperature channel not configured to screen (common)
  2. 2. Temperature out of measurable range (occasional)
  3. 3. Incompatible temperature probe (occasional)
First check

Verify temperature is within measurable range

Test point
Temperature reading
Expected
10-45 °C, 10 to 45 °C

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not power on with AC mains and charged battery installed high voltage

Monitor connected to AC mains with fully charged battery installed but will not turn on

Ranked causes

  1. 1. On/Off key faulty (occasional)
  2. 2. LCD interface board defective (occasional)
  3. 3. Carrier board defective (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Zeroing: Invasive blood pressure channel patient safety

Zeroing procedure for IBP channels with status messages and failure conditions

Ranked causes

  1. 1. Pulsating waveform detected during zeroing (common)
  2. 2. Defective transducer (occasional)
  3. 3. Offset greater than 150 mmHg (occasional)
First check

Monitor for pulsating waveform during zeroing

Test point
pressure channel
Expected
No pulsation; static baseline

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm light does not illuminate patient safety

Alarm light not lit on monitor front panel; audible alarms work and alarm message visible

Ranked causes

  1. 1. Alarm interface cable loose or faulty (common)
  2. 2. Alarm LED board faulty (occasional)
First check

Check alarm interface cable

Test point
Alarm cable connection to LCD interface board
Expected
Cable intact and properly connected

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gas exhaust blockage detected caution

Monitor displays 'Sample gas out' message in digit field when gas exhaust connector is blocked

Ranked causes

  1. 1. Gas exhaust connector blocked or restricted (common)
  2. 2. Backpressure detection sensor malfunction (occasional)
First check

Block gas exhaust connector with finger

Test point
Gas exhaust connector outlet
Expected
Exhaust path blocked

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gas sampling line occlusion detected patient safety

Sampling line blocked; monitor displays 'Sample line blocked' message in digit field within 30 seconds of occlusion

Ranked causes

  1. 1. Sampling line tip blocked or kinked (common)
  2. 2. Occlusion detection sensor malfunction (occasional)
First check

Block sampling line tip with finger

Test point
Gas sampling line inlet
Expected
Line physically blocked

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incorrect system time when not network connected

System clock shows wrong date and time when monitor is not connected to network

Ranked causes

  1. 1. CPU battery empty (common)
  2. 2. Time not configured properly (common)
First check

Check system date and time

Test point
Monitor clock display
Expected
Correct date and time displayed

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keyboard not working or partially working

Keyboard keys unresponsive or only some keys work

Ranked causes

  1. 1. Keyboard cable loose or faulty (common)
  2. 2. Membrane keypad faulty (occasional)
First check

Check keyboard cable connection

Test point
Keyboard cable to LCD interface board
Expected
Cable fully seated and intact

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

MC network port authentication issue

Monitor cannot authenticate on IEEE 802.1X port-based authentication enabled port

Ranked causes

  1. 1. 802.1X authentication not enabled on monitor (common)
  2. 2. Incorrect authentication credentials (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP air leakage

NIBP module exhibits air leakage or loss of pressure integrity

Ranked causes

  1. 1. Hose or cuff leaking (common)
  2. 2. Leakage inside the module (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No MC network traffic

Monitor not communicating on MC network; no traffic detected

Ranked causes

  1. 1. Incorrect network configuration (common)
  2. 2. MC network traffic overload (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No transducer connected or sensor faulty patient safety

Missing or non-functional pressure transducer detected on indicated channel

Ranked causes

  1. 1. Transducer not connected (common)
  2. 2. Sensor faulty (occasional)
First check

Verify transducer connection

Test point
transducer connector on indicated channel
Expected
Transducer physically connected

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sample flow rate out of specification

Measured gas sample flow rate does not meet 150 ± 25 ml/min specification; indicates pump or flow restriction issue

Ranked causes

  1. 1. Sample gas flow rate too low or high (common)
  2. 2. Sampling pump motor bearing wear or malfunction (occasional)
First check

Connect module and allow 5 minute warm-up

Test point
Module operation
Expected
Module powered and stabilized

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touchscreen inoperative

Touchscreen does not respond to touches or clicks

Ranked causes

  1. 1. Touchscreen cable loose (common)
  2. 2. Touchscreen sensor faulty (occasional)
First check

Check touchscreen cable connection

Test point
Touchscreen cable to LCD interface board
Expected
Cable fully seated and intact

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Trim knob not working

Trim knob control is unresponsive

Ranked causes

  1. 1. Trim knob cable loose or faulty (common)
  2. 2. Trim knob defective (occasional)
First check

Check trim knob cable connection

Test point
Trim knob cable to LCD interface board
Expected
Cable fully seated and intact

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Water trap disconnected or missing

Monitor displays 'Check Water Trap' message in digit field; water trap detached from sampling system

Ranked causes

  1. 1. Mini D-fend water trap disconnected or unseated (common)
  2. 2. Water trap detection sensor malfunction (occasional)
First check

Detach Mini D-fend water trap from module

Test point
Water trap gas connector
Expected
Water trap disconnected from module

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ambient pressure sensor reading error

Absolute pressure sensor reading differs from barometer by more than ± 10 mmHg; indicates sensor malfunction

Ranked causes

  1. 1. Pressure sensor calibration drift or failure (occasional)
First check

Access Gas service menu and note ambient pressure reading

Test point
Ambient pressure value in Gas service menu
Expected
Numeric pressure value displayed

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Arrhythmia paused patient safety

ECG channels not available for analysis for 20+ seconds or internal HR not updated for 30+ seconds due to excessive artifact.

Ranked causes

  1. 1. Excessive artifact or poor electrode contact (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Artifact detected in ECG patient safety

Muscle artifact or high/low frequency noise detected in waveform.

Ranked causes

  1. 1. Poor electrode contact or lead placement, patient movement (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery empty

Monitor is battery powered with less than 5 minutes of monitoring time remaining.

Ranked causes

  1. 1. Battery charge depleted (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery low

Monitor is battery powered with less than 20 minutes of monitoring time remaining.

Ranked causes

  1. 1. Battery charge low (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery temperature high caution

Battery temperature is too high.

Ranked causes

  1. 1. Battery overheated (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibrate NIBP message displayed

NIBP calibration required. Message in digit field.

Ranked causes

  1. 1. NIBP calibration required (unknown)

Calibration: CPU battery replacement

Replacement of FRU B40B20 CPU battery at the recommended 5-year interval

Ranked causes

  1. 1. CPU battery maintenance interval reached (common)
First check

CPU battery age verification

Test point
Installation records
Expected
5 years or more since last replacement

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Display brightness adjustment

Set display brightness level according to operational needs

Ranked causes

  1. 1. Display brightness requires adjustment (common)
First check

Verify display brightness adjustment is accessible

Test point
Menu > Brightness control
Expected
Brightness control present and functional

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: E-COP module maintenance check high voltage

Preventive and corrective maintenance checklist for E-COP module including visual inspection and electrical safety validation

Ranked causes

  1. 1. Routine E-COP module maintenance procedure (common)
First check

Module front cover integrity

Test point
E-COP module front cover
Expected
Front cover is intact

5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: E-Entropy module maintenance check patient safety

Verify E-Entropy module functionality, sensor recognition, and electrical safety prior to clinical use

Ranked causes

  1. 1. Module or sensor malfunction requiring verification (unknown)
First check

Visual inspection of module front cover

Test point
Module front cover
Expected
Cover is intact

8 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Invasive blood pressure channel P4

Calibrate pressure transducer on P4 invasive blood pressure measurement channel

Ranked causes

  1. 1. P4 pressure channel calibration drift (unknown)
First check

Verify P4 calibration menu available

Test point
Service interface COP calibrations
Expected
Calibrate P4 option accessible

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Invasive blood pressure channels P1 and P2

Calibrate pressure transducers on invasive blood pressure measurement channels

Ranked causes

  1. 1. Pressure channel calibration drift (unknown)
First check

Verify P1/P2 calibration menu available

Test point
Service interface STP calibration
Expected
Calibrate P1/P2 option accessible

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: NIBP pressure measurement patient safety

NIBP calibration procedure to verify or adjust non-invasive blood pressure measurement accuracy using a digital manometer and reference pressure source.

Ranked causes

  1. 1. NIBP measurement out of specification or calibration check failure (unknown)
First check

Connect adult NIBP hose to NIBP connector

Test point
NIBP connector
Expected
Secure connection established

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Post-FRU functional verification patient safety

Verification procedure after detaching, replacing or upgrading field replaceable units (FRU) in patient monitor to ensure proper installation and system functionality.

Ranked causes

  1. 1. FRU installation requires functional verification (common)
First check

Check all product labelling, marking and symbols

Test point
Device exterior
Expected
Intact and readable

5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Post-maintenance verification patient safety

Visual inspection and functional checks required after detaching, replacing or upgrading FRU components or maintenance parts in the B125/B105 patient monitor

Ranked causes

  1. 1. FRU replacement or upgrade completed (common)
First check

Product labelling marking and symbols intact and readable

Test point
Monitor exterior
Expected
All labels present and legible

5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Sample flow gain

Adjust sampling pump gain value for gas measurement accuracy

Ranked causes

  1. 1. Sample flow measurement drift (unknown)
First check

Verify sample flow gain adjustment available

Test point
Service interface Gas Control menu
Expected
Sample Flow Gain parameter accessible

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Sample flow rate adjustment caution

Sample flow rate must be adjusted if the sample flow rate check failed during checkout procedures.

Ranked causes

  1. 1. Sample flow rate check failed (unknown)
First check

Stop breathing to gas sampling line

Test point
Gas sampling line inlet
Expected
Apnea alarm appears in message field within 30 seconds

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Set national requirements

Activate France specific defaults for ECG heart rate adjustment range and reminder beep behavior

Ranked causes

  1. 1. National requirements configuration needed (occasional)
First check

Access Service menu with authentication

Test point
Service menu
Expected
Service menu accessible after entering credentials

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Set time zone

Configure time zone, daylight savings, time sync source and NTP settings on patient monitor when not connected to network and patient discharged

Ranked causes

  1. 1. Time zone configuration required (common)
First check

Verify monitor is not connected to network

Test point
Network connection status
Expected
Monitor not connected to network

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Slave display picture adjustment

Adjust picture quality on slave display using OSD menu

Ranked causes

  1. 1. Slave display picture requires adjustment (occasional)
First check

Slave display picture quality acceptable

Test point
Slave display screen
Expected
Picture quality clear and properly adjusted per display user manual

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Temperature channels T1 and T2

Calibrate invasive temperature sensor channels T1 and T2 for accurate measurements

Ranked causes

  1. 1. Temperature channel calibration drift (unknown)
First check

Verify T1/T2 calibration menu available

Test point
Service interface STP calibration
Expected
Calibrate T1/T2 option accessible

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Temperature measurement

Temperature channel self-calibration procedure

Ranked causes

  1. 1. Temperature calibration in progress (common)
First check

Monitor temperature self-check status

Test point
temperature channel (DF)
Expected
Self-check completes with no error messages

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Wired CARESCAPE Network configuration

Configure monitor network settings including unit name, bed name, and MCS IP address

Ranked causes

  1. 1. Monitor requires network configuration for first-time use or maintenance (common)
First check

Enter Unit Name in Network Config

Test point
Basic Service > Network menu
Expected
Unit Name field populated and saved

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cardiac output measurement not responding or reading incorrectly patient safety

C.O. manual measurement fails to start, displays error, or returns unexpected values; temperature readings (T injectate, T blood) not close to simulator set values

Ranked causes

  1. 1. Simulator temperature settings not matching configured values or measurement initialization issue (common)
First check

Verify monitor C.O. configuration

Test point
Cardiac output menu
Expected
Catheter Type: User defined, Computation Constant: 0.542, Measurement Type: Manual, Injectate Volume: 10 ml, Ref Measurement: Deselect

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check Device message displayed

Hemo module malfunction, Masimo type only. Message appears in digit field.

Ranked causes

  1. 1. Hemo module malfunction (unknown)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check NIBP message displayed

Systolic and/or diastolic results missing during NIBP measurement. Message in message field.

Ranked causes

  1. 1. Missing NIBP measurement results (unknown)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check Probe message displayed

No detectable SpO2 signal. Sensor is faulty or detached from patient. Message in digit or message field.

Ranked causes

  1. 1. Faulty or detached SpO2 sensor (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check zero drift compensation NIBP

Monitor and assess zero drift compensation in NIBP pneumatics system

Ranked causes

  1. 1. NIBP zero drift exceeds ±10 mmHg limit (occasional)
First check

Display and review zero drift value

Test point
Service interface NIBP module Zero parameter
Expected
Zero drift value between +20 to -20 mmHg, ideally within ±10 mmHg

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Condition battery

Battery is not working properly and requires conditioning.

Ranked causes

  1. 1. Battery requires conditioning or replacement (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

COP module gain adjustment

Adjust gain coefficient to compensate for gain error in cardiac output measurement

Ranked causes

  1. 1. Gain value outside normal range 17000-25000 (occasional)
First check

Monitor gain coefficient value

Test point
Service interface COP module Gain parameter
Expected
Gain value between 17000 and 25000

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

COP module zero offset adjustment

Adjust zero offset compensation value for A/D converter in cardiac output measurement

Ranked causes

  1. 1. Zero offset value outside ±1000 range (occasional)
First check

Monitor zero offset compensation value

Test point
Service interface COP module Zero parameter
Expected
Zero offset within ±1000

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Cuff loose message displayed
  • Cuff occlusion message displayed
  • Cuff overpressure message displayed
  • E-COP module test result documentation
  • ECG measurements removed
  • ECG module error
  • Faulty Probe message displayed
  • Frame has power issue
  • Frame temperature high
  • Gas sampling system leakage detected
  • Hemo parameters removed
  • IBP abnormally low pressure
  • IBP channel zeroed to over 100 mmHg
  • ICP channel must be zeroed separately
  • Identical IP address noticed
  • Identical SpO2 modules message displayed
  • Identical unit and bed name noticed
  • Incompatible NIBP message displayed
  • Incompatible Probe message displayed
  • Incorrect system time when network connected
  • Interference message displayed
  • Invasive pressure channel not zeroed
  • Invasive pressure measurement not zeroing or reading incorrectly
  • Lead off - LA/L, LL/F, RA/R, RL/N, or V/C
  • Leads off
  • License invalid
  • Loading failed
  • Long measurement time message displayed
  • Low Perfusion message displayed
  • Low signal quality message displayed
  • Maintenance: CPU battery replacement
  • Measurement off - impedance respiration
  • Memory ERROR
  • Mode data reset
  • Network down
  • Network down: HL7
  • Network storm causes temporary data loss
  • NIBP cuff loose message displayed
  • NIBP cuff occlusion message displayed
  • NIBP manual message displayed
  • NIBP measurement removed message displayed
  • No battery backup
  • No SpO2 probe message displayed
  • No SpO2 pulse message displayed
  • P1/P2/P4 over range message displayed
  • Parameter module cannot be placed in frame properly
  • Poor Signal message displayed
  • Probe Off message displayed
  • Pulse Search message displayed
  • Px/P2/P4 under range message displayed
  • Recorder cover open
  • Recorder input voltage high or low
  • Recorder out of paper
  • Recorder system error
  • Recorder thermal array overheat
  • Recorder works but no output on paper
  • Replace battery
  • Replacement of planned maintenance parts
  • Restart needed
  • Select cuff size message displayed
  • Small respiration curve
  • Software activation incomplete or unsuccessful
  • SpO2 faulty probe message displayed
  • SpO2 measurement removed message displayed
  • SpO2 module error message displayed
  • SpO2 probe off message displayed
  • Temperature error / T1 temperature error / T2 temperature error
  • Tubing contamination check
  • UMBC communication error
  • Unit does not display Condition battery or Battery failure message after startup
  • Unstable zero pressure message displayed
  • Weak pulsation message displayed
  • Zero valve not functioning

73 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
Alarm setup changed from CentralAlarm setup retrieved from central station or ECG alarms modified from CentralConnection to central station established or ECG alarm configuration changed remotelyinfo
Alarms reset from CentralAlarms were remotely paused from central stationCentral station sends pause commandinfo
Apnea deactivatedMonitor recently admitted patient or measurement just startedPatient admission or measurement initiationinfo
Arrhythmia PausedECG channels unavailable for analysis or HR not updated due to excessive artifactArtifact present for 20+ seconds or HR not updated for 30+ secondswarning
ArtifactMuscle artifact or high/low frequency noise in ECG signalArtifact detected in waveformwarning
CalibratedChannel calibrated successfullyCalibration procedure completed without errorsinfo
CalibratingCalibration of a channel is in progressUser initiates calibration procedureinfo
Call service: Error 0RAM test failureDetected during startup self-testcritical
Call service: Error 1ROM checksum failureDetected during startup self-testcritical
Call service: Error 10RTK 400Hz timer re-entryDetected in real-time kernelcritical
Call service: Error 11RTK 50Hz timer re-entryDetected in real-time kernelcritical
Call service: Error 13RTK overrunDetected in real-time kernelcritical
Call service: Error 14Too early AUTO START according to module checkAUTO START command issued prematurelyerror
Call service: Error 15Calibration data invalid on initialization or unit never calibratedDevice startup when calibration data is invaliderror
Call service: Error 16Communication timeout between main and safety CPUDetected during operation when CPUs fail to communicatecritical
Call service: Error 17Safety CPU report communication timeoutDetected when safety CPU detects communication losscritical
Call service: Error 18Wrong message rate in communication between main and safety CPUMessage rate deviates >±2% from 480 msg/serror
Call service: Error 2Pump on during idle or over current detectedDetected during normal operationerror
Call service: Error 3Startup communication failure with safety CPUDetected at device startupcritical
Call service: Error 5Calibration Not ProtectedDetected during calibration procedureerror
Call service: Error 6Valve stuck closed during cuff typingDuring NIBP cuff inflationerror
Call service: Error 7Could not save calibration dataAfter calibration procedureerror
Call service: Error 8Pressure exceeds limits while idlePT2 >150 mmHg for >15 s while idle, or pressure >15 mmHg for >180 s (Adult) or >5 mmHg for >90 s (Neo)error
Call service: Error 9Determination time too longNIBP measurement takes excessive timeerror
Call service UMBC errorUMBC communication error and UMBC communication disabledUMBC communication failureerror
Default settings returnedMonitor settings have been reset to defaultFactory reset or settings recovery activatedinfo
DEMO MODEDEMO mode has been enabledDEMO mode activatedwarning
Entering standbyStandby mode activation selectedStandby activation command issuedinfo
FailedPressure calibration failure due to time-out or waveform/gain issuesCalibration procedure does not complete within time limit or detects pulsation or gain out of rangeerror
Failed: P<100Calibration reference value entered is less than 100 mmHgUser enters pressure reference below 100 mmHgerror
InvBP's not zeroedAt least one invasive pressure channel has not been zeroedAttempt to measure with unzeroed IBP channelwarning
Lead changedMonitor auto-switched ECG1 waveform to measurable ECG leadCurrent ECG1 waveform not measurableinfo
LearningST algorithm in learning phaseST algorithm startup or ECG measurement startinfo
License invalidLicense is invalid during startupMonitor startup with invalid licenseerror
Loading failedLoading mode from network interruptedNetwork data transfer interruptederror
Loading from networkPatient data being loaded from the networkData transfer from network initiatedinfo
Memory ERRORMemory error after software downloadSoftware download completionerror
Mode data resetError loading settings from flash fileCold start, warm start, upgrade, or revert modeerror
Network downCARESCAPE Network connection failedNetwork connectivity losserror
Network down: HL7No HL7 TCP Client configured to connect to monitorHL7 connection attempt with no client configurederror
Network: HL7HL7 TCP Client application network connection establishedHL7 connection establishedinfo
Network madeCARESCAPE network connection establishedNetwork connection successfulinfo
No breath deactivatedOne of the electrodes is offElectrode disconnection detectedwarning
No x transducerNo transducer connected to indicated channel or sensor is faultyChannel detects missing or non-responsive transducererror
Patient admittedCurrent patient has been admittedPatient admission recordedinfo
Patient dischargedPatient has been dischargedPatient discharge recordedinfo
Performing temp testTemperature is calibrating (self-check in progress)Temperature channel self-calibration initiatedinfo
Printing...Printing is occurringPrint operation initiatedinfo
Printing AlarmAlarm has triggered printingAlarm condition activates printwarning
Recorder: cover openRecorder cover is openRecorder cover openedwarning
Recorder: input voltage high / Recorder: input voltage lowProblems with recorder input voltageRecorder voltage out of specificationerror
Recorder module removedRecorder module has been removedRecorder module disconnectedinfo
Recorder: out of paperRecorder out of paper or cover openNo paper detected or cover openwarning
Recorder system error {0d}Local recorder not workingRecorder communication failureerror
Recorder thermal array overheatProblems with recorder temperatureRecorder temperature exceeds limit during operationwarning
SavingPrinting device not available; records saved for later printingPrint request with no printer availablewarning
Saving to networkSaving trend and information to NetworkNetwork save operation initiatedinfo
STP measurements removedHemodynamic (Hemo) parameters have been removedHemo module disconnected or communication lostwarning
Zero adj > 100 mmHgIBP channel zeroed to over 100 mmHg pressureZero adjustment value exceeds 100 mmHg limiterror
ZeroedZeroing was successfulZeroing procedure completed without errorsinfo
Zero ICP separatelyICP channel must be zeroed separately from other invasive pressuresAttempt to zero ICP with other pressure channelswarning
ZeroingIBP channel is currently being zeroedUser initiates zeroing procedureinfo
Zeroing failedZeroing procedure failed due to pulsation, defective transducer, or excessive offsetZeroing encounters pulsating waveform, detects defective transducer, or offset exceeds 150 mmHgerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI